Rosemount Analytical 150 Specifications

Temperature °F
Product Data Sheet
LIQ-PDS-150 July 2017
Rosemount 150
Insertion/Submersion Conductivity Sensor
SUITABLE FOR INSERTION/SUBMERSION/ FLOW-THROUGH installation.
TEMPERATURE COMPENSATION element is provided as inte­gral part of sensor.
OPERATES IN HIGH PRESSURE SYSTEM to 175 psig at 100°F (38°C).
SENSOR CONSTANTS OF 0.1, 0.5 and 1.0 are available.
FEATURES AND APPLICATIONS
The Rosemount Model 150 Insertion/ Submersion Sensor is available in sensor constants of
0.1, 0.5 or 1.0 for conductivity ranges from 0-10 to 0-10,000 μS when used with Models 54eC, 81C, 1054BC, 1181C, 2081C, and 3081C analyzers/transmit- ters. For optimal loop performance, connect the Model 150 sensor to instrument Models 54eC or 1054BC.
The
Tefzel1 housing has both rear-facing and forward­facing, 1 inch male pipe threads. The sensor can be threaded into a process line or the side of a tank,
or
be
attached to a length of pipe for submersible instal­lation. The Model 150 is available with a 1 inch CPVC tee (Code 16) for flow-through installation. An inte­gral 10 foot cable is supplied for connection to the analyzer/transmitter or junction box.
This sensor is most appropriately used as a set-point detector in applications with a fairly broad control range. This sensor is a good choice for boiler blow­down and open recirculating cooling water. For appli­cations that require high accuracy over a wide
range
of
conductivity, a sensor from the ENDURANCE family (Model 400 series) or the toroidal family (Model 200 series) would be the ideal choice.
SPECIFICATIONS
Available Sensor Constants: 0.1, 0.5 and 1.0 Process Connection: 1 inch MNPT both front and
rear threads
Junction Box: Code 17: ABS, meets NEMA 4X specifi-
cations 210 mm x 127 mm H x 102 mm D (8-1/4 in. x 5 in. x 4 in.)
Integral Cable: 10 ft (250 cm), four conductors (20 ga.)
shielded, provided with spade lugs. For Code 54: 10 ft (250 cm), 5 connectors (22 ga.)
4 drains, shielded, provided with spade lugs.
Interconnect Cable (J-Box to instrument): Specify
length, 5 conductors (22 ga.), 4 drains, shielded.
Weight/Shipping Weight: 0.5 kg/ 0.9 kg (1 lb/2 lb)
1
Tefzel and Viton are registered trademarks of E.I. du Pont de Nemours and Co.
300
200
100
0
Wetted Materials: Electrodes: 316 stainless steel
Insulators: Tefzel and Viton
1
Temperature
Compensation:
0 to 121°C (32 to 250°F)
50
100
Pressure psig
150 200
Pressure/Temperature Rating: (see Figure 1)
Pressure Temperature
1207 kPa abs (175 psig) 38°C (100°F) 345 kPa abs (50 psig) 121°C (250°F)
FIGURE 1. Pressure/Temperature Rating
ORDERING INFORMATION
Model 150 Insertion/Submersion Conductivity Sensor is available in sensor constants of 0.1, 0.5 or 1.0, covering conduc-
tivity ranges from 0-10 to 0-10,000 μS with selected instruments. The Tefzel housing is supplied with both rear facing and for­ward facing 1 in. MNPT’s. The sensor can be threaded into a process line or the side of a tank or attached to a length of pipe for submersion installations. An integral 10 ft cable is supplied for connection to the analyzer/transmitter or junction box. Interconnect cable between J-Box and analyzer/instrument must be ordered separately (specify length). The Model 150 may be optionally supplied in a 1-1/2 in. CPVC tee with 1 in. process connections for flow-through installation (Code 16).
MODEL
150 Contacting Conductivity Sensor
CODE
Cell constant
03
0.1/cm cell constant
06
1.0/cm cell constant
09
0.5/cm cell constant
Tee assembly
OPTIONS
-
No tee assembly included
16
PVC tee with 1 in. FPT process connections
Transmitter compatibility
54
For use with Rosemount 1056, 56, 1066, 1057, and 5081 Transmitters1
Typical model number: 150-03-_-54
MILLIMETER
INCH
FIGURE 2. Model 150 Sensor with optional 1-1/2 in. CPVC tee adapter with 1 in. FNPT process connections
1
Also compatible with legacy transmitter models 54, 1054, 2054, 1055, 2081, 3081, 4081, and XMT.
Loading...