Rosemount Manual: NGA 2000 WCLD Wet NOx Analyzer Module SW 3.6-Rev B | Rosemount Manuals & Guides

Instruction Manual
760009-B November 2002

Model NGA2000 WNX and WCLD

Wet NOx Sample Conversion Module and Wet CLD Analyzer Module
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and in­ternational standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them
normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
to ensure they continue to operate within their
Read all instructions
prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical rep-
resentative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruc-
tion Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel
to install, operate, update, program,
and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts
specified by Rosemount. Unauthorized parts and procedures can affect the product’s per­formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY
.
Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and per­sonal injury.
The information contained in this document is subject to change without notice.
Teflon is a Registered Trademark of E.I. duPont de Nemours and Co., Inc. Kynar is a Registered Trademark of Atochem North America, Inc.
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Model NGA2000 WNX and WCLD

TABLE OF CONTENTS

PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-5
Documentation....................................................................................................................................P-6
Compliances .......................................................................................................................................P-6
Glossary..............................................................................................................................................P-7
1.0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Overview...................................................................................................................................1-1
1-2 Typical Applications..................................................................................................................1-1
1-3 Specifications – Wet NOx Conversion Module ........................................................................1-6
a. Performance ...............................................................................................................1-6
b. Physical.......................................................................................................................1-6
c. Sample ........................................................................................................................1-6
d. Gas Connections ........................................................................................................1-6
1-4 Specifications - Wet CLD Analyzer Module .............................................................................1-7
a. Performance ...............................................................................................................1-7
b. Physical.......................................................................................................................1-7
c. Sample ........................................................................................................................1-7
d. Gas Connections ........................................................................................................1-8
Instruction Manual
760009-B
November 2002
2.0 INSTALLATION ....................................................................................................................2-1
2-1 Unpacking.................................................................................................................................2-1
2-2 Modification of a Conventional CLD for use in a WNX System ...............................................2-1
2-3 Assembly ..................................................................................................................................2-1
2-4 Location ....................................................................................................................................2-1
2-5 Gases .......................................................................................................................................2-2
a. Overview .....................................................................................................................2-2
b. Pneumatic Connections ..............................................................................................2-2
c. Specifications ..............................................................................................................2-2
2-6 Electrical Connections..............................................................................................................2-5
3.0 OPERATION .........................................................................................................................3-1
3-1 Overview...................................................................................................................................3-1
3-2 Displays & Operating Keys.......................................................................................................3-1
a. Menu Lines & Softkey Functionality............................................................................3-1
b. Common Function Keys..............................................................................................3-2
c. Entering & Changing Variables...................................................................................3-3
d. Starting a Function......................................................................................................3-3
e. Measure Mode Display ...............................................................................................3-4
f. Main Menu ...................................................................................................................3-4
3-3 Startup & Initialization...............................................................................................................3-7
a. Startup Procedure.......................................................................................................3-7
b. Warm-up of the WNX Module.....................................................................................3-8
c. Warm-up of the WCLD Module...................................................................................3-8
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Instruction Manual
760009-B November 2002
d. Binding ........................................................................................................................3-8
3-4 Routine Operation ....................................................................................................................3-8
3-5 Basic Controls and Status ........................................................................................................3-8
a. Analyzer Channel Status ............................................................................................3-9
b. Single Component Display..........................................................................................3-9
c. Multi Component Display ............................................................................................3-9
d. Basic Controls.............................................................................................................3-9
3-6 Display Controls .......................................................................................................................3-10
3-7 Analyzer And I/O, Expert Controls And Setup .........................................................................3-11
a. Range Settings ...........................................................................................................3-12
b. Physical Measurements..............................................................................................3-12
c. Concentration Alarms..................................................................................................3-13
d. Linearization Parameters............................................................................................3-13
e. Linearization Functions ...............................................................................................3-14
f. Response Time............................................................................................................3-16
g. Automatic Range Change...........................................................................................3-17
h. Display Units ...............................................................................................................3-17
i. Physical Measurements & Pressure Limits..................................................................3-18
j. Single Component Display Parameters .......................................................................3-19
3-8 Calibration Procedure...............................................................................................................3-20
a. Calibration Setup ........................................................................................................3-20
b. Basic Controls Calibration...........................................................................................3-22
c. Expert Controls Calibration .........................................................................................3-22
d. Unable to Calibrate .....................................................................................................3-24
3-9 System & Network I/O Module Controls (Setup) – System SIO .............................................3-25
a. Analog Output Setup...................................................................................................3-26
b. Serial interface setup ..................................................................................................3-28
c. Relay Outputs Setup ...................................................................................................3-29
3-10 System & Network I/O Module Controls (Setup) – System DIO ...........................................3-30
3-11 System Configuration and Diagnostics ..................................................................................3-30
a. Diagnostic Menus .......................................................................................................3-31
b. Load/Save Module Configuration ...............................................................................3-33
c. Date and Time.............................................................................................................3-33
d. Security Codes............................................................................................................3-34
e. System Reset..............................................................................................................3-35
3-12 Converter Temperature Adjustment.......................................................................................3-35
3-13 Measurement of Converter Efficiency ....................................................................................3-36
a. Test Setup for Measurement of Conversion Efficiency ..............................................3-36
b. Test Procedure ...........................................................................................................3-36
c. Subnormal Conversion Efficiency ...............................................................................3-38
d. Replacement of Converter ..........................................................................................3-38
e. Capillaries ...................................................................................................................3-38
f. TEA Scrubber ..............................................................................................................3-38
Model NGA2000 WNX and WCLD
4.0 MAINTENANCE AND SERVICE – WET NOX .....................................................................4-1
4-1 Overview...................................................................................................................................4-1
4-2 Power Fuse ..............................................................................................................................4-1
4-3 Peristaltic Pump .......................................................................................................................4-3
a. Peristaltic Pump - Replacement .................................................................................4-3
b. Replacing Peristaltic Pump Tube................................................................................4-4
4-4 Oven Components ...................................................................................................................4-5
a. Converter Thermostat .................................................................................................4-5
b. Manifold Thermistor and Case Thermistor .................................................................4-5
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 WNX and WCLD
c. Temperature Sensor ...................................................................................................4-5
d. Converter .....................................................................................................................4-6
e. Oven Fan ....................................................................................................................4-8
f. Regulator/Manifold Assembly ......................................................................................4-9
4-5 LON/Power Module ..................................................................................................................4-11
4-6 Power Module ..........................................................................................................................4-11
4-7 Electronics Assembly ...............................................................................................................4-11
a. Computer Board..........................................................................................................4-13
b. Module Function Board...............................................................................................4-13
4-8 Exhaust Fan .............................................................................................................................4-14
a Exhaust Fan Screen ....................................................................................................4-14
4-9 Intake Fan.................................................................................................................................4-15
4-10 Dehumidifier ...........................................................................................................................4-16
5.0 MAINTENANCE AND SERVICE – WET CLD......................................................................5-2
5-1 Power Fuse Replacement........................................................................................................5-2
5-2 Fans..........................................................................................................................................5-4
a. Fan Replacement........................................................................................................5-4
b. Cleaning EMI Filters....................................................................................................5-4
5-3 Transistor Assembly Replacement ..........................................................................................5-4
5-4 Ozone Generator......................................................................................................................5-4
a. Preventive Maintenance .............................................................................................5-4
b. Replacing ....................................................................................................................5-4
5-5 Ozone Generator Power Supply Replacement......................................................................5-4
5-6 Replacing Printed Circuit Boards .............................................................................................5-6
a. Computer Board..........................................................................................................5-6
b. Signal Board ...............................................................................................................5-7
c. Driver Board ................................................................................................................5-7
d. Power Supply Board ...................................................................................................5-9
e. LON/Power Module.....................................................................................................5-10
5-7 Prom .........................................................................................................................................5-11
5-8 Pressure Switch Replacement .................................................................................................5-12
5-9 Detector ....................................................................................................................................5-13
a. Removing Detector from Analyzer Module .................................................................5-14
b. Replacing Detector Components................................................................................5-14
5-10 Vent and Pump Capillaries.....................................................................................................5-20
5-11 Sample and Ozone Pressure Sensors...................................................................................5-20
5-12 Cleanup of WCLD due to Carryover of Water from WNX Module. ........................................5-21
Instruction Manual
760009-B
November 2002
6.0 REPLACEMENT PARTS ......................................................................................................6-1
6-1 Matrix........................................................................................................................................6-1
6-2 WNX .........................................................................................................................................6-2
a. 656715 Converter .......................................................................................................6-2
b. 656769 Dehumidifier...................................................................................................6-2
6-3 WCLD .......................................................................................................................................6-2
a. Detector.......................................................................................................................6-2
7.0 RETURN OF MATERIAL......................................................................................................7-1
7-1 Return of Material.....................................................................................................................7-1
7-2 Customer Service.....................................................................................................................7-1
7-3 Training.....................................................................................................................................7-1
8.0 INDEX....................................................................................................................................8-1
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
760009-B November 2002
Figure 1-1. WNX Module Flow System ..........................................................................................1-2
Figure 1-2. WNX Module – Major Components .............................................................................1-3
Figure 1-3. WCLD Module Flow Diagram ......................................................................................1-4
Figure 1-4. WCLD Module – Major Components ...........................................................................1-5
Figure 2-1. Rear Panel Connections of the WNX Module..............................................................2-4
Figure 2-2. Rear Panel Connections of the WCLD Module ...........................................................2-4
Figure 2-3. WNX Intake Fan Location ............................................................................................2-4
Figure 2-4. Electrical Connections..................................................................................................2-5
Figure 2-5. WNX - Outline and Mounting Dimensions ...................................................................2-6
Figure 2-6. WCLD - Outline and Mounting Dimensions .................................................................2-7
Figure 2-7. WNX - Wiring Diagram.................................................................................................2-8
Figure 2-8. WCLD – Wiring Diagram............................................................................................2-9
Figure 3-1. Measure Mode Display ................................................................................................3-1
Figure 3-2. The Display Screen......................................................................................................3-3
Figure 3-3. Changing Variables......................................................................................................3-3
Figure 3-4. Function Confirmation..................................................................................................3-3
Figure 3-5. Main Menu Functions...................................................................................................3-4
Figure 3-6. Module Manufacturing Data Displays ..........................................................................3-5
Figure 3-7. Main Menu Sub Menus ................................................................................................3-6
Figure 3-8. Startup Display.............................................................................................................3-7
Figure 3-9. Basic Controls Menu ....................................................................................................3-9
Figure 3-10. Display Controls Menu...............................................................................................3-10
Figure 3-11. Analyzer and I/O Expert Controls and Setup Menu...................................................3-11
Figure 3-12. Expert Controls Menu ................................................................................................3-11
Figure 3-13. Range Settings Menu ................................................................................................3-12
Figure 3-14. Physical Measurements Menu...................................................................................3-12
Figure 3-15. Concentration Alarm Setup Menu..............................................................................3-13
Figure 3-16. Linearization Parameters Menu .................................................................................3-13
Figure 3-17. Linearity Coefficients Menu........................................................................................3-14
Figure 3-19. Polynomial Setup Menu .............................................................................................3-14
Figure 3-20. Gas Concentrations Menu .........................................................................................3-15
Figure 3-21. Midpoint Correction Setup Menu ...............................................................................3-15
Figure 3-22. Response time/delay Parameters..............................................................................3-16
Figure 3-23. Automatic Range Control Menu.................................................................................3-17
Figure 3-24. Actual Switch Levels Menu ........................................................................................3-17
Figure 3-25. Display Units Menu ....................................................................................................3-17
Figure 3-26. Physical Measurements Menu...................................................................................3-18
Figure 3-27. Pressure Limits Menu ................................................................................................3-18
Figure 3-28. Temperature Limits Menu ..........................................................................................3-18
Figure 3-29. Physical Measurements – Manufacturer’s settings Display ......................................3-19
Figure 3-30. Displayed Parameters Menu......................................................................................3-19
Figure 3-31. Calibration Gas List Menu..........................................................................................3-21
Figure 3-32. Calibration Parameters Display .................................................................................3-21
Figure 3-33. Analyzer Zero Display ................................................................................................3-22
Figure 3-34. Expert Controls Menu ................................................................................................3-22
Figure 3-35. Zero/Span Calibration Menu ......................................................................................3-23
Figure 3-36. Analyzer Zero Menu...................................................................................................3-23
Figure 3-37. Zero/Span Diagnostic Data Menu..............................................................................3-23
Figure 3-38. Calibration Factors Menu...........................................................................................3-24
Model NGA2000 WNX and WCLD

LIST OF ILLUSTRATIONS

iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 WNX and WCLD
Figure 3-39. Range Factors Menu..................................................................................................3-24
Figure 3-40. Range Factors Display...............................................................................................3-24
Figure 3-41. System & Network I/O Module Controls Menu ..........................................................3-25
Figure 3-42. System SIO Module Menu .........................................................................................3-25
Figure 3-43. Analog Output Setup Menu........................................................................................3-26
Figure 3-44. Analog Output Setup Menu........................................................................................3-26
Figure 3-45. Signals Menu ............................................................................................................3-26
Figure 3-46. Output Signal If Assigned Module Fails Menu...........................................................3-27
Figure 3-47. Special Scaling for Concentration Signal Menu.........................................................3-28
Figure 3-48. Analog Output Updates per Second Menu ................................................................3-28
Figure 3-49. Serial Interface Setup Menu ......................................................................................3-28
Figure 3-50. AK Protocol Definitions Menu ....................................................................................3-29
Figure 3-51. Relay Outputs Setup Menu........................................................................................3-29
Figure 3-52. Choose Source Module Menu ...................................................................................3-29
Figure 3-53. Choose Signal Menu..................................................................................................3-29
Figure 3-55. System Configuration and Diagnostics Menu............................................................3-30
Figure 3-56. Diagnostic Menu ........................................................................................................3-31
Figure 3-57. Analyzer Diagnostics Menu .......................................................................................3-31
Figure 3-58. Power Supply Voltages Menu....................................................................................3-31
Figure 3-59. Primary Variable Parameters Menu...........................................................................3-31
Figure 3-60. Physical Measurements Menu...................................................................................3-32
Figure 3-61. Temperature Control Menu (Screen 1)......................................................................3-32
Figure 3-62. Temperature Control Menu (Screen 2)......................................................................3-32
Figure 3-63. NO/NOx Flow Balance Menu.....................................................................................3-32
Figure 3-64. Calculate Factor Using Pressure Ratio Menu............................................................3-33
Figure 3-65. Load/Save Configuration (CM/MCA) Menu ...............................................................3-33
Figure 3-66. Date and Time Menu..................................................................................................3-33
Figure 3-67. Security Codes Menu.................................................................................................3-34
Figure 3-68. Define Basic Level Security PIN Menu ......................................................................3-34
Figure 3-69. System Reset Menu...................................................................................................3-35
Figure 3-70. Measuring Efficiency of NO
Figure 3-71. Conversion Efficiency as a Function of Converter Temperature ...............................3-40
Figure 4-1. WNX - Power Fuse Location........................................................................................4-1
Figure 4-2. WNX - Removing Module Cover..................................................................................4-3
Figure 4-3. WNX - Replacing Peristaltic Pump ..............................................................................4-3
Figure 4-4. WNX - Replacing Peristaltic Pump Tube .....................................................................4-4
Figure 4-5. WNX - Removing Converter Thermostat, Manifold Thermistor Temperature Sensor
and Converter..............................................................................................4-6
Figure 4-6. WNX Converter Assembly ...........................................................................................4-7
Figure 4-7. WNX - Removal of Oven Fan ......................................................................................4-8
Figure 4-8. WNX - Removing Rear Panel ......................................................................................4-9
Figure 4-9. WNX - Removing Regulator/Manifold Assembly .........................................................4-10
Figure 4-10. WNX - Regulator/Manifold Assembly.........................................................................4-11
Figure 4-11. WNX - Removing LON/Power Module, Power Module and Electronics Assembly...4-12
Figure 4-12. WNX - Electronics Assembly (Exploded View)..........................................................4-13
Figure 4-13. WNX - Removal of Exhaust Fan and Screen ............................................................4-14
Figure 4-14. WNX - Removal of Intake Fan ...................................................................................4-15
Figure 4-15. WNX - Removal of Dehumidifier ................................................................................4-16
Figure 5-1. WCLD Power Fuse Location........................................................................................5-4
Figure 5-2. WCLD - Removing Cover ............................................................................................5-3
Figure 5-3. WCLD - Removal of Intake and Exhaust Fans ............................................................5-3
Figure 5-4. WCLD - Removal of Ozone Generator, Ozone Generator Power Supply, and
Transistor Assembly ....................................................................................5-6
to NO Converter...........................................................3-39
2
Instruction Manual
760009-B
November 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Contents v
Instruction Manual
760009-B November 2002
Figure 5-5. WCLD - Removing Computer Board and Bracket From Analyzer Module..................5-6
Figure 5-6. WCLD – Removing Computer Board From Bracket ....................................................5-7
Figure 5-7. WCLD – Removing Signal Board, Driver Board ..........................................................5-8
Figure 5-8. WCLD - Removing Power Supply/LON Assembly from Analyzer Module ..................5-9
Figure 5-9. WCLD - Power Supply Board/LON Power Module Assembly .....................................5-10
Figure 5-10. WCLD - PROM Location on Computer Board ...........................................................5-11
Figure 5-11. WCLD - Removing Pressure Switch and Detector from Analyzer Module................5-12
Figure 5-12. WCLD Detector – Cutaway View ...............................................................................5-13
Figure 5-13. WCLD - Removing Heater Thermostat Assembly from Detector ..............................5-14
Figure 5-14. WCLD - Removing Reaction Chamber and Sapphire Window from Detector...........5-15
Figure 5-15. WCLD - Removing the Photodiode from the Detector...............................................5-16
Figure 5-16. WCLD - Photodiode/Socket (Exploded View)............................................................5-17
Figure 5-17. WCLD - Capillary Assembly.......................................................................................5-18
Figure 5-18. WCLD - Removing the Sample Capillary from the Detector......................................5-19
Figure 5-19. WCLD - Replacing Pressure Sensors........................................................................5-20
Model NGA2000 WNX and WCLD
vi Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model NGA2000 WNX and WCLD
November 2002

PREFACE

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the NGA 2000 CLD and the System Accessories of the NGA 2000 System.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
760009-B
DANGER .
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure,
condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications P-1
Instruction Manual
760009-B November 2002
Model NGA2000 WNX and WCLD

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be im­paired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel. For safety and proper per­formance this instrument must be connected to a properly grounded three-wire source of power.
WARNING .
TOXIC AND OXIDIZING GAS
This module generates ozone which is toxic by inhalation and is a strong irritant to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.
The module's exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and may contain other constituents of the sample gas which may be toxic. Such gases include various ox­ides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion re­actions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of consciousness, and death.
Avoid inhalation of the ozone produced within the module, and avoid inhalation of the sample and exhaust products transported within the module. Avoid inhalation of the combined exhaust prod­ucts at the exhaust fitting.
Keep all tube fittings tight to avoid leaks The user is responsible for testing for leakage only at the inlet and outlet fittings on the rear panel. Periodically, the user should do an internal leak test (with a test procedure chosen by the user).
Connect rear exhaust outlet to outside vent with stainless steel or Teflon line. Check vent line and connections for leakage.
P-2 Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model NGA2000 WNX and WCLD
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
WARNING.
OVERBALANCE HAZARD
This analyzer module may tip instrument over if it is pulled out too far and the Platform is not prop­erly supported.
WARNING.
INTERNAL ULTRAVIOLET LIGHT
Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at the ultraviolet source. Use of ultraviolet filtering glasses is recommended.
November 2002
760009-B
WARNING.
TOXIC CHEMICAL HAZARD
The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure. Mercury is highly toxic if absorbed through the skin or ingested, or if vapors are inhaled.
Handle lamp assembly with extreme care. If the lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill.
Immediately clean up and dispose of the mercury spill and lamp residue as follows:
Wearing rubber gloves and goggles, collect all droplets of mercury by means of a suction pump
and aspirator bottle with a long capillary tube. (Alternately, a commercially available mercury spill clean-up kit, such as J.T. Baker product No. 4439-01, is recommended.)
Carefully sweep any remaining mercury and lamp debris into a dust pan. Carefully transfer all
mercury, lamp residue and debris into a plastic bottle which cab be tightly capped.
Label and return to hazardous material reclamation center. Do not place in the trash, incinerate
or flush down the sewer.
Cover any fine droplets of mercury in non-accessible crevices with calcium polysulfide and sul-
fur dust
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications P-3
Instruction Manual
760009-B November 2002
Do not place hands or fingers in the Platform front handles when front panel is open. Dropping the front panel of the Platform while hand or fingers are inside either handle can cause serious injury.
This module requires periodic use of pressurized gas. See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5
STATIC SENSITIVE COMPONENTS
Circuit boards in this instrument are static-sensitive. Take all static precautions when handling them.
Model NGA2000 WNX and WCLD
WARNING.
HAND INJURY HAZARD
CAUTION .
PRESSURIZED GAS
CAUTION.
NOTICE
Software compatibility is necessary for all NGA 2000 components in your system to work together. The version of your Platform's software must be equal to or greater that the version of any other module(s) for successful compatibility. If it is not, contact Rosemount Analytical at 800-433-6076 to order software upgrade kit PN NL657150 for the Platform.
You can locate the version of each NGA 2000 component as follows:
Platform Controller Board Turn power ON.
The display will show "Control Module V2. ...". This is the software version.
Analyzer Module Located on the right side of the Analyzer Module case.
I/O Module Located on the backplane connector of the module. If no label is present, the module is Version
2.0.
P-4 Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
760009-B
Model NGA2000 WNX and WCLD
November 2002
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when lo­cated in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suit­able hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is at­tached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125 permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precau­tions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should never be
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications P-5
Instruction Manual
9
6
760009-B November 2002
Model NGA2000 WNX and WCLD

DOCUMENTATION

The following NGA 2000 WNX and WCLD instruction materials are available. Contact Customer Service Center or the local representative to order.
760009 Instruction Manual (this document)

COMPLIANCES

The Wet NOx Conversion Module and the Wet CLD Analyzer Module carry approvals from several certifying agencies, including the Canadian Standards Association (which is also an OSHA accredited, Nationally Recognized Testing Laboratory), and TUV for use in non­hazardous, indoor locations.
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe.
These products comply with the standard level of NAMUR EMC. Recommendation (May
1993).
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand.
®
97-C219
NAMUR
N
P-6 Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model NGA2000 WNX and WCLD
November 2002

GLOSSARY OF TERMS

Analyzer Module
The module that contains all sensor/detector components for development of a Primary Variable signal; in­cludes all signal conditioning and temperature control circuitry.
Backplane
The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module power and net­work cables, I/O Modules and Expansion Modules plug into.
Control Module
The Operator Interface plus the Controller Board.
Controller Board
The Computer Board that serves as the Network Manager and operates the Display and Keypad.
Distribution Assembly
760009-B
The Backplane and the card cages that hold I/O and Expansion Modules.
Expansion Module
A circuit board that plugs into the Backplane from the front of the Platform and performs special features not related to I/O functions.
I/O Module
A circuit board that plugs into the Backplane from the rear of the Platform. Has a connector terminal for communication with external data acquisition devices and provides an input/output function.
Operator Interface
The Display and Keyboard.
Platform
Any workable collection of the following: Controller Board, Power Supply, Distribution Assembly, Enclosure and Operator Interface.
Power Supply
Any of a variety of components that provides conditioned power to other NGA 2000 components, from the Power Supply Board that plugs into the front of the Backplane in a stand-alone instrument to several larger ones that can power larger collections of modules and components.
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
Data placed on the network by a module regarding current status, e.g., sample flow, source voltage and other diagnostic information.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications P-7
Instruction Manual
760009-B November 2002
Softkeys
The five function keys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software.
System
Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s).
Model NGA2000 WNX and WCLD
P-8 Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 WNX and WCLD
SECTION 1

DESCRIPTION AND SPECIFICATIONS

Instruction Manual
760009-B
November 2002

1-1 OVERVIEW

This manual describes the Wet NOx (WNX) Analysis System, a member of Rosemount Analytical's NGA 2000 Series of gas analysis components. The system consists of two modules: (1) a WNX Conversion Module (P/N
195009) and a WET CLD (WCLD) Chemilu­minescense Analyzer Module (195110).
The WCLD Module P/N 195110 contains only the functionality that is required in order to supplement the functionality of the WNX Module. An existing CLD Module P/N 195005 can be used in place of the WCLD Module P/N 195110 after being suitably modified in the field. This five-step modification disables unrequired functionality. It is described in Sec­tion 2-2.
The WNX Module (Figure 1-1 and Figure 1-2) accept a hot, wet sample stream, cools it, and dehumidifies it for use by the WCLD Analyzer Module (Figure 1-3 and Figure 1-4). The hot sample is introduced to a converter (a heated bed of vitreous carbon) which converts all ox­ides of nitrogen (NO The converter can be bypassed if the sample contains just NO. Once the sample contains only NO, it is directed to the dehumidifier, a Peltier device located outside of the heated compartment. The water vapor is cooled and condensed, and the condensate is removed periodically with a peristaltic pump. The cool,
x) to nitric oxide (NO).
dry sample gas is sent to the WCLD for analysis.
The WCLD Module uses chemiluminescense detection technology, a technique that is based on the reaction between nitric oxide (NO) and ozone (O oxygen (O2). Some of the NO2 molecules thus produced are in an electronically excited state (NO
2*). These revert immediately to the
ground state, with emission of photons (es­sentially, red light). The reactions involved are:
NO + O
NO
The intensity of the emitted red light is propor­tional to the concentration of NO in the origi­nal gas sample and is measured by a photodiode.

1-2 TYPICAL APPLICATIONS

The major application of the WNX System is analysis of internal combustion engine ex­haust.
3 NO2* + O2
2* → NO2 + red light
3) to produce NO2 and
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
Y
g
760009-B November 2002
SAMPLE OUT
FLOW BALANCE
EXHAUST GAS
EXHAUST LIQUID
SAMPLE IN
RESTRICTOR
Needle Valve
Restrictor 70 cc/min.
Model NGA2000 WNX and WCLD
M
Out
F
MANIFOLD
ASSEMBL
Pressure Re
ulator
Figure 1-1. WNX Module Flow System
Solenoid Valve
B
G
H
Top
CONVERTER
DEHUMIDIFIER
L
In
Drain
PERISTALTIC
PUMP
Top
K
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 WNX and WCLD
Peristaltic Pump
Power Module
LON/Power Module
Computer Board
Module Function Board
Converter
Manifold Assembly
Module cover and oven cover removed for clarity.
Figure 1-2. WNX Sample Conversion Module – Major Components
Instruction Manual
760009-B
November 2002
Intake Fan
Dehumidifier
Exhaust Fan
Rear Panel – See Figure 2-1 for connections
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Instruction Manual
PRESSURE
PUMP CAPILLARY
SAMPLE CAPILLARY
PRESSURE SENSOR
760009-B November 2002
OZONE AIR
REDUCING BULKHEAD 1/4T - 1/8T BRASS
SWITCH
CONNECTOR 1/8T - 1/8FPT BRASS
TEE 1/8T BRASS
EXHAUST
REDUCING BULKHEAD 1/4T - 1/8T SS
PORT CONNECTOR 1/8T CRES
TUBE ADAPTER BRASS
TEE 1/8T CRES
CONNECTOR 1/8T - 3/8MPT BRASS
UNION 1/8T SS
REDUCING BULKHEAD 1/4T - 1/8T SS
SAMPLE
Note: For the SAMPLE INLET, a 1/4T - 1/8T Reducer Fitting is supplied in the ship kit.
Model NGA2000 WNX and WCLD
OZONE
GENERATOR
RESTRICTOR 1/8T - 1/8MPT
PRESSURE SENSOR
1/8 OD TEFLON
CROSS 1/8T SS
CONNECTOR 1/8T - 3/8MPT SS
UNION 1/8T SS
UNION 1/8T SS
REDUCING UNION 1/8T - 1/16T SS
70 cc/min OR 200 cc/min
OZONE
EXHAUST
REACTION CHAMBER
SAMPLE
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Figure 1-3. WCLD Module Flow Diagram
Model NGA2000 WNX and WCLD
Instruction Manual
760009-B
November 2002
Driver Board
Ozone Generator
Sample Pressure Sensor
Ozone Pressure Sensor
Signal Board
Inlet/Outlet Fittings
Fans
Detector
Sample Capillary
Pressure Switch
Power Supply Board
Flow Balance Needle Valve
Computer Board
Figure 1-4. WCLD Module – Major Components
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
760009-B November 2002
1-3 SPECIFICATIONS – WET NOX CONVERSION MODULE

a. Performance

Measurement Species:.................. NO, NOx
Ranges: ......................................... 0 to 10 ppm through 0 to 10,000 ppm NOx maximum
Delay Time: ................................... 5 seconds, maximum (Does not include response time of compan-
Ambient Temperature:................... 0°C to 45°C (32°F to 113°F)
Power Requirements:.................... 24V DC 192W
For the following parameters, the Wet NOx Sample Conditioning Module and the Wet CLD Analyzer Module together shall meet the performance specifications of the companion CLD Module.
Repeatability Response Time Minimum Detectable Level Drift Noise Effect of Temperature Linearity

b. Physical

Model NGA2000 WNX and WCLD
ion CLD)
Case Classification:....................... General purpose for installation in weather-protected area
Dimensions:................................... See Figure 2-5
Weight: .......................................... 10.6 kg (23.5 lbs.)
Mounting:....................................... Custom-installed in a panel (not used in a Platform)
Max. Length of LON Cable:........... 1600m (1 mile)

c. Sample

Temperature: ................................ 70°C or less
Flow Rate: ................................... 750 to 2200 ml/min.
Pressure ........................................ 344 to 620 hPa ( 5 to 9 psig)
Particles: ...................................... Filtered to <2 microns
Moisture: .........................................less than 15% water vapor by volume, no entrained liquid
Materials in contact with Sample:.. Stainless steel, Teflon, glass

d. Gas Connections

Sample In ...................................... 1/4” O.D. tube fitting, stainless steel
Exhaust Gas .................................. 1/4” O.D. tube fitting, stainless steel
Sample Out.................................... 1/8” O.D. tube fitting, stainless steel
Exhaust Liquid............................... 1/4” O.D. tube fitting, stainless steel
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 WNX and WCLD

1-4 SPECIFICATIONS - WET CLD ANALYZER MODULE

a. Performance

See the Preface section of the Platform Components manual for specifications regarding Platform­related components (e.g., case dimensions) and the Preface of the I/O Module manual for specifica­tions regarding I/O (e.g., relay outputs).
Measurement Species................... NO
Ranges .......................................... 0 to 10 ppm through 0 to 10,000 ppm; 500 ppm NO
Repeatability.................................. ±0.5% of fullscale (at constant temperature)
Minimum Detectable Level ............ 0.1 ppm
Noise ............................................. <1% of fullscale, peak to peak
Linearity ......................................... ±1% of fullscale
Response Time <3 sec. for 90% of fullscale for ranges of less than 25 ppm
Drift (at constant temperature) ...... Zero - ±1% of fullscale/24 hours, <2% of fullscale/week
Span - <±1% of fullscale/24 hours, <3% of fullscale/week
Effect of Temperature.................... <2% of fullscale (over any 10°C ambient temperature variation in-
Ambient Temperature.................... 10°C to 40°C (50°F to 104°F)
Power ............................................ 24V DC 70W

b. Physical

1
............................ <1 sec. for 90% of fullscale for ranges of 25 ppm or greater
terval for a rate of change no greater than 10°C/hour)
Instruction Manual
760009-B
November 2002
Case Classification ........................ General purpose for installation in weather-protected area
Maximum Length of LON Cable.... 1600m (1 mile)
Dimensions.................................... See Figure 2-6
Weight ........................................... 8.1 kg (18 lbs.)
Mounting........................................ Inside a Platform or custom-installed in a panel

c. Sample

Temperature .................................. 0°C to 55°C (32°F to 131°F)
Flow rate........................................ Approximately 400 ml/min.
Pressure ........................................ 482 hPa (±138 hPa), 7 psig (±2 psig)
Particles ........................................ Filtered to <2 microns
Dewpoint........................................ 5.5°C below ambient temperature, no entrained liquid
Materials in contact with Sample... Stainless steel, glass, Teflon, Kynar
Ozonator Gas ................................ Clean, dry air or oxygen; flow rate: 1 L/min.
Max. pressure: 1702 hPa to 2047 hPa (10 to 15 psig)
1
Does not include delay time introduced by Wet NOx Conversion Module
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
Instruction Manual
760009-B November 2002

d. Gas Connections

Ozone Air....................................... 1/4” O.D. tube fitting, stainless steel
Exhaust.......................................... 1/4” O.D. tube fitting, stainless steel
Sample
1
........................................ 1/8” O.D. tube fitting, stainless steel
Model NGA2000 WNX and WCLD
1
A 1/4T - 1/8T Reducer is supplied in the ship kit
1-8 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 WNX and WCLD
SECTION 2

INSTALLATION

Instruction Manual
760009-B
November 2002

2-1 UNPACKING

If either the WNX Sample Conversion Module or the WCLD Chemiluminescense Analysis Module is received as a separate unit, care­fully examine the shipping carton and con­tents for signs of damage. Immediately notify the shipping carrier if the carton or contents is damaged. Retain the carton and packing ma­terial until all components associated with the module are operational.
2-2 MODIFICATION OF A CONVENTIONAL CLD
FOR USE IN A WNX SYSTEM
A WCLD Module, P/N 195110, or its’ func­tional equivalent, is required for use in a WNX System. Five steps must be taken to modify a conventional CLD module, P/N 195005, to be functionally equivalent. These five steps apply just to the CLD Module. Refer to the instruc­tions below and to the manual for that prod­uct.
Step 1 - There must not be a flow restrictor in the SAMPLE IN fitting of the CLD. If there is, the bulkhead fitting in which it is contained must be removed and replaced with a similar bulkhead fitting that does not contain a flow restrictor. Then, for the remainder of this preparation, temporarily place a restriction in the sample line to the CLD to limit flow to less than 2200 cc/min. This restriction must be re­moved before plumbing into a WNX System.
Step 2 - With zero gas flowing, adjust the back-pressure regulator that controls capillary head pressure in the CLD to 5 - 6 psig (344 ­413 hPa). Do not change this adjustment later.
Step 5 - Now that the CLD is operating prop­erly, disconnect power from it, turn off all gases, and remove the restriction that was placed in the sample line. The CLD is then ready for use in a WNX System.

2-3 ASSEMBLY

If the WCLD module requires assembly with other components (e.g., a Multi Module En­closure), do so at this time. Following the guides on the bottom of the enclosure, care­fully slide the module halfway into place. The WNX module may not be used in a Platform.
CAUTION.
HAND INJURY HAZARD
Do not place hands or fingers in enclosure front handles when the front panel is open. Dropping front panel while hand or fingers are inside either handle can cause serious injury.
Lift the spring-loaded pins on the front of the module, and carefully slide it the rest of the distance. Secure the module in position by re­leasing the pins, which seat in the available holes in the bottom of the case. Connect one of the network cables to either NETWORK 1 or NETWORK 2 connections on the front panel of the module. Connect one of the power cables to the power connection on the front panel also. (See Figure 2-4.)
Secure the front panel with the six screws provided.

2-4 LOCATION

Step 3 - Select the NO mode. Do not change this adjustment later.
Step 4 - Verify that the CLD responds properly to an appropriate span gas.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Install the modules in a clean, level, weather-proofed, vibration-free location free from extreme temperature variations and moisture. For best results, install the Wet NOx Module near the sample stream and the CLD Module near the Wet NOx Module to minimize sample transport time.
Instruction Manual
760009-B November 2002
Model NGA2000 WNX and WCLD
Operating ambient temperature is 10°C to 40°C (50°F to 104°F). Temperature change should not exceed 10°C (18°F) per hour. The same restriction on temperature range applies to the location of the air and span gas cylin­ders.
NOTE
Unrestricted air flow at the rear of each module is critical to its performance and reliability (See Figure 2-1 and Figure 2-2).
Unrestricted air flow at the left side (air in­take) of the WNX is critical to its perform­ance and reliability (Figure 2-3).

2-5 GASES

a. Overview

The WNX Module is used in concert with a WCLD Analyzer Module, either a stan­dard CLD or one specially configured for use with a WNX Module. In either case, certain functions are retained by the CLD Module while other functions are as­sumed by the WNX Module, and these functions determine gas connections to each module.

b. Pneumatic Connections

All pneumatic connectors on both mod­ules are ferrule-type compression fittings.
The SAMPLE OUT connection of the WNX Module is 1/8-inch. The SAMPLE IN connection of the CLD Module is either 1/4-inch (WCLD P/N 191110 and CLD P/N 194005). The 1/4-inch to 1/8-inch re­ducing fitting supplied with the WCLD (or CLD) should be used, as close as practi­cal to the CLD Module. Interconnect the two modules using 1/8-inch teflon or stainless steel tubing. The tubing should be as short as practical.
Connect lines for SAMPLE-IN, GAS EX­HAUST, and EXHAUST LIQUID to ap­propriately labeled fittings on the rear panel of the WNX Module. A heated line with materials of construction that are suitable for the sample must be used for the sample line to the WNX Module. The remaining lines may be teflon or stainless steel.
Connect lines for AIR/OXYGEN and EX­HAUST to appropriately labeled fittings on the rear panel of the CLD Module.
The hot, wet sample and all calibration standard gases are introduced to the WNX Module at the SAMPLE IN connec­tion. The dry sample gas emerging from the SAMPLE OUT connection of the WNX Module is conveyed to the CLD SAMPLE IN connection via appropriate tubing.
The CLD retains all ozonator functions, including introduction of breathing grade air or oxygen at its OZONATOR AIR con­nection.
Calibration gases and ozonator-source gas should be supplied from a cylinder equipped with a clean, non-corrosive, two-stage regulator. A shutoff valve is recommended.
The exhaust lines should be 1/4 inch (6.3 mm) or larger. These lines must not be restricted.

c. Specifications

Calibration Gases
Either U.S.P. breathing grade air or nitro­gen is recommended for use as zero gas. A mixture of NO or NO of nitrogen is recommended as span gas. For maximum accuracy, the concentration of NO in the span gas should be about the same as that in the expected sample stream.
Sample
The sample to the WNX Module must be clean and hot before entering the WNX Module. Sample should be filtered for par­ticulates down to two microns. The sam-
2 in a background
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 WNX and WCLD
Instruction Manual
760009-B
November 2002
ple to the CLD Module should be clean and dry.
airborne contaminants, the user should periodically do an internal leak test (with a test procedure chosen by the user).
Ozonator Source Gas
For analyzers with ranges less than 0 to 2,500 ppm, U.S.P. breathing grade air should be used for generation of the ozone required for the chemilumines­cence reaction. For ranges greater than 0 to 2,500 ppm, breathing grade oxygen is required.
Should be introduced at the AIR/OXYGEN fitting of the CLD Module at between 690 and 1035 hPa-gauge (10 and 15 psig) of ozonator air (or oxygen, if appropriate)
Pressure
Zero gas, span gas and sample should be introduced at the SAMPLE IN fitting of the WNX Module at between 345 and 620 hPa-gauge (5 and 9 psig). Bypass flow rate should be approximately 1000 cc/min. with backpressure regulator pres-
Limits and health effects of some of the airborne contaminates are shown below.
Federal health standards are established to provide an adequate margin of safety in protecting public health. As specified by Federal air quality standards, the maxi­mum permissible outdoor air level is:
Ozone (O
3) (one hour average):
Not to exceed 0.12 ppm (one hour average) more than one day per year.
Nitrogen Oxides (NOx) includes Nitric Oxide (NO) and Nitrogen Dioxide (NO
2).
Nitrogen Oxides make up a major part of PM
10 - fine particle pollution
(less than 10 micrometers in size). Nitric Oxide (NO) exposed to Ozone (O
3) produces Nitrogen Dioxide
(NO
2) and Oxygen (O2).
sure adjusted to 344 hPa (5 psig).
Nitrogen Dioxide (one year average):
NOTE
At no time should flow of zero gas, span gas, or sample exceed 2200 cc/min. Damage to internal com­ponents may occur if this flow level is exceeded.
NOTE
At no time should ozonator gas pres­sure exceed 2070 hPa-gauge (30 psig). Damage to internal components may occur if this pressure level is ex­ceeded.
Not to exceed 0.053 ppm for one year average
Evidence suggests that repeated expo­sure to high ozone levels may cause irre­versible damage to the lung. Health effects of ozone are complicated by the presence of other pollutants, such as ni­tric acid, PM
10 and nitrogen dioxide. In
animals, ozone mixed with acids has been found to be more harmful than ozone alone. The harm depends on the situation: The person who is exposed to pollution (health, age, sex, nutritional and smoking status, sensitivity), the type and
Leak Test
Each module is completely tested at the factory for gas leakage. During installa­tion, the user is responsible for leak test­ing only at the inlet and outlet fittings on the rear panel. To prevent exposure to
concentration of pollutant, the place, time, temperature and humidity. Inhaled fine particles pose a health risk when they penetrate deeply into respiratory tract. Most particle penetration into lungs occur during mouth breathing.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
760009-B November 2002
Model NGA2000 WNX and WCLD
SAMPLE OUT
EXHAUST LIQUID
FLOW BALANCE
SAMP LE IN
EXHAUST GAS
CONVERT. VENT
RESTRICTOR
EXHAUST FAN
Figure 2-1. Rear Panel Connections of the WNX Module
OZONATOR AIR
EXHAUST
SAMPLE IN
EXHAUST FAN
INTAKE FAN
Figure 2-2. Rear Panel Connections of the WCLD Module
INTAKE FAN
Figure 2-3. WNX Intake Fan Location
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 WNX and WCLD

2-6 ELECTRICAL CONNECTIONS

Two electrical connections are required on each module: POWER and NETWORK. These connections are located on the front panel of each module, Figure 2-4, Figure 2-5 and Figure 2-6. On each module, two
Instruction Manual
760009-B
November 2002
NETWORK connectors are available, either of which is appropriate for: 1) interconnection with the network or 2) "daisy-chaining" with other NGA 2000 components. Connect enclo­sure power or external 24 VDC power source to the POWER connection of each module. (The WNX Module may use as much as 8 A power).
NETWORK1
NETWORK2
POWER
FUSE
Figure 2-4. Electrical Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
760009-B November 2002
FRONT VIEW
Model NGA2000 WNX and WCLD
[213]
[208]
[173]
[15]
[13]
SIDE VIEW
.5
[13]
[109]
[28]
22.4
[572]
[72]
[152]
1.3
[33]
REAR VIEW
Dimensions:
INCH
[MM]
3.5
[88]
1.4 [37]
SAMPLE
EXHAUST LIQUID
FLOW BALANCE
SAMPL E IN
EXHAUST GAS
CONVERT. VENT
RESTRICTOR
7.3
[185]
4.2
[106]
4.1
[104]
.6
[14]
1.7
[43]
2.7
[68]
3.6
[92]
Figure 2-5. WNX - Outline and Mounting Dimensions
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Loading...
+ 94 hidden pages