3414 Gas Ultrasonic Flow Meter3411 and 3412 Gas Ultrasonic Flow Meters
HART® Field Device Specification
00825-0400-3240, Rev AA
April 2022
Rosemount™ 3410 Series Gas
Ultrasonic
Meters
Safety and approval information
This Rosemount product complies with all applicable European directives when properly installed in accordance with the
instructions in this manual. Refer to the EU declaration of conformity for directives that apply to this product. The EU declaration
of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions are available
on the internet at www.emerson.com or through your local Emerson support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at http://
www.emerson.com.
For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the user manual.
Product data sheets and manuals are available from the Emerson website at http://www.emerson.com.
Return policy
Follow Emerson procedures when returning equipment. These procedures ensure legal compliance with government
transportation agencies and help provide a safe working environment for Emerson employees. Emerson will not accept your
returned equipment if you fail to follow Emerson procedures. Return procedures and forms are available on our web support site
at www.emerson.com, or by phoning the Emerson Customer Service department.
Emerson Flow customer service
Email:
• Worldwide: http://flow.support@emerson.com
• Asia-Pacific: http://APflow.support@emerson.com
Telephone:
North and South AmericaEurope and Middle EastAsia Pacific
United States800 522 6277U.K.0870 240 1978Australia800 158 727
HART Field Device Specification 3410 Series Ultrasonic Meter
00825-0400-3240
11
1.Introduction
1. Introduction
April 2022
1.1
1.2
1.3
1.4
Scope
The scope of this document is to define the HART revision 7 functional requirements, device
specific commands and all Universal and the supported Common Practice commands
supported by the HART interface for the Rosemount 3410 Series Gas Ultrasonic Flow Meters.
Purpose
The purpose of this document is to list the functional requirements for developing the 3410
Series Gas Ultrasonic Flow Meters HART rev 7 firmware. This document complies with HART
Protocol Revision 7 and Field Device Specification HCF_LIT-18, Revision 11.0 in preparation and
development of engineering regression tests of functionality not covered by the HCF-Kit-192.
Who should use this document?
This specification is designed to be a technical reference for HART capable Host Application
Developers, System Integrators, and knowledgeable End Users. It also provides functional
specifications (e.g., commands, enumerations and performance requirements) used during
Field Device Development, maintenance and testing.
Definitions, acronyms, abbreviations and
references
The following is a list of commonly used definitions used throughout this document.
Scope 1
1. IntroductionHART Field Device Specification 3410 Series Ultrasonic Meter
April 202200825-0400-3240
1.4.1Definitions
Table 1-1 Definiti o n s
NameDefinition
Emerson Field Communicator375 or 475 field communicator - handheld device
HART Field Device Specification 3410 Series Ultrasonic Meter
00825-0400-3240
1.4.4References
The documents referenced within the text of this document are listed in the table below:
Table 1-4 Reference document a t ion
TitleDocument number, revision, date
1. Introduction
April 2022
American Petroleum Institute (API) Manual of Petroleum Measurement Standards (MPMS) Chapter 21 - Flow Measurement Using
Electronic Metering Systems
Section 1 - Electronic Gas Measurement
HART SMART Communications Protocol Specification (also includes
the specifications listed below)
1. IntroductionHART Field Device Specification 3410 Series Ultrasonic Meter
April 202200825-0400-3240
4References
2. Device IdentificationHART Field Device Specification 3410 Series Ultrasonic Meter
00825-0400-3240April 2022
2
2.Device Identification
This section details the Rosemount 3410 Series Ultrasonic Flow Meter identification and
physical description.
2.1
Rosemount 3410 Series Ultrasonic Flow Meter
identification
The Rosemount 3410 Series Ultrasonic Flow Meter Field Device Identification summary is
shown in Table 2-1 below.
Table 2-1 3410 Series Gas Ultrasonic Flow Meter device identification
Rosemount 3410
Series Ultrasonic
Manufacturer name:
Manufacturer ID Code:13 (D Hex)Device Type Code:21 (15 Hex)
HART Protocol Revision:7Device Revision:6
Number of Device Variables:8
Physical Layers Supported:Bell 202 FSK
Physical Device Category:
Emerson
Rosemount 3410 Series
Ultrasonic Fl
ow Meter
Model name(s):
Note:
HART Hardware Revision:4
HART Software Revision:27
Flow Meter
Rosemount 3410 Series Ultrasonic Flow Meter identification
5
HART Field Device Specification 3410 Series Ultrasonic Meter2. Device Identification
April 202200825-0400-3240
2.2Physical description
The Rosemount 3410 Series Ultrasonic Flow Meter electronics enables the meter to easily
communicate with other field devices, and ultimately, communicate key diagnostic
information through PlantWebTM architecture. Any programmable device, such as the FPGA, is
programmed via the CPU Module. The CPU Module part number is P/N 1-360-03-001).
Figure 2-1 3410 Series Gas Ultrasonic Flow Meter electronics - CPU Module
6Physical description
3. Product overviewHART Field Device Specification 3410 Series Ultrasonic Meter
00825-0400-3240April 2022
3
3.Product overview
This section specifies the purpose and application of the Rosemount 3410 Series Ultrasonic
Flow Meter for HART communication.
3.1
3.2
3.3
3.4
Device function, purpose and features
The Rosemount 3410 Series Ultrasonic Flow Meter communicates with other field devices,
and ultimately, communicates key diagnostic information through the PlantWeb®
architecture.
All analog inputs and outputs are isolated from each other and isolated from the system with a
minimum isolation of 500 V.
Process connections
The Rosemount 3410 Series Ultrasonic Flow Meter’s updated electronics includes the
firmware, CPU Module and the Acquisition module allows communications with other field
devices.
External interfaces (electrical and non-electrical)
Any pressure and/or temperature input read via the Rosemount 3410 Series Ultrasonic Flow
Meter is configurable using a hand-held communicator (e.g., Emerson's 375/475 Field
Communicator) and not via the meter such as for device address, device tag, limits, and units.
The Rosemount 3410 Series Ultrasonic Flow Meter is compliant with Asset Management
Solutions, AMSTM Device Manager, software applications that provides operator interface
between the HART enabled field device and a remote PC.
Other required equipment
A RS-232C/RS-485 (half duplex) serial communication port for Modbus communication is
provided as Port A, Port B and Port C.
Any programmable device aboard the Rosemount 3410 Series Ultrasonic Flow Meter (such as
an FPGA) is programmed via the CPU Module.
Device function, purpose and features 7
HART Field Device Specification 3410 Series Ultrasonic Meter3. Product overview
April 202200825-0400-3240
8Other required equipment
4. Product interfacesHART Field Device Specification 3410 Series Ultrasonic Meter
00825-0400-3240April 2022
4.Product interfaces
This section discusses the Rosemount 3410 Series Ultrasonic Flow Meter for HART
communications, electrical interface, and input and output requirements.
4.1
4.1.1
Process interface
The Rosemount 3410 Series Ultrasonic Flow Meter is capable of communicating with a flow
computer or other interface devices (e.g. (AMSTM Device Manager or an Emerson Field
Communicator, etc.) via HART and enables PlantWeb connectivity.
The Rosemount 3410 Series Ultrasonic Flow Meter provides an RS-232C/RS-485 half-duplex
serial communications port (Port A) connected via the CPU Module. Two independent analog
input circuits and 16-bit, 4-20mA analog output circuits are supported.
LED status indicators show 24V power, 24V current limit, TX and RX serial communication ports,
and HART slave communication via Analog Output 1.
Sensor input channels
Rosemount 3410 Series Ultrasonic Flow Meters supports analog and discrete inputs as shown
below.
Name
Analog Input 14 mA- 20 mA Input for temperature device
Analog Input 24 mA- 20 mA Input for pressure device
Functional description
Note: AI-1 and AI-2 are electronically isolated and operate in sink
mode. The input contains a series resistance so HART Communicators
can be connected to configure sensors.
A 24 Volt DC power supply is available to provide power to the sensors.
Digital Input(s)
(Selectable)
Ethernet PortOne Ethernet Port (TCP/IP)
Process interface 9
(1) Single polarity (for flow calibration gating - contact closure)
•Single input for starting and stopping
•Four pulse configurations available
•Up to 100 Mbps
•Modbus TCP
HART Field Device Specification 3410 Series Ultrasonic Meter4. Product interfaces
April 202200825-0400-3240
Name
Frequency/Digital Output(s)Maximum frequency for the frequency outputs
Functional description
•1000Hz
•5000Hz
Mode options:
•Open Collector (requires external excitation supply voltage and
pull-up resistor)
TTL (internally powered by the meter 0-5 VDC signal)
Serial PortOne serial RS-232/RS-485 port
(115 kbps baud rate) (Modbus RTU/ASCII)
Tr an sd u ce rs
•T-11
•T-12
•T-21
22
•T-
•T-32
•T-41
Note: The process temperature must not exceed the operating tem-
perature range of the transducers.
4.1.2Sensor analog output channels
The firmware supports two independently-configurable analog output channels (AO1 and
AO2). For conventional operation, the analog output channels provide identical but separate
configuration parameters including, but not limited to, the currently available AO1
configuration parameters (for content and scaling configuration). These configuration
parameters follow the same naming convention as the AO1related configuration parameters
(which retain their current names). The Boolean data point,
Analog Output 1 (AO1). NOTE: HART communication is available over AO1 only.
IsAO1Avail indicates the presence of
10Sensor analog output channels
4. Product interfacesHART Field Device Specification 3410 Series Ultrasonic Meter
Configuration via the HART slave port, requires device-specific commands.
00825-0400-3240April 2022
Table 4-1 Analog output configuration parameters
Valu es
Direction
Down greater than-3.125%3.5 mA
Linear over-range
Maximum current+106.25%21 mA
Multi-drop current draw
Lift-off voltage7 V@ full scale
Up less than+106.25%21 mA
(percent of range)
4 mA (Available in sink mode only)
Choose the HART selectable output Primary or Secondary Variable (via any serial, Ethernet, or
HART slave port) from among the following choices:
•uncorrected volumetric flow rate
•corrected volumetric flow rate
•average flow velocity
•average sound velocity
Valu es
(e.g., in mA)
•mass rate
•energy rate
The Rosemount 3410 Series Ultrasonic Flow Meter output Third and Fourth Variables (via any
serial, Ethernet, or HART slave port) from among the choices available for the Primary
Variable and additionally the following choices (if applicable):
•live pressure value
•live temperature value
•for each frequency output: maximum frequency, content, relationship to flow
direction, B channel action upon error, A and B channel phase relationship, and output
scaling
•for each digital output: content, and polarity
•for each analog output (conventional 4-20 mA operation): content, relationship to flow
direction, and output scaling
Trim the analog outputs via HART, serial or Ethernet interfaces.
•The HART slave supports the HART rev. 7 commands listed in Section 8 through
Section 12.
•The HART slave does not support transfer functions.
Sensor analog output channels 11
HART Field Device Specification 3410 Series Ultrasonic Meter4. Product interfaces
April 202200825-0400-3240
Each analog output has individually configurable alarm selections. The selections include:
•Very Low (3.5 mA)
•Low (4.0 mA), High (20 mA)
•Very High (20.5 mA)
•Hold Last Value
•None
Trim the analog outputs via, serial or Ethernet interfaces. The AOXActionUponInvalidContent
data point specifies the action for Analog Output 1 current when the content is invalid. No
special action is taken when set to none.
Each analog output is considered saturated if the "pre-trimmed" value is (strictly) outside the
range (3.5, 20.5) mA. Note that a value less than 4 mA should only occur if the output is invalid
and the invalid content is selected to be represented by a fixed 3.5 mA output. The database
point AOXIsSaturated shall be used to indicate the saturation status.
For each analog output, after the saturation determination is made, then the DAC -limits of (3.5,
21) mA is applied to the pre-trimmed value. The resulting value is written to the appropriate
AOXOutput database point (so that the point's meaning is consistent with the pre-HART
firmware).
The analog output trim zero and gain values (stored in database points AOXCurrentTrimZero
and AOXCurrentTrimGain, respectively) are always applied to the analog output's pre-trimmed,
DAC -limited value (i.e., the value stored in the database point AOXOutput) as shown in
Equation 1 (all values in milliamps except the dimensionless gain). The DAC - limits (3.5, 21 mA)
is applied to the resultant trim value (AOXTRIM) and DAC -limited result shall be stored in the
database point AOXOutputTrimmed and output to the DAC.
Equation 4-1 AO(X)Output Trim
4.1.3
AOX
Time Stamp
= AOXCurrentTrimGain
TRIM
× AOXOut()put – 4()4++AOXCurrentTrimZero
A time stamp shall be calculated for each of the HART device variables when the device variable
is updated and the status is good. The time stamps shall be associated with the corresponding
data poi
nts shown in Table 4-2.
Table 4-2 Device variables time stamp
Device variable Associated time stamp data point name
Uncorrected flow rate (QFlow) HARTQFlowUpdateTime
Corrected flow rate (QBase)HARTQBaseUpdateTime
Average flow velocityHARTAvgFlowUpdateTime
Average sound velocityHARTAvgSndVelUpdateTime
Energy rateHARTEnergyRateUpdateTime
Mass rateHARTMassRateUpdateTime
Pressure (FlowPressure)HARTPressureUpdateTime
Temperature (FlowTemperature)HARTTemperatureUpdateTime
12Time Stamp
4. Product interfacesHART Field Device Specification 3410 Series Ultrasonic Meter
00825-0400-3240April 2022
Time stamps shall be calculated for additional device variables that indicate percent of range
and loop current.
Table 4-3 Time stamp calculations for additional device variables
Device variable Associated time stamp data point name
Percent rangeHARTPercentRangeUpdateTime
Analog output 1HARTAO1OutputUpdateTime
Time stamps shall:
•be unsigned 32 bit integer with least significant bit of time value representing 1/32 of
milliseconds.
•indicate time since midnight and shall rollover after 24 hours
•have units of milliseconds
Time Stamp 13
HART Field Device Specification 3410 Series Ultrasonic Meter4. Product interfaces
April 202200825-0400-3240
14Time Stamp
5. Device variablesHART Field Device Specification 3410 Series Ultrasonic Meter
00825-0400-3240April 2022
5
5.Device variables
The Rosemount 3410 Series Ultrasonic Flow Meter does not use Device Family commands.
5.1
Device variable
Number:
Classification:
Device variable 0 - uncorrected flow rate
The flow-condition volumetric flow rate is the result of applying expansion correction and flowprofile correction to the raw volumetric flow rate derived as shown in Equation C -1 subject to
the low-flow cut-off (see Annex C). If the resulting value is below the low-flow cut-off value, it is set to zero. The low-flow cut-off volumetric flow rate (QCutOff) is the specified low-flow velocity
threshold (ZeroCut) converted to a volumetric flow rate.
0
66
Volumetric Flow
Name
Unit Codes
Uncorrected Flow Rate
(see Table 11-5)
5.2Device variable 1 - corrected flow rate
The base-condition volumetric flow rate (QBase) is the result converting the flow-condition
volumetric flow rate to the base pressure-temperature condition. This conversion requires (1)
AGA8 calculations to be either performed internally (i.e., by the meter) or externally (with the
resulting compressibilities specified to the meter via the SpecZFlow and SpecZBase data points),
and (2) the flow-condition temperature and pressure to be live or fixed. If AGA8 calculations are
not performed (i.e., neither internally nor externally) or the flow-condition temperature and/or
pressure are/is not enabled, then the base-condition volumetric flow rate is set to zero. The
base-condition volumetric flow rate is calculated as shown in Equation C -2 (see Annex C).
Device Variable
Number:1NameCorrected Flow Rate
Classification:66
Volumetric Flow
Device variable 0 - uncorrected flow rate 15
Unit Codes
(
Table 11-5)
Loading...
+ 193 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.