Rosemount ROX GT and GP O2 Analyzers-Rev B Manuals & Guides

Page 1
Rosemount Analytical
ROX GT ROX GP
O
XYGEN ANALYZERS
I
NSTRUCTION MANUAL
748375-B
Page 2
OTICE
N
The information contained in this document is subject to change without notice.
Swagelok® is a registered trademark of Crawford Fitting Co.
Manual Part Number 748375-B July 1998 Printed in U.S.A.
Rosemount Analytical Inc.
4125 East La Palma Avenue Anaheim, California 92807-1802
Page 3
C
ONTENTS
PREFACE
PURPOSE/SAFETY SUMMARY........................................................................P-1
SPECIFICATIONS - ROX GT TRACE OXYGEN ANALYZER............................ P-3
SPECIFICATIONS - ROX GP PERCENT OXYGEN ANALYZER......................P-4
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE....P-5
RETURNING PARTS TO THE FACTORY.........................................................P-5
TRAINING ......................................................................................................P-5
DOCUMENTATION............................................................................................P-5
COMPLIANCES .................................................................................................P-6
SECTION 1. INTRODUCTION
1.1 OVERVIEW..............................................................................................1-1
1.2 TYPICAL APPLICATIONS .......................................................................1-1
1.3 THEORY OF TECHNOLOGY...................................................................1-2
S
ECTION
2.1 UNPACKING............................................................................................2-1
2.2 ASSEMBLY..............................................................................................2-1
2.3 LOCATION...............................................................................................2-2
2.4 GASES.....................................................................................................2-2
2.5 ELECTRICAL CONNECTIONS................................................................2-5
2.6 INSTALLATION GUIDELINES .................................................................2-6
2. I
2.4.1 Connections................................................................................2-2
2.4.2 Gas Specifications......................................................................2-3
2.5.1 Line Power Connection...............................................................2-5
2.5.2 Interconnections .........................................................................2-5
2.5.3 Signal Output Connections.........................................................2-6
2.5.4 Alarm Connections......................................................................2-6
NSTALLATION
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
i
Page 4
ONTENTS
C
SECTION 3. STARTUP AND OPERATION
3.1 STARTUP................................................................................................ 3-1
3.2 CALIBRATION......................................................................................... 3-1
3.3 OPERATION............................................................................................ 3-2
3.4 ALARM SETPOINTS ............................................................................... 3-2
SECTION 4. MAINTENANCE AND TROUBLESHOOTING
4.1 MAINTENANCE OVERVIEW .................................................................. 4-1
4.2 OXYGEN SENSOR REPLACEMENT...................................................... 4-1
4.3 POWER FUSE REPLACEMENT............................................................. 4-4
4.4 PRINTED CIRCUIT BOARDS.................................................................. 4-4
4.5 ALARM CONFIGURATION...................................................................... 4-4
4-6 TROUBLESHOOTING............................................................................. 4-6
SECTION 5. REPLACEMENT PARTS
5.1 ROX REPLACEMENT PARTS ................................................................ 5-1
5.1.1 Replacement Sensors................................................................ 5-1
5.1.2 Replacement Parts List.............................................................. 5-1
APPENDIX A. MATERIAL SAFETY DATA SHEETS
MATERIAL SAFETY DATA SHEET (PN 748378) MATERIAL SAFETY DATA SHEET (PN 748380) MATERIAL SAFETY DATA SHEET (PN 748381)
ENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS CYLINDERS
G
ARRA NTY
W
IELD SERVICE AND REP A IR FACILIT IES
F
ii
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 5
FIGURES
1-1 Electrochemical Sensor Technology.........................................................1-2
1-2 ROX GT Front View...................................................................................1-3
1-3 ROX GP Front View..................................................................................1-3
2-1 ROX Top View (cover removed)................................................................2-1
2-2 Back Panel Connections...........................................................................2-3
2-3 Recommended Sample Handling Diagram - Positive Pressure Sample...2-4 2-4 Recommended Sample Handling Diagram - Insufficient Pressure
or Flow ..........................................................................................2-4
4-1 Sensor Installation.....................................................................................4-3
4-2 Power Entry Module..................................................................................4-4
4-3 Alarm Switch Location...............................................................................4-5
TABLES
4-1 Alarm Setting Configuration....................................................................... 4-6
A-1 Sensor Reference Guide ..........................................................................A-1
ONTENTS
C
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
iii
Page 6
ONTENTS
C
NOTES
iv
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 7
P
REFACE
PURPOSE/SAFETY SUMMARY
The purpose of this manual is to provide the procedures for the installation, operation and maintenance of the ROX GT trace oxygen analyzer and ROX GP percent oxygen analyzer..
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this analyzer before operating it. READ THIS INSTRUCTION MANUAL COMPLETELY.
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.
DANGER is used to indicate the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored
WARNING is used to indicate the presence of a hazard which can cause severe personal injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation, or maintenance information which is important but not hazard-related.
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
P-1
Page 8
REFACE
P
WARNING: ELECTRICAL SHOCK HAZARD
The output connection is for use only with equipment which has no live parts which are accessible.
The connecting cable and the equipment to which the output is connected must have insulation rated for at least 150 VAC, since under fault condition the output may be connected to the relay contacts witch themselves may be connected to 150 VAC.
The connection used at the remote end of the output circuit must be such as to be suitable for 150 VAC and must have no accessible live parts.
The equipment connected to the output must either be approved to IEC 1010-1 or equivalent or must be suitable for use with an input that may potentially be connected to 150 VAC, and must not catch fire in this circumstance.
WARNING: POSSIBLE EXPLOSION HAZARD
This equipment is not designed for analysis of flammable samples. Introduction of flammable samples into this equipment could result in explosion causing injury, death, or property damage.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair
CAUTION: CAUSTI C MATER I A L
Tampering with the oxygen sensor may result in leakage of the internal liquid electrolyte. The electrolyte is caustic and can cause severe burns to skin. Do not ingest contents of sensor. Read and understand the Material Safety Data Sheets in the rear of this manual before handling the sensor.
P-2
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 9
ROX GT TRACE OXYGEN ANALYZER SPECIFICATIONS - GENERAL
REFACE
P
RANGES: S
IGNAL OUTPUT
A
LARMS
D
ISPLAY
MIN. R
EPEATABILITY
N
OISE
L
INEARITY
R
ESPONSE TIME
R
ECOVERY TIME
Z
ERO DRIFT
S
PAN DRIFT
P
OWER REQUIREMENTS
: :
DETECTABLE LEVEL
:
:
:
:
:
:
: :
0 - 5, 10, 100, 1000 ppm, 0 - 25% Oxygen 0-1VDC (10K min) and 4-20 mADC Isolated (600 ohm
max) 2 Form “C” contact relays, 1amp @ 30 VDC 1 amp @ 115 VAC
3 1/2 Digit Reflective LCD
0.01 ppm
:
+ 1% of fullscale or .1 ppm whichever is greater + 1% + 1%
90% of fullscale in less than 10 seconds for ranges greater than 0-10 ppm. 90% of fullscale in less than 30 seconds for 0-10 ppm. 90% of fullscale in less than 60 seconds for ranges less than 0-10 ppm.
Typically less than 1 hour (air to less than 10 ppm) Less than 1% per week at constant temperature Less than 1% per week at constant temperature
:
115/230 VAC + 10%, 50/60 Hz, less than 5 W
SPECIFICATIONS - ENVIRONMENTAL
T
EMPERATURE RELATED
D
:
RIFT
A
MBIENT OPERATING
T
EMPERATURE
S
AMPLE TEMPERATURE
S
AMPLE HUMIDITY
S
PECIFICATIONS
A
REA CLASSIFICATION
G
AS CONNECTIONS
M
OUNTING CONFIGURATION
P
ANEL CUTOUT DIMENSIONS
C
ASE DIMENSIONS
W
EIGHT
:
:
:
- P
:
:
+ 2% of fullscale for temperatures between 15 - 35° C; + 5% off fullscale over the entire operating temperature range
5 to 45°C 5 to 45°C
:
0 to 95% relative humidity non-condensing
HYSICAL
:
General Purpose 1/4” Swagelok 1/2 19” or panel mount
:
3.625 X 7.625 inches (92 X 194 mm)
:
5.06 X 8.38 X 5.0 inches (129 X 213 X 127 mm) HxWxD
Approximately 5 lbs (2.27 Kg)
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
P-3
Page 10
REFACE
P
ROX GP PERCENT OXYGEN ANALYZER SPECIFICATIONS - GENERAL
RANGES: S
IGNAL OUTPUT
A
LARMS
D
ISPLAY
M
INIMUM DETECTABLE LEVEL
R
EPEATABILITY
N
OISE
L
INEARITY
R
ESPONSE TIME
Z
ERO DRIFT
S
PAN DRIFT
P
OWER REQUIREMENTS
S
PECIFICATIONS
T
EMPERATURE RELATED
D
RIFT
A
MBIENT OPERATING
T
EMPERATURE
: :
:
:
:
: :
:
:
0 - 5, 10, 25% Oxygen
:
0-1VDC (10K min) and 4-20 mADC Isolated (600 ohm max)
2 Form “C” contact relays, 1amp @ 30 VDC 1 amp @ 115 VAC
3 1/2 Digit Reflective LCD
0.01 % Oxygen
:
+ 1% of fullscale + 1% + 1%
:
90% of fullscale in less than 7 seconds (GP1 sensor) Less than 1% per week at constant temperature Less than 1% per week at constant temperature
:
- E
115/230 VAC + 10%, 50/60 Hz, less than 5 W
NVIRONMENTAL
+ 2% of fullscale for temperatures between 15 - 35°C; + 5% off fullscale over the entire operating temperature range
5 to 45°C
S
AMPLE TEMPERATURE
S
AMPLE HUMIDITY
:
:
5 to 45°C 0 to 95% relative humidity non-condensing
SPECIFICATIONS - PHYSICAL
General Purpose 1/8” Female NPT 1/2 19” or panel mount
:
3.625 X 7.5 inches (92 X 190 mm)
:
5.06 X 8.38 X 5.0 inches (129 X 213 X 127 mm) HxWxD
Approximately 5 lbs (2.27 Kg)
P-4
A
REA CLASSIFICATION
G
AS CONNECTIONS
M
OUNTING CONFIGURATION
P
ANEL CUTOUT DIMENSIONS
C
ASE DIMENSIONS
W
EIGHT
:
:
:
:
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 11
REFACE
P
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE
For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned Materials Authorization (RMA) number. Please have the following information when you call: Model Number, Serial Number, and Purchase Order Number or Sales Order
Number.
Prior authorization by the factory must be obtained before returned materials will be accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive or other hazardous material or used in such a hazardous environment, the user must attach an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
TRAINING
A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service Training Schedule contact the Technical Services Department at:
Rosemount Analytical Inc.
Phone: 1-714-986-7600
FAX: 1-714-577-8006
D
OCUMENTATION
The following ROX GT/ROX GP Oxygen Analyzers instruction materials are available. Contact Customer Service or the local representative to order.
748375 Instruction Manual (this document)
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
P-5
Page 12
REFACE
P
COMPLIANCES
This product is approved by the Canadian Standards Association (which is also an OSHA accredited, Nationally Recognized Testing Laboratory), for use in non­hazardous, indoor locations
®
NRTL/C
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May 1993).
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand.
NAMUR
N96
P-6
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 13
I
NTRODUCTION
1
1.1 OVERVIEW
This manual describes the ROX GT trace oxygen analyzer and the ROX GP percent oxygen analyzer of the ROX Series of oxygen analysis instruments.
The ROX GT is designed to determine continuously the concentration of trace oxygen in a flowing gaseous mixture. The concentration is expressed in parts-per-million by volume. The ROX GP is designed to measure continuously the concentration of percent oxygen in a flowing gaseous mixture.
The ROX GT / GP is designed for panel mount or 1/2 19” rack mount, with gas connections made from the rear. All electronic connections are also made from the rear including the AC power input.
1.2 TYPICAL APPLICATIONS
Typical applications for the ROX GT include:
Monitoring trace oxygen contamination in pure nitrogen or argon streams from air separation facilities.
Determination of trace oxygen content of inerting atmospheres in heat treat furnaces.
Monitoring inert atmosphere glove boxes for oxygen impurity
Typical applications for the ROX GP include:
Measuring percent impurities in pure gases
Controlling inerting atmospheres in heat treat applications
Monitoring oxygen enriching or deficient operations
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
1-1
Page 14
NTRODUCTION
I
1.3 THEORY OF TECHNOLOGY
The ROX GT / GP uses an electrochemical sensor technology to achieve the measurement of oxygen. See Figure 1-1. The sensor is a self contained disposable unit which requires no maintenance. The sensor utilizes the principle of electrochemical reaction to generate a signal proportional to the oxygen concentration in the sample.
The sensor consists of a cathode and anode which are in contact via a suitable electrolyte. The sensor has a gas permeable membrane which covers the cathode allowing gas to pass into the sensor while preventing liquid electrolyte from leaking out.
As the sample diffuses into the sensor, any oxygen present will dissolve in the electrolyte solution and migrate to the surface of the cathode. The oxygen is reduced at the cathode. Simultaneously, an oxidation reaction is occurring at the anode generating four electrons. These electrons flow to the cathode to reduce the oxygen.
The representative half cell reactions are: Cathode:
4e- + 2H2O + O2 4OH
-
Anode:
4OH- + 2Pb 2PbO + 2H2O + 4e
-
The resultant overall cell reaction is:
2Pb + O2 2PbO
This flow of electrons constitutes an electric current which is directly proportional to the concentration of oxygen present in the sample. In the absence of oxygen, no oxidation / reduction reaction occurs and therefore no current is generated. This allows the sensor to have an absolute zero.
Oxygen
Gas Permeable Membrane
Thin Electrolyte Layer
Electrolyte
Anode
OH
-
-
e
Cathode
F
IGURE
1-2
1-1. E
Current Signal
Contact Plates
LECTROCHEMICAL SENSOR TECHNOLOGY
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 15
ALARM SETPOINT ADJUSTMENT
NTRODUCTION
I
F
IGURE
ALARM SETPOINT BUTTON
1-2. ROX GT F
CELL BLOCK COVER
RANGE SELECTOR SWITCH
RONT VIEW
ALARM SETPOINT ADJUSTMENT
LARM ETPOINT UTTON
F
IGURE
748375-B Rosemount Analytical July 1998
1-3. ROX GP F
RONT VIEW
CELL BLOCK COVER
RANGE SELECTOR SWITCH
ROX GT/ROX GP Oxygen Analyzers
1-3
Page 16
NTRODUCTION
I
N
OTES
1-4
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 17
I
NSTALLATION
2
2.1 UNPACKING
Carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or contents is damaged. Retain the carton and packing material until all components associated with the instrument are operational.
2.2 ASSEMBLY
The ROX GT comes fully assembled with sensor installed. Please note that the gas inlet and outlet connections are sealed to prevent exposure of the sensor to air. Prolonged exposure of the sensor to air can cause extended start up time, reduction of performance or damage to the sensor. Do not remove the sealing caps until all
associated sample handling components are installed and the instrument is fully ready for installation.
The ROX GP requires that the sensor be installed after the analyzer has been installed and gas connections made.
F
IGURE
Cell Block
2-1. ROX T
Meter Display Alarm Setting
Power Board
OP VIEW (COVER REMOVED
Switch
Signal Board
Power Entry Module
)
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
2-1
Page 18
NSTALLATION
I
2.3 LOCATION
Install the ROX GT/GP in a clean, weather-protected, non-hazardous, vibration-free location free from extreme temperature variations. For best results, install the analyzer near the sample take off point to minimize sample transport time.
Operating ambient temperature is 5 °C to 45 °C, limited to temperature changes of less than 10 °C/hr. Acceptable dew point range is less than 95% relative humidity, but not in excess of 45 °C wet bulb temperature.
2.4 GASES
During normal operation, the analyzer requires no support gases. Calibration gases of air or an appropriate mixture of oxygen in nitrogen are recommended (ppm level for ROX GT, % level for ROX GP).
After initial startup or startup following a prolonged shutdown, the analyzer may require extended time to recover to the range of measurement. Commonly, this is caused by the introduction of ambient air into the sample and/or vent lines to the sensor. The presence of higher than normal levels of oxygen at the sensor will cause the sensor electrolyte to become saturated with dissolved oxygen. When the
instrument is placed in operation, the sensor must now consume all excess dissolved oxygen above the desired measuring level. This recovery period is required only for the ROX GT.
All new external gas tubing is strongly recommended, preferably pre-cleaned, stainless steel or copper tubing.
Note Do not use plastic tubing for trace oxygen measurements as it can permeate oxygen from the ambient air and cause higher than expected oxygen readings.
ROX GT gas line connections are compression fittings. Do not use pipe thread tape. ROX GP gas line connections are female NPT and the use of pipe thread tape is recommended.
2.4.1 C
ONNECTIONS
See Figure 2-2. Connect sample inlet and outlet lines to appropriately labeled fittings on the rear panel. All connections are 1/4-inch ferrule-type compression fittings on the ROX GT. The ROX GP has 1/8-inch female NPT connections.
If sample is available at positive pressure and flow between .1 and 2 SCFH we recommend following the sample handling configuration shown in Figure 2.3.
2-2
If sample is available at ambient pressure or flow is insufficient to meet minimum flow requirements, we recommend following the sample handling configuration shown in Figure 2.4. The inclusion of fast loop bypass, particulate filtration and moisture removal are not shown in these drawings and are the responsibility of the customer to include as required.
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 19
NSTALLATION
A
A
I
F
IGURE
2.4.2 G
ALIBRATION GASES
C
VENT (ROX GT) 1/4" SWAGELOK (ROX GP) 1/8" FPT
SAMPLE IN (ROX GT) 1/4" SWAGELOK (ROX GP) 1/8" FPT
2-2. B
ACK PANEL CONNECTIONS
AS SPECIFICATIONS
1 2 + -
5 6 + -
7 8 9
C NC NO
10 11 12 C NC NO
LARM CONTACT #2 LARM CONTACT #1
0 - 1VDC OUTPUT
4-20mA ISOLATED OUTPUT
FUSE
POWER SWITCH
Calibration method and gases depends on the desired operating range, and speed of calibration recovery. Due to the characteristics of the sensor, no zero gas is required. In the absence of oxygen, the sensor has an absolute zero.
Span gas can be either air or a certified blend of known ppm concentration of oxygen in nitrogen. All calibration gases are introduced through the sample inlet at the rear of the module
AMPLE GAS
S
Sample gas should be nonflammable (below 100% of the sample's LEL) and inert. Consult factory if sample gas contains CO2, CO, acid gases, or halogen gases. These samples may not be suitable for use with this analyzer.
LOW RATE
F
The sample flow rate can set be between .1 SCFH and 2.0 SCFH (50 - 1000 cc/min) without effecting the accuracy of the sensor. Sample flow should be set at a constant value within these limits which provides the desired response time and sample conservation.
RESSURE/FILTRATION
P
Sample pressure at the inlet should be within the range of 0 to 15 psig (1013 to 2048 hPa).
The analyzer does not contain any integral sample filtration. Sample should be filtered for particulates down to five microns to reduce the risk of internal sample line blockage.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
2-3
Page 20
NSTALLATION
I
Eductor
Vacuum Breaker
Nitrogen or Air In
F
IGURE
2-3. R
S
Sample
Flowmeter
Vent
Sample Inlet
Ambient Air
or Span Gas
Flow Control
Valve
Sensor
Isolation
Valve
ECOMMENDED SAMPLE HANDLING DIAGRAM
AMPLE
ROX GT
(back)
- P
OSITIVE PRESSURE
F
IGURE
2-4
2-4. R
P
Sample Vent
Flowmeter
Sample Inlet
Air or
Span Gas
Sensor
Isolation
Valve
Flow Control
Valve
ECOMMENDED SAMPLE HANDLING DIAGRAM
RESSURE OR FLOW
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
ROX GT
(back)
- I
NSUFFICIENT
Page 21
2.5 ELECTRICAL CONNECTIONS
All electrical connections can be found on the back of the analyzer (Refer to Figure
2.2).
NSTALLATION
I
2.5.1 L
INE POWER CONNECTION
The ROX GT/GP is designed to operate on 115 or 230 VAC 50/60 Hz power. The power entry module is located on the right rear panel of the instrument. The power entry module utilizes a standard IEC 320 connector allowing universal connection in all countries by selecting the appropriate line cord at installation. A standard North American power cord is provided with the analyzer.
The analyzer comes configured from the factory for operation on 115 VAC power. If you need to operate the unit on 230 VAC, check to make sure that the unit is disconnected from line power, open the cover door on the power entry module and extract the fuse-block as explained in Section 4.3. Turn the fuse-block over and reinstall so that the “230V” is now visible through the window in the cover door. Close the cover door prior to connecting power.
2.5.2 I
NTERCONNECTIONS
The analyzer has two kinds of output connections, a signal output and a set of relay contacts. The contacts are rated at 115 VAC, but with the following provisos:
There is only "basic insulation (within the meaning of IEC 1010-1) provided between the contacts and the rest of the circuitry. This means that in the event of a single fault, it is possible that 115 VAC connected to the relay contacts would be connected to the output connections as well. Any equipment connected to the analyzer output may potentially see this 115 VAC.
WARNING: ELECTRICAL SHOCK HAZARD
The output connection is for use only with equipment which has no live parts which are accessible.
The connecting cable and the equipment to which the output is connected must have insulation rated for at least 150 VAC, since under fault condition the output may be connected to the relay contacts witch themselves may be connected to 150 VAC.
The connection used at the remote end of the output circuit must be such as to be suitable for 150 VAC and must have no accessible live parts.
The equipment connected to the output must either be approved to IEC 1010-1 or equivalent or must be suitable for use with an input that may potentially be connected to 150 VAC, and must not catch fire in this circumstance.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
2-5
Page 22
NSTALLATION
I
Note The interconnection terminal strip on the rear of the analyzer is designed to be
unplugged for easier connection. It is polarized so that you cannot plug it in backwards. Make the connections with plenty of service loop and then plug it in. Double check that you have used the correct terminals, and do not apply a high voltage (e.g. 115 V) to the relay contacts before you have verified that the output connections are correct - putting AC on the signal output will destroy the analyzer.
2.5.3 S
IGNAL OUTPUT CONNECTIONS
The analyzer comes standard with both 4-20 mA DC isolated and 0-1 VDC outputs. For ease of installation, the signal output connections are made via a removable terminal strip connector which is shared with the alarm connections (Refer to Figure
2.2).
WARNING: POSSIBLE EXPLOSION HAZARD
Check all wiring connections prior to reconnection of the terminal strip. The application of AC power to the signal output terminals will damage the analyzer.
2.5.4 A
LARM CONNECTIONS
The analyzer comes standard with two alarms both configured high and normally energized. For ease of installation, the alarm connections are made via a removable terminal strip connector which is shared with the signal output connections (Refer to Figure 2.2). Each set of alarm connections are labeled for easy identification to the associated alarm. The connections labeled NO (normally open), NC (normally closed), and C (common) refer to relay positions with the power removed (de­energized) .
Both alarms are configured from the factory for fail-safe mode (normally energized, and loss of power will cause a change in relay status) and high oxygen setpoint (relay status change above setpoint). To change the alarm configuration, refer to Section
4.5.
2.6 INSTALLATION GUIDELINES
Is the analyzer’s location clean, weather-proofed, non-hazardous, vibration-free, and with a stable ambient temperature?
Are gas supply cylinders equipped with a clean, metal diaphragm, hydrocarbon free, two stage regulator with shut off valve?
2-6
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 23
NSTALLATION
I
Are external tubing, regulators, valves, pumps, fittings etc. clean?
Is the oxygen content of the supply gases compatible with the analysis range?
Is the sample non-flammable?
Have all the external gas connections been leaked checked?
Has clean stainless steel or other appropriate metal tubing been used for sample
lines?
Are all electrical connections secure and correct?
Has the sensor been installed in the ROX GP?
Is the proper line input power displayed on the power entry module?
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
2-7
Page 24
NSTALLATION
I
NOTES
2-8
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 25
S
TARTUP AND OPERATION
3
3.1 STARTUP
After completion of the installation procedure in Section Two, proceed as follows:
1. Set range switch to the 25% position (full counterclockwise)
2. Turn on power by placing the switch on the power entry module to the “-” position.
3.2 CALIBRATION
See Section 2.4.2 for a description of the method for choosing calibration zero and span gases.
To calibrate the ROX GT/GP, introduce air or span gas into the SAMPLE INLET, and do the following:
1. Set range switch to the 25% range setting (full counterclockwise) or the appropriate range for the span gas used. It is recommended to use the highest concentration span gas which provides you with the best fit for your requirements. The use of air for calibration provides the least chance of span gas error, but does require extended recovery time down to low ppm level measurement.
2. If you have followed the recommended sample handling diagram in Section 2.4.1. open the sensor isolation valve first to avoid exposing the sensor to excessive sample pressure or vacuum. Flow span gas or air at a floweret between 0.1 and
1.0 L/min. Be careful to not obstruct the outlet as this will cause an increase in sample pressure and a consequent increase in oxygen reading.
Note Alarm setpoints are not defeated during calibration. Take appropriate
precautions during calibration.
3. Wait f o r reading to stabilize. This should only take a few minutes.
4. Locate the SPAN ADJUST potentiometer on the front panel and adjust appropriately to make the display correspond with the span gas concentration. Turning the potentiometer clockwise will increase the display reading while counterclockwise will decrease the reading.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
3-1
Page 26
TARTUP AND OPERATION
S
5. If you have used air to span the instrument, switch to zero gas or sample as soon as possible to avoid prolonged exposure of the sensor to high concentrations of oxygen. The longer the sensor is exposed to air, the longer it will take for the sensor to recover to low ppm levels. W hen installing a new sensor or starting the instrument for the first time, it may take as long as eight hours for the analyzer to purge down to the lowest operating range.
6. Allow the unit to stabilize at the final reading on zero gas or sample. The unit is now ready for operation.
Unit should be calibrated once a week until familiarity is achieved in required calibration frequency. Calibration intervals as long as once a month are possible on stable installations. Processes which have large changes in environmental conditions and sample composition may require more frequent calibrations.
3.3 OPERATION
After calibration, proceed as follows: Supply sample gas to the SAMPLE INLET. Adjust external flow controller or throttle
valve so that flow discharged from the outlet is between 0.1 and 1.0 L/min.
Note Flow indication is not provided with the instrument and must be provided
external to the analyzer. Refer to installation section for proper location.
When shutting down the instrument, always close the sensor isolation valve last to avoid exposing the sensor to excessive pressure or vacuum.
3.4 ALARM SETPOINTS
The ROX GT/GP come standard with two fully adjustable alarms with relay contacts. These alarm setpoints are designed to operate as percent of range as opposed to discrete concentration value. The alarms are configured from the factory as high and failsafe.
To view the alarm setpoints, turn the range switch to the desired measuring range and press the “SET LO ALARM” or “SET HI ALARM” button. The alarm setpoint will be displayed in actual concentration units. Since the alarms are configured as percent of scale, changing ranges will also change the value of the alarm setpoint. For example, if the LO alarm is set for 5 ppm on the 10 ppm range (50% of range), it will read 50 ppm on the 100 ppm range (50% of range).
To adjust the alarm setpoint, press the appropriate alarm setpoint button, then locate and adjust the associated alarm setpoint adjustment potentiometer as indicated in Figure 1-1. Turning the potentiometer clockwise will increase the setpoint while turning the potentiometer counterclockwise will decrease the setpoint. You must keep the alarm setpoint button depressed during the adjustment procedure to continuously view the alarm setpoint value. Always adjust the “LO” alarm for the lowest setpoint value and the “HI” alarm for the higher value.
3-2
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 27
M
AINTENANCE AND TROUBLESHOOTING
4
CAUTION: QUALIFIED PERSONNEL
This equipment should not be adjusted or repaired by anyone except properly qualified personnel.
WARNING: PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect the safety and performance of this product. Use only factory approved components for repair.
4.1. MAINTENANCE OVERVIEW
The ROX GT/GP analyzer components which may require replacement include:
Oxygen sensor
Power fuse
Printed circuit boards
4.2 OXYGEN SENSOR REPLACEMENT
CAUTION: CAUS TIC MATERIAL
The sensor contains electrolyte which is caustic and can cause serious burns to skin. Do not ingest contents of sensor. Read and understand the Material Safety Data Sheet in the rear of this manual before handling sensor.
The oxygen sensor used in the ROX GT/GP analyzers is a consumable item and does require replacement periodically. To determine if the sensor requires replacement, see the troubleshooting portion of this section.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GT Oxygen Analzyers
4-1
Page 28
AINTENANCE AND TROUBLESHOOTING
M
ENSOR INSTALLATION
S
To install a new sensor follow the procedures exactly in the order listed below. Failure to follow the procedure in order may result is extended exposure of the sensor to air causing greater delays in returning to on-line monitoring:
1. Turn off power to unit and turn off sample flow to analyzer.
2. Locate cellblock cover and remove using the key tool provided with instrument. Rotate cover counterclockwise to remove.
3. Inspect cellblock cavity for signs of liquid. If liquid is present, use protective equipment as described in the material safety data sheet in the rear of this manual before proceeding. Extreme care should be used and the liquid should be treated as a corrosive substance.
4. Remove existing sensor and dispose of in accordance with national, federal, state, and local regulations. Use the existing pull-tab on the sensor to remove from the cellblock.
5. Inspect inside of cellblock for signs of residual liquid or deposits on the contact pins. If liquid is present, use appropriate protective equipment and follow the cleanup procedure as described in the material safety data sheet in the rear of this manual. Deposits on the contact pins can be removed by wiping with a damp cloth or using thee eraser from a pencil. Do not use abrasives (i.e. sandpaper) as this will damage the contact pins.
6. If sample is inert (e.g., nitrogen) a low flow of sample (50 cc/min) can be introduced to the analyzer to purge out the internal volume of the cellblock prior to installation of the new sensor. This procedure is not required and is recommended as a step to reduce the purge down time of the new sensor.
CAUTION: MAXIMUM FLOW
Do not flow a volume in excess of 50 cc/min. Exceeding this level may produce an oxygen deficient breathing environment to the technician conducting sensor replacement. Do not flow sample if it is flammable or toxic. Do not use this procedure if the sample handling configuration is for insufficient pressure or flow.
4-2
7. Inspect the package containing the replacement sensor. Do not open the bag. Make sure that it is the correct model type and that there is not visible liquid in the bag. If the bag contains liquid, do not open and contact the factory for a replacement.
July 1998 Rosemount Analytical 748375-BROX GT/ROX GT Oxygen Analzyers
Page 29
AINTENANCE AND TROUBLESHOOTING
M
8. Locate the shorting tab on the back of the sensor. The tab is designed to maintain contact between the two contact plates of the sensor while it is stored in the bag. This contact allows the sensor to continuously consume dissolved oxygen inside the sensor. This clip will need to be removed prior to installing the sensor in the analyzer.
9. Locate the pull-tab portion of the sensor label. W hen installing the sensor in the cellblock, make sure that this tab faces forward for easy sensor removal later.
10. Cut open the bag being careful not to damage the sensing face of the sensor. Quickly remove the sensor from the bag, remove the shorting clip and install in the cellblock with the contact plates on top and the sensing face on the bottom. Make sure that the pull-tab is facing forward and the writing on the label is right side up (see Figure 4-1).
11. Reinstall the cellblock cover turning clockwise until the O-ring just makes contact. Turn no more than 1/8 turn additional.
Note Do not overtighten. Doing so may damage the cover.
12. T urn on power and check reading.
13. Calibrate as described in Section 3.2
Key Slot
Cell Block Cover
Contact Plates
Cell Block (with cover removed)
Oxygen Sensor
Sensing Face
F
IGURE
748375-B Rosemount Analytical July 1998
4-1 S
ENSOR INSTALLATION
Sensor Pull Tab
ROX GT/ROX GT Oxygen Analzyers
4-3
Page 30
AINTENANCE AND TROUBLESHOOTING
M
4.3 POWER FUSE REPLACEMENT
The ROX GT/GP has the main power fuses located in the power entry module. No other fuses are located inside the instrument.
Before replacing the main power fuse, turn off the analyzer and disconnect the power cord from the power entry module.
Using a small blade slotted screwdriver, carefully pry open the f ront door of the power entry module. There is a access slot on the top of the module and the door opens downward (see Figure 4-2). Once the door is opened, you can remove the red power selector / fuseholder by inserting the screwdriver into the slot on top of the fuseholder and carefully prying the fuseholder out. The fuseholder contains two 1A fuses. Replace as required.
Reinstall the fuseholder so that the correct voltage is visible through the small window in the door of the power entry module.
ACCESS SLOT
FUSEHOLDER (RED)
POWER ENTRY MODULE
WITH FRONT DOOR OPEN
F
IGURE
4-2. P
OWER ENTRY MODULE
4.4 PRINTED CIRCUIT BOARDS
Both printed circuit boards can be replaced if necessary. Refer to Figure 2-1 for the location of the power and signal boards.
4.5 ALARM CONFIGURATION
The alarm setpoints are factory set for high normally energized. If you wish to change the configuration of the alarm setpoints you can do so by the following procedure.
1. Verify that the power module is off and the power cord is disconnected from the analyzer
4-4
July 1998 Rosemount Analytical 748375-BROX GT/ROX GT Oxygen Analzyers
Page 31
AINTENANCE AND TROUBLESHOOTING
M
2. Remove the top cover from the analyzer
3. Locate the signal board as indicated in Figure 2-1
4. Locate the alarm setting dip switches on the signal board
5. To change the alarm setting refer to Figure 4-3 and Table 4-1 for desired configuration
In all cases, “LO ALARM” should be adjusted for the lowest setpoint value and “HI ALARM” should be adjusted for the higher setpoint value.
F
IGURE
4-3. A
ON OFF
4 3 2 1
EDGE OF SIGNAL BOARD
EDGE OF SIGNAL BOARD
LARM SWITCH LOCATION
LO ALARM SETTING (S2)
HI ALARM SETTING (S3)
748375-B Rosemount Analytical July 1998
ROX GT/ROX GT Oxygen Analzyers
4-5
Page 32
AINTENANCE AND TROUBLESHOOTING
M
TRIP POINT - FS/NFS SWITCH #
(RELAY STATUS) 1 2 3 4
HIGH - FAILSAFE
OFF ON ON OFF (NORMALLY ENERGIZED) HIGH - NON FAILSAFE
ON OFF OFF ON (NORMALLY DE-ENERGIZED) LOW - FAILSAFE
ON OFF OFF ON (NORMALLY ENERGIZED) LOW - NON FAILSAFE
OFF ON ON OFF
(NORMALLY DE-ENERGIZED)
T
ABLE
4-1. A
LARM SETTING CONFIGURATION
4.6 TROUBLESHOOTING
The following provides a short list of common troubleshooting tips.
NALYZER DOES NOT POWER UP
A
Check to make sure that the power cord is properly installed and the voltage is correctly selected. The voltage configuration of the analyzer is visible through the window of the power entry module. Make sure that the power switch is in the on position (“-”).
If the unit does not power up, turn off power switch, remove power cord and open door on power entry module as described in Section 4.4 above. Inspect fuses and replace as required.
NALYZER FAILS TO PURGE DOWN TO PPM LEVELS
A
The number one problem associated with trace oxygen analyzer installation is the occurrence of leaks in the sample line plumbing. If the oxygen reading will not come down to ppm levels or is reading higher than expected, the sample plumbing prior to the instrument may have a leak. A quick check can be conducted by observing the oxygen reading at two different flow levels; 0.2 and 2.0 SCFH. If the oxygen reading drops significantly when the flow is increased from 0.2 to 2.0 SCFH, this is a good indication that a leak exists.
To check for leaks prior to the sensor, disconnect the analyzer and cap the inlet and outlet fittings on the analyzer to reduce the amount of oxygen exposure to the sensor. Cap the inlet line and pressurize the inlet line to 5 - 10 psig and check all connections with a soapy solution to identify leaks.
4-6
July 1998 Rosemount Analytical 748375-BROX GT/ROX GT Oxygen Analzyers
Page 33
AINTENANCE AND TROUBLESHOOTING
M
WARNING: POSSIBLE SENSOR DAMAGE
Do not pressure check the sample line with the sensor/analyzer connected. Over-pressurization of the sensor can result in damage.
If the reading increases with an increase in flow, the vent line may be obstructed causing back pressure on the sensor. Inspect your vent line and remove any devices which may be causing backpressure.
If the reading slowly decreases over time, the sensor may have been exposed to high levels of oxygen for extended periods of time. The sensor must consume all of the dissolved oxygen in the sensor electrolyte before it will accurately measure ppm le vels of oxygen. In some cases it is not possible for the sensor to recover to low ppm levels and the sensor must be replaced.
NALYZER IS READING TOO LOW
A
Check to make sure that the sensor has been calibrated properly. Flow an appropriate span gas through the instrument and follow the calibration procedure in Section 3.2. If you are unable to achieve span calibration, the sensor has reached the end of its useful life and needs to be replaced (see Section 4.2).
If calibration is possible, but the reading still appears too low, conduct a span check on the 25% range with air. Adjust reading as necessary. If the reading requires adjustment, your bottled span gas may be incorrect or corrupted.
If sensor life seems too short, you may have compounds present in your sample which are poisoning the sensor. Acid gases, CO2, SO2, or H2S can reduce the life of the sensor and should be removed from the sample prior to measuring.
NALYZER ALWAYS READS ZERO
A
Check to make sure that the sensor has been installed correctly with the contact plates on top and the gas permeable membrane facing downward. If the sensor has been installed incorrectly, carefully remove and check to make sure that the membrane has not been damaged. If the sensor is leaking, or shows signs of distress to the membrane, replace with a new sensor. If the sensor is okay quickly reinstall in the correct position.
Verify that the cellblock contact pins are making good contact with the metal surfaces of the sensor. If the pins are not making contact, carefully bend the wire down to reestablish contact. Check that the contact pins are clean and do not show signs of deposits or oxidation. Clean as necessary.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GT Oxygen Analzyers
4-7
Page 34
AINTENANCE AND TROUBLESHOOTING
M
A good sensor will have an output of 150 to 500 micro amps in air. You can check this by connecting a DVM across the contact plates of the sensor while it is removed from the instrument and measuring the current. If the sensor does not produce sufficient output, replace the sensor with a new one.
HE READINGS ARE ERRATIC AND DRIFT NOISILY
T
This is normally an indication that the sensor has reached the end of its useful life. If the analyzer cannot be spanned on a cal gas or air, the sensor must be replaced. If the sensor spans okay and appears stable at high oxygen concentrations, there may be a sample line leak or obstruction of the vent line. Follow the procedure described above in “Analyzer fails to purge down to ppm levels”.
If no sample line problems are found, open the cellblock and inspect the sensor for signs of damage or leakage. Follow the procedure in Section 4.2. for correct method of opening cellblock and precautions for handling a leaking sensor.
4-8
July 1998 Rosemount Analytical 748375-BROX GT/ROX GT Oxygen Analzyers
Page 35
R
EPLACEMENT PARTS
5
5.1 ROX REPLACEMENT PARTS
WARNING: PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect the safety and performance of this product. Use only factory approved components for repair.
5.1.1 R
5.1.2 R
EPLACEMENT SENSORS
Refer to Table A-1 in Appendix A for easy cross reference of the correct oxygen sensor for your specific instrument.
658303 Oxygen Sensor GT1 658304 Oxygen Sensor GT3 658310 Oxygen Sensor GP1 658311 Oxygen Sensor GP3 658312 Oxygen Sensor GP4
EPLACEMENT PARTS LIST
904699 Connector Rear Output 658281 Power Cord 904759 Power Supply Board 904760 Cap. Cell Block 904761 O-Ring, Cell Block Cap 904762 Ribbon Cable 904763 Power Entry Module 099393 Fuse 1A 250V 748375 Instruction Manual
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
5-1
Page 36
EPLACEMENT PARTS
R
ROX GT O
NLY
904768 Cell Block Assembly 904769 Amplifier Board (658301) (658302)
ROX GP O
NLY
904764 Cell Block Assembly 904765 Amplifier Board (658307) 904766 Amplifier Board (658308) 904767 Amplifier Board (658309)
5-2
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 37
A
PPENDIX
A. M
ATERIAL SAFETY DATA SHEETS
A
The following section contains important information regarding the characteristics of the oxygen sensor used in the ROX GT/GP analyzers. Read and understand all information before attempting any replacement or installation of the oxygen sensor. The following sensor reference guide will provide a quick reference of the specific oxygen sensor type used in your specific analyzer model and part number.
T
ABLE
MODEL INSTRUMENT
PART NUMBER
ROX GT 658301 GT1 658303 ROX GT 658302 GT3 658304 ROX GP 658307 GP1 658310 ROX GP 658308 GP3 658311 ROX GP 658309 GP4 658312
A-1. S
ENSOR REFERENCE GUIDE
OXYGEN SENSOR
MODEL
OXYGEN SENSOR
PART NUMBER
748375-A Rosemount Analytical September 1997
ROX GT/ROX GP Oxygen Analyzers
A-1
Page 38
PPENDIX
A
N
OTES
ATERIAL SAFETY DATA SHEETS
A. M
A-2
September 1997 Rosemount Analytical 748375-AROX GT/ROX GP Oxygen Analyzers
Page 39
Rosemount Analytical
Material Safety Data Sheet
748378-A
PRODUCT: PART NUMBER:
XYGEN SENSOR
O
658303, 658310, 658312
24 HOUR EMERGENCY TELEPHONE NUMBER:
CHEMTREC (800) 424-9300
SECTION I - GENERAL
Distributor: Rosemount Analytical Inc.
4125 East La Palma Avenue, CA 92807-1802
714-986-7600 Chemical name and synonyms Trade name and synonyms Chemical family Formula CAS Number
Lead Caustic Potash Inorganic base Inorganic base Pb KOH 749-92-1 1310-58-3
SECTION II – HAZARDOUS INGREDIENTS
Quantity OSHA PEL ACGIH
3-20 gms 1-5 ml
0.05 mg/m
0.15 mg/m
SECTION III – PHYSICAL DATA
Boiling point Melting point Vapor pressure Vapor density (air=1) Specific gravity (H2O=1) % Volatile by volume Evaporation rate (H2O=1) Solubility in water
Appearance and odor
1744°C 1320°C 328°C 360°C NA NA NA NA
11.34 2.04 NA NA NA NA Insoluble Complete
Solid, silver gray, odorless
SECTION IV – FIRE AND EXPLOSION HAZARD DATA
Flash point Extinguishing media
Special fire fighting procedures Unusual fire and explosion hazards
NA Appropriate to surroun ding fi re condition s. No specifi c agents recommended. Wear NIOSH/OSHA approv ed self-co ntained br eathin g apparat us and protective clothing to prevent con tact with skin and eyes. Emits toxic fumes under fire condi tions.
SECTION V – REACTIVITY DATA
Stability Conditions to avoid
Incompatibility (material to avoid) Hazardous decomposition or
byproducts Hazardous polymerization
Stable Avoid contact with acids and hydrogen peroxide >52% Aluminum, organic materials, acid chlorides, acid anhydrides,
magnesium, copper. Toxic fumes Will not occur
Lead Potassium Hydroxide
3 3
None 2 mg/m
White or slightly yellow, odorless
Rosemount Analytical Inc.
4125 East La Palma Avenue Anaheim, California 92807-1802 TEL 714-986-7600 FAX 714-577-8006
April 1997 748378-A Printed in USA
Page 40
MATERIAL SAFETY DATA SHEET
SECTION VI – HEALTH HAZARD DATA
Lead Potassium Hydroxide Threshold limit value Routes of entry
Effects of overexposure
Emergency & first aid procedures
NA NA Inhalation, skin, ingestion Inhalation, skin, ingestion
Chronic exposure to lead may cause disease of the blood and blood forming organs , kidneys and liver, damage to the reproductive systems and decrease i n fertilit y in men and women, and damage to the fetus of a pregnant woman.
Chronic exposure from t he lead contained in this product is extremely unlikely. Eyes and skin
minutes. Remove contaminated clothing. Call a physician.
ORAL INGESTION - DO NOT INDUCE VOMITING
conscious, give quantities of milk or water. Call a physician immediately and take patient to hospital emergency room.
- immediately flush with water for at least fifteen
Acute Effects: Corrosive to tissue. May cause irritation or chemical burns to skin and eyes. Harmful if swallowed, inhaled or absorbe d thro ugh the skin.
Chronic Effects
exposure has a destructive effect on tissue.
: Prolonged
. If person is
SECTION VII – SPILL OR LEAK PROCEDURE
NOTE: The oxygen sensors are sealed, and under normal circumstances, the contents of the sensors do not present a health hazard. The following information is given as a guide in the event that a cell leaks.
Steps to be taken in case material is released or spilled
Waste disposal method
Wipe down the area se veral times, each time with a fresh, wet paper towel. Contaminated paper towels are considered hazardous waste.
Both lead and potassium hydroxide are considered poisonous substances and are regulated under TSCA and SARA Title III. EPA Waste Number: D008 California Waste Number: 181
DOT Information: RQ Hazardous Waste So lid N.O.S. ( Lead), 9 , UN3077 , PG III.
Follow all Federal, State and Local regulations.
SECTION VIII – SPECIAL PROTECTION INFORMATION
Respiratory protection Ventilation Protective gloves Eye protection Other protective equipment
NA NA Rubber gloves. Chemical splash goggles. Apron, face shield
SECTION IX – SPECIAL PRECAUTIONS
Protective measures during cell replacement: Before opening the bag
Precautions to be taken in handling and storing
Other precautions
containing the sensor cell, c heck the sensor cell for leakage. If sensor cell leaks, do not o pen the bag. If liquid is pr esent around cell whi le in the instrument, wear eye and hand protect ion . none
SECTION X – TRANSPORTATION
Must be compliance with federal, state and local regulations
748378-A
NOTICE: WHILE ROSEMOUNT ANALYTICAL BELIEVES THE INFORMATION CONTAINED HEREIN IS VALID AND ACCURATE, ROSEMOUNT ANALYTIC AL MAKES NO W ARRANTY OR REPRESENTATION AS TO ITS VALIDITY, ACC URACY, OR CURRENCY. ROSEMOUNT ANALYTICAL SHALL NOT BE LIABLE OR OTHERWISE RESPON SIBLE IN AN Y WAY FOR U SE OF EIT HER THIS INFO RMATION OR T HE MAT ERIAL TO WHICH IT APPLIES. DIS PO SAL O F HAZ ARDO US M AT ERIA L M A Y B E SU BJE CT TO FEDERAL, ST ATE, OR LOCAL LAWS AND/OR REGULATIONS.
Sheet 2 of 2
April 1997
Page 41
Rosemount Analytical
Material Safety Data Sheet
748380-B
PRODUCT: PART NUMBER:
XYGEN SENSOR
O
658304
, M
ODEL
GT3
24 HOUR EMERGENCY TELEPHONE NUMBER:
CHEMTREC (800) 424-9300
SECTION I - GENERAL
Distributor: Rosemount Analytical Inc.
4125 East La Palma Avenue, CA 92807-1802
714-986-7600 Chemical name and synonyms Trade name and synonyms
Chemical family Formula CAS Number
SECTION II – HAZARDOUS INGREDIENTS
Hazardous mixtures of other liquids, solids or gases
SECTION III – PHYSICAL DATA
Boiling point Melting point Vapor pressure Vapor density (air=1) Specific gravity (H2O=1) % Volatile by volume Evaporation rate (H2O=1) Solubility in water
Appearance and odor
SECTION IV – FIRE AND EXPLOSION HAZARD DATA
Flash point Extinguishing media Special fire fighting procedures Unusual fire and explosion hazards
SECTION V – REACTIVITY DATA
Stability Conditions to avoid Incompatibility (material to avoid)
Hazardous decomposition or byproducts
Hazardous polymerization
acetic acid
(5% v/v)
acetic acid potassium acetate lead inorganic acid inorganic base inorganic base
2H3O2
HC 64-19-7 127-08-2 7439-92-1
none
118°C NA 1744°C
16.6°C 292°C 328°C NA NA NA NA NA NA
1.05 1.57 11.34 NA NA NA NA NA NA Infinite 72% Insoluble
Clear, odorless solution w/strong vinegar-like odor
NA NA NA NA NA NA NA NA NA NA NA NA
In sensor, stable under normal conditions of use. Avoid contact between sensor electrolyte and strong acids and
oxidizing agents. NA
NA will not occur
potassium acetate
(5% w/v)
KC2H3O
White, odorless crystal
2
1
lead
(pure)
Pb
Grey, odorless
1
The electrolyte in the Model GT3 Oxygen Sensor is not flammable. There are not fire or explosion hazards associated with
this sensor.
Rosemount Analytical Inc.
4125 East La Palma Avenue Anaheim, California 92807-1802 TEL 714-986-7600 FAX 714-577-8006
July 1998 748380-B Printed in USA
Page 42
MATERIAL SAFETY DATA SHEET
748380-B
SECTION VI – HEALTH HAZARD DATA
Threshold limit value Routes of entry
Effects of overexposure
Emergency & first aid procedures
10 ppm (TWA) 15 ppm (STEL)
Skin, eyes, ingestion. Liquid inhalation highly unlikely.
- The electrolyte is corrosive; eye contact could result in
Eyes
permanent loss of vision. Skin - The electrolyte is corr osiv e ; skin con ta ct coul d re sul t in a chemical burn.
Ingestion Eyes - Flush eyes with water for at least 15 minutes and get
immediate medical attention.
Skin
contaminated clothing. If burning persists, seek medical attention. Ingestion - Give plenty of cold water. DO NOT INDUCE VOMITING. Seek medical attention.
- The electrolyte could be harmful or fatal if swallowed.
- Wash affected area with plenty of water and remove
NA 0.15 mg/cu.m
SECTION VII – SPILL OR LEAK PROCEDURE
NOTE: THE OXYGEN SENSORS ARE SEALED, AND UNDER NORMAL CIRCUMSTANCES, THE CONTENTS OF THE SENSORS DO NOT PRESENT A HEALTH HAZARD. THE FOLLOWING INFORMATION IS GIVEN AS A GUIDE IN THE EVENT THAT A CELL LEAKS.
Steps to be taken in case material is released or spilled
Waste disposal method
Wipe down the area several times with a wet paper towel. Use a fresh towel each time. Should be in acco rdan ce wi th all ap pli c ab le sta te , local a nd fede r al regulations.
SECTION VIII – SPECIAL PROTECTION INFORMATION
Respiratory protection Ventilation Protective gloves Eye protection Other protective equipment
NA NA Rubber gloves Chemical splash goggles approved working clothes
SECTION IX – SPECIAL PRECAUTIONS
During cell replacement, before opening the bag containing the
Precautions to be taken in handling and storing
Other precautions
sensor cell, check the sensor cell for leakage. If the sensor cell leaks, do not open the bag. If there is liquid around the cell while in the instrument, put on gloves and eye protection before removing the cell. none
SECTION X – TRANSPORTATION
Must be compliance with federal, state and local regulations
NOTICE: WHILE ROSEMOUNT ANALYTICAL BELIEVES THE INFORMATION CONTAINED HEREIN IS VALID AND ACCURATE, ROSEMOUNT ANALYTIC AL MAKES NO W ARRANTY OR REPRESENTATION AS TO ITS VALIDITY, ACC URACY, OR CURRENCY. ROSEMOUNT ANALYTICAL SHALL NOT BE LIABLE OR OTHERWISE RESPON SIBLE IN AN Y WAY FOR U SE OF EIT HER THIS INFO RMATION OR T HE MAT ERIAL TO WHICH IT APPLIES. DIS PO SAL O F HAZ ARDO US M AT ERIA L M A Y B E SU BJE CT TO FEDERAL, ST ATE, OR LOCAL LAWS AND/OR REGULATIONS.
Sheet 2 of 2
July 1998
Page 43
Rosemount Analytical
Material Safety Data Sheet
748381-A
PRODUCT: PART NUMBER:
XYGEN SENSOR
O
658311
, M
ODEL
GP3
24 HOUR EMERGENCY TELEPHONE NUMBER:
CHEMTREC (800) 424-9300
SECTION I - GENERAL
Distributor: Rosemount Analytical Inc.
4125 East La Palma Avenue, CA 92807-1802
714-986-7600
potassium dibasic
Chemical name and synonyms
Trade name and synonyms Chemical family
Formula CAS Number
SECTION II – HAZARDOUS INGREDIENTS
Hazardous mixtures of other liquids, solids or gases
SECTION III – PHYSICAL DATA
Boiling point Melting point Vapor pressure Vapor density (air=1) Specific gravity (H2O=1) % Volatile by volume Evaporation rate (H2O=1) Solubility in water
Appearance and odor
SECTION IV – FIRE AND EXPLOSION HAZARD DATA
Flash point Extinguishing media Special fire fighting procedures Unusual fire and explosion hazards
SECTION V – REACTIVITY DATA
Stability Conditions to avoid Incompatibility (material to avoid)
Hazardous decomposition or byproducts
Hazardous polymerization
phosphate
(1.78% w/v)
potassium dibasic phosphate
inorganic base inorganic base inorganic base
HPO
2
K 16788-57-1 198-14-6 7439-92-1
none
118°C NA 1744°C
16.6°C 292°C 328°C NA NA NA NA NA NA
1.05 1.57 11.34 NA NA NA NA NA NA Infinite 72% Insoluble
Clear, odorless solution w/strong vinegar-like odor
NA NA NA NA
In sensor, stable under normal conditions of use. Avoid contact between sensor electrolyte and strong acids and
oxidizing agents. NA
NA will not occur
O KHCO
4
2
3H
potassium bicarbonate
White, odorless crystal
potassium
bicarbonate
(5.11% w/v)
3
1
lead
(pure)
lead
Pb
Grey, odorless
1
The electrolyte in the Model GP3 Oxygen Sensor is not flammable. There are not fire or explosion hazards associated with
this sensor.
Rosemount Analytical Inc.
4125 East La Palma Avenue Anaheim, California 92807-1802 TEL 714-986-7600 FAX 714-577-8006
May 1997 748381-A Printed in USA
Page 44
MATERIAL SAFETY DATA SHEET 748381-A
SECTION VI – HEALTH HAZARD DATA
Threshold limit value Routes of entry
Effects of overexposure
Emergency & first aid procedures
NA Skin, eyes, ingestion. Liquid inhalation highly unlikely.
Eyes - The electrolyte is corrosive; eye contact could result in permanent loss of vision. Skin - The electrolyte is corr osiv e ; skin con ta ct coul d re sul t in a chemical burn.
Ingestion - The electrolyte could be harmful or fatal if swallowed. Eyes - Flush eyes with water for at least 15 minutes and get
immediate medical attention. Skin - Wash affected area with plenty of water and remove contaminated clothing. If burning persists, seek medical attention. Ingestion - Give plenty of cold water. DO NOT INDUCE VOMITING. Seek medical attention.
SECTION VII – SPILL OR LEAK PROCEDURE
NOTE: THE OXYGEN SENSORS ARE SEALED, AND UNDER NORMAL CIRCUMSTANCES, THE CONTENTS OF THE SENSORS DO NOT PRESENT A HEALTH HAZARD. THE FOLLOWING INFORMATION IS GIVEN AS A GUIDE IN THE EVENT THAT A CELL LEAKS.
Steps to be taken in case material is released or spilled
Waste disposal method
Wipe down the area several times with a wet paper towel. Use a fresh towel each time. Should be in acco rdan ce wi th all ap pli c ab le st ate , local a nd fe der al regulations.
SECTION VIII – SPECIAL PROTECTION INFORMATION
Respiratory protection Ventilation Protective gloves Eye protection Other protective equipment
NA NA Rubber gloves Chemical splash goggles approved working clothes
SECTION IX – SPECIAL PRECAUTIONS
During cell replacement, before opening the bag containing the
Precautions to be taken in handling and storing
Other precautions
sensor cell, check the sensor cell for leakage. If the sensor cell leaks, do not open the bag. If there is liquid around the cell while in the instrument, put on gloves and eye protection before removing the cell.
none
SECTION X – TRANSPORTATION
Must be compliance with federal, state and local regulations
NOTICE: WHILE ROSEMOUNT ANALYTICAL BELIEVES THE INFORMATION CONTAINED HEREIN IS VALID AND ACCURATE, ROSEMOUNT ANALYTIC AL MAKES NO W ARRANTY OR REPRESENTATION AS TO ITS VALIDITY, ACC URACY, OR CURRENCY. ROSEMOUNT ANALYTICAL SHALL NOT BE LIABLE OR OTHERWISE RESPON SIBLE IN AN Y WAY FOR U SE OF EIT HER THIS INFO RMATION OR T HE MAT ERIAL TO WHICH IT APPLIES. DIS PO SAL O F HAZ ARDO US M AT ERIA L M A Y B E SU BJE CT TO FEDERAL, ST ATE, OR LOCAL LAWS AND/OR REGULATIONS.
Sheet 2 of 2
May 1997
Page 45
ENERAL PRECAUTIONS FOR HANDLING AND
G
TORING HIGH PRESSURE GAS CYLINDERS
S
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981 Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202 Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A flame should never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
4125 E
AST LA PALMA AVENUE
• A
Rosemount Analytical Inc.
, C
J
ULY
NAHEIM
ALIFORNIA
1997 • 748525-C • P
92807-1802 • 714-986-7600 • FAX 714-577-8006
RINTED IN
USA
Page 46
(blank)
Page 47
ARRANTY
W
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER W ARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD W ARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control.
Rosemount Analytical
4125 E
AST LA PALMA AVENUE
• A
F
EBRUARY 1997 • 7485189-C • PRINTED IN USA
Rosemount Analytical Inc.
NAHEIM
, C
ALIFORNIA
92807-1802 • 714-986-7600 • FAX 714-577-8006
Page 48
(blank)
Page 49
IELD SERVICE AND REPAIR FACILITIES
F
Field service and repair facilities are located worldwide.
U.S.A.
To obtain field service on-site or assistance with a service problem, contact (24 hours, 7 days a week):
National Response Center
1-800-654-7768
INTERNATIONAL
Contact your local Rosemount Sales and Service office for service support.
FACTORY
For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned Materials Authorization (RMA) number. Please have the following information when you call:
Model Number, Serial Number, and Purchase Order Number or Sales Order Number.
Prior authorization by the factory must be obtained before returned materials will be accepted. Unauthorized returns will be returned to the sende r, f re ight collect.
When return ing any product or compon ent that has been expo sed to a toxic, co rrosive or other hazardous material or used in such a hazardous environment, the user must attach an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
4125 E
AST LA PALMA AVENUE
• A
Rosemount Analytical Inc.
, C
NAHEIM
ULY 1997 • 748190-G • PRINTED IN USA
J
ALIFORNIA
92807-1802 • 714-986-7600 • FAX 714-577-8006
Page 50
(blank)
Page 51
Rosemount Analytical
ADDENDUM
015-748399-A
ODELS
M
ROX GT, ROX GP O
XYGEN ANALYZERS INSTRUCTION MANUAL
748375
1.0 OVERVIEW
This addendum serves as an amendment to the Models ROX GT and ROX GP Oxygen Analyzers instruction manual (PN 748375).
The following information should be considered part of the manual, and supersedes any conflicting information in the body of the manual.
READ THIS INFORMATION AND NOTE THE CONFLICTS
2.0 24 VDC ELECTRICAL CONNECTIONS
The ROX GT 24 VDC Analyzer (PN 658301-24) is designed to operate on 24 VDC input.
Prior to making electrical connections to the analyzer, verify that the power rating on the instrument name-rating label is correct.
The 24 VDC electrical connection is located on the right rear panel of the instrument. The terminal strip is labeled: (Refer to Figure 1.)
+
-
+24 VDC
-24 VDC Ground
WARNING: POSSIBLE ELECTRICAL DAMAGE
Connection of AC voltage to the power input connections will result in serious damage to the analyzer. Verify all electrical input connections for correct voltage and power before powering up analyzer.
Rosemount Analytical Inc.
4125 E
AST LA PALMA AVENUE ANAHEIM
J
ULY
, C
1998
ALIFORNIA
748399-A
92807-1802
PRINTED IN
TEL 714-986-7600
USA
FAX 714-577-8006
Page 52
ADDENDUM
F1A - 24V
FUSE
+
-
24 VDC POWER CONNECTION
FIGURE 1. REAR PANEL ELECTRICAL INPUT CONNECTION
3.0 POWER FUSE REPLACEMENT
The ROX GT 24 VDC Analyzer has the main power fuse located directly above the electrical power input terminal strip on the rear panel. No other fuses are located in the analyzer.
Before replacing the main power fuse, turn off and disconnect electrical power to the analyzer.
The fuseholder utilizes a standard twist-off cap. To replace the fuse, rotate the fuseholder cap counterclockwise until the cap is free from the body. The fuse will come out with the cap. Replace fuse and re-install cap onto fuseholder.
2 of 2
July 1998
748399-ARosemount Analytical
Loading...