Rosemount Tank Gauging System Manuals & Guides

Page 1
Rosemount™ Tank Gauging System Configuration
System configuration
Reference Manual
00809-0300-5100, Rev CB
February 2022
Page 2
Rosemount™ Tank Gauging
NOTICE
Read this manual before working with the product. For personal and system safety, and for optimum product performance, ensure you thoroughly understand the contents before installing, using, or maintaining this product.
For equipment service or support needs, contact your local Emerson Automation Solutions/Rosemount Tank Gauging representative.
Version
This manual is based on the functionality of TankMaster WinSetup version 6.x. For older TankMaster versions all functionality described in this manual may not be available and the Graphical User Interface (GUI) may look different.
2
Page 3
Reference Manual Contents
00809-0300-5100
February 2022

Contents

Chapter 1 Introduction.............................................................................................................. 5
1.1 Manual overview..........................................................................................................................5
1.2 Technical documentation............................................................................................................ 6
1.3 Document structure.................................................................................................................... 8
Chapter 2 Device installation..................................................................................................... 9
2.1 System configuration overview....................................................................................................9
2.2 Communication protocol setup.................................................................................................14
2.3 Preferences................................................................................................................................28
2.4 Field device installation overview...............................................................................................40
2.5 Installing a Rosemount 2460 System Hub..................................................................................41
2.6 Installing a Rosemount 2410 Tank Hub......................................................................................42
2.7 Installing a Rosemount 5900 Radar Level Gauge........................................................................56
2.8 Installing the Rosemount 5900S 2-in-1 version.......................................................................... 72
2.9 Installing Auxiliary Tank Devices................................................................................................ 88
2.10 Installing a Rosemount 5408................................................................................................. 105
2.11 Installing a Rosemount 5300................................................................................................. 116
Chapter 3 Tank installation.................................................................................................... 127
3.1 Installing a tank....................................................................................................................... 127
3.2 Adding a tank.......................................................................................................................... 139
3.3 Calibrating the level gauge...................................................................................................... 150
3.4 Setting up a Tank Capacity Table............................................................................................. 152
3.5 Tank Entry................................................................................................................................154
3.6 Setting up a hybrid system.......................................................................................................155
Chapter 4 Device handling..................................................................................................... 163
4.1 To change device configuration...............................................................................................163
4.2 To uninstall a device.................................................................................................................165
4.3 To uninstall a tank and associated devices............................................................................... 168
Rosemount Tank Gauging 3
Page 4
Contents Reference Manual
February 2022 00809-0300-5100
4 Emerson.com/Rosemount
Page 5
Reference Manual Introduction
00809-0300-5100 February 2022

1 Introduction

1.1 Manual overview

This manual provides information on installation, configuration and maintenance of the Rosemount™ 2410 Tank Hub.
Chapter Introduction provides a brief description of the various components in a Rosemount Tank Gauging system.
Chapter Device installation is a description of how to use TankMaster WinSetup to install devices in a Rosemount Tank Gauging system.
Chapter Tank installation is a description of how to use TankMaster WinSetup to install tanks and associate devices to tanks in a Rosemount Tank Gauging system.
Chapter Device handling is a short description of the basic functions for changing device configuration and how to uninstall devices from the WinSetup workspace.
Rosemount Tank Gauging 5
Page 6
Introduction Reference Manual
February 2022 00809-0300-5100

1.2 Technical documentation

The Rosemount™ Tank Gauging System includes a wide portfolio of user documentation. For a complete list, see product pages on Emerson.com/Rosemount.
Reference manuals
Rosemount Tank Gauging System Configuration Manual (00809-0300-5100)
Rosemount 2460 System Hub (00809-0100-2460)
Rosemount 2410 Tank Hub (00809-0100-2410)
Rosemount 5900S Radar Level Gauge (00809-0100-5900)
Rosemount 5900 Proof Test with Reference Reflector (00809-0200-5900)
Rosemount 5900C Radar Level Gauge (00809-0100-5901)
Rosemount 2240S Multi-Input Temperature Transmitter (00809-0100-2240)
Rosemount 2230 Graphical Field Display (00809-0100-2230)
Rosemount 5300 Guided Wave Radar (00809-0100-4530)
Rosemount 5408 Radar Level Transmitter (00809-0300-4408)
Rosemount Tank Gauging Wireless System (00809-0100-5200)
Rosemount TankMaster WinOpi (00809-0200-5110)
Rosemount TankMaster Software Installation Manual (00809-0400-5110)
Rosemount TankMaster WinSetup (00809-0100-5110)
Rosemount TankMaster Floating Roof Monitoring (00809-0500-5100)
Rosemount TankMaster Network Configuration (303042EN)
Rosemount 5900 Radar Level Gauge and Rosemount 2410 Tank Hub Safety Manual
Option S (00809-0400-5100)
Rosemount TankMaster Mobile User Guide (00809-0100-5120)
Rosemount TankMaster Mobile Installation Manual (00809-0200-5120)
6 Emerson.com/Rosemount
Page 7
Reference Manual Introduction
00809-0300-5100 February 2022
Product data sheets
Rosemount Tank Gauging System (00813-0100-5100)
Rosemount TankMaster Inventory Management Software (00813-0100-5110)
Rosemount TankMaster Mobile Inventory Management Software (00813-0100-5120)
Rosemount 2460 System Hub (00813-0100-2460)
Rosemount 2410 Tank Hub (00813-0100-2410)
Rosemount 5900S Radar Level Gauge (00813-0100-5900)
Rosemount 5900C Radar Level Gauge (00813-0100-5901)
Rosemount 2240S Multi-input Temperature Transmitter (00813-0100-2240)
Rosemount 565/566/765/614 Temperature and Water Level Sensors
(00813-0100-5565)
Rosemount 2230 Graphical Field Display (00813-0100-2230)
Rosemount 5300 Level Transmitter (00813-0100-4530)
Rosemount 5408 Level Transmitter (00813-0100-4408)
Rosemount Tank Gauging 7
Page 8
A B C
D
E
F
G H
I
J
Introduction Reference Manual
February 2022 00809-0300-5100

1.3 Document structure

Figure 1-1: System and User Documentation Structure
A. Rosemount TankMaster WinSetup Reference Manual
B. Rosemount TankMaster WinOpi Reference Manual
C.
Rosemount TankMaster Software Installation Manual
D. Rosemount 2460 System Hub Reference Manual
E. Rosemount Tank Gauging System Configuration Manual F. Rosemount 2410 Reference Manual
G. Rosemount 5900S Reference Manual
8 Emerson.com/Rosemount
H. Rosemount 5900C Reference Manual
I. Rosemount 2240S Reference Manual J. Rosemount 2230 Reference Manual
Page 9
Reference Manual Device installation
00809-0300-5100 February 2022

2 Device installation

This section describes how to install and configure a Rosemount Tank Gauging system by using the Rosemount TankMaster WinSetup configuration program.

2.1 System configuration overview

2.1.1 Preparations

Before installing a Rosemount Tank Gauging system you should ensure that the following information is available:
A plan of all field devices and tanks.
Unit IDs of each device (Unit ID is a unique identifier given to each device at factory).
Modbus® addresses of level devices and ATD devices. The devices are shipped with
default addresses which will be changed at system configuration. The Modbus addresses are configured in the Rosemount 2460 System Hub’s Tank Database as well as the Rosemount 2410 Tank Hub’s Tank Database as described below.
Tank geometry parameters and reference distances such as tank Reference height (R)
and distance between Zero level (datum plate) and tank bottom.
Antenna type for the various level gauges.

2.1.2 Installation procedure

Installation and configuration of a Rosemount Tank Gauging system includes the following steps as briefly described below.
Procedure
1. Setup communication protocols.
Specify communication protocol parameters:
The Modbus® Master Protocol handles communication between a TankMaster
work station and field devices such as the Rosemount 2460 System Hub and the Rosemount 2410 Tank Hub.
The Slave Protocol handles communication with a host computer.
Communication with TankMaster can be supervised by logging various error
types and function codes.
2.
Configure preferences. Specify measurement units, tag prefixes for tank and device labels, inventory
parameters, and parameters to be displayed when viewing tank data.
Rosemount Tank Gauging 9
Page 10
Device installation Reference Manual
February 2022 00809-0300-5100
3. Install and configure the Rosemount 2460 System Hub.
The Rosemount 2460 System Hub has to be installed and configured prior to installing other devices such as a Rosemount 2410 Tank Hub and a Rosemount 5900S Radar Level Gauge.
a) Assign a Modbus communication address
b)
For each communication port, configure protocol and appropriate communication parameters
c) Configure the Tank Database with information about the devices connected
to the fieldbus
4. Install and configure Rosemount 2410 Tank Hubs. The Rosemount 2410 Tank Hub should be installed after the Rosemount 2460
System Hub and prior to the other field devices. In case no system hub is used, the tank hub can be connected directly to a TankMaster work station. Installing a Rosemount 2410 Tank Hub includes the following main steps:
a) Specify a device tag
b) Assign a Modbus communication address
c) Configure the Tank Database to map devices to tanks
d) Optional: Configure the local display
5. Install and configure field devices In a Rosemount Tank Gauging system, the field devices such as level gauges and
temperature transmitters, are installed in TankMaster WinSetup as part of the Rosemount 2410 installation procedure. The devices are configured at a later stage by using the Properties window for each device.
Installation and configuration of devices include the following steps:
a) Communication: specify protocol and address.
b) Configuration: specify tank geometry parameters, device specific
parameters, temperature element positions, and other parameters depending on the device type.
6. Install and configure tanks. Installing a tank includes the following steps:
a) Select tank type. Select one of the available options such as Fixed Roof,
Floating Roof, Sphere LPG, Horizontal LPG, or any other suitable tank type.
b) Specify a tank tag. The name will be used as an identifier in the Workspace
window and other TankMaster windows.
c) Select devices. Associate devices to the tank.
d) Configuration. Specify the available source signals for parameters such as
Free Water Level, Vapor Pressure and Liquid Pressure.
e) Value Entry. Specify an approved value range for Level, Ullage, and Free
Water Level. In case there is no source instrument available, you can specify manual values to be used instead.
10 Emerson.com/Rosemount
Page 11
Reference Manual Device installation
00809-0300-5100 February 2022
7. Calibrate. Once a Rosemount 5900 Radar Level Gauge is installed and configured, the
Calibration Distance parameter may have to be adjusted in order to ensure that measured level and actual product level match. The adjustment should be performed once at the final commissioning.
Rosemount Tank Gauging 11
Page 12
A
B
C
D
E
F
G
TankMaster
Rosemount 2460
Rosemount 2410
Rosemount 5900 Rosemount 2240S
Rosemount 2230
Device installation Reference Manual
February 2022 00809-0300-5100
Installation procedure
Figure 2-1: Rosemount Tank Gauging system installation Procedure
A. Communication setup
B. Preferences
Rosemount 2460 System Hub installation and configuration
C. D. Rosemount 2410 Tank Hub installation and configuration
E. Device installation and configuration F. Tank installation and configuration
G. Calibration
12 Emerson.com/Rosemount
Page 13
Reference Manual Device installation
00809-0300-5100 February 2022

2.1.3 Using the device installation wizard

The device installation wizard guides you step-by-step through the installation procedure.
Procedure
1. In the TankMaster WinSetup workspace, select the
2. Do one of the following:
Click the right mouse button and select Install New from the popup menu.
From the Service menu select Devices Install New.
Devices folder.
Starting device installation wizard from File menu
As an alternative method to start the device installation wizard you can use the following method.
Procedure
In the TankMaster WinSetup workspace, select the server where your system is
1. installed.
2. From the File menu select Install New Device.
Rosemount Tank Gauging 13
Page 14
Device installation Reference Manual
February 2022 00809-0300-5100

2.2 Communication protocol setup

The TankMaster work station can be connected to field devices and host computers by using Master and Slave protocols. The Modbus Master protocol is available as default protocol when the Rosemount TankMaster software is installed on a TankMaster work station. Optional protocols, such as the Modbus® Slave protocol for communication with host systems, can be obtained as well. Please contact Emerson Automation Solutions/ Rosemount Tank Gauging for more information.
A Modbus protocol offers up to eight channels. Enraf and HART protocols support 16 channels. For each channel you can specify various settings such as communication type (for example Modbus TCP or Modbus RTU) and which PC communication port (USB/COM) to connect to, as well as standard communication parameters such as Baud Rate, Parity, and number of Stop Bits.
14 Emerson.com/Rosemount
Page 15
A
B
C
D
ED
H
G
F
I
Rosemount 2410
Rosemount 2460
Rosemount TankMaster
Emerson Wireless 775 THUM Adapter
Emerson Wireless Gateway
Rosemount 2230
Rosemount 5900
Rosemount 2240S
Reference Manual Device installation
00809-0300-5100 February 2022
Figure 2-2: Communication with Various Rosemount Tank Gauging Devices
A. Master protocol (OPC, Modbus RTU)
B. Slave protocol
To host computer (PLC, SCADA, DCS)
C. D. Modbus TCP (Ethernet)
E. Modbus RTU (RS232/RS485)
F. Modbus RTU (TRL2/RS485) G. Tankbus H. HART TCP (Ethernet)
I. Host computer
Rosemount Tank Gauging 15
Page 16
Device installation Reference Manual
February 2022 00809-0300-5100

2.2.1 Master protocol channel configuration

This section describes how to configure the Modbus® Master protocol channel by using Rosemount TankMaster WinSetup.
Procedure
1. Open the
2. Select the icon that corresponds to the particular protocol to be configured (this
example shows the Modbus Master protocol).
3. Click the right mouse button and select Properties, or from the Service menu select
Protocols Properties.
Protocols folder in the Workspace window.
4. The Protocol Properties window lists the available protocol channels. For each
channel the corresponding icon indicates whether the channel is enabled or disabled.
5. Select the desired channel.
16 Emerson.com/Rosemount
Page 17
Reference Manual Device installation
00809-0300-5100 February 2022
6. Click the Properties button to configure the protocol channel.
7. Select the Communication tab. It allows you to configure parameters for
communication between field devices and TankMaster work station.
Modbus TCP; specify IP address for the Rosemount 2460 System Hub that
a)
the TankMaster server is connected to. For redundant system hubs, you may use a separate IP address for the backup server.
b)
Modbus RTU; click the Advanced button in case you need to configure the communication parameters.
Note
If the communication is interrupted and handshaking includes DSR, no query will be sent from the TankMaster Protocol Server. This may result in a Query Timeout.
8. The Communication disabled in backup mode check box can be used for systems
with redundant tank servers. If the check box is selected, the ModbusMaster will not send any queries if the local tank server is in backup mode.
9. Select the
Enable Channel check box to activate the protocol channel.
10. The File Log tab lets you specify what type of information to be logged and saved to disk.
Rosemount Tank Gauging 17
Page 18
Device installation Reference Manual
February 2022 00809-0300-5100
11. Click the OK button to store the current configuration and close the configuration window.
The Modbus Master Channel icon (channel no. 1 in this example) appears in the WinSetup workspace:
Related information
Default Modbus RTU communication parameters Rosemount 2460 Reference Manual
18 Emerson.com/Rosemount
Page 19
Reference Manual Device installation
00809-0300-5100 February 2022
Default Modbus RTU communication parameters
Table 2-1: Default Communication Parameters for Modbus RTU
Parameter Description
Port The COM port that the Rosemount 2180 will be connected to
Baud rate 4800
Stop bits 1
Parity None
Modem Choose FBM 2180 for the Rosemount 2180 Field Bus Modem
Available options:
FBM 2180
FBM 2170/71
RS232
RS485
Handshaking FBM 2180: None
FBM 2170/2171: RTS/CTS/DTS/DSR
RS485: RTS/CTS
RS232: None
Reply timeout 1000 ms
Retries 10
Description Text that describes the configured channel
Related information
Master protocol channel configuration
Rosemount Tank Gauging 19
Page 20
Device installation Reference Manual
February 2022 00809-0300-5100

2.2.2 Slave protocol channel configuration

A Slave protocol allows you to collect data from the TankMaster workstation to a host computer.
Modbus® communication setup
Prerequisites
A hardware key must be installed in order to run a slave protocol server. Host communication needs to be enabled.
Procedure
1. Open the
2. Select the ModbusSlave protocol icon.
3. Click the right mouse button and select Properties, or select Service Protocols
Properties.
Protocols folder in the Workspace window.
4. The Protocol Properties window lists enabled and disabled protocol channels.
20 Emerson.com/Rosemount
Page 21
Reference Manual Device installation
00809-0300-5100
February 2022
5. Select the desired protocol channel and click the Properties button to open the configuration window. .
6. Select the Communication tab.
a) Modbus TCP; verify correct port.
b)
Modbus RTU; click the Advanced button in case you need to configure parameters for proper communication with the host.
Note
If the communication is interrupted and handshaking includes DSR, no query will be sent from the TankMaster Protocol Server. This may result in a Query Timeout.
7. Select the Enable Channel check box to activate the protocol channel.
8. Verify communication address and enter optional description:
Address Set the Modbus address to be used by the host computer to
identify the TankMaster workstation.
Description Optional text to describe the configured channel.
Backup mode
None
Write Commands Rejected means that TankMaster does not
accept any write commands from the host system to device database registers
Rosemount Tank Gauging 21
Page 22
Device installation Reference Manual
February 2022
00809-0300-5100
In Silent mode the Modbus Slave protocol will not send any
replies to requests coming from the host computer while the local tankserver is in backup mode
Note
If the communication is interrupted and handshaking includes DSR, no query will be sent from the TankMaster Protocol Server. This may result in a Query Timeout.
9. Click the OK button to store the current configuration and close the configuration window.
Related information
Advanced Modbus RTU slave protocol configuration
22 Emerson.com/Rosemount
Page 23
Reference Manual Device installation
00809-0300-5100
February 2022
Advanced Modbus® RTU slave protocol configuration
The Advanced Communication window lets you configure modem communication parameters, delay times, and time-outs for the Modbus RTU Slave protocol.
Procedure
1. In the
2. The following default values are used for the TRL2 Modbus Slave protocol:
Slave Protocol Channel Configuration window, select the Communication tab
and click the Advanced button:
Modem Choose the appropriate interface
Baud rate Enter appropriate value for communication with the
host
Stop bits Enter appropriate value for communication with the
host
Parity Enter appropriate value for communication with the
host
Handshaking
Query to Response Delay 10 ms
Query interval 100 ms
Read Query Timeout 400 ms
Write Response Timeout 400 ms
Max. Response Time 800 ms
Rosemount Tank Gauging 23
FBM 2180: None
RS485, RS232: See specifications for the
communication software used on the host system
Page 24
Device installation Reference Manual
February 2022
00809-0300-5100
Tank mapping configuration
The slave protocol allows you to send data from a Rosemount Tank Gauging system to a host computer. In the Tank Mapping window you can specify from which tanks to collect data for the host system.
Procedure
1. In the
2. From the list of tanks that appear in the Available Tanks pane, select the tanks that
3. Click the (right arrow) button to move the selected tanks to the list of mapped
Slave Protocol Channel Configuration window, select the Tank Mapping tab.
the host will connect to.
tanks.
Important
Ensure that the tanks appear in the order required by the host system.
When the host sends a query, TankMaster responds by sending tank data in the order as the tanks are listed in the Mapped Tanks column. You can easily change the position of mapped tanks by using the up arrow and down arrow buttons.
4. Click the OK button to save the current configuration and close the window.
24 Emerson.com/Rosemount
Page 25
Reference Manual Device installation
00809-0300-5100 February 2022

2.2.3 Log file configuration

See the Rosemount TankMaster WinSetup Reference Manual for more information on how to store a communication log on disk.
Rosemount Tank Gauging 25
Page 26
Device installation Reference Manual
February 2022 00809-0300-5100

2.2.4 Changing the current protocol channel configuration

The channel configuration can be changed at any time.
Procedure
1. In the WinSetup Workspace, open the with the enabled channels.
2. Select the desired channel icon.
3. Click the right mouse button and select Properties, or from the Service menu select
Channels Properties.
Protocols folder and the protocol sub folder
4. Select the appropriate tab and change protocol settings as required.
26 Emerson.com/Rosemount
Page 27
Reference Manual Device installation
00809-0300-5100 February 2022

2.2.5 Protocol server configuration

You can specify which protocol servers that will be connected when starting TankMaster WinSetup.
Procedure
1. In the WinSetup workspace select the
2. Click the right mouse button and select Configure.
3. In the Connect column, select check boxes of the protocols to be automatically connected when WinSetup starts up.
Protocols folder.
Disable a protocol server
You may disable a protocol server at any time by using the Disable command.
Procedure
In the WinSetup workspace, open the
1.
2. Click the right mouse button on the desired protocol server icon and select Disable.
Protocols folder.
Rosemount Tank Gauging 27
Page 28
Device installation Reference Manual
February 2022 00809-0300-5100

2.3 Preferences

2.3.1 Measurement units

Specify units for inventory calculations and measured variables such as level and temperature. To change measurement units do the following:
Procedure
1. Select the desired server (e.g. “This Workstation”) in the WinSetup workspace.
2.
Click the right mouse button and select Setup, or from the Service menu select Servers Setup.
3. In the Server Preferences window select the Units tab.
4. Select the desired measurement units for level/ullage, temperature, pressure, volume, density, and weight.
5. Click the OK button to save the current setting and close the window.
Note
Make sure that the desired measuring units are specified before installing new tanks and devices.
28 Emerson.com/Rosemount
Page 29
Reference Manual Device installation
00809-0300-5100 February 2022
Changing measurement units for existing tanks
Configuration of measurement units in the Server Preferences window only affect installation of new tanks. For tanks which are already installed in the WinSetup Workspace you will have to follow this procedure.
Procedure
1. Uninstall the tank.
2.
Change measurement units in the Server Preferences/Units window (or in the TankMaster WinOpi program choose menu option Setup System and change units in the System Setup window).
3. Install the tank again.
Rosemount Tank Gauging 29
Page 30
Device installation Reference Manual
February 2022
00809-0300-5100

2.3.2 Ambient Air Temperature

This section describes how to change the Ambient Air Temperature preferences.
Procedure
1. Select the desired server (e.g. “This Workstation”) in the WinSetup workspace.
2.
Click the right mouse button and select Setup, or from the Service menu select Servers Setup.
3. In the Server Preferences window select the Ambient Air Temperature tab:
4. Choose Auto when there is a temperature sensor available that can be used for Ambient Air Temperature measurements. Otherwise, select the Manual option and type a value for the Ambient Air temperature.
Device Click the browse button and select the device to which a
temperature sensor is connected.
Source Select temperature source associated with the selected device. In a
Rosemount Tank Gauging system the associated temperature transmitter has to be configured in the tank database of the Rosemount 2410 Tank Hub.
Sensor Select a specific sensor to be used for Ambient Air Temperature.
Value Range
5. Click the OK button to save the current setting and close the window.
Related information
Installing a Rosemount 2410 Tank Hub
Value Range defines the minimum and maximum values when Ambient Air Temperature is manually entered.
30 Emerson.com/Rosemount
Page 31
Reference Manual Device installation
00809-0300-5100
February 2022

2.3.3 Inventory

Local Gravity and Ambient Air Density calculations are used for automatic density measurements. This is a description on how to change those Inventory settings.
Procedure
1. Select the desired server (e.g. “This Workstation”) in the WinSetup workspace.
2.
Click the right mouse button and select Setup, or from the Service menu select Servers Setup.
3. In the Server Preferences window select the Inventory tab:
4. Local Gravity is used for density and weight calculations when an optional pressure transmitter is installed.
Select Manual if you want to use a specific value for the Local Gravity.
Select Calculated if you want local gravity to be calculated by TankMaster. In this
case you need to enter Latitude and Elevation of the site.
5. Ambient Air Density is used for calculating Observed Density and Weight in Air (WIA).
Select Manual if you want to use a specific value for Ambient Air Density.
Select Calculated if you want the Ambient Air Density to be calculated by
TankMaster. The calculated value is based on the Base Density and the Ambient Air Temperature.
See the TankMaster WinOpi Reference Manual for more information on inventory parameters and calculations.
Click the OK button to save the current settings and close the window.
6.
Rosemount Tank Gauging 31
Page 32
Device installation Reference Manual
February 2022 00809-0300-5100

2.3.4 Miscellaneous

This is a description of how to change parameters such as type of Tank Capacity Table or Reference Temperature.
Procedure
1. Select the desired server (e.g. “This Workstation”) in the WinSetup workspace.
2.
Click the right mouse button and select Setup, or from the Service menu select Servers Setup.
3. In the Server Preferences window select the Miscellaneous tab.
4. Choose the type of Tank Capacity Table (TCT) to be used as default setting when installing new tanks. The default TCT type will automatically be chosen when strapping tables are created for new tanks. However, the TCT type can be changed when the strapping table is specified in the Tank Capacity Setup window regardless of the settings in the Server Preferences window.
You can choose between TCT type Raw, International, and Northern
Specify the maximum number of
5.
6. Specify the Reference Temperature to be used for inventory calculations. Normally, the standard value 15 °C is used.
7. Click the OK button to save the current settings and close the window.
(1) See the TankMaster WinOpi Reference Manual for further information.
Digital Alarms that will be used.
(1)
.
32 Emerson.com/Rosemount
Page 33
Reference Manual Device installation
00809-0300-5100 February 2022

2.3.5 Setting the name tag prefixes

TankMaster WinSetup allows you to specify default name tag prefixes that will appear automatically when installing new tanks and devices. Note that Tank Tag must begin with a letter. These prefixes can be ignored if you want to use other prefixes instead.
Procedure
1. From the
2. In the Preferences window select the Tag Prefixes tab.
Service menu select Preferences.
3. Type the prefixes to be used for tank names and device names. and click the OK button.
You can change the prefixes later at any time.
Note
Changing tag prefixes will not affect names of existing tanks and devices.
Rosemount Tank Gauging 33
Page 34
Device installation Reference Manual
February 2022
00809-0300-5100

2.3.6 Email configuration

TankMaster WinSetup allows you to setup an e-mail client for alarm notifications and reports.
Procedure
1. From the
2. In the Preferences window select the E-mail Configuration tab.
3. Enter the following information:
SMTP Server Specify an SMTP server for outgoing messages. Contact
Service menu select Preferences.
your LAN administrator or Internet Service Provider (ISP) for details.
Sender Address The e-mail account which will send e-mails from the current
workstation must be located on the specified SMTP server.
SMTP Port Contact your LAN administrator or ISP for details.
Authentication Configuration
Channel Security Configuration
Subject Enter a title for an e-mail alarm notification. This subject is
34 Emerson.com/Rosemount
Select this option and enter a Login name and Password if authentication is required on the mail server. Contact your LAN administrator or ISP for details.
Select this option if the e-mail client requires the use of channel security. Contact your LAN administrator or ISP for details.
only used for Alarm notification and is optional. This subject will not be used for other e-mails sent from the built-in e­mail client.
Page 35
Reference Manual Device installation
00809-0300-5100
February 2022

2.3.7 Tank View layout

The Setup Tank View tab is used to specify variables to be presented in the Tank View window, see the Rosemount TankMaster WinSetup Reference Manual for more information. TankMaster WinSetup allows you to create a new tank view layout and store it on disk, or load an existing table layout from disk.
Procedure
1. From the
2. In the Preferences window select the Setup Tank View tab.
Service menu select Preferences.
3. Click the Load Table button if you wish to edit an existing Table Layout.
4. In the
5. Click the Move button to move it to the Selected Parameters pane on the right-hand
6. Repeat this procedure for each parameter you wish to include. The Move All button
Rosemount Tank Gauging 35
Available Parameters pane on the left-hand side of the Preferences/Setup Tank
View window, select the parameter to be presented in the Tank View window.
side of the Preferences/Setup Tank View window.
allows you to move all variables at once to the Selected Parameters list box.
Page 36
Device installation Reference Manual
February 2022 00809-0300-5100
7. Ensure that all parameters to appear in the Tank View window are included in the Selected Parameters list box as illustrated below:
8. Click the Save As button if you wish to store the current tank view table for future use.
9. Click the
OK button to save the current Tank View settings and close the window.
Note
When clicking the Apply or the OK button, the parameter setup is stored in the table layout that is currently used by the Tank View window.
36 Emerson.com/Rosemount
Page 37
Reference Manual Device installation
00809-0300-5100 February 2022
10. To view the specified tank parameters click the right mouse button and select Open Tank View:
The Tank View window shows the selected parameters as configured in the Setup
Tank View.
Rosemount Tank Gauging 37
Page 38
Device installation Reference Manual
February 2022 00809-0300-5100

2.3.8 Tank visibility

The Tanks Visibility tab lets you configure tanks on a remote tank server to make them visible on the current TankMaster WinOpi client.
Procedure
1. From the
2. In the Preferences window select the Tanks Visibility tab.
3. Select the Enable Tank Visibility function check box.
From the
4. the tanks are installed.
5. In the Select tanks to be visible on this workstation pane, check the tanks you wish to make visible on the current workstation.
Service menu select Preferences.
Select tank server drop-down list select the remote tank server on which
In the default setting all tanks are visible.
6. Select Visible on this workstation in case you would like to make tanks installed on a remote server automatically visible on the current workstation as well.
38 Emerson.com/Rosemount
Page 39
Reference Manual Device installation
00809-0300-5100 February 2022
7. Click the OK button to store the current configuration and close the Preferences window.
Figure 2-3: Tanks Visibility Enabled for All Tanks
Figure 2-4: One Tank Disabled from Visiblity (TK-X)
Rosemount Tank Gauging 39
Page 40
Device installation Reference Manual
February 2022 00809-0300-5100

2.4 Field device installationoverview

The Rosemount TankMaster software supports many field devices, for example:
Rosemount 2460 System Hub
Rosemount 2410 Tank Hub
Rosemount 5900 Radar Level Gauge
Rosemount 2240S Multi-input Temperature Transmitter
Rosemount 2230 Graphical Field Display
Rosemount 5408 Radar Level Transmitter
Rosemount 5300 Guided Wave Radar
Rosemount 3051/3051S Pressure Transmitter
Rosemount 644 Temperature Transmitter

2.4.1 Configuration

A TankMaster installation includes configuration of the Rosemount Tank Gauging system for communication with the TankMaster work station and field devices as well as configuration of device specific parameters.
Configuration of radar level gauges includes:
mapping to a specific tank
communication parameters
device specific parameters
tank geometry
tank environment parameters
Configuration of Auxiliary Tank Devices (ATDs) includes:
communication address
mapping of measurement variables to source devices
Rosemount 2240S Multi-input Temperature Transmitter or other supported
temperature transmitter
temperature elements
water level sensor
Rosemount 2230 Graphical Field Display
other supported field devices
40 Emerson.com/Rosemount
Page 41
Reference Manual Device installation
00809-0300-5100 February 2022

2.5 Installing a Rosemount 2460 System Hub

This is a brief description of how to install and configure a Rosemount 2460 System Hub in a Rosemount Tank Gauging system.
Procedure
1. Ensure that the Rosemount TankMaster WinSetup program is up and running.
2.
Enable and configure a Protocol Channel to establish communication with the appropriate port on the TankMaster PC.
3. Start the installation wizard in TankMaster WinSetup:
a) Click the right mouse button on the Devices folder.
b) Select the Install new option.
4. Select device type 2460 System Hub.
5. Specify a name tag in the 2460 System Hub Tag input field. This tag will be used as
an identifier of the Rosemount 2460 in various windows and dialogs.
Click the Next button to proceed with the installation wizard.
6.
7. Verify that the Host and Field ports are properly configured. Host ports are used for
communication with TankMaster work stations or other host systems. Field ports are used for communication with the Rosemount 2410 Tank Hub, the Rosemount 5900S Radar Level Gauge, and other field devices.
8. Configure the Tank Database. Ensure that the Modbus® Addresses of the
connected devices are properly set. These addresses must correspond to the database settings of the Rosemount 2410 Tank Hub.
Related information
Rosemount 2460 Reference Manual
Rosemount Tank Gauging 41
Page 42
Device installation Reference Manual
February 2022 00809-0300-5100

2.6 Installing a Rosemount 2410 Tank Hub

The installation wizard covers basic configuration of a Rosemount 2410 Tank Hub. If further configuration of Primary Bus, Secondary Bus, Relay Output, and Hybrid Density Calculation is required this must be done separately via the 2410 Tank Hub Properties window.
Related information
Rosemount 2410 Reference Manual

2.6.1 Installing a Rosemount 2410 using the wizard

It is very important that the Tank Database of the Rosemount 2460 System Hub is properly configured prior to configuring the Rosemount 2410 Tank Hub. This ensures that the system hub is able to collect data from the different field devices.
Procedure
1. In the Workspace window select the
2. Click the right mouse button and select Install New, or from the Service menu
choose Devices Install New. Now the Select Device window appears.
Devices folder.
42 Emerson.com/Rosemount
Page 43
Reference Manual Device installation
00809-0300-5100
February 2022
3. From the Device Type drop-down list, choose the 2410 Tank Hub option.
4. Type a name in the 2410 HUB Tag input field. The tag will be used as an identifier for
the Rosemount 2410 Tank Hub in various windows and dialogs.
5. Click the Next button to proceed with communication settings.
Related information
Installing a Rosemount 2460 System Hub
Rosemount Tank Gauging 43
Page 44
Device installation Reference Manual
February 2022
00809-0300-5100
Communication setup
Procedure
1. Specify whether TankMaster communicates directly with the Rosemount 2410
Tank Hub, or via a Rosemount 2460 System Hub.
2. If the tank hub is connected to a system hub, select the appropriate system hub
from the 2460 Tag drop-down list.
3. If the tank hub is connected directly to a TankMaster PC and not via a system hub,
specify the communication protocol channel which is associated with the communication port on the TankMaster workstation. To check which channels are enabled:
in the WinSetup workspace open the Protocols folder
a)
b) click the right mouse button on the ModbusMaster protocol icon
c) choose the Properties option
To check which communication port that is associated with a certain channel:
a) right-click the protocol channel icon
b) open the Communication tab and check which communication port that is
selected.
44 Emerson.com/Rosemount
Page 45
Reference Manual Device installation
00809-0300-5100
4. To verify communication with the tank hub, type the current Modbus address and
click the Verify Communication button. The Unit Id will appear if the correct Modbus address was entered (the tank hub is shipped with the default Modbus address=247).
Note
In case several Rosemount 2410 Tank Hubs are connected using the same default address (247), you will have to change the addresses before the Verify command can be used.
5. In the 2410 Tank Hub Communication window click the Next button to continue the
installation procedure of the tank hub.
Related information
Master protocol channel configuration How to change the Modbus address of the Rosemount 2410
February 2022
How to change the Modbus® address of the Rosemount 2410
Procedure
1. In the
2. Enter the Unit ID and the new Modbus Address. When changing the device
3. Click the OK button to confirm the address settings and close the Change Address
4. In the
2410 Tank Hub Communication window click the Change Address on Device
button to open the Change Address window.
address, the Unit Id is used as a unique identifier of the device. The Unit Id can be found on a label mounted on the device.
Tip
If there is no other device connected that uses the same address as the current device, you can find the Unit Id by typing the current address into the Address field in the 2410 Tank Hub Communication window and clicking the Verify Communication button.
window.
2410 Tank Hub Communication window click the Verify Communication button to check that communication is established between the TankMaster work station and the Rosemount 2410 Tank Hub. The Unit ID will appear when TankMaster finds the tank hub.
Rosemount Tank Gauging 45
Page 46
Device installation Reference Manual
February 2022 00809-0300-5100
Tank Database setup
Each tank is represented by a position in the Rosemount 2410 tank database. Each device connected to the Rosemount 2410 is mapped to a tank position. For each tank position, a name is assigned as an identifier of the tank. The tank database maps field devices to the various tanks, and identifies the devices whenever there is a request for measurement data from a Rosemount 2460 System Hub.
Procedure
1. The
2. In the Tank Position column, map each device to a tank by selecting the
Device Type column lists all devices that communicate on the Tankbus. Ensure that all devices connected to the Tankbus appear in the Device Type list to verify proper communication.
appropriate number from the drop-down list in the tank database.
Example
The example below illustrates a tank connected to a Rosemount 2410 Tank Hub. Note that tank positions which are mapped to devices are enabled for editing. You may change tank name and Modbus® address in the right-hand side of the Tank Hub
Tank Database window.
46 Emerson.com/Rosemount
Page 47
Reference Manual Device installation
00809-0300-5100 February 2022
3. Optional: A multiple tanks version of the Rosemount 2410
(2)
allows you to connect
several tanks to a single tank hub.
Example
The example below illustrates three tanks connected to a tank hub for multiple tanks.
4. Type the desired names in the Tank Name field. These tank names should also be used at a later stage when installing the tanks associated with the current Rosemount 2410 Tank Hub.
5. For each tank, specify a Modbus address in the
Level Modbus Address column to be associated with the level gauge. This must be the same Modbus address as configured in the tank database of the Rosemount 2460 System Hub. The Level Modbus Address is used to identify level gauges when distributing requests for measurement data from the system hub.
(2) Multiple tank version of the Rosemount 2410 Tank Hub requires model code option M. See the Rosemount Tank Gauging
System Data Sheet for more information.
Rosemount Tank Gauging 47
Page 48
Device installation Reference Manual
February 2022 00809-0300-5100
6. The various non-level devices on a tank are represented by a single ATD device in the Rosemount Tank Gauging system.
In tank position 1, the Rosemount Tank Gauging system uses the Modbus
address of the Rosemount 2410 Tank Hub itself as the ATD Modbus address. In the example above, the tank hub has Modbus address 101. This address is automatically used as the ATD Modbus address as well.
For tank positions 2 to 10
(2)
you need to specify Modbus addresses in the ATD
Modbus Address column to be associated with the different ATD devices.
Important
The ATD Modbus addresses must be the same as configured in the Rosemount 2460 System Hub’s tank database.
An empty ATD Modbus address field indicates that no ATD device is mapped to
that particular tank position.
7. Click the
Next button to continue the installation procedure.
Related information
Installing a Rosemount 2460 System Hub Rosemount 2460 Reference Manual Installing a tank
48 Emerson.com/Rosemount
Page 49
Reference Manual Device installation
00809-0300-5100 February 2022
Device Tag setup
Level Tags and ATD Tags are configured automatically based on tank names in the 2410 Tank Hub Tank Database window and the configuration of tag prefixes in the Preferences/Tag Prefix window. However, it is possible to edit the Level Tag and ATD Tag
fields.
Procedure
1. Verify that the
TankMaster Level Tag is correct or type a new one.
2. Verify that the TankMaster ATD Tag is correct or type a new one. In case the ATD tag field is empty and disabled then no ATD device is associated with that tank position.
3. Click the
Related information
Setting the name tag prefixes
Rosemount Tank Gauging 49
Next button to continue the installation procedure.
Page 50
Device installation Reference Manual
February 2022 00809-0300-5100
Local display setup
Choose the parameters to be displayed on the integral display panel on the Rosemount 2410 Tank Hub. The display will alternate between the selected items at a rate given by the Display Toggle Time.
Procedure
1. In the
Units for Display pane, choose the desired measurement units from the drop­down lists. These measurement units will be used by the 2410 local display when presenting various tank variables.
2. In the Display Tanks pane, select check boxes for the tanks that you want to present in the Rosemount 2410 integral display.
3. In the
Display Tank Parameters pane, choose the tank parameters to be displayed by checking the appropriate boxes. See Table 2-2 for more information on available parameters:
Table 2-2: Various Tank Parameters can be Presented on the Local Display
Tank Parameter Description
Level The current product level in the displayed tank.
Distance Distance (ullage) is measured from the Tank Reference
point to the product surface.
Level rate The speed at which the product surface moves when
emptying or filling the tank.
Signal strength The signal strength of the radar level gauge measurement
signal.
50 Emerson.com/Rosemount
Page 51
Reference Manual Device installation
00809-0300-5100 February 2022
Table 2-2: Various Tank Parameters can be Presented on the Local Display
(continued)
Tank Parameter Description
Free water level Water surface level at the bottom of the tank. Available
when a water level sensor is installed in the tank.
Vapor pressure Tank vapor pressure.
Liquid pressure Product liquid pressure.
Air pressure Ambient air pressure.
Air temperature Ambient air temperature.
Vapor temperature Tank vapor temperature.
Product temperature Average temperature of the product.
Tank temperature Average temperature of product and vapor in the tank.
Temperature 1, 2 ... Temperature value measured by element 1, 2, etc.
Observed density The actual product density at the current product
temperature.
Reference density Density at reference temperature (used for inventory
calculations).
Volume Total observed volume.
User defined 1 to 5 Variables for advanced configuration.
4. Enter the Display Toggle Time. Information on the local display alternates between the selected items at a rate given by the Display Toggle Time value.
5. The tank hub can be configured to show different parameters for each tank on the Tank Bus. In the 2410 Tank Hub Local Display window click the
Individual Tank
Configuration button:
6. The Display Tanks pane has a list of all tank positions in the tank database. Tank positions within brackets, for example (Tank Pos 6), are not configured in the Rosemount 2410’s tank database.
Rosemount Tank Gauging 51
Page 52
Device installation Reference Manual
February 2022 00809-0300-5100
7. Select the desired tank position and choose which parameters to show on the display. Repeat this procedure for each tank position.
8. Click the
OK button to store the configuration and return to the 2410 Tank Hub
window.
9. Click the Next button to continue the installation procedure.
52 Emerson.com/Rosemount
Page 53
Reference Manual Device installation
00809-0300-5100
February 2022
Configuration summary
The 2410 Tank Hub Summary window shows information about all devices included in the Rosemount 2410’s tank database.
Procedure
1. Verify that all Modbus® addresses, level tags and ATD tags presented in the Tank Hub Summary window are correct. In case you would like to make any changes, click the Back button until the appropriate configuration window appears.
2410
2. By selecting the Install Level and AUX devices check box, field devices connected to the tank hub will automatically be installed in the TankMaster workspace. The check box is selected by default and this is the recommended setting.
Once installed, the devices need to be configured via the Properties window. The installation procedure will be facilitated by using the Install Level and AUX devices check box. However, the field devices (level gauge and ATD) can be installed at a later stage by using the device installation wizard for the respective device.
Rosemount Tank Gauging 53
Page 54
Device installation Reference Manual
February 2022 00809-0300-5100
3. Click the Finish button to confirm the installation. The installed devices will appear in the Workspace window. The installed devices will appear in the Workspace window.
Figure 2-5: Devices Appear in the WinSetup Workspace Window
Postrequisites
Make sure that devices are properly configured.
Related information
Installing a Rosemount 5900 Radar Level Gauge Installing Auxiliary Tank Devices Using the device installation wizard
54 Emerson.com/Rosemount
Page 55
Reference Manual Device installation
00809-0300-5100 February 2022
Advanced configuration
The installation wizard does not include all configuration options available for a Rosemount 2410 Tank Hub. Further configuration can be done via the 2410 Tank Hub Configuration window:
host communication parameters for the Primary Bus
host communication and emulation parameters for the Secondary Bus
virtual relays
hybrid density calculation
See the Rosemount 2410 Tank Hub Reference Manual (00809-0100-2410) for more information on how to configure a Rosemount 2410 Tank Hub.
Related information
Rosemount 2410 Reference Manual
Rosemount Tank Gauging 55
Page 56
Device installation Reference Manual
February 2022 00809-0300-5100

2.7 Installing a Rosemount 5900 Radar Level Gauge

This is a description of how to setup a Rosemount 5900 Radar Level Gauge by using the TankMaster WinSetup configuration software. It applies to the Rosemount 5900S as well as the Rosemount 5900C. Pictures in the examples below will show the Rosemount 5900S in most cases, but the functionality is the same for the Rosemount 5900C if not otherwise stated.
The Rosemount 5900 Radar Level Gauge is typically installed in TankMaster WinSetup as part of the Rosemount 2410 Tank Hub installation procedure. Once installed, the Rosemount 5900 is configured via the 5900 RLG Properties window. The 5900 RLG Properties window includes tabs for basic and advanced configuration.
When connecting a Rosemount 5900 Radar Level Gauge to a Rosemount 2410 Tank Hub in an existing Rosemount Tank Gauging system, the level gauge needs to be mapped to the appropriate tank in the tank hub’s database. Configuration is performed via the 5900 RLG Properties window.
A Rosemount 5900 is most conveniently installed by utilizing the integrated option included as part of the Rosemount 2410 Tank Hub installation procedure.
It can also be installed and configured by using the WinSetup installation Wizard. This method should only be used in special cases when, for example, the 5900 is connected to the Tankbus at a later stage and not available when installing the Rosemount 2410 Tank Hub.
The following configuration steps are included in the Rosemount 5900 Radar Level Gauge basic configuration:
communication parameters
antenna type
tank geometry
Configuration of a Rosemount may also include:
Tank Scan
Empty Tank Handling
Due to the properties of the product, the tank shape, or other circumstances, further configuration may be needed in addition to the basic configuration. Disturbing objects and turbulent conditions in the tank may require that advanced measures are taken. Advanced configuration options include:
Tank Environment
Tank Shape
Surface Echo Tracking
Filter Settings
Related information
Installing a Rosemount 5900 using the installation wizard Advanced configuration Adding a tank
56 Emerson.com/Rosemount
Page 57
Reference Manual Device installation
00809-0300-5100 February 2022

2.7.1 Configuration via the Properties window

The basic configuration procedure of a Rosemount 5900 Radar Level Gauge is straightforward and easy to perform. All configuration options are available in the Rosemount 5900 RLG Properties window. For a basic configuration of the Rosemount 5900 do the following:
Procedure
1. In the WinSetup Rosemount 5900S Radar Level Gauge icon.
Workspace window, open the Devices folder and select the
Rosemount Tank Gauging 57
Page 58
Device installation Reference Manual
February 2022 00809-0300-5100
2. Click the right mouse button and select Properties, or from the Service menu select Devices Properties.
The Communication, Antenna, and Geometry tabs cover basic configuration of the Rosemount 5900.
3. Select the Communication tab. Note that a red gauge icon means that the device needs to be configured.
4. Verify the tank position.
Connected to 2410 HUB field shows the name of the Rosemount 2410 Tank Hub
The and the tank position that the device is mapped to in the tank hub’s database. The tank position indicates which tank the device is associated with.
In most cases, the Rosemount 2410 single tank version is used for level gauges. In those cases, only one tank position will be used in the tank hub’s database.
In case the multiple tank version of the Rosemount 2410 is used to connect several tanks, the level gauge can be mapped to another tank through the 2410 Tank Hub Properties/Tank Database window:
a) in the WinSetup workspace click the right mouse button on the Rosemount
2410 icon.
b) select Properties and open the Tank Database tab.
c) select the desired tank position.
58 Emerson.com/Rosemount
Page 59
Reference Manual Device installation
00809-0300-5100
February 2022
5. Verify that the Modbus® address is correct. To change the address:
a) Click the
Change button to open the Communication window.
b) Click the Change Address on Device button to open the Change Address
window:
c) Enter the Unit ID. When changing the device address, the Unit Id is used as a
unique identifier of the device. The Unit Id can be found on the main label.
d) Set the desired Modbus address.
e)
Click OK to confirm the settings and close the Change Address window.
f) Click OK to close the Communication window.
6. In the 5900 RLG Properties/Communication window click the Apply button to store the Modbus address in the 5900 holding register.
7. Select the Antenna tab:
8. Select one of the predefined Antenna Types to match the antenna attached to the level gauge. For predefined antennas, a number of parameters are configured automatically in order to optimize measurement performance.
Rosemount Tank Gauging 59
Page 60
Device installation Reference Manual
February 2022 00809-0300-5100
In case there are disturbances close to the nozzle you may need to adjust the Hold Off Distance. By increasing the Hold Off distance, the measurement range is reduced in the upper part of the tank.
For non-standard antennas you may choose one of the User Defined antennas. However, it is recommended that you contact Emerson Automation Solutions/ Rosemount Tank Gauging for advice before using this advanced option.
Click the
9.
Apply button to save the configuration.
10. Select the Geometry tab:
11. Enter the tank geometry parameters: a) Tank Reference Height. This is the distance from the hand dip nozzle (Tank
Reference Point) to the Zero Level (Datum Plate) close to the bottom of the tank.
b) Reference Distance (G). This is the distance from the Gauge Reference Point
to the Tank Reference Point, which is located at the side of the flange that meets the tank nozzle. G is positive if the Tank Reference Point is located above the Gauge Reference Point, otherwise G is negative (see Figure 2-6).
When using a Rosemount 5900 with Still-pipe Array antenna and hinged hatch, the Tank Reference Point is located at the hand-dip plate inside the hatch. The hand-dip plate is also used as the Gauge Reference Point for this antenna type, which means that G=0.
60 Emerson.com/Rosemount
Page 61
G>0
A
B
G<0
Reference Manual Device installation
00809-0300-5100
February 2022
c) Minimum Level Distance (C). This is defined as the distance between the
Zero Level (Dipping Datum Point) and the minimum level for the product surface. By specifying a C-distance the measuring range can be extended to the bottom of the tank.
C>0: the level gauge presents negative level values when the product surface is below the Zero Level. The Show negative level values as zero check box lets you present product levels below the Zero Level (Datum plate) as equal to zero.
C=0: measurements below the Zero Level will not be approved, i.e. the RLG will report “invalid level” if the product level is below the Datum Plate.
Figure 2-6: Reference Distance (G)
A. Tank Reference Point
Gauge Reference Point
B.
12. Enter the Calibration Distance. Use this parameter to calibrate the level gauge so
that measured product levels match hand dipped levels. A minor adjustment may be necessary when the gauge is installed. For example, a minor deviation between the actual tank height and the value stored in the device database may occur if tank dimensions according to drawings are not quite up to date with actual dimensions.
For still-pipe applications the Calibrate function in the TankMaster WinSetup program is a useful tool to configure the Calibration Distance and the Correction Factor.
Optional: Select the
13.
Show negative level as zero check box in case you don’t want
the transmitter to show negative level values.
14. Click the OK button to save the configuration and close the configuration window.
Postrequisites
In addition to the configuration steps described above, a basic configuration of the Rosemount 5900 may include using the Tank Scan function to verify that there are no disturbing objects in the tank that may interfere with the level measurements. Also, the Empty Tank Handling function may be used to optimize measurement performance near the tank bottom.
Related information
Antenna configuration Tank Database setup
Rosemount Tank Gauging 61
Page 62
Device installation Reference Manual
February 2022 00809-0300-5100
Advanced configuration Calibrating the level gauge Rosemount 5900S Reference Manual Rosemount 5900C Reference Manual
62 Emerson.com/Rosemount
Page 63
Reference Manual Device installation
00809-0300-5100 February 2022
Antenna configuration
In the 5900 Properties/Antenna window several antenna types are available. For some of them additional configuration is required, such as antenna size and pipe diameter.
Antenna types:
Horn
Parabolic
Still-Pipe Array Fixed
Still-Pipe Array Hatch
LPG/LNG
Cone (5900C)
Cone Pipe (5900C)
For the LPG/LNG antenna you need to select the appropriate option:
LPG/LNG 150 PSI + Valve
LPG/LNG 150 PSI
LPG/LNG 300 PSI + Valve
LPG/LNG 300 PSI
LPG/LNG 600 PSI + Valve
LPG/LNG 600 PSI
For the Cone antenna the following options are available:
Cone 4" PTFE
Cone 4" Quartz
Cone 6" PTFE
Cone 6" Quartz
Cone 8" PTFE
Cone 8" Quartz
For the Cone Pipe antenna the following options are available:
Cone Pipe PTFE
Cone Pipe Quartz
Pipe diameter
For Still-pipe Array, LPG, and Cone Pipe Pipe Diameter parameter compensates for the lower microwave propagation speed inside a still-pipe. See also “Using the Calibrate function” on page 138 for information on how to calibrate the 5900 for still-pipe installations.
(3) Rosemount 5900C only
Rosemount Tank Gauging 63
(3)
antennas you need to specify Pipe Diameter.
Page 64
Device installation Reference Manual
February 2022
For pipe antennas it is very important that the accurate pipe diameter is input to WinSetup in order to avoid scale factor deviations which may result in inaccurate level readings.
For Cone Pipe antennas supplied by factory the following nominal values are recommended as input in TankMaster WinSetup:
Antenna Nominal Pipe Diameter
Cone Pipe 1 inch 30 mm
Cone Pipe 2 inch 56 mm
Related information
Using the Calibrate function
00809-0300-5100
Antenna size
For Still-pipe Array antennas, various antenna size options are available:
5 inch
6 inch
8 inch
10 inch
12 inch
User Defined antenna
The User Defined option should only be used for special applications when using non­standard antennas.
Procedure
1. Choose the appropriate antenna type:
User Defined Free Propagation
User Defined Still-Pipe
User Defined Still-Pipe Array
2.
Enter the Tank Connection Length in the TCL input field.
3. For still pipe applications, type the inner diameter of the still pipe in the Pipe Diameter input field.
4. In case there are disturbances close to the nozzle you may need to adjust the Hold Off Distance. By increasing the Hold Off distance, the measurement range is reduced in the upper part of the tank.
5. Select Apply to save the configuration.
64 Emerson.com/Rosemount
Page 65
Reference Manual Device installation
00809-0300-5100
February 2022

2.7.2 Advanced configuration

In addition to the basic configuration, there are advanced configuration options available for the Rosemount 5900 Radar Level Gauge. These may be used to optimize measurement performance for certain applications.
Related information
Installing a Rosemount 5900 Radar Level Gauge
Tank Shape
The Tank Type and Tank Bottom Type parameters optimize the Rosemount 5900 for various tank geometries and for measurements close to the tank bottom.
Procedure
1. In the
5900 RLG Properties window select the Tank Shape tab:
2. Select a Tank Type option that is similar to the actual tank. Choose Unknown if there is no applicable option.
3. Select
4. Click the OK button to save the configuration and close the window.
Tank Bottom Type that matches the actual tank. Choose Unknown if there is
no applicable option.
Tank Environment
Certain tank conditions may require additional configuration of the Rosemount 5900 Radar Level Gauge in order to optimize measurement performance. By considering
Rosemount Tank Gauging 65
Page 66
Device installation Reference Manual
February 2022
00809-0300-5100
environmental conditions in the tank, the Rosemount 5900 can compensate for conditions such as rapid level changes, weak echo signals, or varying surface echo amplitudes.
Procedure
1. In the
5900 RLG Properties window select the Environment tab:
2. Select the check boxes that correspond to the conditions in the tank. Use as few options as possible. It is recommended that no more than two options are used simultaneously.
3. Select
Product Dielectric Range from the drop-down list. Select the Unknown option if the correct value range is unknown or if the contents of the tank is changing on a regular basis.
4. Click the OK button to save the configuration and close the window.
Related information
Rosemount 5900S Reference Manual Rosemount 5900C Reference Manual
66 Emerson.com/Rosemount
Page 67
Reference Manual Device installation
00809-0300-5100 February 2022
Advanced Configuration window
The Advanced Configuration window provides further configuration options. The following options are available:
Tank Scan
Empty Tank Handling
Surface Echo Tracking
Safety Alarm (only used for SIL safety systems)
Figure 2-7: The 5900S RLG Properties/Advanced Configuration window
(4)
(4)
Related information
Rosemount 5900S Reference Manual Rosemount 5900C Reference Manual
(4) May also be used in Basic configuration
Rosemount Tank Gauging 67
Page 68
Device installation Reference Manual
February 2022 00809-0300-5100

2.7.3 Installing a Rosemount 5900 using the installation wizard

The installation wizard in TankMaster WinSetup is a tool that may be used to install and configure various devices in a Rosemount Tank Gauging system.
Field devices are typically installed as part of the installation procedure for a Rosemount 2410 Tank Hub. However, the installation wizard lets you install devices in TankMaster separately from the tank hub installation.
Prerequisites
Ensure that the tank databases of the Rosemount 2460 System Hub and the Rosemount 2410 Tank Hub are properly configured for the new Rosemount 5900.
Ensure that the correct Modbus® address is configured for the new device.
Procedure
1. In the Workspace window select the
Devices folder.
68 Emerson.com/Rosemount
Page 69
Reference Manual Device installation
00809-0300-5100
February 2022
2. Click the right mouse button and select Install New, or from the Service menu select Devices Install New.
3. Select Device Type 5900S Radar Level Gauge from the drop-down list.
4. Enter the desired tag for identifying the device.
Click the Next button to open the 5900 RLG Communication window:
5.
6. In the 2410 HUB Tag drop-down list, select the tank hub that the Rosemount 5900 Radar Level Gauge is connected to. Normally, there is one tank hub for each tank.
7. Enter the
Modbus address. This address must be available in the tank databases of
the Rosemount 2460 System Hub and the Rosemount 2410 Tank Hub.
8. Click the Verify Communication button to confirm that the TankMaster PC communicates with the device. The Unit ID will appear when contact is established.
Rosemount Tank Gauging 69
Page 70
Device installation Reference Manual
February 2022
00809-0300-5100
9. Check the Position in 2410 HUB field to verify that the device is mapped to the correct tank position in the Rosemount 2410 tank database. The tank position indicates which tank the device is associated with. Normally, the single tank version of the Rosemount 2410 Tank Hub is used for level gauges. In those cases the Position in 2410 HUB field will be equal to 1 since there is only one tank position that is used in the tank hub's database.
In case the multiple tank version of the Rosemount 2410 Tank Hub is used for several tanks, the level gauge can be mapped to the desired tank via the 2410 Tank Hub Properties/Tank Database window:
In the WinSetup workspace, click the right mouse button on the tank hub
a)
icon.
b)
Select the Properties option.
c) Select the Tank Database tab and map the device to the desired tank
position.
10. Click the Next button to proceed with the Rosemount 5900 configuration.
11. In the 5900 RLG Summary window, click the Finish button to verify the configuration and finish the installation wizard. In case some part of the configuration needs to be changed, click the Back button until the desired window appears.
Postrequisites
Verify that the device communicates with the Rosemount 2410 Tank Hub. Use for example the tank hub’s Device Live List function to see all devices connected to the tank hub. The live list is available by right clicking the tank hub’s device icon.
Further configuration options are available in the 5900 RLG Properties window.
Related information
Installing a Rosemount 5900 Radar Level Gauge
70 Emerson.com/Rosemount
Page 71
Reference Manual Device installation
00809-0300-5100 February 2022
Configuration via the Properties window Advanced configuration Installing a Rosemount 2460 System Hub Installing a Rosemount 2410 Tank Hub Tank Database setup Rosemount 5900S Reference Manual Rosemount 5900C Reference Manual
Rosemount Tank Gauging 71
Page 72
Device installation Reference Manual
February 2022 00809-0300-5100

2.8 Installing the Rosemount 5900S 2-in-1version

Installing a 2-in-1 Rosemount 5900S Radar Level Gauge is in most parts similar to installing a standard 5900S. A significant difference though, is that in the TankMaster operator’s interface the Rosemount 5900S 2-in-1 version will be configured as two separate gauges, and each gauge will be associated with its own tank.

2.8.1 Overview

To install and configure a 2-in-1 version of the Rosemount 5900S level gauge:
Procedure
1. In Rosemount TankMaster WinSetup, configure the Rosemount 2460 System Hub’s tank database by adding two Rosemount 5900S gauges.
2.
In Rosemount TankMaster WinSetup, install the Rosemount 2410 Tank Hub.
3. Configure the new tank hub’s tank database:
a) map the two Rosemount 5900S gauges to two different tank positions, i.e.
configure the Primary and Secondary gauges as if they are installed on two different tanks
b) assign a Level Modbus® address for each Rosemount 5900S
4. Assign tags for the Rosemount 5900S Radar Level Devices and the Auxiliary Tank
5. Install the devices in the TankMaster workspace. This is done automatically by the
6. Configure the level gauges and ATD devices (right-click the device icon and select
7. Install two new tanks to be associated with the primary and secondary Rosemount
8. Assign a Rosemount 5900S and an ATD for the primary tank.
9. Assign a Rosemount 5900S for the secondary tank. Note that the primary and
10. Configure the tanks.
11. Open the TankMaster workspace to verify correct installation of tanks and devices.
Related information
Installing a Rosemount 2410 Tank Hub
(5)
Device transmitters, displays, and pressure transmitters.
installation wizard for the Rosemount 2410 in case the Install Level and AUX
devices check box is selected in the Rosemount 2410 Tank Hub Summary window.
Properties).
5900S radar level gauges.
secondary tanks in TankMaster workspace represent one tank in real life.
(ATD). The ATD includes various non-level devices such as temperature
(5) Auxiliary Tank Device (ATD) is used in the Tank Database of the Rosemount 2410 Tank Hub to designate various devices
such as temperature and pressure transmitters, displays and other non-level devices.
72 Emerson.com/Rosemount
Page 73
A
B
C
Reference Manual Device installation
00809-0300-5100 February 2022
Configuration example of Rosemount 5900S 2-in-1
The Rosemount 5900S 2-in-1 version is installed and configured in TankMaster as a tank gauging system with two tanks. Various configuration options are possible:
1. Primary tank with 5900S and ATD (2240S, 2230 ...). Secondary tank with 5900S.
2.
Primary tank with 5900S and ATD (2240S, 2230 ...). Secondary tank with 5900S. ATD on Primary tank mapped to secondary tank (see Figure 2-8).
3. Primary tank with 5900S and ATD (2240S, 2230...). Secondary tank with 5900S and redundant ATD.
Figure 2-8: Example of a Rosemount 5900S 2-in-1 System
A. Rosemount 5900S 2-in-1 on Primary tank
B. Rosemount 5900S 2-in-1 on Secondary tank
C.
Rosemount 2240S: mapped from primary tank
Rosemount Tank Gauging 73
Page 74
Device installation Reference Manual
February 2022 00809-0300-5100

2.8.2 Installation and configuration

To install and configure the 2-in-1 version of the Rosemount 5900S, follow these steps:
Procedure
1. Start the TankMaster WinSetup program.
2.
Configure the Rosemount 2460 System Hub’s tank database. Since the Rosemount 5900S 2-in-1 version includes two separate gauges, you will have to add two Level Device addresses in the Tank Database.
3. Assign Modbus® addresses for the Rosemount 5900S 2-in-1 level devices.
A. Primary tank
B. Secondary tank
Example
Primary tank TK-1: address=1
Example
Secondary tank TK-1S: address=51
It is recommended that Modbus address 50 + “X” is used for the secondary level
device, where “X” is the address of the primary level device.
For the Primary tank the temperature device will automatically be assigned the
same Modbus address as the Rosemount 2410 Tank Hub.
For the Secondary tank you will not need to enter a “Temp Device Address”.
Do not configure any temperature elements (Number of Temp Elements=0) for
the Secondary temperature device.
In case two temperature transmitters are installed on the tank, one to be associated with the Primary and the other to the Secondary tank, you will need to configure Temp Device Address as well as Number of Temp Elements for both tanks. This configuration is like a standard configuration with one Rosemount 2410 connected to two separate tanks.
74 Emerson.com/Rosemount
Page 75
Reference Manual Device installation
00809-0300-5100 February 2022
4. Proceed with installing a Rosemount 2410 Tank Hub. In the WinSetup workspace, select the Devices folder, click the right mouse button and select File Install New:
5. In the Select Device window, select Device Type 2410 Tank Hub and assign a 2410 Tag.
6. Click the
Next button to proceed to the Tank Database configuration window.
Rosemount Tank Gauging 75
Page 76
Device installation Reference Manual
February 2022 00809-0300-5100
7. Configure the tank hub’s tank database:
A. Primary level gauge on TK-1
B. Secondary level gauge on TK-1S
In the tank database, the Rosemount 5900S 2-in-1 will appear as two separate level gauges. The two gauges are mapped to different tank positions and will be associated with different tanks in the WinSetup workspace.
Example
Assign the same name on the two tanks. Add “S” to the secondary tank name, for example:
primary tank: TK-1
secondary tank: TK-1S
Assign Level Modbus Addresses to the gauges. Example:
TK-1: 1
TK-1S: 51
It is recommended that Modbus address 50 + “X” is used for the secondary device, where “X” is the address of the primary radar level gauge.
Ensure that the same addresses are configured in the tank databases for the Rosemount 2460 System Hub and the Rosemount 2410 Tank Hub, respectively.
8. Click
Next to proceed to the Device Tags window.
76 Emerson.com/Rosemount
Page 77
Reference Manual Device installation
00809-0300-5100 February 2022
9. Verify Level Tags for the primary and secondary Rosemount 5900S gauges, and ATD Tags for the other devices (temperature transmitter, field display, pressure transmitter). The tags can be changed later if needed.
Rosemount Tank Gauging 77
Page 78
Device installation Reference Manual
February 2022 00809-0300-5100
10. In the final step of the Rosemount 2410 installation wizard, check that all device tags and Modbus addresses are correct. Note that if a Rosemount 2460 System Hub is used, the Modbus addresses in the Rosemount 2410’s tank database must match the addresses configured in the Rosemount 2460’s tank database.
Ensure that the Install Level and ATD devices... check box is selected in case you like to enable automatic installation of devices in the TankMaster workspace. This is the recommended setting.
Note
If new devices are added to an existing Rosemount 2410 Tank Hub, the Install Level and ATD devices check box may not be active. Then you will have to add the new devices manually.
78 Emerson.com/Rosemount
Page 79
Reference Manual Device installation
00809-0300-5100 February 2022
11. Verify that the devices appear in the WinSetup workspace. The Rosemount 2410 Tank Hub, the two Rosemount 5900S RLGs, and the ATD are automatically installed in the TankMaster WinSetup workspace if the check box is marked in the 2410 Tank Hub Summary window.
Note
The new device icons are red indicating that these devices need to be configured.
Once the tank hub and associated level and ATD devices are installed in TankMaster you will have to configure each device.
12. Configure the level gauges and ATD devices (right-click the device icon and select Properties).
13. Finally, when devices are installed and configured you will have to install and configure the two tanks to be associated with the 2-in-1 Rosemount 5900S. In the following description one tank will be associated with the primary level gauge (LT­TK-1 in this example), and the other tank will be associated with the secondary level gauge (LT-TK-1S).
a) Select the Tanks folder.
b) Click the right mouse button and select the Install New option.
c) Install and configure the primary tank first and then the secondary tank.
Rosemount Tank Gauging 79
Page 80
Device installation Reference Manual
February 2022 00809-0300-5100
14. For each tank, select Tank Type and assign a Tank Tag. It is recommended that you use the same tank name as specified in the tank database of the Rosemount 2410 Tank Hub.
15. Assign devices to the primary tank TK-1:
a) Primary Rosemount 5900S level gauge (LT-TK-1)
b)
ATD device (ATD-TK-1) in case the tank is equipped with non-level devices such as a Rosemount 2240S Temperature Transmitter, a Rosemount 2230
80 Emerson.com/Rosemount
Page 81
Reference Manual Device installation
00809-0300-5100 February 2022
Graphical Field Display, Water Level Sensor, or a Rosemount 3051S Pressure Transmitter.
Rosemount Tank Gauging 81
Page 82
Device installation Reference Manual
February 2022 00809-0300-5100
16. For the secondary tank TK-1S assign the secondary 5900S level gauge (LT-TK-1S) as shown below:
82 Emerson.com/Rosemount
Page 83
Reference Manual Device installation
00809-0300-5100 February 2022
17. Optional: You may map the ATD device on the primary tank to the secondary tank TK-1S as shown below. This allows you to view temperature data, such as Average Temperature, for the secondary tank as well.
Note
You need to ensure that Show Only Vacant Devices is not selected in order to view all devices.
Rosemount Tank Gauging 83
Page 84
Device installation Reference Manual
February 2022 00809-0300-5100
18. Choose the desired tank measurement variables. For the primary tank, variables associated with the ATD device automatically appears in the Configuration window.
Note
The available source variables may vary depending on the actual devices connected to the Rosemount 2410 Tank Hub.
For the secondary tank, “none” is automatically selected for temperature and pressure variables as well as for free water level. In case the ATD device on the primary tank is mapped to the secondary tank as well, those variables will appear in the Configuration window for the secondary tank.
84 Emerson.com/Rosemount
Page 85
Reference Manual Device installation
00809-0300-5100
February 2022
19. Specify parameter Value Range (Minimum, Maximum) to be used in various windows for presentation of measurement data.
Note
Calculation of the Average Temperature (AVG Temp) requires a valid Level value. In case the Level value is not available, the Average Temperature (AVG Temp) will also be missing.
20. Configure whether to use measurement values (Automatic) from the available instruments or Manual values.
Rosemount Tank Gauging 85
Page 86
Device installation Reference Manual
February 2022 00809-0300-5100
21. In the TankMaster WinSetup workspace verify that the tanks and devices are properly installed. Check that the two Rosemount 5900S gauges and the ATD are associated with the right tanks. Two options are shown below:
ATD (temperature transmitter, etc.) is associated with the primary tank only
ATD on primary tank is mapped to both the primary and secondary tanks
Example
ATD mapped to Primary tank only
Example
ATD mapped to both the primary and secondary tank
86 Emerson.com/Rosemount
Page 87
Reference Manual Device installation
00809-0300-5100 February 2022
Related information
Installing a Rosemount 5900 Radar Level Gauge Installing Auxiliary Tank Devices
Rosemount Tank Gauging 87
Page 88
Device installation Reference Manual
February 2022 00809-0300-5100

2.9 Installing Auxiliary Tank Devices

Auxiliary Tank Devices (ATDs) such as the Rosemount 2240S Multi-input Temperature Transmitter and the Rosemount 2230 Graphical Field Display, are typically installed as part of the installation procedure of the Rosemount 2410 Tank Hub.
ATD devices appear in the TankMaster workspace and are configured via the 22XX ATD window. The 22XX ATD window includes tabs for configuration of temperature sensors, graphical field displays, and water level sensors. It also includes configuration options for mapping outputs of measurement instruments to variables such as Free Water Level and Liquid Pressure. Before starting the ATD installation, read the reference manuals for the various ATD devices, such as the Rosemount 2240S and the Rosemount 2230, to learn more about how to configure these devices.

2.9.1 Opening the Properties window

Procedure
1. In the WinSetup Workspace window, open the icon.
2. Click the right mouse button and select Properties, or from the Service menu select Devices Properties.
3. The 22XX ATD window has a number of configuration tabs for the Rosemount 2240S and the Rosemount 2230.
4. Open the appropriate tabs to configure the ATD devices.
5.
Finish the ATD configuration by clicking the Apply or OK button.
Devices folder and select the ATD
88 Emerson.com/Rosemount
Page 89
Reference Manual Device installation
00809-0300-5100
February 2022

2.9.2 Communication parameter setup

The Communication tab lets you verify the device communication settings. A red icon means that the current ATD needs to be configured.
Procedure
1. In the
2. In the Connected to HUB field, verify that the ATD device is:
22xx ATD window select the Communication tab.
connected to the correct Rosemount 2410 Tank Hub
mapped to the correct position in the tank hub’s tank database
For example, “position 1” means that the ATD device is mapped to tank database position 1 in the tank hub’s database, “position 2” to tank database position 2 etc. You can check the tank database of the current tank hub by opening the 2410 HUB Properties/Tank Database window.
3. The
4. Click the Apply button to store the configuration, or click the OK button to store
Related information
Installing a Rosemount 2410 Tank Hub
Rosemount Tank Gauging 89
Change button lets you change the Modbus® address of the current ATD device in case it is connected to a multiple tank version of the tank hub. Note that the Modbus address can only be changed for ATD devices in tank database position 2 and higher. An ATD device mapped to position 1 in the tank hub’s database uses the same Modbus address as the tank hub itself.
the configuration and close the window.
Page 90
Device installation Reference Manual
February 2022 00809-0300-5100

2.9.3 Temperature sensor configuration

The 2240S MTT Temperature Sensor tab lets you configure a sensor connected to the Rosemount™ 2240S Multi-input Temperature Transmitter. The 2240S MTT Temperature Sensor tab input fields are disabled in case no Rosemount 2240S is connected to the Tankbus.
Procedure
1. In the
2. The Rosemount 2240S is equipped with a DIP switch for automatic configuration of
22xx ATD window, select the 2240S MTT Temperature Sensor tab.
temperature element and wiring. By selecting the Use Auto Sensor Configuration check box, the transmitter is automatically configured according to the DIP switch settings.
In case a temperature element type is used that does not match the Auto Sensor Configuration, the Rosemount 2240S transmitter can be manually configured.
Related information
Temperature sensor manual configuration Rosemount 2240S Reference Manual
90 Emerson.com/Rosemount
Page 91
Reference Manual Device installation
00809-0300-5100
February 2022
Temperature sensor manual configuration
Procedure
1. Ensure that the
2. In the Method input field, select the desired user defined conversion method.
Note
For spot temperature elements, use the Pt100 sensor option to ensure correct average temperature calculation.
3. In the Connection drop-down list, choose the type of sensor connection that is used for the spot sensors.
4. Specify the measurement range of the current temperature element in the
Value and Max Value input fields.
Use Auto Sensor Configuration check box is unchecked.
Min
5. For user defined temperature conversion, such as User defined table, User defined formula, and User defined individual formula, click the configuration button that corresponds to the selected conversion method and configure the temperature sensors.
6. Click the the configuration and close the window.
Related information
Rosemount 2240S Reference Manual User defined linearization table User defined formula User defined individual formula
Rosemount Tank Gauging 91
Apply button to store the configuration, or click the OK button to store
Page 92
Device installation Reference Manual
February 2022 00809-0300-5100
User defined linearization table
When using a resistance temperature sensor, the electrical resistance values can be converted to temperature values by using a table of resistance and temperature values.
Procedure
1. In the
2. Click the Configure User Defined Linearization Table button:
3. Specify the number of conversion points in the Length of Table input field.
4. Type resistance and temperatures values into the Resistance [Ohm] and
5.
2240S MTT Temperature Sensor window, select conversion method User
Defined Linearization Table.
Temperature [C] columns.
Click the OK button to store the linearization table in the temperature transmitter’s database registers.
92 Emerson.com/Rosemount
Page 93
Reference Manual Device installation
00809-0300-5100
February 2022
User defined formula
For a resistance temperature sensor, the relation between temperature and resistance can be specified by a mathematical formula:
R=R0*(1+A*T+B*T2),
where R is the resistance at temperature T, R0 is the electrical resistance at zero degrees Celsius, and A and B are constants.
Procedure
In the
1.
2. Click the Configure User Defined Formula button:
2240S MTT Temperature Sensor window, choose conversion method User
Defined Formula.
3. Enter the parameters R0, A and B in the corresponding input fields.
4. Click the transmitter’s database registers.
OK button to store the R0, A and B parameters in the temperature
Rosemount Tank Gauging 93
Page 94
Device installation Reference Manual
February 2022 00809-0300-5100
User defined individual formula
When using User Defined Individual Formula, a mathematical formula is used for each individual temperature element:
R=R0*(1+AN*T+BN*T2+CN*T3),
where
R is the resistance at temperature T
R0 is the resistance at zero degrees Celsius
A, B, and C are individual constants for each element
N is the number of temperature sensors
Procedure
1. In the
2. Click the Configure User Defined Individual Formula button:
3. Enter parameters R0, A, B, and C for each individual temperature element.
4. Click the
2240S MTT Temperature Sensor window, choose conversion method User
Defined Individual Formula.
OK button to store the formula in the temperature transmitter’s database
registers.
94 Emerson.com/Rosemount
Page 95
Reference Manual Device installation
00809-0300-5100
February 2022

2.9.4 Average Temperature calculation

This section gives a brief description of how to configure average temperature calculations for a temperature sensor connected to the Rosemount 2240S Multi-input Temperature Transmitter.
Procedure
1. In the
2. No of Elements: select the number of temperature elements used. Verify that the
3. Sensor Type: select the type of sensor that is used; Spot or Average.
4. Insert Distance: specify the minimum distance between a temperature sensor and
5. Position: enter the distance from Zero Level to the spot temperature element. If
6.
7. Click the Apply button to store the configuration, or click the OK button to store
22xx ATD window, select the Average Temperature Calculation tab.
input fields for element positions are enabled.
the product surface for temperature sensors included in average temperature calculation.
average temperature elements are used, enter the position where the average element ends.
You may exclude a temperature element from average temperature calculation by selecting the may be useful in case you would like to exclude a malfunctioning temperature element.
the configuration and close the window.
Exclude check box in the column next to the Position field. This option
Related information
Rosemount 2240S Reference Manual
Rosemount Tank Gauging 95
Page 96
Device installation Reference Manual
February 2022
00809-0300-5100

2.9.5 Auxiliary sensor configuration

The 2240S MTT Auxiliary Sensor tab lets you configure a water level sensor connected to a Rosemount 2240S Multi-input Temperature Transmitter.
Procedure
1. In the
22xx ATD window, select the 2240 MTT Auxiliary Sensor tab.
2. Calculate the Level Offset (X) according to the formula X=(R-L1)-(L-L2).
3. Enter the resulting level offset value in the
4. Enter the Upper Dead Zone (UDZ) and the Lower Dead Zone (LDZ) if needed.
5. Click the Apply button to store the configuration, or click the OK button to store the configuration and close the window.
Related information
Rosemount 2240S Reference Manual
96 Emerson.com/Rosemount
Level Offset (X) input field.
Page 97
Reference Manual Device installation
00809-0300-5100
February 2022

2.9.6 Advanced parameter source configuration

The Advanced Parameter Source Configuration tab lets you map the output of various measurement instruments (source devices) to tank measurement variables such as Liquid Pressure and Vapor Pressure. This makes the measurement variables available for configuration in the Tank Configuration window
For tank measurement variables such as Level, Vapor Temperature, and Free Water Level, source devices are automatically mapped and do not need to be configured in the Advanced Parameter Source Configuration tab.
The Rosemount 2410 Tank Hub supports 60 source parameter mappings. Six mappings are reserved for each one of the ten tank database positions.
Note
A warning message will appear in case a certain parameter/source mapping is already used for another tank.
For a source device to be available in the Advanced Parameter Source Configuration tab, it must be mapped to the current tank in the tank hub’s database.
Procedure
1. In the
2. In the Parameter Mapping column select a tank measurement parameter.
3. In the
22xx ATD window select the Advanced Parameter Source Configuration tab.
Source Device Type/ID/No column, choose a measurement device such as a
pressure transmitter or any other type of instrument.
The default setting lists only devices mapped to the current tank position in the tank hub’s database. By using this option you will avoid mapping to a device on the wrong tank by mistake. By selecting the appropriate radio button you can choose to show all available devices, or only the devices mapped to tank position 1 in the 2410 tank database.
4. In the Source Parameter column, select the transmitter variable to be mapped to the tank measurement variable in the first column.
Rosemount Tank Gauging 97
Page 98
Device installation Reference Manual
February 2022
00809-0300-5100
5. In case User Defined is selected, you may type a description in the Description of User Def parameter box.
6. Click the
Apply button to store the configuration, or click the OK button to store
the configuration and close the window.
Related information
Installing a Rosemount 2410 Tank Hub Installing a tank
98 Emerson.com/Rosemount
Page 99
Reference Manual Device installation
00809-0300-5100 February 2022
Using a Rosemount 644 Temperature Transmitter
The Rosemount Tank Gauging system automatically handles source parameter mapping for Rosemount 2240S transmitters to provide correct input for calculation of tank measurement variables such as Vapor Temperature and Average Temperature.
For Rosemount 644 temperature transmitters, tank variables have to be mapped manually to the appropriate source devices. The output from each Rosemount 644 transmitter on the tank is mapped to a temperature tank variable as shown in the following example with three Rosemount 644 transmitters on the tank.
Procedure
1. In the
2. In the Parameter Mapping list select Temperature 1 for the first 644 temperature
3.
4. In the Source Parameter list, select Temperature 1. This is the source parameter
22xx ATD window select the Advanced Parameter Source Configuration tab.
transmitter. This temperature parameter is mapped to the temperature element at position 1 in the tank.
For the second and third 644 transmitters, select Temperature 2 and Temperature 3 in the Parameter Mapping list. These parameters are mapped to the temperature elements at position 2 and 3, respectively.
In the Source Device Type field, for each temperature parameter (Temperature 1, 2, 3) select the actual 644 transmitter to be used as source device.
designation of the temperature output from a Rosemount 644 temperature transmitter.
Note
The Vapor Temperature and Average Temperature variables are not mapped to the source devices.
5. Click the Apply button to store the configuration, or click the OK button to store the configuration and close the window.
Rosemount Tank Gauging 99
Page 100
Device installation Reference Manual
February 2022 00809-0300-5100

2.9.7 Rosemount 2230 Graphical Field Display

This section gives a brief description of how to configure a Rosemount 2230 Graphical Field Display.
All fields in the 2230 Graphical Field Display tab are disabled if no graphical field display is connected to the Tankbus, or if the display is not mapped to the current tank database position of the Rosemount 2410 Tank Hub. The Configure Tank Display 2 and Configure Tank Display 3 buttons are disabled when only one Rosemount 2230 is mapped to the current tank database position.
Procedure
In the
1.
22xx ATD window select the 2230 Graphical Field Display tab.
2. From the drop-down lists in the Units for Display pane, select the desired measurement units to be displayed for the various tank measurement variables.
Note
The Rosemount 2230 converts the unit specified in the Server Preferences window.
3. In the Common Tank Configuration pane, select tanks and tank parameters to be presented by the Rosemount 2230 display by selecting the appropriate check boxes.
100 Emerson.com/Rosemount
Loading...