Short Form Manual
HASXEE-SFM-HS 07/2010
Gas Analyzer Series
Short Form Manual
www.EmersonProcess.com
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specications. The following instructions MUST be adhered to and integrated into your safety program when installing, using and maintaining Emerson Process Management (Rosemount Analytical) products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson Process
Management (Rosemount Analytical) representative for clarication.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install your equipment as specied in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualied personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualied people use replacement parts specied by Emerson Process Management (Rosemount Analytical). Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in re, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualied persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
3rd edition 2010-07
Emerson Process Management GmbH & Co. OHG
Industriestrasse 1 D-63594 Hasselroth Germany T +49 (0) 6055 884-0 F +49 (0) 6055 884-209 Internet: www.EmersonProcess.com
Short Form Manual
HASXEE-SFM-HS 04/2010
X-STREAM XE
Table of ConTenTs
Introduction S-1
Denitions S-1
Terms used in this instruction manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-2
Symbols used on and inside the unit Symbols used in this manual
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-3
Safety Instructions S-5
Intended Use Statement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-5
General safety notice / Residual risk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-5
Authorized personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-6
Additional literature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-6
Operating and maintaining this unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-7
Chapter 1 Technical description 1-1
1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.1.1 The front panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2 Comparison of the various X-STREAM analyzer model . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3 X-STREAM XEGC: ½ 19 inch table-top unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.4 X-STREAM XEGP: 19 inch table-top or rackmount design . . . . . . . . . . . . . . . . . . . . . . .1-6
1.5 X-STREAM XEF/XDF: Field housing with single or dual compartment . . . . . . . . . . . . . .1-8
1.5.1 Field housings for installation in hazardous areas (Ex-zones) . . . . . . . . . . . . . . . . . . 1-11
Table of contents
TOC
Chapter 2 Installation 2-1
2.1 Scope of Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
2.2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.3 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3.1 Common technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3.2 Specic technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.4 Gas conditioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
2.5 Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.6 Detailed Instructions for Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
2.6.1 X-STREAM XEGC, X-STREAM XEGP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
2.6.2 X-STREAM XEF, X-STREAM XDF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-18
2.7 Notes on Wiring Signal Inputs and Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-30
2.7.1 Electrical shielding of cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-30
2.7.2 Wiring inductive loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-33
2.7.3 Driving high-current loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-33
2.7.4 Driving multiple loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-34
Emerson Process Management GmbH & Co. OHG TOC-1
Short Form Manual
X-STREAM XE
Table of Contents
HASXEE-SFM-HS
04/2010
Chapter 3 Startup 3-1
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
3.2 Performing a Leak Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
3.3 Symbols and typographical conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
3.4 Front panel elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3.4.1 Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3.4.2 Status Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3.4.3 Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
3.5 Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
3.5.1 Access levels & codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
3.5.2 Special messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
3.6 Powering up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.6.1 Boot sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
3.6.2 Measurement display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
3.7 Language settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
3.8 Checking the settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
3.8.1 Installed options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
3.8.2 Conguring the display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16
3.8.3 Calibration setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17
3.8.4 Setting the analog outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-19
3.8.5 Setting concentration alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-24
3.8.6 Backup the settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-25
3.9 Perform a calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-27
3.9.1 Manual Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-28
3.9.2 Manual Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-29
Appendix A-1
A.1 Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.2 Assignment of Terminals and Sockets. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-15
A.2.1 Tabletop & Rack mount analyzers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-15
A.2.2 Field Housings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-16
Emerson Process Management GmbH & Co. OHGTOC-2
Short Form Manual
HASXEE-SFM-HS 04/2010
X-STREAM XE
Index of fIgures
Fig. 1-1: X-STREAM front panel (here the X-STREAM XEGP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Fig. 1-2: X-STREAM XEGC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Fig. 1-3: X-STREAM XEGP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Fig. 1-4: Optional screw-type terminal adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Fig. 1-5: X-STREAM XEF / XDF- Front views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Fig. 1-6: X-STREAM XEF - right side and bottom view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Fig. 1-7: X-STREAM XEF - power supply and signal terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Fig. 2-1: X-STREAM XE Analyzers - scope of supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Fig. 2-2: Labelling of gas connectors (example) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Fig. 2-3: Installation in bypass mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Fig. 2-4: X-STREAM XEGC - table top version rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Fig. 2-5: X-STREAM XEGP - rear panel, with optional terminal adapters and front side brackets. . . . . . . . . . 2-11
Fig. 2-6: X-STREAM XEGC - dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Fig. 2-7: X-STREAM XEGP - dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Fig. 2-8: Socket X1 - Analog & Digital Outputs 1-4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Fig. 2-9: Plug X2 - Modbus interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Fig. 2-10: Sockets X4.1 and X4.2 - Pin conguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Fig. 2-11: Plug X5 - analog inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Fig. 2-12: X-STREAM XEF - dimensions for installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Fig. 2-13: X-STREAM XDF - dimensions for installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Fig. 2-14: X-STREAM XE Field housing - Arrangement of terminals, cable glands and gas connectors . . . . . . 2-20
Fig. 2-15: Terminal block X1 - Analog signals and relay outputs 1-4 (XSTA) . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
Fig. 2-16: Terminal block X1 - Modbus interface (XSTA) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Fig. 2-17: X-STREAM XEF - Ethernet connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
Fig. 2-18: Terminals blocks X4 for digital inputs and outputs (XSTD). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Fig. 2-19: Terminal block X5 - Analog input signals (XSTI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
Fig. 2-20: Power supply connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
Fig. 2-21: Shielded signal cable, shielding connected at both ends. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
Fig. 2-22: Shielded signal cable, shielding connected at one end. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
Fig. 2-23: Signal cable with double shielding, shieldings connected at alternate ends. . . . . . . . . . . . . . . . . . . . 2-31
Fig. 2-24: Shield connector terminal with cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-32
Fig. 2-25: Suppressor diode for inductive loads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
Fig. 2-26: Driving high-current loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
Fig. 2-27: Loads in series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
Fig. 2-28: Loads in parallel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
Table of contents
TOC
Fig. 3-1: Leak Testing with U-turn Manometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Fig. 3-2: X-STREAM XE front panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Fig. 3-3: Arrangement of concentration thresholds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Emerson Process Management GmbH & Co. OHG TOC-3
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
04/2010
Emerson Process Management GmbH & Co. OHGTOC-4
Short Form Manual
HASXEE-SFM-HS 04/2010
X-STREAM XE

INTRODUCTION

This short form manual provides information on installation and startup of the X-STREAM® XE series gas analyzers. For comprehensive instructions see the X-STREAM XE series instruction manual.
This manual covers several X-STREAM analyzer models and so may contain information about congurations and/or options not applicable to your analyzer.
The installation and operation of units for use in explosive (hazardous) environments is not covered in this manual.
Analyzers intended to be used in such environments are supplied with further instruction manuals, which should be consulted in addition to this.

DEFINITIONS

The following denitions apply to WARNINGS, CAUTIONS and NOTES found throughout this publication.
HIGHLIGHTS AN OPERATION OR MAINTENANCE PROCEDURE,
PRACTICE, CONDITION, STATEMENT, ETC.
If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Safety Instructions
S
HIGHLIGHTS AN OPERATION OR MAINTENANCE PROCEDURE,
PRACTICE, CONDITION, STATEMENT, ETC.
If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE!
Highlights an essential operating procedure, condition or statement.
Emerson Process Management GmbH & Co. OHG S-1
X-STREAM XE

TERMS USED IN THIS INSTRUCTION MANUAL

Short Form Manual
HASXEE-SFM-HS
04/2010
Explosive Gas(es)
Flammable Gases and gas mixtures in a mix­ture with air within the explosive limits.
Flammable Gas(es)
Gases and gas mixtures are assigned to be ammable if they might become ignitable when in a mixture with air.
Infallible Containment
This term is derived from the standards of explosion protection especially from the re­quirements for pressurized housings: thus an infallible containment can be characterized by no intended leakage out of the gas paths enabling gas to enter the inner compartment of the analyzer housing.
Intrinsically Safe Cell (IS Cell)
Cells supplied with an intrinsically safe power signal, approved by a Test Institute, to operate with explosive gases.
The design ensures the IS cells remains safe even in case of failure and explosive gases are not ignited.
Protection Class IP66 / NEMA 4X
Both terms are used to specify conditions for equipment to be installed outdoor.
IP stands for Ingress Protection, the rst num­ber species protection against solid objects (6. = dust tight) while the second number species the degree of protection against liquids (.6 = heavy seas).
NEMA stands for National Electrical Manuf­acturers Association. 4X species a degree of protection to personnel against incidental contact with the enclosed equipment; to pro­vide a degree of protection against falling dirt, rain, sleet, snow, windblown dust, splashing water, and hose-directed water; and that will be undamaged by the external formation of ice on the enclosure
Upper Explosion Limit (UEL) Volume ratio of ammable gas in air above
which an explosive gas atmosphere will not be formed: the mixture of gas and air is too rich in fuel (decient in oxygen) to burn.
Lower Explosion Limit (LEL)
Volume ratio of ammable gas in air below which an explosive gas atmosphere will not be formed: the mixture of gas and air lacks sufcient fuel (gas) to burn.
NAMUR
NAMUR is an international user association of automation technology in process industries. This organisation has issued experience re­ports and working documents, called recom­mendations (NE) and worksheets (NA).
Emerson Process Management GmbH & Co. OHGS-2
Short Form Manual
HASXEE-SFM-HS 04/2010
Safety Instructions
X-STREAM XE

SYMBOLS USED ON AND INSIDE THE UNIT

Wherever one or more of the following symbols appear on or inside the instrument, be careful and read the instructions given in the accompanying manuals!
Strictly observe the given warnings, instructions and information to minimize hazards!
This symbol at the instrument ... ... means
dangerous voltages may be accessible. Remo­ving covers is permitted only, if the instrument is disconnected from power - and even in this case by qualied personnel only!
hot surfaces may be accessible. Removing covers by qualied personnel is permitted only, if the instrument is disconnected from power. Nevertheless several surfaces may remain hot for a limited time.
more detailled information available: see in­struction manual before proceeding!
more detailled information available: see in­struction manual before proceeding!
Safety Instructions
S
Emerson Process Management GmbH & Co. OHG S-3
Short Form Manual
X-STREAM XE
Safety Instructions
HASXEE-SFM-HS
04/2010

SYMBOLS USED IN THIS MANUAL

Where one or more of the following symbols appear within this manual, carefully read the rela­ted information and instructions!
Strictly observe the given warnings, instructions and information to minimize hazards!
This symbol used in the manual ... ... means
dangerous voltages may be exposed
hot surfaces may be exposed
possible danger of explosion
toxic substances may be present
substances harmful to health may be present
indicates notes relating to heavy instruments
electrical components may be destroyed by
electrostatic discharges
units must be disconnected from the power source
indicates special instructions or information for operation at low temperatures.
indicates basic conditions or procedures are being described.
This symbol may also indicate information impor­tant for achieving accurate measurements.
Emerson Process Management GmbH & Co. OHGS-4
Short Form Manual
HASXEE-SFM-HS 04/2010

Safety Instructions

X-STREAM XE
SAFETY INSTRUCTIONS

INTENDED USE STATEMENT

X-STREAM XE series gas analyzers are intended to be used as analyzers for industrial purposes. They must not be used in medical, diagnostic or life support applications.
Using X-STREAM XE analyzers as safety devices is prohibited where redundancy and/or SIL classication or equivalent is needed.
No independent agency certications or approvals are to be implied as covering such applications!
Safety Instructions
S

GENERAL SAFETY NOTICE / RESIDUAL RISK

If this equipment is used in a manner not specied in these instructions, protective sy­stems may be impaired.
Despite of incoming goods inspections, production control, routine tests and application of state-of-the-art measuring and test methods, an element of risk remains when operating a gas analyzer!
Even when operated as intended and observing all applicable safety instructions some residual risks remain, including, but not limited to, the following:
• An interruption of the protective earth line, e.g. in an extension cable, may result in risk to the user.
• Live parts are accessible when operating the instrument with doors open or covers removed.
• The emission of gases hazardous to health may even be possible when all gas con­nections have been correctly made.
Avoid exposure to the dangers of these residual risks by taking particular care when in­stalling, operating, maintaining and servicing the analyzer.
Emerson Process Management GmbH & Co. OHG S-5
Short Form Manual
X-STREAM XE
Safety Instructions
HASXEE-SFM-HS
04/2010

AUTHORIZED PERSONNEL

In-depth specialist knowledge is an absolutely necessary condition for working with and on the analyzer.
Authorized personnel for installing, operating, servicing and maintaining the analyzer are instructed and trained qualied personnel of the operating company and the manufac­turer.
It is the responsibility of the operating company to
• train staff,
• observe safety regulations,
• follow the instruction manual.
Operators must
• have been trained,
• have read and understood all relevant sections of the instruction manual before commencing work,
• know the safety mechanisms and regulations.
To avoid personal injury and loss of property, do not install, operate, maintain or service this instrument before reading and understanding this instruction manual and receiving appropriate training.

ADDITIONAL LITERATURE

This manual covers aspects important for installation and startup of X-STREAM XE gas analyzers.
For comprehensive information on operating and maintain/service the instrument in a safe manner it is MANDATORY to read all additional instruction manuals, if not provided as printed version, see the accompanying CD-ROM for an electronic version (PDF)!
The following additional instruction manuals are available or referenced within this ma­nual:
HASXEE-IM-HS X-STREAM XE series instruction manual HASICx-IM-H Infallible containment instruction manual Contact your local service center or sales ofce when missing documents.
SAVE ALL INSTRUCTIONS FOR FUTURE USE!
Emerson Process Management GmbH & Co. OHGS-6
Short Form Manual
HASXEE-SFM-HS 04/2010
X-STREAM XE
Safety Instructions

OPERATING AND MAINTAINING THIS UNIT

On leaving our factory, this instrument confor­med to all applicable safety directives.
In order to preserve this state of affairs, the operator must take care to follow all the in­structions and notes given in this manual and on the unit.
Before switching on the unit, ensure that the local nominal mains voltage corresponds to the factory-set operational voltage of this unit.
Any interruption of the protective earth con­nections, whether inside or outside of the unit, may result in exposure to the risk of electricity. Deliberately disconnected the protective earth is therefore strictly forbidden.
Removing covers may expose components conducting electric current. Connectors may also be energised. The unit should therefore be disconnected from the power supply before any kind of maintenance, repair or calibration
work requiring access to the inside of the unit.
Only trained personnel who are aware of the risk involved may work on an open and energized unit.
Fuses may only be replaced by fuses of an identical type and with identical ratings. It is forbidden to use repair fuses or to bypass fuses.
Take note of all applicable regulations when using this unit with an autotransformer or a variable transformer.
Substances hazardous to health may escape from the unit’s gas outlet. This may require additional steps to be taken to guarantee the safety of operating staff.
Safety Instructions
S
Emerson Process Management GmbH & Co. OHG S-7
X-STREAM XE
The units described in this manual may not be used in explosive atmospheres without additional safety measures.
Do not operate without covers secure.
Short Form Manual
HASXEE-SFM-HS
04/2010
Safety Instructions
EXPLOSION HAZARD
ELECTRICAL SHOCK HAZARD
Do not open while energized.
Installation requires access to live parts which can cause death or serious injury.
For safety and proper performace this instrument must be connected to a properly grounded three-wire source of power.
TOXIC GASES
This unit’s exhaust may contain toxic gases such as (but not limited to) e.g. sulfur dioxide. These gases can cause serious injuries. Avoid inhaling exhaust gases.
Connect the exhaust pipe to a suitable ue and inspect the pipes regularly for leaks.
All connections must be airtight to avoid leaks;
X-STREAM XE series
instruction manual on how performing a leak test.
Emerson Process Management GmbH & Co. OHGS-8
Short Form Manual
HASXEE-SFM-HS 04/2010
The models intended for outside and wall mounted use (X-STREAM XEF and XDF) weigh between 25 kg (57 lb) and 45 kg (99 lb), depending on version and options installed.
X-STREAM XE
Safety Instructions
HEAVY INSTRUMENT
Safety Instructions
Two people and/or lifting equipment is required to lift and carry these units.
Take care to use anchors and bolts specied to be used for the weight of the units!
Take care the wall or stand the unit is intended to be installed at is solid and stable to support the weight!
CRUSHING HAZARD
Take care of crushing hazard when closing the front door of analyzer eld housings!
Keep out of the closing area between enclosure cover and base!
OPERATION AT LOW TEMPERATURES
When operating an instrument at temperatures below 0 °C (32 °F), do NOT apply gas nor operate the internal pump before the warmup time has elapsed!
S
Violation may result in condensation inside the gas paths or damaged pump diaphragm!
HIGH TEMPERATURES
Hot parts may be exposed when working on photometers and/or heated components in the unit.
Emerson Process Management GmbH & Co. OHG S-9
X-STREAM XE
Safety Instructions
GASES AND PREPARATION OF GASES
GASES HAZARDOUS TO HEALTH
Follow the safety precautions for all gases (sample and span gases) and gas cylinders.
Before opening the gas lines, they must be purged with air or neutral gas (N2) to avoid hazards from escaping toxic, ammable, explosive or hazardous gases.
Short Form Manual
HASXEE-SFM-HS
04/2010
FLAMMABLE OR EXPLOSIVE GASES
When supplying explosive gases or ammable gases with concentrations of more than 25 % of the lower explosion limit, we RECOMMEND implementing one or more additional safety measures:
• purging the unit with inert gas
• stainless steel internal pipes
• ame arrestors on gas inlets and outlets
• infallible measuring cells.
Emerson Process Management GmbH & Co. OHGS-10
Short Form Manual
HASXEE-SFM-HS 04/2010
POWER SUPPLY
X-STREAM XE
Safety Instructions
CONNECTING UNITS FOR PERMANENT INSTALLATION
Only qualied personnel following all applicable and legal regulations may install the unit and connect it to power and signal cables. Failure to comply may invalidate the unit’s warranty and cause exposure to the risk of damage, injury or death.
This unit may only be installed by qualied personnel familiar with the possible risks.
Working on units equipped with screw-type terminals for electrical connections may require the exposure of energized components.
Wall-mounted units have no power switch and are operational when connected to a power supply. The operating company is therefore required to have a power switch or circuit breaker (as per IEC 60947-1/-3) available on the premises. This must be installed near the unit, easily accessible to operators and labelled as a power cut-off for the analyzer.
HAZARD FROM WRONG SUPPLY VOLTAGE
Ensure that the local power voltage where the unit is to be installed, corresponds to the unit’s nominal voltage as given on the name plate label.
Safety Instructions
S
ADDITIONAL NOTES FOR UNITS WITH SCREW-TYPE TERMINALS
Cables for external data processing must be double-insulated against mains power.
If this is not possible, cables must be laid in such a way as to guarantee a clearance of at least 5 mm from power cables. This clearance must be permanently secured (e.g. with cable ties).
Emerson Process Management GmbH & Co. OHG S-11
Short Form Manual
X-STREAM XE
HASXEE-SFM-HS
04/2010
General Operating Notes
GENERAL OPERATING NOTES
EXPLOSION HAZARD BY EXHAUST GASES
Exhaust gases may contain hydrocarbons, other toxic gases or even explosive gases.
Faulty gas connections may lead to explosion and death.
Ensure that all gas connections are connected as labelled and airtight.
The unit must be installed in a clean and dry area, protected from strong vibrations and frost
.
The unit must not be exposed to direct sunlight and sources of heat. Admissible ambient temperatures (see technical details) must be adhered to.
Gas inlets and outlets must not be interchanged.All gases must be conditioned, before sup­plied to the unit. When using this unit with corrosive sample gases, ensure that these gases do not contain components harmful to the gas lines.
Admissible gas pressure for sample and test gases is 1,500 hPa.
Exhaust lines must be laid inclined downwards, depressurized, protected from frost and according to applicable regulations.
If it is necessary to disconnect the gas lines, the unit’s gas connectors must be sealed with PVC caps to avoid polluting the internal gas lines with condensate, dust, etc.
To ensure electromagnetic compatibility (EMC), only shielded cables (supplied by us on re­quest, or of equivalent standard) may be used. The customer must ensure that the shielding is correctly tted. Shielding and terminal housing must be electrically connected; submin-D plugs and sockets must be screwed to the unit.
When using optional external adapters (submin-D to screw-type terminals), protection against electromagnetic interference is not ensured (CE compliance pursuant to EMC guidelines). In this case the customer or operating company functions as a system builder and must therefore ensure and declare compliance with EMC guidelines.
Emerson Process Management GmbH & Co. OHGS-12
Short Form Manual
HASXEE-SFM-HS 04/2010
Chapter 1
Technical description
The following are the main features of the new Emerson Process Management X-STREAM XE series gas analyzers in brief:
compact design with easily accessible internal components
customizable for a wide range of appli­cations: different housings are available while internal construction remains largely identical
a highly integrated mainboard contains all interfaces and basic functions for the operation of the unit
multilingual microprocessor-controlled user interface with liquid crystal (LCD) to indicate measurement value and status messages
units for outdoor use are optionally sup­plied with an impact tested front panel
widerange power supply unit for worldwide use without modication (1⁄2 19in units with external PSUs)
The new X-STREAM XE gas analyzers can measure up to ve different gas compon­ents using any combination of the following analyzing techniques (restrictions apply to
1
2 19in units): IR = non-dispersive infrared analysis UV = ultraviolet analysis pO2 =
paramagnetic oxygen analysis
X-STREAM XE
eO2 = TC = thermal conductivity analysis Modied resistant measuring cells are avai-
lable for use with corrosive gases and/or gases containing solvents.
Special congurations (e.g. intrinsically safe or infallible measuring cells) for the analysis of combustible gases are also available.
Standard applications
Different housings allow X-STREAM XE analyzers to be tailored to the many different applications:
Tabletop units in 1⁄2 19in modular design,
Tabletop and rackmountable units in 19in
Stainless steel wallmountable eld housing
Cast aluminium wallmountable field
The various analyzer types are described in more detail beginning with 1-3.
electrochemical oxygen analysis
with IP 20 protection class
modular design, with IP 20 protection class
with IP 66 / NEMA 4X protection class for outdoor use (operating temperature 0 (-20) °C to +50°C).
housing with IP 66 / NEMA 4X protection class for outdoor use (operating tempera­ture 0 (-20) °C to +50 °C).
Techn. Description
1
EXPLOSION HAZARD
X-STREAM XE analyzers CAN NOT be used in explosive environments (hazardous areas) without additional safety features.
This instruction manual does NOT describe the special conditions necessary to operate gas analyzers in hazardous areas.
Please refer to the separate instruction manual, supplied with units for use in hazardous areas.
Emerson Process Management GmbH & Co. OHG 1-1
X-STREAM XE

1.1 Overview

Short Form Manual
HASXEE-SFM-HS
04/2010
1.1 Overview
All X-STREAM gas analyzers feature an easy-to-use graphical user interface, which displays measurement values, status and error messages, and menus for the input of parameters.
For ease of use, the operator can select one of three languages for the display: By default any analyzer is congured with English and
1.1.1 The Front Panel
The graphic LCD shows measurement and status information with plain text and sym­bols.
The symbols are designed to indicate the dif­ferent status “Failure”, “Function check”, “Out of specication” and “Maintenance request” as specied by the NE 107 standard. For further information,
X-STREAM XE instruction
manual.
German language sets, while a third can optionally be added. Currently available: French, Italian, Portuguese and Spanish..
The analyzer software is operated by means of only six keys. Depending on the analyzer model, the display can also be protected with an impact tested glass panel, to withstand even harsher conditi­ons and to provide a higher IP protection class of up to IP66 / NEMA Type 4X.
1 3 42
1 Graphic display 2 “Measure“ key 3 “Enter” key 4 4 keys for settings and menu navigation

Fig. 1-1: X-STREAM front panel (here the X-STREAM XEGP)

Emerson Process Management GmbH & Co. OHG1-2
X-STREAM XE
1.2 Comparison of Analyzer Models

1.2 Comparison of the Various X-STREAM Analyzer Model

Short Form Manual
HASXEE-SFM-HS
04/2010
X-STREAM XEGC
1
2 19 in housing, table-top or rackmountable,
protection type: IP 20
1
1 19 in housing, table-top or rackmountable,
protection type: IP 20
X-STREAM XEGP
24V input with external power supply unit Internal wide range power supply unit Internal wide range power supply unit Internal wide range power supply unit Max. 3 channels: 2 photometer + 1 WLD/O
2
max. 6 gas connections,
including 1 optional purge gas connection Options for gas lines: Flow sensor, pressure
sensor, infallible gas lines
Max. 4 channels in any combination max. 8 gas connections,
1 optional extra connection for purge gas Options for gas lines: Flow sensor, pressu-
re sensor, heating for physical components, sample gas pump, 1 or 2 valve blocks, infalli­ble gas lines
1-4 analog outputs, 4 relay outputs, 2 Modbus Ethernet interfaces, 2 USB connectors
optional: 1 interface card with 7 digital inputs and 9 digital outputs 1 interface card with analog inputs
electrical interfaces accessible via sockets on back of unit, optionally: screw-type terminal
adapters
(except for Ethernet & USB)
1-4 analog outputs, 4 relay outputs, 2 Modbus Ethernet interfaces, 2 USB connectors
optional: 1 or 2 interface cards, each with 7 digital inputs and 9 digital outputs 1 interface card with analog inputs
electrical interfaces accessible via sockets on back of unit, optionally: screw-type terminal
adapters
(except for Ethernet & USB)
LCD LCD LCD, impact tested front panel LCD, impact tested front panel Operating ambient temperature:
0 °C to +50 °C (32 °F to 122 °F)
Operating ambient temperature: 0 °C to +50 °C (32 °F to 122 °F)
Size: (DxHxW): max. ca. 440x130x220 mm Weight: ca. 8 - 12 kg (17.6 - 26.5 lb)
Size: (DxHxW): max. ca. 411x133x482 mm Weight: ca. 11 - 16 kg (24.3 - 35.3 lb)
Emerson Process Management GmbH & Co. OHG1-3
Short Form Manual
HASXEE-SFM-HS 04/2010
1.2 Comparison of Analyzer Models
X-STREAM XEF / XDF X-STREAM XEFD
Stainless steel wallmountable eld housing, protection type: IP66 / NEMA 4X
Single or dual compartment design
X-STREAM XE
Techn. Description
1
Cast aluminium wallmountable eld housing, protection type: IP66 / NEMA 4X
Max. 4 channels in any combination max. 8 gas connections,
1 optional extra connection for purge gas Options for gas lines: Flow sensor, pressure
sensor, heating for physical components, sample gas pump, 1 or 2 valve blocks, infalli­ble gas lines
1-4 analog outputs, 4 relay outputs, 2 Modbus Ethernet interfaces, 2 USB connectors
optional: 1 or 2 interface cards, each with 7 digital
inputs and 9 digital outputs 1 interface card with analog inputs
electrical interfaces on internal screw-type ter­minal adapters
Operating ambient temperature:
-20 °C to +50 °C (-4 °F to 122 °F) Models available for use in explosive environ-
ments
Size: (DxHxW): ca. 265x400 (815)x550 mm Weight: max. ca. 25 (45) kg / 55.1 (99.2) lb
(except for Ethernet & USB)
Max. 4 channels in any combination max. 8 gas connections,
including 2 optional purge gas connection Options for gas lines: Flow sensor, pressu-
re sensor, heating for physical components, sample gas pump, 1 or 2 valve blocks, infalli­ble gas lines
1-4 analog outputs, 4 relay outputs, 2 Modbus Ethernet interfaces, 2 USB connectors
optional: 1 or 2 interface cards, each with 7 digital
inputs and 9 digital outputs 1 interface card with analog inputs
electrical interfaces on internal screw-type ter­minal adapters
Operating ambient temperature:
-20 °C to +50 °C (-4 °F to 122 °F) Flameproof enclosure: approved for use in
explosive environments
Size: (DxHxW): max. ca. 222x512x578 mm Weight: max. ca. 63 kg (138.9 lb)
This model is not covered by this manual! See the separate X-STREAM Ex d manual for hazardous area applications!
(except for Ethernet & USB)
Emerson Process Management GmbH & Co. OHG 1-4
Short Form Manual
HASXEE-SFM-HS 04/2010
1.3 X-STREAM XEGC

1.3 X-STREAM XEGC: ½ 19 Inch Table-Top Unit

X-STREAM XE
This compact model for general purposes can be tted with up to three photometer measurement channels (IR or UV) and one additional non-photometer channel (eO
, pO2
2
or WLD). Power is supplied by a separate external power supply unit.
By default the units are congured for tabletop use. For rack mounting an accessory kit can be ordered to either install 2 units alongside, or one single unit together with a blind pla­te.
Connection to power supply
DC 24 V power is supplied via a 3-pin socket at the rear of the unit.
Interfaces
Electrical connections for interface signals are provided via submin-D connectors, Ethernet and USB connectors mounted on the rear panel of the device
(g. 1-5). For applications where screw-type terminals are preferred, optional adapters are available, which are mounted directly onto the submin-D connectors.
Techn. Description
1
1 3 4 5 62
1 Status line 2 Graphic display 3 Messages line 4 home key 5 enter key 6 4 keys for adjustment and menu selection

Fig. 1-2: X-STREAM XEGC

Emerson Process Management GmbH & Co. OHG 1-5
7 Signal connectors (some optional) 8 Gas ttings 9 Power input fuse 10 Power input connector
7 8 9 10
X-STREAM XE
1.4 X-STREAM XEGP

1.4 X-STREAM XEGP: 19 Inch Table-Top Or Rackmount Design

Short Form Manual
HASXEE-SFM-HS
04/2010
This model can be fitted with up to four measurement channels in any combination. The physical components can optionally be encased in a cover. This area can be held at a specic temperature of up to 60 °C to mi­nimize interference from changes in external temperature.
Units congured for rack mounting can be converted for tabletop use by removing the lateral mounting brackets and attaching the four feet supplied as accessories.
Front view
Connection to power supply
Main power is supplied via the IEC chassis plug mounted on the rear panel, with inte­grated power switch and fuse holders. The internal wide range power supply unit enables the analyzers to be used worldwide.
Adaptors are optionally available for signal connections, to provide screw terminals in­stead of submin-d-connectors (
1-7).
1 3 4 5 62
1 Status line 2 Graphic display 3 Messages line
Rear view
1 Gas connector ttings 2 Space for additional ttings 3 Optional purge gas inlet

Fig. 1-3: X-STREAM XEGP

4 home key 5 enter key 6 4 keys for adjustment and menu selection
15 2 4 3
4 Power inlet with lter, fuses & switch 5 Signal input/output connectors (some optional)
Emerson Process Management GmbH & Co. OHG1-6
Short Form Manual
HASXEE-SFM-HS 04/2010
X-STREAM XE
1.4 X-STREAM XEGP
Strain-reliefs, top view details
Techn. Description
1
1
1

Fig. 1-4: Optional screw-type terminal adapters

2
1 Screw-type terminal adapters 2 Strain-reliefs
Emerson Process Management GmbH & Co. OHG 1-7
Short Form Manual
X-STREAM XE
1.5 X-STREAM XE Field Housing

1.5 X-STREAM XEF/XDF: Field Housing With Single Or Dual Compartment

HASXEE-SFM-HS
04/2010
The eld housing model is conceived for outdoor use and wall-mounting. The coa­ted stainless steel housing has a protection class rated IP66 / NEMA Type 4X, offering protection against water and dust entering the device:
IPx6: Water projected in powerful jets against the enclosure from any directions shall have no harmful effects
IP6x: Protection against penetration by dust. Live or internal moving parts are completely protected.
X-STREAM eld housings can be tted with up to four measurement channels in any com­bination. The physical components can op­tionally be encased in a cover. This separate
volume can be held at a specic temperature of up to 60 °C to minimize interference from changes in external temperature.
Front panel
The analyzer ’s display can o
ptionally be protected by an impact tested glass for en­hanced protection against breakage in harsh environments.
Connection to power supply
Mains power is supplied via screw-type termi­nals with integrated fuse holders at the right of the housing, near the front. The wide range power supply unit mounted internally enables the analyzers to be used worldwide.
1 3 4 5 62
1 Status line 2 Graphic display 3 Messages line
4 home key 5 enter key 6 4 keys for adjustment and menu selection

Fig. 1-5: X-STREAM XEF / XDF- Front views

Emerson Process Management GmbH & Co. OHG1-8
Short Form Manual
HASXEE-SFM-HS 04/2010
X-STREAM XE
1.5 X-STREAM XE Field Housing
Techn. Description
1
21 3
1 Cable gland for power cable 2 Cable glands for signal cables 3 4 brackets for wall-mounting 4 Gas in- & outlets (max. 8) 5 Cutouts, to combine 2 housings (here closed)

Fig. 1-6: X-STREAM XEF - right side and bottom view

HEAVY INSTRUMENT
The X-STREAM eld housing, intended for outside and wall mounted use, weighs approx. 26 kg/57 lb (XEF) or 45 kg/99 lb (XDF), depending on options installed.
Two people and/or lifting equipment is required to lift and carry these units.
Take care to use anchors and bolts specied to be used for the weight of the units!
5 4 5
Note!
In case of XDF, the cable glands are located at the upper compartment, while the gas in­& outlets are at the bottom side of the lower compartment. Also only 2 brackets are at each compartment.
Take care the wall or stand the unit is intended to be installed at is solid and stable to support the weight!
Emerson Process Management GmbH & Co. OHG 1-9
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
04/2010
1.7 X-STREAM XE Field Housing
1 32 45
(shown with front
panel removed)
1 Screw-type terminals for signal cables 2 Power line lter 3 Cable glands 4 Power supply terminals with integrated fuses 5 Ethernet and USB connectors

Fig. 1-7: X-STREAM XEF - power supply and signal terminals

Note!
In case of XDF, the terminals and connectors are located at the upper compartment, while physical compo­nents and gas ttings are in the lower compartment.
Emerson Process Management GmbH & Co. OHG1-10