Quick Start Guide
00825-0100-3494, Rev AA
June 2019
Rosemount™ Hx438
Amperometric Steam Sterilizable Dissolved
Oxygen Sensors

Quick Start Guide June 2019
Safety Information
WARNING
High pressure and temperature
Failure to reduce the pressure and temperature may cause serious injury to personnel.
Before removing the sensor, reduce the process pressure to 0 psig and cool down the process
temperature.
Do not exceed temperature and pressure limitations of 266 °F (130 °C) and 43 psig (400 kPa, 4
bar).
WARNING
The electrolyte is highly alkaline.
Protect your hands with gloves and use safety goggles.
Avoid contact of the electrolyte with the skin, eyes, and mucous membranes.
2 Emerson.com

June 2019 Quick Start Guide
1 Plan
1.1 Unpack and inspect
Procedure
1. Inspect the outside of the carton for any damage.
2. If damage is detected, contact the carrier immediately.
3. Inspect the hardware.
4. Make sure all the items in the packing list are present and in good
condition.
5. Notify the factory if any part is missing.
1.2 Specifications
Table 1-1: Rosemount Hx438 dissolved oxygen sensor specifications
Hx438 Dissolved Oxygen Sensor Material and Units
Operating Temperature Range 32 to 266 °F (0 to 130 °C)
Maximum Pressure 43 psig (400 kPa abs, 4 bar)
Measurement Range 0 to 20 ppm or 0 to 250% saturation,
Wetted Materials Stainless steel and EPDM
Process Connections PG 13.5 thread
Cable Connector Standard 4-pin
Cable Compatibility Standard 4-pin connector cable
Compatible Mounting Accessory Insertion Mounting Assembly
Compatible Transmitters Rosemount Transmitter Models 56,
Temperature Compensation 22K NTC
depending on instrument
1056, 5081, and 1066
Quick Start Guide 3

Quick Start Guide June 2019
2 Install
2.1 Preparing the sensor
Procedure
1. Remove the plastic cap carefully which covers the sensing end of
sensor.
2. Avoid causing a vacuum by gently and slowly twisting the sensor and
cap until it is removed.
3. Sensor is shipped dry, so electrolyte must be added prior to
installation.
4. Remove membrane cartridge.
5. Add 1.5 ml of electrolyte to the membrane cartridge.
6. Screw membrane cartridge back onto sensor body.
2.2 Connecting the sensor to the trasmitter
Procedure
1. Wire the sensor to the transmitter.
2. Apply power to the transmitter.
3. Make the sensor ready for calibration after two hours.
The polarization time is necessary to get stable signals from the
sensors
4. If the sensor is disconnected from the transmitter for a short time,
allow the sensor to stabilize for at least two times longer than the
time was disconnected.
Time needed for stabilization should not exceed two hours.
4 Emerson.com

June 2019 Quick Start Guide
2.3 Mount the sensor
Procedure
Use PG 13.5 process thread to mount the sensor.
NOTICE
Do not install the sensor upside down.
Quick Start Guide 5

Quick Start Guide June 2019
Figure 2-1: Dimensional Drawing for Rosemount Hx438-01
A.
Sensor part number
Hx438-01 4.725 120
B. 4-pin connector
C. Free rotating PC 13.5 screw thread
D. O-ring
2.4
6 Emerson.com
Wire the sensor to the transmitter
Procedure
Wire the sensor to the transmitter.
See wiring diagrams below.
A dimension
in. mm