Rosemount Manual: Rosemount Oxymitter DR Manuals & Guides

00809-0100-4884, Rev AC
Rosemount™ Oxymitter DR
In-Situ Oxygen Probe for General Purpose Locations
Reference Manual
April 2022
Warranty
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective portion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of dismantling, re-installation and freight, and the time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESSED OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
Safety instructions
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to ensure compliance with the Low Voltage Directive. Non-EU states should also comply with the following unless superseded by local or national standards.
2
WARNING
Adequate earth connections should be made to all earthing points, internal and external, where provided. After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all earth terminals
must be maintained at all times. Main supply cords should comply with the requirements of IEC227 or IEC245. All wiring shall be suitable for use in an ambient temperature of greater than 221 °F (105 °C). All cable glands used should be of such internal dimensions as to provide adequate cable anchorage. To ensure safe operation of this equipment, connection to the mains supply should only be made through a circuit breaker,
which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker may also include a mechanically operated isolating switch. If it does not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard, such as IEC947. All wiring must conform with any local standards.
Where equipment or covers are marked with the symbol below, hazardous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment - and then only by trained service personnel.
Where equipment or covers are marked with the symbol below, there is a danger of hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Certain surfaces may remain hot to the touch.
Where equipment or covers are marked with the symbol below, refer to the reference manual for instructions.
All graphical symbols used in this product are from one or more of the following standards: EN61010, IEC417, and ISO3864. Where equipment or labels are marked "Do Not Open While Energized" or similar, there is a danger of ignition in areas where
an explosive atmosphere is present. This equipment should only be opened when the power is removed and adequate time as specified on the label or in the Reference Manual has been allowed for the equipment to cool down - and then only by trained service personnel.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
3
4
Reference Manual Contents
00809-0100-4884 April 2022

Contents

Chapter 1 Description and specifications....................................................................................7
1.1 Components checklist................................................................................................................. 7
1.2 Overview................................................................................................................................... 10
1.3 Probe options............................................................................................................................ 13
1.4 Specifications............................................................................................................................ 16
Chapter 2 Install.......................................................................................................................19
2.1 Mechanical installation.............................................................................................................. 19
2.2 Electrical installation..................................................................................................................31
2.3 Pneumatic installation............................................................................................................... 34
2.4 Set up probe with electronics package.......................................................................................36
Chapter 3 Startup and operate................................................................................................. 45
Chapter 4 Maintenance and service..........................................................................................47
4.1 Overview................................................................................................................................... 47
4.2 Calibrate....................................................................................................................................47
4.3 Remove probe........................................................................................................................... 48
4.4 Replace probe............................................................................................................................49
4.5 Replace terminal block...............................................................................................................49
4.6 Replace entire probe (excluding electronics)............................................................................. 50
4.7 Replace heater strut...................................................................................................................50
4.8 Replace cell................................................................................................................................53
4.9 Replace ceramic diffusion element............................................................................................ 55
4.10 Wire termination housing........................................................................................................56
Chapter 5 Troubleshoot........................................................................................................... 59
5.1 No response to oxygen concentration change when heater is cold and thermocouple mV
output is less than setpoint......................................................................................................... 59
5.2 No response to oxygen concentration change when heater is hot and thermocouple mV
output is at setpoint ±0.2 mV......................................................................................................59
5.3 Good response with incorrect indication....................................................................................60
5.4 Probe does not give accurate indication of applied test gas....................................................... 61
5.5 Probe heater temperature unstable........................................................................................... 61
5.6 Probe passes calibration, but still appears to read too high........................................................ 62
5.7 Probe passes calibration, but still appears to read too low..........................................................63
Chapter 6 Return material........................................................................................................65
Chapter 7 Replacement parts................................................................................................... 67
Appendix A Optional accessories................................................................................................ 77
Rosemount Oxymitter DR 5
Contents Reference Manual
April 2022 00809-0100-4884
A.1 Rosemount™ By-Pass Package....................................................................................................77
A.2 Rosemount™ O2 calibration gas sequencer................................................................................ 77
Appendix B Upgrade Rosemount™ Oxymitter DR to full Oxymitter............................................. 79
Appendix C Rosemount Oxymitter™ product certifications.........................................................81
C.1 Directive information................................................................................................................ 81
C.2 Ordinary location certification................................................................................................... 81
C.3 Installing equipment in North America...................................................................................... 81
C.4 Rosemount OXT4A Oxymitter for general purpose locations and OXT4C general locations
certification................................................................................................................................ 81
Appendix D Declaration of Conformity........................................................................................83
Appendix E China RoHS table......................................................................................................85
6 Emerson.com/Rosemount
Reference Manual Description and specifications
00809-0100-4884 April 2022

1 Description and specifications

1.1 Components checklist

Also use the following tables to compare your order number against your unit. The title of the table defines the model. The rest of the table defines the various options and features of the Rosemount Oxymitter DR. Ensure the features and options specified by your order number are on or included with the unit.A typical Rosemount™ Oxymitter DR should contain the items shown in Figure 1-1. Record the part number, serial number, and order number for each component of your system.
Figure 1-1: Typical System Package
A. Quick Start Guide B. Adapter plate with mounting hardware and gasket C. Oxymitter DR
Table 1-1: Rosemount Oxymitter DR Ordering Information
OXT4ADR Oxymitter DR In-Situ Oxygen Transmitter
Code Sensing probe type
1 ANSI (North American standard) probe with ceramic diffuser
Rosemount Oxymitter DR 7
Description and specifications Reference Manual
April 2022 00809-0100-4884
Table 1-1: Rosemount Oxymitter DR Ordering Information (continued)
2 ANSI probe with flame arrestor and ceramic diffuser
3 ANSI probe with snubber diffuser
4 DIN (European standard) probe with ceramic diffuser
5 DIN probe with flame arrestor and snubber diffuser
6 DIN probe with snubber diffuser
7 JIS (Japanese standard) probe with ceramic diffuser
8 JIS probe with flame arrestor and ceramic diffuser
9 JIS probe with snubber diffuser
Code Probe assembly
0 18-in. (457 mm) probe
1 18-in. (457 mm) probe with abrasive shield
(2)
2 3-ft. (0.91 m) probe
3 3-ft. (0.91 m) probe with abrasive shield
(2)
(1)
(1)
(1)
4 6-ft. (1.83 m) probe
5 6-ft. (1.83 m) probe with abrasive shield
(2)
6 9-ft. (2.74 m) probe
7 9-ft. (2.74 m) probe with abrasive shield
(2)
8 12-ft. (3.66 m) probe
9 12-ft. (3.66 m) probe with abrasive shield
A 15-ft. (4.57 m) probe with abrasive shield
B 18-ft. (5.49 m) probe with abrasive shield
(2)
(2)
(2)
Code Mounting hardware: stack side
0 No mounting hardware (You must choose 0 under Mounting
Hardware - Probe Side below.)
1 New installation: square weld plate with studs
2 Mounting to Model 218 mounting plate (with Model 218 shield
removed)
3 Mounting to existing Model 218 support shield
4 Mounting to other mounting
(3)
5 Mounting to Model 132 adapter plate
Code Mounting hardware: probe side
0 No mounting hardware
1 Probe only (ANSI) (North American standard)
2 New bypass or abrasive shield (ANSI)
4 Probe only (DIN) (European standard)
8 Emerson.com/Rosemount
Reference Manual Description and specifications
00809-0100-4884 April 2022
Table 1-1: Rosemount Oxymitter DR Ordering Information (continued)
5 New bypass or abrasive shield (DIN)
7 Probe only (JIS) (Japanese standard)
8 New bypass or abrasive shield (JIS)
Code Housing and customer termination
11 Standard filtered termination
12 Transient protected filtered termination
Code Arrangment: existing electronics
03 For use with existing analog electronics (including
Westinghouse/Rosemount 132/218/225)
04 Westinghouse/Rosemount digital (218A) or universal electronics
05 VeriTrim electronics
07 Model 132 digital electronics
08 For use with Yokogawa electronics (cold junction comp. in probe
junction box)
09 For use with other competitive oxygen transmitter systems
(1) No agency approvals. See Rosemount Hazardous Area Oxymitter 4000 Reference Manual for
hazardous area certified probes.
(2) Recommended usages: High velocity particulates in flue stream, installation within 11.5 ft. (3,5
m) of soot blowers or heavy salt cake buildup. Applications: pulverized coal, recovery boilers, lime kiln. Regardless or application, Emerson recommends abrasive shields with support brackets for 9 ft. (2,74 m), 12 ft. (3,66 m), 15 ft. (4,57 m), and 18 ft. (5,49 m) probe installations, particularly horizontal installations.
(3) Where possible, specify SPS number; otherwise, provide details of the existing mounting plate as
follows. See #unique_6/unique_6_Connect_42_table_x2d_lkk_chb.
Note
1. High sulfur service: For high sulfur applications, please add an additional line item to your purchase order requesting high sulfur cell part number 4847B63G02 in lieu of the standard ZrO2 cell.
Cell replacement kits for high sulfur service are also available.
2. Where possible, specify ANSI or DIN designation; otherwise, provide details of existing mounting plate as follows:
Plate with studs
Plate without studs Bolt circle diameter, number, and arrangement of holes,
Bolt circle diameter, number and arrangement of studs, stud thread, and stud height above mounting plate
threads, and depth of stud mounting plate with accessories
Rosemount Oxymitter DR 9
Description and specifications Reference Manual
April 2022 00809-0100-4884
Table 1-2: Calibration Components
Part number Description
1A99119G01 Calibration gas kit: Contains 0.4% and 8.0% O2 bottles, 145.3 gal.
1A99119G02 Calibration gas regulators kit: Contains two CGA-590 regulators
1A99119G03 Bottle rack
(1) Calibration gas bottles cannot be shipped via air freight.

1.2 Overview

1.2.1 Scope

This Reference Manual provides information needed to install, start up, operate, and maintain the Rosemount™ Oxymitter DR.
You can interface the Oxymitter DR to a number of different earlier model electronics packages. This manual does not cover these electronic packages. For specification information concerning calibration and operation of the system, refer to the Reference Manual applicable to your electronics.
1.2.2

Description

Emerson designed the Oxymitter DR to measure the net concentration of oxygen in an industrial combustion process (i.e., the oxygen remaining after all fuels have been oxidized).
(550 L) each (requires two CGA-590 regulators)
(1)
The probe is permanently positioned within an exhaust duct or stack and performs its task without a sampling system.
The equipment measures oxygen percentage by reading the voltage developed across a heated electrochemical cell, which consists of a small yttria-stabilized, zironcia disc. Both sides of the disc are coated with porous metal electrodes. When operated at the proper temperature, the millivolt output voltage of the cell is given by the following Nernst equation:
EM = KT log10(P1/P2) + C
Where:
P2 is the partial pressure of the oxygen in the measured gas on one side of the cell.
P1 is the partial pressure of the oxygen in the reference air on the opposite side of the ce3ll.
T is the absolute temperature.
C is the cell constant.
K is an arithmetic constant.
When the cell is at an operating temperature and there are unequal oxygen concentrations across the cell, oxygen ions will travel from the high oxygen partial pressure side to the low oxygen partial pressure side of the cell. The resulting logarithmic
10 Emerson.com/Rosemount
Reference Manual Description and specifications
00809-0100-4884 April 2022
output voltage is approximately 50 mV per decade. The output is proportional to the inverse logarithm of the oxygen concentration. Therefore, the output signal increases as the oxygen concentration of the sample gas decreases. This characteristic enables the transmitter to provide exceptional sensitivity at low oxygen concentrations.
The transmitter measures net oxygen concentration in the presence of all the products of combustion, including water vapor. Therefore, you can think of it as an analysis on a wet basis. In comparison with older methods, such as the portable apparatus, which provides an analysis on a dry gas basis, the wet analysis will, in general, indicate a lower percentage of oxygen. The difference is proportional to the water content of the sample stream.

1.2.3 Configuration

Rosemount™ Oxymitter DR Transmitters are available in seven length options, giving the operator flexibility to use an in situ penetration appropriate to the size of the stack or duct. The options are:
18 in. (457 mm)
3 ft. (0,91 m)
1.2.4
6 ft. (1,83 m)
9 ft. (2,7 m)
12 ft. (3,66 m)
15 ft. (4,57 m)
18 ft. (5,49 m)
Emerson offers abrasive shields for applications where abrasive particulates are present. Acid-resistant cells are available for SO2 and HCl environments. Bypass and probe mounting jacket options are available for process temperatures above 1300 °F (705 °C).

Features

The cell output voltage and sensitivity increase as the oxygen concentration decreases.
The Rosemount™ Oxymitter DR includes:
Field replaceable cell, heater, thermocouple, and diffusion element.
Rugged 316L stainless steel wetted parts.
Membrane keypad
Local operator interface (LOI)
Optional HART® interface
Rosemount Oxymitter DR 11
Description and specifications Reference Manual
April 2022 00809-0100-4884

1.2.5 Handling

CAUTION
The Rosemount Oxymitter DR is designed for industrial applications. Some probe components are made from ceramics, which are susceptible to shock when mishandled.
Treat each component of the system with care to prevent physical damage.

1.2.6 System considerations

Prior to installing your transmitter, make sure you have all the components necessary to make the system installation. Ensure all components are properly integrated to make the system functional.
After verifying that you have all the components, select mounting locations and determine how each component will be placed in terms of available line voltage, ambient temperatures, environmental considerations, convenience, and serviceability.
1.2.7
Instrument air for reference air is optional for most applications. Ambient air passively diffuses into the inside of the probe in sufficient quantity for an accurate measurement. Instrument air is required for applications where the ambient air at the probe location may not contain the typical 20.95 percent O2. An example would be an installation into a positive pressure flue gas duct which has many leaks into the surrounding air.
If the calibration gas bottles will be permanently connected, use a blocking valve or check valve next to the calibration fittings on the termination housing. This blocking valve or check valve is to prevent breathing of the calibration gas line and subsequent flue gas condensation and corrosion.
Note
Keep the packaging in which the transmitter arrived from the factory in case you need to ship any components to another site. Emerson has designed this packaging to protect the product.

Upgrading the Rosemount™ Oxymitter DR

You can easily upgrade the Rosemount Oxymitter DR to a full Oxymitter 4000 or 5000.
This provides an economical upgrade path for customers looking to preserve their probe investment upon the eventual failure of the signal conditioning electronics. To upgrade, add a small electronics package to the existing termination housing of the probe. The converted unit will be a full Rosemount Oxymitter 4000 or 5000 Oxygen Transmitter that can provide a 4-20 mA oxygen signal without an external signal conditioning electronics package. The Rosemount Oxymitter electronics comes with HART® or FOUNDATION Fieldbus communications. See Upgrade Rosemount™ Oxymitter DR to full Oxymitter for upgrade information.
12 Emerson.com/Rosemount
Reference Manual Description and specifications
00809-0100-4884 April 2022

1.3 Probe options

1.3.1 Diffusion elements

Ceramic diffusion assembly
The ceramic diffusion assembly is the traditional design for the probe.
Used for over 25 years, the ceramic diffusion assembly provides a greater filter surface area. This element is also available with a flame arrestor and with a dust shield to use with an abrasive shield.
Figure 1-2: Ceramic Diffusion Assembly
Snubber diffusion assembly
The snubber diffusion assembly is satisfactory for most applications. This element is also available with a dust seal for use with an abrasive shield.
Figure 1-3: Snubber Diffusion Assembly
Rosemount Oxymitter DR 13
Description and specifications Reference Manual
April 2022 00809-0100-4884
Cup type diffusion assembly
Use the cup type diffusion assembly in high temperature applications where frequent diffusion element plugging is an option.
Figure 1-4: Cup Type Diffusion Assembly
The assembly is available with either a 5 micron or 40 micron sintered Hastelloy element. This diffusion element offers the greater surface area of the ceramic element, but is much more rugged. This element is also available with a dust seal for use with an abrasive shield.
Abrasive shield assembly
The abrasive shield assembly is a stainless steel tube that surrounds the probe assembly. The shield protects against particle abrasion and condensations, provides a guide for ease of insertion, and acts as a positive support, especially for longer probes. The abrasive shield assembly uses a modified diffusor and vee deflector assembly, fitted with dual dust seal packing.
14 Emerson.com/Rosemount
Reference Manual Description and specifications
00809-0100-4884 April 2022
Figure 1-5: Abrasive Shield Assembly
Dimensions are millimeters with inches in parentheses.
A. View A. B. View B. C. Before welding, butt item 02 with item 01 as shown.
D. Weld on both sides with expanding chill block.
E. On inside Ø break for smooth rounded edge on both ends of chamber. F. Skin cut face for 90°.
G. Ø 0.75 in. (19 mm) through four places equally spaced on Ø 4.75 in. (120.65 mm) B.C.
Note
In highly abrasive applications, rotate the shield 90 degrees at normal service intervals to present a new wear surface to the abrasive flow stream.
Rosemount Oxymitter DR 15
Description and specifications Reference Manual
April 2022 00809-0100-4884

1.4 Specifications

1.4.1 Performance specifications

Temperature limits Process: 32 to 1300 °F (0 to 705 °C), up to 2400 °F (1300 °C) with
optional accessories
Resolution sensitivity 0.01% O2 transmitted signal
Sensing cell repeatability ±0.75% of O2 reading or 0.05% O
Mounting and mounting position
Vertical or horizontal
(1)
2
1.4.2
Lowest detectable limit 0.02% O
System speed of response to calibration gas
Reference air 2 SCFH (1 L/min), clean, dry, instrument quality air (20.95% O2),
Relative humidity 5 to 95% (non-condensing)
Power requirements Normal operation: 175 W
Ambient operating temperature (junction box)
(1) When used with Rosemount Oxymitter signal conditioning electronics.
T T90 < 8 seconds
regulated to 2.5 psi (34 kPa)
Maximum (warmup): 500 W
200 °F (93 °C) 160 °F (71 °C) maximum for YEW replacement
Rosemount has satisfied all obligations coming from the European legislation to harmonize the product requirements in Europe.
< 3 seconds
intial
2

Physical specifications

Probe lengths
18 in. (457 mm)
3 ft. (0.91 m)
6 ft. (1.83 m)
9 ft. (2.74 m)
12 ft. (3.66 m)
15 ft. (4.57 m)
18 ft. (5.49 m)
Materials of construction
Transmitter probe: Process-wetted materials are 316L or 304 stainless steel.
Non-wetted parts: 304 stainless steel, low-copper aluminum Termination housing: Low-copper aluminum
Process connections
16 Emerson.com/Rosemount
2-in. 150# (4.75-in. [121 mm] bolt circle); DIN (5.71-in. [145 mm] bolt circle)
Reference Manual Description and specifications
00809-0100-4884 April 2022
Note
Flanges are flat-faced and for mounting only. Flanges are not pressure-rated. Spool piece part numbers are available to offset electronics housing from hot ductwork. Many adapter flanges are available to mate to existing flanges.
Orientation
Heater voltage
Thermocouple
Vertical or horizontal mount
115 ±10 percent Vac, 50/60 Hz, 200 VA
Type K
Table 1-3: Probe Lengths and Approximate Shipping Weights
Length Weight
18 in. (457 mm) 16 lb. (7.3 kg)
3 ft. (0.91 m) 21 lb. (9.5 kg)
6 ft. (1.83 m) 27 lb. (12.2 kg)
9 ft. (2.74 m) 33 lb. (15.0 kg)
12 ft. (3.66 m) 39 lb. (17.7 kg)
15 ft. (4.57 m) 45 lb. (20.5 kg)
18 ft. (5.49 m) 51 lb. (23 kg)
Rosemount Oxymitter DR 17
Description and specifications Reference Manual
April 2022 00809-0100-4884
18 Emerson.com/Rosemount
Reference Manual Install
00809-0100-4884 April 2022

2 Install

WARNING
Failure to follow safety instructions could result in serious injury or death.
Before installing the equipment, read Safety instructions.

2.1 Mechanical installation

2.1.1 Select location

The location of the transmitter in the stack or flue is most important for maximum accuracy in the oxygen analyzing process.
If the probe will be installed into an existing location, proceed to Install probe.
Procedure
1. Position the transmitter so that the gas it measures is representative of the process. You will obtain best results by positioning the transmitter near the center of the
duct (40 to 60 percent insertion). Longer ducts may require several transmitter units since the O2 can vary due to stratification. A point too near the wall of the duct or the inside radius of a bend may not provide a representative sample because of the very low flow conditions. Select the sensing point so the process gas temperature falls within a range of 32 to 1300 °F (0 to 704 °C).
The ambient temperature of the termination housing must not exceed 200 °F (93 °C).
2. Check the flue or stack holes for air leakage. Make any necessary repairs or install the transmitter upstream of any leakage.
Ducts and stacks that operate under negative pressure will draw air in through any holes or torn seals, substantially affecting the oxygen reading.
3. Ensure that the area is clear of internal and external obstructions that will interfere with probe installation and access to the membrane keypad or LOI. Allow adequate clearance for removal of the transmitter.
Table 2-1: Calibration Gas and Reference Air
Calibration gas Reference air
ANSI 6.34 (¼) tube
DIN 6.35 (¼) tube
Table 2-2: Mounting Flange (Gasket Included)
ANSI DIN
Flange diameter 190 (7.5) 210 (8.25)
Rosemount Oxymitter DR 19
Install Reference Manual
April 2022 00809-0100-4884
Table 2-2: Mounting Flange (Gasket Included) (continued)
ANSI DIN
Hole diameter 19 (0.75) 18 (0.71)
4 holes equally spaced on BC 152,4 (6.00) 170 (6.69)
20 Emerson.com/Rosemount
Reference Manual Install
00809-0100-4884 April 2022
Figure 2-1: Rosemount Oxymitter DR Installation
All dimensions are in inches with millimeters in parentheses.
A. Dimension A B. Dimension B C. 0.062 THK gasket
ANSI: 3535B18H02 JIS: 3535B46H01 DIN: 3535B45H01
D. Insulate if exposed to ambient weather conditions.
E. Calibration gas F. Reference air
ANSI: ¼ (6.35) tube DIN: 0.24 (6) tube JIS: 0.24 (6) tube
G. Electrical connection ¾ NPT H. Cover removal and access
I. Bottom view J. Removal envelope
K. With standard snubber diffuser
L. Add to Dimension A for probe with ceramic diffuser and flame arrestor
M. Add to Dimension A for probe with ceramic diffuser
N. Process flow must be in this direction with respect to deflector 3534B48G01
Table 2-3: Installation/Removal
Probe Dimension A Dimension B
18 in. 16 (405) 31.8 (808)
3 ft. 34 (864) 49.8 (1265)
Rosemount Oxymitter DR 21
Install
Reference Manual
April 2022 00809-0100-4884
Table 2-3: Installation/Removal (continued)
Probe Dimension A Dimension B
6 ft. 70 (1778) 65.8 (2179)
9 ft. 106 (2962) 121.8 (3094)
12 ft. 142 (3607) 157.8 (406)
15 ft. 178 (4521) 193.8 (4923)
18 ft. 214 (5436) 229.8 (5387)
Figure 2-2: Remote Electronics with Membrane Keypad and Blind Cover
Note
All dimensions are in millimeters with inches in parentheses.
A. Pipe mount configuration
22 Emerson.com/Rosemount
Reference Manual Install
00809-0100-4884 April 2022
Figure 2-3: Remote Electronics with LOI and Window Cover, Wall Mount Configuration
Rosemount Oxymitter DR 23
Install Reference Manual
April 2022 00809-0100-4884
Figure 2-4: Rosemount Oxymitter DR with Abrasive Shield
These flat-faced flanges are manufactured to ANSI, DIN, and JIS bolt patterns and are not pressure rated. All dimensions are in inches with millimeters in parentheses.
A. Dimension A B. Dimension B: Removal envelope C. ¾ NPT electrical connection
Table 2-4: Abrasive Shield - 3D98003
D.
Flange ANSI JIS DIN
Flange diameter 229 (9.00) 235 (9.25) 235 (9.25)
Hole diameter 19 (0.75) 19 (0.75) 24 (0.94)
8 holes equal SP on BC 191 (7.50) 190 (7.48) 190 (7.48)
E. Diffuser/dust assembly
F. Deflector assembly G. Snubber/dust seal assembly H. Diameter nominal
Table 2-5: Installation/Removal
Probe Dimension A Dimension B
18 in. 330 (18) 32.5 (826)
3 ft. 31 (787) 50.5 (1283)
6 ft. 67 (1702) 86.5 (2197)
9 ft. 103 (2616) 122.5 (3112)
12 ft. 139 (3531) 158.5 (4028)
15 ft. 175 (4445) 194.5 (4940)
18 ft. 211 (5959) 230.5 (5855)
24 Emerson.com/Rosemount
Reference Manual Install
00809-0100-4884 April 2022
Ref air Cal gas
ANSI ¼-in. tube
DIN ¼-in. tube
(1) Add check valve in cal gas line.
(1)
Figure 2-5: Mounting Plate for Rosemount Oxymitter DR
Note
Dimensions are in millimeters with inches in parentheses.
A. Dimension A B. Dimension B C. Dimension C
D. Four studs, lockwashers, and nuts equally spaced on C diameter B.C.
Table 2-6: Adapter Plate Dimensions for Rosemount Oxymitter DR
Part numbers for adapter plates include attaching hardware.
Dimensions: mm (in.) ANSI (PN
4512C34G01)
A 153 (6.00) 191 (7.5) 165 (5.50)
B thread 0.625-11 (M-16 x 2) (M-12 x 1.75)
C diameter 121 (4.75) 145 (5.708) 130 (4.118)
DIN (PN 4512C36G01) JIS (PN 451235G01)
Rosemount Oxymitter DR 25
Install Reference Manual
April 2022 00809-0100-4884
Figure 2-6: Mounting Plate for Rosemount Oxymitter DR with Abrasive Shield
Note
Dimensions are in inches with millimeters in parentheses.
A. Dimension A. B. Dimension B. C. Dimension C.
D. Crosshatched area in four corners may be used to provide additional holes for field
bolting of plate to outside wall service. E. Eight threaded holes equally spaced on D diameter B.C. F. Abrasive shield flange O.D.
Table 2-7: Adapter Plate Dimensions for Rosemount Oxymitter DR with Abrasive Shield
Dimensions in in. (mm)
A 9.0 (229) 9.25 (235) 9.25 (235)
B: diameter 4.75 (121) 3.94 (100) 125 (4.92)
C: thread 0.625-11 (M-16 x 2) (M-20 x 2.5)
D: diameter B.C. 7.50 (191) 7.48 (190) 7.894 (200)
ANSI (PN 3536B58G02)
DIN (PN 3535B58G08) JIS (PN 3535B58G04)
26 Emerson.com/Rosemount
Loading...
+ 60 hidden pages