Rosemount OCX 4400 O2 / Combustibles Transmitter-Rev 2.0 Manuals & Guides

Page 1
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
OXYGEN/COMBUSTIBLES TRANSMITTER
http://www.processanalytic.com
Page 2

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical repre-
sentative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate In- struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s per­formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Emerson Process Management
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Page 3

HIGHLIGHTS OF CHANGES

Effective January, 2003 Rev. 2.0
Page Summary
Throughout Changed IB revision number and release date. Pages 1-4, 1-5 Identified test gas solenoids as optional equipment. Pages 1-8, 1-9 Updated specifications and specification notes. Page 2-1 Expanded drip loop usage to vertical and horizontal installations. Page 2-2 Component view correcti o n s t o Fi g u re 2-1. Page 2-4
Page 2-5 Expanded drip loop usage to vertical and horizontal installations. Page 2-6
Page 2-8 Corrected electrical wiring diagram discrepancies in Figure 2-7. Page 2-9 Revised paragraph 2-3 to link installation instructions to applicable equip-
Page 2-10
Page 2-11 Deleted CAUTION paragraph from NOTE. Page 3-0 Revised Figure 3-1 to correctly identify location of SW1 and SW2. Page 3-1 Revised “Initial Power Up” procedure. Page 7-3 Corrected electrical wiring diagram discrepancies in Figure 7-2. Page 7-4 Corrected electrical wiring diagram discrepancies in Figure 7-3. Page 7-6 Revised Figure 7-4 to clarify location of exhaust tube during assembly and
Page 7-8 Revised Figure 7-5 to include parts availability note. Page 7-11 Revised Figure 7-6 to include parts availability note. Pages 7-12 thru 7-14
Page 8-1 Revised paragraph 8-3 to describe new troubleshooting table. Page 8-2 Replaced fault listings with troubleshooting table. Pages 10-1, 10-2 Corrected discrepancies in Table 10-1, Replacement Parts. Page 11-1 Revised Section 11 list of Appendices A, B, C, and D. Pages A-3 thru A-6 Revised/corrected LOI menu tree displays. Pages B-1 thru B-6 Added SPA setup instructions. Pages C-1, C-2 Revised/clarified blowback installation instructions and illustration. Pages D-1, D-2 Added IS Barrier installation/wiring instructions and illustration.
Revised adapter plate view, Figure 2-3, to simplify recommendations. Added new Figure 2-4 to show electronics housing wall/pole mounting kit.
Added reference to Appendix D for OCX with IS Barrier wiring. Revised alarm relay contact description.
ment options. Revised/changed Figure 2-8 to show pneumatic installation requirements for OCX with reference air option (only).
Revised Figure 2-9 to show pneumatic installation requirements for OCX with reference air and cal gas solenoid options.
to include parts availability note.
Deleted replacement procedures for electronics package components. Added electronics package replacement procedure. Revised Figure 7-7 and fuse replacement procedure.
Page 4
Page 5
OCX 4400
PREFACE............................................................................................................................1
Definitions ............................................................................................................................1
Safety Instructions ...............................................................................................................2
1-0 DESCRIPTION AND SPECIFICATIONS........................................................................ 1-1
1-1 Component Checklist of Typical System (Package Contents) ........................................ 1-1
1-2 System Overview............................................................................................................. 1-1
1-3 OCX 4400 Specifications................................................................................................. 1-8
2-0 INSTALLATION .............................................................................................................. 2-1
2-1 Mechanical Installation .................................................................................................... 2-1
2-2 Electrical Installation ........................................................................................................ 2-6
2-3 Pneumatic Installation ..................................................................................................... 2-9
2-4 Initial Startup.................................................................................................................. 2-11
3-0 STARTUP AND OPERATION ...................................................................................... 3-1
3-1 General ............................................................................................................................ 3-1
Instruction Manual
IB-106-440 Rev. 2.0
January 2003

TABLE OF CONTENTS

4-0 CALIBRATION ................................................................................................................ 4-1
4-1 General ............................................................................................................................ 4-1
4-2 Calibration Methods......................................................................................................... 4-1
5-0 HART/AMS...................................................................................................................... 5-1
5-1 Overview.......................................................................................................................... 5-1
5-2 HART Communicator Signal Line Connections .............................................................. 5-1
5-3 HART Communicator PC Connections ........................................................................... 5-3
5-4 Off-Line and On-Line Operations .................................................................................... 5-3
5-5 Menu Tree for HART Communicator/ OCX 4400 Applications ....................................... 5-3
5-6 HART Communicator Manual O
5-7 HART Communicator Manual COMB CAL Method......................................................... 5-8
5-8 HART Communicator Manual O
5-9 Automatic Calibration Setup Via HART ......................................................................... 5-10
5-10 Line Voltage Selection using HART Communicator...................................................... 5-10
6-0 OPERATION ...................................................................................................................6-1
6-1 Overview.......................................................................................................................... 6-1
6-2 Operation ......................................................................................................................... 6-1
7-0 MAINTENANCE AND SERVICE .................................................................................... 7-1
7-1 Overview.......................................................................................................................... 7-1
7-2 OCX 4400 Removal/Replacement .................................................................................. 7-1
7-3 Sensor Housing Component Replacement ..................................................................... 7-5
7-4 Electronics Housing Component Replacement............................................................. 7-10
CAL Method ............................................................... 5-7
2
and CO CAL Method .................................................. 5-9
2
8-0 TROUBLESHOOTING .................................................................................................... 8-1
8-1 General ............................................................................................................................ 8-1
8-2 Alarm Output.................................................................................................................... 8-1
8-3 Fault Isolation .................................................................................................................. 8-1
Rosemount Analytical Inc. A Division of Emerson Process Management i
Page 6
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
9-0 RETURN OF MATERIAL .............................................................................................. 9-1
10-0 REPLACEMENT PARTS ............................................................................................. 10-1
11-0 APPENDICES ............................................................................................................... 11-1
Appendix A Local Operator Interface (LOI) ................................................................... A-1
Appendix B SPA with HART Alarm ...............................................................................B-1
Appendix C Blowback....................................................................................................C-1
Appendix D IS Barrier ....................................................................................................D-1
12-0 INDEX............................................................................................................................ 12-1
OCX 4400
ii Rosemount Analytical Inc. A Division of Emerson Process Management
Page 7
OCX 4400
Figure 1-1. Typical System Package ....................................................................................... 1-2
Figure 1-2. System Operation Diagram ................................................................................... 1-5
Figure 1-3. OCX 440 HART Connections and AMS Application.............................................. 1-6
Figure 1-4. Typical System Installation .................................................................................... 1-7
Figure 2-1. OCX 4400 Installation (with Integral Electronics) .................................................. 2-2
Figure 2-2. OCX 4400 Adapter Plate Dimensions ................................................................... 2-3
Figure 2-3. Adapter Plate Installation ....................................................................................... 2-4
Figure 2-4. Wall or Pole Mounting of Electronics Housing....................................................... 2-4
Figure 2-5. Installation with Drip Loops .................................................................................... 2-5
Figure 2-6. Line Voltage and 4-20 mA Connections ................................................................ 2-7
Figure 2-7. Electrical Connections Between Split-Electronics and Sensor Housing ............... 2-8
Figure 2-8. Pneumatic Installation, OCX with Reference Air Set (without Autocalibration) .... 2-9
Figure 2-9. Pneumatic Installation, OCX with Reference Air Set and Solenoids (with
Figure 3-1. OCX 4400 Defaults ................................................................................................ 3-0
Figure 5-1. Signal Line Connections, > 250 Ohms Lead Resistance ...................................... 5-1
Figure 5-2. Signal Line Connections, < 250 Ohms Lead Resistance ...................................... 5-2
Figure 5-3. Menu Tree for HART/AMS on the OCX 4400 (Sheet 1 of 3)................................. 5-4
Figure 7-1. Electronics Housing Terminal Blocks .................................................................... 7-2
Figure 7-2. Sensor Housing Terminal Blocks .......................................................................... 7-3
Figure 7-3. Split-Electronics Housing Cable Connections ....................................................... 7-4
Figure 7-4. Sensor Housing, Exploded View ........................................................................... 7-6
Figure 7-5. O
Figure 7-6. Electronics Housing Exploded View .................................................................... 7-11
Figure 7-7. Fuses, Exploded View ......................................................................................... 7-12
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
LIST OF ILLUSTRATIONS
Autocalibration) .................................................................................................... 2-10
Assembly, Exploded View ................................................................................ 7-8
2

LIST OF TABLES

Table 1-1. OCX 4400 Specifications ....................................................................................... 1-8
Table 1-2. Product Matrix......................................................................................................1-10
Table 1-3. Accessories.......................................................................................................... 1-11
Table 8-1. Troubleshooting ..................................................................................................... 8-2
Table 10-1. Replacement Parts .............................................................................................. 10-1
Rosemount Analytical Inc. A Division of Emerson Process Management iii
Page 8
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
iv Rosemount Analytical Inc. A Division of Emerson Process Management
Page 9
OCX 4400
The purpose of this manual is to provide a comprehensive understanding of the OCX 4400 Oxygen/Combustibles Transmitter components, functions, installation, and maintenance.
This manual is designed to provide information about the OCX 4400. We recommend that you thoroughly familiarize yourself with the Introduction and Installation sections before installing your transmitter.
The introduction presents the basic principles of the transmitter along with its performance characteristics and components. The remaining sections contain detailed procedures and information necessary to install and service the transmitter.
Before contacting Rosemount concerning any questions, first consult this manual. It de­scribes most situations encountered in your equipment’s operation and details necessary action.
Instruction Manual
IB-106-440 Rev. 2.0
January 2003

PREFACE

DEFINITIONS
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
NOTE
Highlights an essential operating procedure, condition, or statement.
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETIN
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illus­tration number. It is not a part number, and is not related to the illustration in any technical manner.
Rosemount Analytical Inc. A Division of Emerson Process Management P-1
Page 10
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
FOR THE WIRING AND INSTALLATION
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non­EU states should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external,
where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced.
The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
OCX 4400
IMPORTANT

SAFETY INSTRUCTIONS

OF THIS APPARATUS
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable
anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be
made through a circuit breaker which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker may also include a mechanically operated isolating switch. If not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazard-
ous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment — and then only by trained service personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a
danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Cer­tain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to
the Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the follow-
ing standards: EN61010-1, IEC417, and ISO3864.
P-2 Rosemount Analytical Inc. A Division of Emerson Process Management
Page 11
OCX 4400
1
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 1

DESCRIPTION AND SPECIFICATIONS

1-1 COMPONENT CHECKLIST OF TYPICAL
SYSTEM (PACKAGE CONTENTS)
A typical Rosemount OCX 4400 Oxygen/ Combustibles Transmitter package should con­tain the items shown in Figure 1-1. Record the part number, serial number, and order number for each component of your system in the table located on the back cover of this manual.
Use the product matrix in Table 1-2 at the end of this section to verify your order number. The first part of the matrix defines the model. The last part defines the various options and fea­tures of the OCX 4400. Check the model num­ber against the transmitter features and options, making sure options specified by this number are on or included with the unit. Use this com­plete model number for any correspondence with Rosemount.

1-2 SYSTEM OVERVIEW

a. Scope
This Instruction Bulletin supplies details needed to install, startup, operate, and maintain the Rosemount OCX 4400 Oxy­gen/ Combustibles Transmitter. Signal con­ditioning electronics outputs separate 4-20 mA signals representing oxygen (O combustibles (COMB) values. This informa­tion, plus additional details, can be ac­cessed with the HART Model 275 handheld communicator or Asset Management Solu­tions (AMS) software. An optional local op­erator interface (LOI) is available to provide local communication with the electronics.
b. System Description
The Rosemount OCX 4400 Oxygen/ Combustibles Transmitter is designed to measure oxygen and combustible concen­trations in flue gas temperatures up to
) and
2
2600°F (1427°C). The transmitter is close coupled to the process and requires mini­mal sample conditioning requirements.
The equipment measures oxygen percent­age by reading the voltage developed across a heated electrochemical cell, which consists of a small yttria-stabilized, zirconia disc. Both sides of the disc are coated with porous metal electrodes. When operated at the proper temperature, the millivolt output of the cell is given by the following Nernst equation:
EMF = KT log10 (P
Where:
1. P
is the partial pressure of the oxygen in
2
the measured gas on one side of the cell.
2. P
is the partial pressure of the oxygen in
1
the reference air on the opposite side of the cell.
3. T is the absolute temperature.
4. C is the cell constant.
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry in­strument air (20.95% oxygen) as the reference air.
When the cell is at operating temperature and there are unequal oxygen concentra­tions across the cell, oxygen ions will travel from the high oxygen partial pressure side to the low oxygen partial pressure side of the cell. The resulting logarithmic output voltage is approximately 50 mV per decade. The output is proportional to the inverse logarithm of the oxygen concentration. Therefore, the output signal increases as the oxygen concentration of the sample gas decreases. This characteristic enables the OCX 4400 to provide exceptional sensitivity at low oxygen concentrations.
1/P2
) + C
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Page 12
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
5
HART
MAN4275A00
October1994
Communicator
o
FISHER-ROSEMOUNT
OCX 4400
1
English
TM
4
3
1. Instruction Bulletin
2. HART
Communicator Package (Optional)
3. Adapter Plate with Mounting Hardware and Gasket
4. OCX 4400 (Shown with Integral Electronics)
5. Reference Air and Calibration Set
6. Blowback Hardware (Optional)
Figure 1-1. Typical System Package
2
6
37020025
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 13
OCX 4400
1
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
The OCX 4400 measures net oxygen con­centration in the presence of all the prod­ucts of combustion, including water vapor. Therefore, it may be considered an analysis on a “wet” basis. In comparison with older methods, such as the portable apparatus, which provides an analysis on a “dry” gas basis, the “wet” analysis will, in general, in­dicate a lower percentage of oxygen. The difference will be proportional to the water content of the sampled gas stream.
The OCX 4400 combustibles sensor is a catalytic sensor consisting of two Resistant Temperature Detectors (RTD). One RTD is the reference element covered with an inert coating. The other RTD element is active, coated with a catalyst. As the sample gases flow by the sensor, the combustible gases oxidize on the surface of the active element. The oxidation that occurs produces heat and a temperature rise in the active ele­ment. The temperature difference produces a resistance relationship between the two elements that is directly proportional to the concentration of combustibles in the sample gases.
The catalyst is specifically designed to de­tect carbon monoxide (CO), but the sensor responds to other combustible gases. The sensor is calibrated using carbon monoxide (CO), thus the output should be expressed in terms of CO. However, since the sensor detects other combustible gases, the output cannot just be labeled CO. The response of the sensor to other combustible gases gives an output that is equivalent to the sensor detecting CO. The term COMB is used in this manual to describe the sensor output. This term indicates that the sensor is cali­brated in terms of CO, and that the sensor output is equivalent to CO but not specific to CO.
c. System Configuration
Transmitters are available in four length op­tions, giving the user the flexibility to use a penetration appropriate to the size of the stack or duct. The length options are 18 in. (457 mm), 3 ft (0.91 m), 6 ft (1.83 m), or 9 ft (2.7 m). Probes are available in three mate­rial options, 316L stainless steel, inconel 600, and ceramic to accommodate higher temperatures.
The electronics are contained in a separate housing from the sensors and may be inte­grally mounted next to the sensor housing on the probe, or split mounted up to 150 feet away from the probe.
The electronics control both sensor tem­peratures and provide individual 4-20 mA isolated outputs that are proportional to the measured oxygen and combustibles con­centrations. The power supply can accept voltages of 100 to 240 VAC and 50 to 60 Hz. The electronics accepts millivolt signals generated by the sensors and produces the outputs to be used by remotely connected devices. The outputs are isolated 4-20 mA linearized currents. See 3-1c and 3-1d for specific instructions upon initial power up.
d. System Features
1. The O tivity increase as the oxygen concen­tration decreases.
2. HART communication is standard. To use the HART capability, you must have either:
(a) A HART Model 275
cell output voltage and sensi-
2
Communicator.
Dilution air is provided to the COMB sensor to ensure there is adequate oxygen to fully oxidize any combustible gases regardless of the concentration of oxygen in the process.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
(b) Asset Management Solutions
(AMS) software for the PC.
3. Oxygen cell and heater/thermocouple assembly are field replaceable.
Page 14
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
4. Electronics are automatically config­ured for line voltages from 100 to 240 VAC.
5. An operator can calibrate and diagnos­tically troubleshoot the OCX 4400 in one of two ways:
(a) HART Interface. Each of the OCX
4400’s 4-20 mA output lines transmit an analog signal propor­tional to oxygen or combustible levels detected. The HART output is superimposed on the oxygen 4­20 mA output line only. This infor­mation can be accessed through the following:
1 Rosemount Model 275 Hand-
held Communicator - The handheld communicator re­quires Device Description (DD) software specific to the OCX 4400. The DD software will be supplied with many Model 275 units, but can also be programmed into existing units at most Fisher­Rosemount service offices. Refer to Section 5, HART/AMS, for additional information.
2 Personal Computer (PC) -
The use of a personal com­puter requires AMS software available from Fisher­Rosemount.
3 Selected Distributed Control
Systems - The use of distrib­uted control systems requires input/output (I/O) hardware and AMS software which permit HART communica­tions.
(b) Optional Local Operator Interface
(LOI). The optional LOI is mounted to the end of the electronics mod­ule and allows local communica­tions with the electronics. Refer to Appendix A, Local Operator Inter­face (LOI), for more information.
6. Optional Blowback System. The blow­back system periodically blows instru­ment air back through the inlet filter and out the inlet tube. This clears out particulate and keeps the inlet filter from clogging.
e. System Operation
Figure 1-2 shows the relationship between the components of the OCX 4400. The sen­sors and the electronics are contained in separate housings. The sensor housing and probe mounts to a duct or process wall so that the probe protrudes into the flue gas stream. An air powered eductor continu­ously pulls samples of the process flue gas through the probe to a chamber in front of the sensor housing where the sample passes the O
sensor and continues on to
2
the COMB sensor. Dilution air is provided to the COMB sensor and reference air to the O
sensor. After the gas sample flows past
2
the O
sensor and through the COMB sen-
2
sor, it is drawn through the eductor where it mixes with the eductor air and exits through exhaust back into the system. The elec­tronics housing contains the CPU which converts the sensor inputs into 4-20 mA analog output signals. The CPU can also initiate and perform calibrations. Three test gasses and instrument air can be turned on and off by solenoids. Test gas flow to the sensors is regulated by a rotometer be­tween the electronics and sensor housings. Instrument air is separated into eductor air, reference air, and dilution air. The instru­ment air solenoid does not allow air flow until the heaters are up to temperature. This minimizes the amount of sampled process flue gas being pulled into cold sensors causing condensation.
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 15
OCX 4400
1
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SAMPLE
GAS
PROBE
EXHAUST
SENSOR
HOUSING
COMBUSTIBLES
SENSOR
O2
SENSOR
EDUCTOR
ROTOMETER
5 SCFH
ELECTRONICS
PRESSURE
SENSOR
OPTIONAL TEST GAS
SOLENOIDS
INSTRUMENT AIR
SOLENOID
EDUCTOR AIR
HOUSING
CPU
POWER SUPPLY
LOW O2
TEST GAS
HIGH O2
TEST GAS
HIGH CO
TEST GAS
INSTRUMENT
AIR
REFERENCE AIR
DILUTION AIR
ROTOMETER
50 CC/MIN. (0.1 SCFH)
Figure 1-2. System Operation Diagram
37020001
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Page 16
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
f. Handling the OCX 4400
It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to pre­vent possible equipment damage.
The OCX 4400 is designed for indus­trial application. Treat each compo­nent of the system with care to avoid physical damage. The probe may con­tain components made from ceramics, which are susceptible to shock when mishandled.
g. System Considerations
Prior to installing your Rosemount OCX 4400 Oxygen/Combustibles Transmitter, make sure you have all the components necessary to make the system installation. Ensure all the components are properly in­tegrated to make the system functional.
After verifying that you have all the compo­nents, select mounting locations and deter­mine how each component will be placed in terms of available line voltage, ambient temperatures, environmental considera­tions, convenience, and serviceability. Figure 1-3 shows a typical system wiring. A typical system installation is illustrated in Figure 1-4.
A source of instrument air is required at the OCX 4400 for reference air, dilution air, and eductor air. Since the OCX 4400 is equipped with an in-place calibration fea­ture, provision should be made for con­necting test gas tanks to the OCX 4400 when it is to be calibrated.
NOTE
The electronics module is designed to meet NEMA 4X (IP55) and the elec­tronic components are rated to tem­peratures up to 185°F (85°C).
Retain packaging in which the unit ar­rived from the factory in case any components are to be shipped to an­other site. This packaging has been designed to protect the product.
INSTRUMENT
AIR
OCX 4400 O2/COMBUSTIBLES TRANSMITTER
WITH INTEGRAL ELECTRONICS
4-20 MA OUTPUT (TWISTED PAIRS)
3 CALIBRATION GAS LINES BY CUSTOMER [300 FT (90M) MAX]
Figure 1-3. OCX 440 HART Connections and AMS Application
TERMINATION IN
CONTROL ROOM
HART MODEL 275
HANDHELD INTERFACE
CUSTOMER’S LAPTOP
WITH ASSET MANAGEMENT
ASSET MANAGEMENT SOLUTIONS
SOLUTIONS
34070002
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 17
OCX 4400
1
INTEGRAL ELECTRONICS
TEST
GAS
ROTOMETER
GASES
STACK
DILUTION AIR
ROTOMETER
ADAPTER
PLATE
DUCT
Instruction Manual
4-20 mA OUTPUTS (2)
LINE VOLTAGE
INSTRUMENT AIR
SUPPLY (REF. GAS)
PRESSURE
REGULATOR
SPAN COMBUSTIBLES TEST GAS LOW O2 TEST GAS
HIGH O2 TEST GAS
IB-106-440 Rev. 2.0
January 2003
ELECTRONICS
CABLE
(UP TO 150FT)
SPLIT-MOUNTED ELECTRONICS
GASES
DUCT
STACK
TEST
GAS
ROTOMETER
DILUTION AIR
ROTOMETER
4-20 mA OUTPUTS (2)
LINE VOLTAGE
INSTRUMENT AIR
SUPPLY (REF. GAS)
PRESSURE
REGULATOR
SPAN COMBUSTIBLES TEST GAS LOW O2 TEST GAS
HIGH O2 TEST GAS
34070003
Figure 1-4. Typical System Installation
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
Page 18
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400

1-3 OCX 4400 SPECIFICATIONS

transmitter before choosing the mounting loca-
tion. Table 1-2 contains the product matrix. Use Table 1-1 contains information about the OCX 4400 Oxygen/Combustibles Transmitter oper­ating characteristics. Use the table to make sure that available conditions are suitable for the
the product matrix for ordering information and
specific model number information. Table 1-3
contains information on accessories available
with the OCX 4400.
Table 1-1. OCX 4400 Specifications
Net O
Range .................................................. 0-1% to 0-40% O
2
, Fully field selectable
2
Combustibles ................................................. 0-1000 ppm to 0-5%, Fully field selectable
Accuracy:
Oxygen..................................................... +0.75% of reading or 0.05% O
, whichever is greater
2
Combustibles .......................................... +2% F.S.
System Response to Test Gas:
Oxygen..................................................... 10 sec T90
Combustibles .......................................... 25 sec T90
Temperature Limits:
Process .................................................... 32° to 2600°F (0° to 1427°C)
Sensors Housing .................................... 32° to 392°F (0° to 200°C)
Electronics Housing ............................... 32° to 185°F (0° to 85°C)
2
2
Nominal and Approximate Shipping Weights:
18 in. (457 mm) package ........................ 54 lbs (24.5 kg)
3 ft (0.91 m) package .............................. 55 lbs (24.5 kg)
6 ft (1.83 m) package .............................. 57 lbs (26 kg)
9 ft (2.74 m) package .............................. 59 lbs (26.8 kg)
1
Mounting and Mounting Position:
Sensors.................................................... Flange
Electronics............................................... Wall/Pipe
Materials:
Probes...................................................... 316L stainless steel
Inconel 600 Ceramic
Enclosures............................................... Low-copper aluminum
Calibration ...................................................... Semi-automatic or automatic
Calibration Gas Mixtures Recommended.... 0.4% O
(Ref. test gas bottles kit #1A99119G04) 8% O2, Balance N
1000 ppm CO, Balance N
NOTES:
(1)
Combustibles sensor is calibrated with CO gas. Sensor may be calibrated with H2 gas. Use in methane service is not rec­ommended.
(2)
Operating temperature of electronics inside of electronics housing, as measured by a HART Communicator, or Emerson Process Management Asset Management Solutions (AMS) Software, or local operator interface. Operation below 0°C (32°F) requires that sensor housing be insulated.
, Balance N
2
2
2
2
1-8 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 19
OCX 4400
1
Table 1-1. OCX 4400 Specifications (Continued)
Calibration Gas Flow..................................... 5 scfh (2.5 l/m)
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
Reference Air.................................................. 2 scfh (1 l/m), clean, dry instrument-quality air (20.95% O
regulated to 35 psi (241 kPa)
Eductor Air...................................................... 5 scfh (2.5 l/m), clean, dry, instrument-quality air
(20.95% O
), regulated to 35 psi (241 kPa)
2
Dilution Air...................................................... 0.1 scfh (0.5 l/m), clean, dry, instrument-quality air
(20.95% O
Blowback Air (Optional) ................................ Clean, dry, instrument-quality air (20.95% O
), regulated to 35 psi (241 kPa)
2
),
2
regulated to 55 psi (374 kPa)
Sensors Housing ........................................... NEMA 4X, IP55 with fitting and pipe on reference exhaust
port to clean, dry atmosphere, two 3/4-14 NPT conduit ports
Electronics Housing ...................................... NEMA 4X, IP55 with fitting and pipe on reference exhaust
port to clean, dry atmosphere, two 3/4-14 NPT conduit ports
Electrical Noise .............................................. Meets EN 50082-2 Electromagnetic Compatibility Generic
Immunity Std., Part II Includes ENG 1000 4-R for Electrostatic Discharge 4 Kv contact, 8 Kv in air Optionally ENG 1000 4-R “Namur-Increased” 8 Kv contact, 16 Kv in air Includes IEC 801-4 fast transients-2 Kv on power supply and control lines
Line Voltage.................................................... Universal 100 to 240 VAC, 50 to 60 Hz,
software selectable, 3/4-14 NPT conduit port
),
2
Isolated Output:
Oxygen..................................................... 4-20 mAdc, 950 ohm max, with HART capability for O
COMB
Combustibles.......................................... 4-20 mAdc, 950 ohm max
Alarm Output Relay ....................................... No. 2, form C, double-pole, double-throw, 2A, 30 Vdc,
(software assignable)
Logic Signals.................................................. SPA HART Alarm Module (optional)
Low Oxygen Alarm
High Combustibles Alarm Calibration Status Unit Failure
Power Consumption Limits:
Heaters..................................................... 650 W nominal max
Electronics............................................... 50 W nominal max
Certification.................................................... CE
FM(Pending) CSA(Pending) CENELEC(Pending)
NOTE:
All static performance characteristics are with operating variables constant. Specifications subject to change without notice.
and
2
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-9
Page 20
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
Table 1-2. Product Matrix
OCX44A O2 / Combustibles Transmitter
O2 / Combustibles Transmitter IB-106-440
Code Probe Length and Material
00 No Probe or Exhaust Tube 11 18 in. (457 mm) 316 SST up to 1300°F (704°C) 21 18 in. (457 mm) Inconel 600 up to 1832°F (1000°C) 31 18 in. (457 mm) Ceramic up to 2600°F (1426°C) 12 3 ft (0.91 m) 316 SST up to 1300°F (704°C) 22 3 ft (0.91 m) Inconel 600) up to 1832°F (1000°C) 32 3 ft (0.91 m) Ceramic up to 2600°F (1426°C) 13 6 ft (1.83 m) 316 SST up to 1300°F (704°C) 23 6 ft (1.83 m) Inconel 600 up to 1832°F (1000°C) 14 9 ft (2.7 m) 316 SST up to 1300°F (704°C)
Code Probe Mounting Assembly
10 (ANSI 2 in. 150 lb) 6" dia. flange, 4.75" BC with 4 x 0.75" dia. holes 20 (DIN) 185mm dia. flange, 145mm BC with 4 x 18mm dia. holes 30 (JIS) 155mm dia. flange, 130mm BC with 4 x 15mm dia. holes
OCX 4400
Code Mounting Hardware - Stack Side
0 No Adapter Plate ("0" must also be chosen under "Mounting Adapter - Probe Side" below) 1 New Installation - Square weld plate with studs 2 Model 218/240 Mounting Plate (with Model 218/240 Shield Removed) 3 Existing Model 218/240 Support Shield 4 Mounting to Other Mounting 5 Model 132 Adapter Plate 6 Mounting to Ametek WDG-IVC 7 Mounting to Servomex Xendos 2700 8 Mounting to Elsag Bailey SMA-90
Code Mounting Hardware - Probe Side
0 No Adapter Plate 1 Probe Only (ANSI) 4 Probe Only (DIN) 7 Probe Only (JIS)
OCX44A 11 10 1 1 (Cont’d) Example
(1)
1-10 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 21
OCX 4400
1
Table 1-2. Product Matrix (Continued)
Cont'd Code Electronic Housing - NEMA 4X, IP55 HART Communications
01 Basic Unit 02 Local Operator Interface 03 Calibration Solenoids 04 IS Barrier 05 Local Operator Interface and Calibration Solenoids 06 Local Operator Interface and IS Barrier 07 Calibration Solenoids and IS Barrier 08 Local Operator Interface and Calibration Solenoids and IS Barrier
Code Electronic Mounting
3
01 Integral to Sensor Housing Electronics 02 Split Electronics and no cable 03 Split Electronics and 20 ft (6m) cable 04 Split Electronics and 40 ft (12m) cable 05 Split Electronics and 60 ft (18m) cable 06 Split Electronics and 80 ft (24m) cable 07 Split Electronics and 100 ft (30m) cable 08 Split Electronics and 150 ft (45m) cable
2
3
3
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
3
Code Accessories
00 None 01 Cal. Gas Rotometers, Flow Rotometers, and Ref. Gas Set 11 Blowback Option and Cal. Gas Rotometers, Flow Rotometers, and Ref. Gas Set
Cont’d 08 01 11 Example
NOTES:
(1)
Where possible, specify SPS number; otherwise, provide details of the existing mounting plate as follows:
Plate with studs Bolt circle diameter, number, and arrangement of studs, stud thread, stud height above mounting plate. Plate without studs Bolt circle diameter, number, and arrangement of holes, thread, depth of stud mounting plate with accessories.
(2)
If the Local Operator Interface (LOI) is not implemented, remote access and functionality available via HART Communica­tions (Model 275 Handheld Communicator) with Oxygen/Combustibles Device Description (DD) required.
(3)
If the IS Barrier option is selected, customer must also purchase the Moore Industries SPA (1A99292H01), current repeater backplane and adpater (intrinsic safety barrier kit p/n 1A99698G01) to maintain the 4 – 20 mA combustibles output.
Table 1-3. Accessories
PART NUMBER DESCRIPTION
1A99119H01 Oxygen Bottle 0.4%, balance N
1A99119H02 Oxygen Bottle 8.0%, balance N
1A99119H06 CO Bottle 1000 PPM and air, balance N
2
2
2
1A99120H02 Regulator for Oxygen (may need 2)
1A99120H03 Regulator for COMB
1A99119G06 Wall mount bracket for gas bottles
1A99119G05 Test gas regulators kit
1A99119G04 Test gas bottles kit
1A99292H01
Moore Industries SPA for Low O
Alarm, High
2
COMB Alarm, Calibration Status, and Unit Fail
275D9EI5D0000 275 HART Communicator with 12 Meg buffer
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-11
Page 22
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
1-12 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 23
OCX 4400
2
Before installing this equipment, read the “Safety instructions for the wiring and installation of this apparatus” at the front of this Instruction Bulletin. Failure to follow the safety instruc­tions could result in serious injury or death.
2-1 MECHANICAL INSTALLATION
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 2

INSTALLATION

readings. Therefore, either make the necessary repairs or install the trans­mitter upstream of any leakage.
3. Ensure the area is clear of internal and external obstructions that will interfere with installation and maintenance ac­cess to the unit. Allow adequate clear­ance for the removal of the OCX 4400.
a. Selecting Location
1. The location of the OCX 4400 in the stack or flue is most important for maximum accuracy in the oxygen analyzing process. The probe must be positioned so the gas it measures is representative of the process. Best re­sults are normally obtained if the transmitter is positioned near the cen­ter of the duct (40-60% insertion). Longer ducts may require several transmitters since the oxygen and combustibles can vary due to stratifica­tion. A point too near the wall of the duct, or the inside radius of a bend, may not provide a representative sam­ple because of the very low flow condi­tions. The sensing point should be selected so the process gas tempera­ture falls within the range of probe material used. Figure 2-1 through Figure 2-5 provide mechanical installa­tion references. The ambient tem­perature inside the electronics housing must not exceed 185°F (85°C).
2. Check the flue or stack for holes and air leakage. The presence of this con­dition will substantially affect the accu­racy of the oxygen and combustibles
Do not allow the temperature of the electronics housing to exceed 185°F (85°C) or damage to the electronics may result.
b. Installation
1. Ensure all components are available to install the OCX 4400.
2. The OCX 4400 may be installed intact as it is received.
3. Weld or bolt adapter plate (Figure 2-3) onto the duct.
4. For systems with split electronics, use the pole or wall mounting hardware as shown in Figure 2-4 to mount the elec­tronics housing. Choose a location not to exceed the length of the electronics cable ordered.
5. For integral electronics installations, ensure the system cable drops verti­cally from the OCX 4400 and the con­duit is routed below the level of the housings. Drip loops minimize the pos­sibility that moisture will damage the electronics (Figure 2-5).
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Page 24
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
NOTE
If process temperatures will exceed 392°F (200°C), use anti-seize com­pound on stud threads to ease future removal of the OCX 4400.
6. Insert sample and exhaust tubes through the opening in the mounting flange and bolt the unit to the flange.
NOTE:
ALL DIMENSIONS ARE IN INCHES WITH MILLIMETERS IN PARENTHESES.
INSULATE IF EXPOSED TO AMBIENT WEATHER CONDITIONS
Uninsulated stacks or ducts may cause ambient temperatures in the electronics housing to exceed 185°F (85°C) and damage the electronics.
7. If insulation is removed to access the duct for OCX 4400 mounting, make sure to replace insulation afterward.
0.062 THK GASKET
3535B18H02
ANSI
3535B46H01
JIS
3535B45H01
DIN
TABLE 2. INSTALLATION/REMOVAL
PROBE
18 IN.
3 FT
6 FT
9 FT
DIM "A"
(406)
(864)34(1280)
(1778)
106
(2692)
DIM "B"
16
70
(2184)
122.38
(3108)
32.38
(822)
50.38
86.38
Figure 2-1. OCX 4400 Installation (with Integral Electronics)
TOP VIEW
DIM “A”DIM “B”
FLANGE DIA
HOLE DIA
(4) HOLES EQ SP ON B.C. DIA.
TABLE 1. MOUNTING FLANGE
ANSI
4512C17H01
6.00
(153)
0.75
(20)
4.75
(121)
DIN
4512C19H01
7.28
(185)
0.71
(18)
5.71
(145)
JIS
4512C18H01
6.10
(155)
0.59
(15)
5.12
(130)
37020002
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 25
OCX 4400
2
DIMENSIONS
IN.
(mm)
ADAPTER PLATE* DIMENSIONS FOR OCX 4400
ANSI
(P/N 4512C34G01)
DIN
(P/N 4512C36G01)
(P/N 4512C35G01)
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
JIS
"A"
"B"
THREAD
"C"
DIA
*PART NUMBERS FOR ADAPTER PLATES INCLUDE
ATTACHING HARDWARE.
NOTE: DIMENSIONS ARE
IN INCHES WITH MILLIMETERS IN PARENTHESES.
2.500 DIA (63.5)
ADAPTER PLATE FOR OCX 4400 INSTALLATION. SEE FIGURE 2-1.
6.00
(153)
0.625-11
4.75
(121)
B
7.5
(191)
(M-16 x 2)
5.708 (145)
A
o
45
C
6.50
(165)
(M-12 x 1.75)
5.118 (130)
A
4 STUDS, LOCKWASHERS AND NUTS EQUALLY SPACED ON C DIA B.C.
29790013
Figure 2-2. OCX 4400 Adapter Plate Dimensions
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Page 26
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
INSTALLATION FOR METAL
WALL STACK OR DUCT
NOTE: ONLY ADAPTER PLATE IS FURNISHED BY ROSEMOUNT.
WELD OR BOLT ADAPTER
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIRTIGHT.
DIMENSIONS IN INCHES WITH MILLIMETERS IN PARENTHESES.
2.50 [63.5] MIN DIA HOLE IN WALL
STACK OR DUCT
METAL WALL
Figure 2-3. Adapter Plate Installation
INSTALLATION FOR MASONRY
WALL STACK CONSTRUCTION
BOLT ADAPTER
PLATE TO OUTSIDE
WALL SURFACE
JOINT MUST
BE AIRTIGHT
MASONRY
STACK WALL
FIELD WELD
PIPE TO
ADAPTER PLATE
PIPE 3.00
SCHEDULE 40,
SLEEVE LENGTH
OPTIONAL
37020003
NOTE:
* INDICATES ITEMS ARE INCLUDED
IN MOUNTING KIT P/N 4851B40G01.
*POLE
MOUNTING
CLAMP
*WALL OR POLE
MOUNT BRACKET
AND SCREWS
1/4-20 UNC X 2 IN. M-F STANDOFFS
RECOMMENDED FOR REAR PIPING
CLEARANCE WHEN WALL MOUNTING
(4 EACH BY CUSTOMER)
WALL MOUNTING BOLTS OR LAG SCREWS (4 EACH BY CUSTOMER)
Figure 2-4. Wall or Pole Mounting of Electronics Housing
ELECTRONICS
HOUSING
37020021
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 27
OCX 4400
2
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
CONDUIT
DRIP LOOPS
CONDUIT
DRIP LOOP
Figure 2-5. Installation with Drip Loops
DUCT
WALL
CONDUIT
DRIP LOOP
DUCT
WALL
37020004
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Page 28
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400

2-2 ELECTRICAL INSTALLATION

All wiring must conform to local and national codes. For reference, factory wired solenoid power connections are shown in Figure 2-6.
Disconnect and lock out power before connecting the unit to the power supply.
Install all protective equipment covers and safety ground leads after installa­tion. Failure to install covers and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation of this equipment, con­nection to the main electrical power supply must be made through a circuit breaker (min 10 A) which will discon­nect all current-carrying conductors during a fault situation. This circuit breaker should also include a me­chanically operated isolating switch. If not, then another external means of disconnecting the supply from the equipment should be located close by. Circuit breakers or switches must comply with a recognized standard such as IEC 947.
NOTE
To maintain CE compliance, ensure a good connection exists between the mounting flange bolts and earth.
a. Connect Line Voltage
The OCX 4400 is software configured for 100 to 240 VAC line voltage at 50 to 60 Hz. The power supply requires no setup. Con­nect the line, or L wire, to the L terminal, and the neutral, or N wire, to the N terminal on the AC power input terminal block in the electronics housing. Connect the ground, or G wire, to the ground stud in the electronics
housing. Secure the connectio n with two nuts. Attach a separate line from the ground stud to the G terminal on the power input terminal block.
NOTE
Make sure line voltage is correctly set. See paragraph 3-1c for instructions.
b. Connect 4-20 mA Signals
Connect the 4-20 mA current loop to the 4-20 mA signal output terminals in the elec­tronics housing. If the isolation barrier
option is provided, refer to Appendix D for alternate wiring instructions.
1. Oxygen (O
) 4-20 mA Signal. One
2
4-20 mA signal represents the O value. Superimposed on the O2 signal is the HART information accessible through a Model 275 Handheld Com­municator or AMS software. The O signal is at the AOUT 1 terminals.
2. Combustibles (COMB) 4-20 mA Signal. Another 4-20 mA signal at the AOUT 2 terminals represents the COMB value. HART information is not available on the COMB signal.
c. Alarm Output Relay
Connect any customer supplied relay input to the alarm output relay terminal. The alarm output relay terminal is a set of dry, no. 2, form C, double-pole, double-throw contacts with 30 mA and 24 VDC capacity.
d. Split-Electronics Connections
to Sensor Housing
If the OCX 4400 is configured with split­electronics, make the following connections between the electronics and sensor hous­ings with the electronics cable ordere d with the package (Figure 2-7). Braided cable is available in lengths up to 150 ft. (45 m).
1. Thermocouple Connections. Connect the electronics housing sensor/therm­ocouple terminals to the corresponding terminals in the sensor housing. The twisted wire pairs are numbered on the inner plastic wrapper. Keep tw isted pairs together and match the numbers and wire colors shown in Figure 2-7.
2
2
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 29
OCX 4400
2
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
Connect the twisted pair wire shields to the shield terminals.
2. Heater Power Connections. Use the purple, orange, black, gray, brown, and red stranded wires in the electronics cable to connect power to the three
SOLENOID POWER
TERMINAL B L OCK (FA CTORY WIRED)
GND
GND
LO O2
HI COMB
HI O2
ALARM OUTPUT
RELA Y TERMINAL
GND
GND
INST AIR
NC
NO
COMMON
SIGNAL OUTPUT
TERMINAL B L OCK
4-20 mA
ELECTRONICS HOUSING
(COVER REMO VED)
SENSOR INPUT
HEATER THERMOCOUPLE,
AND HEATER POWER
INPUT PORT
(FROM SENSOR HOUSING)
heaters in the sensor housing. Match the wire colors to the corresponding heater power terminal blocks in the sensor and electronics housings as shown in Figure 2-7.
HI O2
PRESSURE
SENSOR
HI COMB
SOLENOID
SOLENOID
SOLENOID
LO O2
INST AIR
SOLENOID
AOUT 2-
AOUT 2+
COMB 4-20 mA SIGNAL
AOUT 1-
AOUT 1+
O2 4-20 mA SIGNAL
AC POWER INPU T TERMINAL B L OCK
F1 POWER
SUPPLY FUSE
L
F2 HEATER
N
G
FUSE
Figure 2-6. Line Voltage and 4-20 mA Connections
4-20 mA SIGNAL
OUTPUT
PORT
AC POWER
INPUT PORT
GROUND STUD
TWO NUTS
34070021
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Page 30
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
SPLIT-ELECTRONICS HOUSING
HEATER POWER
TERMINAL BLOCK
GRAY
L2
BROWN
COMB CELL HEATER
RED
L2
ORANGE
BLACK
PURPLE
L2
SL HEATER
O2 CELL HEATER
ELECTRONICS CABLE
(UP TO 150 FT)
NOTE: KEEP TWISTED
PAIRS TOGETHER
(COVER REMOVED)
SENSOR/THERMOCOUPLE
TERMINAL BLOCK
(1)
YELLOW
O2 T/C +
RED
O2 T/C -
(2)
SHIELD
YELLOW
GND
SL T/C +
RED
SL T/C -
(3)
SHIELD
WHITE
BLUE
GND
O2 MV -
O2 MV +
PRESSURE
SENSOR
ELECTRONICS
CABLE PORT
GROUND STUD
(5)
(4)
YELLOW
WHITE
GREEN
COMB ACTIVE +
WHITE
COMB REF -
COMB REF +
COMB ACTIVE -
CABLE
(6)
WHITE
SHIELD
RED
GND
COMB DRIVE
COMB RETURN
PRESSURE
SENSOR
FACTORY
WIRED
WHITE
RED
GREEN
+ 10 PR SRC
PR SENSOR -
PR SENSOR +
BLACK
PR SRC RTN
(5)
(6)
WHITE
WHITE
RED
YELLOW
BLUE
COMB REF -
COMB REF +
COMB DRIVE
COMB RETURN
SENSOR/THERMOCOUPLE
TERMINAL BLOCK
ELECTRONICS
CABLE PORT
PROBE
L2
L2
SL HEATER
O2 CELL HEATER
COMB CELL HEATER
GRAY
BLACK
PURPLE
ORANGE
HEATER POWER
TERMINAL BLOCK
BROWN
L2
RED
(1)
YELLOW
O2 T/C +
RED
O2 T/C -
(2)
YELLOW
SL T/C +
RED
SL T/C -
(4)
(3)
WHITE
BLUE
O2 MV -
O2 MV +
WHITE
GREEN
RED
WHITE
COMB ACTIVE -
COMB ACTIVE +
SENSOR HOUSING
(COVER REMOVED)
37020005
Figure 2-7. Electrical Connections Between Split-Electronics and Sensor Housing
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 31
OCX 4400
2
Instruction Manual
IB-106-440 Rev. 2.0
January 2003

2-3 PNEUMATIC INSTALLATION

Pneumatic system connections depend on which OCX 4400 options are equipped on your transmitter. Refer to paragraphs 2-3a, 2-3b, and 2-3c, and select the paragraph that applies to your OCX transmitter configuration.
a. Reference Air Set Option (only)
When no options or only the reference air set option is equipped, use the following procedure to install the pneum atic system components.
1. Refer to Figure 2-8. Connect the refer­ence air set (regulator/filter and pres­sure gage) to the instrument air inlet on the electronics housing and to the inlet side of the dilution air rotometer.
2. Connect the dilution air rotometer out­put to the DILUTION AIR inlet fitting on the sensor housing.
3. Install an air line between the instru­ment air outlet fitting on the electronics housing and the tee fitting on the sen­sor housing.
Do not use 100% nitrogen as an O2 low gas. It is suggested that O2 low gas be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon concentra­tions of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.
4. One CO gas and two O2 gases are used to calibrate the OCX 4400: CO - 1000 ppm or 4%, O
0.4%, O
high gas - 8%. Connect the
2
low gas -
2
output of the test gas sources to the inlet port of the CAL GAS rotometer. Install an air line between the rotome­ter outlet port and the CAL GAS inlet fitting on the sensor housing.
SENSOR
HOUSING
EDUCTOR
ELECTRONICS
HOUSING
AIR IN
CAL GAS IN
REFERENCE
AIR IN
DILUTION
AIR IN
DILUTION
AIR FLOW
METER
0.1 SCFH
REGULATOR/FILTER
HI CO
INSTRUMENT
AIR SUPPLY
PRESSURE
35 PSIG
CAL GAS
FLOW METER
5 SCFH
20-30 PSIG
(
RECOMMENDED
2-STAGE
REGULATORS
2
2
HI O
LO O
(
37020006
Figure 2-8. Pneumatic Installation, OCX with Reference Air Set (without Autocalibration)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Page 32
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
SENSOR
HOUSING
CAL GAS IN
CAL GAS
FLOW METER
5 SCFH
20-30 PSIG
(
RECOMMENDED
ELECTRONICS
HOUSING
HI CO
(
2
HI O
EDUCTOR
AIR IN
2-STAGE
REGULATORS
2
LO O
REFERENCE
AIR IN
DILUTION AIR IN
DILUTION AIR FLOW METER
0.1 SCFH
INSTRUMENT
AIR SUPPLY
PRESSURE
REGULATOR/FILTER
35 PSIG
37020007
Figure 2-9. Pneumatic Installation, OCX with Reference Air Set and Solenoids
(with Autocalibration)
2-10 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 33
OCX 4400
2
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
b. Reference Air Set and Solenoids
When the reference air set and test gas solenoids are included with your OCX, use the following procedure to install the pneumatic system components.
1. Install the reference air set according to the instructions in paragraph 2-3a, steps 1 through 3.
2. Refer to Figure 2-9. Connect the O low gas source to the CAL GAS LO O
inlet fitting on the electronics
2
housing. Install a shutoff valve and pressure regulator with gage in the O
low supply line, as shown.
2
3. Connect the O the CAL GAS HI O
high gas source to
2
inlet fitting. In-
2
stall a shutoff valve and pressure regulator with gage in the O
2
supply line.
4. Connect the CO high gas to the CAL GAS HI COMB inlet fitting. Install a shutoff valve and pressure regulator with gage in the CO high supply line.
5. Connect the CAL GAS outlet fitting of the electronics housing to the inlet port of the CAL GAS rotometer. In­stall an air line between the rotome­ter outlet port and the CAL GAS inlet fitting on the sensor housing.
high
c. Reference Air, Solenoids, and
Blowback
When the reference air set, calibration gas solenoids, and blowback options are included with your OCX transmitter, use the following procedure to install the pneumatic system components.
1. Connect the calibration gas sources according to the instructions in para­graph 2-3b, steps 2 through 5.
2
2. Use the instructions in Appendix C to install the reference air set and blowback system components.
Upon completing installation, make sure that the OCX 4400 is turned on and operating prior to firing up the combustion process. Damage can result from having a cold OCX 4400 exposed to the process gases.
NOTE
During outages, and whenever pos­sible, leave OCX 4400 units running to prevent condensation and prema­ture aging from thermal cycling.
2-4 INITIAL STARTUP
Observe the preceding Caution and Note. Refer to Section 3, STARTUP AND OPERA­TION, for OCX 4400 startup information.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11
Page 34
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
SWITCH
DEFAULT POSITIONS
SHOWN
SW3
CLOSED
112
OCX 4400
SECTION 3
2
COMB
O
21.1 mA
3.5 mA
2
3
4
4
3
O221.1 mA/3.5 mA: O24-20 mA SIGNAL
RAIL LIMITS: HIGH - 21.1 mA LOW - 3.5 mA
COMB 21.1 mA/3.5 mA:
COMB 4-20 mA SIGNAL RAIL LIMITS: HIGH - 21.1 mA LOW - 3.5 mA
FUSE F2
INTERNAL:
COMB 4-20 mA IS
INTERNALLY
POWERED (DEFAULT)
SW2
Figure 3-1. OCX 4400 Defaults
INTERNAL:
O 4-20 mA IS
2
INTERNALLY
POWERED
(DEFAULT)
EXTERNAL:
COMB 4-20 mA
REQUIRES
AN EXTERNAL
POWER SUPPLY
SW1
EXTERNAL: O 4-20 mA
2
REQUIRES
AN EXTERNAL
POWER SUPPLY
37020008
3-0 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 35
OCX 4400
3
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 3

STARTUP AND OPERATION

Install all protective equipment covers and safety ground leads before equipment startup. Failure to install covers and ground leads could result in serious injury or death.
3-1 GENERAL
a. Verify Installation
Ensure the OCX 4400 is installed correctly. Verify mechanical installation and all electri­cal and pneumatic connections (Section 2, INSTALLATION).
NOTE
During outages, and whenever possi­ble, leave all OCX 4400 units running to prevent condensation and prema­ture aging from thermal cycling.
Make sure that the OCX 4400 is turned on and operating prior to firing up the combustion process. Damage can re­sult from having a cold OCX 4400 ex­posed to the process gases.
b. Verify OCX 4400 Configuration
There are three switches on the microproc­essor board which are user configurable for the OCX 4400 (Figure 3-1). SW1 deter­mines if the O or externally powered. SW2 determines if the COMB 4-20 mA signal is internally or externally powered. SW3 sets the rail limits
4-20 mA signal is internally
2
for the O configures the sample line heater control circuit. All switches are accessible through holes in the electronics box.
Remove power from the OCX 4400 be­fore changing defaults. If defaults are changed under power, damage to the electronics may occur.
and COMB 4-20 mA signals and
2
1. SW1. The two settings are internally or externally powering the O signal. The factory setting is for the O 4-20 mA signal to be internally pow­ered.
2. SW2. The two settings are internally or externally powering the COMB 4-20 mA signal. The factory setting is for the COMB 4-20 mA signal to be internally powered.
3. SW3. The factory sets this switch as follows:
(a) Position 1 determines the O
4-20 mA signal rail limit. The set­tings are high, 21.1 mA, or low, 3.5 mA. The factory setting is at low,
3.5 mA.
(b) Position 2 determines the COMB
4-20 mA signal rail limit. The set­tings are high, 21.1 mA, or low,
3.5 mA. The factory setting is at high, 21.1 mA.
(c) Position 3 is not used.
(d) Position 4 is not used.
4-20 mA
2
2
2
Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-1
Page 36
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
c. Initial Power Up
Allow adequate time for the heaters to begin operation and for the OCX transmitter to reach normal operating temperature on power up. Normal operating temperature for the O
cell is 736°C. Normal operating tem-
2
perature for the combustibles cell is 300°C. The normal sample line temperature is 130°C.
Upon initial power up of the OCX 4400, and any time supply voltage to the unit changes, the power input voltage must be verified in the software to ensure proper operation of the heaters. The input voltage is set as fol­lows:
1. Remove the housing cover from the electronics housing.
2. Remove fuse F2 from the electronics box. See Figure 3-1 for the location of fuse F2.
3. Replace the electronics housing cover.
4. Apply power to the OCX 4400 Oxy­gen/Combustibles Transmitter.
10. Remove the electronics housing cover and install fuse F2.
11. Install the electronics housing cover and restore power to the OCX 4400 electronics.
d. Set Test Gas Values
Using HART/AMS or optional LOI, set the test gas values for calibration. Refer to Section 5, HART/AMS or Appendix A, Local Operator Interface for more information.
1. HART/AMS. If using HART/AMS, from the DETAILED SETUP menu, select O
CALIBRATION.
2
(a) From O
CALIBRATION, select
2
High TG. Enter the percentage of O
used for the high O2 test gas.
2
(b) From O
CALIBRATION, select
2
Low TG. Enter the percentage of O
used for the low O2 test gas.
2
(c) From the DETAILED SETUP
menu, select COMB CALIBRATION.
5. If using the HART handheld communi­cator, connect the communicator to the transmitter.
6. If using the HART communicator, se­lect: Device Setup > Detailed Setup > Line Voltage.
If using the LOI, select: Setup > Pa­rameters > Line Voltage.
7. Set the line voltage to the appropriate value: 100, 110, or 220.
8. If using the HART communicator, press F4 for Enter and then F2 for send.
If using the LOI, press the right arrow to enter the value.
9. Remove power from the OCX 4400 electronics.
(d) From COMB CALIBRATION, se-
lect High TG. Enter the percentage of CO used for high COMB test gas.
2. LOI. If using the LOI, from the SETUP menu select CALIBRATION.
(a) From CALIBRATION, select O
2
High Gas. Enter the percentage of O
used for the high O2 test gas.
2
(b) Select the down arrow and the
next selection will be O Enter the percentage of O for the low O
test gas.
2
Low Gas.
2
used
2
(c) Select the down arrow again and
the next selection will be Comb High Gas. Enter the percentage of O
used for the combustibles span
2
gas.
3-2 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 37
OCX 4400
4
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 4

CALIBRATION

4-1 GENERAL
During a calibration, two calibration gases with known O gas with a known COMB concentration are ap­plied to the OCX 4400. Slope and constant val­ues are calculated to determine if the OCX 4400 is correctly measuring net concentrations of O and combustibles in the industrial process.
Before calibrating the OCX 4400, verify that the calibration gas parameters are correct by setting the test gas values used when calibrating the unit (Refer to Section 3, STARTUP AND OP­ERATION).

4-2 CALIBRATION METHODS

There are three types of calibration methods available to the OCX 4400; automatic, operator­initiated automatic, and manual. Calibration commands and menus can be accessed by HART/AMS or by the optional LOI.
a. Automatic Calibration
concentrations and one calibration
2
(a) Select DIAG/SERVICE from DE-
VICE SETUP menu.
(b) Select CALIBRATE from the
DIAG/SERVICE menu.
(c) Select PERFORM CAL from CALI-
2
BRATE menu.
(d) Select CAL from PERFORM CAL
menu.
(e) Select the type of calibration de-
sired: O method, or O method from CAL menu.
2. LOI. To initiate a calibration using the LOI, perform the following steps on the LOI menu tree. Refer to APPENDIX A for complete LOI menu tree.
(a) From the CALIBRATE menu, use
the right-pointing arrow to select DO CALIBRATION.
Cal method, COMB Cal
2
and COMB Cal
2
If the OCX 4400 is equipped with calibration solenoids, the unit can be programmed to automatically calibrate without any operator action. Refer to Section 5, HART/AMS for information on automatic calibration setup via HART/AMS. Refer to Appendix A for in­formation on automatic calibration setup via LOI.
b. Operator Initiated-Automatic Calibration
An operator can initiate an automatic cali­bration at any time provided that the unit is equipped with calibration solenoids.
1. HART/AMS. To initiate an automatic calibration using HART/AMS, perform the following steps on the HART/AMS menu tree. Refer to Figure 5-3 for the complete HART/AMS menu tree.
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration 4-1
c. Manual Calibration
(b) Select O2, COMBUSTIBLES, or
BOTH sensors for calibration. Use the right-pointing arrow to access the calibration window.
(c) At the prompt, use the right-
pointing arrow to initiate automatic calibration.
If a unit is not equipped with calibration so­lenoids, a calibration must be performed by an operator following prompts from the unit. Refer to Section 5, HART/AMS for informa­tion on manual calibration with HART/AMS. Refer to APPENDIX A for information on manual calibration with LOI.
Page 38
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
4-2 Calibration Rosemount Analytical Inc. A Division of Emerson Process Management
Page 39
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 5

HART/AMS

5-1 OVERVIEW
The HART Communicator is a handheld com­munications interface device. It provides a common communications link to all microproc­essor-based instruments that are HART com­patible. The handheld communicator contains an 8 x 21 character liquid crystal display (LCD) and 25 keys. A pocket-sized manual, included with the HART Communicator, details the spe­cific functions of all the keys.
To interface with the OCX 4400, the HART Communicator requires a termination point along the O
4-20 mA current loop and a mini-
2
mum load resistance of 250 ohms between the communicator and the power supply. The HART Communicator accomplishes its task using a frequency shift keying (FSK) technique. With the use of FSK, high-frequency digital communica­tion signals are superimposed on the 4-20 mA oxygen output signal. The communicator does not disturb the 4-20 mA signal since no net en­ergy is added to the loop. HART information is not available on the COMB output signal.
4-20 mA TERMINAL BLOCK
IN ELECTRONICS HOUSING
The HART Communicator may be interfaced with a personal computer (PC), providing spe­cial software has been installed. To connect the HART Communicator to a PC, an interface adapter is required. Refer to the proper HART Communicator documentation in regard to the PC interface option.
5-2 HART COMMUNICATOR SIGNAL LINE
CONNECTIONS
The HART Communicator can connect to the OCX 4400 oxygen analog output signal line at any wiring termination in the O
4-20 mA current
2
loop. There are two methods of connecting the HART Communicator to the signal line. For ap­plications in which the signal line has a load re­sistance of 250 ohms or more, refer to method
1. For applications in which the signal line load resistance is less than 250 ohms, refer to method 2.
a. Method 1, For Load Resistance 250
Ohms.
Refer to Figure 5-1 and the following steps to connect the HART Communicator to a signal line 250 ohms or more of load resistance.
RL 250
A OUT 1 -
A OUT 1 +
LOOP CONNECTORS
SERIALPORT & BATTERY
CHARGERMUST
NOTBE USED IN
HAZARDOUSAREAS
LOOPCONNECTORS
HART COMMUNICATOR
REAR PANEL
O2 4-20 mA SIGNAL LINE
USEINTERFACE 002750013 ONLY
SERIAL PORT
HART
COMMUNICATOR
ANALOG OUTPUT DEVICE
LEAD SET
29790018
Figure 5-1. Signal Line Connections, > 250 Ohms Load Resistance
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-1
Page 40
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
Explosions can result in death or seri­ous injury. Do not make connections to the HART Communicator’s serial port, 4-20 mA signal lines, or NiCad recharger jack in an explosive atmosphere.
Using the supplied lead set, connect the HART Communicator in parallel to the OCX
4400. Use any wiring termination points in the oxygen analog output 4-20 mA signal line.
b. Method 2, For Load Resistance
250 Ohms
Refer to Figure 5-2 and the following steps to connect the HART Communicator to a
OCX 4400
signal line with less than 250 ohms load resistance.
Explosions can result in death or seri­ous injury. Do not make connections to the HART Communicator’s serial port, 4-20 mA signal lines, or NiCad recharger jack in an explosive atmosphere.
1. At a convenient point, break the oxy­gen analog output 4-20 mA signal line and install the optional 250 ohm load resistor.
2. Plug the load resistor into the loop connectors (located on the rear panel of the HART Communicator).
IN ELECTRONICS HOUSING
A OUT 1 ­A OUT 1 +
NOTE:
THE SIGNAL LOOP MUST BE BROKEN TO INSERT THE OPTIONAL 250 OHM
4-20 mA TERMINAL BLOCK
LOOP CONNECTORS
SERIALPORT & BATTERY
CHARGERMUST
NOTBE USED IN
HAZARDOUSAREAS
HART COMMUNICATOR
REAR PANEL
LOOPCONNECTORS
O2 4-20 mA SIGNAL LINE
USEINTERFACE 002750013 ONLY
SERIAL PORT
LOAD RESISTOR.
RL < 250
ANALOG OUTPUT DEVICE
HART
COMMUNICATOR
Figure 5-2. Signal Line Connections, 250 Ohms Load Resistance
250 OHM
LOAD
RESISTOR
(NOTE 1)
29790017
5-2 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management
Page 41
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
5-3 HART COMMUNICATOR PC
CONNECTIONS
There is an option to interface the HART Com­municator with a personal computer. Load the designated AMS software into the PC. Then link the HART Communicator to the PC using the interface PC adapter that connects to the serial port (on the communicator rear panel). Refer to the proper HART Communicator documentation in regard to the PC interface option.

5-4 OFF-LINE AND ON-LINE OPERATIONS

The HART Communicator can be operated both off-line and on-line.
a. Off-line operations are those in which the
communicator is not connected to the OCX
4400. Off-line operations can include inter­facing the HART Communicator with a PC. (Refer to applicable HART documentation regarding HART/PC applications.)
b. In the on-line mode, the communicator is
connected to the oxygen 4-20 mA analog output signal line. The communicator is connected in parallel to the OCX 4400 or in parallel to the 250 ohm load resistor.
c. The opening menu displayed on the LCD is
different for on-line and off-line operations. When powering up a disconnected (off-line) communicator, the LCD will display the Main Menu. When powering up a connected (on­line) communicator, the LCD will display the On-line Menu. Refer to the HART Commu­nicator manual for detailed menu informa­tion.
5-5 MENU TREE FOR HART COMMUNICATOR/
OCX 4400 APPLICATIONS
This section consists of a menu tree for the HART Communicator. This menu is specific to OCX 4400 applications.
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-3
Page 42
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
PROCESS VARIABLES
DEVICE SETUP O2 O2 AO COMB COMB AO
DIAG/SERVICE
VIEW FLD DEV VARS
VIEW FLD DEV mV
STATUS
O2 O2 temp Electronics temp COMB COMB temp SL temp Pressure
O2 mV O2 temp mV Elec temp V COMB Act mV COMB Ref mV Pressure V SL temp mV
Status Group 1
Status Group 2
Status Group 3
Status Group 4
OCX 4400
O2 Thermocouple Error O2 Cell Temperature Low O2 Cell Temperature High O2 Cell Temperature Very High O2 Heater Open O2 Cell Open O2 Cell Bad O2 Slope Error
O2 Constant Error O2 Calibration Failed Landusky Error High Electronics Temperature EEPROM Corrupt
Sample Line Thermocouple Error Sample Line Temperature Low Sample Line Temperature High Sample Line Temperature Very High Sample Pressure Low Absolute
Combustibles Cell Error Combustibles Cell Chamber Temperature Low Combustibles Cell Chamber Temperature High Combustibles Cell Chamber Temperature
Very High Combustibles Slope Error CO Constant Error CO Cal Failed
O2 AO Saturated COMB AO Saturated
O2 AO Fixed COMB AO Fixed
Cal methods O2 Optrak TG? COMB Optrak TG? CalState
CalState TimeRemain Present O2 Present COMB
O2 Cal slope O2 Cal Const O2 Cell Imp COMB Slope COMB Const
Reset O2 constants Reset COMB constants
(CONTINUED ON
SHEET 2)
LOOP TEST
CALIBRATE
D/A TRIM
AO Saturated
AO Fixed
Max Electronics Temp
O2 Loop test method COMB Loop test
method
PERFORM CAL
CAL STATUS
LAST CAL CONSTANTS
RESET CAL CONSTANTS
O2 D/A trim method COMB D/A trim method
Figure 5-3. Menu Tree for HART/AMS on the OCX 4400 (Sheet 1 of 3)
O2 Cal method COMB Cal method O2 and COMB Cal
method
34070014
5-4 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management
Page 43
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
(CONTINUED FROM
SHEET 1)
BASIC SETUP
DETAILED SETUP
Ta g
SELECT RANGE
DEVICE INFORMATION
S/W VERSION INFO
SENSORS
O2 URV COMB URV
Dev id Descriptor Message Date Final asmbly num O2 Snsr s/n COMB Snsr s/n
Ver Chk sum Bld num Bld date
O2
COMB
O2 O2 Temp Electronics temp O2 mV O2 Temp mV Elec temp V
COMB COMB TEMP COMB ACT MV COMB REF MV SL TEMP SL temp mV Pressure Pressure V
(CONTINUED ON
SHEET 3)
SIGNAL CONDITION
OUTPUT CONDITION
LINE VOLTAGE
(CONTINUED ON
SHEET 3)
O2 % range COMB % range
ANALOG OUTPUT
HART OUTPUT
100 115 220
O2 Analog Output
COMB Analog Output
Poll addr Num req preams
O2 AO
O2 Alrm typ
O2 Range
COMB AO
COMB Alrm typ
COMB Range
Figure 5-3. Menu Tree for HART/AMS on the OCX 4400 (Sheet 2 of 3)
3.5 mA
21.1 mA
4-20 20-4
3.5 mA
21.1 mA
4-20 20-4
37020023
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-5
Page 44
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
(CONTINUED FROM
SHEET 2)
DETAILED SETUP
(CONTINUED FROM
SHEET 2)
O2 SETUP
O2 CALIBRATION
COMB SETUP
COMB CALIBRATION
ALARM RELAY
BLOWBACK
SLOPE
O2 CONST O2 T90
High TG Low TG OPtrak TG? TG Time Purge Time Sol. Present Cal Intrvl h Next Cal Time h
COMB SLOPE
COMB CONST COMB T90
High TG OPtrak TG? TG Time Purge Time Sol. Present Cal Intrvl h Next Cal Time h
EVENT 1 EVENT 2 EVENT 3 Alarm State
Enabled Interval Duration Purge Time State Do Blowback
OP Locks
OP Tracks
OP Locks
OP Tracks
None In Cal O2 Temp O2 Heater O2 Cell Bad Cal Failed Hi Elect Temp Unit Fail SL Temp SL Pressure Comb Temp All
REVIEW
DEVICE INFORMATION
CAL INFO
DEVICE CONFIG
OUTPUT CONFIG
Manufacturer Model Dev id Ta g Descriptor Message Date Final asmbly num O2 Snsr s/n COMB Snsr s/n Fld dev rev Hardware rev Software rev Universal rev
O2 Cal Info
COMB Cal Info
O2 Device Config
COMB Device Config
O2 URV COMB URV Poll addr
Num req preams
OPtrak TG? TG Time Purge Time Low TG High TG Slope O2 const O2 cell Imp
OPtrak TG? TG Time Purge Time High TG COMB slope COMB const
SLOPE
O2 CONST O2 T90
COMB SLOPE
COMB CONST COMB T90
34070016
Figure 5-3. Menu Tree for HART/AMS on the OCX 4400 (Sheet 3 of 3)
5-6 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management
Page 45
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
5-6 HART COMMUNICATOR MANUAL O2 CAL
METHOD
To perform a manual O
calibration using the
2
HART Communicator, use the following proce­dure. If necessary, use the menu tree in Figure 5-3 (sheet 1 of 3) for reference.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right ar­row key or use the number keypad to select the menu item number.
To return to a preceding menu, press the left arrow key.
a. From the PERFORM CAL screen, select
menu item 1, CAL, then menu item 1, O
2
Cal method to access the O2 calibration method.
Failure to remove the OCX 4400 from automatic control loops prior to per­forming this procedure may result in a dangerous operating condition.
calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O2, as the calibration status updates.
f. When CALSTATE displays APPLY HIGH O
GAS, return to the O2 CAL procedure. When the “Loop should be removed from auto­matic control” warning appears, press OK.
g. At the APPLY HIGH O
GAS status prompt,
2
select menu item 2, NEXT CAL STEP. When the status displays HIGH O
GAS
2
FLOW, select menu item 4, EXIT, to leave the O
CAL procedure.
2
h. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O
, as the
2
calibration status updates.
i. When CALSTATE displays STOP GAS, re-
turn to the O
CAL procedure.
2
2
b. In the first O
CAL method screen, a “Loop
2
should be removed from automatic control” warning appears. Remove the OCX 4400 from any automatic control loops to avoid a potentially dangerous operating condition and press OK.
c. The next several screens indicate the cali-
bration status. At each of the following status prompts, select menu item 2, NEXT CAL STEP:
COMPLETE APPLY LOW O LOW O
GAS FLOW
2
GAS
2
d. At this point, select menu item 4, EXIT, to
leave the O2 CAL procedure.
e. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
j. When the “Loop should be returned to
automatic control” message appears, return the OCX 4400 to the automatic control loops previously removed and press OK.
k. At the STOP GAS status prompt, select
menu item 2, NEXT CAL STEP. When the status displays PURGING, select menu item 4, EXIT, to leave the O
CAL procedure.
2
l. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O
, as the
2
calibration status updates.
m. When CALSTATE displays COMPLETE,
the calibration is finished.
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-7
Page 46
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
5-7 HART COMMUNICATOR MANUAL COMB
CAL METHOD
To perform a manual combustibles calibration using the HART Communicator, use the follow­ing procedure. If necessary, use the menu tree in Figure 5-3 (sheet 1 of 3) for reference.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right ar­row key or use the number keypad to select the menu item number.
To return to a preceding menu, press the left arrow key.
a. From the PERFORM CAL screen, select
menu item 1, CAL, then select menu item 2, COMB Cal method to access the combusti­bles calibration method.
Failure to remove the OCX 4400 from automatic control loops prior to per­forming this procedure may result in a dangerous operating condition.
calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 4, PRESENT COMB, as the calibration status updates.
f. When CALSTATE displays APPLY COM-
BUSTIBLES GAS, return to the O procedure. When the “Loop should be re­moved from automatic control” warning ap­pears, press OK.
g. At the APPLY COMBUSTIBLES GAS status
prompt, select menu item 2, NEXT CAL STEP. When the status displays COM­BUSTIBLES GAS FLOW, select menu item 4, EXIT, to leave the COMB CAL procedure.
h. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 4, PRESENT COMB, as the calibration status updates.
i. When CALSTATE displays STOP GAS, re-
turn to the COMB CAL procedure.
CAL
2
b. In the first COMB CAL method screen, a
“Loop should be removed from automatic control” warning appears. Remove the OCX 4400 from any automatic control loops to avoid a potentially dangerous operating condition and press OK.
c. The next several screens indicate the cali-
bration status. At each of the following status prompts, select menu item 2, NEXT CAL STEP:
COMPLETE APPLY HIGH O HIGH O2 GAS FLOW
d. At this point, select menu item 4, EXIT, to
leave the COMB CAL procedure.
e. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
GAS
2
j. When the “Loop should be returned to
automatic control” message appears, return the OCX 4400 to the automatic control loops previously removed and press OK.
k. At the STOP GAS status prompt, select
menu item 2, NEXT CAL STEP. When the status displays PURGING, select menu item 4, EXIT, to leave the COMB CAL procedure.
l. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT COMB, as the calibration status updates.
m. When CALSTATE displays COMPLETE,
the calibration is finished.
5-8 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management
Page 47
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
5-8 HART COMMUNICATOR MANUAL O2 AND
CO CAL METHOD
To perform a manual O
and COMB calibration
2
using the HART Communicator, use the follow­ing procedure. If necessary, use the menu tree in Figure 5-3 (sheet 1 of 3) for reference.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right ar­row key or use the number keypad to select the menu item number.
To return to a preceding menu, press the left arrow key.
a. From the PERFORM CAL screen, select
menu item 1, CAL, then menu item 3, O and COMB Cal method to access the O
2
2
and COMB calibration method.
Failure to remove the OCX 4400 from automatic control loops prior to per­forming this procedure may result in a dangerous operating condition.
CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O
, as the
2
calibration status updates.
f. When CALSTATE displays APPLY HIGH O
GAS, return to the O2 and COMB CAL pro­cedure. When the “Loop should be removed from automatic control” warning appears, press OK.
g. At the APPLY HIGH O
GAS status prompt,
2
select menu item 2, NEXT CAL STEP. When the status displays HIGH O
GAS
2
FLOW, select menu item 4, EXIT, to leave the O
and COMB CAL procedure.
2
h. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O
, as the
2
calibration status updates.
i. When CALSTATE displays APPLY COM-
BUSTIBLES GAS, return to the O
and
2
COMB CAL procedure. When the “Loop should be removed from automatic control” warning appears, press OK.
2
b. In the first O
CAL method screen, a “Loop
2
should be removed from automatic control” warning appears. Remove the OCX 4400 from any automatic control loops to avoid a potentially dangerous operating condition and press OK.
c. The next several screens indicate the cali-
bration status. At each of the following status prompts, select menu item 2, NEXT CAL STEP:
COMPLETE APPLY LOW O LOW O
GAS FLOW
2
GAS
2
d. At this point, select menu item 4, EXIT, to
leave the O
CAL procedure.
2
e. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1,
j. At the APPLY COMBUSTIBLES GAS status
prompt, select menu item 2, NEXT CAL STEP. When the status displays COM­BUSTIBLES GAS FLOW, select menu item 4, EXIT, to leave the O
and COMB CAL
2
procedure.
k. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 4, PRESENT COMB, as the calibration status updates.
l. When CALSTATE displays STOP GAS, re-
turn to the O
and COMB CAL procedure.
2
m. When the “Loop should be returned to
automatic control” message appears, return the OCX 4400 to the automatic control loops previously removed and press OK.
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-9
Page 48
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
n. At the STOP GAS status prompt, select
menu item 2, NEXT CAL STEP. When the status displays PURGING, select menu item 4, EXIT, to leave the O procedure.
o. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the calibration status as it updates. Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; menu item 3, PRESENT O 4, PRESENT COMB, as the calibration status updates.
and COMB CAL
2
, and menu item
2
OCX 4400
When setting automatic calibration times, the CAL INTRVL and the NEXT CAL TIME should be set so that O and combustibles are NOT calibrated simultaneously.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right ar­row key or use the number keypad to select the menu item number.
2
p. When CALSTATE displays COMPLETE,
the calibration is finished.
5-9 AUTOMATIC CALIBRATION SETUP
VIA HART
NOTE
Automatic calibration is only available on units equipped with calibration solenoids.
Use the following procedure to specify a time interval (in hours) at which the OCX 4400 will automatically calibrate the OCX 4400.
To disable the automatic calibration feature for oxygen or combustibles, enter a 0 for the CAL INTRVL for both O
and COMB.
2
If desired, the NEXT CAL TIME can be changed to synchronize a calibration at a specific day or time.
To return to a preceding menu, press the left arrow key.
a. From the DEVICE SETUP screen, select
DETAILED SETUP.
b. From the DETAILED SETUP screen, select
O
CALIBRATION. From the O2 CALIBRA-
2
TION screen, select menu item 7, CAL INTRVL.
c. At the prompt, input a time interval (in
hours) at which an automatic O
calibration
2
will occur and press ENTER.
d. From the DETAILED SETUP screen, select
COMB CALIBRATION.
e. From the COMB CALIBRATION screen,
select menu item 7, CAL INTRVL.
f. At the prompt, input a time interval (in
hours) at which an automatic COMB cali­bration will occur and press ENTER.
5-10 LINE VOLTAGE SELECTION USING HART
COMMUNICATOR
Use the procedure in paragraph 3-1c to select or change the line voltage configuration using the HART Communicator.
5-10 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management
Page 49
OCX 4400
6
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 6

OPERATION

6-1 OVERVIEW
Once the OCX 4400 is installed, set up, and calibrated properly, the unit is ready for operation.
6-2 OPERATION
Before the OCX 4400 can begin normal opera­tion, all heaters must be up to temperature be-
fore the unit begins to take measurements of the sample gas. During normal operation, the unit will perform continuous measurements of the sample gas. The unit is performing properly if there are no alarms present. If the unit is equipped with an LOI, the normal operating dis­play will show a value for Oxygen and Combus­tible concentrations measured. If an LOI is not used, values for Oxygen and Combustibles are available through HART.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 6-1
Page 50
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
6-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 51
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 7

MAINTENANCE AND SERVICE

7-1 OVERVIEW
This section contains the procedures to main­tain and service the OCX 4400.
Install all protective equipment covers and safety ground leads after equip­ment repair or service. Failure to in­stall covers and ground leads could result in serious injury or death.
It is recommended that the OCX 4400 be removed from the stack for all service activities. The unit should be allowed to cool and be taken to a clean work area. Failure to comply may cause severe burns.
Disconnect and lock out power before working on any electrical components. There may be voltage up to 264 VAC.
(e) Disconnect and remove the two
power leads from the AC power in­put terminal block and the ground lead from the ground stud.
(f) Disconnect and remove the O
COMB 4-20 mA signal leads from the 4-20 mA signal output terminal block.
(g) If used, disconnect and remove the
external relay leads from the alarm output relay terminal.
(h) Remove insulation to access the
mounting bolts. Unbolt the OCX 4400 from the stack and take it to a clean work area.
(i) Allow the unit to cool to a comfort-
able working temperature.
2. Replace.
(a) Bolt the OCX 4400 to the stack
and install insulation. Remove cover on electronics housing.
2
and

7-2 OCX 4400 REMOVAL/REPLACEMENT

a. OCX 4400 with Integrally Mounted
Electronics
1. Remove.
(a) Turn off power to the system.
(b) Shut off the calibration gases at
the cylinders and the instrument air.
(c) Disconnect the calibration gas and
instrument air lines from the elec­tronics housing.
(d) Remove the cover from the elec-
tronics housing to expose the electronics housing terminal blocks, Figure 7-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-1
(b) If used, connect external relay
leads to the alarm output relay terminal, Figure 7-1.
(c) Connect the O
mA signal leads to the 4-20 mA signal output terminal block.
(d) Connect the line, or L wire, to the L
terminal, and the neutral, or N wire, to the N terminal on the AC power input terminal block in the elec­tronics housing. Connect the ground, or G wire, to the ground stud in the electronics housing. Secure the connection with two nuts. Attach a separate line from the ground stud to the G terminal on the power input terminal block.
and COMB 4-20
2
Page 52
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
(e) Replace cover on electronics
housing.
(f) Connect the calibration gasses and
instrument air lines to the elec­tronics housing.
4-20 mA
SIGNAL OUTPUT
TERMINAL BLOCK
ELECTRONICS HOUSING
AC POWER INPUT
TERMINAL BLOCK
L
N
G
(COVER REMOVED)
(g) Turn on the calibration gasses at
the cylinders and turn on instru­ment air.
(h) Restore power to the system.
GROUND STUD
TWO NUTS
ALARM OUTPUT
RELAY TERMINAL
NC
NO
AOUT 2+
COMB 4-20 mA SIGNAL
AOUT 2-
AOUT 1-
O2
4-20 mA
SIGNAL
AOUT 1+
COMMON
34070013
Figure 7-1. Electronics Housing Terminal Blocks
7-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 53
OCX 4400
7
L2
SL HEATER
O2 CELL HEATER
SENSOR HOUSING
(COVER REMOVED)
EMI
BOOT
HEATER POWER
TERMINAL BLOCK
L2
L2
COMB CELL HEATER
ELECTRONICS CABLE
(TO ELECTRONICS HOUSING)
SENSOR/THERMOCOUPLE
TERMINAL BLOCK
(1)
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
(5)
YELLOW
(6)
WHITE
RED
WHITE
(2)
RED
YELLOW
RED
YELLOW
(3)
BLUE
(4)
WHITE
GREEN
WHITE
RED
GRAY
BLACK
PURPLE
ORANGE
BROWN
Figure 7-2. Sensor Housing Terminal Blocks
b. OCX 4400 with Split Mounted
Electronics
1. Remove Sensor Housing.
(a) Turn power off to the system.
(b) Shut off the calibration gases at the
cylinders and the instrument air.
(c) Disconnect calibration gas, refer-
ence air, eductor air, and dilution air lines from the sensor housing.
(d) Remove the cover from the sensor
housing to expose the sensor housing terminal blocks, Figure 7-2.
(e) Disconnect the electronics cable
from sensor/thermocouple terminal block and the heater power termi­nal block.
(f) Remove insulation to access the
mounting bolts. Unbolt the sensor
ELECTRONICS
CABLE PORT
O2 T/C +
PROBE
O2 T/C -
SL T/C +
SL T/C -
O2 MV +
RED
O2 MV -
COMB ACTIVE +
WHITE
COMB ACTIVE -
BLUE
COMB REF -
COMB REF +
COMB DRIVE
37020009
housing from the stack and take it to a clean work area.
(g) Allow the sensor housing to cool
to a comfortable working temperature.
2. Replace Sensor Housing
(a) Bolt the sensor housing to the
stack and install insulation.
(b) With the cover removed, connect
the electronics cable to sensor/ thermocouple terminal block and the heater power terminal block, Figure 7-2. Install sensor housing cover.
(c) Connect the calibration gas, refer-
ence air, eductor air, and dilution air lines to the sensor housing.
(d) Turn on the calibration gases at the
cylinders and instrument air.
(e) Restore power to the system.
COMB RETURN
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-3
Page 54
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
HEATER POWER
TERMINAL BLOCK
GRAY
L2
BROWN
COMB CELL HEATER
RED
L2
ORANGE
BLACK
PURPLE
L2
SL HEATER
O2 CELL HEATER
ELECTRONICS CABLE
(UP TO 150 FT)
NOTE: KEEP TWISTED
PAIRS TOGETHER
SPLIT-ELECTRONICS HOUSING
(COVER REMOVED)
ELECTRONICS
CABLE PORT
SENSOR/THERMOCOUPLE
TERMINAL BLOCK
(2)
(1)
YELLOW
O2 T/C +
RED
O2 T/C -
GND
YELLOW
SL T/C +
RED
SL T/C -
(3)
WHITE
BLUE
GND
O2 MV -
O2 MV +
(4)
WHITE
GREEN
COMB ACTIVE -
COMB ACTIVE +
(5)
YELLOW
COMB REF +
Figure 7-3. Split-Electronics Housing Cable Connections
(6)
WHITE
RED
WHITE
GND
COMB REF -
COMB DRIVE
COMB RETURN
PRESSURE
SENSOR
FACTORY
WIRED
+ 10 PR SRC
PR SENSOR -
PR SENSOR +
37020010
PR SRC RTN
3. Remove Split-Electronics Housing.
(a) Turn power off to the system.
(b) Shut off the calibration gases at the
cylinders and the instrument air.
put terminal block. Remove the ground lead from the ground stud.
(f) Disconnect and remove the O
COMB 4-20 mA signal leads from
and
2
the 4-20 mA signal output terminal
(c) Disconnect the three calibration
block. gas input lines and instrument air line from the split-electronics housing. Disconnect the calibration gas outlet and instrument air outlet
(g) If used, disconnect and remove the
external relay leads from the alarm
output relay terminal. lines from the split-electronics
housing.
(d) Remove the cover from the elec-
tronics housing to expose the electronics housing terminal
(h) Disconnect the electronics cable
from sensor/thermocouple terminal
block and the heater power termi-
nal block, Figure 7-3.
blocks, Figure 7-1.
(i) Remove the split-electronics
(e) Disconnect and remove the two
power leads from the AC power in-
7-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
housing from mounting and move
to a suitable work area.
Page 55
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
4. Replace Split-Electronics Housing.
(a) Mount split-electronics housing on
wall or pole within distance of electronics cable used. Remove cover.
(b) Connect electronics cable to the
sensor/thermocouple and heater power terminal blocks, Figure 7-3.
(c) Connect the line, or L wire, to the L
terminal, and the neutral, or N wire, to the N terminal on the AC power input terminal block in the elec­tronics housing. Connect the ground, or G wire, to the ground stud in the electronics housing. Secure the connection with two nuts. Attach a separate line from the ground stud to the G terminal on the power input terminal block.
(d) Connect the O
and COMB 4-20
2
mA signal leads to the 4-20 mA signal output terminal block.
(e) If used, connect external relay
leads to the alarm output relay terminal.
formance have been considered. If cell re­placement is needed, order the O
cell
2
replacement kit (Table 10-1).
The O
cell replacement kit contains an O
2
cell and flange assembly, gaskets, socket head cap screws, and anti-seize compound. The items are carefully packaged to pre­serve precise surface finishes.
Do not remove items from the packaging until they are ready to be used.
Use heat-resistant gloves and clothing when removing the OCX 4400. Do not attempt to work on any sensor hous­ing components until they have cooled to room temperature. Components can be as hot as 2600°F (1427°C). This can cause severe burns.
Disconnect and lock out power before working on any electrical components. There may be voltage up to 264 VAC.
2
(f) Replace cover on electronics
housing.
(g) Connect the three calibration gas
input lines and the instrument air input line to the split-electronics housing. Connect the calibration gas outlet line and instrument air outlet line to the split-electronics housing.
(h) Turn on the calibration gasses at
the cylinders and instrument air.
(i) Restore power to the system.
7-3 SENSOR HOUSING COMPONENT
REPLACEMENT
Each of the following procedures details how to remove and replace a specific component lo­cated in the sensor housing of the OCX 4400.
a. O
Cell Replacement
2
This paragraph covers oxygen sensing cell replacement. Do not attempt to replace the cell until all other possibilities for poor per-
Do not remove the O2 cell unless cer­tain it needs to be replaced. Removal may damage the cell and platinum pad. Only remove the O2 cell if it is to be replaced.
1. Follow the instructions in paragraph 7-2a.1 to remove the OCX 4400 from the stack or duct. If removing a split­mounted unit, follow the instructions in paragraph 7-2b.1 to remove the sensor housing.
2. Remove sensor housing cover (1, Figure 7-4).
3. Remove reference air tube (2) from sensor housing (24).
4. Disconnect and tag all O
assembly (3)
2
wire leads from the sensor housing terminal blocks.
5. Remove the O
assembly (3) from sen-
2
sor housing (24).
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-5
Page 56
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
1. Cover
2. Reference Air Tube Assembly
3. O
2
4. Gasket
5. Internal Ring
6. Eductor Air Tube
7. Dilution Air Tube
8. Screw
9. Washer
10. Bracket
11. COMB Extractive
Tube
12. Heater
13. Screw
14. Plate
15. COMB Sensor
16. Gasket
17. Preheater
18. Thermocouple
19. Heater
20. Heater Clamp
21. Connector
22. Eductor
23. COMB Sensor
Holder
24. Sensor Housing
25. Sample Tube
26. Sample Filter
27. Adapter
28. Check Valve
(w/ blowback only)
29. Plug
(w/out blowback)
30. Exhaust Tube
30
BLOWBACK
FILTER
HOUSING
27
28
NOTES: NOT ALL ITEMS IDENTIFIED ARE AVAILABLE FOR SALE.
1. SEE SECTION 10, REPLACEMENT PARTS, FOR A LIST OF AVAILABLE PARTS.
2.
EXHAUST TUBE (30) MUST BE INSTALLED IN FLANGE PORT NEAREST TO THE BLOWBACK FILTER HOUSING.
1
2
3
8
9
4
10
11
5
21
22
26
29
6
5
12
17
23
20
24
7
13
14
15
16
8
9
18
19
25
37020011
Figure 7-4. Sensor Housing, Exploded View
7-6 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 57
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
6. Remove four allen cap screws (14, Figure 7-5) from O move O
cell and gasket (12). The cell
2
cell (13) and re-
2
flange has a notch that may be used to gently pry the flange away from heater tube (11). Note that the pad on the end of contact/thermocouple assembly (4) will sometimes fuse to O
cell is fused to the contact pad,
the O
2
push the O
cell back into the heater
2
cell (13). If
2
tube (against spring pressure) and quickly twist the O
cell. The cell and
2
contact pad should separate. If the contact pad stays fused to the cell, a new contact/thermocouple assembly must be installed.
7. Remove and discard gasket (12). Clean the mating surface of heater tube (11). Remove burrs and raised surfaces with a block of wood and cro­cus cloth.
8. Rub a small amount of anti-seize com­pound on both sides of new gasket (12).
9. Assemble O
cell (13), gasket (12), and
2
heater tube (11). Make sure the cali­bration gas passage holes line up with each other in all components. Apply a small amount of anti-seize compound to the screw threads and use screws (14) to secure assembly. Torque to 35 in-lbs (4 N·m).
10. Apply anti-seize compound to threads of O
assembly (3, Figure 7-4) and
2
sensor housing (24). Replace graphoil gasket (4) with new gasket in kit. Rein­stall O
11. Reconnect the wire leads from O
assembly in sensor housing.
2
2
as­sembly (3) to the sensor housing ter­minal blocks. Refer to Figure 7-2 if necessary.
12. Replace reference air tube (2) in sen­sor housing (24). Ensure that the open end of reference air tube extends into O
assembly (3).
2
13. Follow the instructions in paragraph 7-2a.2 to install the OCX 4400 into the stack or duct. If installing a split­mounted unit, follow the instructions in paragraph 7-2b.2.
14. Turn on power and after the OCX 4400 stabilizes, calibrate the unit. If new components have been installed, re­peat calibration after 24 hours of operation.
b. O
Heater Strut Assembly and
2
Contact/Thermocouple Assembly Replacement
This paragraph covers O
heater strut and
2
contact/thermocouple replacement. If heater strut or contact/thermocouple replacement is needed, order the replacement parts from Table 10-1.
Use heat-resistant gloves and clothing when removing the OCX 4400. Do not attempt to work on any sensor hous­ing components until they have cooled to room temperature. Components can be as hot as 2600°F (1427°C). This can cause severe burns.
Disconnect and lock out power before working on any electrical components. There may be voltage up to 264 VAC.
1. Follow the instructions in paragraph 7-2a.1 to remove the OCX 4400 from the stack or duct. If removing a split­mounted unit, follow the instructions in paragraph 7-2b.1 to remove the sensor housing.
2. Remove sensor housing cover (1, Figure 7-4).
3. Remove reference air tube (2) from sensor housing (24).
4. Disconnect and tag O
assembly (3)
2
wire leads from the sensor housing terminal blocks.
5. Remove four screws (1, Figure 7-5) and washers (2) securing O
heater
2
strut assembly (3) to heater tube (11). The spring tension should release and the heater strut assembly moves up.
6. Carefully slide heater strut assembly (3) out of heater tube (11).
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-7
Page 58
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
NOTE: NOT ALL ITEMS IDENTIFIED
4
1
2
5
6
7
8
9
10
ARE AVAILABLE FOR SALE. SEE SECTION 10, REPLACEMENT PARTS, FOR A LIST OF AVAILABLE PARTS.
3
1. Screw
2. Washer Heater Strut Assembly
3. O
2
4. Contact/Thermocouple
Assembly
5. Strut Bracket
6. Spring
7. Spring Clip
8. Screw
9. Washer
10. Heater Assembly
11. Heater Tube
12. Gasket Cell
13. O
2
14. Cap Screw
15. Cell Kit (includes 12, 13,
and 14)
11
12
15
13
14
Figure 7-5. O2 Assembly, Exploded View
37020012
7-8 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 59
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
Use care when handling contact and thermocouple assembly. The ceramic rod in this assembly is fragile.
7. If replacing contact/thermocouple as­sembly (4), use a pencil to mark loca­tion of spring clip (7) before removing. Squeeze tabs on spring clip to remove. Retain spring clip and spring (6); re­place if damaged.
8. While carefully handling new contact and thermocouple assembly (4), lay old assembly next to new one. Transfer pencil marks to new assembly.
9. Carefully guide new contact and ther­mocouple assembly (4) through strut bracket (5), spring (6), and spring clip (7) until spring clip reaches pencil mark.
10. Carefully slide new O
heater strut as-
2
sembly (3) into heater tube (11). Press down on the back plate of strut bracket (5) to ensure spring (6) tension is pres­ent to hold contact pad against O (13). Secure with four screws (1) and washers (2).
11. Reconnect the O
assembly wire leads
2
to the sensor housing terminal blocks. Refer to Figure 7-2 if necessary.
cell
2
Use heat-resistant gloves and clothing when removing the OCX 4400. Do not attempt to work on any sensor hous­ing components until they have cooled to room temperature. Components can be as hot as 2600°F (1427°C). This can cause severe burns.
Disconnect and lock out power before working on any electrical components. There may be voltage up to 264 VAC.
1. Follow the instructions in paragraph 7-2a.1 to remove the OCX 4400 from the stack or duct. If removing a split­mounted unit, follow the instructions in paragraph 7-2b.1 to remove the sensor housing.
2. Remove sensor housing cover (1, Figure 7-4).
3. Disconnect and tag the four wire leads from COMB sensor (15) to the sen­sor/thermocouple terminal block.
4. Remove two screws (13) securing plate (14) to holder (23). Remove the plate.
5. Remove old COMB sensor (15) and gasket (16). Replace with new COMB sensor and gasket from replacement kit.
12. Replace reference air tube (2, Figure 7-4) in sensor housing (24). Ensure the open end of the reference air tube ex­tends into O
assembly (3).
2
13. Replace sensor housing cover (1).
14. Follow the instructions in paragraph 7-2a.2 to install the OCX 4400 into the stack or duct. If installing a split­mounted unit, follow the instructions in paragraph 7-2b.2.
c. COMB Sensor Replacement
This paragraph covers combustible sensor replacement. Do not attempt to replace the sensor until all other possibilities for poor performance have been considered. If sen­sor replacement is needed, order the com­bustibles sensor replacement kit from Table 10-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-9
Use care when handling the new com­bustibles sensor. The RTDs are fragile.
6. Secure new COMB sensor (15) with plate (14) to holder (23) with two screws (13).
7. Reconnect the four wire leads from COMB sensor (15) to the sen­sor/thermocouple terminal block. Refer to Figure 7-2 if necessary.
8. Replace sensor housing cover (1, Figure 7-4).
9. Follow the instructions in paragraph 7-2a.2 to install the OCX 4400 into the stack or duct. If installing a split­mounted unit, follow the instructions in paragraph 7-2b.2.
Page 60
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
10. Turn on power and monitor COMB sensor output. After the OCX 4400 stabilizes, calibrate the unit. If new components have been installed, re­peat calibration after 24 hours of operation.
7-4 ELECTRONICS HOUSING COMPONENT
REPLACEMENT
OCX 4400
When installing a solenoid, do not overtighten. Damage to the solenoid may occur.
4. Install the new solenoid base. Be careful not to overtighten. Install the new washer and coil assembly and se­cure with the top nut.
Each of the following procedures details how to remove and replace a specific component lo­cated in the electronics housing of the OCX
4400.
Disconnect and lock out power before working on any electrical components. There may be voltage up to 264 VAC.
a. Solenoid Replacement
The OCX 4400 has three optional calibra­tion gas solenoids (HI COMB, HI O
) and a standard, 3-way instrument air
O
2
solenoid. Use the following procedure to re­place one of four solenoids (11, Figure 7-6) in the electronics housing. The replacement procedure for all solenoids is similar.
1. Follow the instructions in paragraph 7-2a to remove the OCX 4400 from the stack or duct. If removing a split­mounted unit, follow the instructions in paragraph 7-2b.3 to remove the elec­tronics housing.
, LO
2
5. Connect the leads to the proper termi­nations on the solenoid power terminal block. Refer to Figure 2-6 if necessary.
6. Follow the instructions in paragraph 7-2a.2 to install the OCX 4400 into the stack or duct. If installing a split­mounted unit, follow the instructions in paragraph 7-2b.4 to install electronics housing.
b. Pressure Sensor Replacement
Use the following procedure to replace pressure sensor (10, Figure 7-6).
1. Follow the instructions in paragraph 7-2a to remove the OCX 4400 from the stack or duct. If removing a split­mounted unit, follow the instructions in paragraph 7-2b.3 to remove the elec­tronics housing.
2. Tag and remove pressure sensor (10) leads from sensor/thermocouple termi­nal block.
3. Place a 3/4 in. deep socket over pres­sure sensor (10) and remove from electronics housing base (12).
2. Tag and disconnect wire leads from solenoid (11, Figure 7-6) to be re­placed at the solenoid power terminal block.
3. Remove top nut of solenoid (11) se­curing the coil assembly and washer to the solenoid base. Remove the coil as­sembly, including the leads, and washer. Place a 13/16 in. deep socket over the solenoid base and remove from electronics housing base (12).
7-10 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
When installing the pressure sensor, do not overtighten. Damage to the pressure sensor may occur.
4. Install new pressure sensor (10) into electronics housing base (12).
5. Connect leads to the proper termina­tions on the sensor/thermocouple ter­minal block. Refer to Figure 2-7 if necessary.
Page 61
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
1. Cover
2. Screw
3. Local Operator Interface
4. Washers
5. Spacer
6. Connector
7. Electronics Package
8. Screw
9. Washer
10. Solenoid
11. Housing Base
12. Fitting
13. Screw
14. Nameplate
15. Instrument Air Solenoid
16. Vent Tube
NOTES: NOT ALL ITEMS IDENTIFIED
ARE AVAILABLE FOR SALE. SEE SECTION 10, REPLACEMENT PARTS, FOR A LIST OF AVAILABLE PARTS.
2
4
5
6
16
15
14
13
1
3
7
8
9
11
10
12
37020013
Figure 7-6. Electronics Housing Exploded View
6. Follow the instructions in paragraph 7-2a.2 to install the OCX 4400 into the stack or duct. If installing a split­mounted unit, follow the instructions in paragraph 7-2b.4 to install electronics housing.
c. Electronics Package Replacement
1. Follow the instructions in paragraph 7-2a.1 to remove the OCX 4400 from the stack or duct. If removing a split­mounted unit, follow the instructions in paragraph 7-2b.3 to remove the elec­tronics housing.
2. Disconnect and tag wire leads from solenoids and the pressure sensor to
Use the following procedure to replace the electronics package (7, Figure 7-6).
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-11
the solenoid power and sensor/ thermocouple terminal blocks.
Page 62
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
3. If replacing components on an integral electronics mounted unit, disconnect and tag wire leads from the sensor housing at the sensor/thermocouple, heater power, and solenoid power ter­minal blocks.
4. Remove four screws (8, Figure 7-6) and washers (9) securing electronics box (7) to electronics housing base (12). Remove the electronics package.
5. If equipped, install local operator inter­face (3, Figure 7-6) to electronics package (7) with three screws (2), double washers (4), spacers (5), and connector (6).
6. If replacing components on an integral electronics mounted unit, connect wire leads from the sensor housing to sen­sor/thermocouple, heater power, and solenoid power terminal blocks.
7. Connect wire leads from the solenoids and the pressure sensor to the sole­noid power and sensor/thermocouple terminal blocks.
8. Follow the instructions in paragraph 7-2a.2 to install the OCX 4400 into the stack or duct. If installing a split­mounted unit, follow the instructions in
paragraph 7-2b.4 to install electronics housing.
d. Fuse Replacement
The OCX 4400 has two fuses (1 and 3, Figure 7-7). The power supply fuse, F1, is to the left of the AC power terminal block. The heater fuse, F2, is to the right of the AC power terminal block. Both fuses are acces­sible through holes in the electronics box. Refer to Table 10-1 for replacement fuse specifications. Perform the following proce­dure to check or replace fuses.
Disconnect and lock out power before working on any electrical components. There may be voltage up to 264 VAC.
1. Turn off power to the system.
2. Remove electronics housing cover (1, Figure 7-6).
3. Remove fuseholder (2, Figure 7-7) by pushing in and turning 1/4 turn coun­terclockwise. Remove fuse (1 or 3).
4. After checking or replacing fuse (1 or
3), install fuseholder (2) by pushing in and turning 1/4 turn clockwise.
ELECTRONICS
HOUSING
1
1. Fuse – Heater, F2
2. Fuse Holder
3. Fuse – Power Supply, F1
3
2
37020014
Figure 7-7. Fuses, Exploded View
7-12 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 63
OCX 4400
8
Install all protective equipment covers and safety ground leads after trouble­shooting. Failure to install covers and ground leads could result in serious injury or death.
8-1 GENERAL
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 8

TROUBLESHOOTING

Before troubleshooting the system, ensure all ICs are fully seated.
d. Electrostatic Discharge
Electrostatic discharge can damage the ICs used in the electronics. Before removing or handling the processor board or the ICs, ensure you are at ground potential.
The troubleshooting section describes how to identify and isolate faults that may develop in the OCX 4400. When troubleshooting the OCX 4400, reference the following information.
a. Grounding
It is essential that adequate grounding pre­cautions are taken when installing the sys­tem. Thoroughly check both the probe and electronics to ensure the grounding quality has not degraded during fault finding. The system provides facilities for 100% effective grounding and the total elimination of ground loops.
b. Electrical Noise
The OCX 4400 has been designed to oper­ate in the type of environment normally found in a boiler room or control room. Noise suppression circuits are employed on all field terminations and main inputs. When fault finding, evaluate the electrical noise being generated in the immediate circuitry of a faulty system. Also, ensure all cable shields are connected to earth.
c. Loose Integrated Circuits
The OCX 4400 uses a microprocessor and supporting integrated circuits (IC). If the electronics are handled roughly during in­stallation or located where subjected to se­vere vibration, the ICs could work loose.

8-2 ALARM OUTPUT

a. The OCX 4400 is equipped with a set of
alarm relay contacts on the microprocessor board in the electronics housing. This set of dry contacts can be connected to any cus­tomer supplied relay device, 30 VDC, 2 Amp maximum. Any fault condition in the OCX 4400 will trip the alarm relay.
b. The optional SPA with HART programmable
alarm indicates LOW O Calibration Status, and Unit Failure. For more information refer to Appendix B, SPA with HART Alarm.

8-3 FAULT ISOLATION

Faults in the OCX 4400 Transmitter are indi­cated by messages displayed on the HART communicator or LOI. Fault indications that can appear are listed in the PROBLEM column of Table 8-1, Troubleshooting.
If a fault is indicated on the HART communicator or LOI, locate the fault indication in the PPO­BLEM column of Table 8-1. For each fault listed, there are related PROBABLE CAUSES and RECOMMENDED CORRECTIVE ACTIONS. The PROBABLE CAUSES are listed in the order of most probable to least probable. Starting with the most probable cause, inspect and test the unit to isolate the actual cause, then use the RECOMMENDED CORRECTIVE ACTION listed to correct the problem.
, HIGH COMB,
2
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 8-1
Page 64
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
Table 8-1. Troubleshooting
PROBLEM PROBABLE CAUSE RECOMMENDED CORRECTIVE ACTION
OCX 4400
O
2
Thermocouple Error
(thermocouple voltage out of limits)
O2 Cell Temperature Low
(< 726°C)
O2 Cell Temperature High (> 746°C) or Very High (> 820°C)
O2 thermocouple or thermocouple circuit open
O2 thermocouple or thermocouple circuit shorted
O
thermocouple
2
wires reversed
Sensor housing ex-
Check O2 thermocouple and circuit wires for breaks or loose connection. Repair break or loose connection or replace failed thermocouple. Cycle OCX power to resume operation.
Check O
thermocouple and thermocouple circuit wires for
2
short circuit condition. Repair shorted wiring or replace faulty thermocouple. Cycle OCX power to resume operation.
Check O2 thermocouple wiring per Figure 2-7. Correct reversed-wires fault. Cycle OCX power to resume operation.
Install sensor housing insulating jacket. posed to wind or very cold temperatures.
O2 cell heater open Check O2 cell heater circuit for broken wire or loose con-
nection. Repair broken wire or loose connection.
Heater electronics failure
Check heater fuse F2 in electronics housing. If open, locate
and correct cause of overload. If F2 is not open or if cause
of overload cannot be found, replace electronics package.
Transmitter in warm­up mode
High noise in OCX power supply
Electronics package
During warm-up, heater temperature rise may surpass
heater control response. Cycle power to resume operation.
Check power supply for line noise or voltage surges. Install
high quality line filter if high noise or voltages are indicated.
Replace electronics package. power supply failure
O2 Heater Open
O2 Cell Circuit Open
or
O
Cell Bad
2
O2 heater or heater circuit wiring open.
Check O2 cell heater circuit for broken wire or loose con-
nection. Repair broken wire or loose connection.
Check resistance of the O
cell heater; heater resistance
2
should be less than 100 ohms. Replace heater if heater coil
circuit is open.
Transmitter in power­up or warm-up mode
Loose or open O2 cell lead or circuit wire
During power-up and warm-up, low heater temperature give
false O
Check O
indication. Cycle power to resume operation.
2
cell tip and cell circuit wires for breaks or loose
2
connection. Repair cell lead wire break or loose connection. connection
O2 cell failure Check O2 cell impedance value at LOI (see Figure A-3,
sheet 2). If cell impedance is zero, replace O
cell.
2
If cell impedance is less than 500 ohms, check for cell
housing ground fault. Repair ground fault.
If cell impedance is greater than 500 ohms and no ground
Very low O
concen-
2
fault is indicated, replace O
Cycle power to resume operation.
cell.
2
tration in process gas
8-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Page 65
OCX 4400
8
Table 8-1. Troubleshooting (Continued)
PROBLEM PROBABLE CAUSE RECOMMENDED CORRECTIVE ACTION
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
O2 Slope Error
slope
(O
2
> 57.5 or < 34.5 mV/Dec)
O2 Constant Error
(> 20 mV or < –20 mV)
or
O
Calibration
2
Failed
High Electronics Temperature
Calibration gas sup­ply low or gas con­nection leaking
Check calibration gas supplies and connections. Adjust gas
pressure and flow. Replenish low calibration gas supplies
and tighten or repair loose or leaking connections. When
calibration gas supplies are adequate, recalibrate.
O2 cell degraded Recalibrate the OCX.
O
cell failure Check O2 cell impedance value at LOI (see Figure A-3,
2
sheet 2). If cell impedance is zero, replace O
cell.
2
If cell impedance is less than 500 ohms, check for cell
housing ground fault. Repair ground fault.
If cell impedance is greater than 500 ohms and no ground
Calibration gas sup­ply low or gas con­nection leaking
fault is indicated, replace O
Check calibration gas supplies and connections. Adjust gas
pressure and flow. Replenish low calibration gas supplies
and tighten or repair loose or leaking connections. When
cell.
2
calibration gas supplies are adequate, recalibrate.
O2 cell degraded Recalibrate the OCX.
O
cell failure Check O2 cell impedance value at LOI (see Figure A-3,
2
sheet 2). If cell impedance is zero, replace O
cell.
2
If cell impedance is less than 500 ohms, check for cell
housing ground fault. Repair ground fault.
If cell impedance is greater than 500 ohms and no ground
Electronics housing exposed high ambi­ent temperature.
fault is indicated, replace O
Insulate housing from source of high temperature and/or
install cooling fan to remove heat from housing. Cycle OCX
power to resume operation.
cell.
2
(> 85°C)
EEPROM Corrupt
Unit powered down during calibration
With power on, wait five minutes, cycle power and recali-
brate the OCX. parameter storage
EEPROM failure Replace EEPROM.
Sample Line Thermocouple Error
(voltage out of limits)
Sample line thermo­couple wires reversed.
Sample line thermo­couple or thermo­couple circuit open.
Check thermocouple wiring per Figure 2-7. Correct
reversed-wires fault. Cycle OCX power to resume operation.
Check thermocouple and thermocouple circuit wires for
breaks or loose connection. Repair break or loose connec-
tion or replace failed thermocouple. Cycle OCX power to
resume operation.
Sample line thermo­couple or wiring cir­cuit shorted.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 8-3
Check thermocouple and thermocouple circuit wires for
short circuit condition. Repair shorted wiring or replace
faulty thermocouple. Cycle OCX power to resume operation.
Page 66
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
Table 8-1. Troubleshooting (Continued)
PROBLEM PROBABLE CAUSE RECOMMENDED CORRECTIVE ACTION
OCX 4400
Sample Line Temperature Low
(< 155°C; unit not in warm­up mode)
Sample Line Temperature High (> 195°C) or Very High (> 220°C)
Combustibles Cell Error
(cell temp. < –40°C;
84.27 mV or > 400°C;
247.04 mV ref. side or active side)
Sensor housing ex­posed to wind or very cold temperatures.
Sample line heater open
Heater electronics failure
Transmitter in warm­up mode
High noise in OCX power supply
Electronics package power supply failure
COMB sensor circuit wiring
COMB sensor RTD shorted or open (reference or active)
Electronics package power supply failure
Install sensor housing insulating jacket.
Check sample line heater circuit for broken wire or loose
connection. Repair broken wire or loose connection.
Check heater fuse F2 in electronics housing. If open, locate
and correct cause of overload. If F2 is not open or if cause
of overload cannot be found, replace electronics package.
During warm-up, heater temperature rise may surpass
heater control response. Cycle power to resume operation.
Check power supply for line noise or voltage surges. Install
high quality line filter if high noise or voltages are indicated.
Replace electronics package.
Check COMB sensor circuit for broken wire or loose con-
nection. Repair broken wire or loose connection.
Check RTD resistance (lead to ground) of reference and
active RTDs. If either RTD is open or shorted to ground,
replace COMB sensor.
Replace electronics package.
Combustibles Temperature Low
(< 290°C; unit not in warm­up mode)
Combustibles Temperature High (> 310°C) or Very High (> 400°C)
8-4 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Sensor housing ex­posed to wind or very cold temperatures.
COMB cell heater open
Heater electronics failure
Transmitter in warm­up mode
Large rise in COMB concentration.
High noise in OCX power supply
Electronics package power supply failure
Install sensor housing insulating jacket.
Check COMB cell heater circuit for broken wire or loose
connection. Repair broken wire or loose connection.
Check heater fuse F2 in electronics housing. If open, locate
and correct cause of overload. If F2 is not open or if cause
of overload cannot be found, replace electronics package.
During warm-up, heater temperature rise may surpass
heater control response. Cycle power to resume operation.
High temperature rise in cell may surpass heater control re-
sponse. Cycle power to resume operation.
Check power supply for line noise or voltage surges. Install
high quality line filter if high noise or voltages are indicated.
Replace electronics package.
Page 67
OCX 4400
8
Table 8-1. Troubleshooting (Continued)
PROBLEM PROBABLE CAUSE RECOMMENDED CORRECTIVE ACTION
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
Combustibles Slope Error
(< 0 or > 3000 ppm/mV)
or
Combustibles Constant Error
or
Combustibles Calibration Failed
(slope error during last calibration)
Calibration gas sup­ply low or gas con­nection leaking
Check calibration gas supplies and connections. Adjust gas
pressure and flow. Replenish low calibration gas supplies
and tighten or repair loose or leaking connections. When
calibration gas supplies are adequate, recalibrate.
O2 cell degraded Recalibrate the OCX.
O
cell failure Check O2 cell impedance value at LOI (see Figure A-3,
2
sheet 2). If cell impedance is zero, replace O
cell.
2
If cell impedance is less than 500 ohms, check for cell
housing ground fault. Repair ground fault.
If cell impedance is greater than 500 ohms and no ground
fault is indicated, replace O
cell.
2
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 8-5
Page 68
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
8-6 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Page 69
OCX 4400
9
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 9

RETURN OF MATERIAL

9-1 If factory repair of defective equipment is re-
quired, proceed as follows:
a. Secure a return authorization number from
a Rosemount Analytical Sales Office or rep­resentative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
In no event will Rosemount be responsible for equipment returned without proper authorization and identification.
b. Carefully pack defective unit in a sturdy box
with sufficient shock absorbing material to insure that no additional damage will occur during shipping.
c. In a cover letter, describe completely:
1. The symptoms from which it was de­termined that the equipment is faulty.
2. The environment in which the equip­ment has been operating (housing, weather, vibration, dust, etc.).
5. Complete shipping instructions for re­turn of equipment.
6. Reference the return authorization number.
d. Enclose a cover letter and purchase order
and ship the defective equipment according to instructions provided in Rosemount Re­turn Authorization, prepaid, to:
Rosemount Analytical Inc. RMR Department 1201 N. Main Street Orrville, Ohio 44667
If warranty service is requested, the defec­tive unit will be carefully inspected and tested at the factory. If failure was due to conditions listed in the standard Rosemount warranty, the defective unit will be repaired or replaced at Rosemount's option, and an operating unit will be returned to the cus­tomer in accordance with shipping instruc­tions furnished in the cover letter.
3. Site from which equipment was removed.
4. Whether warranty or nonwarranty service is requested.
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 9-1
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.
Page 70
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
9-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Page 71
Instruction Manual
10
IB-106-440 Rev. 2.0
OCX 4400
SECTION 10

REPLACEMENT PARTS

Table 10-1. Replacement Parts
FIGURE and
INDEX No. PART NUMBER DESCRIPTION
1-1, 2 275D9EI5D0000 275 HART Communicator with 12 meg buffer 1-1, 3 3535B30G02 Adapter Plate 218 w/shield to ANSI OCX 1-1, 3 3535B30G08 FLG Adapter Assembly, 132 to ANSI OCX 1-1, 3 3535B31G01 OCX/132 Adapter Plate (DIN) 1-1, 5 263C152G01 Reference Gas Set, includes:
771B635H01 Flowmeter, 0.2-10 SCFH 771B635H08 Flowmeter, 0.05-0.5 SCFH 1A99422H01 Pressure Regulator/Filter, 0 – 60 psig 275431-003 Pressure Gage, 2 in., 0 – 60 psi
1-1, 6 1A99518G01 Blowback Kit
1-4 4851B41G01 Cable, 20 ft. (6m) NEMA 1-4 4851B41G02 Cable, 40 ft. (12m) NEMA 1-4 4851B41G03 Cable, 60 ft. (18m) NEMA 1-4 4851B41G04 Cable, 80 ft. (24m) NEMA 1-4 4851B41G05 Cable, 100 ft. (30m) NEMA 1-4 4851B41G06 Cable, 150 ft. (45m) NEMA
January 2003
2-1 3535B18H02 ANSI Gasket 2-8 1A99119G05 Test Gas Regulator Kit 2-8 1A99119G04 Test Gas Bottles Kit 2-8 1A99119H01 Oxygen Bottle 0.4%, Balance N 2-8 1A99119H02 Oxygen Bottle 8.0%, Balance N 2-8 1A99119H06 CO Bottle 1000 ppm, Balance N 2-8 1A99120H02 Regulator for Oxygen 2-8 1A99120H03 Regulator for Combustibles
1A99119G06 Wall Mount Bracket for Gas Bottles (not shown)
7-4, 1 5R10219G01 Cover, Sensor Housing 7-4, 3 and 7-5, 15 4851B45G01 O2 Cell Replacement Kit 7-4, 12 1A99371H01 Combustibles Chamber Heater 7-4, 13 thru 16 4851B46G01 Combustibles Sensor Replacement Kit 7-4, 18 1A98151H03 Sample Line Thermocouple Assembly 7-4, 19 1A99371H02 Sample Line Heater 7-4, 25 5R10183H02 Tube, Sample, 18 in. (457mm) 316 Stainless steel 7-4, 25 5R10183H06 Tube, Sample, 18 in. (457mm) Inconel 600 7-4, 25 5R10227G01 Tube, Sample, 18 in. (457mm) Ceramic 7-4, 25 5R10183H03 Tube, Sample, 3 ft. (0.91m) 316 Stainless steel 7-4, 25 5R10183H07 Tube, Sample, 3 ft. (0.91m) Inconel 600 7-4, 25 5R10227G02 Tube, Sample, 3 ft. (0.91m) Ceramic 7-4, 25 5R10183H04 Tube, Sample, 6 ft. (1.83m) 316 Stainless steel 7-4, 25 5R10183H08 Tube, Sample, 6 ft. (1.83m) Inconel 600 7-4, 25 5R10183H05 Tube, Sample, 9 ft. (2.7m) 316 Stainless steel 7-4, 26 5R10246H02 Filter, 20 micron 7-4, 30 5R10183H01 Tube, Exhaust
2
2
2
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 10-1
Page 72
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
Table 10-1. Replacement Parts (Continued)
FIGURE and
INDEX No. PART NUMBER DESCRIPTION
7-5, 3 5R10211G01 Heater Strut Assembly 7-5, 4 4851B44G01 Contact and Thermocouple Assembly
7-6, 1 5R10199G01 Cover, Electronics Housing 7-6, 3 4851B49G02 Local Operator Interface 7-6, 7 5R10214G01 Electronics Package 7-6, 11 1A97905H01 Solenoid Valve, Cal. Gas 7-6, 16 1A97905H02 Solenoid Valve, Instrument Air
7-7, 1 1A97913H18 Heater Fuse, Slo-Blo, 6.3 amp, 250 VAC, F2 7-7, 3 1A97913H04 Power Supply Fuse, Slo-Blo, 1.0 amp, 250 VAC, F1
B-1 1A99292H01 Moore Industries SPA for HART Alarm C-1 1A99339H03 Blowback Valve, Air Operated D-1 1A99376H01 IS Barrier Module, Model MTL 5046
D-1 1A99698G01 Intrinsic Safety Barrier Kit
OCX 4400
10-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Page 73
OCX 4400
11
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 11

APPENDICES

APPENDIX A LOCAL OPERATOR INTERFACE (LOI)
APPENDIX B SPA WITH HART ALARM
APPENDIX C BLOWBACK
APPENDIX D IS BARRIER
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 11-1
Page 74
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
11-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Page 75
OCX 4400
A
Instruction Manual
Appendix A Rev 2.0
January 2003

APPENDIX A

LOCAL OPERATOR INTERFACE (LOI)

A-1 FORWARD
This appendix describes the installation and op­eration of the Local Operator Interface (LOI) for the OCX 4400.
CONNECTOR
A-2 INSTALLATION
The LOI mounts to the LOI connector on the LOI board located at the end of the electronics box in the electronics housing of an OCX 4400 Oxygen/Combustibles Transmitter Figure A-1. There are four matching connectors on the back of the LOI assembly that allow the LOI to be ori­ented as desired by the user.
ELECTRONIC HOUSING
(COVER REMOVED)
LOI
LOI
ASSEMBLY
ELECTRONICS
LOI
BOARD
BOX
29790024
Figure A-1. LOI Assembly Mounting
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-1
Page 76
Instruction Manual
Appendix A Rev. 2.0 January 2003
OCX 4400
A-3 OPERATION
The LOI assembly, Figure A-2, allows local communication with the electronics. The LOI uses infrared sensors near the selection arrows to move the cursor between menus and menu items. The LCD display window shows the op­erating display, which consists of the O COMB ppm, current menu, or item selected. When a blinking cursor is displayed, the UP or DOWN arrows can be used to increase or de­crease the value for that item.
SELECTION
ARROWS
% and
2
LCD DISPLAY WINDOW
A-4 LOI MENU TREE
This section consists of a menu tree for the LOI on the OCX 4400, Figure A-3. This menu is specific to the OCX 4400.
a. First Column Sub-Menus
From the operating display, O COMB ppm, the left-pointing arrow is the only option to move into the first column sub-menus of the LOI menu tree. The first column contains three sub-menus: SENSOR DATA, Figure A-3 sheet 1 of 3, CALIBRATE, sheet 2 of 3, and SETUP, sheet 3 of 3. From the operating display, SENSOR DATA is displayed when the right arrow is selected. Use the up or down­pointing arrow to move to the other first col­umn sub-menus.
b. Second Column Sub-Menus
From the first column sub-menus, selecting the right-pointing arrow moves the display into the second column sub-menus. The up and down-pointing arrows allow the display to move to the second column sub-menus of the first column sub-menu selected. The left-pointing arrow moves the display back to the first column sub-menu.
% and
2
c. Third, and Fourth Column Sub-Menus
SELECTION
ARROWS
Figure A-2. Local Operator Interface (LOI)
Assembly
From the second column sub-menus, se­lecting the right-pointing arrow moves the display into the third column sub-menus. The third column sub-menu may be another menu or a list of parameters. The up and down-pointing arrows allow the display to move to the different parameters or menus. The third or fourth column may be a pa­rameter list. When a parameter list is dis­played, the cursor will blink. The up and down-pointing arrows select the value for the parameter displayed.
A-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Page 77
OCX 4400
A
O2 X.XX% Comb XXX ppm
Instruction Manual
Appendix A Rev 2.0
January 2003
O2 Sensor _____dgC
Te mp / Pressure
%
%
Sensor Data
Voltages
Output Values
Comb Sensor _____dgC
Sample Block Temp _____dgC
Sample Block Press _____psi
Electronics _____
Max Elect Temp _____dgC
O2 Sensor _____mV
O2 Sensor T/C _____mV
Comb Delta _____mV
Comb Reference _____mV
Elect Temp _____V
Smpl Blok T/C _____mV
Smpl Line Pres _____V
O2 Analog _____%
Comb Analog _____%
O2 Analog mA ____mA
Comb Analog mA ____mA
(CONTINUED ON
SHEET 2)
36390006
Figure A-3. Menu Tree for Local Operator Interface on the OCX 4400 (Sheet 1 of 3)
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-3
Page 78
Instruction Manual
Appendix A Rev. 2.0 January 2003
OCX 4400
(CONTINUED FROM
SHEET 1)
Calibrate
(CONTINUED ON
SHEET 3)
NOTE:
“Hit E when ready” prompt is displayed during semi­automatic calibration only (when Setup Calibration value “Got Solenoids” = “n”).
Do Calibrate
View Constants
Calib Status
O2
Combustibles
Both
Latest Calib
Prev Calib
Calibrate Step Calibrate Time O2 Next Calib O2 Slope O2 Constant Comb Next Cal Comb Slope Comb Constant
Apply Lo O2 Gas
Hit E when ready
Flow O2 Lo _____s
_____% _____mV
Read O2 Lo _____s
_____% _____mV
Apply Hi O2 Gas
Hit E when ready
Flow O2 Hi _____s
_____% _____mV
Read O2 Hi _____s
_____% _____mV
Hit E when ready
Purge _____s O2 Calibration Pass/Fail
Apply Lo CO Gas
Hit E when ready
Flow CO Lo _____s
_____% _____mV
Read CO Lo _____s
_____% _____mV
Apply Hi CO Gas
Hit E when ready
Flow CO Hi _____s
_____% _____mV
Read CO Hi _____s
_____% _____mV
Hit E when ready
Purge _____s
Apply Lo O2 Gas
Hit E when ready
Flow O2 Lo _____s
_____% _____mV
Read O2 Lo _____s
_____% _____mV
Apply Hi O2 Gas
Hit E when ready
Flow O2 Hi _____s
_____% _____mV
Read O2 Hi _____s
_____% _____mV
Hit E when ready
Purge _____s
Apply Lo CO Gas
Hit E when ready
Flow CO Lo _____s
_____% _____mV
Read CO Lo _____s
_____% _____mV
Apply Hi CO Gas
Hit E when ready
Flow CO Hi _____s
_____% _____mV
Read CO Hi _____s
_____% _____mV
Hit E when ready
Purge _____s
O2 Slope _____mV/D O2 Constant _____mV O2 Cell Imped _____ Comb Slope _____ppm/mv Comb Constant _____mV
Pre O2 Slope _____mV/D Pre O2 Const _____mV Pre Co Slope _____ ppm/mV Pre Co Const _____mV
36390007
Figure A-3. Menu Tree for Local Operator Interface (LOI) on the OCX 4400 (Sheet 2 of 3)
A-4 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Page 79
OCX 4400
A
(CONTINUED FROM
SHEET 2)
SETUP
CALIBRATION
PARAMETERS
BLOWBACK
O2 High Gas _____% O2
O2 Low Gas _____% O2
Comb Test Gas _____% O2
Got Solenoids Yes/No O2 Output Tracks Yes/No Comb
Out Tracks Yes/No
O2 Cal Interval XDXH
O2
Next Cal XH
Comb Cal Interval XDXH
Comb Next Cal XH
Gas Time 0:00
Purge Time 0:00
Line voltage _______
O2 Slope _____ mV/D
O2 Constant _____ mV
Comb Slope _____
Comb Constant _____
Blow Bk Present Yes/No
Blow Bk Intvl 99999 min
Blow Bk Durat 99999 sec
Blow Bk Purge 999 sec
Do Blow Back Yes/No
Instruction Manual
Appendix A Rev 2.0
January 2003
ALARM RELAY
ANALOG OUTPUTS
SOFTWARE
Figure A-3. Menu Tree for Local Operator Interface (LOI) on the OCX 4400 (Sheet 3 of 3)
A-5 AUTOMATIC CALIBRATION SETUP
Use the following procedure to setup the OCX 4400 with the LOI for automatic calibration. If necessary, use the menu tree in Figure A-3 for reference. The unit must be equipped with calibration solenoids to use automatic calibra­tion.
Trigger 1 Event Trigger 2 Event Trigger 3 Event
O2 Type 20-4 mA
Comb Type 20-4 mA
O2 Range 0-VAR (40% Max.)
Comb Range 50,000 ppm
Do O2 Out Trim Do COMB Out Trim
Version Checksum Build Number Build Date Test Code
4-20 mA
4-20 mA
HART
a. From the operating display use the right-
b. From the SETUP first column sub-menu,
Off All Comb Temp Sample Pr Sen Sample Ln Temp Unit Fail Hi Elect Temp Calib Fail O2 Cell Bad O2 Htr Open O2 Cell Temp In Calibrtn Off
37020024
pointing arrow to select SETUP first col­umn sub-menu.
use the right-pointing arrow to select the CALIBRATION second column sub-menu.
NOTE
c. From the CALIBRATION second column
sub-menu, use the right-pointing arrow to
Automatic calibration is only avail-
select the third column parameter list.
able on units equipped with calibra­tion solenoids.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-5
Page 80
Instruction Manual
Appendix A Rev. 2.0 January 2003
OCX 4400
d. Scroll down to the forth item Got Sole-
noids. If the unit is equipped with calibra­tion solenoids and timed automatic calibration is desired, select Yes.
e. Use the down-pointing arrow to select the
next item O
Output Tracks. Select Yes or
2
No to determine if updates to the O value will take place.
f. Use the down-pointing arrow to select the
next item Comb Out Tracks. Select Yes or No to determine if updates to the COMB lock value will take place.
g. Use the down-pointing arrow to select the
next item O
Cal Interval. Enter the
2
amount of time in days and hours that is desired between automatic calibrations.
h. Use the down-pointing arrow to select the
next item O
Next Cal. Enter the amount
2
of time in hours until the next automatic calibration. Select the left-pointing arrow three times to move back to the operating display.
A-6 MANUAL CALIBRATION WITH LOI
Use the following procedure to perform a manual calibration with the LOI. If necessary, refer to the menu tree in Figure A-3 if needed. Once the manual calibration procedure is ini­tiated by the LOI, a series of prompts will ap­pear giving instructions to the operator.
a. Use the right-pointing arrow to select the
CALIBRATE first column sub-menu.
lock
2
e. Use the right-pointing arrow to start the
calibration. Turn on the low O when prompted by “Apply Lo O
test gas,
2
Gas”.
2
f. Press the right-pointing arrow when the
low O
test gas is applied. The calibration
2
data changes as the calibration proceeds.
g. Press the right-pointing arrow when the
low O
reading is stable. Turn off the low
2
O
test gas and turn on the high O2 test
2
gas as prompted by “Apply Hi O
Gas”.
2
h. Press the right-pointing arrow when the
high O
test gas is applied. The calibration
2
data changes as the calibration proceeds.
i. Press the right-pointing arrow when the
high O O to start the high O
reading is stable. Turn off the high
2
test gas. Press the right-pointing arrow
2
gas purge.
2
j. When the Purge period expires, the LOI
displays “O
Calibration Pass or Fail”.
2
(Calibration Fail exits the program.)
k. Press the right-pointing arrow to start
combustibles calibration. Turn on the low CO test gas, when prompted.
l. Press the right-pointing arrow when the
low CO test gas is applied. The calibration data changes as the calibration proceeds.
m. Press the right-pointing arrow when the
low CO reading is stable. Turn off the low CO test gas and turn on the high CO test gas when prompted.
b. From the CALIBRATE sub-menu use the
right-pointing arrow to select the DO CALIBRATION second column sub-menu.
n. Press the right-pointing arrow when the
high CO test gas is applied. The calibra­tion data changes as the calibration pro­ceeds.
c. From the DO CALIBRATION sub-menu
use the right-pointing arrow to select the third column O
d. Remain at the O
sensor calibrate option.
2
sensor option or use the
2
down-pointing arrow to calibrate COM­BUSTIBLES or BOTH sensors. (The fol­lowing sequence applies when BOTH sensors are selected.)
A-6 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
o. Turn off the high CO test gas and press
the right-pointing arrow to start the high CO
gas purge.
p. When the Purge period expires, the LOI
displays “CO Calibration Pass or Fail
”. A failure reapplies the last good calibration parameters for O
/combustibles sensing.
2
Page 81
OCX 4400
B
B-1 FORWARD
Instruction Manual
Appendix B Rev 2.0
January 2003

APPENDIX B

SPA WITH HART ALARM

This appendix describes the SPA with HART Alarm option available for the OCX 4400.
B-2 DESCRIPTION
The Moore Industries SPA with HART Alarm, Figure B-1, is a 4-wire (line or mains powered), site-programmable, digital process alarm. It connects to a standard HART field device, and provides up to four, fully configurable, contact­closure outputs based on readings of the HART digital data. The four outputs from the OCX 4400 which are recognized by the SPA are Low O
, High COMB, Calibration Status, and OCX
2
Unit Failure.
B-3 INSTALLATION
Refer to Figure B-2 for the typical interface con­nections for the OCX 4400 and the SPA with HART alarm. Refer to the Moore Industries user’s manual provided with the SPA for addi­tional information concerning SPA installation, setup, and operation.
Electrostatic discharge (ESD) protec­tion is required to avoid damage to the SPA electronic circuits.
1. Refer to Figure B-3. Turn the SPA over and slide the access cover out. Before changing any jumper or switch posi­tion, take adequate precautions to avoid an electrostatic discharge.
B-4 SETUP
Setup of the SPA for communication with the OCX 4400 includes setting internal jumpers and dip switches and configuring the SPA operating parameters via a menu-driven selection and calibration procedure.
a. Jumper and Switch Settings
37020018
SPA jumper and switch settings are shown in Figure B-3. If the SPA with HART was factory­configured by Rosemount Analytical for opera­tion with your OCX 4400, jumper and switch setting adjustments are not required. However, you may use the following procedure to verify that the jumper and switch settings are correct. Adjust or verify jumper and switch settings as follows:
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices B-1
Figure B-1. SPA With HART Alarm
Page 82
Instruction Manual
Appendix B Rev 2.0 January 2003
>
250
R
<
950
OCX 4400
HART MODEL 275 HANDHELD
COMMUNICATOR
4-20 mA
O SIGNAL LOOP
2
F1 F2 F3 F4
789
456
123
0
ANALOG OUTPUT
OCX 4400
CUSTOMERS
DCS, PLC,
or
PC with AMS
SOFTWARE
IN IN
4-20 mA
or
0-24 vdc
EVENT RECORDER,
AUDIBLE ALARM,
OR OTHER
ANALOG DEVICE
Figure B-2. OCX 4400 and SPA Interface Connections
NOTICE
REMOVE THIS PANEL
FOR SERVICE ACCESS.
STATIC SENSITIVE.
USE CAUTION WHEN
CHANGING JUMPERS
ACCESS COVER
37020019
SPA
BOTTOM
VIEW
SOURCE CURRENT
DIP SWITCHES
PASSWORD
FAILSAFE
DIP SWITCHES
JUMPER PINS
SWITCHES SHOWN IN
CORRECT POSITION
PASSWORD
ON
PASSWORD
OFF
SWITCHES SHOWN IN
CORRECT POSITION
37020022
Figure B-3. SPA Jumper and Dip Switch Settings
B-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Page 83
OCX 4400
B
Instruction Manual
Appendix B Rev 2.0
January 2003
2. Verify that the PASSWORD jumper is set to the OFF position. If the jumper is in the ON position, reposi­tion the jumper.
3. Check the position of the FAILSAFE dip switches. Position the dip switches as shown in Figure B-3.
4. Check the position of the SOURCE CURRENT dip switches. Position the dip switches as shown in Figure B-3.
5. Reinstall the SPA access cover.
b. Configuration/Calibration.
Prior to operation, the SPA operating pa­rameters must be configured via a menu­driven setup procedure. At the end of the con­figuration procedure, the SPA analog output signal is calibrated to insure valid communi­cations.
1. See Figure B-4. Connect a calibrated ammeter (Fluke Model 87 or
equivalent, accurate to ±0.025%) to the SPA analog output terminals. Observe polarity.
2. Connect a 90 to 260 VAC or 22 to 300 VDC power source to the SPA power terminals. When connecting an AC power source, use the AC and ACC (AC Common) terminals. For a DC source, use the AC and Ground terminals.
3. If desired, you can connect the 4 to 20 mA O
signal wires from the OCX
2
4400 analog output terminal block to the SPA INput terminals. (The OCX must be operational to transmit the O
signal. Observe polarity.)
2
NOTE
The O
signal connection is not re-
2
quired for SPA configuration or cali­bration. The OCX interface will allow you to observe the O
signal level
2
when the SPA configuration proce­dure is completed.
FLUKE MODEL 87
MULTIMETER
OCX 4400
OPTIONAL
>
250
950
<
IN IN
OR EQUIVALENT
AC
ACC
GND
AC or DC
POWER INPUT
mA
37020020
Figure B-4. SPA Setup for Calibration
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices B-3
Page 84
Instruction Manual
Appendix B Rev 2.0 January 2003
OCX 4400
4. Observe the front panel of the SPA, Figure B-5:
(a) A process value display in the
SPA display window indicates that the SPA is operational. Four pushbuttons are located below the display window.
(b) Pressing a left-hand pushbutton
scrolls up (Ï) or down (Ð) through the SPA command menu, a submenu, or parameter values list.
(c) Pressing the VIEW pushbutton
displays rail limits and alarm re-
lay configurations. There are five sequential displays in the VIEW mode. While in the VIEW mode, the up (Ï), down (Ð), and SE­LECT pushbuttons are disabled.
NOTE
In the VIEW mode, you can scroll through and display the output zero and full scale settings and the alarm relay trip points and configurations.
(d) Pressing the SELECT pushbut-
ton selects the displayed menu or submenu command or selects a displayed parameter variable.
SPA
DISPLAY
WINDOW
LEFT-HAND
PUSHBUTTONS
RIGHT-HAND
PUSHBUTTONS
37020017
Figure B-5. SPA Front Panel
B-4 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Page 85
OCX 4400
B
Instruction Manual
Appendix B Rev 2.0
January 2003
5. Figure B-6 shows the SPA menu, submenus, and parameter values that must be selected to configure the SPA for use with the OCX 4400.
6. Use the following instructions and selections shown in Figure B-6 to properly configure the SPA.
(a) Press the SELECT pushbutton.
Observe the display window on the SPA front panel. The display window should read SET HART.
(b) See the SET HART command in
Figure B-6. To the right of the command window is the SET HART submenu and related pa­rameter values that must be se­lected via the front panel pushbuttons on the SPA.
(c) In the submenu views shown in
Figure B-6, Ð means press the down pushbutton. Ï means press the up pushbutton.
(d) Â means press the SELECT
pushbutton one time. Ú means press the Ï or Ð pushbutton un­til the desired parameter value is shown in the SPA window.
(e) Proceed through the SPA menu,
selecting the menu commands and parameter values indicated in Figure B-6. After completing the sequence in the first column, go to the top of the second col­umn and continue.
(f) To exit the menu, repeatedly
press SELECT to display any main menu command. Then, press Ï or Ð until CONF EXIT is displayed. Select CONF EXIT.
(g) Detailed instructions concerning
the configuration menu and the submenu structure for each main command are provided in the SPA user’s manual.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices B-5
Page 86
Instruction Manual
Appendix B Rev 2.0 January 2003
OCX 4400
SET
HART
SGNL SRCE
FLT
SRCE
CONF
OPTS
SET MSTR Â Ú PRIM
  SET FUCT Â Ú NRML  NUM VARS Â Ú 02  NUM TRYS Â Ú 03  EXIT HART  SGNL SRCE
 AOUT SRCE Â Ú SV  AL2 SRCE Â Ú PV
AL3 SRCE Â Ú PV
Â
AL4 SRCE Â Ú PV
  EXIT SRCE  FLT SRCE
Ð CONF OPTS
 SET LINR Â Ú LINR OFF  PV SCLE Â Ú AUTO  DSPL EGU Â Ú CSTM Â Ú PPM
(select P, select P, select M)
 DSPL SRCE Â Ú SV  AL2 SEL  AL2 FLT Ð SET FAIL  FAIL LOW Ð EXIT OPTS  SET EGU
SCLE INPT
SCLE DSPL
SCLE
OUT
TRIM
OUT
CONF ALRM
 PV DSPL  SET ZERO Â Ú 0000 PCT  SET FULL Â Ú 25 PCT  SV DSPL  SET ZERO Â Ú 0000 PPM  SET FULL Â Ú 1000 PPM  EXIT DSPL  SCLE OUT
SET ZERO Â Ú 4.000 MA
Â
SET FULL Â Ú 20.00 MA
Â
EXIT Z/FS
  TRIM OUT
 TRIM ZERO Â Ú 4.000 MA
TRIM FULL Â Ú 20.00 MA
  EXIT TRIM  CONF ALRM
Refer to SPA manual for settings or
PASS WORD
Ð
SET
EGU
SCLE
INPT
SCLE DSPL
 PV EGU Â Ú PCT  SV EGU Â Ú CSTM Â Ú PPM
(select P, select P, select M)
EXIT EGU Â SCLE INPT
Â
 PV SCLE  PV ZERO Ú 0000 PCT  PV FULL Ú 25 PCT  SCLE SV  SET ZERO Ú 0000 PPM
 SET FULL Ú 1000 PPM  EXIT SCLE  SCLE DSPL
PASS
WORD
CONF
EXIT
Ð CONF EXIT
 SPA setup complete;
observe process value display.
Figure B-6. SPA Configuration Menu for OCX 4400 Communication
B-6 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Page 87
OCX 4400
C
Instruction Manual
Appendix C Rev 2.0
January 2003

APPENDIX C

BLOWBACK

C-1 FORWARD
This appendix describes the operation of the blowback option for the OCX 4400 Oxygen/ Combustibles Transmitter.
The blowback system blows instrument air back through the inlet filter and out the inlet tube of the transmitter. This removes built up dirt and particulate from the filer and inlet line. The blowback option is normally used in systems that have a dirty process stream.
C-2 INSTALLATION
Installing an OCX 4400 Oxygen/Combustibles Transmitter with the blowback option requires the addition of air operated blowback valve, regulator and gage, and check valve.
Figure C-1 shows the piping arrangement for the OCX 4400 Oxygen/Combustibles Transmit­ter with the blowback and autocalibration op­tions. Figure C-2 shows the piping arrangement for the OCX 4400 Oxygen/Combustibles Transmitter with the blowback option without autocalibration (without test gas solenoids).
1. Connect a clean, dry, instrument­quality supply of air (20.95% O 35 psig and 55 psig pressure regula­tors. The inlet to the 35 psig regulator accepts a 1/8” NPT fitting. The inlet to the 55 psig regulator accepts a 1/4” NPT fitting.
2. See the upper leg of the instrument air supply. Connect the output of the 35 psi regulator/filter to one port of the normally-closed air-operated solenoid valve, and to the inlet side of the dilu­tion air rotometer.
3. Connect the dilution air rotometer out­put to the DILUTION AIR inlet fitting on the sensor housing.
) to the
2
4. Install an instrument air line between the open port of the normally-open air­operated solenoid valve and the tee fit­ting on the sensor housing.
5. Connect the output of the 55 psi regu­lator/filter to one port of the normally­open air-operated solenoid valve, and to the instrument air inlet on the back of the electronics housing.
6. Install an air line between the open port of the normally-closed air-operated solenoid valve and the check valve in­let fitting on the sensor housing.
7. Install an air line between the instru­ment air outlet fitting on the electronics housing and the control air inlet fitting on the air-operated solenoid valve.
C-3 SETUP
There are three settings that need to be speci­fied to set up the blowback option. These are the blowback interval, duration, and purge time. These settings are available through HART from the Device Setup > Detailed Setup > Blowback menu. Using the LOI, these settings are acces­sible from the Setup > Blowback menu.
Interval – Length of time between blow-
back events. (60 minutes rec­ommended.)
Duration – Length of time blowback air is
activated. (5 seconds recom­mended.)
Purge Time – Length of time after blowback is
complete before oxygen/ combustibles readings are con­sidered valid. (Set as required by the application.)
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices C-1
Page 88
Instruction Manual
Appendix C Rev 2.0 January 2003
OCX 4400
CAL GAS
FLOW METER
5SCFH
20-30 PSIG
(
RECOMMENDED
CHECK
VALV E
ELECTRONICS
HOUSING
(
EDUCTOR
AIR IN
SENSOR
HOUSING
CAL GAS IN
REFERENCE
AIR IN
INSTRUMENT AIR
BLOWBACK
OPERATED
Pressure regulator with 1/8” inlet port is factory set for 35 psig. Regulator with 1/4” inlet port is factory set for 55 psig. If regulators are not installed in correct locations, the OCX 4400 will not work.
DILUTION AIR IN
VALVE,
AIR
NORMALLY
**
OPEN
SOLENOID
VALV E
NORMALLY
CLOSED
SOLENOID
VALV E
DILUTION AIR
FLOW METER
0.1 SCFH
AIR PRESSURE
REGULATOR/FILTER
35 PSIG
INSTRUMENT
55 PSIG
AIR SUPPLY
HI CO
2-STAGE
REGULATORS
2
HI O
2
LO O
NOTE: DURING OPERATION, STATE OF NORMALLY-OPEN AND NORMALLY-CLOSED SOLENOID VALVES CHANGE.
Figure C-1. Piping Arrangement, Blowback with Autocalibration
AIR PRESSURE
REGULATOR/FILTER
37020015
C-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Page 89
OCX 4400
C
CHECK
VALV E
ELECTRONICS
HOUSING
SENSOR
HOUSING
EDUCTOR
AIR IN
(
RECOMMENDED
CAL GAS IN
REFERENCE
AIR IN
INSTRUMENT AIR
BLOWBACK
OPERATED
CAL GAS
FLOW METER
5 SCFH
20-30 PSIG
DILUTION AIR IN
VALVE,
AIR
(
2-STAGE
REGULATORS
DILUTION AIR FLOW METER
AIR PRESSURE
REGULATOR/FILTER
Instruction Manual
Appendix C Rev 2.0
January 2003
2
HI CO
0.1 SCFH
HI O
2
LO O
Pressure regulator with 1/8inlet port is factory set for35 psig. Regulator with 1/4inlet port is factory set for 55 psig. If regulators are not installed in correct locations, the OCX 4400 will not work.
Figure C-2. Piping Arrangement, Blowback without Autocalibration
NORMALLY
**
OPEN
SOLENOID
VALV E
NOTE: DURING OPERATION, STATE OF NORMALLY-OPEN AND NORMALLY-CLOSED SOLENOID VALVES CHANGE.
NORMALLY
CLOSED
SOLENOID
VALV E
35 PSIG
55 PSIG
AIR PRESSURE
REGULATOR/FILTER
INSTRUMENT
AIR SUPPLY
37020026
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices C-3
Page 90
Instruction Manual
Appendix C Rev 2.0 January 2003
OCX 4400
C-4 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Page 91
OCX 4400
D
Instruction Manual
Appendix D Rev 2.0
January 2003

APPENDIX D

IS BARRIER

D-1 DESCRIPTION
A twisted pair of signal wires provide communi­cation from the OCX 4400 Transmitter to the customer’s analog devices (alarms, PC, DCS, etc.). In normal operation, the signal wires carry a low-voltage, 4 to 20 milliamp analog signal.
An OCX equipment fault, however, could gen­erate a higher energy signal at the transmitter output terminals. Where the OCX analog signal is transmitted through hazardous areas, failure to use IS barriers could allow higher energy sig­nals to ignite combustible gases or materials.
The OCX 4400 Transmitter is available with an IS barrier option. The IS barrier limits the energy (voltage and current) that can be produced at the analog output terminals of the transmitter. Whenever the OCX with IS barrier protection is selected, the customer’s devices must also be
IS barrier protected to avoid transmitting a feed­back energy surge through the OCX signal wires.
The recommended interface for connecting the OCX 4400 with IS barrier protection to the cus­tomer’s analog devices is shown in Figure D-1. The MTL Model 44041A Current Repeater with adapter and BPS 04 Backplane (intrinsic barrier kit) has been selected by Rosemount Analytical to protect against feedback from the customer’s equipment through the O
Notice that an SPA with HART alarm is included in the IS barrier interconnect diagram. The SPA unit is necessary to decode the 4 to 20 mA combustibles (COMB) concentration signal or line fault detection signals superimposed on the O
analog signal. Without the SPA, only the O
2
analog signal would be available for input to the customer’s analog devices.
signal loop.
2
2
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices D-1
Page 92
Instruction Manual
Appendix D Rev 2.0 January 2003
OCX 4400
OCX 4400
HART MODEL 275 HANDHELD
COMMUNICATOR
F1 F2 F3 F4
789
456
123
0
12
345 6
500
NON-HAZARDOUS
AREA
ISOLATING DRIVER
MTL 5046
PWR
HAZARDOUS
AREA
CUSTOMER’S
2
O SIGNAL LOOP 4 - 20 mA
EQUIPMENT
NON-HAZARDOUS
AREA
789101112
OCX 4400 4 - 20 mA
SIGNAL OUTPUT
TERMINAL BLOCK
SITE-PROGRAMMABLE HART
LOOP MONITOR AND ALARM
13
14
24 VDC
AOUT 2
AOUT 2
MOORE INDUSTRIES
IN IN
ELECTRONICS
HOUSING
2
O SIGNAL
AOUT 1
AOUT 1
ANALOG OUTPUT
ADAPTER
INTRINSIC SAFETY
BARRIER KIT
P/N 1A99698G01
12345 6
>
250
R
<
950
I
9 1011121314
78
CURRENT
REPEATER
MTL 4041A
I
MTL BACKPLANE
28 VDC PWR IN
O SIGNAL 4 - 20 mA
2
BPS 04
AC/DC
PWR IN
EVENT
RECORDER,
AUDIBLE ALARM
OR OTHER
ANALOG
DEVICE
DCS or
PC with AMS
SOFTWARE
37020016
Figure D-1. IS Barrier Interface Diagram
D-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Page 93
OCX 4400
12
This index is an alphabetized listing of parts, terms, and procedures having to do with the Haz­ardous Area Oxygen/Combustibles Transmitter. Every item listed in this index refers to a location in the manual by one or more page numbers.
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
SECTION 12

INDEX

A
Accessories, 1-11 Alarm Output, 2-6, 8-1 AMS, 1-1, 1-3, 1-4 Analog Outputs, 1-4
C
Calibration Methods, 4-1 COMB Sensor Replacement, 7-9 Contact/Thermocouple, 7-7
D
Dilution Air, 1-6
E
Eductor, 1-6 Eductor Air, 1-6 Electronics Housing, 1-3
F
Faults, 8-1 Fuses, 7-12
H
HART, 1-3, 1-4 HART Communicator, 1-1, 1-4, 5-1 HART Menu Tree, 5-3, 5-4 Heater Strut, 7-7
I
Installation
Electrical, 2-6 Mechanical, 2-1 Pneumatic, 2-9
Instrument Air, 1-6
L
Line Voltage, 1-4, 2-6 Local Operator Interface (LOI), 1-4
M
Maintenance, 7-1
N
Nernst Equation, 1-1
O
O2 Cell Replacement, 7-5 Operation, 6-1
P
Package Contents, 1-1 Pressure Sensor Replacement, 7-10 Probe Lengths, 1-3 Probe Materials, 1-3 Product Matrix, 1-1, 1-10, 1-11
R
Reference Air, 1-6, 2-11 Removal/Replacement, 7-1 Replacement Parts, 10-1
S
Sensor Housing, 1-3, 7-6 Sensors, 1-3, 1-6 Solenoid Replacement, 7-10 Solenoids, 1-6, 2-11 SPA with HART Alarm, 8-1 Specifications, 1-8 Switch Configuration, 3-1 System Features, 1-3
T
Test Gas Values, 3-2
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 12-1
Page 94
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
12-2 Index Rosemount Analytical Inc. A Division of Emerson Process Management
Page 95

ROSEMOUNT WARRANTY

Rosemount warrants that the equipment manufactured and sold by it will, upon shipment, be free of defects in workmanship or material. Should any failure to conform to this warranty become ap­parent during a period of one year after the date of shipment, Rosemount shall, upon prompt written notice from the purchaser, correct such nonconformity by repair or replacement, F.O.B. factory of the defective part or parts. Correction in the manner provided above shall constitute a fulfillment of all liabilities of Rosemount with respect to the quality of the equipment.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WAR­RANTIES OF QUALITY WHETHER WRITTEN, ORAL, OR IMPLIED (INCLUDING ANY WARRANTY OF MERCHANTABILITY OF FITNESS FOR PURPOSE).
The remedy(ies) provided above shall be purchaser's sole remedy(ies) for any failure of Rosemount to comply with the warranty provisions, whether claims by the purchaser are based in contract or in tort (including negligence).
Rosemount does not warrant equipment against normal deterioration due to environment. Factors such as corrosive gases and solid particulates can be detrimental and can create the need for re­pair or replacement as part of normal wear and tear during the warranty period.
Equipment supplied by Rosemount Analytical Inc. but not manufactured by it will be subject to the same warranty as is extended to Rosemount by the original manufacturer.
At the time of installation it is important that the required services are supplied to the system and that the electronic controller is set up at least to the point where it is controlling the sensor heater. This will ensure, that should there be a delay between installation and full commissioning that the sensor being supplied with ac power and reference air will not be subjected to component deterioration.
3702/1-03
Page 96
Instruction Manual
IB-106-440 Rev. 2.0 January 2003
OCX 4400
OCX 4400
Pa rt n o. _______________ Se ria l n o. _______________ Ord er no. _______________
Emerson Process Management
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.
1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
© Rosemount Analytical Inc. 2003
Fisher-Rosemount GmbH & Co.
Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd.
Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount
Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...