Rosemount NGA 2000 FID2 Hydrocarbon Analyzer Module SW 3.3-Rev E Manuals & Guides

Page 1
Instruction Manual
748412-E March 2003
Model NGA2000 FID2
Flame Ionization Detector Module
http://www.processanalytic.com
Page 2

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and in­ternational standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
to ensure they continue to operate within their
Read all instructions
prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical repre- sentative for clarification.
Follow all warnings, cautions, and instructions
marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and mainte- nance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate In- struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel
to install, operate, update, program,
and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s per-
formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY
.
Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc. Kynar® is a registered trademark of Pennwalt, Inc.
Emerson Process Management Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Page 3
Model NGA2000 FID2

TABLE OF CONTENTS

Instruction Manual
748412-E
March 2003
PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-5
Documentation....................................................................................................................................P-6
Compliances .......................................................................................................................................P-6
Glossary of Terms ............................................................................................................................P-7
Quick Startup Procedure ....................................................................................................................P-9
1.0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Overview................................................................................................................................1-1
1-2 Typical Applications...............................................................................................................1-1
1-3 Theory of Technology............................................................................................................1-2
1-4 Gas Safety Features..............................................................................................................1-5
1-5 Fuel Gas Option ....................................................................................................................1-5
1-6 Specifications.........................................................................................................................1-7
a. General ...........................................................................................................................1-7
b. Physical...........................................................................................................................1-7
c. Gas Requirements ..........................................................................................................1-8
d. Gas Connections.............................................................................................................1-8
2.0 INSTALLATION ....................................................................................................................2-1
2-1 Unpacking..............................................................................................................................2-1
2-2 Assembly ...............................................................................................................................2-1
2-3 Location .................................................................................................................................2-1
2-4 Gases ....................................................................................................................................2-1
a. Overview .........................................................................................................................2-1
b. Connections ....................................................................................................................2-2
c. Gas Specifications ..........................................................................................................2-2
2-5 Electrical Connections ...........................................................................................................2-5
2-6 Analytical Leak Check ...........................................................................................................2-6
a. Flow Indicator Method.....................................................................................................2-6
b. Manometer Method.........................................................................................................2-6
c. Troubleshooting Leaks....................................................................................................2-6
2-7 Installation Guidelines ...........................................................................................................2-7
3.0 OPERATION .........................................................................................................................3-1
3-1 Overview................................................................................................................................3-1
3-2 Displays .................................................................................................................................3-1
a. Run Mode Display...........................................................................................................3-1
b. Menu Displays.................................................................................................................3-1
c. Help Displays ..................................................................................................................3-3
3-3 Startup Procedure .................................................................................................................3-4
3-4 Optimization Procedure .........................................................................................................3-7
3-5 Binding...................................................................................................................................3-8
3-6 Calibration..............................................................................................................................3-8
3-7 Calibration Details .................................................................................................................3-10
3-8 Routine Operation .................................................................................................................3-11
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Page 4
Instruction Manual
748412-E March 2003
3-9 Shut Down Procedure ...........................................................................................................3-12
3-10 Safety System .......................................................................................................................3-12
3-11 Alarm Indications ...................................................................................................................3-13
3-12 Configuration Storage............................................................................................................3-14
4.0 MAINTENANCE AND SERVICE ..........................................................................................4-1
4-1 General ..................................................................................................................................4-1
4-2 Overview................................................................................................................................4-1
4-3 Disassembly ..........................................................................................................................4-2
4-4 Fuses .....................................................................................................................................4-2
4-5 Burner Block Removal And Installation .................................................................................4-4
4-6 Burner Startup And Troubleshooting.....................................................................................4-4
4-7 Maintenance Schedule ..........................................................................................................4-8
5.0 REPLACEMENT PARTS......................................................................................................5-1
5-1 Matrix .....................................................................................................................................5-1
5-2 Replacement Parts ................................................................................................................5-2
6.0 RETURN OF MATERIAL......................................................................................................6-1
6-1 Return Of Material .................................................................................................................6-1
6-2 Customer Service ..................................................................................................................6-1
6-3 Training..................................................................................................................................6-1
Model NGA2000 FID2
7.0 APPENDIX A. MENU SCREENS.........................................................................................7-1
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Page 5
Model NGA2000 FID2

LIST OF ILLUSTRATIONS

Figure 1-1. Flame Ionization Detection Technology ................................................................ 1-1
Figure 1-2. FID2 Analyzer Flow Diagram................................................................................. 1-3
Figure 1-3. FID2 Wiring Diagram ............................................................................................. 1-4
Figure 2-1. FID2 Outline and Mounting Dimensions................................................................ 2-3
Figure 2-2. FID2 Rear Panel .................................................................................................... 2-4
Figure 2-3. FID2 Front Panel .................................................................................................. 2-5
Figure 2-4. Leak Test - Flow Indicator Method ........................................................................ 2-6
Figure 2-5. Leak Test - Manometer Method............................................................................. 2-6
Figure 3-1. Run Mode Display ................................................................................................. 3-1
Figure 3-2. Main Menu ............................................................................................................. 3-1
Figure 3-3. Basic Controls Menu.............................................................................................. 3-2
Figure 3-4. Expert Controls Menu ............................................................................................ 3-2
Figure 3-5. Analyzer Module Setup Menu................................................................................ 3-2
Figure 3-6. Typical Help Menu ................................................................................................. 3-3
Figure 3-7. Analyzer Diagnostics Menu ................................................................................... 3-4
Figure 3-8. Self Test Results Menu.......................................................................................... 3-4
Figure 3-9. Light Flame Menu .................................................................................................. 3-5
Figure 3-10. Typical Module Response vs. Sample Pressure Setting....................................... 3-6
Figure 3-11. Typical Module Response vs. Fuel Pressure Setting ............................................ 3-6
Figure 3-12. Typical Curves of Module Response vs. Pressure Setting on Air Pressure
Regulator ............................................................................................................... 3-7
Figure 3-13. Typical Calibration Gas List Menu......................................................................... 3-8
Figure 3-14. Zero and Span Calibration Menu........................................................................... 3-8
Figure 3-15. Calibration Parameters Menu................................................................................ 3-8
Figure 3-16. Zero/Span Diagnostic Data.................................................................................... 3-8
Figure 3-17. Analyzer Manufacturing Data Menu .................................................................... 3-11
Figure 3-18. Store/Restore User Settings Menu...................................................................... 3-14
Figure 3-19. Store Historical Data Menu.................................................................................. 3-14
Figure 4-1. Location of Major Components.............................................................................. 4-1
Figure 4-2. Removal of FID2 Cover ......................................................................................... 4-2
Figure 4-3. Main Power Fuse Location .................................................................................... 4-3
Figure 4-4. Fuse Locations on Module Board.......................................................................... 4-3
Figure 4-5. Physical Measurements Menu............................................................................... 4-4
Figure 4-6. Removal of Oven Cover ........................................................................................ 4-5
Figure 4-7. FID2 – Exploded View ........................................................................................... 4-6
Figure 4-8. Burner Block -Exploded View ................................................................................ 4-7
Figure 4-9. Burner .................................................................................................................... 4-8
Instruction Manual
748412-E
March 2003

LIST OF TABLES

Table 1-1. Gas Flow Rates ..................................................................................................... 1-5
Table 1-2. Analyzer Characteristics Relative to Fuel Gas ...................................................... 1-6
Table 2-1. Gas Supply Pressures ........................................................................................... 2-4
Table 3-1. FID2 Analyzer Module Alarms............................................................................. 3-13
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Page 6
Instruction Manual
748412-E March 2003
Model NGA2000 FID2
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Page 7
Model NGA2000 FID2
The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the NGA2000 FID2 and the System Accessories of the NGA 2000 System.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
Instruction Manual
748412-E
March 2003

PREFACE

DEFINITIONS

DANGER .
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Page 8
Instruction Manual
748412-E March 2003
Model NGA2000 FID2

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be im­paired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel. For safety and proper per­formance this instrument must be connected to a properly grounded three-wire source of power.
WARNING .
POSSIBLE EXPLOSION HAZARD
This equipment is used in the analysis of sample gases which may be flammable, and the burner fuel used in the ionization process IS flammable. A system of intrinsically safe electronics and an explosion proof tower are used to prevent any ignition of a flammable gas leak. For this to be effec­tive, the module MUST be placed in a well-ventilated area, with unobstructed air flow around it.
DO NOT place it within another enclosure without assuring this ventilation. DO NOT obstruct the vent holes on the top and sides of the module. DO NOT place the FID module within another enclosure unless the latter has a guaranteed air circu-
lation such as to dilute a worst case fuel or sample leak below 25% of the LEL. Doing so will negate the safety features and may result in an explosion, serious injury, property damage and death.
WARNING .
FLAMMABLE SAMPLES
Consult the factory if flammable samples will be measured.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 9
Model NGA2000 FID2
Tampering or unauthorized substitution of components may adversely affect safety of this produc t. Use only factory documented components for repair.
Do not place hands or fingers in the Platform front handles when front panel is open. Dropping the front panel of the Platform while hand or fingers are inside either handle can cause serious injury.
Ensure that all gas connections are made as labeled and are leak free. Improper gas connections could result in explosion or death.
Instruction Manual
748412-E
March 2003
WARNING.
PARTS INTEGRITY
WARNING.
HAND INJURY HAZARD
WARNING.
POSSIBLE EXPLOSION HAZARD
WARNING.
POSSIBLE EXPLOSION HAZARD
Protection against explosion depends upon a special fuel flow restrictor in the fuel inlet fitting. DO NOT REMOVE THE FUEL INLET RESTRICTOR. Use the correct fuel flow restrictor for the fuel being used. Do not use 100% hydrogen fuel in a 40% H2/60% He configured FID module. Replace with factory supplied fitting only.
CAUTION .
PRESSURIZED GAS
This module requires periodic use of pressurized gas. See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5.
CAUTION .
OVERBALANCE HAZARD
This analyzer module may tip instrument over if it is pulled out too far and the Platform is not prop­erly supported.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Page 10
Instruction Manual
748412-E March 2003
Model NGA2000 FID2
CAUTION.
CONTROLLED ENVIRONMENT
This equipment is for use in a controlled environment. Refer to Section 1-6 Specifications (page 1-
7) in this manual for environmental conditions.
CAUTION.
HOT OVEN COMPONENTS
The oven and sample manifold are controlled to 80°C. Allow the analyzer to cool down before touching any of these components.
CAUTION.
OVER-VOLTAGE SPIKING
If this analyzer module is used with a non-Rosemount Analytical power supply, adding Rosemount Analytical PN 903341 Current Protector in series with the 24V positive power line will prevent over­voltage spiking and resultant fuse blowing when powering up the instrument.
CAUTION .
STATIC ELECTRICITY
Circuit boards in this instrument are static-sensitive. Take all static precautions when handling the circuit boards
NOTE
This Analyzer Module is completely leak-tested at the factory for gas leakage. The user is responsi­ble for testing for leakage at the inlet and outlet fittings on the rear panel (with a test procedure chosen by the user). The user is also responsible for leak-testing periodically and if any internal pneumatic components are adjusted or replaced. See leak test instructions in Section 2-6 on page 2-6.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 11
Instruction Manual
748412-E
Model NGA2000 FID2
March 2003
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against ex­tremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are preva­lent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125 should never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Page 12
Instruction Manual
9
6
748412-E March 2003
Model NGA2000 FID2

DOCUMENTATION

The following NGA2000 FID2 instruction materials are available. Contact Customer Service Center or the local representative to order.
748412 Instruction Manual (this document)

COMPLIANCES

This product may carry approvals from several certifying agencies, like The Canadian Standards Associa­tion (CSA), which is also an OSHA accredited Nationally Recognized Testing Laboratory (NRTL), and LCIE
- a French Notified Body.
The certification marks appear on the product name-rating plate.
®
NRTL /C
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May 1993).
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand.
LCIE 98 ATEX 6004 X EEx d ib IIB (+H 0°C Ta +40°C Date of Manufacture:
0081
NAMUR
) T6
2
II 2 G
N
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 13
Model NGA2000 FID2

GLOSSARY OF TERMS

Analyzer Module
The module that contains all sensor/detector components for development of a Primary Variable signal; in­cludes all signal conditioning and temperature control circuitry.
Backplane
The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module power and net­work cables, I/O Modules and Expansion Modules plug into.
Control Module
The Operator Interface plus the Controller Board.
Controller Board
The computer board that serves as the Network Manager and operates the Display and Keypad.
Distribution Assembly
Instruction Manual
748412-E
March 2003
The Backplane and the card cages that hold I/O and Expansion Modules.
Expansion Module
A circuit board that plugs into the Backplane from the front of the Platform and performs special features not related to I/O functions.
I/O Module
A circuit board that plugs into the Backplane from the rear of the Platform. Has a connector terminal for communication with external data acquisition devices and provides an input/output function.
Operator Interface
The Display and Keyboard.
Platform
Any workable collection of the following: Controller Board, Power Supply, Distribution Assembly, Enclosure and Operator Interface.
Power Supply
Any of a variety of components that provides conditioned power to other NGA 2000 components, from the Power Supply Board that plugs into the front of the Backplane in a stand-alone instrument to several larger ones that can power larger collections of modules and components.
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
Data placed on the network by a module regarding current status, e.g., sample flow, source voltage and other diagnostic information.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7
Page 14
Instruction Manual
748412-E March 2003
Model NGA2000 FID2
Softkeys
The five function keys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software.
System
Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s).
P-8 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 15
Model NGA2000 FID2

QUICK STARTUP PROCEDURE

Instruction Manual
748412-E
March 2003
The purpose of this reference guide is to pro­vide a easy to follow, step-by-step procedure through initial start-up an ignition of the FID2 Analyzer Module. This procedure assumes that the customer has already made all nec­essary electrical and gas connections and es­tablished the proper network connections.
1. Turn on power to the instrument. The #1 LED (POWER) will illuminate. The #3 LED (BLOCK) will begin flashing.
2. If sample gas has been connected and the sample pressure to the analyzer is sufficient to provide and accurate read­ing, the #4 LED (SAMPLE) will be illu­minated.
3. Allow the analyzer module to warm-up and the burner block temperature to reach the proper minimum ignition tem­perature (50°C). When the burner block temperature reaches the mini­mum ignition temperature, the #5 LED (IGNITE OK) will come on.
4. The instrument is now ready to be lit. Lighting the burner can be conducted in one of two methods: a) manual ignition from the front panel of the Analyzer Module or b) auto-ignite from the Plat­form.
a. To light the instrument from the
Analyzer Module, hold the “FUEL OVERRIDE/IGNITE” switch (located to the left of the indicator lights) in the up (FUEL OVERRIDE) position for 30 seconds. Immediately move the switch to the down (IGNITE) po­sition. The “IGNITE” mode is auto­matically set to stay on for a preset time period and does not require the switch to be held down. If the light­ing procedure was successful, the #2 LED (FLAME) will begin flashing as the flame temperature rises to the correct operating temperature. Once the LED becomes solidly lit, the flame has reached operating temperature.
b. To light the instrument from the Plat-
form using the auto-ignite mode, simply press the “light” softkey shown in the “Light Flame” menu of the Platform. The Analyzer Module will begin to go through an auto­mated sequence of enrichment and ignition similar to the manual mode described in step 5. If the burner fails to light the first try, the Analyzer Module will perform 2 more tries be­fore terminating the auto-ignite se­quence. If the Analyzer Module fails to light after 3 attempts, an error message will be displayed showing the cause of the fault.
5. If the burner fails to light, check all gas connections for proper gas composition and pressure, block temperature, and outlets for obstructions. Repeat step 4.
6. If the flame is lit, the #2 LED will begin flashing. Once the flame temperature has reached the correct operating tem­perature, the LED will remain on solid.
7. If the fuel and air pressures and ratios are within proper operating parameters to support a continuous flame opera­tion, the #6 LED (FUEL/AIR) will illumi­nate. This light will not be on before or during flame ignition.
8. Once the burner block temperature reaches the control temperature of 80°C, the #3 LED will stay on solid.
9. If the instrument has be successfully lit, the temperatures are up to proper op­erating levels, and the fuel, air, and sample gases are properly adjusted to support the flame and achieve reliable results, all 6 indicator lights will be lit solid.
The instrument is now ready for calibration or burner optimization.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-9
Page 16
Instruction Manual
748412-E March 2003
Model NGA2000 FID2
P-10 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 17
Model NGA2000 FID2

DESCRIPTION AND SPECIFICATIONS

1-1 OVERVIEW

Instruction Manual
748412-E
March 2003
SECTION 1
This manual describes the Flame Ionization Detector (FID2) Analyzer Module of Rose­mount Analytical's NGA 2000 Series of gas analysis components.
The FID2 Analyzer Module is designed to use a flame ionization technique to measure the total concentration of hydrocarbon (including certain oxygenated hydrocarbons) compo­nents within the sample stream.
The entire FID2 Analyzer Module is designed as a module with electrical connections at its front, and gas connections made from the rear. All electronics relative to sample control and signal conditioning are included in this module.
Igniter
Positive Electrode

1-2 TYPICAL APPLICATIONS

Typical applications for the FID2 Analyzer Module include:
The monitoring of atmospheric air for low­level total hydrocarbon contaminants
Determining the total hydrocarbon content of exhaust emissions from internal combustion engines
Carbon bed monitoring
Determining the total hydrocarbons content of process and product gases from air separa­tion plants.
Exhaust
­+
-
Ions
+
­+
­+
-
Negative Electrode
+
Flame
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Air
Fuel + Sample
Figure 1-1. Flame Ionization Detection Technology
Page 18
Instruction Manual
748412-E March 2003
Model NGA2000 FID2

1-3 THEORY OF TECHNOLOGY

This Analyzer Module uses the flame ioniza­tion method of detection. The sensor is a burner in which a regulated flow of gas sam­ple passes through a flame sustained by regu­lated flows of a fuel gas (a hydrogen/diluent mixture) and air.
Within the flame, the hydrocarbon compo­nents of the sample stream undergo a com­plex ionization that produces electrons and positive ions. Polarized electrodes collect these ions, causing current to flow through an electronic measuring circuit.
The ionization current is proportional to the rate at which carbon atoms enter the burner, and is therefore a measure of the concentra­tion of hydrocarbons in the sample.
The gas pressures are continuously moni­tored and controlled through electronic pres­sure transducers.
The measurement of concentration is placed on the network, where it can be shown on the Platform Display or on other data acquisition devices.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 19
Model NGA2000 FID2
A
FLOW CONTROL MANIFOLD ASSEMBLY 659043
5 REAR PANEL IS INCLUDED IN FLOW CONTROL MANIFOLD ASSEMBLY 659043.
4 MANIFOLD ASSEMBLY 659043 MUST BE RETURNED TO FACTORY WHEN REPLACMEN T OF AIR AND/OR FUEL FLOW FLUISTER VALVES IS REQUIRED .
3 15 PSI 2 L/MIN.: 659178 2 PSI 1 L/MIN.: 659 073 NO RESTRICTOR : 008435
2 STD MIXED FU EL: 659514
1 STD MIXED FUEL: 658146 FUEL CAP ILLARY, 659031 SAMP LE CAPILLARY
FID EXHAUST
FTG, BULKHEAD W/RESTRICTOR 1/4T-1/8MP T
FTG, MALE CONN 1/4T-1/8MPT 008435 W/FILTER 017154
REAR PANEL5
SAMPLE IN
FTG, BULKH EAD
SAMPLE OUT
FTG, MALE CONN 1/4T-1/8MP T 008435
FUEL IN
AIR IN
FUEL FLOW CONTROL FLUISTER VALVE 659541
2
IR FLOW CONTROL FLUISTER VALVE 659541
3
4
4
MANIFOLD
SENSOR, AIR PRESSURE 0-30 PSIG 659498
SENSOR, FUEL PRESSURE 0-30 PSIG 659498
BACK PRESSURE REGULATOR 659063
FTG, ELBOW 1/16T-1/8M PT 904582 (3)
659072
SENSOR, SAMPLE PRESSURE 0-15 PSIG 659497
659038
659037
FTG, UNION 1/16T 818270
FTG ASSEMBLY 659173
FTG, FLBOW 1/4T-1/4MP T 902147
Figure 1-2. FID2 Analyzer Flow Diagram
FTG, UNION 1/16T 818270
Instruction Manual
748412-E
March 2003
FTG, UNION 1/16T 818270
FTG, UNION 1/16T 818270
FUEL
CAPILLARY
AIR
CAPILLARY
SAMPLE
CAPILLARY11
FLAME IONIZATION DETECTOR
(FID)
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Page 20
Instruction Manual
748412-E March 2003
FLOW CONTROL
FLUISTOR
CABLE
CONTINUITY
NCNC
J7J2
J9
J16
J8 J5
IGNITER
POLARIZED
COLLECTOR
BURNER BLOCK
BURNER BLOCK ASSEM BLY
1
CONNECTORS J5 AND J6 ON INTRINSICALLY SAFE BOARD ARE INTERCHANGEABLE.
FACTORY REPLACEMENT.
2
REAR SECTION OF ANALYZER FRONT SECTION OF ANALYZER
FUEL
FLOW CONTROL
FLUISTOR
AIR
659070
FLAME
THERMISTOR
BURNER
BLOCK HEATER
2
J1
FUEL
PRESSURE
SENSOR
0-30 PSIG
J12 J13J3 J11 J4
2
AIR
PRESSURE
SENSOR
0-30 PSIG
INTRINSICALLY SAFE BOARD
1
RTD
MIDDLE
SECTION OF
ANALYZER
SAMPLE
RESSURE
SENSOR
0-15 PSIG
THIS SECTION OF INTRINSICALLY
SAFE BOARD CONTAINS CURRENT-
LIMITING RESISTORS
J6
POLARIZED
COLLECTOR
1
J15
Figure 1-3. FID2 Wiring Diagram
Model NGA2000 FID2
THIS SECTION OF
INTRINSICALLY SAFE BOARD
CONTAINS ±15 SUPPLIES,
+90V SUPPLY AND LOW-
LEVEL ANALOG CIRCUITS
J15
J15 J14 J25
MODULE BOARD 659060
J6 J5
COMPUTER BOARD
J7
J8J11
NCNC
658350
J14 J17
J1J4
J2
J3
J22
J24
J22
LON1
LON2
24V POWER
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 21
Model NGA2000 FID2
Instruction Manual
748412-E
March 2003

1-4 GAS SAFETY FEATURES

The FID2 module is divided into two parts - a pneumatic section and an electronic section. The two sections are separated by a pair of solid partitions to prevent any leak of gas in the pneumatic section from reaching the elec­tronics. The electrical connections into the pneumatic section are made intrinsically safe by a series of over-voltage protection devices and current limiting resistors. The burner it­self is an explosion-proof assembly. The combination of these two techniques allows the analyzer to meet international safety stan­dards without the use of an expensive con­tinuous-dilution purge - but ONLY when it is installed in a general purpose area with good air circulation.
WARNING
POSSIBLE EXPLOSION HAZARD
Hydrocarbon concentration(s) in the sam­ple gas must be below the Lower Explo­sion Limit (LEL).
The FID2 is designed to use 40% H fuel at a maximum inlet pressure of 3446 hPa­gauge (50 psig).
1
WARNING
POSSIBLE EXPLOSION HAZARD
Protection against explosion depends upon a special fuel flow restrictor at the fuel inlet. DO NOT REMOVE THE FUEL INLET RESTRICTOR.

1-5 FUEL GAS OPTION

The standard FID2 Analyzer Module requires 40% hydrogen/60% helium burner fuel gas.
For monitoring internal combustion exhaust emissions or other sample gas with varying oxygen content, mixed fuel is preferable. In fact, a hydrogen/helium mixture is more desir­able than a hydrogen/nitrogen mixture. With this type of sample, the use of mixed fuel gas minimizes the error introduced by oxygen synergism.
/60% He
2
All tubing ahead of the burner is rigid metallic tubing assembled with ferrule/nut type com­pression fittings. However, should an internal fuel leak occur, a worst-case leak would be dissipated below 25% of the LEL of hydrogen by natural dilution outside of the pneumatic section before it could be ignited by any ex­ternal ignition source, and there is nothing within the pneumatic section to ignite it.
Changes in the burner air flow rate have little effect on signal strength. For a given flow, the signal can be optimized by adjusting the fuel flow rate.
Refer to Table 1-1. Gas Flow Rates below.
GAS FLOW MIXED FUEL
FUEL
SAMPLE
AIR
100 cc/min
10 cc/min
400 cc/min
Table 1-1. Gas Flow Rates
1
The fuel restrictor is part of the Flow Control Manifold
Assembly, which is specific to an application.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Page 22
Instruction Manual
748412-E March 2003
ANALYZER CHARACTERISTICS
Full Scale Sensitivity
Fuel Consumption
Operating Setting For Sample Pressure Regulator
Model NGA2000 FID2
40% H
4 ppm, CH
100 to 110 cc/min
345 hPa-gauge (5 psig)
Table 1-2. Analyzer Characteristics Relative to Fuel Gas
to <1%, CH4
4
/60% He
2
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 23
Model NGA2000 FID2

1-6 SPECIFICATIONS

a. General

Measurement Species .................. Total hydrocarbons
Range H
Repeatability.................................. 1% of fullscale at a constant temperature, sample flow and fuel,
Min. Detectable Level ................... 0.04 ppm H
Noise ............................................ <1% of fullscale, peak to peak
Linearity ........................................ ≤ ±1% of fullscale
Response Time ............................
Zero Drift ....................................... ≤ ±1% of fullscale/24 hours at constant temperature, hydrocarbon
Span Drift ...................................... ≤ ±1% of fullscale/24 hours at constant temperature, hydrocarbon
Effect of Temperature ................... ≤ ±2% of fullscale for any temperature change of 10°C and rate of
Operating Temperature ................ 41°F to 104°F (5°C to 40°C)
Operating Humidity........................ <95% relative humidity, non-condensing
Power Requirements .................... +24 VDC ±5%, 120 W max.. direct to analyzer module;

b. Physical

/He fuel
2
Low range.............................. 0 to 4 ppm CH
High range ............................. 0 to 50 ppm CH
Ripple and Noise ................... <100 mV peak to peak
Line and Load Regulations.... <±1%
Instruction Manual
748412-E
March 2003
, through 0 to 1% CH4
4
, through 0 to <5% CH4
4
burner air and sample pressure
/He fuel – methane equivalent
2
≤1 sec. for bypass flow rate of 500 cc/min (for a sample change at
the rear panel connector of the instrument)
concentration of supply gases, sample flow and fuel, burner air and sample pressure
concentration of supply gases, sample flow and fuel, burner air and sample pressure
change less than 10°C/hour
Case Classification ........................ General purpose for installation in weather-protected area
Dimensions ................................... See Outline and Mounting Dimensions, Figure 2-1, on page 2-3
Weight .......................................... 10.43 kg (23 lbs.)
Mounting ....................................... Inside a Platform or custom-installed in a panel
Max. Length of LON Cable ........... 1600m (1 mile) between Analyzer Module and Platform
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
Page 24
Instruction Manual
748412-E March 2003

c. Gas Requirements

Sample .......................................... Non-flammable, below 100% LEL
Temperature ......................... 32°F to 248°F (0°C to 120°C), <20°C variance/24 hours, <10°C
Flow Rate............................... 0.5 to 2.0 ml/min.
Supply Pressure .................... 483 to 1035 hPa-gauge (7 to 15 psig)
Particles ................................. Filtered to <2 microns
Dewpoint ............................... <45°C
Materials in contact
with Sample ................ Stainless steel, Teflon, glass-filled Teflon, Viton
Fuel Gas ....................................... Premixed 40% hydrogen and 60% helium
Flow Rate............................... 75 to 110 ml/min.
THC ....................................... 0.5 ppm, CH
Supply Pressure .................... 3101 to 3450 hPa-gauge (45 to 50 psig)
variance/hour
Model NGA2000 FID2
4
DO NOT USE PURE HYDROGEN FUEL. An explosion resulting in severe personal injury or death could occur. Also, each Analyzer Module is factory-configured for mixed, and can­not use the fuel for which it was not configured unless field reconfiguration is done.
Burner Air ...................................... Hydrocarbon-free grade air
Flow Rate............................... 350 to 400 ml/min.
THC ....................................... 0.1 ppm, CH
Supply Pressure .................... 1725 to 3450 hPa-gauge (25 to 50 psig)

d. Gas Connections

Sample In ...................................... 1/4” O.D. tube fitting
Burner Air In .................................. 1/4” O.D. tube fitting
Fuel In............................................ 1/4” O.D. tube fitting
Bypass Out .................................... 1/4” O.D. tube fitting
Burner Exhaust Out....................... 3/8” O.D. tube slip-fit connection, tygon or equivalent (this connec-
WARNING.
POSSIBLE EXPLOSION HAZARD
4
tion shall slope downward 6° minimum from horizontal)
NOTE
The burner exhaust and bypass out shall be vented to atmospheric pressure and to a non­classified location. .
See the Preface section of the Platform manual for specifications regarding Platform-related components (e.g., case dimensions) and the I/O Module manual for specifications regarding I/O (e.g., relay outputs).
1-8 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 25
Model NGA2000 FID2
Instruction Manual
748412-E
March 2003
SECTION 2
INSTALLATION

2-1 UNPACKING

When the FID2 Analyzer Module is received, carefully examine the shipping carton and contents for signs of damage. Immediately no­tify the shipping carrier if the carton or con­tents is damaged. Retain the carton and packing material until all components associ­ated with the Analyzer Module are opera­tional.

2-2 ASSEMBLY

The FID2 analyzer module MUST NOT be placed within a conventional NGA platform, single module enclosure or dual module en­closure since the latter would not allow free flow of air around the module, thus violating its safety certification. The enclosure is de­signed so that this would be very hard to do anyway.
There is a special platform specifically de­signed to accept this module; consult the fac­tory for details.

2-3 LOCATION

WARNING.
POSSIBLE EXPLOSION HAZARD
Do not place the FID2 module within an­other enclosure unless the latter has a guaranteed air circulation such as to dilute a worst case fuel or sample leak below 25% of the LEL. Failure to will negate the safety features and may result in explo­sion, serious injury, material damage and death. Also, do not cover the vent holes on the top and sides of the module.
Install the Analyzer Module in a clean, weather-proofed, non-hazardous, vibration­free location free from extreme temperature variations. For best results, install the Ana­lyzer Module near the sample stream to mini­mize sample transport time.
Operating ambient temperature is 5 °C to 40 °C, limited to temperature changes of less than 10 °C/hr. Acceptable dew point range is less than 95% relative humidity, but not in ex­cess of 40°C wet bulb temperature.
The cylinders of fuel, air, and calibration gases should be located in an area of rela­tively constant ambient temperature.
CAUTION .
PRESSURIZED GAS
See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5.

2-4 GASES

a. Overview

During normal operation, the Analyzer Module requires fuel and air to maintain the burner flame as well as suitable stan­dard gases for. Refer to the criteria for selection of these gases in Section 2-4c on page 2-2.
After initial startup or after startup follow­ing a prolonged shutdown, the analyzer may display baseline drift for a consider­able period of time, particularly on the most sensitive range.
Commonly, the drift is caused by small amounts of organics (such as hydrocar­bons) in the inner walls of the tubing in both the internal flow system and the ex­ternal gas supply system. Drift results from any factor influencing the equilibrium of these adsorbed hydrocarbons, such as temperature or pressure. Hydrocarbons adsorbed within the analyzer in the gas passageways (or in the fuel or air lines) will elevate the overall baseline.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Page 26
Instruction Manual
748412-E March 2003
Note that this type of drift occurs only when the flame is burning. If drift occurs when the flame is extinguished, the elec­tronic circuitry is at fault or the burner or cabling is contaminated with a conductive film. To minimize drift, use clean fuel and air, keep the analyzer clean, and locate the gas cylinders in an area of relatively constant ambient temperature.
The cylinders supplying all gases each should be equipped with a clean, hydro­carbon-free, two-stage regulator and a shutoff valve.
All new external gas tubing (except for SAMPLE BYPASS) is strongly recom­mended, preferably pre-cleaned, stainless steel, gas chromatograph-grade tubing. Thoroughly clean before use (if a hydro­carbon-based cleaning solvent such as acetone is used, purge tubing with dry ni­trogen or helium for several minutes be­fore using.)
Gas line connections are compression fit­tings. Do not use pipe thread tape on such fittings.
Since the oxidation of hydrogen is ac­companied by the formation of water va­por, the exhaust tubing always should be slanted downward at least 6 degrees from horizontal. Otherwise, water may accumu­late in the line, causing back pressure and noisy readings, or may back up in the line and flood the burner.
If the sample is toxic or noxious, or is to be reclaimed, connect the Bypass outlet to a suitable disposal system. Do not use any device that may cause back pressure in the line.
Model NGA2000 FID2
exhaust and bypass must be vented at atmospheric pressure to a non-classified location in accordance with ANSI/NFPA-
496.

c. Gas Specifications

Fuel Gas — Standard analysis usually
requires mixed fuel, i.e., 40% (±2 %) hy­drogen and 60% helium. H is recommended over H of better linearity in concentration output. Such blends are supplied by many gas vendors specifically for this use, with a guaranteed maximum total hydrocarbon content of 0.5 ppm, measured as meth­ane. This specification should be used when obtaining these mixtures.
Burner Air — In order to ensure a low
background signal, hydrocarbon free grade air with less than 1 ppm maximum total hydrocarbon content is highly rec­ommended. An alternative source for
burner air and zero gas (see Calibration Gases below) is a combination diaphragm
pump and heated palladium catalyst. This process continuously removes moderate amounts of hydrocarbons and carbon monoxide from ambient air.
Calibration Gases — Calibration method
and gases depends on the type of fuel gas used, the operating range, and the desired measurement accuracy. In all methods, zero and span gases are used, and are introduced through the sample inlet at the rear of the module.
Zero Gas It is recommended that the
gas should have a composition as close to the background composition of the sample as possible.
/He mixed fuel
2
fuel because
2/N2

b. Connections

Refer to Figure 2-2 on page 2-4. Con- nect inlet and outlet lines for sample, burner fuel and air, exhaust, and bypass to appropriately labeled fittings on the rear panel. All connections are 1/4-inch fer­rule-type compression fittings. Burner
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Span Gas
specified concentration of methane and other hydrocarbons in a background gas such as nitrogen.
Span gas consists of a
Page 27
Model NGA2000 FID2
Sample Gas — Sample gas must be non-
flammable (below 100% of the sample's LEL).
Flow Rate — The sample flow rate must
be between 0.5 L/min. and 2 L/min.
Instruction Manual
748412-E
March 2003
sample be filtered for particulates down to 0.1 microns. A suitable filter is
the Balston type 95S6 with 0.1 micron fil­ter element. It should normally be re­placed on a two week schedule, depending on the sample.
Pressure/Filtration — See Table 2-1 on
page 2-4 for input pressure specifications. Noncompliance with these specifications could cause over-pressure damage to the module.
Leak Test — The Analyzer Module is
completely tested at the factory for gas leakage. The user is responsible for test­ing for leakage at the inlet and outlet fit­tings on the rear panel. The user is also
responsible for internal leak testing peri­All internal pressure settings are preset at the factory, but the operator should check
for accuracy. It is essential that the
odically and if any internal pneumatic
components are adjusted or replaced
(with a test procedure chosen by the
user).
8.2
[208]
4.3
[110]
22.6
[573]
2.9 [73]
.8
[19]
[17]
.6
[16]
.7
1.2
[31]
[23]
.9
3.1
[78]
1.0
[25]
2.2
[55]
[23]
2.8
.8
[71]
[20]
.9
Figure 2-1. FID2 Outline and Mounting Dimensions
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Page 28
Instruction Manual
748412-E March 2003
MIXED FUEL IN
AIR IN
Gas supply External Pressure Internal Pressure
Fuel
Burner air
Sample
Model NGA2000 FID2
SAMPLE IN
Figure 2-2. FID2 Rear Panel
50 – 55 psig 1035 - 2070 hPa 50 - 55 psig 1380 - 4140 hPa 5 - 10 psig 345 - 690 hPa
Table 2-1. Gas Supply Pressures
SAMPLE BYPASS OUT
BURNER EXHAUST
5 psig 345 hPa 15 psig 1035 hPa 5 psig 345 hPa
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 29
Model NGA2000 FID2

2-5 ELECTRICAL CONNECTIONS

NOTE
Electrical installation must be in compli­ance with National Electrical Code (NEC/NFPA 70) and/or any state or local codes.
Two electrical connections are required on the Analyzer Module: POWER and NETWORK. See Figure 2-3 below. On the Analyzer Mod­ule, two NETWORK connectors are available, either of which is appropriate for: 1) intercon­nection with the Back plane of the Platform or
2) "daisy-chaining" with other NGA 2000
components, or 3) connection to a PC via a suitable LONTALK adapter and software such as the NGA DDE server and client. Connect Analyzer Module POWER to Back plane POWER or external 24 VDC power source.
Connect the network cable to either the NETWORK 1 or NETWORK 2 connection on the Analyzer Module front panel, and the NETWORK connection on the LON I/O mod­ule if used with a Platform, or directly to a computer using appropriate LONTALK adapter hardware and software such as the NGA DDE server. Connect the power cable to both the Analyzer Module front panel and to a 24V 5A minimum power supply.
FUSE
Figure 2-3. FID2 Front Panel
24V POWER
Instruction Manual
748412-E
March 2003
NETWORK
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Page 30
Instruction Manual
748412-E March 2003

2-6 ANALYTICAL LEAK CHECK

Model NGA2000 FID2

b. Manometer Method

The FID2 Analyzer Module is completely tested at the factory for gas leakage. The user is responsible for testing for leakage only at the inlet and outlet fittings on the rear panel.
CAUTION
SENSOR PRESSURE
Do not expose the sensor to pressure in excess of 1.0 psig as this may cause dam­age.

a. Flow Indicator Method

Supply air or inert gas such as nitrogen, at 1 psig (6.8 hPa), to the analyzer through a flow indicator with a range of 0 to 250 cc/min. Install a shut-off valve at the sample gas outlet. Set the flow rate to 125 cc/min.
FID2 Analyzer Module
Install a water-filled U-tube manometer at
the sample gas outlet. Install a shut-off
valve at the sample gas inlet. Admit air or
inert gas to the inlet shut-off valve until
the analyzer is pressurized to approxi-
mately 50 hPa. The water column will be
about 500 mm.
FID2 Analyzer Module
Inlet Outlet
Overpressure
Water
N
2
approx. 50 hPa
N
2
1 psig
(6.9 kPa)
Inlet Outlet
Flow Meter
Figure 2-4. Leak Test - Flow Indicator
Method
Close the outlet shut-off valve and notice that the flow reading drops to zero. If the flow reading does not drop to zero, the system is leaking and must be corrected before the introduction of sample gas or the application of power.
Outlet
Gas
Figure 2-5. Leak Test - Manometer Method
Close the inlet shut-off valve and, follow-
ing a brief period for pressure equilibrium,
verify that the height of the water column
does not drop over a period of about 5
minutes. If the water column height drops,
the system is leaking and must be cor-
rected before the introduction of any
flammable sample gas or application of
power.

c. Troubleshooting Leaks

Liberally cover all fittings, seals, and other
possible sources of leakage with a suit-
able leak test liquid such as SNOOP (part
837801). Bubbling or foaming indicates
leakage. Checking for bubbles will locate
most leaks but could miss some, as some
areas are inaccessible to the application
of SNOOP. For positive assurance that
system is leak free, perform one of the
tests in Section 2-6 above.
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 31
Model NGA2000 FID2

2-7 INSTALLATION GUIDELINES

Is the Analyzer’s Location clean, weather­proofed, non-hazardous, vibration-free, and with a stable ambient temperature?
Instruction Manual
748412-E
March 2003
Is the burner exhaust tube slanted down a minimum of 6 degrees from horizontal?
Have all the external gas connections been leaked checked?
Are gas supply cylinders equipped with a clean, hydrocarbon free two stage regula­tor and shut off valve?
Are external tubing, regulators, valves, pumps, fittings etc. clean?
Is the correct fuel type being used?
Is the THC content of the supply gases
compatible with the analysis range?
Are the burner exhaust and bypass vented to atmospheric pressure? Is the vent pressure constant?
Has the dead volume for external sample and fuel lines been minimized?
Has clean stainless steel tubing been used for fuel and sample lines?
Is a suitable 0.1 micron filter used in the sample line?
Is the sample line and filter heated?
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Page 32
Instruction Manual
748412-E March 2003
Model NGA2000 FID2
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 33
Model NGA2000 FID2
pp
Instruction Manual
748412-E
March 2003
SECTION 3

OPERATION

3-1 OVERVIEW

Prior to initial startup, the user should leak test the module as outlined in Section 2.
For the remainder of this section, the Ana­lyzer Module interconnection with a Plat­form or some interfacing component will be assumed. Display and Keypad information refers to use of this module with the Plat­form.
(For a complete description of Platform front panel controls and indicators, see the Platform instruction manual.)

3-2 DISPLAYS

Three kinds of Display screens are avail­able to the user:
Run Mode
Menu
Help

a. Run Mode Display

FID2
-19.4
0 Range: 2 50
RAW SIGNAL 556320 PRESSURE 14.7 psia
CASE TEMPERATURE 45.3 C NOISE LEVEL 0.811 ppm
DISPLAY
PARMS MENU NEXT INFO
Figure 3-1. Run Mode Display
The Run Mode is the normal mode of operation. In this mode, the display will show the gas measurement of the selected analyzer or subnode (the Control module may be connected to
m THC
many analyzers at once, but only one may be display at a time as shown), the component of interest, user­selectable (up to four) secondary vari­ables, the current operation of the soft­keys, and a graphic bar representing the displayed concentration as a per­cent of fullscale.
It is also possible to show up to five analyzers or subnodes on the screen at once, each analyzer having its own line on the display.

b. Menu Displays

The following is a brief description of the menus shown on the Platform Con­trol Module as they apply to the FID2 analyzer. Much more detail is avail­able in the Platform manual as well as the NGA Reference Manual.
The first menu shown for any of the subnodes is as follows:
Basic controls… Expert controls and set up…
(Operational configuration)
Technical level configuration… (Diagnostic and manufacturing/service)
DISPLAY
Figure 3-2. Main Menu

Main Menu

PARMS NEXT LOCK INFO
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
Page 34
Instruction Manual
748412-E March 2003
The Main Menu is subdivided into three levels of control based generally on which personnel is likely to use it:
Basic Controls – Operators, Expert Controls and set up – System Engi- neers, and Technical level configura­tion – Analyzer technicians. Many
layers of the menu structure are de­scribed at appropriate places through­out this manual.
From the Run Mode display, press the MENU softkey to gain access to the Main Menu.
The Basic controls menu is as follows:
Measurement range number: Range upper limit:
Range and functi o n al c o ntrol: Bypass samp le flow: Ranges with valid calibration: Calibration status: If it won’t calibrate… Flame condition… Light flame…
HOME SPAN
Figure 3-3. Basic Controls Menu
This menu allows the user to change the range for all subnodes, to control the operation of the GC (single shot or continuous operation) and to see its current operational status, to allow re­mote control of the range change, and allows links to other menus to light the flame, zero and span the analyzer.
In the menu figure above, the ital­ized/capital words are the names of the network variables whose values are in fact shown on the screen.
Basic Controls
ESCAPE
CRANGE
CURRENTRNGHI
CONTROL
FLOW_IS(1)
CAL_VALIDITY
CALSTAT
DIGDIAG (5)
ZERO INFO
NOTE
Model NGA2000 FID2
The Expert controls menu is as fol­lows:
Measurement range number: Range upper limit: Range settings… Linearizer… Range and functi o n al c o ntrol: Zero/span calibration… Ranges with valid calibration: Physical measurements… Flame condition: Light flame…
HOME CALDATA
Figure 3-4. Expert Controls Menu
This menu is almost the same as the
Basic controls menu but with the addi-
tion of a few extra links.
The analyzer may be configured
through the Analyzer set up menu, un­der Expert controls and set up.
Calibration gas list… Calibrati on parameters…
Gas measurement parameters… Analyzer parameter list … Physical measurement parameters…
Displayed parameters… Analyzer tag:
HOME
Figure 3-5. Analyzer Module Setup Menu
This menu contains links to many other menus used to configure the operation of the analyzer.
Expert controls
ESCAPE
CAL INFO
Analyzer module set up
ESCAPE
STORE
CRANGE
CURRENTRNGHI
CURRENTSTAT
CONTROL
CAL_VALIDITY
DIGDIA
TAG
INFO
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 35
Model NGA2000 FID2

c. Help Displays

A typical help menu:
Select the aspect of the analyzer to configure. Set up the calibration gas values in the calibration gas list. Set up the other calibration parameters. Linearization, filtering and other functions are set up in menus under measurement parameters. The analyzer parameter list simply lists all the settable parameters in order. Physical measurements show flow, pressure etc. and associated limits. Displayed parameters show what is displayed on the four auxiliary lines on the single component display.
HOME
Figure 3-6. Typical Help Menu
Analyzer module set up
ESCAPE
Instruction Manual
748412-E
March 2003
INFO
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3
Page 36
Instruction Manual
748412-E March 2003

3-3 STARTUP PROCEDURE

WARNING
VENTILATION
For safety, the Analyzer Module should be installed in a non-confined, venti­lated space. Do not block any of the ventilation holes as they are part of the safety system.
1. Connect supply gases and outlets to/from module.
2. Connect the LON cable(s) and the +24VDC power cable.
3. Turn power ON.
Model NGA2000 FID2
RAM test: Power supply test:
HOME
Figure 3-8. Self Test Results Menu
Descriptions of the tests performed fol­low:
RAM Test
Checks the RAM on the Analysis Com­puter PCB.
ESCAPE
Self test results
TEST
SELFTEST(3) SELFTEST(4)
INFO
4. Check the LED’s. The power green LED should be illuminated. The Flame LED should be OFF. The block LED should be blinking or ON.
5. Allow the network to initialize. Perform any binding of I/O modules required – see the Platform manual for details.
6. Check the general health of the ana­lyzer by reviewing the status of the Self Tests. All “Pass” conditions should be obtained.
These test results can be found by se-
lecting the following from the Main
Menu: Technical level configuration, Diagnostic menus, Analyzer module diagnostics, Self test results. All tested
parameters should indicate “Pass.”
Power supply voltages… Primary variable parameters… Physical measurement par a mete rs … Temperature control parameters… Miscellaneous control parameters… Trend display control… Auto ignition parameters… Self test results… Software diagnostics… Start up analyzer…
Analyzer diagn ost ic s
HOME
ESCAPE
Figure 3-7. Analyzer Diagnostics Menu
INFO
Power Supply Test
Verifies that all internal DC voltages are within the required tolerances.
The self test can be repeated at any time by activating the TEST softkey in
the Self test results menu.
7. Introduce the remaining supply gases. Perform leak check. (See Specifica­tions on page 1-7)
8. Set and verify the internal gas pres­sures.
9. Allow the block to warm up to 50°C, approximately 30 minutes.
10. Note the six LED’s on the front panel of the Analyzer Module. They provide necessary information for proper igni­tion procedure. The LED’s, when illu­minated, denote the following information:
Power – Unit powered up.
Flame – Flame on. If the module is
trying to light the flame, with fuel flowing but no flame detected, the LED will flash.
3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 37
Model NGA2000 FID2
Block – Continuous illumination im­plies the block temperature is within 5% of its operating temperature setpoint; otherwise the LED will blink. If the oven temperature is too high the LED blinks at double speed.
Instruction Manual
748412-E
March 2003
13. Before ignition and operation, Fuel Flow must be set to ON in “Light
Flame” menu screen under Basic Con-
trols.
14. The front panel ignition switch must be
manipulated in the following ways:
Sample – Sample pressure is within ±15% capillary requirement.
Ignite OK – The block temperature is 50°C allowing proper ignition to occur.
Fuel/Air – Proper fuel and air condi­tions exist to support a flame. Fuel pressure is between 400 and 675 hPa. Air pressure is between 800 and 1200 hPa. The ratio of fuel/air is between 35% and 65%. This light will not come on until success­ful ignition.
11. Auto-Ignite the flame. The Flame LED should then be continuously illumi­nated when the flame is successfully lit.
12. Auto-Ignition provides fuel override and three attempted ignitions (default setting), if necessary,
a. Press up and hold for 30 seconds.
This puts the system into the en-
riched mode, with lower air flow.
b. Press down to turn on the burner
flow plug for up to 10 seconds.
c. Repeat as necessary (if fuel and
air sources are farther away than 10 feet, several more attempts
may be necessary).
d. Flame on is defined as true when
the flame temperature is greater than the block temperature by the amount contained in the variable
FLAME_DELTA.
e. If the flame has been lit, but the
flame temperature increases
slowly, perform the following steps:
(1) After igniting flame, wait for 2
seconds.
Flame condition: Auto-ignition: Ignition system enable: Number of ignition attempts so far: Time on this cycle – secs: Fuel supply pressure: Burner air pressure: Sample pressure: Flame temperature:
Status:
HOME
ABORT
Light flame
LIGHT
DIGDIAG(5) AUTOIGNITE FUEL_FLOW
TIME_LEFT(2) TIME_LEFT(1)
PRES_IS(3)
PRES_IS(2)
PRES_IS(1)
TEMP_IS(2)
FID_MSGE
INFO
15. Check and re-adjust the internal pres-
(2) Press switch down momentar-
ily.
(3) Repeat release switch, wait
and press down steps as nec-
essary.
sures if required. The Fuel/Air light
must be lit for proper operation.
Figure 3-9. Light Flame Menu
16. The unit is now ready for first binding as described in Section 3-5 on page 3­8 and then optimization as described in
Section 3-4 on page 3-7.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5
Page 38
Instruction Manual
A
748412-E March 2003
Model NGA2000 FID2
RESPONSE
(100 ppm CH
fullscale)
4
Figure 3-10. Typical Module Response vs. Sample Pressure Setting
RESPONSE
(100 ppm CH
fullscale)
4
Figure 3-11. Typical Module Response vs. Fuel Pressure Setting
1.0
0.8
0.6
0.4
0.2
1.0
0.8
0.6
0.4
0.2
SAMPLE: 100 ppm CH FUEL: 10 psig (690 hPa) H
IR: 18 psig (1242 hPa)
0
0
0
0
2
13.76
3
20.64
SAMPLE PRESSURE
FUEL PRESSURE
4
27.52
SAMPLE: 100 ppm CH
5
34.4
psig
hPa
at 5 psig (344 hPa)
5
345
psig
hPa
4
41.28
in N
7.5
518
in N
4
2
2
6
2
7
48.16
10
690
3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 39
Instruction Manual
Model NGA2000 FID2
RESPONSE
(100 ppm CH
fullscale)
4
Figure 3-12. Typical Curves of Module Response vs. Pressure Setting on Air Pressure Regulator
1.0
0.8
0.6
0.4
0.2
SAMPLE: 100 ppm CH
at 5 psig (344 hPa)
0
0
10
690
14
966
AIR PRESSURE
20
1376
psig
hPa
in N
4
2
748412-E
March 2003

3-4 OPTIMIZATION PROCEDURE

Although the module has been set up for best operation at the factory, settings can change and your application may be different. The following explains how to optimize the opera­tion of the FID2 for your application.
1. Ignite the flame of the FID2 using the startup procedure in Section 3-3 on page 3-4.
2. Verify that the mixed fuel supply pres­sure at the Analyzer’s rear panel bulk­head is between 49 and 50 psig.
3. Allow the analyzer module response to stabilize. Typically allow 1 to 4 hours.
4. Select the desired range to optimize. For best results use a span gas with a concentration of 100 ppm CH greater. This will minimize the effects of THC contamination in the fuel and burner air supply.
5. Set the internal sample pressure to the desired operating level. The sample pressure must be kept constant throughout the optimization procedure.
4
6. Set the internal burner air pressure to between 965 and 1103 hPa-gauge (14 to 16 psig). The burner air pressure must be kept constant throughout the optimization procedure.
7. Set the internal fuel pressure to 345 hPa-gauge (5 psig). Calibrate the in­strument as stated in step 8.
8. Introduce the span gas and monitor the reading until it is stable. Increase the in­ternal fuel supply setting in the following sequence: 600 hPa-gauge (8.7 psig), 625 hPa-gauge (9.1 psig), 650 hPa­gauge (9.4 psig), 675 hPa-gauge (9.8 psig), 700 hPa-gauge (10.1 psig), 725 hPa-gauge (10.5 psig). Monitor the reading at each fuel pressure setting. Wait at least 2 minutes between fuel setting changes. Record all readings.
or
9. Review the readings for each fuel pres­sure setting. Select the fuel pressure setting that produces a reading that is within 1% from the maximum. For this condition the FID2 is operating at its op­timized plateau.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Page 40
Instruction Manual
748412-E March 2003

3-5 BINDING

To achieve full coordination between Analyzer Modules and associated I/O Modules, the user must bind those components together in
the System Set Up portion of the Technical Configuration menu. (See the Platform man-
ual for binding instructions.)
NOTE
If binding is attempted after ignition, the flame may be extinguished. If this occurs, it must be re-ignited.

3-6 CALIBRATION

See Section 2-4c on page 2-2 for a descrip­tion of the method for choosing calibration zero and span gases.
Menus used for calibration include the follow­ing:
Zero gas – range 1: Span gas – range 1: Zero gas – range 2: Span gas – range 2: Zero gas – range 3: Span gas – range 3: Zero gas – range 4: Span gas – range 4: Calibration gas HC response factor:
Calibration…
HOME
Figure 3-13. Typical Calibration Gas List Menu
Measurement range number: Zero gas concentration: Span gas concentration: Bypass samp le flow: Flame condition: Raw measurement signal:
Status: Result… Calibration adjustment limits:
HOME
Figure 3-14. Zero and Sp an Calibration Menu
Calibration gas list
ESCAPE
Zero/span calibration
FACTORS
CARBON_ATOMS
CURRENTZERO CURRENTSPAN
CALCHKLIMITS
ZERO SPAN
ZEROGAS(1) SPANGAS(1) ZEROGAS(2) SPANGAS(2) ZEROGAS(3) SPANGAS(3) ZEROGAS(4) SPANGAS(4)
INFO
CRANGE
FLOW_IS(1)
DIGDIAG(5)
RAW_SIGNAL
CALSTAT
INFO
Model NGA2000 FID2
Calibration adjustme nt limits: Calibration averaging time: Calibration failure alarm: Cal failure error allowed: Calibration time out:
Zero ranges: Span ranges:
HOME
Calibration parameters
ESCAPE
Figure 3-15. Calibration Parameters Menu
Date of last zero: Error message for last zero: Error percentage for last zero: Raw signal at last zero: Last zero gas would read: Date of las t span: Error message for last span: Error percentage for last span: Raw signal at last span: The last span gas would read:
HOME

Zero/spa n diagnostic data

ESCAPE
FACTORS
Figure 3-16. Zero/Span Diagnostic Data
To calibrate the Analyzer Module, introduce span gas into the SAMPLE INLET, and do the following:
1. If more than one Analyzer Module is func­tional and the split Run Mode display is shown, press the the desired Analyzer's Run Mode display is acquired.
2. Press the
MENUS softkey to enter the
Main Menu.
3. Verify the fuel type in the Miscellaneous Control Parameters menu (under the
Technical Configuration menu structure,
select the following from the Main Menu:
Diagnostic menus, Analyzer Module Di­agnostics and then Miscellaneous Control Parameters).
CALCHKLIMITS
CALTIMEOUT
ZERORNGS
CALRANGES
CALDATE_Z
CAL_ERR_MSG(1)
CALRESULT(1)
CAL_RAWSIG(1)
LASTZERO
CALDATE_IS
CAL_ERR_MSG(2)
CALRESULT(2)
CAL_RAWSIG(2)
LASTSPAN
DISPLAY softkey until
CALTIME
CALFAIL
CALFPC
INFO
INFO
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 41
Model NGA2000 FID2
4. In the Calibration Gas List menu (from the
Main Menu, select Expert Controls and Setup, Analyzer Module Setup, then Cali­bration Gas List), enter necessary data,
including the Calibration Gas HC Re­sponse Factor. Common HC factors are: methane (CH propane (C only compensate the reading, but are used to select the proper preamp sense resistor.
5. Note that you can go straight to the zero and span calibration screens from here.
6. Press
HOME to re-enter the Main Menu,
enter the Basic Controls menu, introduce
zero gas and allow its response to stabi-
), 1.0, ethane (C2H6 ), 1.90,
4
), 3.00. These factors not
3H8
Instruction Manual
lize, press the Analyzer Zero menu, press and wait.
7. Press the left arrow key to return to the previous menu, change ranges to the next range, and repeat the zero.
8. Press the lyzer Span menu, introduce span gas and allow its response to stabilize, press
SPAN again and wait.
9. Press the Main Menu.
10. Press the
Mode display.
ZERO softkey to enter the
SPAN softkey to enter the Ana-
HOME softkey to re-enter the
DISPLAY softkey for the Run
748412-E
March 2003
ZERO again
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
Page 42
Instruction Manual
748412-E March 2003

3-7 CALIBRATION DETAILS

The range change resistor in the preamp has some effect on the way the spans work. In most NGA analyzers it is possible to make them zero and span all ranges together or separately. If this option is selected, the ana­lyzer attempts to determine if a zero or span is appropriate using the supplied zero and span gases, and if the gases are suitable it will go ahead and zero or span as many ranges as it can. The FID2 however may switch gains be­tween ranges, and if so it will not be able to zero or span the ranges on the other side of the switching point.
The switching point is determined by a num­ber of factors, including the gas response fac­tor, the sample pressure, the type of fuel and the capillary used. In general it will occur at about 600ppm of methane on a low range in­strument. This means that if the ranges are set at 10, 100, 1000, 5000, the analyzer will be able to span the lower two, or the higher two at once, but not all four.
Generally, it is best to operate the analyzer on a single range and calibrate that, or to cali­brate the ranges individually.
If the user is unable to calibrate the Analyzer Module (i.e., when ZERO or SPAN is initiated, nothing happens), several possible solutions may be tried. One solution relates to the use of an incorrect gas for zeroing or spanning (e.g., using a high concentration gas to zero or a zero gas to span the Analyzer Module).
Model NGA2000 FID2
Simply recalibrating with the appropriate gas(es) may not correct the problem because the ZERO OFFSET or SPAN FACTOR has been set to an extreme value in the process.
To remedy the problem, do the following:
1. Verify that correct zero and span calibra-
tion gases are being used properly. If so, attempt to recalibrate according to in­structions at the beginning of the previ­ous section, ensuring that case temperature and displayed measurement reading are stable before initiating the calibration routine. If incorrect gases were used in the initial, failed calibration, skip to Step 2.
2. Make the following selections from the
Main Menu: Expert Controls and Setup, Analyzer Module Setup, then Calibration
Parameters. Disable Calibration Adjust­ment Limits.
3. Recalibrate the analyzer module accord-
ing to instructions at the beginning of the previous section, ensuring that case temperature and displayed measurement reading are stable before initiating the calibration routine.
4. Enable Calibration Adjustment Limits in
the Calibration Parameters menu.
5. For further information and troubleshoot-
ing tips, refer to the NGA Reference Manual.
3-10 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 43
Model NGA2000 FID2
A

3-8 ROUTINE OPERATION

After binding and calibration, proceed as fol­lows:
Supply sample gas to the SAMPLE INLET. Adjust external flow controller or throttle valve so that flow discharged from the BYPASS out­let is between 0.5 and 1.0 L/min. The reading on the SAMPLE pressure gauge should be the same as that used during adjustment of the span control. Adjust, if necessary.
Adjust the Range Number setting.
If maximum sensitivity is required from the FID2 Analyzer Module, use an optimum com­bination of settings on the SAMPLE, FUEL, and AIR pressure controllers. Settings must be determined experimentally, but the curves in Figure 3-10 on page 3-6, Figure 3-11 on page 3-6, and Figure 3-12 on page 3-7 may be used as guides for the FID2 optimization procedure above.
The Analyzer Module will not allow the user to increase the upper limit of a range beyond the ,.maximum range" software setting. To change the "maximum range" value, select
the following from the Main Menu: Technical
Configuration Menu, Service Menus, Manu­facturing Data, and Analyzer Manufacturing Data. Select Minimum1maximum ranges and tags... and then Maximum Range, and use the
arrow keys to scroll the indicated values for each applicable measurement. The same ap-
plies for Minimum Range settings.
Instruction Manual
748412-E
March 2003
nalyzer module s/n:
Manufacturing date code: Hardware revision number:
Software revision number: Minimum range: Maximum range: Measured gas:
HOME
Analyzer manufacturing data
ESCAPE
RESET STORE
Figure 3-17. Analyzer Manufacturing Data Menu
During shutdown, always turn off fuel gas first, then the air and sample gases. The flame can also be turned off by setting Ignition System
Enable to "Off" in the Light Flame menu (un­der Basic Controls). Subsequently, remember
to set Ignition System Enable to "On" before attempting to ignite the flame.
After initial startup, or startup following a pro­longed shutdown, the Analyzer Module re­quires about one day's continuous operation to stabilize. For several days afterwards, cali­brate daily. The frequency of subsequent cali­brations can be reduced as experience dictates, consistent with the accuracy re­quirements of the particular application.
AMMFGDATE
MINRANGE
MAXRANGE
AMSN
AMHR AMSR
GAS
INFO
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-11
Page 44
Instruction Manual
748412-E March 2003

3-9 SHUT DOWN PROCEDURE

If the FID2 is to be left in the powered down state for any length of time greater than a few hours, it is important that the flame tower be dried out before it cools off. This is to avoid condensation causing corrosion. If the sample contains halogenated compounds this is even more important.
Flush the flame tower with burner air for around an hour after the flame has been turned off. Simply leave the burner air flowing. When no evidence of condensation can be seen in the exhaust tubing, it is safe to switch off the FID2 power.
If this precaution is not taken, corrosion can occur making it hard or impossible to light the flame.

3-10 SAFETY SYSTEM

The FID2 Analyzer Module safety system is entirely passive. The pneumatic section of the analyzer is designed to be intrinsically safe, except for the flame tower itself which is de­signed to be explosion proof. Intrinsic safety design ensures that any energy released in the hazardous area (the pneumatic section) will not be enough to ignite any hydrogen that may have leaked. This is achieved by limiting the voltage and the current in all the circuitry
that enters this area. It is therefore essential
Model NGA2000 FID2
that NO modifications are done to the cir­cuit boards. Any repairs should be performed
at the factory where the correct components will be used.
The flame tower is designed to be explosion proof. This is achieved by careful control of the spacing and clearances of the assembly, and the presence of the various flow restric-
tive devices and the outlet flame arrestor. It is
essential that no changes be made to this device that can impact any of these clear­ances, and thus that any repairs of the de­vice are such as to maintain its designed clearances. It is recommended that any re-
pairs of the flame tower be performed in the factory or at other authorized Rosemount ser­vice centers.
A temperature safety system activates when the burner flame temperature ex­ceeds 225°C for over 60 seconds. This will
shut off the fuel and display a message on the
light flame menu screen. After cooling to be-
low 225°C, the unit can be re-ignited normally.
Verify that the proper fuel is being used. Do not use H2 in a unit configured for mixed fuel.
Since the proper methane - non-methane process should be run under 200°C, you should not experience this failure mode. If it occurs, contact Rosemount Analytical factory service.
3-12 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 45
Model NGA2000 FID2

3-11 ALARM INDICATIONS

NGA Analyzers can also report specific alarm conditions through the 1/0 Modules. The FID2 Module is capable of reporting a flame out
DISPLAY MESSAGE DESCRIPTION TYPE
NORMAL No problem found
MAINTENANCE REQUEST Warning alarm noted Warning
FAILURE Failure alarm noted Failure
CAL. IN PROGRESS Performing a zero or span calibration
SPAN IN PROGRESS Performing a span calibration
Instruction Manual
748412-E
March 2003
condition and a valve failure as specific alarm indications.
The specific alarms are shown in Table 3-1 below.
ZERO FAILED The latest zero calibration failed Warning
SPAN FAILED The latest span calibration failed Warning
RANGE OVERFLOW The reading is higher than the current range upper limit Warning
RANGE UNDERFLOW
FLOW TOO LOW Sample pressure is below lower limit Warning
FLOW TOO HIGH Sample pressure is above upper limit Warning
RANGE 1, 2, 3, 4 Indicates current range
FLAME OUT Flame is off
FUEL SHUTOFF
The reading is lower than the next range down's upper limit
Fuel is shut off due to flame over 225°C. This is not a normal operating condition.
Table 3-1. FID2 Analyzer Module Alarms
Action
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-13
Page 46
Instruction Manual
748412-E March 2003

3-12 CONFIGURATION STORAGE

The FID2 analyzer module allows you to store a complete set of configuration variables once the unit has been set up to your satisfaction. This is in addition to the manufacturing vari­ables and the so-called "history" variables al­ready provided by earlier NGA analyzers.
If the configuration is damaged in some way it is possible to press a single softkey and re­place the damaged variables with those from the stored set.
Storing the variables can be done by entering
the Expert controls and set up menu, and
pressing the "Are you sure?" menu, giving you a chance to change your mind. If you want to store the variables, press the menu. To restore them all from the stored val­ues, press
Are you sure? STORE button saves var i ou s user settings for later usage.
RESTORE will restore the pre vio usl y sa ve d user se tti ng s.
HOME
Figure 3-18. Store/Restor e User Settings Menu
STORE button. This leads to an
STORE button in this
RESTORE.
Store/Restore user settings
ESCAPE
STORE RESTORE
INFO
Model NGA2000 FID2
There is another set of stored variables (Just as complete as this set) accessed through the
manufacturing data menu (under Technical set up, Service menus, Manufacturing data, Analyzer manufacturing data). Pressing the
STORE button in this menu leads to a similar
"Are you sure?" screen, with the same func­tionality. This set of variables is intended to store the initial set up as stored at the factory, so be wary of using it.
Are you sure? STORE button saves var i ou s user settings for later usage.
RESTORE will restore the pre viously saved user settings.
Store/Restore user settings
HOME
ESCAPE
STORE RESTORE
Figure 3-19. Store His torical Data Menu
For those who use a computer and the LON network to control this analyzer, the storage variables have the same names as the current variables, but with the addition of "ZZ" at the beginning for the user storage type, and "YY" at the beginning for the factory storage type.
INFO
3-14 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 47
Model NGA2000 FID2

MAINTENANCE AND SERVICE

Instruction Manual
748412-E
March 2003
SECTION 4
Operate this equipment only when covers are secured. Servicing requires access to live parts which can cause death or seri­ous injury. Refer servicing to qualified personnel.

4-1 GENERAL

When removing the cover of the FID2 Ana­lyzer Module for service or maintenance, re­move only the five screws along the bottom of each side, the seven screws on the front panel and the seven screws on the rear panel. Do not remove the three screws on the
of the front panel and rear panel. See Figure 4-2 on page 4-2.
REAR
Burner Block
DANDER
ELECTRICAL SHOCK HAZARD
Flow Control Manifold
Oven (Cover removed)
Figure 4-1. Location of Major Components
bottom
Regulator

4-2 OVERVIEW

The FID2 Analyzer Module requires very little maintenance during normal operation.
Service is usually limited to fuse replacement and burner repair or replacement. Preventa­tive maintenance is limited to checking and possible replacement of the exhaust tubing.
Refer to Figure 1-2. FID2 Analyzer Flow Dia­gram on page 1-3 and Figure 1-3. FID2 Wir­ing Diagram on page 1-4.
Intrinsic Safety Board
Computer Board
FRONT
Module Board
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-1
Page 48
Instruction Manual
748412-E March 2003

4-3 DISASSEMBLY

Refer to Figure 4-2. Removal of FID2 Cover below and Figure 4-6 on page 4-5 for illustra­tions of disassembly.

4-4 FUSES

The analyzer module has five fuses, all of which are located on the Module Board.
The main power fuse is accessible through the front panel of the instrument. See Figure 4-3. Main Power Fuse Location on page 4-3.
Model NGA2000 FID2
The remaining fuses are located on the solder side of the Module Board (the side facing downward). To access these four fuses, the Module Board must be removed from the ana­lyzer (see Figure 4-4. Fuse Locations on Module Board on page 4-3).
NOTE
Before replacing the fuses, remove power to the Analyzer Module.
Figure 4-2. Removal of FID2 Cover
4-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 49
Model NGA2000 FID2
FUEL OVERRIDE
IGNITE
POWER FLAME BLOCK SAMPLE IGNIT E FUEL/AIR OK
Figure 4-3. Main Power Fuse Location
R32
Figure 4-4. Fuse Locations on Module Board
Fuse
3 2 1 LON2 LON1
T 6A
24V
250 V
F3
F5
F4
Instruction Manual
748412-E
March 2003
FID 2
F2
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-3
Page 50
Instruction Manual
748412-E March 2003
4-5 BURNER BLOCK REMOVAL AND
INSTALLATION
CAUTION
Model NGA2000 FID2
8. Loosen the four hex nuts holding the burner block assembly in the oven. Slide the block towards the rear of the module and lift out.
QUALIFIED PERSONNEL
The burner block assembly should not be adjusted, cleaned or repaired by anyone except factory service personnel. Failure to observe this caution will void agency approvals.
CAUTION
BURNER
If a burner problem occurs, remove the burner block assembly from the oven and return it to Rosemount Analytical for main­tenance and/or repair (See Return Of Mate­rial on page 6-1).
The block is temperature controlled at 70°C. Allow unit to cool before touching any of these components.
Removal
1. Power OFF the module
Installation is the reverse of removal.
4-6 BURNER STARTUP AND
TROUBLESHOOTING
If burner startup is not achievable, check the following:
1. All supply gas cylinder pressures are within specification (see Specifications on page 1-7).
2. Correct gases are being supplied to each back panel inlet.
3. Air, not zero gas (if nitrogen), is being supplied to burner.
4. Burner exhaust is being vented to atmos­pheric pressure, and is not tied to either purge air exhaust or another FID exhaust.
5. Burner exhaust continuously slopes downward until reaching atmospheric pressure vent.
2. Shut off gas, air and sample gases.
3. Refer to Figure 4-2 on page 4-2. Remove the 24 screws securing the module cover. Remove the cover.
4. Refer to Figure 4-7 on page 4-6. Remove the four screws securing the oven cover. Remove the oven cover.
5. Refer to Figure 4-8 on page 4-7, and Figure 4-9 on page 4-8. Disconnect the RTD Cable, 90V Cable, Thermistor Cable, Signal Cable and Ignitor Cable.
6. Disconnect the Air, Fuel and Sample Cap-
6. IS parameters match WAS parameters in the Physical Measurement screen (Figure
4-5 below); use
MORE softkey to view all
parameters.
- Physical Measurement Parameters –
Sample capillary pressure: Sample capillary pressure was: Fuel supply pressure: Fuel supply pressure was: Carrier gas pressure: Carrier gas pressure was: Burner air pressure: Burner air pressure was: Pressure limits… Temperature limits…
HOME
ESCAPE INFOMORE
PRES_IS(1)
P_WAS(1)
PRES_IS(3)
P_WAS(3)
PRES_IS(5)
P_WAS(5)
PRES_IS(2)
P_WAS(2)
illaries.
Figure 4-5. Physical Measurements Menu
7. Disconnect the exhaust tube at the oven wall outlet.
4-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 51
Model NGA2000 FID2
Front Panel
Intrinsic Safety Board
Figure 4-6. Removal of Oven Cover
Module Board
Instruction Manual
748412-E
March 2003
Oven Cover
Computer Board
Regulator
Burner Block
Rear Panel
Flow Control Manifold
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-5
Page 52
Instruction Manual
r
748412-E March 2003
Front Panel
Mounting Bracket, Regulato
Regulator
Tube, Sample In
Model NGA2000 FID2
Computer Board
Module Board
Tube, Sample Bypass Out
Figure 4-7. FID2 – Exploded View
Block Heater
Burner Block
Tube, Exhaust
Burner Exhaust Out
Sample Bypass Out
Sample In
Flow Control Manifold
4-6 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 53
Model NGA2000 FID2
r
RTD Cable
90V Cable
Signal Cable
RTD
Base
Instruction Manual
748412-E
March 2003
Cove
Burner (see Figure 4-9)
Cover Plate
Heater
Mounting Plate
NOTE
This exploded view of the burner block is for information only. All servicing of the burner block must be performed by Rosemount Analytical.
Figure 4-8. Burner Block -Exploded View
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-7
Page 54
Instruction Manual
748412-E March 2003
Exhaust
Ignitor Cable (green)
90V Cable (coax)
Model NGA2000 FID2
Thermistor Cable (yellow)
Sample Capillary
Air Capillary
Fuel Capillary
NOTE
This view of the burner is for information only. All servicing of the burner must be performed by Rosemount Analytical.
Figure 4-9. Burner

4-7 MAINTENANCE SCHEDULE

ITEM EVENT TIME
Replace Exhaust Tubing If plastic cracked Approx. 5 years
4-8 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 55
Instruction Manual
748412-E
Model NGA2000 FID2
March 2003
SECTION 5

REPLACEMENT PARTS

WARNING
PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair.

5-1 MATRIX

Each analyzer is configured per the customer sales order. Below is the FID2 sales matrix which lists the various configurations avail­able.
FIDII NGA 2000 HYDROCARBON ANALYZER - FIDII (ANALYZER MODULE)
01 Current Software
02 2.5 Version Software
03 3X Version Software - specify version
Code Software Version
A1 0-10 to 0-5000 ppm CH4: Inlet Pressure 12.5 PSI, Restrictor 7.5#
A2 0-10 to 0-5000 ppm CH4: Inlet Pressure 7 PSI, Restrictor 5#
A3 0-10 to 0-5000 ppm CH4: Inlet Pressure 5 PSI, No Restrictor
A4 0-100 to 0-5000 ppm CH4: Inlet Pressure 12.5 PSI, Restrictor 7.5#
A5 0-100 to 0-5000 ppm CH4: Inlet Pressure 7 PSI, Restrictor 5#
A6 0-100 to 0-5000 ppm CH4: Inlet Pressure 5 PSI, No Restrictor
A7 0-100 to 0-10000 ppm CH4: Inlet Pressure 12.5 PSI, Restrictor 7.5#
A8 0-100 to 0-10000 ppm CH4: Inlet Pressure 7 PSI, Restrictor 5#
A9 0-100 to 0-10000 ppm CH4: Inlet Pressure 5 PSI, No Restrictor
G1 0-10 to 0-5000 ppm CH4: Inlet Pressure 2 PSI, No Restrictor
G2 0-100 to 0-5000 ppm CH4: Inlet Pressure 2 PSI, No Restrictor
G3 0-100 to 0-10000 ppm CH4: Inlet Pressure 2 PSI, No Restrictor
99 Special Factory Calibrated Ranges
Code Configurations, Mixed Fuel, Std Input Pressures, 5#Reg
To identify the configuration of an analyzer, locate the analyzer name-rating plate. The sales matrix identifier number appears on the analyzer name-rating plate.
Configurations, Mixed Fuel, Low Input Pressures, 2# Reg
00 None
G1 Customer Option
99 Special
FIDII 01 A1 00 Example
Code Special Requirements
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 5-1
Page 56
Instruction Manual
748412-E March 2003

5-2 REPLACEMENT PARTS

658350 Board Assembly, Computer 659060 Board Assembly, Module 659070 Board Assembly, Intrinsically safe 903347 Fuse, 6A (Main Power) 903823 Fuse, 4A (F2, F3 on Module Board) 903824 Fuse, 5A (F4, F5 on Module Board) 659063 Regulator 659225 Burner/Block Assembly 658144 Heater, Block 659043 Flow Control Manifold Assembly
The following components are selectable per application. Consult factory.
Fuel Capillary Restrictor, Flow Sample Capillary
Model NGA2000 FID2
5-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Page 57
Model NGA2000 FID2

RETURN OF MATERIAL

6-1 RETURN OF MATERIAL

If factory repair of defective equipment is re­quired, proceed as follows:
1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
2. Rosemount CSC will provide the shipping address for your instrument.
3. In no event will Rosemount be responsi­ble for equipment returned without proper authorization and identification.
4. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping.
5. In a cover letter, describe completely:
The symptoms that determined the
equipment is faulty.
The environment in which the equip-
ment was operating (housing, weather, vibration, dust, etc.).
Site from where the equipment was
removed.
Whether warranty or non-warranty
service is expected.
Complete shipping instructions for the
return of the equipment.
6. Enclose a cover letter and purchase order and ship the defective equipment accord­ing to instructions provided in the Rose­mount Return Authorization, prepaid, to the address provided by Rosemount CSC.
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
SECTION 6
Instruction Manual
If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the condi­tions listed in the standard Rosemount war­ranty, the defective unit will be repaired or replaced at Rosemount’s option, and an oper­ating unit will be returned to the customer in accordance with the shipping instructions fur­nished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.

6-2 CUSTOMER SERVICE

For order administration, replacement Parts, application assistance, on-site or factory re­pair, service or maintenance contract informa­tion, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076

6-3 TRAINING

A comprehensive Factory Training Program of operator and service classes is available. For
a copy of the Current Operator and Service
Training Schedule
vices Department at:
Rosemount Analytical Inc.
Customer Service Center
contact the Technical Ser-
1-800-433-6076
748412-E
March 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 6-1
Page 58
Instruction Manual
748412-E March 2003
Model NGA2000 FID2
6-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Page 59
Model NGA2000 FID2

APPENDIX A. MENU SCREENS

Menu: 0 ANALOP
Instruction Manual
748412-E
March 2003
SECTION 7
Menu: 1 ANALSET
Menu: 2 EXP_CAL
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Screens 7-1
Page 60
Instruction Manual
748412-E March 2003
Menu: 3 FLOCHEK
Model NGA2000 FID2
Menu: 4 ZEROI1
Menu: 5 FLOCHEKI1
7-2 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 61
Model NGA2000 FID2
Menu: 6 ANALOPI1
Instruction Manual
748412-E
March 2003
Menu: 7 ACALSET
Menu: 8 APARLST
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Screens 7-3
Page 62
Instruction Manual
748412-E March 2003
Menu: 9 ANALSETI1
Model NGA2000 FID2
Menu: 10 CALLISTI1
Menu: 11 ACALSETI1
7-4 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 63
Model NGA2000 FID2
Menu: 12 APARLSTI1
Instruction Manual
748412-E
March 2003
Menu: 13 AMMAN
Menu: 14 AMMANI1
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Screens 7-5
Page 64
Instruction Manual
748412-E March 2003
Menu: 15 AMSVC
Model NGA2000 FID2
Menu: 16 AMSVCI1
Menu: 17 ADIAG
7-6 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 65
Model NGA2000 FID2
Menu: 18 AMPWR
Instruction Manual
748412-E
March 2003
Menu: 19 AM1V
Menu: 20 AMTEMP
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Screens 7-7
Page 66
Instruction Manual
748412-E March 2003
Menu: 21 AMMISC
Model NGA2000 FID2
Menu: 22 AMTREND
Menu: 23 ADIAGI1
7-8 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 67
Model NGA2000 FID2
Menu: 24 RANGESETAM
Instruction Manual
748412-E
March 2003
Menu: 25 RANGESSETI1
Menu: 26 LINRANGE1
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Screens 7-9
Page 68
Instruction Manual
748412-E March 2003
Menu: 27 LINRANGE2
Model NGA2000 FID2
Menu: 28 LINRANGE3
Menu: 29 LINRANGE4
7-10 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 69
Model NGA2000 FID2
Menu: 30 LINRANGE0
Instruction Manual
748412-E
March 2003
Menu: 31 AMPWRI1
Menu: 32 FLOCHEK1I1
Page 70
Instruction Manual
748412-E March 2003
Menu: 33 FILTER
Model NGA2000 FID2
Menu: 34 AM1VI1
Menu: 35 AMTEMPI1
7-12 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 71
Model NGA2000 FID2
Menu: 36 AM2VA
Instruction Manual
748412-E
March 2003
Menu: 37 PLIMITSA
Menu: 38 TLIMITSA
Page 72
Instruction Manual
748412-E March 2003
Menu: 39 AMMISCI1
Model NGA2000 FID2
Menu: 40 ANALSIMPLE
Menu: 41 FILTERI1
7-14 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 73
Model NGA2000 FID2
Menu: 42 LINRANGE0I1
Instruction Manual
748412-E
March 2003
Menu: 43 PLIMITSAI1
Menu: 44 CALFACTORS
Page 74
Instruction Manual
748412-E March 2003
Menu: 45 R1FACTORS
Model NGA2000 FID2
Menu: 46 RN2FACTORS
Menu: 47 RN3FACTORS
7-16 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 75
Model NGA2000 FID2
Menu: 48 RN4FACTORS
Instruction Manual
748412-E
March 2003
Menu: 49 RFACTORSI
Menu: 50 AMHELPINDEX
Page 76
Instruction Manual
748412-E March 2003
Menu: 51 LINRANGE1I1
Model NGA2000 FID2
Menu: 52 AMTRENDI1
Menu: 53 CALFACTORSI1
7-18 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 77
Model NGA2000 FID2
Menu: 54 APARLST2
Instruction Manual
748412-E
March 2003
Menu: 55 APARLST4
Menu: 56 APARLST5
Page 78
Instruction Manual
748412-E March 2003
Menu: 57 APARLST6
Model NGA2000 FID2
Menu: 58 AMMISC2
Menu: 59 LIGHTFLAME
7-20 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 79
Model NGA2000 FID2
Menu: 60 IGNITION
Instruction Manual
748412-E
March 2003
Menu: 61 LISTNOTES
Menu: 62 LIGHTFLAMEI1
Page 80
Instruction Manual
748412-E March 2003
Menu: 63 AUTOFLAME
Model NGA2000 FID2
Menu: 64 AUTOFLAMEI1
Menu: 65 DISPLAY
7-22 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 81
Model NGA2000 FID2
Menu: 66 MPARMS
Instruction Manual
748412-E
March 2003
Menu: 67 MPARMS2
Menu: 68 ANALSET2
Page 82
Instruction Manual
748412-E March 2003
Menu: 69 MPARMSI1
Model NGA2000 FID2
Menu: 70 TLIMITSAI1
Menu: 71 IGNITIONI1
7-24 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 83
Model NGA2000 FID2
Menu: 72 SELFTEST
Instruction Manual
748412-E
March 2003
Menu: 73 AMHELPINDEX2
Menu: 74 SOFT_DIAG
Page 84
Instruction Manual
748412-E March 2003
Menu: 75 CALI1
Model NGA2000 FID2
Menu: 76 AM2VC
Menu: 77 AM2VD
7-26 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 85
Model NGA2000 FID2
Menu: 78 RFHIST
Instruction Manual
748412-E
March 2003
Menu: 79 RFHIST2
Menu: 80 RFHIST3
Page 86
Instruction Manual
748412-E March 2003
Menu: 81 RFHIST4
Model NGA2000 FID2
Menu: 82 LINFUNCT
Menu: 83 POLYSETUP
7-28 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 87
Model NGA2000 FID2
Menu: 84 MIDPOINT1
Instruction Manual
748412-E
March 2003
Menu: 85 ANALSETI2
Menu: 86 POLYGAS1
Page 88
Instruction Manual
748412-E March 2003
Menu: 87 POLYSETI1
Model NGA2000 FID2
Menu: 88 POLYGAS2
Menu: 89 POLYGAS3
7-30 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 89
Model NGA2000 FID2
Menu: 90 POLYGAS4
Instruction Manual
748412-E
March 2003
Menu: 91 POLYGAS5
Menu: 92 POLYGAS6
Page 90
Instruction Manual
748412-E March 2003
Menu: 93 POLYGAS7
Model NGA2000 FID2
Menu: 94 POLYGAS8
Menu: 95 POLYGAS9
7-32 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 91
Model NGA2000 FID2
Menu: 96 POLYGAS0
Instruction Manual
748412-E
March 2003
Menu: 97 MIDPOINT2
Menu: 98 MIDPOINT3
Page 92
Instruction Manual
748412-E March 2003
Menu: 99 MIDPOINT4
Model NGA2000 FID2
Menu: 100 ZEROI2
Menu: 101 EXP_CAL_DATA
7-34 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 93
Model NGA2000 FID2
Menu: 102 EXP_CAL_DATI
Instruction Manual
748412-E
March 2003
Menu: 103 UNITS
Menu: 104 UNITSI1
Page 94
Instruction Manual
748412-E March 2003
Menu: 105 POLYSETI2
Model NGA2000 FID2
Menu: 106 ANALSETI3
Menu: 107 POLYSETI3
7-36 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 95
Model NGA2000 FID2
Menu: 108 RESET
Instruction Manual
748412-E
March 2003
Menu: 109 STORE
Menu: 110 REBOOT
Page 96
Instruction Manual
748412-E March 2003
Menu: 111 SW_DIAGI1
Model NGA2000 FID2
Menu: 112 ZERO_NOW2
Menu: 113 SPAN_NOW2
7-38 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 97
Model NGA2000 FID2
Menu: 114 CALLIST
Instruction Manual
748412-E
March 2003
Menu: 115 ABOUT
Menu: 116 ABOUT1
Page 98
Instruction Manual
748412-E March 2003
Menu: 117 ALARM1
Model NGA2000 FID2
7-40 Appendix A. Menu Screens Rosemount Analytical Inc. A Division of Emerson Process Management
Page 99

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and ma­terial under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the com­mercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WAR­RANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DE-
LAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WAR­RANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF AC­TION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDEN­TAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVE­NUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond
Seller's direct control.
Page 100
Instruction Manual
A
748412-E March 2003
Model NGA2000 FID2
EMERSON PROCESS MAN
Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd. 1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
GEMENT
EUROPEAN TECHNOLOGY CENER Fisher-Rosemount GmbH & Co. Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd. Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...