Rosemount™ 402 and 402VP
Contacting Conductivity Sensors
Quick Start Guide
0825-0100-3402
0
Rev. AA
May 2019

Essential Instructions
Read this page before proceeding!
merson designs, manufactures and tests its products to meet many national and international stan-
E
dards. Because these sensors are sophisticated technical products, you MUST properly install, use,
and maintain them to ensure they continue to operate within their normal specifications. The
following instructions MUST be adhered to and integrated into your safety program when installing,
using, and maintaining Rosemount products. Failure to follow the proper instructions may cause any
one of the following situations to occur: loss of life; personal injury; property damage; damage to this
sensor; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson representative for
clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance
of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate
Quick Start Guide and per applicable local and national codes. Connect all products to
the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update,
program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's
performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY.
Third-party substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and
personal injury.
The information contained in this document is subject to change without notice.
CAUTION
Sensor/Process Application Compatibility
The wetted sensor materials may not be compatible with process composition and operating
conditions. Application compatibility is entirely the responsibility of the user.
CAUTION
The 402 and 402VP sensors are retractable. BEFORE RETRACTING THE SENSOR, be absolutely certain the
process pressure is less than 64 psig (542 kPa abs) and the process temperature is at a safe level!

Quick Start Guide Table of Contents
00825-0100-3402 May 2019
Contents
Section 1: Specifications
1.1 Specifications ......................................................................................................1
Section 2: Installation
2.1 Unpacking and Inspection ...................................................................................3
2.2 Sensor Installation ...............................................................................................3
2.3 Installation and Retraction of Sensor....................................................................4
Section 3: Wiring
3.1 Wiring for Rosemount 402/402VP .......................................................................7
Section 4: Calibration and Maintenance
4.1 Cleaning the Sensor...........................................................................................11
4.2 Calibrating the Sensor........................................................................................11
Section 5: Troubleshooting
5.1 Troubleshooting ................................................................................................13
Section 6: Accessories
6.1 Accessories........................................................................................................15
Table of Contents i

Table of Contents Quick Start Guide
May 2019 00825-0100-3402
ii Table of Contents

Quick Start Guide Specifications
00825-0100-3402 May 2019
Section 1: Specifications
1.1 Specifications
Table 1-1: Rosemount 402/402VP contacting conductivity sensor specifications
Wetted Materials
Electrodes Titanium
Insulator Glass Filled PEEK
Sensor tube 316 Stainless Steel
O-ring EPDM
Washer Neoprene
Temperature range
Standard 32 to 212 °F (0 to 100 °C)
Maximum pressure
200 psig (1481 kPa abs)
Maximum retraction pressure
64 psig (542 kPa abs)
Vacuum
At 1.6 in. Hg (5.2 kPa) air leakage is less than 0.005 SCFM (0.00014 m3/min)
Cell constants
0.01, 0.1, and 1.0/cm
Process connection
Sensor is inserted through 1¼ in. NPT full port ball valve
Cable
10 ft (3.1 m) standard; for longer cable lengths, choose option -60 (integral junction box) and order interconnecting cable
separately; 402VP Interconnecting VP6 cables sold separately. (See
Accessories).
Table 1-2: Weights and shipping weights
Sensor model With 10 ft. (3.1 m) cable With junction box
Rosemount 402 Contacting Conductivity Sensor 3 lb. (1.5 kg) 4 lb. (2.0 kg) 4 lb. (2.0 kg) 5 lb. (2.5 kg)
Rosemount 402VP Contacting Conductivity Sensor 3 lb. (1.5 kg) 4 lb. (2.0 kg)
1. Rounded up to the nearest 1 lb or 0.5 kg
Specifications 1
(1)
Weight Shipping
weight
Weight Shipping
weight

Specifications Quick Start Guide
May 2019 00825-0100-3402
Table 1-3: Retraction Assembly Specifications
Assembly includes
Ball valve, retraction body, and pipe nipple.
Wetted materials
all valve
B
ipple
N
acking rings
P
Packing bushing 303 Stainless Steel
Retraction body 316 Stainless Steel
Process connection
Ball valve 1¼ in. FNPT; Nipple 1¼ in. MNPT
Temperature
32 to 212 °F (0 to 100 °C)
Pressure
200 psig (1481 kPa abs) maximum
Maximum retraction pressure
64 psig (542 kPa abs) maximum
Vacuum
At 1.6 in. Hg (5.2 kPa) air leakage is less than 0.005 SCFM (0.00014 m3/min)
2 to 212 °F (0 to 100 °C)
3
16 Stainless Steel
3
raphite
G
2 Specifications