Page references of
this style work as hyperlinks: Simply click
on the page number to
open that page.
To open a specic
page you can now
enter the logical page
number just as it appears in the footer.
Page 2
Page 3
Short Form Manual
HASXEE-SFM-HS
07/2010
Gas Analyzer Series
Short Form Manual
www.EmersonProcess.com
Page 4
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Emerson Process Management (Rosemount Analytical) designs, manufactures and
tests its products to meet many national and international standards. Because these
instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specications.
The following instructions MUST be adhered to and integrated into your safety program
when installing, using and maintaining Emerson Process Management (Rosemount
Analytical) products. Failure to follow the proper instructions may cause any one of the
following situations to occur: Loss of life; personal injury; property damage; damage to
this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson Process
Management (Rosemount Analytical) representative for clarication.
• Follow all warnings, cautions, and instructions marked on and supplied with the
product.
• Inform and educate your personnel in the proper installation, operation, and
maintenance of the product.
• Install your equipment as specied in the Installation Instructions of the
appropriate Instruction Manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualied personnel to install, operate, update,
program, and maintain the product.
• When replacement parts are required, ensure that qualied people use replacement
parts specied by Emerson Process Management (Rosemount Analytical).
Unauthorized parts and procedures can affect the product’s performance, place the
safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike
substitutions may result in re, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place,
except when maintenance is being performed by qualied persons, to prevent
electrical shock and personal injury.
The information contained in this document is subject to change without notice.
3rd edition 2010-07
Emerson Process Management GmbH & Co. OHG
Industriestrasse 1
D-63594 Hasselroth
Germany
T +49 (0) 6055 884-0
F +49 (0) 6055 884-209
Internet: www.EmersonProcess.com
This short form manual provides information on installation and startup of the
X-STREAM® XE series gas analyzers. For comprehensive instructions see the X-STREAM
XE series instruction manual.
This manual covers several X-STREAM analyzer models and so may contain information
about congurations and/or options not applicable to your analyzer.
The installation and operation of units for use in explosive (hazardous) environments is
not covered in this manual.
Analyzers intended to be used in such environments are supplied with further instruction
manuals, which should be consulted in addition to this.
DEFINITIONS
The following denitions apply to WARNINGS, CAUTIONS and NOTES found throughout
this publication.
HIGHLIGHTS AN OPERATION OR MAINTENANCE PROCEDURE,
PRACTICE, CONDITION, STATEMENT, ETC.
If not strictly observed, could result in injury, death, or long-term health
hazards of personnel.
Safety Instructions
S
HIGHLIGHTS AN OPERATION OR MAINTENANCE PROCEDURE,
PRACTICE, CONDITION, STATEMENT, ETC.
If not strictly observed, could result in damage to or destruction of equipment,
or loss of effectiveness.
NOTE!
Highlights an essential operating procedure, condition or statement.
Emerson Process Management GmbH & Co. OHG S-1
Page 10
X-STREAM XE
TERMS USED IN THIS INSTRUCTION MANUAL
Short Form Manual
HASXEE-SFM-HS
04/2010
Explosive Gas(es)
Flammable Gases and gas mixtures in a mixture with air within the explosive limits.
Flammable Gas(es)
Gases and gas mixtures are assigned to be
ammable if they might become ignitable
when in a mixture with air.
Infallible Containment
This term is derived from the standards of
explosion protection especially from the requirements for pressurized housings: thus an
infallible containment can be characterized
by no intended leakage out of the gas paths
enabling gas to enter the inner compartment
of the analyzer housing.
Intrinsically Safe Cell (IS Cell)
Cells supplied with an intrinsically safe power
signal, approved by a Test Institute, to operate
with explosive gases.
The design ensures the IS cells remains safe
even in case of failure and explosive gases
are not ignited.
Protection Class IP66 / NEMA 4X
Both terms are used to specify conditions for
equipment to be installed outdoor.
IP stands for Ingress Protection, the rst number species protection against solid objects
(6. = dust tight) while the second number
species the degree of protection against
liquids (.6 = heavy seas).
NEMA stands for National Electrical Manufacturers Association. 4X species a degree
of protection to personnel against incidental
contact with the enclosed equipment; to provide a degree of protection against falling dirt,
rain, sleet, snow, windblown dust, splashing
water, and hose-directed water; and that will
be undamaged by the external formation of
ice on the enclosure
Upper Explosion Limit (UEL)
Volume ratio of ammable gas in air above
which an explosive gas atmosphere will not
be formed: the mixture of gas and air is too
rich in fuel (decient in oxygen) to burn.
Lower Explosion Limit (LEL)
Volume ratio of ammable gas in air below
which an explosive gas atmosphere will not
be formed: the mixture of gas and air lacks
sufcient fuel (gas) to burn.
NAMUR
NAMUR is an international user association of
automation technology in process industries.
This organisation has issued experience reports and working documents, called recommendations (NE) and worksheets (NA).
Emerson Process Management GmbH & Co. OHGS-2
Page 11
Short Form Manual
HASXEE-SFM-HS
04/2010
Safety Instructions
X-STREAM XE
SYMBOLS USED ON AND INSIDE THE UNIT
Wherever one or more of the following symbols appear on or inside the instrument, be careful
and read the instructions given in the accompanying manuals!
Strictly observe the given warnings, instructions and information to minimize hazards!
This symbol at the instrument ...... means
dangerous voltages may be accessible. Removing covers is permitted only, if the instrument is
disconnected from power - and even in this case
by qualied personnel only!
hot surfaces may be accessible. Removing
covers by qualied personnel is permitted only,
if the instrument is disconnected from power.
Nevertheless several surfaces may remain hot
for a limited time.
more detailled information available: see instruction manual before proceeding!
more detailled information available: see instruction manual before proceeding!
Safety Instructions
S
Emerson Process Management GmbH & Co. OHG S-3
Page 12
Short Form Manual
X-STREAM XE
Safety Instructions
HASXEE-SFM-HS
04/2010
SYMBOLS USED IN THIS MANUAL
Where one or more of the following symbols appear within this manual, carefully read the related information and instructions!
Strictly observe the given warnings, instructions and information to minimize hazards!
This symbol used in the manual ...... means
dangerous voltages may be exposed
hot surfaces may be exposed
possible danger of explosion
toxic substances may be present
substances harmful to health may be present
indicates notes relating to heavy instruments
electrical components may be destroyed by
electrostatic discharges
units must be disconnected from the power
source
indicates special instructions or information for
operation at low temperatures.
indicates basic conditions or procedures are
being described.
This symbol may also indicate information important for achieving accurate measurements.
Emerson Process Management GmbH & Co. OHGS-4
Page 13
Short Form Manual
HASXEE-SFM-HS
04/2010
Safety Instructions
X-STREAM XE
SAFETY INSTRUCTIONS
INTENDED USE STATEMENT
X-STREAM XE series gas analyzers are intended to be used as analyzers for industrial
purposes. They must not be used in medical, diagnostic or life support applications.
Using X-STREAM XE analyzers as safety devices is prohibited where redundancy and/or
SIL classication or equivalent is needed.
No independent agency certications or approvals are to be implied as covering such
applications!
Safety Instructions
S
GENERAL SAFETY NOTICE / RESIDUAL RISK
If this equipment is used in a manner not specied in these instructions, protective systems may be impaired.
Despite of incoming goods inspections, production control, routine tests and application
of state-of-the-art measuring and test methods, an element of risk remains when operating
a gas analyzer!
Even when operated as intended and observing all applicable safety instructions some
residual risks remain, including, but not limited to, the following:
• An interruption of the protective earth line, e.g. in an extension cable, may result in
risk to the user.
• Live parts are accessible when operating the instrument with doors open or covers
removed.
• The emission of gases hazardous to health may even be possible when all gas connections have been correctly made.
Avoid exposure to the dangers of these residual risks by taking particular care when installing, operating, maintaining and servicing the analyzer.
Emerson Process Management GmbH & Co. OHG S-5
Page 14
Short Form Manual
X-STREAM XE
Safety Instructions
HASXEE-SFM-HS
04/2010
AUTHORIZED PERSONNEL
In-depth specialist knowledge is an absolutely necessary condition for working with and
on the analyzer.
Authorized personnel for installing, operating, servicing and maintaining the analyzer are
instructed and trained qualied personnel of the operating company and the manufacturer.
It is the responsibility of the operating company to
• train staff,
• observe safety regulations,
• follow the instruction manual.
Operators must
• have been trained,
• have read and understood all relevant sections of the instruction manual before
commencing work,
• know the safety mechanisms and regulations.
To avoid personal injury and loss of property, do not install, operate, maintain or service
this instrument before reading and understanding this instruction manual and receiving
appropriate training.
ADDITIONAL LITERATURE
This manual covers aspects important for installation and startup of X-STREAM XE gas
analyzers.
For comprehensive information on operating and maintain/service the instrument in a
safe manner it is MANDATORY to read all additional instruction manuals, if not provided
as printed version, see the accompanying CD-ROM for an electronic version (PDF)!
The following additional instruction manuals are available or referenced within this manual:
HASXEE-IM-HS X-STREAM XE series instruction manual
HASICx-IM-H Infallible containment instruction manual
Contact your local service center or sales ofce when missing documents.
SAVE ALL INSTRUCTIONS FOR FUTURE USE!
Emerson Process Management GmbH & Co. OHGS-6
Page 15
Short Form Manual
HASXEE-SFM-HS
04/2010
X-STREAM XE
Safety Instructions
OPERATING AND MAINTAINING THIS UNIT
On leaving our factory, this instrument conformed to all applicable safety directives.
In order to preserve this state of affairs, the
operator must take care to follow all the instructions and notes given in this manual and
on the unit.
Before switching on the unit, ensure that the
local nominal mains voltage corresponds to
the factory-set operational voltage of this
unit.
Any interruption of the protective earth connections, whether inside or outside of the unit,
may result in exposure to the risk of electricity.
Deliberately disconnected the protective earth
is therefore strictly forbidden.
Removing covers may expose components
conducting electric current. Connectors may
also be energised. The unit should therefore
be disconnected from the power supply before
any kind of maintenance, repair or calibration
work requiring access to the inside of the
unit.
Only trained personnel who are aware of
the risk involved may work on an open and
energized unit.
Fuses may only be replaced by fuses of an
identical type and with identical ratings. It is
forbidden to use repair fuses or to bypass
fuses.
Take note of all applicable regulations when
using this unit with an autotransformer or a
variable transformer.
Substances hazardous to health may escape
from the unit’s gas outlet. This may require
additional steps to be taken to guarantee the
safety of operating staff.
Safety Instructions
S
Emerson Process Management GmbH & Co. OHG S-7
Page 16
X-STREAM XE
The units described in this manual may not be used in explosive atmospheres
without additional safety measures.
Do not operate without covers secure.
Short Form Manual
HASXEE-SFM-HS
04/2010
Safety Instructions
EXPLOSION HAZARD
ELECTRICAL SHOCK HAZARD
Do not open while energized.
Installation requires access to live parts which can cause death or serious
injury.
For safety and proper performace this instrument must be connected to a
properly grounded three-wire source of power.
TOXIC GASES
This unit’s exhaust may contain toxic gases such as (but not limited to)
e.g. sulfur dioxide. These gases can cause serious injuries.
Avoid inhaling exhaust gases.
Connect the exhaust pipe to a suitable ue and inspect the pipes regularly
for leaks.
All connections must be airtight to avoid leaks;
X-STREAM XE series
instruction manual on how performing a leak test.
Emerson Process Management GmbH & Co. OHGS-8
Page 17
Short Form Manual
HASXEE-SFM-HS
04/2010
The models intended for outside and wall mounted use (X-STREAM XEF and
XDF) weigh between 25 kg (57 lb) and 45 kg (99 lb), depending on version
and options installed.
X-STREAM XE
Safety Instructions
HEAVY INSTRUMENT
Safety Instructions
Two people and/or lifting equipment is required to lift and carry these
units.
Take care to use anchors and bolts specied to be used for the weight of
the units!
Take care the wall or stand the unit is intended to be installed at is solid
and stable to support the weight!
CRUSHING HAZARD
Take care of crushing hazard when closing the front door of analyzer eld
housings!
Keep out of the closing area between enclosure cover and base!
OPERATION AT LOW TEMPERATURES
When operating an instrument at temperatures below 0 °C (32 °F), do
NOT apply gas nor operate the internal pump before the warmup time has
elapsed!
S
Violation may result in condensation inside the gas paths or damaged
pump diaphragm!
HIGH TEMPERATURES
Hot parts may be exposed when working on photometers and/or
heated components in the unit.
Emerson Process Management GmbH & Co. OHG S-9
Page 18
X-STREAM XE
Safety Instructions
GASES AND PREPARATION OF GASES
GASES HAZARDOUS TO HEALTH
Follow the safety precautions for all gases (sample and span gases) and
gas cylinders.
Before opening the gas lines, they must be purged with air or neutral gas (N2)
to avoid hazards from escaping toxic, ammable, explosive or hazardous
gases.
Short Form Manual
HASXEE-SFM-HS
04/2010
FLAMMABLE OR EXPLOSIVE GASES
When supplying explosive gases or ammable gases with concentrations of
more than 25 % of the lower explosion limit, we RECOMMEND implementing
one or more additional safety measures:
• purging the unit with inert gas
• stainless steel internal pipes
• ame arrestors on gas inlets and outlets
• infallible measuring cells.
Emerson Process Management GmbH & Co. OHGS-10
Page 19
Short Form Manual
HASXEE-SFM-HS
04/2010
POWER SUPPLY
X-STREAM XE
Safety Instructions
CONNECTING UNITS FOR PERMANENT INSTALLATION
Only qualied personnel following all applicable and legal regulations may
install the unit and connect it to power and signal cables. Failure to comply
may invalidate the unit’s warranty and cause exposure to the risk of damage,
injury or death.
This unit may only be installed by qualied personnel familiar with the
possible risks.
Working on units equipped with screw-type terminals for electrical
connections may require the exposure of energized components.
Wall-mounted units have no power switch and are operational when
connected to a power supply. The operating company is therefore required
to have a power switch or circuit breaker (as per IEC 60947-1/-3) available
on the premises. This must be installed near the unit, easily accessible to
operators and labelled as a power cut-off for the analyzer.
HAZARD FROM WRONG SUPPLY VOLTAGE
Ensure that the local power voltage where the unit is to be installed,
corresponds to the unit’s nominal voltage as given on the name plate
label.
Safety Instructions
S
ADDITIONAL NOTES FOR UNITS WITH SCREW-TYPE TERMINALS
Cables for external data processing must be double-insulated against mains
power.
If this is not possible, cables must be laid in such a way as to guarantee
a clearance of at least 5 mm from power cables. This clearance must be
permanently secured (e.g. with cable ties).
Emerson Process Management GmbH & Co. OHG S-11
Page 20
Short Form Manual
X-STREAM XE
HASXEE-SFM-HS
04/2010
General Operating Notes
GENERAL OPERATING NOTES
EXPLOSION HAZARD BY EXHAUST GASES
Exhaust gases may contain hydrocarbons, other toxic gases or even
explosive gases.
Faulty gas connections may lead to explosion and death.
Ensure that all gas connections are connected as labelled and airtight.
• The unit must be installed in a clean and dry area, protected from strong vibrations and
frost
.
• The unit must not be exposed to direct sunlight and sources of heat. Admissible ambient
temperatures (see technical details) must be adhered to.
• Gas inlets and outlets must not be interchanged.All gases must be conditioned, before supplied to the unit. When using this unit with corrosive sample gases, ensure that these gases
do not contain components harmful to the gas lines.
• Admissible gas pressure for sample and test gases is 1,500 hPa.
• Exhaust lines must be laid inclined downwards, depressurized, protected from frost and
according to applicable regulations.
• If it is necessary to disconnect the gas lines, the unit’s gas connectors must be sealed with
PVC caps to avoid polluting the internal gas lines with condensate, dust, etc.
• To ensure electromagnetic compatibility (EMC), only shielded cables (supplied by us on request, or of equivalent standard) may be used. The customer must ensure that the shielding
is correctly tted. Shielding and terminal housing must be electrically connected; submin-D
plugs and sockets must be screwed to the unit.
• When using optional external adapters (submin-D to screw-type terminals), protection against
electromagnetic interference is not ensured (CE compliance pursuant to EMC guidelines).
In this case the customer or operating company functions as a system builder and must
therefore ensure and declare compliance with EMC guidelines.
Emerson Process Management GmbH & Co. OHGS-12
Page 21
Short Form Manual
HASXEE-SFM-HS
04/2010
Chapter 1
Technical description
The following are the main features of the new
Emerson Process Management X-STREAM
XE series gas analyzers in brief:
• compact design with easily accessible
internal components
• customizable for a wide range of applications: different housings are available
while internal construction remains largely
identical
• a highly integrated mainboard contains
all interfaces and basic functions for the
operation of the unit
• multilingual microprocessor-controlled
user interface with liquid crystal (LCD) to
indicate measurement value and status
messages
• units for outdoor use are optionally supplied with an impact tested front panel
• widerange power supply unit for worldwide
use without modication (1⁄2 19in units with
external PSUs)
The new X-STREAM XE gas analyzers can
measure up to ve different gas components using any combination of the following
analyzing techniques (restrictions apply to
lable for use with corrosive gases and/or
gases containing solvents.
Special congurations (e.g. intrinsically safe
or infallible measuring cells) for the analysis
of combustible gases are also available.
Standard applications
Different housings allow X-STREAM XE
analyzers to be tailored to the many different
applications:
• Tabletop units in 1⁄2 19in modular design,
• Tabletop and rackmountable units in 19in
• Stainless steel wallmountable eld housing
• Cast aluminium wallmountable field
The various analyzer types are described in
more detail beginning with 1-3.
electrochemical oxygen analysis
with IP 20 protection class
modular design, with IP 20 protection
class
with IP 66 / NEMA 4X protection class
for outdoor use (operating temperature
0 (-20) °C to +50°C).
housing with IP 66 / NEMA 4X protection
class for outdoor use (operating temperature 0 (-20) °C to +50 °C).
Techn. Description
1
EXPLOSION HAZARD
X-STREAM XE analyzers CAN NOT be used in explosive environments
(hazardous areas) without additional safety features.
This instruction manual does NOT describe the special conditions necessary
to operate gas analyzers in hazardous areas.
Please refer to the separate instruction manual, supplied with units for use
in hazardous areas.
Emerson Process Management GmbH & Co. OHG 1-1
Page 22
X-STREAM XE
1.1 Overview
Short Form Manual
HASXEE-SFM-HS
04/2010
1.1 Overview
All X-STREAM gas analyzers feature an
easy-to-use graphical user interface, which
displays measurement values, status and
error messages, and menus for the input of
parameters.
For ease of use, the operator can select one
of three languages for the display: By default
any analyzer is congured with English and
1.1.1 The Front Panel
The graphic LCD shows measurement and
status information with plain text and symbols.
The symbols are designed to indicate the different status “Failure”, “Function check”, “Out
of specication” and “Maintenance request” as
specied by the NE 107 standard. For further
information,
X-STREAM XE instruction
manual.
German language sets, while a third can
optionally be added. Currently available:
French, Italian, Portuguese and Spanish..
The analyzer software is operated by means
of only six keys.
Depending on the analyzer model, the display
can also be protected with an impact tested
glass panel, to withstand even harsher conditions and to provide a higher IP protection class
of up to IP66 / NEMA Type 4X.
1342
1 Graphic display
2 “Measure“ key
3 “Enter” key
4 4 keys for settings and menu navigation
Fig. 1-1: X-STREAM front panel (here the X-STREAM XEGP)
Emerson Process Management GmbH & Co. OHG1-2
Page 23
X-STREAM XE
1.2 Comparison of Analyzer Models
1.2 Comparison of the Various X-STREAM Analyzer Model
Short Form Manual
HASXEE-SFM-HS
04/2010
X-STREAM XEGC
1
⁄2 19 in housing, table-top or rackmountable,
protection type: IP 20
1
⁄1 19 in housing, table-top or rackmountable,
protection type: IP 20
X-STREAM XEGP
24V input with external power supply unitInternal wide range power supply unitInternal wide range power supply unitInternal wide range power supply unit
Max. 3 channels: 2 photometer + 1 WLD/O
2
max. 6 gas connections,
including 1 optional purge gas connection
Options for gas lines: Flow sensor, pressure
sensor, infallible gas lines
Max. 4 channels in any combination
max. 8 gas connections,
1 optional extra connection for purge gas
Options for gas lines: Flow sensor, pressu-
re sensor, heating for physical components,
sample gas pump, 1 or 2 valve blocks, infallible gas lines
1-4 analog outputs, 4 relay outputs, 2 Modbus
Ethernet interfaces, 2 USB connectors
optional:
1 interface card with 7 digital inputs and
9 digital outputs
1 interface card with analog inputs
electrical interfaces accessible via sockets on
back of unit, optionally: screw-type terminal
adapters
(except for Ethernet & USB)
1-4 analog outputs, 4 relay outputs, 2 Modbus
Ethernet interfaces, 2 USB connectors
optional:
1 or 2 interface cards, each with 7 digital
inputs and 9 digital outputs
1 interface card with analog inputs
electrical interfaces accessible via sockets on
back of unit, optionally: screw-type terminal
adapters
(except for Ethernet & USB)
LCDLCDLCD, impact tested front panelLCD, impact tested front panel
Operating ambient temperature:
0 °C to +50 °C (32 °F to 122 °F)
Operating ambient temperature:
0 °C to +50 °C (32 °F to 122 °F)
Size: (DxHxW): max. ca. 440x130x220 mm
Weight: ca. 8 - 12 kg (17.6 - 26.5 lb)
Size: (DxHxW): max. ca. 411x133x482 mm
Weight: ca. 11 - 16 kg (24.3 - 35.3 lb)
Max. 4 channels in any combination
max. 8 gas connections,
1 optional extra connection for purge gas
Options for gas lines: Flow sensor, pressure
sensor, heating for physical components,
sample gas pump, 1 or 2 valve blocks, infallible gas lines
1-4 analog outputs, 4 relay outputs, 2 Modbus
Ethernet interfaces, 2 USB connectors
optional:
1 or 2 interface cards, each with 7 digital
inputs and 9 digital outputs
1 interface card with analog inputs
electrical interfaces on internal screw-type terminal adapters
Operating ambient temperature:
-20 °C to +50 °C (-4 °F to 122 °F)
Models available for use in explosive environ-
ments
Size: (DxHxW): ca. 265x400 (815)x550 mm
Weight: max. ca. 25 (45) kg / 55.1 (99.2) lb
(except for Ethernet & USB)
Max. 4 channels in any combination
max. 8 gas connections,
including 2 optional purge gas connection
Options for gas lines: Flow sensor, pressu-
re sensor, heating for physical components,
sample gas pump, 1 or 2 valve blocks, infallible gas lines
1-4 analog outputs, 4 relay outputs, 2 Modbus
Ethernet interfaces, 2 USB connectors
optional:
1 or 2 interface cards, each with 7 digital
inputs and 9 digital outputs
1 interface card with analog inputs
electrical interfaces on internal screw-type terminal adapters
Operating ambient temperature:
-20 °C to +50 °C (-4 °F to 122 °F)
Flameproof enclosure: approved for use in
explosive environments
Size: (DxHxW): max. ca. 222x512x578 mm
Weight: max. ca. 63 kg (138.9 lb)
This model is not covered by this manual!
See the separate X-STREAM Ex d manual
for hazardous area applications!
(except for Ethernet & USB)
Emerson Process Management GmbH & Co. OHG 1-4
Page 25
Short Form Manual
HASXEE-SFM-HS
04/2010
1.3 X-STREAM XEGC
1.3 X-STREAM XEGC: ½ 19 Inch Table-Top Unit
X-STREAM XE
This compact model for general purposes
can be tted with up to three photometer
measurement channels (IR or UV) and one
additional non-photometer channel (eO
, pO2
2
or WLD). Power is supplied by a separate
external power supply unit.
By default the units are congured for tabletop
use. For rack mounting an accessory kit can
be ordered to either install 2 units alongside,
or one single unit together with a blind plate.
Connection to power supply
DC 24 V power is supplied via a 3-pin socket
at the rear of the unit.
Interfaces
Electrical connections for interface signals are
provided via submin-D connectors, Ethernet
and USB connectors mounted on the rear
panel of the device
(g. 1-5).
For applications where screw-type terminals
are preferred, optional adapters are available,
which are mounted directly onto the submin-D
connectors.
Techn. Description
1
134562
1 Status line
2 Graphic display
3 Messages line
4 home key
5 enter key
6 4 keys for adjustment and menu selection
Fig. 1-2: X-STREAM XEGC
Emerson Process Management GmbH & Co. OHG 1-5
7 Signal connectors (some optional)
8 Gas ttings
9 Power input fuse
10 Power input connector
78910
Page 26
X-STREAM XE
1.4 X-STREAM XEGP
1.4 X-STREAM XEGP: 19 Inch Table-Top Or Rackmount Design
Short Form Manual
HASXEE-SFM-HS
04/2010
This model can be fitted with up to four
measurement channels in any combination.
The physical components can optionally be
encased in a cover. This area can be held at
a specic temperature of up to 60 °C to minimize interference from changes in external
temperature.
Units congured for rack mounting can be
converted for tabletop use by removing the
lateral mounting brackets and attaching the
four feet supplied as accessories.
Front view
Connection to power supply
Main power is supplied via the IEC chassis
plug mounted on the rear panel, with integrated power switch and fuse holders. The
internal wide range power supply unit enables
the analyzers to be used worldwide.
Adaptors are optionally available for signal
connections, to provide screw terminals instead of submin-d-connectors (
1-7).
134562
1 Status line
2 Graphic display
3 Messages line
Rear view
1 Gas connector ttings
2 Space for additional ttings
3 Optional purge gas inlet
Fig. 1-3: X-STREAM XEGP
4 home key
5 enter key
6 4 keys for adjustment and menu selection
15243
4 Power inlet with lter, fuses & switch
5 Signal input/output connectors (some optional)
Emerson Process Management GmbH & Co. OHG1-6
Page 27
Short Form Manual
HASXEE-SFM-HS
04/2010
X-STREAM XE
1.4 X-STREAM XEGP
Strain-reliefs, top view details
Techn. Description
1
1
1
Fig. 1-4: Optional screw-type terminal adapters
2
1 Screw-type terminal adapters
2 Strain-reliefs
Emerson Process Management GmbH & Co. OHG 1-7
Page 28
Short Form Manual
X-STREAM XE
1.5 X-STREAM XE Field Housing
1.5 X-STREAM XEF/XDF: Field Housing With Single Or Dual Compartment
HASXEE-SFM-HS
04/2010
The eld housing model is conceived for
outdoor use and wall-mounting. The coated stainless steel housing has a protection
class rated IP66 / NEMA Type 4X, offering
protection against water and dust entering
the device:
IPx6: Water projected in powerful jets against
the enclosure from any directions shall have
no harmful effects
IP6x: Protection against penetration by dust.
Live or internal moving parts are completely
protected.
X-STREAM eld housings can be tted with
up to four measurement channels in any combination. The physical components can optionally be encased in a cover. This separate
volume can be held at a specic temperature
of up to 60 °C to minimize interference from
changes in external temperature.
Front panel
The analyzer ’s display can o
ptionally be
protected by an impact tested glass for enhanced protection against breakage in harsh
environments.
Connection to power supply
Mains power is supplied via screw-type terminals with integrated fuse holders at the right
of the housing, near the front. The wide range
power supply unit mounted internally enables
the analyzers to be used worldwide.
134562
1 Status line
2 Graphic display
3 Messages line
4 home key
5 enter key
6 4 keys for adjustment
and menu selection
Fig. 1-5: X-STREAM XEF / XDF- Front views
Emerson Process Management GmbH & Co. OHG1-8
Page 29
Short Form Manual
HASXEE-SFM-HS
04/2010
X-STREAM XE
1.5 X-STREAM XE Field Housing
Techn. Description
1
213
1 Cable gland for power cable
2 Cable glands for signal cables
3 4 brackets for wall-mounting
4 Gas in- & outlets (max. 8)
5 Cutouts, to combine 2 housings (here closed)
Fig. 1-6: X-STREAM XEF - right side and bottom view
HEAVY INSTRUMENT
The X-STREAM eld housing, intended for outside and wall mounted use,
weighs approx. 26 kg/57 lb (XEF) or 45 kg/99 lb (XDF), depending on options
installed.
Two people and/or lifting equipment is required to lift and carry these units.
Take care to use anchors and bolts specied to be used for the weight of
the units!
545
Note!
In case of XDF, the cable glands are located
at the upper compartment, while the gas in& outlets are at the bottom side of the lower
compartment.
Also only 2 brackets are at each compartment.
Take care the wall or stand the unit is intended to be installed at is solid
and stable to support the weight!
Emerson Process Management GmbH & Co. OHG 1-9
Page 30
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
04/2010
1.7 X-STREAM XE Field Housing
13245
(shown with front
panel removed)
1 Screw-type terminals for signal cables
2 Power line lter
3 Cable glands
4 Power supply terminals with integrated fuses
5 Ethernet and USB connectors
Fig. 1-7: X-STREAM XEF - power supply and signal terminals
Note!
In case of XDF, the terminals and
connectors are located at the upper
compartment, while physical components and gas ttings are in the lower
compartment.
Emerson Process Management GmbH & Co. OHG1-10
Page 31
Short Form Manual
HASXEE-SFM-HS
04/2010
1.7 Field Housings in Hazardous Areas
1.5.1 Field Housings for Installation in Hazardous Areas (Ex-Zones)
EXPLOSION HAZARD
X-STREAM XEF and XDF models MUST NOT be used in explosive environments (hazardous areas) without additional safety features.
X-STREAM XE
This instruction manual does NOT describe the special conditions necessary
to operate gas analyzers in hazardous areas.
Please refer to the separate instruction manual supplied with units for use
in hazardous areas.
Special X-STREAM eld housing analyzer variations can be used in Ex-zones 1, 2 or Div 2:
X-STREAM XEFN/XDFN:
These analyzer variations feature a protection
concept called "non-incendive", which means
that non-sparking and non-arcing components, installed within a protecting enclosure,
enable installation in an European Ex-zone 2
for measuring non-ammable gases. No further
measures, such as a supply of protective gas,
are necessary.
X-STREAM XEFS/XDFS:
Equipped with a simplied pressurization system,
these eld housings can be used to measure
non-ammable gases in European Ex-zone 2.
A protective gas (e.g. pressurized air) must be
supplied when operating this model.
Techn. Description
1
X-STREAM XEFZ/XDFZ:
Equipped with a simplied pressurization system, these models can be used to measure
non-ammable gases in American zone Div 2.
A protective gas (e.g. pressurized air) must be
supplied when operating this model.
Please contact your local EMERSON Process
Management ofce if you require analyzers
for use in hazardous areas.
Emerson Process Management GmbH & Co. OHG 1-11
Page 32
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
04/2010
1 Technical Description
Emerson Process Management GmbH & Co. OHG1-12
Page 33
Short Form Manual
HASXEE-SFM-HS
07/2010
X-STREAM XE
Chapter 2
Installation
This chapter describes the correct installation procedure for the various X-STREAM analyzer
versions.
On receipt, check the packaging and its contents thoroughly for damage.
Inform the carrier immediately of any damage to packaging or contents, and keep dama-
ged parts until clarication.
Store the instrument at a dry and clean place, considering the acceptable environmental
conditions.
We recommend to keep the packaging available for future transportation, because only
the original packaging ensures proper protection!
2.1 Scope of Supply
Compare the content of your package with the pictures below.
Analyzers for hazardous areas need additional parts, described in the
accompanying documentation refering to hazardous area installations.
Call your local sales ofce if something is missing, and DO NOT continue
to install your analyzer, until all parts are at hand!
Installation
2
HAZARDS FROM MISSING INFORMATION
Manuals, some of which either as paper
Analyzer
Fig. 2-1: X-STREAM XE Analyzers - scope of supply
Emerson Process Management GmbH & Co. OHG 2-1
or CD-ROM version:
• this short form manual
• X-STREAM XE series manual
• infallible containment instruction manual (if applicable to your instrument)
Page 34
X-STREAM XE
2.2 Introduction
Before connecting the analyzer to mains power, please read the chapter on
safety warnings and the following instructions carefully.
Disregarding may cause death, personnel injury or damage to equipment!
Short Form Manual
HASXEE-SFM-HS
07/2010
2.2 Installation - Introduction
ELECTRICAL SHOCK HAZARD
The place of installation must be clean, dry and protected against strong
vibrations and frost. Please observe the admissible operating temperatures
given in the technical data section.
Units must not be subjected to direct sunlight or sources of heat.
For outdoor installation it is recommended to install the unit in a cabinet.
It should at least be protected against rainfall.
In order to comply with regulations on electromagnetic compatibility, it is recommended
to use only shielded cables which can be
supplied by Emerson Process Management.
The customer must ensure that the shielding
is correctly connected to the signal cable plug
housing. Submin-d plugs and sockets must
be screwed to the analyzer.
The use of external submin-d to screw-type
terminal adapters affects electromagnetic
compatibility. In such a case the customer
must take appropriate measures to comply
with the regulations, and must declare conformity when this is legally required (e.g.
European EMC guidelines).
Emerson Process Management GmbH & Co. OHG2-2
Page 35
Short Form Manual
HASXEE-SFM-HS
07/2010
2.3 Installation - Technical Data
2.3 Technical Data
Before intending to start to install the analyzer,
verify the site of installation meets the specications, given in the following sections!
2.3.1 Common Technical Data
Site of installation
Humidity
(non-condensing)
Degree of pollution2
Installation categoryII
Elevation0 to 2000 m (6560 ft) above sea level
Ambient atmosphere
Certication
Electrical safety
< 90 % RH at +20 °C (68 °F)
< 70 % RH at +40 °C (104 °F)
Units may not be operated in corrosive,
ammable or explosive environments
without additional safety measures.
X-STREAM XE
Installation
2
CAN / USA
Europe
Electromagnetic
compatibility
EuropeCE, based on EN 61326
AustraliaC-Tick
othersNAMUR
CSA-C/US, based on CAN/CSA-C22.2 No.
61010-1-04 / UL 61010-1, 2nd edition
CE, based on EN 61010-1
Emerson Process Management GmbH & Co. OHG 2-3
Page 36
X-STREAM XE
2.3 Installation - Technical Data
2.3.2 Specic Technical Data
Temperatures
operational
Short Form Manual
HASXEE-SFM-HS
XEGCXEGPXEF, XDF
0 ... +50 °C
32 ... 122 °F
0 ... +50 °C
32 ... 122 °F
0 (-20) ... +50 °C
32 (-4) ... 122 °F
07/2010
storage
Weight, max
IP or Type rating
-20 ... +70 °C
-4 ... -158 °F
8 ... 12 kg
17.6 ... 26.5 lb
IP 20
indoor use,
protected against
dripping water and
direct sun light
-20 ... +70 °C
-4 ... -158 °F
12 ... 16 kg
26.5 ... 35.3 lb
IP 20
indoor use ,
protected against
dripping water and
direct sun light
-20 ... +70 °C
-4 ... 158 °F
XEF: 25 kg / 55.1 lb
XDF: 45 kg / 99.2 lb
IP 66 / Type 4X
outdoor use ,
protected against
direct sun light
Gas connections
max number688
max for purging
(incl. / separate)
material
sizes6/4 mm;
Power supply unit
2 incl.1 separate, 1 incl.1 separate, 1 incl.
PVDF; stainless
steel (opt.)
1
⁄4"6/4 mm; 1⁄4"6/4 mm; 1⁄4"
wide range,
external
PVDF; stainless
steel (opt.)
wide range,
internal
stainless steel
wide range,
Power supply
internal
nominal voltageDC 24 V
voltage rangeDC 10 ... 30 V
100 - 240 V
50 / 60 Hz
85 - 264 V
47 - 63 Hz
nominal input current
standard, max2.5 A1.3 - 0.7 A
w/ temperature control, maxn.a.3 - 1.5 A
Power input fuses
AC 230 V
T 3.15 A
5x20 mm
AC 230 V
T 4 A
5x20 mm
Emerson Process Management GmbH & Co. OHG2-4
100 - 240 V
50 / 60 Hz
85 - 264 V
47 - 63 Hz
XEF: 1.3 - 0.7 A
XDF: 1.5 - 0.8 A
XEF: 3 - 1.5 A
XDF: 5.5 - 3 A
AC 230 V
T 6.3 A
5x20 mm
Page 37
Short Form Manual
HASXEE-SFM-HS
07/2010
2.4 Gas Conditioning
X-STREAM XE
2.4 Installation - Gas Conditioning
In order to ensure trouble-free operation, special attention must be paid to the preparation
of the gases:
All gases must be conditioned
before supplying to the analyzer, to be
• dry,
If moisture cannot be avoided, it is necessary
to ensure that the dew point of the gases is at
least 10 °C (18 °F) below the ambient temperature to avoid condensate in the gas lines.
The X-STREAM eld housings can optionally
be tted with heated piping to enable the use
of gases with a maximum dew point of 25 °C
(77 °F).
• free of dust and
• free of any aggressive com-
See the X-STREAM XE instruction manual,
chapter 3 for more information.
ponents which may damage
the gas lines (e.g. by corrosion or solvents) .
Flammable gases must not be
supplied without additional
protective measures.
Summary of the Most Important Gas Parameters
NDIR/UV/VISOxygen Sensor (PO2 and EO2)Thermal Conductivity
Permissible gas ow0.2 - 1.5 l/min.0.2 - 1.0 l/min
Maximum gas pressure
Permissible ambient
temperature
Thermostat control
Warm-up time15 to 50 minutes
1
Paramagnetic oxygen measurement (PO2)
2
Electrochemical oxygen measurement (EO2)
3
Special conditions apply to model XEFD
9
< 1,500 hPa abs. (< 7 psig)
0 to +50 °C (32 to 122 °F) 0 to +50 °C (32 to 122 °F)
6
Optionally 60 °C (140 °F) 5 60 °C (140 °F) 1 / None
The purging medium (e.g. to minimize CO
interference or for enhanced safety when
measuring corrosive or poisonous gases)
must be dry, clean and free of corrosives
or components containing solvents.
To minimize cross interferences the purge
2
Its temperature must correspond to the ambient temperature of the analyzer, but be at
least within the range 20–35 °C (68–95 °F).
For information about values for pressure and
ow, please contact your nearest EMERSON
Process Management sales ofce.
gas also has to be free of components to
be measured.
Emerson Process Management GmbH & Co. OHG 2-5
Page 38
X-STREAM XE
Take care that all external gas pipes are connected in the described way
and that they are gastight to avoid leakages!
Faulty connected gas pipes lead to explosion hazard or even to mortal
danger!
Don‘t take a breathe of the emissions! Emissions may contain hydrocarbons
or other toxic components (e.g. carbon monoxide)! Carbon monoxide may
cause headache, sickness, unconsciousness and death.
Short Form Manual
HASXEE-SFM-HS
07/2010
2.4 Installation - Gas Conditioning
TOXIC GAS HAZARDS
Do not confuse gas inlets and outlets. All gases supplied must be prepared
beforehand. When supplying aggressive gases, ensure that the gas lines
are not damaged.
Max. admissable pressure: 1500 hPa / 7 psig!
Exhaust lines must be installed to incline downwards and be unpressurized
and protected against frost, and conform to legal requirements.
The number of gas connections and their
conguration may vary according to analyzer
version and installed options.
All gas connectors are labelled and can be
found on the
analyzer’s rear panel (X-STREAM XEGP,
•
X-STREAM XEGC)
• underside of the analyzer (X-STREAM eld
housings)
Should it be necessary to open the gas lines,
the gas connectors should be sealed with
PVC caps to prevent pollution by moisture,
dust, etc.
Fig. 2-2: Labelling of gas connectors (example)
Emerson Process Management GmbH & Co. OHG2-6
Page 39
Short Form Manual
HASXEE-SFM-HS
07/2010
X-STREAM XE
2.4 Installation - Gas Conditioning
The analyzer should be mounted close to the
sample gas source to minimize transport time.
A sample gas pump can be used to reduce the
reaction time; this requires that the analyzer
be operated in bypass mode or tted with a
Pressure control valve
Filter
Sample gas pump
Fig. 2-3: Installation in bypass mode
pressure control valve to protect against excessive gas ow and pressure (g. 1-2).
Exhaust
Analyzer
Flow sensor
Exhaust
Installation
2
Emerson Process Management GmbH & Co. OHG 2-7
Page 40
X-STREAM XE
2.5 Installation - Electrical Connections
2.5 Electrical Connections
ELECTRICAL SHOCK HAZARD BY INSTALLATION
Only qualied personnel, observing all applicable technical and legal
requirements, may install these devices and connect power and signal
cables.
Failure to comply may render the guarantee void and cause exposure to
risk of damage, injury or death.
The devices may only be installed by personnel who are aware of the
possible risks. Working on units with screw-type terminals for electrical
connections may require exposure to energized components.
Short Form Manual
HASXEE-SFM-HS
07/2010
Wall-mountable X-STREAM analyzers are not tted with power switches and
are operational as soon as they are connected to a power supply. A switch
or circuit breaker (IEC 60947-1/-3) must be installed on the premises. The
switch or breaker must be located near the analyzer, easily accessible and
labelled as a power supply cut-off for the analyzer.
Units with screw-type terminals must be de-energized by unplugging it or
operating the separate cut-off switch or circuit breaker, when working on
the power connections.
To avoid the risk of electrical shock, all units must be earthed. For this
reason, a power cable with protective earth conductor must be used.
Any break in the earth wire inside or outside the unit may cause exposure
to the risk of electrocution and is therefore prohibited.
Emerson Process Management GmbH & Co. OHG2-8
Page 41
Short Form Manual
HASXEE-SFM-HS
07/2010
2.6 Detailed Instructions for Installation
2.6 Detailed Instructions for Installation
Important note for X-STREAM XEFD!
Due to the special conditions which must be observed when installing units in EX zones, the
installation of the ameproof X-STREAM XEFD version is described in a separate instruction manual HASXEDE-IM-EX.
Even if you do not install your X-STREAM XEFD in an EX zone, please install the unit according
to the instructions in the separate manual.
Notes for wiring signal inputs and outputs page 2-30
Emerson Process Management GmbH & Co. OHG 2-9
Page 42
X-STREAM XE
2.6.1 Installation - X-STREAM XEGC, X-STREAM XEGP
2.6.1 X-STREAM XEGC, X-STREAM XEGP
Short Form Manual
HASXEE-SFM-HS
07/2010
Plugs and sockets required for the electrical
connections are on the rear panel of the units
(g. 2-4 & 2-5).
X-STREAM XEGC analyzers are powered by
an external 24 V power supply unit which can
be supplied with the unit as an option. If the
PSU is not included in delivery, another unit
can be used instead, provided it conforms to
the specications given in the X-STREAM XE
instruction manual.
Digital in- /
Service interface
Analog outputs / relay
outputs
USB connectors
RJ45 ethernet connectors
outputs
Gas inlets
X-STREAM XEGP analyzers provide an internal widerange power supply for worldwide
use.
X-STREAM XEGC / XEGP analyzers should
be operated in a horizontal position.
The optional brackets enable the units to be
mounted in a rack; this is accomplished by
means of four screws (g. 2-5 & 2-6).
HAZARD BY INSTALLATION
The brackets are not designed
to carry the weight of the
instrument!
Support the instrument, when
rack mounting!
Disregarding may cause personal injury and damaged
equipment.
Fuse
Power in
interface
Modbus 485)
Modbus interface
(RS 485 / RS 232 /
Fig. 2-4: X-STREAM XEGC - rear panel
Serial CAN
Gas outlets
213
1: ME
2: + 24 V
3: 0 V ( )
Emerson Process Management GmbH & Co. OHG2-10
Page 43
Short Form Manual
HASXEE-SFM-HS
07/2010
2.6.1 Installation - X-STREAM XEGC, X-STREAM XEGP
USB connectors
Analog outputs / NA-
Services interface
MUR status relays
Digital inputs /
RJ45 ethernet connectors
outputs
Gas inlets
Gas outlets
X-STREAM XE
(optional)
Purge gas inlet
Installation
2
4 screw holes for rack mounting
(7.5 x 10.5 mm / 0.29" x 0.41")
Fuses
interface
Modbus 485)
Modbus interface
(RS 485 / RS 232 /
Fig. 2-5: X-STREAM XEGP - rear panel, shown with optional terminal adapters and
front side brackets for rack mounting
Serial CAN
for signal cables
Strain reliefs and
shielding contacts
Power switch
Power in
The number and conguration of the gas inlets and outlets vary from model to model and are
indicated on the notice on the rear of the instrument.
To simplify installation, we recommend labelling the gas lines as in the gures above (1, 2, 3, ...).
This avoids confusion in case the analyzer ever has to be disconnected.
X2GCX2GP
Gas connections
max number68
max for purging (incl. / separate)2 incl.1 incl. & 1 separate
materialPVDF; stainless steel (opt.)
1
sizes6/4 mm;
⁄4"
Emerson Process Management GmbH & Co. OHG 2-11
Page 44
X-STREAM XE
2.6.1 Installation - X-STREAM XEGC, X-STREAM XEGP
Drawings show tabletop version
with strain-reliefs and screwtype terminals.
Short Form Manual
HASXEE-SFM-HS
07/2010
Analyzer with
blindplate to the left
Analyzer with
blindplate to the right
Two analyzers in a
rack
Fig. 2-6: X-STREAM XEGC - dimensions
approx. values in mm [in]
Emerson Process Management GmbH & Co. OHG2-12
Page 45
Short Form Manual
HASXEE-SFM-HS
07/2010
2.6.1 Installation - X-STREAM XEGC, X-STREAM XEGP
approx. values in mm [in]
X-STREAM XE
Installation
2
RACK INSTALLATION HAZARD
The brackets are not designed
to car ry the weight of th e
instrument!
Support the instrument, when
rack mounting!
Disregarding may cause person al inj ury and damage d
equipment.
Strain relief bracket, detail
(model with clamping adapters)
Fig. 2-7: X-STREAM XEGP - dimensions
Emerson Process Management GmbH & Co. OHG 2-13
Page 46
Short Form Manual
X-STREAM XE
HASXEE-SFM-HS
07/2010
2.6.1 Installation - X-STREAM XEGC, X-STREAM XEGP
Signal inputs and outputs
The number of signal outputs actually available varies according to the unit’s conguration.
Analog signals
Relay outputs
Analog signals and relay outputs are located on a shared 25-pin submin socket X1,
or on an optional terminals adaptor XSTA (
page 2-23).
Specication of analog signal outputs: 4 (0) - 20 mA; burden: RB ≤ 500 Ω
Specication of relay outputs 1-4:
Electrical specication:
Dry relay change-over contacts can be used as
NO or NC.
max. 30 VDC, 1 A, 30 W
Note!
Consider the installation notes in section 2.7.
SignalPin
Channel 1, (+) 4 (0) - 20 mA1
Channel 2, (+) 4 (0) - 20 mA 2
Channel 3, (+) 4 (0) - 20 mA3
Analog
outputs
Channel 4, (+) 4 (0) - 20 mA4
Channel 5, (+) 4 (0) - 20 mA5
unused6
unused7
Output1 (Failure) NC8
Output1 (Failure) NO9
Output1 (Failure) COM10
Output3 (Off spec) NC11
Output3 (Off spec) NO12
Relay contacts
(Status signals)
Output3 (Off spec) COM13
PinSignal
14Channel 1, GND
15Channel 2, GND
16Channel 3, GND
17Channel 4, GND
18Channel 5, GND
19unused
20Output2 (Maintenance request) NC
21Output2 (Maintenance request) NO
22Output2 (Maintenance request) COM
23Output4 (Function check) NC
24Output4 (Function check) NO
25Output4 (Function check) COM
Analog
Relay contacts
Note!
Conguration of relay contacts
as per standard factory setting (NAMUR
status signals)
outputs
(Status signals)
Fig. 2-8: Socket X1 - Analog & Digital Outputs 1-4
Emerson Process Management GmbH & Co. OHG2-14
Page 47
Short Form Manual
HASXEE-SFM-HS
07/2010
2.6.1 Installation - X-STREAM XEGC, X-STREAM XEGP
Serial interface
For specications and notes on control, see
chapter 7 of the X-STREAM XE in-
struction manual.
X-STREAM XE
Pin 1
Pin
6
7
8
9
Pin no.
Fig. 2-9: Plug X2 - Serial interface
MOD 485/
2 wire
1Common Common Common
2not usednot usedRXD
3not usednot usedTXD
4not usedRXD1(+) not used
5D1(+)TXD1(+) Common
6not usednot usednot used
7not usednot usednot used
8not usedRXD0(-)not used
9D0(-)TXD0(-)not used
MOD 485/
4 wire
Pin
1
2
3
4
5
RS 232
Pin 8
Pin no.
other
Ethernet connector
Signal
1TX+
2TX3RX+
6RX-
not
used
Installation
2
Notes!
Consider the installation notes in section 2.7.
When terminal adapters are used, the Serial
X-STREAM analyzers are classified DTE
(Data Terminal Equipment).
interface terminals are located on the same
adapter as those for the analog signal outputs
( Appendix ).
Then a at exible cable, attached to the
terminal adapter, is used for connecting the
illustrated 9-pole plug.
Emerson Process Management GmbH & Co. OHG 2-15
Page 48
X-STREAM XE
2.6.1 Installation - X-STREAM XEGC, X-STREAM XEGP
Digital In- and Outputs
Short Form Manual
HASXEE-SFM-HS
07/2010
7 or 14 digital inputs
(X-STREAM XEGC: max. 7 inputs)
9 or 18 additional relay outputs
(X-STREAM XEGC: max. 9 add.
outputs)
electrical
specication
electrical
specication
max. 30 V, internally limited to 2.3 mA
HIGH: min. 4 V; LOW: max. 3 V
common GND
Dry relay change-over contacts can be
used as NO or NC
max. load. 30 V; 1 A; 30 W resistive
Digital in- & outputs are located on shared 37-pin submin sockets (X4.1 or X4.2),
or on optional terminals adaptor XSTD ( page 2-26)
Notes!
Depending on model and conguration, an
analyzer may be tted with up to 2 of these
sockets (the unit is thus equipped with 14
digital inputs and 18 digital outputs).
SignalPin
Input 11
Input 22
Input 33
inputs
Digital
Input 44
GND for all digital inputs5
unused6
unused7
Output 5 NC8
Output 5 NO9
Output 5 COM10
Output 6, NC11
Output 6, NO12
Output 6, COM13
Output 7, NC14
Digital
outputs
Output 7, NO15
Output 7, COM16
Output 8, NC17
Output 8, NO18
Output 8, COM19
PinSignal
20Input 5
21Input 6
22Input 7
23Output 9, NC
24Output 9, NO
25Output 9, COM
26Output 10, NC
27Output 10, NO
28Output 10, COM
29Output 11, NC
30Output 11, NO
31Output 11, COM
32Output 12, NC
33Output 12, NO
34Output 12, COM
35Output 13, NC
36Output 13, NO
37Output 13, COM
To aid identication, the sockets are labelled X4.1 and X4.2.
Consider the installation notes in section 2.7.
inputs
Digital
Note!
Digital
outputs
The conguration illustrated here
is that of the rst socket, labelled
X4.1.
Inputs 8-14 and outputs 14-22
are on the second socket (X4.2),
if installed.
Fig. 2-10: Sockets X4.1 and X4.2 - Pin conguration
Emerson Process Management GmbH & Co. OHG2-16
Page 49
Short Form Manual
HASXEE-SFM-HS
07/2010
2.6.1 Installation - X-STREAM XEGC, X-STREAM XEGP
Analog Inputs
X-STREAM XE
0 - 1 V, 0 - 10 V (software selectable) R
= 100 kΩ
in
optional (requires to t wire bridges, see gures):
2 analog inputs
electrical
specication
4 (0) - 20 mA ; R
= 50Ω
in
optically isolated from analyzer GND
protected against overload up to
±15 V or ±20 mA
Analog inputs are located on a 9-pole submin-D-connector (X5) or on an optional terminals adaptor XSTI ( page 2-27).
Note!
Consider the installation notes in section 2.7.
BR 2 t a wire bridge here to apply analog
{ }
BR 2 signal in current mode to input 2 **)
BR 1 t a wire bridge here to apply analog
{ }
BR 1 signal in current mode to input 1 *)
SignalPin
unused3
5
4
2
1
PinSignal
9Input 2 high (+)
8Input 2 low (-)
7Input 1 low (-)
6Input 1 high (+)
Installation
2
*) alternatively set jumper P2 on electronics board XASI
**) alternatively set jumper P1 on electronics board XASI
Fig. 2-11: Plug X5 - analog inputs
Emerson Process Management GmbH & Co. OHG 2-17
Page 50
X-STREAM XE
2.6.2 Installation - X-STREAM XE Field Housings
2.6.2 X-STREAM XEF, X-STREAM XDF
Short Form Manual
HASXEE-SFM-HS
07/2010
Fitted with four eyebolts and featuring IP66/
Type 4X protection, the X-STREAM XE eld
Gas ttings
Cable glands
Note!
Take care to reserve space at the
right side of the instrument for laying
the cables!
Connector
for potential
equalization
housing can be mounted in the open air on a
wall or frame with no extra work.
All dimensions in mm
[inches in brackets]
Fig. 2-12: X-STREAM XEF - dimensions for installation
HEAVY INSTRUMENT
X-STREAM XEF / XDF analyzers, intended to be wall mounted or outdoor
installed, weigh up to 45 kg (99 lbs), depending on the chosen options!
Use two persons and/or suitable tools for transportation and lifting these
instruments!
Take care to use anchors and bolts specied to be used for the weight of
the instruments!
Assure that the wall / device for installation is sufciently attached and
stable to carry the instrument!
Emerson Process Management GmbH & Co. OHG2-18
Page 51
Short Form Manual
HASXEE-SFM-HS
07/2010
2.6.2 Installation - X-STREAM XE Field Housings
Note!
Take care to reserve space
at the right side of the instrument for laying the cables!
X-STREAM XE
Gas ttings
Connector
for potential
equalization
Cable glands
Installation
2
All dimensions in mm
[inches in brackets]
Fig. 2-13: X-STREAM XDF - dimensions for installation
Emerson Process Management GmbH & Co. OHG 2-19
Page 52
X-STREAM XE
2.6.2 Installation - X-STREAM XE Field Housings
Short Form Manual
HASXEE-SFM-HS
07/2010
Power and signal cables are connected using
internal screw-type terminals. This requires
opening the unit, which in turn requires releasing the fasteners on the housing.
Gas connectors are accessible from the outside, on the underside of the instrument.
The number and conguration of the gas
inlets and outlets depends on the analytical
18423
application, and is noted on a sticker on the
underside of the instrument next to the connectors.
To simplify installation, we recommend labelling the gas lines in accordance with these
markings. This avoids confusion should the
analyzer need to be disconnectedfor maintenance.
Note on XDF!
In case of the dual compartment
version XDF, the electrical connections are established in the
upper compartment, and the
gas connections to ttings at the
lower compartment.
Besides this, the design and
layout of terminals and ttings
are the same as with the single
compartment version XEF.
1
4
7
Gas inlets
1 Terminals for signal cables
2 Mains lter
3 Power connections with integrated fuses
4 Glands for power cable
Fig. 2-14: X-STREAM XE Field housing - Arrangement of terminals, cable glands and gas connectors
6
Gas outlets
5
5 Glands for signal cables
6 Gas inlets and outlets
7 Plugs for openings to connect housings
8 Ethernet connectors (optional)
Emerson Process Management GmbH & Co. OHG2-20
Page 53
Short Form Manual
HASXEE-SFM-HS
07/2010
2.6.2 Installation - X-STREAM XE Field Housings
Gas connections
Gas connections
max number8
max for purging (incl. / separate)2 incl.
materialPVDF; stainless steel (opt.)
1
sizes6/4 mm;
⁄4"
Signal in- and outputs
Preparation of signal cables
X-STREAM XE
Installation
All signal cables are connected to screw-type
terminals located inside the housing. Access
to the internal components is gained by releasing the two fasteners at the top of the unit
and opening the front panel downwards.
All cables must be fed through cable glands
and secured with a gland nut.
Installing cable glands with shielded cables
1. Strip the cable
2. Expose braided
shield
3. Feed cable through
dome nut and clamping insert
Properly installed, the glands act as a strain
relief and guarantee EMC (electromagnetic
compatibility):
6. Push clamping insert
into body and tighten
dome nut
7. Assemble into housing
and you´re done!
2
4. Fold braided shield
over clamping insert
5. Make sure that braided shield overlaps
the O-ring by 2 mm
(3⁄32“)
Emerson Process Management GmbH & Co. OHG 2-21
Page 54
X-STREAM XE
2.6.2 Installation - X-STREAM XE Field Housings
Verify the power supply at installation site meets the specication given on
the analyzer´s nameplate label, before installing the instrument!
Verify power cables are disconnected and/or instrument is de-energized
prior to working at the terminals!
Verify the power cord is layed with a distance of at least 1 cm (0.4 in) to any
signal cable to ensure proper insulation from signal circuits!
Short Form Manual
HASXEE-SFM-HS
07/2010
ELECTRICAL SHOCK HAZARD
The number of actually available signal outputs, and also the number of built-in modules
with screw-type terminals, varies according to
the conguration of the unit.
A maximum of three modules with 36 terminals each can be tted.
The terminals can be accessed by opening
the front panel of the instrument.
Characteristics of terminals:
2
Accepted wire gauge:
0.14 ... 1.5 mm
end sleeves not required
(AWG 26 ... AWG 16),
Skinning length:5 mm (0.2")
Thread:M2
Min. tightening torque:0.25 Nm (2.21 in.lb)
Emerson Process Management GmbH & Co. OHG2-22
Page 55
Short Form Manual
HASXEE-SFM-HS
07/2010
X-STREAM XE
2.6.2 Installation - X-STREAM XE Field Housings
Analog signals
Relay outputs 1-4
Terminals for analog signals and relay outputs
1 - 4 are located on the outer left module
(terminal block X1).
Specication of analog signal outputs: 4 (0) - 20 mA; burden: RB ≤ 500 Ω
Specication of relay outputs:
Electrical specication:
Dry relay change-over contacts can be used as
NO or NC.
max. 30 VDC, 1 A, 30 W
Installation
Note!
Consider the installation notes in section 2.7
and the notes on installing cable glands on
page 2-21.
P3.12Output 4 (Function check), NC
P4.1Output 4 (Function check), NO
P4.2Output 4 (Function check), COM
P4.3not used
P4.4
P4.5
P4.6
P4.7
P4.8
P4.9
P4.10
P4.11
P4.12
P3.3Output 1 (Failure), NC
P3.4Output 1 (Failure), NO
P3.5Output 1 (Failure), COM
P3.6Output 2 (Maintenance Request), NC
P3.7Output 2 (Maintenance Request), NO
P3.8Output 2 (Maintenance Request), COM
P3.9Output 3 (Out of Spec), NC
P3.10Output 3 (Out of Spec), NO
P3.11Output 3 (Out of Spec), COM
P3.1not used
P3.2not used
P2.7Channel 4, (+) 4 (0) - 20 mA
P2.8Channel 4, GND
P2.9Channel 5, (+) 4 (0) - 20 mA
P2.10Channel 5, GND
P2.11not used
P2.12not used
PinSignal
P2.1Channel 1, (+) 4 (0) - 20 mA
P2.2Channel 1, GND
P2.3Channel 2, (+) 4 (0) - 20 mA
P2.4Channel 2, GND
P2.5Channel 3, (+) 4 (0) - 20 mA
P2.6Channel 3, GND
2
Serial Interface
*)
Relay Outputs
**)
Conguration of relay output termi-
**)
Analog Outputs
nals as per standard factory setting
(NAMUR status signals)
Fig. 2-15: Terminal block X1 - Analog signals and relay outputs 1-4 (XSTA)
Emerson Process Management GmbH & Co. OHG 2-23
Page 56
X-STREAM XE
2.6.2 Installation - X-STREAM XE Field Housings
Serial interface
Specication and interface control:
X-STREAM XE instruction manual,
chapter 9
P4.3not used
P4.4
P4.5
P4.6
P4.7
P4.8
P4.9
P4.10
P4.11
P4.12
P3.5Output 1 (Failure), COM
P3.6Output 2 (Maintenance Request), NC
P3.7Output 2 (Maintenance Request), NO
P3.8Output 2 (Maintenance Request), COM
P3.9Output 3 (Out of Spec), NC
P3.10Output 3 (Out of Spec), NO
P3.11Output 3 (Out of Spec), COM
P3.12Output 4 (Function check), NC
P4.1Output 4 (Function check), NO
P4.2Output 4 (Function check), COM
Short Form Manual
HASXEE-SFM-HS
07/2010
The 9 terminals on the left (28 - 36) of the right
most strip carry the serial interface signals.
X
PinSignal
P2.1Channel 1, (+) 4 (0) - 20 mA
P2.2Channel 1, GND
P2.3Channel 2, (+) 4 (0) - 20 mA
P2.4Channel 2, GND
P2.5Channel 3, (+) 4 (0) - 20 mA
P2.6Channel 3, GND
P2.7Channel 4, (+) 4 (0) - 20 mA
P2.8Channel 4, GND
P2.9Channel 5, (+) 4 (0) - 20 mA
P2.10Channel 5, GND
P2.11not used
P2.12not used
P3.1not used
P3.2not used
P3.3Output 1 (Failure), NC
P3.4Output 1 (Failure), NO
Serial Interface
*)
See table
*)
Relay Outputs
Notes!
Consider the installation notes in section 2.7
and the notes on installing cable glands on
page 2-21.
X-STREAM analyzers are classified DTE
(Data Terminal Equipment).
Fig. 2-16: Terminal block X1 - Serial interface
Terminal
P4.4SER1Common Common Common
**)
MOD 485/
2 wire
MOD 485/
4 wire
Analog Outputs
RS 232
P4.5SER2not usednot usedRXD
P4.6SER3not usednot usedTXD
P4.7SER4not usedRXD1(+)not used
P4.8SER5D1(+)TXD1(+) Common
P4.9SER6not usednot usednot used
P4.107not usednot usednot used
P4.118not usedRXD0(-)not used
P4.129D0(-)TXD0(-)not used
Your analyzer´s type of serial interface is
marked on a label nearby the terminals (see
sample above)
Emerson Process Management GmbH & Co. OHG2-24
Page 57
Short Form Manual
HASXEE-SFM-HS
07/2010
2.6.2 Installation - X-STREAM XE Field Housings
RJ45 connection
The RJ45 connectors are located on an
electronics board in the card cage section
of the unit.
To install this connection, a cable must be
fed through the cable entry without a connector.
The connector can be wired on when the free
end has been fed into the instrument:
We recommend the VARIOSUB RJ45 QUICKON connector (PHOENIX CONTACT), which
is supplied with the unit and requires no special tools. Wiring instructions can be found
in the separate manual supplied with the
connector.
X-STREAM XE
Note!
Please note that, although the Modbus terminals
( previous page) are still installed,
they are not connected!
Installation
2
Pin 1Pin 8
Pin no.
othernot used
Fig. 2-17: X-STREAM XEF - Ethernet connector
Emerson Process Management GmbH & Co. OHG 2-25
Signal
1TX+
2TX-
3RX+
6RX-
Page 58
Short Form Manual
X-STREAM XE
HASXEE-SFM-HS
2.6.2 Installation - X-STREAM XE Field Housings
Digital inputs and relay outputs (option)
Terminals for these signals are located on the terminals board XSTD (terminal block X4).
max. 30 V, internally limited to 2.3 mA
7 or 14 digital inputs
HIGH: min. 4 V; LOW: max. 3 V
common GND
Dry relay change-over contacts can be used as NO or
9 or 18 additional relay outputs
NC
max. load. 30 V; 1 A; 30 W resistive
Notes!
Depending on configuration, an analyzer
can be tted with up to two of these terminal blocks (the unit will then feature 14
Consider the installation notes in section 2.7
and the notes on installing cable glands on
page 2-21.
digital inputs and 18 digital outputs). To aid
identication, the sockets are labelled X4.1
and X4.2 (see sample of label to the right).
07/2010
P3.11Output 5, NO
P3.12Output 5, COM
P4.1Output 6, NC
P4.2Output 6, NO
P4.3Output 6, COM
P4.4Output 7, NC
P4.5Output 7, NO
P4.6Output 7, COM
P4.7Output 8, NC
P4.8Output 8, NO
P4.9Output 8, COM
P4.10Output 9, NC
P4.11Output 9, NO
P4.12Output 9, COM
P2.1Output 10, NC
P2.2Output 10, NO
P2.3Output 10, COM
P2.4Output 11, NC
P2.5Output 11, NO
P2.6Output 11, COM
P2.7Output 12, NC
P2.8Output 12, NO
P2.9Output 12, COM
P2.10Output 13, NC
P2.11Output 13, NO
P2.12Output 13, COM
Digital outputs
Fig. 2-18: Terminals blocks X4 - digital inputs and outputs
P3.7Input 7
P3.8GND for inputs 1-7
P3.9unused
P3.10Output 5, NC
P3.4Input 4
P3.5Input 5
P3.6Input 6
Digital inputs
P4
PinSignal
P3.1Input 1
P3.2Input 2
P3.3Input 3
Note!
The conguration illustrated here is that of the rst
adapter (X4.1).
Inputs 8-14 and outputs
14-22, if available, are on
the second adapter (X4.2).
Emerson Process Management GmbH & Co. OHG2-26
Page 59
Short Form Manual
HASXEE-SFM-HS
07/2010
X-STREAM XE
2.6.2 Installation - X-STREAM XE Field Housings
Analog inputs (option)
Terminals for analog input signals are located on the terminals board XSTI (terminal block X5).
0 - 1 V, 0 - 10 V (software selectable) R
optional (requires to t wire bridges, see gure):
2 analog inputs
4 (0) - 20 mA ; R
= 50Ω
in
optically isolated from analyzer GND
protected against overload up to
Note!
Consider the installation notes in section 2.7
and the notes on installing cable glands on
page 2-21.
P2.3 Input 1 high (+)
P2.4 Input 1 high (+)
P2.5 Input 1 low (-)
P2.6 Input 1 low (-)
P2.7 BR 2 t a wire bridge here to apply analog
P2.8 BR 2 signal in current mode to input 2
P2.9Input 2 high (+)
P2.10Input 2 high (+)
P2.11Input 2 low (-)
P2.12Input 2 low (-)
1
= 100 kΩ
in
±15 V or ±20 mA
P2.1 BR 1 t a wire bridge here to apply analog
P2.2 BR 1 signal in current mode to input 1
P3.11not used
P3.12not used
P3.9not used
P3.10not used
P3.7not used
P3.8not used
P3.5not used
P3.6not used
P3.3not used
P3.4not used
P3.1 not used
P3.2not used
Installation
2
2
PinSignal
{
{
1 Screw-type terminals
2 Reserved for future use
}
IN2
Mode
Analog Inputs
}
IN1
Mode
Fig. 2-19: Terminal block X5 - Analog input signals
Emerson Process Management GmbH & Co. OHG 2-27
Page 60
X-STREAM XE
2.6.2 Installation - X-STREAM XE Field Housings
Connecting the power cord
The power cord is connected to screw-type
terminals located inside the housing.
Skinning length:8 mm (0.315")
Thread:M3
Min. tightening torque:0.5 Nm (4.4 in.lb)
min. 1.5 mm² (AWG 15),
end sleeves not required
Short Form Manual
HASXEE-SFM-HS
07/2010
ELECTRICAL SHOCK HAZARD
Verify the power supply at installation site meets the specication given on
the analyzer´s nameplate label, before installing the instrument!
Verify power cables are disconnected and/or instrument is de-energized
prior to working at the terminals!
Verify the power cord is layed with a distance of at least 1 cm (0.5“) to any
signal cable to ensure proper insulation from signal circuits!
Feed the power cable through the cable
gland at the instrument´s right side and strip
the outer insulation. Strip the individual wires
and connect to the terminals (a label is located next to the terminals on the mains lter
housing).
Finally, tighten the outer dome nut to secure
the power cable.
Live L
Neutral N
Earth PE
Fig. 2-20: Power supply connections
Emerson Process Management GmbH & Co. OHG2-28
Power supply
cable gland
Page 61
Short Form Manual
HASXEE-SFM-HS
07/2010
2.6.2 Installation - X-STREAM XE Field Housings
ELECTRICAL SHOCK HAZARD
Before completing the electrical connection of the instrument, verify cables
are inserted and connected in correct manner!
Ensure the earthing conductor (protective earth; PE) is connected!
When all connections are correctly made and
checked,
•close the front panel and secure with the
two fasteners.
X-STREAM XE
Installation
2
Emerson Process Management GmbH & Co. OHG 2-29
Page 62
X-STREAM XE
2.7 Installation - Notes on Wiring
2.7 Notes on Wiring Signal Inputs and Outputs
Short Form Manual
HASXEE-SFM-HS
07/2010
Emerson Process Managament has made
every effort during the development process
to ensure that the X-STREAM analyzer series
ensures electromagnetic compatibility (EMC)
with respect to emission and interference
resistance, as conrmed by EMC measurements.
2.7.1 Electrical Shielding of Cables
In order to minimise ambient electromagnetic interference, it is necessary to take care
making all electrical connections between the
analyzer and any other devices:
•We recommend using only shielded
signal cables. The shielding must be
connected at both ends to the housing
(g. 1-24).
However, EMC is not only inuenced by the
design of the instrument, but to a large degree
by the on-site installation process. Please
observe the following sections and precautions to guarantee the safe and problem-free
operation of this analyzer.
Fig. 2-21: Shielded signal cable, shielding connected at both ends.
Emerson Process Management GmbH & Co. OHG2-30
Page 63
Short Form Manual
HASXEE-SFM-HS
07/2010
X-STREAM XE
2.7 Installation - Notes on wiring
On-site conditions often differ from test
•The shielding is connected only at one
environments and may require special precautions. Such a case arises when strong
electromagnetic elds which could induce
an interference current in the shielding. This
type of current creates a potential difference
between the connected housings.
Two possible methods of eliminating this are
described here. Fitters familiar with EMC
problems must decide which method should
be emplyed.
Fig. 2-22: Shielded signal cable, shielding connected at one end.
end (connecting to the analyzer is recommended): this gives better protection
against external interference, and interference currents are prevented because
the ground loop is interrupted.
Installation
2
•Cables with double shielding: in this
case, one shielding is connected to the
analyzer housing, the other shielding
to the external device. This is advan-
This method is more expensive, but gi-
ves the best protection against external
interference and against interference
currents
.
tageous when both units are supplied
from different grids (e.g. when installed
in different buildings).
Fig. 2-23: Signal cable with double shielding, shieldings connected at alternate ends.
Emerson Process Management GmbH & Co. OHG 2-31
Page 64
X-STREAM XE
2.7 Installation - Notes on wiring
X-STREAM XEGP/XEGC with terminal
adapters
In order to avoid measured values being inuenced by external interference signals when
terminal adapters are in use, the signal cable
shieldings must be connected to the analyzer housing by means of shield connector
terminals:
• Strip the signal cable to a length of 20 cm
(8"). Take care to not damage the braided
shield!
• Pull up the contact part of the shield connector terminal,
• feed the cable through as illustrated in
g. 2-24, and
• release the contact part down onto the
braided shield.
The result is a secure contact with the cable
shield, improving the unit’s immunity against
interference from other electronic devices.
Finally connect the individual wires as described before.
Short Form Manual
HASXEE-SFM-HS
07/2010
Fig. 2-24: Shield connector terminal with cable
The shield connector must be ordered to t the cable diameter, and can be retrotted:
Ø 1.5 ... 6.5 mm (0.06" ... 0.25") part # ETC02019
Ø 5 ... 11 mm (0.2" ... 0.43")part # ETC02020
Ø 10 ... 17 mm (0.4" ... 0.66")part # ETC02021
Ø 16 ... 24 mm (0.63" ... 0.94")part # ETC02022
When an inductive load (e.g. relay, valve) is
switched off, the magnetic eld resists the
change in current; this induces a high voltage across the coil contacts (several hundred
volts). This impulse propogates through the
connected cables and can influence any
electrical devices nearby or destroy signal
inputs and outputs. This can be avoided with
a simple precaution:
•A silicon diode is connected in parallel to
the load’s contacts. The induced impulse is thus short-circuited at its source.
The cathode must be connected to the
positive end of the coil, the anode to the
negative end (g. 1-28).
Compatible lter components for standard
valves are available on request.
2.7.3 Driving High-Current Loads
Loads which draw a current in excess of the
specications for X-STREAM series analyzer
outputs (>30 mA / >1 A) may not be directly
driven from digital or relay outputs.
Such loads require external relays serving as
de-coupling modules: the X-STREAM output
drives the external relay, which in turn drives
the load.
In order to avoid interference, we recommend
supplying the analyzer and the high-current
loads from different sources (g. 1-29).
As previously described, the use of suppressor diodes for inductive loads is highly
recommended.
Last
External relay
Installation
2
Analyzer output
Fig. 2-26: Driving high-current loads
Fig. 2-25: Suppressor diode for inductive loads.
Emerson Process Management GmbH & Co. OHG 2-33
Page 66
X-STREAM XE
2.7 Installation - Notes on wiring
2.7.4 Driving Multiple Loads
Frequently, several loads in one system are
controlled by several analyzer outputs, whereby the power for the loads derives from a
common source.
Special care is needed when wiring the loads
to minimize interference from switching these
loads:
•avoid connecting the loads in series:
Short Form Manual
HASXEE-SFM-HS
07/2010
•It is recommended the loads to be
wired in parallel, and each load connected separately to the power supply. Beginning at the distribution point,
both the + and the - wires for each
load are laid together to the load (g.
2-28). Interference is further reduced if a
twisted multi-core cable is used.
L
Fig. 2-27: Loads in series
Distribution point
Lay cables to the loads
together as far as possible
(using twisted-pair cables
if possible)
Fig. 2-28: Loads in parallel
J
Emerson Process Management GmbH & Co. OHG2-34
Page 67
Short Form Manual
HASXEE-SFM-HS
04/2010
3.1 Introduction
X-STREAM XE
Chapter 3
Startup
Once the unit has been unpacked and installed, we recommend to rst check the
settings, and if necessary adjust them to the
user’s needs. e.g:
•What hardware is installed?
•Is the unit configured to your needs
(alarms, inputs, outputs, etc.)
In order for the information in this chapter
to be of any relevance, the unit must have
been installed according to the instructions
in chapter 4.
The following pages describe how to perform
a leak test, navigate through the menus and
what is to be observed when conguring the
unit. For the rst startup after installation,
follow the step-by-step instructions for navigating the menus, allowing you to familiarise
yourself with the unit and its software, and if
necessary adjust the settings to your needs.
Startup
3
Emerson Process Management GmbH & Co. OHG 3-1
Page 68
X-STREAM XE
3.2 Leak Test
3.2 Performing a Leak Test
To achieve best and proper measuring results
you must ensure the gas path system does
not have leaks.
The following procedure describes how to
perform a leak test with focus on the instrument.
The gas path system should be leak tested
at least on a bimonthly basis and after maintenance, replacement or repair of gas path
parts.
Note!
We recommend to include external equipment (e.g. cooler, dust lters, etc.) into a leak
test!
Short Form Manual
HASXEE-SFM-HS
04/2010
HAZARD FROM GASES
Before opening gas paths
they must be purged with
ambient air or neutral gas (N2)
to avoid hazards caused by
toxic, ammable, explosive
or harmful to health sample
gas components!
Required tools
•U-turn manometer for max. 1.45 psi
(100 mbar)
•Stop valve
Procedure
•Connect the water lled u-turn manome-
ter to the analyzer‘s sample gas output
(disconnect external gas lines).
•Install the stop valve between gas input
tting and a nitrogen (N
) supply.
2
•Open the stop valve until the internal
gas path is under pressure of approx.
0.725 psi/50 mbar (corresponding to
19.7 inch/500 mm water column)
•Close the stop valve. After a short time
for the water to balance, the water level
must not change over a time period of
approx. 5 minutes!
Analyzer
overpressure
stop
valve
Water
Fig. 3-1: Leak Testing with U-turn Manometer
approx.
0.725 psi /
50 mbar
Max. pressure 7.25 psig
(500 mbar)!
Multi channel instruments:
Analyzers with parallel tubing
require separate leak tests for
each gas path !
Emerson Process Management GmbH & Co. OHG3-2
Page 69
Short Form Manual
HASXEE-SFM-HS
04/2010
3.3 Symbols Used
3.3 Symbols and Typographical Conventions
In the following sections, the symbols and
typographical conventions explained below
are used to describe the software menus and
navigation.
X-STREAM XE
SymbolDescription
Within Process Descriptions
Setup..
Setup..
Analog outputs..
Analog outputs..
Output1...5
Setup..
In-/Outputs..
Analog outputs..
Output1...5
Menu title
Parent (SETUP) and current menu (ANALOG OUTPUTS)
As an example, the menu
for Output1 is shown; the
menus for outputs 2 to 5
look identical
To access the current
menu, access level code
3 has to be entered somewhere in the menu history
Access levels:
Access level 1
(user)
Access level 2
(expert)
Convention Description
Within Text
(MENU TITLE)
6-12
CONTROL
CONTROL - RANGES
"Valves"
"Control.."
Never, 1 min
0 ... 2000
enter
For a detailed description
of MENU, see page 6-12.
Identies the CONTROL
menu, e.g. "press
open CONTROL"
From within the CONTROL
menu, select the RANGES
menu.
Parameter or menu line
name
Values to be selected
Value to be entered
press key (here: ent er
key)
enTer to
Startup
3
Access level 3
(administrator)
Access level 4
(service level)
Screen shot
Control..
Setup..
Status..
Info..
Service..
Emerson Process Management GmbH & Co. OHG 3-3
(here: MAIN MENU)
Page 70
X-STREAM XE
3.4 Front Panel Elements
Short Form Manual
HASXEE-SFM-HS
04/2010
3.4 Front Panel Elements
All X-STREAM XE gas analyzers have an
graphic display to show measuring and status
information, and the easy-to-use menu-based
user interface for entering parameters. For
ease of understanding, the user can at any
time select one of three languages stored in
Ch1CO
Ch2CO
Ch3CO
Ch4CO
0.000
0.000
0.000
ppm
ppm
ppm
ppm
0.000
Fig. 3-2: X-STREAM XE front panel
3.4.1 Display
Ch1CO
Ch2CO
Ch3CO
Ch4CO
0.000
0.000
0.000
0.000
ppm
ppm
ppm
ppm
Input
Ch1
10.0000
Output
Ch2
10.0000
Flow 10.00 l/min
ppm
CO
ppm
CO
the unit (currently available: English, French,
German, Italian, Polish, Portuguese and Spanish in various combinations).
Units are operated by six keys on the front
panel.
The graphic display has various layouts to
support diplaying the measuring results as
well as status information and software menus. The gure to the left shows the basic
measurement displays with 4 channel layout
as well as a 2 channel layout.
3.4.2 Status Line
Ch1CO
Ch2CO
Ch3CO
Ch4CO
10.00
5.000
7.000
9.000
ppm
ppm
ppm
ppm
Note!
During an active webserver session,
the heart symbol is replaced by a network symbol:
Status information is provided by different
symbols in the display´s rst line:
Bell = 'Alarm'
Cross = 'Failure'
Question mark = 'Off spec'
Oil can = 'Maintenance request'
Tool = 'Function check'
Heart = the analyzer´s 'heart beat',
indicating the instrument is operating.
Clear text messages are provided in the
display´s bottom line
Emerson Process Management GmbH & Co. OHG3-4
Page 71
Short Form Manual
HASXEE-SFM-HS
04/2010
3.4.3 Keys
X-STREAM XE
3.4 Front Panel Elements
Six keys enable the use of the menu system.
Depending on the operational mode (measuring, browsing menus, editing) they have the
following functions:
enter key:
ModeFunction
MeasuringEnter main menu
Browsing
Open submenu (..) or
execute command (!)
EditingConrm new entry
home key:
ModeFunction
Measuring(no function)
Browsing
Return to measurement display
EditingAbort entry
up / down keys:
ModeFunction
MeasuringEnter main menu
Highlight next menu line
Browsing
Open the previous/next page,
when currently a line beginning
with / is highlighted
EditingChange current parameter
Startup
3
Emerson Process Management GmbH & Co. OHG 3-5
Page 72
X-STREAM XE
3.4 Front Panel Elements
left key:
ModeFunction
Measuring
Browsing
Editing
Short Form Manual
HASXEE-SFM-HS
04/2010
Enter main menu or
nd
open 2
measurement display
page (if congured)
Go up 1 level or page in menu
system
Move cursor 1 char to the left
Leave channel selection
Cancel editing of given para-
meter
Go to previous menu page,
shows in rst menu line
right key:
ModeFunction
MeasuringEnter main menu or
nd
open 2
measurement display
page (if congured)
BrowsingOpen submenu (..)
Go to next menu page, when
Editing
shows in last menu line
Move cursor 1 char to the right
if
Emerson Process Management GmbH & Co. OHG3-6
Page 73
Short Form Manual
HASXEE-SFM-HS
04/2010
3.5 Software
3.5 Software
The analyzer software displays measurement
results and status messages, allows parameters to be set and edited, and maintenance
functions (e.g. calibration) to be carried out.
The software is organised hierarchically: The
topmost level is called MEASUREMENT DISPLAY, followed by a MAIN MENU; all other
menus and submenus are arranged below.
Menu lines can perform different functions,
to be distinguished by the following characteristics:
FunctionDescription
Text
Simple text (not selectable with
cursor)
A variable description shows a
colon; the line can be made up
of up to 3 elements:
1. description
2. value (number or text)
3. unit (optional)
Examples:
Span gas: 2000 ppm
Editable
Tol.Check: Off
variables /
parameters
Pressing
enter in an editable
variable line highlights the value to be changed.
X-STREAM XE
FunctionDescription
A command line text ends in
an exclamation mark; pressing
enter with such a line highligh-
Executable
command
Selectable
submenu
ted, the command is executed,
e.g. a calibration procedure.
Example:
Start calibration !
A menu line text ends in two
dots. Press
enter with a menu
line highlighted to open the
submenu.
Example:
Setup..
Startup
3
The optional unit can only
be changed utilizing a setup
menu.
Variables shown without a colon cannot be edited, they are
for information only.
Emerson Process Management GmbH & Co. OHG 3-7
Page 74
X-STREAM XE
Browsing
Program Sequence 1of8
Action1:END-OF-PGRM
Node1:All
Action2:END-OF-PGRM
Node2:All
Program Sequence 4of8
Action3:END-OF-PGRM
Action13:END-OF-PGRM
Node3:All
Node13:All
Action4:END-OF-PGRM
Action14:END-OF-PGRM
Node4:
Node14:All
Program Sequence 8of8
Action15:END-OF-PGRM
Action29:END-OF-PGRM
Node15:All
Node29:All
Action16:END-OF-PGRM
Action30:END-OF-PGRM
Node16:
Node30:All
All
Page 1
All
3.5 Software
Page 4
Page 8
Short Form Manual
HASXEE-SFM-HS
04/2010
Some menus have more entries than can displayed at once. In these menus, an indicator
in the last (
) and/or rst () line indicates the
direction the menu continues in.
In the example to the left
• page 1
• page 4
continues downwards
continues upwards and down-
wards
• page 8 continues upwards.
To show the next page (indicator
)
• place the cursor in the last accessible line
and press down or
• press right, irrespective of where the cursor
is located.
To show the previous page (indicator
)
• place the cursor in the rst accessible line
and press up or
• press left, irrespective of where the cursor
is located.
Emerson Process Management GmbH & Co. OHG3-8
Page 75
Short Form Manual
HASXEE-SFM-HS
04/2010
X-STREAM XE
3.5 Software
Editing
The editing mode enables changing parameters. It is initiated by pressing
enter.
If the selection is a parameter list, the current
entry is highlighted and may be changed by
up and down.
If the selection is a value, the cursor is placed
over the last character. Use up and down to
change it.
Use
left and right to select another charac-
ter.
The type of available characters depends on
the position of the cursor:
• It is not possible to select the minus sign or
decimal point as the last character
.
• It is not possible to select the decimal point
in integer values
.
• For decimal numbers, the decimal point can
be placed anywhere within certain limits.
There are two ways to exit the data entry
mode:
enter: the entry is veried.If it is accepted,
it is saved and the new value dis played; if not, an error message is
displayed
.
Component selection menu
Within the analyzer software, one can distinguish between analyzer related and component related menus: While the rst contain
entries, relevant for the analyzer (e.g. time
setting), the second contain entries relevant
for a specic component (channel) only (e.g.
calibrating a channel).
For single channel analyzers, editing any
channel specic parameter will only effect
this one channel.
Different for multi-channel analyzers: Such
instruments require selecting a channel prior
to changing channel related parameters.
When a channel related menu entry is selected, automatically a SELECT COMPONENT
menu shows up, to select the component of
interest, or to cancel the current action.
Select Component
Component:Ch1
Press
to return
Startup
3
home: Cancel: all settings and changes are
reset to their former values.
Select the component / channel you want to
work with, and press
enter.
This menu does not show on single-channel
units.
Within menu descriptions, the following points
out, that for multi-channel instruments a selection is required:
Multi-channel unit:
In SELECT COMPONENT select the channel to be ...
Emerson Process Management GmbH & Co. OHG 3-9
Page 76
X-STREAM XE
3.5.1 Access Levels & Codes
Short Form Manual
HASXEE-SFM-HS
04/2010
3.5 Software
Access levels can be used to prevent changes
to parameters by unauthorised personnel.
The X-STREAM menu system supports four prioritized access levels, which can be activated and deactivated separately, and should
be supplied with their own access codes.
Level four has the highest priority and is
used for factory settings — only qualied
EMERSON service personnel have access
to this level.
Level three gives access to system admin
parameters, e.g. for conguring data acquisition systems communication.
Level two covers the expert settings, e.g.
basic settings for calibrations and measurements.
Level one is the user level and includes
• parameters which should be set by trained
personnel only.
• functions, not to be started by any person
(e.g. start calibrations).
We recommend to set new access codes, if you want to use
this option!
Notes!
If a low level is locked (status On), all higher
levels will also be locked.
If a high level is accessible (code entered
when requested by a menu page), automatically all lower levels are also accessible.
For above reasons, it is always possible to
enter a higher code than requested, to gain
access to a menu (e.g. if access code 1 is requested, you may also enter access code 2).
Entering access codes
If an access code is required for a menu, a
message like the following appears:
All menus not assigned to one of these levels
are not editable or of minor relevance.
Within this manual, the descriptions of the
menus and procedures also indicate, which
level the menus are assigned. These assignments cannot be changed.
Access codes for levels 1 to 3 can be dened, activated and deactivated by the client.
The analyzer is delivered with the following
settings:
Level Access code Status
1 00000001 Off
2 00000002 Off
3 00000003 Off
Access Code 1
Code00000000
Press
to return
To enter the code, press
• up/down to change the currently selected
digit,
• left/rightto select a different digit,
• enter to submit the code
or
• home to exit the edit mode and return to
the previous display.
Emerson Process Management GmbH & Co. OHG3-10
Page 77
Short Form Manual
HASXEE-SFM-HS
04/2010
X-STREAM XE
3.6 Powering Up
3.5.2 Special Messages
Depending on the last action performed by the
user, one of the following messages may be
displayed to assist or inform the user.
Information on incorrect entry:
The value entered by the user is outside
valid limits. The display indicates what limits
apply.
Press to return to the previous screen to
enter an acceptable value.
Wrong input
Range0.1 ... 50
Press
to return
Conrmation of execution of function:
Conrms that a function or procedure (e.g.
calibration) has been started, or cancelled.
The message automatically dissapears after
a few seconds.
Function Executing
3.6 Powering Up
3.6.1 Boot Sequence
When the unit is powered up, a series of internal tests is automatically performed. During
this time the front panel keys are disabled,
while the remaining time counts down in the
display.
3.6.2 Measurement Display
The measurement display is shown
•automaticaly on completion of the boot
sequence
•when
home is pressed
•automatically after a set period of time
of inactivity (i.e. with no keys being
pressed).
The information displayed in the four lines of
the measurement display can be determined
by the operator:
•Sample gas components, measuring
results and measuring units for each
channel
•secondary measurements, e.g. pressure, gas ow, temperature
•nothing (empty line)
The factory settings are as follows:
Line 1: measured value of channel 1
Line 2: measured value of channel 2
Line 3: measured value of channel 3
Line 4: measured value of channel 4
Startup
3
Note!
Press
to return
If less than four channels are installed in the
unit, only the measureands for these channels
are available for selection.
Emerson Process Management GmbH & Co. OHG 3-11
Page 78
X-STREAM XE
3.6 Powering Up
SETUP enables several additional congurations, e.g.
• 2 pages measurement display
• different font sizes
Short Form Manual
HASXEE-SFM-HS
04/2010
The very rst display line shows
• a ashing heart, showing the instrument is
operating
• one or more status pictograms, if (NAMUR)
status are active. Some of these are explained by a text message in the last line
• a channel indicator, if the current menu
page is related to a specic channel only.
The display´s bottom line shows plain text
status information (errors, maintenance
requests, function checks or off-spec performance).
Active messages are stored in an internal
buffer. If there is more than one message in
the buffer, the display will cycle through.
Most messages also activate a NAMUR relay
(if a relay has been assigned to that NAMUR
function;
nual
).
X-STREAM XE instruction ma-
Ch1CO
Ch2CO
Ch3CO
Ch4CO
4 lines display
Input
Ch1
Output
Ch2
Flow 10.00 l/min
10.0000
10.0000
1.000
500
2.400
1.900
ppm
ppm
ppm
ppm
ppm
CO
ppm
CO
Note!
There are also functions, that do activate a
relay, but are not shown on the display (e.g.
concentration alarms). In such cases, check
the status menu for more information.
2 lines display with additional secondary parameter line
MEASUREMENT DISPLAY
Emerson Process Management GmbH & Co. OHG3-12
Page 79
Short Form Manual
HASXEE-SFM-HS
04/2010
3.7 Language Settings
X-STREAM XE
3.7 Language Settings
If the analyzer is operational and it becomes
clear that the incorrect language has been
set, which is unintelligible to the operator, the
following sequence of keypresses (starting at
the measurement display) can be used to set
the language.
If the system has been set up accordingly, the
code for access level 1 must be entered at this
point to enable access to the following menu.
Note!
The factory setting for this unit is "no code
required”. For ease of operation, it is recommended to use the factory settings for access
codes while setting up the unit for the rst
time. In the following sections, therefore, no
more reference will be made to any need for
entering a code.
Startup
3
Note!
Pressing
enter the 3
rd
time in this sequence
highlights the "Language” line.
• down changes the language.
• enter sets this language and the display is
updated accordingly.
• If the selected language is not the intended,
repeate the last three steps until the intended language is set.
Emerson Process Management GmbH & Co. OHG 3-13
Page 80
X-STREAM XE
3.8 Checking the Settings
Short Form Manual
HASXEE-SFM-HS
04/2010
3.8 Checking the Settings
The following sections are structured so that
the user can work through them one by one
after powering up the unit. After completing
Contrast..
Language:English
Phrase VersionEN1.04B
Measurements..
Measurement Display..
Menu Access..10 Min
Auto Home:10 Min
these steps, the unit will be congured to the
user’s needs and properly functioning.
Starting with the measurement display (
3-12), pressing any key except home will access the MAIN MENU; from here, follow these
steps
:
(If the display is showing anything other than
the measurement display, press
home to return
to the measurement display rst).
Note!
If you are unfamiliar with the language set:
3-13 shows the sequence of keys to set
a different language.
If the system has been set up accordingly,
the code for access level 1 must be entered
at this point to enable access to the following
menu.
Note!
The factory setting for this unit is "no code
required”. For ease of operation, it is recommended to use the factory settings for access
codes while setting up the unit for the rst
time. In the following sections, therefore, no
more reference will be made to any need for
entering a code.
If need be, set the preferred language for the
software.
Emerson Process Management GmbH & Co. OHG3-14
Page 81
Short Form Manual
HASXEE-SFM-HS
04/2010
3.8.1 Installed Options
X-STREAM XE
3.8 Checking the Settings
All X-STREAM gas analyzers can be tted
with a variety of optional components: follow
these steps to see which options are installed
on your analyzer.
Do not edit any entries in these
menus without special knowledge.
Incorrect entries may result in
incorrect results or impair the
performance of the unit.
Initial acces s to this menu
should be to gain information
on the configuration of the
unit.
This 2 pages menu indicates, which of the
possible optional components are installed
in the unit. The values displayed on your unit
may differ from those illustrated here.
Note!
Multichannel instruments require to select
a component (channel) to enter the second
menu page.
Startup
3
Licenses
Key 1:0
Key 2:0
Key 3:0
PackageNone
Trial Days
Emerson Process Management GmbH & Co. OHG 3-15
„Licenses..“ opens another menu where you
can check or enter license codes to unlock
optional software features.
Page 82
X-STREAM XE
3.8.2 Conguring the Display
Short Form Manual
HASXEE-SFM-HS
04/2010
3.8 Checking the Settings
Display 1 Assignments
Line 1:Comp1
Line 2:Comp2
Line 3:Comp3
Line 4:Comp4
Line 5:Blank
Labels..
Press left to return to SETUP
.
Check the settings for the measurement
display, temperature and pressure units, and
for menu access: press
enter to open DIS-
PLAY.., select "Measurement display.." and
press
enter.
If a setting does not meet your requirements,
access that menu and adjust the parameter.
Select the value to be displayed in each line
of the measurement display. The following
options are available:
Note!
X-STREAM currently supports one pressure
sensor only. Values Press1 to Press5 thus
refer to the same sensor.
Line 1:
Line 2:
Line 3:
Line 4:
Line 5:
Display 1 Labels
When entering LABELS.., you may change
the channel´s label, that is the rst text phrase in a line showing a measurement value: If
here nothing is entered, the default phrases
(Ch1 ... Ch4) are used.
Note!
Notice the headlines of the menus showing
a "1": This indicates that you can setup more
than 1 measurement display page.
Emerson Process Management GmbH & Co. OHG3-16
Page 83
Short Form Manual
HASXEE-SFM-HS
04/2010
3.8.3 Calibration Setup
Setup..
Calibration..
Ch1
Calibration
Gases..
Tolerances..
Procedure..
Valve Assignment..
Program Sequence..
Interval Time..
Deviations..
3.8 Checking the Settings
Once the display settings have been chekked, press left to return to SETUP and
open CALIBRATION to check the calibration
settings.
Multi-channel unit:
Select the component to be set in the gas
component selection menu.
Note!
For more detailed information about calibration procedures,
tion manual.
In CALIBRATION - GASES, enter the values
for zero and span gas:
• See gas supplier´s certicate for correct
values.
• Values must be correctly set for results to
be accurate.
• Multi-channel units: the values for each
channel must be entered separately.
X-STREAM XE
X-STREAM XE instruc-
Startup
3
Gases
Zero Gas:0.000 ppm
Span Gas:50000.000 ppm
Range Gases..
Current Range:Range 1
Ch2
Tolerances
DeviatToler.:Off
Zero Limit:20.0 %
Span Limit:20.0 %
Press
left to return to CALIBRATION, and
enter TOLERANCES .
By default the 'Deviation Tolerance check"'
option is set inactive (Off).
With "DeviatToler" set to On,
• during calibration the analyzer checks
whether the values set for zero and span
gas conform to the concentration of the gas
currently being supplied.
• If the concentrations differ more than the
percentage of range entered in the following lines, the calibration is aborted.This
prevents calibration from being performed
when the incorrect gas is supplied (e.g.
span gas calibration using zero gas), which
would result in an incorrectly congured
unit.
Signal damping (set in DAMPING) allows
smoothing the output signal, but also affects
the response time of outputs and display:
• The factory setting is 0 seconds.
• The maximum possible t90 time is limited by
the size of the internal sampling buffer and
the sampling rates of the installed measuring principles/sensors.
• Multi-channel units: the value for each
channel must be entered separately.
The last line enables the user
• to enter the current ambient pressure ma-
nually, if no pressure sensor is installed,
or
• to view the current pressure, if a sensor is
installed ( INSTALLED OPTIONS).
If no pressure sensor is installed, enter the
current ambient pressure here and adjust
it, when signicant changes take place: this
improves the accuracy of the instrument.
Emerson Process Management GmbH & Co. OHG3-18
Page 85
Short Form Manual
HASXEE-SFM-HS
04/2010
3.8 Checking the Settings
3.8.4 Setting the Analog Outputs
X-STREAM XE
Analog outputs..
Output1..
Output2..
Output3..
Output4..
Output5..
Setup..
In/Outputs..
Analog Outputs
Press
left to return to SETUP, and then open
IN/OUTPUTS. and from there enter ANALOG
OUTPUTS.
Select the analog output you like to check.
Note!
The following section only in brief describes
the entries currently of interest!
X-STREAM XE instruction manual,
Chapter 6 for a more detailed description.
Startup
3
Signal:Comp1
OutRange:0-20 mA
Low Scale:0.00
Max Scale:100.00
AutoScale:Yes
FailMode:Live
0/4 mA:0.00
20 mA:100.00
Hold:No
Emerson Process Management GmbH & Co. OHG 3-19
"Signal” species the value associated with
the selected output. The following options
(partly dependent on the number of measuring channels and sensors installed) are
available:
Page 86
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
04/2010
3.8 Checking the Settings
None
*)
The analog output is deactivated
Signal
A 0 or 4 mA signal is generated, e.g. to check the signal processing in an
0 mA
external system. Whether a 0 or 4 mA signal is generated, is set by the "Out range” line ( next page).
20 mA
A 20 mA signal is generated, e.g. to check the signal processing in an external system.
Comp1 ... 5Gas concentration
Temp1 ... 5Temperature
Press1 ... 5Pressure
Flow1 ... 5Flow
CalcA ... DResult of calculator
RawVal1 ... 5Raw value
RangeID1 ... 5 ID of selected range
*) Numbers 1 to 5 refer to components [channels] 1 to 5: In case of secondary parameters,
this means, the selected value is that of the
sensor assigned to the given component
(Press2 is the pressure value of the sensor
assigned to component 2).
Description
In contrast, capital letters A to D imply that
these calculator results are component
[channel] independent (Calc C is the result
of calculator C).
Emerson Process Management GmbH & Co. OHG3-20
Page 87
Short Form Manual
HASXEE-SFM-HS
04/2010
X-STREAM XE
3.8 Checking the Settings
Signal:Comp1
OutRange:0-20 mA
Low Scale:0.00
Max Scale:100.00
AutoScale:Yes
FailMode:Live
0/4 mA:0.00
20 mA:100.00
Hold:No
Next, select the output range:
• 0-20 mA (dead zero) generates a 20 mA
signal, if a concentration is measured at
the upper limit of the signal range. A 0 mA
signal is generated if the sample gas concentration equals the value specied with
"LowScale" .
• 4-20 mA (life zero): A 4 mA signal is generated if the concentration equals the value
specied with "Low Scale", thus enabling
to detect e.g. a broken cable.
If „Signal“ is assigned a concentration signal ,
and „Auto scale“ is set to No, in the next lines
enter the concentrations to output 0/4 mA
(„Low scale“) or 20 mA („High scale“).
If either „Auto scale“ is Yes, or „Signal“ not
assigned a concentration signal, the levels
are set with default values and not editable.
Startup
3
"FailMode" selects the output´s behaviour under failure conditions, considering or not, the
NAMUR recommendation NE 43.
Available options:
Track: NE 43 not considered; output always
correlates with the measured value.
HIGH + 10%: NE 43 failure signal level:
'above'.
LOW - 10%: NE 43 failure signal level:
'below'.
Further information on NE 43 is provided on
the next page.
Emerson Process Management GmbH & Co. OHG 3-21
Page 88
X-STREAM XE
3.8 Checking the Settings
Operation Modes acc. NAMUR NE 43
Short Form Manual
HASXEE-SFM-HS
04/2010
If "OutRange" is set to 0-20 mA, a 20
mA signal
is generated, if the measured concentration
is equal to "Max Scale"
. A 0 mA signal is generated if the sample gas concentration is 0
(dead zero).
However, a severed cable also results in a
signal value of 0. An external data acquisition
system thus cannot detect such an failure and
accepts a gas concentration of 0.
The commonly used method of detecting a
severed cable is to apply an offset: a concentration corresponding to the lower range value
is assigned an analog signal of 4 mA, enabling
to detect a severed or disconnected cable.
This live zero mode is activated by setting “Out
Range” to 4-20 mA.
Operation modes conforming to
NAMUR 43
(NE 43) recommendations
The operation modes described above do not
generate a signal which enables detection of a
failure in the measurement system. In such cases the behaviour of the output signal is undened: either the last value is held, or a random
value is sent. System failures thus cannot be
detected by an external data acquisition system.
NE 43 contains recommendations for setting
analog outputs in order to avoid these situations. These are implemented by X-STREAM
analyzers as follows:
Setting "FailMode" to HIGH + 10% or LOW - 10% denes specic analog output signals in
case of a failure. Since these values do not
occur under normal operation conditions, a data
acquisition system is enabled to distinguish
between the following situations (table 3-1):
• Valid signal (signal within valid range; column C)
• Signal out of range (signal rises or falls
slowly to the limits given in columns D or E
and holds this value until the concentration
returns to a valid level).
• Failure (signal out of range; column F)
• Severed cable (no signal; column G)
Output signal, if
ColumnABCDEFG
Failure
"OutRange" "FailMode"
0-20 mATrack-0 ... 20 mA< -19 mA> 21 mAundened0 mA
4-20 mATrack-4 ... 20 mA< -19 mA> 21 mAundened0 mA
Tab. 3-1: Analog output signals settings and operational modes
signal level
acc. NE 43
Measured
value is
valid
Measured value is
below lower limit
("Low scale")
Measured value is
above upper limit
("High scale")
Emerson Process Management GmbH & Co. OHG3-22
An internal
failure
occured
Cable is
severed
Page 89
Short Form Manual
HASXEE-SFM-HS
04/2010
Signal:Comp1
OutRange:0-20 mA
Low Scale:0.00
Max Scale:100.00
AutoScale:Yes
FailMode:Live
0/4 mA:0.00
20 mA:100.00
Hold:No
3.8 Checking the Settings
"0/4 mA" and „20 mA“ enable to netune the
analog output to compensate possible deviations based on electronics tolerances:
• Set "Signal" to 0 mA or 20 mA
• measure the output current,
• adjust it to the expected value.
Accepted range: -10,000 ... +10,000
"Hold" selects the output´s behaviour during
calibrations.
If set to Yes,
• the analog output is xed to the last measured value
• concentration alarms, which may otherwise
be triggered by the concentrations of the
calibration gases, are supressed.
If set to No,
• the analog output signal always corresponds to the actual measured value during
calibration; this may trigger alarms if limits
are exceeded.
Note!
This behaviour may be undesireable if e.g.
the unit is connected to a data acquisition
system.
X-STREAM XE
Startup
3
;
Emerson Process Management GmbH & Co. OHG 3-23
Page 90
X-STREAM XE
3.8 Checking the Settings
3.8.5 Setting Concentration Alarms
Short Form Manual
HASXEE-SFM-HS
04/2010
Note!
If concentration alarms are not being used,
continue with
3-25.
Ch1
Concentration
Alarm Monitor:On
LoLo Level:50.000 ppm
Lo Level:100.000 ppm
Hi Level:400.000 ppm
HiHi Level:600.000 ppm
Hysteresis:10.000 ppm
States..
Press
left until SETUP is displayed, then
open ALARMS - CONCENTRATION
Multi-channel unit:
Select the channel to be setup in SELECT
COMPONENT.
Four concentration limits can be set for each
channel:
"Lo" and "Hi" enframe the expected gas
concentration,
"LoLo" equals or is below "Lo",
"HiHi" equals or is above "Hi".
See the gure to the left for an explanation.
If you enter values for any threshold, the
above given order has to be considered. A
message is displayed, if an entered value
does no comply with this condition.
Fig. 3-3: Arrangement of concentration thresholds
Should the measured concentration go beyond one of the threshold levels (areas B, C, D
& E in the gure), a message is displayed in
the message line of the measurement display,
the NAMUR pictogram appears (bell) and a
corresponding digital output is activated, if
assigned.
A hysteresis avoids oscillating alarms in
case the concentration is uctuating around
a threshold.
Emerson Process Management GmbH & Co. OHG3-24
Page 91
Short Form Manual
HASXEE-SFM-HS
04/2010
X-STREAM XE
3.8 Checking the Settings
Ch1
Concentration Alarms
LoLo Alarm:Off
Lo AlarmOn
Hi AlarmOff
HiHi AlarmOff
Concentration75.000 ppm
3.8.6 Backup the Settings
You may turn the alarm function On or Off se-
parately for each channel ("Alarm monitor").
In case of an alarm, you may like to enter
the STATES submenu, to check which one
is triggered.
Startup
3
The most important settings have now been
checked and the unit is congured to suite
your needs.
A backup copy of these conguration data
can now be saved.
For now, make a local backup:
Enter LOCAL BACKUP and then select
"Save..".
Conrm the operation (select "Yes!").
Copying Data
Busy0
Progress (0..1000)1000
Press
to return
Wait until "Progress (0..1000)" shows 1000,
then press
enter to return to LOCAL BACK-
UP.
Congratulations!
You´ve now successfully completed check-
ing the unit’s setup!
• Press home to return to the MEASURE-
MENT DISPLAY.
Emerson Process Management GmbH & Co. OHG3-26
Page 93
Short Form Manual
HASXEE-SFM-HS
04/2010
3.9 Perform a Calibration
3.9 Perform a Calibration
We recommend to perform at least a zero
calibration, after startup of the instrument, to
ensure proper measuring results.
The following steps describe, how to perform a manual calibration. If your instrument
features a valve block,
instruction manual for a comprehensive description of calibration procedures.
X-STREAM XE
X-STREAM XE
Before performing an y actions, make sure the required
calibration gas is applied and
owing!
Supply all calibration gases
with the same ow as the sample gas (recommeded approx.
1 l/min), pressureless and utilizing
the correct gas tting.
Startup
3
Ensure the warm-up time after
switching on has elapsed!
Warm-up time is 15 to 50 minutes
depending on installed measuring system and conguration!
OPERATION AT LOW TEMPERATURES
When operating an instrument at temperatures below 0 °C (32 °F), do
NOT apply gas nor operate the internal pump before the warmup time has
elapsed!
Violation may result in condensation inside the gas paths or damaged
pump diaphragm!