Rosemount Model 500 Gas Chromatograph Hardware Reference Rev K Manuals & Guides

Model 500
Gas Chromatograph
Hardware Reference Manual
Part Number 3-9000-537
Revision K
JULY 2010
Model 500 Gas Chromatograph
System Reference Manual
ROSEMOUNT (“SELLER”) SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS IN THIS MANUAL OR OMISSIONS FROM THIS MANUAL. SELLER MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL SELLER BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, ETC.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES.
THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
SELLER DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND MAINTENANCE OF ANY SELLER PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.
THE ROSEMOUNT LOGO IS A REGISTERED TRADEMARK OF ROSEMOUNT, INC. THE EMERSON LOGO IS A TRADEMARK AND SERVICE MARK OF EMERSON ELECTRIC CO.
COPYRIGHT
All rights reserved. No part of this work may be reproduced or copied in any form or by any means - graphic, electronic, or mechanical — without first receiving the written permission of Rosemount, Inc. Shakopee, MN. U.S.A.
©
2010 BY Rosemount, HOUSTON, TX U.S.A.
WARRANTY
1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise expressly provided herein, Rosemount, (“Seller”) warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of the Services. Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller rep-resentative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller. Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESSED OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. It
is understood that corrosion or erosion of materials is not covered by our guarantee.
2. LIMITATION OF REMEDY AND LIABILITY: SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY HEREUNDER SHALL BE LIMITED TO REPAIR, CORRECTION, REPLACEMENT OR REFUND OF PURCHASE PRICE UNDER THE LIMITED WARRANTY CLAUSE IN SECTION 1 HEREIN. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED IN CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERWISE), SHALL SELLER'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED OR SERVICES PROVIDED BY SELLER GIVING RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES. THE TERM "CONSEQUENTIAL DAMAGES" SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE AND COST OF CAPITAL.
IMPORTANT INSTRUCTIONS
Read all instructions prior to installing, operating, and servicing this product.
Follow all warnings, cautions, and instructions marked on and supplied with this product.
Inspect the equipment packing case and if damage exists, notify your local carrier for liability.
Open the packing list and carefully remove equipment and spare or replacement parts from the case. Inspect all equipment for damage and missing parts.
If items are damaged or missing, contact the manufacturer at 1 713 396 8880 for instructions about receiving replacement parts.
Install equipment as specified per the installation instructions and per applicable local and national codes. All connections shall be made to proper electrical and pressure sources.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent personal injury.
Use of this product for any purpose other than its intended purpose may result in property damage and/or serious injury or death.
Before opening the flameproof enclosure in a flammable atmosphere, the electrical circuits must be interrupted.
Repairs must be performed using only authorized replacement parts as specified by the manufacturer. Use of unauthorized parts can affect the product's performance and place the safe operation of the product at risk.
When installing or servicing ATEX-certified units, the ATEX approval applies only to equipment without cable glands. When mounting the flameproof enclosures in a hazardous area, only flameproof cable glands certified to IEC 60079-1 must be used.
Technical assistance is available 24 hours a day, 7 days a week by calling 1 713 396 8880.
This page is intentionally left blank.
TABLE OF CONTENTS
DESCRIPTION 1.1 PURPOSE OF THIS MANUAL ..........................1-1
1.1.1 Section 1 Description ...................................1-1
1.1.2 Section 2 Equipment Description....................1-1
1.1.3 Section 3 Installation and Startup...................1-1
1.1.4 Section 4 Operation......................................1-1
1.1.5 Section 5 Maintenance .................................1-2
1.1.6 Section 6 Recommended Spare Parts..............1-2
1.1.7 Appendices ..................................................1-2
1.2 INTRODUCTION............................................1-2
1.3 FUNCTIONAL DESCRIPTION...........................1-4
1.4 MODES OF OPERATION.................................1-6
1.4.1 User Interfaces .............................................1-6
EQUIPMENT
DESCRIPTION
1.4.2 Capabilities...................................................1-6
1.5 THEORY OF OPERATION ...............................1-7
1.5.1 The Analyzer Detector ...................................1-7
1.5.2 Data Acquisition .........................................1-10
1.5.3 Peak Detection ...........................................1-11
1.5.4 Basic Analysis Computations ........................ 1-12
1.6 GLOSSARY ................................................1-16
2.1 SAMPLING SYSTEM......................................2-1
2.1.1 Sampling Point Location.................................2-2
2.1.2 Sample Volume and Flow Rate........................2-2
2.1.3 Sample Conditioning......................................2-3
2.1.4 Contamination Precautions .............................2-3
2.1.5 Valving ........................................................2-3
2 TABLE OF CONTENTS
Model 500
2.1.6 Calibration Gas .............................................2-4
2.2 ANALYZER...................................................2-5
2.2.1 Physical Description ......................................2-5
2.2.2 Chromatograph Valves................................... 2-7
2.2.3 Detector Subsystem ...................................... 2-8
2.2.4 Analyzer Preamplifier Unit ..............................2-9
2.2.5 Analyzer Specifications ..................................2-9
2.2.6 Utility Gas Requirements .............................. 2-10
2.3 CONTROLLER ............................................. 2-10
2.3.1 Controller Hardware Configurations ...............2-12
2.3.2 Optional Keypad and Display ........................ 2-25
2.3.3 Alarm Specifications.................................... 2-26
INSTALLATION AND
SETUP
3.1 PRECAUTIONS AND WARNINGS .................... 3-3
3.1.1 Hazardous Environments ................................3-4
3.1.2 Power Source Wiring .....................................3-5
3.1.3 Signal Wiring ................................................3-6
3.1.4 Electrical and Signal Ground ...........................3-8
3.1.5 Electrical Conduit ........................................ 3-10
3.1.6 Sample Systems Requirements ..................... 3-11
3.2 PREPARATION............................................ 3-12
3.2.1 Introduction................................................ 3-12
3.2.2 Site Selection .............................................3-13
3.2.3 Unpacking the Unit...................................... 3-14
3.2.4 Necessary Tools and Components................. 3-15
3.2.5 Optional Tools and Components.................... 3-16
3.3 INSTALLING THE ANALYZER .......................3-18
3.3.1 Point-to-point Wiring Guide,
Analyzer-Controller...................................... 3-18
Installation and Setup JULY 2010
Model 500
TABLE OF CONTENTS 3
3.3.2 Analyzer AC Power Wiring ........................... 3-26
3.3.3 Sample and Gas Lines..................................3-27
3.4 INSTALLING THE GC CONTROLLER ..............3-30
3.4.1 Modbus Slave Address (COM ID) Setup .........3-30
3.4.2 Controller-Analyzer Wiring ............................3-35
3.4.3 Controller PC Wiring (Serial Connections) .......3-37
3.4.4 CPU and COM4A Serial Communications
Setups .......................................................3-45
3.4.5 Controller Printer Wiring ...............................3-69
3.4.6 Discrete (Digital) I/O Wiring ..........................3-71
3.4.7 Analog I/O Wiring........................................3-74
3.4.8 Controller AC Power Wiring.......................... 3-77
OPERATION FROM
LOCAL KEYPAD AND
DISPLAY
3.5 ANALYZER LEAK CHECKS AND PURGING FOR
FIRST CALIBRATION ...................................3-78
3.5.1 Analyzer Leak Checks.................................. 3-78
3.5.2 Purging Carrier Gas Lines ............................. 3-80
3.5.3 Purging Calibration Gas Lines........................3-83
3.6 SYSTEM START-UP ....................................3-85
4.1 INTERFACE COMPONENTS FOR LOCAL DATA
DISPLAY AND ENTRY....................................4-3
4.1.1 Light Emitting Diode (Led) Indicators................ 4-3
4.1.2 Liquid Crystal Display (LCD) ...........................4-4
4.1.3 Keypad ........................................................4-4
4.2 LOGGING ON TO VIEW OR EDIT DATA ...........4-6
4.2.1 First Time Log-On..........................................4-6
4.2.2 Subsequent Log-On .......................................4-7
4.2.3 Start / Halt an Auto Sequence Analysis............4-8
4.2.4 Editing Procedures....................................... 4-10
4.2.5 Validity Checks of Data Entries .....................4-11
JULY 2010 Operation from Local Keypad and Display
4 TABLE OF CONTENTS
MAINTENANCE 5.1 TROUBLESHOOTING AND REPAIR CONCEPT ...5-2
Model 500
4.3 LOCAL DISPLAY MENUS ............................. 4-12
4.3.1 Main Menu .................................................4-14
4.3.2 Hardware Menu ..........................................4-14
4.3.3 Operator Entries Menu ................................. 4-15
4.3.4 Alarms Menu ..............................................4-16
4.3.5 Chromatogram Menu ................................... 4-16
4.3.6 GC Control Menu ........................................ 4-16
4.3.7 Data Records Menu ..................................... 4-17
4.3.8 Config Rpt - Maint. Log Menu....................... 4-17
5.2 ROUTINE MAINTENANCE...............................5-3
5.2.1 Model 500 Maintenance Checklist...................5-4
5.2.2 Routine Maintenance Procedures.....................5-5
5.2.3 Contract Service ...........................................5-5
5.3 LOCATING AND GAINING ACCESS TO EQUIPMENT
ELEMENTS ...................................................5-5
5.3.1 Analyzer Electrical/Electronic Units .................. 5-5
5.3.2 Detector Elements, Heater Elements, Valves and
Columns ......................................................5-7
5.4 PRECAUTIONS FOR HANDLING PRINTED CIRCUIT
ASSEMBLIES ................................................ 5-7
5.5 SERVICE, TROUBLESHOOTING, AND REPAIR
INSTRUCTIONS ............................................5-8
5.5.1 Preamplifier ................................................5-10
5.5.2 Temperature Control.................................... 5-10
5.5.3 Decoder .....................................................5-10
5.5.4 Analyzer Troubleshooting Guide .................... 5-13
5.5.5 Chromatograph Valves................................. 5-22
5.5.6 Detector Bridge Balance ...............................5-23
Maintenance JULY 2010
Model 500
TABLE OF CONTENTS 5
5.5.7 Temperature Measurements .........................5-25
5.5.8 Measure Vent Flow (MV) .............................5-27
5.5.9 Analog Inputs ............................................. 5-27
5.6 GC CONTROLLER MAINTENANCE................. 5-28
5.6.1 GC Controller Access...................................5-29
5.7 COMMUNICATIONS ....................................5-30
5.7.1 GC Controller Address Change...................... 5-31
5.8 ANALOG INPUTS AND OUTPUTS ................. 5-31
5.8.1 Analog Output Dialog Description.................. 5-32
5.8.2 Changing a Variable.....................................5-34
5.8.3 Changing the Bargraph.................................5-35
5.8.4 Performing a Manual Calibration.................... 5-37
RECOMMENDED SPARE
PARTS
5.8.5 Performing an Automated Calibration............. 5-39
5.8.6 Analog Loopback Test Circuits......................5-41
5.8.7 Upgrading Analog Outputs ...........................5-43
5.9 DISCRETE (DIGITAL) INPUTS AND OUTPUTS . 5-44
5.9.1 Digital Loopback Test Circuit ........................5-44
5.10 FUSE PROTECTION ..................................... 5-45
5.11 ANALYZER-CONTROLLER INTERCONNECT ....5-46
5.11.1 Function Codes........................................... 5-48
6.1 ANALYZER SPARES ......................................6-1
6.1.1 Printed Circuit Card Assemblies (Analyzer) .......6-1
6.1.2 Electrical and Mechanical Assemblies (Analyzer)6-1
6.2 GC CONTROLLER SPARES .............................6-3
6.2.1 Printed Circuit Card Assemblies
(GC Controller)..............................................6-3
6.2.2 Electrical and Mechanical Components (GC
Controller) ....................................................6-3
JULY 2010 Recommended Spare Parts
6 TABLE OF CONTENTS
Model 500
APPENDIX A,
SUPPLEMENTAL
WIRING GUIDE - SERIAL
COMMUNICATIONS
A.1 GC SERIAL PORT AND CABLE CONFIGURATIONS
FOR RS-232 .................................................A-1
A.2 RS-232 CONNECTION FROM GC CONTROLLER TO
PC...............................................................A-4
A.2.1 DB-9 Serial Port of GC to DB-9 Port of PC........A-4
A.2.2 DB-9 Serial Port of GC to DB-25 Port of PC ......A-5
A.2.3 Phoenix Plug Port of GC to DB-9 Port of PC......A-6
A.2.4 Phoenix Plug Port of GC to DB-25 Port of PC....A-7
A.3 RS-232 CONNECTION FROM GC CONTROLLER TO
EXTERNAL MODEM ......................................A-8
A.3.1 DB-9 Serial Port of GC to DB
Modem
........................................................A-8
A.3.2 Phoenix Plug Port of GC to DB-25 Port
Modem
........................................................A-9
-25 Port of
of
A.4 EXAMPLE RS-422 CONNECTION FROM PC TO
GC ............................................................A-10
APPENDIX B,
MANIFOLD FOR TWO
CARRIER GAS BOTTLES
TO GC SYSTEM
APPENDIX C, GUIDE TO
TRANSIENT
PROTECTION MODULES
A.5 EXAMPLE RS-485 CONNECTION FROM PC TO
GC ............................................................A-12
A.6 JUMPER-BASED SERIAL CHANNEL
CONFIGURATIONS......................................A-15
B.1 ILLUSTRATION ............................................. B-2
B.2 INSTALLATION AND LINE PURGING................ B-3
B.3 REPLACING CARRIER CYLINDER ....................B-4
C.1 PURPOSE OF THE TRANSIENT PROTECTION
MODULES....................................................C-1
C.2 PART APPLICATIONS, NUMBERS, AND
DESCRIPTIONS.............................................C-2
Appendix A, Supplemental Wiring Guide - Serial Communications JULY 2010
Model 500
APPENDIX D, INTERNAL
MODEM FOR GC
CONTROLLER
APPENDIX E, SETTING
SOLENOID PURGE
FLOWS
TABLE OF CONTENTS 7
C.3 TROUBLESHOOTING TRANSIENT PROTECTION
MODULES ....................................................C-4
APPENDIX F,
UPGRADING FROM
2251 TO 2350A GC
CONTROLLER
APPENDIX G, 2350 TO
2350A CPU RETROFIT
INSTRUCTIONS
F.1 HALT CURRENT ANALYSIS AND POWER
DOWN .........................................................F-1
F.2 NOTE EXISTING WIRING CONNECTIONS TO
2251 ...........................................................F-2
F.3 REMOVE CABLES, REPLACE CONTROLLER, AND
RECONNECT ................................................ F-5
G.1 2350A CPU ASSEMBLY INTRODUCTION........ G-1
G.2 CONVERSION PROCESS ............................... G-2
G.3 BASIC 2350A CONFIGURATION .................... G-7
G.4 2350A OPTIONS.......................................... G-9
G.4.1 The COM4A Board ....................................... G-9
G.5 TO ADD A MODEM .....................................G-12
G.6 TO ADD AN ETHERNET CARD......................G-15
APPENDIX H, 2350A
H.1 INTRODUCTION........................................... H-2
(LX-800) CPU RETROFIT
INSTRUCTIONS
H.2 REMOVING THE OLD CPU CARD ................... H-4
H.3 INSTALLING THE NEW CPU BOARD ............... H-6
H.4 INSTALLING A KEYBOARD AND DISPLAY WITH
COM4A BOARD........................................... H-7
JULY 2010 Appendix D, Internal Modem for GC Controller
8 TABLE OF CONTENTS
APPENDIX F, SHIPPING
AND LONG-TERM
STORAGE
RECOMMENDATIONS
ADDENDUM 1
ADDENDUM 2
Model 500
H.5 INSTALLING A KEYBOARD AND DISPLAY
WITHOUT COM4A BOARD.............................H-9
H.6 RETROFITTING THE ANALOG BOARD ...........H-11
H.6.1 Identifying your Analog Board .......................H-11
H.6.2 Installing a New Analog Board ......................H-13
H.7 REASSEMBLING THE 2350A ........................H-13
Appendix F, Shipping and long-term storage recommendations JULY 2010
DESCRIPTION
1.1 PURPOSE OF THIS MANUAL
This manual (P/N 3-9000-537) is intended as a user's guide to accompany the Model 500 gas chromatograph.
For software operation instructions, see the MON2000 Software for Gas Chromatographs User Manual (P/N 3-9000-522)
This manual provides the following information:
1.1.1 Section 1 Description
A general description of the Model 500 gas chromatograph (GC) and its components, their configurations and functions.
A brief description of the GC's software, user interface, and capabilities.
Introduction to GC theory of operation and terminology.
1.1.2 Section 2 Equipment Description
Guidelines for sampling system and gas connections.
Descriptions of Analyzer subsystems and components.
Descriptions of GC Controller subsystems and components.
1.1.3 Section 3 Installation and Startup
Instructions for installing the GC hardware.
1.1.4 Section 4 Operation
Instructions for operating the GC by means of its built-in keypad and liquid crystal display (LCD), if provided.
1-2 Description
1.1.5 Section 5 Maintenance
Instructions for regular maintenance and care of the GC hardware.
Instructions for troubleshooting, repair, and service of the GC hardware.
1.1.6 Section 6 Recommended Spare Parts
List of boards, valves, and other components suggested as spare parts.
1.1.7 Appendices
Appendices with additional, helpful reference materials and drawings.
1.2 INTRODUCTION
The Model 500 GC is a high-speed gas chromatograph that is factory engineered to meet specific field application requirements based on typical stream composition and the anticipated concentration of the components of interest. The Model 500 typically consists of three major components, the Analyzer Assembly, the Controller, and the Sample Conditioning System:
Model 500 Gas Chromatograph
Analyzer Assembly (Model 500 Series) - Located near the sample tap. The Analyzer includes GC columns, detectors, preamplifier, stream switching valves, and solenoids. The Model 500 Analyzer is housed in a National Electrical Manufacturers Association (NEMA) 7, National Electrical Code (NEC) Class I, Division 1, Groups B, C, and D approved enclosure, for use in a hazardous environment.
GC Controller - Located no further than 2000 feet (610 meters) away from the Analyzer. The GC Controller includes electronics and ports for signal processing, instrument control, data storage, personal computer (PC) interface, and telecommunications. The GC Controller is available in various enclosures and configurations, as follows:
Explosion Proof - NEMA 4X (weatherproof and corrosion resistant) and NEMA 7, NEC Class I, Division 1, Groups B, C, and D approved enclosure, for use in a hazardous environment. Available with or without a built-in keypad and liquid crystal display (LCD).
Section 5 Maintenance JULY 2010
Model 500 Gas Chromatograph
Rack Mount - Suitable for use in a nonhazardous environment. Made for mounting on a standard 19 inch rack. Available with or without a built-in keypad and LCD.
Retrofit - Suitable for use in a nonhazardous environment. Made for mounting on a 12-inch rack previously sold for 2251 GC Controller. The Retrofit enclosure is not available with a built-in keypad and LCD (therefore, a PC is required for operating).
Sample Conditioning System (SCS) - Located between the process stream and the Analyzer sample inlet, usually mounted on the lower portion of the Analyzer stand. The standard configuration SCS includes a mounting plate, block (or shutoff) valves, and filters. Optionally, the SCS can be configured with Genie® bypass filters, liquid shut-off valves, and optional solenoids for stream switching: all of which can be enclosed in an electric (heat tape design) oven.
Description 1-3
In its standard configuration, the Model 500 Analyzer can handle up to five streams: typically, four for sample and one for calibration. With an optional stream switch assembly added, the Model 500 Analyzer can switch up to twelve streams, maximum.
The GC Controller, is designed to be operated primarily from a personal computer (PC) running the MON2000 Software package. This provides the user with the greatest capability, ease-of-use, and flexibility. One PC running MON2000 can connect with up to 32 gas chromatographs. The PC is used to display chromatograms and reports, which can then be stored as files on the PC, or printed from either the PC's or the GC's printer.
Also, each individual GC can be operated from its built-in keypad and LCD (if installed in that configuration); however, this method offers more limited functions. Display of the chromatograms on the LCD is accomplished in scrolling strip chart fashion.
Since neither the PC nor a normal printer can be placed in a hazardous area, serial port and Modbus communications links are provided for connecting the Model 500 Analyzer to the PC, other computers, printers, chromatographs, and Controllers.
JULY 2010 INTRODUCTION
1-4 Description
1.3 FUNCTIONAL DESCRIPTION
A functional block diagram of a typical Model 500 Analyzer installation is shown in Figure 1-1. A sample of the gas to be analyzed is taken from the process stream by a sample probe installed in the process line. The sample passes through a sample line to the sample conditioning system where it is filtered or otherwise conditioned. After conditioning, the sample flows to the Analyzer for separation and detection of the components of the gas.
The chromatographic separation of the sample gas into its components is accomplished in the Analyzer in the following manner. A precise volume of sample gas is injected into one of the unit's analytical columns. The column contains a stationary phase (packing) that is either an active solid (adsorption partitioning) or an inert solid support that is coated with a liquid phase (absorption partitioning). The gas sample is moved through the column by means of a mobile phase (carrier gas). Selective retardation of the components of the sample takes place in the column that causes each component to move through the column at a different rate. This action separates the sample into its constituent gases and vapors.
Model 500 Gas Chromatograph
A detector located at the outlet of the analytical column senses the elution of components from the column and produces electrical outputs proportional to the concentration of each component. Outputs from the Analyzer detectors are amplified in the Analyzer electronics, then transmitted to the GC Controller for further processing. See also, Section
1.5, “Theory of operation” on page 7.
Output from the GC Controller is normally displayed on a remotely located personal computer (PC) or a printer. Connection between the GC Controller and the PC can be accomplished via a direct serial line, the Modbus-compatible communication interface, modem or ethernet card.
Multiple chromatograms may be displayed on the PC monitor, and compared or contrasted with separate color schemes. This allows a stored chromatogram to be compared/contrasted with a current or another stored chromatogram. This could be of great assistance when changing parameters or isolating a problem.
FUNCTIONAL DESCRIPTION JULY 2010
Model 500 Gas Chromatograph
Use of a PC for configuration and troubleshooting procedures is essential in most instances. Basic operations can also be performed from a keypad and liquid crystal display that are built into certain versions of the GC Controller. The PC may be remotely connected via telephone, radio, ethernet or satellite communications. Once installed and configured, the Model 500 Analyzer can operate independently for long periods of time.
Description 1-5
Figure 1-1 Model 500 Analyzer Functional Block Diagram
JULY 2010 FUNCTIONAL DESCRIPTION
1-6 Description
1.4 MODES OF OPERATION
1.4.1 User Interfaces
You have at least one, and optionally two, user interfaces from which to operate the gas chromatograph (GC) system:
PC connected to the GC and running MON2000 - The PC connected to the GC and running MON2000 offers the greatest amount of capability and flexibility.
Find complete user instructions for MON2000 in the program's online HELP screens and in the program user's manual, MON2000 Software for Gas Chromatographs User Manual (P/N 3-9000-522).
or
The GC Controller's built-in keypad and LCD - The GC Controller's built-in keypad and LCD offer essential startup and operation functions. They are useful in a hazardous environment or if no PC is available.
Model 500 Gas Chromatograph
This feature is optional on all standalone models of the GC, except the portable Compact BTU GC.
1.4.2 Capabilities
Individual gas chromatograph Controller functions that can be initiated or controlled by the GC and its software, MON2000, are listed in the MON2000 Software for Gas Chromatographs User Manual (P/N 3-9000-
522).
Modes of operation JULY 2010
Model 500 Gas Chromatograph
1.5 THEORY OF OPERATION
See also Section 1.6, “GLOSSARY” on page 16, for definitions of some of the terminology used in the following explanations.
1.5.1 The Analyzer Detector
The Analyzer detector subsystem is a thermal conductivity detector that consists of a balanced bridge network with heat-sensitive thermistors in each leg of the bridge. Each thermistor is enclosed in a separate chamber of the detector block. One thermistor is designated the reference element and the other the measurement element. A schematic diagram of the thermal conductivity detector is shown in Figure 1-2.
Description 1-7
JULY 2010 Theory of operation
1-8 Description
Model 500 Gas Chromatograph
Figure 1-2 Schematic Diagram of Analyzer Detector Bridge
In the quiescent condition (prior to injecting a sample), both legs of the bridge are exposed to pure carrier gas. In this condition, the bridge is balanced and the bridge output is electrically nulled. (The bridge can be balanced by the fine and coarse adjustment potentiometers located on the preamplifier circuit board.)
The analysis begins when a fixed volume of sample is injected into the column by operation of the sample valve. The sample is moved through the column by the continuous flow of carrier gas. As successive components elute from the column, the temperature of the measurement element changes. The temperature change unbalances the bridge and produces an electrical output proportional to the component concentration. The differential signal developed between the two thermistors is amplified by the preamplifier.
The Analyzer Detector JULY 2010
Model 500 Gas Chromatograph
Figure 1-3 illustrates the change in detector electrical output during
elution of a component.
Figure 1-3 Detector output during component elution
1. Detector bridge balanced.
Description 1-9
2. First component begins to elute from column and to be sensed by the
measurement thermistor.
3. Peak concentration of first component.
4. Second component begins to elute from column and to be sensed by the
measurement thermistor.
5. Peak concentration of second component.
In addition to amplifying the differential signal developed between the detector's two thermistors, the preamplifier also supplies drive current to the detector bridge. The preamplifier also supplies drive current to the detector bridge. The voltage signal is converted to a 4 to 20-milliamp (mA) current loop for transmission to the GC Controller. The signal is proportional to the concentration of a component detected in the gas sample. The preamplifier provides four different gain channels as well as compensation for baseline drift. The signals from the preamplifier are sent to the GC Controller for computation, recording on a printer, or viewing on a PC monitor or LCD.
JULY 2010 The Analyzer Detector
1-10 Description
1.5.2 Data Acquisition
Every second, exactly 40 equi-spaced data samples are taken for analysis by the GC Controller (i.e., once every 25 milliseconds). Each data sample, after having been precision-amplified, is subjected to a sixteen bit analog to digital (A/D) conversion. The sampling frequency of 40 Hertz (Hz) was chosen to reduce 60 Hz normal mode noise.
After each point on the chromatograph signal is sampled, the resulting number is stored in a buffer area in the GC Controller's memory for processing. During the analysis, only the last 256 data points are available for processing. Because the data analysis is done as the signal is sampled (in real-time), only a limited number of past data samples is required to analyze any signal.
As a part of the data acquisition process, groups of incoming data samples are averaged together before the result is stored to the Controller's memory for processing. Non-overlapping groups of N samples are averaged and stored, and thus reduce the effective incoming data rate to 40/N samples/second. For example, if N = 5, then a total of 40/5 or 6 (averaged) data samples are stored every second. The value for the variable N is determined by the selection of a Peak Width parameter (PW). The relationship is:
Model 500 Gas Chromatograph
N PW ondssec=
where PW is given in seconds. All the various details in the analysis process are independent of the value of N. Allowable values of N are 1 to 63, which corresponds to values of PW from 1 to 63 seconds.
The variable N is known as the integration factor. This term is used because N determines how many points are averaged or integrated to form a single value. The integration of data upon input, before storing, serves two purposes. First, the statistical noise on the input signal is reduced by the square root of N. In the case of N = 4, a noise reduction of two would be realized. Secondly, the integration factor controls the bandwidth of the chromatograph signal. It is necessary to match the bandwidth of the input signal to that of the analysis algorithms in the GC
Data Acquisition JULY 2010
Model 500 Gas Chromatograph
Controller. This prevents small, short duration perturbations from being recognized as true peaks by the program. It is therefore important to choose a Peak Width corresponding to the narrowest peak in a group under consideration.
1.5.3 Peak Detection
For normal area or peak height concentration evaluation, the determination of a peak's start, peak point, and end is automatic. The manual determination of start and end points is used only for area calculations in the Forced Integration mode. Automatic determination of peak onset or start is initiated whenever Integrate Inhibit is turned off. Analysis is started in a region of signal quiescence and stability, such that the signal level and activity can be considered as baseline values. It is important that this be the case because this assumption is made by the GC Controller.
Description 1-11
Having initiated a peak search by turning Inhibit off, the GC Controller performs a point by point examination of the signal slope. This is achieved by using a digital slope detection filter which is a combination low pass filter and differentiator. The output of this detector is constantly compared to a system constant entered by the operator called Slope Sensitivity. A default value of 8 is assumed if no entry is made. Lower values make peak onset detection more sensitive, and higher values make detection less sensitive. Higher values (20 to 100) would be appropriate for noisy signals, e.g. high amplifier gain.
Peak termination is determined by the same application of this detector to the signal, but in the reverse sense. Onset is defined where the detector output exceeds the baseline constant, but termination is defined subsequently where the detector output is less than the same baseline constant.
Sequences of fused peaks are also automatically handled. This is done by testing each termination point to see if the region immediately following it satisfies the criteria of a baseline. A baseline region must have a slope detector value less than the magnitude of the baseline constant for a number of sequential points. When a baseline region is found, this terminates a sequence of peaks.
JULY 2010 Peak Detection
1-12 Description
Model 500 Gas Chromatograph
A zero reference line for peak height and area determination is established by extending a line from the point of the onset of the peak sequence to the point of the termination. The values of these two points are found by averaging the four integrated points just prior to the onset point and just after the termination points, respectively. The zero reference line will, in general, be non-horizontal, and thus compensates for any linear drift in the system from the time the peak sequence starts until it ends.
In a single peak situation, peak area is the area of the component peak between the curve and the zero reference line. The peak height is the distance from the zero reference line to the maximum point on the component curve. The value and location of the maximum point is determined from quadratic interpolation through the three highest points at the peak of the discrete valued curve stored in the GC Controller.
For fused peak sequences, this interpolation technique is used both for peaks as well as valleys (minimum points). In the latter case, lines are dropped from the interpolated valley points to the zero reference line to partition the fused peak areas into individual peaks. The use of quadratic interpolation improves both area and height calculation accuracy and eliminates the effects of variations in the integration factor on these calculations.
For calibration, the GC Controller may average several analyses of the calibration stream.
1.5.4 Basic Analysis Computations
Two basic analysis algorithms are included in the GC Controller. These are:
Area Analysis - Calculates area under component peak
Peak Height Analysis - Measures height of component peak
Concentration Analysis by Using Response Factor
Concentration calculations require a unique response factor foreach component in an analysis. These response factors may be manually entered by an operator or determined automatically by the system
Basic Analysis Computations JULY 2010
Model 500 Gas Chromatograph
through calibration procedures (with a calibration gas mixture that has known concentrations).
Response factor calculation: (using the external standard)
or
where:
ARF
HRF
n
=
=
n
Area
-------------- -
Cal
Ht
-----------
Cal
Description 1-13
n
n
n
n
ARF
HRF
Area
Ht
Cal
Area response factor for component n in area per mole percent (%).
n
Height response factor for component n.
n
Area associated with component n in calibration gas.
n
Height associated with component n in mole percent in calibration gas.
n
Amount of component n in mole percent in calibration gas.
n
Calculated response factors are stored by the GC Controller for use in the concentration calculations, and are printed out in the configuration and calibration reports.
Average response factor is calculated as follows:
k
RF
i
i 1=
------------------
RFAVG
=
n
k
JULY 2010 Basic Analysis Computations
1-14 Description
where:
RFAVGnArea or height average response factor for component n.
Model 500 Gas Chromatograph
Rf
i
Area or height response factor for component n from the calibra­tion run.
k Number of calibration runs actually used to calculate the response
factors.
The percent deviation of new RF averages from old RF average is calculated in the following manner:
% deviation
new
------------------------------------
RF
old
old
100×=
RF
RF
where the absolute value of % deviation for alarm has been previously entered by the operator.
Concentration Calculations in Mole % without Normalization
Once response factors have been determined by the GC Controller or entered by the operator, component concentrations are determined for each analysis by using the following equations:
Area
n
-------------- -
CONC
=
n
ARF
n
or
Ht
n
CONC
--------------
=
n
HRF
n
where:
CONC
Area
Basic Analysis Computations JULY 2010
Concentration of component n in mole percent.
n
Area of component n in unknown sample
n
Model 500 Gas Chromatograph
Description 1-15
ARF
Response factor of component n calculated from area of calibration
n
sample. Units are area per mole percent.
Ht
n
HRF
Peak height of component n in unknown sample
Response factor of component n calculated from peak height of cal-
n
ibration sample. Units are height per mole percent.
Note that the average concentration of each component will also be calculated when data averaging is requested.
Component concentrations may be input through analog inputs 1 - 4 or may be fixed. If a fixed value is used, the calibration for that component is the mole % that will be used for all analyses.
Concentration Calculations with Normalization
CONC
CONCN
n
----------------------------
i 1=
k
CONC
n
100×=
i
where:
CONCN
CONC
CONC
k Number of components to be included in the normalization.
Normalized concentration of component n in percent of total gas
n
concentration.
Non-normalized concentration of component n in mole percent.
n
Non-normalized concentration (in mole percent) from each of the k
i
components to be grouped into this normalization.
JULY 2010 Basic Analysis Computations
1-16 Description
For additional information about other calculations that are performed by the GC Controller and software, see the MON2000 Software for Gas Chromatographs user manual (P/N 3-9000-522).
1.6 GLOSSARY
Auto Zero: Automatic zeroing of the preamplifier. May be entered into the Controller to take place at any time during the analysis when either the component is not eluting or the baseline is steady.
Chromatogram: A permanent record of the detector output. A chromatograph is obtained from a PC interfaced with the detector output through the GC Controller. A typical chromatogram displays all component peaks, and gain changes. It may be viewed in color as it is processed on a PC VGA display. Tick marks recorded on the chromatogram by the GC Controller indicate where timed events take place.
Model 500 Gas Chromatograph
Component: Any one of several different gases that may appear in a sample mixture. For example, natural gas usually contains the following components: nitrogen, carbon dioxide, methane, ethane, propane, isobutane, normal butane, isopentane, normal pentane, and hexanes plus.
Condulet: Fitting resembling a pipe or a box with a removable cover for access to electric conduits.
CTS: Clear to send (a serial port pin assignment).
DCD: Data carrier detect; see also, RLSD (a serial port pin assignment).
DSR: Data set ready (a serial port pin assignment).
DTR: Data terminal ready (a serial port pin assignment).
GLOSSARY JULY 2010
Model 500 Gas Chromatograph
Response Factor: Correction factor for each component as determined by the calibration. See “Concentration Analysis by Using Response
Factor” on page 12 for more information.
Retention Time: The time (in seconds) that elapses between start of analysis (0 seconds) and the sensing of the maximum concentration of each component by the Analyzer detector.
RI: Ring indicator (a serial port pin assignment).
RLSD: Received line signal detect (a digital simulation of carrier detect);
see also, DCD (a serial port pin assignment).
RTS: Request to send (a serial port pin assignment).
RxD, RD, or SIN: Receive data, or signal in (a serial port pin
assignment).
Description 1-17
TxD, TD, or S
assignment).
: Transmit data, or signal out (a serial port pin
OUT
JULY 2010 GLOSSARY
1-18 Description
This page is intentionally left blank.
Model 500 Gas Chromatograph
GLOSSARY JULY 2010
EQUIPMENT DESCRIPTION
This section provides descriptions of the various subsystems and components that make up the Model 500 gas chromatograph. This section is organized as follows:
Sampling System
- Sampling Point Location
- Sample Volume and Flow Rate
- Sample Conditioning
- Contamination Precautions
-Valving
- Calibration Gas
•Analyzer
- Physical Description
- Chromatograph Valves
- Detector Subsystem
- Analyzer Preamplifier Unit
- Analyzer Specifications
- Utility Gas Requirements
Controller
- Controller Hardware Configurations
- Optional Keypad and Display
- Alarm Specifications
2.1 SAMPLING SYSTEM
A well designed, properly adjusted sampling system is essential to optimum performance of any gas chromatograph. If a good sample is not obtained for analysis, the whole purpose of the system is compromised.
The purpose of the sample handling system is not to transfer an exact sample of process fluid to the chromatograph. Rather, the purpose is to transfer a representative sample of the fluid--after it has been
2-2 Equipment description
Model 500 Gas Chromatograph
conditioned--that is compatible with chromatograph sample requirements. This statement encompasses a big difference and is very important to remember.
The Sample Conditioning System (SCS) is located between the process stream and the Analyzer, and is usually mounted on the lower portion of the Analyzer stand. It serves these purposes:
Extracts final sample from the fast loop,
Performs final filtration,
Performs stream switching for a multi-stream Analyzer, and
Adjusts the final pressure, temperature, and flow control on the selected sample flowing to the sample valve.
The following points should be considered in selecting and installing a sampling system.
2.1.1 Sampling Point Location
Gas samples must be representative of the process stream and must be taken from a location where stratification or separation of components does not occur. The sampling point should be as close as feasible to the Analyzer.
2.1.2 Sample Volume and Flow Rate
An adequate response time for sample analysis requires that sample volumes should generally be as small as possible, and the flow rate between the sampling point and the Analyzer should be as high as possible, consistent with accuracy. To minimize time lag and to prevent back diffusion, dryers and filters in the sampling line should be as small as possible. When long sampling lines cannot be avoided, flow velocity in the line can be increased by decreasing the downstream pressure.
Typically, pressure is reduced at the sample point with a pressure regulating sample probe. The input pressure to the Analyzer can be adjusted between 10 and 20 pounds per square inch, gauge (psig). Reducing the pressure at the sample point avoids the problem of heavy liquid dropout in the sample line during cold weather. The flow rate in
Sampling Point Location JULY 2010
Model 500 Gas Chromatograph
the sample line is set at 50 cubic centimeters (cc) per minute with the restrictor valve at the Analyzer.
Use this general rule to approximate sample lag time caused by the length of sample line. Sample line constructed of 1/8-inch tubing contains approximately 1 cc of volume per foot. Therefore, with a flow rate of 50 cc per minute, the lag time of the sample between the sample point and the Analyzer is calculated by dividing the length of the line (in feet) by 50. For example, the sample in a 100 foot sample line will take 2 minutes to travel the length of the line.
2.1.3 Sample Conditioning
Sample systems should contain at least one filter to remove solid particles from the sample stream. Most applications require fine-element filters upstream of the Analyzer.
Equipment description 2-3
2.1.4 Contamination Precautions
Several precautions are recommended to minimize the possibility of contaminating samples. Except in special applications, filters should be of either the ceramic or the porous metallic type to avoid the absorption losses characteristic of fiber or paper filters. Pressure regulators and flow controllers containing cork or felt filters or absorbent diaphragms should not be used. Sampling lines for noncorrosive streams should be stainless steel tubing and must be clean and free of grease. Lines must be pressure tight to prevent diffusion of moisture or atmospheric gases into the sample. Pipe threads should be dressed only with Teflon tape on pipe threads and never with pipe thread compounds (dope).
2.1.5 Valving
A block valve should be installed immediately downstream of the sample takeoff point to permit shutdown of the system for maintenance. Block valves should be either needle valves or cocks of the proper material and packing, and should be rated for the process line pressure. Tight seating of all connections is essential.
JULY 2010 Sample Conditioning
2-4 Equipment description
2.1.6 Calibration Gas
A calibration gas used for Process analysis should be blended of gases specified as Primary Standards. Primary Standard gases are blended using weights that are traceable to the National Institute of Standards and Technology (N.I.S.T). The calibration gas should not have any component that could drop out at the coldest temperature to which the gas will be subjected. A typical C6+ blend for a temperature of zero degrees Fahrenheit (0 °F) is listed in the following table. No dropout will occur in this calibration gas if it is blended at a pressure below 250 psig.
Table 2-1 Contents of Example Calibration Gas
Gas Mole Percent
Nitrogen 2.5 Carbon dioxide 0.5
Model 500 Gas Chromatograph
Methane Balance Ethane 5.0 Propane 1.0 Isobutane 0.3 N-butane 0.3 Neopentane 0.1 Isopentane 0.1 N-pentane 0.1 N-hexane 0.03
The sampling system should be carefully planned for the best chromato­graphic analyses.
Calibration Gas JULY 2010
Model 500 Gas Chromatograph
2.2 ANALYZER
2.2.1 Physical Description
The Analyzer is physically divided into two major sections (see Figure 2-
1). The upper heated section is temperature controlled and contains the
following components:
Pneumatically actuated valves that control the flow of the sample and
carrier gases
Detector elements
Analytical columns
A temperature-controlled heater block
The lower section consists of two explosion-proof housings that contain printed-circuit assemblies for the following functions:
Equipment description 2-5
Valve control
Heater-block temperature control
Detector drive
Detector output signal preamplifier.
The Analyzer assembly, sample valves, and associated piping are mounted in a self-supporting rack that may be placed at or near the sample tap. Under most environmental conditions, the Analyzer assembly requires no additional shelter.
JULY 2010 ANALYZER
2-6 Equipment description
Model 500 Gas Chromatograph
Figure 2-1 Analyzer Subsystems
Physical Description JULY 2010
Model 500 Gas Chromatograph
2.2.2 Chromatograph Valves
A chromatograph valve is shown in Figure 2-2 in exploded view. Its pistons are pneumatically actuated in both switching directions by the actuating assemblies located below the primary plate.
Equipment description 2-7
Figure 2-2 Chromatograph Valve
JULY 2010 Chromatograph Valves
2-8 Equipment description
Model 500 Gas Chromatograph
Primary Plate
The primary plate contains precisely machined internal passages that enter and exit the valve at top ports, each of which is connected to the top and/or bottom of the plate within the valve. The primary plate, which is the only metallic element that comes in contact with the sample, is isolated from the remainder of the valve by specially formulated diaphragms.
GC valve should be torqued to 30 ft. lbs.
Actuating Subassemblies
Below the primary plate, pistons are operated by pneumatic pressure applied to actuating diaphragms through ports in the base plate.
Operation
When pneumatic pressure is applied to the actuating diaphragms, the pistons are actuated, thus forcing the sealing diaphragm against the primary plate. This closes the passages that are connected at the bottom of the plate. When pressure is removed, the pistons are free to move, and flow is resumed through the passages.
2.2.3 Detector Subsystem
The operation of the Analyzer detector subsystem was previously discussed in the "Theory of Operation" section of this manual.
Detector Subsystem JULY 2010
Model 500 Gas Chromatograph
2.2.4 Analyzer Preamplifier Unit
The electrical output from the detector is amplified by the Analyzer preamplifier unit. The preamplifier also supplies drive current to the detector bridge. The voltage signal is converted to a 4 to 20-milliamp (mA) current loop for transmission to the GC Controller. The signal is proportional to the concentration of a component detected in the gas sample. The preamplifier provides four different gain channels and compensates for baseline drift. The signals from the preamplifier are then sent to the GC Controller, where they provide the basis for analysis computations and a chromatographic trace, or chromatogram.
2.2.5 Analyzer Specifications
Power Requirements: 120 volts, alternating current (VAC), +10/­15VAC; 50 to 60 Hertz (Hz); single phase; 10 amperes (A) (maximum during warmup) (additional 10 amps if unit has a Sample System Oven).
Equipment description 2-9
Ambient Temperature Range: -18 degrees Celsius ( C) to +55 C (0 F to +130 F)
Humidity: 0 to 95 percent relative humidity, noncondensing
Vibration: Designed for mounting on process piping or other field
structures subject to normal process vibrations
National Electrical Code (NEC) Area Classification: Suitable for NEC Class 1, Division 1, Groups B, C, and D
Rack Size:
Height: 58 inches (147.3 centimeters [cm])
Width: 18 inches (45.7 cm) maximum
Depth: 18 inches (45.7 cm)
Weight: Approximately 125 pounds (56.8 kilograms [kg]), including mounting hardware.
JULY 2010 Analyzer Preamplifier Unit
2-10 Equipment description
Sample Requirements:
Fluid Phase - Vapor
Pressure - 15 to 30 psig, regulated to ±10 percent
Flow Rate - 50 cc/min, typical
Analyzer Output Signal: Four different gain channels to provide a 4 to 20 mA signal to the Controller.
Transient Over Voltages Category: Installation Category (Over Voltage Category II)
Cleaning requirements: Restricted to the 6-port valve (refer to “Valve
Cleaning” on page 5-22).
2.2.6 Utility Gas Requirements
Model 500 Gas Chromatograph
Carrier Gas: Typically zero grade helium (99.995% pure, with less than 5 ppm water, and less than 0.5 ppm hydrocarbons).
Valve Actuation Gas: Typically zero grade, 99.995% pure helium at 115 psig. Consumption is 100cc per analysis cycle. Clean, dry air also may be used for valve actuation. Carrier Gas and Valve Actuation Gas are normally supplied from a common cylinder, since overall gas consumption is minimal.
2.3 CONTROLLER
The Model 500 Controller is a microprocessor-based device that provides the Model 500 Analyzer with highly accurate timing, precision calculations, pertinent report generation, and an interface with other devices. The Controller provides both analog outputs and a direct digital link with output devices through RS-232C, RS-422, or RS-485 ports. Volatile portions of the program are protected by a lithium battery backup if power is lost or turned off at the unit.
Utility Gas Requirements JULY 2010
Model 500 Gas Chromatograph
The Controller can be packaged for side-by-side use with the Model 500 Analyzer in a hazardous area, or for remote use in a safe area in a 19-inch rack mounting. Also, a retrofit kit is available to replace the older model GC Controller (Model 2251) used with the Model 500 Analyzer. The retrofit kit is suitable for 12-inch rack mounting.
The Model 500 Controller can be linked directly to a PC by serial connection or by a telecommunication link that uses Modbus protocol. This provides the preferred method for operating the GC System. Limited control of the GC System is also possible through a built-in keypad and display, which are optional components of the explosion­proof, hazardous environment GC Controller package. The local alphanumeric keypad and display allow for maintenance and minor adjustments in a hazardous environment.
Equipment description 2-11
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Do not operate a PC or printer in a hazardous environment.
Failure to observe all safety precautions could result in serious injury or death.
The 19-inch rack mounted, 12-inch rack retrofit kit, and the explosion­proof NEMA 4X, Groups B, C, and D, Controllers all operate identically.
JULY 2010 CONTROLLER
2-12 Equipment description
2.3.1 Controller Hardware Configurations
The Controller may be provided for hazardous area mounting, 19-inch rack mounting, or used in a 12-inch rack retrofit kit. See Figure 2-3 through Figure 2-5. The unit consists of an STD-bus based computer and related boards, including boards for terminating field wiring. The enclosure for hazardous areas qualifies as flameproof (explosion-proof NEMA 4X, Groups B, C, and D). Connections to the enclosure are through one 2-inch (50mm) hole (reduced to 3/4 inch with bushing) and two 1-inch (25 mm) conduit fittings located in the bottom. These accept matching conduit or cable entries. Field connections are made through explosion-proof conduit or flameproof glands.
Model 500 Gas Chromatograph
Figure 2-3 GC Controller, Explosion-Proof Version
Controller Hardware Configurations JULY 2010
Model 500 Gas Chromatograph
Equipment description 2-13
Figure 2-4 GC Controller, 19-inch Rack Mounted Version
JULY 2010 Controller Hardware Configurations
2-14 Equipment description
Model 500 Gas Chromatograph
Figure 2-5 Model 2251 Retrofit Kit (12-inch rack) for the GC Controller
For operating a printer in a nonhazardous area at the GC site, a DB-25 parallel port is available on the GC's Terminal Board for field wiring.
For connecting a PC to the GC at the GC site (for setup, operation, or maintenance in a nonhazardous area), a DB-9 serial port connector is available on the Controller's front panel.
Controller Hardware Configurations JULY 2010
Model 500 Gas Chromatograph
The STD-Bus Card Cage inside the GC Controller is equipped with two cards. Card slots are preassigned so that cables can be consistently routed. However, the COM4A board and the modem board (and Ethernet card when using a 16-bit CPU board) can be piggy-backed in any order on the CPU board. If a Radicom modem is used, it must be the top board in the card cage assembly.
An optional stream switching assembly (with either AC or DC solenoids) can be controlled by the GC Controller, allowing for switching up to 12 streams.
Analog Inputs and Outputs
The GC Controller can accommodate eight fully differential analog 4 to 20 mA input signals. Four of the analog inputs are used by the associated Analyzer, and they are filtered with transient protection. The additional four input ports provide the ability to accept signals from other Analyzers, so that the analytical report of the chromatograph can include other information on the gas stream, such as water or sulphur content. Transient protection and shield terminations are available for these inputs.
Equipment description 2-15
There is capability for a maximum of ten analog outputs. Two analog outputs are available as standard components of the Controller; the other eight analog outputs are optional. All ten analog outputs are current type: 4-20 mA, not isolated. Also, all ten analog outputs can be calibrated by the MON2000 software.
Digital Inputs and Outputs
The Controller has the capability of sixteen digital inputs used as follows:
5 - to read a Modbus address, as defined by DIP switch positions.
2 - to indicate presence and type of front panel as defined by switch positions
1 - Spare
1 - temperature sensor input to shut off LCD backlight
1 - GC alarm, optically isolated, with transient protection
JULY 2010 Controller Hardware Configurations
2-16 Equipment description
Model 500 Gas Chromatograph
5 - stream flow alarms, optically isolated, with transient protection
1 - photocell detector, front panel backlight (night on, day off)
The Controller has the capability of 22 digital outputs used as follows:
6 - Analyzer control
8 - driver outputs for DC air solenoids (stream switching, 12 total streams)
5 - alarms, optically isolated, with transient protection
3 - front panel indicators (green, yellow, red)
The digital transient-protected discrete outputs can furnish up to 50 mA. If more current is required (up to 0.5A), a special transient protection plug-in module should be installed (see Appendix C, this manual, for transient protection module details).
Communications
There are 3 to 8 communication ports externally available, depending on options package selected. The communications ports can use either RS­232, RS-422, or RS-485 protocol, which, for the LX-800 CPU board, can be set via MON20/20, and for the 6117 CPU board, must be set by Data Interface Chips resident on the board.
The communications ports are configured at the factory, as specified by the customer's communications requirements. The Modbus switch positions are also normally set at the factory as specified by the customer. If any changes need to be made in the field, refer to the drawings in the rear of this manual.
Controller Hardware Configurations JULY 2010
Model 500 Gas Chromatograph
Driver Outputs
The Controller has eight stream switch outputs, 120 mA continuous current, which can be used to control optional AC or DC solenoid switch boards. This increases stream switch capability from the standard capability of four streams and 1 CAL to a maximum capability of twelve streams.
The optional stream switching assembly (AC or DC solenoid systems) has provisions for eight stream routes, but adding this option has the net effect of adding only seven more stream routes to the standard stream capacity of five (thus giving a total stream capacity of twelve). The reason is, one of the Analyzer's standard five stream routes becomes dedicated to the optional stream switch assembly when the optional stream switch assembly is installed.
Equipment description 2-17
General Controller Specifications
Power requirements (without current outputs): 63.25VA typical for basic instrument
Voltage options:
115 VAC ±15 percent, 50 to 60 Hz @ 0.33 A
230 VAC ± 15 percent, 50 to 60 Hz @ 0.275 A
Temperature:
Operating range: -18 C to 55 C (0 F to 131 F)
Storage range: -40 C to 85 C (-40 F to 185 F)
Humidity: 0 to 95 percent relative humidity, noncondensing
JULY 2010 Controller Hardware Configurations
2-18 Equipment description
Model 500 Gas Chromatograph
Explosion-proof NEMA 4X, Groups B, C, and D, enclosure dimensions:
Height:13 inches
Width:14 inches
Depth: 14 inches
Rack mounted dimensions (standard 19-inch):
Height:8.75 inches
Width: 19 inches
Depth: 8.5 inches
Retrofit kit enclosure dimensions (12-inch rack, "panel mount"):
Height:8.5 inches
Width:11 inches
Depth:11 inches
Weight: Approximately 74 pounds for Explosion-Proof NEMA 4X, Groups B, C, and D, version (not including stand)
Electrical/Mechanical Safety and Integrity - Certifications and Classifications
Both the Analyzer and the GC Controller, when housed inside explosion­proof enclosures, meet these certifications and classifications for electrical and/or mechanical safety and integrity:
National Electrical Manufacturers Association (NEMA) 7 for National Electrical Code (NEC) Class I, Division 1, Groups B, C, and D areas. Meets Underwriters Laboratories Inc. (UL) 1203, "Explosion-Proof and Dust-Proof Electrical Equipment of use in hazardous (Classified) locations" for NEC Class I, Division 1, Groups B, C, and D, and Canadian Standards Association (C.S.A.) 22.2 No. O-M1962, Part II and C.S.A. 22.2 No. 30-M1986 for NEC Class I, Division 1, Groups B, C, and D.
Controller Hardware Configurations JULY 2010
Model 500 Gas Chromatograph
EEx d IIB T6 - Meets CENELEC EN 50 014, and EN 50 018, "Electrical Apparatus for Potentially Explosive Atmospheres...", Parts 1 and 5, as flameproof for Group II, Subdivision B, Temperature Class T6.
The GC Controller, when housed inside explosion-proof enclosure, meets these certifications and classifications for electrical and/or mechanical safety and integrity:
NEMA 4X - Meets NEMA 250, "Enclosures for Electrical Equipment (1000 volts maximum)", for type 4X, Canadian Electrical Code, Part II, Rule 2-400 1 d, and C.S.A. C22.2 No. 94-1967 as C.S.A. enclosure 4, and International Electrotechnical Commission (IEC) 144, "Degrees of protection of enclosures of Switchgear...", for IP 65.
Both the Analyzer, when housed inside explosion-proof enclosure, and the GC Controller, when housed inside (a) the explosion-proof enclosure, (b) the rack mount enclosure, or the retrofit enclosure, meet the Federal Communications Commission (FCC) Part 15, Subparts A and B classification for control against excessive radio frequency emissions.
Equipment description 2-19
GC Controller Circuit Board List
The GC Controller circuit boards are inserted or attached to an STD-bus card cage assembly. The Controller has two boards inserted into the card cage, and two of the boards are attached to the card cage outside.
The 6117 CPU board has three optional piggy-back boards that can be attached to it via the PC 104 bus:
Modem
COM4A serial ports 5, 6, 7, and 8
Ethernet
JULY 2010 Controller Hardware Configurations
2-20 Equipment description
Model 500 Gas Chromatograph
There are two optional piggy-back boards that can be attached to the LX­800 CPU board via the PC 104 bus:
•Modem
COM4A serial ports 5, 6, 7, and 8
The LX-800's onboard Ethernet port can be used instead of the optional Ethernet board, unless the LX-800 is installed with application revision 3.99 or earlier (16 bit BOS), in which case, the Ethernet board must be used.
onboard ethernet port
Controller Hardware Configurations JULY 2010
Model 500 Gas Chromatograph
The inserted circuit boards of the GC Controller perform these functions:
Table 2-2 Functions of Inserted Circuit Boards, GC Controller Card Cage Assembly
Equipment description 2-21
Subsystems
32-bit CPU microprocessor board
or
16-bit CPU microprocessor board
COM4A Board (CPU Daughter board)
Modem (CPU Daughter board)
Analog I/O board
[requires MON2000 software, version
2.3 or later]
Handle Label or Part Number
LX800
MCM/LPM-6117
Analog* Control of eight analog inputs
Function(s)
Microprocessor; control of parallel printer port; control of COM1, COM2, COM3, and COM4 communication ports; system memory; RS-422 serial protocol; three timers, digital I/O, and CompactFlash. Additional memory for higher capacity data archives.
Control of COM5 - COM8 BE-20767
telephone modem N/A
(4 for user applications and 4 for Analyzer-Controller inter­connect) and two, six, or ten analog outputs
See drawing number...
DE-20782
BE-18044
Ethernat Card (16-bit CPU daughter board)
PCM-NE 2000 Flexible, high-performance
networking capability; broad spectrum of software support.
N/A
There are two circuit boards attached to the outside of the card cage:
The System Interface and Driver board, and
The GC Controller's Terminal Board for Field Wiring.
The GC Controller's Terminal Board for Field Wiring provides termination connections for these items:
Communication ports (COM1, COM2, COM3, COM4, COM5, COM6,
COM7, and COM8),
JULY 2010 Controller Hardware Configurations
2-22 Equipment description
Model 500 Gas Chromatograph
Analog inputs and outputs,
Digital inputs and outputs,
Controller-Analyzer interconnections,
Parallel printer port, and
Optional stream switching assemblies.
See drawing DE-20782 for an illustration of the GC Controller's Terminal Board for Field Wiring.
The GC Controller's Terminal Board for Field Wiring also has sockets for transient protection modules, and a 250 VAC, 2A fuse (5 x 20 mm) that protects all of the boards from transient surges.
See Appendix C and drawing CE-18115 for a list of transient suppression modules that are installed for various configurations of the GC Controller and its communication, analog output, and stream-switching options.
The System Interface and Driver board provides these functions:
Drivers for switching the eight optional valve solenoids,
Location for 8-position DIP switch to set the Modbus address,
Opto-isolation circuits for the discrete inputs and outputs,
Switching power supply and temperature shutdown circuit for the LCD display,
RS-232 to RS-422 conversion for the LCD display, and
Voltage-to-current conversion for the analog outputs.
Controller Hardware Configurations JULY 2010
Model 500 Gas Chromatograph
Jumper for selecting driving voltage source for the 4-20 mA circuit.
See drawing CE-18118 for an illustration of the System Interface and Driver board.
Also see Figure 2-6, which illustrates, through a block diagram, the function and placement of the GC Controller circuit boards.
Equipment description 2-23
JULY 2010 Controller Hardware Configurations
2-24 Equipment description
Model 500 Gas Chromatograph
Figure 2-6 Block Diagram of the GC Controller Circuit Boards
Controller Hardware Configurations JULY 2010
Model 500 Gas Chromatograph
2.3.2 Optional Keypad and Display
A keypad and liquid crystal display (LCD), optionally built into the enclosure's front panel, are available for the explosion-proof and rack mount versions of the GC Controller. (See Figure 2-5 for an illustration of the explosion-proof Controller with optional keypad and LCD.) The built­in keypad and LCD are especially useful for the explosion-proof version. They permit onsite display, control, and data entry at a GC Controller that is situated in a hazardous environment. Note, however, that the control capabilities offered through the built-in keypad and LCD are more limited than those available through a PC connected to the GC Controller.
For details on using the Controller's built-in keypad and LCD, see
“Operation from Local Keypad and Display” on page 4-1.
Equipment description 2-25
Keypad
The front panel keypad is an 18-Key data/function entry device arranged so that the ALT key causes the lower key markings to be displayed/ entered. The designations marked on the top of the keys will be displayed/entered when the ALT key is not pressed.
Display
The Controller display, measuring 5.5 x 2 inches, is capable of 8 lines by 41 characters. It is certified for use with a explosion-proof NEMA 4X, Groups B, C, and D, enclosure. The display is capable of producing the complete alphabet and numbers from the keypad. The video display on the Controller may show a truncated (or curtailed) version of the displays available at the PC. Although the keypad/display located at the Controller site can do many of the functions that the PC software can do remotely, any extensive operations are more conveniently performed through MON2000 on the larger screen and keypad of a PC. A few adjustments will be more conveniently accomplished at the Controller site.
JULY 2010 Optional Keypad and Display
2-26 Equipment description
2.3.3 Alarm Specifications
The GC Controller has the capacity for 36 alarms. There are also operator defined alarms. Some of the alarms are active only if the configuration of the Controller requires the function that is associated with those alarms. Active alarms are shown in the ALARMS menu.
Status Indicators
Three colored LED status indicators are located at the side of the display on the front panel. The indicators are arranged from yellow, green, and red. When illuminated, the LED status indicators signify the following:
Yellow LED: When illuminated, the Yellow LED indicates that an out-of-tolerance value or an alarm condition was entered into the Controller memory for printout with the analysis. The Controller memory retains the alarm for printout until the operator clears the alarm(s). This indicator is partially application controlled and may be set at different out-of-tolerance levels with different applications.
Model 500 Gas Chromatograph
Green LED: When illuminated, the Green LED indicates that the Controller is operating. If the green LED is illuminated and the Model 500 does not accept changes, a password lockout may have been entered. If a password has been entered, the password must be re­entered before the Model 500 program can be changed.
Red LED: When illuminated, the Red LED indicates an out-of­tolerance value or an alarm condition in the RUN mode that requires operator action. Alarm contacts are closed under these conditions. The Red LED is automatically turned off and contacts opened by the Controller at the start of the next analysis.
Alarm Specifications JULY 2010
Model 500 Gas Chromatograph
This page is intentionally left blank.
Equipment description 2-27
JULY 2010 Alarm Specifications
2-28 Equipment description
Model 500 Gas Chromatograph
Alarm Specifications JULY 2010
INSTALLATION AND SETUP
Because the Model 500 Gas Chromatograph system is available in different configurations, not all of the instructions in this section may apply. In most cases, however, to install and set up a Model 500 Gas Chromatograph system, it is recommended that you follow the instructions in nearly the same order as presented in this manual. (Also see Table 3-1 for a summary of installation and setup steps.)
This section provides instructions for installing and setting up the Model 500 Gas Chromatograph system. This section is organized as follows:
Precautions and Warnings
- Hazardous Environments
- Power Source Wiring
- Signal Wiring
- Electrical and Signal Ground
- Electrical Conduit
- Sample Systems Requirements
Preparation
-Introduction
- Site Selection
- Unpacking the Unit
- Necessary Tools and Components
- Optional Tools and Components
Installing the Analyzer
- Point-to-Point Wiring Guide, Analyzer-Controller
- Analyzer AC-Power Wiring
- Sample and Gas Lines
3-2 Installation and Setup
Model 500 Gas Chromatograph
Installing the GC Controller
- Modbus Slave Address (COM ID) Setup
- Controller-Analyzer Wiring
- Controller-PC Wiring (Serial Connections)
- CPU and COM4A Serial Communications Setups
- Controller-Printer Wiring
- Discrete (Digital) I/O Wiring
- Analog I/O Wiring
- Controller AC-Power Wiring
•Analyzer Leak Checks and Purging for First Calibration
-Analyzer Leak Checks
- Purging Carrier Gas Lines
- Purging Calibration Gas Lines
System Start-up
Summary of Installation and Setup Steps
1. Observe Precautions and Warnings (See “Precautions and Warnings”
on page 3-3)
2. Plan Site Location (See “Preparation” on page 3-12)
3. Obtain Supplies and Tools (See “Necessary Tools and Components” on
page 3-15)
4. Install Analyzer Wiring (See “Point-to-point Wiring Guide, Analyzer-
Controller” on page 3-18)
5. Install Analyzer Sample & Gas Lines (See “Sample and Gas Lines” on
page 3-27)
6. Install GC Controller Wiring (See “Installing the GC Controller” on
page 3-30)
7. Perform Leak Checks (See “Analyzer Leak Checks” on page 3-78)
8. Purge Carrier Gas Lines (See “Purging Carrier Gas Lines” on page 3-
80)
9. Purge Calibration Lines (See “Purging Calibration Gas Lines” on
page 3-83)
JULY 2010
Model 500 Gas Chromatograph
10.Start Up GC System (See “System Start-Up” on page 3-85)
3.1 PRECAUTIONS AND WARNINGS
The Analyzer and GC Controller, when housed inside explosion­proof enclosures, meet the certifications and classifications identified in “Electrical/Mechanical Safety and Integrity -
Certifications and Classifications” on page 2-18. Emerson does
not, however, accept any responsibility for installations of these, or any attached equipment, in which the installation or operation thereof has been performed in a manner that is negligent and/or non-compliant with applicable safety requirements.
Installation and Setup 3-3
EQUIPMENT DAMAGE OR PERSONAL INJURY
The responsible body shall operate the equipment as designed and specified by the manufacturer.
Failure to do so may cause personal injury or damage to the equipment.
JULY 2010 Precautions and Warnings
3-4 Installation and Setup
3.1.1 Hazardous Environments
Observe Precautions and
1
Warnings
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Follow these precautions if installing or operating the Model 500 Analyzer and the GC Controller instrumentation in a hazardous
area:
1. Install and operate only the explosion-proof version of the GC Controller in a hazardous area.
Model 500 Gas Chromatograph
2. Do not operate in a hazardous area any printer or personal computer (PC) that is connected to the GC Controller. To interface with a GC Controller in a hazardous area, use the Controller's keypad and liquid crystal display (LCD) that are built into the explosion-proof housing as options. Or, alternatively, use a PC that is located in a nonhazardous area and remotely connected to the GC Controller.
3. Ensure that field connections to the Analyzer and the GC Controller are made through explosion-proof conduit or flameproof glands.
EQUIPMENT DAMAGE OR PERSONAL INJURY
Observe ALL applicable regulations when installing explosion-proof GC units.
Failure to observe all regulations when installing explosion-proof GC units may result in noncompliance, equipment damage or personal injury.
Hazardous Environments JULY 2010
Model 500 Gas Chromatograph
The explosion-proof GC Controller and Analyzer housings are certified for use in locations where fire and explosion hazards may exist, specifically, areas that are classified by the National Electronics Code (NEC) as Class I, Division 1, Groups B, C, and D. However, other regulations apply. For example, all interconnecting runs of cable through conduit must be sealed at least 18 inches beyond the conduit's point of entry into certified explosion-proof housing. Consult your company's policies and procedures and other applicable requirements documents to determine appropriate wiring and installation practices.
3.1.2 Power Source Wiring
1 2
3
Installation and Setup 3-5
Observe Precautions and Warnings
Plan Site Location
Obtain Supplies and Tools
Install Analyzer Wiring
4
Follow these precautions when installing AC power source wiring to the Model 500 Analyzer and the GC Controller instru­mentation:
1. All wiring must conform to the National Electric Code, local state or
other jurisdiction, and company standards and practices.
2. Provide single-phase, three-wire, AC power at 115 or 230 volts AC, 50-
60 Hz.
3. A switch or circuit breaker shall be included in the building
installation in a safe area.
(a) The switch or circuit breaker is marked as the power disconnect
device.
(b) For Rack mount units, the power disconnect switch shall be in
close proximity to the equipment and easily accessible to the operator.
4. Provide 20 ampere circuit breaker protection so that the major
components of the Model 500 Analyzer system–the Analyzer, the GC
JULY 2010 Power Source Wiring
3-6 Installation and Setup
Controller, and any optionally installed sample oven or stream switching devices--are all protected by one circuit breaker.
5. Use multi-stranded copper conductor wire according to the following recommendations:
(a) For power feed distances up to 250 feet (76 meters), use wire size
American Wire Gauge (AWG) 14 (18 Metric Wire Gauge, stranded).
(b) For power feed distances 250 feet to 500 feet (76 meters to 152
meters), use wire size AWG 12 (25 Metric Wire Gauge, stranded).
(c) For power feed distances 500 feet to 1000 feet (152 meters to 305
meters), use wire size AWG 10 (30 Metric Wire Gauge, stranded).
3.1.3 Signal Wiring
Observe Precautions and Warnings
1
Plan Site Location
2
Model 500 Gas Chromatograph
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Follow these general precautions for field wiring digital and analog input/output (I/O) lines:
1. Metal conduit must be used for all process signal wiring.
2. Metal conduit used for process signal wiring must be grounded at conduit support points (grounding the conduit at multiple points helps prevent induction of magnetic loops between the conduit and cable shielding).
3. Use suitable lubrication for wire pulls in conduit to prevent wire stress.
4. All process signal wiring should be a single, continuous length between field devices and the GC Controller. If, however, length or conduit runs require that multiple wiring pulls be made, the individual conductors must be interconnected with suitable terminal blocks.
Signal Wiring JULY 2010
Model 500 Gas Chromatograph
5. Use separate conduits for AC voltage and DC voltage circuits (see
Figure 3-1).
6. Do not place digital or analog I/O lines in same conduit as A-C power
circuits (see Figure 3-1).
7. Use only shielded cable for digital I/O line connections.
(a) Ground the shield at only one end. (b) Shield-drain wires must not be more than two AWG sizes smaller
than the conductors for the cable.
Installation and Setup 3-7
Figure 3-1. Seperate Conduit Entries
JULY 2010 Signal Wiring
3-8 Installation and Setup
Model 500 Gas Chromatograph
8. When inductive loads (relay coils) are driven by digital output lines, the inductive transients must be diode clamped directly at the coil.
9. Any auxiliary equipment wired to the GC Controller must have its signal common isolated from earth/chassis ground.
Applicable to the digital and analog I/O lines connecting to the GC Controller, including the Analyzer-Controller Interconnect lines: Any loop of extra cable left for service purposes inside the GC Controller explosion-proof housing must not be placed near the conduit entry for AC power.
If the above precaution is not followed, the data and control signals to and from the GC Controller can be adversely affected.
3.1.4 Electrical and Signal Ground
Observe Precautions and
1
Warnings
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Follow these general precautions for grounding electrical and signal lines:
1. For shielded signal conducting cables, shield-drain wires must not be more than two AWG sizes smaller than the conductors for the cable. Shielding is grounded at only one end.
2. Metal conduit used for process signal wiring must be grounded at conduit support points (Grounding the conduit at multiple points helps prevent induction of magnetic loops between the conduit and cable shielding).
3. A clamp type ground lug (color green) is located on the inside bottom front of the GC Controller's case. Chassis ground conductors (color
Electrical and Signal Ground JULY 2010
Model 500 Gas Chromatograph
code green) inside the Controller's enclosure should be stranded, insulated copper wire. These device chassis ground conductors should all be connected to the clamp type ground lug.
4. A clamp type ground lug is located on the outside of the GC
Controller's case at the rear of the lower right (facing the operator panel) casting rib. This ground point should be connected to a copper ground rod as described next.
5. A single-point ground (the outside case ground lug) must be connected
to a copper-clad, 10-foot long, 0.75" diameter steel rod, which is buried, full-length, vertically into the soil as close to the equipment as is practical. (Grounding rod not provided by Rosemount Customer
Care.)
6. Resistance between the copper-clad steel ground rod and the earth
ground must not exceed 25 Ohms.
7. The equipment-grounding conductors used between the GC Controller
and the copper-clad steel ground rod must be sized according to the following specifications:
Installation and Setup 3-9
-length, 15 feet or less-
(4.6 meters) AWG 8, stranded, insulated copper wire
-length, 15 to 30 feet-
(4.6 to 9.1 meters) AWG 6, stranded, insulated copper wire
-length, 30 to 100 feet-
(9.1 to 30.5 meters) AWG 4, stranded, insulated copper wire
8. All inter-enclosure equipment-grounding conductors must be
protected by metal conduit.
9. External equipment, such as data printers, that are connected to the
GC Controller should be powered via isolation transformers to minimize the ground loops caused by the internally shared safety and chassis grounds.
JULY 2010 Electrical and Signal Ground
3-10 Installation and Setup
3.1.5 Electrical Conduit
Observe Precautions and
1
Warnings
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Model 500 Gas Chromatograph
Follow these general precautions for conduit installation
:
1. Conduit cutoffs must be square. Cutoffs must be made by a cold cutting tool, hacksaw, or by some other approved means that does not deform the conduit ends or leave sharp edges.
2. All conduit fitting threads, including factory-cut threads, must be coated with a metal-bearing conducting grease, such as Crouse-Hinds STL or equivalent, prior to assembly.
3. Temporarily cap the ends of all conduit run runs immediately after installation to prevent accumulation of water, dirt, or other contaminants. If necessary, swab out conduits prior to installing the conductors.
4. Install drain fittings at the lowest point in the conduit run; install seals at the point of entry to the GC Controller's explosion-proof housing to prevent vapor passage and accumulation of moisture.
5. Use liquid-tight conduit fittings, such as Myers® Scru-tite® or similar, for conduit which is exposed to moisture.
When conduit is installed in hazardous areas (e.g., areas classified as NEC Class I, Division 1, Groups B, C, and D), follow these general precautions for conduit installation:
1. All conduit runs must have an explosion-proof sealing (potting) fitting located within 18 inches (45.5 centimeters) distance from the conduit entrance to explosion-proof housings.
2. The conduit installation must be vapor tight, with threaded hub fittings, sealed conduit joints and gaskets on covers, or other approved vapor-tight conduit fittings.
Electrical Conduit JULY 2010
Model 500 Gas Chromatograph
EQUIPMENT DAMAGE OR PERSONAL INJURY
Consult your company's policies and procedures and other applicable requirements documents to determine wiring and installation practices that are appropriate for hazardous areas.
Failure to do so may cause personal injury or damage to equipment.
3.1.6 Sample Systems Requirements
Observe Precautions and
1
Warnings
Installation and Setup 3-11
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Observe the following guidelines for installing GC sample systems:
Sample Line Length: If possible, avoid long sample lines. In case of a
long sample line, flow velocity can be increased by decreasing downstream pressure and using by-pass flow via a speed loop.
Sample Line Tubing Material:
Use stainless steel tubing for noncorrosive streams.
Ensure tubing is clean and free of grease.
Dryers and Filters in Sample Line:
Use small sizes to minimize time lag and prevent back diffusion.
JULY 2010 Sample Systems Requirements
3-12 Installation and Setup
Model 500 Gas Chromatograph
Install a minimum of one filter to remove solid particles. Most applications require fine-element filters upstream of the Analyzer.
Do use ceramic or porous metallic type filters. Do not use cork or felt filters.
Pressure Regulators and Flow Controllers in Sample Line: Do not use types containing cork or felt filters, or absorbent diaphragms.
Pipe Threads, Dressing: Do use Teflon tape. Do not use pipe thread compounds (dope).
Valving:
Install a block valve downstream of sample takeoff point for maintenance and shutdown.
Block valve should be needle valve or cock valve type, of proper material and packing, and rated for process line pressure.
3.2 PREPARATION
3.2.1 Introduction
Your Model 500 Analyzer was started and checked out before it left the factory. Program parameters were installed in the system and documented in the "PC Config Report" furnished with your Model 500 Analyzer.
Preparation JULY 2010
Model 500 Gas Chromatograph
3.2.2 Site Selection
Observe Precautions and
1
Warnings
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Installation and Setup 3-13
Follow these guidelines for site selection
:
1. Provide adequate access space for performing maintenance and
adjustments.
(a) Allow a minimum of 16 inches (41 cm) in front for enclosure
opening and access.
(b) Allow a minimum of 15 inches (38 cm) at the rear and left side for
case removal.
(c) If possible, mount the Analyzer components in a vertical stack
configuration; it provides the greatest operator convenience.
2. Install the Analyzer as close as possible to the sample stream.
3. Install the GC Controller no further than 2000 feet (610 meters) away
from the Analyzer.
(a) In a hazardous environment, you can install the explosion-proof
version of the GC Controller near the Analyzer, on either a 19-inch or 12-inch rack.
(b) Observe the same recommendations for minimum clearance: 16
inches (41 cm) in front enclosure opening and access, and 15 inches (38 cm) at the rear and left side for case removal.
(c) Refer to drawing CE-23878 in the Analyzer drawings addendum of
this manual.
4. Ensure that exposure to radio frequency (RF) interference is minimal.
JULY 2010 Site Selection
3-14 Installation and Setup
3.2.3 Unpacking the Unit
Observe the following checklist for unpacking the unit and inspecting for damage:
1. Unpack the equipment:
(a) Model 500 series Analyzer (b) GC Controller
Observe Precautions and
1
Warnings
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Model 500 Gas Chromatograph
2. Ensure that all documentation and software are included:
(a) This manual, the Model 500 Gas Chromatographs Hardware
Reference Manual, P/N 3-9000-537.
(b) The software manual, MON2000 Software for Gas
Chromatographs User Manual, P/N 3-9000-522.
(c) CD(s) with the MON2000 software program and GC Applications.
Installation and startup of the Model 500 Analyzer should proceed only if all required materials are on hand and free from obvious defects. If any parts or assemblies appear to have been damaged in shipment, first file a claim with the carrier. Next, complete a full report of the nature and extent of the damage and forward the report immediately to Rosemount Customer Care for further instructions. Include complete model number information. Disposition instructions will be returned immediately by Rosemount Customer Care. Refer to the Customer Repair Report in the back of this manual.
Unpacking the Unit JULY 2010
Model 500 Gas Chromatograph
3.2.4 Necessary Tools and Components
Observe Precautions and
1
Warnings
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Observe the following checklist of tools and components that you will need for installing the Analyzer and GC Controller:
1. Chromatographic grade carrier gas: zero grade helium or nitrogen
(99.995% pure, with less than 5 ppm water, and less than 0.5 ppm hydrocarbons).
Installation and Setup 3-15
2. High pressure dual-stage regulator for the carrier gas cylinder, high
side up to 3000 pounds per square inch, gauge (psig), low side capable of controlling pressure up to 150 psig.
3. Calibration standard gas with correct number of components and
concentrations (see Section 2.1.6, “Calibration Gas” on page 4).
4. Dual-stage regulator for the calibration gas cylinder, low pressure side
capable of controlling pressure up to 30 psig.
5. Sample probe (fixture for procuring the stream, or sample gas for
chromatographic analysis).
6. 1/8-inch stainless steel (SS) tubing for connecting calibration standard
to analyzer, 1/4-inch SS tubing for connecting helium carrier to the analyzer, 1/8-inch SS tubing for connecting stream gas to the analyzer.
7. Miscellaneous Swagelok tube fittings, tubing benders and tubing
cutter.
8. 14 American Wire Gauge (AWG) (18 Metric Wire Gauge) or larger
electrical wiring and conduit to provide 115 or 230 volts AC, single phase, 50 to 60 Hertz (Hz), from an appropriate circuit breaker and power disconnect switch. (See previous guidelines in Section 3.1.2,
“Power Source Wiring” on page 5.)
JULY 2010 Necessary Tools and Components
3-16 Installation and Setup
9. Liquid leak detector (SNOOP® or equivalent).
10.Digital volt-ohm meter with probe-type leads.
11.A flow measuring device such as Alltech Digital Flow Check™ Flowmeter.
3.2.5 Optional Tools and Components
Observe Precautions and
1
Warnings
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Observe the following checklist of tools and components you may need for installing and using the GC System:
Model 500 Gas Chromatograph
1. For operation in a nonhazardous area only: Printer and printer paper.
2. For operation in a nonhazardous area only: An IBM-compatible PC and a "straight-through" serial cable connection between the external DB-9 (female) serial ports of the GC Controller and a serial port of the PC. (For details, see Section 3.4.3, “Controller PC Wiring (Serial
Connections)” on page 37.).
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Do not operate a PC or printer in a hazardous environment.
Failure to observe all safety precautions could result in serious injury or death.
Optional Tools and Components JULY 2010
Model 500 Gas Chromatograph
If you are working in a hazardous area and need to perform routine operations, use the optional keypad and LCD that are built into the explosion-proof GC Controller. To make more significant changes, however, use a remotely connected PC installed with MON2000. See the MON2000 Software for Gas Chromatographs User Manual, P/N 3­9000-522, for more information.
Serial cable specifications:
Straight-through serial cable with the following terminations:
DB-9, maleconnects to GC Controller external serial port
DB-9 or DB-25, femaleconnects to PC serial port
3. Interconnect Cable, P/N 6-4618-122, if not already installed between
the Analyzer and GC Controller. This is a shielded, computer and communications grade, 15-conductor cable for making the thirteen interconnections between the Analyzer and the GC Controller. This cable, if not enclosed in conduit between the Analyzer and the GC Controller, is suitable only for use in non-hazardous environments. The cable must be run inside conduit for hazardous environments. (For cable termination details, see Section 3.3.1, this manual.)
Installation and Setup 3-17
4. Direct Serial Connect Cable, P/N 3-2350-068, to connect the PC or an
external modem directly to the one of the GC Controller's serial ports on the GC Controller's Terminal Board for Field Wiring (TB). This cable comes in a customer-specified length. It is terminated with a DB-9 female plug at one end, for connection to a PC or external modem's serial port, and six exposed leads at the other end for connection to the one of the GC Controller's serial ports on the TB. (See Section 3.4.3.3 for directions on how to install this cable.)
5. Items necessary for connecting the GC Controller to an external
modem, a multi-drop serial network, or other type of remote data transfer system (an example item might be an RS-232/RS-485 conversion box for long distance serial transmission).
JULY 2010 Optional Tools and Components
3-18 Installation and Setup
3.3 INSTALLING THE ANALYZER
3.3.1 Point-to-point Wiring Guide, Analyzer-Controller
Observe Precautions and
1
Warnings
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
Model 500 Gas Chromatograph
This section applies only to GC systems which have not been shipped "prewired." In most cases, the explosion-proof system will already have had the Analyzer-Controller connections made. If your system has already been wired, skip this section, and proceed to the next section.
To make wiring connections between the Analyzer and the GC Controller, follow these steps:
1. Disconnect all electrical power to both the Analyzer and the GC Controller.
INSTALLING THE ANALYZER JULY 2010
Model 500 Gas Chromatograph
2. At the Analyzer site, locate the lower explosion-proof box (its cover is
marked with original equipment manufacturer's catalog number "XJT"). Remove its threaded Condulet cover.
Installation and Setup 3-19
Figure 3-2. Lower Explosion-Proof (XJT) Box at Analyzer
3. You will need to feed the Interconnect Cable through the inlet on the
upper left side of the XJT box, and make connections to the interconnect Terminal Board (TB-4) which lies behind Valve Driver board (see Figure 3-3).
(a) The Interconnect Cable is computer and control applications grade,
15-conductor, shielded cable. Individual conductors are stranded tinned copper, #22 AWG-(7x30). (Also see description in Section
3.2.5, “Optional Tools and Components” on page 16.)
JULY 2010 Point-to-point Wiring Guide, Analyzer-Controller
3-20 Installation and Setup
(b) Maximum length of Interconnect Cable (or, the maximum distance
between the Analyzer and the GC Controller) should not exceed 2000 feet (610 meters).
Model 500 Gas Chromatograph
Figure 3-3. Behind the Condulet Cover is the Valve Driver Board, then TB-4
4. With the Condulet cover removed, loosen and remove the four (4) thumbscrews that hold the Valve Driver board.
5. Carefully edge the Valve Driver board off the holding screws. Do not disconnect the Valve Driver board from the cable; merely let the board rest face down, secured by the cable (see Figure 3-4).
6. With the Analyzer TB-4 now exposed, connect thirteen of the Interconnect Cable's fifteen leads to terminals 11 through 23. See
Table 3-1 and Figure 3-5 for purposes and destinations of leads. Also
see "CAUTION", step (6).
Point-to-point Wiring Guide, Analyzer-Controller JULY 2010
Model 500 Gas Chromatograph
Installation and Setup 3-21
Figure 3-4. Valve Driver Board Resting Face Down from its Cable Allows Access to TB-4
JULY 2010 Point-to-point Wiring Guide, Analyzer-Controller
3-22 Installation and Setup
Table 3-1. Analyzer and GC Controller Interconnect Leads
Board Acronyms:
• Interconnect Terminal Board of Analyzer (TB-4)
• Terminal Board for Field Wiring at Controller (TB)
Model 500 Gas Chromatograph
Analyzer (TB-4)
Terminal 11 Function code 1 J19, Terminal 1
Terminal 12 Function code 2 J19, Terminal 2
Terminal 13 Function code 4 J19, Terminal 3
Terminal 14 Function code 8 J19, Terminal 4
Terminal 15 Function code strobe J20, Terminal 1
Terminal 16 Common - function codes J19, Terminal 5
Terminal 17 Auto Zero (AZ) J20, Terminal 2
Terminal 18 Preamp gain channel 1 J18, Terminal 1
Terminal 19 Preamp gain channel 2 J18, Terminal 4
Terminal 20 Preamp gain channel 3 J18, Terminal 7
Terminal 21 Preamp gain channel 4 J18, Terminal 10
Terminal 22 Common - preamp gain J18, Terminal 11
Terminal 23 Alarm function (AF) J20, Terminal 3
color color
Controller (TB)
Connect the interconnect cable SHIELD to one terminal; specifically, terminal 12 of J18, on the GC Controller TB.
Point-to-point Wiring Guide, Analyzer-Controller JULY 2010
Model 500 Gas Chromatograph
EQUIPMENT DAMAGE OR PERSONAL INJURY
Do not apply AC electrical power to the Analyzer or the GC Controller until all electrical power, interconnection, and external signal connections have been verified, and proper grounds have been made. Refer to Section 3.1.3 for general precautions concerning signal wiring.
Failure to properly connect the GC unit may result in serious equipment damage or personal injury.
7. Access the GC Controller's Terminal Board for Field Wiring (TB), and
connect the other leads of the Interconnect Cable to the GC Controller's TB (see instructions in Section 3.4.2, “Controller-Analyzer
Wiring” on page 35). Ensure that the connections correspond to those
listed in Table 3-1 and Figure 3-5.
Installation and Setup 3-23
At this time, you may also want to complete the remainder of all electrical connections at the GC Controller. If so, see Section 3.4,
“Installing the GC Controller” on page 30.
JULY 2010 Point-to-point Wiring Guide, Analyzer-Controller
3-24 Installation and Setup
Model 500 Gas Chromatograph
Figure 3-5. Analyzer and GC Controller Interconnect Leads
Point-to-point Wiring Guide, Analyzer-Controller JULY 2010
Model 500 Gas Chromatograph
8. After confirming that all Interconnect Cable terminations are correct
between the Analyzer and the GC Controller, lift the Valve Driver board from its resting position and place it over the four holding screws.
Reinstall the four thumb screws to secure the Valve Driver board in place.
9. If necessary, complete wiring connections between the Analyzer's
Valve Driver board and any optional stream switch boards.
10.If necessary, complete wiring for connecting AC power to the
Analyzer, with proper connections to hot, neutral, and ground; but do not turn on AC power to the Analyzer yet (see CAUTION below; see details for AC power connection to Analyzer in Section 3.3.2,
“Analyzer AC Power Wiring” on page 26).
Installation and Setup 3-25
EQUIPMENT DAMAGE OR PERSONAL INJURY
Do not apply AC electrical power to the Analyzer or the GC Controller until all electrical power, interconnection, and external signal connections have been verified, and proper grounds have been made.
Failure to properly connect the GC unit may result in serious equipment damage or personal injury.
11.Leave the Analyzer's lower XJT box open if you need to connect
sample and gas lines. (You will need to manually operate the sample valve switches on the Valve Driver board.) Otherwise, reinstall the Condulet cover of the Analyzer's lower XJT box.
12.If necessary, proceed to Section 3.3.3, “Sample and Gas Lines” on
page 27 for instructions on connecting sample and gas lines to the
Analyzer.
JULY 2010 Point-to-point Wiring Guide, Analyzer-Controller
3-26 Installation and Setup
3.3.2 Analyzer AC Power Wiring
Observe Precautions and Warnings
1
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
To connect 115 volts AC Power to the Analyzer, follow these steps:
1. Locate the three leads for connecting 115 volts AC-power to the Analyzer.
(a) Leads are "pig-tailed" from the Analyzer power supply through
conduit to a customer power NPT connection directly behind the lower XJT box of the Analyzer (see drawing CE-10492, Analyzer drawings addendum of this manual).
Model 500 Gas Chromatograph
(b) Leads are colored as follows:
- BLACK: hot
- WHITE: neutral
- GREEN: ground
2. Connect Analyzer AC-power leads to a properly controlled 115 volts AC-power source (i.e., with circuit breaker and power disconnect switch).
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Do not connect AC power leads without first ensuring that AC power source is switched OFF.
Failure to observe all safety precautions could result in serious injury or death.
Analyzer AC Power Wiring JULY 2010
Model 500 Gas Chromatograph
Make power line splices and conduit seals that comply with applicable wiring requirements (for hazardous environments).
EQUIPMENT DAMAGE OR PERSONAL INJURY
Do not apply AC electrical power to the Analyzer or the GC Controller until all electrical power, interconnection, and external signal connections have been verified, and proper grounds have been made.
Failure to properly connect the GC unit may result in serious equipment damage or personal injury.
3. If necessary, connect the Analyzer's chassis ground to an external
copper ground rod (at remote locations). See Section 3.1.4, “Electrical
and Signal Ground” on page 8, regarding electrical and signal ground.
Installation and Setup 3-27
3.3.3 Sample and Gas Lines
4 Install Analyzer Wiring
5 Install Analyzer Sample & Gas Lines
6 Install GC Controller Wiring
7 Perform Leak Checks
To install GC sample and gas lines, follow these steps
Use stainless steel tubing. Keep tubing clean and dry internally to avoid contamination. Before connecting the sample and gas lines, flow clean air or gas through them. Blow out internal moisture, dust, or other contaminants.
:
JULY 2010 Sample and Gas Lines
3-28 Installation and Setup
Model 500 Gas Chromatograph
1. Remove the plug from the Analyzer Sample Vent (SV) line (1/16-inch tubing marked "SV", located at left side of Analyzer.
At this stage in the installation, the Analyzer Measure Vent (MV) line (marked "MV") is left plugged until Analyzer leak checks are completed. For regular Analyzer operation, however, the MV line must be unplugged, or open.
HINT: Do not discard the vent line plugs. They are useful at any time when leak-checking the Analyzer and its sample or gas line connections.
(a) If desired, connect "SV" vent line to an external (ambient pressure)
vent. If the vent line is terminated in an area exposed to wind, protect the exposed vent with a metal shield.
(b) Use 1/4-inch or 3/8-inch tubing for vent lines longer than 10 feet.
2. Connect carrier gas to Analyzer. (DO NOT TURN ON GAS AT THIS TIME.)
See Appendix B, this manual, for a description of a dual-cylinder carrier gas manifold (P/N 3-5000-050) with these features:
• Carrier gas is fed from two bottles.
• When one bottle is nearly empty (100 psig), the other bottle becomes the primary supply.
• Each bottle can be disconnected for refilling without interrupting GC operation.
(a) Use 1/4-inch stainless steel tubing to conduct carrier gas. (b) Use dual-stage regulator: high side capacity 3000 psig; low side
capacity 150 psig.
(c) Analyzer carrier gas inlet is a 1/4-inch fitting located behind the
lower XJT Condulet box.
Sample and Gas Lines JULY 2010
Model 500 Gas Chromatograph
3. Connect calibration standard gas to Analyzer. (DO NOT TURN ON GAS AT THIS TIME.)
(a) Use 1/8-inch stainless steel tubing to conduct calibration standard
gas. (b) Use dual-stage regulator: low side capacity up to 30 psig. (c) Calibration gas inlet is identified in the applicable Sample
Conditioning System ("S.C.S.") drawing in the Analyzer drawings
addendum to this manual. (See drawings CE-16120, CE-16220,
CE-16320, CE-16420, CE-16180, CE-16278, CE-24324, CE-24416,
or CE-24513.)
When installing the calibration standard gas line, take care to follow the proper "S.C.S." drawing in order to make the correct tubing connection to the auto-cal solenoid. Drawings are included in this manual which address stream and column gas connections. Choose the drawing that applies to your installation.
Installation and Setup 3-29
4. Connect sample gas stream(s) to Analyzer. (DO NOT TURN ON GAS AT THIS TIME.)
(a) Use 1/8-inch or 1/4-inch stainless steel tubing to conduct
calibration standard gas.
(b) Ensure that pressure of sample line is regulated to maintain 15-30
psig ±10%.
(c) Sample gas stream inlet(s) are identified in the applicable Sample
Conditioning System ("S.C.S.") drawing in the Analyzer drawings addendum to this manual. (See drawings CE-16120, CE-16220, CE-16320, CE-16420, or CE-16520.)
5. After all lines have been installed, proceed with Controller wiring connections (see next section).
Leak check procedures for the GC sample and gas lines are given in
Section 3.5.1, “Analyzer Leak Checks” on page 78. They require AC
power to be turned on at the Analyzer.
JULY 2010 Sample and Gas Lines
3-30 Installation and Setup
3.4 INSTALLING THE GC CONTROLLER
3.4.1 Modbus Slave Address (COM ID) Setup
4 Install Analyzer Wiring
5 Install Analyzer Sample & Gas Lines
6 Install GC Controller Wiring
7 Perform Leak Checks
The GC Controller's COM ID is determined by dual inline package (DIP) switch settings. In most cases, the COM ID setup made at the factory will not require changes. (Unless otherwise specified by the customer, the DIP switch settings made at the factory give the Controller a COM ID of 1 (one)).
Model 500 Gas Chromatograph
This section applies only to GC systems that have not been shipped "prewired" or may not have had the COM ID set according to customer specifications.
Follow the steps in this section only if you wish to do the following:
1. Change the GC Controller's COM ID, or
2. Visually inspect and verify the COM ID as determined by the DIP switch settings.
Installing the GC Controller JULY 2010
Model 500 Gas Chromatograph
To inspect or change the GC Controller's COM ID setup, follow these steps:
1. At the GC Controller site, locate the DIP switch as described in the following steps.
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Before removing the unit cover from the GC Controller, make certain the power supply switch is OFF and the AC power cord is disconnected. Observe all safety precautions when you are working in a hazardous environment.
Failure to observe all safety precautions could result in serious injury or death.
Installation and Setup 3-31
2. For the explosion-proof Controller, the front panel is secured by 16 screws. Remove those screws first.
Then carefully lower the front panel on its bottom hinges. The front panel is heavy, so make sure it does not drop and cause damage. The DIP switch is located on the lower left side of the front panel (see
Figure 3-6).
JULY 2010 Modbus Slave Address (COM ID) Setup
3-32 Installation and Setup
Model 500 Gas Chromatograph
Figure 3-6. Explosion-Proof Controller DIP Switch
3. For rack mount and panel mount Controllers, use a flat head screw driver to remove the access panel on the right side of the card cage assembly (see Figure 3-7).
Figure 3-7. Right Side View of Rack Mount and Panel Mount Controllers
Modbus Slave Address (COM ID) Setup JULY 2010
Model 500 Gas Chromatograph
4. Inspect or change the DIP switch settings as necessary.
(a) See Table 3-2 as a guide. (b) Make sure you record in the GC Controller's maintenance records
any changes you make to the switch settings.
5. When finished with the inspection or changes, reassemble the
Controllers using the following steps.
(a) For explosion-proof Controllers, close the front panel and replace
the screws.
(b) For Rack mount and Panel mount Controllers, reattach the right
side Access Panel and secure with the four flat head screws.
Installation and Setup 3-33
Figure 3-8. COM ID DIP Switch
Explanation of DIP Switch Setting
Switches "1" through "5" form a 5-bit binary number for setting the
Modbus slave address (also known as COM ID or Device ID.)
Switch number "1" is the least significant bit, and switch number "5"
is the most significant bit.
Switch to ON = 1
Switch to OFF = 0
Switch "6" is a spare for future use. Switches "7" and "8" are set as
needed for the presence of an optional LOI (Local Operator Interface)
JULY 2010 Modbus Slave Address (COM ID) Setup
3-34 Installation and Setup
Model 500 Gas Chromatograph
connected via COM8 When the COM4A Board is installed. If the COM4A Board is not installed, the LOI is connected via COM4.
Table 3-2. Modbus Slave Address (COMID) DIP Switch Settings
Dip Switch Settings Switch Positions
COM ID12345
1 ON OFF OFF OFF OFF
2 OFF ON OFF OFF OFF
3 ON ON OFF OFF OFF
4 OFF OFF ON OFF OFF
5 ON OFF ON OFF OFF
6 OFF ON ON OFF OFF
7 ONONONOFFOFF
8 OFF OFF OFF ON OFF
RAM CLEAR
Dip Switch Setting Switch Positions
8
Clears RAM when unit powered down ON
Keeps RAM when unit powered down OFF
Modbus Slave Address (COM ID) Setup JULY 2010
Model 500 Gas Chromatograph
3.4.2 Controller-Analyzer Wiring
4 Install Analyzer Wiring
5 Install Analyzer Sample & Gas Lines
6 Install GC Controller Wiring
7 Perform Leak Checks
This section applies only to GC systems that have not been shipped "prewired." In most cases, the explosion-proof system will already have had the Controller-Analyzer connections made. If your system has already been wired, skip this section, and proceed to the next section.
Installation and Setup 3-35
Applicable to the digital and analog I/O lines connecting to the GC Controller, including the Analyzer-Controller Interconnect lines: Any loop of extra cable left for service purposes inside the GC Controller explosion-proof housing must not be placed near the conduit entry for AC power.
If the above precaution is not followed, the data and control signals to and from the GC Controller can be adversely affected.
To make wiring connections between the GC Controller and Analyzer, follow these steps:
1. Disconnect all electrical power to both the Analyzer and the GC
Controller.
2. Ensure that Interconnect Cable wiring connections to the Analyzer
have been made as explained earlier in Section 3.3.1, this manual.
JULY 2010 Controller-Analyzer Wiring
3-36 Installation and Setup
Model 500 Gas Chromatograph
3. At the GC Controller site, remove the Controller enclosure's front panel.
(a) For the explosion-proof Controller, the front panel is secured by 16
screws. Remove those screws first.
(b) Then carefully lower the front panel on its bottom hinges. The
front panel is heavy, so make sure it does not drop and cause damage.
(c) For the Rack mount and the Panel mount Controllers, the rear of
the enclosure is open; and an Access Panel on the right side of the unit allows access for most field wiring procedures without removing the enclosure.
4. Locate the GC Controller's Terminal Board for Field Wiring (TB). The TB is attached to the GC Controller's card cage assembly, facing the enclosure's front panel. (In the Rack mount Controller, the TB faces outward toward the rear of the enclosure.)
5. Route the Analyzer-Controller Interconnect Cable appropriately, especially in the case of the explosion-proof Controller enclosure.
Figure 3-9. Seperate Conduit Entries for Cable In/Out of GC Controller
Controller-Analyzer Wiring JULY 2010
Model 500 Gas Chromatograph
6. Make Interconnect Cable wiring connections to the GC Controller TB
as listed earlier (see Section 3.3.1, “Point-to-point Wiring Guide,
Analyzer-Controller” on page 18 and Table 3-1).
3.4.3 Controller PC Wiring (Serial Connections)
4 Install Analyzer Wiring
5 Install Analyzer Sample & Gas Lines
6 Install GC Controller Wiring
7 Perform Leak Checks
A preferred method for operating a Model 500 Analyzer System is from a connected personal computer (PC). The PC must be:
Running MON2000 software
Installation and Setup 3-37
Connected to the GC System by a serial link
This section of the manual addresses the various possibilities for wiring a serial connection between a PC and the GC System.
Before Connecting
Before connecting a PC to the GC Controller, determine the following:
JULY 2010 Controller PC Wiring (Serial Connections)
3-38 Installation and Setup
Model 500 Gas Chromatograph
What serial ports are available at the PC? When you select one, consider these points:
- Standard PC serial ports are type RS-232.
- Usually there are two external serial port jacks on a PC, located on
the rear panel. Most often, they are either DB-9 or DB-25 male (see below).
- PC serial ports can be designated as "COM1" through "COM8," and
they can be used by other peripheral equipment attached to the PC, such as printers, mice, or modems, etc.
You will need to connect the GC Controller to one of the PC's available, or unused serial ports.
To determine which PC serial ports are already being used by other equipment and which port can be used for connecting to the GC Controller, note existing serial connections, refer to your PC user's manual, and use diagnostic software (such as Norton Utilities™).
Controller PC Wiring (Serial Connections) JULY 2010
Model 500 Gas Chromatograph
What serial ports are available at the GC Controller? When you select
one, consider these points:
- The GC Controller's COM1 serial channel is usually reserved for connecting a PC, especially for service or troubleshooting purposes, since the GC Controller's easy-access front panel serial port is connected to serial channel COM1.
Installation and Setup 3-39
Figure 3-10. The Front Panel Serial Port is Connected to GC Controller’s Serial Channel COM1
- COM8 is used for the display/keypad when unit has COM5-8 option (COM4A Board). However, when the unit has a display/ keypad but does not have the COM8 option, COM4 must be RS­232, and is used for the display/keypad. Therefore, there is no COM4 output on the field Terminal Board.
- The optional Modem piggy-backs on to the CPU 104 BUS (See Internal Modem for the Model 2350A Gas Chromatograph, drawing BE-20767).
JULY 2010 Controller PC Wiring (Serial Connections)
3-40 Installation and Setup
- Any one of the Controller's eight serial channels could also be reserved for connecting to a Data Collection Systems (DCS) or multi-drop serial data highway system (and thus, would be unavailable for a serial PC connection).
Is the connection to be made in a...
- Nonhazardous environment?
- Short distance between the PC and Controller?
- With temporary or permanent cable connection?
See “PC-to-GC, Front Panel Quick and Easy RS-232” on page 40.
Is the connection to be made in a...
- Hazardous or nonhazardous environment?
- Short distance between the PC and Controller?
- Permanent cable connection?
Model 500 Gas Chromatograph
See “PC-to-GC, Permanent Cable Connection for Short Distance RS-
232” on page 41.
Is the connection to be made with a...
- Long distance between the PC and Controller?
- Permanent cable connection?
See “PC-to-GC, Long Distance with RS-422 or RS-485” on page 43.
PC-to-GC, Front Panel Quick and Easy RS-232
The easiest way to connect a PC to the GC Controller is with an 'off-the­shelf', straight-through serial cable connected to the GC Controller's front panel DB-9 serial port jack.
Controller PC Wiring (Serial Connections) JULY 2010
Loading...