Rosemount Analytical designs, manufactures and tests its products to meet many national and
international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to
follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Rosemount Analytical repre-
sentative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate In-
struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program,
and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY.
Look-alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except
when maintenance is being performed by qualified persons, to prevent electrical shock
and personal injury.
The information contained in this document is subject to change without notice.
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
HIGHLIGHTS OF CHANGES
Effective October, 2000 Rev. 1.0.
PageSummary
ThroughoutAdded information on the Blowback option.
Page 1-7Corrected Figure 1-4.
Page 1-8Updated specifications.
Page 1-11Updated Accessories section of Product Matrix.
Page 2-7Updated Figure 2-5 to include ground stud.
Page 2-8Updated Figure 2-6 to include shield connections and additional wire
colors.
Page 2-11Corrected Figure 2-8 piping arrangement.
Page 3-1Added Initial Power Up instructions for voltage selection.
Page 5-4 through
5-6
Page 5-10Added Line Voltage Selection procedure.
Page 7-1Added ground stud to removal/replacement instructions.
Page 7-2Updated Figure 7-1 to include ground stud.
Page 7-4Updated Figure 7-3 to include shield connections and additional wire
Page 7-6Updated Figure 7-4 to include new heater configuration and blowback
Page 7-11Updated Figure 7-6 to include new instrument air solenoid and vent
Section 8Updated troubleshooting information and fault indications.
Section 9Updated replacement parts table and added new replacement parts.
11-3 through 11-5Updated Figure 11-3, menu trees for Local Operator Interface.
Appendix CAdded Appendix C.
Updated Figure 5-3, menu trees for HART communicators.
Table 10-1. Replacement Parts .............................................................................................. 10-1
Rosemount Analytical Inc. A Division of Emerson Process Managementiii
OCX 4400
The purpose of this manual is to provide a comprehensive understanding of the OCX 4400
Oxygen/Combustibles Transmitter components, functions, installation, and maintenance.
This manual is designed to provide information about the OCX 4400. We recommend that
you thoroughly familiarize yourself with the Introduction and Installation sections before
installing your transmitter.
The introduction presents the basic principles of the transmitter along with its performance
characteristics and components. The remaining sections contain detailed procedures and
information necessary to install and service the transmitter.
Before contacting Rosemount concerning any questions, first consult this manual. It describes most situations encountered in your equipment’s operation and details necessary
action.
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
PREFACE
DEFINITIONS
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this
publication.
Highlights an operation or maintenance
procedure, practice, condition, statement, etc. If not strictly observed, could
result in injury, death, or long-term
health hazards of personnel.
NOTE
Highlights an essential operating procedure,
condition, or statement.
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETIN
Highlights an operation or maintenance
procedure, practice, condition, statement, etc. If not strictly observed, could
result in damage to or destruction of
equipment, or loss of effectiveness.
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, and is not related to the illustration in any technical
manner.
Rosemount Analytical Inc. A Division of Emerson Process ManagementP-1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
FOR THE WIRING AND INSTALLATION
The following safety instructions apply specifically to all EU member states. They should
be strictly adhered to in order to assure compliance with the Low Voltage Directive. NonEU states should also comply with the following unless superseded by local or National
Standards.
1. Adequate earth connections should be made to all earthing points, internal and external,
where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced.
The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
OCX 4400
IMPORTANT
SAFETY INSTRUCTIONS
OF THIS APPARATUS
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable
anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be
made through a circuit breaker which will disconnect all circuits carrying conductors during a
fault situation. The circuit breaker may also include a mechanically operated isolating switch.
If not, then another means of disconnecting the equipment from the supply must be provided
and clearly marked as such. Circuit breakers or switches must comply with a recognized
standard such as IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazardous voltages are likely to be present beneath. These covers should only be
removed when power is removed from the equipment — and then only by
trained service personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a
danger from hot surfaces beneath. These covers should only be removed by
trained service personnel when power is removed from the equipment. Certain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to
the Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the following standards: EN61010-1, IEC417, and ISO3864.
P-2Rosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1-1 COMPONENT CHECKLIST OF TYPICAL
SYSTEM (PACKAGE CONTENTS)
A typical Rosemount OCX 4400 Oxygen/
Combustibles Transmitter package should contain the items shown in Figure 1-1. Record the
part number, serial number, and order number
for each component of your system in the table
located on the front page of this manual.
Use the product matrix in Table 1-2 at the end
of this section to verify your order number. The
first part of the matrix defines the model. The
last part defines the various options and features of the OCX 4400. Check the model number against the transmitter features and options,
making sure options specified by this number
are on or included with the unit. Use this complete model number for any correspondence
with Rosemount.
1-2 SYSTEM OVERVIEW
a. Scope
This Instruction Bulletin supplies details
needed to install, startup, operate, and
maintain the Rosemount OCX 4400 Oxygen/ Combustibles Transmitter. Signal conditioning electronics outputs separate 4-20
mA signals representing oxygen (O
combustibles (COMB) values. This information, plus additional details, can be accessed with the HART Model 275 handheld
communicator or Asset Management Solutions (AMS) software. An optional local operator interface (LOI) is available to provide
local communication with the electronics.
b. System Description
The Rosemount OCX 4400 Oxygen/
Combustibles Transmitter is designed to
measure oxygen and combustible concentrations in flue gas temperatures up to
) and
2
2600°F (1427°C). The transmitter is close
coupled to the process and requires minimal sample conditioning requirements.
The equipment measures oxygen percentage by reading the voltage developed
across a heated electrochemical cell, which
consists of a small yttria-stabilized, zirconia
disc. Both sides of the disc are coated with
porous metal electrodes. When operated at
the proper temperature, the millivolt output
of the cell is given by the following Nernst
equation:
EMF = KT log10 (P1/P2) + C
Where:
1. P2 is the partial pressure of the oxygen in
the measured gas on one side of the cell.
2. P1 is the partial pressure of the oxygen in
the reference air on the opposite side of
the cell.
3. T is the absolute temperature.
4. C is the cell constant.
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry instrument air (20.95% oxygen) as the
reference air.
When the cell is at operating temperature
and there are unequal oxygen concentrations across the cell, oxygen ions will travel
from the high oxygen partial pressure side
to the low oxygen partial pressure side of
the cell. The resulting logarithmic output
voltage is approximately 50 mV per decade.
The output is proportional to the inverse
logarithm of the oxygen concentration.
Therefore, the output signal increases as
the oxygen concentration of the sample gas
decreases. This characteristic enables the
OCX 4400 to provide exceptional sensitivity
at low oxygen concentrations.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
1
5
MAN4275A00
English
October1994
HART
Communicator
o
4
3
1. Instruction Bulletin
2. HART
Communicator Package (Optional)
3. Adapter Plate with Mounting Hardware and Gasket
4. OCX 4400 (Shown with Integral Electronics)
5. Reference Air and Calibration Set
6. Blowback Hardware (Optional)
FISHER-ROSEMOUNT
6
TM
2
34070023
Figure 1-1. Typical System Package
1-2 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
The OCX 4400 measures net oxygen concentration in the presence of all the products of combustion, including water vapor.
Therefore, it may be considered an analysis
on a “wet” basis. In comparison with older
methods, such as the portable apparatus,
which provides an analysis on a “dry” gas
basis, the “wet” analysis will, in general, indicate a lower percentage of oxygen. The
difference will be proportional to the water
content of the sampled gas stream.
The OCX 4400 combustibles sensor is a
catalytic sensor consisting of two Resistant
Temperature Detectors (RTD). One RTD is
the reference element covered with an inert
coating. The other RTD element is active,
coated with a catalyst. As the sample gases
flow by the sensor, the combustible gases
oxidize on the surface of the active element.
The oxidation that occurs produces heat
and a temperature rise in the active element. The temperature difference produces
a resistance relationship between the two
elements that is directly proportional to the
concentration of combustibles in the sample
gases.
The catalyst is specifically designed to detect carbon monoxide (CO), but the sensor
responds to other combustible gases. The
sensor is calibrated using carbon monoxide
(CO), thus the output should be expressed
in terms of CO. However, since the sensor
detects other combustible gases, the output
cannot just be labeled CO. The response of
the sensor to other combustible gases gives
an output that is equivalent to the sensor
detecting CO. The term COMB is used in
this manual to describe the sensor output.
This term indicates that the sensor is calibrated in terms of CO, and that the sensor
output is equivalent to CO but not specific to
CO.
c. System Configuration
Transmitters are available in four length options, giving the user the flexibility to use a
penetration appropriate to the size of the
stack or duct. The length options are 18 in.
(457 mm), 3 ft (0.91 m), 6 ft (1.83 m), or 9 ft
(2.7 m). Probes are available in three material options, 316L stainless steel, inconel
600, and ceramic to accommodate higher
temperatures.
The electronics are contained in a separate
housing from the sensors and may be integrally mounted next to the sensor housing
on the probe, or split mounted up to 150
feet away from the probe.
The electronics control both sensor temperatures and provide individual 4-20 mA
isolated outputs that are proportional to the
measured oxygen and combustibles concentrations. The power supply can accept
voltages of 100 to 240 VAC and 50 to 60
Hz. The electronics accepts millivolt signals
generated by the sensors and produces the
outputs to be used by remotely connected
devices. The outputs are isolated 4-20 mA
linearized currents. See 3-1b for specific instructions upon initial power up.
d. System Features
1. The O
tivity increase as the oxygen concentration decreases.
2. HART communication is standard. To
use the HART capability, you must
have either:
(a) A HART Model 275
cell output voltage and sensi-
2
Communicator.
Dilution air is provided to the COMB sensor
to ensure there is adequate oxygen to fully
oxidize any combustible gases regardless
of the concentration of oxygen in the
process.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
(b) Asset Management Solutions
(AMS) software for the PC.
3. Oxygen cell and heater/thermocouple
assembly are field replaceable.
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
4. Electronics are configurable for line
voltages from 100 to 240 VAC. See
Section 3-1b.
5. An operator can calibrate and diagnostically troubleshoot the OCX 4400 in
one of two ways:
(a) HART Interface. Each of the OCX
4400’s 4-20 mA output lines
transmit an analog signal proportional to oxygen or combustible
levels detected. The HART output
is superimposed on the oxygen 420 mA output line only. This information can be accessed through
the following:
1 Rosemount Model 275 Hand-
held Communicator - The
handheld communicator requires Device Description
(DD) software specific to the
OCX 4400. The DD software
will be supplied with many
Model 275 units, but can also
be programmed into existing
units at most FisherRosemount service offices.
Refer to Section 5,
HART/AMS, for additional
information.
2 Personal Computer (PC) -
The use of a personal computer requires AMS software
available from FisherRosemount.
3 Selected Distributed Control
Systems - The use of distributed control systems requires
input/output (I/O) hardware
and AMS software which
permit HART
communications.
(b) Optional Local Operator Interface
(LOI). The optional LOI is mounted
to the end of the electronics module and allows local communications with the electronics. Refer to
Appendix A, Local Operator Interface (LOI), for more information.
6. Optional Blowback System. The blowback system periodically blows instrument air back through the inlet filter
and out the inlet tube. This clears out
particulate and keeps the inlet filter
from clogging.
e. System Operation
Figure 1-2 shows the relationship between
the components of the OCX 4400. The sensors and the electronics are contained in
separate housings. The sensor housing and
probe mounts to a duct or process wall so
that the probe protrudes into the flue gas
stream. An air powered eductor continuously pulls samples of the process flue gas
through the probe to a chamber in front of
the sensor housing where the sample
passes the O
sensor and continues on to
2
the COMB sensor. Dilution air is provided to
the COMB sensor and reference air to the
O
sensor. After the gas sample flows past
2
the O
sensor and through the COMB sen-
2
sor, it is drawn through the eductor where it
mixes with the eductor air and exits through
exhaust back into the system. The electronics housing contains the CPU which
converts the sensor inputs into 4-20 mA
analog output signals. The CPU also initiates and performs calibrations. Three test
gasses and instrument air are controlled by
solenoids. Test gas flow to the sensors is
regulated by a rotometer between the electronics and sensor housings. Instrument air
is separated into eductor air, reference air,
and dilution air. The instrument air solenoid
does not allow air flow until the heaters are
up to temperature. This minimizes the
amount of sampled process flue gas being
pulled into cold sensors causing
condensation.
1-4 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SAMPLE
GAS
PROBE
EXHAUST
SENSOR
HOUSING
COMBUSTIBLES
SENSOR
O2
SENSOR
EDUCTOR
ROTOMETER
5 SCFH
ELECTRONICS
PRESSURE
SENSOR
TEST GAS
SOLENOIDS
INSTRUMENT AIR
SOLENOID
EDUCTOR AIR
HOUSING
CPU
POWER
SUPPLY
LOW O2
TEST GAS
HIGH O2
TEST GAS
HIGH CO
TEST GAS
INSTRUMENT
AIR
REFERENCE AIR
DILUTION AIR
ROTOMETER
50 cc/Min.
(0.106 SCFH)
Figure 1-2. System Operation Diagram
34070001
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
f. Handling the OCX 4400
It is important that printed circuit
boards and integrated circuits are
handled only when adequate antistatic
precautions have been taken to prevent possible equipment damage.
The OCX 4400 is designed for industrial application. Treat each component of the system with care to avoid
physical damage. The probe may contain components made from ceramics,
which are susceptible to shock when
mishandled.
g. System Considerations
Prior to installing your Rosemount OCX
4400 Oxygen/Combustibles Transmitter,
make sure you have all the components
necessary to make the system installation.
Ensure all the components are properly integrated to make the system functional.
After verifying that you have all the components, select mounting locations and determine how each component will be placed in
terms of available line voltage, ambient
temperatures, environmental considerations, convenience, and serviceability.
Figure 1-3 shows a typical system wiring. A
typical system installation is illustrated in
Figure 1-4.
A source of instrument air is required at the
OCX 4400 for reference gas, dilution air,
and eductor air. Since the OCX 4400 is
equipped with an in-place calibration feature, provision should be made for connecting test gas tanks to the OCX 4400
when it is to be calibrated.
NOTE
The electronics module is designed to
meet NEMA 4X (IP66) and the electronic components are rated to temperatures up to 185°F (85°C).
Retain packaging in which the unit arrived from the factory in case any
components are to be shipped to another site. This packaging has been
designed to protect the product.
INSTRUMENT
AIR
OCX 4400 O2/COMBUSTIBLES TRANSMITTER
WITH INTEGRAL ELECTRONICS
4-20 MA OUTPUT
(TWISTED PAIRS)
3 CALIBRATION GAS LINES
BY CUSTOMER
[300 FT (90M) MAX]
Figure 1-3. OCX 440 HART Connections and AMS Application
TERMINATION IN
CONTROL ROOM
HART MODEL 275
HANDHELD
INTERFACE
CUSTOMER’S LAPTOP
WITH ASSET MANAGEMENT
ASSET MANAGEMENT SOLUTIONS
SOLUTIONS
34070002
1-6 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
1
INTEGRAL ELECTRONICS
TEST
GAS
ROTOMETER
GASES
STACK
DILUTION AIR
ROTOMETER
ADAPTER
PLATE
DUCT
Instruction Manual
4-20 mA OUTPUTS (2)
LINE VOLTAGE
INSTRUMENT AIR
SUPPLY (REF. GAS)
PRESSURE
REGULATOR
SPAN COMBUSTIBLES TEST GAS
LOW O2 TEST GAS
HIGH O2 TEST GAS
IB-106-440 Rev. 1.0
October 2000
ELECTRONICS
CABLE
(UP TO 150FT)
SPLIT-MOUNTED ELECTRONICS
GASES
DUCT
STACK
TEST
GAS
ROTOMETER
DILUTION AIR
ROTOMETER
4-20 mA OUTPUTS (2)
LINE VOLTAGE
INSTRUMENT AIR
SUPPLY (REF. GAS)
PRESSURE
REGULATOR
SPAN COMBUSTIBLES TEST GAS
LOW O2 TEST GAS
HIGH O2 TEST GAS
34070003
Figure 1-4. Typical System Installation
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
1-3 OCX 4400 SPECIFICATIONS
transmitter before choosing the mounting loca-
tion. Table 1-2 contains the product matrix. Use
Table 1-1 contains information about the OCX
4400 Oxygen/Combustibles Transmitter operating characteristics. Use the table to make sure
that available conditions are suitable for the
the product matrix for ordering information and
specific model number information. Table 1-3
contains information on accessories available
with the OCX 4400.
Table 1-1. OCX 4400 Specifications
Net O
Range ..................................................0-1% to 0-40% O
2
, Fully field selectable
2
Combustibles .................................................0-1000 ppm to 0-5%, Fully field selectable
Accuracy:
Oxygen.....................................................+0.75% of reading or 0.05% O
Calibration ......................................................Semi-automatic or automatic
Calibration Gas Mixtures Recommended....0.4% O
8% O
, Balance N2
2
, Balance N2
2
1000 ppm CO, Balance N2
(Ref. test gas kit #1A99119G04)
Calibration Gas Flow .....................................5 scfh (2.5 l/m)
Reference Air..................................................2 scfh (1 l/m), clean, dry instrument-quality air (20.95% O
regulated to 35 psi (238 kPa)
Eductor Air......................................................5 scfh (2.5 l/m), clean, dry, instrument-quality air
(20.95% O
), regulated to 35 psi (238 kPa)
2
Dilution Air......................................................0.1 scfh (0.5 l/m), clean, dry, instrument-quality air
(20.95% O
Blowback Air (Optional) ................................Clean, dry, instrument-quality air (20.95% O
), regulated to 35 psi (238 kPa)
2
), regulated to
2
55 psi (374 kPa)
1
Operating temperature of electronics inside of electronics housing, as measured by a HART Communicator, or Fisher-
Rosemount Asset Management Solutions Software, or local operator interface.
),
2
1-8 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
1
IB-106-440 Rev. 1.0
OCX 4400
Table 1-1. OCX 4400 Specifications (Continued)
Sensors Housing ...........................................NEMA 4X, IP66 with fitting and pipe on reference exhaust
port to clean, dry atmosphere
Two 3/4-14 NPT conduit ports
Electronics Housing ......................................NEMA 4X, IP66 with fitting and pipe on reference exhaust
port to clean, dry atmosphere
Two 3/4-14 NPT conduit ports
Electrical Noise ..............................................Meets EN 50082-2 Electromagnetic Compatibility Generic
Immunity Std., Part II
Includes ENG 1000 4-R for Electrostatic Discharge 4 Kv
contact, 8 Kv in air
Optionally ENG 1000 4-R “Namur-Increased” 8 Kv contact,
16 Kv in air
Includes IEC 801-4 fast transients-2
Kv on power supply and control lines
Line Voltage....................................................Universal 100 to 240 VAC, 50 to 60 Hz, Software selectable,
3/4-14 NPT conduit port
Isolated Output:
Oxygen.....................................................4-20 mAdc, 950 ohm max, with HART capability for O
COMB
Combustibles ..........................................4-20 mAdc, 950 ohm max
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-9
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
Table 1-2. Product Matrix
OCX44AO2 / Combustibles Transmitter
O2 / Combustibles Transmitter IB-106-440
CodeProbe Length and Material
00No Probe or Exhaust Tube
1118 in. (457 mm) 316 SST up to 1300°F (704°C)
2118 in. (457 mm) Inconel 600 up to 1832°F (1000°C)
3118 in. (457 mm) Ceramic up to 2600°F (1426°C)
123 ft (0.91 m) 316 SST up to 1300°F (704°C)
223 ft (0.91 m) Inconel 600) up to 1832°F (1000°C)
323 ft (0.91 m) Ceramic up to 2600°F (1426°C)
136 ft (1.83 m) 316 SST up to 1300°F (704°C)
236 ft (1.83 m) Inconel 600 up to 1832°F (1000°C)
149 ft (2.7 m) 316 SST up to 1300°F (704°C)
Code Probe Mounting Assembly
10(ANSI 2 in. 150 lb) 6" dia. flange, 4.75" BC with 4 x 0.75" dia. holes
20(DIN) 185mm dia. flange, 145mm BC with 4 x 18mm dia. holes
30(JI S ) 155mm dia. flange, 130mm BC with 4 x 15mm dia. holes
OCX 4400
CodeMounting Hardware - Stack Side
0No Adapter Plate ("0" must also be chosen under "Mounting Adapter - Probe Side" below)
1New Installation - Square weld plate with studs
2Model 218/240 Mounting Plate (with Model 218/240 Shield Removed)
3Existing Model 218/240 Support Shield
4Mounting to Other Mounting
5Model 132 Adapter Plate
6Mounting to Ametek WDG-IVC
7Mounting to Servomex Xendos 2700
8Mounting to Elsag Bailey SMA-90
CodeMounting Hardware - Probe Side
0No Adapter Plate
1Probe Only (ANSI)
4Probe Only (DIN)
7Probe Only (JIS)
OCX44A111011(Cont’d)Example
(1)
1-10 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
1
Table 1-2. Product Matrix (Continued)
Cont'dCodeElectronic Housing - NEMA 4X, IP66 HART Communications
01Basic Unit
02Local Operator Interface
03Calibration Solenoids
04IS Barrier
05Local Operator Interface and Calibration Solenoids
06Local Operator Interface and IS Barrier
07Calibration Solenoids and IS Barrier
08Local Operator Interface and Calibration Solenoids and IS Barrier
CodeElectronic Mounting
01Integral to Sensor Housing Electronics
02Split Electronics and no cable
03Split Electronics and 20 ft (6m) cable
04Split Electronics and 40 ft (12m) cable
05Split Electronics and 60 ft (18m) cable
06Split Electronics and 80 ft (24m) cable
07Split Electronics and 100 ft (30m) cable
08Split Electronics and 150 ft (45m) cable
2
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
CodeAccessories
00None
01Cal. Gas Rotometers, Flow Rotometers, and Ref. Gas Set
11Blowback Option and Cal. Gas Rotometers, Flow Rotometers, and Ref. Gas Set
Cont’d080111Example
NOTES:
(1)
Where possible, specify SPS number; otherwise, provide details of the existing mounting plate as follows:
Plate with studsBolt circle diameter, number, and arrangement of studs, stud thread, stud height above mounting plate.
Plate without studs Bolt circle diameter, number, and arrangement of holes, thread, depth of stud mounting plate with accessories.
(2)
If the Local Operator Interface (LOI) is not implemented, remote access and functionality avai l abl e via HART Communicati ons (M o del 275
Handheld Communicator) with Oxygen/Combustibles Device Desc ri ption (DD) requi red.
Table 1-3. Accessories
PART
NUMBERDESCRIPTION
1A99119H01Oxygen bottles 0.4% balance N2
1A99119H02Oxygen Bottles 8.0% balance N2
1A99119H06CO Bottle 1000 PPM and air balance N2
1A99120H02Regulator for Oxygen (may need 2)
1A99120H03Regulator for COMB
1A99119G063 wide wall mount bracket for gas bottles
1A99119G05Calibration gas regulators kit
1A99119G04Calibration gas bottles kit
1A99292H01Moore Industries SPA for Low O2 Alarm, High
COMB Alarm, Calibration Status, and Unit Fail
275D9EI5D0000275 HART Communicator with 12 Meg buffer
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-11
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
1-12 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
2
Before installing this equipment, read
the “Safety instructions for the wiring
and installation of this apparatus” at
the front of this Instruction Bulletin.
Failure to follow the safety instructions could result in serious injury or
death.
2-1 MECHANICAL INSTALLATION
a. Selecting Location.
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 2
INSTALLATION
3. Ensure the area is clear of internal and
external obstructions that will interfere
with installation and maintenance access to the unit. Allow adequate clearance for the removal of the OCX 4400.
Do not allow the temperature of the
electronics housing to exceed 185°F
(85°C) or damage to the electronics
may result.
1. The location of the OCX 4400 in the
stack or flue is most important for
maximum accuracy in the oxygen
analyzing process. The probe must be
positioned so the gas it measures is
representative of the process. Best results are normally obtained if the
transmitter is positioned near the center of the duct (40-60% insertion).
Longer ducts may require several
transmitters since the oxygen and
combustibles can vary due to stratification. A point too near the wall of the
duct, or the inside radius of a bend,
may not provide a representative sample because of the very low flow conditions. The sensing point should be
selected so the process gas temperature falls within the range of probe
material used. Figure 2-1 through
Figure 2-3 provide mechanical installation references. The ambient temperature inside the electronics housing
must not exceed 185°F (85°C).
2. Check the flue or stack for holes and
air leakage. The presence of this condition will substantially affect the accuracy of the oxygen and combustibles
readings. Therefore, either make the
necessary repairs or install the transmitter upstream of any leakage.
b. Installation
1. Ensure all components are available to
install the OCX 4400.
2. The OCX 4400 may be installed intact
as it is received.
3. Weld or bolt adapter plate (Figure 2-3)
onto the duct.
4. In vertical installations for systems with
integral electronics, ensure the system
cable drops vertically from the OCX
4400 and the conduit is routed below
the level of the electronics housing.
This drip loop minimizes the possibility
that moisture will damage the electronics (Figure 2-4).
NOTE
If process temperatures will exceed
392°F (200°C), use anti-seize compound on stud threads to ease future
removal of the OCX 4400.
5. Insert probe through the opening in the
mounting flange and bolt the unit to the
flange.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
NOTE:
ALL DIMENS IONS ARE IN
INCHES WITH MILLIMETERS
IN PARENTHESES.
INSULATE IF EXPOSED TO
AMBIENT WEATHER CONDITIONS
0.062 THK GASKET
3535B18H02
ANSI
3535B46H01
JIS
3535B45H01
DIN
DIM “A”DIM “B”
OCX 4400
TABLE 2. INSTALLATION/REMOVAL
PROBE
18 IN.
3 FT
6 FT
9 FT
DIM "A"
16
(406)
34
(864)
70
(1778)
106
(2692)
DIM "B"
32.38
(822)
50.38
(1280)
86.38
(2184)
122.38
(3108)
Figure 2-1. OCX 4400 Installation (with Integral Electronics)
BOTT OM VIEW
FLANGE
DIA
HOLE
DIA
(4) HOLES
EQ SP
ON BC
TABLE 1. MOUNTING FLANGE
ANSI
4512C17H01
6.00
(153)
0.75
(20)
4.75
(121)
DIN
4512C19H01
7.28
(185)
0.71
(18)
5.71
(145)
JIS
4512C18H01
6.10
(155)
0.59
(15)
5.12
(130)
29790008
2-2 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
2
DIMENSIONS
IN.
(mm)
ADAPTER PLATE* DIMENSIONS FOR OCX 4400
ANSI
(P/N 4512C34G01)
DIN
(P/N 4512C36G01)
(P/N 4512C35G01)
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
JIS
"A"
"B"
THREAD
"C"
DIA
*PART NUMBERS FOR ADAPTER PLATES INCLUDE
ATTACHING HARDWARE.
NOTE:DIMENSIONS ARE
IN INCHES WITH
MILLIMETERS IN
PARENTHESES.
2.500 DIA
(63.5)
ADAPTER PLATE
FOR OCX 4400
INSTALLATION. SEE
FIGURE 2-1.
6.00
(153)
0.625-11
4.75
(121)
B
7.5
(191)
(M-16 x 2)
5.708
(145)
A
o
45
C
6.50
(165)
(M-12 x 1.75)
5.118
(130)
A
4 STUDS,
LOCKWASHERS AND
NUTS EQUALLY
SPACED ON
C DIA B.C.
29790013
Figure 2-2. OCX 4400 Adapter Plate Dimensions
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
INSTALLATION FOR METAL
WALL STACK OR DUCT
CONSTRUCTION
OCX 4400
INSTALLATION FOR MASONRY
WALL STACK CONSTRUCTION
MTG HOLES
SHOWN ROTATED
o
45 OUT OF
TRUE POSITION
WELD OR BOLT ADAPTER
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIRTIGHT.
0.50 [13]
3.75 [95]
MIN DIA HOLE
IN WALL
STACK OR DUCT
METAL WALL
0.50 [13]
BOLT ADAPTER
PLATE TO OUTSIDE
WALL SURFACE
FIELD WELD
PIPE TO
ADAPTER PLATE
MTG HOLES
SHOWN ROTATED
o
45 OUT OF
TRUE POSITION
JOINT MUST
BE AIRTIGHT
OUTSIDE WALL
SURFACE
NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT
ADAPTER PLATE NOT FURNISHED BY ROSEMOUNT.
BOLT ADAPTER
PLATE TO OUTSIDE
WALL SURFACE
4.50 [114]
O.D. REF
PIPE 4.00 SCHED 40
PIPE SLEEVE (NOT
BY ROSEMOUNT)
LENGTH BY CUSTOMER
MASONRY
STACK WALL
FIELD WELD PIPE
TO ADAPTER PLATE
3.50 [89]
O.D. REF
PIPE 3.00 SCHED 40
PIPE SLEEVE (NOT BY
ROSEMOUNT) LENGTH
BY CUSTOMER
MASONRY
STACK WALL
29790014
WELD OR BOLT ADAPTER
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIRTIGHT.
2.50 [63.5]
MIN DIA HOLE
IN WALL
STACK OR DUCT
METAL WALL
NOTE: DIMENSIONS IN INCHES WITH
MILLIMETERS IN PARENTHESES.
JOINT MUST
BE AIRTIGHT
OUTSIDE WALL
SURFACE
Figure 2-3. Adapter Plate Installation
2-4 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
2
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
4-20 mA
SIGNALS
DRIP
LOOP
INSULATION
OCX 4400
ELECTRONICS
REPLACE
INSULATION
AFTER
INSTALLING
OCX 4400
ADAPTER
PLATE
STACK OR DUCT
METAL WALL
29790015
Figure 2-4. Installation with Drip Loop and Insulation Removal
LINE
VOLT AGE
DRIP
LOOP
Uninsulated stacks or ducts may
cause ambient temperatures around
the electronics to exceed 185°F (85°C),
which may cause overheating damage
to the electronics.
6. If insulation is being removed to access
the duct work for OCX 4400 mounting,
make sure the insulation is replaced
afterward (Figure 2-4).
c. Split-Electronics Installation
If OCX 4400 is configured for split electronics mounting, choose a location not to exceed the length of the electronics cable
ordered. Split-mounted electronics modules
may be pole or wall mounted.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
2-2 ELECTRICAL INSTALLATION
All wiring must conform to local and national
codes. For reference, factory wired solenoid
power connections are shown in Figure 2-5.
Disconnect and lock out power before
connecting the unit to the power
supply.
Install all protective equipment covers
and safety ground leads after installation. Failure to install covers and
ground leads could result in serious
injury or death.
To meet the Safety Requirements of
IEC 1010 (EC requirement), and ensure
safe operation of this equipment, connection to the main electrical power
supply must be made through a circuit
breaker (min 10 A) which will disconnect all current-carrying conductors
during a fault situation. This circuit
breaker should also include a mechanically operated isolating switch. If
not, then another external means of
disconnecting the supply from the
equipment should be located close by.
Circuit breakers or switches must
comply with a recognized standard
such as IEC 947.
NOTE
To maintain CE compliance, ensure a
good connection exists between the
mounting flange bolts and earth.
a. Connect Line Voltage
The OCX 4400 is software configured for
100 to 240 VAC line voltage at 50 to 60 Hz.
The power supply itself requires no setup.
Connect the line, or L wire, to the L terminal,
and the neutral, or N wire, to the N terminal
on the AC power input terminal block in the
electronics housing. Connect the ground, or
G wire, to the ground stud in the electronics
housing. Secure the connection with two
nuts. Attach a separate line from the ground
stud to the G terminal on the power input
terminal block.
NOTE
Make sure line voltage is correctly set.
See Section 3-1b for further
instructions.
b. Connect 4-20 mA Signals
Connect the 4-20 mA current loop to the 420 mA signal output terminals in the electronics housing.
1. Oxygen (O
) 4-20 mA Signal. One 4-
2
20 mA signal represents the O
Superimposed on the O
4-20 mA sig-
2
nal is the HART information that is accessible through a Model 275
Handheld Communicator or AMS software. The O
4-20 mA signal is at the
2
AOUT 1+ and AOUT 1- terminals.
2. Combustibles (COMB) 4-20 mA Signal.
The other 4-20 mA signal represents
the COMB value. HART information is
not available on the COMB signal. The
COMB 4-20 mA signal is at the AOUT
2+ and AOUT 2- terminals.
c. Alarm Output Relay
Connect any customer supplied relay input
to the alarm output relay terminal. The
alarm output relay terminal is a set of dry
contacts with 30 mA and 30 VDC capacity.
d. Split-Electronics Connections
to Sensor Housing
If the OCX 4400 is configured with splitelectronics, make the following connections
between the electronics and sensor housings with the electronics cable ordered with
the package (Figure 2-6). Steel braided cable is available in lengths up to 150 ft.
(45 m).
1. Thermocouple Connections. Connect
the sensor/thermocouple terminal block
in the electronics housing to the corresponding sensor/thermocouple terminal block in the sensor housing. The
twisted wire pairs are numbered on the
inner plastic wrapper. Keep the twisted
pairs together and match the numbers
and wire colors shown in Figure 2-6.
Connect the twisted pair wire shields to
the shield terminals.
value.
2
2-6 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
2
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SOLENOID POWER
TERMINAL B L OCK
(FA CTORY WIRED)
GND
GND
LO O2
HI COMB
HI O2
ALARM OUTPUT
RELA Y TERMINAL
GND
GND
INST AIR
NC
NO
COMMON
SIGNAL OUTPUT
TERMINAL B L OCK
4-20 mA
ELECTRONICS HOUSING
(COVER REMO VED)
SENSOR INPUT
HEATER THERMOCOUPLE,
AND HEATER POWER
INPUT PORT
(FROM SENSOR HOUSING)
PRESSURE
SENSOR
HI COMB
SOLENOID
4-20 mA
SIGNAL
OUTPUT
PORT
HI O2
SOLENOID
LO O2
SOLENOID
INST AIR
SOLENOID
AOUT 2-
AOUT 2+
AOUT 1-
AOUT 1+
COMB
4-20 mA
SIGNAL
O2
4-20 mA
SIGNAL
AC POWER IN P U T
TERMINAL B L OCK
F1 POWER
SUPPLY FUSE
Figure 2-5. Line Voltage and 4-20 mA Connections
2. Heater Power Connections. Use the
purple, orange, black, gray, brown, and
red stranded wires in the electronics
cable to connect power to the three
heaters in the sensor housing. Match
L
F2 HEATER
N
G
AC POWER
INPUT PORT
FUSE
GROUND STUD
TWO NUTS
34070021
the wire colors to the corresponding
heater power terminal blocks in the
sensor and electronics housings as
shown in Figure 2-6.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
SPLIT-ELECTRONICS HOUSING
HEATER POWER
TERMINAL B L OCK
ORANGE
GRA Y
BLACK
PURPLE
SL HEATER
L2
ELECTRONICS CABLE
(UP TO 150 FT)
NOTE: KEEP TWISTED
PAIRS TOGETHER
O2 CELL HEATER
L2
BROWN
COMB CELL HEATER
(COVER REMO VED)
RED
L2
SENSOR/THERMOCOUPLE
TERMINAL B L OCK
(1)
YELLOW
O2 T/C +
RED
O2 T/C -
SHIELD
GND
(2)
YELLOW
SL T/C +
RED
SL T/C -
(3)
BLUE
O2 MV +
SHIELD
WHITE
GND
O2 MV -
PRESSURE
SENSOR
ELECTRONICS
CABLE PORT
GROUND STUD
(5)
(4)
YELLOW
WHITE
GREEN
COMB ACTIVE +
WHITE
COMB REF -
COMB REF +
COMB ACTIVE -
CABLE
(6)
SHIELD
GND
WHITE
RED
RTD DRIVE
RTD RETURN
PRESSURE
SENSOR
FACTORY
WIRED
WHITE
RED
GREEN
+ 10 PR SRC
PR SENSOR -
PR SENSOR +
BLACK
PR SRC RTN
(5)
(6)
WHITE
WHITE
RED
YELLOW
WHITE
RTD DRIVE
COMB REF -
COMB REF +
SENSOR/THERMOCOUPLE
TERMINAL B L OCK
RTD RETURN
ELECTRONICS
CABLE PORT
PROBE
O2 CELL HEATER
SL HEATER
L2
PURPLE
ORANGE
HEATER POWER
TERMINAL B L OCK
COMB CELL HEATER
L2
GRA Y
BLACK
BROWN
L2
RED
(1)
YELLOW
O2 T/C +
RED
O2 T/C -
(2)
YELLOW
SL T/C +
RED
SL T/C -
(3)
BLUE
O2 MV +
WHITE
O2 MV -
(4)
WHITE
GREEN
BLUE
COMB ACTIVE -
COMB ACTIVE +
RED
SENSOR HOUSING
(COVER REMO VED)
34070010
Figure 2-6. Electrical Connections Between Split-Electronics and Sensor Housing
2-8 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
2
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
2-3 PNEUMATIC INSTALLATION
a. Connections to Electronics Housing
1. Reference Air Set. After the OCX 4400
is mechanically installed, connect the
reference air set (regulator and pressure gage) to the instrument air inlet on
the electronics housing. The reference
air set should be installed in accordance with Figure 2-7.
Do not use 100% nitrogen as an O2 low
gas. It is suggested that O2 low gas be
between 0.4% and 2.0% O2. Do not use
gases with hydrocarbon concentrations of more than 40 parts per million.
Failure to use proper gases will result
in erroneous readings.
2. Calibration Gas. Two O2 calibration
gas concentrations and one COMB
calibration gas are used with the OCX
4400: O
8%, and COMB high gas - 1000 ppm
or 4%. Connect O
GAS LO O
housing. Connect O
CAL GAS HI O
ics housing. Connect COMB high gas
to the CAL GAS HI COMB input on the
low gas - 0.4%, O2 high gas -
2
low gas to the CAL
2
input on the electronics
2
2
input on the electron-
2
high gas to the
electronics housing. Refer to Figure
2-8.
3. Instrument Air Regulator. Connect a
regulator to the instrument air inlet on
the electronics housing as shown in
Figure 2-8. Set the regulator to 35 psig.
Instrument air is separated into three
air lines after exiting the instrument air
regulator. Connect a rotometer to the
dilution air line as shown in Figure 2-8.
Set the dilution air rotometer to 0.1
scfh.
4. Calibration Gas Rotometer. Connect a
rotometer to the calibration gas outlet
on the electronics housing as shown in
Figure 2-8. Set the rotometer to 5 scfh.
b. Connections to Sensor Housing
1. Calibration Gas. Connect the outlet of
the calibration gas rotometer to the
CAL GAS input on the sensor housing
(Figure 2-8).
2. Instrument Air. Connect the three instrument air lines to their proper connections on the sensor housing (Figure
2-8). Connect dilution air rotometer
output to DILUTION AIR input on the
sensor housing. Connect the instrument air regulator output to REF AIR
input and EDUCTOR AIR input on the
sensor housing.
NOTE
!
Upon completing installation, make sure that the OCX 4400 is turned on and operating prior
to firing up the combustion process. Damage can result from having a cold OCX 4400 exposed to the process gases.
During outages, and if possible, leave all OCX 4400 units running to prevent condensation
and premature aging from thermal cycling.
Upon completing installation, make sure that the OCX 4400 is turned on and operating prior
to firing up the combustion process. Damage can result from having a cold OCX 4400 exposed to the process gases.
2-4 STARTUP
Refer to Section 3, STARTUP AND OPERATION, for startup information.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
1
DRAIN VALVE
ELECTRONICS
HOUSING
OUTLET
0.250 OR 6mm O.D. TUBING
(SUPPLIED BY CUSTOMER)
2
INSTRUMENT
AIR INLET
3.12 (79.25) MAX
2.250 (57.15)
0.25-18 NPT FEMALE
INLET CONNECTION
2.0
(50.80)
1.50
(38.10)
12" PRESSURE GAGE0-60 PSIG275431-003
2COMBINATION FILTER-REG.0-60 PSIG
NOTE: DIMENSIONS ARE IN INCHES WITH
2 MOUNTING HOLES
3.19 (81.03) LG
THROUGH BODY FOR
0.312 (7.92) DIA BOLTS
INSTRUMENT AIR SUPPLY
10-225 PSIG MAX PRESSURE
SET TO 35 PSIG
MILLIMETERS IN PARENTHESES.
8.50
(215.90)
MAX
SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OCX 4400 ELECTRONICS HOUSING.
34070004
Figure 2-7. Air Set, Plant Air Connection
2-10 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
2
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
ELECTRONICS
HOUSING
CAL GAS
HI COMB
INPUT
CAL GAS
ROTOMETER
5 SCFH
SENSOR
HOUSING
CAL GAS
HI O2
INPUT
CALIBRA TION
GAS OUTLET
INSTR UMENT
AIR IN
INSTRUMENT
AIR OUT
CAL GAS
LO O2
INPUT
CAL GAS
INPUT
SET REGULATOR TO 35 PSIG
0 - 60 PSIG
DILUTION AIR
ROTOMETER
0.1 SCFH
EDUCTOR
AIR INPUT
REF AIR
INPUT
DILUTION AIR
INPUT
34070006
Figure 2-8. OCX 4400 Gas Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
ON
SW3
112
C
O
M
O
B
2
21.1 mA
3.5 mA
2
3
3
4
SECTION 3
DEFAULT POSITION -
3,4 – NOT USED
4
COMB 21.1 mA/3.5 mA:
INTERNAL:
COMB 4-20 mA IS
INTERNALLY
POWERED
(DEFA U LT)
21.1 mA
O2 21.1 mA/3.5 mA: O2 4-20 mA SIGNAL
RAIL LIMITS:
HIGH - 21.1 mA
LOW - 3.5 mA
COMB 4-20 mA SIGNAL
RAIL LIMITS:
HIGH - 21.1 mA
LOW - 3.5 mA
SW2
EXTERNAL:
COMB 4-20 mA
REQUIRES
AN EXTERNAL
POWER SUPPLY
FUSE F2
INTERNAL:
O2 4-20 mA IS
INTERNALLY
POWERED
(DEFA U LT)
SW1
Figure 3-1. OCX 4400 Defaults
EXTERNAL:
O2 4-20 mA
REQUIRES
AN EXTERNAL
POWER SUPPLY
34070007
3-0 Startup and OperationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
3
STARTUP AND OPERATION
Install all protective equipment covers
and safety ground leads before
equipment startup. Failure to install
covers and ground leads could result
in serious injury or death.
3-1 GENERAL
a. Verify Installation
Ensure the OCX 4400 is installed correctly.
Verify mechanical installation and all electrical and pneumatic connections (Section 2,
INSTALLATION).
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 3
4. Apply power to the OCX 4400 Oxygen/Combustibles Transmitter.
5. If using the HART handheld communicator, connect the communicator to the
transmitter.
6. If using the HART communicator, select: Device Setup > Detailed Setup >
Line Voltage.
If using the LOI, select: Setup > Parameters > Line Voltage.
7. Set the line voltage to the appropriate
value: 100, 110, or 220.
Setting the power input voltage to the
wrong value can damage the heaters
in the sensor housing or prevent the
transmitter from operating properly.
Always ensure that the power input
voltage is set properly after installing
the transmitter.
b. Initial Power Up
The OCX 4400 Oxygen/Combustibles
Transmitter has a 12 minute delay before
the heaters begin operation. Upon initial
power up of the OCX 4400, and any time
supply voltage to the unit changes, the
power input voltage must be verified in the
software to ensure proper operation of the
heaters. The input voltage is set as follows:
1. Remove the housing cover from the
electronics housing.
2. Remove fuse F2 from the electronics
box. See Figure 3-1 for the location of
fuse F2.
3. Replace the electronics housing cover.
8. If using the HART communicator, press
F4 for Enter and then F2 for send.
If using the LOI, press the right arrow
to enter the value.
9. Remove power from the OCX 4400
electronics.
10. Remove the electronics housing cover
and install fuse F2.
11. Install the electronics housing cover
and restore power to the OCX 4400
electronics.
c. Verify OCX 4400 Configuration
There are three switches on the microprocessor board which are user configurable for
the OCX 4400 (Figure 3-1). SW1 determines if the O
or externally powered. SW2 determines if
the COMB 4-20 mA signal is internally or
externally powered. SW3 sets the rail limits
for the O
switches are accessible through holes in the
electronics box.
4-20 mA signal is internally
2
and COMB 4-20 mA signals. All
2
Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
Remove power from the OCX 4400 before changing defaults. If defaults are
changed under power, damage to the
electronics may occur.
1. SW1. The two settings are internally or
externally powering the O
signal. The factory setting is for the O
4-20 mA signal to be internally powered.
2. SW2. The two settings are internally or
externally powering the COMB 4-20
mA signal. The factory setting is for the
COMB 4-20 mA signal to be internally
powered.
3. SW3. The factory sets this switch as
follows:
(a) Position 1 determines the O
4-20 mA signal rail limit. The settings are high, 21.1 mA, or low, 3.5
mA. The factory setting is at high,
21.1 mA.
4-20 mA
2
2
OCX 4400
Section 5, HART/AMS or Appendix A, Local
Operator Interface (LOI) for additional
information.
1. HART/AMS. If using HART/AMS, from
the DETAILED SETUP menu, select
O
CALIBRATION.
2
(a) From O
2
High TG. Enter the percentage of
O
(b) From O
Low TG. Enter the percentage of
O
(c) From the DETAILED SETUP
menu, select COMB
CALIBRATION.
(d) From COMB CALIBRATION, se-
lect High TG. Enter the percentage
of CO used for high COMB test
gas.
2. LOI. If using the LOI, from the SETUP
menu select CALIBRATION.
CALIBRATION, select
2
used for the high O2 test gas.
2
CALIBRATION, select
2
used for the low O2 test gas.
2
(b) Position 2 determines the COMB
4-20 mA signal rail limit. The settings are high, 21.1 mA, or low,
3.5 mA. The factory setting is at
high, 21.1 mA.
(c) Position 3 is not used.
(d) Position 4 is not used.
d. Set Test Gas Values
Using HART/AMS or optional LOI, set the
test gas values used for calibration. Refer to
(a) From CALIBRATION, select O
2
High Gas. Enter the percentage of
O
used for the high O2 test gas.
2
(b) Select the down arrow and the
next selection will be O
Enter the percentage of O
for the low O
test gas.
2
Low Gas.
2
used
2
(c) Select the down arrow again and
the next selection will be Comb
High Gas. Enter the percentage of
O
used for the combustibles span
2
gas.
3-2 Startup and OperationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
4
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 4
CALIBRATION
4-1 GENERAL
During a calibration, two calibration gases with
known O
gas with a known COMB concentration are applied to the OCX 4400. Slope and constant values are calculated to determine if the OCX 4400
is correctly measuring net concentrations of O
and combustibles in the industrial process.
Before calibrating the OCX 4400, verify that the
calibration gas parameters are correct by setting
the test gas values used when calibrating the
unit (Section 3, STARTUP AND OPERATION).
4-2 CALIBRATION METHODS
There are three types of calibration methods
available to the OCX 4400; automatic, operatorinitiated automatic, and manual. Calibration
commands and menus can be accessed by
HART/AMS or by the optional LOI.
a. Automatic Calibration
concentrations and one calibration
2
If the OCX 4400 is equipped with calibration
solenoids, the unit can be programmed to
automatically calibrate without any operator
action. Refer to Section 5, HART/AMS for
information on automatic calibration setup
via HART/AMS. Refer to Appendix A for information on automatic calibration setup via
LOI.
HART/AMS menu tree. Refer to Figure
5-3 for the complete HART/AMS menu
tree.
2. Select DIAG/SERVICE from DEVICE
SETUP menu.
2
(a) Select CALIBRATE from the
DIAG/SERVICE menu.
(b) Select PERFORM CAL from CALI-
BRATE menu.
(c) Select CAL from PERFORM CAL
menu.
(d) Select the type of calibration de-
sired: O
method, or O
method from CAL menu.
3. LOI. To initiate a calibration using the
LOI, perform the following steps on the
LOI menu tree. Refer to Figure 11-3 for
complete LOI menu tree.
(a) From the CALIBRATE SENSORS
menu, select PERFORM
CALIBRATION.
(b) At the prompt, select the right ar-
row to initiate automatic calibration.
Cal method, COMB Cal
2
and COMB Cal
2
b. Operator Initiated-Automatic Calibration
An operator can initiate an automatic calibration at any time provided that the unit is
equipped with calibration solenoids.
1. HART/AMS. To initiate an automatic
calibration by using HART/AMS, perform the following steps on the
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration 4-1
c. Manual Calibration
If a unit is not equipped with calibration solenoids, a calibration must be performed by
an operator following prompts from the unit.
Refer to Section 5, HART/AMS for information on manual calibration with HART/AMS.
Refer to APPENDIX A for information on
manual calibration with LOI.
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
4-2 CalibrationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 5
HART/AMS
5-1 OVERVIEW
The HART Communicator is a handheld communications interface device. It provides a
common communications link to all
microprocessor-based instruments that are
HART compatible. The handheld communicator
contains an 8 x 21 character liquid crystal display (LCD) and 25 keys. A pocket-sized manual,
included with the HART Communicator, details
the specific functions of all the keys.
To interface with the OCX 4400, the HART
Communicator requires a termination point
along the O
4-20 mA current loop and a mini-
2
mum load resistance of 250 ohms between the
communicator and the power supply. The HART
Communicator accomplishes its task using a
frequency shift keying (FSK) technique. With the
use of FSK, high-frequency digital communication signals are superimposed on the 4-20 mA
oxygen output signal. The communicator does
not disturb the 4-20 mA signal since no net energy is added to the loop. HART information is
not available on the COMB 4-20 mA output
signal.
4-20 mA TERMINAL BLOCK
IN ELECTRONICS HOUSING
The HART Communicator may be interfaced
with a personal computer (PC), providing special software has been installed. To connect the
HART Communicator to a PC, an interface
adapter is required. Refer to the proper HART
Communicator documentation in regard to the
PC interface option.
5-2 HART COMMUNICATOR SIGNAL LINE
CONNECTIONS
The HART Communicator can connect to the
OCX 4400 oxygen analog output signal line at
any wiring termination in the O
4-20 mA current
2
loop. There are two methods of connecting the
HART Communicator to the signal line. For applications in which the signal line has a load resistance of 250 ohms or more, refer to method
1. For applications in which the signal line load
resistance is less than 250 ohms, refer to
method 2.
a. Method 1, For Load Resistance > 250
Ohms. Refer to Figure 5-1 and the following steps to connect the HART Communicator to a signal line 250 ohms or more of
load resistance.
≥
RL 250Ω
A OUT 1 A OUT 1 +
LOOP CONNECTORS
SERIALPORT & BATTERY
CHARGERMUST
NOTBE USED IN
HAZARDOUSAREAS
LOOPCONNECTORS
HART COMMUNICATOR
REAR PANEL
O2 4-20 mA SIGNAL LINE
USEINTERFACE
002750013 ONLY
SERIAL PORT
HART
COMMUNICA TOR
ANALOG OUTPUT DEVICE
LEAD SET
29790018
Figure 5-1. Signal Line Connections, > 250 Ohms Lead Resistance
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
Explosions can result in death or serious injury. Do not make connections
to the HART Communicator’s serial
port, 4-20 mA signal lines, or NiCad
recharger jack in an explosive
atmosphere.
Using the supplied lead set, connect the
HART Communicator in parallel to the OCX
4400. Use any wiring termination points in
the oxygen analog output 4-20 mA signal
line.
b. Method 2, For Load Resistance
< 250 Ohms
Refer to Figure 5-2 and the following steps
to connect the HART Communicator to a
signal line with less than 250 ohms load
resistance.
Explosions can result in death or serious injury. Do not make connections
to the HART Communicator’s serial
port, 4-20 mA signal lines, or NiCad
recharger jack in an explosive
atmosphere.
1. At a convenient point, break the oxygen analog output 4-20 mA signal line
and install the optional 250 ohm load
resistor.
2. Plug the load resistor into the loop
connectors (located on the rear panel
of the HART Communicator).
5-3 HART COMMUNICATOR PC
CONNECTIONS
There is an option to interface the HART Communicator with a personal computer. Load the
designated AMS software into the PC. Then link
the HART Communicator to the PC using the
interface PC adapter that connects to the serial
port (on the communicator rear panel). Refer to
the proper HART Communicator documentation
in regard to the PC interface option.
IN ELECTRONICS HOUSING
A OUT 1 A OUT 1 +
NOTE:
THE SIGNAL LOOP MUST BE BROKEN
TO INSERT TH E O PTIONAL 250 OH M
4-20 mA TERMINAL BLOCK
LOOP CONNECTORS
SERIALPORT & BATTERY
CHARGERMUST
NOTBE USED IN
HAZARDOUSAREAS
HART COMMUNICATOR
REAR PANEL
LOOPCONNECTORS
O2 4-20 mA SIGNAL LINE
USEINTERFACE
002750013 ONLY
SERIAL PORT
LOAD RESISTOR.
RL < 250Ω
ANALOG OUTPUT DEVICE
HART
COMMUNICA TOR
Figure 5-2. Signal Line Connections, < 250 Ohms Lead Resistance
250 OHM
LOAD
RESISTOR
(NOTE 1 )
29790017
5-2 HART/AMSRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
5-4 OFF-LINE AND ON-LINE OPERATIONS
The HART Communicator can be operated both
off-line and on-line.
a. Off-line operations are those in which the
communicator is not connected to the OCX
4400. Off-line operations can include interfacing the HART Communicator with a PC
(refer to applicable HART documentation
regarding HART/PC applications.
b. In the on-line mode, the communicator is
connected to the oxygen 4-20 mA analog
output signal line. The communicator is
connected in parallel to the OCX 4400 or in
parallel to the 250 ohm load resistor.
c. The opening menu displayed on the LCD is
different for on-line and off-line operations.
When powering up a disconnected (off-line)
communicator, the LCD will display the
Main Menu. When powering up a connected
(on-line) communicator, the LCD will display
the On-line Menu. Refer to the HART
Communicator manual for detailed menu
information.
5-5 MENU TREE FOR HART COMMUNICATOR/
OCX 4400 APPLICATIONS
This section consists of a menu tree for the
HART Communicator. This menu is specific to
OCX 4400 applications.
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-3
O2 mV
O2 temp mV
Elec temp V
COMB Act mV
COMB Ref mV
Pressure V
SL temp mV
Status Group 1
Status Group 2
Status Group 3
Status Group 4
OCX 4400
O2 Thermocouple Error
O2 Cell Temperature Low
O2 Cell Temperature High
O2 Cell Temperature Very High
O2 Heater Open
O2 Cell Open
O2 Cell Bad
O2 Slope Error
O2 Constant Error
O2 Calibration Failed
Landusky Error
High Electronics Temperature
EEPROM Corrupt
Sample Line Thermocouple Error
Sample Line Temperature Low
Sample Line Temperature High
Sample Line Temperature Very High
Sample Pressure Low Absolute
Combustibles Cell Error
Combustibles Cell Chamber Temperature Low
Combustibles Cell Chamber Temperature High
Combustibles Cell Chamber Temperature
Very High
Combustibles Slope Error
CO Constant Error
CO Cal Failed
O2 AO Saturated
COMB AO Saturated
O2 AO Fixed
COMB AO Fixed
Cal methods
O2 Optrak TG?
COMB Optrak TG?
CalState
CalState
TimeRemain
Present O2
Present COMB
O2 Cal slope
O2 Cal Const
O2 Cell Imp
COMB Slope
COMB Const
Reset O2 constants
Reset COMB constants
(CONTINUED ON
SHEET 2)
LOOP TEST
CALIBRATE
D/A TRIM
AO Saturated
AO Fixed
Max Electronics Temp
O2 Loop test method
COMB Loop test
method
PERFORM
CAL
CAL
STATUS
LAST CAL
CONSTANTS
RESET CAL
CONSTANTS
O2 D/A trim method
COMB D/A trim method
Figure 5-3. Menu Tree for HART/AMS on the OCX 4400 (Sheet 1 of 3)
O2 Cal method
COMB Cal method
O2 and COMB Cal
method
34070014
5-4 HART/AMSRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
(CONTINUED FROM
SHEET 1)
BASIC SETUP
DETAILED
SETUP
Ta g
SELECT
RANGE
DEVICE
INFORMATION
S/W VERSION
INFO
SENSORS
O2 URV
COMB URV
Dev id
Descriptor
Message
Date
Final asmbly num
O2 Snsr s/n
COMB Snsr s/n
Figure 5-3. Menu Tree for HART/AMS on the OCX 4400 (Sheet 2 of 3)
3.5 mA
21.1 mA
4-20
20-4
3.5 mA
21.1 mA
4-20
20-4
34070015
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
(CONTINUED FROM
SHEET 2)
DETAILED
SETUP
(CONTINUED FROM
SHEET 2)
O2 SETUP
O2
CALIBRATION
COMB SETUP
COMB
CALIBRATION
ALARM
RELAY
BLOWBACK
SLOPE
O2 CONST
O2 T90
High TG
Low TG
OPtrak TG?
TG Time
Purge Time
Sol. Present
Cal Intrvl h
Next Cal Time h
COMB SLOPE
COMB CONST
COMB T90
High TG
OPtrak TG?
TG Time
Purge Time
Sol. Present
Cal Intrvl h
Next Cal Time h
EVENT 1
EVENT 2
EVENT 3
Alarm State
Enabled
Interval
Duration
Purge Time
State
Do Blowback
OP Locks
OP Tracks
OP Locks
OP Tracks
None
In Cal
O2 Temp
O2 Heater
O2 Cell Bad
Cal Failed
Hi Elect Temp
Unit Fail
SL Temp
SL Pressure
Comb Temp
All
REVIEW
DEVICE
INFORMATION
CAL
INFO
DEVICE
CONFIG
OUTPUT
CONFIG
Manufacturer
Model
Dev id
Ta g
Descriptor
Message
Date
Final asmbly num
O2 Snsr s/n
COMB Snsr s/n
Fld dev rev
Hardware rev
Software rev
Universal rev
O2 Cal Info
COMB Cal Info
O2 Device Config
COMB Device Config
O2 URV
COMB URV
Poll addr
Num req preams
OPtrak TG?
TG Time
Purge Time
Low TG
High TG
Slope
O2 const
O2 cell Imp
OPtrak TG?
TG Time
Purge Time
High TG
COMB slope
COMB const
SLOPE
O2 CONST
O2 T90
COMB SLOPE
COMB CONST
COMB T90
34070016
Figure 5-3. Menu Tree for HART/AMS on the OCX 4400 (Sheet 3 of 3)
5-6 HART/AMSRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
5-6 HART COMMUNICATOR MANUAL O2 CAL
METHOD
To perform a manual O
calibration using the
2
HART Communicator, use the following procedure. If necessary, use the menu tree in Figure
5-3 (sheet 1 of 3) for reference.
NOTE
To select a menu item, either use the
up and down arrow keys to scroll to
the menu item and press the right arrow key or use the number keypad to
select the menu item number.
To return to a preceding menu, press
the left arrow key.
a. From the PERFORM CAL screen, select
menu item 1, CAL, then menu item 1, O
2
Cal method to access the O2 calibration
method.
Failure to remove the OCX 4400 from
automatic control loops prior to performing this procedure may result in a
dangerous operating condition.
Or, access the CALIBRATE screen and select menu item 2, CAL STATUS, to view
menu item 1, CALSTATE; menu item 2,
TIMEREMAIN; and menu item 3, PRESENT
O
, as the calibration status updates.
2
f. When CALSTATE displays APPLY HIGH O
GAS, return to the O2 CAL procedure. When
the “Loop should be removed from automatic control” warning appears, press OK.
g. At the APPLY HIGH O
GAS status prompt,
2
select menu item 2, NEXT CAL STEP.
When the status displays HIGH O
GAS
2
FLOW, select menu item 4, EXIT, to leave
the O
CAL procedure.
2
h. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
calibration status as it updates. Or, access
the CALIBRATE screen and select menu
item 2, CAL STATUS, to view menu item 1,
CALSTATE; menu item 2, TIMEREMAIN;
and menu item 3, PRESENT O
, as the
2
calibration status updates.
i. When CALSTATE displays STOP GAS, re-
turn to the O
CAL procedure.
2
2
b. In the first O
CAL method screen, a “Loop
2
should be removed from automatic control”
warning appears. Remove the OCX 4400
from any automatic control loops to avoid a
potentially dangerous operating condition
and press OK.
c. The next several screens indicate the cali-
bration status. At each of the following
status prompts, select menu item 2, NEXT
CAL STEP:
COMPLETE
APPLY LOW O
LOW O
GAS FLOW
2
GAS
2
d. At this point, select menu item 4, EXIT, to
leave the O
CAL procedure.
2
e. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
calibration status as it updates.
j. When the “Loop should be returned to
automatic control” message appears, return
the OCX 4400 to the automatic control
loops previously removed and press OK.
k. At the STOP GAS status prompt, select
menu item 2, NEXT CAL STEP. When the
status displays PURGING, select menu item
4, EXIT, to leave the O
CAL procedure.
2
l. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
calibration status as it updates. Or, access
the CALIBRATE screen and select menu
item 2, CAL STATUS, to view menu item 1,
CALSTATE; menu item 2, TIMEREMAIN;
and menu item 3, PRESENT O
, as the
2
calibration status updates.
m. When CALSTATE displays COMPLETE,
the calibration is finished.
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
5-7 HART COMMUNICATOR MANUAL COMB
CAL METHOD
To perform a manual combustibles calibration
using the HART Communicator, use the following procedure. If necessary, use the menu tree
in Figure 5-3 (sheet 1 of 3) for reference.
NOTE
To select a menu item, either use the
up and down arrow keys to scroll to
the menu item and press the right arrow key or use the number keypad to
select the menu item number.
To return to a preceding menu, press
the left arrow key.
a. From the PERFORM CAL screen, select
menu item 1, CAL, then select menu item 2,
COMB Cal method to access the combustibles calibration method.
Failure to remove the OCX 4400 from
automatic control loops prior to performing this procedure may result in a
dangerous operating condition.
calibration status as it updates. Or, access
the CALIBRATE screen and select menu
item 2, CAL STATUS, to view menu item 1,
CALSTATE; menu item 2, TIMEREMAIN;
and menu item 4, PRESENT COMB, as the
calibration status updates.
f. When CALSTATE displays APPLY COM-
BUSTIBLES GAS, return to the O
procedure. When the “Loop should be removed from automatic control” warning appears, press OK.
g. At the APPLY COMBUSTIBLES GAS status
prompt, select menu item 2, NEXT CAL
STEP. When the status displays COMBUSTIBLES GAS FLOW, select menu item
4, EXIT, to leave the COMB CAL procedure.
h. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
calibration status as it updates. Or, access
the CALIBRATE screen and select menu
item 2, CAL STATUS, to view menu item 1,
CALSTATE; menu item 2, TIMEREMAIN;
and menu item 4, PRESENT COMB, as the
calibration status updates.
i. When CALSTATE displays STOP GAS, re-
turn to the COMB CAL procedure.
CAL
2
b. In the first COMB CAL method screen, a
“Loop should be removed from automatic
control” warning appears. Remove the OCX
4400 from any automatic control loops to
avoid a potentially dangerous operating
condition and press OK.
c. The next several screens indicate the cali-
bration status. At each of the following
status prompts, select menu item 2, NEXT
CAL STEP:
COMPLETE
APPLY HIGH O
HIGH O2 GAS FLOW
d. At this point, select menu item 4, EXIT, to
leave the COMB CAL procedure.
e. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
5-8 HART/AMSRosemount Analytical Inc. A Division of Emerson Process Management
GAS
2
j. When the “Loop should be returned to
automatic control” message appears, return
the OCX 4400 to the automatic control
loops previously removed and press OK.
k. At the STOP GAS status prompt, select
menu item 2, NEXT CAL STEP. When the
status displays PURGING, select menu item
4, EXIT, to leave the COMB CAL procedure.
l. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
calibration status as it updates. Or, access
the CALIBRATE screen and select menu
item 2, CAL STATUS, to view menu item 1,
CALSTATE; menu item 2, TIMEREMAIN;
and menu item 3, PRESENT COMB, as the
calibration status updates.
m. When CALSTATE displays COMPLETE,
the calibration is finished.
OCX 4400
5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
5-8 HART COMMUNICATOR MANUAL O2 AND
CO CAL METHOD
To perform a manual O
and COMB calibration
2
using the HART Communicator, use the following procedure. If necessary, use the menu tree
in Figure 5-3 (sheet 1 of 3) for reference.
NOTE
To select a menu item, either use the
up and down arrow keys to scroll to
the menu item and press the right arrow key or use the number keypad to
select the menu item number.
To return to a preceding menu, press
the left arrow key.
a. From the PERFORM CAL screen, select
menu item 1, CAL, then menu item 3, O
and COMB Cal method to access the O
2
2
and COMB calibration method.
Failure to remove the OCX 4400 from
automatic control loops prior to performing this procedure may result in a
dangerous operating condition.
CALSTATE; menu item 2, TIMEREMAIN;
and menu item 3, PRESENT O
, as the
2
calibration status updates.
f. When CALSTATE displays APPLY HIGH O
GAS, return to the O2 and COMB CAL procedure. When the “Loop should be removed
from automatic control” warning appears,
press OK.
g. At the APPLY HIGH O
GAS status prompt,
2
select menu item 2, NEXT CAL STEP.
When the status displays HIGH O
GAS
2
FLOW, select menu item 4, EXIT, to leave
the O
and COMB CAL procedure.
2
h. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
calibration status as it updates. Or, access
the CALIBRATE screen and select menu
item 2, CAL STATUS, to view menu item 1,
CALSTATE; menu item 2, TIMEREMAIN;
and menu item 3, PRESENT O
, as the
2
calibration status updates.
i. When CALSTATE displays APPLY COM-
BUSTIBLES GAS, return to the O
and
2
COMB CAL procedure. When the “Loop
should be removed from automatic control”
warning appears, press OK.
2
b. In the first O
CAL method screen, a “Loop
2
should be removed from automatic control”
warning appears. Remove the OCX 4400
from any automatic control loops to avoid a
potentially dangerous operating condition
and press OK.
c. The next several screens indicate the cali-
bration status. At each of the following
status prompts, select menu item 2, NEXT
CAL STEP:
COMPLETE
APPLY LOW O
LOW O
GAS FLOW
2
GAS
2
d. At this point, select menu item 4, EXIT, to
leave the O
CAL procedure.
2
e. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
calibration status as it updates. Or, access
the CALIBRATE screen and select menu
item 2, CAL STATUS, to view menu item 1,
j. At the APPLY COMBUSTIBLES GAS status
prompt, select menu item 2, NEXT CAL
STEP. When the status displays COMBUSTIBLES GAS FLOW, select menu item
4, EXIT, to leave the O
and COMB CAL
2
procedure.
k. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
calibration status as it updates. Or, access
the CALIBRATE screen and select menu
item 2, CAL STATUS, to view menu item 1,
CALSTATE; menu item 2, TIMEREMAIN;
and menu item 4, PRESENT COMB, as the
calibration status updates.
l. When CALSTATE displays STOP GAS, re-
turn to the O
and COMB CAL procedure.
2
m. When the “Loop should be returned to
automatic control” message appears, return
the OCX 4400 to the automatic control
loops previously removed and press OK.
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-9
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
n. At the STOP GAS status prompt, select
menu item 2, NEXT CAL STEP. When the
status displays PURGING, select menu item
4, EXIT, to leave the O
procedure.
o. From the PERFORM CAL screen, view
menu item 4, CALSTATE, to monitor the
calibration status as it updates. Or, access
the CALIBRATE screen and select menu
item 2, CAL STATUS, to view menu item 1,
CALSTATE; menu item 2, TIMEREMAIN;
menu item 3, PRESENT O
4, PRESENT COMB, as the calibration
status updates.
p. When CALSTATE displays COMPLETE,
the calibration is finished.
5-9 AUTOMATIC CALIBRATION SETUP
VIA HART
NOTE
Automatic calibration is only available
on units equipped with calibration
solenoids.
Use the following procedure to specify a time
interval (in hours) at which the OCX 4400 will
automatically calibrate the OCX 4400.
and COMB CAL
2
, and menu item
2
OCX 4400
When setting automatic calibration
times, the CAL INTRVL and the NEXT
CAL TIME should be set so that O
and combustibles are NOT calibrated
simultaneously.
NOTE
To select a menu item, either use the
up and down arrow keys to scroll to
the menu item and press the right arrow key or use the number keypad to
select the menu item number.
To return to a preceding menu, press
the left arrow key.
a. From the DEVICE SETUP screen, select
DETAILED SETUP.
b. From the DETAILED SETUP screen, select
O
CALIBRATION. From the O2 CALIBRA-
2
TION screen, select menu item 7, CAL
INTRVL.
c. At the prompt, input a time interval (in
hours) at which an automatic O
will occur and press ENTER.
d. From the DETAILED SETUP screen, select
COMB CALIBRATION.
2
calibration
2
To disable the automatic calibration feature for
oxygen or combustibles, enter a 0 for the CAL
INTRVL for both O
and COMB.
2
e. From the COMB CALIBRATION screen,
select menu item 7, CAL INTRVL.
f. At the prompt, input a time interval (in
hours) at which an automatic COMB cali-
If desired, the NEXT CAL TIME can be changed
bration will occur and press ENTER.
to synchronize a calibration at a specific day or
time.
5-10 LINE VOLTAGE SELECTION USING HART
COMMUNICATOR
Use the procedure in Section 3-1b to select or
change the line voltage configuration using the
HART Communicator.
5-10 HART/AMSRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
6
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 6
OPERATION
6-1 OVERVIEW
Once the OCX 4400 is installed, set up, and
calibrated properly, the unit is ready for
operation.
6-2 OPERATION
Before the OCX 4400 can begin normal operation, all heaters must be up to temperature be-
fore the unit begins to take measurements of
the sample gas. During normal operation, the
unit will perform continuous measurements of
the sample gas. The unit is performing properly
if there are no alarms present. If the unit is
equipped with an LOI, the normal operating display will show a value for Oxygen and Combustible concentrations measured. If an LOI is not
used, values for Oxygen and Combustibles are
available through HART.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 6-1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
6-2 OperationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 7
MAINTENANCE AND SERVICE
7-1 OVERVIEW
This section contains the procedures to maintain and service the OCX 4400.
Install all protective equipment covers
and safety ground leads after equipment repair or service. Failure to install covers and ground leads could
result in serious injury or death.
It is recommended that the OCX 4400
be removed from the stack for all
service activities. The unit should be
allowed to cool and be taken to a clean
work area. Failure to comply may
cause severe burns.
Disconnect and lock out power before
working on any electrical components.
There may be voltage up to 264 VAC.
(e) Disconnect and remove the two
power leads from the AC power input terminal block and the ground
lead from the ground stud.
(f) Disconnect and remove the O
COMB 4-20 mA signal leads from
the 4-20 mA signal output terminal
block.
(g) If used, disconnect and remove the
external relay leads from the alarm
output relay terminal.
(h) Remove insulation to access the
mounting bolts. Unbolt the OCX
4400 from the stack and take it to a
clean work area.
(i) Allow the unit to cool to a comfort-
able working temperature.
2. Replace.
(a) Bolt the OCX 4400 to the stack
and install insulation. Remove
cover on electronics housing.
2
and
7-2 OCX 4400 REMOVAL/REPLACEMENT
a. OCX 4400 with Integrally Mounted
Electronics
1. Remove.
(a) Turn off power to the system.
(b) Shut off the calibration gases at
the cylinders and the instrument
air.
(c) Disconnect the calibration gas and
instrument air lines from the electronics housing.
(d) Remove the cover from the elec-
tronics housing to expose the
electronics housing terminal
blocks, Figure 7-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-1
(b) If used, connect external relay
leads to the alarm output relay
terminal, Figure 7-1.
(c) Connect the O
mA signal leads to the 4-20 mA
signal output terminal block.
(d) Connect the line, or L wire, to the L
terminal, and the neutral, or N wire,
to the N terminal on the AC power
input terminal block in the electronics housing. Connect the
ground, or G wire, to the ground
stud in the electronics housing.
Secure the connection with two
nuts. Attach a separate line from
the ground stud to the G terminal
on the power input terminal block.
and COMB 4-20
2
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
(e) Replace cover on electronics
housing.
(f) Connect the calibration gasses and
instrument air lines to the electronics housing.
4-20 mA
SIGNAL OUTPUT
TERMINAL B L OCK
AC POWER INPUT
TERMINAL B L OCK
(g) Turn on the calibration gasses at
the cylinders and turn on instrument air.
(h) Restore power to the system.
ELECTRONICS HOUSING
(COVER REMOVED)
ALARM OUTPUT
RELA Y TERMINAL
AOUT 2+
COMB
4-20 mA
SIGNAL
AOUT 2-
AOUT 1-
AOUT 1+
O2
4-20 mA
SIGNAL
G
NL
Figure 7-1. Electronics Housing Terminal Blocks
NO
NC
COMMON
29790029
7-2 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
7
O2 CELL HEATER
SL HEATER
L2
SENSOR HOUSING
(COVER REMO VED)
EMI
BOOT
HEATER POWER
TERMINAL B L OCK
COMB CELL HEATER
L2
L2
ELECTRONICS CABLE
(TO ELECTRONICS HOUSING)
SENSOR/THERMOCOUPLE
TERMINAL B L OCK
(1)
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
(5)
(2)
YELLOW
YELLOW
RED
RED
(3)
BLUE
WHITE
(4)
WHITE
GREEN
(6)
WHITE
WHITE
RED
YELLOW
RED
GRA Y
BLACK
PURPLE
ORANGE
BROWN
Figure 7-2. Sensor Housing Terminal Blocks
b. OCX 4400 with Split Mounted
Electronics
1. Remove Sensor Housing.
(a) Turn power off to the system.
(b) Shut off the calibration gases at the
cylinders and the instrument air.
(c) Disconnect calibration gas, refer-
ence air, eductor air, and dilution
air lines from the sensor housing.
(d) Remove the cover from the sensor
housing to expose the sensor
housing terminal blocks, Figure
7-2.
(e) Disconnect the electronics cable
from sensor/thermocouple terminal
block and the heater power termi-
nal block.
(f) Remove insulation to access the
mounting bolts. Unbolt the sensor
ELECTRONICS
CABLE PORT
O2 T/C +
PROBE
O2 T/C -
SL T/C +
SL T/C -
BLUE
RED
O2 MV -
O2 MV +
COMB ACTIVE -
COMB ACTIVE +
WHITE
RTD DRIVE
COMB REF -
COMB REF +
34070011
housing from the stack and take it
to a clean work area.
(g) Allow the sensor housing to cool
to a comfortable working
temperature.
2. Replace Sensor Housing
(a) Bolt the sensor housing to the
stack and install insulation.
(b) With the cover removed, connect
the electronics cable to sensor/
thermocouple terminal block and
the heater power terminal block,
Figure 7-2. Install sensor housing
cover.
(c) Connect the calibration gas, refer-
ence air, eductor air, and dilution
air lines to the sensor housing.
(d) Turn on the calibration gases at the
cylinders and instrument air.
(e) Restore power to the system.
RTD RETURN
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-3
put terminal block. Remove the
ground lead from the ground stud.
(f) Disconnect and remove the O
COMB 4-20 mA signal leads from
and
2
the 4-20 mA signal output terminal
(c) Disconnect the three calibration
block.
gas input lines and instrument air
line from the split-electronics
housing. Disconnect the calibration
gas outlet and instrument air outlet
(g) If used, disconnect and remove the
external relay leads from the alarm
output relay terminal.
lines from the split-electronics
housing.
(d) Remove the cover from the elec-
tronics housing to expose the
electronics housing terminal
(h) Disconnect the electronics cable
from sensor/thermocouple terminal
block and the heater power termi-
nal block, Figure 7-3.
blocks, Figure 7-1.
(i) Remove the split-electronics
(e) Disconnect and remove the two
power leads from the AC power in-
7-4 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
housing from mounting and move
to a suitable work area.
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
4. Replace Split-Electronics Housing.
(a) Mount split-electronics housing on
wall or pole within distance of
electronics cable used. Remove
cover.
(b) Connect electronics cable to the
sensor/thermocouple and heater
power terminal blocks, Figure 7-3.
(c) Connect the line, or L wire, to the L
terminal, and the neutral, or N wire,
to the N terminal on the AC power
input terminal block in the electronics housing. Connect the
ground, or G wire, to the ground
stud in the electronics housing.
Secure the connection with two
nuts. Attach a separate line from
the ground stud to the G terminal
on the power input terminal block.
(d) Connect the O
and COMB 4-20
2
mA signal leads to the 4-20 mA
signal output terminal block.
(e) If used, connect external relay
leads to the alarm output relay
terminal.
formance have been considered. If cell replacement is needed, order the O
cell
2
replacement kit (Table 10-1).
The O
cell replacement kit contains an O
2
cell and flange assembly, gaskets, socket
head cap screws, and anti-seize compound.
The items are carefully packaged to preserve precise surface finishes.
Do not remove items from the packaging
until they are ready to be used. Special
tools needed for this procedure are part of
an available O
Cell Disassembly Kit (Table
2
10-1).
Use heat-resistant gloves and clothing
when removing the OCX 4400. Do not
attempt to work on any sensor housing components until they have cooled
to room temperature. Components can
be as hot as 2600°F (1427°C). This can
cause severe burns.
Disconnect and lock out power before
working on any electrical components.
There may be voltage up to 264 VAC.
2
(f) Replace cover on electronics
housing.
(g) Connect the three calibration gas
input lines and the instrument air
input line to the split-electronics
housing. Connect the calibration
gas outlet line and instrument air
outlet line to the split-electronics
housing.
(h) Turn on the calibration gasses at
the cylinders and instrument air.
(i) Restore power to the system.
7-3 SENSOR HOUSING COMPONENT
REPLACEMENT
Each of the following procedures details how to
remove and replace a specific component located in the sensor housing of the OCX 4400.
a. O
Cell Replacement
2
This paragraph covers oxygen sensing cell
replacement. Do not attempt to replace the
cell until all other possibilities for poor per-
Do not remove the O2 cell unless certain it needs to be replaced. Removal
may damage the cell and platinum
pad. Only remove the O2 cell if it is to
be replaced.
1. Follow the instructions in paragraph
7-2a.1 to remove the OCX 4400 from
the stack or duct. If removing a splitmounted unit, follow the instructions in
paragraph 6-2b1 to remove the sensor
housing.
2. Remove sensor housing cover
(1, Figure 7-4).
3. Remove reference air tube (2) from
sensor housing (23).
4. Disconnect and tag all O
assembly (3)
2
wire leads from the sensor housing
terminal blocks.
5. Remove the O
assembly (3) from sen-
2
sor housing (23).
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
1
2
1. Cover
2. Reference Air Tube
Assembly
3. O
2
4. Gasket
5. Internal Ring
6. Eductor Air Tube
7. Dilution Air Tube
8. Screw
9. Washer
10. Bracket
11. COMB Extractive Tube
12. Heater
13. Screw
14. Plate
15. COMB Sensor
16. Gasket
17. Preheater
18. Heater
19. Connector
20. Eductor
21. COMB Sensor Holder
22. Thermocouple
23. Sensor Housing
24. Probe Tube
25. Exhaust Tube
26. Filter
27. Adapter
28. Check Valve (with
blowback option only)
29. Plug (without blowback option)
30. Screw
24
28
27
12
6
5
13
17
21
23
7
14
15
16
30
18
25
34070022
3
8
9
4
10
11
5
19
8
9
22
26
29
20
Figure 7-4. Sensor Housing, Exploded View
7-6 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
6. Remove four allen cap screws (14,
Figure 7-5) from O
move O
cell and gasket (12). The cell
2
cell (13) and re-
2
flange has a notch that may be used to
gently pry the flange away from heater
tube (11). Note that the pad on the end
of contact/thermocouple assembly (4)
will sometimes fuse to O
cell is fused to the contact pad,
the O
2
push the O
cell back into the heater
2
cell (13). If
2
tube (against spring pressure) and
quickly twist the O
cell. The cell and
2
contact pad should separate. If the
contact pad stays fused to the cell, a
new contact/thermocouple assembly
must be installed.
7. Remove and discard gasket (12).
Clean the mating surface of heater
tube (11). Remove burrs and raised
surfaces with a block of wood and crocus cloth.
8. Rub a small amount of anti-seize compound on both sides of new gasket
(12).
9. Assemble O
cell (13), gasket (12), and
2
heater tube (11). Make sure the calibration gas passage holes line up with
each other in all components. Apply a
small amount of anti-seize compound
to the screw threads and use screws
(14) to secure assembly. Torque to
35 in-lbs (4 N·m).
10. Apply anti-seize compound to threads
of O
assembly (3, Figure 7-4) and
2
sensor housing (23). Replace graphoil
gasket (4) with new gasket in kit. Reinstall O
11. Reconnect the wire leads from O
assembly in sensor housing.
2
2
assembly (3) to the sensor housing terminal blocks. Refer to Figure 7-2 if
necessary.
12. Replace reference air tube (2) in sensor housing (23). Ensure that the open
end of reference air tube extends into
O
assembly (3).
2
13. Follow the instructions in paragraph
7-2a.2 to install the OCX 4400 into the
stack or duct. If installing a splitmounted unit, follow the instructions in
paragraph 7-2b.2.
14. Turn on power and after the OCX 4400
stabilizes, calibrate the unit. If new
components have been installed, repeat calibration after 24 hours of
operation.
b. O
Heater Strut Assembly and
2
Contact/Thermocouple
Assembly Replacement
This paragraph covers O
heater strut and
2
contact/thermocouple replacement. If heater
strut or contact/thermocouple replacement
is needed, order the replacement parts from
Table 10-1.
Use heat-resistant gloves and clothing
when removing the OCX 4400. Do not
attempt to work on any sensor housing components until they have cooled
to room temperature. Components can
be as hot as 2600°F (1427°C). This can
cause severe burns.
Disconnect and lock out power before
working on any electrical components.
There may be voltage up to 264 VAC.
1. Follow the instructions in paragraph
7-2a.1 to remove the OCX 4400 from
the stack or duct. If removing a splitmounted unit, follow the instructions in
paragraph 7-2b.1 to remove the sensor
housing.
2. Remove sensor housing cover
(1, Figure 7-4).
3. Remove reference air tube (2) from
sensor housing (23).
4. Disconnect and tag O
assembly (3)
2
wire leads from the sensor housing
terminal blocks.
5. Remove four screws (1, Figure 7-5)
and washers (2) securing O
heater
2
strut assembly (3) to heater tube (11).
The spring tension should release and
the heater strut assembly moves up.
6. Carefully slide heater strut assembly
(3) out of heater tube (11).
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
4
1. Screw
2. Washer
Heater Strut Assembly
3. O
2
4. Contact/Thermocouple
Assembly
5. Strut Bracket
6. Spring
7. Spring Clip
8. Screw
9. Washer
10. Heater Assembly
11. Heater Tube
12. Gasket
Cell
13. O
2
14. Cap Screw
15. Cell Kit (includes 12,
13, and 14)
1
2
5
3
6
7
8
9
10
11
12
13
14
15
34070009
Figure 7-5. O2 Assembly, Exploded View
7-8 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
Use care when handling contact and
thermocouple assembly. The ceramic
rod in this assembly is fragile.
7. If replacing contact/thermocouple assembly (4), use a pencil to mark location of spring clip (7) before removing.
Squeeze tabs on spring clip to remove.
Retain spring clip and spring (6); replace if damaged.
8. While carefully handling new contact
and thermocouple assembly (4), lay old
assembly next to new one. Transfer
pencil marks to new assembly.
9. Carefully guide new contact and thermocouple assembly (4) through strut
bracket (5), spring (6), and spring clip
(7) until spring clip reaches pencil
mark.
10. Carefully slide new O
heater strut as-
2
sembly (3) into heater tube (11). Press
down on the back plate of strut bracket
(5) to ensure spring (6) tension is present to hold contact pad against O
(13). Secure with four screws (1) and
washers (2).
11. Reconnect the O
assembly wire leads
2
to the sensor housing terminal blocks.
Refer to Figure 7-2 if necessary.
cell
2
Use heat-resistant gloves and clothing
when removing the OCX 4400. Do not
attempt to work on any sensor housing components until they have cooled
to room temperature. Components can
be as hot as 2600°F (1427°C). This can
cause severe burns.
Disconnect and lock out power before
working on any electrical components.
There may be voltage up to 264 VAC.
1. Follow the instructions in paragraph
7-2a.1 to remove the OCX 4400 from
the stack or duct. If removing a splitmounted unit, follow the instructions in
paragraph 7-2b.1 to remove the sensor
housing.
2. Remove sensor housing cover (1,
Figure 7-4).
3. Disconnect and tag the four wire leads
from COMB sensor (15) to the sensor/thermocouple terminal block.
4. Remove two screws (13) securing
plate (14) to holder (21). Remove the
plate.
5. Remove old COMB sensor (15) and
gasket (16). Replace with new COMB
sensor and gasket from replacement
kit.
12. Replace reference air tube (2, Figure
7-4) in sensor housing (23). Ensure the
open end of the reference air tube extends into O
assembly (3).
2
13. Replace sensor housing cover (1).
14. Follow the instructions in paragraph
7-2a.2 to install the OCX 4400 into the
stack or duct. If installing a splitmounted unit, follow the instructions in
paragraph 7-2b.2.
c. COMB Sensor Replacement
This paragraph covers combustible sensor
replacement. Do not attempt to replace the
sensor until all other possibilities for poor
performance have been considered. If sensor replacement is needed, order the combustibles sensor replacement kit from Table
10-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-9
Use care when handling the new combustibles sensor. The RTDs are fragile.
6. Secure new COMB sensor (15) with
plate (14) to holder (21) with two
screws (13).
7. Reconnect the four wire leads from
COMB sensor (15) to the sensor/thermocouple terminal block. Refer
to Figure 7-2 if necessary.
8. Replace sensor housing cover
(1, Figure 7-4).
9. Follow the instructions in paragraph
7-2a.2 to install the OCX 4400 into the
stack or duct. If installing a splitmounted unit, follow the instructions in
paragraph 7-2b.2.
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
10. Turn on power and monitor COMB
sensor output. After the OCX 4400
stabilizes, calibrate the unit. If new
components have been installed, repeat calibration after 24 hours of
operation.
7-4 ELECTRONICS HOUSING COMPONENT
REPLACEMENT
OCX 4400
When installing a solenoid, do not
overtighten. Damage to the solenoid
may occur.
4. Install the new solenoid base. Be
careful not to overtighten. Install the
new washer and coil assembly and secure with the top nut.
Each of the following procedures details how to
remove and replace a specific component located in the electronics housing of the OCX
4400.
Disconnect and lock out power before
working on any electrical components.
There may be voltage up to 264 VAC.
a. Solenoid Replacement
The OCX 4400 has three optional calibration gas solenoids (HI COMB, HI O
) and a standard, 3-way instrument air
O
2
solenoid. Use the following procedure to replace one of four solenoids (11, Figure 7-6)
in the electronics housing. The replacement
procedure for all solenoids is similar.
1. Follow the instructions in paragraph
7-2a to remove the OCX 4400 from the
stack or duct. If removing a splitmounted unit, follow the instructions in
paragraph 7-2b.3 to remove the electronics housing.
, LO
2
5. Connect the leads to the proper terminations on the solenoid power terminal
block. Refer to Figure 2-5 if necessary.
6. Follow the instructions in paragraph
7-2a.2 to install the OCX 4400 into the
stack or duct. If installing a splitmounted unit, follow the instructions in
paragraph 7-2b.4 to install electronics
housing.
b. Pressure Sensor Replacement
Use the following procedure to replace
pressure sensor (10, Figure 7-6).
1. Follow the instructions in paragraph
7-2a to remove the OCX 4400 from the
stack or duct. If removing a splitmounted unit, follow the instructions in
paragraph 7-2b.3 to remove the electronics housing.
2. Tag and remove pressure sensor (10)
leads from sensor/thermocouple terminal block.
3. Place a 3/4 in. deep socket over pressure sensor (10) and remove from
electronics housing base (12).
2. Tag and disconnect wire leads from
solenoid (11, Figure 7-6) to be replaced at the solenoid power terminal
block.
3. Remove top nut of solenoid (11) securing the coil assembly and washer to
the solenoid base. Remove the coil assembly, including the leads, and
washer. Place a 13/16 in. deep socket
over the solenoid base and remove
from electronics housing base (12).
7-10 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
When installing the pressure sensor,
do not overtighten. Damage to the
pressure sensor may occur.
4. Install new pressure sensor (10) into
electronics housing base (12).
5. Connect leads to the proper terminations on the sensor/thermocouple terminal block. Refer to Figure 2-6 if
necessary.
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
6. Follow the instructions in paragraph
7-2a.2 to install the OCX 4400 into the
stack or duct. If installing a split-
1. Cover
2. Screw
3. Local Operator Interface
4. Washers
5. Spacer
6. Connector
7. Electronics Box
8. Screw
9. Washer
10. Pressure Sensor
11. Solenoid
12. Electronics Housing Base
13. Screw
14. Nameplate
15. Fitting
16. Instrument Air Solenoid
17. Vent Tube
mounted unit, follow the instructions in
paragraph 7-2b.4 to install electronics
housing.
1
2
3
4
5
6
7
17
8
16
9
14
13
10
12
34070005
11
15
Figure 7-6. Electronics Housing Exploded View
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-11
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
c. Interface Board Replacement
Use the following procedure to replace interface board (22, Figure 7-7).
1. Follow the instructions in paragraph
7-2a.1 to remove the OCX 4400 from
the stack or duct. If removing a splitmounted unit, follow the instructions in
paragraph 7-2b.3 to remove the electronics housing.
2. If equipped, remove local operator interface (3, Figure 7-6) by removing
three screws, double washers (4),
spacers (5) and connector (6) securing
local operator interface to electronics
box assembly (7).
3. Remove four screws (24, Figure 7-7)
and lock washers (23) securing interface board (22) to the electronics assembly. Install new interface board to
electronics box assembly (7, Figure
7-6) with four screws and lock
washers.
4. If equipped, install local operator interface (3, Figure 7-6) to electronics box
assembly (7) with three screws (2),
double washers (4), spacers (5), and
connector (6).
5. Follow the instructions in paragraph
7-2a.2 to install the OCX 4400 into the
stack or duct. If installing a splitmounted unit, follow the instructions in
paragraph 7-2b.4 to install electronics
housing.
d. Power Supply Board, Microprocessor
Board and Analog Board Replacement
Use the following procedure to replace
power supply board (14, Figure 7-7), microprocessor board (11), or analog board (6).
Replace boards as needed.
1. Follow the instructions in paragraph
7-2a.1 to remove the OCX 4400 from
the stack or duct. If removing a splitmounted unit, follow the instructions in
paragraph 7-2b.3 to remove the electronics housing.
2. Follow the instructions in paragraph
7-4c to remove interface board (22,
Figure 7-7).
3. Disconnect and tag wire leads from
solenoids and the pressure sensor to
the solenoid power and sensor/
thermocouple terminal blocks.
4. If replacing components on an integral
electronics mounted unit, disconnect
and tag wire leads from the sensor
housing at the sensor/thermocouple,
heater power, and solenoid power terminal blocks.
5. Remove four screws (8, Figure 7-6)
and washers (9) securing electronics
box (7) to electronics housing base
(12). Remove the electronics box.
6. Remove four screws (1 and 20, Figure
7-7) and washers (2 and 19) securing
top (3) and bottom (21) of electronics
box together. Power supply board (14),
microprocessor board (11), and analog
board (6) stack can now be removed.
7. Remove four screws (25 and 32), lock
washers (26 and 31) and standoffs (27)
securing power supply board (14) and
micro-processor board (11).
8. Remove two screws (4), washers (5),
and standoffs (10) securing analog
board (6) and microprocessor board
(11) to power supply board (14). Note
board orientation for proper reassembly. Separate analog board and microprocessor board from power supply
board. Note orientation of connector
(28).
9. Remove two screws (29), washers
(30), nuts (35), washers (34), and
standoffs (33) securing analog board
(6) to microprocessor board (11). Note
board orientation for proper reassembly. Separate analog board (6) from
microprocessor board (11). Note orientation of connectors (7, 8 and 9).
7-12 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
1. Screw
2. Washer
3. Electronics Box, Top
4. Screw
5. Washer
6. Analog Board
7. Connector
8. Connector
9. Connector
10. Standoff
11. Microprocessor Board
12. Covers
13. Standoff
14. Power Supply Board
15. Fuse - Heater, F2
16. Fuse Holder
17. Washer
18. Standoff
19. Washer
20. Screw
21. Electronics Box, Bottom
22. Interface Board
23. Washer
24. Screw
25. Screw
26. Washer
27. Standoff
28. Connector
29. Screw
30. Washer
31. Washer
32. Screw
33. Standoff
34. Washer
35. Nut
36. Fuse - Power Supply, F1
32
31
27
26
34
33
35
1
2
3
4
5
6
28
30
7
8
12
29
9
10
11
13
14
15
16
24
23
25
21
22
36
16
17
18
19
20
34070012
Figure 7-7. Electronics Box, Exploded View
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 7-13
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
10. Assemble analog board (6) to microprocessor board (11) in proper orientation with connectors (7, 8 and 9), two
screws (29), washers (30), standoffs
(33), washers (34), nuts (35), two
screws (4), washers (5), and standoffs
(10).
11. Assemble analog board (6) microprocessor board (11) stack to power supply
board (14) with four screws (25 and
32), lock washers (26 and 31) and
standoffs (27). Install two standoffs
(18) and washers (17) through bottom
of power supply board (14) into standoffs (13).
12. Place assembled electronics board
stack into electronics box bottom (21).
Assemble bottom and top (3) of electronics box with four screws (1 and 20)
and washers (2 and 19).
13. Assemble interface board (22) to the
assembled electronics box with four
screws (24) and washers (23).
17. Follow the instructions in paragraph
7-2a.2 to install the OCX 4400 into the
stack or duct. If installing a splitmounted unit, follow the instructions in
paragraph 7-2b.4 to install electronics
housing.
e. Fuse Replacement
The OCX 4400 has two fuses (15 and 36,
Figure 7-7) on power supply board (14). The
power supply fuse, F1 is to the left of the AC
power terminal block. The heater fuse, F2 is
to the right of the AC power terminal block.
Both fuses are accessible through holes in
the electronics box. Refer to Table 10-1 for
replacement fuse specifications. Perform
the following procedure to check or replace
fuses.
Disconnect and lock out power before
working on any electrical components.
There may be voltage up to 264 VAC.
14. If equipped, install local operator interface (3, Figure 6-6) to electronics box
assembly (7) with three screws (2),
double washers (4), spacers (5), and
connector (6).
15. If replacing components on an integral
electronics mounted unit, connect wire
leads from the sensor housing to sensor/thermocouple, heater power, and
solenoid power terminal blocks.
16. Connect wire leads from the solenoids
and the pressure sensor to the solenoid power and sensor/thermocouple
terminal blocks.
3. Remove fuseholder (16, Figure 7-7) by
pushing in the top and turning 1/4 turn
counterclockwise. Remove fuse (15).
4. After checking or replacing fuse (15),
install fuseholder (16) by pushing in the
top and turning 1/4 turn clockwise.
5. Repeat for fuse (36).
7-14 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
8
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 8
TROUBLESHOOTING
d. Electrostatic Discharge
Install all protective equipment covers
and safety ground leads after troubleshooting. Failure to install covers and
ground leads could result in serious
injury or death.
8-1 GENERAL
The troubleshooting section describes how to
identify and isolate faults that may develop in
the OCX 4400. When troubleshooting the OCX
4400, reference the following information.
a. Grounding
It is essential that adequate grounding precautions are taken when installing the system. Thoroughly check both the probe and
electronics to ensure the grounding quality
has not degraded during fault finding. The
system provides facilities for 100% effective
grounding and the total elimination of
ground loops.
b. Electrical Noise
The OCX 4400 has been designed to operate in the type of environment normally
found in a boiler room or control room.
Noise suppression circuits are employed on
all field terminations and main inputs. When
fault finding, evaluate the electrical noise
being generated in the immediate circuitry
of a faulty system. Also, ensure all cable
shields are connected to earth.
c. Loose Integrated Circuits
Electrostatic discharge can damage the ICs
used in the electronics. Before removing or
handling the processor board or the ICs,
ensure you are at ground potential.
e. Line Voltage Configuration
See Section 3-1b for specific instructions.
8-2 ALARM OUTPUT
a. The OCX 4400 is equipped with a set of
alarm relay contacts on the microprocessor
board in the electronics housing. This set of
dry contacts can be connected to any customer supplied relay device, 30 VDC, 30
mA maximum. Any fault condition in the
OCX 4400 will trip the alarm relay.
b. The optional SPA with HART programmable
alarm indicates LOW O
, HIGH COMB,
2
Calibration Status, and Unit Failure. For
more information refer to Appendix B, SPA
with HART Alarm.
8-3 FAULT INDICATIONS
Faults in the OCX 4400 are indicated by messages displayed on the HART communicator or
LOI. The following faults are as they appear on
the HART communicator. Messages on the LOI
may appear slightly different.
a. O
Thermocouple Error
2
O
thermocouple voltage is not within speci-
2
fied operating parameters.
The OCX 4400 uses a microprocessor and
b. O
Cell Temperature Low
2
supporting integrated circuits (IC). If the
O
electronics are handled roughly during in-
cell temperature is less than 726°C.
2
stallation or located where subjected to severe vibration, the ICs could work loose.
c. O
Cell Temperature High
2
Before troubleshooting the system, ensure
O
all ICs are fully seated.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 8-1
cell temperature is greater than 746°C.
2
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
d. O2 Cell Temperature Very High
O
cell temperature is greater than 820°C.
2
The OCX 4400 requires a power reset.
Heater Open
e. O
2
The O
f. O
The O
cell heater is not operating properly.
2
Cell Open
2
cell voltage is not within specified
2
parameters.
g. O
Cell Bad
2
Unit is not in warm up or power up mode
and O
cell impedance value is greater than
2
500 ohms.
Slope Error
h. O
2
The newly calculated O
slope is greater
2
than 57.5 or less than 34.5 mV/Dec.
Constant Error
i. O
2
The newly calculated O
constant is greater
2
than 20 mV or less than –20 mV.
o. Sample Line Temperature High
The unit is not in warm up mode and the
sample line temperature is greater than
383°F (195°C).
p. Sample Line Temperature Very High
Sample line temperature is greater than
220°C. The OCX 4400 requires a power
reset.
q. Sample Pressure Low Absolute
The sample pressure is low.
r. Comb Cell Error
The temperature of the reference side or the
active side of the comb cell is below -40°F
(-40°C) (84.27 mV) or above 752°F (400°C)
(247.04 mV).
s. Comb Cell Chamber Temperature Low
The unit is not in warm up mode and the
combustible cell chamber is less than 554°F
(290°C).
t. Comb Cell Chamber Temperature High
Calibration Failed
j. O
2
Slope Error or an O2 Constant Error
An O
2
occurred during the last calibration.
k. High Electronics Temperature
The electronics temperature is above 85°C.
l. EEPROM Corrupt
The EEPROM has been corrupted by some
means.
m. Sample Line Thermocouple Error
Sample line thermocouple voltage is not
within specified parameters.
n. Sample Line Temperature Low
Unit is not in warm up mode and the sample
line temperature is less than 311°F (155°C).
The unit is not in warm up mode and the
combustible cell chamber temperature is
greater than 590°F (310°C).
u. Comb Cell Chamber Temperature
Very High
Combustible cell chamber temperature is
greater than 400°C. The OCX 4400 requires
a power reset.
v. Comb Slope Error
The newly calculated combustible slope is
greater than 3000 or less than 0 ppm/mV.
w. Comb Constant Error
The newly calculated combustible constant
is not within acceptable parameters.
x. Comb Calibration Failed
The Comb Slope Error has occurred during
the last calibration.
8-2 TroubleshootingRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
9
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 9
RETURN OF MATERIAL
9-1 If factory repair of defective equipment is re-
quired, proceed as follows:
a. Secure a return authorization number from
a Rosemount Analytical Sales Office or representative before returning the equipment.
Equipment must be returned with complete
identification in accordance with Rosemount
instructions or it will not be accepted.
In no event will Rosemount be responsible
for equipment returned without proper
authorization and identification.
b. Carefully pack defective unit in a sturdy box
with sufficient shock absorbing material to
insure that no additional damage will occur
during shipping.
c. In a cover letter, describe completely:
1. The symptoms from which it was determined that the equipment is faulty.
2. The environment in which the equipment has been operating (housing,
weather, vibration, dust, etc.).
5. Complete shipping instructions for return of equipment.
6. Reference the return authorization
number.
d. Enclose a cover letter and purchase order
and ship the defective equipment according
to instructions provided in Rosemount Return Authorization, prepaid, to:
Rosemount Analytical Inc.
RMR Department
1201 N. Main Street
Orrville, Ohio 44667
If warranty service is requested, the defective unit will be carefully inspected and
tested at the factory. If failure was due to
conditions listed in the standard Rosemount
warranty, the defective unit will be repaired
or replaced at Rosemount's option, and an
operating unit will be returned to the customer in accordance with shipping instructions furnished in the cover letter.
3. Site from which equipment was
removed.
4. Whether warranty or nonwarranty
service is requested.
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 9-1
For equipment no longer under warranty,
the equipment will be repaired at the factory
and returned as directed by the purchase
order and shipping instructions.
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
9-2 Return of MaterialRosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
10
IB-106-440 Rev. 1.0
OCX 4400
SECTION 10
REPLACEMENT PARTS
Table 10-1. Replacement Parts
FIGURE and
INDEX No.PART NUMBERDESCRIPTION
6-4, 245R10183H02Pipe, 18 in. (457mm) 316 Stainless steel
6-4, 245R10183H06Pipe, 18 in. (457mm) Inconel 600
6-4, 245R10227G01Pipe, 18 in. (457mm) Ceramic
6-4, 245R10183H03Pipe, 3 ft. (0.91m) 316 Stainless steel
6-4, 245R10183H07Pipe, 3 ft. (0.91m) Inconel 600
6-4, 245R10227G02Pipe, 3 ft. (0.91m) Ceramic
6-4, 245R10183H04Pipe, 6 ft. (1.83m) 316 Stainless steel
6-4, 245R10183H08Pipe, 6 ft. (1.83m) Inconel 600
6-4, 245R10183H05Pipe, 9 ft. (2.7m) 316 Stainless steel
6-4, 255R10183H01Pipe, Exhaust
6-4, 15R10219G01Housing Cover, Blind
6-4, 4 and 6-5, 154851B45G01O2 Cell Replacement Kit, all sizes
4851B45G02O2 Cell Disassembly Kit
6-4, 265R10246H02Filter, 20 micron
October 2000
6-5, 35R10211G01Heater Strut Assy.
6-5, 44851B44G01Contact and Thermocouple Assy.
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 10-1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
Table 10-1. Replacement Parts (Continued)
FIGURE and
INDEX No.PART NUMBERDESCRIPTION
1-1, 5263C152G01Reference Gas Set
1A99119H01Oxygen Bottles 0.4% Balance N2
1A99119H02Oxygen Bottles 8.0% Balance N2
1A99119H06CO Bottle 1000 ppm, Balance N2
1A99120H02Regulator for Oxygen
1A99120H03Regulator for Combustibles
1A99119G06Wall Mount Bracket for Gas Bottles
1-1, 51A99119G05Test Gas Regulator Kit
1-1, 51A99119G04Test Gas Bottles Kit
1-1, 61A99518G01Blowback Kit
1-1, 2275D9EI5D0000275 HART Communicator with 12 meg buffer
1-44851B41G01Cable, 20 ft. (6m) NEMA
1-44851B41G02Cable, 40 ft. (12m) NEMA
1-44851B41G03Cable, 60 ft. (18m) NEMA
1-44851B41G04Cable, 80 ft. (24m) NEMA
1-44851B41G05Cable, 100 ft. (30m) NEMA
1-44851B41G06Cable, 150 ft. (45m) NEMA
OCX 4400
11-41A99292H01Moore Industries SPA for HART Alarm
10-2 Replacement PartsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
11
APPENDIX A. LOCAL OPERATOR INTERFACE (LOI)
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SECTION 11
APPENDICES
11-1 FORWARD
This appendix describes the installation and operation of the Local Operator Interface (LOI) for
the OCX 4400.
CONNECTOR
11-2 INSTALLATION
The LOI mounts to the LOI connector on the
LOI board located at the end of the electronics
box in the electronics housing of an OCX 4400
Oxygen/Combustibles Transmitter Figure 11-1.
There are four matching connectors on the back
of the LOI assembly that allow the LOI to be oriented as desired by the user.
ELECTRONIC HOUSING
(COVER REMO VED)
LOI
LOI
ASSEMBLY
ELECTRONICS
BOX
LOI
BOARD
29790024
Figure 11-1. LOI Assembly Mounting
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 11-1
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
11-3 OPERATION
The LOI assembly, Figure 11-2, allows local
communication with the electronics. The LOI
uses infrared sensors near the selection arrows
to move the cursor between menus and menu
items. The LCD display window shows the operating display, which consists of the O
COMB ppm, current menu, or item selected.
When a blinking cursor is displayed, the UP or
DOWN arrows can be used to increase or decrease the value for that item.
SELECTION
ARROWS
% and
2
LCD
DISPLA Y
WINDOW
11-4 LOI MENU TREE
This section consists of a menu tree for the LOI
on the OCX 4400, Figure 11-3. This menu is
specific to the OCX 4400.
a. First Column Sub-Menus
From the operating display, O
COMB ppm, the right arrow is the only option to move into the first column sub-menus
of the LOI menu tree. The first column contains three sub-menus: SENSOR DATA,
Figure 11-3 sheet 1 of 3, CALIBRATE SENSORS, Figure 11-3 sheet 2 of 3, and
SETUP, Figure 11-3 sheet 3 of 3. From the
operating display, SENSOR DATA is displayed when the right arrow is selected. Use
the down arrow to move to the other first
column sub-menus.
b. Second Column Sub-Menus
From the first column sub-menus, selecting
the right arrow moves the display into the
second column sub-menus. The up and
down arrows allow the display to move to
the second column sub-menus of the first
column sub-menu selected. The left arrow
moves the display back to the first column
sub-menu selected.
% and
2
c. Third, and Fourth Column Sub-Menus
SELECTION
ARROWS
Figure 11-2. Local Operator Interface (LOI)
Assembly
From the second column sub-menus, selecting the right arrow moves the display
into the third column sub-menus. The third
column sub-menu may be another menu or
a list of parameters. The up and down arrows allow the display to move to the different parameters or menus. The third or
fourth column may be a parameter list.
When a parameter list is displayed, the cursor will blink. The up and down arrows allow
the value for the parameter displayed to be
changed.
11-2 AppendicesRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
11
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
o
o
o
o
o
C
C
o
o
C
34070017
O2 X.XX%
COMB XXX ppm
SENSOR
DATA
(CONTINUED ON
SHEET 2)
TEMP/
PRESSURES
VOLTAGES
OUTPUT
VALUES
O2 Sensor _____ C
Comb Sensor _____ C
Sample Block Temp _____ C
Sample Block Press _____
Elect Temp _____ C
Max Elect Temp _____
O2 Sensor ___mV
O2 Sensor T/C ___mV
Comb Delta ___
Comb Reference ___mV
Elect Temp ___mV
Sample Block T/C ___mV
Sample Line Pressure ___V
O2 Analog _____%
Comb Analog _____%
O2 Analog ____mA
Comb Analog ____mA
Figure 11-3. Menu Tree for Local Operator Interface on the OCX 4400 (Sheet 1 of 3)
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 11-3
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
(CONTINUED FROM
SHEET 1)
CALIBRATE
(CONTINUED ON
SHEET 3)
Figure 11-3. Menu Tree for Local Operator Interface (LOI) on the OCX 4400 (Sheet 2 of 3)
DO
CALIBRATION
VIEW
CONSTANTS
CALIBRATION
STATUS
If Auto Cal in Setup, YES
Press --> To Start
Automatic Calibration
Start Auto Cal O2
High O2 Gas
Time Left: 0:00 _____% O2
O2 Cell mV_____mV
Low Gas
Time Left: 0:00 _____% O2
O2 Cell mV_____mV
Calibration Complete
Purging 0:00
O2 Calibration Pass/Fail
Start Auto Cal Comb
Low Gas
Time Left: 0:00 _____Comb
Comb Cell Output _____
High Gas
Time Left: 0:00 _____Comb
Comb Cell Output _____
Calibration Complete
Purging 0:00
Comb Calibration
If Auto CAL in Setup, NO
--> To Start O2
Press
Manual Calibration
Switch on High Test Gas
--> When ready
Press
High Gas _____% O2
-->
PressWhen ready
Turn Off High Test Gas and
On Low Test Gas
-->
PressWhen ready
Low Gas ______% O2
--> When O2
Press
reading is stable
Turn Off Low Test Gas
-->
PressWhen ready
O2 Calibration Pass/Fail
--> When Probe has
Press
returned to process
--> To Start Comb
Press
Manual Calibration
Switch On Low Test Gas
-->
PressWhen Ready
Low Gas _____ Comb
--> When Comb
Press
reading is stable
Turn Off Low Test Gas and
on High Test Gas
-->
PressWhen ready
High Gas _____Comb
--> When Comb
Press
reading is stable
Turn Off High Test Gas
-->
PressWhen ready
Comb Calibration Pass/Fail
--> When Probe has
Press
returned to process
LATEST
CALIBRATION
PREVIOUS
CALIBRATION
Calibrate Step
Calibrate Time
O2 Next Calib
O2 Slope
O2 Constant
Comb Next Calib
Comb Slope
Comb Constant
Pass/Fail
OCX 4400
O2 Slope _______
O2 Constant _______
O2 Cell Imped _______
Comb Slope _______
Comb Constant _______
Pre O2 Slope _______
Pre O2 Constant _______
Pre Comb Slope _______
Pre Comb Constant _______
34070018
11-4 AppendicesRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
11
(CONTINUED FROM
SHEET 2)
SETUP
CALIBRATION
PARAMETERS
BLOWBACK
O2 High Gas _____% O2
O2 Low Gas _____% O2
Comb Test Gas _____% O2
Got Solenoids Yes/No
O2 Output Tracks Yes/No
Comb
Out Tracks Yes/No
O2 Cal Interval XDXH
O2
Comb Cal Interval XDXH
Next CalXH
Comb Next CalXH
Gas Time0:00
Purge Time0:00
Line Voltage _____
O2 Slope _____ mV/D
O2 Constant _____ mV
Comb Slope _____
Comb Constant _____
Blow Bk Present Yes/No
Blow Bk Intvl 99999 min
Blow Bk Durat 99999 sec
Blow Bk Purge 999 sec
Do Blow Back Yes/No
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
ALARM RELAY
ANALOG
OUTPUTS
SOFTWARE
Figure 11-3. Menu Tree for Local Operator Interface (LOI) on the OCX 4400 (Sheet 3 of 3)
11-5 AUTOMATIC CALIBRATION SETUP
Use the following procedure to setup the OCX
4400 with the LOI for automatic calibration. If
necessary, use the menu tree in Figure 11-3 for
reference. The unit must be equipped with calibration solenoids to use automatic calibration.
Trigger 1 Event
Trigger 2 Event
Trigger 3 Event
O2 Type 20-4 mA
Comb Type 20-4 mA
O2 Range 0-VAR (40% Max.)
Comb Range50,000 ppm
Do O2 Out Trim
Do COMB Out Trim
Version
Checksum
Build Number
Build Date
Test Code
4-20 mA
4-20 mA
HART
a. From the operating display select SETUP
first column sub-menu.
b. From the SETUP first column sub-menu,
select CALIBRATION second column submenu.
Off
All
Comb Temp
Sample Pr Sen
Sample Ln Temp
Unit Fail
Hi Elect Temp
Calib Fail
O2 Cell Bad
O2 Htr Open
O2 Cell Temp
In Calibrtn
Off
34070019
NOTE
c. From the CALIBRATION second column
sub-menu, select the third column parame-
Automatic calibration is only available
ter list by selecting the right arrow.
on units equipped with calibration
solenoids.
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 11-5
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
d. Scroll down to the fifth item Auto Cal. If the
unit is equipped with calibration solenoids
and timed automatic calibration is desired,
select Yes.
e. Use the down arrow to select the next item
O
Output Tracks. Select Yes or No to de-
2
termine if updates to the O
lock value will
2
take place.
f. Use the down arrow to select the next item
CO Output Tracks. Select Yes or No to determine if updates to the COMB lock value
will take place.
g. Use the down arrow to select the next item
Cal Interval. Enter the amount of time in
days and hours that is desired between
automatic calibrations.
h. Use the down arrow to select the next item
Next Cal. Enter the amount of time in hours
until the next automatic calibration. Select
the left arrow three times to move back to
the operating display.
11-6 MANUAL CALIBRATION WITH LOI
Use the following procedure to perform a manual calibration with the LOI. If necessary, refer to
the menu tree in Figure 11-3 if needed. Once
the manual calibration procedure is initiated by
the LOI, a series of prompts will appear giving
instructions to the operator.
a. From the operating display select CALI-
BRATE SENSORS first column sub-menu.
b. From the CALIBRATE SENSORS first col-
umn sub-menu select the PERFORM
CALIBRA-TION second column sub-menu.
e. “Switch on High Test Gas”. Turn on the high
O
test gas.
2
f. Press the right arrow when the high O
test
2
gas reading is stable.
g. Turn off the high O
low O
test gas. Press the right arrow when
2
test gas and turn on the
2
ready.
h. Press the right arrow when the low O
test
2
gas reading is stable.
i. Turn off the low O
test gas. Press the right
2
arrow when ready.
j. The LOI will display “O
Calibration
2
Pass/Fail”.
k. Press the right arrow when the probe has
returned to the process.
l. Press the right arrow to start manual com-
bustibles calibration.
m. Turn on the high O
test gas. High O2 test
2
gas is used for the low COMB test gas.
Press the right arrow when ready.
n. Press the right arrow when the low com-
bustibles reading is stable.
o. Turn off low COMB test gas and turn on
high COMB test gas. Press the right arrow
when ready.
p. Press the right arrow when the high COMB
reading is stable.
q. Turn off the high COMB test gas. Press the
right arrow when ready.
c. From the PERFORM CALIBRATION sec-
ond column sub-menu select the third column parameter list by selecting the right
r. “Comb Calibration Pass/Fail” will appear.
Press the right arrow when the probe has
returned to the process.
arrow.
d. Scroll down to the “If Auto CAL in Setup,
NO” parameter and select the right arrow to
start manual O
11-6 AppendicesRosemount Analytical Inc. A Division of Emerson Process Management
calibration.
2
OCX 4400
11
APPENDIX B. SPA WITH HART ALARM
11-7 FORWARD
This appendix describes the Moore Industries
SPA with HART Alarm option available for the
OCX 4400.
11-8 DESCRIPTION
The Moore Industries SPA with HART, Figure
11-4, is a 4-wire (line or mains powered), siteprogrammable, digital process alarm. It
connects to a standard HART field device, and
provides up to four, fully configurable, contactclosure outputs based on readings of the HART
digital data. The four outputs from the OCX
4400 which are recognized by the SPA are Low
O
, High COMB, Calibration Status, and Unit
2
Failure.
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
11-9 INSTALLATION
Refer to the Users’ Reference Manual for the
SPA with HART provided with the SPA for detailed installation, setup, and operation
information.
Figure 11-4. SPA With HART Alarm
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 11-7
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
APPENDIX C. BLOWBACK
11-10 FORWARD
This appendix describes the operation of the
blow-back option for the OCX 4400
Oxygen/Combustibles Transmitter.
The blowback system blows instrument air back
through the inlet filter and out the inlet tube of
the transmitter. This removes built up dirt and
particulate from the filer and inlet line. The
blowback option is normally used in systems
that have a dirty process stream.
11-11 INSTALLATION
Installing an OCX 4400 Oxygen/Combustibles
Transmitter with the blowback option is identical
to installation without the blowback option with
the exception of the pneumatic piping. The
blowback option requires the addition of a blowback solenoid, regulator and gage, and check
valve.
Refer to Figure 11-5 for the piping arrangement
for the OCX 4400 Oxygen/Combustibles
Transmitter with the blowback option.
11-12 SETUP
There are three settings that need to be specified to set up the blowback option. These are
the blowback interval, duration, and purge time.
These settings are available through HART from
the Device Setup > Detailed Setup > Blowback
menu. Using the LOI, these settings are accessible from the Setup > Blowback menu.
Interval– Length of time between blow-
backs.
Duration– Length of time blowback air is
activated.
Purge Time – Length of time after blowback is
complete before oxygen/
combustibles readings are considered valid.
11-8 AppendicesRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400
11
BLOWBACK
INPUT
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
SENSOR
HOUSING
CAL GAS
INPUT
CAL GAS
ROTOMETER
ELECTRONICS
CAL GAS
HI COMB
INPUT
CHECK
VALVE
5 SCFH
HOUSING
CAL GAS
HI O2
INPUT
EDUCTOR
AIR INPUT
CALIBRA TION
GAS OUTLET
INSTR UMENT
INSTRUMENT
CAL GAS
LO O2
INPUT
AIR IN
AIR OUT
REF AIR
INPUT
DILUTION AIR
INPUT
BLOWBACK
SOLENOID
NORMALLY
OPEN
DILUTION AIR
ROTOMETER
0.1 SCFH
SET REGU L ATOR TO 35 PSIG
NORMALLY
CLOSED
SET REGULATOR TO 55 PSIG
0 - 60 PSIG
0 - 60 PSIG
34070020
Figure 11-5. Blowback Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 11-9
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
APPENDIX D. IS BARRIER
11-13 FORWARD
The IS BARRIER feature is currently under
development.
OCX 4400
11-10 AppendicesRosemount Analytical Inc. A Division of Emerson Process Management
SECTION 12
12
INDEX
This index is an alphabetized listing of parts, terms, and procedures having to do with the Hazardous Area Oxygen/Combustibles Transmitter. Every item listed in this index refers to a location
in the manual by one or more page numbers.
12-2 IndexRosemount Analytical Inc. A Division of Emerson Process Management
ROSEMOUNT WARRANTY
Rosemount warrants that the equipment manufactured and sold by it will, upon shipment, be free
of defects in workmanship or material. Should any failure to conform to this warranty become apparent during a period of one year after the date of shipment, Rosemount shall, upon prompt
written notice from the purchaser, correct such nonconformity by repair or replacement, F.O.B.
factory of the defective part or parts. Correction in the manner provided above shall constitute a
fulfillment of all liabilities of Rosemount with respect to the quality of the equipment.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES OF QUALITY WHETHER WRITTEN, ORAL, OR IMPLIED (INCLUDING ANY
WARRANTY OF MERCHANTABILITY OF FITNESS FOR PURPOSE).
The remedy(ies) provided above shall be purchaser's sole remedy(ies) for any failure of
Rosemount to comply with the warranty provisions, whether claims by the purchaser are based
in contract or in tort (including negligence).
Rosemount does not warrant equipment against normal deterioration due to environment. Factors
such as corrosive gases and solid particulates can be detrimental and can create the need for repair or replacement as part of normal wear and tear during the warranty period.
Equipment supplied by Rosemount Analytical Inc. but not manufactured by it will be subject to the
same warranty as is extended to Rosemount by the original manufacturer.
At the time of installation it is important that the required services are supplied to the system and
that the electronic controller is set up at least to the point where it is controlling the sensor heater.
This will ensure, that should there be a delay between installation and full commissioning that the
sensor being supplied with ac power and reference air will not be subjected to component
deterioration.
3582/8-01
Instruction Manual
IB-106-440 Rev. 1.0
October 2000
OCX 4400
OCX 4400
Part no._______________
Serial no. _______________
Order no. _______________
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
E gas.csc@emersonprocess.com
ASIA - PACIFIC
Fisher-Rosemount
Singapore Private Ltd.
1 Pandan Crescent
Singapore 128461
Republic of Singapore
T 65-777-8211
F 65-777-0947