Rosemount World Class 3000 O2 Analyzer CENELEC with IFT 3000 Intelligent Field Transmitter-Rev 4.2 Manuals & Guides

Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
Oxygen Analyzer (CENELEC) with IFT 3000 Intelligent Field Transmitter (CENELEC)
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical repre-
sentative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate In- struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s per­formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
HIGHLIGHTS OF CHANGES
Effective June, 1997 Rev. 4
Page Summary
--- General. Added snubber version of probe to manual. Text and art changed as necessary to reflect new style of probe.
Effective February, 1998 Rev. 4.1
Page Summary
Page 2-2 Figure 2-1. Change calibration gas tube dimensions.
Page 3-10 Add note on test gas flowmeter.
Effective January, 2002 Rev. 4.2
Page Summary
Page 2-13 Updated Figure 2-8.
Page 2-14 Updated analog output current/voltage mode selection procedure.
HIGHLIGHTS OF CHANGES
APPENDIX AX
Effective June, 1997 Rev. 2
Page Summary
-- General. Added snubber version of probe to manual. Text and art changed as necessary to reflect new style probe.
Effective February, 1998 Rev. 2.1
Page Summary
Page A-12 Changed screw torque in paragraph A-3h.
HIGHLIGHTS OF CHANGES
APPENDIX BX
Effective February, 1995 Rev. 1.1
Page Summary
B-3 Figure B-3. Updated for IB consistency.
Effective January, 1997 Rev. 1.2
Page Summary
Page B-2 Figure B-2. Added fuse locations.
Page B-5 Insert protective cover and ground lead warning.
Page B-8 Insert protective cover and ground lead warning.
Page B-11 Table B-2. Add expanded fuse specifications to replacement parts.
HIGHLIGHTS OF CHANGES
APPENDIX DX
Effective February, 1995 Rev. 2
Page Summary
-- General. Updated appendix with new version of MPS.
Effective January, 1997 Rev. 2.1
Page Summary
Page D-5 Insert protective cover and ground lead warning.
Page D-7 Insert protective cover and ground lead warning. Add fuse
specifications and clarify fuse replacement.
Page D-11 Add fuse specifications to replacement parts
HIGHLIGHTS OF CHANGES
APPENDIX EX
Effective February, 1995 Rev. 1.1
Page Summary
Page E-4 Figure E-2. Updated for IB consistency.
Page E-7 Figure E-4. Updated Flowchart.
Effective May, 1995 Rev. 1.2
Page Summary
Page E-4 Figure E-2. Added callout text “Heater Power Supply (Optional)”.
Effective January, 1997 Rev. 1.3
Page Summary
Page E-5 Insert protective cover and ground lead warning.
Page E-9 Insert protective cover and ground lead warning.
Page E-15 Added expanded fuse specifications to replacement parts
HIGHLIGHTS OF CHANGES
APPENDIX JX
Effective January, 1997 Rev. 1.0
Page Summary
Page J-4 Insert warning concerning protective equipment covers and safety
ground leads.
Page J-11 Insert warning concerning protective equipment covers and safety
ground leads.
World Class 3000
PREFACE.........................................................................................................................P1
Definitions ......................................................................................................................... P1
Safety Instructions ........................................................................................................... P2
1-0 DESCRIPTION ................................................................................................................ 1-1
1-1 Component Checklist Of Typical System (Package Contents) ....................................... 1-1
1-2 System Overview............................................................................................................. 1-1
2-0 INSTALLATION .............................................................................................................. 2-1
2-1 Oxygen Analyzer (Probe) Installation .............................................................................. 2-1
2-2 Intelligent Field Transmitter (IFT) Installation .................................................................. 2-9
2-3 Heater Power Supply Installation .................................................................................. 2-16
2-4 Multiprobe Test Gas Sequencer Installation ................................................................. 2-20
3-0 SETUP AND OPERATION ........................................................................................... 3-1
3-1 Overview.......................................................................................................................... 3-1
3-2 IFT with GUI and LDP Front Panel Controls and Indicators ........................................... 3-2
3-3 Help Key .......................................................................................................................... 3-2
3-4 Status Line....................................................................................................................... 3-3
3-5 Quick Reference Chart .................................................................................................... 3-3
3-6 Main Menu ....................................................................................................................... 3-3
3-7 Probe Data Sub-Menu .................................................................................................... 3-6
3-8 Calibrate O
3-9 Setup Sub-Menu.............................................................................................................. 3-6
3-10 System Calibration........................................................................................................... 3-9
ub-Menu ..................................................................................................... 3-6
2
Instruction Manual
106-300NFX Rev. 4.2
January 2002
TABLE OF CONTENTS
4-0 LDP OPERATION .......................................................................................................... 4-1
4-1 Overview.......................................................................................................................... 4-1
4-2 IFT with LDP Front Panel Controls and Indicators .......................................................... 4-1
4-3 LDP Displays ................................................................................................................... 4-1
4-4 LDP Defaults.................................................................................................................... 4-2
4-5 Calibration........................................................................................................................ 4-2
5-0 TROUBLESHOOTING .................................................................................................... 5-1
5-1 Overview.......................................................................................................................... 5-1
5-2 Special Troubleshooting Notes........................................................................................ 5-1
5-3 System Troubleshooting.................................................................................................. 5-2
6-0 RETURN OF MATERIAL .............................................................................................. 6-1
7-0 APPENDICES ................................................................................................................. 7-1
8-0 INDEX.............................................................................................................................. 8-1
9-0 DRAWINGS AND SCHEMATICS................................................................................. 9-1
Rosemount Analytical Inc. A Division of Emerson Process Management i
Instruction Manual
106-300NFX Rev. 4.2 January 2002
Figure 1-1. Typical System Package ....................................................................................... 1-1
Figure 1-2. Typical System Installation .................................................................................... 1-5
Figure 1-3. World Class 3000 Typical Application with Intelligent Field Transmitters -
Figure 2-1. Probe Installation ...................................................................................................2-2
Figure 2-2. Orienting the Optional Vee Deflector ..................................................................... 2-7
Figure 2-3. Air set, Plant Air Connection.................................................................................. 2-8
Figure 2-4. Outline of Intelligent Field Transmitter ................................................................... 2-9
Figure 2-5. Power Supply Board Jumper Configuration ........................................................ 2-10
Figure 2-6. IFT Power Supply Board Jumpers....................................................................... 2-11
Figure 2-7. Wiring Layout for IFT 3000 (CENELEC approved) System without HPS............ 2-12
Figure 2-8. IFT Microprocessor Board Jumper Configuration................................................ 2-13
Figure 2-9. IFT Microprocessor Board Jumpers .................................................................... 2-14
Figure 2-10. Interconnect Board Jumper Configuration........................................................... 2-14
Figure 2-11. IFT Interconnect Board Output Connections ....................................................... 2-15
Figure 2-12. Outline of CENELEC Approved Heater Power Supply........................................ 2-16
Figure 2-13. Wiring layout for IFT 3000 (CENELEC approved) with HPS ............................... 2-17
Figure 2-14. CENELEC Approved Heater Power Supply Wiring Connections ........................ 2-19
Figure 2-15. Jumper Selection Label. ...................................................................................... 2-19
Figure 2-16. Jumpers on HPS Motherboard ............................................................................ 2-20
Figure 2-17. MPS Module ........................................................................................................ 2-21
Figure 2-18. MPS Gas Connections ........................................................................................ 2-22
Figure 2-19. MPS Electrical Connections ................................................................................ 2-23
Figure 3-1. IFT with GUI and LDP Front Panel ........................................................................ 3-2
Figure 3-2. Typical Calibration Setup..................................................................................... 3-11
Figure 3-3. Portable Rosemount Oxygen Test Gas Kit......................................................... 3-12
Figure 3-4. Typical Portable Test Calibration Setup .............................................................. 3-13
Figure 3-5. Typical Automatic Calibration System ................................................................. 3-15
Figure 4-1. IFT with LDP Front Panel ...................................................................................... 4-1
World Class 3000
LIST OF ILLUSTRATIONS
CENELEC Approved ............................................................................................. 1-6
LIST OF TABLES
Table 3-1. Sample HELP Messages....................................................................................... 3-2
Table 3-2. Main Menu .............................................................................................................3-3
Table 3-3. PROBE DATA Sub-Menu. ................................................................................... 3-3
Table 3-4. CALIBRATION O
Table 3-5. SETUP Sub-Menu ................................................................................................. 3-8
Table 3-6. Efficiency Constants .............................................................................................. 3-9
Table 4-1. LDP Defaults.........................................................................................................4-3
ii Rosemount Analytical Inc. A Division of Emerson Process Management
Sub-Menu ................................................................................ 3-7
2
World Class 3000
The purpose of this manual is to provide information concerning the components, func­tions, installation and maintenance of this particular oxygen analyzer.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this publication.
Instruction Manual
106-300NFX Rev. 4.2
January 2002
PREFACE
DEFINITIONS
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an essential operating procedure, condition, or statement.
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETIN
NOTE TO USERS
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
The number in the lower right corner of each illustration in this publication is a manual illus­tration number. It is not a part number, and is not related to the illustration in any technical manner.
Rosemount Analytical Inc. A Division of Emerson Process Management P-1
Instruction Manual
106-300NFX Rev. 4.2 January 2002
FOR THE WIRING AND INSTALLATION
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non­EU states should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external, where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
World Class 3000
IMPORTANT

SAFETY INSTRUCTIONS

OF THIS APPARATUS
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be made through a circuit breaker which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker may also include a mechanically operated isolating switch. If not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazard­ous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment — and then only by trained service personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Cer­tain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to the Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the follow­ing standards: EN61010-1, IEC417, and ISO3864.
P-2 Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Instruction Manual
106-300NFX Rev. 4.2
January 2002
SECTION 1

DESCRIPTION

1-1 COMPONENT CHECKLIST OF TYPICAL
SYSTEM (PACKAGE CONTENTS)
A typical Rosemount World Class 3000 Oxygen Analyzer (CENELEC approved) with IFT 3000 Intelligent Field Transmitter (CENELEC approved) should contain the items shown in Figure 1-1. Record the Part Number, Serial Number, and Order Number for each compo­nent of your system in the table located on the cover of this manual.
The IFT 3000, Oxygen Analyzer (Probe), and probe abrasive shield are heavy. Lifting and carrying procedures should take account of this weight.
1-2 SYSTEM OVERVIEW
a. Scope
This Instruction Bulletin has been designed to supply details needed to install, start up, oper­ate, and maintain the Rosemount World Class 3000 Oxygen Analyzer (CENELEC approved) with IFT 3000 Intelligent Field Transmitter (CENELEC approved). The Intelligent Field Transmitter (IFT) can be interfaced with one World Class 3000 probe. The IFT provides all necessary intelligence for controlling the probe and optional MPS 3000 Multiprobe Gas Sequencer.
1
3
1. Intelligent Field Transmitter (CENELEC approved)
2. Instruction Bulletin
3. Multiprobe Test Gas Sequencer (Optional)
4. Heater Power Supply (CENELEC approved)(Optional)
5. Oxygen Analyzer (Probe) (CENELEC approved)
6. System Cable
7. Mounting Plate with Mounting Hardware and Gasket
8. Reference AIR set (optional)
R
O
S
E
M
O
U
N
T
8
7
6
Figure 1-1. Typical System Package
2
4
E
N
E
E L I
R
H
G I
W
S
E
N
D
E
P O T
G
O
N
N
I N
O
E
D
P
O
S
E
E
R
E
O
L
F
A
E
B
B
E
L
5
26030001
Rosemount Analytical Inc. A Division of Emerson Process Management Description 1-1
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
NOTE
The Rosemount encode sheets (Prod­uct Ordering Matrix) allow a customer to order either the hazardous area ver­sion of the IFT 3000 or the non­hazardous area version. The hazard­ous area version has the symbol "EExd" on the apparatus nameplate. The non-hazardous area version does not. Ensure that if you have received the non-hazardous version that you do not install it in a potentially explosive atmosphere. This also applies to the hazardous/non-hazardous versions of the HPS 3000.
b. System Description
The Rosemount Oxygen Analyzer (Probe) is designed to measure the net concentration of oxygen in an industrial process; i.e., the oxygen remaining after all fuels have been oxidized. The probe is permanently posi­tioned within an exhaust duct or stack and performs its task without the use of a sam­pling system.
The equipment measures oxygen percent­age by reading the voltage developed across a heated electrochemical cell, which consists of a small Yttria-stabilized, Zirconia disc. Both sides of the disc are coated with porous metal electrodes. When operated at the proper temperature, the millivolt output voltage of the cell is given by the following Nernst equation:
EMF = KT log
10(P1/P2
) + C
Where:
is the partial pressure of the oxygen
1. P
2
in the measured gas on one side of the cell,
2. P
is the partial pressure of the oxygen
1
in the reference gas on the other side,
3. T is the absolute temperature,
4. C is the cell constant,
5. K is an arithmetic constant.
For best results, use clean, dry, in­strument air (20.95% oxygen) as a ref­erence gas.
When the cell is at operating temperature, and there are unequal oxygen concentra­tions across the cell, oxygen ions will travel from the high partial pressure of oxygen side to the low partial pressure side of the cell. The resulting logarithmic output voltage is approximately 50 mV per decade. Because the magnitude of the output is proportional to the logarithm of the inverse of the sample of the oxygen partial pres­sure, the output signal increases as the oxygen concentration of the sample gas decreases. This characteristic enables the oxygen analyzer to provide exceptional sensitivity at low oxygen concentrations.
Oxygen analyzer equipment measures net oxygen concentration in the presence of all the products of combustion, including water vapor. Therefore, it may be considered an analysis on a "wet" basis. In comparison with older methods, such as the Orsat apparatus, which provides an analysis on a "dry" gas basis, the "wet" analysis will, in general, indicate a lower percentage of oxygen. The difference will be proportional to the water content of the sampled gas stream.
c. System Configuration
The equipment discussed in this manual consists of three major components: the oxygen analyzer (CENELEC approved) (probe), the intelligent field transmitter (CENELEC approved) (IFT), and an op­tional heater power supply (CENELEC approved) (HPS). The HPS is required when the cable run between the electronics and the probe exceeds 45 m (150 ft). There is also an optional multiprobe test gas sequencer (MPS), which can be used to facilitate the automatic calibration of a multiple probe configuration.
1-2 Description Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Instruction Manual
106-300NFX Rev. 4.2
January 2002
CENELEC approved probes are available in three length options, giving the user the flexibility to use an in situ penetration ap­propriate to the size of the stack or duct. The options on length are 457 mm (18 in.),
0.91 m (3 ft), and 1.83 m (6 ft). The probe is certified EExd IIB T1 [370°C (698°F)] to CENELEC standards EN50014 and EN50018.
The IFT contains electronics that control probe temperature (in conjunction with the optional HPS) and supply power, and pro­vide isolated outputs that are proportional to the measured oxygen concentration. The oxygen sensing cell is maintained at a con­stant temperature by modulating the duty cycle of the probe heater. The IFT accepts millivolt signals generated by the sensing cell and produces outputs to be used by remotely connected devices. The IFT output is isolated and selectable to provide linearized voltage or current.
The heater power supply CENELEC approved (HPS) can provide an interface between the IFT and the probe. The HPS contains a transformer for supplying proper voltage to the probe heater. The unit is certified EExd IIC T6 to CENELEC stan­dards EN50014 and EN50018.
Systems with multiprobe and multiple IFT applications may employ an optional MPS 3000 Multiprobe Test Gas Sequencer. The MPS 3000 provides automatic test gas sequencing for up to four probes and IFTs to accommodate automatic calibration. The MPS 3000 must be installed in a non­hazardous, explosive-free environment.
3. User friendly, menu driven operator interface with context-sensitive on-line help.
4. Field replaceable cell.
5. Analyzer constructed of rugged 316 LSS for all wetted parts.
6. The intelligent field transmitter (IFT) can be located up to 45 m (150 ft) from the probe when used without optional heater power supply (HPS). When the system includes the optional HPS, the HPS can be located up to 45 m (150 ft) from the probe and the IFT may be lo­cated up to 364 m (1200 ft) from the HPS.
7. All electronic modules are adaptable to 120, 220, and 240 line voltages.
8. Five languages may be selected for use with the IFT. These are:
English French German Italian Spanish
9. An operator can set up, calibrate, or troubleshoot the IFT in one of two ways:
(a) Optional General User Interface
(GUI). The GUI is housed within the IFT electronics enclosure and makes use of an LCD display and keypad.
d. System Features
1. Unique and patented electronic cell protection action that automatically protects sensor cell when the analyzer detects reducing atmospheres.
2. Output voltage and sensitivity increase as the oxygen concentration de­creases.
Rosemount Analytical Inc. A Division of Emerson Process Management Description 1-3
(b) Optional LED Display Panel (LDP).
The LED display and a limited function keypad permit calibration only.
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
e. Handling the Oxygen Analyzer
It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to pre­vent possible equipment damage.
The oxygen analyzer is designed for industrial application. Treat each component of the system with care to avoid physical damage. The probe contains components made from ce­ramics, which are susceptible to shock when mishandled. See Safety Data Sheets 1M03243, 1M03226, and 1M03296 for safety related informa­tion.
NOTE
Retain packaging in which the oxygen analyzer arrived from the factory in case any components are to be shipped to another site. This packag­ing has been designed to protect the product.
f. System Considerations
Once you have verified that you have all the components, select mounting locations and determine how each component will be placed in terms of available power supply, ambient temperatures, environmental con­siderations, convenience, and serviceability. A typical system installation is illustrated in Figure 1-2. Figure 1-3 shows a typical system wiring. For details on installing the individual components of the system, refer to Section 2, Installation.
After selecting the probe mounting location, provision should be made for a platform where the probe can be easily serviced. The intelligent field transmitter (IFT) can be located up to 45 m (150 ft) cabling distance from the probe when used without optional heater power supply (HPS). When the sys­tem includes the optional HPS, the HPS can be located up to 45 m (150 ft) cabling distance from the probe and the IFT may be located up to 364 m (1200 ft) cabling distance from the HPS.
A source of instrument air is required at the probe for reference gas use. Since the probe is equipped with an in-place calibra­tion feature, provision should be made for connecting test gas tanks to the oxygen analyzer when the probe is to be calibrated.
Prior to installation of your Rosemount CENELEC approved World Class 3000 Oxygen Analyzer with Intelligent Field Transmitter make sure that you have all of the components necessary to make the system installation. Ensure that all the com­ponents are properly integrated to make the system functional.
1-4 Description Rosemount Analytical Inc. A Division of Emerson Process Management
If the test gas bottles will be permanently hooked up, a check valve must be con­nected to the calibration gas fitting on the probe junction box. This is to prevent breathing of calibration gas line and subse­quent gas condensation and corrosion. The check valve is in addition to the stop valve in the test gas kit or the solenoid valve in the multiprobe test gas sequencer units.
World Class 3000
1
Instruction Manual
106-300NFX Rev. 4.2
January 2002
CALIBRATION
INSTRUMENT
AIR SUPPLY
(REF. GAS)
FIELD TRANSMITTER
GAS
PRESSURE
REGULATOR
FLOWMETER
INTELLIGENT
EExd IIB T6
GASES
STACK
STANDARD
DUCT
OXYGEN ANALYZER EExd IIB T1 (370°C) (PROBE)
HEATER POWER SUPPLY EExd IIC T6
ADAPTER PLATE
LINE VOLTAGE
}
GASES
OPTIONS
DUCT
*MULTIPROBE TEST
GAS SEQUENCER
(NON-HAZARDOUS AREA)
*NOTE: THE MPS 3000 MUST BE
INSTALLED IN A NON-HAZARDOUS EXPLOSIVE FREE ENVIRONMENT.
ADAPTER
CALIBRATION
SUPPLY
INST. AIR
TEST GAS 1
TEST GAS 2
REFERENCE AIR
FIELD TRANSMITTER
STACK
PLATE
GAS
INTELLIGENT
EExd IIB T6
OXYGEN ANALYZER EExd IIB T1 (370°C) (PROBE)
HEATER POWER SUPPLY EExd IIC T6
}
LINE VOLTAGE
23800017
Figure 1-2. Typical System Installation
Rosemount Analytical Inc. A Division of Emerson Process Management Description 1-5
Instruction Manual
106-300NFX Rev. 4.2 January 2002
Stack Thermocouple
(optional)
2-Conductor T/C
Wire [46 (150) max]
(optional)
(OPTIONAL)
Line Voltage
World Class 3000
4 Twisted Pair Plus 2 Twisted Pair
for Options [366 (1200) max]
Line Voltage
World Class 3000
Probe
World Class 3000
Probe
7-Conductor Cable
[46 (150) max]
2-Pneumatic Lines
by Customer
[91 (300) max]
HPS 3000
HPS 3000
Explosion Proof Required only for Hazardous Area Applications, otherwise use NEMA 4X. Lengths Exceeding 150 feet.
(OPTIONAL)
*MPS 3000
TEST GAS
SEQUENCER
Modular Design
Up to 4 Probes
[HPS not required for lengths of less than 46 (150) max]
7-Conductor Cable
[46 (150) max]
Line Voltage
Test Gas
by
Customer
IFT 3000
Intelligent Field Transmitter
Line Voltage 100 to 120 Volt 220 to 240 Volt
5 Conductor
[305 (1000) max]
Line Voltage
Stack Thermocouple
(optional)
*NOTE 1:
NOTE 2:
2-Conductor T/C
Wire [46 (150) max]
(optional)
THE MPS 3000 MUST BE INSTALLED IN A NON-HAZARDOUS EXPLOSIVE FREE ENVIRONMENT.
ALL DIMENSIONS APPEAR IN METERS WITH FEET IN PARENTHESES.
IFT 3000
Intelligent Field Transmitter
Line Voltage 100 to 120 Volt 220 to 240 Volt
P00003
Figure 1-3. World Class 3000 Typical Application
with Intelligent Field Transmitters - CENELEC Approved
1-6 Description Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
106-300NFX Rev. 4.2
January 2002
SECTION 2

INSTALLATION

2-1 OXYGEN ANALYZER (PROBE)
INSTALLATION
Before probe installations, consult probe Safety Data Sheet 1M03226.
The probe and probe abrasive shield are heavy. Use proper lifting and carrying procedures to avoid personnel injury.
Install all protective equipment covers and safety ground leads after installa­tion. Failure to install covers and ground leads could result in serious injury or death.
a. Selecting Location
provides you with mechanical installa­tion references.
2. Check the flue or stack for holes and air leakage. The presence of this condition will substantially affect the accuracy of the oxygen reading. Therefore, either make necessary repairs or install the probe upstream of any leakage.
3. Ensure that the area is clear of obstructions internal and external that will interfere with installation. Allow adequate clearance for removal of probe (Figure 2-1).
Do not allow the temperature of the probe junction box to exceed 150°C (302°F) or damage to the unit may result. If the probe junction box tem­perature exceeds 150°C (302°F), the user must fabricate a heat shield or provide adequate cooling air to the probe junction box.
1. The location of the probe in the stack or flue is most important for maximum accuracy in the oxygen analyzing pro­cess. The probe must be positioned so that the gas it measures is representa­tive of the process. Best results are normally obtained if the probe is posi­tioned near the center of the duct (40 to 60% insertion). A point too near the edge or wall of the duct may not pro­vide a representative sample because of the possibility of gas stratification. In addition, the sensing point should be selected so that the process gas tem­perature falls within a range of 10° to 704°C (50° to 1300°F). Figure 2-1
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
b. Mechanical Installation
1. Ensure that all components are avail­able for installation of the probe. Check the ceramic filter to ensure that it is not damaged and that the system cable is the required length.
2. The probe may be installed intact as it is received. It is recommended that you disassemble the adapter plate for each installation.
3. Weld or bolt adapter plate (Figure 2-1) onto the duct.
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
o
DIM
DIM
MM
DIMENSIONS
TABLE II. REMOVAL/INSTALLATION
DIN
"B"
755
"A"
459
(IN.)
1U05680G01
210
1209
(29.7)
913
(18.07)
18 IN. PROBES
1U05680G02
18
(8.25)
2126
(47.6)
1831
(35.95)
3 FT PROBES
1U05680G03
170
(0.708)
(83.7)
EXTERNAL EARTH
(72.09)
6 FT PROBES
(6.693)
(HARDWARE)
INTERNAL EARTH
VIEW B-B
(TERMINAL 6)
AT BOTTOM
(ROTATED 90 CCW)
INSTALL WITH CONNECTIONS
DIM "B"
REMOVAL ENVELOPE
200 (7.87)
B
110 (4.33)
1.57 (0.062) THK GASKET
(P/N 1M03237H01)
ELECTRICAL
CONNECTION
B
BOTTTOM VIEW
DIM "A"
MM
(IN.)
DIMENSIONS
TABLE I. MOUNTING FLANGE
FLANGE
DIA
HOLE SIZE
DIA
4 HOLES EQ
SP ON B.C. DIA
PROCESS FLOW MUST BE IN
THIS DIRECTION WITH RESPECT
TO VEE SHIELD WHEN USING
OPTIONAL CERAMIC DIFFUSOR.
76 (3.0)
DIA MAX
CENELEC APPROVED WORLD CLASS PROBE WITH SNUBBER DIFFUSOR
A
23800006
VIEW A-A
INCHES IN PARENTHESES UNLESS
NOTE: DIMENSIONS ARE IN MILLIMETERS WITH
A
OTHERWISE INDICATED.
Figure 2-1. Probe Installation (Sheet 1 of 5)
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
WITH INCHES IN PARENTHESES
UNLESS OTHERWISE INDICATED.
NOTE: DIMENSIONS ARE IN MILLIMETERS
DIM "D"
(REMOVAL
ENVELOPE)
378
(14.9)
DETAILS
SEE SHEET 1
FOR CONNECTION
Instruction Manual
106-300NFX Rev. 4.2
January 2002
ELECTRICAL
CONNECTION
CHECK
VALV E
REF, VENT,
AND CAL GAS
CONNECTIONS
DIM "D"
(REMOVAL
DIM
REMOVAL/INSTALLATION
MM
DIMENSIONS
769
(30.3)
ENVELOPE)
"C"
387
(15.3)
(IN.)
18 IN. PROBE 1U05680G04
SHIELD 1N04966H01
1224
2144
(48.2)
(84.4)
843
1762
(33.2)
(69.4)
3 FT PROBE 1U05680G05
SHIELD 1N04966H02
6 FT PROBE 1U05680G06
SHIELD 1N04966H03
DIM "C"
178
(7.00)
(P/N 4507C52H03)
1.52 (0.06) THK GASKET
FURNISHED IN
HARDWARE PACKAGE
SEAL HUB
DIFFUSOR/DUST
(P/N 1U05677G03)
ABRASIVE SHIELD INSTALLATION WITH CENELEC APPROVED WORLD CLASS 3000
23800007
Figure 2-1. Probe Installation (Sheet 2 of 5)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
o
DIM
DIM
MM
DIMENSIONS
TABLE II. REMOVAL/INSTALLATION
DIN
"B"
755
"A"
554
(IN.)
1U05680G01
210
1209
(29.7)
1008
(21.81)
18 IN. PROBES
1U05680G02
18
(8.25)
2126
(47.6)
1926
(39.68)
3 FT PROBES
1U05680G03
170
(0.708)
(83.7)
EXTERNAL EARTH
(75.83)
6 FT PROBES
(6.693)
(HARDWARE)
VIEW B-B
(TERMINAL 6)
INTERNAL EARTH
AT BOTTOM
(ROTATED 90 CCW)
INSTALL WITH CONNECTIONS
DIM "B"
REMOVAL ENVELOPE
200 (7.87)
B
110 (4.33)
1.57 (0.062) THK GASKET
(P/N 1M03237H01)
ELECTRICAL
CONNECTION
B
BOTTTOM VIEW
DIM "A"
MM
(IN.)
DIMENSIONS
TABLE I. MOUNTING FLANGE
FLANGE
DIA
HOLE SIZE
DIA
4 HOLES EQ
SP ON B.C. DIA
OPTIONAL CERAMIC DIFFUSOR WITH VEE DEFLECTOR
PROCESS FLOW MUST BE IN
THIS DIRECTION WITH RESPECT
TO VEE SHIELD.
VIEW A-A
INCHES IN PARENTHESES UNLESS
NOTE: DIMENSIONS ARE IN MILLIMETERS WITH
OTHERWISE INDICATED.
76 (3.0)
DIA MAX
A
A
23800011
Figure 2-1. Probe Installation (Sheet 3 of 5)
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
106-300NFX Rev. 4.2
January 2002
"A"
"A"
PROBE TO STACK
(NEW INSTALLATIONS)
TABLE IV. MOUNTING PLATE:
MOUNTING PLATE OUTLINE
ABRASIVE SHIELD TO STACK
(NEW INSTALLATIONS)
TABLE III. MOUNTING PLATE:
215
DIN
MM
DIMENSIONS
DIN
MM
DIMENSIONS
(IN.)
(IN.)
"A"
235
"A"
(8.5)
(9.25)
M-16x2
"B"
THREAD
100
(3.94)
"B"
DIA
170
(6.7)
"C"
DIA
M-20 x 2.5
"C"
THREAD
18 IN., 3 FT AND6FT
MOUNTING PLATE FOR
SEE SHEET 2.
ABRASIVE SHIELD INSTALLATIONS
MOUNTING PLATE FOR
CENELEC WORLD CLASS 3000
SEE SHEET 1.
PROBE INSTALLATIONS
"A"
AND NUTS, EQUALLY
SPACED ON "C" DIA B.C.
4 STUDS, LOCKWASHERS,
"B"
DIA
"C" DIA
82.5 (3.25)
WITH INCHES IN PARENTHESES
UNLESS OTHERWISE INDICATED.
"A"
"C"
NOTE: DIMENSIONS ARE IN MILLIMETERS
o
22.5
8 THREADED HOLES
189.992 (7.48) DIA B.C.
EQUALLY SPACED ON
P00006
Figure 2-1. Probe Installation (Sheet 4 of 5)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
106-300NFX Rev. 4.2 January 2002
INSTALLATION FOR METAL
WALL STACK OR DUCT
CONSTRUCTION
World Class 3000
INSTALLATION FOR MASONRY
WALL STACK CONSTRUCTION
MTG HOLES
SHOWN ROTATED
o
22.5 OUT OF
TRUE POSITION
WELD OR BOLT
MOUNTING PLATE TO
METAL WALL OF STACK
OR DUCT. JOINT MUST
BE AIR TIGHT.
13 (0.50)
95 (3.75)
MIN DIA HOLE IN WALL
STACK OR DUCT METAL WALL
BOLT ADAPTOR
PLATE TO OUTSIDE
WALL SURFACE
FIELD WELD
PIPE TO
ADAPTOR PLATE
MTG HOLES
SHOWN ROTATED
o
22.5 OUT OF
TRUE POSITION
JOINT MUST
BE AIRTIGHT
OUTSIDE WALL
SURFACE
NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT
MOUNTING PLATE NOT FURNISHED BY ROSEMOUNT.
ABRASIVE SHIELD MOUNTING
13 (0.50)
114 (4.50) O.D. REF
PIPE 4.00 SCHED 40 PIPE SLEEVE (NOT BY ROSEMOUNT) LENGTH BY CUSTOMER
MASONRY STACK WALL
WELD OR BOLT MOUNTING
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIR TIGHT.
82.5 (3.25)
MIN DIA HOLE IN WALL
STACK OR DUCT METAL WALL
BOLT MOUNTING
PLATE TO OUTSIDE
WALL SURFACE
BE AIRTIGHT
OUTSIDE WALL
PROBE MOUNTING
JOINT MUST
SURFACE
FIELD WELD PIPE TO ADAPTOR PLATE
102 (4.0) O.D. REF
PIPE 3.5 SCHED 40 PIPE SLEEVE (NOT BY ROSEMOUNT) LENGTH BY CUSTOMER
MASONRY STACK WALL
P00007
Figure 2-1. Probe Installation (Sheet 5 of 5)
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
106-300NFX Rev. 4.2
January 2002
4. If using the optional ceramic diffuser element, the vee deflector must be cor­rectly oriented. Before inserting the probe, check the direction of flow of the gas in the duct. Orient the vee deflector on the probe so that the apex points upstream toward the flow (Figure 2-2). This may be done by loosening the setscrews, and rotating the vee de­flector to the desired position. Retighten the setscrews.
5. In horizontal installations, the probe cover should be oriented so that the system cable drops vertically from the probe cover. In a vertical installation, the system cable can be oriented in any direction.
6. If the system has an abrasive shield, check the diffusion element dust seal packings. The joints in the two pack­ings must be staggered 180°. Also, make sure that the packings are in the hub grooves as the probe slides into the 15° forcing cone in the abrasive shield.
GAS FLOW DIRECTION
VEE
APEX
FILTER
DEFLECTOR
CERAMIC
DIFFUSION
ELEMENT
VEE
DEFLECTOR
SETSCREW
23800008
Figure 2-2. Orienting the Optional Vee Deflector
7. Insert the probe through the opening in the mounting flange and bolt the unit to the flange.
8. Ensure that probe is properly earthed by way of both the internal and external points.
NOTE
If process temperatures will exceed 1000°F (538°C), use anti-seize com­pound on stud threads to ease future removal of probe.
9. Ensure that the installation does not obscure the messages on either the probe nameplate or the junction box lid.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
c. Reference Air Package
After the oxygen analyzing (probe) unit is installed, connect the reference gas air set to the probe cover. The reference gas air set should be installed in accordance with Figure 2-3.
d. Service Required.
1. Power input: 44 VAC from HPS 3000 or IFT 3000.
2. Compressed air: 68.95 kPa (10 psig) minimum, 1551.38 kPa (225 psig)
0.125-27 NPT FEMALE
OUTLET CONNECTION
122.17 (4.81)
FLOW SET
POINT KNOB
1
2
OUTLET
30.22 (1.19)
DRAIN VALVE
maximum at 56.6 L/hr (2 scfh) maxi­mum; supplied by one of the following (less than 40 parts-per-million total hy­drocarbons).
(a) Instrument air - clean, dry.
(b) Bottled standard air with step-down
regulator.
(c) Bottled compressed gas mixture
(20.95% oxygen in nitrogen).
(d) Other equivalent clean, dry, oil-free
air supply.
79.25 (3.12) MAX
3
57.15 (2.250)
50.80 (2.0)
38.10 (1.50)
0.25-18 NPT FEMALE INLET CONNECTION
NOTE: DIMENSIONS ARE IN MILLIMETERS
WITH INCHES IN PARENTHESES.
215.90 MAX (8.50)
2 MOUNTING HOLES
81.03 (3.19) LG
THROUGH BODY FOR
7.92 (0.312) DIA BOLTS
254 REF
(10.0)
6 (0.250) OD
TUBE COMPRESSION
FITTING (SUPPLIED BY WECO)
TO PROBE
JUNCTION BOX
6 (0.250) OD TUBING
(SUPPLIED BY CUSTOMER)
SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYGEN ANALYZER PROBE HEAD.
REF GAS SET
1 FLOWMETER 0.2-2.0 SCFH 771B635H02
2 2" PRESSURE GAGE 0-15 PSIG 275431-006
3 COMBINATION FILTER-REG. 0-30 PSIG 4505C21G01
COMPRESSED AIR SUPPLY 10-225 PSIG MAX PRESSURE
263C152G01
NOTE: DIMENSIONS ARE IN MILLIMETERS
WITH INCHES IN PARENTHESES.
Figure 2-3. Air set, Plant Air Connection
17300016
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
2-2 INTELLIGENT FIELD TRANSMITTER (IFT)
INSTALLATION
The Rosemount Encode Sheet (Prod­uct Ordering Matrix) allows a customer to order either the hazardous area ver­sion of the IFT 3000 or the non­hazardous area version. The hazard­ous area version has the symbol "EExd" on the apparatus nameplate. The non-hazardous area version does not. Ensure that if you have the non­hazardous area version that you do not install it in a potentially explosive environment. This warning applies equally to the hazardous area and non-hazardous area versions of the HPS 3000.
Instruction Manual
106-300NFX Rev. 4.2
January 2002
The IFT 3000 is heavy. Lifting and car­rying procedures should take account of this weight.
a. Mechanical Installation
The outline drawing of the IFT module (CENELEC approved) in Figure 2-4 shows mounting centers and clearances. The en­closure is designed to be mounted on a wall. The IFT should be installed no more than 364 m (1200 ft) from the optional HPS or 45 m (150 ft) from the probe if HPS is not installed in the system. Ambient tempera­ture must be between 0°C and 50°C (32°F and 122°F).
NOTE
Fuse specifications are included in Figure 2-4.
Before IFT 3000 Installation, consult Safety Data Sheet 1M03296.
310 (12.2)
270 (10.6)
EXTERNAL
EARTH
FUSES
INTERNAL
EARTH
350 (13.8)
NOTES: ALL DIMENSIONS ARE IN MILLIMETERS WITH INCHES IN PARENTHESES UNLESS
OTHERWISE INDICATED.
FUSES SHOWN (F3 THROUGH F6) ARE 5 AMP, ANTI-SURGE, TYPE T TO IEC127 (ROSEMOUNT PART NUMBER 1L01293H02). IF INTERNAL HEATER IS INSTALLED, TWO ADDITIONAL 5 AMP FUSES (F1 AND F2) ARE USED.
320
(12.6)
18 (0.7)
350
(13.8)
300
(11.8)
225
(8.9)
35870001
Figure 2-4. Outline of Intelligent Field Transmitter
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Instruction Manual
106-300NFX Rev. 4.2 January 2002
ALWAYS DISCONNECT LINE VOLTAGE FROM INTELLIGENT FIELD TRANSMITTER BEFORE CHANGING JUMPERS.
PROBE HEATER
VOLTAGE SELECTION
WORLD CLASS PROBE (44V)
LINE VOLTAGE
SELECTION
120 V.A.C.
220 V.A.C.
240 V.A.C.
JUMPER
(INSTALL)
JM8, JM7, JM1
JM6, JM5, JM2
JM6, JM5, JM1
JUMPER
CONFIGURATION
If you reconfigure the equipment for a line voltage other than the one marked on the serial label and the mains filter of the power supply then you should change the marking on the serial label and the mains filter to state the new line voltage.
World Class 3000
JUMPER
(INSTALL)
JM10
P00010
If incorrect heater voltage is selected, damage to the probe may occur. For HPS voltage selection jumper, refer to Figure 2-15.
Figure 2-5. Power Supply Board Jumper Configuration
b. Electrical Connections
1. The IFT can be configured for 100, 120, 220, or 240 line voltages. For 120 Vac usage, install JM8, JM7, and JM1. For 220 Vac usage, install jumpers JM6, JM5, JM2 (refer to Figure 2-5 and
Do not install jumper JM6 on the mi­croprocessor board, or JM1 on the in­terconnect board, if an HPS is installed in the system. This will result in system failure.
Figure 2-6).
3. The IFT must have JM6 on the micro­processor board (Figure 2-8 and Figure
If you reconfigure the equipment for a line voltage other than the one marked on the serial label and the mains filter of the power supply you should change the marking on the serial label and the mains filter to state the new line voltage.
2-9) and JM1 on the interconnect board (Figure 2-10 and Figure 2-11) installed if an HPS is not installed in the system.
4. If an MPS is not used in the system, wire jumper between CAL RET and NO GAS must be installed on the intercon­nect board. Remove wire jumper if
2. The IFT can be configured to connect directly to a probe or to an optional
MPS is installed in the system. Refer to
Figure 2-7, note 6. HPS. The electrical connections for a non-HPS equipped system should be made as described in the electrical in­stallation diagram, Figure 2-7.
5. The power cable should comply with the safety regulations in the user's country and should not be smaller than 16 gauge, 3 amp.
2-10 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
106-300NFX Rev. 4.2
January 2002
THIS TERMINAL BLOCK IS A 44VAC OUTPUT SPECIFICALLY FOR POWERING THE
!
WC 3000 PROBE.
P00011
Figure 2-6. IFT Power Supply Board Jumpers
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11
Instruction Manual
106-300NFX Rev. 4.2 January 2002
PROBE INTERIOR
GN CELL -VE
OR CELL +VE
YE CHROMEL
RD ALUMEL
123456 78
RD
GN
WH
SM. BR
PROBE JUNCTION
BOX WIRING
GN
GN/YE
World Class 3000
3D3911BG REV
HEATER
}
BK
BK
BL
LG. BR
LINE VOLTAGE SECTION
120 V.A.C. 220 V.A.C. 240 V.A.C.
JUMPER CONFIGURATION
JUMPER
(INSTALL) JM8, JM7, JM1 JM6, JM5, JM2 JM6, JM5, JM1
J1
3D39122G REV
POWER SUPPLYBOARD
ALWAYS DISCONNECT LINE VOLTAGE FROM INTELLIGENT FIELD TRANSMITTER BEFORE CHANGING JUMPERS.
PROBE HEATER
VOLTAGE SECTION
WORLD CLASS PROBE
218 PROBE
JUMPER
(INSTALL)
JM10
JM9
MICROPROCESSOR BOARD
J2
J3
J4
J1
3D39120G REV
INTERCONNECT BOARD
JM5JM6
WC 3000 PROBE (CENELEC APPROVED)
NOTES
STACK TC WIRING AS REQUIRED
SPECIAL PROBE CABLE BETWEEN PROBE AND IFT BY ROSEMOUNT (FITTED WITH EExd GLANDS)
INSTALL JM1 ON INTERCONNECT BOARD
INSTALL JM6 ON MICROPROCESSOR BOARD
IF STACK TEMPERATURE NOT USED
IF MPS 3000 NOT USED
1 RELAY PER PROBE AVAILABLE FOR CALIBRATION STATUS INDICATION (48 V max, 100 mA max)
CABLE COLORS SHOWN HERE APPLY TO ROSEMOUNT SUPPLIED SPECIAL CABLE FITTED WITH EExd GLANDS (P/N 1U03066)
J5 J6
H
L
EEN
LINE
VOLTAGE
LG. BR BL
GN/YE SHIELD
SM. BR
WH GN
RD
INTELLIGENT FIELD TRANSMITTER IFT3000
LINE OUT
ER
PROBE 1
J10
J11
J19 J20 J21 J22
PROBE 1
HI GAS
IN CAL
NC NC NCNC C C CCNONONONO
PROBE 2
MPS3000 MULTI GAS SEQUENCER (OPTIONAL)
CAL RET
CAL RET NO GAS
LO GAS HI GAS IN GAS
SHIELD
5 CONDUCTOR SHIELDED CABLE PER PROBE #16 AWG BY CUSTOMER
3D39064G REV
NO GAS
LO GAS
PROBE 2
IN CAL
HI GAS
PROBE 3
CAL RET
LO GAS
NO GAS
PROBE 4
PROBE 3
HI GAS
MPS TERMINATION BOARD
J5
J6
J7
SHIELD STACK TC ­STACK TC +
J8
SHIELD
PROBE TC -
GN RD
PROBE TC +
J9
SHIELD
WH
PROBE MV -
PROBEMV+
SM.BR
PROBE 4
NO GAS
LO GAS
HI GAS
IN CAL
CAL RET
NO GAS
IN CAL
CAL RET
J13 J14 J15 J16 J17 J18
JM1
J9
J8J7J6J5J4J3J2J1
LINE IN
LL
NN
LO GAS
L
LINE VOLTAGE
N
J12
35870007
Figure 2-7. Wiring Layout for IFT 3000 (CENELEC approved) System without HPS
2-12 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
106-300NFX Rev. 4.2
January 2002
JM7
JM7
CURRENT/VOLTAGE SELECTOR SWITCH
SW1
J4
JM6
TO I/O
BOARD
JM6
TP5 TP6
TP4
+30VISO-C GNDC +5VISO-C
TP3
-15V
TP8
+5V
TP2
TP7
+15V
TP1
-5V
J1
TO POWER
SUPPLY CARD
J2
3D39513G REV
TO LDP CARD
J3
TO GUI CARD
Figure 2-8. IFT Microprocessor Board Jumper Configuration
29850004
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-13
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
OUTPUT JUMPER
HPS Probe (No HPS)
(See Figure 2-8 for jumper locations.)
Figure 2-9. IFT Microprocessor Board Jumpers
OUTPUT JUMPER
HPS Probe (No HPS)
Figure 2-10. Interconnect Board Jumper
Configuration
6. Before supplying power to the IFT, ver­ify that the jumpers have been properly set in the IFT, Figure 2-5, Figure 2-8, and Figure 2-10.
7. Terminal strip J5 on the power supply board is used for supplying the IFT with power. Terminal strip J6 on the power supply board is used to supply the probe heater with power if an HPS is not used (Figure 2-6).
8. Ensure that the IFT 3000 is properly earthed by way of both the internal and external earthing hardware.
Remove JM6 Install JM6
Remove JM1 Install JM1
c. Analog Output and Relay Output
Connections
1. The microprocessor board has a switch to select voltage or current op­erations. Figure 2-8 shows the switch location. In voltage mode, output is 0­10 V. In the current mode, the output can be configured from the setup menu to be 0-20 mA or 4-20 mA.
2. The analog output and relay outputs are programmed by the user as needed. The analog output is typically sent to recording equipment such as chart recorders. Relay outputs are typi­cally sent to annunciators.
3. Relays K1 and K2 are user configur­able from the PROBE SETUP sub­menu (Table 3-5). Typically these are used to indicate O below specified tolerances. OK relay is energized when unit is functioning properly.
4. All wiring must conform to local and national codes.
5. Connect the analog output and relay outputs as shown in Figure 2-11.
values above or
2
9. Ensure that the installation does not obscure the message on either the IFT nameplate or the IFT lid.
2-14 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
NOTES:
DENOTES SHIELD CONNECTION.
OK RELAY IS ENERGIZED WHEN UNIT IS FUNCTIONING PROPERLY.
OK-NC
K1-NC
K2-NC
CAL INIT-2
CAL INIT-1
CALRET
NOGAS
LOGAS
HIGAS
INCAL
RELAY
RELAY+
AD590
AD590+
TRIAC
TRIAC+
Instruction Manual
106-300NFX Rev. 4.2
January 2002
24
24 23
22
21 20 19
18 17 16 15
14 13 12
11 10
9
8
-
7 6
-
5
4 3 2
-
1
OK-COM
OK-NO
23
22
K1-COM
K1-NO
21
K2-COM
20 19
K2-NO
18
17
16
15
14 13 12
ANOUT-
11
ANOUT+
10
9
STACK T/C
8
STACK T/C
7 6
PROBE T/C
5
4
PROBE T/C
3 2
PROBE MV
1
PROBE MV+
-
JM1 (UNDER SHIELD)
730002
Figure 2-11. IFT Interconnect Board Output Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-15
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000

2-3 HEATER POWER SUPPLY INSTALLATION

The Rosemount encode sheets (Prod­uct Ordering Matrix) allow a customer to order either the hazardous area ver­sion of the HPS 3000 or the non­hazardous area version. The hazard­ous area version has the symbol "EExd" on the apparatus nameplate. The non-hazardous area version does not. Ensure that if you have received the non-hazardous version that you do not install it in a potentially explosive atmosphere. This also applies to the hazardous/non-hazardous version of the IFT 3000.
Before HPS installation, consult Safety Data Sheet 1M03243.
264.0
(10.39)
253
(9.96)
233
(9.17)
EExd IIC T6
14.22 (0.56) DIA
MTG HOLE (2 PLS)
ENCLOSURE
120
(4.72)
NOTE: DIMENSIONS ARE IN MILLIMETERS
WITH INCHES IN PARENTHESES UNLESS OTHERWISE INDICATED.
216.0
(8.50)
157
(6.18)
219005
a. Mechanical Installation
The outline drawing of the CENELEC ap­proved heater power supply enclosure in Figure 2-12, shows mounting centers and clearances. The CENELEC approved en­closure is designed to be mounted on a wall or bulkhead. The heater power supply should be installed no further than 45 m (150 ft) from the probe. The heater power supply must be located in a location free from significant ambient temperature changes and electrical noise. Ambient tem­perature must be between 0° to 60°C (32° to 140°F).
b. Electrical Connections
1. Electrical connections should be made as described in the electrical installa­tion diagram, Figure 2-13. The wiring terminals are divided into two layers: the bottom (FROM PROBE) terminals should be connected first, the top (FROM ELECTRONICS) terminals should be connected last (Figure 2-14). Each terminal strip has a protective cover which must be removed when making connections. To remove the
Figure 2-12. Outline of CENELEC Approved Heater
Power Supply
terminal covers, remove two slotted screws holding cover in place. Always reinstall terminal covers after making connections.
2. Power Input: 120, 220 or 240 Vac. For 120 Vac usage, install jumpers JM4 and JM1 and remove JM5 if installed. For 220 or 240 Vac usage, install jumper JM5 and remove JM1 and JM4 if installed (see label, Figure 2-15).
If you reconfigure the equipment for a line voltage other than the one marked on the serial label and the mains filter of the power supply then you should change the marking on the serial label and the mains filter to state the new line voltage.
NOTE
Fuse specifications are shown in Figure 2-14.
2-16 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
106-300NFX Rev. 4.2
January 2002
PROBE INTERIOR
HEATER
GN CELL -VE
OR CELL +VE
YE CHROMEL
RD ALUMEL
GN
}
BK
BK
12345678
RD
WH
GN
SM. BR
PROBE JUNCTION
BOX WIRING
GN/YE
BL
LG. BR
WC PROBE 3000 CENELEC APPROVED
LINE VOLTAGE
SELECTION
120 V.A.C.
220/240 V.A.C.
PROBE HEATER
VOLTAGE SECTION
WORLD CLASS PROBE JM7
JUMPER
(INSTALL)
JM4, JM1
JM5
JUMPER
(INSTALL)
JUMPER
CONFIGURATIONS
NOTES
RELAY WIRE IS OPTIONAL; RELAY CAN BE PERMANENTLY ENABLED WITH JUMPER IF NOT USED
STACK TC WIRING AS REQUIRED
ALL WIRES #16-#22 AWG TWISTED PAIR WITH SHIELD BY CUSTOMER EXCEPT AS NOTED
SPECIAL PROBE CABLE BETWEEN PROBE AND HPS BY ROSEMOUNT
REMOVE JM1 ON INTERCONNECT BOARD (IFT 3000)
REMOVE JM6 ON MICROPROCESSOR BOARD
IF RELAY WIRE OF NOTE 1 INSTALLED THEN REMOVE JM2 ON HPS 3000
IF STACK TEMPERATURE NOT USED
IF MPS 3000 NOT USED
1 RELAY PER PROBE AVAILABLE FOR CALIBRATION STATUS INDICATION (48 V max, 100 mA max)
CABLE COLORS SHOWN HERE APPLY TO ROSEMOUNT SUPPLIED SPECIAL CABLE FITTED WITH EExd GLANDS (P/N 1U03066)
ALWAYS DISCONNECT LINE VOLTAGE FROM HEATER POWER SUPPLY AND ANALOG ELECTRONICS (IF USED) BEFORE CHANGING JUMPERS.
HEATER POWER
REMOTE
ON
ELECTRONICS
SELECTION
NEW GENERATION
ELECTRONICS
JUMPER
REMOVE JM2
INSTALL JM2
JUMPER
REMOVE JM3, JM6
TOP
TRIAC RELAY
J9
+
+++
-
STACKTCANALOG
---
HEATER
BK
WH
2 TWISTED PAIR SHIELDED #22 AWG BY CUSTOMER (OPTIONAL)
A
PROBE
SM. BR CELL+ WH CELL -
RD HTR TC +
GN HTR TC -
BL LG. BR GN/YE
SHIELD
PROBE
MV
J8
+
-
BOTTOM
PROBE
MV
J3
+
-
STACK
TC
J2 J1
+
-
TC
PROBE
TC
+
PROBE
HEATER
RH
AD590
+
-
-
B
4 TWISTED PAIR SHIELDED #22 AWG BY CUSTOMER
LINE
L
N
LINE VOLTAGE
HPS 3000 INTERFACE MODULE
35870008
Figure 2-13. Wiring layout for IFT 3000 (CENELEC approved) with HPS (Sheet 1 of 2)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-17
Instruction Manual
106-300NFX Rev. 4.2 January 2002
3D39118G
MICROPROCESSOR
BOARD
World Class 3000
5
JM6 JM5
LINE VOLTAGE SECTION
100 V.A.C. 120 V.A.C. 200 V.A.C. 220 V.A.C. 240 V.A.C.
J1
ALWAYS DISCONNECT LINE VOLTAGE FROM INTELLIGENT FIELD TRANSMITTER BEFORE CHANGING JUMPERS.
JUMPER CONFIGURATION
JUMPER
(INSTALL) JM3, JM7, JM2 JM8, JM7, JM1 JM4, JM5, JM2 JM6, JM5, JM2 JM6, JM5, JM1
PROBE HEATER
VOLTAGE SECTION
3D39122G REV
POWER SUPPLYBOARD
JUMPER
NOT USED REMOVE
(INSTALL)
JM9, JM10
J2
J3
J4
J1
3D39120G REV
INTERCONNECT BOARD
J5
CAL RET
J5 J6
NOT USED
L
EN
LINE
VOLTAGE
A
NO GAS
LO GAS HI GAS IN CAL
SHIELD RELAY – RELAY + SHIELD AD590 –
AD590 + SHIELD TRIAC –
TRIAC +
J6
J7
J8
J9
SHIELD STACK TC – STACK TC +
SHIELD
PROBE TC – PROBE TC +
SHIELD
PROBE MV – PROBE MV +
JM1
B
INTELLIGENT FIELD
TRANSMITTER IFT 3000
PROBE 1
LINE OUT LINE IN
L
IN CAL
HI GAS
N
J10
NC C NO NC C NO NC C NO
J19
J11
J20 J21
PROBE 1
MPS3000 MULTIPROBE CALIBRATION GAS SEQUENCER (OPTIONAL)
J3
J2J1
PROBE 2 PROBE 3 PROBE 4
IN CAL
HI GAS
LO GAS
NO GAS
CAL RET
PROBE 2 PROBE 3
CAL RET
MPS TERMINATION BOARD
5 CONDUCTOR SHIELDED CABLE PER PROBE #16 AWG BY CUSTOMER
3D390646 REV
LO GAS
NO GAS
NC C NO
J22
PROBE 4
IN CAL
HI GAS
CAL RET
PROBE 1
SOLENOID
J13 J14 J15 J16 J17 J18
NO GAS
PROBE 2
LO GAS
SOLENOID
PROBE 3
SOLENOID
HI GAS
IN CAL
PROBE 4
SOLENOID
J8J7J6J5J4
LO GAS
NO GAS
CAL RET
HIGH GAS
SOLENOID
LOW GAS
SOLENOID
L
N
SWITCH
PRESSURE
J12
J9
L
LINE
E
VOLTAGE
N
34990011
Figure 2-13. Wiring layout for IFT 3000 (CENELEC approved) with HPS (Sheet 2 of 2)
2-18 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
106-300NFX Rev. 4.2
January 2002
SCREW
(2 PER COVER)
JM8
JM7
TERMINAL
COVERS
(PROVIDED)
INTERNAL
EARTHING
HARDWARE
EXTERNAL
EARTHING
HARDWARE
J7
JM4
JM5
JM2
1 JM
J2
FUSE
FUSES
FRONT VIEW SIDE VIEW
NOTE: FUSES SHOWN (F1 THROUGH F4) ARE 5 AMP, ANTI-SURGE,
TYPE T TO IEC127 (ROSEMOUNT PART NUMBER 1L01293H02).
TRANSFORMER
TERMINAL STRIP (FROM ELECTRONICS)
TERMINAL STRIP (FROM PROBE)
219006-1
Figure 2-14. CENELEC Approved Heater Power Supply Wiring Connections
ALWAYS DISCONNECT LINE VOLTAGE FROM HEATER POWER SUPPLY AND ANALOG ELECTRONICS (IF USED) BEFORE CHANGING JUMPERS.
HEATER
POWER
*ON
REMOTE
ELECTRONICS
SELECTION
NEW GENERATION
ELECTRONICS
JUMPER
INSTALL JM2
REMOVE JM2
JUMPER
REMOVE JM3, JM6
219007
LINE VOLTAGE
SELECTION
120 V.A.C.
220/240 V.A.C.
PROBE HEATER
VOLTAGE SELECTION
*WORLD CLASS PROBE
(44V)
JUMPER
(INSTALL)
JM4, JM1
JM5
JUMPER
(INSTALL)
JM7
JUMPER
CONFIGURATIONS
Figure 2-15. Jumper Selection Label.
If you reconfigure the equipment for a line voltage other than the one marked on the serial la­bel and the mains filter of the power supply then you should change the marking on the serial label and the mains filter to state the new line voltage.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-19
Instruction Manual
106-300NFX Rev. 4.2 January 2002
3. The power cable should comply with safety regulations in the user's country and should not be smaller than 16 gauge, 3 amp.
Before supplying power to the heater power supply, verify that jumpers JM3 and JM6 are removed, and JM7 is in­stalled. If relay wire (Figure 2-13, Note
1) is installed, JM2 must be removed from HPS Motherboard (Figure 2-16).
4. Before supplying power to the heater power supply, verify that the jumpers on the motherboard, Figure 2-16, are properly configured. Jumpers JM3, JM6, should be removed and JM7 should be installed. Additionally, make sure that the proper jumper for your line voltage is installed, Figure 2-15. If relay wire (Figure 2-13, note 1) is not installed, JM2 should be installed on the HPS Motherboard (Figure 2-16).
World Class 3000
Figure 2-16. Jumpers on HPS Motherboard
2-4 MULTIPROBE TEST GAS SEQUENCER
INSTALLATION
5. Ensure that the HPS 3000 is properly earthed by way of both the internal and external earthing points.
6. Ensure the installation does not ob­scure the messages on either the HPS nameplate or HPS lid.
NOTE
Refer to Figure 2-8 and Figure 2-10 for proper IFT jumper configuration. IFT microprocessor and interconnect board jumper configurations must be set correctly in order for HPS to work properly.
The MPS 3000 Multiprobe Test Gas Sequencer must be installed in a non­hazardous, explosive-free environ­ment.
NOTE
A Z-Purge option is available for the MPS 3000. Appendix DX contains in­formation concerning the Z-Purge.
a. Mechanical Installation
The outline drawing of the MPS module in Figure 2-17 shows mounting centers and clearances. The box is designed to be mounted on a wall or bulkhead. The MPS module should be installed no further than 91 m (300 ft) piping distance from the probe, and no more than 303 m (1000 ft) cabling distance from the IFT. Install the MPS module in a location where the ambi­ent temperature is between -30° and 71°C (-20° and 160°F).
2-20 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
304.80
(12.00)
254.00
(10.00)
Instruction Manual
106-300NFX Rev. 4.2
January 2002
304.80
(12.00)
LOWCAL
HIGHCAL
GASIN
GASIN
INSTR
AIR
355.60 (14.00) REF
b. Gas Connections
Figure 2-18 shows the bottom of the MPS where the gas connections are made. 1/4 in. threaded fittings are used.
1. Connect the reference air supply to INSTR. AIR IN. The air pressure regulator valve is set at the factory to
PROBE1 PROBE 2 PROBE 3 PROBE4
TESTGAS
TESTGAS
TESTGAS
REFAIR
TESTGAS
OUT
OUT
OUT
REFAIR
REFAIR
OUT
OUT
OUT
REFAIR
OUT
OUT
21.34 (0.84)
49.78 (1.96)
78.49 (3.09)
106.93 (4.21)
133.35 (5.25)
140.72 (5.54)
Figure 2-17. MPS Module
DIMENSIONS ARE IN
NOTE:
MILLIMETERS WITH INCHES IN PARENTHESES.
35870002
138 kPa (20 psi). If the reference air pressure should need readjustment, turn the knob on the top of the valve until the desired pressure is obtained.
2. Connect the high O
test gas to HIGH
2
GAS. The test gas pressure should be set at 138 kPa (20 psi).
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-21
Instruction Manual
106-300NFX Rev. 4.2 January 2002
DRAIN
PROBE 1 PROBE2 PROBE 3 PROBE4
LOW CAL
TEST GAS
TEST GAS
TEST GAS
HIGH CAL
GAS IN
INSTR
GAS IN
OUT
OUT
REF AIR
REF AIR
OUT
OUT
AIR
OUT
REF AIR
OUT
TEST GAS
OUT
REF AIR
OUT
Figure 2-18. MPS Gas Connections
World Class 3000
LINE IN
SIGNAL IN
35870003
Do not use 100% nitrogen as a low (zero) gas. It is suggested that the low gas be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon con­centrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.
3. Connect the low O2 test gas to LOW GAS. The test gas pressure should be set at 138 kPa (20 psi).
4. Connect the REF AIR OUT to the ref­erence gas fitting on the probe junction box.
5. Connect the TEST GAS OUT to the calibration gas fitting on the probe junction box.
6. If the MPS is configured for multiple probes (up to four), repeat steps 4 and 5 for each additional probe.
A check valve is required for each probe connected to an MPS to prevent condensation of flue gas in the cali­bration gas lines. The check valve must be located between the calibra­tion fitting and the gas line.
c. Electrical Connections
Electrical connections should be made as described in the electrical installation dia­gram, Figure 2-19. All wiring must conform to local and national codes. The electrical connections will exist only between the electronics package and the MPS to enable automatic and semiautomatic calibration. If more than one probe system is being used, the additional probes and electric packages would be wired similarly to the first probe.
NOTE
MPS power supply fuse locations and specifications are shown in Figure 2-19.
2-22 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
106-300NFX Rev. 4.2
January 2002
Figure 2-19. MPS Electrical Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-23
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
1. Run the line voltage through the bulk­head fitting on the bottom of the MPS where marked LINE IN, Figure 2-18. Refer to Figure 2-19. Connect the line voltage to the LINE IN terminal on the MPS terminal board located inside the unit. Tighten the cord grips to provide strain relief.
2. The MPS can accommodate up to four probes. The terminal strips on the MPS termination board are marked PROBE 1, PROBE 2, PROBE 3, and PROBE 4.
NOTE
!
Upon completing installation, make sure that the probe is turned on and operating prior to firing up the combustion process. Damage can result from having a cold probe exposed to the process gases.
Select PROBE 1 if this is the first probe and electronic package installed on the MPS.
3. Make the connections from the MPS to the IFT as shown in Figure 2-19. Run wires from the MPS Termination Board inside the unit through the bulkhead fit­ting on the bottom of the unit where marked SIGNAL IN, Figure 2-18. After the connections are made, tighten the cord grips to provide strain relief.
During outages, and if possible, leave all probes running to prevent condensation and premature aging from thermal cycling.
If the ducts will be washed down during outage, MAKE SURE to power down the probes and remove them from the wash area.
2-24 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
See Safety Data Sheet 1M03296 for safety related information.
Instruction Manual
106-300NFX Rev. 4.2
January 2002
SECTION 3

SETUP AND OPERATION

3-1 OVERVIEW
Ensure that the oxygen analyzer, heater power supply, and intelligent field transmitter have been properly connected. It is important to check that grounding and screening of termina­tions are correctly made to prevent the introduc­tion of ground loops. The IFT is equipped with noise suppression circuitry on the power supply and signal input lines. Proper grounding at in­stallation will ensure accuracy of function.
The following five languages can be selected within the IFT:
English Italian French Spanish German
The Intelligent Field Transmitter may be sup­plied with any one of three configurations. These configurations are the blind version, dis­play only version, and the deluxe version. The three versions differ as follows.
a. Blind Version
HART connections must be made out­side of the hazardous area. Because the Hart option is not protected by en­ergy limiting barriers, it must not be interfaced from within a hazardous area. The signal cables should be routed outside the hazardous area and the connections made external to the hazardous area.
The blind version has no display and no keypad. With this version, an external HART communications device is required. Refer to Appendix JX regarding the HART Commu­nications option.
b. Display Only Version (LDP)
The display only version is also known as the LED Display Panel (LDP) version. This IFT contains a bright LED display and a four-key pad. The LDP version provides for calibration capability only.
c. Deluxe Version (GUI)
The deluxe version is also known as the General User Interface (GUI) version. This IFT contains an LED display, liquid crystal display panel, and an eight-key pad that al­lows probe and electronics configuration, calibration, and troubleshooting of the probe and electronics.
This section of the manual deals with operator controls and displays available with the GUI equipped IFT. Operating parameters are listed and instructions are included for viewing and changing them.
Operating instructions for the IFT equipped with the LDP and four membrane keys are included in Section 4.
Any procedures not associated with normal op­eration are included in Section 2, Installation, or Section 5, Troubleshooting.
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-1
Instruction Manual
106-300NFX Rev. 4.2 January 2002
5
CAL TGH
4
3
2
1
ESC
13 12 11 10
TGL
SETUP
CALHELP
DATA
ENTER
Figure 3-1. IFT with GUI and LDP Front Panel
3-2 IFT WITH GUI AND LDP FRONT PANEL
CONTROLS AND INDICATORS (See Figure 3-1.)
Fig. 3-1
Index
No.
Control/
LED Description
1 HELP Context sensitive HELP is
displayed when this key is pressed.
2 DATA DATA key is used to access
DATA menu.
3LCD
Display
Top line displays system status, menu and probe number.
4LED
Display
Indicates current O value (only in LDP equipped units).
5 CAL Calibration in progress
indicator light (only in LDP equipped units).
6 TGH High test gas indicator light.
High test gas is being used in calibration process (only in LDP equipped units).
7 TGL Low test gas indicator light.
Low test gas is being used in calibration process (only in LDP equipped units).
34990016
or test
2
6
7
8
9
World Class 3000
8 CAL CAL key used to access
CALIBRATE menu.
9 SETUP SETUP key used to access
SETUP menu.
10 ENTER The ENTER key is used to
select a lower level menu, initiate calibration, or select a parameter to change.
11
12
13 ESC The escape key is used to
3-3 HELP KEY
The HELP key will display explanatory informa­tion about a menu, sub-menu, or parameter that the asterisk is next to when pressed. The HELP key is not available during calibration routines. Refer to Table 3-1 for sample HELP messages.
Table 3-1. Sample HELP Messages
MENU, SUB-MENU,
HELP OR
PARAMETER NAME MESSAGE
PROBE DATA Press ENTER key
CALIBRATE O
2
SETUP The SETUP menu
The increase key is used to move the cursor (asterisk) when scrolling through lists or to increase a parameter value.
The decrease key is used to move the cursor (asterisk) when scrolling through lists or to decrease a parameter value.
exit to a high level menu or to abort a parameter change.
to access DATA menu.
The CAL menu is used to start cali­bration and view calibration.
is used to configure the IFT-3000.
3-2 Setup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
Instruction Manual
106-300NFX Rev. 4.2
January 2002

3-4 STATUS LINE

The top line of the LCD display (3, Figure 3-1) is a status line that always displays system status, menu name, and O
level. The system status
2
displays are:
a. OK - System is functioning correctly.
b. CAL - Calibration in progress.
c. C Err - Calibration error.
d. H Err - Heater error.
e. TGLow - Test gas is low.
f. HiO
- O
value is above the high alarm
2
2
limit.
- O
g. LoO
value is below the low alarm limit.
2
2
h. R Hi - Resistance is above the high limit.
i. Off - The probe has been turned off be-
cause the IFT cannot control the heater temperature.

3-5 QUICK REFERENCE CHART

The quick reference chart on pages 3-4 and 3-5 is designed to help you determine how to get where you want to be in the menu system. The chart shows all the available menu and sub­menu options for the IFT. Follow the lines to determine which choices to make. Moving down a level on the chart is accomplished by use of the ENTER key. To move up a level on the chart, press the ESCAPE key.

3-6 MAIN MENU

When power is first applied to the IFT, the MAIN menu (Table 3-2) is initially displayed. It is from the MAIN menu that the PROBE DATA (Table 3-3), CALIBRATE O
(Table 3-4), and SETUP
2
(Table 3-5) menus can be accessed.
Table 3-2. Main Menu
MENU SELECTION DESCRIPTION
PROBE DATA Refer to Table 3-3.
CALIBRATE O
2
Refer to Table 3-4.
j. PRBE - The probe is disconnected, cold, or
SETUP Refer to Table 3-5.
leads are reversed.
Table 3-3. PROBE DATA Sub-Menu.
SUB-MENU
SELECTION PARAMETER DESCRIPTION
Process Data O
2
__% Efficiency ENA/DIS Stack Temp __DegC
O
value for the probe.
2
Enable/Disable efficiency display. Stack temperature.
Diagnostic Data
Temperature Cell __DegC
Stack __DegC Cold Junct __DegC
Voltages Cell __mV
Cell T/C __mV Stk T/C __mV Cold Jnt __mV
Output Values Analog __% FS
K1 State OFF/ON K2 State OFF/ON
Cell temperature of the probe. Stack temperature. Cold junction temperature.
Cell voltage of the probe. Cell thermocouple voltage of the probe. Stack thermocouple voltage. Cold junction voltage.
Analog output voltage. Status of relay 1. Status of relay 2
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-3
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
QUICK REFERENCE CHART
MAIN MENU
A
PROBE DATA
PROCESS DATA DIAGNOSTIC DATA
O
2
EFFICIENCY
STACK TEMP
TEMPERATURE
CELL
STACK
COLD JUNCT
VOLTAGES
CELL
CELL T/C
STK T/C
COLD JNT
OUTPUT VALUES
ANALOG
K1 STATE
K2 STATE
CALIBRATE O
PERFORM
CALIBRATION
2
VIEW
CONSTANTS
SLOPE
CONSTANT
RESIST
CALIBRATION
STATUS
NEXT CAL
SLOPE
CONSTANT
RESIST
3-4 Setup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
A
SETUP
Instruction Manual
106-300NFX Rev. 4.2
January 2002
CALIBRATION
HIGH GAS
LOW GAS
AUTO CAL
OUTPUT TRACKS
CAL INTRVL
NEXT CAL
GAS TIME
PURGE TIME
RES ALARM
EFFICIENCY CALC
ENABLE CALC
K1 VALUE
K2 VALUE
K3 VALUE
O CALCULATION
SLOPE
CONSTANT
SET POINT
LOAD CONSTANTS
RELAY OUTPUT
K1 SETUP
EVENT 1
EVENT 2
EVENT 3
O ALARMS
K2 SETUP
EVENT 1
EVENT 2
EVENT 3
22
HI ALARM
LO ALARM
ALARM DB
ANALOG OUTPUTS
SOURCE
AOUT TYPE
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-5
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000

3-7 PROBE DATA SUB-MENU

The PROBE DATA sub-menu is a list of all the parameters of the system as it is currently con­figured. To access the PROBE DATA sub­menu, press the DATA key at any time. The in­crease and decrease keys are used to scroll through the list. The PROBE DATA sub-menu can be viewed but not changed. The operator must use the SETUP menu to change any of the parameters.
There are two selections available on the PROBE DATA sub-menu; Process Data and Diagnostic Data. Refer to Table 3-3 for contents of the sub-menu.
3-8 CALIBRATE O
The CALIBRATE O
SUB-MENU
2
sub-menu (Table 3-4) is
2
used to enter the calibration mode. To access the CALIBRATE O
sub-menu, press the CAL
2
key at any time. The increase and decrease keys are used to scroll through the list.
The CALIBRATE O
sub-menu has three selec-
2
tions available: Perform Calibration, View Con­stants, and Calibration Status. Refer to Table 3-4 for contents of the sub-menus.
Perform Calibration has two options depending on how Auto Cal is selected in Probe Setup. Refer to SETUP Setting in Table 3-4.
3-9 SETUP SUB-MENU
The SETUP sub-menu is used to enter all op­erator set variables into the system. To access the SETUP sub-menu, press the SETUP key at any time. To select the parameter to be changed, move the cursor to the desired pa­rameter using the arrow keys. Press ENTER to select that parameter. To change the value for that parameter, use the arrow keys to increase or decrease the value. Press ENTER to save changes.
There are six selections available on the SETUP sub-menu: Calibration, O O
Alarms, Efficiency Calc., Relay Outputs, and
2
Analog Outputs. Refer to Table 3-5 for the con­tents of the SETUP sub-menu.
NOTE
Dip shunt is to be set for voltage or current operation. Refer to Figure 2-7.
Calculation,
2
3-6 Setup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
Instruction Manual
106-300NFX Rev. 4.2
January 2002
Table 3-4. CALIBRATION O2 Sub-Menu
SUB-MENU
SELECTION
Perform Calibration
View Constants
Calibration Status
SETUP SETTING (SEE TABLE 3-4) DISPLAY DESCRIPTION
Auto Cal in Probe Setup is YES
Auto Cal in Probe Setup is NO.
(N/A) Slope _____mV/D
(N/A) Slope _____
Press ENTER to start Auto Calibration. MPS will start calibration probe.
Starting Automatic Calibration
High Gas _____%O Time Left 0:00
Cell mV ______mV Cell voltage of the probe.
Low Gas _____%O Time Left 0:00
Cell mV ______mV Cell voltage of the probe.
Resistance Check Time Left 0:00
Cell _____mV _____C Cell voltage and probe temperature.
Calibration Complete Purging 0:00
Cell _____mV _____C Cell voltage and probe temperature.
Calibration Complete
Press ENTER to start Manual Calibration. Manual calibration sequence will begin
Switch ON high test gas. Press ENTER when ready.
High gas ______%O
Press ENTER when O2 reading is stable.
Turn off high test gas and ON low test gas. Press ENTER when ready.
Low gas ______%O
Press ENTER when O2 reading is stable.
Resistance Check Resistance check in progress.
Turn off low test gas. Press ENTER when ready.
Press ENTER when probe has returned to process.
Constant _____mV Resist _____ohms
Next Cal XD XH Time until next calibration in number of
Constant _____ Resist _____
2
2
2
2
Value for high O Amount of time necessary to complete the current testing phase in min:sec.
Value for low O Amount of time necessary to complete the current testing phase in min:sec.
Resistance check in progress.
Gas lines are being purged of test gas.
when ENTER is pressed.
High O2 test gas value.
Low O2 test gas value.
Slope for probe. Offset for probe. Resistance for probe.
days and number of hours.
Status of the slope. Status of the offset. Status of the resistance.
test gas.
2
test gas.
2
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-7
Instruction Manual
106-300NFX Rev. 4.2 January 2002
Table 3-5. SETUP Sub-Menu
SUB-MENU SELECTION PARAMETERS DESCRIPTION
World Class 3000
Calibration High Gas ____%O
Low Gas ____%O Auto Cal YES/NO Output Tracks YES/NO Cal Intrvl XD XH
Value of high O
2
Value of low O
2
If system has MPS, select YES or NO. Select output tracks. Select time between calibrations in number of
test gas.
2
test gas.
2
days and number of hours. (1 year max.)
Next Cal XH
Time until next calibration in number of hours. (1 year max.)
Gas Time 0:00
Amount of time the test gases will be turned on in number of minutes and seconds, allow
enough time for signal values to stabilize. Purge Time 0:00 Res Alarm _______
O2 Calculation Slope ____ mV/D
Constant ____ mV Set Point ____°C
Amount of time for gas lines to clear of test gas.
Resistance alarm set from 50 to 10,000 ohms.
Set value between 34.5 and 57.5.
Set value between -20.0 and +20.0 mV.
Set either 736 for World Class 3000 probes or
843 for 218 probes.
Ensure the correct voltage is selected when using HPS 3000 with either WC 3000 probes or 218 probes. Refer to Figure 2-14, Jumper Selection Label for proper voltage selections. If in­correct SET POINT is selected, damage to the probe may occur.
Load Constants Press ENTER to load constants from last
calibration.
O2 Alarms
Efficiency Calc.
Relay Outputs
K1 Setup
K2 Setup
Analog Outputs
Hi Alarm ____% O Lo Alarm ____% O Alarm DB ____% O
Enable Calc. YES/NO K1 Value _____ K2 Value _____ K3 Value _____
Event 1 _____ Event 2 _____ Event 3 _____
Event 1 _____ Event 2 _____ Event 3 _____
Source (EFF or O
) 0-100%
2
Aout Type 0-10V
Set value for high alarm limit.
2
Set value for low alarm limit.
2
Set value for alarm dead band.
2
Select YES to enable, NO to disable.
Set between 0.0000 and 1.0. Refer to Table 3-6.
Set between 0.0000 and 1.0. Refer to Table 3-6.
Set between 1.000 and 20.0. Refer to Table 3-6.
Set event to activate relay.
Set event to activate relay.
Set event to activate relay.
Set event to activate relay.
Set event to activate relay.
Set event to activate relay.
Source can be set to 0-1%, 0-5%, 0-10%, 0-
25%, 0-100% of O
, or 0-100% Efficiency
2
Aout Type can be set to 0-10 V, 0-20 mA, or
4-20 mA
3-8 Setup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
Instruction Manual
106-300NFX Rev. 4.2
January 2002
Table 3-6. Efficiency Constants
UNITED STATES
CONSTANT
GAS OIL
K1 0.407 0.432 0.66 0.69
K2 0.0 0.0 0.0082 0.0051
K3 5.12 5.12 12.28 8.74
3-10 SYSTEM CALIBRATION
a. Overview
The primary purpose of an oxygen analyzer is to give an accurate representation of the percentage of O
in the gas stream. The
2
system should be calibrated periodically to maintain an accuracy which may otherwise reduce over time due to cell aging.
A requirement for calibration is a set of two accurate test gases spanning the oxygen range of most interest. For example, 0.4% and 8% for a 0-10% oxygen range.
Under normal conditions the probe should not need frequent calibration. Because cali­bration is necessary, the system can be equipped with the optional MPS 3000 Multi­probe Test Gas Sequencer for fully auto­matic calibration at regular intervals. Without an MPS, the probes must be cali­brated manually (semi-automatically).
b. Probe Calibration
1. Previous Calibration Constants Func­tionality. Three sets of registers are used to store calibration constants. These are: Latest Calibration, Previous Calibration, and Calculation. Only the values in the Calculation registers are used to calculate the oxygen value for display and representation on the analog output signal. These values may be changed in two ways:
(a) The operator may change the Cal-
culation values by entering the SETUP menu and then entering
EUROPE
GAS OIL
the O
Calculations sub-menu. The
2
operator may adjust the slope and constant individually or reset both to the values calculated during the last good calibration. Simultane­ously resetting both values is done by selecting Load Constants and pressing ENTER.
(b) The IFT will automatically change
the values after each successful calibration.
The values in the Latest Calibration registers are updated after every complete calibration even if the cali­bration is not successful. If the calibra­tion is successful, the values in the Latest Calibration registers are copied into the Previous Calibration registers. This function is accomplished prior to the update of the Latest Calibration registers. The values from successful calibrations are automatically loaded into the Calculation registers for use in future O
calculations. If a calibration
2
fails, the Previous Calibration registers and Calculation registers retain their existing values, while the Latest Cali­bration registers record the values of the failed calibration.
2. Calibration Methods. There are three calibration methods: manual (semi­automatic), manually initiated auto­matic, and fully automatic. Manual (semiautomatic) calibration is done without an MPS unit. Test gases are switched on and off by the operator and the IFT is sequenced through the calibration procedure by the operator with the front panel keyboard. The IFT prompts the operator for the correct action. Manually initiated automatic calibration is done with an MPS. The operator manually initiates the calibra­tion at the IFT or through a remote switch, and the IFT controls the opera­tion of the MPS unit and the calibration sequencing. Fully automatic calibration requires no action from the operator. The setup is the same as semiauto­matic except the IFT automatically
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-9
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
initiates the calibration at a fixed cali­bration interval. In this mode the op­erator can also manually initiate calibrations between the intervals in the same manner as semiautomatic calibrations.
c. Manual (Semiautomatic) Calibration
1. Test Gases for Manual (Semiauto­matic) Calibration. There are two op­tions for supplying test gases to the probe during semiautomatic calibration. The first "A" uses refillable bottles and adjustable 2-stage pressure regulators; the second, "B" uses disposable bot­tles and a fixed single-stage regulator to provide a mixed flow. Normally, the first (method "A") will have a higher cost and not be portable. The second ("B") is less costly, portable, and weighs about 4.54 kg (10 lbs).
Test Method "A" Fixed Tanks and Manifolds.
(a) Required Equipment.
700 South Alameda Street Los Angeles, California 90058 213/585-2154
767 Industrial Road San Carlos, California 94070 415/592-7303
9950 Chemical Road Pasadena, Texas 77507 713/474-4141
12054 S.W. Doty Avenue Chicago, Illinois 60628 312/568-8840
603 Bergen Street Harrison, New Jersey 07029 201/485-1995
255 Brimley Road Scarborough, Ontario, Canada 416/266-3161
SCOTT ENVIRONMENTAL TECHNOLOGY, INC. SCOTT SPECIALTY GASES
Do not use 100% nitrogen as a zero gas. It is suggested that gas for the zero be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon con­centrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.
1 Two tanks of precision cali-
bration gas mixtures. Rec­ommended calibration gases are nominally 0.4 percent and
8.0 percent oxygen in nitro­gen.
Two sources of calibrated gas mixtures are:
LIQUID CARBONIC GAS CORP. SPECIALTY GAS LABORATORIES
2600 Cajon Blvd. San Bernardino, CA 92411 714/887-2571 TWX: 910-390-1159
1290 Combermere Street Troy, MI 48084 314/589-2950
Route 611 Plumsteadville, PA 18949 215/766-8861 TWX: 510-665-9344
2616 South Loop West, Suite 100 Houston, TX 77054 713/669-0469
2 A check valve is required at
the probe (between the cali­bration fitting and the gas line) to prevent the migration of process gases down the cali­bration gas line.
3-10 Setup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
Instruction Manual
106-300NFX Rev. 4.2
January 2002
3 Two, 2-stage pressure regu-
lators with stainless steel dia­phragms for tanks. Maximum output required: 138 kPa (20 psi).
4 One instrument air pressure
regulator: 138 kPa (20 psi) maximum and a supply of clean, dry instrument air.
5 Two zero-leakage shutoff
valves.
6 Miscellaneous oil-free tubing
and fittings.
(b) Calibration
1 A typical calibration setup is
shown in Figure 3-2. Care must be taken that all fittings are tight and free from oil or other organic contaminants. Small openings can cause back diffusion of oxygen from the atmosphere even though positive pressures are main­tained in the lines.
NOTE
The probe calibration gas fitting has a seal cap which must be in place at all times except during calibration.
In addition to the precision calibration gas mixtures, clean, dry, oil-free instrument air should be used for calibra­tion.
For optimum accuracy, this calibration should be run with the process at normal temperature and operating conditions.
When the calibration gas line exceeds 1.8 m (6 ft) in length from the leak tight valves, a
check valve, Rosemount P/N 6292A97H02, should be in­stalled next to the calibration gas connection on the probe to prevent breathing of the line with the process gas and subsequent gas condensation and corrosion.
CALIBRATE
IN-PLACE
FITTING
FLOW METER
LEAK TIGHT
VALV ES
NOTE:
PROBE CALIBRATION GAS FITTING HAS A SEAL CAP THAT MUST BE IN PLACE AT ALL TIMES EXCEPT DURING CALIBRATION.
PROBE (END VIEW)
REFERENCE AIR
CONNECTION
5 SCFH
REG
0.4% O
2
CHECK VALV E
8.0% O
2
Figure 3-2. Typical Calibration Setup
2 SCFH
REFERENCE
AIR
SET
INSTR. AIR IN
730013
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-11
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
NOTE
Only set the test gas flowmeter upon initial installation and after changing the diffusion element. A slightly lower test gas flow rate may indicate a plugged diffusion element.
2 Set the test gas pressure
regulators and the flow meter for a flow of 5 scfh at (20 psig) 138 kPa for both gases. The reference gas should be flowing as in normal opera­tion.
3 Refer to paragraph 3-10d of
this section for Manual (Semi­automatic) Calibration setup and procedure using the IFT.
4 Test gases will be switched
on and off using the shutoff valves.
Test Method "B" Rosemount Oxygen Test Gas and Service Kit.
(a) Required Equipment
2 Extra gas bottles are available
at:
Rosemount Analytical Inc. Box 901 Orrville, Ohio 44667 U.S.A.
Rosemount Limited Burymead Road Hitchin, Herts. U.K.
Rosemount Italy VIA Guido Cavalcanti 8 20127 Milan, Italy
Rosemount Spain Saturnino Calleja 6 28002 Madrid, Spain
Rosemount France 165 Boulevard de Vallmy
92706, Colombes, France
Rosemount Part Number 3530B07G01 for probe 0.4% oxygen in nitrogen in dispos­able bottle.
Do not use 100% nitrogen as a zero gas. It is suggested that gas for the zero be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon con­centrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.
1 Portable Rosemount Oxygen
Test Gas Kits (Figure 3-3), Rosemount Part Number
TEST GAS
KIT #1
(P/N 6296A27G1)
6296A27G01, containing 8% and 0.4% gases in a port­able carrying case with regu­lator, built-in valve, hose and hose connecting adapter to the calibration gas connec­tion.
35870004
Figure 3-3. Portable Rosemount
Oxygen Test Gas Kit
3-12 Setup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
Instruction Manual
106-300NFX Rev. 4.2
January 2002
Rosemount Part Number 3530B07G02 for probe 8% oxygen in nitrogen in dispos­able bottle.
3 A check valve is required at
the probe (between the cali­bration fitting and the gas line) to prevent the migration of process gases down the cali­bration gas line.
(b) Calibration with a Portable Rose-
mount Oxygen Test Gases Kit.
1 A typical portable test calibra-
tion setup is shown in Figure 3-4. For Manual (semiauto­matic) calibration, remove cap plug from the calibrate-in­place fitting. The cap plug must be retained to seal this fitting after calibration is com­plete; failure to do so may render the probe useless if the system pressure is slightly negative. The reference gas should be flowing as in normal operation.
2 Refer to paragraph 3-10d of
this section for Manual (Semi­automatic) Calibration setup and procedure using the IFT.
3 Screw the push button regu-
lator with contents gage on to the test gas of choice and in­ject the test gas by opening the valve. Gas is on continu­ously when the valve is opened.
d. Manual (Semiautomatic) Calibration
Procedure
Consult Safety Data Sheet 1M03296 before performing any work on the CENELEC approved IFT 3000.
The following procedure relates to an op­erator initiated calibration selected at the IFT by pressing the CAL key. The calibra­tion is manually performed by the operator upon data queues from the IFT. Any system without an MPS 3000 Multiprobe Test Gas Sequencer must follow these steps.
CALIBRATE
IN PLACE
CONNECTION
CHECK
VALV E
PUSHBUTTON
REGULATOR
WITH CONTENTS
GAGE - SET 5 SCFH
NOTE: PROBE CALIBRATION GAS
FITTING HAS A SEAL CAP THAT MUST BE IN PLACE EXCEPT DURING CALIBRATION.
REFERENCE
AIR
CONNECTION
TEST HOSE CONNECTS TO CHECK VALV E
0.4 %
O
2
23800010
8.0 %
O
2
1. Press SETUP to display the SETUP menu. Select PROBE CALIBRATION sub-menu. Ensure that Auto Cal is dis­abled. Set the cursor on Auto Cal. Press ENTER. Set Auto Cal to NO if not already done.
2. Press the CAL key. Select PERFORM CALIBRATION sub-menu. "Press EN­TER to start Manual Calibration" will appear on the LCD display. Press EN­TER to start. Follow the data queues. Refer to Table 3-4. CALIBRATE O
2
Menu.
Figure 3-4. Typical Portable Test Calibration Setup
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-13
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
e. Fully Automatic Calibration
Consult Safety Data Sheet 1M03296 before performing any work on the CENELEC approved IFT 3000.
1. Test Gases for Fully Automatic Cali­bration. For fully automatic calibration, an MPS 3000 Multiprobe Test Gas Se­quencer is required as well as the two types of test gas.
Do not use 100% nitrogen as a zero gas. It is suggested that gas for the zero be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon con­centrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.
Two tanks of precision calibration gas mixtures. Recommended calibration gases are nominally 0.4 percent and
8.0 percent oxygen in nitrogen.
A typical automatic calibration system is shown in Figure 3-5.
2. Fully Automatic Calibration Setup. In order for the IFT system to calibrate automatically, the following parameters from the CALIBRATE sub-menu in the IFT have to be entered. Refer to Table 3-5. SETUP Sub-Menu.
(a) Auto Cal YES/NO
Set to YES
Set the desired time between cali­brations in number of days and hours. (1 year max.)
(d) Next Cal. XD XH
Displays the time left to the start of the next calibration. Set the de­sired time until the start of the next calibration. If nothing is entered here, the unit will automatically enter the cal intrvl and count down from that. (1 year max.)
(e) Gas Time 0:00
Set the amount of time for the test gasses to be turned on in number of minutes and seconds, allow enough time for signal values to stabilize.
(f) Purge Time 0:00
Set the amount of time for the gas lines to clear in number of minutes and seconds.
(g) Abort Time 0:00
Set the amount of time allowed between key functions before the calibration procedure is aborted in number of minutes and seconds.
(h) Res Alarm _____
Set the desired resistance alarm between 50 - 10000 ohms.
Once these parameters have been set, the system will initiate calibration with­out operator intervention as set by the CAL INTVL parameter.
(b) Output Tracks YES/NO
Set as desired to configure analog output tracking.
(c) Cal Intrvl XD XH
3-14 Setup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
3. Manually Initiated Fully Automatic Cali­bration Procedure. The following pro­cedure relates to an operator initiated calibration, either by a remote switch (CAL INIT on interconnect board) or selected at the IFT by pressing the CAL key using an MPS 3000 Multi­probe Gas Sequencer.
World Class 3000
3
Instruction Manual
106-300NFX Rev. 4.2
January 2002
(a) Press SETUP to display the
SETUP sub-menu. Select Calibra­tion. Ensure that Auto Cal is en­abled. Set the cursor on Auto Cal. Press ENTER. Set Auto Cal to YES if not already done.
CHECK
VALV E
TEST GAS
PROBE
(END VIEW)
REFERENCE AIR
HPS
PROBE SIGNAL CONNECTIONS
(b) Press the CAL key. Select Perform
Calibration. "Press ENTER to start Automatic Calibration" will appear on the LCD display. Press ENTER to start. Refer to Table 3-4. CALI­BRATE O
Sub-Menu.
2
IFT
MPS
INSTRUMENT
AIR IN
NOTE: 1. THE MPS CAN BE USED WITH UP
TO FOUR PROBES. ONLY ONE PROBE CAN BE CALIBRATED AT A TIME. PROBE CALIBRATIONS MUST BE SCHEDULED IN MULTIPLE PROBE APPLICATIONS.
2. HPS OPTION SHOWN.
MPS-IFT SIGNAL CONNECTIONS
TEST GAS 1
(HIGH O )
2
TEST GAS 2
(LOW O )
2
730014
Figure 3-5. Typical Automatic Calibration System
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-15
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
3-16 Setup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
Consult Safety Data Sheet 1M03296 before performing any work on the CENELEC approved IFT 3000.
4-1 OVERVIEW
Ensure that the oxygen analyzer, heater power supply (if included with the system), and intelli­gent field transmitter have been properly con­nected. It is important to check that grounding and screening of terminations are correctly made to prevent the introduction of ground loops. The IFT is equipped with noise suppres­sion circuitry on the power supply and signal in­put lines. Proper grounding at installation will ensure accuracy of function.
This section of the manual deals with operator controls and displays available for the IFT equipped with LDP and four membrane keys.
Operating instructions for the GUI equipped IFT are included in Section 3.
Any procedures not associated with normal op­erations are included in Section 2, Installation, or Section 5, Troubleshooting.
1
2
SECTION 4

LDP OPERATION

4-2 IFT WITH LDP FRONT PANEL CONTROLS
AND INDICATORS (Figure 4-1.)
Fig. 4-1
Index
No.
1
2 CAL Calibration in progress
3 TGH High test gas indicator light.
4 TGL Low test gas indicator light.
5 ENTER The ENTER key is used to
6
7
8SEL
Instruction Manual
106-300NFX Rev. 4.2
Control/
LED Description
LED
Display
Indicates current O gas value.
indicator light.
High test gas is being used in calibration process.
Low test gas is being used in calibration process.
initiate calibration or select a test gas parameter to change.
The increase key is used to increase the value of the test gas parameters.
The decrease key is used to decrease the value of the test gas parameters.
The select key is used to scroll through the list of parameters.
January 2002
or test
2
CAL TGH TGL
8
ENTERSEL
3
4
5

4-3 LDP DISPLAYS

Consult Safety Data Sheet 1M03296 before performing any work on the CENELEC approved IFT 3000.
7
6
The LDP equipped IFT contains four selectable displays. The displays are selected by pressing the SEL key. The display is advanced once each time the SEL key is depressed and con­tinually scrolls. The only parameters which the
34990019
Figure 4-1. IFT with LDP Front Panel
Rosemount Analytical Inc. A Division of Emerson Process Management LDP Operation 4-1
operator is permitted to change change are H (high test gas) and L (low test gas). When a pa­rameter is to be changed, press the ENTER
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
key. To change the parameter, the increase and decrease arrows are depressed until the proper value is displayed. Depress the ENTER key to accept the new value, or the SEL key to abort the change. The four displays are as follows:
a. O
2
The O2 selection will display the O2 value on the LED display when ENTER key is de­pressed.
b. H
The high test gas display allows the value of the high test gas parameter to be changed for calibration purposes.
c. L
The low test gas display allows the value of the low test gas parameter to be changed for calibration purposes.
d. CAL
tion. Calibration must be manually initiated. Information on test gases and hardware re­quirements may be found in paragraph 3-10, Calibration.
b. Manual Calibration
The following procedures relate to an op­erator initiated calibration. The calibration is manually performed by the operator upon data queues from the IFT. Any system with­out an MPS 3000 multiprobe test gas se­quencer must follow these steps.
Consult Safety Data Sheet 1M03296 before performing any work on the CENELEC approved IFT 3000.
1. Press the SEL key until CAL is shown on the LED display.
2. Press ENTER. The CAL light is now on. Turn on high test gas.
The calibration selection allows the operator to initiate the calibration process when EN­TER key is depressed.

4-4 LDP DEFAULTS

The LDP equipped IFT is programmed at the factory with the defaults indicated in Table 4-1. For a description of the defaults refer to Table 3-4. SETUP Sub-Menu.
4-5 CALIBRATION
a. Overview
The LDP equipped IFT is configured at the factory for manual (semiautomatic) calibra-
3. Press ENTER. The TGH light is now flashing.
4. When the value shown on the LED display has stabilized, press ENTER. Turn off high test gas.
5. Turn on low test gas. Press ENTER. The TGL light is now flashing.
6. When the value shown on the LED display has stabilized, press ENTER.
7. Turn off low test gas. Press ENTER.
8. All indicator lights are off. Calibration complete.
4-2 LDP Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
4
106-300NFX Rev. 4.2
World Class 3000
Table 4-1. LDP Defaults
PARAMETER DEFAULT PARAMETER DEFAULT
January 2002
Probe Calibration
High Gas Low Gas Auto Cal Output Tracks Cal Interval Next Cal Gas Time Purge Time Resistance Alarm
O2 Calculation
Slope Constant Htr Set Point
O2 Alarms
HI Alarm LO Alarm Alarm DB
8.0%
0.4% NO YES OFF Disabled 5:00 MIN 5:00 MIN 1000 ohms
____(value from calibration) ____(value from calibration) 736°C (when implemented)
30%
0.3%
0.0%
Efficiency Calc.
Enable Calc K1 Value K2 Value K3 Value
Relay Outputs (K1)
Event 1 Event 2 Event 3
Relay Outputs (2)
Event 1 Event 2 Event 3
Analog Output
Source Aout Type
NO
0.0
0.0
0.0
Heater Fail Cal. Fail INCAL
LO O
2
OFF OFF
O
0-10%
2
4-20mA
Rosemount Analytical Inc. A Division of Emerson Process Management LDP Operation 4-3
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
4-4 LDP Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5
Instruction Manual
106-300NFX Rev. 4.2
January 2002
SECTION 5

TROUBLESHOOTING

b. Electrical Noise
Consult Safety Data Sheet 1M03296 before performing any work on the CENELEC approved IFT 3000.
Install all protective equipment covers and safety ground leads after trouble­shooting. Failure to install covers and ground leads could result in serious injury or death.
5-1 OVERVIEW
The system troubleshooting describes how to identify and isolate faults which may develop in the Oxygen Analyzer System. Refer to Probe, IFT, HPS, and MPS Appendices.

5-2 SPECIAL TROUBLESHOOTING NOTES

a. Grounding
It is essential that adequate grounding pre­cautions are taken when the system is be­ing installed. A very thorough check must be made at both the probe and electronics to ensure that the grounding quality has not degraded during fault finding. The system provides facilities for 100% effective grounding and the total elimination of ground loops.
The IFT has been designed to operate in the type of environment normally found in a boiler room or control room. Noise suppres­sion circuits are employed on all field termi­nations and main inputs. When fault finding, the electrical noise being generated in the immediate circuitry of a faulty system should be evaluated. All cable shields must be connected to earth.
c. Loose Integrated Circuits
The IFT uses a microprocessor and sup­porting integrated circuits. Should the elec­tronics unit receive rough handling during installation in a location where it is sub­jected to severe vibration, an Integrated Circuit (IC) could work loose. The fault find­ing guides in paragraph 5-3a. and Table E-2, Appendix EX, show the resulting vari­ety of failure modes. It is recommended that all IC's be confirmed to be fully seated be­fore troubleshooting on the system begins.
d. Electrostatic Discharge
Electrostatic discharge can damage the IC's used in the electronics unit. It is essential before removing or handling the processor board or the IC's used on it, that the user ensure he/she is at ground potential.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-1
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
5-3 SYSTEM TROUBLESHOOTING
The IFT provides system failure information with two different error message formats. The error messages vary due to system configuration. Refer to Appendix EX, IFT 3000 Troubleshoot­ing.
a. GUI Equipped IFT
The status line of the GUI equipped IFT will display one of ten conditions: OK, CAL (calibration), C Err (calibration error), H Err (heater error), TGLow (test gas low), HiO (high O2 level), LoO2 (low O2 level), and R Hi (high resistance level), Off and PRBE. Refer to Table E-2, Appendix EX for addi­tional troubleshooting information on the GUI equipped IFT.
Consult Safety Data Sheet 1M03296 before performing any work on the CENELEC approved IFT 3000.
1. OK - The system is operating normally.
2. CAL - The system is currently undergoing calibration.
3. C Err - An error was detected during the calibration process.
4. H Err - There is a fault within the heater system.
5. TGLow - There is no test gas pressure.
2
6. HiO
- The O2 value is above the high
2
alarm limit.
7. LoO
- The O2 value is below the low
2
alarm limit.
8. R Hi - The cell resistance is above the high limit.
9. Off - The probe has been turned off because the IFT cannot control the heater temperature.
10. PRBE - The probe is disconnected, cold, or leads are reversed.
5-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
6
Instruction Manual
106-300NFX Rev. 4.2
January 2002
SECTION 6

RETURN OF MATERIAL

6-1 If factory repair of defective equipment is re-
quired, proceed as follows:
a. Secure a return authorization number from
a Rosemount Analytical Sales Office or Representative before returning the equip­ment. Equipment must be returned with complete identification in accordance with Rosemount instructions, or it will not be ac­cepted.
In no event will Rosemount be responsible for equipment returned without proper authorization and identification.
b. Carefully pack defective unit in a sturdy box
with sufficient shock absorbing material to insure that no additional damage will occur during shipping.
c. In a cover letter, describe completely:
1. The symptoms from which it was de­termined that the equipment is faulty.
2. The environment in which the equip­ment has been operating (housing, weather, vibration, dust, etc.).
5. Complete shipping instructions for re­turn of replacement or repaired equip­ment to you.
6. Reference the return authorization number.
d. Enclose a cover letter and purchase order
and ship the defective equipment, according to instructions provided in Rosemount Re­turn Authorization, prepaid, to:
Rosemount Analytical Inc. RMR Department 1201 N. Main Street Orrville, Ohio 44667
If warranty service is requested, the defec­tive unit will be carefully inspected and tested at the factory. If failure was due to conditions listed in the standard Rosemount warranty, the defective unit will be repaired or replaced at Rosemount's option, and an operating unit will be returned to the cus­tomer in accordance with shipping instruc­tions furnished in the cover letter.
3. The site from which equipment was removed.
4. Whether warranty or nonwarranty service is requested.
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 6-1
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by your purchase order and shipping instructions.
Instruction Manual
106-300NFX Rev. 4.2 January 2002
World Class 3000
6-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
7
106-300NFX Rev. 4.2
World Class 3000
January 2002
SECTION 7

APPENDICES

APPENDIX AX. WORLD CLASS 3000 OXYGEN ANALYZER PROBE (CENELEC
APPROVED VERSION)
APPENDIX BX. HPS 3000 HEATER POWER SUPPLY FIELD MODULE (CENELEC
APPROVED VERSION)
APPENDIX DX. MPS 3000 MULTIPROBE TEST GAS SEQUENCER
APPENDIX EX. IFT 3000 INTELLIGENT FIELD TRANSMITTER
APPENDIX JX. HART COMMUNICATOR MODEL 275D9E IFT 3000 APPLICATIONS
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 7-1
Instruction Manual
Appendix AX Rev. 2.1 February 1998
World Class 3000
NOTES:
1.
SEE SAFETY DATA SHEET 1M03226 FOR A LIST OF PROBE REPAIRS A CUSTOMER MAY MAKE.
1
2
2.
ITEM B , CALIBRATION GAS TUBE, FITS INTO HOLE A WHEN PROBE IS ASSEMBLED.
4
5
6
3
NOT ALL PARTS SHOWN ARE AVAILABLE
3. FOR PURCHASE SEPARATELY. FOR A LIST OF AVAILABLE PARTS, SEE TABLE A-3.
7
9
10
B
8
A
13
1. Snubber Diffuser
2. Retainer Screw
3. Calibration Tube Insert
4. Retainer
5. Cell and Flange Assembly
6. Corrugated Seal
7. Probe Tube and Junction Box Assembly
8. Tube Nut, Ferrule, and Tie-Rap
9. Hose Clamp
10. Label
11. Terminal Block
12. Label Attaching Hardware Mounting Plate
13. Gas Connection
14. Terminal Block Marker
15. Seal Cap
16. Terminal Block
17. V-Strut Assembly
16
18
19
20
18. Terminal Block Screw
19. Ground Wires
20. Junction Box Gasket
21. Hose
22. Junction Box Lid
APPENDIX AX
23. Probe Head Hardware
24. Cover Setscrew
17
15
22
14
21
23
12
11
E
N
E
E
L
I
R
H
G
I
W
S
E
N
D
E
P
O
T
G
O
N
N
I
N
O
E
D
P
O
S
E
E
R
E
O
L
F
A
E
B
B
E
L
24
19780003
Figure A-1. Oxygen Analyzer (Probe - CENELEC Approved) Exploded View
A-0 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
AX
WORLD CLASS 3000 OXYGEN ANALYZER PROBE
(CENELEC APPROVED VERSION)
Instruction Manual
Appendix AX Rev. 2.1
February 1998
APPENDIX AX, REV. 2.1
DESCRIPTION
a. Cell and Flange Assembly
Consult Safety Data Sheet 1M03226 for probe safety related information.
A-1 OXYGEN ANALYZER (PROBE) – GENERAL
The CENELEC approved Oxygen Analyzer (Probe), Figure A-1 consists of three component groups (Figure A-2): probe exterior, inner probe, and probe head. Specifications for the CENELEC approved probe are contained in Table A-1.
A-2 PROBE ASSEMBLY EXTERIOR
Primary probe exterior components include a flange-mounted zirconium oxide cell, mounted on a tube assembly and protected by a flame arrestor and snubber diffuser.
PROBE
EXTERIOR
The primary component in the cell and flange assembly, Figure A-3, is a yttria­stabilized zirconium oxide cell. It creates an electrical signal when theoxygen level on one side is out of balance with the oxygen level on the other side. This signal is pro­portional to the difference in oxygen levels.
b. Probe Tube Assembly
Screws and a retainer ring secure the cell and flange assembly, Figure A-3, to the probe tube assembly. When in place, the cell is inside the tube. The retainer ring is high temperature chrome plated to prevent galling and seizing. It is also coated with anti-seize compound to help avoid seizing.
PROBE
TUBE
CORRUGATED
SEAL
RETAINER
JUNCTION
BOX
E
N
E
E
L
I
R
H
G
I
W
S
E
N
PROBE
INTERIOR
D
E
P
O
T
G
O
N
N
I
N
O
E
D
P
O
S
E
E
R
E
O
L
F
A
E
B
B
E
L
CELL AND
FLANGE
ASSEMBLY
CALIBRATION
19780004
TUBE INSERT
730009
Figure A-2. Main Probe Components
Figure A-3. Cell and Tube Assemblies
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-1
Instruction Manual
Appendix AX Rev. 2.1 February 1998
World Class 3000
Table A-1. Specifications for Oxygen Analyzing Equipment
1, 2
Probe lengths, nominal ................................................... 457 mm (18 in.), 0,91 m (3 ft), 1,83 m (6 ft),
depending on duct dimensions.
Temperature limits in process measurement area ......... 10-704°C (50-1300°F)
Standard/current output .................................................. 4-20 mA dc signal (factory set)
O
indication (Digital display and analog output) ............ 0.1% O2 or ±3% of reading, whichever is
2
greater using Rosemount test gases.
Cell speed of response ................................................... 1 millisecond
System speed of response ............................................. less than 3 seconds (amplifier output)
Resolution sensitivity ...................................................... 0.01% O
transmitted signal
2
Probe reference air flow.................................................. 56,6 L/hr (2 scfh) clean, dry, instrument quality
air (20.95% O
)
2
Calibration gas mixtures ................................................. Rosemount Hagan Test Gas Kit Part No.
6296A27G01 contains 0.4% O
2N2
Nominal and 8% O2N2 Nominal
Calibration gas flow......................................................... 141,6 L/hr (5 scfh)
Probe heater power supply............................................. 44 Vac from HPS 3000
HPS 3000 power requirement ........................................ 225 VA
Ambient operating temperature of probe junction box.... 0°-150°C (32°-302°F)
HPS 3000 ambient operating temperature ..................... 0°-50°C (32°-120°F)
Approximate shipping weights:
457 mm (18 in.) package...................................... 30 kg (66 lbs)
0,91 m (3 ft) package............................................ 40 kg (88 lbs)
1,83 m (6 ft) package............................................ 55 kg (121 lbs)
1
All static performance characteristics are with operating variables constant.
2
Temperatures over 537°C (1000°F) may affect the ease of field cell replaceability.
The tube assembly includes a flange which mates with a stack-mounted flange. Studs on the stack flange make installation easy. There is also a tube to carry calibration gas from the probe head to the process side of the cell during calibration.
c. Flame Arrestor Diffuser
The flame arrestor diffuser, Figure A-4, protects the cell from heavy particles and isolates the cell from changes in tempera­ture. The assembly consists of a flame arre­stor and a snubber diffuser. The flame arrestor and diffuser thread onto the probe tube. Pin spanner wrenches (probe disas-
sembly kit 1L03825G01) are applied to holes in the diffusion element hub to re­move or install the diffuser assembly.
The Flame Arrestor and Flame Arre­stor Hub are among the critical com­ponents in this type of protection (Flameproof Enclosure Type ‘D'). See Safety Data Sheet 1M03226.
Systems that use an abrasive shield require a special flame arrestor and diffuser as­sembly with a hub that is grooved to accept two dust seal gaskets.
A-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
AX
Instruction Manual
Appendix AX Rev. 2.1
February 1998
PIN
WRENCH
SNUBBER
DIFFUSOR
19780006
Figure A-4. Flame Arrestor Diffuser Assembly
d. Cell – General
The components which make up the cell are machined to close tolerances and assem­bled with care to provide accurate oxygen measurements. Any replacement requires attention to detail and care in assembly to provide good results.
Failure to follow the instructions in this manual could cause danger to personnel and equipment. Read and follow instructions in this manual carefully.
The oxygen probe includes an inner elec­trode for the cell assembly. It consists of a platinum pad and a platinum/inconel com­posite wire which produces the cell constant offset voltage described in the Nernst equa­tion.
With this pad and wire, the constant will be between -10 to +15 mV. The cell constant is noted in the calibration data sheet supplied with each probe.
HEATER
REFERENCE
GAS TUBE
CERAMIC
SUPPORT TUBE
V-ST RU T
P00005A
Figure A-5. Inner Probe Assembly
A-3 INNER PROBE ASSEMBLY
The inner probe assembly, Figure A-5, consists of six main parts:
a. Ceramic support rod with four holes running
through the length. The holes serve as in­sulated paths for the cell signal wire and thermocouple wires.
b. A heater that is helically wrapped on a
quartz support cylinder and insulated.
c. A chromel-alumel thermocouple which acts
as the sensing element for the temperature controller. (Not visible in Figure A-4; located within ceramic support rod.)
d. A platinum screen pad which forms electri-
cal contact with the inner electrode of the electrochemical cell. (Not visible in Figure A-5; located at end of ceramic support rod.) The pad is attached to an inconel wire which carries the signal to the terminal strip.
e. A V-strut assembly to give support to the
inner probe assembly.
Every probe should be calibrated and checked after repair or replacement of cell, pad and wire, heater, or thermocouple, or after disassembly of the probe.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-3
f. A tube to carry reference gas to the cell.
Turn to Maintenance and Service for repair pro­cedures for probe components.
Instruction Manual
Appendix AX Rev. 2.1 February 1998
World Class 3000
TERMINAL
STRIP
CALIBRATION
GAS FITTING
VENT
REFERENCE
GAS FITTING
Figure A-6. Junction Box
A-4 JUNCTION BOX
The junction box, Figure A-6, is positioned at the external end of the probe and contains a terminal strip for electrical connections and fit­tings for reference and calibration gases. Fit­tings are for 0.250 inch stainless steel tubing. The calibration fitting has a seal cap which must remain in place except during calibration. A tubing fitting is also supplied to be used with the calibration gas supply during calibration.
If the calibration gas bottles will be permanently hooked up to the probe, an optional check valve is recommended to prevent condensation of flue gas in the calibration lines.
During operation and calibration, reference gas is supplied through the reference gas fitting to the reference side of the cell. This gives the system a known quantity of oxygen with which to compare the oxygen level in the process gas. Unlike the non-CENELEC World Class 3000 Probe, reference air must be used in conjunc­tion with the CENELEC World Class 3000 probe.
P00006A
During calibration, two gases of different known oxygen concentrations are injected one at a time through the calibration gas fitting. Stainless steel tubing delivers this gas to the process side of the cell. In a healthy cell, the difference in oxygen pressure from the process side to the reference side of the cell will cause a millivolt output proportional to the difference in oxygen levels. The electronics unit can use the two millivolt outputs caused by the two calibration gases for either automatic or semi-automatic calibration.
Do not attempt to remove a process gas sample through either gas fitting. Hot gases from the process would damage gas hoses in the probe head.
A-5 CABLE ASSEMBLY
Cable used to interconnect apparatus must conform to the applicable codes of practice in the country of installation (example: BS4345 in Great Britain). Rosemount can supply a cable (P/N 1U03066) which is fitted with EExd IIC barrier glands. The installer should note that on some earlier versions of the cable, the glands, although EExd IIC approved, are not of the bar­rier gland variety. These glands should not be used. Rosemount can supply a gland kit for the corresponding barrier gland (P/N 1U03066G07). Each kit contains one pair of glands. The Rosemount supplied cable is a 7 conductor cable to connect the probe to the HPS 3000, and to connect the HPS 3000 to the electronics package. Standard length for this cable is 6 m (20 ft), but lengths up to 45 m (150 ft) are avail­able. The 7 conductors include 1 shielded pair of wires for the cell millivolt signal, 1 shielded pair of type K wires for the thermocouple, and 3 individual 16-gauge wires for the heater and for ground. All metal shields are isolated at the probe end and connect by drain wires to ground at the electronics. The cable is suitable for use in ambient temperatures up to 90°C (194°F).
A-4 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
AX
Instruction Manual
Appendix AX Rev. 2.1
February 1998
BC A
GROUP NOTE
PARTS LIST
UNITS: INCHES
PARTS LIST
G03
1
G02
1
G01
GROUP
1
MAT'L CODE
PART NUMBER
OR REF DWG
1U05680G20
) PROBE 18"
SIZE-REFERENCE
INFORMATION
DEFINER
DWG
PART NAME
CENELEC
01
ITEM
NOTE
1
1
4507C26G07
4507C26G08
)3FT
)6FT
DWG
DWG
WORLD CLASS 3000
GAS PICK-UP
GAS PICK-UP
02
03
1
1
1
1
1
1
1
1
1
1
4507C26G09
1M03241H01
4507C51H03
1L03826G02
)9FT
DWG
DWG
DWG
DWG
GAS PICK-UP
ANALYZER HOUSING
MTG HARDWARE
ASSY HARDWARE
04
05
06
07
GAS TUBE PICK-UP
GROUP NOTE
3FT
A
GAS TUBE PICK-UP
6FT
B
CAL
GAS
ELECT
CABLE
REF
GAS
100
(3.93)
OF
FLOW
DIRECTION
VIEW ‘A-A'
REFERENCE AIR SUPPLY
CONNECTION BITE TYPE
FITTING (PARKER CPI) FOR
1.
NOTES:
0.250 O.D. TUBING. 2 SCFH
AT 3 PSIG MAX. CLEAN DRY
A
A
AIR REQUIRED. FITTING IS
LOCATED ON FAR SIDE.
CABLE
693.67
CALIBRATION AND PURGE
GAS CONNECTION. BITE
TYPE FITTING (PARKER CPI)
10 SCFH AT 32 PSIG MAX.
CALIBRATION GAS REQUIRED.
2.
(27.31)
REQUIRED
INSERTION
CLEARANCE
FOR PROBE
oo
LAG TO ENSURE GAS
TEMPERATURE DOES NOT
GO BELOW DEW POINT
3.
FLANGE
AND
REMOVAL
oo o o
WELD BY
CUSTOMER
RECOMMENDED 2 IN. THK
INSULATION. THERMAL CONDUCTIVITY
K EQUALS 0.5 FOR INSULATION.
DIMENSIONS ARE IN MILLIMETERS WITH INCHES
INSTALL WITH ANALYZER IN A VERTICALLY
OR EXCEED 500 C (932 F).
4.
GASKET
DOWNWARD DIRECTION ONLY.
AND
FLUE GAS OPERATING TEMPERATURE
RANGE 650 - 980 C (1200 - 1800 F).
5.
HARDWARE
6.
IN PARENTHESES UNLESS OTHERWISE INDICATED.
7.
GASKET
AND HARDWARE
584.0
Figure A-7. Bypass Probe Option
(23.0)
DRAIN
165.10 (6.50) REF
940 (37) REF ON 3 FT (4507C26G07)
1850 (73) REF ON 6 FT (4507C26G08)
1727.2 (68) REF ON 9 FT PICK-UP (4507C26G09)
825.5 (32.5) REF ON 3 FT AND6FTPICK-UP (4507C26G07 AND G08)
19780009
2770 (109) REF ON 9 FT (4507C26G09)
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-5
Instruction Manual
Appendix AX Rev. 2.1 February 1998
World Class 3000
A-6 PROBE OPTIONS
a. Abrasive Shield Assembly
The abrasive shield assembly (IB-106­300NX Series, Figure 2-1) is a stainless­steel tube that surrounds the probe assem­bly. The shield protects the probe against particle abrasion and corrosive condensa­tions, provides a guide for ease of insertion, and acts as a probe position support, espe­cially for longer length probes. The abrasive shield assembly uses a modified flame ar­restor and diffuser filled with dual dust seal packing.
b. Bypass Probe Options
For processes where the flue gas exceeds the maximum allowable temperature of
704°C (1300°F) a bypass sensor package can be employed. The bypass system uses an 18 inch probe mounted externally on the stack or duct. The process or exhaust gases are directed out to the probe through an extension/return duct. The bypass ar­rangement does not require the use of aspi­ration air and the gas which flows past the probe is returned to the stack or duct.
The bypass probe package is normally used for process temperatures of 704°C (1300°F) to 980°C (1800°F). "Inconel 600" has an operating range up to 980°C (1800°F).
Overall dimensions and mounting details of the bypass system are shown in Figure A-7.
TROUBLESHOOTING
Before conducting any work on the probe, consult probe Safety Data Sheet 1M03226.
Install all protective equipment covers and safety ground leads after trouble­shooting. Failure to replace covers and ground leads could result in seri­ous injury or death.
A-7 OVERVIEW
The probe troubleshooting section describes how to identify and isolate operating faults which may develop in the probe assembly.
A-8 PROBE TROUBLESHOOTING
a. Probe Faults
The three symptoms of probe failure are:
1. The system does not respond to changes in the oxygen concentration.
2. The system responds to oxygen changes but does not give the correct indication.
3. The system does not give an accept­able indication of the value of the oxy­gen test gas being applied during calibration.
b. Table A-2 provides a guide to fault finding
for the above symptoms.
c. Figure A-8 and Figure A-9 provide an alter-
nate approach to finding probe related problems.
A-6 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
AX
SYMPTOM CHECK FAULT REMEDY
1. No response to oxygen concentration change when:
Instruction Manual
Appendix AX Rev. 2.1
February 1998
Table A-2. Fault Finding
Heater is cold and TC mV output is less than setpoint.
Heater is hot and T/C mV output is at setpoint ±0.2 mV.
2. System responds to oxygen concentration changes but does not give correct indication.
Thermocouple continuity Thermocouple failure Replace thermocouple
or return probe to Rosemount.
Heater cold resistance to be 11 ohm - 14 ohm
Triac O/P to heater Failure of electronics Check HPS and
Recorder chart Recorder failure See Recorder Instruc-
Cell mV input to electron­ics and cell mV at probe junction box
Heater failure Replace heater or return
probe to Rosemount.
electronics package.
tion Manual.
No cell mV at probe when test gas applied
Probe cell mV OK but no input to electronics
Cell mV satisfactory both at probe junction box and input to electronics ­failure of electronics
Replace cell or return probe to Rosemount.
Check out cable connection.
Check electronics package.
Good response, with incorrect indication.
3. Probe does not give accurate indication of applied test gas.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-7
Recorder or remote indicator
System calibration Calibration error Recalibrate system.
Probe mounting and condition of duct
Cell mV input to electronics
Test gas input port Blocked port Clean port.
Ceramic diffusion element Diffusion element cracked,
Calibration error Recalibrate recorder or
indicator, reference Recorder Instruction Manual.
Air ingress into duct Stop air leaks or resite
probe.
Failure of electronics Check electronics
package.
Replace diffusion
broken, or missing
element.
Instruction Manual
Appendix AX Rev. 2.1 February 1998
CHECK CELL MV TERMINALS 1 & 2 IN PROBE JUNCTION BOX
DOES CELL MV CORRESPOND WITH EXPECTED O CONCENTRATION
World Class 3000
2
YES
MEANS LOW O OR COMBUSTIBLES IS GAS STREAM. RECONFIRM THE O
CONCENTRATION
2
CHECK ELECTRONICS AND CELL MV I/P TO ELECTRONICS.
CHECK CABLE
PULL PROBE AND CHECK FOR CRACKED DIFFUSION ELEMENT OR CRACKED CELL
?
NO - HIGH
2
OK
NO - ZERO OR VERY LOW
?
LOW
CHECK CELL RESISTANCE
?
ZERO
TOO HIGH
REPLACE CELL
CHECK THE CRIMP WHICH CONNECTS THE INCONEL WIRE TO ITS EXTENSION WIRE IN THE PROBE JUNCTION BOX.
BAD
?
GOOD
CONTACT
CHECK THAT PLATINUM WIRE IS MAKING CONTACT WITH CELL. IF THE PAD ASSEMBLY CAN BE PUSHED FORWARD THEN THERE WAS POOR CONTACT
CONTACT
REPLACE CRIMP
BAD
?
GOOD
CONTACT
REPLACE CELL
CONTACT
DISASSEMBLE AND CHECK FOR BROKEN CERAMICS RODS, SPRING, OR OTHER RESTRICTIONS.
P00009B
Figure A-8. Flowchart of Probe Related Problems, #1
A-8 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
AX
CHECK THERMOCOUPLE MV AT TERMINALS 3 AND 4 IN PROBE JUNCTION BOX.
IS IT ABOUT SETPOINT MV ±0,2 MV? [THE MV MEASURED IS NOT YET COLD. JUNCTION COMPENSATED, SO IT WILL VARY BY -0,022 MV PER DEGREE FAHRENHEIT RISE IN THE TEMPERATURE OF THE ELECTRONICS BOX LOCATION. NORMAL IS SETPOINT MV AT 25°C (77 ).]°F
Instruction Manual
Appendix AX Rev. 2.1
February 1998
ALLOW ADDITIONAL TIME
1.
FOR TEMP CONTROL TO SETTLE.
2.
COLD JUNCTION TEMP AT ELECTRONICS BOX MAY BE MUCH LOWER THAN 25°C (77°F).
?
NO
IS IT HIGHER, LOWER, OR ZERO?
?
ZERO
YES
LOWERHIGHER
HEATER, THERMOCOUPLE AND CONTROL ARE SATISFACTORY. REFER TO HEATER POWER SUPPLY TROUBLESHOOTING.
HEATER
THERMOCOUPLE
CIRCUIT
OPEN?
IS
NO
HEATER MAY BE DEAD. THERMOCOUPLE
YES
MAY BE READIN STACK TEMPERATURE.
1.
CHECK HEATER FUSE AND TRIAC OUTPUT TO HEATER.
3.
CONFIRM MV VALUE ON ELECTRONICS UNIT. IF OK, RESOLVE DIFFERENCE BETWEEN VOLTMETER AND ELECTRONICS.
REPLACE CONTACT
AND THERMOCOUPLE
ASSEMBLY
2.
CHECK HEATER CONDITION. TURN OFF POWER. RESISTANCE BETWEEN TERMINALS 7 AND 8 SHOULD BE 11-14 OHMS. RESISTANCE BETWEEN TERMINALS 6 AND 7 AND 6 AND 8 SHOULD BE OPEN CIRCUIT. IF FAILED, REPLACE STRUT.
P00009A
Figure A-9. Flowchart of Probe Related Problems, #2
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-9
Instruction Manual
Appendix AX Rev. 2.1 February 1998
World Class 3000
MAINTENANCE AND SERVICE
NOTE
!
Upon completing installation, make sure that the probe is turned on and operating prior to firing up the combustion process. Damage can result from having a cold probe exposed to the process gases.
During outages, and if possible, leave all probes running to prevent condensation and prema­ture aging from thermal cycling.
If the ducts will be washed down during outage, MAKE SURE to power down the probes and remove them from the wash area.
A-10 PROBE RECALIBRATION
Before carrying out any service or maintenance on the probe, consult Safety Data Sheet 1M03226.
When working on this equipment on the laboratory bench, be aware that the probe, probe tube, and flame arre­stor hub can be hot [up to 370°C (698°F)] in the region of the probe heater.
Install all protective equipment covers and safety ground leads after equip­ment repair or service. Failure to in­stall covers and ground leads could result in serious injury or death.
A-9 OVERVIEW
This section describes routine maintenance of the oxygen analyzer probe. Spare parts referred to are available from Rosemount. Probe disas­sembly kit 1LO3825G01 contains the required spanner and hex wrenches. Refer to Replace­ment Parts of this appendix for part numbers and ordering information.
The oxygen analyzer system should be cali­brated when commissioned. Under normal circumstances the probe will not require fre­quent calibration. When calibration is required, follow the procedure described in the Instruction Bulletin applicable to your electronics package.
A-11 CELL REPLACEMENT
This paragraph covers oxygen sensing cell replacement. Do not attempt to replace the cell until all other possibilities for poor per­formance have been considered. If cell re­placement is needed, order cell replacement kit, Table A-3.
The cell replacement kit contains a cell and flange assembly, corrugated seal, calibra­tion tube insert, setscrews, socket head cap screws, and anti-seize compound. Items are carefully packaged to preserve precise surface finishes. Do not remove items from packaging until they are ready to be used. Spanner wrenches and hex wrenches needed for this procedure are part of an available special tools kit, Table A-3.
A-10 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
AX
Before carrying out this procedure, consult Safety Data Sheet 1M03226.
Wear heat resistant gloves and cloth­ing to remove probe from stack. Normal operating temperatures of diffuser and vee deflector are approximately 316° to 427°C (600° to 800°F). They could cause severe burns.
Do not remove cell unless it is certain that replacement is needed. Removal may damage cell and platinum pad. Go through complete troubleshooting procedure to make sure cell needs replacement before removing it.
Instruction Manual
Appendix AX Rev. 2.1
February 1998
a. Follow the power down procedure outlined
in Safety Data Sheet 1M03226 and the offi­cial "Codes of Practice" for your country of installation. Shut off and disconnect refer­ence gas and calibration gas supplies from probe junction box, Figure A-10. Wearing heat resistant gloves and clothing, remove probe assembly from stack carefully and allow to cool to room temperature. Do not attempt to work on unit until it has cooled to a comfortable working temperature.
b. The vee deflector is an option used with the
ceramic diffuser. Figure A-11 shows a probe with a snubber diffuser. If the probe has the vee deflector, remove vee deflector and hub setscrews and remove vee deflector. Use spanner wrenches from probe disassembly tools kit, Table A-3, to turn hub free from probe tube. If applicable, inspect optional ceramic diffusion element. If damaged, re­place element.
CELL EXTENSION
WIRE (ORANGE)
THERMOCOUPLE WIRE (+)
(YELLOW CHROMEL)
GROUND WIRE
(GREEN)
EXTERNAL
EARTHING
HARDWARE
CALIBRATION
GAS FITTING
CABLE
REFERENCE GAS FITTING
THERMOCOUPLE WIRE (-)
(RED ALUMEL)
GROUND WIRE (GREEN)
HEATER WIRES (BLACK)
INTERNAL EARTHING TERMINAL (TERMINAL 6)
P00010A
Figure A-10. Cell Wiring Installation
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-11
Instruction Manual
Appendix AX Rev. 2.1 February 1998
World Class 3000
c. Loosen 4 socket head screws from retainer.
Pull off retainer and calibration tube in­sert, Figure A-12. Discard calibration tube insert.
d. Press the cell flange against the probe end
flange and twist through 90° in both the clockwise and counterclockwise directions. This breaks the bond between the cell and the platinum pad. Once the bond is broken, remove the cell.
e. View the platinum pad through the probe
end flange. Reform the platinum pad into a shape to match the cell electrode by gath­ering in any loose strands which may have become untidy. This may be done with a small flat screwdriver.
f. Remove and discard corrugated seal. Clean
mating faces of probe tube and retainer. Remove burrs and raised surfaces with block of wood and crocus cloth.
g. Rub one or two drops of anti-seize com-
pound #3535B53G01 between fingers. Smear on both sides of corrugated seal.
h. Install new calibration tube insert in retainer,
with the short arm of the tube insert pene­trating the calibration gas passage in the outer ring of the retainer. Assemble retainer, cell and flange assembly, corrugated seal,
and probe tube. Make sure the calibration tube insert lines up with the calibration gas passage in each component. Firmly seat the calibration tube insert in the assembly. Apply a small amount of anti-seize com­pound to screw threads and use screws to secure assembly. Torque to 4 N·m (55 in-lbs).
i. To install snubber diffuser, apply anti-seize
compound to threads of probe tube, Figure A-11, and snubber diffuser. Reinstall dif­fuser on probe tube. Using pin spanner wrenches, torque to 14 N·m (10 ft-lbs).
j. If using the optional vee deflector and ce-
ramic hub, apply anti-seize compound to threads of probe end flange, hub, and setscrews. Reinstall hub on probe tube. Using pin spanner wrenches, torque to 14 N·m (10 ft-lbs). Reinstall vee deflector, orienting apex toward gas flow. Secure with setscrews and anti-seize compound. Torque to 2.8 N·m (25-in-lbs). Secure hub retaining setscrew.
k. On systems equipped with an abrasive
shield, install dust seal gaskets, with joints 180° apart.
PROBE TUBE
(NOT INCLUDED KIT)
PROBE
TUBE
SNUBBER
DIFFUSOR
Figure A-11. Removal of Snubber Diffuser
A-12 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
PIN SPANNER
WRENCH
PROBE
END
FLANGE
CALIBRATION
TUBE PASSAGE
19780017
Figure A-12. Cell Replacement Kit
CORRUGATED
SEAL
FLANGE ASSEMBLY
CELL END
RETAINER
SOCKET
HEAD CAP
SCREW
CALIBRATION TUBE INSERT
730010
World Class 3000
AX
l. Reinstall probe and gasket on stack flange.
If there is an abrasive shield in the stack, make sure dust seal gaskets are in place as they enter 15° reducing cone.
m. Consult Safety Data Sheet 1M03226 then
turn power on to electronics and monitor thermocouple output. It should stabilize at
29.3 mV ±0.2 mV. Set reference air flow at
56.6 L/hr (2 scfh). After probe stabilizes, calibrate probe per Instruction Bulletin ap­plicable to your electronics package. If new components have been installed, repeat calibration after 24 hours of operation.
Instruction Manual
Appendix AX Rev. 2.1
February 1998
Wear heat resistant gloves and cloth­ing to remove probe from stack. Normal operating temperatures of diffuser and vee deflector are approximately 316° to 427°C (600° to 800°F). They can cause severe burns.
Before carrying out this procedure, consult Safety Data Sheet 1M03226.
A-12 CERAMIC DIFFUSION ELEMENT
REPLACEMENT
NOTE
This refers to ceramic diffuser element only.
a. General
The diffusion element protects the cell from particles in process gases. It does not nor­mally need to be replaced because the vee deflector protects it from particulate erosion. In severe environments, the filter may be broken or subject to excessive erosion. Ex­amine the ceramic diffusion element when­ever removing the probe for any purpose. Replace if damaged.
Damage to the ceramic diffusion element may become apparent during calibration. Compare probe response with previous re­sponse. A broken diffusion element will cause a slower response to calibration gas.
Hex wrenches needed to remove setscrews and socket head screws in the following procedure are available as part of a Probe Disassembly Kit, Table A-3.
Do not remove cell unless it is certain that replacement is necessary. Cell cannot be removed for inspection without damaging it.
b. Replacement Procedure
1. Follow the power down procedure out­lined in Safety Data Sheet 1M03226 and the official "Codes of Practice" for your country of installation. Disconnect cable conductors and remove cable, Figure A-10. Shut off and disconnect reference gas and calibration gas sup­plies from probe junction box. Wearing heat resistant gloves and clothing, carefully remove probe assembly from stack.
2. Loosen setscrews, Figure A-13, using hex wrench from Probe Disassembly Kit, Table A-3 and remove vee deflec­tor. Inspect setscrews. If damaged, re­place with stainless setscrews coated with anti-seize compound.
3. On systems equipped with abrasive shield, remove dual dust seal gaskets.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-13
Instruction Manual
Appendix AX Rev. 2.1 February 1998
World Class 3000
4. Use spanner wrenches from Probe Disassembly Kit, Table A-3, to turn hub free from retainer.
5. Put hub in vise. Break out old ceramic diffusion element with chisel along ce­ment line and 9.5 mm (3/8 in.) pin punch through cement port.
6. Break out remaining ceramic diffusion element by tapping lightly around hub with hammer. Clean grooves with pointed tool if necessary.
7. Replace ceramic diffusion element, using replacement kit in Table A-3. This consists of a diffusion element, cement, setscrews, anti-seize com­pound and instructions.
8. Test fit replacement ceramic diffusion element to be sure seat is clean.
Do not get cement on ceramic diffu­sion element except where it touches the hub. Any cement on ceramic diffu­sion element blocks airflow through element. Wiping wet cement off of ce­ramic only forces cement into pores. Also do not get any cement onto the flame arrestor element.
9. Thoroughly mix cement and insert tip of squeeze bottle into cement port. Tilt bottle and squeeze while simultane­ously turning ceramic diffusion element into seat. Do not get any cement on upper part of ceramic diffusion ele­ment. Ensure complete penetration of cement around 3 grooves in hub. Ce­ment should extrude from opposite hole. Wipe excess material back into holes and wipe top fillet of cement to form a uniform fillet. (A Q-Tip is useful for this.) Clean any excess cement from hub with water.
PIN
WRENCH
RETAINER
OPTIONAL CERAMIC
DIFFUSION ELEMENT
SETSCREW
HUB
CEMENT
PORT
CEMENT
FILLET
VEE
DEFLECTOR
21240026
Figure A-13. Removal of Optional Ceramic
Diffuser and Vee Deflector
10. Allow filter to dry at room temperature overnight or 1 to 2 hours at 93°C (200°F).
11. Wipe a heavy layer of anti-seize com­pound onto the threads and mating surfaces of the flame arrestor, diffusion hub, and probe tube.
12. Assemble flame arrestor and diffusion hub with two pin spanner wrenches. Torque to 14 N·m (10 ft-lbs). Secure with hub retaining setscrew.
13. On systems equipped with abrasive shield, install dust seal gaskets with joints 180° apart.
14. Reinstall vee deflector, orienting apex toward gas flow. Apply anti-seize com­pound to setscrews and tighten with hex wrench.
15. Reinstall probe on stack flange.
A-14 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
AX
Instruction Manual
Appendix AX Rev. 2.1
February 1998
16. Consult Safety Data Sheet 1M03226 then turn power on to electronics and monitor thermocouple output. It should stabilize at 29.3 mV ±0.2 mV. Calibrate probe per Instruction Bulletin applica­ble to your electronics package.
A-13 REPLACEMENT OF CONTACT AND
THERMOCOUPLE ASSEMBLY
Use heat resistant gloves and clothing when removing probe junction box and inner probe assembly. Do not at­tempt to work on these components until they have cooled to room tem­perature. Probe components can be as hot as 800°C (1500°F). This can cause severe burns.
Before carrying this procedure, con­sult Safety Data Sheet 1M03226.
a. Follow the cell removal procedure in para-
graph A-3, steps a, b, c, and d. Using heat resistant gloves and clothing, remove cover setscrew (24, Figure A-1) and junction box lid (23). Squeezing tabs on hose clamps, remove hoses from inner probe assembly, Figure A-14. Remove 4 screws which se­cure the inner probe assembly to the junc­tion box. Disconnect all inner probe assembly wires from the terminal block (Figure A-10). Pull inner probe assembly free from junction box. Set on bench and allow to cool to room temperature.
HOSE
CLAMP
PROBE
JUNCTION BOX
HOSE
BACKPLATE
MOUNTING
SCREW
P00013A
Figure A-14 Probe Junction Box Mechanical
Connections
HEATER
SCREWS
STRUT
CONTACT AND
THERMOCOUPLE
ASSEMBLY
GUIDE
LOOPS
SPRING
CLIP
b. Use a pencil to mark locations of spring
clips on ceramic rod, Figure A-15.
c. Pry or squeeze tabs on spring clips, and
pull contact and thermocouple assembly out
Figure A-15. Inner Probe Replacement (Heater,
V-Strut, and Backplate Assembly)
BACKPLATE
MOUNTING
HARDWARE
P00014A
of probe assembly. Retain spring clips and spring; replace if damaged.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-15
Instruction Manual
Appendix AX Rev. 2.1 February 1998
Be very careful when handling contact
and thermocouple assembly. The ce-
ramic rod in this assembly is fragile.
d. While very carefully handling new contact
and thermocouple assembly, lay old as­sembly next to new one. Transfer pencil marks to new rod. Throw away old contact and thermocouple assembly.
e. Carefully guide new contact and thermo-
couple assembly through V-strut assembly leaf spring (3, Figure A-16), spring (8), spring clips (9) (held open by squeezing tabs), tube supports and heater support un­til spring clip reaches pencil mark.
World Class 3000
Do not bend wires closer than 6.4 mm (1/4 in.) from end of ceramic rod. Dress wires so they do not touch sides of probe junction box.
h. Reinstall the cell per paragraph A-11, steps
f, g, h, i, j, and k.
i. Consult Safety Data Sheet 1M03226 then
power up system. Monitor thermocouple output. It should stabilize at setpoint mV ±0.2 mV. Recalibrate probe per Instruction Bulletin applicable to your electronics pack­age.
A-14 REPLACEMENT OF HEATER, V-STRUT
AND BACKPLATE ASSEMBLY (Inner Probe Assembly; Includes Contact and Thermocouple Assembly)
f. Slide assembled inner probe assembly into
junction box and probe tube. To align cali­bration gas tube with corresponding hole in backplate (A, B, Figure A-1) insert scriber through hole in backplate and into calibra­tion gas tube. Secure with screws. Reinstall hoses.
NOTE
The letter "A" is stamped inside the junction box close to the calibration tube. The calibration tube may fall over towards the opposite side. If it has, bring it back towards the letter "A" before inserting the inner probe assembly. Also note that there are grooves cut in the heater plates that fit over the calibration gas tube. Insert inner probe assembly gently to avoid kinking the calibration gas tube.
g. Connect color-coded wires to proper termi-
nals as shown in Figure A-10. Rosemount recommends connecting the thermocouple wires directly to the terminal strip. This is because the junction of different metals at the wires and lugs and at the lugs and the terminal strip could act as additional ther­mocouple junctions. This could produce a voltage that would affect the thermocouple output signal.
Figure A-17 is a cross-sectional view of the CENELEC approved oxygen analyzer (probe). Use Figure A-17 and the following procedure to replace heater, v-strut, and backplate assembly.
Before carrying out this procedure, consult Safety Data Sheet 1M03226.
Use heat resistant gloves and clothing when removing probe junction box and inner probe assembly. Do not at­tempt to work on these components until they have cooled to room tem­perature. Probe components can be as hot as 800°C (1500°F). This can cause severe burns.
a. Follow the cell removal procedure in para-
graph A-3, steps a, b, c, and d. Using heat resistant gloves and clothing, remove cover setscrew (24, Figure A-1) and junction box lid (23). Squeezing tabs on hose clamps, remove hoses from inner probe assembly, Figure A-13. Remove 4 screws which se­cure the inner probe assembly to the junc­tion box. Disconnect all inner probe assembly wires from the terminal block (Figure A-10). Pull inner probe assembly free from junction box. Set on bench and allow to cool to room temperature.
A-16 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
AX
Instruction Manual
Appendix AX Rev. 2.1
February 1998
12
1. Heater Ceramic Rod
2. Contact and Thermocouple Assembly
3. Leaf Spring
Figure A-16. Heater, Strut, and Backplate Assembly (Inner Probe Assembly)
Not all parts shown are available for purchase separately. For a list of available parts, see Table A-3.
1
36
3
5
POOO15A
4. Extension Wire
5. Ring Lug
6. Connector
7. Backplate
8
91011
4
73
6
8. Spring
9. Spring Clip
10. Strut
11. Heater
NOTE
5
89
74
211
1. Snubber Diffuser
2. Calibration Tube Insert
3. Socket Head Cap Screw
0.25 In.-28 x 0.63 (16 mm)
4. Cell and Flange Assembly
13
5. Corrugated Seal
6. Probe Tube Assembly
7. Gasket
8. Cover Head Assembly
9. Hose Clamp
10
12
19780022
10. Heater Strut
11. Retainer
12. Junction Box Setscrew
13. Flame Arrestor Hub Setscrew
Figure A-17. Oxygen Analyzer (Probe), Cross-Sectional View
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-17
Instruction Manual
Appendix AX Rev. 2.1 February 1998
World Class 3000
b. Slide new assembled inner probe assembly
into junction box and probe tube. To align calibration gas tube with corresponding hole in backplate (A, B, Figure A-1), insert scriber through hole in backplate and into calibration gas tube. Secure with screws. Reinstall hoses.
NOTE
The letter "A" is stamped inside the junction box close to the calibration tube. The calibration tube may fall over towards the opposite side. If it has, bring it back towards the letter "A" before inserting the inner probe assembly. Also note that there are grooves cut in the heater plates that fit over the calibration gas tube. Insert inner probe assembly gently to avoid kinking the calibration gas tube.
c. Connect color-coded wires to proper termi-
nals as shown in Figure A-10. Rosemount
recommends connecting the thermocouple wires directly to the terminal strip. This is because the junction of different metals at the wires and lugs and at the lugs and the terminal strip could act as additional ther­mocouple junctions. This could produce a voltage that would affect the thermocouple output signal.
Do not bend wires closer than 6.4 mm (1/4 in.) from end of ceramic rod. Dress wires so they do not touch sides of probe junction box.
d. Reinstall the cell per paragraph A-3, steps
f, g, h, i, j, and k.
e. Consult Safety Data Sheet 1M03226, then
power up system. Monitor thermocouple output. It should stabilize at setpoint mV ±0.2 mV. Recalibrate probe per Instruction Bulletin applicable to your electronics package.
A-18 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
AX
Instruction Manual
Appendix AX Rev. 2.1
February 1998
REPLACEMENT PARTS
Table A-3. Replacement Parts for Probe
FIGURE and
INDEX No.
A-1, 13 3D39149G06 A-1, 13 3D39149G07 A-1, 13 3D39149G08 A-16, 2 3534B56G04 A-16, 2 3534B56G05 A-16, 2 3534B56G06
PART
NUMBER DESCRIPTION
1
1
1
2
2
2
1U03066G07V-Strut Assembly (18 in.) V-Strut Assembly (3 ft) V-Strut Assembly (6 ft) Contact and Thermocouple Assembly (18 in.) Contact and Thermocouple Assembly (3 ft)
Contact and Thermocouple Assembly (6 ft) 4847B61G19 Cell Replacement Kit (18 in.) 4847B61G20 Cell Replacement Kit (3 ft) 4847B61G21 Cell Replacement Kit (6ft) 4847B61G24 Cell Replacement Kit (No Inconel and Platinum Pad Assembly) 1U05677G01 Probe Disassembly Kit
A-11 1U05677G04 F/A Diffuser Hub Assembly (Snubber Diffuser)
1U05677G06 F/A Diffuser Hub Assembly (For use with Abrasive Shield) 6292A74G02 Ceramic Diffusion Element Replacement Kit 1N04966H02 Abrasive Shield Assembly (3 ft) 1N04966H03 Abrasive Shield Assembly (6 ft) 1M03241H01 90° Elbow for Bypass
A-7, 2 4507C26G07 Bypass Gas Pickup Tube (3 ft) A-7, 3 4507C26G08 Bypass Gas Pickup Tube (6 ft) A-7, 4 4507C26G09 Bypass Gas Pickup Tube (9 ft)
263C152G01 Reference Gas Air Set
771B635H01 Calibration Gas Rotameter
1L03650H01 F/A Diffusion Hub Setscrew IB-106-300NFX Instruction Bulletin (IFT) IB-106-300NCX Instruction Bulletin (CRE) IB-106-300NEX Instruction Bulletin (Upgrade - CENELEC Digital Electronics)
1U03066G07 EExd Barrier Gland Kit (one pair of glands per kit)
1
V-Strut assembly includes contact and thermocouple assembly.
2
Contact and thermocouple assembly includes platinum pad and inconel wire.
NOTE
The replacement parts listed above must be obtained only from the manufacturer or his agent.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-19
Instruction Manual
Appendix AX Rev. 2.1 February 1998
World Class 3000
A-20 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
BX
HPS 3000 HEATER POWER SUPPLY FIELD MODULE
(CENELEC APPROVED VERSION)
Consult Safety Data Sheet 1M03243 for safety related information.
B-1 DESCRIPTION
The Rosemount CENELEC approved HPS 3000 Heater Power Supply Field Module (Figure B-1) acts as an interface between probe and elec­tronics and supplies power to the probe heater. The unit allows the use of probes with a number of different electronics packages.
Instruction Manual
Appendix BX Rev. 1.2
January 1997
APPENDIX BX, REV. 1.2
DESCRIPTION
P00001B
Figure B-1. CENELEC Approved HPS 3000
Heater Power Supply Field Module
The CENELEC approved HPS 3000 is certified EExd IIC T6 to CENELEC standards EN50014 and EN50018.
The heater power supply, Figure B-2, consists of a motherboard, daughterboard, and a trans­former for supplying correct voltage to the probe heater. The mother- and daughter­boards contain terminal strips for connecting probe, electronics, and power supply.
The HPS is jumper configurable for 120, 220, or 240 Vac.
If you reconfigure the equipment for a line voltage other than the one marked on the serial label, then you should change the marking on the serial label to state the new line voltage.
B-2 THEORY OF OPERATION
The HPS 3000 Heater Power Supply may per­form slightly different functions, depending upon which electronics package with which it is used. Figure B-3 shows a functional block diagram of the unit. The HPS contains a transformer for converting line voltage to 44 volts needed to power the probe heater. The relay can be used to remotely turn the probe on or off manually.
A triac module is used to turn the heater on or off, depending on probe temperature.
When used with the CRE 3000 Control Room Electronics or IFT 3000 Intelligent Field Trans­mitter, the HPS uses a cold junction tempera­ture compensation feature. This allows for the use of a less expensive cable between the HPS and CRE or HPS and IFT. The HPS and elec­tronics package can be located up to 364 m (1200 ft) apart.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices B-1
Instruction Manual
Appendix BX Rev. 1.2 January 1997
World Class 3000
TERMINAL
COVERS
(PROVIDED)
INTERNAL
EARTHING
HARDWARE
EXTERNAL
EARTHING
HARDWARE
SCREW
(2 PER COVER)
FRONT VIEW SIDE VIEW
TRANSFORMER
JM8
JM7
J7
JM4
JM5
JM2
1 JM
J2
Figure B-2. Heater Power Supply, Interior
TERMINAL STRIP (FROM ELECTRONICS)
TERMINAL STRIP (FROM PROBE)
P00002B
Table B-1. Specifications for Heater Power Supply
Power Supply .............................................................................. 110/115/220 VAC ± 10% at 50/60 Hz
Power Requirement .................................................................... 200 VA
Humidity Range........................................................................... 95% Relative Humidity
Ambient Temperature Range...................................................... 0° to 60°C (32° to 140°F)
Vibration ...................................................................................... 5 m/sec
2
, 10 to 500 xyz plane
External Electrical Noise ............................................................. Minimum Interference
Installation Category (Overvoltage Category)............................. IEC 664 Category II
Cabling Distance Between HPS 3000 and Probe....................... Maximum 45 m (150 ft)
Cabling Distance Between HPS 3000 and CRE 3000................ Maximum 364 m (1200 ft)
Cabling Distance Between HPS 3000 and IFT 3000.................. Maximum 364 m (1200 ft)
Approximate Shipping Weight..................................................... 20 kg (44 lbs)
B-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
BX
LINE
RELAY
TRIAC
TRANSFORMER
ZERO
CROSSING
DETECTOR
Instruction Manual
Appendix BX Rev. 1.2
January 1997
TRIACRELAY
TO HEATER
FROM
IFT OR
CRE
AD590
PROBE TC
STACK TC
CELL MV
COLD
JUNCTION
TEMP.
Figure B-3. Heater Power Supply Block Diagram
The standard cable, between probe and HPS, is thermocouple compensated. This prevents the additional junctions between thermocouple and cablefrom producing a voltage which would af­fect the thermocouple output signal. A tem­perature sensor in the HPS monitors the temperature at the junction and sends a voltage signal to the CRE and IFT. The CRE and IFT uses this signal to compensate the probe ther­mocouple reading for the temperature at the junction between the compensated and uncompensated cables.
In operation, when connected to the CRE 3000 Control Room Electronics, line voltage passes through the relay (when on) and is converted into 44 volts by the transformer. If the probe thermocouple indicates that the probe has dropped below operating temperature, a signal from the CRE triggers the triac. The triac then supplies voltage to the probe heater, warming the cell. Conversely, if the probe thermocouple indicates that the probe heater has reached the upper limit of operating temperature, the CRE deactivates the triac, shutting off power to the heater.
B-3 CONNECTIONS TO NEW GENERATION
ELECTRONICS (IFT 3000 AND CRE 3000)
Use the following information about input and output connections if your system includes CENELEC approved new generation electronics.
PROBE TC
STACK TC
CELL
a. Probe Thermocouple
PBTC+ and PBTC- (J3 pins 4 and 5, J8 pins 3 and 4). This thermocouple provides feed­back from the probe heater to the electron­ics.
b. Stack Thermocouple (Optional)
STTC+ and STTC- (J2 pins 1 and 2, J9 pins 5 and 6). The stack thermocouple is a ther­mocouple separate from the CENELEC ap­proved World Class 3000 Probe. It is used to measure stack temperature.
c. AD590
AD590+ and AD590- (J8 pin 5, J8 pin 6). The AD590 is a temperature measuring de­vice used to measure temperature inside the CENELEC approved Heater Power Supply HPS 3000.
d. Zirconium Cell
PBMV+ and PBMV- (J3 pins 1 and 2, J8 pins 1 and 2). The voltage signal from the zirconium cell.
e. Probe Heater
44 Vac output (J2 pins 4, 5, and 6). The 44 Vac output to power the probe heater.
TO PROBE
730004
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices B-3
Instruction Manual
Appendix BX Rev. 1.2 January 1997
World Class 3000
f. TRIAC
TRIAC+ and TRIAC- (J9 pins 1 and 2). The signal by which the new generation elec­tronics controls the triac of Heater Power Supply HPS 3000. By controlling this triac, the new generation electronics modulates the 44 V waveform to the probe heater.
g. Relay
Relay+ and Relay- (J9 pins 3 and 4). By ac­tivating or deactivating this relay, the new generation electronics can switch off or switch on the 44 Vac from the Heater Power Supply HPS 3000 to the probe heater. This feature is jumper selectable (JM2).
h. Line Voltage
(J1 pins 1, 2, and 3). Jumper selectable (JM1, JM4, and JM5) for 120 Vac or 220/240 Vac.
i. Shield (Cable Armour)
a. Probe Thermocouple
PBTC+ and PBTC- (J3 pins 4 and 5, J8 pins 3 and 4). This thermocouple provides feed­back from the probe heater to the electron­ics.
b. Stack Thermocouple
(Optional) STTC+ and STTC- (J2 pins 1 and 2, J9 pins 5 and 6). The stack thermocouple is a thermocouple separate from the WC3000 probe. It is used to measure stack temperature.
c. Zirconium Cell
PBMV+ and PBMV- (J3 pins 1 and 2, J8 pins 1 and 2). The voltage signal from the zirconium cell.
d. Probe Heater
44 Vac out (J2 pins 4, 5, and 6). This is the 44 Vac output to power the probe heater.
The cable armour, as well as providing me­chanical protection, acts as a shield. This shield is directly connected to the Heater Power Supply HPS 3000 housing through the cable gland.
NOTE
When using the HPS 3000 with an ex­isting electronics package, such as Models 218, 218A, or 225, the elec­tronics will not have the input/output capacity to support all of the functions mentioned in this section. Refer to In­struction Bulletin IB-106-300NEX.
B-4 CONNECTION TO OLD GENERATION
ELECTRONICS
Use the following information about input and output connections if your system does not in­clude CENELEC approved new generation digital electronics:
e. Line Voltage Pins
This is a modulated 115 Vac from the old generation electronics. Bridge rectifier (BR1) converts this modulated 115 Vac to the TRIAC+ and TRIAC- low voltage signal. This signal controls the triac of the Heater Power Supply HPS 3000. A modulated 44 Vac is thus sent to the probe heater.
f. Shield (Optional)
(J8 pin 7.) The PBMV+/PBMV- and the PBTC+/PBTC- lines to the old generation electronics may be shielded against electri­cal noise through these connections.
g. Shield (Cable Armour)
The cable armour, as well as providing me­chanical protection, acts as a shield. This shield is directly connected to the Heater Power Supply HPS 3000 housing through the cable gland.
B-4 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
BX
Instruction Manual
Appendix BX Rev. 1.2
January 1997
TROUBLESHOOTING
B-5 OVERVIEW
Before carrying out any work on the CENELEC approved HPS 3000, consult Safety Data Sheet 1M03243.
Install all protective equipment covers and safety ground leads after trouble­shooting. Failure to replace covers and ground leads could result in seri­ous injury or death.
The HPS 3000 troubleshooting section de­scribes how to identify and isolate faults which may develop in the HPS 3000 assembly.
B-6 HPS 3000 TROUBLESHOOTING
The HPS 3000 troubleshooting may overlap with the probe in use in the system. Faults in either system may cause an error to be dis­played in the electronics package. Figure B-4 and Figure B-5 provide troubleshooting informa­tion.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices B-5
Instruction Manual
Appendix BX Rev. 1.2 January 1997
SYMPTOM
World Class 3000
HEATER DOES NOT HEAT UP (DOES NOT INCREASE IN OUTPUT).
SET METER* FOR 50 VAC. PLACE PROBES ON TERMINAL BLOCK J2, “FROM PROBE”, PROBE HEATER.
METER INDICATES PULSATING NOMINAL 44 VAC.
YES
DISCONNECT POWER FROM HPS & PROBE ELECT. SET METER ON RX1.
MEASURE RESISTANCE OF HEATER BY PLACING PROBES ON TERMINAL BLOCK J2, “FROM PROBE”, PROBE HEATER.
NO
SET METER TO 250 VAC. PLACE METER PROBES ON J9, “FROM ELECTRONICS”, ANALOG HEATER.
METER SHOULD REGISTER A PULSATING NOMINAL 115 VAC.
YES
DISCONNECT POWER TO HPS AND PROBE ELECTRONICS. CHECK ALL FOUR FUSES IN HPS.
FUSES BLOWN IN HPS.
YES YES
CHECK THAT LINE VOLTAGE IS CORRECT BY SELECTING CORRECT JUMPERS ACCORDING TO CHART ON INSIDE OF HPS COVER.
NO
NO
CHECK FUSES IN PROBE ELECTRONICS.
IF FUSES OK, POSSIBLE PROBE ELECTRONICS TRIAC FAILURE.
CHECK JUMPER JM2 IS INSTALLED.
POSSIBLE TRIAC FAILURE. REPLACE HPS MOTHERBOARD.
RESISTANCE MEASURED SHOULD BE NOMINALLY 12 OHMS.
*SIMPSON MODEL 260 OR EQUIVALENT MULTIMETER.
NO
REPLACE BLOWN FUSES.
HEATER IS OPEN. REPLACE HEATER.
34990015
Figure B-4. HPS Troubleshooting Flowchart, #1
B-6 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
BX
SYMPTOM
HEATER OVERHEATS. NOTE: ON INITIAL STARTUP THE
TEMPERATURE OF THE PROBE MAY OVERHEAT TO A NOMINAL TEMP OF 800°C(480°F).
Instruction Manual
Appendix BX Rev. 1.2
January 1997
CHECK IF THE YELLOW WIRE IS CONNECTED TO POSITIVE AND THE RED TO NEGATIVE ON BOTH FROM ELECTRONICSAND FROM PROBETERMINAL BLOCKS. (MODELS 218 AND 225 ONLY)
YES
SET METER* ON 250 VAC SCALE.
CHECK FROM ELECTRONIC, ANALOG HEATER TERMINALS IF THERE IS A PULSATING NOMINAL 115 VAC ON J9.
YES
SET METER* ON 50 VAC SCALE.
NO
NO
CONNECT THE WIRING ACCORDING TO THE DECALS ON THE TERMINAL COVERS.
IF THE VOLTAGE IS CONSTANTLY ON, THEN THE TRIAC IN THE PROBE ELECTRONICS IS SHORTED.
CHECK FROM PROBE, TERMINALS AT PROBE HEATER IF THERE IS A PULSATING NOMINAL 44 VOLTS ON J2.
YES
CHECK PROBE ELECTRONICS SETPOINT HAS BEEN CHANGED ACCORDING TO APPLICABLE ELECTRONICS PACKAGE IB.
YES
CHECK IF PROBE HEATER VOLTAGE SELECTION JUMPER JM7 INSTALLED (REMOVE JM8).
*SIMPSON MODEL 260 OR EQUIVALENT MULTIMETER.
NO
IF THE VOLTAGE IS CONSTANTLY ON, THEN THE TRIAC MODULE IN THE HPS IS SHORTED.
34990014
Figure B-5. HPS Troubleshooting Flowchart, #2
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices B-7
Instruction Manual
Appendix BX Rev. 1.2 January 1997
MAINTENANCE AND SERVICE
Consult Safety Data Sheet 1M03243 before performing any work on the CENELEC approved HPS 3000.
Install all protective equipment covers and safety ground leads after equip­ment repair or service. Failure to in­stall covers and ground leads could result in serious injury or death.
B-7 OVERVIEW
This section describes service and routine maintenance of the HPS 3000 Heater Power Supply Field Module. Replacement parts re­ferred to are available from Rosemount. Refer to Replacement Parts of this appendix for part numbers and ordering information.
World Class 3000
b. Loosen setscrew (13, Figure B-6) retaining
HPS cover (14). Remove HPS cover.
c. Remove hex nut (2), flat washer (3), and
hex head bolt (4) that retain transformer to mounting bracket.
d. Disconnect the 6-pin transformer wiring
harness connector from J7 on the HPS motherboard (25). Disconnect the 2-pin wiring harness connector from the jumper that was selected to configure the HPS for the proper voltage.
NOTE
The transformer connector is keyed making connection possible in only one direction.
e. Remove old toroidal transformer (11). Place
new transformer in position and reconnect wiring harness connectors as noted in step d.
B-8 FUSE REPLACEMENT
The heater power supply's motherboard (25, Figure B-6) contains four identical 5 amp fuses (24) (5 amp anti-surge, Type T to IEC127, ROSEMOUNT Part No. IL01293H02). To check or replace a fuse, simply unscrew the fuse holder cap (23) with a flat head screwdriver and remove fuse. After checking or replacing a fuse, reinstall fuse holder cap.
B-9 TRANSFORMER REPLACEMENT
Consult Safety Data Sheet 1M03243 before performing any work on the CENELEC approved HPS 3000.
a. Follow the power down procedure outlined
in safety data sheet 1M03243 and the offi­cial "Codes of Practice" for your country of installation.
f. Place insulating disc (10) (one on either
side of transformer) and metal disc (12) on transformer.
g. Tighten hex nut (2) and hex head bolt (4)
only enough to firmly hold transformer in place.
h. Reinstall HPS cover (14) and secure with
setscrew (13).
B-10 MOTHERBOARD REPLACEMENT.
Consult Safety Data Sheet 1M03243 before performing any work on the CENELEC approved HPS 3000.
a. Follow the power down procedure outlined
in safety data sheet 1M03243 and the offi­cial "Codes of Practice" for your country of installation.
B-8 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
BX
Instruction Manual
Appendix BX Rev. 1.2
January 1997
NOTE:
NOT ALL PARTS SHOWN ARE AVAILABLE FOR PURCHASE SEPARATELY. FOR A LIST OF AVAILABLE PARTS, SEE TABLE B-2.
1
10
12
10
17
27
9
2 3
5
1. HPS Housing
2. Hex Nut
3. Flat Washer
4. Hex Head Bolt
5. Transformer Mounting Bracket
6. Screw
7. Lockwasher
8. Flat Washer
8
7
6
9. Hex Standoff
10. Insulating Disc
11. Toroidal Transformer
12. Metal Disc
11
2-PIN CONNECTOR
6-PIN CONNECTOR
13. Setscrew
14. HPS Cover
15. O-Ring Seal
16. Slotted Screw
17. Lockwasher
18. Flat Washer
19. Terminal Strip Cover (Probe)
20. Terminal Strip Cover (Electronics)
4
21. HPS Daughterboard PCB
22. Hex Stand Off
23. Fuse Holder Cap
24. Fuse
25. HPS Motherboard PCB
26. Hex Nut
27. Hex Standoff
28. Earthing Hardware
18
17
26
17
27
28
18
22
21
18
25
18
18
20
17
18
16
19
24
17
22
16
18
23
17
15
16
14
13
P00007
Figure B-6. Heater Power Supply, Exploded View
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices B-9
Instruction Manual
Appendix BX Rev. 1.2 January 1997
World Class 3000
b. Loosen setscrew (13) retaining HPS cover
(14). Remove cover.
c. Disconnect transformer wiring harness con-
nectors as described in the previous para­graph from motherboard (25).
d. Remove slotted screw (16), lockwasher
(17), and flat washer (18) attaching terminal strip covers (19) and (20). Remove terminal strip covers.
e. Unplug ribbon cable from the receptacle on
the daughterboard (21). Take note of loca­tion of black and white wires connected to pin 8 and pin 9 of terminal strip J8. Discon­nect these wires from J8.
f. Unscrew stand offs (22) on either side of
the daughterboard. Remove daughterboard (21).
g. Unscrew four stand offs (22) that support
the daughterboard.
h. Making a note of the location and color of
each wire, disconnect wires from terminal strip on HPS motherboard (25).
i. Remove hex nut (26), lockwasher (17), and
flat washer (18) securing the HPS mother­board to the enclosure and transformer mounting bracket (5).
j. Remove motherboard (25).
n. Plug ribbon cable back into receptacle on
daughterboard and reconnect black and white wires. The black wire goes to pin 8 and the white to pin 9 of terminal block J8 on daughter board. Reinstall terminal cov­ers.
o. Reconnect transformer wiring harness con-
nectors to motherboard. Note that in any case, the transformer's connector is keyed so that it can only be inserted one way.
p. Reinstall HPS cover (14) and secure with
setscrew (12).
B-11 DAUGHTERBOARD REPLACEMENT
Consult Safety Data Sheet 1M03243 before performing any work on the CENELEC approved HPS 3000.
a. Follow the power down procedure outlined
in safety data sheet 1M03243 and the offi­cial "Codes of Practice" for your country of installation.
b. Loosen setscrew (13, Figure B-6) retaining
HPS cover (14). Remove cover.
c. Remove slotted screw (16), lockwasher
(17), and flat washer (18) securing terminal strip covers (19) and (20). Remove terminal strip covers.
k. Position new motherboard on stand offs and
reinstall hex nut, lockwasher, and flat washer removed in step i.
l. Reconnect wires to terminal strip in posi-
tions noted in step h. Cross check with wir­ing diagram shown on Figure 2-4 of the main text (IB-106-300NX Series).
m. Reinstall four stand offs removed in step g.
Position daughterboard (21) on stand offs and reinstall stand offs removed in step f.
B-10 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
d. Making a note of the location and color of
each wire, disconnect wires from the termi­nal strip on the daughterboard (21).
e. Unplug ribbon cable from receptacle on
daughterboard. Take note of location of black and white wires connected to pin 8 and pin 9 of terminal strip J8. Disconnect these wires from J8.
f. Unscrew two stand offs (22) from daughter-
board. Remove daughterboard (21).
World Class 3000
BX
Instruction Manual
Appendix BX Rev. 1.2
January 1997
g. Position new daughterboard on four stand
offs (22) on motherboard (25). Reinstall the stand offs removed in step f.
h. Plug ribbon cable back into receptacle on
daughterboard and reconnect black and white wires. The black wire goes to pin 8 and the white wire to pin 9 of terminal block J8 on daughterboard. Reinstall terminal covers.
i. Reconnect wires to terminal strip in posi-
tions noted in step d. Cross check with wir­ing diagram shown on Figure 2-4 of the main text (IB-106-300NX Series). Reinstall terminal covers.
j. Reinstall HPS cover (14) and secure with
setscrew (13).
REPLACEMENT PARTS
Table B-2. Replacement Parts for Heater Power Supply
FIGURE and
INDEX No.
B-1 1U05667G01 CENELEC HPS 3000 (120V) B-1 1U05667G03 CENELEC HPS 3000 (220V/240V)
B-6, 24 1L01293H02 Fuse 5A @ 250 Vac, anti-surge, case size; 5 x 20 mm, type T
B-6, 25 3D39080G02 Motherboard B-6, 21 3D39078G01 Daughterboard B-6, 11 1M02961G05 Toroidal Transformer (120V/240V)
PART
NUMBER DESCRIPTION
to IEC127, Schurter
IB-106-300NEX Instruction Bulletin Upgrade and Digital Electronics IB-106-300NFX Instruction Bulletin (IFT) IB-106-300NCX Instruction Bulletin (CRE)
NOTE
The replacement parts listed above must be obtained only from the manufacturer or his agent.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices B-11
Instruction Manual
Appendix BX Rev. 1.2 January 1997
World Class 3000
B-12 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Loading...