Rosemount NGA 2000 FID Hydrocarbon Analyzer Module SW 3.6 including Hardware-Rev A Manuals & Guides

Instruction Manual
760001-A October 2002
Model NGA2000 FID
Flame Ionization Detector Module
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and in­ternational standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them
normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
to ensure they continue to operate within their
Read all instructions
prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical representative for clarification.
Follow all warnings, cautions, and instructions
marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruc- tion Manual and per applicable local and national codes. Connect all products to the proper elec-
trical and pressure sources.
To ensure proper performance, use qualified personnel
to install, operate, update, program, and
maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY
. Look-alike substitutions may result
in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Teflon and Viton are registered trademarks of E.I. duPont de Nemours and Co., Inc. Kynar® is a registered trademark of Pennwalt, Inc. Tygon is a registered trademark of Saint-Gobain Performance Plastics
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Model NGA2000 FID

TABLE OF CONTENTS

Instruction Manual
760001-A
October 2002
PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-5
Documentation....................................................................................................................................P-6
Compliances .......................................................................................................................................P-6
Glossary of Terms ............................................................................................................................P-7
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Overview................................................................................................................................1-1
1-2 Typical Applications...............................................................................................................1-1
1-3 Theory of Technology............................................................................................................1-1
1-4 Gas Safety Features..............................................................................................................1-3
1-5 Fuel Gas Option ....................................................................................................................1-3
1-6 Specifications.........................................................................................................................1-5
a. General ...........................................................................................................................1-5
b. Gas Requirements ..........................................................................................................1-6
c. Physical...........................................................................................................................1-7
d. Gas Connections.............................................................................................................1-7
2-0 INSTALLATION ....................................................................................................................2-1
2-1 Unpacking..............................................................................................................................2-1
2-2 Assembly ...............................................................................................................................2-1
2-3 Location .................................................................................................................................2-2
2-4 Gases ....................................................................................................................................2-3
a. Overview .........................................................................................................................2-3
b. Connections ....................................................................................................................2-3
c. Specifications ..................................................................................................................2-5
d. Leak Test ........................................................................................................................2-6
2-5 Electrical Connections ...........................................................................................................2-6
3-0 OPERATION .........................................................................................................................3-1
3-1 Overview................................................................................................................................3-1
3-2 Displays .................................................................................................................................3-1
3-3 Run Mode Display .................................................................................................................3-1
3-4 Menu Displays .......................................................................................................................3-1
3-5 Help Displays.........................................................................................................................3-1
3-6 Startup Procedure .................................................................................................................3-7
3-7 Binding...................................................................................................................................3-9
3-8 Calibration..............................................................................................................................3-9
3-9 Routine Operation .................................................................................................................3-10
3-10 Safety System .......................................................................................................................3-10
4-0 MAINTENANCE AND SERVICE ..........................................................................................4-1
4-1 Burner disassembly and cleaning .........................................................................................4-1
4-2 Component Replacement......................................................................................................4-2
4-3 Burner Startup Troubleshooting ............................................................................................4-2
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Instruction Manual
760001-A October 2002
5-0 REPLACEMENT PARTS ......................................................................................................5-1
5-1 Replacement Parts ................................................................................................................5-1
a. Electronics.......................................................................................................................5-1
b. Burner Assembly and Components ................................................................................5-1
c. Electro-Mechanical..........................................................................................................5-1
d. O-Rings and Gaskets......................................................................................................5-1
e. Pneumatics .....................................................................................................................5-1
5-2 Matrix .....................................................................................................................................5-2
6-0 RETURN OF MATERIAL ......................................................................................................6-1
6-1 Return Of Material .................................................................................................................6-1
6-2 Customer Service ..................................................................................................................6-1
6-3 Training..................................................................................................................................6-1
7-0 APPENDIX A – MENU SCREENS .......................................................................................7-1
8-0 INDEX....................................................................................................................................8-1
Model NGA2000 FID
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

LIST OF ILLUSTRATIONS

Figure 1-1. Flame Ionization Detection Technology..........................................................................1-1
Figure 1-2. FID Component Locations – Top View ...........................................................................1-2
Figure 2-1. Analyzer Module Installation into Platform......................................................................2-1
Figure 2-2. Back Panel Connections.................................................................................................2-2
Figure 2-3. FID Module Flow Diagram ..............................................................................................2-4
Figure 2-4. Front Panel Connections, Controls and Indicators .........................................................2-4
Figure 2-5. Outline and Mounting Dimensions..................................................................................2-7
Figure 2-6. FID Wiring Diagram........................................................................................................2-8
Figure 3-1. Run Mode Display...........................................................................................................3-1
Figure 3-2. Main Menu Display..........................................................................................................3-2
Figure 3-3. Basic Controls Menu Display..........................................................................................3-2
Figure 3-4. Expert Controls and Setup Menu Display.......................................................................3-2
Figure 3-5. Technical Configuration Menu Display ...........................................................................3-3
Figure 3-6. Typical Help Screen........................................................................................................3-3
Figure 3-7. Typical Curves of Module Response vs. Pressure Setting on Sample Pressure
Regulator .......................................................................................................................3-5
Figure 3-8. Typical Curves of Module Response vs. Pressure Setting on Fuel Pressure
Regulator .......................................................................................................................3-5
Figure 3-9. Typical Curves of Module Response vs. Pressure Setting on Air Pressure
Regulator .......................................................................................................................3-6
Figure 3-10. Front Panel Torque Sequence........................................................................................3-11
Figure 4-1. FID Burner Assembly – Exploded View..........................................................................4-3
Instruction Manual
760001-A
October 2002

LIST OF TABLES

Table 1-1. Analyzer Characteristics Relative to Fuel Gas ...............................................................1-4
Table 3-1. FID Analyzer Module Alarms ..........................................................................................3-4
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
760001-A October 2002
Model NGA2000 FID
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the NGA2000 FID and the System Accessories of the NGA2000 System.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
Instruction Manual
760001-A
October 2002

PREFACE

DEFINITIONS

DANGER .
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
760001-A October 2002
Model NGA2000 FID

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be im­paired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel. For safety and proper per­formance this instrument must be connected to a properly grounded three-wire source of power.
WARNING .
POSSIBLE EXPLOSION HAZARD
This equipment is used in the analysis of sample gases which may be flammable, and the burner fuel used in the ionization process is flammable. A continuous dilution purge system is fac­tory-installed (in accordance with Standard ANSI/NFPA 496-1993, Chapter 6, and it must be func­tional at all times during operation. Do not disable this purge system.
WARNING .
FLAMMABLE SAMPLES
Consult the factory if flammable samples will be measured.
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Do not place hands or fingers in the Platform front handles when front panel is open. Dropping the front panel of the Platform while hand or fingers are inside either handle can cause serious injury.
Ensure that all gas connections are made as labeled and are leak free. Improper gas connections could result in explosion or death.
This module requires periodic use of pressurized gas. See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5.
Instruction Manual
760001-A
October 2002
WARNING.
HAND INJURY HAZARD
WARNING.
POSSIBLE EXPLOSION HAZARD
CAUTION .
PRESSURIZED GAS
CAUTION .
PURGE AIR REQUIREMENT
This Analyzer Module must be used in conjunction with a device (Control Module or PC Interface) that can actively monitor network variables related to pressure or flow of the continuous dilution purge, or the front panel LEDs of the Analyzer Module, as installed, must be visible. The purpose of this requirement is to maintain adherence to ANSI/NFPA 496 standard which assures the continued viability of the purge system. Under no circumstances should any pressure or flow indicator be connected to the PURGE AIR OUT outlet of the Analyzer Module because this may affect the sealing performance of the module.
CAUTION.
OVER-VOLTAGE SPIKING
If this Analyzer Module is used with a non-Rosemount Analytical power supply, adding Rosemount P/N 903341 Current Protector in series with the 24 V positive power line will prevent over-voltage spiking and resultant fuse blowing when powering up the instrument.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
760001-A October 2002
Model NGA2000 FID
CAUTION .
STATIC ELECTRICITY
Circuit boards in this instrument are static-sensitive. Take all static precautions when handling the circuit boards
NOTE
This Analyzer Module is completely leak-tested at the factory for gas leakage. The user is responsi­ble for testing for leakage at the inlet and outlet fittings on the rear panel (with a test procedure chosen by the user). The user is also responsible for leak-testing periodically and if any internal pneumatic components are adjusted or replaced. See leak test instructions in Section2-4d on page 2-6.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
760001-A
Model NGA2000 FID
October 2002
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of
weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or
bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a
suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is
attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125
never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding,
precautions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
9
6
760001-A October 2002
Model NGA2000 FID

DOCUMENTATION

The following NGA2000 FID instruction materials are available. Contact Customer Service Center or the local representative to order.
760001 Instruction Manual (this document)

COMPLIANCES

This product may carry approvals from several certifying agencies, including Factory Mutual and the Cana­dian Standards Association (which is also an OSHA accredited Nationally Recognized Testing Laboratory, NRTL) for use in non-hazardous, indoor locations.
The certification marks appear on the product name-rating plate.
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May 1993).
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand.
FM
APPROVED
®
97-C219
NAMUR
N
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

GLOSSARY OF TERMS

Analyzer Module
The module that contains all sensor/detector components for development of a Primary Variable signal; in­cludes all signal conditioning and temperature control circuitry.
Backplane
The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module power and net­work cables, I/O Modules and Expansion Modules plug into.
Control Module
The Operator Interface plus the Controller Board.
Controller Board
The computer board that serves as the Network Manager and operates the Display and Keypad.
Instruction Manual
760001-A
October 2002
Distribution Assembly
The Backplane and the card cages that hold I/O and Expansion Modules.
Expansion Module
A circuit board that plugs into the Backplane from the front of the Platform and performs special features not related to I/O functions.
I/O Module
A circuit board that plugs into the Backplane from the rear of the Platform. Has a connector terminal for communication with external data acquisition devices and provides an input/output function.
Operator Interface
The Display and Keyboard.
Platform
Any workable collection of the following: Controller Board, Power Supply, Distribution Assembly, Enclosure and Operator Interface.
Power Supply
Any of a variety of components that provides conditioned power to other NGA2000 components, from the Power Supply Board that plugs into the front of the Backplane in a stand-alone instrument to several larger ones that can power larger collections of modules and components.
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
Data placed on the network by a module regarding current status, e.g., sample flow, source voltage and other diagnostic information.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Softkeys
The five function keys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software.
System
Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s).
P-8 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
A

DESCRIPTION AND SPECIFICATIONS

1-1 OVERVIEW

Instruction Manual
760001-A
October 2002
SECTION 1
This manual describes the Flame Ionization Detector (FID) Analyzer Module of Rose­mount Analytical's NGA2000 Series of gas analysis components. See Figure 1-1 below and Figure 1-2 on page 1-2.
The FID Analyzer Module is designed to con­tinuously determine the concentration of hy­drocarbons in a flowing gaseous mixture. The concentration is expressed in parts-per-million or percent of volume.
The entire FID Analyzer Module is designed as a slide-in module (if configured in stand-alone instrument fashion), removable from the front of the Platform, with gas con­nections made from the rear. All electronics relative to sample detection and conditioning are included in this module.

1-2 TYPICAL APPLICATIONS

The monitoring of atmospheric air for low-level hydrocarbon contaminants and determining the hydrocarbon content of exhaust emissions from internal combustion engines are exam­ples of typical applications for the FID Ana­lyzer Module.

1-3 THEORY OF TECHNOLOGY

This Analyzer Module uses the flame ioniza­tion method of detection. The sensor is a burner in which a regulated flow of sample gas passes through a flame sustained by regulated flows of a fuel gas (hydrogen or a hydrogen/diluent mixture) and air.
Within the flame, the hydrocarbon compo­nents of the sample stream undergo a com­plex ionization that produces electrons and positive ions. Polarized electrodes collect
these ions, causing current to flow through an electronic measuring circuit
The ionization current is proportional to the rate at which carbon atoms enter the burner, and is therefore a measure of the concentra­tion of hydrocarbons in the sample. This measure of concentration is placed on the network, where it is can be shown on the Plat­form Display or on other data acquisition de­vices.
Negative Ion Collection Ring
+90V
Sample
Figure 1-1. Flame Ionization Detection
Technology
Signal Conditioning
Positive Carbon Ions
ir
Fuel
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
A
760001-A October 2002
Gas Lines (See Figure 2-2)
Fuel Shutoff Solenoid Valve
Capillary
Case Heater Silicon Pad (On Base)
Purge Air Flow Switch
Fuel Pressure Sensor
Air Pressure Sensor
Ignite/Operate Mode Burner-Air Flow Solenoid Valve
ir Circulation Fan
Purge Air Heater Cartridge
Sample Pressure Sensor
Burner Air and Fuel Regulators (Stacked Vertically)
Sample Back Pressure Regulators
Model NGA2000 FID
Purge Air Pressure Regulator
Purge Air Outlet Pressure Switch
Preamp Board (In Shield)
Detector (Burner)
Computer Board
Sample Bypass Flow Sensor
DC-DC Converter
Safety Board
Power Supply Board
Network & Power Module
Manual Ignite Switch
Figure 1-2. FID Component Locations – Top View
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

1-4 GAS SAFETY FEATURES

The FID Analyzer Module is designed with a factory installed continuous dilution purge sys­tem in accordance with standard ANSI/NFPA 496 - 1993, Chapter 6. Front-panel LEDs indi­cate that the burner flame is lit and that the purge system is enabled. In addition, fuel gas is automatically shut off when a flame-out condition occurs or the safety system is dis­abled.
The purge system is enabled only if there is proper purge gas flow in, purge gas pressure, and internal case pressure, and after five times the case volume has been exchanged.
All tubing ahead of the burner is rigid metallic tubing assembled with ferrule/nut type com­pression fittings. However, should an internal fuel leak occur, a worst-case leak would be dissipated below 25% of the LEL of hydrogen through the combination of an inlet fuel flow restrictor and purge gas flow.
This module is designed to use 100% hydro­gen fuel or 40% H mum inlet pressure of 3446 hPa-gauge (50 psig). A different flow restrictor is used for each fuel type.
A standard FID Analyzer Module is only equipped to analyze a non-flammable sample, below 100% of the LEL.
POSSIBLE EXPLOSION HAZARD
Protection against explosion depends upon a special fuel flow restrictor in the fuel inlet fitting. Do not remove fuel inlet restrictor. Use the correct fuel flow restric­tor for the fuel being used. Do not use 100% hydrogen fuel in a 40% H2/60% He configured Analyzer Module. Replace only with factory supplied fitting.
2/60% He fuel at a maxi-
WARNING
Instruction Manual
760001-A
October 2002
equipped to use 100% hydrogen fuel. The particular application and characteristics of the sample gas to be measured will dictate the preferred type of fuel. The following guide­lines can be used for determining fuel gas type:
1. For measuring low-level hydrocarbons in ambient air or in other sample gas with relatively constant oxygen content, 100% hydrogen is preferable. It provides the highest obtainable sensitivity and maxi­mum stability. Zero drift caused by ambi­ent temperature variations of the fuel cylinder is somewhat lower for 100% hydrogen than for mixed fuel. (With either fuel, it is desirable to maintain a constant cylinder temperature.)
2. For monitoring internal combustion ex­haust emissions or other sample gas with varying oxygen content, mixed fuel is preferable. In fact, a hydrogen/helium mixture is more desirable than a hydro­gen/nitrogen mixture. With this type of sample, the use of mixed fuel gas mini­mizes the error introduced by oxygen synergism.
An effective way to reduce the effect of inter­nal oxygen is to dilute it with an inert gas. This can be accomplished with a constant dilution of sample and calibration gases upstream from the burner. But it is simpler and more ac­curate to provide that diluent in the form of premixed fuel. Both nitrogen and helium have been used as a diluent, but helium has proven to be most effective in improving the quality of response to the various species of hydrocar­bons.
As indicated earlier the flame output signal is optimum when the ratio of hydrogen flow to inert flow is about 40/60. Therefore, this is the chosen composition for hydrogen/helium pre­mixed fuel.
The sample flow is kept low to maximize the

1-5 FUEL GAS OPTION

The standard FID Analyzer Module requires 40% hydrogen/60% helium burner fuel gas. As an option, the analyzer module can be
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
dilution effect while still providing adequate sensitivity. The burner air flow is normally about four times the fuel flow, and changes have little effect on signal strength. For a
Instruction Manual
760001-A October 2002
given flow, the signal can be optimized by ad­justing the fuel flow rate.
Typical flow rates with premixed fuel:
Model NGA2000 FID
with the latter. However, in any application where the sample contains more than one species of hydrogen and/or a varying concen­tration of oxygen, mixed fuel is preferred.
Fuel 100 cc/min
Sample 7 cc/min
Air 400 cc/min
Note that with a 40/60 premixed fuel, the above flow rates amount to 40 cc (8%) hydro­gen, 67 cc (13%) inert plus sample and 400 cc (79%) air, which compare closely to the 30 cc (8%) hydrogen, 45 cc (12%) inert/sample and 300 cc (80%) air noted earlier for straight hydrogen fuel.
Since sample flow in the case of mixed fuel operation is only about 1/6 of that with straight hydrogen fuel, higher sensitivity is obtained
ANALYZER FUEL GAS
CHARACTERISTICS 100% H2 40% H2/60% He
Fullscale Sensitivity 1 ppm, CH4 to 2%, CH4 4 ppm, CH4 to <5%, CH4
Fuel Consumption 35 to 40 cc/min 75 to 80 cc/min
Operating Range 276 to 345 hPa-gauge (4 to 5 psig) 207 to 345 hPa-gauge (3 to 5 psig)
Table 1-1. Analyzer Characteristics Relative to Fuel Gas
The mixed fuel is recommended, not only for sample containing variable concentrations of oxygen, but also for a specific pure gas appli­cation. If straight oxygen samples are used with straight hydrogen fuel, the mixture enter­ing the burner is essentially 40% H which tends to produce an unstable signal. The mixed fuel works better. Note that the choice of fuel determines certain analyzer characteristics, as shown in Table 1-1 below.
2/60% O2,
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

1-6 SPECIFICATIONS

a. General

Measurement Species................... Total hydrocarbons
Ranges .......................................... 0 to 100 ppm (output scalable down to 0-2 ppm
H2/He fuel.............................. low range: 0 to 4 ppm CH
H2 fuel
Repeatability.................................. 1% of fullscale at a constant temperature, sample
Minimum Detectable Level
H2/He fuel.............................. 0.04 ppm
H2 fuel
Noise ............................................. 1% of fullscale, peak to peak
Linearity ......................................... ≤ ±1% of fullscale for H
Response Time
CEMS .................................... 30 sec. For 10% to 90% of fullscale, with sample
ICEE ...................................... 1 sec. For 10% to 90% of fullscale, with sample
Zero Drift........................................ ≤ ±1% of fullscale/24 hours at constant
Span Drift....................................... ≤ ±1% of fullscale/24 hours at constant
Effect of Temperature.................... ≤ ±2% of fullscale for any temperature change of
Operating Temperature ................. 32°F to 113°F (0°C to 45°C)
Power Requirements..................... +24 VDC ±5%, 120 W max.. direct to analyzer
1
.................................. low range: 0 to 1 ppm CH4, through 0 to 2500 ppm
1
................................. 0.01 ppm H2 fuel
Instruction Manual
fullscale)
4, through 0 to 1% CH4
high range: 0 to 50 ppm CH4, through 0 to <5% CH
4
CH4 high range: 0 to 10 ppm CH CH4
flow and fuel, burner air and sample pressure
bypass flow at 0.5 L/min. (non-flammable sample)
bypass flow at 0.5 L/min. (non-flammable sample)
temperature, hydrocarbon concentration of supply gases, sample flow and fuel, burner air and sample pressure
temperature, hydrocarbon concentration of supply gases, sample flow and fuel, burner air and sample pressure
10°C and rate of change less than 10°C/hour
module; Ripple and Noise: <100 mV peak to peak
Line and Load Regulations: <±1%
4, through 0 to 2%
2/He fuel and He fuel
760001-A
October 2002
1
Option
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
760001-A October 2002

b. Gas Requirements

Sample .......................................... Non-flammable, below 100% of LEL
Flow rate ................................ 0.5 to 40 ml/min.
THC ....................................... 0.5 ppm, CH
Supply pressure..................... 483 to 1035 hPa-gauge (7 to 15 psig)
Temperature .......................... 32°F to 132°F (0°C to 55°C), <20°C variance/24
Particulates............................ Filtered to <2 microns
Dewpoint................................ <45°C
Purge Air........................................ Instrument air, nitrogen or other non-flammable
Flow rate: .............................. 16 to 18 L/min.
Supply pressure:.................... 689 to 1378 hPa-gauge (10 to 20 psig)
Fuel Gas (Standard)...................... Premixed 40% hydrogen and 60% helium
Flow rate ................................ 75 to 80 ml/min.
THC ....................................... 0.5 ppm, CH
Supply pressure..................... 2415 to 3450 hPa-gauge (35 to 50 psig)
Model NGA2000 FID
4
hours, <10°C variance/hour
gas (refer to ANSI/NFPA 496 for the requirements for the Protective Gas System)
4
WARNING
POSSIBLE EXPLOSION HAZARD
Unless this Analyzer Module is factory- or field-configured specifically for using 100% hydrogen fuel, DO NOT USE PURE HYDROGEN FUEL. An explosion resulting in severe personal injury or death could occur. Also, each Analyzer Module is fac­tory-configured for either mixed or pure hydrogen fuel, and cannot use the fuel for which it was not configured unless field reconfiguration is done.
Fuel Gas (H2 option) ..................... Zero-grade hydrogen
Flow rate ................................ 35 to 40 ml/min.
THC ....................................... 0.5 ppm, CH
Supply pressure..................... 2415 to 3450 hPa-gauge (35 to 50 psig)
Burner Air ...................................... Zero-grade air
Flow rate ................................ 350 to 400 ml/min
THC ....................................... ppm, CH
Supply pressure............................. 1725 to 3450 hPa-gauge (25 to 50 psig)
4
4
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

c. Physical

Case Classification:....................... General purpose for installation in
Maximum Separation .................... 1600m (1 mile) from Analyzer Module to Platform
Materials in Contact With Sample . Stainless steel, Teflon, glass-filled Teflon, brass,
Dimensions.................................... See Outline and Mounting Dimensions, Figure 2-5
Weight ........................................... 10.43 kg (23 lbs.)
Mounting........................................ Horizontal, inside a Platform or custom installed in

d. Gas Connections

Sample In: ..................................... 1/4" O.D. tube fitting, stainless steel
Regulated Air In: ..............................1/4" O.D. tube fitting, brass
Burner Air In: ....................................1/4" O.D. tube fitting, brass
Fuel In:..............................................1/4" O.D. tube fitting, stainless steel
Purge Air In: .....................................3/8" O.D. tube fitting, brass
Purge Air Out: ..................................3/8" O.D. tube fitting, brass
Bypass Out:......................................1/4" O.D. tube fitting, stainless steel
Burner Exhaust Out: ........................1/2 inch O.D. tube connection, Tygon or equivalent
Pressure Relief Valve
Instruction Manual
760001-A
October 2002
weather-protected area
neoprene, Kynar
on page 2-7
a panel
(this connection shall slope downward 6° minimum from horizontal)
CAUTION
PRESSURE RELIEF VALVE RESTRICTION
No connection shall be made to this fitting. If this caution is ignored, damage to the case seals may occur, and the instrument will not operate properly.
See the Preface section of the Platform Components manual for specifications re­garding Platform-related components and the Preface of the I/O Module manual for specifications regarding I/O (e.g., relay outputs).
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
Instruction Manual
760001-A October 2002
Model NGA2000 FID
1-8 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
A

2-1 UNPACKING

If the FID Analyzer Module is received as a separate unit, carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or contents is damaged. Retain the carton and packing material until all components associated with the Analyzer Module are operational.

2-2 ASSEMBLY

If the Analyzer Module requires assembly with other components (e.g., the Platform and associated I/O Modules), do so at this time.
PIN SEATS
Instruction Manual
760001-A
October 2002
SECTION 2

INSTALLATION

Following the guides on the bottom left and bottom center of the Platform, carefully slide the Analyzer Module halfway into place.
CAUTION
HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is open. Dropping front panel while hand or fingers are inside either handle can cause serious injury.
NALYZER MODULE GUIDES
DISENGAGED FRONT PANEL
Figure 2-1. Analyzer Module Installation Into Platform
Rosemount Analytical Inc. A Division of Emerson Process Management Installation
2-1
Instruction Manual
760001-A October 2002INSTALLATION
Lift the spring loaded pins on the front of the module, and carefully slide it the rest of the distance. Secure the module in position by releasing the pins, which seat in the available holes in the bottom of the case (see Figure 2-1 on page 2-1). If the module and Platform are difficult to assemble, remove the module, ensure the top cover of the module is firmly seated on the hold down screws, and repeat the assembly procedure.
Connect the network cable to either the NETWORK 1 or NETWORK 2 connection on the Analyzer Module (see Figure 2-4 on page 2-4), and the NETWORK connection on the Backplane (see Platform manual). Connect the power cable to both the Analyzer Module front panel and to the Backplane.
Install I/O Module(s) according to guidelines in the I/O manual. After startup and calibration have been performed, secure the Front Panel with the six screws provided.
Model NGA2000 FID
WARNING
INSTALLATION RESTRICTIONS
For safety, the Analyzer Module should be installed in a non-confined, ventilated space. Do not block any of the rear panel outlets as they are part of the safety system.
Operating ambient temperature is 0°C to 45°C, limited to temperature changes of less than 10°C/hr. Acceptable dewpoint range is less than 95% relative humidity, but not in excess of 45°C wet bulb temperature.
The cylinders of fuel, air, and calibration gas(es) and the source of purge air should be located in an area of relatively constant ambient temperature.

2-3 LOCATION

Install the Analyzer Module in a clean, weather-proofed, non-hazardous, vibration free location free from extreme temperature variations. For best results, install the Analyzer Module near the sample stream to minimize sample transport time.
PURGE
AIR OUT
PRESSURE
RELIEF
VALVE
MIXED
FUEL
IN
BURNER
SAMPLE
IN
PURGE
AIR IN
AIR IN
BYPASS OUT
VENT TO SAFE AREA
SLOPE DOWNWARD
6 MINIMUM
MAXIMUM INPUT PRESSURE
BURNER AIR: 50 PSIG (3450 hPa)
PURGE AIR: 20 PSIG (1378 hPa)
BURNER EXHAUST OUT
FUEL: 50 PSIG (3450 hPa)
SAMPLE: 15 PSIG (1035 hPa)
Figure 2-2. Back Panel Connections
Installation Rosemount Analytical Inc. A Division of Emerson Process Management
2-2
Model NGA2000 FID

2-4 GASES

a. Overview

During normal operation, the Analyzer Module requires fuel and air to maintain the burner flame as well as suitable stan­dard gases for calibration and instrument air for purge requirements. Criteria for se­lection of these gases follow in 2-4c on page 2-5.
After initial startup or after startup follow­ing a prolonged shutdown, the analyzer may display baseline drift for a consider­able period of time, particularly on the most sensitive range. Commonly, the drift is caused by small amounts of hydrocar­bons in the inner walls of the tubing in both the internal flow system and the ex­ternal gas supply system. Drift results from any factor influencing the equilibrium of these absorbed hydrocarbons, such as temperature or pressure.
Note that this type of drift occurs only when the flame is burning. If drift occurs when the flame is extinguished, the elec­tronic circuitry is at fault. To minimize drift, use clean fuel and air, keep the analyzer clean, and locate the gas cylinders in an area of relatively constant ambient tem­perature.
The cylinders supplying all gases each should be equipped with a clean, hydro­carbon free, two stage regulator and a shutoff valve.
All new external gas tubing (except for PURGE IN/OUT and SAMPLE BYPASS) is strongly recommended, preferably pre­cleaned, stain-less steel, gas chromatograph grade tubing. Thoroughly clean before use. if a hydrocarbon based cleaning solvent such as acetone is used, purge tubing with dry nitrogen or helium for several minutes before using.
Instruction Manual
760001-A
October 2002
Since the oxidation of hydrogen is ac­companied by the formation of water va­por, the Exhaust tubing always should be slanted downward at least 6 degrees from horizontal. Otherwise, water may accumu­late in the line, causing back pressure and noisy readings, or may back up in the line and flood the burner.
If the sample is toxic or noxious, or is to be reclaimed, connect the Bypass outlet to a suitable disposal system. Do not use any device that may cause back pressure in the line.
Purge air and burner air should be sup­plied from separate sources.

b. Connections

Reference Figure 2-2 on page 2-2. Con­nect inlet and outlet lines for sample, burner fuel and air, exhaust, bypass, and purge to appropriately labeled fittings on the rear panel. All connections are 1/4 inch ferrule type compression fittings ex­cept the PURGE AIR IN and OUT con­nections, which are 3/8 inch compression fittings. The BURNER EXHAUST OUT is a 1/2 inch connection. Burner exhaust, bypass and purge air out must be vented at atmospheric pressure to a non­classified location in accordance with ANSI/NFPA-496.
It is recommended that no connection be made to the PURGE AIR OUT port. If, however, the analyzer's location requires interconnection with a venting system, the 3/8" O.D. line should be kept as short as possible, and no longer than four feet.
CAUTION
POSSIBLE INSTRUMENT DAMAGE
No connection should be made to the PRESSURE RELIEF VALVE fitting. Doing so may cause damage to the instrument.
Gas line connections are compression fit­tings. Do not use pipe thread tape.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
V
760001-A October 2002
BURNER
EXHAUST
OUT
REGULATOR
Restrictor
PURGE AIR
Plug 1/4MPT Brass
Run Tee 1/4T - 1/8MPT Brass
Tubing, Teflon 1/4” OD
PURGE
FLOW
SWITCH
Bulkhead 3/8T Brass
Tubing, SS 3/8” OD
PURGE AIR
SENSOR
PURGE AIR HEATER
Model NGA2000 FID
SAMPLE
IN
IN
CAPILLARY
Tubing, Viton 1/4” OD
PURGE PRESS
SAMPLE
PRESS
SENSOR
Branch Tee 1/8T - 1/8FPT SS
SAMPLE
BACK PRESS
REGULATOR
Figure 2-3. FID Module Flow Diagram
BURNER
AIR
!
!
FUEL
Figure 2-4. Front Panel Connections, Controls and Indicators
BYPASS
SAMPLE
Elbow 1/8T - 1/8MPT Kynar
Connector 1/8T - 1/8MPT Kynar
WARNING
ATTENTION
OUT
Bulkhead Reducer 1/4T - 1/8T Brass
Tubing, Teflon 1/8” OD
FLOW
SENSOR
Tubing, Teflon 1/8” OD
Elbow 1/8T - 1/4MPT SS
BURNER AIR PRESSURE REGULATOR
SAMPLE
FUEL OVERRIDE
BURNER
AIR IN
Bulkhead Reducer 1/4T - 1/8T SS
GA
IN OUT
BURNER
AIR
PRESS
SENSOR
POWER
HEAT
FLAME
ON
PURGE
Filter
LON
LON
1 +
AIR
2 -
3 GND T 6A 260V
COM
3-WAY VALVE
1
2
IGNITE
Run Tee 1/4T - 1/4FPT Brass
PURGE AIR
OUT
NC
Run Tee 1/8T - 1/8MPT SS
FID
WARNING
!
ATTE NTION
!
Bulkhead Connector 1/4T - 1/8NPT SS
Fuel Restrictor
Purge Exit Restrictor
PRESS
SW
Connector 1/16T - 1/8MPT SS
BURNER
MIXED FUEL
IN
SOLENOID
ALVE
Tubing, Viton 1/4” OD
FUEL
PRESS
SENSOR
Fuel Restrictor
NETWORK 1
NETWORK 2
POWER
FUSE
Connector 1/8T - 1/8MPT SS
FUEL PRESSURE REGULATOR
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

c. Specifications

Fuel Gas
Standard analysis usually requires mixed fuel, i.e., 40% (±2%) hydrogen and 60% helium. H2/He mixed fuel is recom­mended over H2/N2 fuel because of bet­ter linearity in concentration output. Such blends are supplied by many gas vendors specifically for this use, with a guaranteed maximum total hydrocarbon content of 0.5 ppm, measured as methane. This specifi­cation should be used when obtaining these mixtures.
NOTE
The fuel restrictor is marked with a red dot, and the sample capillary is marked with a red or green dot for mixed fuel applications.
Some applications require the use of 100% hydrogen fuel. When using this option, always ensure that sample pressure (4 to 5 psig) is present when fuel flow is present. Otherwise, the de­tector tip may be damaged.
The fuel restrictor and sample capillary are marked with a white dot for 100% hydrogen fuel applications.
Burner Air
In order to ensure a low background sig­nal, burner air should contain less than 1 ppm maximum total hydrocarbon content. An alternate source for burner air and zero gas (see CALIBRATION GASES be­low) is a combination diaphragm pump and heated palladium catalyst. This proc­ess continuously removes moderate amounts of hydrocarbons and carbon monoxide from ambient air.
Purge Air
Instrument quality air, nitrogen, or other nonflammable gas is required for the safety purge system.
Instruction Manual
760001-A
October 2002
Calibration Gases
Calibration method and gases depends on the type of fuel gas used, the operating range, and the desired measurement ac­curacy. In all methods, zero and span gases are used, and are introduced through the sample inlet at the rear of the module.
Z
ERO GAS - Analysis is affected by the
background gas of the sample. Therefore, it is recommended to use zero gas with as close to the background composition of the sample as possible. Normally less than 0.5 THC as CH
If the burner fuel is 100% hydrogen, the zero gas, background gas of the sample or background gas of the span gas cannot be hydrogen or oxygen. These gases combined with pure hydrogen fuel would generate excessive heat in the burner, causing deterioration of the internal com­ponents of the burner.
S
PAN GAS - Span gas consists of a speci-
fied concentration of methane or other hydrocarbon in a background gas such as nitrogen. Analysis is affected by the
background gas of the sample. There­fore, span gas containing the same background gas as the sample is rec­ommended. Then, the background ef­fect is canceled out.
S
AMPLE GAS - Sample gas should be non-
flammable (below 100% of the sample's LEL). For high sensitivity applications re­quiring background gas compensation, contact the factory.
F
LOW RATE - The sample flow rate must
be between 0.5 L/min. and 2 L/min. Flow rate for purge air should be 16 to 18 L/min.
P
RESSURIZATION/FILTRATION - Sample
pressure at the SAMPLE inlet should be within the range of 483 to 1035 hPa-gauge (7 to 15 psig), and internally, should be 345 hPa-gauge (5 psig) nomi­nally. Burner fuel pressures should be:
4 is sufficient.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
760001-A October 2002
1725 to 3450 hPa-gauge (25 to 50 psig) for cylinder regulator, 1518 to 1723 hPa-gauge (22 to 25 psig) internal. Burner air pressures should be: 1725 to 3450 hPa-gauge (25 to 50 psig) for cylin­der regulator, 965 to 1103 hPa-gauge (14 to 16 psig) internal. Purge air (external supply) pressure should be between 689 and 1378 hPa-gauge (10 and 20 psig), 689 to 827 hPa-gauge (10 to 12 psig) nominal. The internal purge air regulator pressure is factory preset at a nominal setting of 551 hPa-gauge (8 psig) with a supply pressure of 689 hPa-gauge (10 psig). Noncompliance with these specifi­cations, particularly those concerning purge air, could cause over-pressure damage to the module. The nominal in­ternal case pressure is about 0.5 to 1.0 inch of water, and the pressure relief valve is set at 1/3 psig (nominal).
At the very least, the module's safety sys­tem, which requires a certain volume of purge air flowing through the case before allowing burner ignition, will not allow the instrument to operate.
Model NGA2000 FID
for accuracy. Sample should be filtered for particulates down to two microns.

d. LEAK TEST

The Analyzer Module is completely tested at the factory for gas leakage. The user is responsible for testing for leakage at the inlet and outlet fittings on the rear panel. The user is also responsible for internal leak testing periodically and if any internal pneumatic components are adjusted or replaced (with a test procedure chosen by the user).

2-5 ELECTRICAL CONNECTIONS

Two electrical connections are required on the Analyzer Module: POWER and NETWORK. See Figure 2-4 on page 2-4. On the Analyzer Module, two NETWORK connectors are available, either of which is appropriate for: 1) interconnection with the Backplane of the Platform or 2) "daisy-chaining" with other NGA2000 components. Connect Analyzer Module POWER to Backplane POWER or ex­ternal 24 VDC power source.
All internal pressure settings are preset at the factory, but the operator should check
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
.25
[6.3]
Dimensions:
INCHES
[MM]
4.3
[109.7]
1.1
[27.9]
.9
[22.5]
1.9
[49.4]
Figure 2-5. Outline and Mounting Dimensions
[208.2]
[152.4]
17.75
[450.9]
20.0
[510.0]
8.2
6.0
[143.6]
5.7
4.3
[109.1]
FID
2.9
[73.6]
1.4
[35.6]
.8
[20.9]
D
Instruction Manual
760001-A
October 2002
1.7
[43.4]
1.8
[46.3]
3.1
[78.0]
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
A
N
R
E
A
A
A
760001-A October 2002
Model NGA2000 FID
CARTRIDGE HEATER
CABLE ASSEMBLY, FLAT 3 COND
LON/POWER
MODULE
J5
P5
E2 24VOLT
GR
RE
BL
N
D
K
E1 RTN
E3
CABLE ASSEMBLY, FLAT 10 COND
J6
CABLE ASSEMBLY, FLAT 24COND
P1
J4
P4
P6
J6
J1
POWER
SUPPLY
BOARD
J5
P5
P9
J9
OVEN HEATER
FUEL SOL.
++ +
J10
P10
YEL
BRN
RED
ORN
GRN
P3
J2
P2
J3
J8
P8
IR
BLU
P15
P11
FAN
J7
P7 J11
BLK
BLK
RED
RED
IR
J15
3-WAY SOLENOID
J14
P14
SENSOR,
CASE TEMP
HARNESS, CHASSIS
ASSEMBLY
THERMOSTAT
HEATER
CABLE ASSEMBLY,
COND
FLAT 3
CABLE ASSEMBLY, FLAT 16 COND
2-WAY
SOLENOID
J16
P16
FUEL IN
HARNESS, GLOW PLUG
P4
J4
SAFETY BOARD
J3
P3
SSY,
SWITCH
PRESSURE
SWITCH
CABLE
P2
J2
1 2 3 4 5 6 7 8
FLO SW
J1
P1
POWER RESISTOR
FLOW
SENSOR
CASE
PRESS
FLO
SW
FLO
SW
FLO SEN
HARNESS
J11
J10
J12
CHASSIS GND
SHIELD
E3
R37
R38
J13
P13
CABLE ASSEMBLY, ANODE
SENSOR, FLAME-OUT
CABLE ASSEMBLY, FLAT 20 COND
MTG STUD
GND STRAP
BOARD
J2P2J3
P7
J7
P3
POL VOL
GRY
CABLE ASSEMBLY, CATHODE
PUR
P1
J1
E2
EI
SHIELD
PREAMP
BOARD
P6
J6
COMPUTER
ANALYSIS
J5
P5
J4
P4
J1
P1
Figure 2-6. FID Wiring Diagram
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
POWER
J1
P1
JP1
RTN
J3
P3
SENSOR
AIR
P11
PRESS
SENSOR
FUEL
P10
PRESS
SENSOR SAMPLE
P12
PRESS
BURNERGND
GLOW PLUG ASSEMBLY
J1
P1
OVEN
HEAT
FLAME O
P2
J2
POWER MODULE
ASSEMBLY
J2
J4
CATHODE
BURNER ASSEMBLY
PURGE AI
SWITCH ASSEM MANUAL IGNIT
LED INDICATOR
ASSEMBLY
PWR MOD
J1
J6
J1
J5 +10V REF
NODE
Model NGA2000 FID
Secondary Variab
Secondary Variable
Instruction Manual
760001-A
October 2002
SECTION 3

OPERATION

3-1 OVERVIEW

Prior to initial startup, the user should leak test the module as outlined in Section 2.
For the remainder of this section, Analyzer Module interconnection with a Platform or some interfacing component will be assumed. Display and Keypad information shall refer to that which the user can expect to see and do with regard to the Front Panel of the Platform.
(For a complete description of Platform Front Panel controls and indicators, see Section 1 of the Platform instruction manual.)

3-2 DISPLAYS

Three kinds of Display screens are available to the user:
Run Mode
Menu
Help

3-3 RUN MODE DISPLAY

The Run Mode is the normal mode of opera­tion. In this mode, the display (see Figure 3-1 below) will show current gas measurement, the component of interest, user-selectable (up to four) secondary variables, the current op­erations of the softkeys, and a graphic bar representing the displayed concentration as a percent of fullscale.
If more than one Analyzer Module is con­nected to the system, another Run Mode dis­play will show up to four gas measurements at once. Alarm messages may also appear on the display (See Table 3-1 on page 3-4).
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1

3-4 MENU DISPLAYS

The Main Menu structure enables the user to access data and functions, and put informa­tion onto the network.
The Main Menu (see Figure 3-2 on page 3-2) is subdivided into three levels of control based generally on which personnel is likely to use it: Basic Controls, Expert Controls, and Techni­cal Controls. (See Figure 3-3through Figure 3-5.) Many layers of the menu structure are described at appropriate places throughout this manual.
From the Run Mode display, press the MENUS softkey to gain access to the Main Menu. (See Figure 3-2 on page 3-2.)

3-5 HELP DISPLAYS

The Help structure is intended to be an on-line "tutorial," context-sensitive and topic-interconnected, so that the user can practically operate NGA2000 without need of an instruction manual. (See Figure 3-6 on page 3-3.)
23.2 ppm HC
0 ppm 50
Display Parms. Menu Dual Info
F1
Figure 3-1. Run Mode Display
le: XXX
: XXXX
F2 F3 F4 F5
Analyzer PQ 322-14
Instruction Manual
p
p
(
)
(
)
760001-A October 2002
Model NGA2000 FID
23.2 ppm HC Analyzer XXXXXXXX
Main Menu
Basic Controls
Expert controls and setup ...
Operational configuration
Technical level configuration ...
Diagnostic and manufacturing/service
Delete alarm message!
Display Parms. Info
F1
F2 F3 F4 F5
Figure 3-2. Main Menu Display
23.2 ppm HC Analyzer XXXXXXXX Basic Controls
Measurement range numbers:
Range upper limit: 10 ppm Range and functional control: Local Bypass sample flow: 1000 ml/min Ranges with valid calibration 1&2 Calibration status: Ready If it won’t calibrate… Flame condition: On Light flame…
Home Escape Zero Span Info
F1
F2 F3 F4 F5
Figure 3-3. Basic Controls Menu Display
23.2 ppm HC Analyzer XXXXXXXX
Ex
ert controls and set u
Expert analyzer controls ...
Auxiliary module controls ... System set up ... Analyzer module set up ... Auxiliary module set up ... Local I/O set up ...
Home Escape Info
F1
F2 F3 F4 F5
Figure 3-4. Expert Controls and Setup Menu Display
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
g
760001-A
Model NGA2000 FID
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3
23.2 ppm HC Analyzer XXXXXXXX Technical confi
System set up ...
Service menus... Diagnostic menus... Other module diagnostic menus... Listing of all modules...
uration menu
Home Escape Cal Info
F1
Figure 3-5. Technical Configuration Menu Display
23.2 ppm HC Analyzer XXXXXXXX
The Main Menu for the analyzer system. Note that this menu refers to the particular analyzer selected from the run screen, when used in a system. The softkey marked “HOME” will always return you to this screen. Help menu system... Help on help... Keyboard controls... Editing controls...
F2 F3 F4 F5
Main Menu Help
Home Escape Map
F1
F2 F3 F4 F5
Figure 3-6. Typical Help Screen
October 2002
Instruction Manual
760001-A October 2002
DISPLAY MESSAGE DESCRIPTION TYPE
AIR FET FID Air FET current WARNING
AIR PRESS FID Air Pressure WARNING
BAIR FLOW Burner Air Flow WARNING
BAROMETER System Barometer WARNING
BFUEL FLOW Burner Fuel Flow WARNING
BLOCK FET Heater current WARNING
CASE TEMP Case Temperature WARNING
CRUDE NOISE Calculated Noise WARNING
CURRENTRNGHI Current, High Range WARNING
CURRENTRNGLO Current, Low Range WARNING
CURRENTSFAC Current Range WARNING
FLAME TEMP Flame Temperature WARNING
FUEL PRES Fuel Pressure WARNING
LIN ERROR Linearizer Error WARNING
N15 VOLTS Power Supply -15V WARNING
P10 VOLTS Power Supply +10V REF WARNING
P15 VOLTS Power Supply +15V WARNING
POL VOLTS Polarizing Volts WARNING
SAMP PRES Sample Pressure WARNING
CALRESULT Calibration Error FAILURE
PURGE AIR FID Purge Air FAILURE
SW ERROR Software Error FAILURE
Model NGA2000 FID
Table 3-1. FID Analyzer Module Alarms
3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
)
)
)
Model NGA2000 FID
Figure 3-7. Typical Curves of Module Response vs. Pressure Setting on Sample Pressure Regulator
1.0
0.8
0.6
RESPONSE
(100 ppm CH
4 fullscale)
0.4
0.2
0
0
2
13.76
3
20.64
SAMPLE: 100 ppm CH4 in N2 FUEL: 25 psig (1726 hPa) H2 AIR: 25 psig (1726 hPa)
4
27.52
5
34.4
41.28
psig
RESPONSE
(100 ppm CH
4 fullscale)
1.0
0.8
0.6
0.4
0.2
0
SAMPLE PRESSURE
SAMPLE: 100 ppm CH4 in N2 at 5 psig (344 hPa)
0
5
344
10
688
15
1032
hPa
20
1376
AIR: 30 psig (2064 hPa
AIR: 20 psig (1376 hPa
AIR: 10 psig (688 hPa
25
1726
psig
FUEL PRESSURE
hPa
Figure 3-8. Typical Curves of Module Response vs. Pressure Setting on Fuel Pressure Regulator
October 2002
6
48.16
30
2064
760001-A
7
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5
Instruction Manual
psig (
760001-A October 2002
Model NGA2000 FID
RESPONSE
(100 ppm CH
4
fullscale)
Figure 3-9. Typical Curves of Module Response vs. Pressure Setting on Air Pressure Regulator
1.0
0.8
0.6
0.4
0.2
FUEL: 30
FUEL: 25 psig (1726 hPa) H
FUEL: 20 psig (1376 hPa) H
SAMPLE: 100 ppm CH
0
0
5
344
10
688
AIR PRESSURE
15
1032
2064 hPa) H
at 5 psig (344 hPa)
20
1376
psig
hPa
2
2
2
4
25
1726
in N
2
30
2064
3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

3-6 STARTUP PROCEDURE

Apply power to the FID Analyzer Module. If it is associated with a Platform, do this by plug­ging in the Platform to a power source. The Platform has no ON/OFF power button. Once power has been supplied to the Platform, the FID Analyzer Module will be energized.
If the user's system contains only one Ana­lyzer Module, all system components, the Controller Board and the network "self-install" (bind together) during initial startup. If the sys­tem contains more than one Analyzer Module, the startup sequence will interrogate the net­work to locate and identify all components on the network. The user will have to bind appro­priate combinations of components after the startup sequence (see Section 3-7 on page 3-
9).
After the warm-up period (about one hour for the FID Analyzer module), all modules are completely functional.
Before introducing gases to the FID, the user should check the general health of the ana­lyzer module's electronics by reviewing the results of its self test sequence. These test results can be found by selecting the following from the Main Menu: Technical Level Configu­ration, Diagnostic Menus, Analyzer Module Diagnostics, Self Test. All tested parameters should indicate "Pass."
Descriptions of the tests performed follow:
Instruction Manual
760001-A
October 2002
20 bit ADC test - Checks the 20-bit ADC on the Analysis Computer PCB by sending a DC signal through the Preamp PCB and reading the signal back with the 20-bit ADC.
12 bit ADC test - Checks the 12-bit ADC on the Analysis Computer PCB by sending a DC signal and reading the signal back with the 12-bit ADC.
Power Supply PCB test - Checks the presence of the Power Supply PCB by activating the 3-way air solenoid.
Safety PCB test - Checks the pres- ence of the Safety PCB by sending a command and reading it back.
Case temperature test - Compares the temperature read between the Preamp temperature sensor and the case temperature sensor. They must be within 10°C of each other. This test sometimes fails if the case is opened. The sensor in the Preamp will take longer to cool off since it is in an en­closure. Re-running the self-test after thermal equilibrium will produce a positive result if the sensors are work­ing properly.
The self-test can be repeated at any time by activating the TEST softkey in the Self Test Results menu.
EEPROM test - Checks the EEPROM on the Analysis Computer PCB.
EPROM test - Checks the EPROM on the Analysis Computer PCB.
RAM test - Checks the RAM on the Analysis Computer PCB.
Power supply test - Verifies that all internal DC voltages are within the re­quired tolerances.
Network test - Checks the internal network interface.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Set the Range Number (Expert Controls sub-menu) to the desired setting. Introduce all gases at suitable pressures (see Section 1-6 Specifications, on page 1-5).
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Check/set internal pressure regulators accord- ing to the following specifications:
Internal Pressure
Regulator
Burner Air 965 to 1103 hPa-gauge (14 to 16 psig) Fuel 1516 to 1723 hPa-gauge (22 to 25 psig) Sample (100% H2) 276 to 345 hPa-gauge (4 to 5 psig) Sample (Mixed Fuel) 207 to 345 hPa-gauge (3 to 5 psig)
Typical Operating Pressures
Purge air of the following specifications must be present:
Flow: 16 to 18 L/min.
Supply Pressure: 689 to 1378 hPa-gauge
(10 to 20 psig).
Noncompliance could cause damage to the module. At the very least, the module's safety system, which requires a certain volume of purge air flowing through the case before al­lowing burner ignition, will not allow the in­strument to operate. The lowest purge air flow/pressure setting possible during burner operation is preferable. Thus, the user should set the external purge air pressure initially at 689 hPa-gauge (10 psig). Check the Miscella­neous Control Parameters screen under Technical Diagnostics, and note whether the Purge Gas (switch) variable is "ON." If it is "OFF," increase purge air supply by 69 hPa-gauge (1 psig), and recheck the Purge Gas variable until it reads "ON." DO NOT EXCEED 1378 hPa-GAUGE (20 PSIG). If the maximum setting is reached, and the Purge Gas variable does not read "ON," contact fac­tory. If the safety system is initiated success­fully (Purge Gas variable is "ON"), continue with the remainder of the startup procedure.
either ignition procedure. The LEDs, when illu­minated, denote the following information:
Green - unit powered on
Amber - constant illumination indicates case temp. is within 5% of operating temp. setpoint (i.e., 50°C with burner off, 54°C with burner on);otherwise, LED will blink
Green - Flame on
Green - purge air system intact (it has filled five volumes of the module interior)
Auto-ignition provides fuel override and three attempted ignitions (default setting), if neces­sary. Manual ignition requires that the Plat­form front panel, if used, be disengaged.
CAUTION
HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is open. Dropping front panel while hand or fingers are inside either handle can cause serious injury.
NOTE
Do not restrict the PURGE OUT port and the pressure relief valve. They must be vented to atmospheric pressure.
Two methods of burner ignition are possible: auto-ignition and manual ignition. Note the four LEDs on the front panel of the Analyzer Module. They provide necessary information for
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
The manual ignition switch on the Analyzer Module front panel must be manipulated in the following ways:
1. Press up and hold for one minute. This opens burner fuel and air solenoids.
2. Press down to ignite burner glow plug for up to 10 seconds.
Model NGA2000 FID
3. Repeat as necessary (if fuel and air sources are farther away than 10 feet, several more attempts may be neces­sary).
4. Flame on is defined true when the flame temperature exceeds 115°C (239°F).
5. If the flame has been lit, but the flame temperature increases slowly, perform the following steps:
a. After igniting flame, release switch for
2 sec.
b. Press switch down for 2 sec.
c. Repeat release switch and press
down steps as necessary.

3-7 BINDING

To achieve full coordination between Analyzer Modules and associated I/O Modules, the user must bind those components together in the System Set Up portion of the Technical Configuration Menu in software.

3-8 CALIBRATION

See Section 2-4c on page 2-5 for a descrip­tion of the method for choosing calibration zero and span gases.
To calibrate the Analyzer Module, introduce zero gas into the SAMPLE INLET, and do the following:
1. If more than one Analyzer Module is func­tional and the split Run Mode display is shown, press the DISPLAY softkey until the desired Analyzer's Run Mode display is acquired.
2. Press the MENUS softkey to enter the Main Menu
3. Verify the fuel type in the Miscellaneous Control Parameters menu (under the Technical Configuration menu structure, select the following from the Main Menu Diagnostic menus, Analyzer Module Di-
.
Instruction Manual
760001-A
October 2002
agnostics and then Miscellaneous Control Parameters).
4. Verify the capillary type in the Analyzer Manufacturing Data menu (under the Technical Configuration menu structure, select the following from the Main Menu Technical Level Configuration, Service Menus, Manufacturing Data, Analyzer Module Data).
5. In the Calibration Gas List Main Menu, select Expert Controls and Setup, Analyzer Module Setup, then Cali­bration Gas List), enter necessary data, including the Operational Sample Pres­sure and the Calibration Gas HC Re­sponse Factor. Common HC factors are: methane (CH propane (C3H8), 3.00. These factors are not used to compensate the reading, but are used to select the proper preamp sense resistor.
6. Press HOME to re-enter the Main Menu enter the Basic Controls menu, introduce zero gas and allow its response to stabilize, press the ZERO softkey to enter the Ana­lyzer Zero menu, press ZERO again and wait.
7. Press the SPAN softkey to enter the Ana­lyzer Span menu, introduce span gas and allow its response to stabilize, press SPAN again and wait.
8. Press the HOME softkey to re-enter the Main Menu.
9. Press DISPLAY softkey for the Run Mode display.
If the user is unable to calibrate the Analyzer Module (i.e., when ZERO or SPAN is initiated, nothing happens), several possible solutions present themselves. One solution relates to the use of an incorrect gas for zeroing or spanning (e.g., using a high concentration gas to zero or a zero gas to span the Analyzer Module). Simply recalibrating with the appro-
:
priate gas(es) will not correct the problem be­cause the ZERO OFFSET or SPAN FACTOR
4), 1.0, ethane (C2H6), 1.90,
menu (from the
:
,
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
Instruction Manual
760001-A October 2002
has been set to an extreme value in the proc­ess.
To remedy the problem, do the following:
1. Verify that correct zero and span calibra­tion gases are being used properly. If so, attempt to recalibrate according to instruc­tions at the beginning of this section, en­suring that case temperature and displayed measurement reading are sta­ble before initiating the calibration routine. If incorrect gases were used in the initial, failed calibration, skip to Step 2.
2. Make the following selections from the Main Menu Analyzer Module Setup, then Calibration Parameters. Disable Calibration Adjust­ment Limits.
3. Recalibrate the analyzer module accord­ing to instructions at the beginning of this section, ensuring that case temperature and displayed measurement reading are stable before initiating the calibration rou­tine.
: Expert Controls and Setup,
Model NGA2000 FID
If maximum sensitivity is required from the FID Analyzer Module, use an optimum combination of settings on the SAMPLE, FUEL, and AIR pressure regulators. Settings must be determined experimentally, but the curves in Figure 3-7 on page 3-5, Figure 3-8 on page 3-5and Figure 3-9 on page 3-6 may be used as guides.
The Analyzer Module will not allow the user to increase the upper limit of a range beyond the "maximum range" software setting. To change the "maximum range" value, select the follow­ing from the Main Menu tion Menu, Service Menus, Manufacturing Data, and Analyzer Module Data. Select Maximum Range, and use the arrow keys to scroll the indicated value. The same applies for Minimum Range settings.
During shutdown, always turn off fuel gas first, then the air and sample gases. The flame can also be turned off by setting Ignition System Enable to "Off" in the Light Flame sequently, remember to set Ignition System Enable to "On" before attempting to ignite the flame.
: Technical Configura-
menu. Sub-
4. Enable Calibration Adjustment Limits in the Calibration Parameters menu.

3-9 ROUTINE OPERATION

After binding and calibration, proceed as fol­lows:
Supply sample gas to SAMPLE INLET. Adjust external flow controller or throttle valve so that flow discharged from the BYPASS outlet is between 0.5 and 2.0 L/min. The reading on the SAMPLE BYPASS pressure gauge should be the same as that used during adjustment of the span control. Adjust, if necessary.
Adjust the Range Number setting. The Ana­lyzer Module will now automatically and con­tinuously output the measured hydrocarbon content of the sample. Output is in terms of the particular hydrocarbon present in the span gas. Note that readings obtained during op­eration depend on the concentration of total hydrocarbons in the sample.
After initial startup, or startup following a pro­longed shutdown, the Analyzer Module re­quires about one day's continuous operation to stabilize. For several days afterwards, cali­brate daily. The frequency of subsequent cali­brations can be reduced as experience dictates, consistent with the accuracy re­quirements of the particular application.

3-10 SAFETY SYSTEM

The FID Analyzer Module safety system will not allow ignition or continuous burner func­tion unless the following conditions are pre­sent:
Internal purge gas pressure is at least 380 hPa-gauge (5.5 psig). (Monitor display message, Purge Gas Pressure in Physical Measurements menu, for the proper set­ting.)
Internal fuel pressure is less than 2064 hPa-gauge (30 psig). (Monitor dis­play messages, Fuel Supply Pressure in
3-10 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Physical Measurements menu and Fuel Pressure Status "ON" in Miscellaneous Control Parameters menu for proper set­tings. If the internal fuel pressure has ex­ceeded 2064 hPa-gauge [30 psig], check that the external pressure is less than 3450 hPa-gauge [50 psig]. The +24 VDC power to the Analyzer Module must be cy­cled to continue operation.)
Flow rate for purge air in is at least 16 L/min. and case pressure is greater than 0.5 inch of water. (Monitor display message, Purge Gas "ON" in Miscellane­ous Control Parameters menu for correct state. Proper sealing hardware must be used in order to obtain the required purge air in flow rate and case pressure.)
Five case volumes of purge air have been achieved and the three above conditions are present. The time duration to achieve a safe system is a minimum of 6 min. The
614
5
3
2
Figure 3-10. Front Panel Torque Sequence
Instruction Manual
elapsed time can be monitored in the Technical Startup Analyzer the Purge Air Green LED "ON," Purge Control Status "ON," or Purge Air Alarm for indication of the state of the safety sys­tem.)
As stated above, proper sealing hardware is crucial to the successful operation of the safety system. Therefore, a specific torque sequence (as shown in Figure 3-10 below) must be followed when the front panel of the module is being reinstalled after removal. All front and rear panel screws must be installed.
NOTE
Do not over-torque rear panel screws.
Torque Sequence:
Screw #1, 4 to 5 turns Screw #2, 4 to 5 turns Screw #3, 4 to 5 turns Screw #4, 4 to 5 turns Screw #5, 4 to 5 turns Screw #6, 4 to 5 turns
Repeat torque sequence until all screws are tight.
The gasket must fill in between the front panel plate and the enclosure.
760001-A
October 2002
menu. (Monitor
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-11
Instruction Manual
760001-A October 2002
Model NGA2000 FID
3-12 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

MAINTENANCE AND SERVICE

Instruction Manual
760001-A
October 2002
SECTION 4

4-1 BURNER DISASSEMBLY AND CLEANING

Disassemble the burner only if contaminants must be removed. Combustion products or other contaminants which accumulate inside the burner may form electrical leakage paths between the collector and the burner contact, resulting in noisy readings.
If the instrument is to be operated at the high­est sensitivity, traces of such contaminants can cause erroneous readings. For best per­formance, clean the burner per the following procedure:

Disassembly

1. Power OFF the module.
2. Shut fuel gas, air and sample gases OFF.
3. Slide Analyzer Module from the Platform
(if applicable).
CAUTION
HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is open. Dropping the panel while hand or fingers are inside either handle can cause serious injury.
7. Disconnect the exhaust tube.
8. Remove the burner assembly by loosen­ing the two securing nuts.
9. Unscrew the burner cap retainer ring, and remove burner cap.
10. Loosen clamp; lifting straight up, remove combustion chamber-chimney assembly from the manifold.
NOTE
If the old burner tip assembly is to be used again, do not touch it with bare hands or any materials likely to contaminate it with hydrocarbons, salt, etc. Additionally, all items used for cleaning (tweezers, swabs, etc.) must be absolutely free of contamina­tion.
11. Unscrew and remove burner tip assem­bly.

Cleaning

1. Clean chimney assembly, combustion chamber, and burner tip assembly with acetone or methyl ethyl ketone.
2. Follow cleaning with a distilled water wash.
4. On the combustion chamber, disconnect the polarizing voltage cable and amplifier input cable.
5. Disconnect the four connectors attached to the burner (one to glow plug, two to preamp board , and one to thermistor).
6. Disconnect the four gas supply tubes (two air, one fuel and one sample).
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-1

Reassembly

Using care not to touch any internal parts, re­assemble the burner per the following proce­dure.
1. Holding with clean tissue, screw burner tip assembly finger-tight into manifold.
2. Push combustion chamber/chimney as­sembly down onto manifold, taking care
Instruction Manual
760001-A October 2002
not to hit burner tip. Tighten clamp on combustion chamber chimney assembly.
3. Replace burner cap (and flameout sensor assembly), with flameout sensor assem­bly in-line with burner exhaust.
4. Install burner onto module base.
Model NGA2000 FID

4-3 BURNER STARTUP TROUBLESHOOTING

If unable to startup burner, check the follow­ing:
1. All supply gas cylinder pressures are within specifications (see Section 1-6 Specifications on page 1-5).
5. Connect all leads and tubing. Replace module cover and install module into Plat­form.
NOTE
Fittings using Teflon pipe seal tape: Re­place Teflon pipe seal tape to maintain a tightly sealed burner assembly.

4-2 COMPONENT REPLACEMENT

Components which may require replacement include; sample capillary, fan, fuses, EPROM, printed circuit boards and front panel LED's.
Replacement of each of these components require sliding out the bottom chassis from the enclosure.
Three fuses that may be replaced are located on; 1) the Network/Power Assembly, 2) the Safety Board and 3) the Power Supply Board. If the thermal fuse on the Safety Board or the Power Supply Board require replacement, this indicates a overheating condition, which de­notes a wider problem requiring troubleshoot­ing.
EPROM's should only be replaced by quali­fied electronics personnel because special tools and knowledge are required.
For accessibility when replacing printed circuit boards, the printed circuit board mounting panel has been designed to swing down.
2. Correct gases are being supplied to each back panel inlet.
3. Air, not zero gas (if nitrogen), is being supplied to burner.
4. Burner exhaust is being vented to atmos­pheric pressure, and is not tied to either purge air exhaust or another FID exhaust.
5. Burner exhaust continuously slopes downward until reaching atmospheric pressure vent.
6. Parameters "IS/WAS" match in the Physi­cal Measurement Parameters screen (see page 7-13); use NEXT softkey to view all parameters.
NOTE
As a last resort, check burner fittings for leaks (purge air pressure is higher than the pressure in any of the lines connected to the burner; thus, any leak would be into the fitting rather than out of it).
4-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
GLOW PLUG
Instruction Manual
760001-A
October 2002
FLAMEOUT SENSOR
O-RING 903227
SCREW 2-56 X 3/16 LOCK WASHER NO.2 FLAT WASHER NO. 2
JUMPER
CLAMP
TEE FITTING
MOUNTING BRACKET
SCREW (2) M3X0.5 x 6mm
Figure 4-1. FID Burner Assembly – Exploded View
BURNER CAP NUT
BURNER CAP
CHIMNEY
ADAPTER EXHAUST
O-RING 903227
COMBUSTION CHAMBER
BURNER TIP
O-RING 903256
BURNER FUEL RESTRICTOR
CAPILLARY
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-3
Instruction Manual
760001-A October 2002
Model NGA2000 FID
4-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

REPLACEMENT PARTS

WARNING
Instruction Manual
760001-A
October 2002
SECTION 5

c. Electro-Mechanical

PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair.

5-1 REPLACEMENT PARTS

a. Electronics

655588 Computer Board 655796 Preamp Board 655596 Safety Board 655764 Power Supply Board 657540 DC-DC Power Board 813344 Fuse 6A 903107 Fuse Thermal 657029 LED Indicator Assembly 657053 Switch Assembly 656026 Case Temperature Sensor 655468
DC-DC Converter (+5V, ±15V)

b. Burner Assembly and Components

655779 Burner Assembly, Mixed Fuel 655784 Burner Assembly, 100% Fuel
1
2
655803 Flameout Sensor 655801 Glow Plug 100194 Gasket, Connector - Sample 810156 Fitting, Tee 655782 Burner Cap 656217 Burner Manifold 656364 Chimney 630354 Exhaust Adapter 652995 Connector, Restrictor - Mixed Fuel 073616 Burner Tip 656365 Combustion Chamber 903227 O-Ring, Viton 903256 O-Ring, Viton
1
Select per application fuel. Restrictor and sample capillary
must display the same color code.
2
Select per application. The sample capillary displays a green
color marking. The restrictor displays a red color marking (mixed fuel).
902833 Fan 655760 Case Heater 656347 Purge Air Heater

d. O-Rings and Gaskets

011167 O-Ring, Buna-N (LED) 010177 O-Ring, Buna-N (Switch) 903124 O-Ring, Buna-N (1/4" Bulkhead) 008025 O-Ring, Buna-N (Fuel, Air
Regulator, 3/8" Bulkhead)
656167 Front Panel Gasket

e. Pneumatics

657448 Kit, Capillary - Mixed Fuel Low
Range
657777 Kit, Capillary - Mixed Fuel High
Range
657776 Kit, Capillary - 100% Fuel Low
Range 017154 Filter, Burner Air & Sample 902832 Regulator, Fuel and Burner Air 655269 Regulator, Sample - Brass 655513 Regulator, Sample - Stainless
steel 871672 Regulator, Purge Air 615598 Restrictor, Bulkhead - Mixed 656890 Restrictor, Bulkhead - 100%
Fuel 655794 Restrictor, Air Ignite 656888 Restrictor, Air Measurement 630669 Restrictor, Burner - Mixed Fuel 630662 Restrictor, Burner - 100% Fuel 902931 Sensor, Flow 400- 2000 ml/min. 656444 Sensor, Fuel Pressure 656443 Sensor, Burner Air Pressure 656418 Sensor, Sample Pressure 656218 Solenoid - Fuel 656219 Solenoid, 3-Way - Air 657126 Switch, Flow - Purge Air 903690 Switch, Pressure - Purge Air 903128 Valve, Pressure Relief
1
2
1
1
1
1
1
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 5-1
Instruction Manual
760001-A October 2002

5-2 MATRIX

Each analyzer is configured per the customer sales order. Below is the FID sales matrix which lists the various configurations avail­able.
FID FID HYDROCARBON ANALYZER
Model NGA2000 FID
To identify the configuration of an analyzer, locate the analyzer name-rating plate. The sales matrix identifier number appears on the analyzer name-rating plate.
01 Current Software
02 2.2.2 Version Software
03 3X Version Software - specify version
Code Software Version
A Mixed Fuel, Calibrated Ranges 0-10, 0-25, 0-100, 0-250 ppm
B Mixed Fuel, Calibrated Ranges 0-10, 0-30, 0-100, 0-300 ppm
C Mixed Fuel, Calibrated Ranges 0-4, 0-10, 0-40, 0-100 ppm
D Hydrogen Fuel, Calibrated Ranges 0-1, 0-2.5, 0-10, 0-25 ppm
E Mixed Fuel, Calibrated Ranges 0-250, 0-1000, 0-2500, 0-10000 ppm
F Mixed Fuel, Calibrated Ranges 0-300, 0-1000, 0-3000, 0-10000 ppm
G Mixed Fuel, Calibrated Ranges 1000, 2500, 10000, 25000
9 Special Calibration Ranges
1 Brass and Neoprene Back Pressure Regulator 5 psig
2 Brass and Neoprene Back Pressure Regulator 2 psig
3 Stainless Steel and Viton Back Pressure Regulator 5 psig
4 Stainless Steel and Viton Back Pressure Regulator 2 psig
9 Special
Code Configurations
Code Materials
01 Standard (no flow sensor)
02 Integral Flow Sensor (400 – 2000 cc/min.)
99 Special
00 None
01 Customer Option
99 Special
FID 01 A1 01 00 Example
Code Flow Path
Code Special Requirements
5-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID

RETURN OF MATERIAL

6-1 RETURN OF MATERIAL

If factory repair of defective equipment is re­quired, proceed as follows:
1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
2. Rosemount CSC will provide the shipping address for your instrument.
3. In no event will Rosemount be responsi­ble for equipment returned without proper authorization and identification.
4. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping.
5. In a cover letter, describe completely:
The symptoms that determined the
equipment is faulty.
The environment in which the equip-
ment was operating (housing, weather, vibration, dust, etc.).
Site from where the equipment was
removed.
Whether warranty or non-warranty
service is expected.
Complete shipping instructions for the
return of the equipment.
6. Enclose a cover letter and purchase order and ship the defective equipment accord­ing to instructions provided in the Rose­mount Return Authorization, prepaid, to the address provided by Rosemount CSC.
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
Instruction Manual
760001-A
October 2002
SECTION 6
If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the condi­tions listed in the standard Rosemount war­ranty, the defective unit will be repaired or replaced at Rosemount’s option, and an operating unit will be returned to the customer in accordance with the shipping instructions furnished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.

6-2 CUSTOMER SERVICE

For order administration, replacement Parts, application assistance, on-site or factory re­pair, service or maintenance contract informa­tion, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076

6-3 TRAINING

A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service Training Schedule contact the Technical Ser- vices Department at:
Rosemount Analytical Inc.
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 6-1
Instruction Manual
760001-A October 2002
Model NGA2000 FID
6-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
APPENDIX A – MENU SCREENS
Instruction Manual
760001-A
October 2002
SECTION 7
Menu: 0 ANALOP
Menu: 1 ANALSET
Menu: 2 EXP_CAL
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-1
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 3 FLOCHEK
Menu: 4 ZEROI1
Menu: 5 FLOCHEKI1
7-2 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 6 ANALOPI1
Menu: 7 ACALSET
Menu: 8 APARLST
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-3
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 9 ANALSETI1
Menu: 10 CALLISTI1
Menu: 11 ACALSETI1
7-4 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 12 APARLSTI1
Menu: 13 AMMAN
Menu: 14 AMMANI1
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-5
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 15 AMSVC
Menu: 16 AMSVCI1
Menu: 17 ADIAG
7-6 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 18 AMPWR
Menu: 19 AM1V
Menu: 20 AMTEMP
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-7
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 21 AMMISC
Menu: 22 AMTREND
Menu: 23 ADIAGI1
7-8 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 24 RANGESETAM
Menu: 25 RANGESSETI1
Menu: 26 LINRANGE1
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-9
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 27 LINRANGE2
Menu: 28 LINRANGE3
Menu: 29 LINRANGE4
7-10 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 30 LINRANGE0
Menu: 31 AMPWRI1
Menu: 32 FLOCHEK1I1
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-11
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 33 FILTER
Menu: 34 AM1VI1
Menu: 35 AMTEMPI1
7-12 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 36 AM2VA
Menu: 37 PLIMITSA
Menu: 38 TLIMITSA
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-13
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 39 AMMISCI1
Menu: 40 ANALSIMPLE
Menu: 41 FILTERI1
7-14 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 42 LINRANGE0I1
Menu: 43 PLIMITSAI1
Menu: 44 CALFACTORS
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-15
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 45 R1FACTORS
Menu: 46 RN2FACTORS
Menu: 47 RN3FACTORS
7-16 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 48 RN4FACTORS
Menu: 49 RFACTORSI
Menu: 50 AMHELPINDEX
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-17
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 51 LINRANGE1I1
Menu: 52 AMTRENDI1
Menu: 53 CALFACTORSI1
7-18 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 54 APARLST2
Menu: 55 APARLST4
Menu: 56 APARLST5
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-19
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 57 APARLST6
Menu: 58 AMMISC2
Menu: 59 LIGHTFLAME
7-20 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 60 IGNITION
Menu: 61 LISTNOTES
Menu: 62 LIGHTFLAMEI1
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-21
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 63 AUTOFLAME
Menu: 64 AUTOFLAMEI1
Menu: 65 DISPLAY
7-22 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 66 MPARMS
Menu: 67 MPARMS2
Menu: 68 ANALSET2
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-23
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 69 MPARMSI1
Menu: 70 TLIMITSAI1
Menu: 71 IGNITIONI1
7-24 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 72 SELFTEST
Menu: 73 AMHELPINDEX2
Menu: 74 SOFT_DIAG
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-25
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 75 CALI1
Menu: 76 AM2VC
Menu: 77 AM2VD
7-26 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 78 RFHIST
Menu: 79 RFHIST2
Menu: 80 RFHIST3
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-27
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 81 RFHIST4
Menu: 82 LINFUNCT
Menu: 83 POLYSETUP
7-28 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 84 MIDPOINT1
Menu: 85 ANALSETI2
Menu: 86 POLYGAS1
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-29
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 87 POLYSETI1
Menu: 88 POLYGAS2
Menu: 89 POLYGAS3
7-30 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 90 POLYGAS4
Menu: 91 POLYGAS5
Menu: 92 POLYGAS6
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-31
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 93 POLYGAS7
Menu: 94 POLYGAS8
Menu: 95 POLYGAS9
7-32 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 96 POLYGAS0
Menu: 97 MIDPOINT2
Menu: 98 MIDPOINT3
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-33
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 99 MIDPOINT4
Menu: 100 ZEROI2
Menu: 101 EXP_CAL_DATA
7-34 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 102 EXP_CAL_DATI
Menu: 103 UNITS
Menu: 104 UNITSI1
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-35
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 105 POLYSETI2
Menu: 106 ANALSETI3
Menu: 107 POLYSETI3
7-36 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 108 RESET
Menu: 109 STORE
Menu: 110 REBOOT
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-37
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 111 SW_DIAGI1
Menu: 112 ZERO_NOW2
Menu: 113 SPAN_NOW2
7-38 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Instruction Manual
760001-A
October 2002
Menu: 114 CALLIST
Menu: 115 ABOUT
Menu: 116 ABOUT1
Rosemount Analytical Inc. A Division of Emerson Process Management Menu Screens 7-39
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Menu: 117 ALARM1
7-40 Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
SECTION 8
12 bit ADC test, 3-7 20 bit ADC test, 3-7
A
air
connections, 2-4
Analyzer Module interconnection with a Platform, 3-1, 3-
12, 3-12
B
back pressure, 2-4, 2-8 Basic Controls, 3-1, 3-12 burner
auto-ignition, 3-8 auto-ignition default, 3-8 cleaning, 4-1 contaminants, 4-1 disassembly, 4-1 electrical leakage, 4-1 erroneous readings, 4-1 maintaining seal integrity, 4-2 manual ignition, 3-8 manual-ignition switch, 3-8 reassembly, 4-1
unable to start, 4-2 Burner Air, 2-5, 2-8 BURNER EXHAUST OUT, 2-4 burner fuel
connections, 2-4 burner fuel determining type, 1-3 burner tip, 4-1 bypass
connections, 2-4
C
Calibration Gases, 2-5, 2-8 Case temperature test, 3-7 Controller Board, 3-7
D
dewpoint range, 2-2 drift, 1-3, 2-4

INDEX

E
EEPROM test, 3-7 electrical connections, 2-6, 2-8 EPROM
exhaust
Exhaust tubing, 2-4 Expert Controls, 3-1 external 24 VDC power source, 2-6 external gas tubing, 2-4
F
fan
flame output signal, 1-3 F Flow rate for purge air, 3-11 front panel LED's
fuel gas
Fuel Gas, 2-5, 1-8 fuel, air, cal and purge air installation location, 2-2, 3-12,
fuses
I
I/O Module, 2-2 I/O Modules, 7, 2-1, 3-9 initial startup, 2-4, 2-8 Installation location, 2-2 INSTALLATION RESTRICTIONS, 2-2, 2-8 Internal fuel pressure, 3-10 internal pressure settings, 2-6 Internal purge gas pressure, 3-10
Instruction Manual
760001-A
October 2002
replacement, 4-2 test, 3-7
connections, 2-4
replacement, 4-2, 2-8, 1-1
LOW RATE, 2-5
replacement, 4-2
option, 1-3 standard, 1-3
3-12, 3-12
location, 4-2 replacement, 4-2
Rosemount Analytical Inc. A Division of Emerson Process Management Index 8-1
Instruction Manual
760001-A October 2002
M
maximum range software setting, 3-10 maximum sensitivity required, 3-10, 3-12 Menu screen
Analyzer diagnostics (info), 7-8
Analyzer diagnostics (miscellaneous control
parameters), 7-8 Analyzer diagnostics (power supply voltages info), 7-11 Analyzer diagnostics (power supply voltages), 7-7 Analyzer diagnostics (Primary variable parameters), 7-7 Analyzer diagnostics (temperature control), 7-7 Analyzer diagnostics (trend display control), 7-8 Analyzer diagnostics, 7-6 Analyzer manufacturing data (info), 7-5 Analyzer manufacturing data, 7-5 Analyzer module help (flame ionization detector), 7-17 Analyzer module help, 7-25 Analyzer module service history (info), 7-6 Analyzer module service history, 7-6 Analyzer module service notes, 7-21 Analyzer module setup (info), 7-4 Analyzer module setup, 7-1 Analyzer module version information, 7-39 Analyzer parameter list (calibration gases), 7-20 Analyzer parameter list (calibration parameters), 7-19 Analyzer parameter list (info), 7-5 Analyzer parameter list (miscellaneous control
parameters), 7-20 Analyzer parameter list (primary variable parameters)
(more), 7-19 Analyzer parameter list (primary variable parameters),
7-19 Analyzer parameter list, 7-3 Analyzer span, 7-38 Analyzer starting up (info), 7-22 Analyzer starting up, 7-22 Analyzer zero, 7-38 Auto ignition parameters (info), 7-24 Auto ignition parameters, 7-21 Basic controls, 7-14 Calculated physical parameters, 7-26 Calibration factors (info), 7-18 Calibration factors, 7-15 Calibration gas list (info), 7-4 Calibration gas list, 7-39 Calibration info, 7-26 Calibration parameters, 7-3, 7-4 Concentration alarm setup, 7-40, 3-12 Current measurement parameters (more), 7-23 Current measurement parameters help, 7-24 Current measurement parameters, 7-23 Displayed parameters, 7-22 Expert controls, 7-1 Filter and delay parameters, 7-14 Gas concentrations (point 1, point 2), 7-29 Gas concentrations (point 3, point 4), 7-30 Gas concentrations (point 5, point 6), 7-30 Gas concentrations (point 7, point 8), 7-31
Model NGA2000 FID
Gas concentrations (point 9, point 10), 7-31 Gas concentrations (point 11, point 12), 7-31 Gas concentrations (point 13, point 14), 7-32 Gas concentrations (point 15, point 16), 7-32 Gas concentrations (point 17, point 18), 7-32 Gas concentrations (point 19, point 20), 7-33 Gas measurement parameters, 7-23 Help, 7-34 Light flame, 7-20 Light flame (info), 7-21 Linearity coefficients (curve 1), 7-9 Linearity coefficients (curve 2), 7-10 Linearity coefficients (curve 3), 7-10 Linearity coefficients (curve 4), 7-10 Linearity coefficients, 7-18 Linearization functions, 7-28 Linearization parameters (info), 7-15 Linearization parameters, 7-11 Manufacturer information, 7-39 Measurement function help, 7-3 Midpoint correction setup (info), 7-29 Midpoint correction setup (info), 7-36 Midpoint correction setup (range 1), 7-29 Midpoint correction setup (range 2), 7-33 Midpoint correction setup (range 3), 7-33 Midpoint correction setup (range 4), 7-34 Miscellaneous control parameters (info), 7-14 Physical measurement parameters, 7-13 Physical measurement parameters, 7-26 Physical measurements (info), 7-11 Physical measurements (info), 7-2 Physical measurements, 7-2 Polynomial setup (info), 7-30 Polynomial setup (info), 7-36 Polynomial setup (info), 7-36 Polynomial setup, 7-28 Pressure limits (info), 7-15 Pressure limits, 7-13 Primary variable parameters, 7-12 Range 1 factors(manufacturer’s settings), 7-27 Range 1 factors, 7-16 Range 2 factors(manufacturer’s settings), 7-27 Range 2 factors, 7-16 Range 3 factors(manufacturer’s settings), 7-27 Range 3 factors, 7-16 Range 4 factors (manufacturer’s settings), 7-28 Range 4 factors, 7-17 Range factors (info), 7-17 Range settings (info), 7-9 Range settings, 7-9 Re-initialize the analyzer, 7-37 Reset, 7-37 Response time/delay parameters, 7-12 Self test results, 7-25 Software diagnostics (info), 7-38 Software diagnostics, 7-25 Store historical data, 7-37
8-2 Index Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 FID
Temperature control (info), 7-12 Temperature limits (info), 7-24 Temperature limits, 7-13 Trend display control (info), 7-18 Units (info), 7-35 Units, 7-35 Zero/span calibration help, 7-2 Zero/span calibration, 7-1 Zero/span diagnostic data (info), 7-35 Zero/span diagnostic data, 7-34
N
network
access data, 3-1 access functions, 3-1 enter data, 3-1
network cable, 2-2 NETWORK connectors, 2-6 Network test, 3-7 noisy readings, 2-4
Instruction Manual
760001-A
October 2002
R
RAM test, 3-7 reduce the effect of internal oxygen, 1-3 Return Authorization, 6-1
S
Safety PCB test, 3-7 safety system, 1-3, 2-2, 2-6, 3-8, 3-10, 3-11 sample
connections, 2-4
sample capillary
replacement, 4-2, 2-8, 2-8
S
AMPLE GAS, 2-5, 2-8
sample is toxic or noxious, 2-4, 2-8 self test sequence, 3-7 shutdown, 3-10, 3-12 SPAN GAS, 2-5, 2-8, 3-12, 3-12 system contains more than one Analyzer Module, 3-7 system contains only one Analyzer Module, 3-7
O
Operating ambient temperature, 2-2, 2-8
P
Platform, P-7, P-8, 1-1, 1-6, 1-7, 2-1, 2-2, 2-6, 3-1, 3-7, 3-
8, 3-12, 4-1, 4-2
power cable, 2-2 Power Supply PCB test, 3-7 Power supply test, 3-7, 3-12, 3-12
RESSURIZATION/FILTRATION, 2-5
P printed circuit boards
replacement, 4-2
prolonged shutdown, 2-4 proper sealing hardware, 3-11 purge
connections, 2-4
Purge air
specifications
flow, 3-8
Purge Air, 2-5 Purge air specifications
supply pressure, 3-8
purge system, 2, 3, 1-3, 2-5
T
Technical Controls, 3-1 Teflon pipe seal tape, 4-2, 3-12 the PURGE AIR OUT port, 2-4 torque sequence, 3-11 Typical flow rates with premixed fuel, 1-4
U
unable to calibrate, 3-9
W
warm-up period, 3-7
Z
ZERO GAS, 2-5, 2-8
Rosemount Analytical Inc. A Division of Emerson Process Management Index 8-3
Instruction Manual
760001-A October 2002
Model NGA2000 FID
8-4 Index Rosemount Analytical Inc. A Division of Emerson Process Management

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy.
DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure.
Seller's direct control.
Seller shall not be liable for failure to perform due to labor strikes or acts beyond
SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY
Instruction Manual
760001-A October 2002
Model NGA2000 FID
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.
1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
EUROPEAN TECHNOLOGY CENTER Fisher-Rosemount GmbH & Co.
Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd.
Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount
Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...