Rosemount SPS 4000 Single Probe Autocalibration Sequencer-Rev 1.1 Manuals & Guides

Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
Single Probe Autocalibration Sequencer
http://www.processanalytic.com
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical repre-
sentative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate In- struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s per­formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Emerson Process Management
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com

HIGHLIGHTS OF CHANGES

Effective Feb., 1999 Rev. 1.0
Page Summary
Page 1-1 Added note concerning the Oxymitter 5000.
Page 1-2 Added the Oxymitter 5000 to the product matrix in Table 1-1. Removed
the disposable gas bottles and flow regulators from the product matrix in Table 1-1 and created Table 1-2 to distinguish these components as separate order items because the calibration gas bottles cannot be shipped via airfreight.
Page 7-1 Added Table 7-2 to list the calibration gas bottles and flow regulators
as replacement parts.
Effective April, 2001 Rev. 1.1
Page Summary
Page 7-1 Table 7-1; changed part number of solenoid, items 24 and 30.
SPS 4000
PREFACE........................................................................................................................ P-1
Definitions ........................................................................................................................P-1
Safety Instructions .......................................................................................................... P-2
1-0 DESCRIPTION AND SPECIFICATIONS........................................................................ 1-1
1-1 Component Checklist ..................................................................................................... 1-1
1-2 Overview.......................................................................................................................... 1-1
1-3 Specifications................................................................................................................... 1-4
1-4 Physical Description ....................................................................................................... 1-5
1-5 Theory of Operation....................................................................................................... 1-6
2-0 INSTALLATION .............................................................................................................. 2-1
2-1 Overview.......................................................................................................................... 2-1
2-2 Mechanical Installation ................................................................................................... 2-1
2-3 Gas Connections ............................................................................................................ 2-1
2-4 Electrical Connections .................................................................................................... 2-3
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001

TABLE OF CONTENTS

3-0 OPERATION ...................................................................................................................3-1
3-1 Overview.......................................................................................................................... 3-1
3-2 Calibration Requirements ............................................................................................... 3-1
3-3 Calibration Gas Flow Setup .......................................................................................... 3-2
3-4 Automatic Calibration ..................................................................................................... 3-2
3-5 Semi-Automatic Calibration ............................................................................................ 3-2
4-0 MAINTENANCE AND SERVICE .................................................................................. 4-1
4-1 Overview.......................................................................................................................... 4-1
4-2 Fuse Replacement ......................................................................................................... 4-1
4-3 Board Replacement ........................................................................................................ 4-1
4-4 Solenoid Replacement ................................................................................................... 4-3
4-5 Pressure Switch Replacement ...................................................................................... 4-5
4-6 Check Valve Replacement ............................................................................................ 4-6
4-7 Pressure Regulator (Optional) Maintenance ................................................................ 4-6
4-8 Flowmeter Adjustments .................................................................................................. 4-6
4-9 Flowmeter Replacement................................................................................................. 4-6
5-0 TROUBLESHOOTING .................................................................................................... 5-1
5-1 Overview.......................................................................................................................... 5-1
5-2 SPS 4000 Troubleshooting............................................................................................ 5-1
6-0 RETURN OF MATERIAL .............................................................................................. 6-1
7-0 REPLACEMENT PARTS ............................................................................................... 7-1
8-0 INDEX.............................................................................................................................. 8-1
Rosemount Analytical Inc. A Division of Emerson Process Management i
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
Figure 1-1. Typical SPS 4000 Package ................................................................................... 1-1
Figure 1-2. Autocalibration System Installation Options .......................................................... 1-3
Figure 1-3. SPS 4000 Components ......................................................................................... 1-5
Figure 1-4. SPS 4000 Calibration Setup .................................................................................. 1-7
Figure 2-1. Installation.............................................................................................................. 2-2
Figure 2-2. Electrical Connections ........................................................................................... 2-4
Figure 4-1. SPS 4000, Exploded View ..................................................................................... 4-2
Figure 4-2. Board Connections ................................................................................................ 4-4
Figure 5-1. SPS 4000 Troubleshooting Flowchart (Sheet 1 of 2) ............................................ 5-3
Table 1-1. Product Matrix........................................................................................................ 1-2
Table 1-2. Calibration Components ........................................................................................ 1-2
Table 5-1. SPS 4000 Fault Finding ......................................................................................... 5-2
Table 7-1. SPS 4000 Replacement Parts ............................................................................... 7-1
Table 7-2. Calibration Replacement Parts .............................................................................. 7-1
SPS 4000

LIST OF ILLUSTRATIONS

LIST OF TABLES

ii Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
The purpose of this manual is to provide information concerning the components, func­tions, installation and maintenance of the SPS 4000 Single Probe Autocalibration Sequencer.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this publication.
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
PREFACE
DEFINITIONS
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an essential operating procedure, condition, or statement.
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETIN
NOTE TO USERS
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
The number in the lower right corner of each illustration in this publication is a manual illus­tration number. It is not a part number, and is not related to the illustration in any technical manner.
Rosemount Analytical Inc. A Division of Emerson Process Management P-1
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
FOR THE WIRING AND INSTALLATION
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non­EU states should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external,
where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced.
The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
SPS 4000
IMPORTANT
SAFETY INSTRUCTIONS
OF THIS APPARATUS
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable
anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be
made through a circuit breaker which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker may also include a mechanically operated isolating switch. If not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazard-
ous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment — and then only by trained service personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a
danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Cer­tain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to
the Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the follow-
ing standards: EN61010-1, IEC417, and ISO3864.
P-2 Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
1
IB-106-340AC Rev. 1.1
SPS 4000
SECTION 1
DESCRIPTION AND SPECIFICATIONS
NOTE
The SPS 4000 Single Probe Autocalibration Sequencer operates exactly the same with either the Oxymitter 4000 Oxygen Transmitter or the Oxymitter 5000 Oxygen Transmitter with
OUNDATION
F instruction bulletin also include the Oxymitter 5000. When referred to an instruction bulletin for more information, reference IB-106-340 for the Oxymitter 4000 and IB-106-340-FB for the Oxymitter 5000.
fieldbus Communications. Any references to the Oxymitter 4000 throughout this
April 2001

1-1 COMPONENT CHECKLIST

A typical SPS 4000 Single Probe Autocalibra­tion Sequencer should contain the items shown in Figure 1-1. Record the part number, serial number, and order number for the SPS 4000 on the first page of this manual.
Also, use the product matrix in Table 1-1 to compare your order number against your unit. The first part of the matrix defines the model. The last part defines the various options and features of the sequencer. Ensure the features and options specified by your order number are on or included with the unit.
1-2 OVERVIEW
The SPS 4000 provides the capability of per­forming automatic, timed or on demand, calibra­tions of a single Oxymitter 4000 without sending a technician to the probe site.
The SPS 4000 can be mounted either directly to an In situ Oxymitter 4000 or at a remote location if space is limited. See Figure 1-2. However, this instruction bulletin only covers remote mounted sequencers. For information regarding integrally mounted sequencers, refer to the Oxymitter 4000 Oxygen Transmitter Instruction Bulletin. For information on equipping your existing Oxymitter 4000 with an integrally mounted SPS 4000, contact Rosemount.
1
3
1. Instruction Bulletin
2. SPS 4000 (shown with optional reference air com­ponents)
3. Optional Mounting Hardware (for pipe mounting)
Figure 1-1. Typical SPS 4000 Package
2
26010001
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
Table 1-1. Product Matrix
SPS 4000 B Autocalibration System for Oxymitter 4000 or Oxymitter 5000. Mounted separate from the probe.
SPS 4000 Autocalibration System - Instruction Bulletin
Code Oxygen Analyzer System
20 Used with Oxymitter 4000 or Oxymitter 5000 system (remote mounted only)
SPS 4000
Code Reference Air
1 No reference air required
2 Reference air provided
Code Fittings and Tubing
1 Brass Fittings, Teflon Tubing
2 Stainless Steel Fittings and Tubing
Code Electrical Classification
10 NEMA 4X
20 Hazardous Area Classifications - Cenelec EExd IIB + H2
30 Hazardous Area Classifications (Class I, Div. I, Group B,C,D) - PENDING
SPS 4000 B 20 1 1 10 Example
Notes:
(1)
Reference air is recommended with 9 ft (2.74 m) and 12 ft (3.66 m) long probes. Reference air is also recommended when ambient air may not contain the normal 20.95% O unit nearby with leaks.
2)
Customer to pipe from remote SPS 4000 to probe.
3)
Hazardous area classifications require stainless steel fittings and tubing.
, such as when the probe is mounted into a positive pressure duct with leaks or where there is a process
2
(1)
(2)
(3)
(3)
Table 1-2. Calibration Components
Part Number Description
1A99119G01 Two disposable calibration gas bottles—0.4% and 8%
O
, balance nitrogen—550 liters each, includes bottle
2
rack*
1A99119G02 Two flow regulators for calibration gas bottles
*Calibration gas bottles cannot be shipped via airfreight.
When the bottles are used with “CALIBRATION RECOMMENDED” features, the bottles should provide 2 to 3 years of calibrations in normal service.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
1
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
OXYMITTER 4000
INTEGRAL OR
REMOTE
INTEGRALLY
MOUNTED SPS 4000
SINGLE PROBE
AUTOCALIBRATION
SEQUENCER
(1 PROBE)
REFER TO IB-106-340
INTELLIGENT MULTIPROBE
IMPS 4000
TEST GAS SEQUENCER
FOR USE WITH OXYMITTER 4000
(1 TO 4 PROBES)
Figure 1-2. Autocalibration System Installation Options
In addition to the SPS 4000, multiprobe se­quencers are also available as shown in Figure 1-2. Rosemount has offered multiprobe auto­calibration sequencer systems for many years. These autocalibration systems are most cost
REMOTE MOUNTED SPS 4000
SINGLE PROBE
AUTOCALIBRATION SEQUENCER
FOR USE WITH OXYMITTER 4000
(EXPLO VERSIONS MUST BE REMOTE MOUNTED)
(1 PROBE)
26010003
effective for boilers and other combustion proc­esses that utilize many probes. Users with only one probe per combustion process can now take advantage of Rosemount’s autocalibration capability by utilizing the SPS 4000.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
1-3 SPECIFICATIONS
Mounting ............................................................. Integral to Oxymitter 4000
Remote from Oxymitter 4000
Materials of Construction
Manifold/electronics enclosure ................. Aluminum
Mounting brackets .................................... 316 stainless steel (SS)
Pneumatic fittings ..................................... 1/8 in. brass NPT (SS optional)
Pneumatic tubing ...................................... 1/4 in. Teflon (SS optional)
Assembly hardware .................................. Galvanized and stainless steel
Humidity range .................................................... 100% relative humidity
Ambient temperature range ................................ -40° to 149°F (-40° to 65°C)
Electrical classification ........................................ NEMA 4X (IP56)
Explosion-proof option (both pending) ................ CENELEC EExd IIB + H2
(Class 1, Div. 1, Group B, C, D)
SPS 4000
Electrical feedthroughs........................................ 1/2 in. NPT
Input power.......................................................... 90 to 250VAC, 50/60Hz
Power consumption............................................. 5VA maximum
External electrical noise ...................................... EN 50 082-2, includes 4KV electrostatic discharge
Handshake signal to/from
Oxymitter 4000 (self-powered) ................. 5V (5mA maximum)
Cal initiate contact input from control room......... 5 to 30VDC, Form A (SPST)
(one “In-Cal”, one “Cal Failed”)
Cabling distance between SPS 4000 and
Oxymitter 4000 ......................................... Maximum 1000 ft (303 m)
Piping distance between SPS 4000 and
Oxymitter 4000 ......................................... Maximum 300 ft (91 m)
Approximate shipping weight .............................. 10 lbs (4.5 kg)
Fisher-Rosemount has satisfied all obligations coming from the European legislation to harmonize the product requirements in Europe.
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
1
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001

1-4 PHYSICAL DESCRIPTION

The main components of the SPS 4000 are de­scribed in the following paragraphs and illus­trated in Figure 1-3.
a. Manifold
The manifold provides a mounting platform for the circuit board(s) and terminations and contains the electrical feedthroughs. Also, calibration gases are piped into and se­quenced through solenoids mounted on the manifold.
b. Calibration Gas Solenoids
The calibration gas solenoids sequence the calibration gases. One solenoid controls calibration gas 1 (high calibration gas), and the other controls calibration gas 2 (low
calibration gas). The solenoids activate and deactivate to allow the calibration gases to flow between the sequencer and Oxymitter
4000.
c. Pressure Switch
The pressure switch detects if the pressure of a calibration gas is low, which can be caused by an empty gas bottle, a discon­nected gas line, etc. Calibration is prohib­ited when calibration gas pressure is low.
d. Power Supply Board
This board converts the incoming line volt­age from AC to DC for use by the solenoids, terminations, and the programmable logic device. The power supply board also has a 5 A, 250 V, slow blow fuse.
REFERENCE AIR PRESSURE REGULATOR (OPTIONAL)
NOTE:
CALIBRATION GAS
FLOWMETER
MANIFOLD COVER IS REMOVED TO SHOW INTERNAL COMPONENTS.
ALSO, BOARD COMPONENTS ARE NOT SHOWN FOR CLARITY.
PRESSURE
SWITCH
CALIBRATION GAS 1
(HIGH CAL GAS)
SOLENOID
INTERFACE
BOARD
Figure 1-3. SPS 4000 Components
REFERENCE AIR FLOWMETER (OPTIONAL)
MANIFOLD
CALIBRATION GAS 2 (LOW CAL GAS) SOLENOID
POWER SUPPLY BOARD
TERMINAL COVER
26010002
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
e. Interface Board
The interface board contains a programma­ble logic device (PLD) that has the elec­tronics to energize and deenergize the solenoids based on a signal from the Oxy­mitter 4000.
f. Calibration Gas Flowmeter
The calibration gas flowmeter indicates the flow rate of calibration gas flowing to the Oxymitter 4000.
g. Reference Air Flowmeter (Optional)
The reference air flowmeter indicates the amount of reference air continuously flowing to the Oxymitter 4000.
h. Pressure Regulator (Optional)
The pressure regulator ensures the instru­ment air (reference air) flowing to the Oxy­mitter 4000 is at a constant pressure [20 psi (138 kPa)]. The regulator also has a filter to remove particulates in the reference air and a drain valve to bleed the moisture that col­lects in the filter bowl.
i. Terminal Strip
The terminal strip housed within the termi­nal cover provides convenient access for all signal and power user connections.
describe how an Oxymitter 4000 is autocali­brated when used with the SPS 4000.
a. In addition to the calibration methods avail-
able via the Oxymitter 4000 keypad, HART communicator, AMS software, or a remote contact, the SPS 4000 works in conjunction with the Oxymitter 4000’s CAL RECOM­MENDED feature to perform an autocalibra­tion. This feature automatically per-forms an impedance check every hour on the Oxy­mitter 4000. If a calibration is recommended and its contact output signal is set for “handshaking” with the sequencer, the Oxymitter 4000 sends a signal to the se­quencer. The sequencer automatically per­forms a calibration upon receiving the signal.
Thus, no human interface is required for the automatic calibration to take place.
b. When a calibration is required, the Oxymit-
ter 4000 sends a signal to the programma­ble logic device (PLD) on the interface board of the sequencer. The PLD energizes the calibration gas 1 (high O2) solenoid. Calibration gas 1 then flows through the se­quencer to the Oxymitter 4000. The Oxy­mitter 4000 measures the oxygen content of calibration gas 1 and sends a signal to the sequencer indicating that it received the gas. When the sequencer receives the sig­nal, the PLD deenergizes the calibration gas 1 solenoid.

1-5 THEORY OF OPERATION

The Oxymitter 4000 is one of the few instru­ments found in industry that permit the perma­nent piping of a calibration standard into the probe. Most instruments measuring pressure, flow, or temperature require that a calibration standard be brought to the instrument or that the instrument be taken to the calibration source in the instrument shop.
The permanent calibration gas connections al­low for auto-calibrations to occur without op­erator intervention. The following paragraphs
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
c. Next, the PLD energizes the calibration gas
2 (low O2) solenoid, and calibration gas 2 then flows through the sequencer to the Oxymitter 4000. The Oxymitter 4000 meas­ures the oxygen content of calibration gas 2 and sends a signal to the sequencer indi­cating that it received the gas. After meas­uring the two calibration gases, the Oxymitter 4000 automatically makes an in­ternal calibration adjustment and sends the signal to the sequencer. When the se­quencer receives the signal, the PLD deenergizes the calibration gas 2 solenoid.
SPS 4000
1
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
OXYMITTER 4000 LOGIC I/O
CALIBRATION GAS
CHECK VALVE
(SEE NOTE 1)
SPS 4000
OXYMITTER 4000 (BOTTOM VIEW)
CALIBRATION GAS FITTING
REFERENCE AIR (SEE NOTE 2)
INSTRUMENT
AIR IN
(SEE NOTE 2)
NOTE 1.
CALIBRATION
GAS 1
(HIGH O )
A CHECK VALVE IS REQUIRED AT THE OXYMITTER 4000 (BETWEEN THE CALI­BRATION GAS FITTING AND THE GAS LINE) TO PREVENT THE MIGRATION OF PROCESS GASES DOWN THE CALIBRATION GAS LINE.
CLEAN, DRY INSTRUMENT AIR IS
2. RECOMMENDED FOR REFERENCE AIR. NO REFERENCE AIR IS REQUIRED IF AMBIENT AIR CONDITIONS CONTAIN
20.95% OXYGEN.
2
CALIBRATION
GAS 2
(LOW O )
2
26010004
Figure 1-4. SPS 4000 Calibration Setup
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
1-8 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
2
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
SECTION 2
INSTALLATION
2-1 OVERVIEW
This section describes the installation of the SPS 4000.
Before starting to install this equip­ment, read the “Safety Instructions for the Wiring and Installation of this Ap­paratus” at the front of this Instruction Bulletin. Failure to follow the safety instructions could result in injury or death.
Install all protective equipment covers and safety ground leads after installa­tion. Failure to install covers and ground leads could result in serious injury or death.
2-2 MECHANICAL INSTALLATION
a. Reference Air (Figure 2-1)
1. For units with the optional reference air components, connect the instrument air supply to the IN port of the pressure regulator.
2. The pressure regulator is factory set at 20 psi (138 kPa). If necessary, readjust by turning the knob on the top of the regulator until the desired pressure is obtained.
3. Next, connect the reference air from the upper 1/4 in. tube fitting on the ref­erence air flowmeter to the REF GAS port on the Oxymitter 4000.
b. Calibration Gas (Figure 2-1)
1. Connect O2 calibration gas 1 (high calibration gas) to the HIGH CAL GAS IN 1/4 in. tube fitting on the top of the manifold. Ensure the calibration gas pressure is set at 20 psi (138 kPa).
The outline drawing in Figure 2-1 shows mounting centers and clearances of the SPS
4000. The unit is designed to mount on a wall, bulkhead, or pipe. Ensure the unit is installed according to the following specifications.
a. Install the unit no further than 300 ft (91 m)
from the Oxymitter 4000 and no further than 1000 ft (303 m) from the electronics pack­age or any customer-supplied remote input or relay output connections in the control room.
b. Locate the unit where the ambient tem-
perature is between -40° and 149°F (-40° and 65°C).

2-3 GAS CONNECTIONS

Use the following procedure to connect the cali­bration gases and reference air.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instrument air is not recommended for the high calibration gas. Do not use 100% nitrogen as a low gas (zero gas). It is suggested that the low (zero) gas be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon concen­trations of more than 40 parts per mil­lion. Failure to use proper gases will result in erroneous readings.
2. Connect O2 calibration gas 2 (low cali­bration gas) to the LOW CAL GAS IN 1/4 in. tube fitting on the top of the manifold. Ensure the calibration gas pressure is set at 20 psi (138 kPa).
3. Connect the calibration gas from the upper 1/4 in. tube fitting on the calibra­tion gas flowmeter to the check valve connected to the CAL GAS port on the Oxymitter 4000.
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
1/4 IN. TUBE FITTING
FOR HIGH CALIBRATION
0.94 (23.88) 0.94 (23.88)
1/4 IN. TUBE FITTING FOR LOW CALIBRATION GAS IN
GAS IN
1/4 IN. TUBE TO CALIBRATION GAS FLOWMETER
SPS 4000
CALIBRATION GAS FLOWMETER
(CALIBRATION GAS OUT
TO OXYMITTER 4000)
1/2 IN. NPT CONDUIT PORTS
(SEE NOTE 2)
2.00
(50.80)
13.50
(342.90)
4.25
(107.95)
9.00 (228.60) NOMINAL
14.81
(376.17)
NOMINAL
REFERENCE AIR FLOWMETER (OPTIONAL) (REFERENCE AIR OUT TO OXYMITTER 4000)
1/2 IN. NPT
CONDUIT PORT
(SEE NOTE 2)
2.00 (50.80) NOMINAL
CLEARANCE
TO REMOVE
COVER
NOTES: 1.
DIMENSIONS ARE IN INCHES WITH MILLIMETERS IN PARENTHESES.
2.
THREE 1/2 IN. NPT PORTS ARE PROVIDED FOR LOGIC I/O AND SIGNAL CONNECTIONS. THE CABLE ROUTING WILL BE DETERMINED BY THE CUSTOMER.
1/4 IN. TUBE FITTING FOR INSTRUMENT AIR IN (OPTIONAL)
1/2 IN. CONDUIT FITTING FOR LINE VOLTAGE
6.12 (155.45) NOMINAL
26010006
Figure 2-1. Installation
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
2
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001

2-4 ELECTRICAL CONNECTIONS

All wiring must conform to local and national codes. Use the following procedure to connect an SPS 4000 to an Oxymitter 4000.
Disconnect and lock out power before connecting the unit to the power supply.
NOTE
Ensure the Oxymitter 4000 is set up to handshake with the sequencer by con­figuring the logic I/O to mode 8. Refer to the Oxymitter 4000 Oxygen Trans­mitter Instruction Bulletin for more in­formation.
a. Remove screws (20, Figure 4-1) securing
terminal cover (19) and remove the cover.
b. Route the line voltage leads into the mani-
fold through the lower 1/2 in. conduit fitting on the right side of the manifold (Figure 2-1) and out through the bottom of the manifold. Connect the incoming 90 to 250 VAC, 50/60 Hz line voltage leads to the terminal strip as indicated in Figure 2-2.
c. Route the handshake logic I/O wires
through one of the 1/2 in. NPT conduit ports
on the manifold (Figure 2-1) and out through the bottom of the manifold. Connect the 5V (5 mA maximum) logic I/O leads from the Oxymitter 4000 to the terminal strip as indicated in Figure 2-2.
d. To set up the SPS 4000 to initiate a calibra-
tion from a remote location, route the 5 VDC calibration initiate contact input through one of the 1/2 in. NPT conduit ports on the manifold (Figure 2-1) and out through the bottom of the manifold. Connect the input leads to the terminal strip as shown in Figure 2-2.
e. Relay output connections are available on
the unit to signal when the Oxymitter 4000 is in calibration or when calibration failed. Relay outputs can be connected to either indicator lights or a computer interface. The relay contacts are capable of handling a 5 to 30 VDC maximum power source. The cabling requirement is 1000 ft (303 m) maximum. Route the relay output wires through one of the 1/2 in. NPT conduit ports on the manifold (Figure 2-1) and out through the bottom of the manifold. Connect the relay output wires to the terminal strip as shown in Figure 2-2.
f. Once all connections are made, install ter-
minal cover (19, Figure 4-1) and secure with screws (20).
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
5V(5mA
MAXIMUM)
LOGIC I/O
5 VDC (SELF-POWERED)
TO REMOTE CONTACT
INPUT CONNECTION
CAL INITIATE CAL FAIL
NOT USED
WHITE
NOT USED
BLACK
HANDSHAKE
CONNECTION
TO
OXYMITTER
4000
YELLOW
BROWN
5-30VDCTO
RELAY OUTPUT
CONNECTIONS
RED
BLUE
IN CAL
++++
ORANGE
GREEN
LINE IN
GROUND
NEUTRAL
----
90 - 250 VAC, 50/60 HZ LINE VOLTAGE INPUT
FACTORY WIRING TO INTERFACE BOARD
Figure 2-2. Electrical Connections
FACTORY WIRING TO
POWER SUPPLY BOARD
26010007
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
3
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
SECTION 3
OPERATION
3-1 OVERVIEW
This section specifies the requirements to set up an Oxymitter 4000 calibration and how to verify the calibration gas flow setup. It also explains the differences between automatic and semi­automatic calibrations and how to initiate them.
3-2 CALIBRATION REQUIREMENTS
a. Two tanks of precision calibration gas mix-
tures are required. Recommended calibra­tion gases are nominally 0.4% and 8.0% oxygen in nitrogen.
Do not use 100% nitrogen as a low gas (zero gas). It is suggested that gas for the low (zero) be between 0.4% and
2.0% O2. Do not use gases with hy­drocarbon concentrations of more than 40 parts per million. Failure to use proper gases will result in errone­ous readings.
12054 S.W. Doty Avenue Chicago, Illinois 60628 312/568-8840
603 Bergen Street Harrison, New Jersey 07029 201/485-1995
255 Brimley Road Scarborough, Ontario, Canada 416/266-3161
SCOTT ENVIRONMENTAL TECHNOLOGY, INC. SCOTT SPECIALTY GASES
2600 Cajon Blvd. San Bernardino, California 92411 714/887-2571 TWX: 910-390-1159
1290 Combermere Street Troy, Michigan 48084 314/589-2950
In addition to the optional disposable gas bottles available from Rosemount, two ad­ditional sources of calibrated gas mixtures are:
LIQUID CARBONIC GAS CORP. SPECIALTY GAS LABORATORIES
700 South Alameda Street Los Angeles, California 90058 213/585-2154
767 Industrial Road San Carlos, California 94070 415/592-7303
9950 Chemical Road Pasadena, Texas 77507 713/474-4141
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
Route 611 Plumsteadville, Pennsylvania 18949 215/766-8861 TWX: 510-665-9344
2616 South Loop West Suite 100 Houston, Texas 77054 713/669-0469
b. A check valve is required at the Oxymitter
4000 (between the calibration fitting and the gas line) to prevent the migration of process gases down the calibration gas line.
A typical calibration setup for the Oxymitter 4000 is shown in Figure 1-4.
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000

3-3 CALIBRATION GAS FLOW SETUP

After installing the SPS 4000 as described in Section 2, calibrate the Oxymitter 4000 to verify SPS 4000 operation and the communication link between the sequencer and Oxymitter 4000.
a. Verify that both calibration gases are con-
nected to the SPS 4000. Also verify that the pressure regulators on both calibration gas bottles are set to 20 psig (138 kPa gage).
b. Initiate a semi-automatic calibration using
one of the methods specified in paragraph 3-5.
NOTE
Only set the calibration gas flowmeter upon initial installation and after changing the diffusion element in the Oxymitter 4000. Refer to the flowmeter adjustments in Section 4 for more in­formation.
c. As the Oxymitter 4000 and SPS 4000 apply
the first calibration gas, set the calibration gas flowmeter to 5 scfh. During the applica­tion of the second calibration gas, verify that the flowmeter reads 5 scfh. If not, adjust the pressure regulator on the second calibration gas bottle so the 5 scfh flow is provided.

3-4 AUTOMATIC CALIBRATION

AMS. Refer to the logic I/O information in the HART/AMS section of the Oxymitter 4000 Oxygen Transmitter Instruction Bulle­tin for more information.
b. Timed Interval
An automatic calibration can also be pro­grammed to occur at a specific time interval, in hours, using the HART communicator or AMS software. Refer to the HART/AMS section of the Oxymitter 4000 Oxygen Transmitter Instruction Bulletin for this procedure.

3-5 SEMI-AUTOMATIC CALIBRATION

Semi-automatic calibrations are operator initi­ated and can be performed using the Oxymitter 4000 keypad, HART handheld communicator/ AMS software, or a remote contact. In addition, the calibration gases must be permanently piped to the Oxymitter 4000.
a. Oxymitter 4000 Keypad
A semi-automatic calibration can be initiated by pressing the CAL button on the Oxymit­ter 4000 keypad. For more information, re­fer to the Oxymitter 4000 Oxygen Transmitter Instruction Bulletin.
b. HART Handheld Communicator/AMS
Software
Automatic calibrations require no operator ac­tion and can be performed through the Oxymit­ter 4000 CAL RECOMMENDED feature or through scheduled time intervals that can be programmed through the HART/AMS for the Oxymitter 4000. In addition, the calibration gases must be permanently piped to the Oxy­mitter 4000.
a. CAL RECOMMENDED
If the Oxymitter 4000 is configured for handshake mode with the SPS 4000, the Oxymitter 4000 can initiate a calibration by sending a signal to the sequencer when the CAL RECOMMENDED LED activates. To enable handshake mode, the Oxymitter 4000 logic I/O must be set for mode 8.
Handshake mode is configured at the fac­tory or can be accessed through HART/
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
A semi-automatic calibration can be initiated by connecting the HART handheld commu­nicator, or AMS software, to the Oxymitter 4000 4-20 mA signal line and using the HART communicator keypad or computer keyboard to access the applicable calibra­tion menu. Refer to the Oxymitter 4000 Oxygen Transmitter Instruction Bulletin or the available HART documentation for more information.
c. Remote Contact
A semi-automatic calibration can be initiated using a remote contact such as a cus­tomer’s control system. The remote contact processes the calibration command on a PC and sends the signal to the Oxymitter
4000. For more information on remote-site calibrations, refer to the documentation for the system in use.
SPS 4000
4
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
SECTION 4
MAINTENANCE AND SERVICE
4-1 OVERVIEW
This section describes service and routine maintenance of the SPS 4000. Replacement parts referenced are available from Rosemount. Refer to Section VI for part numbers and order­ing information.
Install all protective equipment covers and safety ground leads after equip­ment repair or service. Failure to in­stall covers and ground leads could result in serious injury or death.
4-2 FUSE REPLACEMENT
The SPS 4000 has a fuse on the power supply board. Refer to Table 7-1 for replacement fuse specifications. Perform the following procedure to check or replace the fuse.
Disconnect and lock out power before working on any electrical components.

4-3 BOARD REPLACEMENT

Perform the following procedure to replace power supply board (23, Figure 4-1) or interface board (31).
Disconnect and lock out power before working on any electrical components.
a. Turn off power to the system.
b. Remove screw (22) securing manifold cover
lock (21) and remove the lock.
c. Remove manifold cover (28).
d. Remove two screws (32) attaching spacers
(35) to manifold (38).
e. Being careful not to disconnect the board
wiring, carefully lift power supply board (23) and interface board (31) from manifold (38) and set aside. Do not lose o-rings (36) from the bottom of spacers (35).
f. For the board to be replaced, tag all leads
to simplify installation.
a. Turn off power to the system.
b. Remove screw (22, Figure 4-1) securing
manifold cover lock (21) and remove the lock.
c. Remove manifold cover (28).
d. Remove fuseholder (26) by pushing in the
top and turning 1/4 turn counterclockwise. Remove fuse (25).
e. After checking or replacing fuse (25), install
fuseholder (26) by pushing in the top and turning 1/4 turn clockwise.
f. Install manifold cover (28) and secure with
manifold cover lock (21) and screw (22).
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-1
g. If replacing the power supply board, refer to
Figure 4-2. Remove the line voltage input leads from connector J7. Also, unplug cali­bration gas 1 solenoid leads from connector J5, calibration gas 2 solenoid leads from connector J4, and pressure switch leads from connector J2.
h. If replacing the interface board, refer to
Figure 4-2. Remove the CAL INITIATE leads from connector J3, CAL FAIL and IN CAL leads from connector J4, and logic I/O handshake connection from connector J5.
i. Remove stop nuts (33, Figure 4-1), washers
(34), and screws (37) securing power sup­ply board (23) and interface board (31) to spacers (35).
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
46
47
45
48
49
50
43
44
42
51
41
40
52
39
12
10
38
3
1
5
4
2
6
8
9
10
7
28
29
30
33
32
31
34
26
27
25
35
36
37
12
11
21
22
23
24
NOTE: A STANDARD SPS 4000 IS EQUIPPED WITH TEFLON
20
TUBING AND BRASS FITTINGS. OPTIONAL STAINLESS STEEL TUBING AND FITTINGS ARE ALSO AVAILABLE. REFER TO SECTION VI FOR ORDERING INFORMATION.
14 15
16
17
18
19
12
13
26010009
Figure 4-1. SPS 4000, Exploded View
4-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
4
1. Reference Air Pressure Regulator (Optional)
2. Straight Fitting
3. Mounting Bracket
4. Flat Washer
5. Lockwasher
6. Hex Nut
7. Screw
8. Tube
9. Straight Fitting
10. Elbow Fitting
11. Hex Head Plug
12. Square Head Plug
13. Conduit Fitting
14. Ground Nut
15. Terminal Base
16. Terminal Strip
17. Screw
LEGEND FOR FIGURE 4-1
18. Terminal Cover Gasket
19. Terminal Cover
20. Screw
21. Manifold Cover Lock
22. Screw
23. Power Supply Board
24. Calibration Gas 2 Solenoid
25. Fuse
26. Fuseholder
27. Pressure Switch
28. Manifold Cover
29. O-Ring
30. Calibration Gas 1 Solenoid
31. Interface Board
32. Screw
33. Stop Nut
34. Washer
35. Spacer
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
36. O-Ring
37. Screw
38. Manifold
39. Elbow Fitting (Optional)
40. Tubing (Optional)
41. Elbow Fitting (Optional)
42. Screw
43. Elbow Street Fitting (Optional)
44. Elbow Fitting
45. Reference Air Flowmeter (Optional)
46. Calibration Gas Flowmeter
47. Elbow Fitting
48. Elbow Street Fitting (Optional)
49. Straight Fitting (Optional)
50. Flowmeter Bracket
51. Bracket
52. Screw
j. Carefully separate boards (23 and 31).
k. Connect replacement board to board (23 or
31).
l. Install screws (37), washers (34), and stop
nuts (33) to secure interface board (31) and power supply board (23) to spacers (35).
m. Install all applicable leads in the appropriate
locations on the power supply board or in­terface board as shown in Figure 4-2.
n. Install power supply board (23, Figure 4-1)
and interface board (31) into manifold (38). Align spacers (35) with the mounting holes on the manifold and secure with screws (32). Ensure o-rings (36) are installed be­tween the spacers and the manifold surface.
o. Install manifold cover (28) and secure with
manifold cover lock (21) and screw (22).
4-4 SOLENOID REPLACEMENT
Use this procedure to replace either calibration gas 1 (high calibration gas) solenoid (30, Figure 4-1) or calibration gas 2 (low calibration gas) solenoid (24).
Disconnect and lock out power before working on any electrical components.
a. Turn off power to the system.
b. Shut off the calibration gases at the
cylinders.
c. Remove screw (22) securing manifold cover
lock (21) and remove the lock.
d. Remove manifold cover (28).
e. Remove two screws (32) attaching spacers
(35) to manifold (38).
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-3
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
TO PRESSURE
SWITCH
LINE IN
LINE
VOLTAGE
INPUT
NEUTRAL
GROUND
SPS 4000
POWER SUPPLY BOARD
TO CALIBRATION GAS 1 SOLENOID
TO CALIBRATION GAS 2 SOLENOID
GROUND
VAC INPUT TO OXYMITTER 4000
L1
N
INTERFACE BOARD
BLACK WHITE
RED BLUE ORANGE GREEN
YELLOWYELLOW
BROWN
+
CAL INITIATE
-
+
CAL FAIL
-
+
IN CAL
-
LOGIC I/O
+
HANDSHAKE TO
-
OXYMITTER 4000
REMOTE CONTACT INPUT
STATUS INDICATOR OUTPUT
26010012
Figure 4-2. Board Connections
4-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
4
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
f. Being careful not to disconnect the board
wiring, carefully lift the board and spacer assembly from manifold (38) and set aside. Do not lose o-rings (36) from the bottom of spacers (35).
g. Tag and unplug solenoid (30 or 24) leads
from power supply board (23). Refer to Figure 4-2. Calibration gas 1 solenoid wires connect to connector J5, and calibration gas 2 solenoid wires connect to connector J4.
h. Remove the top nut of solenoid (30 or 24)
securing the coil assembly and washer to the base. Remove the coil assembly, in­cluding the leads, and washer. Place a 13/16 in. deep socket over the solenoid base and remove.
When installing a solenoid, do not overtighten. Damage to the solenoid may occur.

4-5 PRESSURE SWITCH REPLACEMENT

Use the following procedure to replace pressure switch (27, Figure 4-1).
a. Turn off power to the system.
b. Shut off the calibration gases at the
cylinders.
c. Remove screw (22) securing manifold cover
lock (21) and remove the lock.
d. Remove manifold cover (28).
e. Remove two screws (32) attaching spacers
(35) to manifold (38).
f. Being careful not to disconnect the board
wiring, carefully lift the board and spacer assembly from manifold (38) and set aside. Do not lose o-rings (36) from the bottom of spacers (35).
g. Tag and remove the leads from pressure
switch (27).
i. Install the new solenoid base. Be careful not
to overtighten. Install the new washer and coil assembly and secure with the top nut. Connect the leads to the proper connector on power supply board (23). Refer to Figure 4-2 if necessary.
j. Carefully install the board and spacer as-
sembly into manifold (38, Figure 4-1) by aligning spacers (35) with the mounting holes on the manifold and securing with screws (32). Ensure o-rings (36) are in­stalled between the spacers and the mani­fold surface.
k. Install manifold cover (28), and secure with
manifold cover lock (21) and screw (22).
l. Turn on the calibration gases at the
cylinders.
h. Place a 1-1/16 in. 6-point socket over pres-
sure switch (27) and remove.
When installing the pressure switch, do not overtighten. Damage to the so­lenoid may occur.
i. Install new pressure switch (27). Be careful
not to overtighten. Connect the leads to the proper terminals on the pressure switch.
j. Carefully install the board and spacer as-
sembly into manifold (38, Figure 4-1) by aligning spacers (35) with the mounting holes on the manifold and securing with screws (32). Ensure o-rings (36) are in­stalled between the spacers and the mani­fold surface.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-5
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
k. Install manifold cover (28), and secure with
manifold cover lock (21) and screw (22).
l. Turn on the calibration gases at the
cylinders.

4-6 CHECK VALVE REPLACEMENT

The check valve may stick or become plugged over time. Replace when necessary. If conden­sation deposits are noted upon removal, con­sider insulating the check valve.
4-7 PRESSURE REGULATOR (OPTIONAL)
MAINTENANCE
a. Pressure Adjustments
Reference air pressure regulator (1, Figure 4-1) is factory set to 20 psi (138 kPa). Ad­just using the knob on top of the pressure regulator if necessary.
In applications with a heavy dust loading, the O
probe diffusion element may become
2
plugged over time, causing a slower speed of response. The best way to detect a plugged diffusion element is to note the time it takes the Oxymitter 4000 to return to the normal process reading after the last cali­bration gas is removed and the calibration gas line is blocked off. A plugged element also can be indicated by a slightly lower reading on the flowmeter.
Change the diffusion element when the calibration gas flowmeter reads slightly lower during calibration or when the re­sponse time to the process flue gases be­comes very slow. Each time the diffusion element is changed, reset the calibration gas flowmeter to 5 scfh and calibrate the Oxymitter 4000. For more information on changing the diffusion element, refer to the instruction bulletin for the Oxymitter 4000 in use.
b. Reference Air Flowmeter (Optional)
Do not use fingers to release valve stem. The valve may release air at high pressures and cause injury.
b. Condensation Drain
To drain excess moisture from the filter bowl of reference air pressure regulator (1), use a screwdriver or comparable tool to periodi­cally release the valve stem on the bottom of the pressure regulator.
4-8 FLOWMETER ADJUSTMENTS
a. Calibration Gas Flowmeter
Calibration gas flowmeter (46, Figure 4-1) regulates the calibration gas flow and must be set to 5 scfh. However, only adjust the flowmeter to 5 scfh after placing a new dif­fusion element on the end of the Oxymitter
4000. Adjusting the flowmeter at any other time can pressurize the cell and bias the calibration.
Reference air flowmeter (45) regulates the reference air and must be set to 2 scfh. Adjust the flow with the knob on the bottom of the reference air flowmeter when necessary.

4-9 FLOWMETER REPLACEMENT

Use this procedure to replace either reference air flowmeter (45, Figure 4-1) or calibration gas flowmeter (46).
a. Turn off power to the system.
b. Shut off the calibration gases at the
cylinders.
c. Loosen, but do not remove, four screws
(42) securing flowmeter bracket (50) to mounting bracket (3).
d. Flex the bottom of flowmeter bracket (50)
downward and away to disengage and re­move the flowmeter bracket from mounting bracket (3).
4-6 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
4
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
e. For reference air flowmeter (45), remove
pressure regulator (1) by disconnecting tubing (40) from elbow fitting (39). Also, dis­connect the tubing between the Oxymitter 4000 and sequencer from straight fitting (49).
For calibration gas flowmeter (46), discon­nect the tubing between the Oxymitter 4000 and the sequencer at elbow fitting (47). Also, disconnect tube (8) from elbow fitting (44).
f. Remove screws (52) and bracket (51) se-
curing flowmeter (45 or 46) to flowmeter bracket (50).
g. Remove flowmeter (45 or 46), with installed
fittings, from flowmeter bracket (50).
h. For reference air flowmeter (45), remove el-
bow street fittings (43 and 48). It is not nec­essary to remove fittings (39 and 49) from the street fittings.
For calibration gas flowmeter (46), remove elbow fittings (44 and 47).
i. Apply pipe thread sealant to the threads of
top fitting (48 or 47) and bottom fitting (43 or
44) and install fittings into new flowmeter (45 or 46).
j. Position flowmeter (45 or 46) into flowmeter
bracket (50) and secure with bracket (51) and screw (52).
k. For reference air flowmeter (45), connect
tubing (40) to elbow fitting (39) and install pressure regulator (1). Also, connect the tubing between the Oxymitter 4000 and se­quencer to straight fitting (49).
For calibration gas flowmeter (46), connect tube (8) to elbow fitting (44) and connect the gas tubing between the Oxymitter 4000 and sequencer to elbow fitting (47).
l. Slide the top slots of flowmeter bracket (50)
onto screws (42). Flex the bottom of the bracket downward and toward mounting bracket (3) to engage the bottom bracket slots and screws. Tighten screws.
m. Turn on the calibration gases at the
cylinders.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-7
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
4-8 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
5
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
SECTION 5
TROUBLESHOOTING
5-1 OVERVIEW
This section describes the SPS 4000 trouble­shooting procedures. Additional troubleshooting information can be found in the Oxymitter 4000 Oxygen Transmitter Instruction Bulletin.
Install all protective equipment covers and safety ground leads after trouble­shooting. Failure to replace covers and ground leads could result in seri­ous injury or death.

5-2 SPS 4000 TROUBLESHOOTING

Use the CAL FAIL and IN CAL relay outputs to identify possible SPS faults.
a. If a calibration was not successfully com-
pleted, the SPS 4000 sends a CAL FAIL contact indication to the control room. To determine if the SPS 4000 caused the failed calibration, go to the Oxymitter 4000 site to view the keypad. Or, access the HART/AMS menus. For more information on HART/ AMS, refer to the HART/AMS section in the Oxymitter 4000 Oxygen Transmitter Instruc­tion Bulletin.
1. If no alarms are indicated on the key­pad or in the HART/AMS STATUS sub­menu, the calibration did not fail be­cause of an Oxymitter 4000 fault. Therefore, a calibration gas flow prob­lem occurred. Refer to Table 5-1 or Figure 5-1 to troubleshoot the SPS
4000.
2. If the LAST CAL FAILED alarm is indi­cated on the keypad or in the HART/ AMS STATUS sub-menu, the failure is
due to either a bad Oxymitter 4000 cell or a calibration gas flow problem.
(a) Verify your calibration setup per
paragraph 2-3 in Section 2, IN­STAL-LATION; Section 3, OP­ERATION; and paragraph 4-8 in Section 4, MAINTENANCE AND SERVICE.
(b) Perform another calibration and
monitor the process. If the calibra­tion fails before both calibration gases finish sequencing, a gas flow problem exists. Refer to Table 5-1 or Figure 5-1 to troubleshoot the SPS 4000.
If the calibration setup is correct and the Oxymitter 4000 indicates an invalid slope fault (fault 12) be­fore the gases are purged and a last calibration failed fault (fault 14) after the gases are purged, replace the Oxymitter 4000 cell per the Oxymitter 4000 Oxygen Transmit­ter Instruction Bulletin.
b. If a semi-automatic or manual calibration is
being performed but no 5 - 30 VDC relay output contact (IN CAL or CAL FAIL) is be­ing received by the control room, the inter­face board relays are malfunctioning. Replace the interface board per paragraph 4-3 in Section 4, MAINTENANCE AND SERVICE.
NOTE
If the unit is performing frequent auto­calibrations, investigate at the Oxymit­ter 4000 site or using HART/AMS. This condition may indicate an aging cell in the Oxymitter 4000.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-1
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
Table 5-1. SPS 4000 Fault Finding
Symptom Check Fault Remedy
SPS 4000
No calibration gas flow
Wiring Improper wire connections,
loose connections, or dam­aged wiring
Logic I/O Oxymitter 4000 logic I/O not
set for calibration handshak­ing with SPS 4000
Calibration gas lines be­tween cylinders and manifold
Calibration gas flowmeter knob
Calibration gas line between manifold and calibration gas flowmeter
Fuse on power supply board Blown fuse Replace fuse per paragraph
Interface board option Interface board not sending
Check valve Clogged check valve Replace check valve per
Clogged calibration gas line Replace clogged calibration
Flowmeter knob not turned counterclockwise to allow flow
Clogged calibration gas line Replace clogged calibration
signals
Properly connect wiring or secure loose wiring connec­tions; replace damaged wir­ing if necessary.
Set logic I/O to mode 8 via HART/AMS.
gas line.
Turn calibration gas flow­meter knob counter­clockwise to allow calibration gas to flow.
gas line.
4-2.
Replace interface board per paragraph 4-3.
paragraph 4-6.
Calibration gas line between calibration gas flowmeter and check valve
Calibration gas flowmeter Clogged flowmeter Replace flowmeter per para-
Power supply output Power supply failure Replace power supply board
Solenoid Solenoid failure Replace solenoid per para-
Pressure switch Pressure switch failure Replace pressure switch per
Clogged calibration gas line Replace calibration gas line.
graph 4-9.
per paragraph 4-3.
graph 4-4.
paragraph 4-5.
5-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
5
SYMPTOM — NO CALIBRATION GAS FLOW
CHECK ALL WIRING BETWEEN OXYMITTER 4000 AND SPS 4000.
IS
WIRING
PROPERLY
CONNECTED
AND
SECURE?
YES
CHECK LOGIC I/O SETTING VIA HART/AMS.
PROPERLY CONNECT WIRING OR
NO
SECURE LOOSE WIRING CON­NECTIONS; REPLACE DAMAGED WIRING.
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
IS
LOGIC I/O
SET FOR
MODE 8?
YES
DISCONNECT CAL GAS INPUT LINES AT MANIFOLD.
IS THERE
FLOW?
YES
ENSURE CAL GAS FLOWMETER KNOB IS TURNED COUNTER­CLOCKWISE TO ALLOW FLOW.
DOES
CAL GAS
FLOWMETER
REGISTER
FLOW?
YES
CHECK FUSE F1 ON POWER SUPPLY BOARD.
NO
NO
NO
SET LOGIC I/O TO MODE 8 VIA HART/AMS.
REPLACED CLOGGED CAL GAS LINE BETWEEN CAL GAS CYLINDER AND MANIFOLD.
REPLACED CLOGGED CAL GAS LINE BETWEEN MANIFOLD AND CAL GAS FLOWMETER.
F1
J2
J5
J4
J3
LO GAS
HI GAS
NO GAS
CAL RET
IS
FUSE
BLOWN?
NO
CONTINUED
ON SHEET
2OF2
YES
REPLACE FUSE PER PARAGRAPH 4-2.
POWER SUPPLY BOARD
35970001
Figure 5-1. SPS 4000 Troubleshooting Flowchart (Sheet 1 of 2)
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-3
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SYMPTOM NO CALIBRATION GAS FLOW (CONTINUED)
SPS 4000
CONTINUED
FROM SHEET
1OF2
PLACE JUMPER BETWEEN CAL RET TERMINAL AND EITHER HI GAS OR LO GAS TERMINAL OF J3. SEE NOTE 1.
IS THERE
FLOW?
NO
USE METER (SEE NOTE 2) TO CHECK FOR SHORT BETWEEN CAL RET AND NO GAS TERMINALS OF J3.
IS THERE A SHORT?
NO
DISCONNECT SOLENOID FROM POWER SUPPLY BOARD AND USE METER TO MEASURE ACROSS TWO OUTER PINS OF BOARD CONNECTOR. SEE NOTE 4. CHECK BOTH SOLENOIDS.
IS THERE +30VDC?
YES
DISCONNECT CAL GAS LINE AT MANIFOLD OUTPUT PORT.
YES
YES
NO
INTERFACE BOARD IS NOT SENDING SIGNAL. REPLACE INTERFACE BOARD PER PARAGRAPH 4-3.
DISCONNECT CAL GAS LINE AT CHECK VALVE.
REPLACE POWER SUPPLY BOARD PER PARAGRAPH 4-3.
NOTE 1:
NOTE 2:
NOTE 3:
NOTE 4:
DISCONNECT CAL GAS LINE AT TOP FITTING OF CAL GAS FLOWMETER.
REPLACE FAULTY CAL GAS FLOWMETER PER PARAGRAPH 4-9.
SECURELY TIGHTEN ALL J3 SCREW TERMINALS ON POWER SUPPLY BOARD TO MAKE CONNECTIONS.
USE A SIMPSON MODEL 260 OR EQUIVALENT MULTIMETER.
IF REPLACING THE CHECK VALVE DOES NOT CORRECT THE PROBLEM, A CLOG COULD EXIST IN THE RED SILICON GAS TUBE WITHIN THE PROBE.
IF CHECKING CAL GAS 1 SOLENOID CONNECTOR J5, ENSURE CAL RET TERMINAL IS JUMPERED TO HI GAS TERMINAL OF J3. IF CHECKING CAL GAS 2 SOLENOID CONNECTOR J4, ENSURE CAL RET TERMINAL IS JUMPERED TO LO GAS TERMINAL OF J3.
IS THERE
FLOW?
NO
IS THERE
FLOW?
NO
YES
YES
REPLACE CHECK VALVE PER PARAGRAPH 4-6. SEE NOTE 3.
REPLACE CLOGGED CAL GAS LINE BETWEEN CAL GAS FLOWMETER AND CHECK VALVE.
NO
IS THERE
FLOW?
YES
REPLACE PRESSURE SWITCH PER PARAGRAPH 4-5.
REPLACE SOLENOID PER PARAGRAPH 4-4.
35970002
Figure 5-1. SPS 4000 Troubleshooting Flowchart (Sheet 2 of 2)
5-4 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
6
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
SECTION 6
RETURN OF MATERIAL
6-1 If factory repair of defective equipment is re-
quired, proceed as follows:
a. Secure a return authorization from a Rose-
mount Analytical Sales Office or Repre­sentative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
In no event will Rosemount be responsible for equipment returned without proper authorization and identification.
b. Carefully pack the defective unit in a sturdy
box with sufficient shock absorbing material to ensure no additional damage occurs during shipping.
c. In a cover letter, describe completely:
1. The symptoms that determined the equipment is faulty.
2. The environment in which the equip­ment was operating (housing, weather, vibration, dust, etc.).
3. Site from where the equipment was removed.
4. Whether warranty or nonwarranty service is requested.
5. Complete shipping instructions for the return of the equipment.
d. Enclose a cover letter and purchase order
and ship the defective equipment according to instructions provided in a Rosemount Return Authorization, prepaid, to:
Rosemount Analytical Inc. RMR Department 1201 N. Main Street Orrville, Ohio 44667
If warranty service is requested, the defec­tive unit will be carefully inspected and tested at the factory. If failure was due to conditions listed in the standard Rosemount warranty, the defective unit will be repaired or replaced at Rosemount's option, and an operating unit will be returned to the cus­tomer in accordance with shipping instruc­tions furnished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 6-1
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
6-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
7
IB-106-340AC Rev. 1.1
SPS 4000
SECTION 7
REPLACEMENT PARTS
Table 7-1. SPS 4000 Replacement Parts
Figure and Index No. Part Number Description
Figure 1-4 6292A97H03 Check Valve Figure 4-1, 29 1A99089H01 Cover O-ring Figure 4-1, 46 771B635H01 Flowmeter Assembly, Calibration Gas Figure 4-1, 45 771B635H02 Flowmeter Assembly, Reference Air (Optional) Figure 4-1, 1 1A99094H01 Reference Air Pressure Regulator (Optional) Figure 4-1, 25 1A97913H03 Fuse, 5A, 250V, 5  20 mm, Slow Blow Figure 4-1, 31 4850B56G01 Interface Board Figure 4-1, 23 4850B54G01 Power Supply Board Figure 4-1, 27 7305A67H01 Pressure Switch Figure 4-1, 24 and 30 3D39435G01 Solenoid Figure 4-1, 18 4850B75H01 Terminal Cover Gasket Figure 4-1, 16 1A99147H01 Terminal Strip
April 2001
Table 7-2. Calibration Replacement Parts
Figure and Index No. Part Number Description
Figure 1-4 1A99119G01 Calibration Gas Bottles — 0.4% and 8% O2, balance nitrogen
— 550 liters each, includes bottle rack*
Figure 1-4 1A99119G02 Two flow regulators (for calibration gas bottles)
*Calibration gas bottles cannot be shipped via airfreight.
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-1
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
7-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
SPS 4000
I
This index is an alphabetized listing of parts, terms, and procedures having to do with the Haz­ardous Area Oxygen/Combustibles Transmitter. Every item listed in this index refers to a location in the manual by one or more page numbers.
Instruction Manual
IB-106-340AC Rev. 1.1
April 2001
SECTION 8

INDEX

A
Ambient Temperature Range, 1-4, 2-1 AMS, 1-6, 3-2, 5-1, 5-2, 5-3 Autocalibration, 1-1, 1-2, 1-3, 1-6 Automatic Calibration, 3-2
C
Calibration Gas, 1-5, 1-6, 2-1, 3-1, 3-2, 4-1, 4-6, 4-7,
5-2, 5-3 Calibration Gas Bottles, 1-2, 7-1 Check Valve, 2-1, 3-1, 4-6, 5-2, 5-4 Condensation Drain, 4-6 Contact Input, 1-4, 2-3
D
Drain Valve, 1-6
E
Electrical Classification, 1-2, 1-4 Electrical Connections, 2-3, 2-4 Electrical Feedthroughs, 1-4, 1-5 Electrical Noise, 1-4 Explosion-proof Rating, 1-4
F
Flow Regulators, 1-2, 7-1 Flowmeter, 1-6, 2-1, 3-2, 4-3, 4-6, 4-7, 5-2, 5-3, 5-4 Fuse, 4-1, 5-2, 5-3
G
Gas Connections, 2-1
H
Handshake Signal, 1-4, 2-3, 3-2 HART, 1-6, 3-2, 5-1, 5-2, 5-3 Humidity, 1-4
I
Input Power, 1-4 Interface Board, 1-6, 2-4, 4-1, 4-3, 4-4, 5-1, 5-2, 5-4
L
Line Voltage, 2-3 Logic I/O, 2-3, 2-4, 3-2, 5-2, 5-3
M
Manifold, 1-4, 1-5, 2-1 Mechanical Installation, 2-1 Mounting, 1-4, 2-1 Multiprobe Sequencers, 1-3
P
Piping Distance, 1-4 PLD, 1-6 Power Consumption, 1-4 Power Supply Board, 1-5, 4-1, 4-3, 4-4, 5-2, 5-3, 5-4 Pressure Regulator, 1-6, 2-1, 3-2, 4-3, 4-6, 4-7 Pressure Switch, 1-5, 4-3, 4-5, 5-2, 5-4 Product Matrix, 1-1, 1-2
R
Reference Air, 1-1, 1-2, 1-6, 2-1, 4-3, 4-6, 4-7 Relay Output, 2-1, 2-3 Remote Contact, 1-6, 3-2 Replacement Parts, 7-1 Return of Material, 6-1
S
Semi-automatic Calibration, 3-1, 3-2 Solenoid, 1-5, 1-6, 4-3, 4-5, 5-2, 5-4 Specifications, 1-4
T
Terminal Strip, 1-6 Troubleshooting, 5-1, 5-2, 5-3, 5-4
W
Weight, 1-4
Rosemount Analytical Inc. A Division of Emerson Process Management Index 8-1
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
8-2 Index Rosemount Analytical Inc. A Division of Emerson Process Management

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the com­mercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WAR­RANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DE­LAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WAR­RANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF AC­TION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDEN­TAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVE­NUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control.
3491
3536/4-01
Instruction Manual
IB-106-340AC Rev. 1.1 April 2001
SPS 4000
SPS 4000
Emerson Process Management
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
Pa rt n o. ____________
Se ria l no . ____________
Or d er no. ____________
Fisher-Rosemount GmbH & Co.
Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.
1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
© Rosemount Analytical Inc. 2001
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd.
Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount
Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...