Rosemount Quick Start Guide: Configure a File Transfer Protocol (FTP) Data Port Range With Rosemount MON2020 Manuals & Guides

Quick Start Guide
ECO-5008259
00825-0300-2020, Rev AA
March 2022
Configure a File Transfer Protocol (FTP) Data Port Range
With Rosemount MON2020
Quick Start Guide March 2022
port range
When gas chromatographs (GCs) are installed behind a router through port forwarding, it can be useful to configure an FTP data port range. With this feature, GCs will be differentiated by their ports, but will have the same Internet protocol (IP) address.
Note
This feature has been added to Rosemount 370XA, 700XA, and 1500XA GCs with firmware version 4.1.1 or newer.
If using port forwarding (for example, if the GC is connected behind a cellular router), set the FTP data port range when Rosemount is connected locally to a GC with a DHCP connection.
This feature will only be enabled if:
Rosemount is connected to a GC.
The user logged into Rosemount has admin rights.
The Security switch is unlocked.
The Designate Port option is selected.
2 Emerson.com/Rosemount
March 2022 Quick Start Guide
Prerequisites
Make sure the Designate Port option is selected in the Ethernet 1 Connection Properties for a DHCP dialog.
Figure 1-1: Ethernet 1 Connection Properties for a DHCP dialog
Quick Start Guide 3