Rosemount ROX GT and GP O2 Analyzers-Rev C Manuals & Guides

Page 1
Rosemount Analytical
ROX GT ROX
O
XYGEN ANALYZERS
I
NSTRUCTION MANUAL
748375-C
GP
Page 2
OTICE
N
The information contained in this document is subject to change without notice.
Swagelok® is a registered trademark of Crawford Fitting Co.
Manual Part Number 748375-C August 2000 Printed in U.S.A.
Rosemount Analytical Inc.
4125 East La Palma Avenue Anaheim, California 92807-1802
Page 3
C
PREFACE
PURPOSE/SAFETY SUMMARY........................................................................1
SPECIFICATIONS - ROX GT TRACE OXYGEN ANALYZER............................ 3
SPECIFICATIONS - ROX GP PERCENT OXYGEN ANALYZER......................4
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE............5
RETURNING PARTS TO THE FACTORY.........................................................5
TRAINING ......................................................................................................5
DOCUMENTATION............................................................................................5
COMPLIANCES .................................................................................................6
ONTENTS
SECTION 1. INTRODUCTION
1.1 OVERVIEW.................................................................................................1
1.2 TYPICAL APPLICATIONS ..........................................................................1
1.3 THEORY OF TECHNOLOGY......................................................................2
SECTION 2. INSTALLATION
2.1 UNPACKING...............................................................................................5
2.2 ASSEMBLY................................................................................................. 5
2.3 LOCATION..................................................................................................6
2.4 GASES ......................................................................................................6
2.4.1 Connections...................................................................................6
2.4.2 Gas Specifications.........................................................................7
2.5 ELECTRICAL CONNECTIONS...................................................................9
2.5.1 Line Power Connection..................................................................9
2.5.1.1 115/230 VAC Operation.................................................9
2.5.1.2 24 VDC Operation..........................................................9
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
i
Page 4
ONTENTS
C
2.5.2 Interconnections............................................................................ 10
2.5.3 Signal Output Connections............................................................ 11
2.5.4 Concentration Alarms.................................................................... 11
2.5.5 Off Range Contact ....................................................................... 11
2.6 INSTALLATION GUIDELINES.................................................................... 12
SECTION 3. STARTUP AND OPERATION
3.1 STARTUP................................................................................................... 13
3.2 CALIBRATION............................................................................................ 13
3.3 OPERATION............................................................................................... 14
3.4 ALARM SETPOINTS .................................................................................. 14
SECTION 4. MAINTENANCE AND TROUBLESHOOTING
4.1 MAINTENANCE OVERVIEW .................................................................... 17
4.2 OXYGEN SENSOR REPLACEMENT......................................................... 17
4.3 POWER FUSE REPLACEMENT................................................................ 20
4.3.1 115/230V AC................................................................................. 20
4.3.2 24V DC ..................................................................................... 20
4.4 PRINTED CIRCUIT BOARDS..................................................................... 21
4.5 ALARM CONFIGURATION......................................................................... 21
4.6 TROUBLESHOOTING................................................................................ 22
S
ECTION
5.1 ROX REPLACEMENT PARTS ................................................................... 25
5. R
5.1.1 Replacement Sensors................................................................... 25
5.1.2 Replacement Parts List.................................................................25
EPLACEMENT PARTS
APPENDIX A. MATERIAL SAFETY DATA SHEETS
MATERIAL SAFETY DATA SHEET (PN 748378) MATERIAL SAFETY DATA SHEET (PN 748380) MATERIAL SAFETY DATA SHEET (PN 748381)
ii
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 5
ONTENTS
C
ENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS CYLINDERS
G
ARRA NTY
W
IELD SERVICE AND REP A IR FACILIT IES
F
FIGURES
1-1. Electrochemical Sensor Technology..........................................................2
1-2. ROX GT Front View....................................................................................3
1-3. ROX GP Front View...................................................................................3
2-1. ROX Top View (cover removed..................................................................5
2-2. Back Panel Connections............................................................................7
2-3. Recommended Sample Handling Diagram -
Positive Pressure Sample.............................................................8
2-4. Recommended Sample Handling Diagram -
Insufficient Pressure or Flow........................................................8
2-5. Rear Panel Electrical Input Connection.....................................................10
4-1. Sensor Installation......................................................................................19
4-2. ROX GT, GP 115/230V AC Fuse Location.................................................20
4-3. ROX GT 24V DC Fuse Location.................................................................21
4-4. Alarm Switch Location...............................................................................22
TABLES
4-1. Alarm Setting Configuration........................................................................ 22
A-1. Sensor Reference Guide ............................................................................A1
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
iii
Page 6
ONTENTS
C
NOTES
iv
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 7
I
NTRODUCTION
1
1.1 OVERVIEW
This manual describes the ROX GT trace oxygen analyzer and the ROX GP percent oxygen analyzer of the ROX Series of oxygen analysis instruments.
The ROX GT is designed to determine continuously the concentration of trace oxygen in a flowing gaseous mixture. The concentration is expressed in parts-per-million by volume. The ROX GP is designed to measure continuously the concentration of percent oxygen in a flowing gaseous mixture.
The ROX GT / GP is designed for panel mount or 1/2 19” rack mount, with gas connections made from the rear. All electronic connections are also made from the rear including the AC power input.
1.2 TYPICAL APPLICATIONS
Typical applications for the ROX GT include:
Monitoring trace oxygen contamination in pure nitrogen or argon streams from air separation facilities.
Determination of trace oxygen content of inerting atmospheres in heat treat furnaces.
Monitoring inert atmosphere glove boxes for oxygen impurity
Typical applications for the ROX GP include:
Measuring percent impurities in pure gases
Controlling inerting atmospheres in heat treat applications
Monitoring oxygen enriching or deficient operations
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
1
Page 8
NTRODUCTION
I
1.3 THEORY OF TECHNOLOGY
The ROX GT / GP uses an electrochemical sensor technology to achieve the measurement of oxygen. See Figure 1-1. The sensor is a self contained disposable unit which requires no maintenance. The sensor utilizes the principle of electrochemical reaction to generate a signal proportional to the oxygen concentration in the sample.
The sensor consists of a cathode and anode which are in contact via a suitable electrolyte. The sensor has a gas permeable membrane which covers the cathode allowing gas to pass into the sensor while preventing liquid electrolyte from leaking out.
As the sample diffuses into the sensor, any oxygen present will dissolve in the electrolyte solution and migrate to the surface of the cathode. The oxygen is reduced at the cathode. Simultaneously, an oxidation reaction is occurring at the anode generating four electrons. These electrons flow to the cathode to reduce the oxygen.
The representative half cell reactions are: Cathode:
4e- + 2H2O + O2 4OH
-
Anode:
4OH- + 2Pb 2PbO + 2H2O + 4e
-
The resultant overall cell reaction is:
2Pb + O2 2PbO
This flow of electrons constitutes an electric current which is directly proportional to the concentration of oxygen present in the sample. In the absence of oxygen, no oxidation / reduction reaction occurs and therefore no current is generated. This allows the sensor to have an absolute zero.
Oxygen
Gas PermeableMembrane
Thin Electrolyte Layer
Electrolyte
Anode
OH
-
-
e
Cathode
F
IGURE
2
1-1. E
Current Signal
Contact Plates
LECTROCHEMICAL SENSOR TECHNOLOGY
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 9
ALARM SETPOINT ADJUSTMENT
NTRODUCTION
I
F
IGURE
ALARM SETPOINT BUTTON
1-2. ROX GT F
CELL BLOCK COVER
RANGE SELECTOR SWITCH
RONT VIEW
ALARM SETPOINT ADJUSTMENT
LARM ETPOINT UTTON
F
IGURE
748375-C Rosemount Analytical August 2000
1-3. ROX GP F
RONT VIEW
CELL BLOCK COVER
RANGE SELECTOR SWITCH
ROX GT/ROX GP Oxygen Analyzers
3
Page 10
NTRODUCTION
I
N
OTES
4
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 11
I
NSTALLATION
2
2.1 UNPACKING
Carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or contents is damaged. Retain the carton and packing material until all components associated with the instrument are operational.
2.2 ASSEMBLY
The ROX GT comes fully assembled with sensor installed. Please note that the gas inlet and outlet connections are sealed to prevent exposure of the sensor to air. Prolonged exposure of the sensor to air can cause extended start up time, reduction of performance or damage to the sensor. Do not remove the sealing caps until all
associated sample handling components are installed and the instrument is fully ready for installation.
The ROX GP requires that the sensor be installed after the analyzer has been installed and gas connections made.
F
IGURE
Cell Block
2-1. ROX T
Meter Display Alarm Setting
Power Board
OP VIEW (COVER REMOVED
)
Switch
Signal Board
Power Entry Module
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
5
Page 12
NSTALLATION
I
2.3 LOCATION
Install the ROX GT/GP in a clean, weather-protected, non-hazardous, vibration-free location free from extreme temperature variations. For best results, install the analyzer near the sample take off point to minimize sample transport time.
Operating ambient temperature is 5 °C to 45 °C, limited to temperature changes of less than 10 °C/hr. Acceptable dew point range is less than 95% relative humidity, but not in excess of 45 °C wet bulb temperature.
2.4 GASES
During normal operation, the analyzer requires no support gases. Calibration gases of air or an appropriate mixture of oxygen in nitrogen are recommended (ppm level for ROX GT, % level for ROX GP).
After initial startup or startup following a prolonged shutdown, the analyzer may require extended time to recover to the range of measurement. Commonly, this is caused by the introduction of ambient air into the sample and/or vent lines to the sensor. The presence of higher than normal levels of oxygen at the sensor will cause the sensor electrolyte to become saturated with dissolved oxygen. When the instrument is placed in operation, the sensor must now consume all excess dissolved oxygen above the desired measuring level. This recovery period is required only for the ROX GT.
All new external gas tubing is strongly recommended, preferably pre-cleaned, stainless steel or copper tubing.
Note Do not use plastic tubing for trace oxygen measurements as it can permeate oxygen from the ambient air and cause higher than expected oxygen readings.
ROX GT gas line connections are compression fittings. Do not use pipe thread tape. ROX GP gas line connections are female NPT and the use of pipe thread tape is recommended.
2.4.1 C
ONNECTIONS
See Figure 2-2. Connect sample inlet and outlet lines to appropriately labeled fittings on the rear panel. All connections are 1/4-inch ferrule-type compression fittings on the ROX GT. The ROX GP has 1/8-inch female NPT connections.
If sample is available at positive pressure and flow between .1 and 2 SCFH we recommend following the sample handling configuration shown in Figure 2.3.
If sample is available at ambient pressure or flow is insufficient to meet minimum flow requirements, we recommend following the sample handling configuration shown in
6
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 13
NSTALLATION
A
I
Figure 2.4. The inclusion of fast loop bypass, particulate filtration and moisture removal are not shown in these drawings and are the responsibility of the customer to include as required.
F
IGURE
2.4.2 G
ALIBRATION GASES
C
VENT (ROX GT) 1/4" SWAGELOK (ROX GP) 1/8" FPT
SAMPLE IN (ROX GT) 1/4" SWAGELOK (ROX GP) 1/8" FPT
2-2. B
ACK PANEL CONNECTIONS
AS SPECIFICATIONS
1 2
+ -
3 4
C NO
5 6 + -
7 8 9
C NC NO
10 11 12
C NC NO
LARM CONTACT #1
0 - 1VDC OUTPUT OFF RANGE CONTACT
4-20mA ISOLATED OUTPUT
FUSE
POWER SWITCH
Calibration method and gases depends on the desired operating range, and speed of calibration recovery. Due to the characteristics of the sensor, no zero gas is required. In the absence of oxygen, the sensor has an absolute zero.
Span gas can be either air or a certified blend of known ppm concentration of oxygen in nitrogen. All calibration gases are introduced through the sample inlet at the rear of the module
AMPLE GAS
S
Sample gas should be nonflammable (below 100% of the sample's LEL) and inert. Consult factory if sample gas contains CO2, CO, acid gases, or halogen gases. These samples may not be suitable for use with this analyzer.
LOW RATE
F
The sample flow rate can set be between .1 SCFH and 2.0 SCFH (50 - 1000 cc/min) without effecting the accuracy of the sensor. Sample flow should be set at a constant value within these limits which provides the desired response time and sample conservation.
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
7
Page 14
NSTALLATION
I
RESSURE/FILTRATION
P
Sample pressure at the inlet should be within the range of 0 to 15 psig (1013 to 2048 hPa).
The analyzer does not contain any integral sample filtration. Sample should be filtered for particulates down to five microns to reduce the risk of internal sample line blockage.
Sample Vent
F
IGURE
2-3. R
S
Flowmeter
Sample Inlet
Air or Span
Gas
Sensor
Isolation
Valve
Flow Control
Valve
ECOMMENDED SAMPLE HANDLING DIAGRAM
AMPLE
Vacuum Breaker
Eductor
ROX GT
(back)
- P
OSITIVE PRESSURE
Nitrogen or Air In
F
IGURE
8
2-4. R
P
Flowmeter
Sample
Vent
Sample Inlet
Ambient Air or
Span Gas
Flow Contr ol
Valve
Sensor
Isolation
Valve
ECOMMENDED SAMPLE HANDLING DIAGRAM
RESSURE OR FLOW
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
ROX GT
(back)
- I
NSUFFICIENT
Page 15
2.5 ELECTRICAL CONNECTIONS
All electrical connections can be found on the back of the analyzer (Refer to Figure
2.2).
NSTALLATION
I
2.5.1 L
INE POWER CONNECTION
2.5.1.1 115/230 VAC O
The ROX GT/GP is designed to operate on 115 or 230 VAC 50/60 Hz power. The power entry module is located on the right rear panel of the instrument. The power entry module utilizes a standard IEC 320 connector allowing universal connection in all countries by selecting the appropriate line cord at installation. A standard North American power cord is provided with the analyzer.
The analyzer comes configured from the factory for operation on 115 VAC power. If you need to operate the unit on 230 VAC, check to make sure that the unit is disconnected from line power, open the cover door on the power entry module and extract the fuse-block as explained in Section 4.3. Turn the fuse-block over and reinstall so that the “230V” is now visible through the window in the cover door. Close the cover door prior to connecting power.
2.5.1.2 24 VDC O
The ROX GT 24 VDC Analyzer (PN 658301-24) is designed to operate on 24 VDC input.
PERATION
PERATION
Prior to making electrical connections to the analyzer, verify that the power rating on the instrument name-rating label is correct.
The 24 VDC electrical connection is located on the right rear panel of the instrument, see Figure 2-5. The terminal strip is labeled:
+
-
+24 VDC
-24 VDC Ground
WARNING: POSSIBLE ELECTRICAL DAMAGE
Connection of AC voltage to the power input connections will result in serious damage to the analyzer. Verify all electrical input connections for correct voltage and power before powering up analyzer.
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
9
Page 16
NSTALLATION
I
F1A - 24V
FUSE
+
-
24 VDC POWER CONNECTION
F
IGURE
2.5.2 I
2-5. R
NTERCONNECTIONS
EAR PANEL ELECTRICAL INPUT CONNECTION
The analyzer has two kinds of output connections, a signal output and a set of relay contacts. The contacts are rated at 115 VAC, but with the following provisos:
There is only "basic insulation (within the meaning of IEC 1010-1) provided between the contacts and the rest of the circuitry. This means that in the event of a single fault, it is possible that 115 VAC connected to the relay contacts would be connected to the output connections as well. Any equipment connected to the analyzer output may potentially see this 115 VAC.
10
WARNING: ELECTRICAL SHOCK HAZARD
The output connection is for use only with equipment which has no live parts which are accessible.
The connecting cable and the equipment to which the output is connected must have insulation rated for at least 150 VAC, since under fault condition the output may be connected to the relay contacts witch themselves may be connected to 150 VAC.
The connection used at the remote end of the output circuit must be such as to be suitable for 150 VAC and must have no accessible live parts.
The equipment connected to the output must either be approved to IEC 1010-1 or equivalent or must be suitable for use with an input that may potentially be connected to 150 VAC, and must not catch fire in this circumstance.
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 17
NSTALLATION
I
Note The interconnection terminal strip on the rear of the analyzer is designed to be
unplugged for easier connection. It is polarized so that you cannot plug it in backwards. Make the connections with plenty of service loop and then plug it in. Double check that you have used the correct terminals, and do not apply a high voltage (e.g. 115 V) to the relay contacts before you have verified that the output connections are correct - putting AC on the signal output will destroy the analyzer.
2.5.3 S
IGNAL OUTPUT CONNECTIONS
The analyzer comes standard with both 4-20 mA DC isolated and 0-1 VDC outputs. For ease of installation, the signal output connections are made via a removable terminal strip connector which is shared with the alarm connections (Refer to Figure
2.2).
WARNING: POSSIBLE EXPLOSION HAZARD
Check all wiring connections prior to reconnection of the terminal strip. The application of AC power to the signal output terminals will damage the analyzer.
2.5.4 C
ONCENTRATION ALARMS
The analyzer comes standard with two concentration alarms both configured high and normally energized. For ease of installation, the alarm connections are made via a removable terminal strip connector which is shared with the signal output connections (Refer to Figure 2.2). Each set of alarm connections are labeled for easy identification to the associated alarm. The connections labeled NO (normally open), NC (normally closed), and C (common) refer to relay positions with the power removed (de­energized) .
Both alarms are configured from the factory for fail-safe mode (normally energized, and loss of power will cause a change in relay status) and high oxygen setpoint (relay status change above setpoint). To change the alarm configuration, refer to Section
4.5.
2.5.5 O
FF RANGE CONTACT
Additionally, there is an off range contact available. This contact provides the user with a contact that changes status when the range switch is moved out of the desired range. This contact comes from the factory configured normally energized on the 0-10 ppm range. In the event of a loss of power or the range switch is moved to a different position, the contact will open.
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
11
Page 18
NSTALLATION
I
2.6 INSTALLATION GUIDELINES
Is the analyzer’s location clean, weather-protected, non-hazardous, vibration-free, and with a stable ambient temperature?
Are gas supply cylinders equipped with a clean, metal diaphragm, hydrocarbon free, two stage regulator with shut off valve?
Are external tubing, regulators, valves, pumps, fittings etc. clean?
Is the oxygen content of the supply gases compatible with the analysis range?
Is the sample non-flammable?
Have all the external gas connections been leaked checked?
Has clean stainless steel or other appropriate metal tubing been used for sample
lines?
Are all electrical connections secure and correct?
Has the sensor been installed in the ROX GP?
Is the proper line input power displayed on the power entry module?
12
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 19
S
TARTUP AND OPERATION
3
3.1 STARTUP
After completion of the installation procedure in Section Two, proceed as follows:
1. Set range switch to the 25% position (full counterclockwise)
2. Turn on power by placing the switch on the power entry module to the “-” position.
3.2 CALIBRATION
See Section 2.4.2 for a description of the method for choosing calibration zero and span gases.
To calibrate the ROX GT/GP, introduce air or span gas into the SAMPLE INLET, and do the following:
1. Set range switch to the 25% range setting (full counterclockwise) or the appropriate range for the span gas used. It is recommended to use the highest concentration span gas which provides you with the best fit for your requirements. The use of air for calibration provides the least chance of span gas error, but does require extended recovery time down to low ppm level measurement.
2. If you have followed the recommended sample handling diagram in Section 2.4.1. open the sensor isolation valve first to avoid exposing the sensor to excessive sample pressure or vacuum. Flow span gas or air at a floweret between 0.1 and
1.0 L/min. Be careful to not obstruct the outlet as this will cause an increase in sample pressure and a consequent increase in oxygen reading.
Note Alarm setpoints are not defeated during calibration. Take appropriate
precautions during calibration.
3. Wait f o r reading to stabilize. This should only take a few minutes.
4. Locate the SPAN ADJUST potentiometer on the front panel and adjust appropriately to make the display correspond with the span gas concentration.
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
13
Page 20
TARTUP AND OPERATION
S
Turning the potentiometer clockwise will increase the display reading while counterclockwise will decrease the reading.
5. If you have used air to span the instrument, switch to zero gas or sample as soon as possible to avoid prolonged exposure of the sensor to high concentrations of oxygen. The longer the sensor is exposed to air, the longer it will take for the sensor to recover to low ppm levels. W hen installing a new sensor or starting the instrument for the first time, it may take as long as eight hours for the analyzer to purge down to the lowest operating range.
6. Allow the unit to stabilize at the final reading on zero gas or sample. The unit is now ready for operation.
Unit should be calibrated once a week until familiarity is achieved in required calibration frequency. Calibration intervals as long as once a month are possible on stable installations. Processes which have large changes in environmental conditions and sample composition may require more frequent calibrations.
3.3 OPERATION
After calibration, proceed as follows: Supply sample gas to the SAMPLE INLET. Adjust external flow controller or throttle
valve so that flow discharged from the outlet is between 0.1 and 1.0 L/min.
Note Flow indication is not provided with the instrument and must be provided
external to the analyzer. Refer to installation section for proper location.
When shutting down the instrument, always close the sensor isolation valve last to avoid exposing the sensor to excessive pressure or vacuum.
3.4 ALARM SETPOINTS
The ROX GT/GP come standard with two fully adjustable alarms with relay contacts. These alarm setpoints are designed to operate as percent of range as opposed to discrete concentration value. The alarms are configured from the factory as high and failsafe.
To view the alarm setpoints, turn the range switch to the desired measuring range and press the “SET LO ALARM” or “SET HI ALARM” button. The alarm setpoint will be displayed in actual concentration units. Since the alarms are configured as percent of scale, changing ranges will also change the value of the alarm setpoint. For example, if the LO alarm is set for 5 ppm on the 10 ppm range (50% of range), it will read 50 ppm on the 100 ppm range (50% of range).
14
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 21
TARTUP AND OPERATION
S
To adjust the alarm setpoint, press the appropriate alarm setpoint button, then locate and adjust the associated alarm setpoint adjustment potentiometer as indicated in Figure 1-1. Turning the potentiometer clockwise will increase the setpoint while turning the potentiometer counterclockwise will decrease the setpoint. You must keep the alarm setpoint button depressed during the adjustment procedure to continuously view the alarm setpoint value. Always adjust the “LO” alarm for the lowest setpoint value and the “HI” alarm for the higher value.
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
15
Page 22
TARTUP AND OPERATION
S
N
OTES
16
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 23
M
AINTENANCE AND TROUBLESHOOTING
4
CAUTION: QUALIFIED PERSONNEL
This equipment should not be adjusted or repaired by anyone except properly qualified personnel.
WARNING: PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect the safety and performance of this product. Use only factory approved components for repair.
4.1 MAINTENANCE OVERVIEW
The ROX GT/GP analyzer components which may require replacement include:
Oxygen sensor
Power fuse
Printed circuit boards
4.2 OXYGEN SENSOR REPLACEMENT
CAUTION: CAUS TIC MATERIAL
The sensor contains electrolyte which is caustic and can cause serious burns to skin. Do not ingest contents of sensor. Read and understand the Material Safety Data Sheet in the rear of this manual before handling sensor.
The oxygen sensor used in the ROX GT/GP analyzers is a consumable item and does require replacement periodically. To determine if the sensor requires replacement, see the troubleshooting portion of this section.
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
17
Page 24
AINTENANCE AND TROUBLESHOOTING
M
ENSOR INSTALLATION
S
To install a new sensor follow the procedures exactly in the order listed below. Failure to follow the procedure in order may result is extended exposure of the sensor to air causing greater delays in returning to on-line monitoring:
1. Turn off power to unit and turn off sample flow to analyzer.
2. Locate cellblock cover and remove using the key tool provided with instrument. Rotate cover counterclockwise to remove.
3. Inspect cellblock cavity for signs of liquid. If liquid is present, use protective equipment as described in the material safety data sheet in the rear of this manual before proceeding. Extreme care should be used and the liquid should be treated as a caustic substance.
4. Remove existing sensor and dispose of in accordance with national, federal, state, and local regulations. Use the existing pull-tab on the sensor to remove from the cellblock.
5. Inspect inside of cellblock for signs of residual liquid or deposits on the contact pins. If liquid is present, use appropriate protective equipment and follow the cleanup procedure as described in the material safety data sheet in the rear of this manual. Deposits on the contact pins can be removed by wiping with a damp cloth or using the eraser from a pencil. Do not use abrasives (i.e. sandpaper) as this will damage the contact pins.
6. If sample is inert (e.g., nitrogen) a low flow of sample (50 cc/min) can be introduced to the analyzer to purge out the internal volume of the cellblock prior to installation of the new sensor. This procedure is not required and is recommended as a step to reduce the purge down time of the new sensor.
CAUTION: MAXIMUM FLOW
Do not flow a volume in excess of 50 cc/min. Exceeding this level may produce an oxygen deficient breathing environment to the technician conducting sensor replacement. Do not flow sample if it is flammable or toxic. Do not use this procedure if the sample handling configuration is for insufficient pressure or flow.
18
7. Inspect the package containing the replacement sensor. Do not open the bag. Make sure that it is the correct model type and that there is not visible liquid in the bag. If the bag contains liquid, do not open and contact the factory for a replacement.
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 25
AINTENANCE AND TROUBLESHOOTING
M
8. Locate the shorting tab on the back of the sensor. The tab is designed to maintain contact between the two contact plates of the sensor while it is stored in the bag. This contact allows the sensor to continuously consume dissolved oxygen inside the sensor. This clip will need to be removed prior to installing the sensor in the analyzer.
9. Locate the pull-tab portion of the sensor label. W hen installing the sensor in the cellblock, make sure that this tab faces forward for easy sensor removal later.
10. Cut open the bag being careful not to damage the sensing face of the sensor. Quickly remove the sensor from the bag, remove the shorting clip and install in the cellblock with the contact plates on top and the sensing face on the bottom. Make sure that the pull-tab is facing forward and the writing on the label is right side up (see Figure 4-1).
11. Reinstall the cellblock cover turning clockwise until the O-ring just makes contact. Turn no more than 1/4 turn additional.
Note Do not overtighten. Doing so may damage the cover.
12. T urn on power and check reading.
13. Calibrate as described in Section 3.2
Key Slot
Cell Block Cover
Contact Plates
Cell Block (with cover removed)
Oxygen Sensor
Sensing Face
F
IGURE
748375-C Rosemount Analytical August 2000
4-1. S
ENSOR INSTALLATION
Sensor Pull Tab
ROX GT/ROX GP Oxygen Analyzers
19
Page 26
AINTENANCE AND TROUBLESHOOTING
M
4.3 POWER FUSE REPLACEMENT
4.3.1 115/230 V AC
The ROX GT/GP has the main power fuses located in the power entry module. No other fuses are located inside the instrument.
Before replacing the main power fuse, turn off the analyzer and disconnect the power cord from the power entry module.
Using a small blade slotted screwdriver, carefully pry open the f ront door of the power entry module. There is a access slot on the top of the module and the door opens downward (see Figure 4-2). Once the door is opened, you can remove the red power selector / fuseholder by inserting the screwdriver into the slot on top of the fuseholder and carefully prying the fuseholder out. The fuseholder contains two 1A fuses. Replace as required.
Reinstall the fuseholder so that the correct voltage is visible through the small window in the door of the power entry module.
Access slot
Fuseholder (Red)
Power Entry Module
(with front door open)
Rear Panel
F
IGURE
4-2. ROX GT, GP 115/230V AC F
USE LOCATION
4.3.2 24V DC
The ROX GT 24V DC Analyzer has the main power fuse located directly above the electrical power input terminal strip on the rear panel, Figure 4-3. No other fuses are located in the analyzer.
20
Before replacing the main power fuse, turn off and disconnect electrical power to the analyzer.
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 27
AINTENANCE AND TROUBLESHOOTING
M
The fuseholder utilizes a standard twist-off cap. To replace the fuse, rotate the fuseholder cap counterclockwise until the cap is free from the body. The fuse will come out with the cap. Replace fuse and re-install cap onto fuseholder.
F1A - 24V
FUSE
+
-
Rear Panel
F
IGURE
4-3. ROX GT 24V DC F
USE LOCATION
4.4 PRINTED CIRCUIT BOARDS
Both printed circuit boards can be replaced if necessary. Refer to Figure 2-1 for the location of the power and signal boards.
4.5 ALARM CONFIGURATION
The alarm setpoints are factory set for high normally energized. If you wish to change the configuration of the alarm setpoints you can do so by the following procedure.
1. Verify that the power module is off and the power cord is disconnected from the analyzer
2. Remove the top cover from the analyzer
3. Locate the signal board as indicated in Figure 2-1
4. Locate the alarm setting dip switches on the signal board
5. To change the alarm setting refer to Figure 4-4 and Table 4-1 for desired configuration
In all cases, “LO ALARM” should be adjusted for the lowest setpoint value and “HI ALARM” should be adjusted for the higher setpoint value.
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
21
Page 28
AINTENANCE AND TROUBLESHOOTING
M
F
IGURE
4-4. A
ON OFF
4 3 2 1
EDGE OF SIGNAL BOARD
EDGE OF SIGNAL BOARD
LARM SWITCH LOCATION
LO ALARM SETTING (S2)
HI ALARM SETTING (S3)
T
ABLE
TRIP POINT - FS/NFS SWITCH #
(RELAY STATUS) 1 2 3 4
HIGH - FAILSAFE (NORMALLY ENERGIZED) HIGH - NON FAILSAFE (NORMALLY DE-ENERGIZED) LOW - FAILSAFE (NORMALLY ENERGIZED) LOW - NON FAILSAFE (NORMALLY DE-ENERGIZED)
4-1. A
LARM SETTING CONFIGURATION
OFF ON ON OFF
ON OFF OFF ON
ON OFF OFF ON
OFF ON ON OFF
22
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 29
AINTENANCE AND TROUBLESHOOTING
M
4.6 TROUBLESHOOTING
The following provides a short list of common troubleshooting tips.
NALYZER DOES NOT POWER UP
A
Check to make sure that the power cord is properly installed and the voltage is correctly selected. The voltage configuration of the analyzer is visible through the window of the power entry module. Make sure that the power switch is in the on position (“-”).
If the unit does not power up, turn off power switch, remove power cord and open door on power entry module as described in Section 4.4 above. Inspect fuses and replace as required.
NALYZER FAILS TO PURGE DOWN TO PPM LEVELS
A
The number one problem associated with trace oxygen analyzer installation is the occurrence of leaks in the sample line plumbing. If the oxygen reading will not come down to ppm levels or is reading higher than expected, the sample plumbing prior to the instrument may have a leak. A quick check can be conducted by observing the oxygen reading at two different flow levels; 0.2 and 2.0 SCFH. If the oxygen reading drops significantly when the flow is increased from 0.2 to 2.0 SCFH, this is a good indication that a leak exists.
To check for leaks prior to the sensor, disconnect the analyzer and cap the inlet and outlet fittings on the analyzer to reduce the amount of oxygen exposure to the sensor. Cap the inlet line and pressurize the inlet line to 5 - 10 psig and check all connections with a soapy solution to identify leaks.
WARNING: POSSIBLE SENSOR DAMAGE
Do not pressure check the sample line with the sensor/analyzer connected. Over-pressurization of the sensor can result in damage.
If the reading increases with an increase in flow, the vent line may be obstructed causing back pressure on the sensor. Inspect your vent line and remove any devices which may be causing backpressure.
If the reading slowly decreases over time, the sensor may have been exposed to high levels of oxygen for extended periods of time. The sensor must consume all of the dissolved oxygen in the sensor electrolyte before it will accurately measure ppm le vels of oxygen. In some cases it is not possible for the sensor to recover to low ppm levels and the sensor must be replaced.
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
23
Page 30
AINTENANCE AND TROUBLESHOOTING
M
NALYZER IS READING TOO LOW
A
Check to make sure that the sensor has been calibrated properly. Flow an appropriate span gas through the instrument and follow the calibration procedure in Section 3.2. If you are unable to achieve span calibration, the sensor has reached the end of its useful life and needs to be replaced (see Section 4.2).
If calibration is possible, but the reading still appears too low, conduct a span check on the 25% range with air. Adjust reading as necessary. If the reading requires adjustment, your bottled span gas may be incorrect or corrupted.
If sensor life seems too short, you may have compounds present in your sample which are poisoning the sensor. Acid gases, CO2, SO2, or H2S can reduce the life of the sensor and should be removed from the sample prior to measuring.
NALYZER ALWAYS READS ZERO
A
Check to make sure that the sensor has been installed correctly with the contact plates on top and the gas permeable membrane facing downward. If the sensor has been installed incorrectly, carefully remove and check to make sure that the membrane has not been damaged. If the sensor is leaking, or shows signs of distress to the membrane, replace with a new sensor. If the sensor is okay quickly reinstall in the correct position.
Verify that the cellblock contact pins are making good contact with the metal surfaces of the sensor. If the pins are not making contact, carefully bend the wire down to reestablish contact. Check that the contact pins are clean and do not show signs of deposits or oxidation. Clean as necessary.
A good sensor will have an output of 150 to 500 micro amps in air. You can check this by connecting a digital volt-meter across the contact plates of the sensor while it is removed from the instrument and measuring the current. If the sensor does not produce sufficient output, replace the sensor with a new one.
HE READINGS ARE ERRATIC AND DRIFT NOISILY
T
This is normally an indication that the sensor has reached the end of its useful life. If the analyzer cannot be spanned on a cal gas or air, the sensor must be replaced. If the sensor spans okay and appears stable at high oxygen concentrations, there may be a sample line leak or obstruction of the vent line. Follow the procedure described above in “Analyzer fails to purge down to ppm levels”.
If no sample line problems are found, open the cellblock and inspect the sensor for signs of damage or leakage. Follow the procedure in Section 4.2. for correct method of opening cellblock and precautions for handling a leaking sensor.
24
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 31
R
EPLACEMENT PARTS
5
5.1 ROX REPLACEMENT PARTS
WARNING: PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect the safety and performance of this product. Use only factory approved components for repair.
5.1.1 R
5.1.2 R
EPLACEMENT SENSORS
Refer to Table A-1 in Appendix A for easy cross reference of the correct oxygen sensor for your specific instrument.
658303 Oxygen Sensor GT1 658304 Oxygen Sensor GT3 658310 Oxygen Sensor GP1 658311 Oxygen Sensor GP3 658312 Oxygen Sensor GP4
EPLACEMENT PARTS LIST
904699 Connector Rear Output 873155 Power Cord 904759 Power Supply Board 904760 Cap. Cell Block 904761 O-Ring, Cell Block Cap 904762 Ribbon Cable 904763 Power Entry Module 099393 Fuse 1A 250V 748375 Instruction Manual
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
25
Page 32
EPLACEMENT PARTS
R
ROX GT O
NLY
904768 Cell Block Assembly 904769 Amplifier Board (658301) (658302)
ROX GP O
NLY
904764 Cell Block Assembly 904765 Amplifier Board (658307) 904766 Amplifier Board (658308) 904767 Amplifier Board (658309)
26
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Page 33
A
PPENDIX
A. M
ATERIAL SAFETY DATA SHEETS
A
The following section contains important information regarding the characteristics of the oxygen sensor used in the ROX GT/GP analyzers. Read and understand all information before attempting any replacement or installation of the oxygen sensor. The following sensor reference guide will provide a quick reference of the specific oxygen sensor type used in your specific analyzer model and part number.
T
ABLE
MODEL INSTRUMENT
PART NUMBER
ROX GT 658301 GT1 658303 ROX GT 658302 GT3 658304 ROX GP 658307 GP1 658310 ROX GP 658308 GP3 658311 ROX GP 658309 GP4 658312
A-1. S
ENSOR REFERENCE GUIDE
OXYGEN SENSOR
MODEL
OXYGEN SENSOR
PART NUMBER
748375-C Rosemount Analytical August 2000
ROX GT/ROX GP Oxygen Analyzers
A1
Page 34
PPENDIX
A
N
OTES
ATERIAL SAFETY DATA SHEETS
A. M
A2
August 2000 Rosemount Analytical 748375-CROX GT/ROX GP Oxygen Analyzers
Loading...