The information contained in this document is subject to change without notice.
Rosemount Analytical's system of NGA 2000 Modular Gas Analyzers and Controllers are patented, under U.S. Patent
5.787.015.
Viton-A® is a registered trademark of E.I. duPont de Nemours & Co.
Paliney No. 7™ is a trademark of J.M. Ney Co.
Manual Part Number 748419-B
June 2000
Printed in U.S.A.
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
Table 3-2. Calibration Range for Various Zero Based Operating Ranges.........19
Table 3-3. Calibration Range for Various Suppressed Range Operations........19
Table 3-4. Oxygen Equivalents of Common Gases...........................................23
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
iii
CONTENTS
NOTES
Paramagnetic Detector Analyzer Module
iv
Rosemount Analytical NGA 2000
748419-B
June 2000
PREFACE
PURPOSE/SAFETY SUMMARY
The purpose of this manual is to provide the procedures for the installation, operation and
maintenance of this NGA 2000 module.
Some sections may describe equipment not used in your NGA 2000 system configuration.
The user should become thoroughly familiar with the operation of this module before
operating it. Read this instruction manual completely.
To avoid explosion, loss of life, personal injury and damage to this equipment
and on-site property, all personnel authorized to install, operate and service
the this equipment should be thoroughly familiar with and strictly follow the
instructions in this manual. SAVE THESE INSTRUCTIONS.
If this equipment is used in a manner not specified in these instructions,
protective systems may be impaired.
DANGER is used to indicate the presence of a hazard which will cause severe personal
injury, death, or substantial property damage if the warning is ignored
WARNING is used to indicate the presence of a hazard which can cause severe personal
injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor
personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation, or maintenance information which is
important but not hazard-related.
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without covers secure. Servicing requires access to live parts
which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a
properly grounded three-wire source of power.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P1
PREFACE
WARNING: POSSIBLE EXPLOSION HAZARD
This equipment is not designed for and should not be used in the analysis of
flammable samples. Use of this equipment in this way could result in explosion
and death.
NOTE
Apply leak test liquid to cell or detectors only as a last resort.
WARNING: POSSIBLE EXPLOSION HAZARD
Verify that all gas connections are made as labeled and are leak free. Improper
gas connections could result in explosion or death. See Section 2.4.3 for Leak
test procedure.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect
safety of this product. Use only factory documented components for repair
WARNING: OVER-VOLTAGE SPIKING
If this Analyzer Module is used with a non-Rosemount Analytical power supply,
adding Rosemount Analytical PN 90331 Current Protector in series with the 24 V
positive line will prevent over-voltage spiking and resultant fuse blowing when
powering up the instrument.
CAUTION: PRESSURIZED GAS
This module requires periodic calibration with a known standard gas. See
General Precautions for Handling and Storing High Pressure Gas Cylinders in
the rear of this manual.
Paramagnetic Detector Analyzer Module
P2
Rosemount Analytical NGA 2000
748419-B
June 2000
PREFACE
CAUTION: HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is
open. Dropping front panel while hand or fingers are inside either handle can
cause serious injury.
CAUTION: OVERBALANCE HAZARD
This Analyzer Module may tip instrument over if it is pulled out too far and the
Platform is not properly supported.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P3
PREFACE
GLOSSARY
Analyzer Module
Self contained analysis modules that are designed to be installed into the NGA 2000
System. One Analyzer Module can be installed into a Single Enclosure containing the
Platform Module. Two Analyzer Modules can be installed into a Dual Enclosure. The
simplest NGA 2000 System consists of one Analyzer Module.
Backplane
The interconnect circuit board which the Controller Board, Power Supply Board, I/O
Board(s) and Expansion Board(s) are plugged into the Backplane
Control module
The operator interface plus the Controller Board.
Controller Board
The Controller Board in the Platform which runs the software program that operates the
Display, Keypad and Network Manager. The Controller Board plugs into the Backplane.
Distribution Assembly
The Distribution Assembly consists of the Backplane and the card cages in the Platform
Module that contain I/O Module(s) and Expansion Module(s).
Expansion Board
The Expansion Board performs special features not related to I/O functions. The Expansion
Board plugs into the Backplane from the Platform front.
I/O Module
An auxiliary module that provides some sort of interface to the outside world. I/O modules
may include analog outputs, relay contacts, and digital interfaces. In general, they are
mounted in platforms as options.
Operator Interface
The Display and Keyboard of the Platform.
Paramagnetic Detector Analyzer Module
P4
Rosemount Analytical NGA 2000
748419-B
June 2000
PREFACE
Platform
Any combination of the NGA case, the display and computer board, power supply, and I/O
modules. In general, it could be considered to be anything in the NGA system other than
the analyzer modules.
Power Supply
Any of a variety of components that provide conditioned power to other NGA 2000
components, from the Power Supply Board that plugs into the Backplane in a stand-alone
instrument to several larger ones that can power larger collections of modules and
components.
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
The current status data placed on the network by an Analyzer Module. This includes
sample flow, source voltage and other diagnostic information.
Softkeys
The five function keys located below the front panel display. The menu function for each
softkey is displayed directly above it and is controlled by the software.
System
A NGA 2000 System consisting of one (or more) Analyzer Modules, an optional Platform,
one or more optional I/O Boards, an optional Expansion Board and an optional
Supplemental Power Supply.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P5
PREFACE
SPECIFICATIONS
MEASUREMENT SPECIES:
R
ANGES
:
REPEATABILITY:
M
INIMUM DETECTABLE LEVEL
NOISE:
L
INEARITY
R
ESPONSE TIME
D
RIFT (ZERO AND SPAN
E
FFECT OF TEMPERATURE
E
NVIRONMENT
A
MBIENT TEMPERATURE
E
FFECT OF FLOW
P
OWER REQUIREMENTS
:
:
):
:
:
:
Oxygen
0 to 100% oxygen; four fullscale selections, including
suppressed zero ranges in 1% increments
±1% of fullscale (at constant temperature)
0.01% oxygen
:
<1% of fullscale, peak-to-peak; <±1% for suppressed ranges
±1% of fullscale
0 to 90% of fullscale in 20 seconds (±2 seconds)
<±1% of fullscale/24 hours, <±2% of fullscale/week at constant
temperature;
<±2% of fullscale/24 hours, <±4% of fullscale/week of range for
99 to 100% (at constant temperature)
:
:
<±1% of fullscale over any 10°C interval for rate of change no
greater than 10°C per hour
Location - Class B controlled, indoor, non-hazardous
0 to 45°C (32 to 113°F)
<±1% of range when sample flow rate is changed by 20 ml/min.
24 VDC ±5%, 50 W max.; ripple and noise: <100 mV
peak-to-peak; line and load regulations: <±1%
SPECIFICATIONS – SAMPLE
T
EMPERATURE
F
LOW RATE
E
XHAUST PRESSURE
P
ARTICLES
D
EWPOINT
M
ATERIALS IN CONTACT WITH
S
AMPLE
Paramagnetic Detector Analyzer Module
P6
Rosemount Analytical NGA 2000
:
:
:
:
:
Non-flammable;: 10 to 66°C (50 to 150°F)
800 to 1400 ml/min.
:
-345 to 690 hPa-gauge (-5 to 10 psig)
filtered to <2 microns
below 43°C (110°F), no entrained liquid
Glass, 316 stainless steel, titanium, Paliney No. 7, epoxy
resin, Viton A, platinum, nickel, rhodium and MgF
2
748419-B
June 2000
SPECIFICATIONS - PHYSICAL
PREFACE
C
ASE CLASSIFICATION
D
IMENSIONS
W
EIGHT
:
:
MOUNTING:
M
AXIMUM LENGTH OF
C
ABLE
:
:
LON
General purpose for installation in weather-protected
areas
See Outline and Mounting Dimensions, Figure 2-4
8 kg (17.6 lbs.
Inside a Platform or custom-installed in a panel
1600 m (1 mile) between Analyzer Module and Platform
See the Preface section of the Platform Components manual for specifications regarding
Platform-related components (e.g., case dimensions) and the Preface of the I/O Module
manual for specifications regarding I/O (e.g., relay outputs).
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P7
PREFACE
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE
For order administration, replacement Parts, application assistance, on-site or factory
repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when you
call: Model Number, Serial Number, and Purchase Order Number or Sales Order Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive or
other hazardous material or used in such a hazardous environment, the user must attach
an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the
material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
TRAINING
A comprehensive Factory Training Program of operator and service classes is available.
For a copy of the Current Operator and Service Training Schedule contact the Technical
Services Department at:
Rosemount Analytical Inc.
Phone: 1-718-986-7600
FAX: 1-714-577-8006
Paramagnetic Detector Analyzer Module
P8
Rosemount Analytical NGA 2000
748419-B
June 2000
PREFACE
9
6
DOCUMENTATION
The following Paramagnetic Detector Analyzer Module instruction materials are available.
Contact Customer Service or the local representative to order.
748419 Instruction Manual (this document)
COMPLIANCES
This product may carry approvals from several certifying agencies, including Factory Mutual
and the Canadian Standards Association (which is also an OSHA accredited, Nationally
Recognized Testing Laboratory), for use in non-hazardous, indoor locations
®
NRTL /C
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to
harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May
1993).
NAMUR
This product satisfies all obligations of all relevant standards of the EMC framework in
Australia and New Zealand.
97-C219
748419-B
June 2000
N
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P9
PREFACE
NOTES
P10
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
INTRODUCTION
1.1 OVERVIEW
This manual describes the Paramagnetic Detector (PMD) Analyzer Module of Rosemount
Analytical's NGA 200 Series of gas analysis components.
The PMD Analyzer Module is designed to continuously determine the concentration of
oxygen in a flowing gaseous mixture. The concentration is expressed in ppm or percent
volume O2.
The entire Analyzer Module is designed as a slide-in module (if configured in stand-alone
instrument fashion), removable from the front of the Platform, with gas connections made
from the rear. All electronics relative to sample detection and conditioning are included in
this module.
1.2 TYPICAL APPLICATIONS
1
PMD Analyzer Module applications include:
• process control
• continuous emissions monitoring systems (CEMS)
• industrial gas production
• fermentation process monitoring
1.3 THEORY OF TECHNOLOGY
Oxygen is strongly paramagnetic (i.e., capable of becoming a temporary magnet when
placed in a magnetic field) while most other common gases are weakly diamagnetic (i.e.,
tend to be non-magnetic). See Figure 1-1.
The Magnetic susceptibility of the flowing gas sample is sensed in the detector/magnet
assembly. As shown in Figure 1-2, a dumbbell shaped, nitrogen -filled, h ollow gas test body
is suspended on a platinum/nickel alloy ribbon in a non-uniform magnetic field.
Because of a "magnetic buoyancy" effect, the spheres of the test body are subjected to
displacement forces, resulting in a displacement torque proportional to the magnetic
susceptibility of the gas surrounding the test body.
Measurement is accomplished by a null-balance system, whereby the displacement torque
is opposed by an equal and opposite restorative torque. The restoring current is
automatically maintained at the correct level by an electro-optical feedback system. A beam
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
1
INTRODUCTION
1
of light from the source LED is reflected off the square mirror attached to the test body onto
a bi-cell (dual photodiode).
The current required to keep the test body to the null position is a linear function of the total
magnetic susceptibility of the sample gas.
See Figures 4-1, 4-2, and 4-3 for component configuration.
Shaded
Pole
Piece
Sphere
(Magnetic Susceptibility = ko )
F
k
Sample Gas
(Magnetic Susceptibility = k )
Note:
) increases.
k
IGURE
F
1-1. S
As percentage of oxygen in sample gas increases,
displacement force (F
PHERICAL BODY IN NON-UNIFORM MAGNETIC FIELD
1.4 FEATURES
Among the features incorporated into the PMD Analyzer Module is a flow splitter (≈2:1)
that allows for greater sample flow, decreased lag time and faster analyzer response.
The "Time Alignment" feature can be used to delay sending the Primary Variable from
the PMD Analyzer Module for up to 30 seconds in 0.1 second intervals. This feature
allows Primary Variables form more than one PMD Analyzer Module to be "time
aligned" if necessary.
Paramagnetic Detector Analyzer Module
2
Rosemount Analytical NGA 2000
748419-B
June 2000
q
y
Axis
A
Displacement
q
ASSEMBLY
Tor
INTRODUCTION
ue
1
Balancing
Weight
Electromagnetic
Platinum/Nickel Alloy
Suspension Ribbon
TEST BODY DETAIL
Displacement
Torque
Restoring
ue
Tor
Restoring
Current
Mirror
Restoring
Current
Restoring
Torque
Current Loop
Electromagnetic
xis
Balancing Weight
Nitrogen-Filled Hollow Glass
Test Bod
IGURE
F
Magnet
Shaded Pole Pieces (4)
Dual Photocell
1-2. P
ARAMAGNETIC DETECTOR TECHNOLOGY
Test Body
Restoring
LED Source
DETECTOR/MAGNET
Zero
NULL BALANCE SYSTEM
SIGNAL CONDITIONING
Span
SIGNAL OUTPUT
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
3
INTRODUCTION
1
NOTES
Paramagnetic Detector Analyzer Module
4
Rosemount Analytical NGA 2000
748419-B
June 2000
INSTALLATION
2.1 UNPACKING
If the Paramagnetic Analyzer Module is received as a separate unit, carefully examine the
shipping carton and contents for signs of damage. Immediately notify the shipping carrier if
the carton or contents is damaged. Retain the carton and packing material until all
components associated with the Analyzer Module are operational.
2.2 ASSEMBLY
If the Analyzer Module requires assembly with other components (e.g., the Platform and
associated I/O Modules), do so at this time. Following the guides on the bottom left and
bottom center of the Platform, carefully slide the Analyzer Module halfway into place.
2
WARNING: HAND INJURY HAZARD
Do not place hands or fingers in the Platform front handles when front panel is
open. Dropping the front panel of the Platform while hand or fingers are inside
either handle can cause serious injury.
ANALYZER MODULE GUIDES
PIN SEATS
IGURE
F
748419-B
June 2000
2-1. A
DISENGAGED FRONT PANEL
NALYZER MODULE INSTALLATION INTO INSTRUMENT PLATFORM
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
5
2
INSTALLATION
Lift the spring-loaded pins on the front of the Analyzer Module, and carefully slide it the rest
of the distance. Secure the module in position by releasing the pins, which seat in the
available holes in the bottom of the case (see Figure 2-1). If the module and Platform are
difficult to assemble, remove the module, ensure the top cover of the module is firmly
seated on the hold-down screws, and repeat the assembly procedure.
Install I/O Module(s) according to guidelines in the I/O manual. After startup and calibration
have been performed, secure the front panel with the six screws provided.
2.3 LOCATION
Install the Analyzer Module in a clean, non-hazardous, weather protected, vibration free
location free from extreme temperature variations. For best results, either install the module
near the sample stream to minimize sample transport time or supply a flow greater than
necessary and route only the appropriate amount through the Analyzer Module.
Observing these requirements are critical. Note the following:
• Excessive vibration can cause a noisy readout. To minimize vibration effects, the
detector/magnet assembly is enveloped in a shock-mounted compartment.
• The user should ensure, when making any internal electrical connections, that no
cables are placed in contact with the detector assembly or associated internal sample
inlet and outlet tubing.
• Magnetic susceptibilities and partial pressures of gases vary with temperature.
Permissible ambient temperature range is 32°F to 113°F (0°C to 45°C).
• The interior of the Detector Assembly is maintained at approximately 144°F (62°C) by
an electronically controlled heater. Prior to entering the detector assembly, the sample
is heated in a coiled tubing to match the detector's temperature.
Paramagnetic Detector Analyzer Module
6
Rosemount Analytical NGA 2000
748419-B
June 2000
2.4 GASES
INSTALLATION
2
2.4.1 R
Calibration Gases
Analyzer Module calibration requires the establishment of zero and span calibration points.
This requires a zero standard gas to set the zero point span gas to establish a calibration
point at or near the upper range limit.
An oxygen-free gas, typically nitrogen, is required for use as the zero standard gas.
Recommendations for span calibration gases, bases on various operating ranges, are
tabulated in Table 3-1. Air (20.93% oxygen) can be used as span gas regardless of the
ranges used for sampling, although very low ranges may lose accuracy.
Sample Gas
Sample gas should be non-flammable.
Temperature
Sample temperature at the inlet should be from 50°F to 150°F (10°C to 66°C). A maximum
entry temperature of 110°F (43°C) is recommended to prevent cooling of the sample and
possible internal condensation. Such condensation could damage some components of the
Analyzer Module. This recommendation can be ignored if a thoroughly dry sample is
examined.
EQUIREMENTS
Pressure
Sample exhaust pressure limits are -5 to 10 psig (-345 to 690 hPa-gauge). Normal
operation is in the positive range, between 0 and 10 psig (0 and 690 hPa-gauge). Negative
gauge pressures are not normally recommended, but may be used in certain special
applications.
To prevent over-pressurization, insert a pressure relief valve into the sample inlet line. A
check valve should also be placed in the outlet line if the Analyzer Module is connected to a
manifold associated with a flare or other apparatus that does not operate at atmospheric
pressure.
The outlet port is commonly vented to the atmosphere. Any change in barometric pressure
has a directly proportional effect on the indicated percent of oxygen, and should be
neutralized through manual or computer correction of data. Note the following example:
Range = 0% to 5% oxygen
Barometric pressure change after calibration = 1%
The error is more significant for suppressed range 99% to 100%.
An optional barometric pressure compensation board is available to automatically perform
this correction.
A general rule regarding calibration gas pressure is that it should be the same as the
expected sample gas pressure during routine operation.
The above requirement increases the difficulty of operation at negative gauge pressure. A
suction pump can be connected to the outlet port for drawing sample through the Analyzer
Module. Such operation necessitates special precautions to ensure accurate readout,
including the following:
The need for equilibrium between sample and gas calibration pressures.
Any leakage in the sample handling system will decrease readout accuracy.
Flow Rate
Recommended sample flow rate is 800 to 1400 ml/min., ±40 ml/min. Optimum flow rate is
1100 ml/min.
If flow is held to within tolerance and operating pressure remains constant, zero and span
drift will meet specified limits.
Paramagnetic Detector Analyzer Module
8
Rosemount Analytical NGA 2000
748419-B
June 2000
INSTALLATION
(
)
NETWORK 1
NETWORK 2
POWER
FUSE
2
IGURE
F
Note: Reference and purge gas connections are applicable only to certain applications.
IGURE
F
2-2. PMD F
2-3. PMD B
RONT PANEL CONNECTIONS
FAN
ACK PANEL CONNECTIONS
SAMPLE
10 PSI MAX
69 kPa MAX
OUT
IN
2.4.2 C
ONNECTIONS
(See Figure 2-3) Connect inlet and outlet lines for sample gas to appropriately labeled
fittings on the rear panel. Both connections are 1/4 inch ferrule-type compression fittings.
Zero and span gases use the same inlet and outlet as the sample. Figure 2-4 shows a
typical external sample handling manifold for gas selection. Particulates must be filtered
down to two microns, gases generally require pressurization, and flow measurement
metering MUST be present.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
9
2
INSTALLATION
Needle
Valves
PMD Analyzer
Module
IGURE
F
Sample In
Zero
Standard
Gas
Span
Standard
Gas
2-4. I
Two Micron
Filter
Flowmeter
NTERCONNECTION OF TYPICAL GAS MANIFOLD TO
ODULE
M
Flow Splitter
(≈2:1)
PMD A
To Vent
NALYZER
10
2.4.3 L
EAK TEST
The Analyzer Module is thoroughly tested at the factory for gas leakage. The user is
responsible for testing for leakage only at the inlet and outlet fittings on the rear panel. The
user is also responsible for internal leak testing periodically and if any internal pneumatic
components are adjusted or replaced (with a test procedure selected by the user).
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
INSTALLATION
2.5 ELECTRICAL CONNECTIONS
NOTE
Electrical connections must be in compliance with National Electrical Code
(ANSI/NFPA 70) and/or any applicable national or electrical codes.
Two electrical connections are required on the Analyzer Module; POWER and NETWORK.
See Figure 2-2. On the Analyzer Module, two NETWORK connections are available, either
of which is appropriate for : 1) interconnection with Backplane of the Platform (see Platform
instruction manual) or 2) "daisy chaining" with other NGA 2000 components.
Connect Analyzer Module POWER 24 VDC power source, either the Platform or external
power source.
2
J5
J10
FAN
J11
J13
J14
J19
FLOW SENSOR CABLE
PRESSURE
COMPENSATION
BOARD
J6
COMPUTER ANALYSIS
J7
W1
PREAMP CABLE ASSEMBLY
CASE TEMP SENSOR
BOARD
J4
FLOW
SENSOR
J17
J1
J2
J3
J16
MAIN BOARD
J20
SENSOR
ASSEMBLY
J21J22
LED SOURCE
THERMAL CUTOFF
MAGNET HEATER AND
OPTICAL BENCH
ASSEMBLY
J9
J12
J15
IGURE
F
748419-B
June 2000
2-5. PMD W
IRING DIAGRAM
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
11
2
INSTALLATION
A
[213]
[208]
[109]
[28]
20.0
[508]
[157]
[152]
17.41
[442]
[15]
[13]
C
2.8
[71]
D
E
1.6
[40]
.6
[15]
8.4
[213]
E. FUSE.
D. POWER CABLE TO NETWORK.
C. NETWORK CABLE CONNECTIONS TO PLATFORM.
B. SAMPLE OUT: 1/4" O.D. TUBE FITTING.
A. SAMPLE IN: 1/4" O.D. TUBE FITTING.
5. MODULE TO BE INSTALLED WITHIN ±15° OF HO RIZONTAL.
4. POWER REQUIREMENTS: 24 VDC 3.5 A.
3. ELECTRICAL INSTALLATION MUST BE IN COMPLIANCE WITH NATIONAL ELECTRICAL
CODE (ANSI/NFPA 70) AND/OR ANY APPLICABLE NATIONAL OR LOCAL CODES.
2. MODULE IS NOT WEATHERPROOF.
1. APPROXIMATE WEIGHT: 15 LB (6.8 kg).
IGURE
F
2-6 O
UTLINE AND MOUNTING DIMENSIONS
6.7
[170]
B
10 PSI MAX
1.0
[25]
1.6
[41]
DIMENSIONS
INCH
[mm]
Paramagnetic Detector Analyzer Module
12
Rosemount Analytical NGA 2000
748419-B
June 2000
STARTUP AND OPERATION
3.1 OVERVIEW
Prior to initial startup, the user should perform the leak test procedure outlined in Section 2.
For the remainder of this section Analyzer Module interconnection with a Platform or some
interfacing component will be a ssumed. Display and keypad information shall ref er to that
which the user can expect to see and do with regard to the front panel of the Platform.
For a complete description of the Platform front panel controls and indicators, see the
Platform instruction manual.
3.2 DISPLAYS
Three kinds of Display screens are available to the user:
3
• Run Mode
• Menu
• Help
3.3 R
The Run Mode is the normal mode of operation. In this mode, the Display (see Figure 3-1)
will show current gas measurement, the component of interest, the current operations of the
softkeys, and a graphic bar representing the displayed concentration as ppm or as a
percent of oxygen.
If more than one Analyzer Module is connected to the system, the Run Mode display will
show as many as four gas measurements on screen. Alarm messages may also appear on
the display (See Table 3-1).
3.4 M
UN MODE DISPLAY
ENU DISPLAYS
The menu structure enables the user to access data and functions, and put information onto
the network.
The Main Menu (Figure 3-2) is subdivided into three levels of control based generally on
which personnel is likely to use it: Basic Controls, Expert Controls and Setup, and
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
13
3
STARTUP AND OPERATION
Technical Level Configuration. See Figures 3-3, 3-4, and 3-5. Many layers of the menu
structure are described at appropriate places throughout this manual.
From the Run Mode display, press the MENU softkey to enter the Main Menu (Figure 3-2).
MESSAGE DISPLAYDESCRIPTIONTYPE
BAROMETERSystem BarometerWARNING
CASE TEMPCase TemperatureWARNING
CRUDE NOISECalculated NoiseWARNING
CURRENTRNGHICurrent, High RangeWARNING
CURRENTRNGLOCurrent, Low RangeWARNING
DET TEMDetector TemperatureWARNING
FAN FETFan CurrentWARNING
HEATER FETHeater CurrentWARNING
LED CURRENTLED CurrentWARNING
LIN ERRORLinearizer ErrorWARNING
LOOP CURRENTPMD Loop CurrentWARNING
N15 VOLTSPower Supply, -15VWARNING
P15 VOLTSPower Supply, +15VWARNING
P24 VOLTSPower Supply, +24VWARNING
P5 VOLTSPower Supply, +5vWARNING
RAW SIGNALRaw SignalWARNING
SAMP PRESSample PressureWARNING
SVFLOWSample Bypass FlowWARNING
BICELLAPMD Photo SensorFAILURE
BICELLBPMD Photo SensorFAILURE
SW ERRORSoftware ErrorFAILURE
14
ABLE
T
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
3-1. PMD A
NALYZER MODULE ALARMS
748419-B
June 2000
STARTUP AND OPERATION
y
y
y
y
3.5 HELP DISPLAYS
The Help structure is intended to be an on-line "tutorial," context sensitive and topicinterconnected, so that the user can practically operate NGA 2000 without need of an
instruction manual (Figure 3-6).
Expert controls and setup ...
(Operati onal configuration)
Technical level configuration ...
(Diagnostic and manufacturing/service)
Delete alarm message!
Display Parms. Info
50
Main Menu
F1
IGURE
F
748419-B
June 2000
3-2. M
AIN MENU DISPLAY
F2F3F4F5
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
15
3
STARTUP AND OPERATION
g
p
p
23.2 ppm HC Analyzer XXXXXXXX
Measurement range numbers:
Range upper limit: 10 ppm
Range and functional control: Local
Bypass sample flow: 1000 ml/min
Ranges with valid calibration 1&2
Calibration status: Ready
If it won’t calibrate…
Flame condition: On
Light flame…
Home Escape Zero Span Info
F1F2F3F4F5
Basic Controls
IGURE
F
IGURE
F
3-3. B
3-4. E
ASIC CONTROLS MENU
23.2 ppm HC Analyzer XXXXXXXX
Ex
ert controls and set u
Expert analyzer controls ...
Auxiliary module controls ...
System set up ...
Analyzer module set up ...
Auxiliary module set up ...
Local I/O set up ...
Home Escape Info
F1
F2F3F4F5
XPERT CONTROLS AND SETUP MENU
23.2 ppm HC Analyzer XXXXXXXX
Technical confi
System set up ...
Service menus...
Diagnostic menus...
Other module diagnostic menus...
listing of all modules...
uration menu
16
IGURE
F
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
3-5. T
ECHNICAL LEVEL CONFIGURATION MENU
Home Escape Cal Info
F1F2F3F4F5
748419-B
June 2000
STARTUP AND OPERATION
p
23.2 ppm HC Analyzer XXXXXXXX
Main Menu Hel
The Main Menu for the analyzer system.
Note that this menu refers to the particular
analyzer selected from the run screen, when
used in a system. The softkey marked “HOME”
will always return you to this screen.
Help menu system...
Help on help...
Keyboard controls...
Editing controls...
Home Escape Map
3
IGURE
F
3-6. T
F1
YPICAL HELP SCREEN
F2F3F4F5
3.6 STARTUP PROCEDURE
Introduce a suitable on-scale gas (NOT actual sample) into sample inlet.
Apply power to the PMD Analyzer Module. If it is associated with a Platform, do this by
plugging in the Platform to a power source. The Platform has no ON/OFF power switch.
Once power is supplied to the Platform, the PMD Analyzer Module will be energized.
If the user's system contains only one Analyzer Module, all system components, the
Controller Board and the network "self-install" (bind together) during initial startup. If the
system contains more than one Analyzer Module, the startup sequence will interrogate the
network to locate and identify all components on the network. The u ser will have to bind
appropriate combinations of components after the startup sequence.
After the warm-up period, approximately one hour for PMD Module, all modules are
completely functional.
Enter appropriate data in the Calibration Gas List (by making the following display
selections: Main Menu, Expert Controls and Setup [enter security code, if necessary],
Analyzer Module setup, Calibration Gas List). Also, enter appropriate values in the
Calibration Parameters menu (by making the following display selections: Main Menu,
Expert Controls and Setup [enter security code, if necessary], Analyzer Module Setup,
Calibration Parameters), particularly data related to which ranges are to be zeroed together
and how the Analyzer Module is expected to calibrate ranges (separately or otherwise).
The Analyzer Module will not allow the user to increase the upper limit of a range beyond
the "maximum range" software setting. To change the "maximum range" value, select the
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
17
3
STARTUP AND OPERATION
following from the Main Menu: Technical Configuration Menu, Service Menu, Manufacturing
Data, Analyzer Module Data. Select Maximum Range, and use the arrow keys to scroll the
indicated value. The same applies for "minimum range" settings.
3.7 BINDING
To achieve full coordination between Analyzer Modules and associated I/O Modules, the
user must bind those components together in the System Setup portion of the Technical
Configuration Menu in software.
3.8 CALIBRATION
Calibration consists of establishing zero and span calibration points. Generally, zero and
span calibration should be performed on the range that will be used during sample analysis.
To calibrate the PMD Analyzer Module, introduce zero gas into the SAMPLE INLET, and do
the following:
1. If the Multi-Analyzer Module, split Run Mode display is shown, press the DISPLAY
softkey until the desired Analyzer's Run Mode display is acquired.
2. Press the MENUS softkey to enter the Main Menu.
3. Press the ENTER key to enter the Basic Controls menu.
4. Press the ZERO softkey to enter the Analyzer Zero menu.
5. Press the ZERO softkey and wait.
6. Introduce span gas (Table 3-2) into the SAMPLE INLET.
7. Press the SPAN softkey to enter the Analyzer Span menu, press SPAN again and wai t.
8. Press the HOME softkey to re-enter the Main Menu.
9. Press the DISPLAY softkey for the Run Mode display.
18
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
STARTUP AND OPERATION
3
RANGE % OXYGEN
0 to 1Nitrogen0.9% O2, balance N
0 to 2.5Nitrogen2.3% O2, balance N
0 to 5Nitrogen4.5% O2, balance N
0 to 10Nitrogen9% O2, balance N
RECOMMENDED ZERO
STANDARD GAS
RECOMMENDED SPAN
STANDARD GAS
2
2
2
2
0 to 25NitrogenAir (20.93% O2)
2
ABLE
T
0 to 50Nitrogen45% O2, balance N
0 to 100Nitrogen100% O
3-2. C
2
ALIBRATION RANGE FOR VARIOUS ZERO BASED OPERATING RANGES
For users of analyzers with suppressed indicating ranges, it may be desireable to calibrate
the analyzer zero and span points within the suppressed range with gases in the
suppressed range. Table 3-3 shows recommended zero and span standard gases for
suppressed range oxygen indication in an Argon background.
ABLE
T
RANGE % OXYGEN
50 to 10050 - 50.5% O2 Balance AR100% O
70 – 100%70 – 70.5% O2 Balance AR100% O
98 – 100%98 – 98.5% O2 Balance AR100% O
99 – 100%99 – 99.5% O2 Balance AR100% O
3-3. C
ALIBRATION RANGE FOR VARIOUS SUPPRESSED RANGE OPERATIONS
RECOMMENDED ZERO
STANDARD GAS
RECOMMENDED SPAN
STANDARD GAS
2
2
2
2
If the user is unable to calibrate the Analyzer Module (i.e., when ZERO or SPAN is initiated,
nothing happens), a possible solution relates to the use of an incorrect gas for zeroing or
spanning (e.g., using a high concentration gas to zero or a zero gas to span the Analyzer
Module). Recalibrating with the appropriate gas(es) WILL NOT correct the problem
because the ZERO OFFSET or SPAN FACTOR has been set to an extreme value in the
process.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
19
3
STARTUP AND OPERATION
To remedy the problem, do the following:
1. Select the following from the Main Menu: Expert Controls and Setup (enter security
code if necessary), Analyzer Module Setup, and Calibration Parameters.
2. Using the down arrow, select Zero Ranges, press ENTER and, using the up/down
arrows, toggle to SEPARATE. Do the same for the Span Ranges selection. Do not
press ESCAPE at any time unless retention of prior settings is desired.
3. Return to the Main Menu (HOME) and make the following selections: Expert Controls
and Setup (enter security code if necessary), Expert Analyzer Controls, CAL softkey,
FACTORS softkey, and Range 1 (2, 3, 4) Factors. (Do steps 4 and 5 for each range.)
4. Select Zero Offset, press ENTER, adjust the value to 525000 with the up/down arrow
keys, and press ENTER. Do not press ESCAPE at any time unless retention of prior
settings is desired.
5. Select Span Factor, press ENTER, adjust the value to 0.000015 with the up/down arrow
keys, and press ENTER. Do not press ESCAPE unless retention of prior settings is
desired.
6. Attempt to recalibrate the Analyzer Module according to the procedure outlined at the
beginning of Section 3.8. If re-calibration fails, return to the Range Factors menu,
readjust factors, and attempt calibration again.
3.9 BACKGROUND GAS COMPENSATION
Any gas having a composition other than 100% oxygen contains background gas, that is,
non-oxygen components. Sometimes, the PMD Module response to background gas is
significant, depending largely on the span and range used.
If the operator uses zero and span gases that contain the same background gas as the
sample, calibration procedures automatically compensate. No adjustments are necessary.
If the background gas in the sample is different from that in the zero and/or span gases, the
operator must take into consideration background effects to ensure correct readout. During
entry of zero and span gas values in the Calibration Gas List, the instrument is not set to
indicate the true oxygen content of the zero and span standard gases. It is set to indicate a
slightly different value, relative to background gas, calculated to provide correct readout
during subsequent analysis of sample gas.
20
Oxygen Equivalent Values of Gases
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
STARTUP AND OPERATION
For computation of background corrections, the analyzer's response to each component of
the sample must be known. Table 3-4 lists the percentage oxygen equivalent values for
many common gases. For a more comprehensive list of oxygen equivalent values, refer to
a resource text such as the Handbook of Chemistry and Physics for tables of magnetic
susceptibility of substances. The percentage oxygen equivalent of a gas can be determined
by the following equation, assuming both gases are supplied at the same pressure:
Analyzer Response to Gas
% O2 Equivalent of Gas =
Analyzer Response to O
For example, if the analyzer's response to oxygen is +100%, the response to xenon would
be -1.34%.
The oxygen equivalent of a gas mixture is the sum of the contribution of the individual gas
components.
Example: Zero Based Range
X 100%
2
3
At lower range limit (i.e., 0% O2), composition of sample is: 80% CO2, 20% N2.
From Table 3-4, the percent oxygen equivalents are: CO2 -0.623%, N2 -0.358%.
The percent oxygen equivalent of the mixture =
Computing Adjusted Values for Calibration Gas List
Before calibrating the Analyzer Module, values in the Calibration Gas List must be adjusted
to correct for magnetic susceptibility of background gas. In the equation that follows, the
quantities are defined as follows:
• BGGst = oxygen equivalent of background gas in standard gas (Table 3-4).
• BGGs = oxygen equivalent of background gas in sample (Table 3-4).
• OP = operating pressure. Unless special pressure corrections are to be made,
the zero standard, span standard and sample gases must all be admitted at the
same pressure.
Use the following equation to compute the adjusted settings for the Calibration Gas List:
Adjusted percent oxygen for standard gas =
(A)[100 + (B-C)]-100[B-C]
100
Where:
748419-B
June 2000
A = true percent oxygen of standard gas
B = BGGs
C = BGGst
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
21
3
STARTUP AND OPERATION
Example:
Background gs in sample is CO2, oxygen equivalent = -0.623%.
Zero gas is 100% N
Span standard gas is air: 21% O2, 79% N
2.
2.
Background gas in zero and span standard gases is N2, oxygen equivalent = 0.358%.
With N2 zero standard gas flowing, zero gas value in the Calibration Gas List would be
0.265% O2 (as determined by the following):
0[100+(-0.623-(-0.358))] - 100{-0.623-(-0.358)]
100
= 0.265% O
2
With air flowing, span gas value in the Calibration Gas List would be 21.21% oxygen (as
determined by the following):
21(100 - 0.265) - 100 (-0.265)
100
= 21.209% O2 ≅ 21.21 O
2
In two limiting cases, the general equation is reduced to simpler forms.
1. If the span standard gas is 100% oxygen, the adjusted oxygen value is the same as the
true value (i.e., 100% O2).
2. If the zero standard is an oxygen-free zero gas, the adjusted value for setting the ZERO
Control = BGGst-BGGs. (If the oxygen-free zero gas is more diamagnetic than the
background gas in the sample, this difference is negative. The negative value may be
entered in the Calibration Gas List.)
22
Alternately, the user can avoid these compensation calculations by using zero and span
gases which have been specially prepared to contain the expected amounts of background
gas. Calibration of the analyzer module will then factor in background gas effects in the
same proportions as normal run mode measurement.
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
STARTUP AND OPERATION
3
GASEQUIV. % AS O
Acetylene, C2H
Allene, C3H
Ammonia, NH
2
4
3
-0.612Hyd r o g e n B r o m i d e , Hbr-0.968
-0.744Hydrogen Chloride, HC
-0.479Hyd r o g e n F luori d e , H F-0.253
2
GAS
EQUIV. % AS
1
O
2
-0.650
Ar g o n , A-0.569Hyd rog e n I odide, H I-1.403
Bromine, Br
Although normally calibrated for readout in percent oxygen, the PMD Analyzer Module
actually responds to oxygen partial pressure. The partial pressure of the oxygen
component in a gas mixture is proportional to the total pressure of the mixture. Thus
readout is affected by pressure variations.
For instance, assume that an instrument is calibrated for correct readout with a standard
gas containing 5% oxygen, admitted at the normal sea level atmospheric pressure of 14.7
psia (1013 hPa). If the operating pressure now drops to one-half of the original value (i.e.,
to 7.35 psia/506 hPa and the calibration controls are left at the previously established
settings, the display reading for the standard gas will drop to 2.5%.
It is therefore necessary to calibrate the in strument at the same pressure that will be used
during subsequent operation, and to maintain this pressure during operation.
Alternatively, an optional Barometric Pressure Compensation Board, typically used for
suppressed range applications, can perform signal corrections automatically.
24
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
MAINTENANCE AND TROUBLESHOOTING
WARNING: QUALIFIED PERSONNEL
This equipment should not be adjusted or repaired by anyone except properly
qualified service personnel.
4.1 OVERVIEW
PMD Analyzer components that may require replacement include:
• All printed circuit board
4
• Thermal fuse inside Detector
• Case temperature sensor
• Flow sensor
• Power fuse
• Detector
• Module fan
The LED bi-cell assembly source required adjustment (rotation) anytime the Detector is
disassembled. Refer to Figures 4-2 for locations of these components.
4.2 P
Refer to Figure 4-1 for locations of the Signal, Microprocessor, Power Supply and (optional)
Pressure Compensation Boards.
All boards are secured to a side of the analyzer module that folds out while interconnection
wiring is still in place. Remove the securing screws and fold out the entire panel.
RINTED CIRCUIT BOARD REPLACEMENT
To remove individual boards on the fold-out panel, label and unplug all interconnection
wiring, and remove securing hardware. Reverse this procedure for installation.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
25
4
MAINTENANCE AND TROUBLESHOOTING
MODULE BOARD
COMPUTER
ANALYSIS BOARD
PRESSURE
COMPENSATION
BOARD
FLOW SENSOR
NETWORK INPUT
DETECTOR
HOUSING
CASE
TEMPERATURE
SENSOR
IGURE
F
4-1. PMD M
ODULE
AJOR COMPONENTS
- M
4.3 MODULE FAN REPLACEMENT
The Analyzer Module fan assembly is disassembled as shown in Figure 4-2.
26
IGURE
F
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
4-2. M
ODULE FAN ASSEMBLY
748419-B
June 2000
MAINTENANCE AND TROUBLESHOOTING
4.4 THERMAL FUSE REPLACEMENT
See Figure 4-3 for location of the Detector Thermal Fuse. Remove the Detector Assembly
from the detector housing, replace the thermal Fuse. Reassemble in reverse order.
DETECTOR HOUSING
DETECTOR ASSEMBL Y
THERMAL FUSE
4
DETECTOR ASSEMBLY
TOP VIEW
SPLITTER
ASSEMBLY
BOTTOM VIEW
IGURE
F
4-3. D
ETECTOR ASSEMBLY
MAGNET/COIL
ASSEMBLY
THERMAL
FUSE
DETECTOR
ASSEMBLY
LED
OPTICAL
BENCH
ASSEMBLY
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
27
4
MAINTENANCE AND TROUBLESHOOTING
4.5 F
See Figure 4-1 for location of Flow Sensor. To replace the sensor, remove connections to
sample gas line and disconnect securing hardware. Reassemble in reverse order.
4.6 P
The power fuse is located in the Network Input Module and is accessible through the front
panel of the PMD Analyzer. To remove the fuse, push and turn the fuseholder cap 1/4 turn
counterclockwise. Verify that the replacement fuse is the same type and rating.
LOW SENSOR REPLACEMENT
OWER FUSE REPLACEMENT
28
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
REPLACEMENT PARTS
WARNING: PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely
affect safety of this product. Use only factory-approved components for repair.
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
PMD IDENTIFICATION MATRIX
Each analyzer is configured per the customer sales order. Below is the PMD sales matrix
which lists the various configurations available.
To identify the configuration of an analyzer, locate the analyzer name-rating plate. The 12position sales matrix identifier number appears on the analyzer name-rating plate.
P NGA 2000 PARAMAGNETIC DETECTOR
Code Language
A English
X Special
A
Code Configuration Identifier
A10 Calibrated Standard Ranges: 0-5, 0-10, 0-25, 0-100%
A15 Calibrated Standard Ranges: 0-1, 0-5, 0-10, 0-25%
A20
A21Same as A20 for elevations from 501 to 1500 ft above sea level
A22Same as A20 for elevations from 1501 to 2500 ft above sea level
A23Same as A20 for elevations from 2501 to 3500 ft above sea level
A24Same as A20 for elevations from 3501 to 4500 ft above sea level
A25Same as A20 for elevations from 4501 to 5500 ft above sea level
A26Same as A20 for elevations from 5501 to 6500 ft above sea level
A27Same as A20 for elevations from 6501 to 7500 ft above sea level
X99 Special Calibration Ranges
Calibrated Standard Suppressed Ranges/Barometric Pressure
Compensation: 90-100, 95-100, 98-100, 99-100%
Code Detector Type
A00 Rhodium Plated Current Loop
Z00 Titanium Current Loop (Standard)
ZZZ No Selection
Z No Selection
PAA10Z00ZZZZExample
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
A1
PMD IDENTIFICATION MATRIX
A
NOTES
Paramagnetic Detector Analyzer Module
A2
Rosemount Analytical NGA 2000
748419-B
June 2000
ENERAL PRECAUTIONS FOR HANDLING AND
G
TORING HIGH PRESSURE GAS CYLINDERS
S
Edited from selected paragraphs of the Compressed
Gas Association's "Handbook of Compressed Gases"
published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against
extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders
should be stored in the shade when located in areas where extreme temperatures are
prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured
against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be
moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an
empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A
flame should never be permitted to come in contact with any part of a compressed gas
cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric
arc welding, precautions must be taken to prevent striking an arc against the cylinder.
4125 E
AST LA PALMA AVENUE
Rosemount Analytical Inc.
• A
J
, C
NAHEIM
ULY
ALIFORNIA
1997 • 748525-C • P
92807-1802 • 714-986-7600 • FAX 714-577-8006
RINTED IN
USA
(blank)
ARRANTY
W
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from
defects in workmanship and material under normal use and service for a period of twelve (12)
months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid
junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and
material under normal use and service for a period of ninety (90) days from date of shipment by
Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or
material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods,
part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid,
within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of
consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for
replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in
the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall
not operate to condemn such commercial unit when such goods, part(s) and consumables are
capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or
indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any
other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER
GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER W ARRANTIES ARE
GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN
PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL
BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD W ARRANTY CLAUSE. IN
NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S
LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY
SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT
SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL
DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS
OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND
DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE
OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR
PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE
AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH
SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond
Seller's direct control.
Rosemount Analytical
4125 E
AST LA PALMA AVENUE
Rosemount Analytical Inc.
• A
F
EBRUARY 1997 • 7485189-C • PRINTED IN USA
NAHEIM
, C
ALIFORNIA
92807-1802 • 714-986-7600 • FAX 714-577-8006
(blank)
IELD SERVICE AND REPAIR FACILITIES
F
Field service and repair facilities are located worldwide.
U.S.A.
To obtain field service on-site or assistance with a service problem, contact (24 hours, 7
days a week):
National Response Center
1-800-654-7768
INTERNATIONAL
Contact your local Rosemount Sales and Service office for service support.
FACTORY
For order administration, replacement Parts, application assistance, on-site or factory repair,
service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when you call:
Model Number, Serial Number, and Purchase Order Number or Sales Order Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sende r, f re ight collect.
When return ing any product or compon ent that has been expo sed to a toxic, co rrosive or
other hazardous material or used in such a hazardous environment, the user must attach an
appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the material
has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
4125 E
AST LA PALMA AVENUE
Rosemount Analytical Inc.
• A
J
, C
NAHEIM
ULY 1997 • 748190-G • PRINTED IN USA
ALIFORNIA
92807-1802 • 714-986-7600 • FAX 714-577-8006
(blank)
NSTRUCTIONS
I
015-748438-A
Rosemount Analytical
ADDENDUM
ARAMAGNETIC DETECTOR ANALYZER MODULE
P
This addendum serves as an amendment to the Paramagnetic Detector Analyzer Module.
The following information should be considered part of the manual, and supersedes any
conflicting information in the body of the manual.
Read this information and note the conflicts.
ENU DISPLAYS
M
Menu: 0 ANALOP
Menu: 1 ANALSET
AST LA PALMA AVENUE • ANAHEIM
4125 E
UGUST
A
Rosemount Analytical Inc.
ALIFORNIA
1999
, C
•
92807-1802
015-748438-A
• PRINTED IN
(714) 986-7600
•
USA
FAX: (714) 577-8006
•
ADDENDUM
Menu: 2 FLOCHEK
Menu: 3 ZEROI1
2 of 40
Menu: 4 SPANI1
Rosemount AnalyticalAugust 1999748438-A
Menu: 5 FLOCHEKI1
Menu: 6 ANALOPI1
ADDENDUM
Menu: 7 ACALSET
August 1999748438-A
Rosemount Analytical
3 of 40
ADDENDUM
Menu: 8 LINSET
Menu: 9 APARLST
4 of 40
Menu: 10 ANALSETI1
Rosemount AnalyticalAugust 1999748438-A
Menu: 11 CALLIST
Menu: 12 CALLISTI1
ADDENDUM
Menu: 13 ACALSETI1
August 1999748438-A
Rosemount Analytical
5 of 40
ADDENDUM
Menu: 14 APARLSTI1
Menu: 15 AMMAN
6 of 40
Menu: 16 AMMANI1
Rosemount AnalyticalAugust 1999748438-A
Menu: 17 AMSVC
Menu: 18 AMSVCI1
ADDENDUM
Menu: 19 ADIAG
August 1999748438-A
Rosemount Analytical
7 of 40
ADDENDUM
Menu: 20 AMPWR
Menu: 21 AM1V
8 of 40
Menu: 22 AMTEMP
Rosemount AnalyticalAugust 1999748438-A
Menu: 23 AMMISC
Menu: 24 AMTREND
ADDENDUM
Menu: 25 ADIAGI1
August 1999748438-A
Rosemount Analytical
9 of 40
ADDENDUM
Menu: 26 RANGESETAM
Menu: 27 RANGESSETI1
10 of 40
Menu: 28 SPANI2
Rosemount AnalyticalAugust 1999748438-A
Menu: 29 ACALSETI2
Menu: 30 ZEROI2
ADDENDUM
Menu: 31 LINRANGE1
August 1999748438-A
Rosemount Analytical
11 of 40
ADDENDUM
Menu: 32 LINRANGE2
Menu: 33 LINRANGE3
12 of 40
Menu: 34 LINRANGE4
Rosemount AnalyticalAugust 1999748438-A
Menu: 35 LINRANGE0
Menu: 36 AMPWRI1
ADDENDUM
Menu: 37 FLOCHEK1I1
August 1999748438-A
Rosemount Analytical
13 of 40
ADDENDUM
Menu: 38 FILTER
Menu: 39 AM1VI1
14 of 40
Menu: 40 AMTEMPI1
Rosemount AnalyticalAugust 1999748438-A
Menu: 41 AM2VA
Menu: 42 PLIMITSA
ADDENDUM
Menu: 43 TLIMITSA
August 1999748438-A
Rosemount Analytical
15 of 40
ADDENDUM
Menu: 44 AMMISCI1
Menu: 45 ANALSIMPLE
16 of 40
Menu: 46 FILTERI1
Rosemount AnalyticalAugust 1999748438-A
Menu: 47 LINSET1I1
Menu: 48 LINRANGE0I1
ADDENDUM
Menu: 49 PLIMITSAI1
August 1999748438-A
Rosemount Analytical
17 of 40
ADDENDUM
Menu: 50 CALFACTORS
Menu: 51 R1FACTORS
18 of 40
Menu: 52 RN2FACTORS
Rosemount AnalyticalAugust 1999748438-A
Menu: 53 RN3FACTORS
Menu: 54 RN4FACTORS
ADDENDUM
Menu: 55 RFACTORSI
August 1999748438-A
Rosemount Analytical
19 of 40
ADDENDUM
Menu: 56 AMHELPINDEX
Menu: 57 LINRANGE1I1
20 of 40
Menu: 58 CALFACTORSI1
Rosemount AnalyticalAugust 1999748438-A
Menu: 59 APARLST2
Menu: 60 APARLST4
ADDENDUM
Menu: 61 APARLST5
August 1999748438-A
Rosemount Analytical
21 of 40
ADDENDUM
Menu: 62 APARLST6
Menu: 63 DISPLAY
22 of 40
Menu: 64 MPARMS
Rosemount AnalyticalAugust 1999748438-A
Menu: 65 MPARMS2
Menu: 66 MPARMSI1
ADDENDUM
Menu: 67 BAROM_PARMS
August 1999748438-A
Rosemount Analytical
23 of 40
ADDENDUM
Menu: 68 BAROM_PARMI1
Menu: 69 TLIMITSIAI1
24 of 40
Menu: 70 SW_DIAG
Rosemount AnalyticalAugust 1999748438-A
Menu: 71 LISTNOTES
Menu: 72 FLOWLIMITS
ADDENDUM
Menu: 73 AMPATH
August 1999748438-A
Rosemount Analytical
25 of 40
ADDENDUM
Menu: 74 CALI1
Menu: 75 LINFUNCT
26 of 40
Menu: 76 POLYSETUP
Rosemount AnalyticalAugust 1999748438-A
Menu: 77 MIDPOINT1
Menu: 78 POLYGAS1
ADDENDUM
Menu: 79 POLYSETI1
August 1999748438-A
Rosemount Analytical
27 of 40
ADDENDUM
Menu: 80 POLYGAS2
Menu: 81 POLYGAS3
28 of 40
Menu: 82 POLYGAS4
Rosemount AnalyticalAugust 1999748438-A
Menu: 83 POLYGAS5
Menu: 84 POLYGAS6
ADDENDUM
Menu: 85 POLYGAS7
August 1999748438-A
Rosemount Analytical
29 of 40
ADDENDUM
Menu: 86 POLYGAS8
Menu: 87 POLYGAS9
30 of 40
Menu: 88 POLYGAS0
Rosemount AnalyticalAugust 1999748438-A
Menu: 89 MIDPOINT2
Menu: 90 MIDPOINT3
ADDENDUM
Menu: 91 MIDPOINT4
August 1999748438-A
Rosemount Analytical
31 of 40
ADDENDUM
Menu: 92 EXP_CAL
Menu: 93 EXP_CAL_DAT
32 of 40
Menu: 94 EXP_CAL_DATI
Rosemount AnalyticalAugust 1999748438-A
Menu: 95 UNITS
Menu: 96 UNITSI1
ADDENDUM
Menu: 97 POLYSETI2
August 1999748438-A
Rosemount Analytical
33 of 40
ADDENDUM
Menu: 98 POLYSETI3
Menu: 99 RESET
34 of 40
Menu: 100 STORE
Rosemount AnalyticalAugust 1999748438-A
Menu: 101 RFHIST1A
Menu: 102 RFACTORSIA
ADDENDUM
Menu: 103 RFHIST2A
August 1999748438-A
Rosemount Analytical
35 of 40
ADDENDUM
Menu: 104 RFHIST3A
Menu: 105 RFHIST4A
36 of 40
Menu: 106 TWEAKI1
Rosemount AnalyticalAugust 1999748438-A
Menu: 107 ANALSETI3
Menu: 108 ACALSETI1A
ADDENDUM
Menu: 109 INIT
August 1999748438-A
Rosemount Analytical
37 of 40
ADDENDUM
Menu: 110 SWDIAGI1
Menu: 111 STOREDPVA
38 of 40
Menu: 112 ZERO_NOW
Rosemount AnalyticalAugust 1999748438-A
Menu: 113 SPAN_NOW
Menu: 114 CALFAIL
ADDENDUM
Menu: 115 ABOUT
August 1999748438-A
Rosemount Analytical
39 of 40
ADDENDUM
Menu: 116 ABOUT1
Menu: 117 ALARM1
40 of 40
Menu: 118 MANDATA
Rosemount AnalyticalAugust 1999748438-A
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.