Safety Summary ………………………………………………………………………… ………… … 1
Specifications – Analysis Enclosure: General …………………………………………………….. 4
Specifications – Probe/Sample Handling Enclosure: General ..………………………………… 5
Customer Service, Technical Assistance and Field Service ………..………………………….. 6
Returning Parts to the Factory………………………………………………………………………. 6
Training ……………………………………………………………………………… ………………... 7
Preface
Intended Use Statement …………………………………………………………………………….. 1
The µCEM Continuous Emission Monitoring Gas Analyzer is intended for use as an
industrial process measurement device only. It is not intended for use in medical,
diagnostic, or life support applications, and no independent agency certifications or
approvals are to be implied as covering such applications.
SAFETY SUMMARY
DANGER is used to indicate the presence of a hazard which will cause severe
personal injury, death, or substantial property damage if the warning is ignored.
WARNING is used to indicate the presence of a hazard which can cause severe
personal injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor
personal injury or property damage if the warning is ignored.
NOTE
IS IMPORTANT BUT NOT HAZARD RELATED
IS USED TO INDICATE INSTALLATION, OPERATION, OR MAINTENANCE INFORMATION WHICH
.
DANGER: ALL PERSONNEL AUTHORIZED TO INSTALL,
OPERATE AND SERVICE THIS EQUIPMEN
To avoid explosion, loss of life, personal injury and damage to this equipment
and on-site property, do not operate or service this instrument before reading
and understanding this instruction manual and receiving appropriate training.
Save these instructions.
If this equipment is used in a manner not specified in these instructions,
protective systems may be impaired.
WARNING: DEVICE CERTIFICATION(S)
Any addition, substitution, or replacement of components installed on or in this
device, must be certified to meet the hazardous area classification that the
device was certified to prior to any such component addition, substitution, or
replacement. In addition, the installation of such device or devices must meet
the requirements specified and defined by the hazardous area classification of
the unmodified device. Any modifications to the device not meeting these
requirements, will void the product certification(s).
This device may contain explosive, toxic or unhealthy gas components. Before
cleaning or changing parts in the gas paths, purge the gas lines with ambient air
or nitrogen.
+
WARNING: ELECTRICAL SHOCK HAZARD
POSSIBLE EXPLOSION HAZARD
Do not open while energized. Do not operate without dome and covers secure.
Installation requires access to live parts which can cause death or serious injury.
WARNING: ELECTRICAL SHOCK HAZARD
For safety and proper performance this instrument must be connected to a
properly grounded three-wire source of power.
WARNING: POSSIBLE EXPLOSION HAZARD
Ensure that all gas connections are made as labeled and are leak free. Improper
gas connections could result in explosion and death.
WARNING: TOXIC GAS
This unit’s exhaust may contain hydrocarbons and other toxic gases such as
carbon monoxide. Carbon monoxide is highly toxic and can cause headache,
nausea, loss
Avoid inhalation of the exhaust gases at the exhaust fitting.
of consciousness, and death.
Connect exhaust outlet to a safe vent using stainless steel or Teflon line. Check
vent line and connections for leakage.
Keep all tube fittings tight to avoid leaks. See Section 3.3.5 for leak test
information.
Tampering with or unauthorized substitution of components may adversely
affect the safety of this instrument. Use only factory approved components for
repair.
Because of the danger of introducing additional hazards, do not perform any
unauthorized modification to this instrument.
Return the instrument to a Rosemount Analytical Service office for service or
repair to ensure that safety features are maintained.
CAUTION: PRESSURIZED GAS
This unit requires periodic calibration with a known standard gas. It also may
utilize a pressurized carrier gas, such as helium, hydrogen, or nitrogen. See
General Precautions for Handling and Storing High Pressure Gas Cylinders at the
rear of this manual.
CAUTION: HEAVY WEIGHT
U
SE TWO PERSONS OR A SUITABLE LIFTING DEVICE TO MOVE OR
Chemiluminscent (NOx) // Up to three in one analyzer
Mounting: Wall Mount
Area Classification: General Purpose / NEMA 4X Fiberglass Enclosure Compliant
Compliance's: CSA (Pending)
Ambient Temperature Range: -30
0
to 500 Celsius.
PREFACE
Relative Humidity: 5 to 99%
Inputs/Outputs: The complete I/O list with terminal locations is located in section
3.3.4
Digital:
RS-485 Serial Port. (Multi-Drop Network)
RS-232 Serial Port.
LAN, Ethernet 10/100-BaseT
Connectivity Protocols:
HTML (Web Browser) – Status, file transfer Modem / Web browser
TCP/IP, MTTP ASCII String
Microsoft Shared drive
FTP Logs download
TELNET Server
Analog:
Analog Outputs: Qty. 3 Isolated 4-20 mA dc, 500 ohms Max Load (O2, CO or SO2,
NOx)
Mounting: Customer Flange Mount (2 Hole Top) or Wall Mount for High Temp Option
Area Classification: General Purpose / NEMA 4X Fiberglass Enclosure
Compliance's: CSA (Pending)
Ambient Range Temperature: -30
Relative Hum: 5 to 99%
Instrument Weight: 95 lbs Typical
Size: 24“ X 34“ X 12“ (H W D)
Stack Sample Moisture: Up to 25% max
Sample Cooler: Thermo Electric dual pass Chiller. Permeation Tube (-30 degrees C.
Dewpoint. Customer instrument air required @ 5 L/M, -40 degree C dewpoint
Max. Stack Temperature: Standard 400Optional: 600 F (available with elongated spool option)
High Temp: 1400 F (Off Stack Option)
Stack Pressure: Typical -5 to 15 inches H
Sample Flow Rate: 500 to 2500cc/min
Response Time: Maximum distance between Analysis Enclosure and Sample
Conditioning/Probe Enclosure is 300'. (Response time is 30 seconds/100' w/1/4"
tubing)..
Probe Length: 48" length 316 SS Probe with .5 micron sintered filter. Customer to cut
to length in field if necessary. Optional 5’ and 6’ probes.
Mounting Flange: Standard 4“ 150# Raised Face. Shipped Equipped with Gasket
Sample Pump: 316 SS diaphragm type
Instrument Air Requirements: Instrument grade air required. 15 SCFM @ 60 -100 PSIG (30
seconds 2 times per day) Pressure Regulation by Customer
0
to 500 Celsius
0
F.
2
O
CUSTOM ER SERVICE,TECHNICAL ASSIST ANCE AND FIELD SER VICE
For order administration, replacement parts, application assistance, on-site or factory
repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when you
call: Model Number, Serial Number, and Purchase Order Number or Sales Order
Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive
or other hazardous material or used in such a hazardous environment, the user must
attach an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification
that the material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
1201 North Main St.
Orrville, OH 44667
USA
PREFACE
TRAINING
A comprehensive Factory Training Program of operator and service classes is
available. For a copy of the Current Operator and Service Training Schedule contact
the Technical Services Department at:
Rosemount Analytical Inc.
Phone: 1-330-682-9010
COMPLIANCES
This product may carry approvals from several certifying agencies. The certification
marks appear on the product name-rating plate.
This manual describes the Rosemount Analytical Micro Continuous Emission
Monitoring (µCEM) gas Analyzer Module.
The µCEM Analyzer Module is designed to continuously determine the concentration of
O2, CO, SO2, and NOx in a flowing gaseous mixture. The concentration is expressed
in percent or parts-per-million.
The sampled gas is collected from the stack and prepared by the Probe/Sample
Handling Enclosure for analysis and processing by the Analysis Enclosure. The
ANALYSIS ENCLOSURE is a stand alone, computer-controlled unit, utilizing PC/104
as the system bus. The uCEM is enclosed in rugged NEMA 4X, IP65 type enclosures,
for harsh environment. The ANALYSIS ENCLOSURE utilizes convection cooling with
no air intake and air vents. The ANALYSIS ENCLOSURE is modular, general purpose
and easily expandable. It utilizes industry standard components such as PC/104
boards, and modular signal conditioning modules.
Figure 1-2. µCEM Micro Continuous Emission Monitoring Gas Analyzer with Time
Share option.
1.2 Time Shared Option
Provides the functionality to monitor and process sample gases from two streams on a
time-share scheme. This option allows you to connect one uCEM to two Sample
Handling units.
The NOx analyzer continuously analyzes a flowing gas sample for NOx [nitric oxide
(NO) plus nitrogen dioxide (NO
reported as NOx.
The µCEM NOx Analyzer Module uses the chemiluminecence method of detection.
This technology is based on NO’s reaction with ozone (O
(O
). Some of the NO2 molecules produced are in an electronically excited state (NO2*
2
where the * refers to the excitation). These revert to the ground state, with emission of
photons (essentially, red light). The reactions involved are:
NO
+ O3 → NO2* + O
2
NO2* → NO2 + red light
The sample is continuously passed through a heated bed of vitreous carbon, in which
NO
is reduced to NO. Any NO initially present in the sample passes through the
2
converter unchanged, and any NO
amount of NO.
)]. The sum of the concentrations is continuously
2
) to produce NO2 and oxygen
3
2
is converted to an approximately equivalent (95%)
2
1.3.2 CO
The NO is quantitatively converted to NO
by gas-phase oxidation with molecular
2
ozone produced within the analyzer from air supplied by an external source. During the
reaction, approximately 10% of the NO
molecules are elevated to an electronically
2
excited state, followed by immediate decay to the non-excited state, accompanied by
emission of photons. These photons are detected by a photomultiplier tube which
produces an output proportional to the concentration of NOx in the sample.
To minimize system response time, an internal sample bypass feature provides highvelocity sample flow through the analyzer.
The optical bench can selectively measure multiple components in a compact design by using a
unique dual optical bench design. Depending on the application, any two combinations of NDIR
channels can be combined on a single chopper motor/dual source assembly.
Other application-dependent options include a wide range of sample cell materials, optical filters and
solid state detectors. The NDIR Microflow detector consists of two chambers, measurement and
reference with an interconnected path in which an ultra low flow filament sensor is mounted. During
operation, a pulsating flow occurs between the two chambers which is dependent upon: sample gas
absorption, modulation by the chopper motor and the fill gas of the detector chambers. The gas
flow/sensor output is proportional to the measured gas concentration. The optical bench is further
enhanced by a novel “Look-through” detector technique. This design allows two detectors to be
arranged in series --- enabling two different components to be measured on a single optical bench.
The optical bench contains a unique eddy current drive chopper motor and source assembly. This
design incorporates on board “intelligence” to provide continuous “self test” diagnostics.
Paramagnetic: The determination of oxygen is based on the measurement of the magnetic
susceptibility of the sample gas. Oxygen is strongly paramagnetic, while other common gases
are not. The detector used is compact, has fast response and a wide dynamic range. The long
life cell is corrosion resistant, heated and may be easily cleaned. It has rugged self-tensioning
suspension and is of welded Non-Glued construction.
The optical bench can selectively measure multiple components in a compact design by using a
unique dual optical bench design. Depending on the application, any two combinations of NDIR
channels can be combined on a single chopper motor/dual source assembly.
Other application-dependent options include a wide range of sample cell materials, optical filters
and solid state detectors. The NDIR Microflow detector consists of two chambers, measurement
and reference with an interconnected path in which an ultra low flow filament sensor is mounted
during operation, a pulsating flow occurs between the two chambers which is dependent upon:
sample gas absorption, modulation by the chopper motor and the fill gas of the detector
chambers. The gas flow/sensor output is proportional to the measured gas concentration. The
optical bench is further enhanced by a novel “Look-through” detector technique. This design
allows two detectors to be arranged in series --- enabling two different components to be
measured on a single optical bench. The optical bench contains a unique eddy current drive
chopper motor and source assembly. This design incorporates on board “intelligence” to provide
continuous “self test” diagnostics.
The µCEM can employ up to three different measuring methods depending on the
configuration chosen. The methods are: NDIR CO/SO2, Paramagnetic O
Electrochemical O
, and chemiluminescent NOx.
2
2.1 Non-Dispersive Infrared (NDIR)
The non-dispersive infrared method is based on the principle of absorption of infrared
radiation by the sample gas being measured. The gas-specific wavelengths of the
absorption bands characterize the type of gas while the strength of the absorption gives
a measure of the concentration of the gas component being measured.
An optical bench is employed comprising an infrared light source, two analysis cells
(reference and measurement), a chopper wheel to alternate the radiation intensity
between the reference and measurement side, and a photometer detector. The
detector signal thus alternates between concentration dependent and concentration
independent values. The difference between the two is a reliable measure of the
concentration of the absorbing gas component.
Depending on the gas being measured and its concentration, one of two different
measuring methods may be used as follows:
,
2
2.1.1 Interference Filter Correlation Method
With the IFC method the analysis cell is alternately illuminated with filtered infrared
concentrated in one of two spectrally separated wavelength ranges. One of these two
wavelength bands is chosen to coincide with an absorption band of the sample gas and
the other is chosen such that none of the gas constituents expected to be encountered
in practice absorbs anywhere within the band.
The spectral transmittance curves of the interference filters used in the µCEM analyzer
and the spectral absorption of the gases CO and CO
can be seen that the absorption bands of these gases each coincide with the
passbands of one of the interference filters. The forth interference filter, used for
generating a reference signal, has its passband in a spectral region where none of
these gases absorb. Most of the other gases of interest also do not absorb within the
passband of this reference filter.
The signal generation is accomplished with a pyroelectrical (solid-state) detector. The
detector records the incoming infrared radiation. This radiation is reduced by the
absorption of the gas at the corresponding wavelengths. By comparing the
measurement and reference wavelength, an alternating voltage signal is produced.
This signal results from the cooling and heating of the pyroelectric detector material.
Figure 2-1. Absorption Bands of Sample Gas and Transmittance of Interference Filters
2.1.2 Opto-Pneumatic Method
In the opto-pneumatic method, a thermal radiator generates the infrared radiation which
passes through the chopper wheel. This radiation alternately passes through the filter
cell and reaches the measuring and reference side of the analysis cell with equal
intensity. After passing another filter cell, the radiation reaches the pneumatic detector.
The pneumatic detector compares and evaluates the radiation from the measuring and
reference sides of the analysis cell and converts them into voltage signals proportional
to their respective intensity.
The pneumatic detector consists of a gas-filled absorption chamber and a
compensation chamber which are connected by a flow channel in which a Microflow
filament sensor is mounted. This is shown in Figure 2-2 below.
In principle the detector is filled with the infrared active gas to be measured and is only
sensitive to this distinct gas with its characteristic absorption spectrum. The absorption
chamber is sealed with a window which is transparent for infrared radiation. The
window is usually Calcium Fluoride (CaF
).
2
When the infrared radiation passes through the reference side of the analysis cell into
the detector, no pre-absorption occurs. Thus, the gas inside the absorption chamber is
heated, expands and some of it passes through the flow channel into the compensation
chamber.
When the infrared radiation passes through the open measurement side of the analysis
cell into the detector, a part of it is absorbed depending on the gas concentration. The
gas in the absorption chamber is, therefore, heated less than in the case of radiation
coming from the reference side. Absorption chamber gas becomes cooler, gas
pressure in the absorption chamber is reduced and some gas from the compensation
chamber passes through the flow channel into the absorption chamber.
The flow channel geometry is designed in such a way that it hardly impedes the gas
flow by restriction. Due to the radiation of the chopper wheel, the different radiation
intensities lead to periodically repeated flow pulses within the detector.
The Microflow sensor evaluates these flow pulses and converts them into electrical
pulses which are processed into the corresponding analyzer output.
In the case of dual-channel analyzers, the broadband emission from two infrared
sources pass through the chopper wheel. In the case of the Interference Filter
Correlation (IFC) method, the infrared radiation then passes through combinations of
interference filters. In the case of the opto-pneumatic method, the infrared radiation
passes through an optical filter depending on the application and need for reduction of
influences. Then the infrared radiation enters the analysis cells from which it is focused
by filter cells onto the corresponding detector. The preamplifier detector output signal is
then converted into the analytical results expressed directly in the appropriate physical
concentration units such as percent volume, ppm, mg/Nm
2-3 below.
The paramagnetic principle refers to the induction of a weak magnetic field, parallel and
proportional to the intensity of a stronger magnetizing field.
The paramagnetic method of determination of oxygen concentration utilizes nitrogen
filled quartz spheres arranged at opposite ends of a bar, the center of which is
suspended by and free to rotate on a thin platinum wire ribbon in a cell. Nitrogen (N2) is
used because it is diamagnetic or repelled by a magnet.
A small mirror that reflects a light beam coming from a light source to a photodetector,
is mounted on the platinum ribbon. A strong permanent magnet specifically shaped to
produce a strong, highly inhomogeneous magnetic field inside the analysis cell, is
mounted outside the wall of the cell.
When oxygen molecules enter the cell, their paramagnetism will cause them to be
drawn towards the region of greatest magnetic field strength. The oxygen molecules
thus exert different forces on the two suspended nitrogen filled quartz spheres,
producing a torque which causes the mirror to rotate away from its equilibrium position.
The rotated mirror deflects the incident light onto the photodetector creating an
electrical signal which is amplified and fed back to a coil attached to the bar holding the
quartz spheres, forcing the suspended spheres back to the equilibrium position.
The current required to generate the restoring torque to return the quartz bar to its
equilibrium position is a direct measure of the O
concentration in the sample gas.
2
The complete paramagnetic analysis cell consists of an analysis chamber, permanent
magnet, processing electronics, and a temperature sensor. The temperature sensor is
used to control a heat exchanger to warm the measuring gas to about 55 °C.
The electrochemical oxygen sensor incorporates a lead and gold galvanic process with
a lead anode (1) and a gold cathode (2), using an acid electrolyte (3).
Oxygen molecules diffuse through a non-porous Teflon membrane (4) into the
electrochemical cell and are reduced at the gold cathode. Water is the byproduct of this
reaction.
On the anode, lead oxide is formed which is transferred into the electrolyte. The lead
anode is continuously regenerated and, therefore, the electrode potential remains
unchanged for a long time. The rate of diffusion and corresponding response time (t
90
)
of the sensor is dependent on the thickness of the Teflon membrane.
The electric current between the electrodes is proportional to the O
concentration in
2
the sample gas being measured. The resultant signal is measured as a voltage across
the resistor (6) and thermistor (5), the latter of which is used for temperature
compensation. A change in the output voltage (mV) represents oxygen concentration.
NOTE: The electrochemical O
cell requires a minimum internal consumption of
2
oxygen. Sample gases with an oxygen concentration of less than 2% could result in a
reversible detuning of sensitivity and the output will become unstable. The
recommended practice is to purge the cell with conditioned ambient air between
periods of measurement. If the oxygen concentration is below 2% for several hours or
days, the cell must be regenerated for about one day with ambient air. Temporary
flushing with nitrogen (N
) for less than one hour (analyzer zeroing) will have no effect
Installation and servicing of this device requires access to components
which may present electrical shock and/or mechanical hazards. Refer installation
and servicing to qualified service personnel.
CAUTION: CODE COMPLIANCE
Installation of this device must be made in accordance with all
applicable national and/or local codes. See specific references on installation
drawing located in the rear of this manual.
3.1 Specifications
Electrical Power
See Specifications in Preface
Power Cable
AC Operation: 16 gauge, minimum.
Gas Lines
For external gas lines, the use of all new tubing throughout is strongly recommended.
The preferred type is new, Teflon or Stainless Steel tubing, sealed at the ends.
Services
AC as well as input and output digital and analog signals connect through the circular
connectors located on the bottom of the uCEM enclosures.
1. Set MicroCEM Pressure guage (P1)to 5 psig +/- 0.5psig. Pressure set by BPR located behind
gauge in detector section. If CO and NOx response times are sluggish this pressure can be
increased.
2. Set Calibration gas cylinder dual stage pressure regulators to 10 to 20 psig.
3. Set Flowmeter (F1) to 500cc to 1500cc per min.
4. TV1 is used to balance the flow between a probe and local calibration. It is located beside the
solenoid valve manifold.
5. Set Ozone air pressure to 12 psig.
6. Exhaust line should be free of any backpressure. Immediately vent into ½” pipe.
7. Time Share Box:
TV1: Use to equalize cal gas flow between SHU1 and SHU2.
TV2: Use to equalize cal gas flow between SHU1 and SHU2.
TV3: Use to equalize sample flow between SHU1 and SHU2.
TV4: Use to equalize sample flow between SHU1 and SHU2.
8. Pressure Switch: The pressure switch is located beside the pressure gauge. If the sample or
cal gas pressure flow below 2.5 psig the MicroCEM will give trouble alarm. The alarm will turn
off upon pressure above 4 psig.
3.2 Process and Calibration Gas Connection
Besides sample gas, the µCEM requires other gases for operation. In most cases, one
or more Calibration Standards must be provided. These should be cylinders of gas
which closely resemble the expected sample, both in species and concentrations.
These calibration gases are normally introduced into the system as an input to the
Sample Conditioning Plate Option or sample conditioning may be provided by others.
Each gas cylinder should be equipped with a clean, hydrocarbon free two-stage
regulator with indicating gauges of approximately 0 to 3000 psig (0 to 20.7 Mpa) for
cylinder pressure and 0 to 100 psig (0 to 689 Kpa) for delivery pressure. Regulators
should have a metallic as opposed to elastomeric diaphragm, and provide for ¼ inch
compression fitting outlet and should be LOX clean.
NOTE: All connections specified in the Installation Drawing, in conjunction with
the Application Data Sheet, should be made.
All gases must be supplied to the analyzer as conditioned gases! When the system is
used with corrosive gases, it must be verified that there are no gas components which
may damage the gas path components.
The gas conditioning must meet the following conditions:
Free of condensable constituents
Free of dust above 2 µm
Free of aggressive constituents which may damage the gas paths
Temperature and pressure in accordance with the specifications
When analyzing vapors, the dewpoint of the sample gas must be at least 10 °C below
the ambient temperature in order to avoid the precipitation of condensate in the gas
paths.
An optional barometric pressure compensation feature can be supplied for the µCEM.
This requires a pressure sensor with a range of 800 – 1,100 hPa. The concentration
values computer by the detectors will then be corrected to eliminate erroneous
measurements due to changes in barometric pressure.
The gas flow rate must be in the range of 0.5 l/min to a maximum of 1.5 l/min. A
constant flow rate of 1 l/min is recommended. NOTE: The maximum gas flow rate for
paramagnetic oxygen detectors is 1.0 l/min!
Care should be taken if hazardous gases are to be measured or used for
calibration.
Refer to installation drawing supplied with the application data package.
3.3.1 Location
The µCEM is designed to be installed in an outdoor environmental location. It is highly
recommended that the analyzer be located out of direct sunlight and direct rain/snow to
the extent possible to assure longevity and accuracies.
The µCEM analysis enclosure should be installed as near as possible to the
probe/sample handling enclosure, in order to avoid low response time caused by long
sample gas lines.
The enclosure must be grounded to earth by the user or ground loops and computer
lockups are possible.
3.3.2 Limitations
Ambient Temperature: -30° to 50° Celsius (-34° to 122° F)
Relative Humidity: 5% to 99%
3.3.3 Mounting Options
Although the µCEM is enclosed in an environmentally sealed enclosure, it should be
protected from direct sunlight. In areas subjected to harsh winter climates, protection
should be provided from sun, rain and snow. A corrigated awning or other suitable
means can be provided to meet these conditions.
3.3.4 Electrical Connections
NOTE: The enclosure is a NEMA 4x. All entry locations must be sealed.
Connect all required signal cables to the connections at the bottom of the µCEM. The
cable locations are indicated on the inside bottom cover of the µCEM box. The actual
electrical connections will be specified in the Application Data package. All connections
are not necessary for every application.
Cable length for these signals should not exceed 3,000 feet (914 meters), to avoid
excessive capacitance and corresponding signal distortion.
All connections are made through the bottom of the µCEM enclosure using circular
connectors. Mating circular external connectors are provided by Rosemount with a 6’
wire harness pigtail for connections to J1, J3, J5, J6, J7 & J8.
J1 – AC Power Input J2 – CPU I/O
J3 – COM Interface (pocket pc) J4 – Ethernet LAN Port
J5 – EXT I/O Interface J6 – SHU #1 Interface
J7 – SHU #2 Interface (T/S units only) J8 – SSU Power (T/S units only)
3.3.4.2 EXT I/O Interface Connector (J5) - MicroCEM inputs and outputs are specific for
customer use.
The Analog Interface connector has a shell size of 22, 100 contacts. Each pin will accept a wire
size of 26, 24, or 22 AWG. Connector and 6’ pigtail by Rosemount.
3.3.4.3 SHU #1 / #2 Interface Connector (J6 & J7). These wires are to be connected directly to
the MicroCEM sample handling enclosure (SHU) and will control the operation of the
sample pump, drain pump, purge valve and calibration valve respetively. All toggle
switches in sample handling enclosure should be set to “remote” mode upon hookup of wire
so the MicroCEM analysis enclosure will control the full system.
The Digital Interface connector has a shell size of 14, 15 contacts. Each pin will accept a wire
size of 20 AWG. Connector and 6’ pigtail by Rosemount.
PIN NAME DESCRIPTION COLOR Sample Handling
1 SPUMP1/2NO BLK Not Used
2 SPUMP1/2C BRN 1
3 SPUMP1/2NC
4 DPUMP1/2NO ORG Not Used
5 DPUMP1/2C YEL 1
6 DPUMP1/2NC
3.3.4.4 COM Interface Connector (J3) – Pocket PC external connection
The COM Interface connector has a shell size of 10, 13 contacts. Each pin will accept a wire
size of 28, 26, or 24 AWG. Connector and 3’ pigtail by Rosemount.
SIGNAL NAME DEFINITION PIN
DCD (pin 1)
DSR (pin 6)
RxD (pin 2)
RTS (pin 7)
TxD (pin 3)
CTS (pin 8)
DTR (pin 4)
RI (pin 9)
GND (pin 5) Signal Ground, RS232 9
TxD/RxD+ (pin 2)
TxD/RxD- (pin 7)
GND (pin 3)
VCC
Data Carrier Detect Input, RS232
Data Set Ready Input, RS232
Receive Data Input, RS232
Request to Send Output, RS232
Transmit Data Output, RS232
Clear To Send Input, RS232
Data Terminal Ready Output, RS232
Ring Indicator Input, RS232
RS-485 Bidirectional Data
RS-485 Bidirectional Data
Signal Ground
+5V DC
Table 3-3. COM Interface Terminal Assignments
1
2
3
4
5
6
7
8
10
11
12
13
3.3.4.5 Lan Interface Connector (J4) – Customer PC, network or laptop connection
The Lan Interface connector has a shell size of 8, 6 contacts. Each pin will accept a wire size
of 28, 26, 24, or 22 AWG.
SIGNAL NAME DEFINITION PIN
TxD+ (pin 1) 1
TxD- (pin 2)
RxD+ (pin 3) 3
RxD- (Pin 6)
Transmit Data
Receive Data
Not Used
Table 3-4. LAN Interface Terminal Assignments
5-6
2
4
3.3.4.6 CPU I/O Interface Connector (J2) – Rosemount Factory trained port for
communication with CPU hard drive
The CPU I/O Interface connector has a shell size of 14, 19 contacts. Each pin will accept a
wire size of 28, 26, or 24 AWG.
PIN NAME DESCRIPTION
A RED RED CENTER
B GND RED SHIELD
C GREEN GREEN CENTER
D GND GREEN SHIELD
E BLUE BLUE CENTER
F GND BLUE SHIELD
N KBDATA KEYBOARD DATA
P KBCLK KEYBOARD CLOCK
R GND GROUND
S VCC VCC, +5VDC
R GND GROUND
S VCC VCC, +5VDC
T MSDATA MOUSE DATA
U MSCLK MOUSE CLOCK
Table 3-5. CPU I/O Terminal Assignments
INSTALLATION
3.3.4.7 SSU Power Connector, T/S units Only (J8) – T/S enclosure can be located away from
the Analysis enclosure. This cable serves as the connection and is by Rosemount.
The SSU Power connector has a shell size of 8, 3 contacts. Each pin will accept a wire size of
24, 22, or 20 AWG. Connector and 6’ pigtail by Rosemount.
SIGNAL NAME DEFINITION PIN
SSUCtrl
Vbb_rtn
Gnd
SSU Control line
+24V Return
GND
A
B
C
Table 3-6. SSU Power Connection Terminal Assignments
3.3.4.8 AC Power Connector (J1) – Customer 120VAC Power Connection
The AC Power Interface connector has a shell size of 12, 3 contacts. Each pin will accept a
wire size of 16 AWG. Connector and 6’ pigtail by Rosemount.
SIGNAL NAME DEFINITION PIN
L1 A
L2
GND AC Ground B
85-264 VAC, 47-440 Hz
C
Table 3-7. AC Power Connection Terminal Assignments
Connect AC power through a 20A circuit breaker that is to be located close to the
µCEM. The circuit breaker will provide over current protection as well as a means of
disconnecting the power.
Maximum power requirements will be 1000 watts, with most applications requiring less
than this amount
If explosive or hazardous gas samples are being measured with the µCEM, it is
recommended that gas line fittings and components be thoroughly leak-checked prior
to initial application of electrical power, and at bimonthly intervals thereafter, as well as
after any maintenance which involves breaking the integrity of the sample containment
system.
3.3.5.1 Flow Indicator Method
Figure 3-8. Leak Test Flow Method
Supply air or inert gas such as nitrogen, at 10 psig (689 hPa), to the analyzer through a
flow indicator with a range of 0 to 250 cc/min. Install a shut-off valve at the sample gas
outlet. Set the flow rate to 125 cc/min.
Close the outlet shut-off valve and notice that the flow reading drops to zero. If the flow
reading does not drop to zero, the system is leaking and must be corrected before the
introduction of any flammable sample gas or application of power.
3.3.5.2 Manometer Method
Install a water-filled U-tube manometer at the sample gas outlet. Install a shut-off valve
at the sample gas inlet. Admit air or inert gas to the inlet shut-off valve until the analyzer
is pressurized to approximately 50 hPa. The water column will be about 500 mm.
Close the inlet shut-off valve and, following a brief period for pressure equilibrium, verify
that the height of the water column does not drop over a period of about 5 minutes. If
the water column height drops, the system is leaking and must be corrected before the
introduction of any flammable sample gas or application of power.
Liberally cover all fittings, seals, and other possible sources of leakage with a suitable
leak test liquid such as SNOOP
™
(part 837801). Bubbling or foaming indicates
leakage. Checking for bubbles will locate most leaks but could miss some, as some
areas are inaccessible to the application of SNOOP. For positive assurance that
system is leak free, perform one of the preceding tests.
NOTE:
Refer to Specification in Preface for maximum pressure limitations.
For differential measurement, the leak check must be performed for the
measurement and reference side separately.
For analyzers with parallel gas paths, the leak check must be performed
for each gas path separately.
Figure illustrates MicroCEM analysis enclosure (Left) wire connections to the Sample Handling box
Installation,” the analyzer is ready for operation.
Before operating the system, verify that the Leak Checks have been
performed and that the sample handling unit is performing correctly.
MicroCEM analysis enclosure On/Off switch is located inside the door on
the bottom right hand corner. Push switch to the “on” position to start
system.
The unit will immediately run thru a self diagnostic mode. This may take up
to 2 minutes. The user will know the system has passed all diagnostic test
and is “ready” upon the green LED (located above on/off switch) flashing. If
the green LED does not start to flash verify that proper power is connected
to the unit and restart. If AC/Heater fan is running but the green LED still
will not flash then call the factory immediately for help.
NOTE: After startup a warm-up time from 20 to 60 minutes (Depending
upon ambient temp) is required for accurate measuements.
Analyzer operation can be confirmed by the green LED light flashing. The
pocket pc can then be connected for viewing menus. Upon power up, the
analyzer will perform a self-test routine. The test will take approximately 6
minutes.
4.2 Analyzer Operation
4.2.1 User Interface
The µCEM User Interface runs on a Pocket-PC with Windows CE operating system. It
communicates with the µCEM via serial communication port. All input to the Pocket-PC
is done using a pointing device that comes with the Pocket-PC. The Pocket PC can be
plugged into two different ports. The first port is located on the front panel below the
on/off switch inside the front door. The second port is from the bottom of the uCEM via
J3 connector.
The pocket PC can be found behind the door behind the glass piece. Note that upon
shipment the pocket PC may be located in a separate box.
To connect the pocket PC to the: µCEM via the inside connection.
1. Open µCEM door.
2. Plug RS232 plug into adapter located on front panel
3. Plug power supply cable into 5V adapter
4. Turn Pocket PC on
5. In order to assure no other windows are open press the reset button. Reset
button is located on the back of the pocket PC.
6. Go to tools menu (Icon in upper left hand corner) and click on µCEMTS.
7. Unit will display data in 3 to 5 seconds. If unit does not show data in 3 to 5
seconds repeat procedure starting with number 5.
To connect the pocket PC to the: µCEM via the outside connection.
1. Plug the external COM cable into J3 circular connector on the bottom of the
uCEM.
2. Plug pocket pc RS232 plug into the J1 on the external COM cable.
3. Plug power supply cable into 5V plug on the COM cable.
4. Turn Pocket PC on.
5. In order to assure no other windows are open press the reset button. Reset
button is located on the back of the pocket PC.
6. Go to tools menu (Icon in upper left hand corner) and click on µCEMTS.
7. Unit will display data in 3 to 5 seconds. If unit does not show data in 3 to 5
seconds repeat procedure starting with number 5.
Note: The Pocket PC can by used on any MicroCEM TS analysis enclosure regardless
of the MicroCEM units IP address.
4.2.2 µCEM Main Window
The µCEM Main Window shown in Figure 4-1 provides the status of the three emissions
channels. The status includes the current reading (updated approximately every 2
seconds), the last 1-minute average, and the last 15-minute average. The status column
(Sts) indicates the status of the measurement and can be any of the values in the Table
4-1.
Lower left part of the µCEM screen contains three menus, from which all of the µCEM
user-interface functions can be accessed. There are three main menus: File, Tools and
Advanced, presented on Figures 4-2.1, 4-2.2, and 4-2.3.
File Menu: Provides General access to Connect, Log-in, Log Off features
Tools Menu: Provides access to basic µCEM Tools, like alarms and stream switching
Advanced Menu: Provides access to advanced µCEM Features, like Stream Settings
and User
Toolbar Buttons: Shortcuts to Alarms, µCEM Settings, µCEM Admin, Stream Switching
Administration
Tools Menu: Provides
access to all
functionality
Note: Exit will only be
available when
current user has
administrative
access
Figure 4-2.1 - µCEM File Menu
Toolbar Buttons: Shortcuts
to Alarms, µCEM
Settings, µCEM Admin.,
Data Logs and About
The µCEM Alarms dialog shows all the current alarms. A current alarm is one with an
Active status of 1 (active) or an Acknowledged state of 0 (not acknowledged).. To see
the historical Alarms for the last 3 months , the web based µCEM interface must be
used. If one or more alarms are current, the most recent of them will be displayed on
the main display. If more than one alarm is current “(more)” will be displayed after the
name of the most recent alarm on the main window to indicate that more than one alarm
is active. Horizontal scroll bar is be used to see Date and Time of the Alarms. Alarms
can be General and Stream-specific. By selecting the radio buttons on the bottom, user
can view different types of alarms.
Drag the edges of
the columns to
resize the columns
Use the scrollbar
to see the full set
of data
Figure 4-3. Pocket PC Alarms Screen
Alarms with a critical level will cause the System trouble output to become active when
the alarm is active. When all active critical alarms are acknowledged, the System
trouble output will become inactive.
Alarm Name Level Description Type
O2 Calibration
Failed
CO Calibration
Failed
NOx Calibration
Failed
The login dialog appears (Figure 4-4) when first requesting the µCEM Settings or µCEM
Admin. If a valid user name and password are entered, the user logging in will have
permission to use the µCEM Settings and/or the µCEM Administration (Refer to the User
Settings page of the µCEM Settings dialog). After logging in the first time, it is not
required again until the user logs out, or is logged out automatically because of a period
of inactivity (Refer to the Auto Logout page of the µCEM Administration dialog).
Figure 4-4 - µCEM Login
On-screen keyboard is
available at any time
by clicking on the
keyboard button.
Typically a dual stream system is in Automatic Stream Switching mode. That means that
it runs the timing schedule specified in User Settings Configuration file. If Automatic
switching is not desirable, the user may turn it off using Tools-> Stop Auto Switching
menu. In this case the system will remain on the current stream indefinitely. When
Automatic switching is needed again, user may turn it back on with Tools->Start Auto
Switching menu.
The same task can be accomplished remotely, by clicking Stop Auto Switching button on
the µCEM Real-Time Web page.
Note, that this option is sustained even if the system is rebooted.
The operator may also force a switch between the streams at any time whether the
system is in Auto-Switching mode or not. Tools menu has an option “Switch to
StreamName”, where StreamName is a user-specified name of the stream.
The same task can be accomplished remotely by clicking Switch to “StreamName”
button on the µCEM Real-Time Web page.
The µCEM Settings dialog is only available to users with µCEM Settings permission. If a
user is not currently logged in, the login dialog will be displayed. If the current user
doesn’t have µCEM Settings permission, µCEM will not allow Settings screen to appear.
When the µCEM Settings are invoked from the Advanced menu or the µCEM Settings
button, the µCEM Settings tabbed dialog is displayed. The Range page (tab) is
displayed initially.
4.3.1 µCEM Settings-Range
The Range Settings page is used to specify the range for the analog outputs. Setting
Range 2 to a value of 0 (zero) enables single range functionality and disables the
dual range function. For Dual Range applications do not set range 2 equal too or
higher than Range 1 or the system will not calibrate properly. Note that Range 1 can be
changed by the user but must be changed in the webrowser tools. See the Webrowser
user settings section.
The dual range setting will enable the analyzer software and diagnostics to perform two
separate performance curves for each range thus enhancing the measuring capabilities
of the analyzer. A dual range setting is desired for applications burning dual fuels or that
may display high dynamic reading between the low and high of the day. The analog
outputs will also support the dual range mode. When the emission is below the Range 2
value, the analog output will switch to Range 2 mode and the Range 2 value becomes
the full-scale value of the analog output. The range indication digital output will change
to the Range 2 state. When the emission is above the Range 2 value, the output
switches to Range 1 mode and the Range 1 value becomes the full-scale value of the
output. The range indication digital output will change to the Range 1 state.
The Auto-Calibration settings are set on the Auto-Calibration page of the µCEM settings.
If auto calibration is turned to the on position, then the user can select time and/or
frequency of the auto calibration in the Auto Calibration Frequency tab (4.3.3).
Note: Both manual and auto calibration need to be perform with the MicroCEM
enclosure door in the closed position. If the door is opened then critical detector
temperatures will vary which will cause a drift in the calibration. If the door is kept open
long enough for temps to be constant at their setpoints then an open door calibration is
acceptable. See section 4.7 “temp diagnostics”- for details on temperature setpoints.
4.3.3 µCEM Settings - Auto Calibration Time and Frequency
The Auto-Calibration Time and Frequency tab allows specifying time and frequency of
the auto-calibration. Time field requires military time format. The times are displayed in
Military time type.
The emission limits alarms can be set on the Limits page of the µCEM Settings. When a
measured emission exceeds its limit, the emission will have a limit-exceeded status.
This is indicated on the main display and on the Data-Logs display. It is also indicated in
the limit exceeded digital output.
The Calibration Gas values and Gas Bottle allocation may be set on the Calibration Gas
page of the µCEM Settings. This should be set whenever a Calibration Gas container is
replaced or upon Startup of the system.
Calibration Gas Values:
R1Mid: This is typically used for CGA audits and not for daily calibrations. The specific
calibration gas mid value (typically between 40% to 60% of range) is set in this space.
The MicroCEM will perform mid calibration on Range 1 on this gas but will not perform
any corrections. This box should typically be left blank. It is mostly used as a check.
R1Span: The specific calibration gas span value (typically between 80% to 100% of
range) is set in this space for Range 1. A Nox range of 0-100ppm would typically use a
gas bottle with 90ppm NOx balance N2.
R2Mid: This is typically used for CGA audits and not for daily calibrations. The specific
calibration gas mid value (typically between 40% to 60% of range) is set in this space.
The MicroCEM will perform a mid calibration on Range 2 on this gas but will not perform
any corrections. This box should typically be left blank. It is mostly used as a check.
R2Span: The space is allocated for dual range applications. If the MicroCEM range
setting is set for single range then the user will not be able to input any value into this
space. The specific calibration gas span value (typically between 80% to 100% of
range) is set in this space. A Nox range of 0-10ppm would typically use a gas bottle with
9ppm NOx balance N2.
Note that zero values do not have to be input into this page. For all zero calibrations the
user must assure that the calibration gas used does not have any levels of the
measurement gas in the cylinder. For example upon the analyzer zeroing O2, the bottle
used must have 0% O2 in the Bottle. Zeroing the O2 is typically performed by using the
NOx or CO Span gases.
Gas Bottle Allocation:
Gas 1, Gas 2 and Gas 3 are labels for the respective location of where the calibration
gas cylinders are physically located on the external fittings.
Off: Designates that no operation will be performed.
Zero: The MicroCEM will perform a zero calibration.
R1Span: MicroCEM will perform a Span calibration for Range 1.
R2Span: MicroCEM will perform a Span calibration for Range 2. Note that if a second
range is NOT chosen in the range settings menu then user will not be able to input any
value into this space. Range 2 should always be a lower value than range 1 if used.
R1Mid: MicroCEM will perform a Mid Calibration for Range 1.
R2Mid: MicroCEM will perform a Mid Calibration for Range 2. Note that if a second
range is NOT chosen in the range settings menu then user will not be able to input any
value into this space.
Maintenance mode may be selected for any of the emission types on the Maintenance
Mode page of the µCEM Settings.
Choosing maintenance mode will invoke an “M” flag” onto the data. Customer can
perform routine maintenance while in this setting
This mode is typically used when preventive maintenance is being performed. The M
flag signifies to the EPA that the values reported are not valid therefore should not be
applied to emissions reporting.
Upon completion of Maintenance the user must go back into this screen to turn the
Maintenance off. If not, the MicroCEM will continue to show the M flag in the data.
A dry-run Calibration may be initiated from the Manual Calibration page of the µCEM
Settings by pressing the Manual Calibrate All icon. A full zero and span calibration will
be run by the MicroCEM but the end result corrections of the calibration will not be
applied to the O2/Nox/CO measurement values. If desired a partial calibration may be
invoked for one or more of the emission types. While the manual calibration is in
process, a calibration progress dialog will be displayed as shown in Figure 4.24. When
the manual calibration is completed, the results are displayed in the Manual Calibration
Results dialog as shown in Figure 4.10. If the Local Calibration checkbox is checked,
the Local Calibration valve will be used during the calibration rather than the probe
Calibration valve.
Note that “Start Auto Cal now” will invoke a calibration and will apply new correction
factor to all measurement when done.
The Auto Calibration dialog is displayed whenever calibration is in process. It displays
the current emission values and the status of the calibration. The calibration may be
canceled before it completes by pressing the Cancel button.
Note: The title of this dialog will
read either “Auto Calibration” or
“Manual Calibration” to indicate
how the calibration process was
initiated.
The µCEM Administration dialog is only available to users with µCEM Administration
permission. If a user is not currently logged in, the login dialog will be displayed. If the
current user doesn’t have µCEM Administration permission, a message will be displayed
which reads “Permission denied”. When the µCEM Administration is invoked from the
Tools menu or the µCEM Administration button, the µCEM Administration tabbed dialog
is displayed. The User Settings page (tab) is displayed initially.
4.4.1 µCEM Administration-User Settings
The user settings page of the µCEM Administration dialog allows users to be added,
deleted or modified. Each user has a name, password, and permission settings. The
permission settings include Settings permission that allows access to the µCEM Settings
dialog, and Administrative permission that allows access to the µCEM Administration
dialog. The Settings permission also allows a user to access the µCEM remotely using
the web-based interface.
A µCEM Factory and User Settings files are available for use by µCEM technicians to
set parameters in the µCEM or a qualified customer technician.
µCEM Settings are separated into two files: Factory Settings and User Settings. Factory
Settings should be modified by a Rosemount technician only. Note: Some parameters in
this file, if set incorrectly, may cause permanent damage to hardware.
User Settings can be modified by a qualified customer technician. User settings are
accessible through the User Settings Web screen. See section 4.7 for details on
access. Settings files are formatted as a standard Windows INI files. File is organized
in sections (in square brackets). Configuration Parameters are presented in “Name =
Value” format. Comments start with semicolon. User Settings files has three sections
[General], [Stream 1] and [Stream 2].
The list of some settings is shown in Table 4.3 & Table 4.4.
Consult a Rosemount factory person for details.
Stream1Time Stream 1 processing time in minutes when auto switching
Stream2Time Stream 2 processing time in minutes when auto switching
TransitionTime Time to keep the B flag after the switch have occurred, in
seconds
Stream1Name Stream 1 Name to be shown on Pocket PC and Web pages
Stream2Name Stream 2 Name to be shown on Pocket PC and Web pages
CalibrationCurrentLoopOutputs Defines the behavior of Current Loops during Calibrations
1 - Hold the Last Good Value,
2 - Use the User-Specified Value
3 - Follow the Gases as is
CalibrationCurrentLoopOutputsUserValue Value in milliamps. Used when the previous parameter is
set to 2
ByPassCurrentLoopOutputs Defines the behavior of Current Loops during By-Pass
1 - Hold the Last Good Value
2 - Use the User-Specified Value
ByPassCurrentLoopOutputsUserValue Value in milliamps. Used when the previous parameter is
set to 2
AutoCalForcesSwitch Defines what to do, when the scheduled Auto-Calibration
time comes, but the system happens to process another
stream
1 - force a switch to the stream and run the Calibration
2 - wait until the stream is switching occures by itself and
run the Calibration
DigitalOutputsLogic Defines how to control Digital Outputs
DiluentCorrectionPercent Diluent Correction Percent used in calculations for the Stream
O2R1Range Range 1 Setting for O2 (Range 2 can be changed from the Pocket PC)
COR1Range Range 1 Setting for CO (Range 2 can be changed from the Pocket PC)
NOXR1Range Range 1 Setting for NOx (Range 2 can be changed from the Pocket
PC)
PostCalibrationDelay Number of seconds to keep the C(Calibration) flag after the Auto
Calibration process is over
R1O2ZeroDriftLimit O2 Allowed Zero Drift
Limit for Range 1.
R1COZeroDriftLimit CO Allowed Zero Drift
Limit for Range 1.
R1NOXZeroDriftLimit NOx Allowed Zero Drift
Limit for Range 1.
R1OSMidDriftLimit O2 Allowed Mid Drift Limit
for Range 1.
R1COMidDriftLimit CO Allowed Mid Drift
Limit for Range 1.
R1NOXMidDriftLimit NOx Allowed Mid Drift
Limit for Range 1.
R1O2SpanDriftLimit O2 Allowed Span Drift
Limit for Range 1.
R1COSpanDriftLimit CO Allowed Span Drift
Limit for Range 1.
R1NOXSpanDriftLimit NOx Allowed Span Drift
Limit for Range 1.
R2O2ZeroDriftLimit O2 Allowed Zero Drift
Limit for Range 2.
R2COZeroDriftLimit CO Allowed Zero Drift
Limit for Range 2.
R2NOXZeroDriftLimit NOx Allowed Zero Drift
Limit for Range 2.
If the drift exceeds the allowed amount
a drift alarm will occur, and the
readings on the channel will no longer
be valid until a successful calibration is
completed.
If the drift exceeds the allowed amount
a drift alarm will occur, and the
readings on the channel will no longer
be valid until a successful calibration is
completed.
The µCEM maintains a minimum of 3 months of history in three types of data log files.
The first type of log file is the measurement log, which contains emission measurements
(at 1 minute intervals), alarm indications and maintenance mode indications. The
second type of log file is the calibration log file, which contains information on each auto
calibration done. The third is the alarm log file, which records any improperly functioning
hardware. The data will be stored in flat, ASCII, CSV (comma-delineated) files. This file
format can be read directly by MS Excel and imported into many types of software
applications. The following parameters are factory set for each of the log file types.
4.6.1 Maximum Log File Size
This is how large a log file can get (in bytes) before it is closed and a new log file is
opened.
This is how many log files can be created. When the maximum number of log files is
reached, the oldest file is overwritten when new ones are created.
Emissions Log: 6
Calib Log: 6
Alarm Log: 6
4.6.3 Log File Name Format
The log file name uses the date that the file was created. It is of the format
TYYYYMMDD.CSV where T is the log file type (E=Emissions, C=Calibration and
A=Alarm), YYYY is the Year, MM is the month, and DD is the day of the month. For
example, the file name E20010329.csv contains emissions data and was created on
March 29, 2001.
The log file contains data in a flat, ASCII, CSV file. The following are the fields of the
file, in order of occurrence. The log file size will be about 42 bytes per entry. 3 months
of data logs will require about 5,443,200 bytes
(calibration failed or sensor in failed state)
CO CO parts per million 12
CO Limit CO Limit exceeded alarm, 0=inactive,
1=active
CO Status V=Valid, M=Maintenance Mode,
C=Calibration in process, I=Invalid
(calibration failed or sensor in failed state)
Nox NOx parts per million 15
NOx Limit NOx Limit exceeded alarm, 0=inactive,
1=active
NOx Status V=Valid, M=Maintenance Mode,
C=Calibration in process, I=Invalid
(calibration failed or sensor in failed state)
0
V
0
V
0
V
Table 4.7 –Measurement Log File Format
4.6.5 Calibration Log File Format
The log file contains data in a flat, ASCII, CSV file. The following are the fields of the
file, in order of occurrence. The log file size will be about 350 bytes per entry. 3 months
of data logs will require about 32000 bytes (based on Calibration performed every 24
hours).
Name Description Example
Date/Time Calibration
start
Gas 1 Time Time that Gas 1 started, Hours:Minutes:Seconds 10:25:30
Gas 2 Time Time That Gas 2 started, Hours:Minutes:Seconds 10:27:30
Gas 3 Time Time that Gas 3 started, Hours:Minutes:Seconds 10:28:30
Purge Time Time that the final purge started, Hours:Minutes:Seconds 10:30:30
Finish Time Time that the final purge finishes 10:31:00
O2 Expected Zero Expected percent O2 for Zero phase of calibration 0.0
O2 Measured Zero Measured percent O2 for Zero phase of calibration 0.0
O2 Zero Drift Percent drift of O2 zero calibration 0.0
O2 R1 Expected Mid
Span
O2 R1 Measured Mid Measured percent O2 for Range 1 Mid span phase of
Month-day-year Hours:Minutes:Seconds
Expected percent O2 for Range 1 Mid span phase of
calibration
Span calibration
O2 R1 Mid Drift Percent drift of O2 Range 1 mid calibration. 0.4
O2 R1 Expected Span
O2 R1 Measured Span
O2 R1 Span Drift Percent drift of O2 Range 1 span calibration 0.4
O2 R2 Expected Mid
Span
O2 R2 Measured Mid
Span
O2 R2 Mid Drift Percent drift of O2 Range 2 mid calibration. 0.4
O2 R2 Expected Span
O2 R2 Measured Span
O2 R2 Span Drift Percent drift of O2 Range 2 Span calibration 0.4
CO Expected Zero Expected ppm CO for zero phase of calibration 1
CO Measured Zero Measured ppm CO for zero phase of calibration 0
CO Zero Drift Percent drift of CO zero calibration -0.3
CO Expected R1 Mid
Span
CO Measured R1 Mid
Span
CO R1 Mid Span Drift Percent drift of CO Range 1 mid span calibration 0.3
CO R1 Expected Span Expected ppm CO for Range 1 span phase of calibration 45
CO R1 Measured Span Measured ppm CO for Range 1 span phase of calibration 45
CO R1 Span Drift Percent drift of CO Range 1 span calibration 0
CO Expected R2 Mid
Span
CO Measured R2 Mid
Span
CO R2 Mid Span Drift Percent drift of CO Range 2 mid span calibration 0.3
CO R2 Expected Span Expected ppm CO for Range 2 span phase of calibration 45
CO R2 Measured Span Measured ppm CO for Range 2 span phase of calibration 45
CO R2 Span Drift Percent drift of CO Range 2 span calibration 0
NOx Expected Zero Measured ppm NOx for zero phase of calibration 15
NOx Measured Zero Expected ppm NOx for zero phase of calibration 15
NOx Zero Drift Percent drift of NOx zero calibration 0
NOx Expected R1 Mid
Span
NOx Measured R1 Mid
Span
NOx R1 Mid Span Drift Percent drift of NOx Range 1 mid span calibration 0
NOx Expected R1 span
NOx Measured R1 span
Expected percent O2 for Range 1 Span phase of
calibration
Measured percent O2 for Range 1 Span phase of
calibration
Expected percent O2 for Range 2 Mid span phase of
calibration
Measured percent O2 for Range 2 Mid span phase of
calibration
Expected percent O2 for Range 2 Span phase of
calibration
Measured percent O2 for Range 2 Span phase of
calibration
Expected ppm CO for Range 1 mid span phase of
calibration
Measured ppm CO for Range 1 mid span phase of
calibration
Expected ppm CO for Range 2 mid span phase of
calibration
Measured ppm CO for Range 2 mid span phase of
calibration
Measured ppm NOx for Range 1 mid span phase of
calibration
Measured ppm NOx for Range 1 mid span phase of
calibration
Measured ppm NOx for Range 1 span phase of
calibration
Measured ppm NOx for Range 1 span phase of
calibration
NOx R2 Span Drift Percent drift of NOx Range 1 span calibration 0
NOx Expected R2 Mid
Span
NOx Measured R2 Mid
Span
NOx R2 Mid Span Drift Percent drift of NOx Range 2 mid span calibration 0
NOx Expected R2 span
NOx Measured R2 span
NOx R2 Span Drift Percent drift of NOx Range 2 span calibration 0
Table 4.8 – Calibration Log File Format
Measured ppm NOx for Range 2 mid span phase of
calibration
Measured ppm NOx for Range 2 mid span phase of
calibration
Measured ppm NOx for Range 2 span phase of
calibration
Measured ppm NOx for Range 2 span phase of
calibration
30
30
59
59
4.6.6 Alarm Log File Format
The log file contains data in a flat, ASCII, CSV file. The following are the fields of the
file, in order of occurrence. The days or months maintained in the Alarm Log depends
on how often trouble conditions are recorded. If alarms rarely occur, there is enough
space for many years of alarm logs to be recorded.
4.6.7 Accessing the Real-Time ACSII Data String via Ethernet TCP/IP
(DAS)
Remote Real-time data acquisition from the uCEM is done through the TCP/IP enabled
network via the HTTP (Web transport) protocol.
Acquisition software has to request the page form the Web Server running on the uCEM
unit with the desired frequency (real update time is 1 sec).
URL for the real time data is defined as such: http://[uCEM IP]/fetchData.asp
For example: http://127.0.0.1/ fetchData.asp
In response Web Server will return the comma-delimited string that contains current
analyzer data. Note: the response is a plain text not the HTML document.
If the actual analyzer software is running, the response data will be formatted as such:
The result is a single string of data.
DateTime is formatted as such: Month-Day-Year4Digits HoursMilitary:Minutes:Seconds
Example: 02-05-2002 14:58:53
All the current and average gas values are the floating-point numbers and may contain a
sign. Certain rules are defined for the current and average gas values:
If there is a “#” sign in this field – data for this field are not valid. That usually means
there is no data available or the data cannot be converted to the string representation (due
for example to faulty Calibration).
If the value field shows –
denotes that the system haven’t yet initialized the data. That usually happens when uCEM
starts up and 1 minute or 15-minute averages are not yet available (calculated). Note that
regardless of the status, values show the current measured data from the analyzer. “Magic
number” means that the data (usually 15 minute averages) have not been yet calculated.
ExtProcess1 and ExtProcess2 are the values of the Analog Inputs (Mega Watts and Fuel
Flow usually).
DigInput1, DigInput2, DigInput3- show the state of the digital inputs and can take a
value of either 1(On) or 0(Off). DigInput1 is usually interpreted as ProcessOn. DigInput2
– as FlameOn. DigInput3 – as Shutdown.
CurAlarms values show the current state of the emissions limit alarm associated with the
gas. It’s an integer number that equals to 1 when emission limit for the gas is exceeded
and stays 0 if the gas doesn’t gave associated alarm active.
All the Status values are single-character values. Status is defined as such:
V – Valid
I – Invalid
“-555.00” (negative 555.00). That is a “magic number” that
M – Maintenance
C – Calibration
P – Process Off
O – uCEM Off
AlarmsString – is a string data that describes the current Alarms situation with the
uCEM module. It is separated from the rest of the data by a semi-colon. AlarmsString
usually is not parsed and used for the presentation purposes only. Example: “1,NOx
Emission Failed. 13 More ...”
12.45,I,-555.00,I,5.0,3.76,0,1,0;1,NOx Emission Failed. 13 More ...
This string means that the sample was taken February 5 2002 at 2:58PM, O2 values were
all Valid except the 15 Minutes average was not yet calculated, CO process was Off - the
data were not available. NOX data were Invalid and the 15 Minutes average was not yet
calculated. Mega Watts value read from the input was 5.0, Fuel Flow – 3.76
DigInput1(ProcessOn) is set to 0(Off), DigInput2(FlameOn) is set to 1 (On),
DigInput3(Shutdown) is set to 0 (Off)
There were also 13 alarms active, NOx Emission Failed being the most recent one.
If the uCEM analyzer is not currently running the return string will be:
“uCEM is not running. No data Available.”
The Pocket PC Web Browser menu can be accessed via the pocket pc main menu. In
the top upper left hand corner of the menu the name of the unit will be displayed (ucem
XXXX). Point on this name. A drop down menu will appear. Point on Internet Explorer.
A sign on page will then be displayed. User name and password will be the identical as
the normal names used on the administration settings. Very importantly the Web
Browser function allows the user to access all data (calibrations, alarms, emission data
logs, diagnostics) internally stored in the MicroCEM.
The Web Browser will show the following screens/options for the user. Note that these
screens are updated once every 10 seconds unless the refresh bottom is pressed:
Real Time: This screen is identical to the main menu screen normally shown on the
pocket pc.
Emissions: This screen will enable the user to view all internal emission data logs
stored in the MicroCEM. User can choose between 1, 15, 1hr or 24 hour periods. A
designated time frame or most recent data can be choses. The report generator will
display data in a chart type format showing each gas value and associated time along
with data flags. The function is very helpful in very historical data or performing trouble
shooting.
Alarms: This screen will allow the user to display all alarms and time frames. User may
choose time frames or most recent alarms.
Cal: Display of all calibrations with results can be viewed from this page.
User Config: This file contains user selectable files that are typically input at startup
and never changed. See section 4.4 for details on descriptions. Note that reboot of the
MicroCEM may be necessary for system to accept changed for several items in this file.
Factory Config: Do not access this file unless a certified Rosemount technician is
present. Changes to this file may adversely affect or destroy the unit. Changes
made to this file without the written consent of the MicroCEM Product Manager will void
the warranty.
Download: User can easily download all data log files (Emissions, Alarms, Calibrations)
stored in the MicroCEM. This is typically used when user is accessing the MicroCEM via
a separate laptop or tabletop computer. See next section.
Temp Diag: Temperature diagnostics is a very important tool for diagnosing existing
problems or potential issues/problems with the MicroCEM.
The following parameters will be shown: Temperature Parameter, Temp Setting, Actual
Temp and Integral %.
*Zone Temperature: Zone temperature is typically set to 47 degrees C. This is the
temperature of the MicroCEM taken from the detector section thermocouple that is
located behind the pressure gauge. This thermocouple is always used for systems with
no CO. For systems with CO a thermocouple is located on the CO assy detector. The
MicroCEM will typically control temperature to within +/- 1 degree C. Depending upon
the outside ambient temperature the % on time can be from 0 to 100% on. If a negative
value is shown in the integral then cooling is in process. Variations greater than 1
degree C will lead to gas measurement drifting.
*PMD Temp: This is the Temperature of the chassis of the MicroCEM. Thermocouple is
located in the PMD detector. Temperature is typically within 2 degrees C. of the zone. If
the temperature drifts greater than this. Upon first turn on of the MicroCEM this temp
can be monitored. Once this temp is within 2 degrees C of the Zone then the unit is
ready for accurate measurements. Temperature above the 2 degree variance of the
zone may show than the AC/Heater unit fan may have failed or a possible defective
thermocouple.
*PDD Temp: For systems with NOx a cylindrical NOx detector assy is located in the
detector section. Internal to the detector assy a small peltier device is operating and
must operate at 0 degrees C. The temperature should never deviate +/- .05 degrees C
from the setpoint of zero or the NOx readings may drift. Integral will typically run
between 40 to 70%.
*PMT Temp: This temperature is for the detector assy heater core. Setting is set to 52
degrees C. Temperature should not drift more than 0.2 degrees C or NOx drift may
occur. Excessive temperature variation may be caused by either poor zone temperature
control or a faulty heater.
*Conv Temp: This temperature is for the NOx converter assembly. Temperature setting
is 330 degrees C. Temp should not vary more than 1 degrees C. or NOx
measurements will drift. A faulty heater will cause temp variations.
Note that when the enclosure door is opened that all of the above temperature setting
may be affected and will take a short about of time to react and control to the desired
temperatures.
Note: The Real-time, Config and Download are included in the navigation menu but these pages
are intended for remote desktop use. As an enhancement these items could be hidden if the
pages are browsed from a Windows CE version of Internet Explorer.
Alarms and
Calibration data may
also be viewed.
A Date is shown for
1 min or 15 minute
averages. A date
range is shown for 1
hour or greater
averages.
4.8 Viewing µCEM Data with an external PC Web Browser
The MicroCEM internal log files may be accessed using a user PC or laptop with a web
browser that has access to the µCEM over a LAN, serial port connection (PPP) or
Dialup Connection (RAS). The µCEM has Window CE Web Server installed and
provides a Web-based interface to select and download the Data-Log files. The
downloaded Data-Log files will be in a CSV (comma delineated ASCII) format. The log
files may also be viewed as a web page in a tabular format.
1. Connect user PC or laptop to the MicroCEM via Ethernet LAN circular connector
located on J4 connector. The Ethernet cable can then be routed to the users
Ethernet hub where as many PC’s as desired can access the MicroCEM Web
Browser. Customer may also choose to connect the cable directly to the
Ethernet port located on the MicroCEMs PC104 PCB which is internal to the
MicroCEM. Note that a crossover type Ethernet cable must be used if a hub is
not utilized.
2. The user PC or laptop must have the same IP address as the MicroCEM or the
MicroCEM IP address can be changed to the users desired IP address.
Standard IP address of the MicroCEM is: 192.168.1.92
3. Once the IP addresses are matched the user can simply open internet explorer
on their computer and type is the MicroCEMs IP address.
4. Once entered a user ID and password must be entered. These are the identical
user ID and password as input into the administration menus.
5. Once entered the user can then access all pages as specified in section 4.7.
4.8.1 Real-Time Page
The Real-Time page provides a real-time display of the emission values and emission
statuses. The display is refreshed every 10 seconds.
The Emissions Page can be used to view emission history in a tabular web-page format.
This page is used as part of the µCEM User interface as well as by a remote user
(probably from a desktop computer).
If Most Recent is
selected, the month day
and hour do not need to
be selected.
Select the ending hour
to view (applicable only
to 1 minute averages)
The Emission Data-Logs table is displayed (as shown in figure 4.19) after selecting the
Date and Average Period and pressing the Display button. If desired a bookmark or
shortcut may be made to the page displaying the table. In the future, the same table can
be displayed by selecting this bookmark. If Most Recent Data was selected, the bookmarked page will always display Most Recent Data. If a specific date was specified, the
book-marked page will always display the same date.
The download page of the µCEM allows the selection and download of the three types of
Data-Logs. To quickly download recent data, a “Download Most Recent Emissions
Data” selection is provided. For more control over the date range, a “Download
Emissions by Date Range” selection is available. Once the selection is made, press the
Download button to start the HTTP download. The µCEM will create a temporary file
that contains the selected data. Due to memory limitations there is a limit to the number
of files that can be downloaded simultaneously. If this limit is exceeded, a message will
be displayed that reads “The simultaneous download limit has been reached, please try
again later”.
The µCEM Data may be viewed with MS Excel. CSV comma delineated files can be
opened either from the Web browser Session or after the file(s) are saved onto a
workstation. The files may then be opened directly with Excel. These files later can be
converted and saved in MS Excel native format to enable charting and other secondary
analysis functions.
This equipment should not be adjusted or repaired by
anyone except properly qualified service personnel.
WARNING: PARTS INTEGRITY
Tampering with or unauthorized substitution of components
may adversely affect safety of this product. Use only factory-approved
components for repair.
WARNING: ELECTRICAL SHOCK HAZARD
Disconnect power to the module(s) prior to replacing
components.
The uCEM Analyzer Module requires very little maintenance during normal
operation.
5.1 Overview
The uCEM Analyzer Module requires very little maintenance during normal
operation.
Occasionally, the detector's reaction chamber and sapphire window may
require cleaning, refer to Section 5.7.
White crystal deposits on the windows of the reaction chamber and pl ugging
of capillaries and vent are usually due to sample contaminates such as
ammonia reacting with the high ozone levels and NO components. To
eliminate the contaminates, the sampling system should be reworked or a
preventive maintenance program developed (if dropout is not excessive).
Another source of crystalline formation is contaminated air.
Several components may require replacement. These are discussed in the
following sections.
Refer to Figure 5-1, Item 97, and Figure 5-3. To replace the converter or
sensor, disconnect the two pneumatic tubes and two electrical connections.
Unlace the heater blanket, and remove the converter. Reassemble in
reverse order, ensuring that the converter is oriented with the glass cloth at
the bottom and the sensor is oriented correctly inside the heater jacket.
ASSEMBLED SIDE VI EW
Sensor
Heater
Jacket
655228
Converter
Tube 655227
Glass
Cloth
Figure 5-3. Converter Assembly
5.3 Ozonator
Refer to Figure 5-1, item 98.To replace the ozonator, remove the gas
fittings, the two large straps and all tie-wraps, and disconnect the one
electrical connection. Reassemble in reverse order.
5.4 Personality Modules
There are seven different personality modules. Depending on your unit, you
may have three, four five, six, or seven modules installed. These
personality modules are installed on a custom backplane. see figure 5-4 for
more information.
Tag each cable and its location before disconnecting any wiring. This helps
in re-assembly
Wrap with
aluminum foil
Sensor
655282
To remove any on the personality modules. Remove cables from the
module to be removed, there are two screws at the bottom of each module.
You will have to loosen each screw before you can remove the personality
module.
Disconnect the stainless steel tubing lines at the Gyrolok fittings. Remove
the (4) nuts holding the Detector Assembly to the chassis. Disconnect the
plug from connector J1 on the Signal Board and remove the assembly from
the chassis.
Note: Heatsink Compound. Care should be taken to avoid getting
heatsink compound on optical surfaces. If this substance is removed
during the disassembly process, a zinc-oxide-filled, silicone grease
(e.g., Dow Corning 340 or EG&G Wakefield Engineering's Series 120
Thermal Joint Compound) be reapplied in the re-assembly of this
component.
Although the heater and thermostat can be removed to facilitate handling,
contact with the white heatsink compound can be minimized by leaving
these items in place. Remove the (2) screws holding the top plate of the
Detector , and move the plate along the wires and away from the Detector .
Remove the (2) screws holding the tube assembly in place. Hold the tubing
with one hand while inverting the Detector Housing with the other, allowing
the Reaction Chamber O-ring and window to be removed from below.
REACTION CHAMBER INSTALLATION:
To reinstall, hold the housing in the inverted position while sliding the
Reaction Chamber O-ring and window into position and the tubing into the
slot in the housing. Hold the Reaction Chamber in place while rotating the
housing upright. Replace the hold-down screws.
Note: Component Positioning. The procedure described above is for
the purpose of maintaining the relative positions of windows and
O-ring to the Reaction Chamber during installation.
Replace the top cap and screws. Reverse the removal procedure to
reinstall the Detector Assembly into the Analyzer Module.
PHOTODIODE REMOVAL:
Remove the Detector Assembly as described above. Invert the housing to
access the mounting bracket. Remove the (3) screws and shoulder
washers from the bracket. Remove the bracket, insulating disk and bottom
plate as a unit to minimize the spread of the heatsink compound.
Remove the (2) screws holding the lower section of the Detector Housing,
then slide the section along the cable and remove.
Remove the (2) screws holding the socket, thermistor and photodiode in
place, being careful not to lose the washers that are used as shims.
Grasp the socket and photodiode base while slowly rotating to separate the
photodiode from the housing. Some friction will be felt as an O-ring is used
around the photodiode as a seal.
PHOTODIODE INSTALLATION:
To replace the photodiode, carefully remove the diode from the green
socket, and replace with a new one. Before mounting the new diode, the
top cap of the enclosure should be temporarily removed and the (2) screws
holding the Reaction Chamber loosened about two turns. This allows air
which is trapped between the O-ring seals to escape when the diode is
inserted. It also maintains the position of the O-ring and window in the
upper compartment.
The new photodiode should be slowly inserted into the housing while
gradually rotating the body. This allows the O-ring to properly seat.
Continue replacing screws, washers, thermistors, etc., with the thicker shim
(washer) on the opposite side of the socket from the thermistor.
Replace the lower section of the housing, then the bottom cover, insulator
and bracket with the shoulder washers and screws.
Re-tighten the screws in the Reaction Chamber (upper section). Replace
the top cap and its screws.
To reinstall in the Analyzer Module, reverse the procedure for removal as
indicated above.
The CPU is an Embedded Pentium-type AT Computer in 5.75” x 8” form factor. The
peripherals integrated on board are: SVGA, 4 serial ports and one parallel port, Fast
Ethernet ctrl., IDE, Keyboard, Mouse, 2 USB. The module is built around the Intel
Tillamook processor and is equipped with 64MB SDRAM. The module also integrates
one socket for SSD that performs like an HDD unit and can be used to store the
operating system, the user’s programs and the data files. Other peripherals available on
board are the Floppy disk controller, and the parallel port. The CPU is depicted in Figure
5-7.
Figure 5-7. CPU
5.6.1.1 Features
Architecture: PC/AT Compatible
Dimensions: 5.75” x 8”
Processor: Intel Tillamook processor - 266MHz
Memory: 64 MB SDRAM
Ram/Rom disk: 1 x 32 pin socket (max. 288MB)
Operating System: WinNT
BIOS: Standard with embedded extensions
Interfaces: IDE ctrl
2 USB ports
4 RS232 serial ports (one can be 485)
Parallel port (bi-directional EPP-ECP)
Keyboard PS/2
Mouse PS/2
Bus: AT bus according to PC/104 spec.
Power Supply: AT/ATX
Connectors: COM1-4, SVGA, USB 1 and 2, PS/2 Mouse/Keyboard,
ATX Power, Parallel, IDE, Floppy, and Fast Ethernet
5.6.1.2 EMBEDDED ENHANCED BIOS:
- Award, 256KB Flash Bios.The Bios is immediately activated when you first turn on the
system. The Bios reads system configuratio information in CMOS RAM and begins the
process of checking out the system.
5.6.2 Analog/Digital I/O Board
The Analog/Digital IO (ADIO) Board is an off-the-shelf, complete data acquisition system
in a compact PC/104 packaging. The analog section contains 32 input channels,
multiplexed A/D converter with 16 bit resolution and 10uS conversion time. Input ranges
are +/-5v or +/- 10V. It also includes on-board DMA support. The analog output section
includes two 12 bit D/A converters. Both sections features simplified calibration using on
board programmable digital potentiometer. The digital I/O section provides 24 digital I/O
lines, which feature high current TTL drivers. The board requires only +5V from the
system power supply and generates its own +/-15V analog supplies on board. The board
operates over the Extended Temperatures range of -25 to +85C. Figure 3 depicts the
ADIO board and Figure 5-9 depicts the ADIO block diagram.
The ADIO board features automatic calibration of both analog inputs and outputs for
enhanced accuracy and reliability. The potentiometers, which are subject to tampering
and vibration, have been eliminated. Instead, all A/D calibration adjustments are
performed using an octal 8-bit DAC. The DAC values are stored in an EEPROM and are
recalled automatically on power up. The board includes three precision voltage
references for negative full scale, zero, and positive full-scale. A calibration utility
program provided with the board allows you to recalibrate the board anytime, in both
unipolar and bipolar modes, and store the new settings in EEPROM.
Autocalibration applies to the 4 D/A channels as well. The full-scale D/A range is
selected with a jumper block. The analog outputs are fed back to the A/D converter so
they can be calibrated without user intervention. Again, calibration settings are stored in
EEPROM and automatically recalled on power-up.
5.6.2.2 Analog Inputs
The ADIO board provides split configuration capability, with more total input channels
than any other PC/104 analog I/O board. The board can be user-configured in any of
three ways:
A programmable gain amplifier, programmable unipolar/bipolar range, and
programmable 5V/10V full-scale range combine to give the ADIO board a total of 10
different possible analog input ranges. All range settings are controlled in software for
maximum flexibility.
5.6.2.4 Enhanced Trigger and Sampling Control Signals
The ADIO board has an extra A/D trigger and sample control signals in the design.
Seven auxiliary digital I/O lines on the analog I/O connector provide a sample/hold
output signal, A/D trigger in and out lines (to enable synchronization of multiple boards)
and external A/D clocking.
5.6.2.5 Analog Outputs
The ADIO board contains 4 12-bit analog outputs with autocalibration capability. Up to
5mA of output current per channel can be drawn from all channels simultaneously. Both
unipolar and bipolar output ranges are supported with jumper configuration. And on
power up, all outputs are reset to 0V automatically.
Mode
Full-Scale
Unipolar10V0-10V2.44mV
Unipolar5V0-5V1.22mV
Bipolar10V±10V4.88mV
Bipolar5V±5V2.44mV
Output
Range
Resolution
5.6.2.6 FIFO and 16-Bit Bus Interface
An on-board 1024-byte FIFO enables the ADIO board to work with Windows 95 and NT
by dramatically reducing the interrupt overhead. Each interrupt transfers 256 2-byte
samples, or half the buffer, so the interrupt rate is 1/256 the sample rate. FIFO operation
can be disabled at slow sample rates, so there is no lag time between sampling and data
availability. The 16-bit interface further reduces software overhead by enabling all 16
A/D bits to be read in a single instruction, instead of requiring 2 8-bit read operations.
The net result of this streamlined design is that the ADIO board supports gap-free A/D
sampling at rates up to 200,000 samples per second, twice as fast as our previous
boards.
5.6.2.7 Specifications
Analog Inputs
Number of inputs
A/D resolution16 bits (1/65,536 of full scale)
Nonlinearity±3LSB, no missing codes
Conversion rate200,000 samples/sec.max
On-board FIFO1K x 8(512 16-bit samples)
CalibrationAutomatic;values stored in EEPROM
Analog Outputs
Number of outputs4
D/A resolution12 bits (1/4096 of full scale)
Output ranges±5, ±10, 0-5, 0-10
Output current±5mA max per channel
Settling time6µS max to 0.01%
Relative accuracy±1 LSB
Nonlinearity±1 LSB, monotonic
ResetAll channels reset to OV
Calibration
Digital I/O
Main I/O24 programmable I/O
Input current±1µA max
32 single-ended, 16 differential, or
16 SE + 8 DI; user selectable