Hazardous Area OCX 8800 Oxygen/Combustibles Transmitter
with Remote Electronics and Local Operator Interface
http://www.raihome.com
HIGHLIGHTS OF CHANGES
Effective January, 2007 Rev 1.0
PageSummary
Page 2-19Revised Figure 2-11.
Page 3-2Revised the text for SW1 and SW2 to read internally powered.
Page 3-3Revised Figure 3-1. Updated the paragraph Initial Power Up.
Page A-2 thru A-23Added note 11 to the safety data section. Added additional pages/languages.
Page C-1Updated the return of materials address.
Back coverUpdated the address blocks.
Emerson Process Management designs, manufactures and tests its prod uct s
to meet many national and international standards. Because these
instruments are sophisticated technical prod ucts, you MUST properly inst all, use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to
integrated into your safety program when installing, using, and maintaining
Emerson’s Rosemount Analytical products. Failure to follow the proper
instructions may cause any one of the following situations to occur: Loss of
life; personal injury; property damage; damage to this instrument; and
warranty invalidation.
• Read all instructions
product.
• If you do not understand any of the instructions, contact your Emerson Process Management representative for clarification.
• Follow all warnings, cautions, and instructions
supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions
of the appropriate Instruction Manual and pe r a pplicable loca l and
national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel
operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use
replacement parts specified by Emerson Process Management.
Unauthorized parts and procedures can affect the product's
performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire,
electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers
are in place, except when maintenance is being performed by
qualified persons, to prevent electrical shock and personal injury.
prior to installing, operating, and servicing the
marked on and
to install,
and
http://www.raihome.com
The information contained in this document is subject to change without
notice.
If a Model 275/375 Universal HART® Communicator is used with this unit, the software
within the Model 275/375 may require modification. If a software modification is required,
please contact your local Emerson Process Management Service Group or National
Response Center at 1-800-654-7768.
Instruction Manual
N
IM-106-880, Rev 1.0
OCX 8800
January 2007
PREFACEThe purpose of this manual is to provide a comprehensive understanding of
the OCX 8800 components, functions, installation , and maintenance.
We recommend that you thoroughly familiarize yourself with the Introduction
and Installation sections before installing your transmitter.
The introduction presents the basic principles of the transmitter along with its
performance characteristics and components. The remaining sections contain
detailed procedures and information necessary to install and service the
transmitter.
Before contacting Emerson Process Management concerning any questions,
first consult this manual. It describes most situations encountered in your
equipment's operation and details necessary action.
DEFINITIONSThe following definitions apply to WARNINGS, CAUTIONS, and NOTES
found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
SYMBOLS
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in damage to or destruction of equipment, or loss of
effectiveness.
NOTE
Highlights an essential operating procedure, condition, or statement.
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETI
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a
manual illustration number. It is not a part number, and is not related to the
illustration in any technical manner.
A typical OCX 8800 Oxygen/Combustibles Transmitter package should
contain the items shown in Figure 1-1.
Use the product matrix in Table 1-1 at the end of this section to verify your
order number. The first part of the matrix defines the model. The last part
defines the various options and features of the OCX 8800. Check the model
number against the transmitter features and options, making sure options
specified by this number are on or included with the unit. Use this complete
model number for any correspondence with Emerson Process Management.
A list of accessories for use with the OCX 8800 is provided in Table 1-2.
SYSTEM OVERVIEWScope
This Instruction Manual supplies details needed to install, startup, operate,
and maintain the OCX 8800. Signal conditioning electr on ics o utputs separate
4-20 mA signals representing oxygen (O
This information, plus additional details, can be accessed with the HART
Model 275/375 handheld communicator or Rosemount Analytical AMS
software. The local operator interface (LOI) also provides a communications
interface with the electronics.
System Description
The OCX 8800 is designed to measure oxygen and combustible
concentrations in flue gas temperatures up to 2600°F (1427°C). Electrical
connections, power and communications are made through two 3/4 NP T ports
in the flameproof electronics enclosure using fittings and cables provided by
the customer. Cable installation must meet NEC, IEC and/or other applicable
national or local codes for Class I, Zone 1, Group IIB +H2 T3/T6 permanently
mounted equipment. The transmitter is close coupled to the process and
requires minimal sample conditioning requirements.
) and combustibles (COe) values.
2
http://www..raihome.com
OCX 8800
Figure 1-1. Typical System Package
4
MAN4275A00
English
October 1994
HART Communicator
o
FISHER-ROSEMOUNT
Instruction Manual
IM-106-880, Rev 1.0
January 2007
1
A
n
a
l
y
t
i
c
a
l
TM
6
3
1. Instruction Manual
®
2. HART
Communicator Package (optional)
3. Adapter Plate with Mounting Hardware and Gasket
4. Reference Air and Calibration Set
5. Blowback Hardware (optional)
6. OCX 8800 with Remote Electronics
2
5
37390076
1-2
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
The equipment measures oxygen percentage by reading the voltage
developed across a heated electrochemical cell, which consists of a small
yttria-stabilized, zirconia disc. Both sides of the disc are coated with porous
metal electrodes. When operated at the proper temperature, the millivolt
output of the cell is given by the following Nernst equation:
EMF = KT log10 (P
Where:
1. P
is the partial pressure of the oxygen in the mea sured gas on one side
2
of the cell.
is the partial pressure of the oxygen in the reference air on the
2. P
1
opposite side of the cell.
3. T is the absolute temperature.
4. C is the cell constant.
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry instrument air (20.95% oxygen) as the
reference air.
When the cell is at operating temperature and there are unequal oxygen
concentrations across the cell, oxygen ions will travel from the high oxygen
partial pressure side to the low oxygen partial pressure side of the cell. The
resulting logarithmic output voltage is approximately 50 mV per decade. The
output is proportional to the inverse logarithm of the oxygen concentration.
Therefore, the output signal increases as the oxygen concentration of the
sample gas decreases. This characteristic enables the OCX 8800 to provide
exceptional sensitivity at low oxygen concentrations.
1/P2
) + C
The OCX 8800 measures net oxygen concentration in the presence of all the
products of combustion, including water vapor. Therefore, it may be
considered an analysis on a "wet" basis. In comparison with older metho ds,
such as the portable apparatus, which provides an analysis on a "dry" gas
basis, the "wet" analysis will, in general, indicate a lower percentage of
oxygen. The difference will be proportional to the water content of the
sampled gas stream.
The OCX 8800 combustibles sensor is a catalytic sensor consisting of two
Resistance Devices (RTD). One RTD is the reference element covered with
an inert coating. The other RTD element is active, coated with a catalyst. As
the sample gases flow by the sensor, the combustible gases oxidize on the
surface of the active element. The oxidation that occurs produces heat and a
temperature rise in the active element. The temperature differ ence produces
a resistance relationship between the two elements that is directly
proportional to the concentration of combustibles in the sample gases.
1-3
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
The catalyst is specifically designed to detect carbon monoxide (CO), but the
sensor responds to other combustible gases. The sensor is calibrated using
CO, thus the output should be expressed in terms of CO. However, since the
sensor detects other combustible gases, the output cannot just be labeled
CO. The response of the sensor to other combustible gases gives an output
that is equivalent to the sensor detecting CO. The term COe is used in this
manual to describe the sensor output. This term indicates that the sensor is
calibrated in terms of CO, and that the sensor output is equivalent to CO but
not specific to CO.
Dilution air is provided to the COe sensor to ensure there is adequate oxygen
to fully oxidize any combustible gases regardless of the concentration of
oxygen in the process.
System Configuration
Transmitters are available in four lengths, giving the user the flexibility to use
a penetration appropriate to the size of the stack or duct. The length optio ns
are 18 in. (457 mm), 3 ft (0.91 m), 6 ft (1.83 m), or 9 ft (2.7 m). Probes are
available in three material options, 316L stainless steel, inconel 600, and
ceramic to accommodate higher temperatures.
The electronics are contained in a separate housing from the sensors. The
electronics housing may be mounted up to 150 feet away from the sensor
housing.
The electronics control both sensor temperatures and provide individual 4-20
mA isolated outputs that are proportional to the measured o xygen and
combustibles concentrations. The power supply can accept voltages of 100 to
240 V AC and 50 to 60 Hz. The electronics accepts millivolt signals generated
by the sensors and produces the outputs to be used by remotely connected
devices. The outputs are isolated 4-20 mA linearize d currents. Refer to
Section 3, Configuration and Startu p for specific instructions upon initial
power up.
System Features
1. The O
concentration decreases.
2. HART communication is standard. To use the HART capability, you
must have either:
a. HART Model 275/375 Communicator.
b. AMS software for the PC.
3. Oxygen cell and heater/thermocouple assembly are field replaceable.
4. Electronics are automatically configured for line voltages from 100 to
240 VAC.
cell output voltage and sensitivity increase as the oxygen
2
1-4
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
5. An operator can calibrate and diagnostically troubleshoot the OCX 8800
in one of two ways:
a. LOI. The LOI is mounted to the end of the electronics module and
allows local communications with the electronics. Refer to Section 4,
Using the LOI, for more information.
b. Optional HART Interface. Each of the OCX 8800's 4-20 mA output
lines transmit an analog signal proportional to oxygen or combustible
levels detected. The HART output is superimposed on the oxy gen
4-20 mA output line only. This information can be accessed through
the following:
•Model 275/375 Handheld Communicator - The handheld
communicator requires Device Description (DD) software specific
to the OCX 8800. The DD software will be supplied with many
Model 275/375 units, but can also be programmed into existing
units at most Emerson Process Management service offices.
Refer to Section 5, Using HART Communications, for additional
information.
•Personal Computer (PC) - The use of a personal computer
requires AMS software available from Emerson Process
Management.
•Selected Distributed Control Systems - The use of distributed
control systems requires input/output (I/O) hardware and AMS
software which permit HART communications.
6. Optional Blowback System. The blowba ck sys tem pe rio dic ally blows
instrument air back through the sample line filter and out the sample
tube. This clears out particulate and keeps the sample line filter from
clogging.
System Operation
Figure 1-2 shows the relationship between the components of the OCX 8800.
The sensors and the electronics are contained in separate housings. The
sensor housing and probe mounts to a duct or process wall so that the prob e
protrudes into the flue gas stream. An air powered eductor continuously pulls
samples of the process flue gas through the probe to a chamber in front of the
sensor housing where the sample passes the O
the COe sensor. Dilution air is prov ided to the COe sensor and reference air
to the O
the COe sensor, it is drawn through the eductor where it mixes with the
eductor air and exits through exhaust back into the system. The electronics
housing contains the CPU and HART boar ds which con vert the se nsor inpu ts
into 4-20 mA analog output signals. The CPU can also initiate and perform
calibrations. Three test gasses and instrument air can be turned on an d off by
solenoids. Test gas flow to the sensors is regulated by a flow meter between
the electronics and sensor housings. Instrument air is separated into eductor
air, reference air, and dilution air. The instrument air solenoid does not allow
air flow until the heaters are up to temperature. This minimizes the amount of
sampled process flue gas being pulled into cold sensors causing
condensation.
sensor. After the gas sample flows past the O2 sensor and through
2
sensor and continues on to
2
1-5
OCX 8800
(
)
Figure 1-2. System Operation Diagram
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Sample
Gas
Probe
Exhaust
SENSOR
HOUSING
COe
Combustibles
Sensor
O
2
Sensor
Eductor
Flow Meter
Reference Air
Dilution Air
7 scfh
ELECTRONICS
Instrument Air
Eductor Air
Flow Meter
50 cc/min.
0.1 scfh
HOUSING
CPU
HART
Board
Power
Supply
Optional
Test Gas
Solenoids
Solenoid
Low O
Test Gas
High O
Test Gas
CO
Test Gas
Instrument
Air
2
2
37390001
Handling the OCX 8800
It is important that printed circuit boards and integrated circuits are handled only when
adequate antistatic precautions have been taken to prevent possible equipment damage.
The OCX 8800 is designed for industrial application. Treat each component of the system
with care to avoid physical damage. The probe may contain components made from
ceramics, which are susceptible to shock when mishandled.
System Considerations
Prior to installing your OCX 8800, make sure you have all the components
necessary to make the system installation. Ensure all the components are
properly integrated to make the system functional.
1-6
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
After verifying that you have all the component s, select mounting locations
and determine how each component will be placed in terms of available line
voltage, ambient temperatures, environmental considerations, convenience,
and serviceability. Figure 1-3 shows a typical system wiring. Simplified
installations for the OCX 8800 are shown in Figure 1-4.
A source of instrument air is required at the OCX 8800 for reference air,
dilution air, and eductor air. Since the OCX 8800 is equipped with an in-place
calibration feature, provision should be made for connecting test gas tanks to
the OCX 8800 when it is to be calibrated.
NOTE
The electronics module is designed to meet NEMA 4 (IP66) and the electronic
components are rated to temperature s up to 185°F (85°C).
Retain packaging in which the unit arrived from the factory in case any
components are to be shipped to another site. This packaging has been
designed to protect the product.
Figure 1-3. OCX 8800 HART
Connections and AMS Application
OCX 8800
Sensor Housing
Instrument
Air
4-20 mA Output
OCX 8800
Electronics Housing
3 calibration
gas lines by
customer
[300 ft (91 m) max.)
(Twisted Pairs)
Termination in
Control Room
HART
Model 275/375
Handheld
Communicator
Customer’s Laptop
with AMS
AMS
37390065
1-7
OCX 8800
Figure 1-4. Typical System
Installation
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Heater
Power Cable
[up to 150 ft (46 m)]
Signal Cable
[up to 150 ft (46 m)]
Gases
Stack
Dilution
Air
Flow
Meter
Duct
OCX 8800 with
REMOTE
ELECTRONICS
Test Gas
Flow Meter
Pressure
Regulator
High O Test Gas
Low O Test Gas
COTest Gas
4-20 mA Outputs
(2 Twisted Pairs)
Line Voltage
Instrument Air
Supply
(Reference Gas)
2
2
37390064
1-8
Instruction Manual
IM-106-880, Rev 1.0
January 2007
SPECIFICATIONSHazardous Area OCX
Specifications
Range 0-1% to 0-40% O2, fully field selectable
Net O
2
Combustibles 0-1000 ppm to 0-5%, fully field selectable
Accuracy
Oxygen± 0.75% of reading or 0.05% O2 (whichever is greater)
Combustibles± 2% range
System Response to
Test Gas
Oxygen10 sec T90
Combustibles25 sec T90
Temperature Limits
Process32° to 2600°F (0° to 1427°C)
Sensors Housing-40° to 212°F (-40° to 100°C), ambient
Electronics Housing-40° to 149°F (-40° to 65°C), ambient
Blowback Air (optional)Clean, dry, instrument-quality air (20.95% O
OCX 8800
-40° to 185°F (-40° to 85°C), internal - operating temperature of
electronics inside instrument housing, as measured by a HART
communicator or AMS software
-40° to 158°F (-40° to 70°C), ambient
[At temperatures above 158°F (70°C) inside instrument housing,
the infrared keypad will cease to function, but the OCX 8800 will
continue to operate properly.]
Sensors HousingNEMA 4, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports
Electronics HousingNEMA 4, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports
Electrical NoiseEN 61326-1, Class A
CertificationsSensor Housing
0344II 2 G
KEMA 04 ATEX 2308
EEx d IIB+H2 T3
CUS
2005.1602514
CLASS 1, ZONE 1
Ex d IIB+H2 T3
AEx D IIB+H2 T3
TYPE 4 / IP66
APPROVED
CLASS 1, ZONE 1
AEx D IIB+H2 T3
TYPE 4 / IP66
Electronics Housing
0344II 2 G
KEMA 04 ATEX 2308
EEx d IIB+H2 T6 IP66
CUS
2005.1602514
CLASS 1, ZONE 1
Ex d IIB+H2 T6
AEx D IIB+H2 T6
TYPE 4 / IP66
APPROVED
CLASS 1, ZONE 1
AEx D IIB+H2 T3
TYPE 4 / IP66
Line VoltageUniversal 100 to 240 VAC ±10%, 50 to 60 Hz, no switches or
jumpers required, 3/4-14 NPT conduit port
Pollution Degree2
Over Voltage CategoryII
Relative Humidity5 to 95% (non-condensing)
Isolated Output
Oxygen4-20 mAdc, 950 ohm maximum, with HART capability
Combustibles 4-20 mAdc, 950 ohm maximum
AlarmAlarm output relay - dry contact, form C, 30 mA, 30VDC capacity
Power Consumption750 W maximum
NOTE
All static performance characteristics are with operati ng variables const ant. Specific ations subject t o change
without notice.
1-10
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Table 1-1. Prod uct Matrix - Hazardous Area OCX 8800
OCX88CO2/Combustibles Transmitter - Flameproof
Code Probe Length and Material
00No Probe or Exhaust Tube
1118 in. (457 mm) 316 SST up to 1300°F (704°C)
2118 in. (457 mm) Inconel 600up to 2600°F (1427°C)
3118 in. (457 mm) Ceramic up to 2600°F (1427°C)
123 ft (0.91 m) 316 SST up to 1300°F (704°C)
223 ft (0.91 m) Inconel 600up to 1832°F (1000°C)
323 ft (0.91 m) Ceramic up to 2600°F (1427°C)
136 ft (1.83 m) 316 SST up to 1300°F (704°C)
236 ft (1.83 m) Inconel 600 up to 1832°F (1000°C)
149 ft (2.7 m) 316 SST up to 1300°F (704°C)
249 ft (2.7 m) Inconel 600up to 1832°F (1000°C)
Code Probe Mounting Assembly
10(ANSI 2 in. 150 lb) 6" dia. flange, 4.75" BC with 4 x 0.75" dia. holes
20(DIN) 185 mm dia. flange, 145 mm BC with 4 x 18 mm dia. holes
Code Mounting Hardware - Stack Side
0No Adapter Plate (“0” must be chosen under “Mounting Adapter - Probe Side” below)
1New Installation - Square weld plate with studs
2Model 218/240 Mounting Plate (with Model 218/240 Shield Removed)
3Existing Model 218/240 Support Shield
4Special Mounting
5Model 132 Adapter Plate
Code Mounting Hardware - Probe Side
0No Adapter Plate
1Probe Only (ANSI)
2Probe Only (DIN)
Code Electronics Housing - Communications
OCX88C111011H20202Example
NOTES:
(1)
Provide details of the existing mounting plate as follows:
Plate with studsBolt circle diameter, number, and arrangement of studs, stud thread, stud height above mounting plate.
Plate without studsBolt circle diameter, number, and arrangement of holes, thread, depth of stud mounting plate with accessories.
(2)
If the LOI is not implemented, remote access and functionality available via HART Communications (Model 275/375 Handheld Communicator)
with Oxygen/Combustibles Device Description (DD) required.
(1)
H1HART Communications
H2HART Communications with LOI
H3HART Communications with Calibration Solenoids
H4HART Communications with LOI and Calibration Solenoids
Code Electronics Mounting
02Remote Electronics and no cable
Code Accessories
00None
01Flow meters & Ref. Air Set
02In-Situ Filter (Stainless Steel only)
03In-Situ Filter (SST), Flow meters & Ref. Air Set
11Flow meters, and Ref. Air Set with Blowback
12In-Situ Filter (SST) with Blowback
13In-Situ Filter (SST), Flow meters & Ref. Air Set with Blowback
(2)
(2)
OCX 8800
1-11
OCX 8800
Table 1-2. Accessories
Instruction Manual
PART NUMBERDESCRIPTION
1A99119H01Oxygen test gas bottle; 0.4% O
1A99119H02Oxygen test gas bottle; 8.0% O2, balance N
1A 99119H07CO test gas bottle; 1000 ppm CO, balance air
1A99120H02Regulator for Oxygen (may need 2)
1A99120H03Regulator for CO test gas
1A99119G06Wall mount bracket for test gas bottles
1A99119G05Test gas regulators kit
1A99119G04Test gas bottles kit
1A99292H01
1A99339H03Blowback valve, air operated
4851B40G01Wall or Pipe Mounting Kit
1A99784H02
6A00171G01Power line filter kit
Moore Industries SPA for Low O
Calibration Status, and Unit Fail
375 HART Communicator with 12 Megabyte buffer,
model no. 375HR1EKLU
Before installing this equipment, read the "Safety instructions for the wiring and installation
of this apparatus" in Appendix A: Safety Data. Failure to follow the safety instructions could
result in serious injury or death.
OCX 8800
To maintain explosion-proof protection of the OCX 8800 in hazardous areas, all cable entry
devices and blanking elements for unused apertures must be certified flameproof, suitable
for the conditions of use and properly installed.
To maintain explosion-proof protection of the OCX88C in hazardous areas, the sensor
housing must not be mounted to any surface or flange that exceeds 200ºC (392ºF).
To maintain explosion-proof protection of the OCX88C in hazardous areas, the sample
entering the sensor housing must not exceed 200ºC (392ºF).
http://www..raihome.com
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
MECHANICAL
INSTALLATION
Selecting Location
1. The location of the OCX 8800 in the stack or flue is most important for
maximum accuracy in the oxygen analyzing process. The probe must
be positioned so the gas it measures is representative of the process.
Best results are normally obtained if the transmitter is positioned near
the center of the duct (40-60% insertion). Longer ducts may require
several transmitters since the oxygen and combustibles can vary due to
stratification. A point too near the wall of the duct or the inside radius of
a bend, may not provide a representative sample because of the very
low flow conditions. The sensing point should be selected so the
process gas temperature falls within the range of probe material used.
Figure 2-1 through Figure 2-4 provide mechanical installation
references. The ambient temperature inside the electronics housing
must not exceed 185°F (85°C).
2. Check the flue or stack for holes and air leakage. The presence of this
condition will substantially affect the accuracy of the oxygen and
combustibles readings. Therefore, either make the necessary repairs or
install the transmitter up stream of any leakage.
3. Ensure the area is clear of internal and external obstructions that will
interfere with installation and maintenan ce access to the unit. Allow
adequate clearance for the removal of the OCX 8800.
Do not allow the temperature of the electronics housing to exceed 185°F (85°C) or damage
to the electronics may result.
Whenever a positive stack pressure exists at the installation site, be sure to connect all
pneumatic lines prior to installing the OCX 8800 in the stack or ductwork. Failure to connect
the pneumatic lines can allow the flow of contaminants into the OCX 8800 ports.
Before installing the OCX 8800 into a hot stack or ductwork, make sure that the OCX 8800
is turned on and at normal operating temperature. Exposing a cold OCX transmitter to hot
process gases can cause permanent damage to the equipment.
Installation
1. Ensure all components are available to install the OCX 8800.
2. The OCX 8800 may be installed intact as it is received.
3. Weld or bolt adapter plate (Figure 2-2) onto the duct.
2-2
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
4. Use the pipe or wall mounting hardware as shown in Figure 2-3 to
mount the electronics housing. Choose a location not to exceed the
length of the electronics cable ordered.
5. Ensure the conduits drop vertically from the OCX 8800 and the conduit
is routed below the level of the conduit ports on the housing to form a
drip loop. Drip loops minimize the possibility that moisture will damage
the electronics (Figure 2-4).
6. Where a positive stack pressure exists at the inst allation site, connect all
pneumatic lines prior to installing the OCX 8800 in the stack or
ductwork.
NOTE
If process temperatures will exceed 392°F (200 °C), use anti- s eize compou nd
on stud threads to ease future removal of the OCX 8800.
7. Insert sample and exhaust tubes through the opening in the mounting
flange and bolt the unit to the flange.
Uninsulated stacks or ducts may cause ambient temperatures in the electronics housing to
exceed 185°F (85°C) and damage the electronics.
8. If insulation is removed to access the duct for OCX 8800 mountin g ,
make sure to replace insulation afterward.
Enclosures
The OCX 8800 enclosures are designed to meet ingress conditions of IP66.
Each enclosure cover is threaded to its base and sealed with an o-ring that
isolates the threads from external contaminants.
Each cover is secured by a clip attached to the base that engages the cover
between the ribs of the cover sidewall. The clip is held in place by an Allen
head cap screw and lockwasher mounted in a recess. Cover removal and
installation requires an Allen wrench to loosen and tighten the screw.
2-3
OCX 8800
Figure 2-1. Installation, OCX 8800
NOTE
All dimensions are in inches with millimeters in parentheses.
Insulate if exposed to adverse weather or extreme temperature
changes, install a protective housing and/or insulation
around the unit.
SeeTable 1
Flange Dia.
B.C. Dia.
Hole Dia.
Table 1. Mounting Flange
ANSI
Flange
Dia.
Hole
Dia.
(4) Holes
equally
spaced on
B.C.dia
6.00
(152)
0.75
(19)
4.75
(121)
Instruction Manual
IM-106-880, Rev 1.0
January 2007
DIN
5R10244H025R10244H01
7.28
(185)
0.71
(18)
5.71
(145)
Removal Envelope
Dim “B”
SeeTable 2
Insertion Depth
Dim “A”
SENSOR HOUSING
Optional
In Situ FIilter
Table 2. Installation/Removal
Probe
18 in.
3 ft
6 ft
Signal Cable
Heater Power Cable
9 ft
Dim "A”
18
(457)
36
(914)
72
(1829)
108
(2743)
Dim "B”
34
(864)
52
(1321)
88
(2235)
124
(3150)
Allow 9 in.
(229 mm) for
Cover Removal
0.06 In. Thick Gasket
3535B18H02
ANSI
3535B45H01
Signal Cable
DIN
Heater Power Cable
8.3
(211)
2-4
ELECTRONICS HOUSING
37390009
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 2-2. Adapter Plate Installation
OCX 8800
2-5
OCX 8800
Figure 2-3. Wall or Pipe Mounting
of Electronics Housing
Note:
* Indicates items are included
in mounting kit P/N 4851B40G01.
*Pipe
Mounting
Clamp
*Wall or Pipe
Mounting Bracket
and Screws
Wall Mounting Bolts
or Lag Screws
(4 each by customer)
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Electronics
Housing
1/4-20 UNC x 2 in. M-F Standoffs.
Recommended for rear piping
clearance when wall mounting.
(4 each by customer)
37020021
2-6
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 2-4. Installation
with Drip Loops
OCX 8800
Conduit Drip Loops
Conduit Drip LoopConduit Drip Loop
DuctWall
DuctWall
37020004
2-7
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
ELECTRICAL
INSTALLATION
All wiring must conform to local and national codes. For reference, factory
wired solenoid power connections are shown in Figure 2-5.
Disconnect and lock out power before connecting the unit to the power supply. Failure to
lock out power could result in serious injury or death.
Install all protective equipment covers and safety ground leads after installation. Failure to
install covers and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation
of this equipment, connection to the main electrical power supply must be made through a
circuit breaker (min 10 A) in close proximity and marked for this equipment which will
disconnect all current-carrying conductors during a fault situation. This circuit breaker
should also include a mechanically operated isolating switch. If not, then another external
means of disconnecting the supply from the equipment should be located close by. Circuit
breakers or switches must comply with a recognized standard such as IEC 947.
To maintain explosion-proof protection of the OCX 8800 in hazardous areas, all cable entry
devices and blanking elements for unused apertures must be certified flameproof, suitable
for the conditions of use and properly installed.
To maintain explosion-proof protection of the OCX88C in hazardous areas, the sensor
housing must not be mounted to any surface or flange that exceeds 200ºC (392ºF).
To maintain explosion-proof protection of the OCX88C in hazardous areas, the sample
entering the sensor housing must not exceed 200ºC (392ºF).
2-8
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
NOTE
To maintain proper earth grounding, ensure a positive connection exists
between the sensor housing, the electronics housing, and earth. The
connecting ground wire must be 14 AWG minimum. Refer to Figure 2-5.
NOTE
Line voltage, signal, and relay wiring must be rated for at least 105ºC (221ºF ).
Electrical Connections
Electrical connections, power and communications are made to the ele ctronic
enclosure. The connections are made through two 3/4 NPT ports in the
enclosure using fittings and cables provided by the customer. Cable
installation must meet NEC, IEC and/or other applicable national or local
codes for Class I, Zone 1, IIB +H2 T3/T6 permanently mounted equipment.
Connect Line Voltage
The OCX 8800 operates on 100 to 240 VAC line voltage at 50 to 60 Hz. The
power supply requires no setup. Connect the line (L wire) to the L terminal,
and the neutral (N wire) to the N terminal on the AC power input terminal
block in the electronics housing. Connect the ground (G wire) to the ground
stud in the electronics housing as shown in Figure 2-5.
Connect 4-20 mA Signals
Connect the 4-20 mA current loop to the 4-20 mA signal output terminals in
the electronics housing as shown in Figure 2-5. Use individual shielded
twisted wire pairs. Terminate the shield at the electronics housing.
4-20 mA Signal
O
2
One 4-20 mA signal represents the O
signal is the HART information accessible through a Model 275/375
Handheld Communicator or AMS software. The O
terminals.
COe 4-20 mA Signal
Another 4-20 mA signal at the AOUT 2 terminals represents the COe
value. HART information is not available on the COe signal.
Alarm Output Relay
Connect any customer-supplied relay input to the alarm output relay terminal.
Use shielded wire and terminate the shield at the electronics housing. The
alarm output relay terminal is a set of dry, no. 2, form C, contacts with 30 mA,
30 VDC capacity.
value. Superimposed on the O2
2
signal is at the AOUT 1
2
2-9
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Remote Electronics Connections to Sensor Housing
Make the following connections between the electronics and sensor housings
with the electronics cable ordered with the package (Figure 2-6). Braided
cable is available in lengths up to 150 ft. (46 m).
NOTE
Interconnect wiring shown is for Rosemount Analytical supplied cables. For
customer furnished interconnect wiring or cables, refer to Figure 2-7.
Signal Connections
Connect the electronics housing terminals to the corresponding terminals
in the sensor housing. The twisted wire pairs are numbered on the inner
plastic wrapper. Keep twisted pairs together and match the numbers and
wire colors shown in Figure 2-6.
Heater Power Connections
Use the blue, white, orange black, red, and yellow stranded wires in the
heater power cable to connect power to the three heaters in the sensor
housing. Match the wire colors to the corresponding heater power terminal
blocks in the sensor and electronics housings as shown in Figure 2-6.
2-10
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 2-5. Line Voltage, Earth,
and 4-20 mA Connections
OCX 8800
#1
NC
COM
NO
Alarm Output Relay
Terminal Block
COe Signal
O Signal
4-20 mA Signal Output
Terminal Block
AOUT2+
{
AOUT2 -
2
AOUT1 -
{
AOUT1+
EMI Filter
TOP VIEW
(1/2 SIZE)
Ground Stud
Earth Ground
Typical for Electronics and
Sensor Housing
#1
Customer
Wiring
Signal Port
3/4 NPT
Terminal
G
L1
N
Block
G
Ground
Stud
Power Port
3/4 NPT
G
ExternalTooth
Lockwasher
37390013
2-11
OCX 8800
Figure 2-6. Electrical Connections Between Remote Electronics and Sensor Housing
Instruction Manual
IM-106-880, Rev 1.0
January 2007
YEL
RED
BLK
ORG
WHT
BLU
Heater Power
Connector (J3)
RED
BLK
WHT
BLK
GRN
BLK
BLU
BLK
Heater Power Cable
#1
O Sensor and
2
Thermocouple
Connector (J5)
T/C CO+
T/C CO-
T/C SB+
T/C SBT/C O2+
T/C O2-
O2 CELL+
O2 CELL-
SHLD
#1
2HTR CO
1HTR CO
2HTR O
1HTR O
2HTR SB
1HTR SB
To ground
2
2
screw
#1
#1#1
YEL
BRN
BLK
RED
WHT
ORG
BLK
BLK
#1
EXC+
CO ACT+
CO ACT-
CO REF+
CO REFCJC+
CJCEXC-
GRN
SHIELD
To
ground
screw
COe Sensor
and
Cold Junction
Connector (J4)
ELECTRONICS HOUSING
2-12
SENSOR HOUSING
RED
BLK
-
WHT
+
T/C SB
BLK
EXC-
BLK
GRN
BLK
Signal Cable
-
+
BLU
+
T/C O2
O2
-
BLK
-
+
T/C CO
BLK
-
ORG
CJC
CO
BRN
+
ACT
2
WHT
HTR
YEL
+
EXC
SB
1
BLU
To ground screw
GRN
37390014
WHT
RED
BLK
+
REF
2
YEL
HTR
-
CO
1
RED
2
BLK
HTR
-
O2
1
ORG
CO
+
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 2-7. Customer-Furnished
Interconnect Wiring or Cables
OCX 8800
SIGNALWIRING OR CABLE
NOTE: For RFI/CE compliance, the connector
must provide 360 degrees of electrical
contact to the cable shield.
ELECTRONICS END
7.25
±0.10
Strip Wire
Ends 3/16”
Typical
Stud Size
#10
7.25 ±0.10
9.25 ±0.10
0.5
Typ.
Heat Shrink
Tubing
2” Long
1/2” Size
2.0 ±0.25
Typical
7.0” Long TeflonTubing,
0.042” ID (Cut off drain
wire at probe end of
shield).
Pneumatic system connections depend on whether reference air set,
calibration solenoids, and/or blowback equipment options are equipped on
your transmitter. Refer to the following paragraphs and select the option that
applies to your transmitter configuration.
Reference Air Set Option (only)
When no options or only the reference air set option is equipped, use the
following procedure to install the pneumatic system components.
1. Refer to Figure 2-8. Connect the reference air set (regulator/filter and
pressure gage) to the instrument air inlet on the electronics housing an d
to the inlet side of the dilution air flow meter.
2. Connect the dilution air flow meter output to the dilution air inlet fitting on
the sensor housing.
3. Install an air line between the instrument air outlet fitting on the
electronics housing and the tee fitting on the sensor housing.
.
Do not use 100% nitrogen as an O2 low gas. It is suggested that O2 low gas be between
0.4% and 2.0% O
parts per million. Failure to use proper gases will result in erroneous readings.
. Do not use gases with hydrocarbon concentrations of more than 40
2
4. One CO gas and two O2 gases are used to calibrate the OCX 8800:
CO - 1000 ppm or 4%
low gas - 0.4%
O
2
high gas - 8%
O
2
Connect the output of the test gas sources to the inlet port of the CAL
GAS flow meter. Install an air line between the flow meter outlet port and
the CAL GAS inlet fitting on the sensor housing.
2-14
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 2-8. Pneumatic
Installation, OCX with Reference
Air Set (without Autocalibration)
Eductor
Air In
Sensor
Housing
Reference Air In
Dilution Air In
CAL Gas In
OCX 8800
CAL Gas
Flow Meter
7 scfh, 20-30 psig
(
Recommended
(
Electronics
Housing
Dilution Air
Flow Meter
0.1 scfh
Instrument
Air Out
2-Stage
Regulators
2
2
CO
HI O
LO O
Instrument
Air Supply
Pressure Reguator/Filter
35 psig - General Purpose
45 psig - Hazardous Area
37390011
2-15
OCX 8800
Figure 2-9. Pneumatic
Installation, OCX with Reference
Air Set and Solenoids (with
Autocalibration)
CAL Gas
Flow Meter
7 scfh, 20-30 psig
(
Recommended
Electronics
Housing
(
Eductor
Air In
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Sensor
Housing
CAL Gas In
Reference
Air In
Dilution
Air In
Dilution Air
Flow Meter
0.1 scfh
CAL Gas Out
Instrument Air Out
2
LO O
2
HI O
Pressure Regulator/Filter
35 psig - General Purpose
45 psig - Hazardous Area
2-Stage
Regulators
CO
Instrument
Air Supply
Reference Air Set and Solenoids Option
When the reference air set and test gas solenoids are included with your
OCX 8800, use the following procedure to install the pneumatic system
components.
1. Install the reference air set according to th e instructions in Reference Air
Set Option, steps 1 through 3.
2. Refer to Figure 2-9. Connect the O
inlet fitting on the electronics housing. Install a shutoff valve and
O
2
pressure regulator with gage in the O
3. Connect the O
high gas source to the CAL GAS HI O2 inlet fitting.
2
Install a shutoff valve and pressure regulator with gage in the O
low gas source to the CAL GAS LO
2
low supply line, as shown.
2
high
2
supply line.
37390012
2-16
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
4. Connect the CO high gas to the CAL GAS HI COe inlet fitting. Install a
shutoff valve and pressure regulator with gage in the CO high supply
line.
5. Connect the CAL GAS outlet fitting of the electronics housing to the inlet
port of the CAL GAS flow meter. Install an air line between the flow
meter outlet port and the CAL GAS inlet fitting on the sensor housing.
Reference Air Set, Solenoids, and Blowback Option
The blowback system blows instrument air back through the blowback filter
and out the sample tube of the transmitter. This removes built up dirt and
particulate from the filter and sample line. The blowback option is normally
used in systems that have a dirty process stream.
Installing an OCX 8800 with the blowback option requires the addition of air
operated blowback valve, regulator and gage, and check valve.
Figure 2-10 shows the piping arrangement for the OCX 8800 with the
blowback and autocalibration options. Figure 2-11 shows the piping
arrangement for the OCX 8800 with the blowback option, but without
autocalibration (without test gas solenoids).
When the reference air set, calibration gas solenoids, and blowback options
are included with your transmitter, use the following procedure to install the
pneumatic system components.
1. Connect the calibration gas sources according to the instructions in the
previous paragraph “Reference Air Set and Solenoids Option”, steps 2
through 5.
2. Connect a clean, dry , instrument-quality su pply of air (20.95% O
35/45 psig and 55 psig pressure regulators. The inlet to the 35/45 psig
regulator accepts a 1/8" NPT fitting. The inlet to the 55 psig regulator
accepts a 1/4" NPT fitting.
3. See the upper leg of the instrument air supply. Connect the output of the
35/45 psi regulator/filter to one port of the normally-closed air-operated
solenoid valve, and to the inlet side of the dilution air flow meter.
4. Connect the dilution air flow meter output to the DILUTION AIR inlet
fitting on the sensor housing.
5. Install an instrument air line between the open port of the normally-open
air-operated solenoid valve and the tee fitting on the sensor housing.
6. Connect the output of the 55 psi regulator/filter to one port of the
normally-open air-operated solenoid valve, and to the instrument air
inlet on the back of the electronics housing.
7. Install an air line between the open port of the normally-closed
air-operated solenoid valve and the check valve inlet fitting on the
sensor housing.
) to the
2
2-17
OCX 8800
Figure 2-10. Piping Arrangement,
Blowback with Autocalibration
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Cal Gas Flow Meter
7 scfh, 20-30 psig
(
Recommended
Eductor Air In
Electronics
Housing
(
Check
Valve
Dilution Air In
Instrument Air
*Normally
Open
Solenoid
Valve
Tee - (3 x 1/4 in.
Tube Unions)
Check Valve
Sensor
Housing
Reference
Air In
Vacuum
Gage
*Normally
Closed
Solenoid
Valve
CAUTION
Pressure regulator with 1/8” inlet port is factory
set for 35 or 45 psig. Regulator with 1/4” inlet port
is factory set for 55 psig. If regulators are not
installed in correct locations, the OCX 8800 will
not work.
Dilution
Air Flow Meter
0.1 scfh
Pressure Regulator/Filter
35 psig - General Purpose
45 psig - Hazardous Area
CAL Gas Out
2-Stage
Regulators
2-18
2
LO O
2
HI O
CO
Blowback Valve,
Air Operated
*NOTE: During blowback operation, states of
both solenoid valves change.
NOTE: Wall mount the air-operated blowback
valve on a suitable mounting plate.
Pressure
Regulator/Filter
55 psig
Instrument
Air Supply
38850004
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 2-11. Piping Arrangement,
Blowback without Autocalibration
Eductor Air In
CheckValve
Reference Air In
Dilution Air In
Sensor
Housing
CAL Gas In
OCX 8800
CAUTION
Pressure regulator with 1/8” inlet port is factory
set for 35 or 45 psig.Regulator with 1/4” inlet port
is factory set for 55 psig.If regulators are not
installed in correct locations, the OCX 8800 will
not work.
Cal Gas Flow Meter
(
7 scfh, 20-30 psig
Recommended
(
2-Stage
Regulators
Electronics
Housing
Instrument Air
*Normally
Open
Solenoid
Valve
Dilution
Air Flow Meter
0.1 scfh
*Normally
Closed
Solenoid
Valve
BlowbackValve,
Air Operated
*NOTE: During blowback operation, states of
both solenoid valves change.
NOTE: Wall mount the air-operated blowback
valve on a suitable mounting plate.
Pressure Regulator/Filter
35 psig - General Purpose
45 psig - Hazardous Area
Pressure
Regulator/Filter
55 psig
2
LO O
2
CO
HI O
Instrument
Air Supply
37390005
2-19
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
8. Install an air line between the instrument air outlet fitting on the
electronics housing and the control air inlet fitting on the air-operated
solenoid valve.
9. There are three settings that need to be specified to set up the blowback
option. These are the blowback interval, duration, and purge time.
Interval - Length of time between blowback events.
(60 minutes recommended.)
Duration - Length of time blowback air is activated.
(5 seconds recommended.)
Purge - Length of time after blowback is complete before
oxygen/combustibles readings are co ns ide re d valid.
(Set as required by the application.)
These settings are available through HART fro m the DEVICE SETUP >
DETAILED SETUP > OUTPUT CONDITIONS > BLOWBACK menu.
Using the LOI, these settings are accessible from the SYSTEM >
INPUT/OUTPUT > BLOWBACK menu.
INITIAL STARTUPObserve the following Caution and Note. Refer to Section 3: Configuration
and Startup, for OCX 8800 startup information.
Upon completing installation, make sure that the OCX 8800 is turned on and operating prior
to firing up the combustion process. Damage can result from having a cold OCX 8800
exposed to the process gases.
If ducts will be washed down during outages, make sure to power down the OCX 8800 units
and remove them from the wash area.
NOTE
During outages, and whenever possible, leave OCX 8800 units running to
prevent condensation and premature aging from thermal cycling.
Install all protective equipment covers and safety ground leads after installation. Failure to
install covers and ground leads could result in serious injury or death.
V E RIFY INSTALLATIONEnsure the OCX 8800 is installed correctly . Vertify mechanical installation and
all electrical and pneumatic connections. Refer to Section 2, Installation.
Make sure that the OCX 8800 is turned on and operating prior to firing up the combustion
process. Damage can result from having a cold OCX 8800 exposed to the process gases.
NOTE
During outages, and whenever possible, leave all OCX 8800 units running to
prevent condensation and premature aging from thermal cycling.
http://www..raihome.com
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Verify Configuration
There are three switches on the microprocessor board which are user
configurable for the OCX 8800 (Figure 3-1). SW1 determines if the O
mA signal is internally or externally powered. SW2 determines if the COe 4-20
mA signal is internally or externally powered. SW3 sets the rail limits for the
and COe 4-20 mA signals and configures the sample line heater control
O
2
circuit. All switches are accessible through holes in the electronics box.
Remove power from the OCX 8800 before changing defaults. If defaults are changed under
power, damage to the electronics may occur.
Verify that the following switch settings are correct for your OCX 8800
installation:
4-20
2
SW1 The two settings are internally or externally powering the O
mA signal. The factory setting is for the O
4-20 mA signal to be internally
2
4-20
2
powered.
SW2 The two settings are internally or externally powering the COe 4-20
mA signal. The factory setting is for the COe 4-20 mA signal to be
internally powered.
SW3 The factory sets this switch as follows:
• Position 1 determines the O
4-20 mA signal rail limit. The settings are
2
high, 21.1 mA, or low, 3.5 mA. The factory setti ng is low, 3.5 mA.
• Position 2 determines the COe 4-20 mA signal rail limit. The settings
are high, 21.1 mA, or low, 3.5 mA. The factory setting is high, 21.1 mA.
• Positions 3 and 4 must be set as shown for proper software control of
the device heaters.
3-2
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 3-1. OCX 8800 Defaults
Switch
Default Postions
Shown
21.1 mA
3.5 mA
2
COe
O
1223344
OCX 8800
Open
Closed
Open
Closed
SW3
O 21.1 mA/3.5 mA:2O 4-20 mA Signal
Open
Closed
1
1234
COe 21.1 mA/3.5 mA:
Open
Closed
2
Rail Limits:
High - 21.1 mA
Low - 3.5 mA
COe 4-20 mA Signal
Rail Limits:
High- 21.1 mA
Low - 3.5 mA
Internal:
COe 4-20 mA
is Internally
Powered
SW2
External:
COe 4-20 mA
Requires an External
Power Supply
(Default)
Internal:
O 4-20 mA
2
is Internally
Powered
SW1
External:
O 4-20 mA
2
Requires an External
Power Supply
(Default)
INITIAL POWER UPAllow adequate time (approximately 60 minutes) for the heaters to begin
operation and for the OCX 8800 to reach normal operating temperature on
power up. Normal operating temperature for the O
operating temperature for the combustibles cell is 300°C. The norma l sample
line temperature is 170°C. During this time the eductor air solenoid will remain
closed so no sample is pulled through the analyzer. When the OCX reaches
operating temperature the solenoid will energize, eductor air will begin to flow,
and the unit will begin normal operation.
cell is 736°C. Normal
2
37390026
3-3
Instruction Manual
IM-106-880, Rev 1.0
OCX 8800
January 2007
SET TEST GAS VALUESUse HART/AMS or the optional LOI to set test gas values for calibration.
Refer to Section 4, Using the LOI or Section 5, Using HART Communications
for more information.
Setting Test Gas Values with HART
1. Use the HART communicator or AMS software to access the HART
menu.
2. From the DETAILED SETUP menu, select O2 CALIB PARAMS.
3. From O2 CALIB PARAMS, select O2 High Gas. Enter the percent O
used for the high O
4. From O2 CALIB PARAMS, select Low TG. Enter the percent O
for the low O
5. From the DETAILED SETUP menu, select COe CALIB PARAMS.
6. From COe CALIB PARAMS, select COe Test Gas. Enter the CO
concentration (ppm) used for COe test gas.
Setting Test Gas Values with the LOI
1. Use the "Z" pattern to enter the LOI menu tree.
2. From the SYSTEM menu, select Calib Setup.
3. From Calib Setup, select O2 High Gas %. Enter the percent O
for the high O
4. Select the down arrow and the next selection will be O2 Low Gas %.
Enter the percent O
5. Select the down arrow several times to display COe T est Gas. Enter the
CO concentration (ppm) used for COe test gas.
2
test gas.
2
test gas.
test gas.
2
used for the low O2 test gas.
2
used
2
used
2
2
OCX 8800 RESET
PROCEDURE
Whenever you correct an equipment alarm or fault condition, the OCX 8800
will either revert to normal operation or continue to indicate an alarm status
condition. If the equipment does not revert to normal operation when a fault
condition is cleared, or if instructed to do so in Section 8, Troubleshooting,
use the following procedure to reset the OCX 8800.
OCX Reset with the LOI
1. Use the "Z" pattern to enter the LOI menu tree. (Refer to Section 4,
Using the LOI).
2. Select the SYSTEM submenu.
3. From the SYSTEM submenu, select the Status submenu.
4. From the Status submenu, select Reset Device. The OCX 8800 will
reset and the LOI will revert to the normal operation display.
OCX Reset with HART
Remove the OCX 8800 from the process loop and recycle power.
OVERVIEWThis section describes the installation and operation of the LOI module in the
OCX 8800.
DISPLAY ORIENTATIONThe LOI module mounts to a connector on the LOI board. The board is
installed on the end of the electronics stack in the electronics ho using,
Figure 4-1. There are four mating connectors on the back of the LOI module
that allow the LOI to be oriented as desired by the user.
Figure 4-1. LOI Components
Mounting
Electronics Housing
(Cover Removed)
LOI Connector
LOI Board
LOI Module
Electronics Stack
l
a
c
i
t
y
l
a
n
A
37390027
http://www..raihome.com
Instruction Manual
IM-106-880, Rev 1.0
OCX 8800
January 2007
LOI CONTROLS
OverviewThe LOI, shown in Figure 4-2, utilizes a bright blue gas-fluorescent display.
Intensity is adjustable. There is an Infrared LED source and a detector for
each key. The detectors can detect a finger placed above the button through
the glass window. There is no need to open the instrument in bad weather or
in hazardous areas in order to access the electronics.
It should be noted that the Hazardous Area OCX 8800 also utilizes HART
communications, permitting access to all instrument functionality anywhere
the 4-20 mA signal terminates via a HART model 275/375 handheld
communicator.
Figure 4-2. LOI Assembly
Touch
Selection
Arrow
(Enter key)
Confirmation
LED
Selection
Arrow
E
Lockout
Notation
LK
AL
Analytical
Display
Window
Selection
Arrows
Status
Code
37390042
LOI Key FunctionsThe gray (top left) key will move one level higher in the menu structure. When
entering parameter values (numbers), this key moves the cursor to the left.
The left-pointing key also doubles as an Enter key, used after the digits of a
parameter value are entered, and the cursor is moved to it s left-most po sition.
When the Enter key is touched, the new parameter value, if accepted, will
appear in the top line of the display.
The blue (bottom left) key acts as a selector when choo sing from among
several menu items. This right-pointing key also will move the cursor to the
right when entering the digits of a new parameter value.
4-2
The up and down-pointing keys are used to increment up and down when
selecting from a vertical list of menu items. These keys are also used for
incrementing values up and down for new data input.
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 4-3. ‘Z’ Pattern Entry
OCX 8800
Lockout
The LOI has a lockout feature that prevents nuisance actuation by someone
brushing against the glass window, raindrops, dirt, insects, etc. This lockout
mode is automatically established when no buttons are pushed for 30
seconds (default). This countdown to lockout is configurable.
In order to unlock the display, input a "Z" pattern (Figure 4-3). First, touch the
top left (gray) Enter key. Next, touch the top right key, followed by the bottom
left key and the bottom right key. The LK notation in the upper right corner of
the display will disappear. Touch the Enter key once more to enter into the
menu structure. Whenever a key is touched additional time to lockout is
provided, so that the lockout feature does not beco m e a nuisa nc e. This
additional revert time is one hour (default) and is also user configurable.
1
E
Analytical
3
2
4
37390057
LOI Status CodesThe LOI display shows a status code in the lower right hand corner of the
display. There are nine status codes to indicate the existing status of the
device during operation. The status code description s are shown in Table 4-1.
Table 4-1. LOI Status Codes
CODEDESCRIPTION
ALAlarm - The device is in a recoverable alarm state.
BLBloback - A blowback cycle is active.
CACalibration - A calibration cycle is active.
CVCalibration Verify - A calibration verify task is in progress.
NMNormal - The device is in a normal operating mode.
PO
SF
ST
WUWarm Up - The device heaters are ramping up to operating temperature.
Power On - A system level initialization sequence is active. This will
continue for several seconds.
System Fault - The device is in a non-recoverable alarm condition. The unit
must be reset or power must be cycled off and on to resume operation.
Stabilize - The device heater control is stabilizing (after warm up). Sensors
are warming up to operating temperature.
4-3
Instruction Manual
IM-106-880, Rev 1.0
OCX 8800
January 2007
LOI MENU TREEThis section consist s of a menu tree fo r the LOI on the OCX 8800, Fig ure 4-4.
This menu is specific to the OCX 8800.
First Column Submenus
From the operating display (O
% and COe ppm), the left-pointing Enter key is
2
the only option to move into the first column submenus of the LOI menu tree.
The first column contains three submenus: SENSOR DATA, Figure 4-4 sheet
1 of 4, CALIBRATION, sheet 2 of 4, and SYSTEM, sheets 3 and 4 of 4. From
the operating display, SENSOR DATA is displayed when the right-pointing
key is selected. Use the up or down-pointing key to move to the other first
column submenus.
Second Column Submenus
From the first column submenus, selecting the right-pointing key moves the
display into the second column submenus. The up and down-pointing keys
allow the display to move to the second column submenus of the first column
submenu selected. The left-pointing key moves the display back to the first
column submenu.
Third and Fourth Column Submenus
From the second column submenus, selecting the right-pointing key moves
the display into the third column submenus. The third column submenu may
be another menu or a list of parameters. The up- and down-pointing keys
allow the display to move to the different para meters or menus. The third or
fourth column submenu may be a parameter list. When a parameter list is
displayed, the cursor will blink. The up- and down-pointing keys select the
value for the parameter displayed.
Figure 4-4. LOI Menu Tree
(Sheet 1 of 4)
O2X.XX%
%
CombXXX ppm
%
SENSOR DATA
(CONTINUED ON
SHEET 2)
Temperatures
RawValues
Analog Outputs
O2 Temp _____dgC
O2 Temp-MAX _____dgC
COe Temp _____dgC
COe Temp-MAX _____dgC
SB Temp _____dgC
SB Temp-MAX _____dgC
Board Temp _____dgC
Board Temp-MAX _____dgC
CJC Temp _____dgC
CJC Temp-MAX _____dgC
O2 Sensor _____ mV
O2 Sensor R ____ ohm
O2 T/C _____ mV
COe Delta V _____ mV
COe Delta R
COe ReferenceV
COe Reference R
RTD Current _____ mA
Board Temp IC
CJC Temp Signal _____ mV
O2 Output % _____%
O2 Current ____mA
COe Output % _____%
COe Current ____mA
____ ohm
_____ mV
____ ohm
_____ mV
COe T/C
SB T/C _____ mV
_____ mV
37390007
4-4
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 4-4. LOI Menu Tree
(Sheet 2 of 4)
OCX 8800
(CONTINUED FROM
SHEET 1)
CALIBRATION
(CONTINUED ON
SHEET 3)
Cal
Control
Abort Cal
Start Cal
O2
Start Cal
COe
Start Cal
Both
Apply Lo O2 Gas
Hit E when ready
Flow O2 Lo _____s
______% _____mV
Read O2 Lo _____s
______% _____mV
Apply Hi O2 Gas
Hit E when ready
Flow O2 Hi_____s
______% _____mV
Read O2 Hi _____s
______% _____mV
Hit E when ready
Purge_____s
Apply Lo COe Gas
Hit E when ready
Flow COe Lo _____s
______%_____mV
Read COe Lo _____s
______%_____mV
Apply Hi COe Gas
Hit E when ready
Flow COe Hi _____s
______%_____mV
Read COe Hi _____s
______%_____mV
Hit E when ready
Purge_____s
Apply Lo O2 Gas
Hit E when ready
Flow O2 Lo _____s
______% _____mV
Read O2 Lo _____s
______% _____mV
Apply Hi O2 Gas
Hit E when ready
Flow O2 Hi_____s
______% _____mV
Read O2 Hi _____s
______% _____mV
Hit E when ready
Purge_____s
Flow COe Hi _____s
______%_____mV
Read COe Hi _____s
______%_____mV
Hit E when ready
Purge_____s
Cal
Constants
Cal
Status
Cal
Verify
Current
Calib
Prev
Calib
Falied
Calib
Calib Result
Calib Step
Calib Time
Next O2 Cal
Next COe Cal
Flow High Gas
Flow Low Gas
Flow COe Gas
Purge
Status
Pre O2 Slope_____mV/D
Pre O2 Const_____mV
Pre O2 Sensor R _____ohm
Pre COe Slope_____ppm/mV
Pre COe Const_____mV
Bad O2 Slope_____mV/D
Bad O2 Const_____mV
Bad COe Slope _____ppm/mV
Bad COe Const _____mV
NOTE: “Hit E when ready” is displayed during
semi-automatic calibration only (when
Calib Setup value Use Solenoids = n).
37390015
4-5
OCX 8800
Figure 4-4. LOI Menu Tree
(Sheet 3 of 4)
(CONTINUED FROM
SHEET 2)
SYSTEM
Calib Setup
Instruction Manual
IM-106-880, Rev 1.0
January 2007
O2 High Gas%
O2 Low Gas%
O2 Reset ValsYes/No
O2 Out TracksYes/No
O2 Cal IntervalHrs
O2 Next CalHrs
COe Test Gas%
COe Reset ValsYes/No
COe Out TracksYes/No
COe Cal IntervalHrs
COe Next CalHrs
COe Slope Warn%
Use SolenoidsYes/No
Gas TimeSeconds
Purge TimeSeconds
Input/Output
(CONTINUED ON
SHEET 4)
Analog
Alarm Relay
Blowback
O2 Type20-4 mA
O2 Range Hi___%
O2 Range Lo
O2 AlarM Level___mA
Trim O2 Out (procedure)
COe Type20-4 mA
COe Range HI___ppm
COe Range Lo___ppm
COe Alarm Lvl___mA
Trim COe Out (procedure)
Trigger 1 Event
Trigger 2 Event
Trigger 3 Event
Blow Bk Enable __Yes/No
Blow Bk Intrvl ___Minutes
Blow Bk Period ___Seconds
Blow Bk Purge ___Seconds
Blow Bk Status
___%
All
COe Temp
Sample Pr Sen
Sample Ln Temp
Unit Fail
Hi Elect Temp
Calib Fail
O2 Cell Bad
O2 Htr Open
O2 Cell Temp
In Calibrtn
Off
Version
Checksum
Build Number
Build Date
Restart Count
SW Err File
SW Err Line
SW Err Number
Alarms
CPLD Registers
Line Frequency
Line Voltage
PCD Counter
PCN Counter
Reset Device
Sensor Housing
Serial Number
OCX 8800
37390056
4-7
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
D/A TRIM PROCEDURESO
D/A trim procedure using the LOI
2
Use the following procedure to perform the O
D/A trim procedure at the LOI.
2
Refer to the LOI menu tree in Figure 4-4.
To avoid a potentially dangerous operating condition, the OCX 8800 must be removed from
the automatic combustion control loop before you start the D/A trim procedure.
1. From the operating display use the left-pointing key to select the first
column submenu. Use the down-pointing key to select SYSTEM.
2. From the SYSTEM menu, use the down-pointing key to select
Input/Output. Use the right-pointing key to select the Analog
parameters list.
3. Scroll down to the item Trim O2 Out. Touch the right-pointing key to
start the O
trim procedure.
2
NOTE
If you wish to exit D/A Trim with no changes, step through the procedure using
yes responses, and enter no meter readings.
4. Remove the electronics housing cover.
5. Refer to Figure 2-5. Connect a digital multimeter to read the milliamp
output from the O
D/A converter circuit. Connect the positive lead to
2
the AOUT1+ terminal and connect the negative lead to the AOUT1terminal. Then, touch the Enter key at the LOI.
6. The LOI displays 4 mA........Meter. The trim program inputs the
design-equivalent signal for a 4.00 mA output. Read the O
millamp
2
output at the digital multimeter. Use the right-pointing key to select each
digit and use the up- and down-pointing keys to change the value. When
the correct value is displayed, use the Enter key to input the value.
7. The LOI displays 20 mA........Meter. The trim program inputs the
design-equivalent signal for a 20.00 mA output. Read the O
millamp
2
output at the digital multimeter. Use the right-pointing key to select each
digit and use the up- and down-pointing keys to change the value. When
the correct value is displayed, use the Enter key to input the value.
8. The LOI displays a Meter at 4 mA prompt. Use the ri ght- po inting key to
select the letter yes or no. Use the up- or down-pointing key to change
the letter. Then use the Enter key to input the reponse. If no, the
process repeats from step 7.
9. The LOI displays a Meter at 20 mA prompt. Use the right-pointing key
to select the letter yes or no. Use the up- or down-pointing key to
change the letter . Then use the Enter key to input the reponse. If no, the
process repeats from step 8.
10. When the reponses in steps 9 and 10 are yes, the trim procedure is
complete. Exit the LOI menu and return the control loop to automatic
control.
4-8
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
COe D/A trim procedure using the LOI
Use the following procedure to perform the COe D/A trim procedure at the
LOI. Refer to the LOI menu tree in Figure 4-4.
To avoid a potentially dangerous operating condition, the OCX 8800 must be removed from
the automatic combustion control loop before you start the D/A trim procedure.
1. From the operating display use the left-pointing key to select the first
column submenu. Use the down-pointing key to select SYSTEM.
2. From the SYSTEM menu, use the down-pointing key to select
Input/Output. Use the right-pointing key to select the Analog
parameters list.
3. Scroll down to the item Trim COe Out. Touch the right-pointing key to
start the COe trim procedure.
NOTE
If you wish to exit D/A Trim with no changes, step through the procedure using
yes responses, and enter no meter readings.
4. Remove the electronics housing cover.
5. Refer to Figure 2-5. Connect a digital multimeter to read the milliamp
output from the COe D/A converter circuit. Connect the positive lead to
the AOUT2+ terminal and connect the negative lead to the AOUT2terminal. Then, touch the Enter key at the LOI.
6. The LOI displays 4 mA........Meter. The trim program inputs the
design-equivalent signal for a 4.00 mA output. Read the COe millamp
output at the digital multimeter. Use the right-pointing key to select each
digit and use the up- and down-pointing keys to change the value. When
the correct value is displayed, use the Enter key to input the value.
7. The LOI displays 20 mA........Meter. The trim program inputs the
design-equivalent signal for a 20.00 mA output. Read the COe
output at the digital multimeter. Use the right-pointing key to select each
digit and use the up- and down-pointing keys to change the value. When
the correct value is displayed, use the Enter key to input the value.
8. The LOI displays a Meter at 4 mA prompt (question). Use the
right-pointing key to select the letter yes or no. Use the up- or
down-pointing key to change the letter. Then use the Enter key to input
the reponse. If no, the process repeats from step 7.
9. The LOI displays a Meter at 20 mA prompt (question). Use the
right-pointing key to select the letter yes or no. Use the up- or
down-pointing key to change the letter. Then use the Enter key to input
the reponse. If no, the process repeats from step 8.
10. When the reponses in steps 9 and 10 are yes, the trim procedure is
complete. Exit the LOI menu and return the control loop to automatic
control.
OVERVIEWThe HART communicator is a handheld communications interface device. It
provides a common communications link to all microprocessor-based
instruments that are HART comp atible. Th e handheld commu nicator cont ains
an 8 x 21 character liquid crystal display (LCD) and 25 keys. A pocket-sized
manual, included with the HART communicator, details the specific functions
of all the keys.
To interface with the OCX 8800, the HART communicator requires a
termination point along the O
resistance of 250 ohms between the communicator and the power supply.
The HART communicator accomplishes its task using a Frequency Shift
Keying (FSK) technique. With the use of FSK, high-frequency digital
communication signals are superimposed on the 4-20 mA oxygen output
signal. The communicator does not disturb the 4-20 mA signal since no net
energy is added to the loop. HART information is not available on the COe
output signal.
4-20 mA current loop and a minimum load
2
HART COMMUNICATOR
SIGNAL CONNECTIONS
The HART communicator may be interfaced with a personal computer (PC) ,
providing special software has been installed. To connect the HART
communicator to a PC, an interface adapter is required. Refer to the proper
HART communicator documentation in regard to the PC interface option.
The HART communicator can connect to the OCX 8800 oxygen analo g output
signal line at any wiring termination in the O
two methods of connecting the HART communicator to the signal line. For
applications in which the signal line has a load resistance of 250 ohms or
more, refer to method 1. For applications in which the signal line load
resistance is less than 250 ohms, refer to method 2.
4-20 mA current loop. There are
2
http://www..raihome.com
OCX 8800
Figure 5-1. Signal Line Connections,
≥ 250 Ohms Load Resistance
4-20 mA Terminal Block in
Electronics Housing
Instruction Manual
IM-106-880, Rev 1.0
January 2007
A OUT 1 +
A OUT 1 -
1
Loop
Connectors
HART Communicator
Rear Panel
Method 1, for Load Resistance
Refer to Figure 5-1 and the following steps to connect the HART
communicator to a signal line 250 ohms or more of load resistance.
O 4-20 mA Signal Line
2
HART
Communicator
Model 275/375
≥
RL 250Ω
Analog Output Device
Lead Set
≥ 250 Ohms
37390023
5-2
Do not make connections to the HART communicator's serial port, 4-20 mA signal lines, or
NiCad recharger jack in an explosive atmosphere. Explosions can result in death or serious
injury.
Using the supplied lead set, connect the HART communicator in parallel to
the OCX 8800. Use any wiring termination points in the oxygen analog output
4-20 mA signal line.
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 5-2. Signal Line Connections,
< 250 Ohms Load Resistance
OCX 8800
Method 2, for Load Resistance < 250 Ohms
Refer to Figure 5-2 and the following steps to connect the HART
communicator to a signal line with less than 250 ohms load resistance.
Do not make connections to the HART communicator's serial port, 4-20 mA signal lines, or
NiCad recharger jack in an explosive atmosphere. Explosions can result in death or serious
injury.
1. At a convenient point, break the oxygen analog output 4-20 mA signal
line and install the optional 250 ohm load resistor.
2. Plug the load resistor into the loop conne ctors (located on the rear panel
of the HART communicator).
4-20 mA Terminal Block in
Electronics Housing
1
A OUT 1 +
A OUT 1 -
Loop
Connectors
HART Communicator
Rear Panel
O 4-20 mA Signal Line
2
HART
Communicator
Model 275/375
The signal loop must be broken to insert
NOTE:
the optional 250 ohm load resistor.
RL <250Ω
Analog Output Device
250 OHM
Load Resistor
(See Note)
37390021
5-3
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
HART COMMUNICATOR
PC CONNECTIONS
There is an option to interface the HART communicator with a personal
computer. Load the designated AMS software into the PC. Then link the
HART communicator to the PC using the interface PC adapter that connects
to the serial port (on the communicator rear panel). Refer to the proper HART
communicator documentation in regard to the PC interface option.
Off-line and On-line Operations
The HART communicator can be operated both off-line and on-line.
Off-line operations are those in which the communicator is not connected to
the OCX 8800. Off-line operations can include interfacing the HART
communicator with a PC. (Refer to applicable HART document ation regarding
HART/PC applications.)
In the on-line mode, the communicator is connected to the oxygen 4-20 mA
analog output signal line. The communicator is connected in parallel to the
OCX 8800 or in parallel to the 250 ohm load resistor.
The opening menu displayed on the LCD is different for on- line and off-line
operations. When powering up a disconnected (off-line) communicator, the
LCD will display the Main Menu. When powering up a connected (on-line)
communicator, the LCD will display the On-line Menu. Refer to the HART
communicator manual for detailed menu information.
5-4
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
HART MENU TREEThis section consists of a menu tree for the HART communicator. This menu
Use the following procedure to perform the O
HART communicator or AMS. Refer to the HART menu tree, Figure 5-3.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select D/A TRIM. The HART
communicator displays O2 D/A Trim.
2. Press the right arrow key to start the procedure. (If you wish to exit D/A
Trim with no changes, select ABORT.)
3. The HART communicator displays WARNING: Loop should be removed from automatic control. Remove the OCX 8800 from any
automatic control loops to avoid a pote ntially dangerous operating
condition and press OK.
4. Remove the electronics housing cover.
5. Refer to Figure 2-5. Connect a digital multimeter to read the milliamp
output from the O
AOUT1+ terminal and connect the negative lead to the AOUT1terminal. Then, press OK at the HART communicator.
6. The HART communicator displays Setting Fld dev output to 4 mA.
Press OK. Read the O
the reading at the HART communicator and press ENTER. (Select
ABORT to exit without changes).
7. The HART communicator displays Setting Fld dev output to 20 mA.
Press OK. Read the O
the reading at the HART communicator and press ENTER. (Select
ABORT to exit without changes).
8. The HART communicator displays Setting Fld dev output to 4 mA.
Press OK.
9. The HART communicator displays Fld dev output 4.00 mAequal to reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 6.
10. The HART communicator displays Setting Fld dev output to 20 mA.
Press OK.
11. The HART communicator displays Fld dev output 20.00 mAequal to reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 7.
12. The HART communicator displays NOTE: Loop may be returned to automatic control.
D/A coverter circuit. Connect the positive lead to the
2
millamp output at the digital multimeter. Enter
2
millamp output at the digital multimeter. Enter
2
D/A trim procedure using the
2
5-9
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
COe D/A trim procedure using HART/AMS
Use the following procedure to perform the COe D/A trim procedur e using the
HART communicator or AMS. Refer to the HART menu tree, Figure 5-3.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select D/A TRIM. The HART
communicator displays O2 D/A Trim. Press the up or down arrow to
select COe D/A Trim.
2. Press the right arrow key to start the procedure. (If you wish to exit D/A
Trim with no changes, select ABORT.)
3. The HART communicator displays WARNING: Loop should be removed from automatic control. Remove the OCX 8800 from any
automatic control loops to avoid a pote ntially dangerous operating
condition and press OK.
4. Remove the electronics housing cover.
5. Refer to Figure 2-5. Connect a digital multimeter to read the milliamp
output from the COe D/A coverter circuit. Connect the positive lead to
the AOUT2+ terminal and connect the negative lead to the AOUT2terminal. Then, press OK at the HART communicator.
6. The HART communicator displays Setting Fld dev output to 4 mA.
Press OK. Read the COe
the reading at the HART communicator and press ENTER. (Select
ABORT to exit without changes).
7. The HART communicator displays Setting Fld dev output to 20 mA.
Press OK. Read the COe
the reading at the HART communicator and press ENTER. (Select
ABORT to exit without changes).
8. The HART communicator displays Setting Fld dev output to 4 mA.
Press OK.
9. The HART communicator displays Fld dev output 4.00 mAequal to reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 6.
10. The HART communicator displays Setting Fld dev output to 20 mA.
Press OK.
11. The HART communicator displays Fld dev output 20.00 mAequal to reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 7.
12. The HART communicator displays NOTE: Loop may be returned to automatic control.
OVERVIEWDuring a calibration, two calibration gases with known O
one calibration gas with a known COe concentration are applied to the OCX
8800. Slope and constant values are calculated to determ ine if the OCX 8800
and combustibles in the
2
FULLY AUTOMATIC
CALIBRATION
is correctly measuring net concentrations of O
industrial process.
Before calibrating the OCX 8800, verify that the calibration gas parameters
are correct by setting the test gas values used when calibratin g the unit. Refer
to Section 3, Configuration and Startup.
There are three calibration methods available to the OCX 8800; automatic,
operator-initiated automatic, and manual. Calibration commands and menus
can be accessed by HART/AMS or by the LOI.
If the OCX 8800 is equipped with calibration solenoids, the unit can be
programmed to automatically calibrate without any operator action. Refer to
the following paragraphs for using the LOI or HART/AMS to set up the OCX
8800 for fully automatic calibration.
Autocalibration Setup using the LOI
Use the following procedure to set up the OCX 880 0 for automatic calibration.
If necessary , use the LOI me nu tree in Figure 4-4 for reference. The unit must
be equipped with calibration solenoids to use automatic calibration.
NOTE
Automatic calibration is only available on units equipped with calibration
solenoids.
concentrations and
2
http://www..raihome.com
1. From the operating display use the right-pointing key to select SYSTEM
first column submenu.
2. From the SYSTEM first column submenu, use the right-pointing key to
select the Calib Setup second column submenu.
3. From the Calib Setup second column submenu, use the right-pointing
key to select the third column parameter list.
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
4. Scroll down to the last item Use Solenoids. If the unit is equipped with
calibration solenoids and timed automatic calibration is desired, select
Yes.
5. Use the up-pointing key to select the item O2 Out T racks. Select Yes or
No to determine if updates to the O
6. Use the down-pointing key to select the item COe Out Tracks. Select Yes or No to define if updates to the COe lock value will take place.
7. Use the down-pointing key to select the item O2 Cal Interval. Enter the
amount of time in days and hours that is desired between automatic
calibrations.
8. Use the down-pointing key to select the next item O2 Next Cal. Enter
the amount of time in hours until the next automatic calibration. Select
the left-pointing key three times to move back to the LOI operating
display.
Autocalibration Setup using HART
NOTE
Automatic calibration is only available on units equipped with calibration
solenoids.
lock value will take place.
2
Use the following procedure to specify a time interval (in hours) at which the
OCX 8800 will automatically calibrate.
1. From the DEVICE SETUP screen, select DETAILED SETUP.
2. From the DETAILED SETUP screen, select O2 CALIB PARAMS or COE CALIB PARAMS.
3. If the unit is equipped with calibration solenoids and timed automatic
calibrations are desired, select Solenoids, then select Yes. Selct No to
disable the calibration solenoids.
6-2
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
4. Select O2 CalIntrvl (O2 calibration interval) and enter the desired time
in hours between automatic O
enter the desired time between automatic COe calibrations. To disable
automatic calibration for O
parameters.
5. If desired, the O2NxtCalTm and the COeNxtCalTm (next calibration
time) parameters can be changed to synchronize a calibration at a
specific day or time.
When setting automatic calibration times, CalIntrvl and NxtCalTm should be set so that O2
and COe are NOT calibrated simultaneously.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
calibrations. Select COE Callintrvl and
2
and COe, enter 0 for both CalIntrvl
2
To return to a preceding menu, press the left arrow key.
6. From the O2 CALIB PARAMS screen, select CalIntrvl.
7. At the prompt, input a time interval (in hours) at which an automatic O
calibration will occur and press ENTER.
8. From the DETAILED SETUP screen, select COE CALIB PARAMS.
9. From the COE CALIB PARAMS menu, select CalIntrvl.
10. At the prompt, input a time interval (in hours) at which an automatic COe
calibration will occur and press ENTER.
2
6-3
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OPERATOR - INITIATED
AUTOCALIBRATION
An operator can initiate an automatic calibration at any time provided that the
unit is equipped with calibration solenoids.
Autocalibration using the LOI
To initiate a calibration using the LO I, pe rf or m the fo llowin g steps on the LOI
menu tree. Refer to Section 4, Using the LOI, for the LOI menu tree.
1. From the CALIBRATION menu, use the right-pointing arrow to select
the Cal Control menu.
2. Select Start Cal-O2, Start Cal COe, or Start Cal Both to start the
calibration. Select Cal Verify to access the calibration window.
3. At the prompt, use the right-pointing arrow to initiate automatic
calibration.
Autocalibration using HART
To initiate an automatic calibration using HART/AMS, perform the following
steps on the HART menu tree. Refer to Section 5, Using HART
Communications, for the HART menu tree.
1. Select DIAG/SERVICE from DEVICE SETUP menu.
2. Select CALIBRATE from the DIAG/SERVICE menu.
3. Select PERFORM CAL from the CALIBRATE menu.
4. Select CAL METHODS from the PERFORM CAL menu.
5. From the CAL METHODS menu, select the type of calibration desired:
O2 Calibration, COe Calibration, or O2 and COe Calibration.
MANUAL CALIBRATIONIf a unit is not equipped with calibration solenoids, a calibration must be
performed by an operator following prompts from the unit. Refer to the
following paragraphs for manual calibration.
Manual Calibration using the LOI
Use the following procedure to perform a manual calibration with the LOI. If
necessary, refer to the menu tree in Section 4, Using the LOI. Once the
manual calibration procedure is initiated at the LOI, a series of prompts will
appear giving instructions to the operator.
1. Use the right-pointing key to select the CALIBRATION first column
submenu.
2. From the CALIBRATION submenu use the right-pointing key to select
the Cal Control second column submenu.
3. From the Cal Control submenu use the right-pointing key to select the
third column Start Cal O2 option.
4. Remain at the Start Cal O2 option or use the down-pointing key to
select the Start Cal COe option or Start Cal Both option. (The following
sequence applies when Start Cal Both is selected.)
5. Use the right-pointing key to start the calibration. T urn on the low O
gas, when prompted by the Flow Low Gas message.
6. Press the right-pointing key when the low O
calibration data changes as the calibration proceeds.
test gas is applied. The
2
test
2
6-4
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
7. Press the right-pointing key when the low O2 reading is stable. Turn off
the low O
Flow High Gas message.
8. Press the right-pointing key when the high O
calibration data changes as the calibration proceeds.
9. Press the right-pointing key when the high O
the high O
purge.
10. When the purge period expires, the LOI display reverts to the nor mal
operation display. If the calibration failed, the display will indicate an
alarm condition.
11. Press the right-pointing key to start combustibles calibration. Turn on
the CO test gas when prompted.
12. Press the right-pointing key when the CO test gas is applied. The
calibration data changes as the calibration proceeds.
13. Press the right-pointing key when the CO reading is stable.
14. Turn off the CO test gas and press th e right-p ointing key to sta rt the CO
gas purge.
15. When the purge period expires, the LOI display reverts to the nor mal
operation display. If the calibration failed, the display will indicate an
alarm condition.
test gas and turn on the high O2 test gas as prompted by the
2
test gas is applied. The
2
reading is stable. Turn of f
test gas. Press the right-pointing key to start the high O2 gas
2
2
Manual O
To perform a manual O
Calibration using HART
2
calibration using the HART communicator or AMS,
2
use the following procedure. If necessary, refer to Section 5, Using HART
Communications, for the HART menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select CALIBRATION. From the CALIBRATION menu, select PERFORM CAL.
2. From the PERFORM CAL menu, select CAL METHODS. Select O2 CALIBRATION to start O
3. In the first O
Calibration screen, a Loop should be removed from
2
calibration.
2
automatic control warning appears. Remove the OCX 8800 from any
automatic control loops to avoid a pote ntially dangerous operating
condition and press OK.
4. The main Calibration screen should look like the following. Press OK to
continue.
OCX: TAG NAME
STATUS: Idle
TIME REMAIN: 0s
O2: 0.4 %, 85.95mV
OK/NEXT to Select
ABORT/CANCEL to Exit
6-5
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
5. From the SELECT ACTION screen, select START/NEXT CALSTEP to
continue calibration, select ABORT CAL to abort calibration or EXIT
CAL to exit calibration. Select one item from the list and press ENTER.
OCX: TAG NAME
SELECT ACTION
1. START/NEXT CALSTEP
2. ABORT CAL
3. EXIT CAL
6. When the Calibration Status is at the AppO2Low step, switch on O
Low Gas. Verify the O
concentration measured matches the O2 LOW
2
GAS parameter in the Setup. Press OK when ready.
7. Select Start/Next Cal Step to start applying the O
Low Gas. The time
2
to apply the test gas is specified by the Gas Time.
8. The Calibration St atus should be automatically changed to FlowO2Low
and then ReadO2Low for a period of time. During this period, if an
attempt is made to go to the next calibration step by pressing OK and
selecting St art/Next Cal S tep, you will be prompted with Operator step command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
9. When ready, Calibration Status will stop at the AppO2Hi. Switch off the
Low Gas and switch on the O2 High Gas. V erify the O2 concentration
O
2
measured matches the O2 HIGH GAS parameter in the Setup. Press
OK when ready.
10. Select Start/Next Cal Step to start applying the O
High Gas. The time
2
to apply the test gas is specified by the Gas Time.
11. The Calibration Status should be automatically changed to FlowO2Hi
and then ReadO2Hi for a period of time. During this perio d, if an attempt
is made to go the next calibration step by pressing OK and selecting
Start/Next Cal Step, you will be prompted with Operator step
command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
12. When ready, Calibration Status will stop at STOP GAS. Switch off the
High Gas. Press OK when ready . Select St art /Next Cal S tep to st art
O
2
purging gas. The time to purge gas is specified by the Purge Time.
13. When the Purge step is complete, the Calibration Status will be at IDLE
if the calibration is successful or CAL RECOMMENDED if the
calibration has failed. A Calibration Failed alarm will be set if the
calibration has failed.
14. When calibration is complete. Select Exit Cal to exit the calibration
method.
2
6-6
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
Manual COe Calibration using HART
To perform a manual COe calibration using the HART communicator or AMS,
use the following procedure. If necessary, refer to Section 5, Using HART
Communications, for the HART menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select CALIBRATION. From the CALIBRATION menu, select PERFORM CAL.
2. From the PERFORM CAL menu, select CAL METHODS. Select COe CALIBRATION to start COe calibration.
3. In the first COe Calibration screen, a Loop should be removed from automatic control warning appears. Remove the OCX 8800 from any
automatic control loops to avoid a pote ntially dangerous operating
condition and press OK.
4. The main Calibration screen should look like the following. Press OK to
continue.
OCX: TAG NAME
STATUS: Idle
TIME REMAIN: 0s
COe: 0.20 ppm, 0.00 mV
OK/NEXT to Select
ABORT/CANCEL to Exit
5. From the SELECT ACTION screen, select START/NEXT CALSTEP to
continue calibration, select ABORT CAL to abort calibration or EXIT
CAL to exit calibration. Select one item from the list and press ENTER.
OCX: TAG NAME
SELECT ACTION
1. START/NEXT CALSTEP
2. ABORT CAL
3. EXIT CAL
6. The unit samples reference air as the COe
Status sho uld automatically change to ReadCOLow for a period of time.
During this period, if an attempt is made to go to the next calibration
step by pressing OK and selecting Start/Next Cal Step, you will be
prompted with Operator step command is not accep ted at th is time .
The Next Cal Step command is not accepted at this time.
7. When ready, Calibration Status will stop at the AppCOeHi. Switch on
the COe High Gas. Ve rify the COe concentration measured matches the
COe HIGH GAS parameter in the Setup. Press OK when ready.
8. Select Start/Next Cal Step to start applying the COe High Gas. The
time to apply the test gas is specified by the Gas Time.
Low Gas. The Calibration
6-7
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
9. The Calibration Status should be automatically changed to FlowCOeHi
and then ReadCOeHi for a period of time. During this period, if an
attempt is made to go the next calibration step by pressing OK and
selecting St art/Next Cal S tep, you will be prompted with Operator step command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
10. When ready, Calibration Status will stop at STOP GAS. Switch off the
COe High Gas. Press OK when ready. Select Start/Next Cal Step to
start purging gas. The time to purge gas is specified by the Purge T ime.
1 1. When the Purge step is complete, the Calibration Status will be at IDLE
if the calibration is successful or CAL RECOMMENDED if the
calibration has failed. A Calibration Failed alarm will be set if the
calibration has failed.
12. When calibration is complete. Select Exit Cal to exit the calibration
method.
Manual O
To perform a manual O
or AMS, use the following procedure. If necessary, refer to Section 5, Using
HART Communications, for the HART menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select CALIBRATION. From the
2. From the PERFORM CAL menu, select CAL METHODS. Select 02 &
3. In the first Calibration screen, a Loop should be removed from
4. The main Calibration screen should look like the following. Press OK to
and COe Calibration using HART
2
and COe calibration using the HART communicator
2
CALIBRATION menu, select PERFORM CAL.
COe CALIBRATION to start O
automatic control warning appears. Remove the OCX 8800 from any
automatic control loops to avoid a pote ntially dangerous operating
condition and press OK.
continue.
OCX: TAG NAME
STATUS: Idle
TIME REMAIN: 0s
O2: 0.4 %, 85.95mV
COe: 0.20 ppm, 0.00 mV
OK/NEXT to Select
ABORT/CANCEL to Exit
and COe calibration.
2
6-8
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
5. From the SELECT ACTION screen, select START/NEXT CALSTEP to
continue calibration, select ABORT CAL to abort calibration or EXIT
CAL to exit calibration. Select one from the list and press ENTER.
OCX: TAG NAME
SELECT ACTION
1. START/NEXT CALSTEP
2. ABORT CAL
3. EXIT CAL
6. When the Calibration Status is at the AppO2Low step, switch on O
Low Gas. Verify the O
concentration measured matches the O2 LOW
2
GAS parameter in Setup. Press OK when ready.
7. Select Start/Next Cal Step to start applying the O
Low Gas. The time
2
to apply the test gas is specified by the Gas Time.
8. The Calibration St atus should auto matically change to FIowO2Low and
then ReadO2Low for a period of time. During this period, if an attempt
is made to go to the next calibration step by pressing OK and selecting
Start/Next Cal Step, you will be prompted with Operator step
command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
9. When ready, Calibration Status will stop at AppO2Hi. Switch off the O
Low Gas and switch on the O
High Gas. Verify the O2 concentration
2
measured matches the O2 HIGH GAS parameter in Setup. Press OK
when ready.
10. Select Sta rt/Next Cal S tep to apply th e O
High Gas. The time to apply
2
the test gas is specified by the Gas Time.
11. The Calibration Status should automatically change to FlowO2Hi, then
ReadO2Hi, and then ReadCOeLo for a period of time. During this
period, if an attempt is made to go the next calibration step by pressing
OK and selecting Start/Next Cal Step, you will be prompted with
Operator step command is not accepted at this time. The Next Cal
Step command is not accepted at this time.
12. When ready , Calibration St atus will stop at AppCOeHi. Switch off the O
High Gas and switch on the COe Gas. Verify the COe concentration
measured matches the COe TEST GAS parameter in the Setup. Press
OK when ready.
13. Select Start/Next Cal Step to start applying the COe Gas. The time to
apply the test gas is specified by the Gas Time.
14. The Calibration Status should automatically change to FlowCOeHi and
then ReadCOeHi for a period of time. During this period, if an attempt is
made to go the next calibration step by pressing OK and selecting
Start/Next Cal Step, you will be prompted with Operator step
command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
15. When ready, Calibration Status will stop at STOP GAS. Switch off the
COe gas. Press OK when ready. Select Start/Next Cal Step to start
purging gas. The time to purge gas is specified by Purge Time.
OVERVIEWThis section contains the procedures to maintain and service the OCX 8800.
Install all protective equipment covers and safety ground leads after equipment repair or
service. Failure to install covers and ground leads could result in serious injury or death.
OCX 8800 REMOVAL
AND INSTALLATION
It is recommended that the OCX 8800 be removed from the stack for all service activities.
The unit should be allowed to cool and be taken to a clean work area. Failure to comply may
cause severe burns.
Disconnect and lock out power before working on any electrical components. There may be
voltage up to 264 VAC.
Observe housing classification guidelines prior to removing cover. Failure to observe
classification guidelines may result in serious injury or death.
Use the following procedures to remove or install the OCX 8800.
http://www..raihome.com
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX with
Remote Electronics
Figure 7-1. OCX with
Remote Electronics
Remove Sensor Housing
1. Turn off power to the system.
2. Shut off the test gasses at the cylinders and close the instrument ai r
valve.
3. Disconnect the calibration gas, reference air, eductor air, and dilution air
lines from the sensor housing, Figure 7-1.
Duct
Stack
Sensor
Housing
Heater
Power Cable
Signal Cable
Eductor Air
Reference Air
Dilution Air
Test Gas
4-20 mA
Outputs
AC Power
Input
Instrument Air
Electronics
Housing
Instrument Air
(Reference Gas)
High O Test Gas
2
Low O Test Gas
2
COTest Gas
37390044
7-2
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 7-2. Sensor
Housing Terminals
OCX 8800
4. Remove the cover from the sensor housing to expose the sensor
housing terminal blocks, Figure 7-2.
5. Disconnect the signal cable from the O
from the CO and CJC terminal blocks.
6. Disconnect the heater power cable from the HTR terminal blocks.
7. If moving the sensor housing to another work site, disconnect and
remove the power and signal cables from the sensor housing.
8. Remove insulation to access the mounting bolts. Unbolt the sensor
housing from the stack and take it to a clean work area.
9. Allow the sensor housing to cool to a comfortable working temperature.
and T/C terminal blocks, and
2
Heater Power Cable
Signal Cable
BLK
-
NOTE: Wire colors shown are for cables
+
BLU
GRN
BLK
RED
BLK
+
T/C CO
-
-
+
T/C O2
supplied by Emerson Process
Management.
O2
-
BLK
WHT
+
T/C SB
BLK
EXC
BLK
-
ORG
CJC
SENSOR HOUSING
WHT
RED
BLK
BRN
YEL
2
BLK
-
HTR 02
1
ORG
CO REF
+
YEL
+
HTR CO
2
-
1
RED
+
CO ACT
HTR SB
2
WHT
+
EXC
1
BLU
To
Ground
Screw
GRN
7390028
7-3
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Install Sensor Housing
1. Connect the test gas, reference air, eductor air, and dilution air lines to
the sensor housing.
2. Remove the sensor housing cover.
3. If removed, install the power and signal cables and the customer power
and signal conduits and wiring at the sensor housing.
4. Connect the signal cable to O
and CJC terminal blocks, Figure 7-2. Connect the heater power cable to
the HTR terminal blocks.
5. Refer to Figure 7-1 and make sure all test gas lines and electrical
connectors are complete.
6. Install the sensor housing cover.
7. Restore power to the system. Allow OCX to reach normal operating
temperature before installing in a hot process stack.
8. Insert and bolt the sensor housing in the stack and install insulation.
9. Turn on the test gases at the cylinders and open the instrument air
supply valve.
and T/C terminal blocks and to the CO
2
Remove Remote Electronics Housing
1. Turn off power to the system.
2. Shut off the test gases at the cylinders and close the instrument air
supply valve, Figure 7- 1.
3. Disconnect the test gas and instrument air lines from the remote
electronics housing.
4. Remove the cover from the electronics housing to expose the
electronics housing terminal blocks, Figure 7-3.
5. Disconnect and remove the power leads from the AC power input
terminal block. Remove the ground lead from the ground stud.
6. Disconnect and remove the O
signal output terminal block.
7. If used, disconnect and remove the external relay leads from the alarm
output relay terminal block.
8. Disconnect the signal cable leads from O
connector (J4), and from COe and CJC connector (J5), Figure 7-4.
9. Disconnect the heater cable leads from the heater power connector
(J3).
10. If moving the electronics housing to another work site, disconnect and
remove the power and signal cables and customer wiring condu its from
the housing.
11. Remove the remote electronics housing from its mounting and move it
to a suitable work area.
1. Mount remote electronics housing on wall or pipe within distance of
signal and heater cables in use.
2. Remove the electronics housing cover.
3. If removed, install the power and signal cables and the customer power
and signal conduits and wiring at the electronics housing.
4. Connect the signal cable leads to the O
connector (J4), and to the COe and CJC connector (J5), Figure 7-4.
5. Connect the heater cable leads to the heater power connector (J3).
6. Connect the line (L1 wire) to the L1 terminal, and the neutral (N wire) to
the N terminal on the AC power input terminal block, Figure 7-4.
7. Connect the ground lead to the ground stud. Secure the connection with
two nuts. Attach a separate ground lead (G wire) from the ground stud
to the G terminal on the power input terminal block.
8. If used, connect external relay leads to the alarm output relay terminal.
9. Connect the test gas and instrument air lines to the remote electronics
housing. Connect the calibration gas line and instrument air line to the
remote electronics housing.
10. Refer to Figure 7-1 and Figure 7-4. Make sure all test gas lines and
electrical connections are complete.
11. Install the cover on the electronics housing.
12. Turn on the test gasses at the cylinders and open the instrument air
valve.
13. Restore power to the system.
cell and thermocouple
2
7-7
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
REPAIR SENSOR
HOUSING
Sensor Housing
Disassembly
Use the following procedures to remove damaged comp onent s from the OCX
8800 sensor housing and to install new replacement parts. Disassemble the
unit only as needed to replace damaged components. Use the assembly
procedures that apply to install replacement parts and reassemble the unit.
Remove Cover and Terminals Insulator
1. Loosen screw (1, Figure 7-5) and slide locking clip (2) away from cover .
Retighten screw (1).
2. With two hands or strap wrench, turn cover (3) counterclockwise to
loosen. Unthread and remove cover.
3. Inspect cover o-ring (4) for wear or damage. Replace cover o-ring if
damaged.
Figure 7-5. Removal of O2 Cell
and Heater Strut Assembly
1. Screw
2. Locking Clip
3. Cover
4. O-ring
5. Marking Plate
6. Terminal Insulator
7. Reference Air Tube
8. Sensor Housing
9. Heater Strut Assembly
10. Gasket
11. Screw
12. Heater Clamp
13. Heater Rod
14. Insulator
15. Thermal Switch
OCX 8800
3
11
10
4
9
14
5
15
12
6
15
8
13
2
1
7
7
7-9
OCX 8800
Figure 7-6. O2 Cell,
Thermocouple, and
Heater Connections
+
O2
-
Sensor HousingTerminals
ORG
GRN
O CellWires
2
O HeaterWires
2
ThermocoupleWires
RED
YEL
T/C O2
+
T/C CO
+
T/C SB
+
-
EXC
CJC
+
Instruction Manual
IM-106-880, Rev 1.0
January 2007
+
+
++
O Cell and Heater Strut
2
Assembly
Sample
Block
Thermo-
couple
Sample Block
Heater Rods
+
ReturnWire
2
HTR 02
1
2
HTR CO
HTR 02
1
2
HTR SB
1
Remove O
Cell and Heater Strut Assembly
2
CO REF
+
CO ACT
+
+
EXC
Thermal
Switch
1. Remove reference air tube (7, Figure 7-5) from sensor housing (8).
2. See Figure 7-6. Disconnect and tag O
heater wires, O2 cell and return
2
wires, and thermocouple wires at the sensor housing terminals.
3. Remove the O
cell and heater strut assembly (9, Figure 7-5) from
2
sensor housing (8). Remove and discard gasket (10).
Remove Sample Block Heater Rods
1. Disconnect sample block heater rod wires from terminal block. Refer to
Figure 7-6.
37390069
7-10
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 7-7. COe Sensor,
Thermocouple, and Heater
Connections
OCX 8800
2. Loosen screws (1 1, Figur e 7-5) and rotate heater cla mps (12) to release
heater rods (13). One heater clamp secures each heater rod.
3. Slide sample block heater rods (13) out of housing (8).
4. To replace thermal switch (15), remove insulator (14). Disconnect
heater wires. Unscrew and remove thermal switch.
Remove COe Sensor Assembly
1. Disconnect COe heater, thermocouple, and sensor wires from terminal
blocks. Refer to Figure 7-7.
2. Remove insulator (1, Figure 7-8).
O2
+
-
RED
YEL
T/C O2
+
T/C CO
+
Sensor HousingTerminals
HTR 02
HTR CO
HTR 02
HTR SB
CJC Sensor
2
1
2
1
2
1
T/C SB
+
-
EXC
BLU
RED
BLU
WHT
WHT
RED
RED
COe HeaterWires
COe SensorWires
COeThermocouple Wires
CJC
+
CO REF
+
CO ACT
+
+
EXC
NOTE:All wires
at these terminals
are in the CJC
current loop.
COe Sensor
Assembly
37390018
7-11
OCX 8800
Figure 7-8. Removal of
COe Sensor Assembly
Instruction Manual
IM-106-880, Rev 1.0
January 2007
NOTE
For easier access, you may remove two screws from base of terminal block
mounting (13) and move terminal block assembly out of the way.
3. Remove tubes (2, 3, and 4) from COe sensor assembly (5), eductor
fittings (12 and 15), and sensor housing (7).
4. Unfasten bayonet connector of COe thermocouple (8) and remove
thermocouple.
5. Loosen clamp screw of COe band heater (9) until heater rotaters freely
on sensor holder (11).
4
10
3
8
11
12
14
7
5
1
9
2
15
6
13
16
1. Insulator
2. Dilution Air Tube
3. COe Extractive Tube
4. Eductor Air Tube
5. COe Sensor Assembly
6. Eductor
7. Sensor Housing
8. COe Thermocouple
9. Heater Insulator
10. COe Band Heater
11. Sensor Holder
12. Eductor Elbow
13. Terminal Block Mounting
14. Eductor Holder
15. Tube Fitting
16. CJC Sensor
7-12
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 7-9. Alignment of
COe Sensor Assembly
OCX 8800
6. See Figure 7-9. Using straightedge on the sensor holder flat, as shown,
matchmark upper flange of sensor h ousing to show correct alignme nt of
sensor holder.
7. With one wrench holding eductor elbow (12, Figure 7-8), and one
wrench on flats of sensor holder (11), unthread and remove COe sensor
assembly (5). Do not allow eductor elbow to turn.
8. Slide band heater (10) and heater insulator (9) from sensor holder (11).
Matchmark
Straightedge
Sensor
Holder
Flat
Matchmark
37390035
7-13
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Remove Eductor
Figure 7-10. Eductor Alignment
Matchmarks
The O
cell and heater strut assembly (9, Figure 7-5) and the COe sensor
2
assembly (5, Figure 7-8) must be removed before you start this procedure.
1. Use straightedge to matchmark alignment of eductor flat and elbow, as
shown in Figure 7-1 0.
3. Unscrew eductor holder (14) with eductor (6) and fittings (12 and 15)
from sensor housing (7).
4. Clamp flats of eductor (6) in jaws of bench vise.
Use heat resistant gloves when removing mating parts from the eductor. The mating parts
are bonded with a thread sealing compound. The compound softens at 450°F (232°C). The
heated parts can cause severe burns.
5. Use a propane torch to heat the e ductor (6) to 450° F (232°C), minimum.
Apply the heat near the screw threads.
6. While heating the eductor (6), use wrench to apply removal torque to
elbow (12), eductor holder (14), or tube fitting (15) until the thread
sealant softens. Remove mating parts (12, 14, and 15).
7. Use MEK or methylene chloride solvent to clean thread sealant residue
from the pipe threads of the mating parts. Refer to applicable MSDS
sheet for solvent handling precautions.
7-14
Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
Figure 7-1 1. Removal of Sample
and Exhaust Tubes
1
Remove Sample and Exhaust Tubes
Use heat resistant gloves when removing the probe tube or exhaust tube. The tubes are
bonded with a thread sealing compound. The compound softens at 450°F (232°C). The
heated parts can cause severe burns.
1. Secure the sensor housing (1, Figure 7-11) in soft (plastic, wood, or
brass) vice jaws.
2. Use a propane torch to heat the sample tube (2) or exhaust tube (3) to
450°F (232°C), minimum. Apply the heat near the threaded end of the
tube.
3. While heating the tube, use a pipe wrench to apply removal torque to
part being removed. Apply torque until the pipe thread sealant softens.
Remove and discard the used sample tube (2), exhaust tube (3), or
in-situ filter (4).
4. Use MEK or methylene chloride solvent to clean thread sealant residue
from the internal pipe threads in the housing. Refer to applicable MSDS
sheet for solvent handling precautions.
2
1. Sensor Housing
2. Sample Tube
3. Exhaust Tube
4. In Situ Filter
4
3
37390048
7-15
OCX 8800
Figure 7-12. O2 Cell, Heater,
and Thermocouple, Exploded
View
Instruction Manual
IM-106-880, Rev 1.0
January 2007
4
12
11
10
1. Screw
2. O Cell
3. Heater Tube
4. Contact/
5. Gasket
6. Screw
7. Lockwasher
8. Return Wire
9. Heater Strut
10. Spring Clip
11. Spring
12. Strut Bracket
3
Test Gas
Passage
Holes
2
1
6
7
8
9
2
Thermocouple
Assembly
Assembly
5
Do not remove the O2 cell unless you are certain it needs to be replaced. Removal may
damage the cell and platinum pad. Remove the O
Disassemble O
Do not attempt to replace the O
Cell and Heater Strut Assembly
2
cell until all other possibilities for poor
2
cell for cell replacement only.
2
performance have been considered. If cell replacement is needed, order the
cell replacement kit (Refer to Section 8, Replacement Parts).
O
2
The O
cell replacement kit contains an O2 cell and flange assembly , ga skets,
2
socket head cap screws, and anti-seize compound. The items are carefully
packaged to preserve precise surface finishes.
Do not remove items from the package until they are ready to be used.
1. Remove the four allen cap screws (1, Figure 7-12) from the O
Remove the O
cell. The cell flange has a notch that may be used to
2
gently pry the flange away from heater tube (3).
NOTE
The pad on the end of contact/thermocouple assembly (4) will sometimes
fuse to the O2 cell (2).
2. If the O
cell is fused to the contact pad, push the O2 cell back into the
2
heater tube (against spring pressure) and quickly twist the O
cell and contact pad should separate. If the contact pad stays fused to
the cell, a new contact/thermocouple assembly (4) must be installed.
3. Remove and discard gasket (5). Clean the mating surface of heater
tube (3). Remove burrs and raised surfaces with a block of wood and
crocus cloth.
Use care when handling contact and thermocouple assembly. The ceramic rod in this
assembly is fragile.
5. If replacing contact and thermocouple assembly (4), use a pencil to
mark location of spring clip (10) before removing. Squeeze tabs on
spring clip to remove. Retain spring clip and spring (11); replace if
damaged.
6. While carefully handling new contact and thermocouple assembly (4)
lay old assembly next to new one. Transfer match marks to new
37390032
assembly.
7. Carefully guide new contact and thermocouple assembly (4) through
strut bracket (12), spring (11), and spring clip (10) until spring clip
reaches pencil mark.
cell (2).
2
cell. The
2
7-16
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 7-13. COe Sensor,
Exploded View
OCX 8800
Disassemble COe Sensor Assembly
1. Carefully remove screws (1, Figure 7-13), lockwashers (2), and COe
sensor (3) from sensor holder (4). Remove and discard gasket (5).
2. If damaged, use the following procedure to remove thermocouple
adaptor (6) from sensor holder (4):
a. Use a propane torch to heat the thermocouple adaptor to 450°F
(232°C), minimum.
b. While heating, use a flat-head screwdriver to apply removal torque.
Apply torque until the pipe thread sealant softens. Remove and
discard the thermocouple adaptor.
c. Use MEK or methylene chloride solvent to clean thread sealant
residue from the internal pipe threads in the sensor holder. Refer to
applicable MSDS sheet for solvent handling precautions.
1
2
6
5
3
7
9
8
4
1. Screw
2. Lockwasher
3. COe Sensor
4. Sensor Holder
5. Gasket
6. Thermocouple Adapter
7. Pre-Heater
8. Plug
9. Stainless Steel Balls
37390030
7-17
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Always remove the stainless steel balls (approximately 200) from sensor holder before
removing or installing pre-heater. Turning pre-heater in the sensor holder with the stainless
steel balls in place will cause permanent damage to the pre-heater.
3. If pre-heater (7) is to be removed, clamp flat s of sensor holder (4) in vise
jaws with plug (8) pointing up. Remove plug. Unclamp sensor holder
and pour stainless steel balls (9) into a container.
NOTE
Pre-heater should only be removed when pre-heater or sensor holder is
damaged. If removal is not required, leave the pre-heater installed in the
sensor holder.
4. Unthread and remove pre-heater (7).
5. Use a cleaning solvent to thoroughly clean stainless steel balls (9) and
pre-heater chamber in sensor housing (4). Refer to applicable MSDS
sheet for solvent handling precautions.
Sensor Housing
Assembly
Assemble COe Sensor Assembly
Always remove the stainless steel balls (approximately 200) from sensor holder before
removing or installing pre-heater. Turning pre-heater in the sensor holder with the stainless
steel balls in place will cause permanent damage to the pre-heater.
1. If pre-heater (7, Figure 7-13) was removed, apply pipe thread sealant
(Loctite #567) to the external pipe th reads of pre-heater (7) and plu g (8).
Do not apply sealant to the first turn of the pipe threads.
2. Clamp flats of sensor holder (4) in vise jaws with pre-heater port
pointing up.
3. Install and tighten pre-heater (7). Align pre-heater to flat of sensor
holder (4) as shown in Figure 7-14.
4. Invert sensor holder (4, Figure 7-13) in vise and pour stainless steel
balls (9) into plug port. Press down on stainless steel balls and tap
sensor holder with plastic hammer to compact balls in pre-heater
chamber.
5. Install and tighten plug (8).
7-18
Use care when installing the combustibles (COe) sensor. The RTD element s are fragile and
correct alignment in sensor holder is required for proper OCX operation.
6. Lubricate and install COe sensor gasket (5). Apply anti-seize compoun d
to threads of screws (1).
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 7-14. COe Sensor and
Pre-Heater Alignment
THERMOCOUPLE
COe
SENSOR ASSEMBLY
TOP VIEW
Sensor
Holder
OCX 8800
COe Sensor
Flat
Sensor
Holder
Flat
2 to 2-1/4 in.
(51 to 57 mm)
Pre-Heater
Flat
Sensor
Holder
Flat
7. Install COe sensor (3), lockwashers (2), and screws (1). Rotate flat of
COe sensor (3) to center of sensor holder (4).
8. Align COe sensor flat parallel to sensor holder flat, as shown in
Figure 7-14. Tighten screws (1, Figure 7-13).
9. If replacing thermocouple adaptor (6), apply anti-seize to the pipe
threads. Install and tighten thermocouple adaptor.
Assemble O
1. See Figure 7-12. Assemble O
Sensor and Heater Strut Assembly
2
cell (2), gasket (5), and heater tube (3).
2
Make sure the test gas passage holes line up with each other in all
components.
2. Apply a small amount of anti-seize compound to the screw threads and
use screws (1) to secure assembly. Torque to 35 in-lbs (4 N·m).
3. Carefully slide O
heater strut assembly (9) into heater tube (3).
2
4. Press down on the back plate of strut bracket (12) to ensure spring (11)
tension is present to hold contact pad against O
cell (2).
2
5. Secure strut bracket (12) and return wire (8) with four screws (6) and
lockwashers (7). Make sure return wire (8) is tightly fastened. This is the
ground side connection for the O
cell.
2
Install Sample and Exhaust Tubes
37390034
1. See Figure 7-11. Apply pipe thread sealant (Loctite #567) to the
replacement sample tube (2) or exhaust tube (3) pipe threads. Do not
apply sealant to the first turn of the pipe threads.
2. Thread the sample tube (2) or exhaust tube (3) into the ho using (1). Use
a pipe wrench to tighten the tube.
3. If used, install and tighten in-situ filter (4).
7-19
OCX 8800
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Install Eductor
Figure 7-15. Installation of
Eductor and COe Sensor
8
10
11
If installed, the O
cell and heater strut assembly (9, Figure 7-5) must be
2
removed from sensor housing (8), before you install the eductor.
1. Apply pipe thread sealant (Loctite #567) to the external pipe threads of
eductor (6, Figure 7-15). Do not apply sealant to the first turn of the pipe
threads.
4
3
5
1
9
15
6
2
12
14
7
13
16
1. Insulator
2. Dilution Air Tube
3. COe Extractive Tube
4. Eductor Air Tube
5. COe Sensor Assembly
6. Eductor
7. Sensor Housing
8. COe Thermocouple
9. Heater Insulator
10. COe Band Heater
11. Sensor Holder
12. Eductor Elbow
13. Terminal Block Mounting
14. Eductor Holder
15. Tube Fitting
16. CJC Sensor
7-20
37390047
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 7-16. COe Sensor
Parts Alignment
OCX 8800
2. Install and tighten eductor (6) in eductor holder (14).
3. Install and tighten elbow (12) on eductor (6). Male port of elbow must
point up and be in line with long axis of eductor.
4. Apply anti-seize compound to the external pipe threads of eductor
holder (14).
5. Install and tighten eductor holder (14) in sensor housing (7). Align
eductor with matchmarks, as shown in Figure 7-16.
Eductor Flat
Straightedge
Matchmark
37390036
7-21
Instruction Manual
IM-106-880, Rev 1.0
OCX 8800
Install COe Sensor Assembly
Figure 7-17. Band Heater Height1. Apply pipe thread sealant (Loctite #567) to the exposed pipe threads
of eductor elbow (12, Figure 7-15). Do not apply sealant to the first turn
of the pipe threads.
3. With wrenches on eductor elbow (12) and on flats of sensor hol der (11),
tighten sensor holder. Do not allow eductor elbow to turn.
4. Tighten sensor holder (11) to align outside flat with matchmark on
sensor housing flange, as shown in Figure 7-18.
The heater insulator prevents current leakage between the band heater and the sensor
holder. Failure to properly install the insulator may cause the device to trip a ground fault
interrupt circuit.
5. Wrap heater insulator (9) around sensor holder (11). Make sure the
insulator joint lines up with the band gap of the COe band heater (10).
6. Slide COe band heater (10, Figure 7-15) up onto sensor holder (11). Do
not tighten the band heater at this time. Heater must rot ate freely around
sensor holder.
7. Check for proper height of COe heater thermocouple (Figure 7-14).
Thread bayonet connector up or down to adjust height.
8. Install and fasten thermocouple (8, Figure 7-15).
9. Position band heater as shown in Figure 7-17 and Figure 7-18 and
tighten band heater clamp screw . The heater insulator (9) end joint must
line up with the band gap of the COe band heater (10).
0.37 in.
(9,4 mm)
Band Heater
COe Sensor
Insulator
37390058
January 2007
7-22
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 7-18. COe Sensor
Holder Alignment
OCX 8800
Matchmark
Straightedge
Sensor
Holder
Flat
Matchmark
37390035
7-23
OCX 8800
Figure 7-19. COe Sensor,
Thermocouple, and Heater
Connections
Instruction Manual
IM-106-880, Rev 1.0
January 2007
10. Reconnect the COe sensor, thermocouple, and heater wires at the
sensor housing terminal blocks. Refer to Figure 7-19.
1 1. Install and fasten the COe insulator (1, Figure 7-15) around COe sensor
assembly (5). All wiring must remain outside of the insulator.
12. If terminal block mounting (13, Figure 7-15) was moved, reinstall with
two base mounting screws.
O2
Sensor HousingTerminals
HTR 02
HTR CO
HTR 02
HTR SB
+
-
CJC Sensor
2
1
2
1
2
1
RED
YEL
BLU
RED
BLU
WHT
WHT
RED
RED
T/C O2
+
T/C CO
+
T/C SB
+
-
EXC
CJC
+
CO REF
+
CO ACT
+
+
EXC
NOTE:All wires
at these terminals
are in the CJC
current loop.
COe Sensor
Assembly
7-24
COe HeaterWires
COe SensorWires
COeThermocouple Wires
37390018
Instruction Manual
3
390066
IM-106-880, Rev 1.0
January 2007
Figure 7-20. Installation of O2 Cell
and Heater Strut Assembly
1. Screw
2. Locking Clip
3. Cover
4. O-ring
5. Marking Plate
6. Terminal Insulator
7. Reference Air Tube
8. Sensor Housing
9. Heater Strut Assembly
10. Gasket
11. Screw
12. Heater Clamp
13. Heater Rod
14. Insulator
15. Thermal Switch
OCX 8800
3
11
10
4
9
14
5
15
12
6
15
8
13
2
1
7
Install Sample Block Heater Rods
1. Before installing sample block heater rods (13, Figure 7-20), evenly coat
the heater rods with Watlube heater release agent.
7
7-25
OCX 8800
OC
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Figure 7-21. O2 Cell,
Thermocouple, and
Heater Connections
O2
Sensor HousingTerminals
HTR 02
HTR CO
HTR 02
HTR SB
ellWires
2
O HeaterWires
2
ThermocoupleWires
+
+
-
2
1
2
1
2
1
ORG
GRN
RED
YEL
T/C O2
+
T/C CO
+
T/C SB
+
-
EXC
CJC
+
CO REF
+
CO ACT
+
+
EXC
++
O Cell and Heater Strut
2
Assembly
Sample
Block
Thermo-
couple
+
ReturnWire
+
Sample Block
Heater Rods
Thermal
Switch
7-26
37390069
2. If thermal switch (15) was removed, install thermal switch on mating
heater clamp (12), connect wires of heater rods (13), and install
insulator (14).
3. Install the heater rods (13), heater clamps (12), and screws (11).
4. Reconnect the heater rod leads at the sensor housing terminal blocks,
(Figure 7-21).