Rosemount Quick Start Guide: Rosemount Volume 1 Sensor Assembly Manuals & Guides

Quick Start Guide
00825-0300-2654, Rev GB
March 2021
Rosemount™ Volume 1 Sensor Assembly
Quick Start Guide March 2021
Contents
About this guide...........................................................................................................................3
Wiring diagrams...........................................................................................................................4
Rosemount Series 58C sheath cutting ......................................................................................... 6
Drawings......................................................................................................................................7
Rosemount Series 68 Platinum RTD........................................................................................... 10
Rosemount Series 78 Platinum RTD........................................................................................... 12
Rosemount Series 183 Thermocouple........................................................................................15
Rosemount Series 68Q Sanitary Platinum RTD ...........................................................................18
Rosemount Series 58C Platinum RTD......................................................................................... 20
Product certifications................................................................................................................. 21
2 Emerson.com/Rosemount
March 2021 Quick Start Guide
This guide provides basic guidelines for Sensor models. It does not provide instructions for configuration, diagnostics, maintenance, service, troubleshooting, explosion-proof, flameproof, or intrinsically safe (I.S.) installations. If the Rosemount Volume 1 Sensor was ordered assembled to a temperature transmitter, see the appropriate transmitter Quick Start Guide for information on configuration and hazardous locations certifications.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
CAUTION
Refer to Product Certification section of this Quick Start Guide documentation.
Quick Start Guide 3
A
A
B
B
B B
A
C C
D
Quick Start Guide March 2021

2 Wiring diagrams

Figure 2-1: Rosemount Series 68, 68Q, 78, and 58C RTD Wire Colors
Single Element Dual Element
(1) Dual element sensors are only available on Rosemount Series 68Q and 78
Sensors.
A. Red B. White C. Green
D. Black
(1)
Note
For three-wire systems, use one white and two red leads. Do not connect the white leads. Insulate or terminate the unused white lead in a manner that prevents shorting to the ground. For two-wire systems, connect both sets of leads.
4 Emerson.com/Rosemount
+ A
- B
+ C
- B
+ D
- B
+ E
- B
March 2021 Quick Start Guide
Figure 2-2: Rosemount Series 183 Thermocouple Wire Colors
Type J Type E
Type K Type T
A. White B. Red C. Purple
D. Yellow
E. Blue
Note
To distinguish the two sensors in dual Rosemount 183 Sensors, there is an outer insulation wrapped around each pair of sensor wires.
Quick Start Guide 5
Quick Start Guide March 2021

3 Rosemount Series 58C sheath cutting

Procedure
1. Determine the length to which the sheath will be cut. The finished length needs to include an additional 1.5-in. (3.8 cm) for compression fittings or 2.5-in. (6.5 cm) for spring-loaded fittings (see
Figure 4-1).
2. Remove and save the heat shrink tubing from the rear of the sensor.
3. Place the sensor in a vise, taking care not to overtighten, and position the tubing cutter on the sheath.
4. Score the sheath to a depth of approximately 1/64-in. (0.4 cm) To prevent damage to the lead wire insulation, do not cut completely through the sheath.
5. Firmly grasp the end of the sheath with your hand or a pair of pliers. Using a sharp snapping motion, break off and remove the excess sheath material. Take care not to strip or damage the lead wire insulation while removing the excess sheath material.
Note
If you are unable to easily break off excess sheath material, deepen the score and repeat Step 5.
6. Replace the heat shrink tubing.
6 Emerson.com/Rosemount
A
B
C
D
E
F
March 2021 Quick Start Guide

4 Drawings

Figure 4-1: Rosemount Series 58C Sensor
A. Four lead wires 6-in. (152 mm) long. B. X length ±0.25 (±6) C. 0.25 ±0.002 (6.35 ±0.13) diameter
D. Heat shrink tubing
E. Do not cut or bend sheath within 2-in. (51 mm) F. 0.6-in. (15 mm) max. sensing element
Quick Start Guide 7
A
B
D
C
E
F
I
G
H
K
J
L
M
O
N
P
Q
R
S
T
Quick Start Guide March 2021
Figure 4-2: Sensor Assembly
A
Open identification tag
B
Standard adapter sensor assembly
C
Sensor immersion length “X”
D
0.5-in. (13 mm) nominal engagement
E
F
G
H
I
J
8 Emerson.com/Rosemount
Coupling nipple
Extension length
Overall thermowell length
Lead wire extensions and seals
0.5-in. (13 mm) nominal engagement
Extensions
K
Union nipple
L
Flat or extended cover aluminum connection heads
M
Threaded thermowell
N
Thermowells
O
T + 1.75-in. (44.5 mm)
P
Socket weld thermowell
Q
Polypropylene connection head
R
Thermowell immersion length
S
Flanged thermowell
T
Rosemount aluminum connection head
March 2021 Quick Start Guide
Note
Sensor assemblies can be provided without an enclosure or with an enclosure such as the connection heads shown above or assembled to a Rosemount transmitter.
Quick Start Guide 9
Quick Start Guide March 2021

5 Rosemount Series 68 Platinum RTD

5.1 Specifications

5.1.1 Performance specifications

Temperature range
–50 to 400 °C (–58 to 752 °F)
Effect of temperature cycling
±0.05 percent (0.13 °C or 0.23 °F) maximum ice-point resistance shift following 10 cycles over the specified temperature range
Stability
±0.11 percent 0.28 °C or 0.51 °F maximum ice-point resistance shift following 1,000 hours at maximum specified temperature (400 °C)
Maximum hysteresis
±0.1 percent of operating temperature range
Time constant
12 seconds maximum required to reach 63.2 percent sensor response in water flowing at 3 ft/s (0.91 m/s)
Nominal R0 100 Ohm
Nominal alpha 0.00385 Ω/Ω °C

5.1.2 Physical specifications

Material selection
Emerson provides a variety of Rosemount product with various product options and configurations including materials of construction that can be expected to perform well in a wide range of applications. The Rosemount product information presented is intended as a guide for the purchaser to make an appropriate selection for the application. It is the purchaser’s sole responsibility to make a careful analysis of all process parameters (such as all chemical components, temperature, pressure, flow rate, abrasives, contaminants, etc.), when specifying product, materials, options and components for the particular application. Emerson is not in a position to evaluate or guarantee the compatibility of the process fluid or other process parameters with the product, options, configuration or materials of construction selected.
Sheath material
316 SST and 321 SST
10 Emerson.com/Rosemount
Loading...
+ 22 hidden pages