Rosemount Wireless Pressure Gauge with WirelessHART Protocol Manuals & Guides

00809-0100-4045, Rev AC
Rosemount™ Wireless Pressure Gauge
with WirelessHART® Protocol
Reference Manual
September 2019
2
Reference Manual Contents
00809-0100-4045 September 2019

Contents

Chapter 1 Rosemount Wireless Pressure Gauge............................................................................. 5
Chapter 2 Introduction.................................................................................................................. 7
2.1 Using this manual.............................................................................................................................7
2.2 Models covered................................................................................................................................ 7
2.3 Product recycling/disposal................................................................................................................7
Chapter 3 Hardware installation.................................................................................................... 9
3.1 Overview.......................................................................................................................................... 9
3.2 Safety messages...............................................................................................................................9
3.3 Considerations................................................................................................................................. 9
3.4 Installation procedure.................................................................................................................... 11
3.5 Impulse piping considerations........................................................................................................14
3.6 Process connection........................................................................................................................ 15
3.7 Rosemount manifolds.................................................................................................................... 16
Chapter 4 Configuration...............................................................................................................21
4.1 Overview........................................................................................................................................ 21
4.2 Safety messages.............................................................................................................................21
4.3 System readiness............................................................................................................................22
4.4 Configuration basics.......................................................................................................................22
4.5 Basic gauge setup...........................................................................................................................23
4.6 Configuration verification...............................................................................................................24
4.7 Advanced device parameter setup..................................................................................................26
4.8 Notifications and service................................................................................................................ 27
4.9 Advanced configuration................................................................................................................. 28
Chapter 5 Operation and maintenance.........................................................................................33
5.1 Overview........................................................................................................................................ 33
5.2 Safety messages.............................................................................................................................33
5.3 Pressure signal trimming................................................................................................................ 33
5.4 Replacing the battery..................................................................................................................... 38
5.5 Local device status and notifications...............................................................................................38
Chapter 6 Troubleshooting.......................................................................................................... 41
6.1 Service support...............................................................................................................................41
6.2 Local troubleshooting.....................................................................................................................41
6.3 Remote troubleshooting................................................................................................................ 42
Chapter 7 Reference data............................................................................................................. 45
7.1 Product certificatons...................................................................................................................... 45
7.2 Ordering information, specifications, and dimensional drawings....................................................45
Emerson.com/Rosemount iii
Contents Reference Manual
September 2019 00809-0100-4045
Appendix Appendix A Field Communicator Menu Trees................................................................47
A.1 Overview........................................................................................................................................47
Appendix Appendix B Network Design Best Practices...................................................................51
B.1 Overview........................................................................................................................................ 51
B.2 Effective range............................................................................................................................... 51
Appendix Appendix C Device Variable Index................................................................................ 53
C.1 Overview........................................................................................................................................53
C.2 Messages and descriptions.............................................................................................................53
iv Emerson.com/Rosemount
Reference Manual Rosemount Wireless Pressure Gauge
00809-0100-4045 September 2019

1 Rosemount Wireless Pressure Gauge

NOTICE
The products described in this document are NOT designed for nuclear-qualified applications. Using non-nuclear qualified products in applications that require nuclear­qualified hardware or products may cause inaccurate readings.
For information on Emerson Automation Solutions nuclear-qualified products, contact your local Rosemount Sales Representative.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation.
Changes or modification not expressly approved by Rosemount Inc. could void the user’s authority to operate the equipment.
Using the Rosemount Wireless Pressure Gauge in a manner other than what is specified by the manufacturer may impair the protection provided by the equipment.
This device must be installed to ensure a minimum antenna separation distance of 20 cm (8 in.) from all persons.
Shipping considerations The device is shipped with the battery installed.Each device contains one “D” size primary
lithium-thionyl chloride battery. Primary lithium batteries are regulated in transportation by the U.S. Department of Transportation, and are also covered by IATA (International Air Transport Association), ICAO (International Civil Aviation Organization), and ARD (European Ground Transportation of Dangerous Goods). It is the responsibility of the shipper to ensure compliance with these or any other local requirements. Consult current regulations and requirements before shipping.
WARNING
Explosions could result in death or serious injury.
Installation of device in an explosive environment must be in accordance with appropriate local, national, and international standards, codes, and practices.
Ensure device is installed in accordance with intrinsically safe or non-incendive field practices.
Before connecting a communicator in an explosive atmosphere, make sure the instruments in the segment are installed in accordance with intrinsically safe or non­incendive field wiring practices.
Verify the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations certifications.
Emerson.com/Rosemount 5
Rosemount Wireless Pressure Gauge Reference Manual
September 2019 00809-0100-4045
WARNING
Electrical shock could cause death or serious injury.
Care must be taken during transportation of power module to prevent electrostatic charge build-up.
Device must be installed to ensure a minimum antenna separation distance of 8 in. (20 cm) from all persons.
Process leaks could result in death or serious injury.
Handle the device carefully.
Failure to follow these installation guidelines could result in death or serious injury.
Make sure only qualified personnel perform installation or service.
Apply wrench only to the flats, not on housing.
The battery is not replaceable in a hazardous location.
CAUTION
Keep the vent path free of any obstruction, including but not limited to paint, dust, and lubrication by mounting the device so the process can drain away.
Interfering or blocking the atmospheric reference port will cause the device to output erroneous pressure values.
Absolute pressure devices are calibrated at the factory. Trimming adjusts the position of the factory characterization curve. It is possible to degrade performance of the device if any trim is done improperly or with inaccurate equipment.
Individuals who handle products exposed to a hazardous substance can avoid injury if they are informed of and understand the hazard. The product being returned will require a copy of the required Safety Data Sheet (SDS) for each substance must be included with the returned goods.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
6 Emerson.com/Rosemount
Reference Manual Introduction
00809-0100-4045 September 2019

2 Introduction

2.1 Using this manual

The sections in this manual provide information on installing, operating, and maintaining the Rosemount™ Wireless Pressure Gauge with WirelessHART® protocol. The sections are organized as follows:
Hardware installation contains mechanical and electrical installation instructions and
considerations.
Configuration provides instruction on commissioning and operating the gauge.
Information on software functions, configuration parameters, and online variables are also included.
Operation and maintenance contains operation and maintenance techniques.
Troubleshooting provides troubleshooting techniques for the most common operating
problems.
Reference data supplies reference and specification data, as well as ordering information.
Field Communicator menu trees provides full menu trees and abbreviated fast key
sequences for commissioning tasks.
Network design best practices provides information on how to optimize network reliability
and performance.
Device variable index outlines the most important alerts in the HART command 48
additional status field for Rosemount Wireless Pressure Gauge.

2.2 Models covered

This manual covers the Rosemount Wireless Pressure Gauge.
Measures gage/absolute/compound/vacuum pressure up to 10,000 psi (689.5 bar).

2.3 Product recycling/disposal

Recycling of equipment and packaging should be taken into consideration and disposed of in accordance with local and national legislation or regulations.
Emerson.com/Rosemount 7
Introduction Reference Manual
September 2019 00809-0100-4045
8 Emerson.com/Rosemount
Reference Manual Hardware installation
00809-0100-4045 September 2019

3 Hardware installation

3.1 Overview

The information in this section covers installation considerations. A Quick Start Guide is shipped with every device to describe basic installation and startup procedures. Dimensional drawings for the Rosemount™ Smart Pressure Gauge can be found in the
Product Data Sheet.

3.2 Safety messages

Procedures and instructions in this section may require special precautions to ensure the safety of the personnel performing the operation. Information that raises potential safety issues is indicated with a warning symbol ( ). Refer to the following safety messages before performing an operation preceded by this symbol.
WARNING
Explosions could result in death or serious injury.
Installation of device in an explosive environment must be in accordance with appropriate local, national, and international standards, codes, and practices.
Ensure device is installed in accordance with intrinsically safe or non-incendive field practices.
Electrical shock could cause death or serious injury.
Care must be taken during transportation of device to prevent electrostatic charge build­up.
Process leaks could result in death or serious injury.
Handle the device carefully.
Failure to follow safe installation guidelines could result in death or serious injury.
Only qualified personnel should install the equipment.

3.3 Considerations

3.3.1 Pre-installation

Optional: power/device check
The device is designed to be installation-ready. To check device battery prior to installation, perform the following:
Emerson.com/Rosemount 9
1 2 3
64 5
8 709
A
B
C
Hardware installation Reference Manual
September 2019 00809-0100-4045
Procedure
1. Perform Turn on device.
2. Slide the ON/OFF switch to the OFF position until ready for use.
Field Communicator connections
The device needs to be turned on in order for the Field Communicator to interface with the Rosemount Wireless Pressure Gauge. The Field Communicator connection is located to the right of the ON/OFF switch. To communicate with the device, connect the Field Communicator to connections labeled “COMM”. Field communication with this device requires a HART-based tool using the correct Rosemount Smart Pressure Gauge device driver (DD). Refer to Figure 3-1 for instructions on connecting the Field Communicator to the device.
Figure 3-1: Connect to Device
A
B
C
Field Communicator
HART modem
AMS Device Manager

3.3.2 Installation

Measurement performance depends upon proper installation of the device and impulse piping. Mount the device close to the process and use minimal piping to achieve best performance. Also, consider the need for easy access, personnel safety, and a suitable device environment. Install the device to minimize vibration, shock, and temperature fluctuation.
10 Emerson.com/Rosemount
Reference Manual Hardware installation
00809-0100-4045 September 2019

3.3.3 Mechanical

Location
When choosing an installation location and position, take into account the direction of the device for future access to the COMM connections and readability of the analog display.
Electronics cover
The electronics cover is tightened so that polymer contacts polymer. When removing the electronics cover, ensure that there is no damage done to the O-ring. If damaged, replace before reattaching cover, ensuring polymer contacts polymer (i.e. no O-ring visible).

3.3.4 Electrical

Battery
The Rosemount Smart Pressure Gauge is self-powered. The battery contains approximately five grams of lithium. Under normal conditions, the battery materials are self-contained and are not reactive as the battery is maintained inside the enclosure of the device. Care should be taken to prevent thermal, electrical, or mechanical damage. Contacts should be protected to prevent premature discharge.
Use caution when handling the battery, it may be damaged if dropped.
The battery should be stored in a clean, dry area. For maximum battery life, storage temperature should not exceed 86 °F (30 °C).
3.3.5

Environmental

Verify the operating atmosphere of the device is consistent with the appropriate hazardous locations certifications.
Temperature effects
The device will operate within specifications for ambient temperatures as outlined in the specifications section of the Product Data Sheet. Heat from the process is transferred to the device housing. If the process temperature is high, the ambient temperature will need to be lower to account for heat transferred to the device housing. See Temperature limits for process temperature derating.

3.4 Installation procedure

Figure 3-2: Installation Flowchart
Emerson.com/Rosemount 11
OR
Hardware installation Reference Manual
September 2019 00809-0100-4045

3.4.1 Seal and protect threads

3.4.2 Mount device

Note
Use wrench on flats, not on housing.
Mounting orientation
The low side pressure port (atmospheric reference) on the pressure gauge is located in the neck of the device behind the housing. The vent path is between the housing and sensor. See Figure 3-3.
CAUTION
Keep the vent path free of any obstruction, including but not limited to paint, dust, and lubrication by mounting the device so the process can drain away.
12 Emerson.com/Rosemount
A
Reference Manual Hardware installation
00809-0100-4045 September 2019
Figure 3-3: Low Side Pressure Port
A
Low side pressure port (atmospheric reference)

3.4.3 Turn on device

Check to ensure the device and battery are working properly.
Procedure
1. Twist the cover counterclockwise to remove it.
2. Slide the OFF/ON switch to the ON position to initiate the power sequence.
Note
During the power sequence, the dial tests full range of motion and LED flashes amber.
3. Once the power sequence ends, verify the LED flashes green.
Note
The LED may display several colors; see Table 5-2 for device statuses.
Emerson.com/Rosemount 13
Hardware installation Reference Manual
September 2019 00809-0100-4045

3.5 Impulse piping considerations

3.5.1 Best practices

The piping between the process and the device must accurately transfer the pressure to obtain accurate measurements. There are five possible sources of error: leaks, friction loss (particularly if purging is used), trapped gas in a liquid line, liquid in a gas line, and density variations between the legs.
The best location for the device in relation to the process pipe depends on the process itself. Use the following guidelines to determine device location and placement of impulse piping:
Keep impulse piping as short as possible.
For liquid service, slope the impulse piping at least 1in. per ft. (8 cm per m) upward
from the device toward the process connection.
For gas service, slope the impulse piping at least 1in. per ft. (8 cm per m) downward
from the device toward the process connection.
3.5.2
Avoid high points in liquid lines and low points in gas lines.
Make sure both impulse legs are the same temperature.
Use impulse piping large enough to avoid friction effects and blockage.
Vent all gas from liquid piping legs.
When using a sealing fluid, fill both piping legs to the same level.
When purging, make the purge connection close to the process taps and purge
through equal lengths of the same size pipe. Avoid purging through the device.
Keep corrosive or hot (above 250 °F [121 °C]) process material out of direct contact
with the sensor module and flanges.
Prevent sediment deposits in the impulse piping.
Keep the liquid head balanced on both legs of the impulse piping.
Avoid conditions that might allow process fluid to freeze within the process flange.

Mounting requirements

Liquid flow measurement
Place taps to the side of the line to prevent sediment deposits on the process isolators.
Mount the device beside or below the taps so gases vent into the process line.
Mount drain/vent valve upward to allow gases to vent.
Gas flow measurement
Place taps in the top or side of the line.
Mount the device beside or above the taps so to drain liquid into the process line.
14 Emerson.com/Rosemount
A
Reference Manual Hardware installation
00809-0100-4045 September 2019
Steam flow measurement
Place taps to the side of the line.
Mount the device below the taps to ensure that impulse piping will remain filled with
condensate.
Fill impulse lines with water to prevent steam from contacting the device directly and
to ensure accurate measurement start-up.
Note
For steam or other elevated temperature services, it is important that temperatures do not exceed 250 °F (121 °C) for devices with silicone fill. For vacuum service, these temperature limits are reduced to 220 °F (104 °C) for silicone fill.

3.6 Process connection

CAUTION
Interfering or blocking the atmospheric reference port will cause the device to output erroneous pressure values.
Keep the vent path free of any obstruction, including but not limited to paint, dust, and lubrication by mounting the device so the process can drain away.
The low side pressure port (atmospheric reference) on the pressure gauge is located in the neck of the device behind the housing. The vent path is between the housing and sensor. (See Figure 3-3).
Figure 3-4: Low Side Pressure Port
A low side pressure port (atmospheric reference).
WARNING
Do not apply torque directly to the sensor module. Rotation between the sensor module and the process connection can damage the electronics. To avoid damage, apply torque only to the hex-shaped process connection.
Emerson.com/Rosemount 15
Hardware installation Reference Manual
September 2019 00809-0100-4045

3.7 Rosemount manifolds

The Rosemount 306 Integral Manifold mounts directly to the device. The manifold is used with this device to provide block-and-bleed valve capabilities of up to 10,000 psi (689.5 bar).

3.7.1 Installation procedure

The Rosemount 306 Integral Manifold is for use only with a Rosemount Wireless Pressure Gauge.
Assemble the Rosemount 306 Manifold to the device with a thread sealant.
Procedure
1. Place device into holding fixture.
2. Apply appropriate thread paste or tape to threaded instrument end of the manifold.
3. Count total threads on the manifold before starting assembly.
4. Start turning the manifold by hand into the process connection on the device.
3.7.2
Note
If using thread tape, be sure the thread tape does not strip when the manifold assembly is started.
5. Wrench tighten manifold into process connection (minimum torque value is 425 in-
lbs).
6. Count how many threads are still showing (minimum engagement is three
revolutions).
7. Subtract the number of threads showing (after tightening) from the total threads to
calculate the revolutions engaged. Further tighten until a minimum of three rotations is achieved.
8. For block and bleed manifold, verify the bleed screw is installed and tightened. For
2-valve manifold, verify the vent plug is installed and tightened.
9. Leak-check assembly to maximum pressure range of device.

Manifold operation

2-valve and block and bleed style manifolds
Isolating the device
In normal operation the Isolate (block) valve between the process port and device will be open and the Test/Vent valve will be closed. On a block and bleed style manifold, a single block valve provides device isolation and a bleed screw provides drain/vent capabilities.
16 Emerson.com/Rosemount
Device
Test/vent
(closed)
Isolate
Process
(open)
Device
Test/vent
(closed)
Isolate
Process
(closed)
Reference Manual Hardware installation
00809-0100-4045 September 2019
Procedure
1. To isolate the device, close the isolate valve.
2. To bring the device to atmospheric pressure, open the vent valve or bleed screw.
Note
A ¼-in. male NPT pipe plug may be installed in the test/vent port and will need to be removed with a wrench in order to vent the manifold properly.
Emerson.com/Rosemount 17
Device
Test/vent
(open)
Isolate
Process
(closed)
Device
Test/vent
(closed)
Isolate
Process
(closed)
Device
Test/vent
(closed)
Isolate
Process
(open)
Hardware installation Reference Manual
September 2019 00809-0100-4045
3. After venting to atmosphere, perform any required calibration and then close the
test/vent valve or replace the bleed screw.
4. Open the Isolate (block) valve to return the device to service.
18 Emerson.com/Rosemount
Reference Manual Hardware installation
00809-0100-4045 September 2019
Adjusting valve packing
This is a 20-50-word description of the value to the user in this task. REQUIRED.
Over time, the packing material inside a Rosemount manifold may require adjustment in order to continue to provide proper pressure retention. Not all Rosemount manifolds have this adjustment capability. The Rosemount manifold model number will indicate what type of stem seal or packing material has been used.
The following steps are provided as a procedure to adjust valve packing.
Procedure
1. Remove all pressure from device.
2. Loosen manifold valve jam nut.
3. Tighten manifold valve packing adjuster nut ¼ turn.
4. Tighten manifold valve jam nut.
5. Re-apply pressure and check for leaks.
6. Above steps can be repeated, if necessary.
If the above procedure does not result in proper pressure retention, the complete manifold should be replaced.
Emerson.com/Rosemount 19
Hardware installation Reference Manual
September 2019 00809-0100-4045
A Bonnet
B Stem
C Packing
D Ball seat
E Packing adjuster
F Jam nut
G Packing follower
20 Emerson.com/Rosemount
Reference Manual Configuration
00809-0100-4045 September 2019

4 Configuration

4.1 Overview

This section contains information on commissioning and tasks.
Field Communicator and AMS Device Manager Instructions are given to perform configuration functions.
Full Field Communicator menu trees are available in Field Communicator menu trees.

4.2 Safety messages

Procedures and instructions in this section may require special precautions to ensure the safety of the personnel performing the operation. Information that raises potential safety issues is indicated with a warning symbol ( ). Refer to the following safety messages before performing an operation preceded by this symbol.
WARNING
Explosions could result in death or serious injury.
Installation of this device in an explosive environment must be in accordance with the appropriate local, national, and international standards, codes, and practices. Review the approvals section of the Rosemount™ Smart Pressure Gauge Reference Manual for any restrictions associated with a safe installation.
Before connecting a Field Communicator in an explosive atmosphere, make sure the instruments are installed in accordance with intrinsically safe or non-incendive field wiring practices.
Verify that the operating environment of the device is consistent with the appropriate hazardous locations certifications.
NOTICE
This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation.
This device must be installed to ensure a minimum antenna separation distance of 20 cm (8 in.) from all persons.
Emerson.com/Rosemount 21
Measured
process
input
P A/D
Micro Local Display output
Configuration Reference Manual
September 2019 00809-0100-4045

4.3 System readiness

4.3.1 Confirm correct device driver

Verify the latest Device Description (DD/DTM™) is loaded on your systems to ensure proper communications.
Procedure
1. Visit the Emerson Device Install Kits Library or Fieldcommgroup.org.
2. Select desired product.
a) Within Table 4-1, use the HART Universal Revision and Device Revision
numbers to find the correct Device Description.
Table 4-1: Rosemount Smart Pressure Gauge Device Revisions and Files
Software release date
October 2017
(1) NAMUR Software Revision is located on the hardware tag of the device. (2) HART Software Revision can be read using a HART capable configuration tool. (3) Device Driver file names use Device and DD Revision (e.g. 10_01). HART Protocol is
designed to enable legacy device driver revisions to continue to communicate with new HART devices. To access new functionality, the new Device Driver must be downloaded. It is recommended to download new Device Driver files to ensure full functionality.
Identify device Find device driver Review
NAMUR software revision
1.0.0 1.0.0 2 7 1 00809-0100-4145Initial release
(1)
NAMUR software revision
Figure 4-1: Data Flow

4.4 Configuration basics

HART software revision
(2)
HART universal revision
Device revision
instructions
Manual
(3)
document number
Review functionality
Changes to software

4.4.1 Configuration tools

Configuration requires a Field Communicator or AMS Device Manager. Connect the Field Communicator leads to the terminals labeled “COMM” on the front of the device (see
Figure 3-1).
When using a Field Communicator, any configuration changes made must be sent to the device by using the Send key (F2). AMS Device Manager configuration changes are implemented when the Apply button is selected.
22 Emerson.com/Rosemount
Reference Manual Configuration
00809-0100-4045 September 2019

4.4.2 Connection diagrams

Figure 3-1 illustrates the wiring for a field hook-up with a Field Communicator or AMS
Device Manager. The Field Communicator or AMS Device Manager may be connected at “COMM” on the device.

4.5 Basic gauge setup

4.5.1 Eliminate mounting effects

Devices are factory-calibrated. Once installed, it is recommended to perform this step to eliminate potential error caused by mounting position or static pressure. Instructions for using a Field Communicator are listed below:
Procedure
1. Vent the device.
2. Connect the Field Communicator.
3. From the HOME screen, enter the HART Fast Key sequence.
4.5.2
Fast Keys
4. Follow the commands to perform the procedure.
2, 1, 1

Activate wireless

Do not activate wireless until Emerson Wireless Gateway is installed and functioning properly; toggling off and on reduces battery life.
Join device to network.
1. Obtain Network ID and Join Key for the wireless network (available in wireless
gateway).
2. From the HOME screen, enter the HART Fast Key sequence.
Fast Keys
3. Follow the commands to perform the procedure.
4. Select Overview > Status.
5. Verify communication status displays Connected.
Note
Joining the device to the network could take several minutes.
2, 1, 2
Emerson.com/Rosemount 23
Configuration Reference Manual
September 2019 00809-0100-4045

4.5.3 Considerations for devices with percent of range engineering unit

Set range points
The range values command sets the lower and upper range values used for the percent of range engineering unit.
Note
Devices are shipped from Emerson fully calibrated to the factory default of full scale (scale range = upper range limit).
From the HOME screen, enter the Fast Key sequence.
Fast Keys
Procedure
1. Select lower or upper range value as applicable.
2. Follow the commands to perform the procedure.
2, 2, 1, 2

4.6 Configuration verification

The following is a list of factory default configurations that can be viewed by using the Field Communicator or AMS Device Manager. Follow the steps below to review the gauge configuration information.
Note
Information and procedures in this section that make use of Field Communicator Fast Key sequences and AMS Device Manager assume the gauge and communication equipment are connected, powered, and operating correctly.
4.6.1

Review pressure information

From the HOME screen, enter the Fast Key sequence.
Fast Keys
1, 2
Procedure
1. From the Home screen, select 1: Overview.
2. Select 2: Pressure.
4.6.2
24 Emerson.com/Rosemount

Review device information

From the HOME screen, enter the Fast Key sequence.
Fast Keys
1, 9
Reference Manual Configuration
00809-0100-4045 September 2019
Procedure
1. From the Home screen, select 1: Overview.
2. Select 9: Device Information.
3. Select from the corresponding number to view each field:
1 Identification
2 Revisions
3 Materials of Construction
4 Security
5 Dial Faceplate
6 Capabilities
4.6.3
4.6.4

Review radio information

From the HOME screen, enter the Fast Key sequence.
Table 4-2:
Fast Keys 1, 9, 3
1. From the Home screen, select 1: Overview.
2. Select 9: Device Information.
3. Select 3: Radio.
4. Select from the corresponding number to view each field: 1 MAC address 2 Manufacturer 3 Device type 4 Devision revision 5 Software revisions 6 Hardware revision 7 Xmit power level 8 Min brdcst rate

Review operating parameters

The pressure output value in both engineering units and percent of range will reflect it even when it is outside of the configured range as long as it is between the upper and lower range limit of the device. For example, if a scale range 0 - 150 psi (LRL = 0 psi, URL = 150 psi) is ranged from 0 to 100 psi, an applied pressure of 150 psi will return a percent of range output of 150 percent.
From the HOME screen, enter the Fast Key sequence.
Fast Keys
Emerson.com/Rosemount 25
3, 2, 1
Configuration Reference Manual
September 2019 00809-0100-4045
Procedure
1. From the Home screen, select 3: Service Tools.
2. Select 2: Variables.
3. Select 1: All Variables.
The Operating Parameters menu displays the following information pertaining to the device:
All variables:
Pressure
Pressure Quality
Custom Scale
Cust Scale Quality
Percent of Range
Percent of Rng Quality
Sensor Temp
Sensor Temp Quality
Sensor Temp Unit
Supply Voltage
Supply Voltage Quality

4.7 Advanced device parameter setup

4.7.1 Write protect

The device has a software write protect security feature.
From the HOME screen, enter the Fast Key sequence.
Table 4-3:
Fast Keys 2, 2, 4, 1
1. Select Write Protect to enable.
2. Right click on device and select 2: Configure.
3. Select 2: Advanced Setup.
4. Select the tab labeled 4: Security.
5. Select Write Protect to enable this feature.
26 Emerson.com/Rosemount
Reference Manual Configuration
00809-0100-4045 September 2019

4.7.2 Wireless update rate

From the HOME screen, enter the Fast Key sequence.
Fast Keys 2, 2, 3, 2
1. From the Home screen, select 2: Configure.
2. Select 2: Advanced Setup.
3. Select 3: Wireless.
4. Select 2: Update Rate.
5. Follow the commands to perform the procedure.
4.7.3

Dial update rate

From the HOME screen, enter the Fast Key sequence.
Fast Keys
Procedure
1. From the Home screen, select 2: Configure.
2. Select 2: Manual Setup.
3. Select 1: Measurements.
4. Select 1: Dial/Pressure.
5. Select 2: Dial Update Rate.
6. Follow the commands to perform the procedure.
2, 2, 1, 1, 2

4.8 Notifications and service

Notifications and service functions listed below are primarily for the user after field installation. The device simulation feature is designed to verify proper operating functionality, and can be performed either on the bench or in the field.
4.8.1

Simulating device variables

From the HOME screen, enter the Fast Key sequence.
Fast Keys
Procedure
1. From the Home screen, select 3: Service Tools.
2. Select 4: Simulate.
Emerson.com/Rosemount 27
3, 4
Configuration Reference Manual
September 2019 00809-0100-4045
Note
The following parameters pertaining to the device can be simulated: Pressure, sensor temperature, and supply voltage

4.8.2 Device reset

The master reset function will reset the device electronics. To perform a device reset:
From the HOME screen, enter the Fast Key sequence.
Fast Keys 3, 3, 1
Procedure
1. From the Home screen, select 3: Service Tools.
2. Select 3: Maintenance
3. Select 1: Device Reset
4.8.3

Join status

From the HOME screen, enter the Fast Key sequence.
Table 4-4:
Fast Keys 3, 3, 1
1. From the Home screen, select 3: Service Tools.
2. Select 3: Communications.
3. Select 1: Join Status.
Wireless devices join the secure network through a four-step process:
Step 1. Network Found
Step 2. Network Security Clearance Granted
Step 3. Network Bandwidth Allocated
Step 4. Network Join Complete

4.9 Advanced configuration

4.9.1 Overpressure notification

This notification can be used to know if a process pressure higher than 105 percentage of the devices maximum working pressure (MWP) has been measured. The overpressure notification must be configured to latched mode to activate the notification. If this event occurs when the device is configured to latch, the dial will be driven to the Red X and the LED will blink red. Additionally, it is required to acknowledge and reset the overpressure
28 Emerson.com/Rosemount
Reference Manual Configuration
00809-0100-4045 September 2019
notification after an overpressure event before the dial can move back to an on-scale position.
Table 4-5 contains further information on device specific MWP as it correlates to the
device specific scale range.
Table 4-5: Maximum Working Pressure
Scale range Maximum working
pressure (MWP)
Vacuum to 30 psi 30 psi 31.5 psi 750 psi
31–150 psi 150 psi 157.5 psi 1,500 psi
151–800 psi 800 psi 840 psi 1,600 psi
801–4,000 psi 4,000 psi 4,200 psi 6,000 psi
4,001-10,000 psi 10,000 psi 10,500 psi 15,000 psi
105% of MWP Maximum overpressure
limit
Table 4-6 demonstrates the different dial locations based on configuration of the
overpressure notification (Unlatched vs Latched).
Table 4-6: Dial Locations
Measured process pressure
Within scale range
Parameter configuration
Unlatched (factory default) Latched
LED color: Green Dial location: On-scale
Emerson.com/Rosemount 29
LED color: Green Dial location: On-scale
Configuration Reference Manual
September 2019 00809-0100-4045
Table 4-6: Dial Locations (continued)
Measured process pressure
Beyond scale range and <105% of MWP
>105% MWP
Parameter configuration
Unlatched (factory default) Latched
LED color: Green Dial location: Off-scale
LED color: Green Dial location: Off-scale
LED color: Green Dial location: Off-scale
See Local device status and notifications for more information.
From the HOME screen, enter the Fast Key sequence
Fast Keys
2, 2, 1, 1, 3
Procedure
1. From the Home screen, select 2: Configure
2. Select 2: Manual Setup
3. Select 1: Measurements
4. Select 1: Dial/Pressure
5. Select 3: Over-Press Ind
6. Follow the commands to perform the procedure.
LED color: Red Dial location: Red X
30 Emerson.com/Rosemount
Reference Manual Configuration
00809-0100-4045 September 2019
Note
When the parameter has been set to activate, the notification must be acknowledged and cleared for the device to return to normal operation.

4.9.2 Acknowledge and reset overpressure notification

From the HOME screen, enter the Fast Key sequence
Fast Keys 3, 4, 3
1. From the Home screen, select 3: Service Tools.
2. Select 4: Maintenance.
3. Select 3: Acknowledge Over-Pressure.
4. Follow the commands to perform the procedure.
Emerson.com/Rosemount 31
Configuration Reference Manual
September 2019 00809-0100-4045
32 Emerson.com/Rosemount
Reference Manual Operation and maintenance
00809-0100-4045 September 2019

5 Operation and maintenance

5.1 Overview

This section contains information on commissioning and operating Rosemount™ Wireless Pressure Gauges.
Field Communicator and AMS Device Manager instructions are provided for convenience.

5.2 Safety messages

Procedures and instructions in this section may require special precautions to ensure the safety of the personnel performing the operation. Information that raises potential safety issues is indicated with a warning symbol ( ). Refer to the following safety messages before performing an operation preceded by this symbol.

5.3 Pressure signal trimming

Calibrating a Rosemount Wireless Pressure Gauge may include the sensor trim procedure to adjust for mounting effects.
Sensor trimming requires an accurate pressure input and adds additional compensation that adjusts the position of the factory trim to optimize performance over a specific pressure range.
Note
Sensor trimming adjusts the position of the factory trim. It is possible to degrade the performance of the gauge if the trim is done improperly or with inaccurate equipment.
CAUTION
Absolute pressure devices are calibrated at the factory. Trimming adjusts the position of the factory characterization curve. It is possible to degrade performance of the device if any trim is done improperly or with inaccurate equipment.
Table 5-1: Recommended Calibration Tasks
Measurement type Tasks
Gage Compound Vacuum
1. Reconfigure parameters if necessary.
2. Zero trim the device to compensate for mounting effects or static pressure effects.
3. Optional: Perform a sensor trim. (Accurate pressure source required.)
Emerson.com/Rosemount 33
Operation and maintenance Reference Manual
September 2019 00809-0100-4045
Table 5-1: Recommended Calibration Tasks (continued)
Measurement type Tasks
Absolute 1. Reconfigure parameters if necessary.
2. Perform low trim value section of the sensor trim procedure to correct for mounting position effects.
3. Optional: Perform a sensor trim if equipment available (accurate absolute pressure source required), otherwise perform the low trim value section of the sensor trim procedure.
Note
For devices with absolute measurement type, an accurate absolute pressure source is required.
5.3.1
5.3.2

Determining necessary sensor trims

Bench calibrations allow for calibrating the instrument for its desired range of operation. Straightforward connections to pressure source allow for a full calibration at the planned operating points. Exercising the device over the desired pressure range allows for verification of the output value. Sensor trim discusses how the trim operations change the calibration. It is possible to degrade the performance of the device if a trim is done improperly or with inaccurate equipment. The device can be set back to factory settings using the Recall factory trim — sensor trim command in Manifold operation.
For devices that are field installed, the manifolds discussed in Manifold operation allow the device to be zeroed using the zero trim function. This field calibration will eliminate any pressure offsets caused by mounting effects (head effect of the oil fill) and static pressure effects of the process.
Determine the necessary trims with the following steps.
Procedure
1. Apply pressure.
2. Check pressure. If the pressure does not match the applied pressure, perform a digital zero trim. See Manifold operation.

Sensor trim overview

A sensor trim corrects the pressure offset and pressure range to match a pressure standard. The upper sensor trim corrects the pressure range and the lower sensor trim (zero trim) corrects the pressure offset. An accurate pressure standard is required for full calibration. A zero trim can be performed if the process is vented.
Zero trim is a single-point offset adjustment. It is useful for compensating for mounting position effects and is most effective when performed with the device installed in its final mounting position. Since this correction maintains the slope of the characterization curve, it should not be used in place of a sensor trim over the full sensor range.
34 Emerson.com/Rosemount
Reference Manual Operation and maintenance
00809-0100-4045 September 2019
When performing a zero trim, ensure the equalizing valve is open and all wet legs are filled to the correct levels. Line pressure should be applied to the device during a zero trim to eliminate line pressure errors. Refer to Manifold operation.
Note
Do not perform a zero trim on the Rosemount Smart Pressure Gauge with absolute measurement type. Zero trim uses a zero reference against ambient air pressure for gage, vacuum, and compound pressure devices, while absolute pressure devices reference absolute zero. To correct mounting position effects on a Smart Pressure Gauge with absolute measurement type, perform a low trim within the sensor trim function. The low trim function provides an offset correction similar to the zero trim function, but it does not require zero-based input.
Sensor trim is a two-point sensor calibration where two end-point pressures are applied, and output is linearized. Always adjust the low trim value first to establish the correct offset. Adjustment of the high trim value provides a slope correction to the characterization curve based on the low trim value. The trim values allow you to optimize performance over your specified measuring range at the calibration temperature.
During a trim operation, the device is placed in high power refresh mode, which provides frequent pressure measurement updates. This behavior allows for more accurate calibration of the device. When the device is in high power refresh mode, the battery power supply will be depleted more rapidly.
Figure 5-1: Sensor Trim Example
A
B

5.3.3 Sensor trim

When performing a sensor trim, both the upper and lower limits can be trimmed. If both upper and lower trims are to be performed, the lower trim must be done before the upper trim.
Note
Use a pressure input source at least four times more accurate than the device, and allow the input pressure to stabilize for 60 seconds before entering any values.
Before trim
After trim
From the HOME screen, enter the Fast Key sequence
Emerson.com/Rosemount 35
Operation and maintenance Reference Manual
September 2019 00809-0100-4045
Fast Keys 2, 2, 1, 1, 1
Procedure
1. Assemble and power the entire calibration system including the gauge, Field Communicator or AMS Device Manager, power supply, pressure input source, and readout device.
2. From the Home screen, select 2: Configure.
3. Select 2: Manual Setup.
4. Select 1: Measurements.
5. Select 1: Dial/Pressure.
6. Select 1: Verify/Calibrate.
Note
Select pressure points so that lower and upper values are equal to or outside the expected process operation range.
Note
The applied pressure must be within five percent of the selected pressure points when performing sensor trim.
5.3.4
7. Follow the on-screen instructions to complete the adjustment of the lower value.
8. Repeat the procedure for the upper value.
Performing a digital zero trim
Devices are factory-calibrated. Once installed, it is recommended to perform this step to eliminate potential error caused by mounting position or static pressure. Instructions for using a Field Communicator are listed below.
Fast Keys
1. Vent the device.
2. Connect the Field Communicator.
3. From the HOME screen, enter the HART Fast Key sequence.
4. Follow the commands to perform the procedure.
2, 1, 1

Dial adjustment

Dial adjustment can be used to adjust the dial above or below zero and allows for adjustments up to 13 percent of span. Dial adjustment only impacts needle position and does not impact sensor.
Note
Dial adjustment adjusts the position of the factory dial calibration. It is possible to degrade the performance of the gauge if the operation is done improperly or inaccurately.
From the HOME screen, enter the Fast Key sequence
36 Emerson.com/Rosemount
Reference Manual Operation and maintenance
00809-0100-4045 September 2019
Fast Keys 2, 2, 1, 1, 1, 1
Procedure
1. Select 2: Configure.
2. Select 2: Manual Setup.
3. Select 1: Measurements.
4. Select 1: Dial/Pressure.
5. Select 1: Verify/Calibrate.
6. Select 1: Verify/Calibrate Dial+Digital Pressure.
7. Adjust dial indicator until it points to lower endpoint.
The following adjustments are available and can be used to complete the dial adjustment.
Fine counterclockwise (0.1 percent of Span)
Fine clockwise (0.1 percent of Span)
Coarse counterclockwise (0.3 percent of Span)
5.3.5
Coarse clockwise (0.3 percent of Span)
8. Select 5: Save Dial.

Recall factory trim — sensor trim

The recall factory trim—sensor trim command allows the restoration of the as-shipped factory settings of the sensor trim. This command can be useful for recovering from an inadvertent zero trim of an absolute pressure unit or inaccurate pressure source.
Fast Keys
1. From the HOME screen, enter the Fast Key sequence
2. Select 3: Service Tools.
3. Select 4: Restore to Default Settings.
4. Select 2: Restore to Default Settings.
5. Follow the screen prompts to recall sensor and dial trim.
3, 4, 2
Emerson.com/Rosemount 37
Operation and maintenance Reference Manual
September 2019 00809-0100-4045

5.4 Replacing the battery

WARNING
The Rosemount Pressure Gauge shall be used only with the battery (00G45-9000-0001) supplied by Rosemount. This battery has been officially tested with the device as required by the I.S. standards during the assessment of the Rosemount Pressure Gauge.
The battery is not replaceable in a hazardous location.
Dispose of battery in accordance with local and national jurisdictions.
Procedure to replace the battery:
Procedure
1. Remove enclosure cover.
2. Switch the device “OFF”.
3. Loosen the screw holding the electronics assembly to the enclosure.
Note
Use caution as the electronics assembly is connected to the pressure sensor via a cable. Take care not to over stretch this cable as this could damage the device.
4. Release battery connection from electronics board.
5. Loosen the two screws on the battery holder and slide the battery holder to the left.
Note
The screws holding down the electronics board do not need to be removed, just loosened. Take care not to let the battery fall out of the enclosure.
6. Remove battery from enclosure.
7. Installation of new battery is the reverse of the removal.

5.5 Local device status and notifications

The flashing LED indicates device status using the colors described in Table 5-2. For start up considerations, refer to Turn on device.
Table 5-2: Status Descriptions
LED color Device status
Green Functioning properly
Amber Battery is low, battery replacement recommended
38 Emerson.com/Rosemount
Reference Manual Operation and maintenance
00809-0100-4045 September 2019
Table 5-2: Status Descriptions (continued)
LED color Device status
Red Battery replacement required OR Device is
malfunctioning
No color No power, verify ON/OFF switch is in “on” position
If the dial is pointing towards the red “X”, refer to Troubleshooting for more information.
Emerson.com/Rosemount 39
Operation and maintenance Reference Manual
September 2019 00809-0100-4045
40 Emerson.com/Rosemount
Reference Manual Troubleshooting
00809-0100-4045 September 2019

6 Troubleshooting

6.1 Service support

To expedite the return process outside of the United States, contact the nearest Emerson representative.
Contact information for a regional Rosemount™ office is provided on the last page of this document.
The center will ask for product model and serial numbers, and will provide a Return Material Authorization (RMA) number. The center will also ask for the process material to which the product was last exposed.
CAUTION
Individuals who handle products exposed to a hazardous substance can avoid injury if they are informed of and understand the hazard. The product being returned will require a copy of the required Safety Data Sheet (SDS) for each substance must be included with the returned goods.
Emerson representatives will explain the additional information and procedures necessary to return goods exposed to hazardous substances.

6.2 Local troubleshooting

Table 6-1: Interpreting Local Notifications
LED color Dial location Device status Recommended action(s)
Green
Amber
Red
Functioning properly No action required.
Battery is low Battery replacement recommended.
Battery is low, device is malfunctioning
Battery replacement required
OR Device is
malfunctioning
Investigate active notification via a HART Communicator. Replace battery if device is determined to be functioning properly and notifications have been verified.
Investigate active notification via a HART Communicator. Replace battery if device is determined to be functioning properly and notifications have been verified.
®
Emerson.com/Rosemount 41
Troubleshooting Reference Manual
September 2019 00809-0100-4045
Table 6-1: Interpreting Local Notifications (continued)
LED color Dial location Device status Recommended action(s)
Black, no color N/A No power Verify ON/OFF switch is in “ON” position.

6.3 Remote troubleshooting

Table 6-2: Interpreting Plantweb™ Statuses
Plantweb status
Good None Functioning properly No action required
Advisory
Notification Description Recommended action(s)
Database Storage Error
High Power Active
Simulate Active The device is in simulation
Non-Critical User Data
The device has failed to write to the database memory at some point in the past. Any data written during this time nay have been lost.
The device is operating in high power mode. This is not recommended for this device.
mode and may not be reporting actual informaiton.
A user written parameter does not match its expected value.
1. Perform a Device Reset.
2. If logging dynamic data is not needed, this advisory alert can be safely ignored.
3. If the condition persists, replace the device.
1. Disable high power mode.
1. Verify that simulation is no longer required.
2. Disable simulation mode.
3. Reset the device.
1. Restart the device.
2. Reconfirm all configuration items in the device.
3. Restore the default settings and reconfigure device.
4. If the condition persists, replace the device.
Sensor Temperature Out of Limits
Maintenance
42 Emerson.com/Rosemount
The sensor temperature has exceeded its safe operating range.
1. Verify process and ambient temperature is within the device’s operating range.
2. Remote mount the device away from process and environmental conditions.
3. Reset the device.
4. If the condition persists, replace the device.
Reference Manual Troubleshooting
00809-0100-4045 September 2019
Table 6-2: Interpreting Plantweb™ Statuses (continued)
Plantweb status
Notification Description Recommended action(s)
Pressure Out of Limits
The pressure has exceed the maximum measurement range.
1. Check the pressure applied to ensure it is within the sensor limits.
2. Check the device pressure connection to make sure it is not plugged or that the isolating diaphragms are not damaged.
3. If the condition persists, replace the device.
Voltage Conditions Out of Range
Environmental Conditions Out of Range
The supply voltage is low and may soon affect device operation.
The device is outside its normal environmental operating conditions which may affect accuracy and/or proper operation.
1. Replace the battery.
1. Verify process and ambient temperature is within the device’s operating range.
2. Remote mount the device away from process and environmental conditions.
3. Reset the device.
4. If the condition persists, replace the device.
Failure
Capacity Denied
Configuration Alert
The device has failed to acquire the wirless communication bandwidth necessary to support the configured update rates.
The device has detected a configuration error.
1. Obtaining the bandwidth may take some time depending on the configured update rates and other devices in the network. Wait several minutes to see if the error resolves itself.
2. There may be too many devices attached to the WirelessHART
®
network, or the updates rates may be too fast. Try using a different network, or slowing down the update rate on one or more devices.
1. Click on details for more information.
2. Correct the parameter that has a configuration error.
3. Reset the device.
4. If the condition persists, replace the device.
Emerson.com/Rosemount 43
Troubleshooting Reference Manual
September 2019 00809-0100-4045
Table 6-2: Interpreting Plantweb™ Statuses (continued)
Plantweb status
Notification Description Recommended action(s)
Over-pressure Seen
Critical Power Failure
Electronics Failure
Dial Failure The device is no longer able to
The pressure has gone beyond the maximum operating limits of the device, which may have caused permanent damage to the sensor.
The supply voltage is too low for the device to update.
An electronics error that could impact the device measurement reading has occurred.
validate the position of the dial.
1. Check the pressure applied to ensure it is within the sensor limits.
2. Check the device pressure connection to make sure it is not plugged or that the isolating diaphragms are not damaged.
3. Acknowledge the over pressure condition to clear the latched indication, and verify the integrity of the sensor.
4. If the condition persists, replace the device.
1. Replace the battery.
1. Restore device to default settings.
2. Perform a Device Reset.
3. If the condition persists, replace the device.
1. Reset the device.
2. If condition persists, replace the device.
44 Emerson.com/Rosemount
Reference Manual Reference data
00809-0100-4045 September 2019

7 Reference data

7.1 Product certificatons

To view current Rosemount™ Wireless Pressure Gauge Product Certifications, follow these steps:
Procedure
1. Go to Emerson.com/Rosemount/Rosemount-Wireless-Pressure-Gauge.
2. Scroll as needed to the green menu bar and select Documents & Drawings.
3. Click Manuals & Guides.
4. Select the appropriate Quick Start Guide.

7.2 Ordering information, specifications, and dimensional drawings

To view current Rosemount Wireless Pressure Gauge ordering information, s, and dimensional drawings, follow these steps:
Procedure
1. Go to Emerson.com/Rosemount/Rosemount-Wireless-Pressure-Gauge.
2. Scroll as needed to the green menu bar and click Documents & Drawings.
3. For installation drawings, select Drawings & Schematics and select the appropriate document.
4. For ordering information, specifications, and dimensional drawings, select Data Sheets & Bulletins.
5. Select the appropriate Product Data Sheet.
Emerson.com/Rosemount 45
Reference data Reference Manual
September 2019 00809-0100-4045
46 Emerson.com/Rosemount
Reference Manual Field Communicator menu trees
00809-0100-4045 September 2019
A Field Communicator menu trees

A.1 Overview

Figure A-1: Overview
Emerson.com/Rosemount 47
Field Communicator menu trees Reference Manual
September 2019 00809-0100-4045
Figure A-2: Configure
Figure A-3: Service Tools
48 Emerson.com/Rosemount
Reference Manual Field Communicator menu trees
00809-0100-4045 September 2019
Figure A-4: Device Information
Emerson.com/Rosemount 49
Field Communicator menu trees Reference Manual
September 2019 00809-0100-4045
Figure A-5: Device Information (continue)
50 Emerson.com/Rosemount
Reference Manual Network design best practices
00809-0100-4045 September 2019
B Network design best practices

B.1 Overview

All recommended practices should be followed to ensure highest data reliability. Deviation from these best practices may require device repeaters in the network to maintain 99 percent data reliability. The following are guidelines to achieve the best possible wireless network.
Each wireless network field should be scoped to a single process unit.
Minimize the number of hops to the Gateway in order to reduce latency. A minimum of
five wireless instruments should be within effective range of the Emerson Wireless Gateway.
Each device in the network should have at minimum three devices with potential
communication paths. A mesh network gets its reliability from multiple communication pathways. Ensuring each device has multiple neighbors within range will result in the most reliable network.
Have 25 percent of wireless instruments in the network within range of Emerson
Wireless Gateway. Other enhancing modifications include creating a higher percentage of devices within effective range of the gateway to 35 percent or more. This clusters more devices around the gateway and ensures fewer hops and more bandwidth available to WirelessHART® devices with fast scan rates.
Effective range is determined by type of process unit and the density of the
infrastructure that surrounds the network.

B.2 Effective range

Heavy Obstruction: 100 ft. (30 m). Typical heavy density plant environment. Cannot drive a truck or equipment through.
Medium Obstruction: 250 ft. (76 m). Typical light process areas, lots of space between equipment and infrastructure.
Light Obstruction: 500 ft. (152 m). Typical of tank farms. Despite tanks being big obstructions themselves, lots of space between and above makes for good RF propagation.
Line of Sight: 750 ft. (230 m). No obstructions between WirelessHART® devices and devices mounted a minimum of 6 ft. (2 m) above ground or obstructions.
For examples and complete explanations, refer to the IEC62591 WirelessHART System
Engineering Guide.
Emerson.com/Rosemount 51
Network design best practices Reference Manual
September 2019 00809-0100-4045
52 Emerson.com/Rosemount
Reference Manual Device variable index
00809-0100-4045 September 2019
C Device variable index

C.1 Overview

This section outlines the most important alerts in the HART® command 48 additional status field for Rosemount™ Wireless Pressure Gauge. The information in this section can be used by DeltaV™ for notification monitoring, and in the Emerson Wireless Gateways for additional status mapping in Modbus®, OPC, etc.

C.2 Messages and descriptions

A complete list of additional status bits is available in the Gateway.
Table C-1: Device Variable
Device Variable Index
0 Supply voltage
1 Pressure
2 Custom
3 Sensor temperature
Table C-2: Failures
Message Additional status
Radio failure Byte 12 :: Bit 4 Wireless radio has detected a
Configuration alert Byte 2 :: Bit 6 Device has detected a
Over-pressure seen Byte 4 :: Bit 0 Pressure has gone beyond the
Critical power failure Byte 6 :: Bit 2 Supply voltage is too low for
Electronics failure Byte 8 :: Bit 6 Electronics error that could
(1)
Description
failure or stopped communicating.
configuration error.
maximum operating limits of the device, which may have caused permanent damage to the sensor.
the device to broadcast updates.
impact the device measurement reading has occurred.
Dial failure Byte 4 :: Bit 4 Device has detected a failure or
unable to confirm dial location.
(1) Location of the alert in the HART command 48 Additional Status field.
Emerson.com/Rosemount 53
Device variable index Reference Manual
September 2019 00809-0100-4045
Table C-3: Maintenance
Message Additional status
Voltage condition out of range Byte 8 :: Bit 4 Supply voltage is low and may
Pressure out of limits Byte 3 :: Bit 5 Pressure has exceeded the
Capacity denied Byte 12 :: Bit 0 Device has failed to acquire the
(1)
Description
soon affect device operation.
maximum measurement range.
wireless communication bandwidth necessary to support the configured update rates.
Sensor temperature beyond sensor limits
Byte 3 :: Bit 1 Sensor temperature has
exceeded its safe operating range.
Environmental conditions out of range
Byte 8 :: Bit 5 Device is outside its normal
environmental operating conditions which may affect accuracy and/or proper operation.
(1) Location of the alert in the HART command 48 Additional Status field.
Table C-4: Advisory
Message Additional status
Database storage error Byte 0 :: Bit 2 Device has failed to write to the
Simulation active Byte 8 :: Bit 0 Device is in simulation mode
High power active Byte 1 :: Bit 7 Device is operating in a high
(1)
Description
database memory at some point in the past; any data written during this time may have been lost.
and is not reporting actual information.
power mode; this is not recommended for this device.
Non-critical user data warning Byte 2 :: Bit 1 User-written parameter does
not match its expected value.
(1) Location of the alert in the HART command 48 Additional Status field.
54 Emerson.com/Rosemount
Reference Manual
00809-0100-4045 September 2019
Emerson.com/Rosemount 55
00809-0100-4045
Rev. AC
2019
Global Headquarters
Emerson Automation Solutions 6021 Innovation Blvd. Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 204 8889 RFQ.RMD-RCC@Emerson.com
Latin America Regional Office
Emerson Automation Solutions 1300 Concord Terrace, Suite 400 Sunrise, FL 33323, USA
+1 954 846 5030 +1 954 846 5121 RFQ.RMD-RCC@Emerson.com
Asia Pacific Regional Office
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461
+65 6777 8211 +65 6777 0947 Enquiries@AP.Emerson.com
North America Regional Office
Emerson Automation Solutions 8200 Market Blvd. Chanhassen, MN 55317, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 204 8889 RMT-NA.RCCRFQ@Emerson.com
Europe Regional Office
Emerson Automation Solutions Europe GmbH Neuhofstrasse 19a P.O. Box 1046 CH 6340 Baar Switzerland
+41 (0) 41 768 6111 +41 (0) 41 768 6300 RFQ.RMD-RCC@Emerson.com
Middle East and Africa Regional Office
Emerson Automation Solutions Emerson FZE P.O. Box 17033 Jebel Ali Free Zone - South 2 Dubai, United Arab Emirates
+971 4 8118100 +971 4 8865465 RFQ.RMTMEA@Emerson.com
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Youtube.com/user/RosemountMeasurement
©
2019 Emerson. All rights reserved.
Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
Loading...