Rosemount Quick Start Guide: Rosemount 848T FOUNDATION™ Fieldbus High Density Temperature Transmitter Manuals & Guides

Quick Start Guide
00825-0100-4697, Rev WB
December 2021
Rosemount™ 848T FOUNDATION™ Fieldbus High Density Temperature Transmitter
Quick Start Guide December 2021
Introduction.................................................................................................................................3
Mount the transmitter..................................................................................................................7
Wire and apply power................................................................................................................ 15
Verify the tag............................................................................................................................. 22
Verify the transmitter configuration...........................................................................................23
Product certifications................................................................................................................. 24
Declaration of conformity.......................................................................................................... 36
China RoHS................................................................................................................................ 38
2 Rosemount 848T
December 2021 Quick Start Guide

1 Introduction

The Rosemount 848T is optimized for process temperature measurement because it can simultaneously measure eight separate and independent temperature points with one transmitter. You can connect multiple temperature sensor types to each transmitter. In addition, the Rosemount 848T can accept 4-20 mA inputs. The enhanced measurement capability of the Rosemount 848T allows it to communicate these variables to any FOUNDATION™ Fieldbus host or configuration tool.

1.1 About this guide

This guide provides basic guidelines for the Rosemount 848T Temperature Transmitter. It does not provide instructions for detailed configuration, diagnostics, maintenance, service, troubleshooting. Refer to the Rosemount 848T Reference Manual for more instruction. The manual and this guide are also available electronically on Emerson.com/Rosemount.

1.2 Hazard messages

This document uses the following criteria for hazard messages based on ANSI standards Z535.6-2011 (R2017).
DANGER
Serious injury or death will occur if a hazardous situation is not avoided.
WARNING
Serious injury or death could occur if a hazardous situation is not avoided.
CAUTION
Minor or moderate injury will or could occur if a hazardous situation is not avoided.
NOTICE
Data loss, property damage, hardware damage, or software damage can occur if a situation is not avoided. There is no credible risk of physical injury.
Quick Start Guide 3
Quick Start Guide December 2021
Physical access
NOTICE
Unauthorized personnel can potentially cause significant damage and/or misconfiguration of end users' equipment. Protect against all intentional or unintentional unauthorized use.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access to protect users' assets. This is true for all systems used within the facility.

1.3 Safety messages

Read this manual before working with the product. For personal and system safety, and for optimum product performance, ensure you thoroughly understand the contents before installing, using, or maintaining this product.
WARNING
Explosions could result in death or serious injury.
Installation of this transmitter in an explosive environment must be in accordance with the appropriate local, national, and international standards, codes, and practices. Review the approvals section of this manual for any restrictions associated with a safe installation.
Before connecting a Field Communicator in an explosive atmosphere, make sure the instruments in the loop are installed in accordance with intrinsically safe or non-incendive field wiring practices.
WARNING
Failure to follow these installation guidelines could result in death or serious injury.
Ensure the transmitter is installed by qualified personnel and in accordance with applicable code of practice.
WARNING
Process leaks could result in death or serious injury.
Do not remove the thermowell while in operation.
Install and tighten thermowells and sensors before applying pressure.
4 Rosemount 848T
December 2021 Quick Start Guide
WARNING
Electrical shock could cause death or serious injury.
If the sensor is installed in a high-voltage environment and a fault or installation error occurs, high voltage may be present on transmitter leads and terminals.
Use extreme caution when making contact with the leads and terminals.
CAUTION
This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:
This device may not cause harmful interference.
This device must accept any interference received, including
interference that may cause undesired operation.
This device must be installed to ensure a minimum antenna separation
distance of 7.9 in. (20 cm) from all persons.
NOTICE
Battery hazards remain when cells are discharged.
The power module may be replaced in a hazardous area. The power module has surface resistivity greater than one gigaohm and must be properly installed in the wireless device enclosure. Care must be taken during transportation to and from the point of installation to prevent electrostatic charge build-up.
Shipping considerations for wireless products.
The unit was shipped to you without the power module installed.
Remove the power module prior to any re-shipping.
Each power module contains two “C” size primary lithium batteries.
Primary lithium batteries are regulated in transportation by the U. S. Department of Transportation, and are also covered by IATA (International Air Transport Association), ICAO (International Civil Aviation Organization), and ARD (European Ground Transportation of Dangerous Goods). It is the responsibility of the shipper to ensure compliance with these or any other local requirements. Consult current regulations and requirements before shipping.
Quick Start Guide 5
Quick Start Guide December 2021

1.4 Product recycling/disposal

Recycling of equipment and packaging should be taken into consideration and disposed of in accordance with local and national legislation/ regulations.
6 Rosemount 848T
December 2021 Quick Start Guide

2 Mount the transmitter

2.1 Mount to a DIN rail without an enclosure

To mount the transmitter to a DIN rail without an enclosure, follow these steps:
Procedure
1. Pull up the DIN rail mounting clip located on the top back side of the transmitter.
2. Hinge the DIN rail into the slots on the bottom of the transmitter.
3. Tilt the transmitter and place onto the DIN rail. Release the mounting clip. The transmitter should be securely fastened to the DIN rail.
Figure 2-1: Mount the transmitter to a DIN rail
A. Rosemount 848T without installed enclosure B. DIN rail mounting clip C. DIN rail
2.2
Quick Start Guide 7

Mount to a panel from an aluminum junction box

Prerequisites
Use four 1/4-20 x 1.25-in. screws.
Procedure
Mount the transmitter to a panel from inside the junction box using one of the following dimension drawings.
10.21 [259]
A
1.70 [43]
B
6.70
[170]
C
D
3.67 [93]
E
.80 [20]
2.30 [58]
2.27 [58]
1.33 [34]
1.88 [48]
3.98 [101]
8.42 [214]
6.00 [152]
Quick Start Guide December 2021
Figure 2-2: Aluminum junction box with cable glands (option code JA4)
A. Nameplate B. Removable wiring connector C. Security switch
D. Simulation switch
E. External ground screw (optional)
Dimensions are in inches (millimeters).
8 Rosemount 848T
A
10.21 [259]
1.70 [43]
B
6.70 [170]
C
D
.80 [20]
2.30 [58]
3.21 [81]
1.33 [34]
1.88 [48]
3.97 [101]
E
7.14 [181]
6.00 [152]
December 2021 Quick Start Guide
Figure 2-3: Aluminum junction box with plugged holes (option code JA5)
A. Nameplate B. Removable wiring connection C. Security switch
D. Simulation switch
E. External ground screw (optional)
Dimensions are in inches (millimeters).
2.3

Mount to a panel from a stainless steel junction box

Prerequisites
Use two 1/4-20 x 1/2-in. screws.
Procedure
Mount the transmitter to a panel from inside the junction box using one of the following dimension drawings.
Quick Start Guide 9
9.91 (231)
7.7 (196)
C
1.8 (46)
1.1 (28)
1.73 (44)
2.4 (62)
1.2 (30)
1.8 (47)
4.0 (102)
9.14 (232.2)
7.72 (196)
6.61 (168)
A B
D E
Quick Start Guide December 2021
Figure 2-4: Stainless steel junction box with cable glands (option code JS2)
A. Top view B. 3-D view C. Ground screw
D. Front view
E. Side view
Dimensions are in inches (millimeters).
10 Rosemount 848T
9.1 (231)
7.7 (196)
C
A
4.0 (102)
B
1.4 (35)
1.1 (27)
2.8 (70)
4.0 (102)
1.8 (4.7)
2.4 (62)
1.6 (42)
F
D
1.2 (30)
E
December 2021 Quick Start Guide
Figure 2-5: Stainless steel junction box with a conduit entry (option code JS3)
A. Top view B. 3-D view C. Ground screw
D. Front view
E. Side view F. Five plugged 0.86 in (21.8 mm) diameter holes suitable for installing ½-
in. NPT fittings
Dimensions are in inches (millimeters).
2.4

Mount to a 2 in (51 mm) pipe stand

Procedure
Use the optional mounting bracket (option code B6) to mount the transmitter to a 2 in (51 mm) pipe stand when using a junction box.
Quick Start Guide 11
A B
5.08
[129]
7.55
[193]
Quick Start Guide December 2021
Figure 2-6: Mount an aluminum junction box
A. Front view B. Side view
Dimensions are in inches (millimeters)
12 Rosemount 848T
Loading...