Rosemount Analytical 225 Specifications

Toroidal Conductivity Sensor
LIQ-PDS-225
Rosemount™ 225
Toroidal Conductivity Sensors
Product Data Sheet
July 2017
Rosemount 225 toroidal conductivity sensors are intended to be used in many pharmaceutical and food and beverage applications where a sanitary design is required. These corrosion and fouling resistant sensors are ideal for measuring the concentration of CIP solutions, detecting product/water interfaces, checking product quality, and monitoring eleuents in chromatographic separations.
July 2017Toroidal Conductivity Sensor
Meet Various Sanitary Requirements
J USP Class VI compliant J FDA food contact 21CFR177.2415 compliant J 3-A Sanitary Standard 74-06 compliant
Ease of Installation
J 2 inch tri-clamp process connection J Installation exibility with maximum cable lenths up to
200 ft (61 m) between sensor and transmitter when paired
with remote junction box (sold separately)
High Performance and Reliability
J Robust measurements - insensitive to process ow and
direction
J Meet application compatibility requirements with a selection
of chemical resistant body options including PEEK and Tefzel
Contents
Ordering Information........ ................................................................................................................................................................3
Specications........ ............................................................................................................................................................................3
Dimensional Drawing........ ................................................................................................................................................................4
Accessories........................................................................................................................................................................................4
Engineering Specication for 225 Sensor........ ................................................................................................................................... 5
2
www.Emerson.com/RosemountLiquidAnalysis
July 2017 Toroidal Conductivity Sensor
Ordering Information
Rosemount 225 Sanitary Toroidal Conductivity Sensors work well in high conductivity liquids up to 2 S/cm (2,000,000 µS/cm). These sensors include an integral Pt-100 RTD for
temperature compensation and 20 ft of integral cable. Cables can be extended using a remote junction box PN 23550-00 (see Accessories).
Rosemount 225 Sanitary Toroidal Conductivity Sensor
Table 1 -Rosemount 225 Sanitary Toroidal Conductivity Sensor Ordering Information
Model Sensor Type
225 Toroidal Conductivity Sensor
Body Material & Mounting Type
03 Glass-lled PEEK with tri-clamp
07 Unlled PEEK with tri-clamp
08 USP Class VI unlled PEEK with tri-clamp
09 Unlled Tefzel
1
Transmitter Compatibility
54 Standard integral cable
56 Integral cable with additional shielding for improved EMI/RFI protection
Typical model number: 225-03-56
(1) Only available with -56 option
(2) Recommended for use with Rosemount transmitter models 56, 1056, 5081, and 1066
1
2
Specifications
Cell Constant (Nominal): 2.7/cm
Minimum Conductivity: 200 µS/cm (15 µS/cm when used with Rosemount 1056 and 56 transmitters)
Maximum Conductivity: 2 S/cm
Process Connection: 2 inch tri-clamp
Conformance to 3-A Sanitary Standards: Sensors with option -07 meet 3-A sanitary standards for sensors and sensor ttings and
connections used on milk and milk products equipment (74-06).
Compliance with FDA Food Contact Requirements: Sensors with option -07 are molded from PEEK that meet 21CF177.2415.
Compliance with USP Class VI: Sensors with option -08 are molded from PEEK that meet USP Class VI requirements.
Cable Length: 20 ft (6.1 m)
Maximum Cable Length: 200 ft (61 m)
Weight/Shipping Weight: 2 lb/3 lb (1.0 kg/1.5 kg)