Rosemount Analytical 222 Specifications

Page 1
Rosemount™ 222
Toroidal Flow Through Conductivity Sensor
Product Data Sheet
July 2017
LIQ-PDS-222
A high performance sensor with ease of maintenance in mind
Page 2
Rosemount 222 Toroidal Flow Through Conductivity Sensor

Overview

High Performance and Reliability
Robust measurements - insensitive to process flow rate and
direction.
Avoid accumulation of solids with no part of the sensor
protruding into the sample flow.
Corrosion resistant Teflon pipe liner.
Pt-100 RTD included to allow temperature compensated
conductivity measurements.
Operates in process temperatures up to 182 °C (356 °F).
Meet Process Flow Through Requirements
Versatile installation options allows for direct mounting into
1 in. and 2 in. process lines with 150 lb or 300 lb raised face threaded ANSI B16.5 flanges.
Plugging resistant and suitable for liquids containing high
levels of suspended solids.
July 2017
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2
Instrument Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
www.Emerson.com/RosemountLiquidAnalysis
Page 3
July 2017
Rosemount 222 Toroidal Flow Through Conductivity Sensor

Ordering Information

Rosemount 222 Flow Through Toroidal sensors feature Teflon-lined pipe, an external toroid assembly, two unlined carbon steel outer flanges, a Pt-100 RTD temperature compensation element, and 20 ft. of integral cable. Optional stainless steel outer flanges are also available.
Table 1. Rosemount 222 Flow Through Toroidal Conductivity Sensor Ordering Information
Model Sensor type
222 Flow Through Toroidal Conductivity Sensor
(1)
Sizes
01 1 inch, 150 lb flange
02 2 inch, 150 lb flange
05 1 inch, 300 lb flange
06 2 inch, 300 lb flange
Flange option
_ No selection
21 316 stainless steel outer flange
(2)
Cable
54 Standard integral cable
(3)
Typical Model Number: 222-01-_-54
1. Grounding rings are required for proper operation if the outer flanges of the Rosemount 222 sensor are substituted by the customer with non-conductive flanges. Order SQ 7430 and consult the factory for pricing.
2. Only available with option -01.
3. Cables can be extended using remote junction box PN 23550-00 and extension cable. See Accessories .
www.Emerson.com/RosemountLiquidAnalysis
3
Page 4
Rosemount 222 Toroidal Flow Through Conductivity Sensor

Specifications

Table 2. Rosemount 222 sensor specifications
Sensor type
Toroidal Flow Through Conductivity Sensor
Cell constant (Nominal)
1 in. diameter 6/cm
2 in. diameter 4/cm
Process connection
1 in. 150 lb or 300 lb raised face threaded ANSI B16.5 flange
2 in. 150 lb or 300 lb raised face threaded ANSI B16.5 flange
Wetted materials
Teflon-lined carbon steel pipe, with carbon steel after flanges; 316 stainless steel outer flange (option -21)
Cable length
July 2017
20 ft. (6.1 m)
Maximum cable length
100 ft. (30 m)
Table 3. Temperature and pressure
Flange Temperature range Maximum pressure
150 lb
5 to 182 °C (41 to 360 °F)
300 lb 250 psig (1825 kPa [abs])
125 psig (963 kPa [abs])
Table 4. Pressure (for CRN registration only)
Option Flange Diameter Maximum pressure
-01 150 lb
-02 2 in.
-05 300 lb
-06 2 in. 250 psig (1825 kPa [abs])
-01-21 150 lb 1 in. 125 psig (963 kPa [abs])
1 in.
125 psig (963 kPa [abs])
1 in. 200 psig (1480 kPa [abs])
Table 5. Weight/Shipping weight (Weights are rounded up to the nearest 1 lb or 0.5 kg)
Flange Diameter Weight Shipping weight
150 lb
1 in.
300 lb 17 lb (8.0 kg) 20 lb (9.0 kg)
150 lb
2 in.
300 lb 35 lb (16.0 kg) 40 lb (18.0 kg)
4
11 lb (5.0 kg) 14 lb (6.5 kg)
33 lb (15.0 kg) 37 lb (17.0 kg)
www.Emerson.com/RosemountLiquidAnalysis
Page 5
July 2017
Rosemount 222 Toroidal Flow Through Conductivity Sensor

Instrument Compatibility

For more information regarding compatibility with older instruments please call our Customer Care Department at 855-724-2638.
Rosemount 56, 1056, and 1066
Measurement choices: Conductivity, resistivity, total dissolved solids, salinity, and % concentration
Salinity: Uses Practical Salinity Scale
Total Dissolved Solids: Calculated by multiplying conductivity at 25 °C by 0.65
% Concentration Selections
Temperature Compensation Options: Manual slope (X%/°C) and neutral salt (dilute sodium chloride)
Repeatability: ± 0.25% ± 5 μS/cm after zero cal
Input Filter: Time constant 1-999 seconds, default 2 seconds
Response Time: 3 seconds to 100% of final reading
Table 6. Rosemount 56, 1056, and 1066 Transmitter temperature specifications
(1)
: 0-12% NaOH, 0-15% HCl, 0-20% NaCl, and 0-25% or 96-99.7% H2SO
4
Temperature range -25 to 210 °C (-13 to 410 °F)
Temperature accuracy, Pt-100, -25 to 50 °C ± 0.5 °C
Temperature accuracy, Pt-100, 50 to 210 °C ± 1 °C
Table 7. Transmitter loop performance with Rosemount 222 Sensors
Transmitter model Conductivity range Loop performance (following calibration)
Rosemount 1056/56
500 μS/cm to 2000 mS/cm
Rosemount 1066 ± 4% of reading ± 5 mS/cm in recommended range
± 4% of reading in recommended range
Table 8. Rosemount 1066 Transmitter loop performance with Rosemount 222 Sensors
Conductivity range Loop performance (following calibration)
15 μS/cm to 1500 mS/cm ± 1% of reading ± 15 μS/cm in recommended range
1500 mS/cm to 2000 mS/cm ± 5% of reading outside high recommended range
1. The conductivity concentration algorithms for these solutions are fully temperature compensated.
www.Emerson.com/RosemountLiquidAnalysis
5
Page 6
Rosemount 222 Toroidal Flow Through Conductivity Sensor
July 2017
Rosemount 5081
Table 9. Rosemount 5081 Transmitter specifications at 25 °C
Accuracy ± 1.0% of reading
Repeatability ± 0.25% of reading
Stability 0.25% of output range per month, non-cumulative
Ambient temperature coefficient ± 0.1% of reading ± 2 μS/cm per °C
Table 10. Recommended conductivity ranges for Rosemount 222 Sensor with Rosemount 5081 Transmitter
Nominal cell constant 6.0/cm (1 in.) or 4.0/cm (2 in.)
Minimum conductivity 500 μS/cm
Maximum conductivity 2,000,000 μS/cm
Note: Values shown are for 25° conductivity with a temperature slope of 2% per °C. The maximum range value will be lower for solutions with a higher temperature slope.

Dimensional Drawings

Figure 1. Rosemount 222 Dimensional Drawing
6
www.Emerson.com/RosemountLiquidAnalysis
Page 7
July 2017
Table 11. Dimensions
Rosemount 222 Toroidal Flow Through Conductivity Sensor
Option Pipe Diameter Flange
(1)
“A” Dimension
(2)
“B” Dimension
-01 1 in. 150 lb 7.0 in. (178 mm) 8.4 in. (213 mm)
-02 2 in. 150 lb 24.0 in. (610 mm) 26.0 in. (660 mm)
-05 1 in. 300 lb 7.0 in. (178 mm) 9.1 in. (232 mm)
-06 2 in. 300 lb 24.0 in. (610 mm) 26.6 (676 mm)
-01-21 1 in. 150 lb 7.0 in. (178 mm) 8.4 in. (213 mm)
1. Outside flanges are ANSI B16.5 raised face, threaded pipe flanges
2. Dimension is +0.125 in. (3 mm)
3. Approximate dimension

Accessories

Table 12. Rosemount 222 Sensor accessories information
Part number Description
2001492 Stainless steel tag (must specify marking)
23550-00 Remote junction box without preamplifier
23294-00 Interconnecting extension cable, unshielded, prepped (for use with remote junction box)
23294-05 Interconnecting extension cable, shielded, prepped (for use with remote junction box)
9200276 Interconnecting extension cable, shielded, unprepped (for use with remote junction box)
2002557 Insulation kit, 1 in. 150 lb flange, 2 sets
2002558 Insulation kit, 1 in. 300 lb flange, 2 sets
2002559 Insulation kit, 2 in. 150 lb flange, 2 sets
2002560 Insulation kit, 2 in. 300 lb flange, 2 sets
8950101 Pt-100 RTD assembly
(1)
(1)
(1)
(1)
(3)
1. Each insulation kit contains two flange gaskets and sufficient insulating sleeves, insulating washers, and stainless steel washers to replace both flange seals of one sensor. The kit does not contain flange bolts or nuts.
www.Emerson.com/RosemountLiquidAnalysis
7
Page 8
Rosemount 222 Toroidal Flow Through Conductivity Sensor
LIQ-PDS-222
Global Headquarters
Emerson Automation Solutions
8200 Market Blvd Chanhassen, MN 55317
+1 800 999 9307 or +1 952 906 8888
+1 952 949 7001 Liquid.CSC@Emerson.com
North America Regional Office
Emerson Automation Solutions
8200 Market Blvd. Chanhassen, MN 55317, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 949 7001 RMT-NA.RCCRFQ@Emerson.com
Latin America Regional Office
Emerson Automation Solutions
1300 Concord Terrace, Suite 400 Sunrise, FL 33323, USA
+1 954 846 5030 +1 954 846 5121 RFQ.RMD-RCC@Emerson.com
Product Data Sheet
July 2017
Europe Regional Office
Emerson Automation Solutions GmbH
Neuhofstrasse 19a P.O. Box 1046 CH 6340 Baar Switzerland
+41 (0) 41 768 6111 +41 (0) 41 768 6300 RFQ.RMD-RCC@Emerson.com
Asia Pacific Regional Office
Emerson Automation Solutions Asia Pacific Pte Ltd
1 Pandan Crescent Singapore 128461
+65 6777 8211 +65 6777 0947 Enquiries@AP.Emerson.com
Middle East and Africa Regional Office
Emerson Automation Solutions
Emerson FZE P.O. Box 17033, Jebel Ali Free Zone - South 2 Dubai, United Arab Emirates
+971 4 8118100 +971 4 8865465 RFQ.RMTMEA@Emerson.com
Analyticexpert.com
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Youtube.com/user/RosemountMeasurement
Google.com/+RosemountMeasurement
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount and Rosemount logotype are trademarks of Emerson. All other marks are the property of their respective owners. © 2017 Emerson. All rights reserved.
Loading...