Rosemount Manual: OxyBalance Oxygen Display and Averaging System-Rev 1.1 | Rosemount Manuals & Guides

Instruction Manual
IM-106-4050, Rev 1.1 July 2008
OxyBalance
Oxygen Display and Averaging System
http://www.raihome.com
HIGHLIGHTS OF CHANGES
Effective January 2007, Rev 1.0
Page Summary
Page 5-8 Added the procedure: Loading PLC Software. Page 5-9 Added the procedures: Loading HMI Software and Setting HMI Settings. Page 1-2 thru A-14 Added note 11 to the safety data section. Page B-1 Updated the return of materials address. Back Cover Updated the address block.
Page Summary
Page 3-4 Updated Figure 3-4. Added Fgure 3-5, and Probe Names description. Page 3-5 Updated Figure 3-6. Added description of the Failed Average condition. Page 3-7 Descriptions of Warning and Fault conditions corrected. Page 3-9 Description of Average Failed corrected. Appendix A Updated Safety Instructions.
Instruction Manual
IM-106-4050 Rev 1.1 July 2008
OxyBalance Oxygen Display
and Averaging System
Table of Contents
Essential Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
SECTION 1 Description
SECTION 2 Installation
SECTION 3 Setup
SECTION 4 Calibration
Component Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Part Numbers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
OxyBalance System Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Physical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
System Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Mounting, NEMA 4X Enclosure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mounting, Plate Mounted System . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Wiring Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Login. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
OxyBalance Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
SECTION 5 Maintenance and Service
SECTION 6 Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
PanelView Plus 600. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
PanelView Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
PanelView Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
MicroLogix 1200 Programmable Logic Controllers . . . . . . . . . . . . . . . 5-6
PLC Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
PLC Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Loading PLC Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Loading HMI Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Setting HMI Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
I/O Module Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
I/O Module Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Probe Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
TOC-1
OxyBalance Oxygen Display and Averaging System
Instruction Manual
IM-106-4050 Rev 1.1
July 2008
SECTION 7 Recommended Spare Parts
SECTION 8 Optional Accessories
APPENDIX A Safety Data
SECTION B Returning Equipment to the Factory
Spare Parts Listing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
PanelView and MicroLogix Accessories . . . . . . . . . . . . . . . . . . . . . . . 8-1
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Returning Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
TOC-2
Instruction Manual
IIM-106-4050 Rev 1.1 July 2008
OxyBalance Oxygen Display
and Averaging System
OxyBalance
Oxygen Display and Averaging System
READ THIS PAGE BEFORE PROCEEDING!

ESSENTIAL INSTRUCTIONS

Emerson Process Management designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Emerson’s Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson Process Management representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation,
operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson Process Management. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers
are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
http://www.raihome.com
The information contained in this document is subject to change without notice.
OxyBalance Oxygen Display
RISKOFELECTRICAL SHOCK
WARNING:REFER TOINSTRUCTIONMANUAL
PROTECTIVECONDUCT OR TERMINAL
EARTH(GROUND) TERMINAL
:
:
:
:
and Averaging System
If a Model 275/375 Universal HART® Communicator is used with this unit, the software within the Model 275/375 may require modification. If a software modification is required, please contact your local Emerson Process Management Service Group or National Response Center at 1-800-433-6076 or 1-888-433-6829.
Instruction Manual
IM-106-4050 Rev 1.1
July 2008

PREFACE

DEFINITIONS

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the OxyBalance Oxygen Display and Average System.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.

SYMBOLS

ii
NOTE
Highlights an essential operating procedure, condition, or statement.
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, and is not related to the illustration in any technical manner.
Instruction Manual
IM-106-4050 Rev 1.1 July 2008

Section 1 Description

Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-2
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-5
OxyBalance Oxygen Display
and Averaging System

COMPONENT CHECKLIST

PART NUMBERS

Table 1-1. OxyBalance System Part Numbers
Figure 1-1. Typical System Package
A typical Rosemount Analytical OxyBalance Oxygen Display and Averaging System package should contain the items shown in Figure 1-1. Your configuration may not include the NEMA 4X enclosure.
Use the part numbers listed in Table 1-1 to verify your OxyBalance system part number. Copy the part number and serial number from the OxyBalance unit into the chart on the back cover of this manual. Refer to this complete part number for any correspondence with Emerson Process Management.
Part Number Description
6A00203G01 1-4 probes, NEMA 4X enclosure (Figure 1-1) 6A00203G02 5-8 probes, NEMA 4X enclosure (Figure 1-1) 6A00203G03 1-4 probes, plate mounted (not shown) 6A00203G04 5-8 probes, plate mounted (not shown)
http://www.raihome.com
1. Instruction Manual
2. OxyBalance Oxygen Display and Averaging System with NEMA 4X Enclosure
OxyBalance Oxygen Display and Averaging System

SYSTEM OVERVIEW

Instruction Manual
IM-106-4050 Rev 1.1
July 2008

General

The OxyBalance is a display and averaging system that augments the Oxymitter O2 probe product line. It receives up to (8) 4-20 mA signals from Rosemount Analytical, Westinghouse, or other O2 probes, and calculates and transmits up to 4 additional programmable averages (again 4-20 mA). The OxyBalance system also displays the individual and averaged outputs on a color touch screen graphic panel. Trend screens, bar graphs (individual probes only), and alarming indications are provided when a probe goes off line.
The OxyBalance system replaces much of the functionality of older averaging systems offered by Westinghouse and Rosemount Analytical - the Control Room Electronics (CRE), 1500 Controller-based averaging systems, and the Veritrak-based 1290 systems. It should be noted, however, that the heater control and signal conditioning for each individual probe is executed within separate electronics for each probe. Unlike past averaging systems, and most current competitive averaging systems, there is no component failure in the OxyBalance system that will cause the loss of individual probe signals.
The OxyBalance system senses 4-20 mA inputs automatically. Averages are very easy to set up. Any probes that may fail will default to a 3.5 mA signal level, indicated in red, and the OxyBalance system will remove these probes from the average. Likewise, probes that are in calibration will indicate blue and provide a contact closure to the OxyBalance system to remove them from the average while calibrating.
Trends offer adjustable scaling, three time spans (5 min, 8 hrs, and 24 hrs), and panning capability back and forth in history.

OxyBalance System Features

The standard OxyBalance system includes the following features:
• Graphically displays and trends from 1 to 8 O2 measurements.
• Probe inputs may be from Rosemount Analytical, Westinghouse or any competitive O2 probe with a 4-20 mA signal output.
• Generates up to four 4-20 mA signals as programmable averages.
• Failed or calibrating probes are automatically removed from the average.
• Individual probe readings remain autonomous and are unaffected by any potential failure of the OxyBalance system.
• Variable time scale trends for each individual probe value as well as for up to 4 averages.
• Password protection for setup menus.
1-2
Instruction Manual
DCOMM
COMM0
FORCE
FAULT
RUN
POWER
OUT
47IN01235689
47
0123568910111213
1
0
4
3
2
0
1
0
1
3
4
3
4
22
7
6
5
5
6
7
5
6
7
38370016
PanelView
Plus600
MicroLogix1200
Programmable
LogicController
I/OModules
Input
Terminals
Output
Terminals
SignalPass
Through Terminals
Failure/
Warning
Terminals
Power
Terminals
InCal
Terminals
PLC
Failure
Terminals
IM-106-4050 Rev 1.1 July 2008
OxyBalance Oxygen Display
and Averaging System

Physical Description

Aside from the PanelView Plus 600, the OxyBalance system consists of additional components mounted on three DIN rails. The main components are shown in Figure 1-2.
PanelView Plus
The PanelView Plus 600 is the touch screen interface located on the front of the OxyBalance system. All probe readings and averages are displayed through this interface. There is also a power cord and communications cord that connect to this module. Plate mounted systems may be equipped with the optional PanelView Plus 1500 touch screen interface.
MicroLogix 1200 Programmable Logic Controller (PLC)
PLC
The PLC processes the raw data received by the OxyBalance. The processed data is transmitted to the PanelView Plus 600 where it is displayed on the touch screen interface.
I/O Modules
Also known as the expansion I/O these modules act as part of the controller system and are located on the right side of the PLC unit.
Input Terminals
This row of terminals receives the 4-20 mA signals from the O2 probes.
Figure 1-2. OxyBalance Components
1-3
OxyBalance Oxygen Display and Averaging System
Signal Pass Through Terminals
The pass through terminals are located on the input terminals. These terminals provide connection points for passing on the incoming O2 signal to other devices.
Output (Average) Terminals
These terminals provide a connection point for the average O2% 4-20 mA signals.
Failure / Warning Terminals
These terminals provide connection points for the output signal warning and failure contacts.
In Cal Terminals
These terminals provide connection points for In Cal signals from the individual oxygen probes. When a probe is in calibration it is removed from any average calculations.
Power Terminals
Instruction Manual
IM-106-4050 Rev 1.1
July 2008
This terminal provides wire contacts for power distribution in the OxyBalance system.
OxyBalance operating specifications are listed in Table 1-2.
1-4
Instruction Manual
IM-106-4050 Rev 1.1 July 2008
OxyBalance Oxygen Display
and Averaging System

SPECIFICATIONS

Specifications for the OxyBalance system are listed in Table 1-2.
Table 1-2. OxyBalance Specifications
SPECIFICATIONS
Ambient Environment
Temperature specification 23oF to 149o F (-5oC to 65oC) Ambient temperature effect on electronics Less than 0.01% of reading per 10oC Vibration IEC 68-2-6 and ISA S37.3 Shock IEC 68-2-31 and ISA S37.3 Enclosure (if included) NEMA 4X Area Classification General Purpose
Power Requirements 100-240 VAC, 50/60 Hz MicroLogix 1200
Resistive Load on Current Output 0 to 500 Ohms (includes wire resistance) Input Impedance for Current Terminal 275 Ohms
PanelView Plus 600
Minimum On-State Current 2.0 mA at 10V dc Nominal On-State Current 8.9 mA at 24V dc Maximum On-State Current 12.0 mA at 30V dc Current per Group Common 8A 240V ac (Maximum Volts) 2.5A 120V ac (Maximum Volts) 2.5A 125V dc (Maximum Volts) 1.0A (Amperes Continuous) 24V dc (Maximum Volts) 2.0A (Amperes Continuous)
I/O
Signal Inputs 4-20 mA (qty 1 to 8 O2 probes) Signal Outputs 4-20 mA (qty 1 to 8 pass through each probe, qty 4 programmable averages) Signal Output Resolution (for averages) 12 bit Signal Output Resistive Load Less than 500 ohms Discrete Inputs "IN CAL" (qty 2 to 8 from individual probes)
Relay Outputs
Architecture
Logic 4 programmable averages from 2-8 probes.
Individual probes removed from average if
Signal Security
Personnel Security Password protection configuration changes in programmable averages. Color Graphic Display
Size 6" diagonal (optional 15" diagonal for plate mounted units) Type Color Active Matrix, TFT LCD Resolution 320 by 240 min. Touch screen operator interface.
(1)
(2)
(Amperes Continuous)
(2)
(Amperes Continuous)
"LOSS OF PLC" "AVERAGE WARN" if one or more probes drop out of an average. "AVERAGE FAILED" for each average when only one probe in the average remains valid.
Each probe utilizes its own autonomous conditioning electronics including its own power supply. Any failure in the OxyBalance system will not affect the individual 4-20 mA signals going to the control room for each probe.
1) Probe fails (4-20 mA to default condition of 3.5 mA or 21 mA
2) Probe is in cal (SPS / IMPS contact must share with control room) 4-20 mA signals to be wired in series to the PLC and to the DCS I/O marshalling
panel such that the loss of the PLC will not affect the transmission of the individual probe signals to the DCS. If PLC power, I/O cards or processor card(s), are lost 4-20 mA signals from the probes are still transmitted to the control room.
(1)
Specifications are subject to change without notification. Our policy is one of continuous improvement and we reserve the right to change specifications.
(2)
1.5A above 40oC.
1-5
OxyBalance Oxygen Display and Averaging System
Instruction Manual
IM-106-4050 Rev 1.1
July 2008
1-6
Instruction Manual
IM-106-4050 Rev 1.1 July 2008

Section 2 Installation

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1
System Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1
Mounting, NEMA 4X Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Mounting, Plate Mounted System . . . . . . . . . . . . . . . . . . . . . . . . . page 2-3
Wiring Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-5
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-5
Before starting to install this equipment read the "Safety Instructions" in Appendix A: Safety Data. Failure to follow the safety instructions could result in serious injury or death.
OxyBalance Oxygen Display
and Averaging System

OVERVIEW

SYSTEM CONSIDERATIONS

Disconnect all power before installing or replacing components. Failure to disconnect power may result in electrical shock and/or damage to the terminal.
Install all protective equipment covers and safety ground leads after installation. Failure to secure covers and ground leads could result in serious injury or death.
This section describes:
1. Surface-mounting of the OxyBalance system.
2. The necessary wiring schematics for proper connection of the OxyBalance system.
3. Setup procedures conducted through the PanelView Plus 600 on how to configure, assign nomenclature, change/update display settings, and assign the O2 probes to the average outputs.
Prior to installation of the OxyBalance system check for all components necessary to install the system completely. When selecting a mounting location determine where the OxyBalance system will be placed in terms of serviceability, ambient temperatures, environmental considerations, and convenience.
http://www.raihome.com
OxyBalance Oxygen Display and Averaging System
Instruction Manual
IM-106-4050 Rev 1.1
July 2008

MOUNTING, NEMA 4X ENCLOSURE

Figure 2-1. NEMA 4X Enclosure Mounting Dimensions
The OxyBalance Oxygen Display and Averaging System was designed to be wall-mounted or simply placed on a table. Locate the unit where the ambient temperature is between 23oF to 149oF (-5oC to 65oC).
The outline drawing in Figure 2-1 shows mounting dimensions of the OxyBalance system when equipped with the NEMA 4X enclosure.
2-2
Instruction Manual
6.15
(156)
4.86
(123)
15.4
(391)
12.4
(305)
PanelViewPlus600 PanelViewPlus1500
Note: Alldimensionsare
inincheswithmillimeters inparentheses.
38370019
IM-106-4050 Rev 1.1 July 2008
OxyBalance Oxygen Display
and Averaging System

MOUNTING, PLATE MOUNTED SYSTEM

Locate the plate mounted OxyBalance Oxygen Display and Averaging System where the ambient temperature is between 23oF to 149oF (-5oC to 65oC).
An alternate plate mounting is recommended when the NEMA 4X enclosure is not included with the OxyBalance system. Figure 2-2 shows the backplane mounting dimensions for the OxyBalance system without the NEMA 4X enclosure.
The dimensions shown in Figure 2-3 are the cutout dimensions for mounting the PanelView Plus 600 or optional PanelView Plus 1500 touch screen in a customer-supplied equipment panel.
Figure 2-2. Backplane Mounting Dimensions
Figure 2-3. PanelView Plus Cutout Dimensions
2-3
OxyBalance Oxygen Display
DCOMM
COMM0
FORCE
FAULT
RUN
POWER
OUT
47IN01235689
47
0123568910111213
1
0
4
3
2
0
1
0
1
3
4
3
4
22
7
6
5
5
6
7
5
6
7
38370003
Average4-20mA
OutputSignals
Probe1
Probe2
Probe3
Probe4
Probe5
Probe6
Probe7
Probe8
Average1
Average2
Average3
Average4
Incoming4-20mA
O Signals
2
+-+-+-+-+-+-+-+
-
+-+-+-+
-
Line
Voltage
Line
Neutral
Ground
Probe1
InCal
Contacts
Probe2
Probe3
Probe4
Probe5
Probe6
Probe7
Probe8
PLCFailure
Avg1&2FailCommon
Avg1Fail
Avg2Fail
Avg3&4FailCommon
Avg3Fail
Avg4Fail
AvgWarningCommon
Avg1Warning
Avg2Warning
Avg3Warning
Avg4Warning
and Averaging System
Figure 2-4. Installation Wiring Schematic
Instruction Manual
IM-106-4050 Rev 1.1
July 2008
2-4
Instruction Manual
DCOMM
COMM0
FORCE
FAULT
RUN
POWER
OUT
47IN01235689
470123568910111213
1
0
4
3
2
010
1
3
4
3
4
22
7
6
5
5
6
7
5
6
7
38370015
4-20mA
Pass-Through
Terminals
4-20mA
Input
Terminals
IM-106-4050 Rev 1.1 July 2008
OxyBalance Oxygen Display
and Averaging System

WIRING CONNECTIONS

SETUP

Figure 2-5. Pass-Through Terminals
Refer to Figure 2-4 and connect the input and output signal wiring to the OxyBalance as follows:
1. Connect the incoming 4-20 mA signals to the input signal terminals.
2. Connect the 4-20 mA average O2 signals to the output signal terminals.
3. Connect the In Cal input signals to the in cal signal terminals.
4. Connect the warning and failure signal output wiring to the warning and failure signal terminals.
5. If the signal pass-through feature for the incoming O2 signals will be used remove the jumpers from the pass-through terminals. Connect the outgoing O2 signal wires to the raised output signal terminals. (Figure 2-5).
6. Connect 90-240 Vac, 50/60 Hz input power to the power input terminals.
For setup information refer to Section 3, Setup.
2-5
OxyBalance Oxygen Display and Averaging System
Instruction Manual
IM-106-4050 Rev 1.1
July 2008
2-6
Instruction Manual
IM-106-4050 Rev 1.1 July 2008

Section 3 Setup

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Before starting to install this equipment read the "Safety Instructions" in Appendix A: Safety Data. Failure to follow the safety instructions could result in serious injury or death.
Disconnect all power before installing or replacing components. Failure to disconnect power may result in electrical shock and/or damage to the equipment.
OxyBalance Oxygen Display
and Averaging System

OVERVIEW

SETUP

Table 3-1. Signal Interpretation
This section provides information on setup and configuration of the OxyBalance Oxygen Display and Averaging System.
Each O2 probe generates a 4-20 mA output signal that is received by the OxyBalance system. When a probe channel detects a current signal greater than 3.5 mA, that probe is considered active, and is automatically made available. It is not necessary to tell the OxyBalance that a probe is present. It is automatically detected and used. Table 3-1 shows how the OxyBalance system interprets and displays these signals.
4-20 mA Signal Value Status Display Color
<3.5 mA Probe Inactive Not Shown
3.5 mA to 3.6 mA
3.6 mA to 3.8 mA Reading Under Range Yellow
3.8 mA to 20.5 mA Good Reading Green
20.5 mA to 21 mA Reading Over Range Yellow >21 mA Probe Failed Red
After these signals are converted into an O2% they are displayed on the touch screen interface. The rest of this section will outline the necessary steps to configure and operate the OxyBalance system through the touch screen interface.
Probe Failed
Red
http://www.raihome.com
OxyBalance Oxygen Display and Averaging System
All of the operations that are conducted through the touch screen interface are conducted through five subject tabs that appear on the right hand side of the interface.
• Current
• Trend
• Login
• Setup
• Alarm
Instruction Manual
IM-106-4050 Rev 1.1
July 2008

Login

Figure 3-1. Login Required Screen
Figure 3-2. Login Not Required Screen
The operation of the interface can be password protected so either the Login Required, or Login Not Required screen will be displayed once the Login tab has been pressed (Figure 3-1 and Figure 3-2). Initial login requires the user name Oper and the password 0 (the number zero).
3-2
Instruction Manual
IM-106-4050 Rev 1.1 July 2008
OxyBalance Oxygen Display
and Averaging System

Setup

Figure 3-3. Scale Values Setup
The Setup tab contains five sub-tabs displayed along the top of the screen: a General tab and one tab for each of the four averages.
General
Scale Values
Under this tab the values for the Analog full scale, Trend min value and Trend max value can be changed to represent more of an applicable scale when displaying the probe or average outputs (Figure 3-3).
• Analog full scale - This setting controls the O2% value represented by the incoming 4-20 mA signals, and the average O2% output signals. All input signals must represent the same full scale value.
• Trend min value - This value allows the adjustment of the minimum value displayed on the trend graphs on the touch screen display allowing the display to zoom in on the range of interest. This setting only affects the displayed values. It has no affect on the average O2% output signals.
• Trend max value - This value allows the adjustment of the maximum value displayed on the trend graphs on the touch screen display allowing the display to zoom in on the range of interest. This setting only affects the displayed values. It has no affect on the average O2% output signals.
3-3
OxyBalance Oxygen Display and Averaging System
Time and Date
To set the time and date press the "Time/Date" button shown on the General tab of the Setup screen. A new screen will be shown (Figure 3-4). Touch any value in the Actual column to change that value. Once the new value has been entered hold the "Hold To Set Clock" button until the time and date has been updated in the upper right hand corner of the screen.
Figure 3-4. Time and Date Setup
Instruction Manual
IM-106-4050 Rev 1.1
July 2008
Figure 3-5. Probe Names
Probe Names
To set a name for a probe, press the "Probe Names" button at the bottom of the General tab. The Probe Names screen shown in Figure 3-5 will be shown. Each probe can be assigned a three character alpha-numeric identifier.
3-4
Loading...
+ 54 hidden pages