Rosemount Quick Start Guide: Rosemount FCL Free Chlorine System with Rosemount 56 Transmitter Manuals & Guides

Quick Start Guide
00825-0200-3412, Rev AA
January 2020
Rosemount™ FCL
Free Chlorine System with Rosemount 56 Transmitter
Quick Start Guide January 2020
Read this page before proceeding!
Your instrument purchase from Emerson is one of the finest available for your particular application. Emerson designs, manufactures, and tests its products to meet many national and international standards. Experience indicates that its performance is directly related to the quailty of the installation and knowledge of the user in operating and maintaining the instrument. To ensure continued operation to the design specifications, read this Manual thoroughly before proceeding with installation, commissioning, operation, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.
Ensure that you have received the correct model and options from your purchase order. Verify
that this Manual covers your model and options. If not, call 1-800-854-8257 or 949-757-8500 to request the corrrect Manual.
For clarification of instructions, contact your Rosemount representative.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Use only qualified personnel to install, operate, program, and maintain the product.
Install equipment as specified in the installation instructions of the appropriate Reference Manual
and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
Use only factory documented components for repair. Tampering or unauthorized substitution of
parts and procedures can affect the performance and cause unsafe operation of your process.
WARNING
Electrical shock
Making cable connections to and servicing this instrument require access to shock hazard level voltages, which can cause death or serious injury.
Equipment protected throughout by double insulation. Disconnect main power and relay contacts wired to separate power sources before servicing. Do not operate or energize instrument with case open. Non-metallic cable strain reliefs do not provide grounding between conduit connections. Use
grounding type bushings and jumper wires. Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70)
and/or any other national or local codes. Unused cable conduit entries must be securely sealed by non-flammable closures to provide
exposure integrity in compliance with personal safety and environmental protection requirements. Unused conduit openings must be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (IP65).
Operate only with front and rear panels fastened and in place over terminal area. Safety and performance require that this instrument be connected and properly grounded
through a three-wire power source.
WARNING
This product is not intended for use in the light industrial, residential, or commercial environments per the instrument's certification to EN50081-2.
2 Emerson.com/Rosemount
January 2020 Quick Start Guide
CAUTION
Radio interference
This product generates, uses, and can radiate radio frequency energy and thus can cause radio communication interference. Improper installation or operation may increase such interference. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of Class A computing devices, pursuant to Subpart J of Part 15 of FCC rules, which are designed to provide reasonable protection against such interference.
Operation of this equipment in a residential area may cause interference, in which case the operator, at his own expense, will be required to take whatever measures may be required to correct the interference.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
Contents
Specifications...............................................................................................................................5
Install........................................................................................................................................... 7
Wire........................................................................................................................................... 12
EU Declaration of Conformity.....................................................................................................19
China RoHS table........................................................................................................................21
Quick Start Guide 3
Quick Start Guide January 2020
4 Emerson.com/Rosemount
January 2020 Quick Start Guide

1 Specifications

Rosemount 56 Transmitter
For Rosemount 56 Transmitter specifications, see the Rosemount 56 Transmitter Reference Manual on Emerson.com/Rosemount: Manual: Rosemount 56 Advanced Dual-Input Transmitter.
Table 1-1: General Specifications
Characteristic Specifications
Sample requirements
Pressure 3 to 65 psig (122 to 549 kPa abs)
Temperature 32 to 122 °F (0 to 50 °C)
Minimum flow 3 gal/hr (11 L/hr)
Maximum flow 80 gal/hr (303 L/hr); high flow causes the
Sample conductivity > 50 µS/cm at 77 °F (25 °C)
Process connection ¼-in. OD tubing compression fitting (can be
Drain connection ¾-in. barbed fitting. Sample must drain to
Wetted parts Overflow sampler and flow cell: acrylic,
Response time < 80 sec to 95% of final reading for inlet
Weight/shipping weight
(1)
A check valve in the inlet prevents the sensor flow cells from going dry if sample flow is lost. The check valve opens at 3 psig (122 kPa abs). If the check valve is removed, minimum pressure is 1 psig (108 kPa abs).
overflow tube to back up.
removed and replaced with barbed fitting for soft tubing)
open atmosphere.
polycarbonate, Kynar™, nylon, silicone Chlorine sensor: Noryl™, Viton™, wood, silicone, polyethersulfone, polyester, and platinum
pH sensor (3900 VP): stainless steel, glass, Teflon™, polyphenylene sulfide, EPDM, and silicone
sample flow of 3 gph (11 L/hr)
Rosemount FCL-01: 10 lb./13 lb. (4.5 kg/6.0 kg) Rosemount FCL-02: 11 lb./14 lb. (5.0 kg/6.5 kg)
(1) Rounded to the nearest 1 lb. (0.5 kg)
Quick Start Guide 5
Quick Start Guide January 2020
Table 1-2: Sensor Specifications
Characteristics Specifications
Free chlorine range 0 to 10 ppm as Cl2. For higher ranges, consult
pH correction range 6.0 to 9.5. For samples having pH between 9.5
Accuracy Accuracy depends on the accuracy of the
Interferences Monochloramine, permanganate, peroxides
Electrolyte volume 25 mL (approx.)
Electrolyte life 3 months (approx.); for best results, replace
the factory.
and 10.0, consult the factory. If pH < 6.0, correction is not necessary. For manual pH correction, choose option -01. For continuous pH correction, choose option -02.
chemical test used to calibrate the sensor.
electrolyte monthly.
6 Emerson.com/Rosemount
January 2020 Quick Start Guide

2 Install

2.1 Unpack and inspect

Procedure
1. Inspect the shipping container(s). If there is damage, contact the shipper immediately for instructions.
2. If there is no apparent damage, unpack the container(s).
3. Ensure that all items shown on the packing list are present. If items are missing, notify Emerson immediately.

2.1.1 Rosemount™ FCL-01 (free chlorine without continuous pH correction)

The Rosemount FCL-01 consists of the following items mounted on a back plate.
1. Rosemount 56-03-24-38-HT transmitter with sensor cable attached.
2. Constant head overflow sampler with flow cell for chlorine sensor.
The free chlorine sensor (Rosemount 499ACL-01-54-VP), three membrane assemblies, and a bottle of electrolyte solution are in a separate package.

2.1.2 Rosemount™ FCL-02 (free chlorine with continuous pH correction)

The Rosemount FCL-02 consists of the following items mounted on a back plate:
1. Rosemount 56-03-24-38-HT transmitter with sensor cables attached.
2. Constant head overflow sampler with flow cells for pH and chlorine sensors.
3. Stand to hold pH buffer solution during calibration.
The free chlorine sensor (Rosemount 499ACL-01-54-VP), shipped with three membrane assemblies and a bottle of electolyte solution, and the Rosemount 3900VP-02-10 pH sensor are in separate packages.
2.2
Quick Start Guide 7

General installation information

1. Although the system is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperature.
Quick Start Guide January 2020
CAUTION
Hazardous areas
The system is not suitable for use in hazardous areas.
2. Install the system in an area where vibrations and electromagnetic and radio frequency interference are minimized or absent.
3. Be sure there is easy access to the transmitter and sensor(s).

2.3 Sample requirements

Be sure the sample meets the following requirements:
1. Temperature: 32 to 122 °F (0 to 50 °C )
2. Pressure: 3 to 65 psig (122 to 549 kPa abs)
3. Minimum flow: 3 gal/hr (11 L/hr)

2.4 Mounting, inlet, and drain connections

The Rosemount™ FCL is intended for wall mounting only.
Refer to Figure 2-1 or Figure 2-2 for details. The sensor(s) screw into the flow cell adapters as shown in the figures. For Rosemount FCL-02 (free chlorine with continuous pH adjustment), you must also install the pH sensor.
8 Emerson.com/Rosemount
Loading...
+ 16 hidden pages