Rosemount NGA 2000 TFID Hydrocarbon Analyzer Module SW 3.2 Software -1st Ed. Manuals & Guides

Rosemount Analytical
NGA 2000
Software Manual
TFID Analyzer
TFID Analyzer Module
(combined with NGA 2000 platform /
TFID analyzer / MLT analyzer
Software Version 3.2.X
1. Edition 06/98
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
Catalog No.: 90 003 808
Managing The Process Better
Rosemount Analytical
This Operation Manual includes information about the operation of the instrument. Information about the additional indications and notes regarding maintenance, troubleshooting and repair are found in the accompanying Maintenance & Operation Manual.
T roubleshooting, component replacement and internal adjustments must be made by qualified service personnel only.
Fisher-Rosemount GmbH & Co does not take responsibility for an y omissions or errors in this manual. Any liability for direct or indirect damages, which might occur in connection with the deliv ery or the use of this manual, is expressly e xcluded to the extend permitted by applicable law.
This instrument has left the works in good order according to safety regulations. T o maintain this operating condition, the user must strictly follow the instructions and consider the warnings in this manual or provided on the instrument.
Misprints and alterations reserved
©
1998 by FISHER-ROSEMOUNT GmbH & Co. (PAD)
1. Edition: 06/98
Read this operation manual carefully before attempting to operate the analyzer ! For expedient handling of reports of defects , please include the model and serial number which can be read on the instrument identity plate.
Fisher - Rosemount GmbH & Co.
Industriestrasse 1
D - 63594 Hasselroth • Germany
Phone + 49 (6055) 884-0
Telefax + 49 (6055) 884-209
Internet: http://www.processanalytic.com
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98

Contents

1 Introduction 1 - 1
2 Structure of Menus 2 - 1
3 Startup and Operation, General Notes and Main Menu 3 - 1
3.1 Starting and Initializing...................................................................................3 - 1
3.2 Display and Function...................................................................................... 3 - 6
3.3 "TAG" and Operating Keys ............................................................................ 3 - 6
3.4 Lines and Softkey Functionality ..................................................................... 3 - 7
3.5 Important Functions of the Softkeys...............................................................3 - 8
3.6 Entering/Changing of Variables......................................................................3 - 9
3.7 Starting a Function.........................................................................................3 -10
3.8 Main Menu .....................................................................................................3 -11
4 Basic Controls and Analyzer Module Calibration 4 - 1
4.1 Analyzer Channel Status................................................................................ 4 - 3
4.1.1 Status Details – Failures...............................................................................4 - 5
– Maintenance requests ........................................................ 4 - 9
– Function controls ................................................................ 4 -13
– Measurements/Alarms........................................................4 -15
– Events.................................................................................4 -17
– Acknowledge and clear failures..........................................4 -21
– Acknowledge and clear maintenance requests .................. 4 -25
– Acknowledge and clear function controls............................4 -29
4.1.2 Operational Settings.......................................................................................4 -33
4.2 Single Component Display - Change of Channel...........................................4 -35
4.3 Multi Component Display - Change of Channel............................................. 4 -37
4.4 Calibration Procedure Status.........................................................................4 -39
4.5 Zero Calibration..............................................................................................4 -41
4.6 Span Calibration.............................................................................................4 -45
4.7 Flow Zero Gas................................................................................................ 4 -49
4.8 Flow Span Gas...............................................................................................4 -53
4.9 Flow Sample Gas...........................................................................................4 -57
4.10 Flow Test Gas................................................................................................4 -61
4.11 Close all Valves.............................................................................................. 4 -65
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
I
5 Analyzer and I/O-Module Expert Configuration 5 - 1
5.1 Analyzer Module Setup 5 - 3
5.1.1 Calibration Parameters...................................................................................5 - 5
– Span gases.........................................................................................5 - 6
– Tolerances ..........................................................................................5 - 8
– Calibration procedure setup................................................................ 5 -10
– Time controlled calibration..................................................................5 -13
– Calibration...........................................................................................5 -15
– Advanced calibration methods............................................................ 5 -18
– Zero gases..........................................................................................5 -20
5.1.2 Alarm Parameters...........................................................................................5 -21
5.1.3 Range Parameters .........................................................................................5 -25
– Begin and end of ranges.....................................................................5 -27
– Response times (t90) ...........................................................................5 -28
– Autoranging control............................................................................. 5 -29
5.1.4 Cross Interference Compensation..................................................................5 -31
5.1.5 Linearization...................................................................................................5 -33
5.1.6 Programmable Logic Control (PLC)................................................................5 -37
5.1.7 Programmable Calculator...............................................................................5 -45
5.1.8 Measurement Display Configuration...............................................................5 -49
5.1.9 Acknowledgement of Status Reports..............................................................5 -52
5.1.10 Concentration Measurement Parameters.......................................................5 -54
5.1.11 Peak Measurement ........................................................................................5 -57
5.1.12 Differential Measurement ...............................................................................5 -59
5.1.13 Gasflow Setup................................................................................................5 -61
5.1.14 Pressure Compensation.................................................................................5 -62
5.1.15 Flow Measurement......................................................................................... 5 -64
5.1.16 Temperature Measurement............................................................................5 -65
5.1.17 Loading/Saving Configuration Parameters.....................................................5 -66
5.1.18 Inputs and Outputs.........................................................................................5 -69
– Local SIO ............................................................................................5 -70
– Local DIO............................................................................................5 -76
– Signal codes........................................................................................5 -77
5.1.19 Delay and Average.........................................................................................5 -81
5.1.20 Special Functions...........................................................................................5 -83
5.1.21 AK-Protocol Communication ..........................................................................5 -84
II
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
5.2 I/O Module Controls 5 - 85
5.2.1 SIO Module 5 - 86
5.2.2 DIO Module(s) 5 - 95
5.3 I/O Module Setup 5 -101
6 System Configuration 6 - 1
6.1 Diagnostic Menus........................................................................................... 6 - 3
6.1.1 Control Module Diagnostics.................................................................6 - 4
6.1.2 Analyzer Module Diagnostics ..............................................................6 - 5
6.2 Date and Time................................................................................................6 - 6
6.3 Security Codes...............................................................................................6 - 7
6.4 Network Module Binding ................................................................................6 - 9
6.5 System Reset.................................................................................................6 -11
6.6 System Modules.............................................................................................6 -12
7 Display Controls 7 - 1
Index
Supplement: System Calibration
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
III
IV
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
1 Introduction
This software manual describes step by step how to operate successfully with the 19" Thermo-FID (TFID) analyzer module and analyzer of the NGA 2000 Series. Chapter two shows the structure of the TFID software menus. Chapter three describes the display and the keyboard of the analyzer. Chapter four describes the basic controls with detailed illustrations. So you can easily compare the actual display of the analyzer module with the illustrations of the manual. Chapter five describes the expert configurations of the analyzer module and of the Input/ Output Modules (I/O modules). Chapter six describes the system configuration. The layout of both chapters is not as detailed as in chapter four. Normally, the way to a certain menu of the TFID software is described with the software catchwords you have to press to reach this menu. You will find the illustration of the corresponding LCD screen at the end of the catchword listing. After that you can read the meaning of the functions and variables of each expert or system configuration menu. Some contents of the expert configurations are not important for each customer. It depends on the configuration of your NGA 2000 system, relative to the following components:
Control Module CMAnalyzer Module AMInput/Output Modules I/O's (SIO = Standard I/O, DIO = Digital I/O,
System Auto Calibration I/O, Analog Output with 3 Alarms I/O, Auto Calibration I/O)
You can distinguish the following system units and SIO/DIO configurations:
System Unit SIO/DIO-Configuration
TFID analyzer module (AM):
without front panel,
i.e. without control unit
can be combined with a platform,
an MLT analyzer, a TFID analyzer or a customer developed control
1 local SIO and 1 local DIO can be
installed in the TFID analyzer module
SIO and DIO can be configured
for the channels of the analyzer module only
unit
Platform (CM Software):
Control unit with front panel
Without measurement channels
1 SIO and up to 4 DIO's can be
installed in the platform (CM I/O)
SIO and DIO can be configured
for all channels combined with the platform
Section
Page
5.1.18
p. 5-69
5.2
p. 5-85
TFID analyzer (CM and TFID AM software):
Analyzer with front panel
CM and AM software in the same
analyzer, i.e. all functions of the control unit and of the AM are combined in one controller board
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
1 SIO and 1 DIO can be installed
in the TFID analyzer (CM I/O)
SIO and DIO can be configured
for all channels combined with the TFID analyzer
NGA 2000
5.2
p. 5-85
1 - 1
The following illustrations shall make plain the connection between the hardware configu­ration and the software setup of the modules:
NGA 2000 System via Platform
(see 5.1.18)
(see 5.1.18)
Local
I/O's
SIO DIO
SIO DIO
(additional
manuals)
(additional
manuals)
(additional
manuals)
NGA 2000 System via TFID Analyzer
Analyzer
Modules (AM's)
TFID
MLT
CLD
PMD
FID
Control
Module (CM)
Platform
I/O
Modules
1 SIO
4 DIO's
max.
Other I/O's
(see 5.2.1)
(see 5.2.2)
(see 5.1.18)
(see 5.1.18)
1 - 2
Local
I/O's
SIO DIO
SIO DIO
(additional
manuals)
(additional
manuals)
(additional
manuals)
Analyzer
Modules (AM's)
TFID
MLT
CLD
PMD
FID
NGA 2000
Control
Module (CM)
TFID Analyzer
System I/O
Modules
1 SIO
1 DIO
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
(see 5.2.1)
(see 5.2.2)
NGA 2000 System via MLT Analyzer
1 Introduction
(see 5.1.18)
(see 5.1.18)
Local
I/O's
SIO DIO
SIO DIO
(additional
manuals)
(additional
manuals)
(additional
manuals)
Analyzer
Modules (AM's)
MLT
TFID
CLD
PMD
FID
Control
Module (CM)
MLT Analyzer
System I/O
Modules
1 SIO
1 DIO
(see 5.2.1)
(see 5.2.2)
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
1 - 3
1 - 4
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
enu lines ending with three points (...) are followed by submenus with further functions and set-ups.
All set-ups in the menu "Basic controls" are valid for the
At each menu point of the "Expert configuration" and "System configuration" you can find which functionality will be set up:
nalyzer
odule), CM (
odule), I/O (
utput Module).
-
-
-
-
Section 3.8
Display controls...
front panel LCD
- Display controls of the
- Front panel control
(number of digits)
(for the
single component display)
- Auxiliary lines
2 Structure of Menus
Main Menu
Analyzer and I/O-
and diagnostics...
System configuration
module expert
configuration...
(AM, CM, I/O's: SIO, DIO
- System calibration... (CM)
- Analyzer module
and other I/O's)
- Diagnostic menus...
controls... (AM)
(I/O's: SIO, DIO)
- I/O module controls...
(Lock of each level)
(AM, CM, I/O's: SIO, DIO
- Date and time... (CM)
- Security codes...
(AM)
- Analyzer module setup...
- I/O module setup...
and other I/O's)
- Network module binding...
(Other micro processor
- System reset... (CM)
- System modules... (CM)
- System tag (CM)
controlled I/O's)
analyzer module (AM).
Input-/O
Control M
M
Analyzer module
calibration & basic
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
controls... (AM)
Section 4 Section 5 Section 6 Section 7
Calibration procedure
status...
Zero calibration of all
ranges
Span calibration of the
NGA 2000
current range
Current gas valve setting:
- Zero gas flow
- Span gas flow
- Sample gas flow
- Test gas flow
- Closing all valves
Notes:
* M
AM (A
*
*
2 - 1
2 - 2
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98

3 Startup and Operation, General Notes and Main Menu

Check before starting:
All pneumatic supplies and mains, especially the hydrogen mains have to be tight !All gases used must have the right input pressure !All gas supply valves have to be closed !All electrical mains have to be joint and their voltage must operate correctly !The exhaust gas output must be without pressure
(atmosphere pressure or up to 300 mbar less than atmosphere pressure) !
3.1 Starting and Initializing (Look at the operation manual, too !)
1. Switching on
Join your Thermo-FID (TFID) with the supply voltage.Switch on.The following display will appear:
(C) 1998 FISHER-ROSEMOUNT Analytical
NGA-2000 Control Module/Version 3.2.1
Initializing Network
Initializing network interface
LCDReset Abort
F1 F2 F3 F4 F5
After switching on the TFID analyzer or analyzer module (in a platform or part of a NGA network), the initialization procedure will be performed. A self control of the analyzer modules or the analyzer is running. You can see a sequence of several displays. They show the status of initialization, revision notes of the TFID software and the tag. If you press the F1 key during the initializing, you will reset the LCD brightness and contrast to factory settings (see also section 7). Pressing the F3 key will abort the network initializing. Then you will have no connection to any analyzer module. Only the menus of the platform will be available. At the end of the initializing procedure you can see the single component display of channel one (see illustration on next page). It is the origin to all the other channels, menus and submenus.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
3 - 1
The instructions of the basic controls (chapter four) are all beginning with the single com­ponent display. The actual display might differ from the shown one because the customer can configure it according to his requirements (see sections 5.1.8 p. 5-49 and section 7).
TFID-R1
5.40 ppm Propane
0.00 50.00
Failures: Yes Maintenance-Requests: Yes Function-Control: Yes Operation: WarmUp
Display Status... Main... Channel Calib...
Range: 1
F1 F2 F3 F4 F5
2. Input of the correct span gas concentration
Put in the span gas concentration (Vol.-% or ppm) of the span gas certificate, the number of carbon atoms, the name of the span gas and the response factor! The correctness of these values is very important for the automatic calibration!
Change to the menu "Span Gas Definition":
Press the F3 key: The "Main Menu" will appear.
Change to the line "Analyzer and I/O-module expert configuration..." with the ↓↓ -key
Press the 88 -key: The menu "Expert Module Configuration" will appear.
Change to the line "Analyzer module set up..."
Press the 88 -key three times: You will reach the menu "Span Gas Definition":
3 - 2
TFID-R1
-- Span Gas Definition (1/2) --
Actual span gas concentration: 50.0 ppm
Spangas range -1: 50.00 ppm Spangas range -2: 250.0 ppm Spangas range -3: 2500 ppm Spangas range -4: 10000 ppm
Span gas unit: ppm "ppm" --> "mg/Nm3" conversion factor: 1.000
Concentration in span gas units: 37.50 ppm
Measure Back...
37.50 ppm
More...
F1 F2 F3 F4 F5
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
3 Startup and Operation, General Notes and Main Menu
You have to put in the span gas concentration of the span gas certificate in the line "Actual span gas concentration" and in all "Spangas range" lines !
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the 88 -key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the 88 -key or
cancel and go back to the last value with the F2 key.
Change to the second menu page with the F5 key (More...):
TFID-R1
-- Span Gas Definition (2/2) --
Span gas carbon atoms: 3
Span gas name: Propane Response factor: 1.000
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
Set up the parameters in the same way like described above !
Lines of variables "Span gas carbon atoms" and "Response factor": The TFID originally measures the carbon atoms of a gas. So, for a correct conversion into
"ppm" and other units, it is necessary to determine the number of carbon atoms of the span gas. Because the raw measurement also differs with different kinds of hydrocarbons, you have to put in the response factors of the analyzer for the type of hydrocarbons.
Note: You will find the table of response factors in the Thermo-FID operation manual (No.: 90 003
806).
Line of variables "Span gas name": In this line you can put in the real name of the span gas you want to use, e.g. "Propane".
Changing back to the single component display:
Press the F1 key.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
3 - 3
3. Automatic internal calibration
Open the supply gas valves!Set up the supply gas pressure values (see operation manual) !
During the warm-up procedure the failures and maintenance requests will disappear:
TFID-R1
6.20 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No
Function-Control: Yes Operation: WarmUp
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
If all temperatures and pressures are correct, the calculator will start an automatic calibration. The single component display will be:
TFID-R1
300.0
0.00 50.00
Failures: No Maintenance-Requests: No Function-Control: Yes
Operation: Calibration
Range: 1 F.S.
ppm Propane
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Zero and span gas concentrations will be displayed and the operation line shows "Calibration".
After successful calibration the system will be ready for measurement. The valve will be switched to "sample gas". In the line operation of the single component display you will find the message "Ready" and all reports of events will be "No":
3 - 4
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
3 Startup and Operation, General Notes and Main Menu
TFID-R1
2.30 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No
Function-Control: No Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
4. Start manual calibration
Before starting the measurement you should calibrate with zero and span gas to improve the accuracy of measurements.
How to calibrate:
The zero and span gas calibration are described detailed in the sections 4.5 p. 4-41
and 4.6 p. 4-45.
Further informations about calibration you will find in the expert configuration in section
5.1.1 p. 5-5 to 20.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
3 - 5

3.2 Display and Function

The LCD screen shows all measurement values of the analyzer and all customer instruc­tions. You can operate with five function keys, four arrow keys (cursors) and the enter key. The function of each key depends on:
the type of analyzer/analyzer module usedthe optional auxiliary modules (e.g. I/O boards) usedthe individual menu displayed
In case of power failure all customer specific module parameters are saved by a battery­powered buffer.

3.3 "TAG" and Operating Keys

At the top left of each menu page you will find the tag of the current channel.
Typical tags:
MLT/CH1/R1: MLT Analyzer or Analyzer Module / CHannel 1 / Range 1TFID-R1: Thermo FID Analyzer or Analyzer Module - Range 1
In this manual you will find normally "TAG" as general name. But in the specific TFID menu pages you will find "TFID".
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Function Keys:
keys without defined functionsThe current function depends on
the menu selected
The softkey legend is shown on
the display above the key
Range: 1
0.0 100.0
Cursor keys:
↑↑ -key / ↓↓ -key:
Line up / line down
within the same menu
Alteration of numbers,
variables or digits
-key / -key:
Moving back/forwards
between the pages of a menu
Selection of digits
Enter Key:
To confirm a previously entered value
(variable)
To start a selected function
(Alternative: -key)
To go into a menu (via menu line)
3 - 6
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
3 Startup and Operation, General Notes and Main Menu

3.4 Lines and Softkey Functionality

Lines can be selected by the ↓↓ -key or the ↑↑ -key. The selected line is displayed white on black. You have four different types of lines in the menu:
Menu line... / Menu Softkey...
Line/Softkey lettering ending with three dots.You will go to a submenu/further menu by pressing the softkey resp. by pressing the
88 -key or the -key in the menu line.
Function line / Function Softkey !
Line/Softkey lettering ending with an exclamation-mark.You will start a function (e.g. calibration) by pressing the softkey resp. by pressing the
88 -key or the -key in the function line.
Line of variables:
Line ending with a colon.Display of module parameters (variables).Some parameters can be changed (e.g. begin of range), some parameters display only
a status (e.g. temperature) and cannot be changed. These variables will be displayed below a line within the menu.
Text line
Line without any punctuation marks.Only display of informations.
The following illustrations shall make plain the functionalities of lines and softkeys des­cribed above:
Menu Line
Function Lines
Lines of Variables
(These here cannot
be changed)
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
F1 F2 F3 F4 F5
37.50 ppm
Valves...
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
3 - 7
Text Line
TAG
-- Calibration Procedure Status --
Procedure status: Ready Remaining procedure time: 0 s Current/expected gas flow: Samplegas Concentration in span gas units: 37.50 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Fri 05-29-1998 13:32:06 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
F1 F2 F3 F4 F5
37.50 ppm
Function Softkey
Menu Softkeys
3.5 Important Functions of the Softkeys Display
Change from the single component display to the multi component display.F1 in the single component display.
Measure
Change from all menus and submenus to the single component display of the channel
selected.
F1.
Status (see section 4.1 p. 4-3!)
Change to the menu "Analyzer Channel Status":
Display of the most important parameters and information about the status of the current channel or module.
If available: F2.
Main (see section 3.8 p. 3-11!)
Change from the single component display to the main menu.F3 in the single component display.
Channel
Scrolling through the channels in the same menu. In the main menu and the single
component display you can move among all channels of the connected analyzers and analyzer modules. In the submenus you can only move among the channels of the current analyzer or analyzer modules.
If available: F3 (F4 in the single component display).
3 - 8
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
3 Startup and Operation, General Notes and Main Menu
Lock
Lock of all three operation levels from the main menu, if a security code is activated in
the system configuration level (see section 6.3 p. 6-7).
F4 in the main menu.
Calib (see section 4.4 p. 4-39 and 5.1.1 p. 5-15!)
Change from the single component display to the menu "Analyzer module calibration".F5 in the single component display.
MFG Data (see section 3.8 p. 3-11/12!)
Change from the main menu to the menu "Module Manufacturing Data":
Further submenus are available with informations about the control module and analyzer module data, such as address of the manufacturer, serial number of the modules or the software and hardware revisions.
F5 in the main menu.
Back
Moving back to the last menu page selected (Alternative: -key) or
reset of a changed but not confirmed parameter to the former value.
If available: F4 for moving back, F2 for reset.
More
Changing to a further menu page.If available: F5.

3.6 Entering/Changing of Variables

88 -key
If you have already selected a line of variables (displayed white on black) and press the
88 -key, only the parameter will be selected and can be changed.
If you press the 88 -key again, the new value will be confirmed.
↑↑ -key / ↓↓ -key
Function depends on the variable selected: - Changing the parameter values
- Scrolling among variables selected
- Changing of digits or characters
Increasing or decreasing of numbers.
-key / -key
Selection of digits within a number.For some variables you can change the quantity of digits or characters.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
3 - 9

3.7 Starting a function

If you press the 88 -key or the → -key in a function line, you will be asked to confirm the order in the following way:
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
F1 F2 F3 F4 F5
37.50 ppm
If you will press the F2 key, the function will start immediately.If you will press the F4 key, you will go back to the last menu page.
Note: If you don't want to be asked before the start of each function, you can configure it in the menu "Measurement Display Configuration" in the expert configuration level (see section 5.1.8 p. 5-49). Select "No" in the line "Display confirmation menus". Then, each function will start directly after the order is done and no confirmation will be required.
3 - 10
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
3 Startup and Operation, General Notes and Main Menu
3.8 Main Menu If you press the F3 key (Main...) or the → -key in any single component display, you
will change to the "Main Menu". From there you can change to all operating levels of your TFID analyzer or analyzer module to set up and control the parameters of measurement, calibration and data transfer ! Via the F5 key (MFG Data) you can change to several submenus, where you will find a lot of important data about the control module (TFID analyzer or platform) and the analyzer module, such as service address or serial number !
Ramifications from the Main Menu:
TAG
-- Main Menu --
Analyzer module calibration & basic controls...
Analyzer and I/O-module expert configuration... System configuration and diagnostics... Display controls... Time & Date: 14:01:45 25 May 1998
System tag: Fisher-Rosemount
Measure Status... Channel Lock... MFG Data
F1 F2 F3 F4 F5
F1
F2
F3
Changing to the single component display of the current channel
Changing to the menu "Analyzer Channel Status" of the current channel
Scrolling through all channels of the connected analyzers and analyzer modules
95.00 ppm
See chapter 4 ! See chapter 5 ! See chapter 6 ! See chapter 7 !
To setup see 6.2 p. 6-6 Factory Setting
Section 3.1 p. 3-1/2 !
+
Section 4.1 p. 4-3 !
+
See channel tag !
+
F4
F5
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
Lock of any operating level by security code
Changing to the menu "Module Manufacturing Data"
NGA 2000
+
Section 6.3 p. 6-7 !
See next pages !
+
3 - 11
Ramifications from the menu "Module Manufacturing Data":
1. Control module data:
TAG
Control module data...
Analyzer module data...
Measure <<< Back... >>>
-- Module Manufacturing Data --
F1 F2 F3 F4 F5
TAG
(C) Copyright Fisher-Rosemount Analytical Inc., 1998
Manufactured by:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, CA 92807-1802 /USA
Tel: (714) 986-7600
FAX: (714) 577-8739
Measure Back...
F1 F2 F3 F4 F5
2. Analyzer module data:
95.00 ppm
95.00 ppm
TAG
(C) Copyright Fisher-Rosemount GmbH & Co, 1998
Manufactured by:
Fisher-Rosemount GmbH & Co
Industriestrasse 1
D-63594 Hasselroth / Germany
Tel. (+49) 6055 884-0
FAX. (+49) 6055 884-209
Measure Or... Back... More...
F1 F2 F4 F5
TAG
-- Control Module Version Information --
Serial number: CM 4711 Manufacturing date: 14.05.1998
Hardware revision: ACU02 R: 3.2.2 Software revision: 3.2.1/P001 Revision date: May 29 1998 Revision time: 12:35:00
Phrase dictionary version: P001/1 Language: English
Measure Back...
F3
95.00 ppm
95.00 ppm
F1 F2 F3 F4 F5
TAG
Control module data...
Analyzer module data...
Measure <<< Back... >>>
-- Module Manufacturing Data --
95.00 ppm
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Version Information --
Serial number: AM 4712 Manufacturing date: 14.05.1998
Hardware revision: ACU02 R: 3.2.2 Software revision: 3.2.1 / P001 Revision date: May 29 1998 Revision time: 12:41:18
Measure Back...
95.00 ppm
More... Measure Back...
F1 F2 F3 F4 F5
TAG
(C) Copyright Fisher-Rosemount GmbH & Co, 1998
Manufactured by:
Fisher-Rosemount GmbH & Co
Industriestrasse 1
D-63594 Hasselroth / Germany
Tel. (+49) 6055 884-0
FAX. (+49) 6055 884-209
Measure Back... More...
95.00 ppm
F1 F2 F3 F4 F5
TAG
Measurement system: PSV-System RAM-memory: 242748 Bytes Local SIO module installed: Enabled Serial interface adaptor: RS-232 Heater installed: No Local DIO module installed: 2 Sensor system revision:
-- Hardware Configuration --
Channel
95.00 ppm
F1 F2 F3 F4 F5
3 - 12
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4 Basic Controls and Analyzer Module Calibration
In the chapter "Basic Controls and Analyzer Module Calibration" the most important mea­surement and calibration functions of your TFID analyzer or analyzer module are described.
All steps are figured with detailed illustrations and operation instructions. In the left column you can see display and keyboard of the NGA front panel. The keys you have to press are illustrated in black. In the right column you can read the instructions and notes. All instructions will begin with any single component display and will end with the correspond­ing single component display after the setups are done. So you can easily compare the actual display of the analyzer or analyzer module with the illustrations of the manual.
Example: You want to change from the single component display of the TFID (propane)
to the single component display of the MLT-channel 2 (CO).
Picture one shows the starting situation: single component display of propane.
Picture two shows the result you get if you press the F4 key (Channel):
single component display of CO.
Left column: Right column:
Display and keyboard Instructions and notes
TFID-R1
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
MLT25/CH2/R2
95.00 ppm CO
0 250
Temperature: 25.0 C Maintenance-Requests: No Any_Alarms: No Operation: Ready
Range: 1
0.0 100.0
Range: 2
0.0 100.0
Change to the single
component display of another channel
F4
Press
Example: Changing from propane (TFID) to CO (MLT-channel 2)
Next instruction
or step
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 1
4 - 2
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
37.50 ppm Propane
0.00 50.00
Range: 1
4.1 Analyzer Channel Status
Change to the
menu "Analyzer Channel Status"
F2
Press
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
In the menu "Analyzer Channel Status" you can find status informations of the current channel. Via the menu lines "Status details..." and "Operational settings..." you can change to further submenus. (see 4.1.1 p. 4-5...32 and
4.1.2 p. 4-33/34)
Change to the
single component display
F1
Press
Note: Via the F2 key you can change to submenus containing the primary and secondary raw measurements.
F1 F2 F3 F4 F5
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back in the single
component display
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 3
4 - 4
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane

Failures

Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the "List
of Possible Failures (page 1 of 4)"
Press
or
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 5
TAG
-- List of Possible Failures (1/4) --
One or more failures: No Configuration replaced by factory setting: No Hydrogen pressure too low: No No source voltage: No
Sample gas pump #1 fail: No Sample gas pump #2 fail: No Detector temperature too low: No Flame is off: No Ignition canceled: No
Measure Back... More...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- List of Possible Failures (2/4) --
External Input: No Probe temperature too low: No Probe temperature too high: No Flame temperature sensor fail: No Sample under pressure too high: No Capillary #1: No Control of sample gas pump #1 fail: No Sample gas pressure too high: No
37.50 ppm
Change to the
second menu page
F5
Press
Change to the
third menu page
F5
Press
Control of sample gas pump #2 fail: No
Measure Back... More...
F1 F2 F3 F4 F5
TAG
-- List of Possible Failures (3/4) --
Sample gas (Filter/supply/under pressure): No Capillary #3 or H2 supply: No Detector signal communication failed: No Power supply less than 23V: No Air pressure less than 2.9 bar: No Igniter broken: No Igniter short-circuit: No Detector temperature sensor fail: No Sensor temperature sensor fail: No Ambient temperature too high: No
Measure Back... More...
37.50 ppm
F1 F2 F3 F4 F5
Change to the
last menu page
F5
Press
4 - 6
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
-- List of Possible Failures (4/4) --
Sample gas throttle closed: No Burner gas throttle closed: No Under pressure pump fail: No Sample gas pump fail: No Burner gas pump fail: No Sample gas control valve fail: No Burner gas control valve fail: No
Measure Back... (1/4)...
F1 F2 F3 F4 F5
37.50 ppm
Failures
Change to the
single component display
F1
Press
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back in the single
component display of the current channel
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 7
4 - 8
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane

Maintenance Requests

Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Maintenance requests..."
Press or as often as necessary to get the menu line "Maintenance requests..." white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 9
TAG
-- Status Details --
Failures...
Maintenance requests...
Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
37.50 ppm
Change to the
"List of Possible Maintenance Requests (page 1 of 2)"
Press
Measure Back...
F1 F2 F3 F4 F5
TAG
-- List of Possible Maintenance Requests (1/2) --
One or more maintenance requests: No Too many hours of operation: No Zero calibration - deviation too high: No Span calibration - deviation too high: No Span gas pressure less than 1.5bar: No Measurement too noisy for calibration: No Sample gas (Filter/supply/under pressure): No Detector temperature too high: No Calibration-span gas pressure less than 1.5bar: No Flame optimization canceled: No
Measure Back...
37.50 ppm
More...
F1 F2 F3 F4 F5
or
Change to the
second menu page
F5
Press
TAG
-- List of Possible Maintenance Requests (2/2) --
External input: No Probe temperature too low: No Probe temperature too high: No Catalysator temperature too high: No Catalysator temperature too low: No Calibration canceled: No Basic calibration required: No Catalysator temperature fail: No Probe temperature sensor fail: No Air pressure < 2.9 bar: No
Measure Back...
F1 F2 F3 F4 F5
4 - 10
37.50 ppm
NGA 2000
Change to the
single component display
F1
Press
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane
Maintenance Requests
Back in the single
component display of the current channel
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 11
4 - 12
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane

Function Controls

Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Function controls..."
Press or as often as necessary to get the menu line "Function controls..." white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 13
TAG
-- Status Details --
Failures... Maintenance requests...
Function controls...
Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- List of Possible Function Controls (1/1) --
One or more function controls: No Request from external input: No Concentration raw signal simulation: No Sample gas valve closed: No No sample gas flowing: No State - Warmup: No State - Pause: No State - Standby: No Secondary measurement simulation: No External Input: No
Measure Back...
37.50 ppm
Change to the
"List of Possible Function Controls"
Press
or
Change to the
single component display
F1
Press
F1 F2 F3 F4 F5
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back in the single
component display of the current channel
4 - 14
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane

Measurements/Alarms

Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Measurements/ Alarms..."
Press or as often as necessary to get the menu line "Measurements/Alarms..." white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 15
TAG
-- Status Details --
Failures... Maintenance requests... Function controls...
Measurements/Alarms...
Events... Acknowledge and clear failures !
Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Measurements/Alarms --
Concentration...
Concentration average... Flow... Pressure... Temperature... Calculator-1... Calculator-2... Calculator-3... Calculator-4...
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the
"Measurements/
Press
or
Change to the single
component display
F1
Press
Options:
With the ↑↑ or the ↓↓ -key
you can change to each line in the menu.
Press the 88 -key in the line
selected to change to the corresponding sub-menu: There you will find the
status of the four alarms Note: If the hardware of the menu point selected is not available, a corresponding message will appear!
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
4 - 16
Range: 1
0.0 100.0
NGA 2000
Back in the single
component display of the current channel
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane

Events

Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Events..."
Press or as often as necessary to get the menu line "Events" white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 17
TAG
-- Status Details --
Failures... Maintenance requests... Function controls... Measurements/Alarms...
Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Events (1/2) --
Failure: Maintenance request: Function control/Service: Last configuration lost: Concentration alarm: Flow alarm: Linearization underflow: Linearization overflow: Calculator alarms active: Local analog output out of range:
Measure Status... Channel Back... More...
37.50 ppm
Change to the
submenu "Analyzer Module Events (page 1 of 2)"
Press
or
Change to the
second menu page
F5
Press
Note: If any event has happened,
you will find a report of date and time in the corresponding line of this menu.
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Events (2/2) --
Zero calibration started: Zero calibration finished: Span calibration started: Span calibration finished: Concentration average alarm: Temperature alarm: Pressure alarm: New linearization calculated: Sensor calibration:
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the
single component display
F1
Press
4 - 18
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane
Events
Back to the single
component display of the current channel
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 19
4 - 20
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane
Acknowledge and Clear Failures
Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: Yes Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Notes: If you have solved the reasons for the failures reported, you should start this function. The menu "List of Possible Failures" will be ready for new reports !
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Acknowledge and clear failures !"
Press or as often as necessary to get the menu line "Acknowledge and clear failures !" white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 21
TAG
-- Status Details --
Failures... Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures !
Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
Start the function
Press
or
Note: Starting this function here is only possible, if it is enabled in the menu "Acknowledge­ment of Status Reports" (see 5.1.9 p. 5-52) !
Confirm the order
F2
Press to start the function immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Status Details".
F1 F2 F3 F4 F5
TAG
- S U C C E S S -
- The selected function has been started/executed ­(Wait a moment...)
37.50 ppm
F1 F2 F3 F4 F5
Confirmation
message of the function start appears
This message will be dis­played after the function has been started. Then the display will jump automatically to the menu "Status Details".
4 - 22
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
Acknowledge and Clear Failures
Change to the
single component display of the channel selected
F1
Press
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back to the single
component display of the current channel after clearing of failures
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 23
4 - 24
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane
Acknowledge and Clear Maintenance Requests
Change to the menu
"Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: Yes Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Notes: If you have solved the reasons for the maintenance requests reported, you should start this function. The menu "List of Possible Maintenance Requests" will be ready for new reports !
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Acknowledge and clear maintenance requests !"
Press or as often as necessary to get the menu line "Acknowledge and clear maintenance requests !" white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 25
TAG
-- Status Details --
Failures... Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures !
Acknowledge and clear maintenance requests !
Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
Start the function
Press
or
Note: Starting this function here is only possible, if it is enabled in the menu "Acknowledge­ment of Status Reports" (see 5.1.9 p. 5-52) !
Confirm the order
F2
Press to start the function immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Status Details".
F1 F2 F3 F4 F5
TAG
- S U C C E S S -
- The selected function has been started/executed ­(Wait a moment...)
37.50 ppm
F1 F2 F3 F4 F5
Confirmation
message of the function start appears
This message will be dis­played after the function has been started. Then the display will jump automatically to the menu "Status Details".
4 - 26
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
Acknowledge and Clear Maintenance Requests
Change to the
single component display of the channel selected
F1
Press
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back to the single
component display of the current channel after clearing of maintenance requests
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 27
4 - 28
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm Propane
Acknowledge and Clear Function Controls
Change to the menu
"Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Notes: If you have solved the reasons for the function controls reported, you should start this function. The menu "List of Possible Function Controls" will be ready for new reports !
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Acknowledge and clear function controls !"
Press or as often as necessary to get the menu line "Acknowledge and clear function controls!" white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 29
TAG
-- Status Details --
Failures... Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests !
Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
Start the function
Press
or
Note: Starting this function here is only possible, if it is enabled in the menu "Acknowledge­ment of Status Reports" (see 5.1.9 p. 5-52) !
Confirm the order
F2
Press to start the function immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Status Details".
F1 F2 F3 F4 F5
TAG
- S U C C E S S -
- The selected function has been started/executed ­(Wait a moment...)
37.50 ppm
F1 F2 F3 F4 F5
Confirmation
message of the function start appears
This message will be dis­played after the function has been started. Then the display will jump automatically to the menu "Status Details".
4 - 30
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
4.1.1 Analyzer Channel Status - Status Details
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
Acknowledge and Clear Function Controls
Change to the
single component display of the channel selected
F1
Press
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back to the single
component display of the current channel after clearing of function controls
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 31
4 - 32
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98

4.1.2 Analyzer Channel Status - Operational Settings

TAG
37.50 ppm Propane
Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the line
"Operational settings..."
Press once to get the line "Operational settings..." white on black.
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings...
Hours of operations: 164 Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F1 F2 F3 F4 F5
Change to the
menu "Analyzer Operation Settings"
Press
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 33
TAG
-- Analyzer Operation Settings --
Remote control: Enabled Range change control: Manual Range: 1 Range upper limit: 50.00 ppm Span gas concentration: 50.00 ppm t90-time: 2.00 s Hours of operation: 164 Last re-start occured: Fri 05-29-1998 15:32:00 Actual zero gas concentration: 0.00 ppm
Auto-start procedures...
Measure Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Auto-Start Procedures --
Position in auto-start list: 1
Channel tag: ­Procedure type: ­Interval mode: Never Start time: ­Start date: -
Measure Back...
37.50 ppm
Change to the
submenu "Auto­Start Procedures"
Press
Change to the
single component display
Note: In the menu "Auto-Start Procedures" you can control the status of the three kinds of time controlled calibrations. (see 5.1.1 p. 5-13/14)
F1 F2 F3 F4 F5
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back to the single
component display of the current channel
4 - 34
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TFID-R1
37.50 ppm Propane
4.2 Single Component Display
Change of Channel
Change to the single
component display of another channel
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
MLT25/CH2/R2
Range: 1
0.0 100.0
95.00 ppm CO
0 250
Temperature: 25.0 C Maintenance-Requests: No Any_Alarms: No Operation: Ready
Range: 2
0.0 100.0
F4
Press
Example: Changing from propane (TFID) to CO (MLT-channel 2)
Note: A change of channel will only be possible, if the TFID is combined with other modules or analyzers.
Change back to the
single component display of the starting channel
F4
Press
as often as necessary until the display of the channel you want will appear
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
TFID-R1
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Note: You can reach any existing channel by pressing the F4 key for several times.
Single component
display of the starting channel appears
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 35
4 - 36
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TFID-R1
37.50 ppm Propane
4.3 Multi Component Display
Change of Channel
Change to the

multi component display

0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
37.50
95.00
333.0
150.0
20.00
Select Status... Tags Off
TFID
ppm Propane
MLT25/CH1
ppm CO
MLT25/CH2
ppm SO2
MLT25/CH3
ppm NO
MLT25/CH4
%O2
[1]
50.000.00
0.00 250.00
0.00 500.00
0.00 150.00
0.00 100.00
[2]
[2]
[2] F.S.
[2]
LCDReset
F1 F2 F3 F4 F5
F1
Press
Notes:
A change of channel will
only be possible, if the TFID is combined with other modules or analy­zers.
You can change to the
multi component display from each single compo­nent display.
Enable the "selecting
symbol": >
F1
Press
Notes:
In each bargraph you will
find the begin and end of range of the corresponding channel. (F.S. = full scale)
The number in parentheses
shows the number of the selected range.
Option: With the F3 key you can fade out or in the tags.
or
37.50
95.00
333.0
150.0
20.00
Select Status... Tags Off
TFID
ppm Propane
MLT25/CH1
ppm CO
MLT25/CH2
ppm SO2
MLT25/CH3
ppm NO
MLT25/CH4
%O2
0.00 250.00
0.00 500.00
0.00 150.00
0.00 100.00
F1 F2 F3 F4 F5
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
[1]
[2]
[2]
[2] F.S.
[2]
50.000.00
LCDReset
NGA 2000
Select any channel
Press or as often as necessary to
put the > - mark into the line you want to select.
Example: Change from propane (TFID) to CO (MLT-channel 1)
4 - 37
37.50
95.00
333.0
150.0
20.00
Select Status... Tags Off
TFID
ppm Propane
MLT25/CH1
ppm CO
MLT25/CH2
ppm SO2
MLT25/CH3
ppm NO
MLT25/CH4
%O2
[1]
0.00 250.00
0.00 500.00
0.00 150.00
0.00 100.00
[2]
[2]
[2] F.S.
[2]
LCDReset
F1 F2 F3 F4 F5
MLT25/CH1/R2
95.00 ppm CO
Change to the single
component display of
50.000.00
the channel selected
F1
Press
Note: With the F5 key you can enable the LCD-brightness and contrast of the factory settings (see also section 7).
Single component
display of the channel selected appears
0 250
Temperature: 25.0 C Maintenance-Requests: No Any_Alarms: No Operation: Ready
Display Status... Main... Channel Calib...
Range: 2
0.0 100.0
F1 F2 F3 F4 F5
4 - 38
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
37.50 ppm Propane
4.4 Calibration Procedure Status
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
37.50 ppm
Valves...
F1 F2 F3 F4 F5
F5
Press
Change to the
menu "Calibration Procedure Status"
Press
Options: Via the F5 key you can change to a submenu, where you can close all valves or set up each valve separately with:
zero gas or
span gas or
sample gas or
test gas.
(See sections 4.7 to 4.11, p. 4-49...)
TAG
-- Calibration Procedure Status --
Procedure status: Ready Remaining procedure time: 0 s Current/expected gas flow: Samplegas Concentration in span gas units: 37.50 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Fri 05-29-1998 13:32:06 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
F1 F2 F3 F4 F5
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
37.50 ppm
NGA 2000
Change to the menu
page "Calibration Deviations"
F5
Press
4 - 39
TAG
-- Calibration Deviations --
Deviation from zero: -0.05 ppm Sum of zero deviations: -0.41 ppm
Deviation from span: -0.14 ppm Sum of span deviations: -0.57 ppm
Measure Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
37.50 ppm Propane
Change to the single
component display of the current channel
F1
Press
Options:
With the F3 key you can
change to further available channels to check their "Calibration Deviations".
With the F4 key you can
go back to the menu "Calibration Procedure Status".
Back to the single
component display
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
4 - 40
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
37.50 ppm Propane
0.00 50.00
Range: 1
4.5 Zero Calibration
Change to the
menu "Analyzer Module Calibration"
F5
Press
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
37.50 ppm
Valves...
Caution:
Before starting zero calibration, realize that zero gas is available ! (See also section 5.1.1, p. 5-5... !)
Note: The zeroing of all measure­ment ranges of the same channel is running simulta­neously.
Change to the
line "Start zero calibration procedure !"
Press once to get the line "Start zero calibration procedure !" white on black.
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure !
Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
37.50 ppm
Valves...
F1 F2 F3 F4 F5
Start the zero
calibration
Press
or
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 41
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Calibration Procedure Status --
Procedure status: Purging1-Wait Remaining procedure time: 8 s Current/expected gas flow: Zerogas Concentration in span gas units: 37.50 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Fri 05-29-1998 13:32:06 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
37.50 ppm
F1 F2 F3 F4 F5
Confirm the order for
zero calibration
F2
Press
to start
the zeroing immediately.
Option: Press the F4 key if you want to cancel the order.
Notes:
The display of this message
depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
The 3 following illustrations
show the displays you can see after starting the zeroing procedure.
Zeroing:
1. Purging-Wait
Notes:
The procedure time de-
pends on the parameters entered for purge time (see section 5.1.1 p. 5-10).
The purge time must be
long enough to get a stable signal before calibration.
You can cancel the run-
ning calibration procedure at any time with the F2 key.
TAG
-- Calibration Procedure Status --
Procedure status: Zeroing-Wait Remaining procedure time: 97 s Current/expected gas flow: Zerogas Concentration in span gas units: 3.13 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Fri 05-29-1998 13:32:06 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
F1 F2 F3 F4 F5
4 - 42
3.13 ppm
NGA 2000
Zeroing:
2. Zeroing-Wait
Note: The procedure time de­pends on the parameters entered for stability time and averaging time. (see section 5.1.1 p. 5-10)
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
-- Calibration Procedure Status --
Procedure status: Ready Remaining procedure time: 0 s Current/expected gas flow: Samplegas Concentration in span gas units: 0.00 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Mon 06-08-1998 09:21:34 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
0.00 ppm
F1 F2 F3 F4 F5
4.5 Zero Calibration
Change to the
single component display of the current channel
F1
Press
Options:
F4 key: Changing back
to the menu "Analyzer Module Calibration".
F5 key: Changing to
the menu "Calibration Deviations".
TAG
0.00 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Single component
display appears after zeroing is done
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 43
4 - 44
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
37.50 ppm Propane
4.6 Span Calibration
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
37.50 ppm
Valves...
F1 F2 F3 F4 F5
F5
Press
Caution:
Before starting span calibration, realize that spangas with correct concentration is available ! (See also section 5.1.1, p.5-5... !)
Change to the line
"Start span calibration procedure !"
Press twice to get the line "Start span calibration procedure !" white on black.
Notes:
Normally, all measurement
ranges of the same chan­nel are calibrated simulta­neously.
To calibrate seperately you
have to change the para­meters (see 5.1.1 p. 5-10).
TAG
-- Analyzer Module Calibration --
Calibration procedure status... Start zero calibration procedure !
Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
F1 F2 F3 F4 F5
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
37.50 ppm
Valves...
NGA 2000
Start the

span calibration

Press
or
4 - 45
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Calibration Procedure Status --
Procedure status: Purging1-Wait Remaining procedure time: 4 s Current/expected gas flow: Spangas-1 Concentration in span gas units: 37.50 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Mon 06-08-1998 09:21:34 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
37.50 ppm
F1 F2 F3 F4 F5
Confirm the order for
span calibration
F2
Press
to start
the spanning immediately.
Option: Press the F4 key if you want to cancel the order.
Notes:
The display of this message
depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
The 3 following illustrations
show the displays you can see after starting the spanning procedure.
Spanning:
1. Purging-Wait
Notes:
The procedure time de-
pends on the parameters entered for purge time (see 5.1.1 p. 5-10).
The purge time must be
long enough to get a stable signal before calibration.
You can cancel the run-
ning calibration procedure at any time with the F2 key.
TAG
-- Calibration Procedure Status --
Procedure status: Spanning-Wait Remaining procedure time: 108 s Current/expected gas flow: Spangas-1 Concentration in span gas units: 43.57 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Mon 06-08-1998 09:21:34 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
F1 F2 F3 F4 F5
4 - 46
43.57
NGA 2000
Spanning:
2. Spanning-Wait
Note: The procedure time de­pends on the parameters entered for stability time and averaging time. (see 5.1.1 p.5-10)
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
-- Calibration Procedure Status --
Procedure status: Ready Remaining procedure time: 0 s Current/expected gas flow: Samplegas Concentration in span gas units: 50.00 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Mon 06-08-1998 09:21:34 Last span calibration was: Mon 06-08-1998 09:28:13 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
50.00 ppm
F1 F2 F3 F4 F5
4.6 Span Calibration
Change to the
single component display of the current channel
F1
Press
Options:
F4 key: Changing back
to the menu "Analyzer Module Calibration".
F5 key: Changing to
the menu "Calibration Deviations".
TAG
50.00 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1 F.S.
0.0 100.0
Single component
display appears after spanning is done
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 47
4 - 48
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
37.50 ppm Propane
4.7 Flow Zero Gas
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
37.50 ppm
Valves...
F5
Press
Caution:
Before starting zero gas flow, realize that zero gas is available ! (See also section 5.1.1, p.5-15 !)
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Start the zero
gas flow
Press
or
Starting this function will switch on the zero gas valve as current gas valve.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 49
TAG
-- Confirmation Required --
37.50 ppm
Confirm the order to
flow zero gas
F2
Press to start the zero gas flow immediately.
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
F1 F2 F3 F4 F5
TAG
- S U C C E S S -
- The selected function has been started/executed ­(Wait a moment...)
37.50 ppm
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Zerogas Operation status: Ready
Measure Status... Channel Back...
0.00 ppm
F1 F2 F3 F4 F5
Change to the single
component display of the current channel
Options:
Start the gasflow with
span gas, sample gas or test gas.
Close all valves.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
4 - 50
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
0.00 ppm Propane
4.7 Flow Zero Gas
Single component
display appears during the zero gas flow is running
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Note: The valve position "Zerogas" will last 1000 seconds. After that it will go back to "Samplegas" for a moment. Then it will return to "Zero­gas" automatically. This procedure will repeat every 1000 seconds as long as the "Flow zero gas" order is set.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 51
4 - 52
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
37.50 ppm Propane
4.8 Flow Span Gas
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
37.50 ppm
Valves...
F5
Press
Caution:
Before starting span gas flow, realize that spangas is available ! (See also section 5.1.1, p. 5-15 !)
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Flow span gas !"
Press once to get the line "Flow span gas !" white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 53
TAG
Flow zero gas !
Flow span gas !
Flow sample gas ! Flow test gas !
-- Set Gas Valves --
37.50 ppm
Start the span
gas flow
Press
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
or
Starting this function will switch on the span gas valve as current gas valve.
Confirm the order to
flow span gas
F2
Press to start the span gas flow immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
TAG
- S U C C E S S -
- The selected function has been started/executed -
(Wait a moment...)
F1 F2 F3 F4 F5
4 - 54
37.50 ppm
NGA 2000
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Spangas-1 Operation status: Ready
Measure Status... Channel Back...
50.00 ppm
F1 F2 F3 F4 F5
4.8 Flow Span Gas
Change to the single
component display of the current channel
Options:
Start the gasflow with
zero gas, sample gas or test gas.
Close all valves.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
TAG
50.00 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1 F.S.
0.0 100.0
Single component
display appears during the span gas flow is running
Note: The valve position "Spangas" will last 1000 seconds. After that it will go back to "Samplegas" for a moment. Then it will return to "Span­gas" automatically. This procedure will be repea­ted every 1000 seconds as long as the "Flow span gas" order is set.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 55
4 - 56
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
0.00 ppm Propane
4.9 Flow Sample Gas
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Valve position: Zerogas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
0.00 ppm
Valves...
F5
Press
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Zerogas Operation status: Ready
Measure Status... Channel Back...
0.00 ppm
F1 F2 F3 F4 F5
Change to the line
"Flow sample gas !"
Press or as often as necessary to get the line "Flow sample gas !" white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 57
TAG
Flow zero gas ! Flow span gas !
Flow sample gas !
Flow test gas !
-- Set Gas Valves --
0.00 ppm
Start the sample
gas flow
Press
Close all valves !
Valve position: Zerogas Operation status: Ready
Measure Status... Channel Back...
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
0.00 ppm
F1 F2 F3 F4 F5
or
Starting this function will switch on the sample gas valve as current gas valve.
Confirm the order to
flow sample gas
F2
Press to start the sample gas flow immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
TAG
- S U C C E S S -
- The selected function has been started/executed -
(Wait a moment...)
F1 F2 F3 F4 F5
4 - 58
0.00 ppm
NGA 2000
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
4.9 Flow Sample Gas
Change to the single
component display of the current channel
Options:
Start the gasflow with
zero gas, span gas or test gas.
Close all valves.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
TAG
37.50 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Single component
display appears during the sample gas flow is running
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 59
4 - 60
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
37.50 ppm Propane
4.10 Flow Test Gas
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
37.50 ppm
Valves...
F5
Press
Starting this function will
switch an external valve
and flow the TFID with test gas.
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Flow test gas !"
Press or as often as necessary to get the line "Flow test gas !" white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 61
TAG
Flow zero gas ! Flow span gas ! Flow sample gas !
Flow test gas !
-- Set Gas Valves --
37.50 ppm
Start the function to
flow test gas
Press
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
or
Confirm the order
to flow test gas
F2
Press to start the test gas flow immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
TAG
- S U C C E S S -
- The selected function has been started/executed -
(Wait a moment...)
F1 F2 F3 F4 F5
4 - 62
37.50 ppm
NGA 2000
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Testgas Operation status: Ready
Measure Status... Channel Back...
20.00 ppm
F1 F2 F3 F4 F5
4.10 Flow Test Gas
Change to the single
component display of the current channel
Options:
Start the gasflow with
zero gas, span gas or sample gas.
Close all valves.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
TAG
20.00 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Single component
display appears in the standby mode: All valves are closed
Note: After the valves are closed you can start again the measurement with the "Flow sample gas" order (see 4.9). The system will require an auto calibration.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 63
4 - 64
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
37.50 ppm Propane
4.11 Close all Valves
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Propane
Measure Status... Channel Back...
37.50 ppm
Valves...
F5
Press
Starting this function will
close the fuel gas valve
(flame out).
switch off the compressed
air or the burner pump (depending from hardware configuration).
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Close all valves !"
Press or as often as necessary to get the line "Close all valves !" white on black.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 65
TAG
Flow zero gas ! Flow span gas ! Flow sample gas ! Flow test gas !
-- Set Gas Valves --
37.50 ppm
Start the function to
close all valves
Press
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
or
Confirm the order
to close all valves
F2
Press to close all valves immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
TAG
- S U C C E S S -
- The selected function has been started/executed -
(Wait a moment...)
F1 F2 F3 F4 F5
4 - 66
37.50 ppm
NGA 2000
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: AllClosed Operation status: Ready
Measure Status... Channel Back...
0.00 ppm
F1 F2 F3 F4 F5
4.11 Close all Valves
Change to the single
component display of the current channel
Options:
Start the gasflow with
zero gas, span gas, sample gas or test gas.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
TAG
0.00 ppm Propane
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Single component
display appears in the standby mode: All valves are closed
Note: After the valves are closed you can start again the measurement with the "Flow sample gas" order (see 4.9). The system will require an auto calibration.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
4 - 67
4 - 68
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98

5 Analyzer and I/O-Module Expert Configuration

If you press the 88 -key or the → -key in the line "Analyzer and I/O-module expert configuration..." of the "Main Menu", you will change to the following display:
TAG
-- Expert Module Configuration --
Analyzer module controls...
I/O module controls... Analyzer module setup...
I/O module setup...
Measure Channel Back...
F1 F2 F3 F4 F5
37.50 ppm
From the menu "Expert Module Configuration" you can change to several submenus to set up parameters for the measurement and calibration of your analyzer and analyzer module. Besides you can set up the configuration for auxiliary modules. Which part of these menus are important for you depends on the configuration of your NGA 2000 system. In the following table you will find a short overview about the contents of the menus:
Menu Important Contents Section/Page
Analyzer module controls... See "Analyzer module setup" and note below! I/O module controls...
Configuration of the SIO and of the DIO's
installed in a platform or
5.2
p. 85 - 100
in a TFID analyzer (auxiliary I/O's)
Analyzer module setup...
Measurement set up and controlCalibration set up and controlConfiguration of the SIO and of the DIO's
5.1
p. 3 - 84
installed in the TFID analyzer module (Local SIO/DIO)
Programmable logic control (PLC)Programmable calculators
I/O module setup...
Set up for micro processor controlled
5.3 / p. 101
auxiliary I/O modules
Note: All submenus of "Analyzer module controls..." are the same like those of "Analyzer module
setup...", if you use a TFID analyzer or analyzer module (resp. MLT analyzer or analyzer module)! If your analyzer is not such a type, the submenus of "Analyzer module controls..." and "Analyzer module setup..." will look different, partially. Look at the corresponding software manuals.
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
5 - 1
Structure of chapter five:
At the begin of any chapter you will find the way to a certain submenu of the TFID software starting from the line "Analyzer and I/O-module expert configuration" in the "Main Menu". The way will be described with the software catchwords, that you have to enter one after another to reach the corresponding submenu. At the end of the catchword listing you will find the illustration of the LCD screen. After that you can read the set up instructions and explanations. Sometimes the illustration will be completed by sketches or further menu pictures.
Example: You will set up the calibration parameters for the calibration procedure of the
analyzer module.
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
Calibration procedure setup
↓↓
TAG
-- Calibration Procedure Setup --
Purge time: 10 s
Procedure times-out after: 120 s Analog output during calibration: Tracking Span ranges: Together
Valve position: Samplegas
Stability time: 30 s Averaging time: 5 s
Measure Channel Back...
F1 F2 F3 F4 F5
In the menu "Calibration Procedure Setup" you can set up the parameters for the cali­bration procedure of the zeroing and spanning ...
95.00 ppm
Further explanations and instructions will follow !
5 - 2
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
5.1 Analyzer Module Setup
Main Menu — Analyzer and I/O-module expert configuration
↓↓

Analyzer module setup

↓↓
TAG
-- Analyzer Module Setup and Controls (1/3) --
Calibration parameters...
Alarm parameters... Range parameters... Cross interference compensation... Linearization... Programmable logic control (PLC)... Programmable calculator... Measurement display configuration... Acknowledgement of status reports... Concentration measurement parameters...
Measure ManData Channel Back... More...
95.00 ppm
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Setup and Controls (2/3) --
Peak measurement...
Differential measurement... Gasflow setup... Pressure compensation... Flow measurement... Temperature measurement... Loading/saving configuration parameters... Inputs and outputs... Delay and average... Special functions...
Measure Channel Back...
95.00 ppm
More...
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Setup and Controls (3/3) --
AK-protocol communication...
Measure Back...
F1 F2 F3 F4 F5
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
95.00 ppm
5 - 3
From the menu "Analyzer Module Setup and Controls" you can change to all menus and submenus of the expert configuration level via the corresponding menu lines.
Changing to the menus and submenus:
Select the menu line you want with the ↓↓ -key or the ↑↑ -key.
If necessary: Change to a further menu page with the F5 key (More...)
Change to the corresponding menu/submenu with the 88 -key or the -key.
Menu "Load Factory Configuration":
If you press the F2 key (ManData) in the menu "Analyzer Module Setup and Controls (1/3)", you will change to the following menu:
TAG
-- Load Factory Configuration --
- BE CAREFUL with this function -
Replace current configuration with factory settings !
Measure Back...
F1 F2 F3 F4 F5
95.00 ppm
With the function of this menu you can delete the RAM data and load the factory settings from the Flash-EPROM.
Caution: This function is irreversible after starting (and confirmation) ! Your RAM settings will be deleted !
Note:
The same function is available in the menu "Load/Save Analyzer Module Configuration" ! (see section 5.1.17 p. 5-66)
Starting the function:
Press the 88 -key or the -key or
cancel and go back to the last menu page with the F4 key (Back...) or the ← -key.
If asked: Confirm with the F2 key (Yes) or
cancel and go back to the last menu page with the F4 key (Back...) or the -key.
Further Ramifications via softkeys:
F1
F3
5 - 4
Change to the single component display of the current channel !
Change to the other available channels of an analyzer module or of a TFID analyzer !
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
Chapter 4 !
+
See channel tag !
+
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓

Calibration parameters

↓↓
TAG
-- Calibration Parameters --
Span gases...
Tolerances... Calibration procedure setup... Time controlled calibration... Calibration... Advanced calibration methods... Zero gases...
Measure Channel Back...
F1 F2 F3 F4 F5
37.50 ppm
From the menu "Calibration Parameters" you can change to several submenus to set up the zero and span gas calibration parameters and to start the different calibration methods.
Change to the submenus:
Select the menu line you want with the ↓↓ -key or the ↑↑ -key.Change to the corresponding submenu with the 88 -key or the → -key.
Alternatives:
F1: Change to the single component display !F3: Change to the other available channels of a TFID analyzer
analyzer module !
F4: Go back to the last menu page !
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
5 - 5
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Span gases
Main Menu — Analyzer and I/O-module expert configuration
Analyzer module setup
Calibration parameters
Span gases
↓↓ ↓↓ ↓↓ ↓↓
TFID-R1
-- Span Gas Definition (1/2) --
Actual span gas concentration: 50.00 ppm
Spangas range -1: 50.00 ppm Spangas range -2: 250.0 ppm Spangas range -3: 2500 ppm Spangas range -4: 10000 ppm
Span gas unit: ppm "ppm" --> "mg/Nm3" conversion factor: 1.000
Concentration in span gas units: 37.50 ppm
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
More...
In the menu "Span Gas Definition (1/2)" you can set up the default value for each range of the current channel, the actual span gas concentration and the desired span gas unit. The concentration of each range should be a value between 70 and 110% of the end of range value. Furthermore, you can set up the "ppm" mg/Nm3" conversion factor for span gas.
Notes:
The "ppm" "mg/Nm
3
" conversion factor depends on the span gas which is used.
The set-up is necessary if you want to display span gas values in "mg/Nm3" !
To setup the "ppm" "mg/Nm
3
" conversion factor for sample gas you have to change
to the menu "General Concentration Measurement Setup" (see 5.1.10 p. 5-54)
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the 88 -key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the 88 -key or
cancel and go back to the last value with the F2 key.
The last line of variables: In this line you will find the measurement concentration in span gas units.
5 - 6
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
TFID - Span Gas Parameters:
Via the F5 key (More...) you can change to the second menu page of the "Span Gas Definition":
TFID-R1
-- Span Gas Definition (2/2) --
Span gas carbon atoms: 3
Span gas name: Propane Response factor: 1.000
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
In the menu "Span Gas Definition (2/2)" you will find the special span gas parameters of the TFID.
Lines of variables "Span gas carbon atoms" and "Response factor": The TFID originally measures the carbon atoms of a gas. So, for a correct conversion into
"ppm" and other units, it is necessary to determine the number of carbon atoms of the span gas. Because the raw measurement also differs with different kinds of hydrocarbons, you have to put in the response factors of the analyzer for the type of hydrocarbons.
Note: You will find the table of response factors in the Thermo-FID operation manual (No.: 90 003
806).
Line of variables "Span gas name": In this line you can put in the real name of the span gas you want to use, e.g. "Propane".
Note: The same parameters can be setup for sample gas in the menu "General Concentration
Measurement Setup" (see section 5.1.10 p. 5-54/55).
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98
NGA 2000
5 - 7
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Tolerances
Main Menu — Analyzer and I/O-module expert configuration
Analyzer module setup
Calibration parameters
Tolerances
↓↓ ↓↓ ↓↓ ↓↓
TAG
-- Tolerances --
Max. zero calibration deviation: 20.00 %
Max. span calibration deviation: 20.00 % Check deviation: Disabled Stability tolerance - range 1: 100.00 % Stability tolerance - range 2: 100.00 % Stability tolerance - range 3: 100.00 % Stability tolerance - range 4: 100.00 %
Last zero calibration: Success Last span calibration: Success
Measure Channel Back...
F1 F2 F3 F4 F5
95.00 ppm
In the menu "Tolerances" you can set up the parameters of the stability controlled calibration procedure (see p. 5-10/11) in one channel. Besides, you can disable the stability control and the maximum calibration deviation (deviation control deviation tolerance). If you want to set up the parameters of the other channels of a TFID analyzer or analyzer module, press the F3 key.
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the 88 -key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the 88 -key or
cancel and go back to the last value with the F2 key.
Lines "Last zero calibration/Last span calibration": The calibration status of the last calibration will be displayed in these two lines.
5 - 8
NGA 2000
90003808(1) [NGA-e (TFID-Software 3.2.X)] 06/98