Rosemount Manual: RAMAN Process Analyzer-Rev 1.0 | Rosemount Manuals & Guides

RAMAN PROCESS ANALYZER
Instruction Bulletin IB-103-300 Rev 1.0
HIGHLIGHTS OF CHANGES
Effective May, 2001 Rev. 1.0
PAGE SUMMARY
Throughout Added information as needed to meet CE mark certification.
5-1 through 5-5 Added Analog Output section to manual.
ROSEMOUNT WARRANTY
The Raman Process Analyzer is designed for industrial applications. Treat with care to avoid physical damage. THE WARRANTY DOES NOT COVER DAMAGE FROM MISHANDLING.
Rosemount warrants that the equipment manufactured and sold by it will, upon shipment, be free of defects in workmanship or material. Should any failure to conform to this warranty become apparent during a period of one year after the date of shipment, Rosemount shall, upon prompt written notice from the purchaser, correct such nonconformity by repair or replacement, F.O.B. factory of the defective part or parts. Correction in the manner provided above shall constitute a fulfillment of all liabilities of Rosemount with respect to the quality of the equipment.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES OF QUALITY WHETHER WRITTEN, ORAL, OR IMPLIED (INCLUDING ANY WARRANTY OF MERCHANTABILITY OF FITNESS FOR PURPOSE).
The remedy(ies) provided above shall be purchaser’s sole remedy(ies) for any failure of Rosemount to comply with the warranty provisions, whether claims by the purchaser are based in contract or in tort (including negligence).
Rosemount does not warrant equipment against normal deterioration due to environment. Factors such as corrosive gases and solid particulates can be detrimental and can create the need for repair or replacement as part of normal wear and tear during the warranty period.
Equipment supplied by Rosemount Analytical Inc. but not manufactured by it will be subject to the same warranty as is extended to Rosemount by the original manufacturer.
At the time of installation it is important that the required services are supplied to the system. This will ensure, that should there be a delay between installation and full commissioning that the analyzer being supplied with ac power will not be subjected to component deterioration.
IB-103-300
PURPOSE
The purpose of this manual is to provide a comprehensive understanding of the Raman Process Analyzer
components, functions, installation, and maintenance.
This manual is designed to provide information about the Raman Process Analyzer. We recommend that
you thoroughly familiarize yourself with the Description and Installation sections before installing your analyzer.
The description presents the basic principles of the analyzer along with its performance characteristics and components. The remaining sections contain detailed procedures and information necessary to install and service the analyzer.
Before contacting Rosemount concerning any questions, first consult this manual. It describes most situations encountered in your equipment’s operation and details necessary action.
DEFINITIONS
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an essential operating procedure, condition, or statement.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of
equipment, or loss of effectiveness.
NOTE
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, and is not related to the illustration in any technical manner.
IB-103-300
ii
IMPORTANT
LASER SAFETY INSTRUCTIONS
The Raman Process Analyzer uses a Class IV Diode Laser with a maximum output of 1 Watt at a wave­length of 810 nm (invisible, near-infrared light). The light is transmitted through fiber optic cables to and from the probe. The Raman Process Analyzer contains a Laser Safety Device that detects the return inten­sity of the laser light. If the return light diminishes below a certain level, the device reduces the laser power to Class I. This safety feature reduces the chance of personnel injury from contact with the laser. Take the following precautions when working around the Raman Process Analyzer:
Invisible laser light — avoid exposure to operating laser. A Class IV laser is used in this analyzer. The laser has a maximum output of 1 Watt at a wavelength of 810 nm.
1. Only service this product if you have completed formal training in laser safety and safe servicing techniques on this unit.
2. Never look at the probe tip when the probe is connected to the analyzer. Disconnect the fiber optic cables before examining the probe.
3. Always clean the probe tip when removed from process. It may be possible for liq­uid droplets or particles to attach to the tip of the probe, allowing light “reflection” to return to the Laser Safety Device. Use only appropriate material that will not scratch or otherwise damage the probe tip.
4. Do not bring the probe tip in contact with any surfaces when removed from the pro­cess. Any “scattered” light returned by the probe may trigger the laser back to Class IV.
5. All personnel working in the area of the laser must wear laser safety goggles (full­goggle type with side shields). Goggles must have a minimum attenuation factor of 100,000 (optical density 5) for an 810 nm wavelength.
IB-103-300
iii
IMPORTANT
SAFETY INSTRUCTIONS FOR THE WIRING AND
INSTALLATION OF THE LASER
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non-EU states should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external, where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be made through a circuit breaker which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker may also include a mechanically operated isolating switch. If not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard such as IEC947 (properly grounded three wire source of electrical power). All wiring must conform with any local standards.
7. Warning - Electrical Shock Hazard. Where equipment or covers are marked with the symbol to the right, hazardous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment — and then only by trained service personnel.
8. Caution - Hot Surface Hazard. Where equipment or covers are marked with the symbol to the right, there is a danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Certain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to the Operator Manual for instructions.
10. Warning - Laser Light Warning. Where equipment or covers are marked with the symbol to the right, high powered laser light is beneath. These covers should only be removed by trained service personnel when power is removed from the equipment.
11. All graphical symbols used in this product are from one or more of the following standards: EN61010-1, IEC417, and ISO3864.
IB-103-300
iv
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING
Rosemount Analytical designs, manufactures, and tests all its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property damage, damage to the instrument, and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product. If this
Instruction Bulletin is not the correct manual, telephone 1-800-654-7768 and the re­quired manual will be provided. Save this instruction manual for future reference.
If you do not understand any of the instructions, contact your Rosemount represen-
tative for clarification. Refer to the technical support hotline on page ix.
Follow all warnings, cautions, and instructions marked on and supplied with the
product.
Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
!
Install your equipment as specified in the installation instructions of the appropriate
Instruction Bulletin and per applicable local and national codes. Connect all prod­ucts to the proper electrical and pressure sources.
This product must only be used in the manner prescribed by Rosemount. To ensure
proper performance, use qualified personnel to install, operate, update, program, and maintain this product.
The environmental conditions in which this equipment is designed to operate is
within an ambient operating temperature of 32° to 122°F (0° to 50°C) at 20 to 100% relative humidity.
There are no operator-replaceable parts.When replacement parts are required, en-
sure that the qualified people use replacement parts specified by Rosemount. Un­authorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitutions may result in fire, electrical hazards, or improper operation.
To prevent electrical shock and personal injury, ensure that all equipment doors are
closed and protective covers are in place. Only qualified persons are authorized to open equipment doors and remove protective covers for equipment service or maintenance.
IB-103-300
v
WHAT YOU NEED TO KNOW
BEFORE INSTALLING AND WIRING A ROSEMOUNT
RAMAN PROCESS ANALYZER
1. What is the line voltage being supplied to the Raman Process Analyzer?
Write the line voltage here __________ .
2. Is the analyzer being controlled at the analyzer cabinet or by network connection?
3. Is the conduit set-up and all wiring for the analyzer run?
CAN YOU USE THE FOLLOWING
QUICK START GUIDE?
Use the Quick Start Guide if....
1. You are familiar with the Raman Process Analyzer installation requirements.
2. All wiring and conduit are in place for installation of the analyzer.
3. A configuration set exists for the process to be analyzed.
4. You are familiar with the SURE calibration procedures.
If you cannot use the Quick Start Guide, turn to Section II, Installation, in this Instruction Bulletin.
IB-103-300
vi
QUICK START GUIDE
RAMAN PROCESS ANALYZER
Before using the Quick Start Guide, please read “WHAT YOU NEED TO KNOW BEFORE INSTALLING AND WIRING A ROSEMOUNT RAMAN PROCESS ANALYZER” on the preced­ing page.
1. Mount the analyzer cabinet. Refer to Section II, paragraph 2-3.a.
2. Install the Raman probes. Refer to Section II, paragraph 0.
3. Install the fiber optic cable conduits. Refer to Section II, paragraph 0.
4. Connect the fiber optic cables. Refer to Section II, paragraph 1-1.a.
5. Connect the cooling air supply and filters to the analyzer cabinet. Refer to Section
II, paragraph 2-3.e.
6. Connect line voltage to the analyzer cabinet. Refer to Section II, paragraph 2-4.
7. Connect the network cable or computer peripherals to the analyzer cabinet. Refer
to Section II, paragraphs 0 and 2-6.
8. Turn power on to the analyzer cabinet.
9. If using a network connection to control the analyzer, use the NetSupport software
to enable communication with the analyzer. Refer to Section III, paragraph 3-1.
10. Use MAINCFG to select the desired configuration set. Refer to Section III, para-
graph 3-3.a.
11. Perform a dark scan and photometric calibration procedure. Refer to Section III,
paragraph 3-3.b.
12. Select the QUIT option to start the MAIN program and monitor the process.
IB-103-300
vii
QUICK REFERENCE GUIDE
RAMAN PROCESS ANALYZER SOFTWARE
Select a configuration set
1. Close the MAIN program.
2. Start the MAINCFG program.
3. Select SYSTEM CONFIGURATION.
4. Select ENABLE A CONFIGURATION.
5. Select the configuration file to be used.
Calibrate the analyzer
1. Use the MAINCFG program and load the correct configuration set for the process
to be monitored.
2. Remove the probe from the process and ensure the probe is clean of any liquid or
foreign material.
3. Mount the probe in the SURE calibration kit.
4. Select PHOTOMETRIC CALIBRATION from the MAIN Setup Menu screen.
5. A prompt to verify that the probe is connected to the calibration kit is displayed.
Click the CONTINUE button.
6. Select the appropriate channel to be calibrated. Set the number of scans to 1 and
click CONTINUE.
7. The system auto-ranges and acquires the dark scan current, then displays the pho-
tometric curve. Make sure the curve is not saturated (have a flat region). If flat re­gions appear in the scan, back out the calibration kit two or three turns and press RE-SAMPLE. Repeat the calibration kit adjustments and re-sampling until a smooth photometric curve is measured.
8. Make sure the appropriate channel is selected and set the number of scans to 10.
Press RE-SAMPLE, then press CONTINUE.
9. Press STORE to save this scan for future use. The selected channel photometrics
are calibrated.
10. To calibrate the photometrics of another channel, repeat steps 1 through 9.
IB-103-300
viii
Technical Support Hotline:
For assistance with technical problems, please call the Customer Support Center (CSC).
Phone: 1-800-433-6076
In addition to the CSC, you may also contact Emerson Process Management North American Response Center (NARC). North American Response Center coordinates Rosemount’s field service throughout the US and abroad.
Phone: 1-800-654-RSMT (1-800-654-7768)
Rosemount may also be reached via the Internet through e-mail and the World Wide Web:
E-mail: GAS.CSC@emersonprocess.com
World Wide Web: www.processanalytic.com
Manufacturer’s Address
Rosemount Analytical Inc. 1201 N. Main Street Orrville, Ohio 44667
IB-103-300
ix/x
TABLE OF CONTENTS
Section Page
Rosemount Warranty ................................................................................................................................... i
Section I. System Overview....................................................................................................................1-1
1-1. Scope Of Manual...........................................................................................................1-1
1-2. Components Checklist (Package Contents)...................................................................1-1
1-3. Functional Equipment Description................................................................................ 1-1
1-4. Equipment Specifications............................................................................................... 1-5
Section II. Installation .............................................................................................................................2-1
2-1. Overview ......................................................................................................................... 2-1
2-2. Typical Installation......................................................................................................... 2-1
2-3. Mechanical Installation ..................................................................................................2-2
2-4. Electrical Installation...................................................................................................... 2-5
2-5. Computer Network.........................................................................................................2-6
2-6. Computer Peripherals..................................................................................................... 2-6
2-7. Installation Inspections...................................................................................................2-7
Section III. Setup...................................................................................................................................... 3-1
3-1. Communication Software............................................................................................... 3-1
3-2. MAINCFG Program....................................................................................................... 3-3
3-3. Setup................................................................................................................................ 3-8
Section IV. Operation .............................................................................................................................. 4-1
4-1. Overview ......................................................................................................................... 4-1
4-2. Software Usage Conventions ........................................................................................4-1
4-3. MAIN Program Startup................................................................................................. 4-3
Section V. ANALOG OUTPUT .............................................................................................................5-1
5-1. Introduction ..................................................................................................................... 5-1
5-2. Analog Output Scaling Setup ....................................................................................... 5-1
5-3. Access The Scaling Setup Screen................................................................................5-1
5-4. Save (or Replace) Configuration..................................................................................5-3
5-5. How Scaling is Applied................................................................................................... 5-3
5-6. Analog Output Diagnostic Screen................................................................................ 5-4
5-7. Accessing the Diagnostic Screen..................................................................................5-4
5-8. Concentration Values ......................................................................................................5-5
5-9. Outputs ............................................................................................................................5-5
Section VI. Preventive Maintenance .....................................................................................................6-1
6-1. General ............................................................................................................................6-1
6-2. Cleaning ..........................................................................................................................6-1
Section VII. Troubleshooting.................................................................................................................. 7-1
7-1. General ............................................................................................................................7-1
7-2. Alarm Messages .............................................................................................................7-1
7-3. Troubleshooting ..............................................................................................................7-1
Section VIII. Service ................................................................................................................................8-1
8-1. General ............................................................................................................................8-1
8-2. Analyzer Component Replacement............................................................................... 8-1
8-3. Analyzer Calibration ......................................................................................................8-6
IB-103-300
xi/xiv
TABLE OF CONTENTS (Continued)
Section Page
Section IX. Replacement Parts...............................................................................................................9-1
Section X. Returning Equipment to the Factory .............................................................................10-1
10-1. Returning Equipment....................................................................................................10-1
Glossary..........................................................................................................................................................1
Index............................................................................................................................................................... 1
LIST OF ILLUSTRATIONS
Figure Page
Figure 1-1. Typical Raman Process Analyzer System Package..............................................................1-1
Figure 1-2. Functional Equipment Diagram............................................................................................1-2
Figure 1-3. Analyzer Cabinet..................................................................................................................1-4
Figure 2-1. Typical Installation...............................................................................................................2-1
Figure 2-2. Analyzer Cabinet Installation...............................................................................................2-2
Figure 2-3. Raman Probe ........................................................................................................................2-2
Figure 2-4. Cable Pull Box .....................................................................................................................2-3
Figure 2-5. Cable to Pull Tape Connection.............................................................................................2-3
Figure 2-6. Probe Connections................................................................................................................2-4
Figure 2-7. Air Supply Hookup ..............................................................................................................2-5
Figure 2-8. Conduit Ports........................................................................................................................2-5
Figure 2-9. Input Power Terminal...........................................................................................................2-6
Figure 2-10. Computer Connections.........................................................................................................2-6
Figure 3-1. NETSUPPORT CONTROL Screen ...................................................................................3-1
Figure 3-2. KNOWN CLIENTS Screen ...............................................................................................3-1
Figure 3-3. MAINCFG Program Structure .............................................................................................3-2
Figure 3-4. SETUP MENUS Screen.....................................................................................................3-3
Figure 3-5. SYSTEM CONFIGURATION MENU Screen ................................................................3-3
Figure 3-6. SYSTEM CONFIGURATION Screen...............................................................................3-3
Figure 3-7. PLS SETUP Screen ............................................................................................................3-4
Figure 3-8. SYSTEM CONFIGURATION Screen...............................................................................3-4
Figure 3-9. CCD MAP Screen ...............................................................................................................3-5
Figure 3-10. CONFIGURE MODBUS Screen ........................................................................................3-5
Figure 3-11. SYSTEM CONFIGURATION MENU Screen .................................................................3-6
Figure 3-12. SETUP MENUS Screen......................................................................................................3-6
Figure 3-13. Calibration Message.............................................................................................................3-6
Figure 3-14. PHOTOMETRIC CALIBRATION Screen........................................................................3-7
Figure 3-15. DARK CURRENT CALIBRATION Screen ....................................................................3-8
Figure 4-1. MAIN Program Structure.....................................................................................................4-2
Figure 4-2. RAMAN PROCESS ANALYZER Screen........................................................................4-3
Figure 4-3. PASSWORD VERIFICATION Screen..............................................................................4-3
Figure 4-4. OPERATOR MENUS Screen ............................................................................................4-4
Figure 4-5. SETUP MENUS Screen.....................................................................................................4-4
Figure 4-6. ZERO CLIP OPTION Screen............................................................................................4-4
Figure 4-7. RAMAN ANALYZER DATALOG UTILITY Screen....................................................4-5
Figure 4-8. OPERATOR MENUS Screen ............................................................................................4-5
IB-103-300
xii
LIST OF ILLUSTRATIONS (Continued)
Figure Page
Figure 4-9. GRAB SCAN MENUS Screen .........................................................................................4-5
Figure 4-10. GRAB SCAN Screen .......................................................................................................... 4-6
Figure 4-11. SAVE AS Screen ...............................................................................................................4-6
Figure 4-12. OPERATOR MENUS Screen.............................................................................................4-7
Figure 4-13. DIAGNOSTIC MENUS Screen..........................................................................................4-7
Figure 4-14. VIEW CALCULATIONS Screen.......................................................................................4-7
Figure 4-15. DIAGNOSTIC MENUS Screen..........................................................................................4-8
Figure 4-16. ANALOG INPUT MEASUREMENTS Screen.................................................................4-9
Figure 4-17. DIGITAL I/O STATE Screen............................................................................................ 4-9
Figure 4-18. SMART DIAGNOSTICS MENU Screen........................................................................ 4-10
Figure 4-19. ANALYZER SOFTWARE STATUS Screen.................................................................. 4-10
Figure 4-20. ANALYZER ALARMS STATUS Screen....................................................................... 4-11
Figure 4-21. PASSWORD VERIFICATION Screen ............................................................................4-12
Figure 4-22. TRENDS Screen ................................................................................................................4-12
Figure 4-23. SMART DIAGNOSTICS MENU Screen....................................................................... 4-13
Figure 4-24. PERFORMANCE MEASURES Screen ...........................................................................4-13
Figure 4-25. DARK CURRENT & PHOTOMETRIC CORRECTION Screen .................................4-14
Figure 4-26. CALIBRATION PATH & MODEL Screen ..................................................................4-14
Figure 4-27. DIAGNOSTICS MENUS Screen ....................................................................................4-15
Figure 4-28. OPERATOR MENUS Screen...........................................................................................4-15
Figure 4-29. SET SECURITY Screen ...................................................................................................4-16
Figure 4-30. SET PASSWORDS Screen...............................................................................................4-16
Figure 4-31 GRAB DARK SCAN Screen.......................................................................................... 4-16
Figure 4-32. ABOUT Screen .................................................................................................................. 4-16
Figure 5-1. RAMAN PROCESS ANALYZER Screen ........................................................................5-1
Figure 5-2. PASSWORD VERIFICATION Screen ..............................................................................5-2
Figure 5-3. OPERATOR MENUS Screen.............................................................................................5-2
Figure 5-4. SETUP MENUS Screen...................................................................................................... 5-2
Figure 5-5. ANALOG OUTPUT SCALING SETUP Screen ............................................................... 5-2
Figure 5-6. SAVE Prompt ......................................................................................................................5-3
Figure 5-7. DIAGNOSTIC MENUS Screen..........................................................................................5-4
Figure 5-8. ANALOG OUTPUT DIAGNOSTIC Screen ..................................................................... 5-4
Figure 7-1. Raman Troubleshooting Flowchart #1.................................................................................7-2
Figure 7-2. Raman Troubleshooting Flowchart #2.................................................................................7-3
Figure 7-3. Raman Troubleshooting Flowchart #3.................................................................................7-4
Figure 8-1. Fuse Locations .....................................................................................................................8-2
Figure 8-2. CCD Camera Thermocouple Removal.................................................................................8-2
Figure 8-3. Laser Cooling Fan Removal.................................................................................................8-3
Figure 8-4. Laser Removal .....................................................................................................................8-4
LIST OF TABLES
Table Page
Table 1-1. Raman Process Equipment Specifications (Analyzer)*...................................................... 1-5
Table 9-1. Replacement Parts for the Raman Process Analyzer ...........................................................9-1
IB-103-300
xiii/xiv
1
SECTION I. SYSTEM OVERVIEW
1-1. SCOPE OF MANUAL
This instruction bulletin covers installation, setup, operation, troubleshooting, and maintenance of the Raman Process Analyzer manufactured by Rose­mount Analytical. The troubleshooting and service procedures in this manual are limited to those that can be performed by most equipment operators.
1-2. COMPONENTS CHECKLIST
(PACKAGE CONTENTS)
The Raman Process Analyzer system includes the items listed in Figure 1-1. The SURE Calibration Kit contains all the necessary components for quickly calibrating the analyzer.
1-3. FUNCTIONAL EQUIPMENT DESCRIPTION
When properly configured, the analyzer detects and measures the concentration of sample components that inelastically scatter light. Access for the detec­tion of the process components is accomplished with one to four analyzer probes installed in or around the process flow.
Functional components of the analyzer are shown in the electronic diagram of Figure 1-2. The important components of the diagram include the process probes, the diode lasers, the laser safety device, spectrograph, vortex cooler, CCD camera, and the industrial computer. Brief descriptions of the ana­lyzer component functions are provided in the fol­lowing paragraphs.
1. Analyzer
2. Analyzer Test Records
3. Instruction Bulletin
4. Backup Software
5. SURE Calibration Kit
6. Oil Filter
7. Water Filter
8. In Situ Probe (1 per channel)
9. Fiber Optic Cables (3 per probe)
Figure 1-1. Typical Raman Process Analyzer System Package
IB-103-300
1-1
PRESSURE
REGULATOR
AIR
SUPPLY
OIL
FILTER
VORTEX
COOLING
CHAMBER
AIR
VALV E
HOT
AIR
USER
INTERFACE
ANALYZER CABINET
INDUSTRIAL
COMPUTER
LASER
CONTROL
MODULE
WATER FILTER
CONTROLLER
PROCESS PROBES
LASER
SAFETY
BOARD
DIAMOND
REFERENCE 1
FILTER
CCD
CAMERA
1
OPTIC
SPECTROGRAPH
DIODE
LASER 1
SPLITTER
OPTIC
FILTER
DIODE
LASER 2
DIAMOND
REFERENCE 2
CCD
CAMERA
HEAD
SPLITTER
SPLITTER
2
3
4
Figure 1-2. Functional Equipment Diagram
IB-103-300
1-2
SPLITTER
27320024
The following are brief descriptions of analyzer com­ponent functions. A listing of components in the analyzer cabinet is provided in Figure 1-3.
a. Diode Lasers and Probes. The analyzer uses
one or two diode lasers and up to four process probes. Each diode laser generates a light beam. The light beam is split in a 10/90 optic splitter. Fiber optic cables transmit ten percent of the beam to a diamond reference circuit and the re­maining ninety percent to a 50/50 optic splitter. A fiber optic cable from the diamond reference transmits the reference beam to the spectrograph.
The ninety percent portion of the beam is evenly split to provide excitation energy for two process probes. Each beam is filtered at the probe to eliminate stray radiation. The probe emits the filtered laser beam into the process flow.
The process components inelastically scatter dis­tinct wavelengths. Two return cables transmit scattered light from the probe to the analyzer. The unfiltered S cable transmits all scattered light to the Laser Safety Device. The filtered R cable transmits only a specific range of near­infrared wavelengths to the spectrograph.
b. Laser Safety Device. To minimize operator
hazards, the laser safety device monitors the in­tensity of the scattered laser light from the proc­ess flow. If the safety device receives less than an acceptably large part of the scattered light, it reduces the laser power output from Class IV to Class I.
The laser safety device continues to regulate the output of the diode lasers at pulsed Class I power levels until the device detects an acceptable amount of scattered light from the process.
Low light scatter may indicate a broken fiber op­tic cable, a bad cable connection, or removal of the probe from the process. Refer to the “Laser Safety Instructions” in the front of this manual for more information.
c. Vortex Cooler. The vortex cooler helps control
the temperature of the CCD (charge coupled de­vice) camera and the cabinet interior. The main function of the vortex cooler is to maintain the
ambient temperature near the camera at 30°C so that the internal CCD chip temperature will re­main at –30°C.
Incoming instrument air is dried, filtered, and routed through a temperature-controlled air valve to the vortex cooler. The vortex effect separates the forced air into warm and cold air flows. The chilled air is routed to the CCD camera.
As the cabinet air gets warmer, the air valve opens a larger flow of forced air to the vortex chamber, providing more cooling air to the cam­era. As the cabinet air cools, a proportional air valve partly closes to slow the flow of cooling air.
d. Spectrograph and CCD Camera. Filtered light
from the diamond reference and the R probe is transmitted by fiber optic cable to the spectro­graph. The spectrograph separates incoming light into distinct wavelengths. Each wavelength of light is sent to the CCD camera head where the spectral image is decoded and electronically trans­ferred to the CCD camera controller. The con­troller processes the spectral image into numerical data. The data is sent to the industrial computer for interpretation and display.
e. Industrial Computer. The industrial computer
converts data points into light intensity with re­spect to frequency. The histogram of each pre­dicted component is displayed on the CRT. The spectra are processed by the prediction’s algo­rithm in accordance with the calibration file. Each application requires its own calibration file. The output data is stored for further analysis.
f. User Interface. An interface device is required
to set up and operate the analyzer using the MAIN and MAINCFG software programs. The user interface device can be a remote computer
connected through a network modem, Ethernet, RF antenna and laptop computer, or a monitor, keyboard, and mouse connected at the analyzer cabinet.
g. Analyzer Outputs. The available analyzer out-
puts are MODBUS and/or analog, 4 to 20 mA.
h. SURE Calibration Kit. The SURE calibration
kit is a light proof housing that secures the probe tip above a fluorescent glass. The kit uses no external light source or power source. When the fluorescent glass and probe are properly assem­bled in the light-proof housing, the analyzer can be calibrated using the MAINCFG software pro­gram.
IB-103-300
1-3
1. Hoffman Enclosure 10. Temperature Controller
2. Computer Power Supply 11. Enclosure Backplate
3. Computer 12. CCD Camera Controller
4. Disk Drives 13. Spectrograph Assembly
5. Optics Assembly Tray 14. Laser Pedestal Assembly
6. CCD Camera 15. Laser Safety Device
7. CCD Camera Hood 16. Proportional Air Valve
8. Termination Board 17. Raman Power Supply
9. Peripheral Device Connector 18. Vortex Cooler Assembly
Figure 1-3. Analyzer Cabinet
IB-103-300
1-4
1-4. EQUIPMENT SPECIFICATIONS
Table 1-1. Raman Process Equipment Specifications (Analyzer)*
Spectral Range ........................................................................................................... 500 cm
-1
to 2000 cm-1 Raman Shift
Spectral Resolution .................................................................................................................................................. 2 nm
Typical Error ...................................................................................... Less than 1% of full-scale, application dependent
Operating Temperature ............................................................................................................. 32° to 122°F (0° to 50°C)
Storage Temperature ................................................................................................................ 32° to 122°F (0° to 50°C)
Relative Humidity .......................................................................................................................................... 20 to 100%
Warm-up Time ................................................................................................................................................ 60 minutes
Input Power ........................................................................................................... 115 VAC, 60 Hz or 230 VAC, 50 Hz
Power Rating ......................................................................................................................................... 250 VA (250 W)
Electrical Classification ................................................................... General Purpose or Class I, Div II, Type “Z” Purge
Environmental Classification .......................................................................................................................... NEMA-4X
Dimensions (H x W x D) .................................................................................. 48 x 24 x 12 in. (1219 x 610 x 305 mm)
Weight ...................................................................................................................................................... 165 lbs (75 kg)
Certification ........................................................................................................................................................ CE Mark
*Probe specifications may vary from probe to probe. Refer to the material supplied with probe for specifications.
1-5. CERTIFICATIONS
Emerson Process Management has satisfied all obligations coming from the European legislation to harmo­nize the product requirements in Europe.
IB-103-300
1-5/1-6
2
SECTION II. INSTALLATION
2-2. TYPICAL INSTALLATION
Before starting to install this equipment, read the “Safety instructions for the wiring and installation of the laser” at the front of this Instruction Bulletin. Failure to follow the safety instructions could result in serious injury, death, or substantial property damage.
2-1. OVERVIEW
This section covers installation of the Rosemount Raman Process Analyzer. Included are a typical in­stallation, common requirements, what you need to have on hand to start an installation, and the me­chanical and electrical installation instructions. An authorized Rosemount service representative must complete the analyzer installation. You must com­plete the mechanical installation before calling the service representative for final check-out and instal­lation commissioning.
A typical system installation is shown in Figure 2-1. When properly installed, the analyzer cabinet is se­curely mounted to a wall in the process control area. The cabinet should not be mounted in the path of di­rect sunlight.
Process probes are installed in the process flow (usu­ally a sampling flow). Up to four process probes may be installed for each analyzer. The installed probes should be adequately shielded from accidental im­pact, rainfall, and direct sunlight. Ready access for probe removal, cleaning, and replacement is recom­mended.
Fiber optic cable lengths are per customer specifica­tion. Make sure the correct cables are selected for each run. One inch or larger diameter conduits or ca­ble troughs are recommended for shielding the fiber optic cables. Pull boxes are needed when successive cable bend angles exceed 180 degrees. Conduit ends should point down to protect against water accumu­lation. Sharp conduit edges must be avoided and the cables should be lubricated before pulling.
Figure 2-1. Typical Installation
IB-103-300
2-1
2-3. MECHANICAL INSTALLATION
a. Analyzer Cabinet.
Lift the cabinet with the upper brackets. Avoid contact with the vortex cooler. Position and se­curely mount the cabinet using appropriate screws and flat washers.
b. Raman Probe.
Never rest the cabinet standing up. Damage to conduit fittings can occur.
The analyzer is housed in a NEMA-4X cabinet. The NEMA-4X cabinet is suitable for wall mounting in a Class I, Division II environment when equipped with an ISA Type Z-Purge.
When lifting, the cabinet door must be closed and latched. Make sure the mounting brackets are installed and tight.
The analyzer cabinet weighs approximately 165 lbs (75 kg). Analyzer cabinet mounting dimen­sions are provided in Figure 2-2. Mark the posi­tion for cabinet hanger mounting on center with wall studs or a securely mounted plywood back­plate.
24.0
(609.6)
12.0 (304.8)
Each analyzer system includes up to four process probes designed for mounting in a flow line or tank. Install each process probe according to the following instructions:
1.
Remove the protective covers from the probe tip, Figure 2-3.
2. Install a bored through compression fitting
of compatible metallurgy in the process line, tee, or process tank wall.
3.
Insert the probe to the desired depth in the compression fitting.
4. If compatible with the process, apply a suit-
able sealant around the seal diameter of the probe.
5. Tighten the compression fitting to secure
the probe. Do not over-tighten the fitting.
48.0
(1219.2)
NOTE:
DIMENSIONS ARE IN INCHES WITH MILLIMETERS IN PARENTHESIS.
Figure 2-2. Analyzer Cabinet Installation
27320025
IB-103-300
Figure 2-3. Raman Probe
2-2
ENDPLATE
4. Overlay 3 ft (91.4 cm) of the fiber optic ca-
ble onto the pull tape and wrap with nylon string to hold in place. Wrap and tape the fiber optic cable to the pull tape as shown in Figure 2-5.
CABLE PULL
BOX (90 TURN
CONDUIT
FIBER OPTIC
CABLE
Figure 2-4. Cable Pull Box
c. Fiber Optic Cable Conduits.
Fiber optic cables are precision optics devices. Careless handling or installa­tion of a fiber optic cable can result in permanent cable damage.
1.
Install a 3/4 in. (19 mm) diameter minimum conduit from the analyzer cabinet to each process probe. The minimum radius for all conduit bends is 10 in. (254 mm). For fiber optic cable runs, it is not required to con­nect the conduit to the analyzer cabinet.
o
FITTING)
27320060
NYLON STRING
(WRAP TIGHTLY)
PULL TAPE
5.
Carefully pull the cable through the con­duit; pull enough cable to allow for cable slack when connecting at the analyzer cabi­net or probe.
6.
When using cable pull boxes, turn the coiled cable over. Feed the pull tape to the opposite side of the pull box. Wrap and tape the cable to the pull tape and pull the free end through the next section of conduit. Repeat the cable pulling instructions as needed to install all fiber optic cables.
FIBER OPTIC
CABLE
FEED STRING
THROUGH PULL
TAPE EYELET
3FT
(91.4 CM)
WRAP WITH ELECTRICAL
TAPE
2.
Install a cable pull box following a series of conduit bends totaling 180 degrees. A 90 degree turn fitting (Figure 2-4) is recom­mended for use as a cable pull box.
3.
When using pull boxes, coil the fiber optic cable into a figure eight below the pull box. Feed a cable pull tape from one conduit end to the pull box.
27320031
Figure 2-5. Cable to Pull Tape Connection
IB-103-300
2-3
Loading...
+ 63 hidden pages