A high performance sensor with ease of maintenance in mind
The Rosemount 222 toroidal flow through conductivity sensor measures conductivity in highly conductive liquids up to 2 S/cm
(2,000,000 μS/cm). A noninvasive flow through design allows for in-line installations without any obstruction to sample and makes
these sensors ideal for use with viscous or fibrous liquids.
Page 2
Rosemount 222 Toroidal Flow Through Conductivity Sensor
Overview
High Performance and Reliability
Robust measurements - insensitive to process flow rate and
direction.
Avoid accumulation of solids with no part of the sensor
protruding into the sample flow.
Corrosion resistant Teflon pipe liner.
Pt-100 RTD included to allow temperature compensated
conductivity measurements.
Operates in process temperatures up to 182 °C (356 °F).
Meet Process Flow Through Requirements
Versatile installation options allows for direct mounting into
1 in. and 2 in. process lines with 150 lb or 300 lb raised face
threaded ANSI B16.5 flanges.
Plugging resistant and suitable for liquids containing high
Rosemount 222 Toroidal Flow Through Conductivity Sensor
Ordering Information
Rosemount 222 Flow Through Toroidal sensors feature Teflon-lined pipe, an external
toroid assembly, two unlined carbon steel outer flanges, a Pt-100 RTD temperature
compensation element, and 20 ft. of integral cable. Optional stainless steel outer
flanges are also available.
Table 1. Rosemount 222 Flow Through Toroidal Conductivity Sensor Ordering Information
ModelSensor type
222Flow Through Toroidal Conductivity Sensor
(1)
Sizes
011 inch, 150 lb flange
022 inch, 150 lb flange
051 inch, 300 lb flange
062 inch, 300 lb flange
Flange option
_No selection
21316 stainless steel outer flange
(2)
Cable
54Standard integral cable
(3)
Typical Model Number: 222-01-_-54
1.Grounding rings are required for proper operation if the outer flanges of the Rosemount 222 sensor are substituted by the customer with non-conductive flanges.
Order SQ 7430 and consult the factory for pricing.
2.Only available with option -01.
3.Cables can be extended using remote junction box PN 23550-00 and extension cable. See Accessories .
www.Emerson.com/RosemountLiquidAnalysis
3
Page 4
Rosemount 222 Toroidal Flow Through Conductivity Sensor
Specifications
Table 2. Rosemount 222 sensor specifications
Sensor type
Toroidal Flow Through Conductivity Sensor
Cell constant (Nominal)
1 in. diameter6/cm
2 in. diameter4/cm
Process connection
1 in. 150 lb or 300 lb raised face threaded ANSI B16.5 flange
2 in. 150 lb or 300 lb raised face threaded ANSI B16.5 flange
Wetted materials
Teflon-lined carbon steel pipe, with carbon steel after flanges; 316 stainless steel outer flange (option -21)
Cable length
July 2017
20 ft. (6.1 m)
Maximum cable length
100 ft. (30 m)
Table 3. Temperature and pressure
FlangeTemperature rangeMaximum pressure
150 lb
5 to 182 °C (41 to 360 °F)
300 lb250 psig (1825 kPa [abs])
125 psig (963 kPa [abs])
Table 4. Pressure (for CRN registration only)
OptionFlangeDiameterMaximum pressure
-01
150 lb
-022 in.
-05
300 lb
-062 in.250 psig (1825 kPa [abs])
-01-21150 lb1 in.125 psig (963 kPa [abs])
1 in.
125 psig (963 kPa [abs])
1 in.200 psig (1480 kPa [abs])
Table 5. Weight/Shipping weight (Weights are rounded up to the nearest 1 lb or 0.5 kg)
FlangeDiameterWeightShipping weight
150 lb
1 in.
300 lb17 lb (8.0 kg)20 lb (9.0 kg)
150 lb
2 in.
300 lb35 lb (16.0 kg)40 lb (18.0 kg)
4
11 lb (5.0 kg)14 lb (6.5 kg)
33 lb (15.0 kg)37 lb (17.0 kg)
www.Emerson.com/RosemountLiquidAnalysis
Page 5
July 2017
Rosemount 222 Toroidal Flow Through Conductivity Sensor
Instrument Compatibility
For more information regarding compatibility with older instruments please call our Customer Care Department at 855-724-2638.
Rosemount 56, 1056, and 1066
Measurement choices: Conductivity, resistivity, total dissolved solids, salinity, and % concentration
Salinity: Uses Practical Salinity Scale
Total Dissolved Solids: Calculated by multiplying conductivity at 25 °C by 0.65
% Concentration Selections
Temperature Compensation Options: Manual slope (X%/°C) and neutral salt (dilute sodium chloride)
Repeatability: ± 0.25% ± 5 μS/cm after zero cal
Input Filter: Time constant 1-999 seconds, default 2 seconds
Response Time: 3 seconds to 100% of final reading
Table 6. Rosemount 56, 1056, and 1066 Transmitter temperature specifications
(1)
: 0-12% NaOH, 0-15% HCl, 0-20% NaCl, and 0-25% or 96-99.7% H2SO
4
Temperature range-25 to 210 °C (-13 to 410 °F)
Temperature accuracy, Pt-100, -25 to 50 °C± 0.5 °C
Temperature accuracy, Pt-100, 50 to 210 °C± 1 °C
Table 7. Transmitter loop performance with Rosemount 222 Sensors
15 μS/cm to 1500 mS/cm± 1% of reading ± 15 μS/cm in recommended range
1500 mS/cm to 2000 mS/cm± 5% of reading outside high recommended range
1.The conductivity concentration algorithms for these solutions are fully temperature compensated.
www.Emerson.com/RosemountLiquidAnalysis
5
Page 6
Rosemount 222 Toroidal Flow Through Conductivity Sensor
July 2017
Rosemount 5081
Table 9. Rosemount 5081 Transmitter specifications at 25 °C
Accuracy± 1.0% of reading
Repeatability± 0.25% of reading
Stability0.25% of output range per month, non-cumulative
Ambient temperature coefficient± 0.1% of reading ± 2 μS/cm per °C
Table 10. Recommended conductivity ranges for Rosemount 222 Sensor with Rosemount 5081 Transmitter
Nominal cell constant6.0/cm (1 in.) or 4.0/cm (2 in.)
Minimum conductivity500 μS/cm
Maximum conductivity2,000,000 μS/cm
Note: Values shown are for 25° conductivity with a temperature slope of 2% per °C. The maximum range value will be lower for
solutions with a higher temperature slope.
Dimensional Drawings
Figure 1. Rosemount 222 Dimensional Drawing
6
www.Emerson.com/RosemountLiquidAnalysis
Page 7
July 2017
Table 11. Dimensions
Rosemount 222 Toroidal Flow Through Conductivity Sensor
OptionPipe DiameterFlange
(1)
“A” Dimension
(2)
“B” Dimension
-011 in.150 lb7.0 in. (178 mm)8.4 in. (213 mm)
-022 in.150 lb24.0 in. (610 mm)26.0 in. (660 mm)
-051 in.300 lb7.0 in. (178 mm)9.1 in. (232 mm)
-062 in.300 lb24.0 in. (610 mm)26.6 (676 mm)
-01-211 in.150 lb7.0 in. (178 mm)8.4 in. (213 mm)
1.Outside flanges are ANSI B16.5 raised face, threaded pipe flanges
2.Dimension is +0.125 in. (3 mm)
3.Approximate dimension
Accessories
Table 12. Rosemount 222 Sensor accessories information
Part numberDescription
2001492Stainless steel tag (must specify marking)
23550-00Remote junction box without preamplifier
23294-00Interconnecting extension cable, unshielded, prepped (for use with remote junction box)
23294-05Interconnecting extension cable, shielded, prepped (for use with remote junction box)
9200276Interconnecting extension cable, shielded, unprepped (for use with remote junction box)
2002557Insulation kit, 1 in. 150 lb flange, 2 sets
2002558Insulation kit, 1 in. 300 lb flange, 2 sets
2002559Insulation kit, 2 in. 150 lb flange, 2 sets
2002560Insulation kit, 2 in. 300 lb flange, 2 sets
8950101Pt-100 RTD assembly
(1)
(1)
(1)
(1)
(3)
1.Each insulation kit contains two flange gaskets and sufficient insulating sleeves, insulating washers, and stainless steel washers to replace both flange seals of one
sensor. The kit does not contain flange bolts or nuts.
www.Emerson.com/RosemountLiquidAnalysis
7
Page 8