Rosemount Analytical designs, manufactures and tests its products to meet many national and
international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to
follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Rosemount Analytical repre-
sentative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate In-
struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program,
and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY.
Look-alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except
when maintenance is being performed by qualified persons, to prevent electrical shock
and personal injury.
The information contained in this document is subject to change without notice.
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
e-mail: gas.csc@EmersonProcess.com
Table 10-1. Replacement Parts .............................................................................................. 10-1
Rosemount Analytical Inc. A Division of Emerson Process Managementiii
Instruction Manual
IB-106-440C Original Issue
February 2003
OCX 4400 HAZARDOUS AREA
ivRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
PREFACE
The purpose of this manual is to provide a comprehensive understanding of the OCX 4400
Oxygen/Combustibles Transmitter components, functions, installation, and maintenance.
This manual is designed to provide information about the OCX 4400. We recommend that
you thoroughly familiarize yourself with the Introduction and Installation sections before
installing your transmitter.
The introduction presents the basic principles of the transmitter along with its performance
characteristics and components. The remaining sections contain detailed procedures and
information necessary to install and service the transmitter.
Before contacting Rosemount concerning any questions, first consult this manual. It describes most situations encountered in your equipment’s operation and details necessary
action.
DEFINITIONS
Instruction Manual
IB-106-440C Original Issue
February 2003
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this
publication.
Highlights an operation or maintenance
procedure, practice, condition, statement, etc. If not strictly observed, could
result in injury, death, or long-term
health hazards of personnel.
NOTE
Highlights an essential operating procedure,
condition, or statement.
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETIN
Highlights an operation or maintenance
procedure, practice, condition, statement, etc. If not strictly observed, could
result in damage to or destruction of
equipment, or loss of effectiveness.
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, and is not related to the illustration in any technical
manner.
Rosemount Analytical Inc. A Division of Emerson Process ManagementP-1
Instruction Manual
IB-106-440C Original Issue
February 2003
FOR THE WIRING AND INSTALLATION
The following safety instructions apply specifically to all EU member states. They should
be strictly adhered to in order to assure compliance with the Low Voltage Directive. NonEU states should also comply with the following unless superseded by local or National
Standards.
1. Adequate earth connections should be made to all earthing points, internal and external,
where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced.
The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
OCX 4400 HAZARDOUS AREA
IMPORTANT
SAFETY INSTRUCTIONS
OF THIS APPARATUS
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable
anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be
made through a circuit breaker which will disconnect all circuits carrying conductors during a
fault situation. The circuit breaker may also include a mechanically operated isolating switch.
If not, then another means of disconnecting the equipment from the supply must be provided
and clearly marked as such. Circuit breakers or switches must comply with a recognized
standard such as IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazardous voltages are likely to be present beneath. These covers should only be
removed when power is removed from the equipment — and then only by
trained service personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a
danger from hot surfaces beneath. These covers should only be removed by
trained service personnel when power is removed from the equipment. Certain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to
the Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the following standards: EN61010-1, IEC417, and ISO3864.
P-2Rosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
1
SECTION 1
DESCRIPTION AND SPECIFICATIONS
Instruction Manual
IB-106-440C Original Issue
February 2003
1-1 COMPONENT CHECKLIST OF TYPICAL
SYSTEM (PACKAGE CONTENTS)
A typical Rosemount OCX 4400 Oxygen/
Combustibles Transmitter package should contain the items shown in Figure 1-1. Record the
part number, serial number, and order number
for each component of your system in the table
located on the back cover of this manual.
Use the product matrix in Table 1-2 at the end
of this section to verify your order number. The
first part of the matrix defines the model. The
last part defines the various options and features of the OCX 4400. Check the model number against the transmitter features and options,
making sure options specified by this number
are on or included with the unit. Use this complete model number for any correspondence
with Rosemount.
1-2 SYSTEM OVERVIEW
a. Scope
This Instruction Bulletin supplies details
needed to install, startup, operate, and
maintain the Rosemount OCX 4400 Oxygen/ Combustibles Transmitter. Signal conditioning electronics outputs separate 4-20
mA signals representing oxygen (O
combustibles (COMB) values. This information, plus additional details, can be accessed with the HART Model 275 handheld
communicator or Asset Management Solutions (AMS) software. An optional local operator interface (LOI) is available to provide
local communication with the electronics.
b. System Description
The Rosemount OCX 4400 Oxygen/
Combustibles Transmitter is designed to
measure oxygen and combustible concentrations in flue gas temperatures up to
2600°F (1427°C). Electrical connections,
power and communications are made
through two 3/4 NPT ports in the flameproof
electronics enclosure using fittings and
cables provided by the customer. Cable
) and
2
installation must meet NEC, IEC and/or
other applicable national or local codes for
Category 2, Zone 1, Group IIB +H2 T3 permanently mounted equipment. The transmitter is close coupled to the process and
requires minimal sample conditioning requirements.
The equipment measures oxygen percentage by reading the voltage developed
across a heated electrochemical cell, which
consists of a small yttria-stabilized, zirconia
disc. Both sides of the disc are coated with
porous metal electrodes. When operated at
the proper temperature, the millivolt output
of the cell is given by the following Nernst
equation:
EMF = KT log10 (P
Where:
1. P
is the partial pressure of the oxygen in
2
the measured gas on one side of the cell.
2. P
is the partial pressure of the oxygen in
1
the reference air on the opposite side of
the cell.
3. T is the absolute temperature.
4. C is the cell constant.
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry instrument air (20.95% oxygen) as the
reference air.
When the cell is at operating temperature
and there are unequal oxygen concentrations across the cell, oxygen ions will travel
from the high oxygen partial pressure side
to the low oxygen partial pressure side of
the cell. The resulting logarithmic output
voltage is approximately 50 mV per decade.
The output is proportional to the inverse
logarithm of the oxygen concentration.
Therefore, the output signal increases as
the oxygen concentration of the sample gas
decreases. This characteristic enables the
OCX 4400 to provide exceptional sensitivity
at low oxygen concentrations.
1/P2
) + C
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
IB-106-440C Original Issue
February 2003
5
OCX 4400 HAZARDOUS AREA
1
MAN4275A00
English
October1994
HART
Communicator
o
TM
FISHER-ROSEMOUNT
4
3
1. Instruction Bulletin
2. HART
Communicator Package (Optional)
3. Adapter Plate with Mounting Hardware and Gasket
4. OCX 4400 (Shown with Integral Electronics)
5. Reference Air and Calibration Set
6. Blowback Hardware (Optional)
Figure 1-1. Typical System Package
2
6
37020025
1-2 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
1
Instruction Manual
IB-106-440C Original Issue
February 2003
The OCX 4400 measures net oxygen concentration in the presence of all the products of combustion, including water vapor.
Therefore, it may be considered an analysis
on a “wet” basis. In comparison with older
methods, such as the portable apparatus,
which provides an analysis on a “dry” gas
basis, the “wet” analysis will, in general, indicate a lower percentage of oxygen. The
difference will be proportional to the water
content of the sampled gas stream.
The OCX 4400 combustibles sensor is a
catalytic sensor consisting of two Resistant
Temperature Detectors (RTD). One RTD is
the reference element covered with an inert
coating. The other RTD element is active,
coated with a catalyst. As the sample gases
flow by the sensor, the combustible gases
oxidize on the surface of the active element.
The oxidation that occurs produces heat
and a temperature rise in the active element. The temperature difference produces
a resistance relationship between the two
elements that is directly proportional to the
concentration of combustibles in the sample
gases.
The catalyst is specifically designed to detect carbon monoxide (CO), but the sensor
responds to other combustible gases. The
sensor is calibrated using carbon monoxide
(CO), thus the output should be expressed
in terms of CO. However, since the sensor
detects other combustible gases, the output
cannot just be labeled CO. The response of
the sensor to other combustible gases gives
an output that is equivalent to the sensor
detecting CO. The term COMB is used in
this manual to describe the sensor output.
This term indicates that the sensor is calibrated in terms of CO, and that the sensor
output is equivalent to CO but not specific to
CO.
c. System Configuration
Transmitters are available in four length options, giving the user the flexibility to use a
penetration appropriate to the size of the
stack or duct. The length options are 18 in.
(457 mm), 3 ft (0.91 m), 6 ft (1.83 m), or 9 ft
(2.7 m). Probes are available in three material options, 316L stainless steel, inconel
600, and ceramic to accommodate higher
temperatures.
The electronics are contained in a separate
housing from the sensors and may be integrally mounted next to the sensor housing
on the probe, or split mounted up to 150
feet away from the probe.
The electronics control both sensor temperatures and provide individual 4-20 mA
isolated outputs that are proportional to the
measured oxygen and combustibles concentrations. The power supply can accept
voltages of 100 to 240 VAC and 50 to 60
Hz. The electronics accepts millivolt signals
generated by the sensors and produces the
outputs to be used by remotely connected
devices. The outputs are isolated 4-20 mA
linearized currents. See 3-1c and 3-1d for
specific instructions upon initial power up.
d. System Features
1. The O
tivity increase as the oxygen concentration decreases.
2. HART communication is standard. To
use the HART capability, you must
have either:
(a) A HART Model 275
cell output voltage and sensi-
2
Communicator.
Dilution air is provided to the COMB sensor
to ensure there is adequate oxygen to fully
oxidize any combustible gases regardless
of the concentration of oxygen in the
process.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
(b) Asset Management Solutions
(AMS) software for the PC.
3. Oxygen cell and heater/thermocouple
assembly are field replaceable.
Instruction Manual
IB-106-440C Original Issue
February 2003
OCX 4400 HAZARDOUS AREA
4. Electronics are automatically configured for line voltages from 100 to 240
VAC.
5. An operator can calibrate and diagnostically troubleshoot the OCX 4400 in
one of two ways:
(a) HART Interface. Each of the OCX
4400’s 4-20 mA output lines
transmit an analog signal proportional to oxygen or combustible
levels detected. The HART output
is superimposed on the oxygen 420 mA output line only. This information can be accessed through
the following:
1 Rosemount Model 275 Hand-
held Communicator - The
handheld communicator requires Device Description
(DD) software specific to the
OCX 4400. The DD software
will be supplied with many
Model 275 units, but can also
be programmed into existing
units at most FisherRosemount service offices.
Refer to Section 5,
HART/AMS, for additional
information.
2 Personal Computer (PC) -
The use of a personal computer requires AMS software
available from FisherRosemount.
3 Selected Distributed Control
Systems - The use of distributed control systems requires
input/output (I/O) hardware
and AMS software which
permit HART communications.
(b) Optional Local Operator Interface
(LOI). The optional LOI is mounted
to the end of the electronics module and allows local communications with the electronics. Refer to
Appendix A, Local Operator Interface (LOI), for more information.
6. Optional Blowback System. The blowback system periodically blows instrument air back through the inlet filter
and out the inlet tube. This clears out
particulate and keeps the inlet filter
from clogging.
e. System Operation
Figure 1-2 shows the relationship between
the components of the OCX 4400. The sensors and the electronics are contained in
separate flameproof housings. The sensor
housing and probe mounts to a duct or process wall so that the probe protrudes into
the flue gas stream. An air powered eductor
continuously pulls samples of the process
flue gas through the probe to a chamber in
front of the sensor housing where the sample passes the O
sensor and continues on
2
to the COMB sensor. Dilution air is provided
to the COMB sensor and reference air to
the O
sensor. After the gas sample flows
2
past the O
sensor and through the COMB
2
sensor, it is drawn through the eductor
where it mixes with the eductor air and exits
through exhaust back into the system. The
electronics housing contains the CPU which
converts the sensor inputs into 4-20 mA
analog output signals. The CPU can also
initiate and perform calibrations. Three test
gasses and instrument air can be turned on
and off by solenoids. Test gas flow to the
sensors is regulated by a rotometer between the electronics and sensor housings.
Instrument air is separated into eductor air,
reference air, and dilution air. The instrument air solenoid does not allow air flow
until the heaters are up to temperature. This
minimizes the amount of sampled process
flue gas being pulled into cold sensors
causing condensation.
1-4 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
1
Instruction Manual
IB-106-440C Original Issue
February 2003
SAMPLE
GAS
PROBE
EXHAUST
SENSOR
HOUSING
COMBUSTIBLES
SENSOR
O2
SENSOR
EDUCTOR
ROTOMETER
5 SCFH
ELECTRONICS
PRESSURE
SENSOR
OPTIONAL
TEST GAS
SOLENOIDS
INSTRUMENT AIR
SOLENOID
EDUCTOR AIR
HOUSING
CPU
POWER
SUPPLY
LOW O2
TEST GAS
HIGH O2
TEST GAS
HIGH CO
TEST GAS
INSTRUMENT
AIR
REFERENCE AIR
DILUTION AIR
ROTOMETER
50 CC/MIN.
(0.1 SCFH)
Figure 1-2. System Operation Diagram
37020001
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
IB-106-440C Original Issue
February 2003
OCX 4400 HAZARDOUS AREA
f. Handling the OCX 4400
It is important that printed circuit
boards and integrated circuits are
handled only when adequate antistatic
precautions have been taken to prevent possible equipment damage.
The OCX 4400 is designed for industrial application. Treat each component of the system with care to avoid
physical damage. The probe may contain components made from ceramics,
which are susceptible to shock when
mishandled.
g. System Considerations
Prior to installing your Rosemount OCX
4400 Oxygen/Combustibles Transmitter,
make sure you have all the components
necessary to make the system installation.
Ensure all the components are properly integrated to make the system functional.
After verifying that you have all the components, select mounting locations and determine how each component will be placed in
terms of available line voltage, ambient
temperatures, environmental considerations, convenience, and serviceability.
Figure 1-3 shows a typical system wiring. A
typical system installation is illustrated in
Figure 1-4.
A source of instrument air is required at the
OCX 4400 for reference air, dilution air, and
eductor air. Since the OCX 4400 is
equipped with an in-place calibration feature, provision should be made for connecting test gas tanks to the OCX 4400
when it is to be calibrated.
NOTE
The electronics module is designed to
meet NEMA 4X (IP55) and the electronic components are rated to temperatures up to 185°F (85°C).
Retain packaging in which the unit arrived from the factory in case any
components are to be shipped to another site. This packaging has been
designed to protect the product.
INSTRUMENT
AIR
OCX 4400 O2/COMBUSTIBLES TRANSMITTER
WITH INTEGRAL ELECTRONICS
4-20 MA OUTPUT
(TWISTED PAIRS)
3 CALIBRATION GAS LINES
BY CUSTOMER
[300 FT (90M) MAX]
Figure 1-3. OCX 440 HART Connections and AMS Application
TERMINATION IN
CONTROL ROOM
HART MODEL 275
HANDHELD
INTERFACE
CUSTOMER’S LAPTOP
WITH ASSET MANAGEMENT
ASSET MANAGEMENT SOLUTIONS
SOLUTIONS
34070002
1-6 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
1
INTEGRAL ELECTRONICS
ADAPTER
PLATE
DUCT
TEST
GAS
ROTOMETER
GASES
STACK
DILUTION AIR
ROTOMETER
Instruction Manual
IB-106-440C Original Issue
4-20 mA OUTPUTS (2)
LINE VOLTAGE
INSTRUMENT AIR
SUPPLY (REF. GAS)
PRESSURE
REGULATOR
SPAN COMBUSTIBLES TEST GAS
LOW O2 TEST GAS
HIGH O2 TEST GAS
February 2003
ELECTRONICS
CABLE
(UP TO 150FT)
SPLIT-MOUNTED ELECTRONICS
GASES
DUCT
STACK
TEST
GAS
ROTOMETER
DILUTION AIR
ROTOMETER
4-20 mA OUTPUTS (2)
LINE VOLTAGE
INSTRUMENT AIR
SUPPLY (REF. GAS)
PRESSURE
REGULATOR
SPAN COMBUSTIBLES TEST GAS
LOW O2 TEST GAS
HIGH O2 TEST GAS
34070003
Figure 1-4. Typical System Installation
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
Instruction Manual
IB-106-440C Original Issue
February 2003
OCX 4400 HAZARDOUS AREA
1-3 OCX 4400 SPECIFICATIONS
transmitter before choosing the mounting loca-
tion. Table 1-2 contains the product matrix. Use
Table 1-1 contains information about the OCX
4400 Oxygen/Combustibles Transmitter operating characteristics. Use the table to make sure
that available conditions are suitable for the
the product matrix for ordering information and
specific model number information. Table 1-3
contains information on accessories available
with the OCX 4400.
Table 1-1. OCX 4400 Specifications
Net O
Range ..................................................0-1% to 0-40% O
2
, Fully field selectable
2
Combustibles .................................................0-1000 ppm to 0-5%, Fully field selectable
Accuracy:
Oxygen.....................................................+0.75% of reading or 0.05% O
Calibration ......................................................Semi-automatic or automatic
Calibration Gas Mixtures Recommended....0.4% O
(Ref. test gas bottles kit #1A99119G04)8% O2, Balance N
1000 ppm CO, Balance N
NOTES:
(1)
Combustibles sensor is calibrated with CO gas. Sensor may be calibrated with H2 gas. Use in methane service is not
recommended.
(2)
Operating temperature of electronics inside of electronics housing, as measured by a HART Communicator, or Emerson
Process Management Asset Management Solutions (AMS) Software, or local operator interface. Operation below 0°C
(32°F) requires that sensor housing be insulated.
, Balance N
2
2
2
2
1-8 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
1
IB-106-440C Original Issue
OCX 4400 HAZARDOUS AREA
Table 1-1. OCX 4400 Specifications (Continued)
Calibration Gas Flow .....................................5 scfh (2.5 l/m)
Reference Air..................................................2 scfh (1 l/m), clean, dry instrument-quality air (20.95% O
regulated to 45 psi (310 kPa)
Eductor Air......................................................5 scfh (2.5 l/m), clean, dry, instrument-quality air
(20.95% O
Dilution Air......................................................0.1 scfh (0.5 l/m), clean, dry, instrument-quality air
(20.95% O
Blowback Air (Optional) ................................Clean, dry, instrument-quality air (20.95% O
regulated to 55 psi (374 kPa)
Sensor and Electronics Housings ...............NEMA 4X, IP55 with fitting and pipe on reference exhaust
port to clean, dry atmosphere, two 3/4-14 NPT conduit ports
Electrical Noise ..............................................Meets EN 50082-2 Electromagnetic Compatibility Generic
Immunity Std., Part II
Includes ENG 1000 4-R for Electrostatic Discharge 4 Kv
contact, 8 Kv in air
Optionally ENG 1000 4-R “Namur-Increased” 8 Kv contact,
16 Kv in air
Includes IEC 801-4 fast transients-2
Kv on power supply and control lines
), regulated to 45 psi (310 kPa)
2
), regulated to 45 psi (310 kPa)
2
February 2003
),
2
),
2
Line Voltage....................................................Universal 100 to 240 VAC, 50 to 60 Hz,
software selectable, 3/4-14 NPT conduit port
Isolated Output:
Oxygen.....................................................4-20 mAdc, 950 ohm max, with HART capability
for O
and COMB
2
Combustibles ..........................................4-20 mAdc, 950 ohm max
FM (Pending)
CSA (Pending)
CENELEC EEX d IIB+H2, Sensor Housing T3,
Electronics Housing T6
CENELEC EEX d ia IIB+H2, Sensor Housing T3,
Electronics Housing T6 (with IS Barrier option)
NOTE:
All static performance characteristics are with operating variables constant. Specifications subject to change without
notice.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-9
Instruction Manual
IB-106-440C Original Issue
February 2003
Table 1-2. Product Matrix
OCX44CO2 / Combustibles Transmitter - Flameproof
O2 / Combustibles Transmitter IB-106-440C
CodeProbe Length and Material
00No Probe or Exhaust Tube
1118 in. (457 mm) 316 SST up to 1300°F (704°C)
2118 in. (457 mm) Inconel 600 up to 1832°F (1000°C)
3118 in. (457 mm) Ceramic up to 2600°F (1426°C)
123 ft (0.91 m) 316 SST up to 1300°F (704°C)
223 ft (0.91 m) Inconel 600) up to 1832°F (1000°C)
323 ft (0.91 m) Ceramic up to 2600°F (1426°C)
136 ft (1.83 m) 316 SST up to 1300°F (704°C)
236 ft (1.83 m) Inconel 600 up to 1832°F (1000°C)
149 ft (2.7 m) 316 SST up to 1300°F (704°C)
Code Probe Mounting Assembly
10(ANSI 2 in. 150 lb) 6" dia. flange, 4.75" BC with 4 x 0.75" dia. holes
20(DIN) 185mm dia. flange, 145mm BC with 4 x 18mm dia. holes
30(JIS) 155mm dia. flange, 130mm BC with 4 x 15mm dia. holes
OCX 4400 HAZARDOUS AREA
CodeMounting Hardware - Stack Side
0No Adapter Plate ("0" must also be chosen under "Mounting Adapter - Probe Side" below)
1New Installation - Square weld plate with studs
2Model 218/240 Mounting Plate (with Model 218/240 Shield Removed)
3Existing Model 218/240 Support Shield
4Mounting to Other Mounting
5Model 132 Adapter Plate
6Mounting to Ametek WDG-IVC
7Mounting to Servomex Xendos 2700
8Mounting to Elsag Bailey SMA-90
CodeMounting Hardware - Probe Side
0No Adapter Plate
1Probe Only (ANSI)
4Probe Only (DIN)
7Probe Only (JIS)
OCX44C111011(Cont’d)Example
(1)
1-10 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
1
Table 1-2. Product Matrix (Continued)
Cont'dCodeElectronic Housing - NEMA 4X, IP55 HART Communications
01Basic Unit
02Local Operator Interface
03Calibration Solenoids
04IS Barrier
05Local Operator Interface and Calibration Solenoids
06Local Operator Interface and IS Barrier
07Calibration Solenoids and IS Barrier
08Local Operator Interface and Calibration Solenoids and IS Barrier
CodeElectronic Mounting
3
01Integral to Sensor Housing Electronics
02Split Electronics and no cable
03Split Electronics and 20 ft (6m) cable
04Split Electronics and 40 ft (12m) cable
05Split Electronics and 60 ft (18m) cable
06Split Electronics and 80 ft (24m) cable
07Split Electronics and 100 ft (30m) cable
08Split Electronics and 150 ft (45m) cable
2
3
3
Instruction Manual
IB-106-440C Original Issue
February 2003
3
CodeAccessories
00None
01Cal. Gas Rotometers, Flow Rotometers, and Ref. Gas Set
11Blowback Option and Cal. Gas Rotometers, Flow Rotometers, and Ref. Gas Set
Cont’d080111Example
NOTES:
(1)
Where possible, specify SPS number; otherwise, provide details of the existing mounting plate as follows:
Plate with studsBolt circle diameter, number, and arrangement of studs, stud thread, stud height above mounting plate.
Plate without studs Bolt circle diameter, number, and arrangement of holes, thread, depth of stud mounting plate with accessories.
(2)
If the Local Operator Interface (LOI) is not implemented, remote access and functionality available via HART Communications (Model 275 Handheld Communicator) with Oxygen/Combustibles Device Description (DD) required.
(3)
If the IS Barrier option is selected, customer must also purchase the Moore Industries SPA (1A99292H01), current repeater
backplane and adpater kit (p/n 1A99698G01) to maintain the 4 – 20 mA combustibles output.
Table 1-3. Accessories
PART NUMBERDESCRIPTION
1A99119H01Oxygen Bottle 0.4%, balance N
1A99119H02Oxygen Bottle 8.0%, balance N
1A99119H06CO Bottle 1000 PPM and air, balance N
2
2
2
1A99120H02Regulator for Oxygen (may need 2)
1A99120H03Regulator for COMB
1A99119G06Wall mount bracket for gas bottles
1A99119G05Test gas regulators kit
1A99119G04Test gas bottles kit
1A99292H01
Moore Industries SPA for Low O
Alarm, High
2
COMB Alarm, Calibration Status, and Unit Fail
275D9EI5D0000275 HART Communicator with 12 Meg buffer
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-11
Instruction Manual
IB-106-440C Original Issue
February 2003
OCX 4400 HAZARDOUS AREA
1-12 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
2
SECTION 2
INSTALLATION
Before installing this equipment, read
the “Safety instructions for the wiring
and installation of this apparatus” at
the front of this Instruction Bulletin.
Failure to follow the safety instructions could result in serious injury or
death.
2-1 MECHANICAL INSTALLATION
Instruction Manual
IB-106-440C Original Issue
February 2003
readings. Therefore, either make the
necessary repairs or install the transmitter upstream of any leakage.
3. Ensure the area is clear of internal and
external obstructions that will interfere
with installation and maintenance access to the unit. Allow adequate clearance for the removal of the OCX 4400.
a. Selecting Location
1. The location of the OCX 4400 in the
stack or flue is most important for
maximum accuracy in the oxygen
analyzing process. The probe must be
positioned so the gas it measures is
representative of the process. Best results are normally obtained if the
transmitter is positioned near the center of the duct (40-60% insertion).
Longer ducts may require several
transmitters since the oxygen and
combustibles can vary due to stratification. A point too near the wall of the
duct, or the inside radius of a bend,
may not provide a representative sample because of the very low flow conditions. The sensing point should be
selected so the process gas temperature falls within the range of probe
material used. Figure 2-1 through
Figure 2-5 provide mechanical installation references. The ambient temperature inside the electronics housing
must not exceed 185°F (85°C).
2. Check the flue or stack for holes and
air leakage. The presence of this condition will substantially affect the accuracy of the oxygen and combustibles
Do not allow the temperature of the
electronics housing to exceed 185°F
(85°C) or damage to the electronics
may result.
b. Installation
1. Ensure all components are available to
install the OCX 4400.
2. The OCX 4400 may be installed intact
as it is received.
3. Weld or bolt adapter plate (Figure 2-3)
onto the duct.
4. For systems with split electronics, use
the pole or wall mounting hardware as
shown in Figure 2-4 to mount the electronics housing. Choose a location not
to exceed the length of the electronics
cable ordered.
5. For integral electronics installations,
ensure the system cable drops vertically from the OCX 4400 and the conduit is routed below the level of the
housings. Drip loops minimize the possibility that moisture will damage the
electronics (Figure 2-5).
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
IB-106-440C Original Issue
February 2003
OCX 4400 HAZARDOUS AREA
NOTE
If process temperatures will exceed
392°F (200°C), use anti-seize compound on stud threads to ease future
removal of the OCX 4400.
6. Insert sample and exhaust tubes
through the opening in the mounting
flange and bolt the unit to the flange.
Uninsulated stacks or ducts may
cause ambient temperatures in the
electronics housing to exceed 185°F
(85°C) and damage the electronics.
NOTE:
ALL DIMENSIONS ARE IN
INCHES WITH MILLIMETERS
IN PARENTHESES.
INSULATE IF EXPOSED TO
AMBIENT WEATHER CONDITIONS
7. If insulation is removed to access the
duct for OCX 4400 mounting, make
sure to replace insulation afterward.
c. Enclosures
The OCX 4400 enclosures are designed to
meet ingress conditions of IP55. Each enclosure cover is threaded to its base and
sealed with an o-ring that isolates the
threads from external contaminants.
Each cover is secured by a clip attached to
the base that engages the cover between
the ribs of the cover sidewall. The clip is
held in place by an Allen head cap screw
and lockwasher mounted in a recess. Cover
removal and installation requires an Allen
wrench (special tool) to loosen and tighten
the screw.
0.062 THK GASKET
3535B18H02
ANSI
3535B46H01
JIS
3535B45H01
DIN
DIM “A”DIM “B”
TABLE 2. INSTALLATION/REMOVAL
PROBE
18 IN.
3 FT
6 FT
9 FT
DIM "A"
(406)
(864)34(1280)
(1778)
106
(2692)
DIM "B"
16
70
(2184)
122.38
(3108)
32.38
(822)
50.38
86.38
TOP VIEW
FLANGE
DIA
HOLE
DIA
(4) HOLES
EQ SP ON
B.C. DIA.
TABLE 1. MOUNTING FLANGE
ANSI
4512C17H01
6.00
(153)
0.75
(20)
4.75
(121)
DIN
4512C19H01
7.28
(185)
0.71
(18)
5.71
(145)
JIS
4512C18H01
6.10
(155)
0.59
(15)
5.12
(130)
37020002
Figure 2-1. OCX 4400 Installation (with Integral Electronics)
2-2 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
2
ADAPTER PLATE* DIMENSIONS FOR OCX 4400
DIMENSIONS
IN.
(mm)
ANSI
(P/N 4512C34G01)
(P/N 4512C36G01)
DIN
IB-106-440C Original Issue
JIS
(P/N 4512C35G01)
Instruction Manual
February 2003
"A"
"B"
THREAD
"C"
DIA
*PART NUMBERS FOR ADAPTER PLATES INCLUDE
ATTACHING HARDWARE.
NOTE: DIMENSIONS ARE
IN INCHES WITH
MILLIMETERS IN
PARENTHESES.
2.500 DIA
(63.5)
ADAPTER PLATE
FOR OCX 4400
INSTALLATION. SEE
FIGURE 2-1.
6.00
(153)
0.625-11
4.75
(121)
B
7.5
(191)
(M-16 x 2)
5.708
(145)
A
o
45
C
6.50
(165)
(M-12 x 1.75)
5.118
(130)
A
4 STUDS,
LOCKWASHERS AND
NUTS EQUALLY
SPACED ON
C DIA B.C.
29790013
Figure 2-2. OCX 4400 Adapter Plate Dimensions
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
IB-106-440C Original Issue
February 2003
OCX 4400 HAZARDOUS AREA
INSTALLATION FOR METAL
WALL STACK OR DUCT
NOTE: ONLY ADAPTER PLATE IS FURNISHED BY ROSEMOUNT.
WELD OR BOLT ADAPTER
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIRTIGHT.
DIMENSIONS IN INCHES WITH MILLIMETERS IN PARENTHESES.
2.50 [63.5]
MIN DIA HOLE
IN WALL
STACK OR DUCT
METAL WALL
Figure 2-3. Adapter Plate Installation
INSTALLATION FOR MASONRY
WALL STACK CONSTRUCTION
BOLT ADAPTER
PLATE TO OUTSIDE
WALL SURFACE
JOINT MUST
BE AIRTIGHT
MASONRY
STACK WALL
FIELD WELD
PIPE TO
ADAPTER PLATE
PIPE 3.00
SCHEDULE 40,
SLEEVE LENGTH
OPTIONAL
37020003
NOTE:
* INDICATES ITEMS ARE INCLUDED
IN MOUNTING KIT P/N 4851B40G01.
*POLE
MOUNTING
CLAMP
*WALL OR POLE
MOUNT BRACKET
AND SCREWS
1/4-20 UNC X 2 IN. M-F STANDOFFS
RECOMMENDED FOR REAR PIPING
CLEARANCE WHEN WALL MOUNTING
(4 EACH BY CUSTOMER)
WALL MOUNTING
BOLTS OR LAG SCREWS
(4 EACH BY CUSTOMER)
Figure 2-4. Wall or Pole Mounting of Electronics Housing
ELECTRONICS
HOUSING
37020021
2-4 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
2
Instruction Manual
IB-106-440C Original Issue
February 2003
CONDUIT
DRIP LOOPS
CONDUIT
DRIP LOOP
Figure 2-5. Installation with Drip Loops
DUCT
WALL
CONDUIT
DRIP LOOP
DUCT
WALL
37020004
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
IB-106-440 Rev. 2.0
January 2003
OCX 4400
2-2 ELECTRICAL INSTALLATION
All wiring must conform to local and national
codes. For reference, factory wired solenoid
power connections are shown in Figure 2-6.
Disconnect and lock out power before
connecting the unit to the power
supply.
Install all protective equipment covers
and safety ground leads after installation. Failure to install covers and
ground leads could result in serious
injury or death.
To meet the Safety Requirements of
IEC 1010 (EC requirement), and ensure
safe operation of this equipment, connection to the main electrical power
supply must be made through a circuit
breaker (min 10 A) which will disconnect all current-carrying conductors
during a fault situation. This circuit
breaker should also include a mechanically operated isolating switch. If
not, then another external means of
disconnecting the supply from the
equipment should be located close by.
Circuit breakers or switches must
comply with a recognized standard
such as IEC 947.
housing. Secure the connectio n with two
nuts. Attach a separate line from the ground
stud to the G terminal on the power input
terminal block.
NOTE
Make sure line voltage is correctly set.
See paragraph 3-1c for instructions.
b. Connect 4-20 mA Signals
Connect the 4-20 mA current loop to the
4-20 mA signal output terminals in the electronics housing. If the isolation barrier
option is provided, refer to Appendix D
for alternate wiring instructions.
1. Oxygen (O
) 4-20 mA Signal. One
2
4-20 mA signal represents the O
value. Superimposed on the O2 signal
is the HART information accessible
through a Model 275 Handheld Communicator or AMS software. The O
signal is at the AOUT 1+ and AOUT 1terminals.
2. Combustibles (COMB) 4-20 mA Signal.
The other 4-20 mA signal represents
the COMB value. HART information is
not available on the COMB signal. The
COMB signal is at the AOUT 2+ and
AOUT 2- terminals.
c. Alarm Output Relay
Connect any customer supplied relay input
to the alarm output relay terminal. The
alarm output relay terminal is a set of dry,
no. 2, form C, double-pole, double-throw
contacts with 30 mA and 24 VDC capacity.
2
2
NOTE
To maintain CE compliance, ensure a
good connection exists between the
mounting flange bolts and earth.
d. Split-Electronics Connections
to Sensor Housing
If the OCX 4400 is configured with splitelectronics, make the following connections
between the electronics and sensor hous-
a. Connect Line Voltage
The OCX 4400 is software configured for
100 to 240 VAC line voltage at 50 to 60 Hz.
The power supply requires no setup. Connect the line, or L wire, to the L terminal,
and the neutral, or N wire, to the N terminal
on the AC power input terminal block in the
electronics housing. Connect the ground, or
G wire, to the ground stud in the electronics
ings with the electronics cable ordere d with
the package (Figure 2-7). Braided cable is
available in lengths up to 150 ft. (45 m).
1. Thermocouple Connections. Connect
the electronics housing sensor/thermocouple terminals to the corresponding
terminals in the sensor housing. The
twisted wire pairs are numbered on the
inner plastic wrapper. Keep tw isted
pairs together and match the numbers
2-6 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
2
Instruction Manual
IB-106-440C Original Issue
February 2003
SOLENOID POWER
TERMINAL BLOCK
(FACTORY WIRED)
GND
GND
LO O2
HI COMB
HI O2
ALARM OUTPUT
RELAY TERMINAL
GND
GND
INST AIR
NC
NO
COMMON
SIGNAL OUTPUT
TERMINAL BLOCK
4-20 mA
ELECTRONICS HOUSING
(COVER REMOVED)
SENSOR INPUT
HEATER THERMOCOUPLE,
AND HEATER POWER
INPUT PORT
(FROM SENSOR HOUSING)
PRESSURE
SENSOR
HI COMB
SOLENOID
4-20 mA
SIGNAL
OUTPUT
PORT
HI O2
SOLENOID
LO O2
SOLENOID
INST AIR
SOLENOID
AOUT 2-
AOUT 2+
AOUT 1-
AOUT 1+
COMB
4-20 mA
SIGNAL
O2
4-20 mA
SIGNAL
AC POWER INPUT
TERMINAL BLOCK
F1 POWER
SUPPLY FUSE
L
F2 HEATER
N
G
FUSE
Figure 2-6. Line Voltage and 4-20 mA Connections
AC POWER
INPUT PORT
GROUND STUD
TWO NUTS
34070021
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
IB-106-440C Original Issue
February 2003
OCX 4400 HAZARDOUS AREA
SPLIT-ELECTRONICS HOUSING
HEATER POWER
TERMINAL BLOCK
GRAY
L2
BROWN
COMB CELL HEATER
RED
L2
ORANGE
BLACK
PURPLE
L2
SL HEATER
O2 CELL HEATER
ELECTRONICS CABLE
(UP TO 150 FT)
NOTE: KEEP TWISTED
PAIRS TOGETHER
(COVER REMOVED)
SENSOR/THERMOCOUPLE
TERMINAL BLOCK
(1)
YELLOW
O2 T/C +
RED
O2 T/C -
(2)
SHIELD
YELLOW
GND
SL T/C +
RED
SL T/C -
(3)
SHIELD
WHITE
BLUE
GND
O2 MV -
O2 MV +
PRESSURE
SENSOR
ELECTRONICS
CABLE PORT
GROUND STUD
(5)
(4)
YELLOW
WHITE
GREEN
COMB ACTIVE +
WHITE
COMB REF -
COMB REF +
COMB ACTIVE -
CABLE
(6)
WHITE
SHIELD
RED
GND
COMB DRIVE
COMB RETURN
PRESSURE
SENSOR
FACTORY
WIRED
WHITE
RED
GREEN
+ 10 PR SRC
PR SENSOR -
PR SENSOR +
BLACK
PR SRC RTN
(5)
(6)
WHITE
WHITE
RED
YELLOW
BLUE
COMB REF -
COMB REF +
COMB DRIVE
COMB RETURN
SENSOR/THERMOCOUPLE
TERMINAL BLOCK
ELECTRONICS
CABLE PORT
PROBE
L2
L2
SL HEATER
O2 CELL HEATER
COMB CELL HEATER
GRAY
BLACK
PURPLE
ORANGE
HEATER POWER
TERMINAL BLOCK
BROWN
L2
RED
(1)
YELLOW
O2 T/C +
RED
O2 T/C -
(2)
YELLOW
SL T/C +
RED
SL T/C -
(4)
(3)
WHITE
BLUE
O2 MV -
O2 MV +
WHITE
GREEN
RED
WHITE
COMB ACTIVE -
COMB ACTIVE +
SENSOR HOUSING
(COVER REMOVED)
37020005
Figure 2-7. Electrical Connections Between Split-Electronics and Sensor Housing
2-8 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
OCX 4400 HAZARDOUS AREA
2
Instruction Manual
IB-106-440C Original Issue
February 2003
e. Split-Electronics Connections
to Sensor Housing
When remotely mounted, two interconnecting cables are made through 3/4 NPT ports.
The interconnecting Power and Sensor cables are supplied as part of this instrument.
Cable installation in two conduits will be
provided by the customer and must meet
NEC, IEC and/or other applicable national
or local codes for Category 2, Zone 1,
Group IIB +H2 T3 permanently mounted
equipment. When the OCX 4400 is configured with split-electronics, make the following connections between the electronics
and sensor housings with the electronics
cable ordered with the package (Figur e
2-7). Steel braided cable is available in
lengths up to 150 ft. (45 m).
1. Thermocouple Connections. Connect
the electronics housing sensor/thermocouple terminals to the corresponding
terminals in the sensor housing. The
twisted wire pairs are numbered on the
inner plastic wrapper. Keep tw isted
pairs together and match the numbers
and wire colors shown in Figure 2-7.
Connect the twisted pair wire shields to
the shield terminals.
2. Heater Power Connections. Use the
purple, orange, black, gray, brown, and
red stranded wires in the electronics
cable to connect power to the three
heaters in the sensor housing. Match
the wire colors to the corresponding
heater power terminal blocks in the
sensor and electronics housings as
shown in Figure 2-7.
2-3 PNEUMATIC INSTALLATION
Pneumatic system connections depend on
which OCX 4400 options are equipped on your
transmitter. Refer to paragraphs 2-3a, 2-3b, and
2-3c, and select the paragraph that applies to
your OCX transmitter configuration.
a. Reference Air Set Option (only)
When no options or only the reference air
set option is equipped, use the following
procedure to install the pneum atic system.
1. Refer to Figure 2-8. Connect the reference air set (regulator/filter and pressure gage) to the instrument air inlet on
the electronics housing and to the inlet
side of the dilution air rotometer.
SENSOR
HOUSING
EDUCTOR
ELECTRONICS
HOUSING
AIR IN
CAL GAS IN
REFERENCE
AIR IN
DILUTION
AIR IN
DILUTION
AIR FLOW
METER
0.1 SCFH
REGULATOR/FILTER
HI CO
INSTRUMENT
AIR SUPPLY
PRESSURE
45 PSIG
CAL GAS
FLOW METER
5 SCFH
20-30 PSIG
(
RECOMMENDED
2-STAGE
REGULATORS
2
2
HI O
LO O
(
35330001
Figure 2-8. Pneumatic Installation, OCX with Reference Air Set (without Autocalibration)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Loading...
+ 65 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.