July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 7
P
REFACE
PURPOSE/SAFETY SUMMARY
The purpose of this manual is to provide the procedures for the installation, operation
and maintenance of the ROX GT trace oxygen analyzer and ROX GP percent oxygen
analyzer..
Some sections may describe equipment not used in your configuration. The user
should become thoroughly familiar with the operation of this analyzer before operating
it. READ THIS INSTRUCTION MANUAL COMPLETELY.
To avoid explosion, loss of life, personal injury and damage to this equipment
and on-site property, all personnel authorized to install, operate and service this
equipment should be thoroughly familiar with and strictly follow the instructions
in this manual. SAVE THESE INSTRUCTIONS.
If this equipment is used in a manner not specified in these instructions,
protective systems may be impaired.
DANGER is used to indicate the presence of a hazard which will cause severe
personal injury, death, or substantial property damage if the warning is ignored
WARNING is used to indicate the presence of a hazard which can cause severe
personal injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor
personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation, or maintenance information which is
important but not hazard-related.
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to
live parts which can cause death or serious injury. Refer servicing to qualified
personnel.
For safety and proper performance this instrument must be connected to a
properly grounded three-wire source of power.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
P-1
Page 8
REFACE
P
WARNING: ELECTRICAL SHOCK HAZARD
The output connection is for use only with equipment which has no live parts
which are accessible.
The connecting cable and the equipment to which the output is connected must
have insulation rated for at least 150 VAC, since under fault condition the output
may be connected to the relay contacts witch themselves may be connected to
150 VAC.
The connection used at the remote end of the output circuit must be such as to
be suitable for 150 VAC and must have no accessible live parts.
The equipment connected to the output must either be approved to IEC 1010-1
or equivalent or must be suitable for use with an input that may potentially be
connected to 150 VAC, and must not catch fire in this circumstance.
WARNING: POSSIBLE EXPLOSION HAZARD
This equipment is not designed for analysis of flammable samples. Introduction
of flammable samples into this equipment could result in explosion causing
injury, death, or property damage.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect
safety of this product. Use only factory documented components for repair
CAUTION: CAUSTI C MATER I A L
Tampering with the oxygen sensor may result in leakage of the internal liquid
electrolyte. The electrolyte is caustic and can cause severe burns to skin. Do
not ingest contents of sensor. Read and understand the Material Safety Data
Sheets in the rear of this manual before handling the sensor.
P-2
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 9
ROX GT TRACE OXYGEN ANALYZER
SPECIFICATIONS - GENERAL
+ 1% of fullscale or .1 ppm whichever is greater
+ 1%
+ 1%
90% of fullscale in less than 10 seconds for ranges greater
than 0-10 ppm.
90% of fullscale in less than 30 seconds for 0-10 ppm.
90% of fullscale in less than 60 seconds for ranges less
than 0-10 ppm.
Typically less than 1 hour (air to less than 10 ppm)
Less than 1% per week at constant temperature
Less than 1% per week at constant temperature
:
115/230 VAC + 10%, 50/60 Hz, less than 5 W
SPECIFICATIONS - ENVIRONMENTAL
T
EMPERATURE RELATED
D
:
RIFT
A
MBIENT OPERATING
T
EMPERATURE
S
AMPLE TEMPERATURE
S
AMPLE HUMIDITY
S
PECIFICATIONS
A
REA CLASSIFICATION
G
AS CONNECTIONS
M
OUNTING CONFIGURATION
P
ANEL CUTOUT DIMENSIONS
C
ASE DIMENSIONS
W
EIGHT
:
:
:
- P
:
:
+ 2% of fullscale for temperatures between 15 - 35° C; +
5% off fullscale over the entire operating temperature range
5 to 45°C
5 to 45°C
:
0 to 95% relative humidity non-condensing
HYSICAL
:
General Purpose
1/4” Swagelok
1/2 19” or panel mount
:
3.625 X 7.625 inches (92 X 194 mm)
:
5.06 X 8.38 X 5.0 inches (129 X 213 X 127 mm)
HxWxD
Approximately 5 lbs (2.27 Kg)
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
P-3
Page 10
REFACE
P
ROX GP PERCENT OXYGEN ANALYZER
SPECIFICATIONS - GENERAL
RANGES:
S
IGNAL OUTPUT
A
LARMS
D
ISPLAY
M
INIMUM DETECTABLE LEVEL
R
EPEATABILITY
N
OISE
L
INEARITY
R
ESPONSE TIME
Z
ERO DRIFT
S
PAN DRIFT
P
OWER REQUIREMENTS
S
PECIFICATIONS
T
EMPERATURE RELATED
D
RIFT
A
MBIENT OPERATING
T
EMPERATURE
:
:
:
:
:
:
:
:
:
0 - 5, 10, 25% Oxygen
:
0-1VDC (10K min) and 4-20 mADC Isolated (600 ohm
max)
90% of fullscale in less than 7 seconds (GP1 sensor)
Less than 1% per week at constant temperature
Less than 1% per week at constant temperature
:
- E
115/230 VAC + 10%, 50/60 Hz, less than 5 W
NVIRONMENTAL
+ 2% of fullscale for temperatures between 15 - 35°C;
+ 5% off fullscale over the entire operating temperature
range
5 to 45°C
S
AMPLE TEMPERATURE
S
AMPLE HUMIDITY
:
:
5 to 45°C
0 to 95% relative humidity non-condensing
SPECIFICATIONS - PHYSICAL
General Purpose
1/8” Female NPT
1/2 19” or panel mount
:
3.625 X 7.5 inches (92 X 190 mm)
:
5.06 X 8.38 X 5.0 inches (129 X 213 X 127 mm)
HxWxD
Approximately 5 lbs (2.27 Kg)
P-4
A
REA CLASSIFICATION
G
AS CONNECTIONS
M
OUNTING CONFIGURATION
P
ANEL CUTOUT DIMENSIONS
C
ASE DIMENSIONS
W
EIGHT
:
:
:
:
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 11
REFACE
P
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE
For order administration, replacement Parts, application assistance, on-site or factory
repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when
you call: Model Number, Serial Number, and Purchase Order Number or Sales Order
Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive
or other hazardous material or used in such a hazardous environment, the user must
attach an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification
that the material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
TRAINING
A comprehensive Factory Training Program of operator and service classes is
available. For a copy of the Current Operator and Service Training Schedule contact
the Technical Services Department at:
Rosemount Analytical Inc.
Phone: 1-714-986-7600
FAX: 1-714-577-8006
D
OCUMENTATION
The following ROX GT/ROX GP Oxygen Analyzers instruction materials are available.
Contact Customer Service or the local representative to order.
748375 Instruction Manual (this document)
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
P-5
Page 12
REFACE
P
COMPLIANCES
This product is approved by the Canadian Standards Association (which is also an
OSHA accredited, Nationally Recognized Testing Laboratory), for use in nonhazardous, indoor locations
®
NRTL/C
Rosemount Analytical Inc. has satisfied all obligations from the
European Legislation to harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC.
Recommendation (May 1993).
This product satisfies all obligations of all relevant standards of the EMC framework in
Australia and New Zealand.
NAMUR
N96
P-6
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 13
I
NTRODUCTION
1
1.1 OVERVIEW
This manual describes the ROX GT trace oxygen analyzer and the ROX GP percent
oxygen analyzer of the ROX Series of oxygen analysis instruments.
The ROX GT is designed to determine continuously the concentration of trace oxygen
in a flowing gaseous mixture. The concentration is expressed in parts-per-million by
volume. The ROX GP is designed to measure continuously the concentration of
percent oxygen in a flowing gaseous mixture.
The ROX GT / GP is designed for panel mount or 1/2 19” rack mount, with gas
connections made from the rear. All electronic connections are also made from the
rear including the AC power input.
1.2 TYPICAL APPLICATIONS
Typical applications for the ROX GT include:
• Monitoring trace oxygen contamination in pure nitrogen or argon streams from
air separation facilities.
• Determination of trace oxygen content of inerting atmospheres in heat treat
furnaces.
• Monitoring inert atmosphere glove boxes for oxygen impurity
Typical applications for the ROX GP include:
• Measuring percent impurities in pure gases
• Controlling inerting atmospheres in heat treat applications
• Monitoring oxygen enriching or deficient operations
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
1-1
Page 14
NTRODUCTION
I
1.3 THEORY OF TECHNOLOGY
The ROX GT / GP uses an electrochemical sensor technology to achieve the
measurement of oxygen. See Figure 1-1. The sensor is a self contained disposable
unit which requires no maintenance. The sensor utilizes the principle of
electrochemical reaction to generate a signal proportional to the oxygen concentration
in the sample.
The sensor consists of a cathode and anode which are in contact via a suitable
electrolyte. The sensor has a gas permeable membrane which covers the cathode
allowing gas to pass into the sensor while preventing liquid electrolyte from leaking
out.
As the sample diffuses into the sensor, any oxygen present will dissolve in the
electrolyte solution and migrate to the surface of the cathode. The oxygen is reduced
at the cathode. Simultaneously, an oxidation reaction is occurring at the anode
generating four electrons. These electrons flow to the cathode to reduce the oxygen.
The representative half cell reactions are:
Cathode:
4e- + 2H2O + O2 → 4OH
-
Anode:
4OH- + 2Pb → 2PbO + 2H2O + 4e
-
The resultant overall cell reaction is:
2Pb + O2 → 2PbO
This flow of electrons constitutes an electric current which is directly proportional to
the concentration of oxygen present in the sample. In the absence of oxygen, no
oxidation / reduction reaction occurs and therefore no current is generated. This
allows the sensor to have an absolute zero.
Oxygen
Gas Permeable Membrane
Thin Electrolyte Layer
Electrolyte
Anode
OH
-
-
e
Cathode
F
IGURE
1-2
1-1. E
Current Signal
Contact Plates
LECTROCHEMICAL SENSOR TECHNOLOGY
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 15
ALARM SETPOINT ADJUSTMENT
NTRODUCTION
I
F
IGURE
ALARM
SETPOINT
BUTTON
1-2. ROX GT F
CELL BLOCK
COVER
RANGE SELECTOR SWITCH
RONT VIEW
ALARM SETPOINT ADJUSTMENT
LARM
ETPOINT
UTTON
F
IGURE
748375-B Rosemount Analytical July 1998
1-3. ROX GP F
RONT VIEW
CELL BLOCK
COVER
RANGE SELECTOR SWITCH
ROX GT/ROX GP Oxygen Analyzers
1-3
Page 16
NTRODUCTION
I
N
OTES
1-4
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 17
I
NSTALLATION
2
2.1 UNPACKING
Carefully examine the shipping carton and contents for signs of damage. Immediately
notify the shipping carrier if the carton or contents is damaged. Retain the carton and
packing material until all components associated with the instrument are operational.
2.2 ASSEMBLY
The ROX GT comes fully assembled with sensor installed. Please note that the gas
inlet and outlet connections are sealed to prevent exposure of the sensor to air.
Prolonged exposure of the sensor to air can cause extended start up time, reduction
of performance or damage to the sensor. Do not remove the sealing caps until all
associated sample handling components are installed and the instrument is
fully ready for installation.
The ROX GP requires that the sensor be installed after the analyzer has been
installed and gas connections made.
F
IGURE
Cell Block
2-1. ROX T
Meter DisplayAlarm Setting
Power Board
OP VIEW (COVER REMOVED
Switch
Signal Board
Power Entry Module
)
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
2-1
Page 18
NSTALLATION
I
2.3 LOCATION
Install the ROX GT/GP in a clean, weather-protected, non-hazardous, vibration-free
location free from extreme temperature variations. For best results, install the analyzer
near the sample take off point to minimize sample transport time.
Operating ambient temperature is 5 °C to 45 °C, limited to temperature changes of
less than 10 °C/hr. Acceptable dew point range is less than 95% relative humidity, but
not in excess of 45 °C wet bulb temperature.
2.4 GASES
During normal operation, the analyzer requires no support gases. Calibration gases
of air or an appropriate mixture of oxygen in nitrogen are recommended (ppm level for
ROX GT, % level for ROX GP).
After initial startup or startup following a prolonged shutdown, the analyzer may
require extended time to recover to the range of measurement. Commonly, this is
caused by the introduction of ambient air into the sample and/or vent lines to the
sensor. The presence of higher than normal levels of oxygen at the sensor will cause
the sensor electrolyte to become saturated with dissolved oxygen. When the
instrument is placed in operation, the sensor must now consume all excess dissolved
oxygen above the desired measuring level. This recovery period is required only for
the ROX GT.
All new external gas tubing is strongly recommended, preferably pre-cleaned,
stainless steel or copper tubing.
Note
Do not use plastic tubing for trace oxygen measurements as it can permeate
oxygen from the ambient air and cause higher than expected oxygen readings.
ROX GT gas line connections are compression fittings. Do not use pipe thread tape.
ROX GP gas line connections are female NPT and the use of pipe thread tape is
recommended.
2.4.1 C
ONNECTIONS
See Figure 2-2. Connect sample inlet and outlet lines to appropriately labeled fittings
on the rear panel. All connections are 1/4-inch ferrule-type compression fittings on the
ROX GT. The ROX GP has 1/8-inch female NPT connections.
If sample is available at positive pressure and flow between .1 and 2 SCFH we
recommend following the sample handling configuration shown in Figure 2.3.
2-2
If sample is available at ambient pressure or flow is insufficient to meet minimum flow
requirements, we recommend following the sample handling configuration shown in
Figure 2.4. The inclusion of fast loop bypass, particulate filtration and moisture
removal are not shown in these drawings and are the responsibility of the customer to
include as required.
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 19
NSTALLATION
A
A
I
F
IGURE
2.4.2 G
ALIBRATION GASES
C
VENT
(ROX GT)
1/4" SWAGELOK
(ROX GP)
1/8" FPT
SAMPLE IN
(ROX GT)
1/4" SWAGELOK
(ROX GP)
1/8" FPT
2-2. B
ACK PANEL CONNECTIONS
AS SPECIFICATIONS
1 2
+ -
5 6
+ -
7 8 9
C NC NO
10 11 12
C NC NO
LARM CONTACT #2
LARM CONTACT #1
0 - 1VDC OUTPUT
4-20mA ISOLATED OUTPUT
FUSE
POWER
SWITCH
Calibration method and gases depends on the desired operating range, and speed of
calibration recovery. Due to the characteristics of the sensor, no zero gas is required.
In the absence of oxygen, the sensor has an absolute zero.
Span gas can be either air or a certified blend of known ppm concentration of oxygen
in nitrogen. All calibration gases are introduced through the sample inlet at the rear of
the module
AMPLE GAS
S
Sample gas should be nonflammable (below 100% of the sample's LEL) and inert.
Consult factory if sample gas contains CO2, CO, acid gases, or halogen gases. These
samples may not be suitable for use with this analyzer.
LOW RATE
F
The sample flow rate can set be between .1 SCFH and 2.0 SCFH (50 - 1000 cc/min)
without effecting the accuracy of the sensor. Sample flow should be set at a constant
value within these limits which provides the desired response time and sample
conservation.
RESSURE/FILTRATION
P
Sample pressure at the inlet should be within the range of 0 to 15 psig (1013 to
2048 hPa).
The analyzer does not contain any integral sample filtration. Sample should be
filtered for particulates down to five microns to reduce the risk of internal sample line
blockage.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
2-3
Page 20
NSTALLATION
I
Eductor
Vacuum
Breaker
Nitrogen or
Air In
F
IGURE
2-3. R
S
Sample
Flowmeter
Vent
Sample
Inlet
Ambient Air
or Span Gas
Flow Control
Valve
Sensor
Isolation
Valve
ECOMMENDED SAMPLE HANDLING DIAGRAM
AMPLE
ROX GT
(back)
- P
OSITIVE PRESSURE
F
IGURE
2-4
2-4. R
P
Sample
Vent
Flowmeter
Sample
Inlet
Air or
Span Gas
Sensor
Isolation
Valve
Flow Control
Valve
ECOMMENDED SAMPLE HANDLING DIAGRAM
RESSURE OR FLOW
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
ROX GT
(back)
- I
NSUFFICIENT
Page 21
2.5 ELECTRICAL CONNECTIONS
All electrical connections can be found on the back of the analyzer (Refer to Figure
2.2).
NSTALLATION
I
2.5.1 L
INE POWER CONNECTION
The ROX GT/GP is designed to operate on 115 or 230 VAC 50/60 Hz power. The
power entry module is located on the right rear panel of the instrument. The power
entry module utilizes a standard IEC 320 connector allowing universal connection in all
countries by selecting the appropriate line cord at installation. A standard North
American power cord is provided with the analyzer.
The analyzer comes configured from the factory for operation on 115 VAC power. If
you need to operate the unit on 230 VAC, check to make sure that the unit is
disconnected from line power, open the cover door on the power entry module and
extract the fuse-block as explained in Section 4.3. Turn the fuse-block over and
reinstall so that the “230V” is now visible through the window in the cover door. Close
the cover door prior to connecting power.
2.5.2 I
NTERCONNECTIONS
The analyzer has two kinds of output connections, a signal output and a set of relay
contacts. The contacts are rated at 115 VAC, but with the following provisos:
There is only "basic insulation (within the meaning of IEC 1010-1) provided between
the contacts and the rest of the circuitry. This means that in the event of a single fault,
it is possible that 115 VAC connected to the relay contacts would be connected to the
output connections as well. Any equipment connected to the analyzer output may
potentially see this 115 VAC.
WARNING: ELECTRICAL SHOCK HAZARD
The output connection is for use only with equipment which has no live parts
which are accessible.
The connecting cable and the equipment to which the output is connected must
have insulation rated for at least 150 VAC, since under fault condition the output
may be connected to the relay contacts witch themselves may be connected to
150 VAC.
The connection used at the remote end of the output circuit must be such as to
be suitable for 150 VAC and must have no accessible live parts.
The equipment connected to the output must either be approved to IEC 1010-1
or equivalent or must be suitable for use with an input that may potentially be
connected to 150 VAC, and must not catch fire in this circumstance.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
2-5
Page 22
NSTALLATION
I
Note
The interconnection terminal strip on the rear of the analyzer is designed to be
unplugged for easier connection. It is polarized so that you cannot plug it in
backwards. Make the connections with plenty of service loop and then plug it
in. Double check that you have used the correct terminals, and do not apply a
high voltage (e.g. 115 V) to the relay contacts before you have verified that the
output connections are correct - putting AC on the signal output will destroy the
analyzer.
2.5.3 S
IGNAL OUTPUT CONNECTIONS
The analyzer comes standard with both 4-20 mA DC isolated and 0-1 VDC outputs.
For ease of installation, the signal output connections are made via a removable
terminal strip connector which is shared with the alarm connections (Refer to Figure
2.2).
WARNING: POSSIBLE EXPLOSION HAZARD
Check all wiring connections prior to reconnection of the terminal strip. The
application of AC power to the signal output terminals will damage the analyzer.
2.5.4 A
LARM CONNECTIONS
The analyzer comes standard with two alarms both configured high and normally
energized. For ease of installation, the alarm connections are made via a removable
terminal strip connector which is shared with the signal output connections (Refer to
Figure 2.2). Each set of alarm connections are labeled for easy identification to the
associated alarm. The connections labeled NO (normally open), NC (normally
closed), and C (common) refer to relay positions with the power removed (deenergized) .
Both alarms are configured from the factory for fail-safe mode (normally energized,
and loss of power will cause a change in relay status) and high oxygen setpoint (relay
status change above setpoint). To change the alarm configuration, refer to Section
4.5.
2.6 INSTALLATION GUIDELINES
• Is the analyzer’s location clean, weather-proofed, non-hazardous, vibration-free, and
with a stable ambient temperature?
• Are gas supply cylinders equipped with a clean, metal diaphragm, hydrocarbon free,
two stage regulator with shut off valve?
2-6
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 23
NSTALLATION
I
• Are external tubing, regulators, valves, pumps, fittings etc. clean?
• Is the oxygen content of the supply gases compatible with the analysis range?
• Is the sample non-flammable?
• Have all the external gas connections been leaked checked?
• Has clean stainless steel or other appropriate metal tubing been used for sample
lines?
• Are all electrical connections secure and correct?
• Has the sensor been installed in the ROX GP?
• Is the proper line input power displayed on the power entry module?
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
2-7
Page 24
NSTALLATION
I
NOTES
2-8
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 25
S
TARTUP AND OPERATION
3
3.1 STARTUP
After completion of the installation procedure in Section Two, proceed as follows:
1. Set range switch to the 25% position (full counterclockwise)
2. Turn on power by placing the switch on the power entry module to the “-” position.
3.2 CALIBRATION
See Section 2.4.2 for a description of the method for choosing calibration zero and
span gases.
To calibrate the ROX GT/GP, introduce air or span gas into the SAMPLE INLET, and
do the following:
1. Set range switch to the 25% range setting (full counterclockwise) or the appropriate
range for the span gas used. It is recommended to use the highest concentration
span gas which provides you with the best fit for your requirements. The use of air
for calibration provides the least chance of span gas error, but does require
extended recovery time down to low ppm level measurement.
2. If you have followed the recommended sample handling diagram in Section 2.4.1.
open the sensor isolation valve first to avoid exposing the sensor to excessive
sample pressure or vacuum. Flow span gas or air at a floweret between 0.1 and
1.0 L/min. Be careful to not obstruct the outlet as this will cause an increase in
sample pressure and a consequent increase in oxygen reading.
Note
Alarm setpoints are not defeated during calibration. Take appropriate
precautions during calibration.
3. Wait f o r reading to stabilize. This should only take a few minutes.
4. Locate the SPAN ADJUST potentiometer on the front panel and adjust
appropriately to make the display correspond with the span gas concentration.
Turning the potentiometer clockwise will increase the display reading while
counterclockwise will decrease the reading.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GP Oxygen Analyzers
3-1
Page 26
TARTUP AND OPERATION
S
5. If you have used air to span the instrument, switch to zero gas or sample as soon
as possible to avoid prolonged exposure of the sensor to high concentrations of
oxygen. The longer the sensor is exposed to air, the longer it will take for the
sensor to recover to low ppm levels. W hen installing a new sensor or starting the
instrument for the first time, it may take as long as eight hours for the analyzer to
purge down to the lowest operating range.
6. Allow the unit to stabilize at the final reading on zero gas or sample. The unit is
now ready for operation.
Unit should be calibrated once a week until familiarity is achieved in required
calibration frequency. Calibration intervals as long as once a month are possible on
stable installations. Processes which have large changes in environmental conditions
and sample composition may require more frequent calibrations.
3.3 OPERATION
After calibration, proceed as follows:
Supply sample gas to the SAMPLE INLET. Adjust external flow controller or throttle
valve so that flow discharged from the outlet is between 0.1 and 1.0 L/min.
Note
Flow indication is not provided with the instrument and must be provided
external to the analyzer. Refer to installation section for proper location.
When shutting down the instrument, always close the sensor isolation valve last to
avoid exposing the sensor to excessive pressure or vacuum.
3.4 ALARM SETPOINTS
The ROX GT/GP come standard with two fully adjustable alarms with relay contacts.
These alarm setpoints are designed to operate as percent of range as opposed to
discrete concentration value. The alarms are configured from the factory as high and
failsafe.
To view the alarm setpoints, turn the range switch to the desired measuring range and
press the “SET LO ALARM” or “SET HI ALARM” button. The alarm setpoint will be
displayed in actual concentration units. Since the alarms are configured as percent of
scale, changing ranges will also change the value of the alarm setpoint. For example,
if the LO alarm is set for 5 ppm on the 10 ppm range (50% of range), it will read 50
ppm on the 100 ppm range (50% of range).
To adjust the alarm setpoint, press the appropriate alarm setpoint button, then locate
and adjust the associated alarm setpoint adjustment potentiometer as indicated in
Figure 1-1. Turning the potentiometer clockwise will increase the setpoint while
turning the potentiometer counterclockwise will decrease the setpoint. You must keep
the alarm setpoint button depressed during the adjustment procedure to continuously
view the alarm setpoint value. Always adjust the “LO” alarm for the lowest setpoint
value and the “HI” alarm for the higher value.
3-2
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 27
M
AINTENANCE AND TROUBLESHOOTING
4
CAUTION: QUALIFIED PERSONNEL
This equipment should not be adjusted or repaired by anyone except properly
qualified personnel.
WARNING: PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely
affect the safety and performance of this product. Use only factory approved
components for repair.
4.1. MAINTENANCE OVERVIEW
The ROX GT/GP analyzer components which may require replacement include:
• Oxygen sensor
• Power fuse
• Printed circuit boards
4.2 OXYGEN SENSOR REPLACEMENT
CAUTION: CAUS TIC MATERIAL
The sensor contains electrolyte which is caustic and can cause serious burns to
skin. Do not ingest contents of sensor. Read and understand the Material
Safety Data Sheet in the rear of this manual before handling sensor.
The oxygen sensor used in the ROX GT/GP analyzers is a consumable item and does
require replacement periodically. To determine if the sensor requires replacement,
see the troubleshooting portion of this section.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GT Oxygen Analzyers
4-1
Page 28
AINTENANCE AND TROUBLESHOOTING
M
ENSOR INSTALLATION
S
To install a new sensor follow the procedures exactly in the order listed below. Failure
to follow the procedure in order may result is extended exposure of the sensor to air
causing greater delays in returning to on-line monitoring:
1. Turn off power to unit and turn off sample flow to analyzer.
2. Locate cellblock cover and remove using the key tool provided with instrument.
Rotate cover counterclockwise to remove.
3. Inspect cellblock cavity for signs of liquid. If liquid is present, use protective
equipment as described in the material safety data sheet in the rear of this manual
before proceeding. Extreme care should be used and the liquid should be treated
as a corrosive substance.
4. Remove existing sensor and dispose of in accordance with national, federal, state,
and local regulations. Use the existing pull-tab on the sensor to remove from the
cellblock.
5. Inspect inside of cellblock for signs of residual liquid or deposits on the contact
pins. If liquid is present, use appropriate protective equipment and follow the
cleanup procedure as described in the material safety data sheet in the rear of this
manual. Deposits on the contact pins can be removed by wiping with a damp cloth
or using thee eraser from a pencil. Do not use abrasives (i.e. sandpaper) as this
will damage the contact pins.
6. If sample is inert (e.g., nitrogen) a low flow of sample (50 cc/min) can be
introduced to the analyzer to purge out the internal volume of the cellblock prior to
installation of the new sensor. This procedure is not required and is recommended
as a step to reduce the purge down time of the new sensor.
CAUTION: MAXIMUM FLOW
Do not flow a volume in excess of 50 cc/min. Exceeding this level may produce
an oxygen deficient breathing environment to the technician conducting sensor
replacement. Do not flow sample if it is flammable or toxic. Do not use this
procedure if the sample handling configuration is for insufficient pressure or
flow.
4-2
7. Inspect the package containing the replacement sensor. Do not open the bag.
Make sure that it is the correct model type and that there is not visible liquid in the
bag. If the bag contains liquid, do not open and contact the factory for a
replacement.
July 1998 Rosemount Analytical 748375-BROX GT/ROX GT Oxygen Analzyers
Page 29
AINTENANCE AND TROUBLESHOOTING
M
8. Locate the shorting tab on the back of the sensor. The tab is designed to maintain
contact between the two contact plates of the sensor while it is stored in the bag.
This contact allows the sensor to continuously consume dissolved oxygen inside
the sensor. This clip will need to be removed prior to installing the sensor in the
analyzer.
9. Locate the pull-tab portion of the sensor label. W hen installing the sensor in the
cellblock, make sure that this tab faces forward for easy sensor removal later.
10. Cut open the bag being careful not to damage the sensing face of the sensor.
Quickly remove the sensor from the bag, remove the shorting clip and install in the
cellblock with the contact plates on top and the sensing face on the bottom. Make
sure that the pull-tab is facing forward and the writing on the label is right side up
(see Figure 4-1).
11. Reinstall the cellblock cover turning clockwise until the O-ring just makes contact.
Turn no more than 1/8 turn additional.
Note
Do not overtighten. Doing so may damage the cover.
12. T urn on power and check reading.
13. Calibrate as described in Section 3.2
Key Slot
Cell Block
Cover
Contact Plates
Cell Block (with
cover removed)
Oxygen Sensor
Sensing Face
F
IGURE
748375-B Rosemount Analytical July 1998
4-1 S
ENSOR INSTALLATION
Sensor Pull Tab
ROX GT/ROX GT Oxygen Analzyers
4-3
Page 30
AINTENANCE AND TROUBLESHOOTING
M
4.3 POWER FUSE REPLACEMENT
The ROX GT/GP has the main power fuses located in the power entry module. No
other fuses are located inside the instrument.
Before replacing the main power fuse, turn off the analyzer and disconnect the power
cord from the power entry module.
Using a small blade slotted screwdriver, carefully pry open the f ront door of the power
entry module. There is a access slot on the top of the module and the door opens
downward (see Figure 4-2). Once the door is opened, you can remove the red power
selector / fuseholder by inserting the screwdriver into the slot on top of the fuseholder
and carefully prying the fuseholder out. The fuseholder contains two 1A fuses.
Replace as required.
Reinstall the fuseholder so that the correct voltage is visible through the small window
in the door of the power entry module.
ACCESS SLOT
FUSEHOLDER
(RED)
POWER ENTRY MODULE
WITH FRONT DOOR OPEN
F
IGURE
4-2. P
OWER ENTRY MODULE
4.4 PRINTED CIRCUIT BOARDS
Both printed circuit boards can be replaced if necessary. Refer to Figure 2-1 for the
location of the power and signal boards.
4.5 ALARM CONFIGURATION
The alarm setpoints are factory set for high normally energized. If you wish to change
the configuration of the alarm setpoints you can do so by the following procedure.
1. Verify that the power module is off and the power cord is disconnected from the
analyzer
4-4
July 1998 Rosemount Analytical 748375-BROX GT/ROX GT Oxygen Analzyers
Page 31
AINTENANCE AND TROUBLESHOOTING
M
2. Remove the top cover from the analyzer
3. Locate the signal board as indicated in Figure 2-1
4. Locate the alarm setting dip switches on the signal board
5. To change the alarm setting refer to Figure 4-3 and Table 4-1 for desired
configuration
In all cases, “LO ALARM” should be adjusted for the lowest setpoint value and “HI
ALARM” should be adjusted for the higher setpoint value.
F
IGURE
4-3. A
ON OFF
4
3
2
1
EDGE OF SIGNAL BOARD
EDGE OF SIGNAL BOARD
LARM SWITCH LOCATION
LO ALARM SETTING
(S2)
HI ALARM SETTING
(S3)
748375-B Rosemount Analytical July 1998
ROX GT/ROX GT Oxygen Analzyers
4-5
Page 32
AINTENANCE AND TROUBLESHOOTING
M
TRIP POINT - FS/NFSSWITCH #
(RELAY STATUS)1234
HIGH - FAILSAFE
OFFONONOFF
(NORMALLY ENERGIZED)
HIGH - NON FAILSAFE
ONOFFOFFON
(NORMALLY DE-ENERGIZED)
LOW - FAILSAFE
ONOFFOFFON
(NORMALLY ENERGIZED)
LOW - NON FAILSAFE
OFFONONOFF
(NORMALLY DE-ENERGIZED)
T
ABLE
4-1. A
LARM SETTING CONFIGURATION
4.6 TROUBLESHOOTING
The following provides a short list of common troubleshooting tips.
NALYZER DOES NOT POWER UP
A
Check to make sure that the power cord is properly installed and the voltage is
correctly selected. The voltage configuration of the analyzer is visible through the
window of the power entry module. Make sure that the power switch is in the on
position (“-”).
If the unit does not power up, turn off power switch, remove power cord and open door
on power entry module as described in Section 4.4 above. Inspect fuses and replace
as required.
NALYZER FAILS TO PURGE DOWN TO PPM LEVELS
A
The number one problem associated with trace oxygen analyzer installation is the
occurrence of leaks in the sample line plumbing. If the oxygen reading will not come
down to ppm levels or is reading higher than expected, the sample plumbing prior to
the instrument may have a leak. A quick check can be conducted by observing the
oxygen reading at two different flow levels; 0.2 and 2.0 SCFH. If the oxygen reading
drops significantly when the flow is increased from 0.2 to 2.0 SCFH, this is a good
indication that a leak exists.
To check for leaks prior to the sensor, disconnect the analyzer and cap the inlet and
outlet fittings on the analyzer to reduce the amount of oxygen exposure to the sensor.
Cap the inlet line and pressurize the inlet line to 5 - 10 psig and check all connections
with a soapy solution to identify leaks.
4-6
July 1998 Rosemount Analytical 748375-BROX GT/ROX GT Oxygen Analzyers
Page 33
AINTENANCE AND TROUBLESHOOTING
M
WARNING: POSSIBLE SENSOR DAMAGE
Do not pressure check the sample line with the sensor/analyzer connected.
Over-pressurization of the sensor can result in damage.
If the reading increases with an increase in flow, the vent line may be obstructed
causing back pressure on the sensor. Inspect your vent line and remove any devices
which may be causing backpressure.
If the reading slowly decreases over time, the sensor may have been exposed to high
levels of oxygen for extended periods of time. The sensor must consume all of the
dissolved oxygen in the sensor electrolyte before it will accurately measure ppm le vels
of oxygen. In some cases it is not possible for the sensor to recover to low ppm levels
and the sensor must be replaced.
NALYZER IS READING TOO LOW
A
Check to make sure that the sensor has been calibrated properly. Flow an
appropriate span gas through the instrument and follow the calibration procedure in
Section 3.2. If you are unable to achieve span calibration, the sensor has reached the
end of its useful life and needs to be replaced (see Section 4.2).
If calibration is possible, but the reading still appears too low, conduct a span check on
the 25% range with air. Adjust reading as necessary. If the reading requires
adjustment, your bottled span gas may be incorrect or corrupted.
If sensor life seems too short, you may have compounds present in your sample which
are poisoning the sensor. Acid gases, CO2, SO2, or H2S can reduce the life of the
sensor and should be removed from the sample prior to measuring.
NALYZER ALWAYS READS ZERO
A
Check to make sure that the sensor has been installed correctly with the contact
plates on top and the gas permeable membrane facing downward. If the sensor has
been installed incorrectly, carefully remove and check to make sure that the
membrane has not been damaged. If the sensor is leaking, or shows signs of distress
to the membrane, replace with a new sensor. If the sensor is okay quickly reinstall in
the correct position.
Verify that the cellblock contact pins are making good contact with the metal surfaces
of the sensor. If the pins are not making contact, carefully bend the wire down to
reestablish contact. Check that the contact pins are clean and do not show signs of
deposits or oxidation. Clean as necessary.
748375-B Rosemount Analytical July 1998
ROX GT/ROX GT Oxygen Analzyers
4-7
Page 34
AINTENANCE AND TROUBLESHOOTING
M
A good sensor will have an output of 150 to 500 micro amps in air. You can check this
by connecting a DVM across the contact plates of the sensor while it is removed from
the instrument and measuring the current. If the sensor does not produce sufficient
output, replace the sensor with a new one.
HE READINGS ARE ERRATIC AND DRIFT NOISILY
T
This is normally an indication that the sensor has reached the end of its useful life. If
the analyzer cannot be spanned on a cal gas or air, the sensor must be replaced. If
the sensor spans okay and appears stable at high oxygen concentrations, there may
be a sample line leak or obstruction of the vent line. Follow the procedure described
above in “Analyzer fails to purge down to ppm levels”.
If no sample line problems are found, open the cellblock and inspect the sensor for
signs of damage or leakage. Follow the procedure in Section 4.2. for correct method
of opening cellblock and precautions for handling a leaking sensor.
4-8
July 1998 Rosemount Analytical 748375-BROX GT/ROX GT Oxygen Analzyers
Page 35
R
EPLACEMENT PARTS
5
5.1 ROX REPLACEMENT PARTS
WARNING: PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely
affect the safety and performance of this product. Use only factory approved
components for repair.
5.1.1 R
5.1.2 R
EPLACEMENT SENSORS
Refer to Table A-1 in Appendix A for easy cross reference of the correct oxygen
sensor for your specific instrument.
July 1998 Rosemount Analytical 748375-BROX GT/ROX GP Oxygen Analyzers
Page 37
A
PPENDIX
A. M
ATERIAL SAFETY DATA SHEETS
A
The following section contains important information regarding the characteristics of
the oxygen sensor used in the ROX GT/GP analyzers. Read and understand all
information before attempting any replacement or installation of the oxygen sensor.
The following sensor reference guide will provide a quick reference of the specific
oxygen sensor type used in your specific analyzer model and part number.
September 1997 Rosemount Analytical 748375-AROX GT/ROX GP Oxygen Analyzers
Page 39
Rosemount Analytical
Material Safety Data Sheet
748378-A
PRODUCT:
PART NUMBER:
XYGEN SENSOR
O
658303, 658310, 658312
24 HOUR EMERGENCY TELEPHONE NUMBER:
CHEMTREC (800) 424-9300
SECTION I - GENERAL
Distributor:Rosemount Analytical Inc.
4125 East La Palma Avenue, CA 92807-1802
714-986-7600
Chemical name and synonyms
Trade name and synonyms
Chemical family
Formula
CAS Number
LeadCaustic Potash
Inorganic baseInorganic base
PbKOH
749-92-11310-58-3
SECTION II – HAZARDOUS INGREDIENTS
Quantity
OSHA PEL
ACGIH
3-20 gms 1-5 ml
0.05 mg/m
0.15 mg/m
SECTION III – PHYSICAL DATA
Boiling point
Melting point
Vapor pressure
Vapor density (air=1)
Specific gravity (H2O=1)
% Volatile by volume
Evaporation rate (H2O=1)
Solubility in water
Appearance and odor
1744°C1320°C
328°C360°C
NANA
NANA
11.342.04
NANA
NANA
InsolubleComplete
Solid, silver gray, odorless
SECTION IV – FIRE AND EXPLOSION HAZARD DATA
Flash point
Extinguishing media
Special fire fighting procedures
Unusual fire and explosion hazards
NA
Appropriate to surroun ding fi re condition s. No specifi c agents
recommended.
Wear NIOSH/OSHA approv ed self-co ntained br eathin g apparat us
and protective clothing to prevent con tact with skin and eyes.
Emits toxic fumes under fire condi tions.
SECTION V – REACTIVITY DATA
Stability
Conditions to avoid
Incompatibility (material to avoid)
Hazardous decomposition or
byproducts
Hazardous polymerization
Stable
Avoid contact with acids and hydrogen peroxide >52%
Aluminum, organic materials, acid chlorides, acid anhydrides,
magnesium, copper.
Toxic fumes
Will not occur
LeadPotassium Hydroxide
3
3
None
2 mg/m
White or slightly yellow,
odorless
Rosemount Analytical Inc.
4125 East La Palma Avenue • Anaheim, California 92807-1802 • TEL 714-986-7600 • FAX 714-577-8006
April 1997 • 748378-A • Printed in USA
Page 40
MATERIAL SAFETY DATA SHEET
SECTION VI – HEALTH HAZARD DATA
LeadPotassium Hydroxide
Threshold limit value
Routes of entry
Chronic exposure to lead may
cause disease of the blood and
blood forming organs , kidneys and
liver, damage to the reproductive
systems and decrease i n fertilit y in
men and women, and damage to
the fetus of a pregnant woman.
Chronic exposure from t he lead
contained in this product is
extremely unlikely.
Eyes and skin
minutes. Remove contaminated clothing. Call a physician.
ORAL INGESTION - DO NOT INDUCE VOMITING
conscious, give quantities of milk or water. Call a physician
immediately and take patient to hospital emergency room.
- immediately flush with water for at least fifteen
Acute Effects: Corrosive to
tissue. May cause irritation or
chemical burns to skin and
eyes. Harmful if swallowed,
inhaled or absorbe d thro ugh the
skin.
Chronic Effects
exposure has a destructive
effect on tissue.
: Prolonged
. If person is
SECTION VII – SPILL OR LEAK PROCEDURE
NOTE: The oxygen sensors are sealed, and under normal circumstances, the contents of the
sensors do not present a health hazard. The following information is given as a guide in the event
that a cell leaks.
Steps to be taken in case
material is released or spilled
Waste disposal method
Wipe down the area se veral times, each time with a fresh, wet paper
towel. Contaminated paper towels are considered hazardous waste.
Both lead and potassium hydroxide are considered poisonous
substances and are regulated under TSCA and SARA Title III.
EPA Waste Number: D008
California Waste Number: 181
DOT Information: RQ Hazardous Waste So lid N.O.S. ( Lead), 9 , UN3077 , PG III.
Follow all Federal, State and Local regulations.
SECTION VIII – SPECIAL PROTECTION INFORMATION
Respiratory protection
Ventilation
Protective gloves
Eye protection
Other protective equipment
NA
NA
Rubber gloves.
Chemical splash goggles.
Apron, face shield
SECTION IX – SPECIAL PRECAUTIONS
Protective measures during cell replacement: Before opening the bag
Precautions to be taken in
handling and storing
Other precautions
containing the sensor cell, c heck the sensor cell for leakage. If sensor
cell leaks, do not o pen the bag. If liquid is pr esent around cell whi le in
the instrument, wear eye and hand protect ion .
none
SECTION X – TRANSPORTATION
Must be compliance with federal, state and local regulations
748378-A
NOTICE: WHILE ROSEMOUNT ANALYTICAL BELIEVES THE INFORMATION CONTAINED HEREIN IS
VALID AND ACCURATE, ROSEMOUNT ANALYTIC AL MAKES NO W ARRANTY OR REPRESENTATION AS
TO ITS VALIDITY, ACC URACY, OR CURRENCY. ROSEMOUNT ANALYTICAL SHALL NOT BE LIABLE OR
OTHERWISE RESPON SIBLE IN AN Y WAY FOR U SE OF EIT HER THIS INFO RMATION OR T HE MAT ERIAL
TO WHICH IT APPLIES. DIS PO SAL O F HAZ ARDO US M AT ERIA L M A Y B E SU BJE CT TO FEDERAL, ST ATE,
OR LOCAL LAWS AND/OR REGULATIONS.
Sheet 2 of 2
April 1997
Page 41
Rosemount Analytical
Material Safety Data Sheet
748380-B
PRODUCT:
PART NUMBER:
XYGEN SENSOR
O
658304
, M
ODEL
GT3
24 HOUR EMERGENCY TELEPHONE NUMBER:
CHEMTREC (800) 424-9300
SECTION I - GENERAL
Distributor:Rosemount Analytical Inc.
4125 East La Palma Avenue, CA 92807-1802
714-986-7600
Chemical name and synonyms
Trade name and synonyms
Chemical family
Formula
CAS Number
SECTION II – HAZARDOUS INGREDIENTS
Hazardous mixtures of other liquids,
solids or gases
SECTION III – PHYSICAL DATA
Boiling point
Melting point
Vapor pressure
Vapor density (air=1)
Specific gravity (H2O=1)
% Volatile by volume
Evaporation rate (H2O=1)
Solubility in water
Appearance and odor
SECTION IV – FIRE AND EXPLOSION HAZARD DATA
Flash point
Extinguishing media
Special fire fighting procedures
Unusual fire and explosion hazards
SECTION V – REACTIVITY DATA
Stability
Conditions to avoid
Incompatibility (material to avoid)
Hazardous decomposition or
byproducts
Hazardous polymerization
acetic acid
(5% v/v)
acetic acidpotassium acetatelead
inorganic acidinorganic baseinorganic base
- The electrolyte is corrosive; eye contact could result in
Eyes
permanent loss of vision.
Skin - The electrolyte is corr osiv e ; skin con ta ct coul d re sul t in a
chemical burn.
Ingestion
Eyes - Flush eyes with water for at least 15 minutes and get
immediate medical attention.
Skin
contaminated clothing. If burning persists, seek medical attention.
Ingestion - Give plenty of cold water. DO NOT INDUCE
VOMITING. Seek medical attention.
- The electrolyte could be harmful or fatal if swallowed.
- Wash affected area with plenty of water and remove
NA0.15 mg/cu.m
SECTION VII – SPILL OR LEAK PROCEDURE
NOTE: THE OXYGEN SENSORS ARE SEALED, AND UNDER NORMAL CIRCUMSTANCES, THE
CONTENTS OF THE SENSORS DO NOT PRESENT A HEALTH HAZARD. THE FOLLOWING
INFORMATION IS GIVEN AS A GUIDE IN THE EVENT THAT A CELL LEAKS.
Steps to be taken in case material is
released or spilled
Waste disposal method
Wipe down the area several times with a wet paper towel. Use a
fresh towel each time.
Should be in acco rdan ce wi th all ap pli c ab le sta te , local a nd fede r al
regulations.
SECTION VIII – SPECIAL PROTECTION INFORMATION
Respiratory protection
Ventilation
Protective gloves
Eye protection
Other protective equipment
NA
NA
Rubber gloves
Chemical splash goggles
approved working clothes
SECTION IX – SPECIAL PRECAUTIONS
During cell replacement, before opening the bag containing the
Precautions to be taken in handling
and storing
Other precautions
sensor cell, check the sensor cell for leakage. If the sensor cell
leaks, do not open the bag. If there is liquid around the cell while
in the instrument, put on gloves and eye protection before
removing the cell.
none
SECTION X – TRANSPORTATION
Must be compliance with federal, state and local regulations
NOTICE: WHILE ROSEMOUNT ANALYTICAL BELIEVES THE INFORMATION CONTAINED HEREIN IS
VALID AND ACCURATE, ROSEMOUNT ANALYTIC AL MAKES NO W ARRANTY OR REPRESENTATION AS
TO ITS VALIDITY, ACC URACY, OR CURRENCY. ROSEMOUNT ANALYTICAL SHALL NOT BE LIABLE OR
OTHERWISE RESPON SIBLE IN AN Y WAY FOR U SE OF EIT HER THIS INFO RMATION OR T HE MAT ERIAL
TO WHICH IT APPLIES. DIS PO SAL O F HAZ ARDO US M AT ERIA L M A Y B E SU BJE CT TO FEDERAL, ST ATE,
OR LOCAL LAWS AND/OR REGULATIONS.
Sheet 2 of 2
July 1998
Page 43
Rosemount Analytical
Material Safety Data Sheet
748381-A
PRODUCT:
PART NUMBER:
XYGEN SENSOR
O
658311
, M
ODEL
GP3
24 HOUR EMERGENCY TELEPHONE NUMBER:
CHEMTREC (800) 424-9300
SECTION I - GENERAL
Distributor:Rosemount Analytical Inc.
4125 East La Palma Avenue, CA 92807-1802
714-986-7600
potassium dibasic
Chemical name and synonyms
Trade name and synonyms
Chemical family
Formula
CAS Number
SECTION II – HAZARDOUS INGREDIENTS
Hazardous mixtures of other liquids,
solids or gases
SECTION III – PHYSICAL DATA
Boiling point
Melting point
Vapor pressure
Vapor density (air=1)
Specific gravity (H2O=1)
% Volatile by volume
Evaporation rate (H2O=1)
Solubility in water
Appearance and odor
SECTION IV – FIRE AND EXPLOSION HAZARD DATA
Flash point
Extinguishing media
Special fire fighting procedures
Unusual fire and explosion hazards
SECTION V – REACTIVITY DATA
Stability
Conditions to avoid
Incompatibility (material to avoid)
In sensor, stable under normal conditions of use.
Avoid contact between sensor electrolyte and strong acids and
oxidizing agents.
NA
NA
will not occur
OKHCO
4
2
•3H
potassium
bicarbonate
White, odorless
crystal
potassium
bicarbonate
(5.11% w/v)
3
1
lead
(pure)
lead
Pb
Grey, odorless
1
The electrolyte in the Model GP3 Oxygen Sensor is not flammable. There are not fire or explosion hazards associated with
this sensor.
Rosemount Analytical Inc.
4125 East La Palma Avenue • Anaheim, California 92807-1802 • TEL 714-986-7600 • FAX 714-577-8006
May 1997 • 748381-A • Printed in USA
Page 44
MATERIAL SAFETY DATA SHEET 748381-A
SECTION VI – HEALTH HAZARD DATA
Threshold limit value
Routes of entry
Effects of overexposure
Emergency & first aid procedures
NA
Skin, eyes, ingestion. Liquid inhalation highly unlikely.
Eyes - The electrolyte is corrosive; eye contact could result in
permanent loss of vision.
Skin - The electrolyte is corr osiv e ; skin con ta ct coul d re sul t in a
chemical burn.
Ingestion - The electrolyte could be harmful or fatal if swallowed.
Eyes - Flush eyes with water for at least 15 minutes and get
immediate medical attention.
Skin - Wash affected area with plenty of water and remove
contaminated clothing. If burning persists, seek medical attention.
Ingestion - Give plenty of cold water. DO NOT INDUCE
VOMITING. Seek medical attention.
SECTION VII – SPILL OR LEAK PROCEDURE
NOTE: THE OXYGEN SENSORS ARE SEALED, AND UNDER NORMAL CIRCUMSTANCES, THE
CONTENTS OF THE SENSORS DO NOT PRESENT A HEALTH HAZARD. THE FOLLOWING
INFORMATION IS GIVEN AS A GUIDE IN THE EVENT THAT A CELL LEAKS.
Steps to be taken in case material is
released or spilled
Waste disposal method
Wipe down the area several times with a wet paper towel. Use a
fresh towel each time.
Should be in acco rdan ce wi th all ap pli c ab le st ate , local a nd fe der al
regulations.
SECTION VIII – SPECIAL PROTECTION INFORMATION
Respiratory protection
Ventilation
Protective gloves
Eye protection
Other protective equipment
NA
NA
Rubber gloves
Chemical splash goggles
approved working clothes
SECTION IX – SPECIAL PRECAUTIONS
During cell replacement, before opening the bag containing the
Precautions to be taken in handling
and storing
Other precautions
sensor cell, check the sensor cell for leakage. If the sensor cell
leaks, do not open the bag. If there is liquid around the cell while
in the instrument, put on gloves and eye protection before
removing the cell.
none
SECTION X – TRANSPORTATION
Must be compliance with federal, state and local regulations
NOTICE: WHILE ROSEMOUNT ANALYTICAL BELIEVES THE INFORMATION CONTAINED HEREIN IS
VALID AND ACCURATE, ROSEMOUNT ANALYTIC AL MAKES NO W ARRANTY OR REPRESENTATION AS
TO ITS VALIDITY, ACC URACY, OR CURRENCY. ROSEMOUNT ANALYTICAL SHALL NOT BE LIABLE OR
OTHERWISE RESPON SIBLE IN AN Y WAY FOR U SE OF EIT HER THIS INFO RMATION OR T HE MAT ERIAL
TO WHICH IT APPLIES. DIS PO SAL O F HAZ ARDO US M AT ERIA L M A Y B E SU BJE CT TO FEDERAL, ST ATE,
OR LOCAL LAWS AND/OR REGULATIONS.
Sheet 2 of 2
May 1997
Page 45
ENERAL PRECAUTIONS FOR HANDLING AND
G
TORING HIGH PRESSURE GAS CYLINDERS
S
Edited from selected paragraphs of the Compressed
Gas Association's "Handbook of Compressed Gases"
published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against
extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders
should be stored in the shade when located in areas where extreme temperatures are
prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured
against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be
moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an
empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A
flame should never be permitted to come in contact with any part of a compressed gas
cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric
arc welding, precautions must be taken to prevent striking an arc against the cylinder.
4125 E
AST LA PALMA AVENUE
• A
Rosemount Analytical Inc.
, C
J
ULY
NAHEIM
ALIFORNIA
1997 • 748525-C • P
92807-1802 • 714-986-7600 • FAX 714-577-8006
RINTED IN
USA
Page 46
(blank)
Page 47
ARRANTY
W
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from
defects in workmanship and material under normal use and service for a period of twelve (12)
months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid
junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and
material under normal use and service for a period of ninety (90) days from date of shipment by
Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or
material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods,
part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid,
within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of
consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for
replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in
the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall
not operate to condemn such commercial unit when such goods, part(s) and consumables are
capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or
indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any
other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER
GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER W ARRANTIES ARE
GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN
PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL
BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD W ARRANTY CLAUSE. IN
NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S
LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY
SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT
SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL
DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS
OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND
DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE
OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR
PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE
AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH
SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond
Seller's direct control.
Rosemount Analytical
4125 E
AST LA PALMA AVENUE
• A
F
EBRUARY 1997 • 7485189-C • PRINTED IN USA
Rosemount Analytical Inc.
NAHEIM
, C
ALIFORNIA
92807-1802 • 714-986-7600 • FAX 714-577-8006
Page 48
(blank)
Page 49
IELD SERVICE AND REPAIR FACILITIES
F
Field service and repair facilities are located worldwide.
U.S.A.
To obtain field service on-site or assistance with a service problem, contact (24 hours, 7
days a week):
National Response Center
1-800-654-7768
INTERNATIONAL
Contact your local Rosemount Sales and Service office for service support.
FACTORY
For order administration, replacement Parts, application assistance, on-site or factory repair,
service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when you call:
Model Number, Serial Number, and Purchase Order Number or Sales Order Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sende r, f re ight collect.
When return ing any product or compon ent that has been expo sed to a toxic, co rrosive or
other hazardous material or used in such a hazardous environment, the user must attach an
appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the material
has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
4125 E
AST LA PALMA AVENUE
• A
Rosemount Analytical Inc.
, C
NAHEIM
ULY 1997 • 748190-G • PRINTED IN USA
J
ALIFORNIA
92807-1802 • 714-986-7600 • FAX 714-577-8006
Page 50
(blank)
Page 51
Rosemount Analytical
ADDENDUM
015-748399-A
ODELS
M
ROX GT, ROX GP O
XYGEN ANALYZERS INSTRUCTION MANUAL
748375
1.0 OVERVIEW
This addendum serves as an amendment to the Models ROX GT and ROX GP
Oxygen Analyzers instruction manual (PN 748375).
The following information should be considered part of the manual, and supersedes
any conflicting information in the body of the manual.
READ THIS INFORMATION AND NOTE THE CONFLICTS
2.0 24 VDC ELECTRICAL CONNECTIONS
The ROX GT 24 VDC Analyzer (PN 658301-24) is designed to operate on 24 VDC
input.
Prior to making electrical connections to the analyzer, verify that the power rating on
the instrument name-rating label is correct.
The 24 VDC electrical connection is located on the right rear panel of the instrument.
The terminal strip is labeled: (Refer to Figure 1.)
+
-
+24 VDC
-24 VDC
Ground
WARNING: POSSIBLE ELECTRICAL DAMAGE
Connection of AC voltage to the power input connections will result in
serious damage to the analyzer. Verify all electrical input connections for
correct voltage and power before powering up analyzer.
Rosemount Analytical Inc.
4125 E
AST LA PALMA AVENUE • ANAHEIM
J
ULY
, C
1998
ALIFORNIA
748399-A
•
92807-1802 •
• PRINTED IN
TEL 714-986-7600 •
USA
FAX 714-577-8006
Page 52
ADDENDUM
F1A - 24V
FUSE
+
-
24 VDC POWER CONNECTION
FIGURE 1. REAR PANEL ELECTRICAL INPUT CONNECTION
3.0 POWER FUSE REPLACEMENT
The ROX GT 24 VDC Analyzer has the main power fuse located directly above the
electrical power input terminal strip on the rear panel. No other fuses are located in
the analyzer.
Before replacing the main power fuse, turn off and disconnect electrical power to the
analyzer.
The fuseholder utilizes a standard twist-off cap. To replace the fuse, rotate the
fuseholder cap counterclockwise until the cap is free from the body. The fuse will
come out with the cap. Replace fuse and re-install cap onto fuseholder.
2 of 2
July 1998
748399-ARosemount Analytical
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.