Rosemount Manual: Rosemount 3150 Series Nuclear Pressure Transmitters including the Rosemount 3152, 3153 and 3154 Manuals & Guides

Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Rosemount 3150 Series Nuclear Pressure Transmitters Incl uding the Rosemount 3152, 3153, and 3154
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BF
August 2019
ii
This page intentionally left blank
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Title Page
iii
NOTICE
Read this manual before working with the product. For personal and system safety, and for optimum
Chan.RNII-CustomerFeedback@Emerson.com
Rosemount 3150 Series Nuclear Pressure Transmitters Including the Rosemou nt 3152, 3153 and 3154
performance, make sure you thoroughly understand the contents of this manual before installing, using or maintaining this product.
For Assistance: Within the United States, contact Rosemount Nuclear Instruments, Inc. (Rosemount Nuclear) at
1-952-949-5210. Outside the United States, contact the nearest Rosemount representative. Customer Feedback: Your feedback is important to us, please send comments or suggestions to:
Rosemount Nuclear satisfies all obligations coming from legislation to harmonize product requirements in the European Union
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BF
August 2019
iv
Title Page
RETURN OF MATERIAL
Authorization for return is required from Rosemount Nuclear prior to shipment. Contact Rosemount
USA
IMPORTANT
Rosemount 3152, 3153 and 3154 Series Pressure Transmitters are designed for Nuclear Class 1E
IEEE Std 323TM-1974, -1983 and -2003
RCC-E-2002
IEEE Std 344TM-1975, -1987 and -2004
KTA 3505-2005
and liability for transmitters for which the foregoing has not been complied with by the user.
Rosemount Nucl ear Instruments, Inc. Warranty and Limitations of Remedy
The warranty and limitations of remedy applicable to this Rosemount equipment are as stated on the reverse side of the current Rosemount quotation and customer acknowledgment forms.
Nuclear (1-952-949-5210) for details on obtaining Return Material Authorization (RMA). Rosemount Nuclear will not accept any returned material without a Return Material Authorization. Material returned without authorization is subject to return to customer.
Material returned for repair, whether in or out of warranty, should be shipped prepaid to: Rosemount Nuclear Instruments, Inc.
8200 Market Boulevard Chanhassen, MN 55317
usage, and have been tested to the standards shown below:
These transmitters are manufactured under a quality system that meets the requirements of 10CFR50 Appendix B, 10CFR Part 21, ISO 9001, NQA-1, KTA 1401, KTA 3507, CSA N285.0, CSA Z299.1 and the applicable portions of IAEA-50-C-Q. During qualification testing, interfaces were defined between the transmitter and its environment that are essential to meeting requirements of the qualification standards listed above. Specifically, to ensure compliance with 10CFR Part 21, the transmitter must comply with the requirements herein and in the applicable Rosemount qualification report(s) throughout its installation, operation and maintenance. It is incumbent upon the user to ensure that Rosemount Nuclear’s component traceability program is continued throughout the life of the transmitter.
To maintain the qualified status of the transmitter, the essential environmental interfaces must not be compromised. Performance of any operations on the transmitter other than those specifically authorized in this manual have the potential for compromising an essential environmental interface.
Where the manual uses the terms requirement, mandatory, must or required, the instructions so referenced must be carefully followed. Rosemount Nuclear expressly disclaims all responsibility
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Title Page
v
Page
(Rev BE)
Page
(Rev BF)
Cover,
throughout
Cover,
throughout
10
10
Figure 2-3 updated to show correct polarity.
Figure 2-5 updated to show straight carbon ste el gus set plat e and that diamond or square hole patterns are acceptable alternatives.
54
54
Figure 5-2 updated to show correct polarity.
NOTE
complete comparison details.
NOTE
experience/usage.
Revision Status
Changes from February 2018 (Rev BE) to August 2019 (Rev BF)
Changes
Document revision change from February 2018 to August 20 19, Rev BE to Rev BF
16 16
The above Revision Status list summarizes the changes made. Please refer to both manuals for
Revision of the Reference Manual has no impact to form, fit, or function and does not impact transmitter qualification. Updates were made to provide clarity and improve customer
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BF
August 2019
vi
Title Page
This page intentionally left blank
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Contents
vii
SECTION 1
Using this Manual.....................................................................................
1
Introduction
SECTION 2
Overview...................................................................................................
3
Installation
Safety Messages......................................................................................
3
General Considerations............................................................................
4
Mechanical Considerations.......................................................................
4
Process Connections and Interfaces...........................................
5
Impulse Piping.............................................................................
7
Mounting Configurations..............................................................
8
Conduit Connections...................................................................
9
Electronics Housing.....................................................................
10
Electrical Considerations..........................................................................
10
Signal Integrity.............................................................................
13
Wiring Connections......................................................................
14
Installation Procedures.............................................................................
14
Mechanical – Transmitter............................................................
14
Mechanical – Conduit..................................................................
18
Electrical......................................................................................
18
SECTION 3
Overview...................................................................................................
23
Calibration
Safety Messages......................................................................................
23
Calibration Overview.................................................................................
24
Calibration Considerations...........................................................
24
Definitions....................................................................................
25
Span Adjustment Range..............................................................
27
Zero Adjustment Range...............................................................
27
Calibration Procedures.............................................................................
28
Span and Zero Adjustment..........................................................
28
Zero Based Calibration Procedure (LRV is Zero)...........
28
Elevated or Suppressed Zero Calibration Procedure.....
29
Coarse Zero Select Jumper Position Selection
Procedure.......................................................................
33
Damping Adjustment...................................................................
35
Correction for High Static Line Pressure.....................................
36
High Static Line Pressure Span Effect on Rage Codes
1, 2 and 3 DP Transmitters.............................................
36
High Static Line Pressure Span Correction for Range
Code 4 and 5 DP Transmitters.......................................
36
High Static Line Pressure Zero Correction for DP
Transmitters (All Ranges)...............................................
42
Linearity.......................................................................................
43
Table of Contents
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BF
August 2019
viii
Contents
SECTION 4
Overview...................................................................................................
45
Operation
Transmitter Theory of Operation...............................................................
46
The Sensor Cell........................................................................................
47
Demodulator.............................................................................................
48
Oscillator...................................................................................................
48
Voltage Regulator.....................................................................................
48
Current Control.........................................................................................
49
Current Limit.............................................................................................
49
Reverse Polarity Protection......................................................................
49
SECTION 5
Overview...................................................................................................
51
Maintenance and
Safety Messages......................................................................................
51
Troubleshooting
General Considerations............................................................................
52
Test Terminal............................................................................................
54
Electronics Assembly Checkout...............................................................
54
Sensor Module Checkout.........................................................................
55
Disassembly Procedure............................................................................
56
Process Flange Removal.............................................................
57
Electronics Housing Disassembly................................................
57
Reassembly Procedure............................................................................
59
Electronics Housing Reassembly................................................
59
Process Flange Reassembly.......................................................
63
Post Assembly Tests................................................................................
64
SECTION 6
Overview...................................................................................................
69
Transmitter Spare
Safety Messages......................................................................................
69
Parts
General Considerations............................................................................
71
Spare Parts Storage and Shelf Life..........................................................
71
Impact on Transmitter Qualified Life.........................................................
72
Transmitter Spare Parts List.....................................................................
73
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Contents
ix
This page intentionally left blank
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BF
August 2019
SECTION 1:
Introduction
USING THIS MANUAL
This manual is designed to assist in installing, operating and
sections:
Section 2: Installation
considerations.
Section 3: Calibration
Provides transmitter calibration procedures.
Section 4: Operation
Provides a description of how the transmitter operates.
Section 5: Maintenance and Troubleshooting
assembly tests.
Section 6: Transmitter Spare Pa rts
Provides order information for transmitter spare parts.
NOTE
dimensional drawings for each model.
Figure 1-1 shows the standard transmitter nameplate and Nameplate material is stainless steel.
Figure 1-1 – Standard Transmitter Nameplate
maintaining the Rosemount 3150 Series Pressure Transmitters. Instructions for the 3152, 3153 and 3154 models are included in this manual. Where differences in instructions between the models exist, they are noted with in those instructions. The manual is organized into the following
Provides general, mechanical, and electrical installation
Provides basic hardware troubleshooting considerations including disassembly and reassembly procedures and post
Refer to the applicable Rosemount Qualification/Test Reports, Product Data Sheets and/or Specification Drawing 03153­2003 and Specificat ion Dra w ing 03154-2003 (applicable to RCC qualified transmitters only) for details on testing, performance specifications, ordering information, and
where transmitter information is stamped onto the nameplate.
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Introduction
2
This page intentionally left blank
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BF
August 2019
SECTION 2:
Installation
Overview..................................................................
page 3
Safety Messages....................................................
page 3
General Considerations.........................................
page 4
Mechanical Considerations...................................
page 4
Electrical Considerations......................................
page 10
Installation Procedures..........................................
page 14
OVERVIEW
This section contains the following installation considerations:
Electrical
SAFETY MESSAGES
Procedures and instructions in this section may require special
performing an operation preceded by this symbol:
WARNING
Explosions can result in death or injury.
parameters.
General Considerations
Mechanical Considerations → Process Connections Impulse Piping Mounting Configurations Conduit Connections Electronics Housing
Electrical Considerations → Signal Integrity Wiring Connections
Installation Procedures → Mechanical
precautions to ensure the safety of the personnel performing the operation. Refer to the following safety messages before
Do not remove the transmitter covers in explosive
environments when the circuit is live.
Verify that the operating atmosphere of the transmitter
is consistent with the appropriate qualification
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Installation
4
WARNING
Electrical shock can result in death or serious injury.
the bracket.
GENERAL
Measurement accuracy depends upon proper installation of
vibration and shock during normal operation.
MECHANICAL
This section contains information you should consider when
qualification.
WARNING
Do not attempt to loosen or remove flange bolts while the transmitter is in service.
Avoid contact with the leads and terminals.
Process leaks could result in death or serious injury.
Install and tighten all four flange bolts before applying
pressure.
Do not attempt to loosen or remove flange bolts while
the transmitter is in service.
Replacement equipment or spare parts not approved by Rosemount Nuclear for use could reduce the pressure retaining capabilities of the transmitter and may render the instrument dangerous or adversely impact i ts qualified status.
Use only components supplied with the Rosemount
3152, 3153 or 3154 transmitter or designated by Rosemount Nuclear as spare parts for the 3152, 3153 or 3154.
Improper assembly of mounting bracke t to tra ditional process flange can damage sensor module.
For safe assembly of bracket to transmitter traditional
process flange, bolts must break back plane of flange web (i.e. bolt hole), but must not contact module housing. Use only the approved bolts supplied with
CONSIDERATIONS
CONSIDERATIONS
the transmitter and its associated impulse piping and valves. Mount the transmitter close to the process and use a minimum of piping to achieve best accuracy. For flow measurement, proper installation of the primary element is also critical to accuracy. Also, consider the need for easy access, personnel safety, practical field calibration and a suitable transmitter environment. Transmitter installation should minimize the effects of temperature gradients and fluctuations, and avoid
preparing to mount the transmitter. Read this section carefully before proceeding to the mechanical installation procedure. Proper installation is mandatory to assure seismic
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BF
August 2019
5
Installation
Mount the Rosemount 3150 Series transmitter to a rigid
effect caused by m ounti ng pos it ion.
NOTE
transmitter to cancel this effect per Section 3: Calibration.
Mount the process flanges with sufficient clearance for
draining, etc.).
Process Connections and
Process tubing must be installed to prevent any added
suitable thickness for the pressure involved.
support (i.e. one with a fundamental mechanical resonant frequency of 40 Hz or greater). Two mounting options are qualified for the transmitter: panel mount or 2-inch pipe mount. A stainless steel panel bracket is provided with the 3154. For the 3152 and 3153 transmitters, the user has the option of specifying either the stamped carbon steel panel bracket or the stainless steel panel bracket.
Refer to Figure 2-5 for qualified mounting configurations for both the panel and pipe mount options.
Orientation with respect to gravity is not critical to qualification. For maximum accuracy, zero the transmitter after installation to cancel any zero shift that may occur due to liquid head
The transmitter is calibrated in an upright position at the factory. Mounting the transmitter in another position may cause the zero point to shift by an amount equivalent to the internal liquid head within the sensor module induced by the varied mounting position. For maximum accuracy, zero the
Interfaces
process connections. For safety reasons, place the drain/vent valves so the process fluid is directed away from possible human contact when the vents are used. Also, consider that access to the vent/drain valve(s) and process connection(s) may be required for plant specific operations (i.e. calibration,
mechanical stress on the transmitter under seismic conditions. Use stress-relief loops in the process tubing or separately support the process tubing close to the transmitter.
Typical connections on the transmitter flanges are 1/4-18 NPT
®
and 1/4 inch or 3/8 inch Swagelok
. Use your plant-approved, qualified thread sealant when making threaded connections. The end-user is responsible for the qualification of the threaded seal interface on all 1/4-18 NPT interfaces.
Transmitters with flange options including 1/4 inch or 3/8 inch
®
Swagelok
are shipped with front ferrule, rear ferrule and nut. Place these fittings on the tubing with the orientation and relative position shown in Figure 2-1. Use process tubing with 1/4 inch or 3/8 inch outside diameter respectively, an d of
Reference Manual
August 2019
Rosemount 3150 Series
Installation
6
Figure 2-1 – Swagelok® Compression
Dimensions are nominal in inches (mm)
Assembly
Maintenance and Troubleshooting when closing.
00809-0100-4835 Rev BF
Fitting Detail
The Swagelok® tube fittings come completely assembled and are ready for immediate use. Do not disassemble them before use because dirt or foreign materials may get into the fitting
®
and cause leaks. Insert the tubing into the Swagelok
tube fitting, make sure the tubing rests firmly on the shoulder of the fitting and the nut is finger tight. Tigh ten the nut on e-and-one­quarter turns. Do not over-tighten.
Re-assembly
To reconnect, insert the tubing with pre-swaged ferrules into the fitting until the front ferrule sits in the fitting. Tighten the nut by hand, then rotate one-quarter turn more or to the originally secured position. Then tighten the nut slightly.
For more detailed information regarding the specifications and
®
use of Swagelok
tube fittings, refer to:
Fittings Catalog MS-01-140 “Gaugeable Tube Fittings and Adapter Fittings”
www.swagelok.com
Drain/Vent Valves
If drain/vent valves are opened to bleed process lines, torque drain/vent valve stems to the value in Table 5-2 in Section 5:
Rosemount 3150 Series
Reference Manual
August 2019
7
Installation
Impulse Piping
The piping between the process and the transmitter must
details).
For gas service, slope the impuls e piping at least 1
to be compatible with anticipated pressure.
00809-0100-4835 Rev BF
accurately transfer the pressure to obtain accurate measurements. There are five possible sources of error: pressure transfer (such as obstruction), leaks, friction loss (particularly if purging is used), trapped gas in a liquid line or liquid in a gas line and density variations between the legs.
The best location for the transmitter in relation to the process pipe depends on the process itself. Use the following guidelines to determine transmitter location and placement of impulse piping:
Keep impulse piping as short as possible.
For liquid service, slope the impulse piping at least 1
inch per foot (8 cm per meter) upward from the transmitter toward the process tap (see Figure 2-2 for
inch per foot (8 cm per meter) downward from the transmitter toward the process tap (see Figure 2-2 for details).
Avoid high points in liquid lines and low points in gas lines.
Make sure both impulse legs are the same temperature.
Use impulse piping of large enough diameter to avoid friction effects and blockage.
Vent all gas from liquid piping legs and internal to transmitter process flange.
Drain all liquid from gas piping legs and internal to transmitter process flange.
When using a sealing fluid, fill both piping legs to the same level.
When purging, make the purge connection close to the process taps and purge through equal lengths of the same size pipe – avoid purging through the transmitter.
Keep corrosive or hot process material out of direct contact with the transmitter.
Prevent sediment deposits in the impulse piping.
Keep the liquid balanced on both legs of the impulse
piping.
Avoid conditions that might allow process fluid to freeze within the process flange.
Make sure the impulse piping is of adequate strength
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Installation
8
Mounting Configuration
Refer to Figure 2-2 for examples of the following mounting
negligible.
NOTE
and refill the lines with water before resuming measurement.
NOTE
installation practices should be used.
NOTE
220°F (104°C).
configurations: Liquid Flow Measurement
Place taps to the side of the line to prevent sediment deposits on the process isolators.
Mount the transmitter beside or below the taps so gases vent into the process lines.
Gas Flow Measurement
Place taps in the top or side of the line.
Mount the transmitter beside or above the taps to
drain liquid into the process line.
Steam Flow Measurement
Place taps to the side of the line.
Mount the transmitter below the taps to ensure that
impulse piping will remain filled with condensate.
Fill impulse lines with water to prevent steam from contacting the transmitter directly and to ensure accurate measurement start-up. Condensate chambers are not typically necessary since the volumetric displacement of the transmitter is
For steam service, do not blow down impulse piping through the transmitter. Flush the lines with the transmitter isolated
The mounting configurations described above and depicted in Figure 2-2 are based on general industry “best practice” recommendations. Where applicable, specific plant approved
In steam or other elevated temperature services, it is important that temperatures at the process flanges not exceed 250°F (121°C). In vacuum service, these limits are reduced to
Rosemount 3150 Series
Reference Manual
August 2019
9
Installation
Figure 2-2 – Transmitter Installation
Please note that transmitters depicted in Figure 2-2 are intended for reference only.
Conduit Connections
The conduit connections to the transmitter are threaded.
conduit connection threads.
IMPORTANT
For all 3152 and 3153 transmitters, install the conduit plug
and Troubleshooting.
The 3154 has one conduit connection.
instructions or use the procedure in this section.
NOTE
Rosemount Nuclear for ordering information.
Examples (liquid, gas or steam)
00809-0100-4835 Rev BF
Options available are 1/2-14 NPT, M20 x 1.5, PG 13.5 and G1/2. Housings with M20 x 1.5 threads are marked with “M20 x 1.5”. Two openings are available on the 3152 and 3153 transmitter housings for convenient installation. Close off the unused opening with a compatible thread type stainless steel plug. Use your plant-approved, qualified thread sealant on the
(provided with the transmitter) in the unused conduit opening per the torque values in Table 5-2 in Section 5: Maintenance
Use a qualified conduit seal at the conduit entry to prevent moisture from entering the terminal side of the housing during accident conditions. Certain option codes provide a factory­installed, qualified electrical connector. To prevent excessive mechanical stress during seismic disturbances, support the conduit/connector cable near the transmitter. Install the conduit seal in accordance with the manufacturer’s
A Swagelok® rotatable two-piece 90 degre e elbow is available in two configurations: fully assembled to a 3150 Series pressure transmitter or in a ready to assemble kit (Spare Parts number 03152-0702-0001; see Manual Supplement 00809-0400-4835 for installation instructions). Please contact
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Installation
10
Electronics Housing
The standard transmitter orientation is shown in dimensional
housing.
ELECTRICAL
This section contains information you should consider when
computer).
Figure 2-3 – Typical Transmitter Wiring Connection
The power supply versus load limit relationship is shown in
ripple appears in the output signal.
drawings found in this manual (see Figure 2-6). While rotation of the electronics housing in the field is possible with special instructions, it is not recommended. Please contact Rosemount Nuclear prior to any attempt to rotate the electrical
CONSIDERATIONS
preparing to make electrical connections to the transmitter. Read this section carefully before proceeding to the electrical installation procedure.
Rosemount 3150 Series transmitters provide a 4-20 mA signal when connected to a suitable dc power source. Figure 2-3 illustrates a typical signal loop consisting of a transmitter, power supply, and various receivers (controller, indicator,
Figure 2-4. See qualification reports for additional details. The loop load is the sum of the resistance of the signal leads and the load resistance of the receivers. Any power supply
Rosemount 3150 Series
Reference Manual
August 2019
11
Installation
Figure 2-4 – Transmitter Supply Voltage vs. Load
Figure 2-4a – IEEE Qualified and Design 3154N models only)
POWER SUPPLY (VDC)
2500
1500
2000
500
1000
12
20
25
30
35
40
45
50
48
13.5
15
38
55
IEEE
DESIGN
LOAD (OHMS)
53
2150
1725
60
43
0
Regions (applicable to 3152N, 3153N and
00809-0100-4835 Rev BF
REGION
QUALIFIED
REGION
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Installation
12
Figure 2-4b – KTA Qualified and Design models only)
Figure 2-4c – RCC-E Qualified and Design models only)
2150
POWER SUPPLY (VDC)
2500
1500
500
12
1000
20
25
30
35
40
45
50
15
55
KTA
REGION
DESIGN
REGION
LOAD (OHMS)
53
60
43
0
1900
2000
45
43
POWER SUPPLY (VDC)
2500
2000
1500
500
12
1000
20
25
30
35
40
50
15 55
RCC
DESIGN
LOAD (OHMS)
53
2150
2050
60
0
38
750
Regions (applicable to 3152K and 3154K
QUALIFIED
Regions (applicable to 3153K and 3154K
REGION
QUALIFIED
REGION
Rosemount 3150 Series
Reference Manual
August 2019
13
Installation
Signal Integrity
Signal wiring need not be shielded, but twisted pairs yield the
ordering information.
best results. Shielded cable should be used for best results in electrically noisy environments. Do not run signal wiring in conduit or open trays with AC power wiring, or near heavy electrical equipment.
For installations with EMC performance requirements, consult the applicable EMC test reports for additional details regarding recommended practices for electrical wiring per various national and international codes and regulations.
The capacitance sensing element uses alternating current to generate a capacitance signal. This alternating current is developed in an oscillator circuit with a nominal frequency of 110 kHz +/- 11 kHz. This 110 kHz signal is capacitively­coupled to the transmitter case ground through the sensing element. Because of this coupling, a voltage may be imposed across the load, depending on choice of grounding.
This impressed voltage, which is seen as high frequency noise, has no effect on most instruments. Computers with short sampling times in a circuit where the negative transmitter terminal is grounded can detect a significant noise signal. Filter this signal out by using a large capacitor (1 uf) or a 110 kHz LC filter across the load. Signal loops grounded at any other point are negligibly affected by this noise and do not need filtering.
Signal wiring may be ungrounded (floating) or grounded at any one point in the signal loop.
The transmitter case may be grounded or ungrounded. Grounding should be completed in accordance with national and local electrical codes. Transmitter case can be grounded using either the internal or external ground connection.
00809-0100-4835 Rev BF
Internal Ground Connection: The Internal Ground Connection screw is inside the terminal side of the electronics housing (see Figure 2-8). The screw is
identified by a ground symbol ( ), and is standard on all 3150 Series transmitters.
External Ground Assembly: The External Ground location is indicated by the ground symbol ( ) on the
module (see Figure 2-6). An External Ground Assembly kit can be ordered as an option on the 3150 Series transmitter. This kit can also be ordered as a spare part. Please contact Rosemount Nuclear for
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Installation
14
Wiring Connections
The transmitter terminal block and ground screw terminals are
recommendations with proper tooling.
INSTALLATION
Installation consists of mounting the transmitter and connections. The procedures for each operation follow.
Mechanical – Transmitter
WARNING
Improper assembly of mounting bracket to transmitter
supplied with the bracket.
1. Attach the mounting bracket to the mounting location
.
Troubleshooting.
designed to accommodate wir e sizes from 24 AWG to 14 AWG. The screw terminals are also compatible with stud size #6 (M3.5) or #8 (M4) crimp terminals. Crimped connections shall be performed in accordance with manufacturers’
PROCEDURES
conduit/connector and making electrical and process
traditional process flange can damage sensor module. For safe assembly of bracket to traditional flange, bolts must
break back plane of flange web (i.e. bolt hole), but must not contact module housing. Use only the approved bolts
as follows:
Panel Mount Mount the bracket to a panel or other flat surface (for illustration see Figure 2-5). Please note that the bolts required for this step are customer supplied hardware. Based on qualification tests performed by Rosemount, the bolts listed in Table 2-1 are recommended for the bracket-to-customer interface. Torque each bolt to the value shown in
Table 5-2 in Section 5: Maintenance and Troubleshooting
Pipe Mount Assemble the bracket kit to a 2-inch pipe (for illustration see Figure 2-5). Torque each bolt to the value shown in Table 5-2 in Section 5: Maintenance and Troubleshooting.
2. Attach the transmitter to the mounting bracket (for illustration see Figure 2-5). Use the four 7/16-20 x 3/4-inch bolts with washers supplied with the transmitter. Torque each bolt to the value shown in Table 5-2 in Section 5: Maintenance and
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BF
August 2019
15
Installation
Table 2-1 – Recommended Bolts for Bracket-to-Customer Interface
Recommended Bolt for
Bracket-to-Customer Interface
0
No Bracket Supplied
N/A
1
(2)
Carbon Steel Panel Bracket
5/16-18 UNC 2A Grade 2
2
(3)
SST Panel Bracket
3/8-24 UNF 2A Grade 2
3
(4)
SST 2-inch Pipe Mount Bracket
2-inch pipe U-bolts provided
5
(2)
Carbon Steel Panel Bracket
5/16-18 UNC 2A Grade 2
7
(3)
SST Panel Bracket
3/8-24 UNF 2A Grade 2
8
(4)
SST 2-inch Pipe Mount Bracket
2-inch pipe U-bolts provided
(1) T he Bracket Code can be found in the 13th position of the 3152, 3153 and 3154 model strings.
bracket combined with the 2-inch pipe mount hardware (Rosemount P/N 01154-0044-0003)
Bracket Code
(2) Bracket Codes 1 and 5 are available on 3152 and 3153 transmitters only. (3) Bracket Codes 2 and 7 are available on 3152, 3153 and 3154 transmitters. (4) Bracket Codes 3 and 8 are available on 3152, 3153 and 3154 transmitters; this bracket code includes the listed SST panel
(1)
Bracket Type
Reference Manual
00809-0100-4835 Rev BF
August 2019
Rosemount 3150 Series
Installation
16
Figure 2-5 – Typical Transmitter Traditional Flange
Carbon Steel
Panel Mount
Stainless Steel
Panel Mount
Stainless Steel
2-Inch Pipe Mount
NOTE: All dimensions are nominal in inches (millimeters)
(not shown)
(3) A pipe-mount kit with three (3) U-bolts (not shown) is also available. The use of a third U-bolt is necessary to meet KTA seismic
Data Sheet (PDS) for ordering information.
Mounting Bracket Configuration,
(1) (2)
Bracket Options
3152
&
3153
(3)
3154
(1) T ransmitter and bracket orientation with respect to gravity will not impact qualification (2) T ransmitters can alternatively be mounted inside bracket (as shown below) or with process connection positioned adjacent to bracket
Carbon Steel Panel Mount Bracket is
not available for the 3154
requirements in installations where the KTA Airplane Crash (APC) value of 8g is applicable. Please consult the applicable Product
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BF
August 2019
17
Installation
Figure 2-6 – Transmitter Dimensional Drawings
Figure 2-6a – 3152, 3153 Traditional Flange
NOTE: All dimensions are nominal in inches (millimeters)
Figure 2-6b – 3154 Traditional Flange
NOTE: All dimensions are nominal in inches (millimeters)
Loading...
+ 58 hidden pages