The information contained in this document is subject to change without notice.
Rosemount Analytical's system of NGA 2000 Modular Gas Analyzers and Controllers are patented, under U.S. Patent
5.787.015.
Viton-A® is a registered trademark of E.I. duPont de Nemours & Co.
Paliney No. 7™ is a trademark of J.M. Ney Co.
Manual Part Number 748419-B
June 2000
Printed in U.S.A.
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
Table 3-2. Calibration Range for Various Zero Based Operating Ranges.........19
Table 3-3. Calibration Range for Various Suppressed Range Operations........19
Table 3-4. Oxygen Equivalents of Common Gases...........................................23
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
iii
CONTENTS
NOTES
Paramagnetic Detector Analyzer Module
iv
Rosemount Analytical NGA 2000
748419-B
June 2000
PREFACE
PURPOSE/SAFETY SUMMARY
The purpose of this manual is to provide the procedures for the installation, operation and
maintenance of this NGA 2000 module.
Some sections may describe equipment not used in your NGA 2000 system configuration.
The user should become thoroughly familiar with the operation of this module before
operating it. Read this instruction manual completely.
To avoid explosion, loss of life, personal injury and damage to this equipment
and on-site property, all personnel authorized to install, operate and service
the this equipment should be thoroughly familiar with and strictly follow the
instructions in this manual. SAVE THESE INSTRUCTIONS.
If this equipment is used in a manner not specified in these instructions,
protective systems may be impaired.
DANGER is used to indicate the presence of a hazard which will cause severe personal
injury, death, or substantial property damage if the warning is ignored
WARNING is used to indicate the presence of a hazard which can cause severe personal
injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor
personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation, or maintenance information which is
important but not hazard-related.
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without covers secure. Servicing requires access to live parts
which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a
properly grounded three-wire source of power.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P1
PREFACE
WARNING: POSSIBLE EXPLOSION HAZARD
This equipment is not designed for and should not be used in the analysis of
flammable samples. Use of this equipment in this way could result in explosion
and death.
NOTE
Apply leak test liquid to cell or detectors only as a last resort.
WARNING: POSSIBLE EXPLOSION HAZARD
Verify that all gas connections are made as labeled and are leak free. Improper
gas connections could result in explosion or death. See Section 2.4.3 for Leak
test procedure.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect
safety of this product. Use only factory documented components for repair
WARNING: OVER-VOLTAGE SPIKING
If this Analyzer Module is used with a non-Rosemount Analytical power supply,
adding Rosemount Analytical PN 90331 Current Protector in series with the 24 V
positive line will prevent over-voltage spiking and resultant fuse blowing when
powering up the instrument.
CAUTION: PRESSURIZED GAS
This module requires periodic calibration with a known standard gas. See
General Precautions for Handling and Storing High Pressure Gas Cylinders in
the rear of this manual.
Paramagnetic Detector Analyzer Module
P2
Rosemount Analytical NGA 2000
748419-B
June 2000
PREFACE
CAUTION: HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is
open. Dropping front panel while hand or fingers are inside either handle can
cause serious injury.
CAUTION: OVERBALANCE HAZARD
This Analyzer Module may tip instrument over if it is pulled out too far and the
Platform is not properly supported.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P3
PREFACE
GLOSSARY
Analyzer Module
Self contained analysis modules that are designed to be installed into the NGA 2000
System. One Analyzer Module can be installed into a Single Enclosure containing the
Platform Module. Two Analyzer Modules can be installed into a Dual Enclosure. The
simplest NGA 2000 System consists of one Analyzer Module.
Backplane
The interconnect circuit board which the Controller Board, Power Supply Board, I/O
Board(s) and Expansion Board(s) are plugged into the Backplane
Control module
The operator interface plus the Controller Board.
Controller Board
The Controller Board in the Platform which runs the software program that operates the
Display, Keypad and Network Manager. The Controller Board plugs into the Backplane.
Distribution Assembly
The Distribution Assembly consists of the Backplane and the card cages in the Platform
Module that contain I/O Module(s) and Expansion Module(s).
Expansion Board
The Expansion Board performs special features not related to I/O functions. The Expansion
Board plugs into the Backplane from the Platform front.
I/O Module
An auxiliary module that provides some sort of interface to the outside world. I/O modules
may include analog outputs, relay contacts, and digital interfaces. In general, they are
mounted in platforms as options.
Operator Interface
The Display and Keyboard of the Platform.
Paramagnetic Detector Analyzer Module
P4
Rosemount Analytical NGA 2000
748419-B
June 2000
PREFACE
Platform
Any combination of the NGA case, the display and computer board, power supply, and I/O
modules. In general, it could be considered to be anything in the NGA system other than
the analyzer modules.
Power Supply
Any of a variety of components that provide conditioned power to other NGA 2000
components, from the Power Supply Board that plugs into the Backplane in a stand-alone
instrument to several larger ones that can power larger collections of modules and
components.
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
The current status data placed on the network by an Analyzer Module. This includes
sample flow, source voltage and other diagnostic information.
Softkeys
The five function keys located below the front panel display. The menu function for each
softkey is displayed directly above it and is controlled by the software.
System
A NGA 2000 System consisting of one (or more) Analyzer Modules, an optional Platform,
one or more optional I/O Boards, an optional Expansion Board and an optional
Supplemental Power Supply.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P5
PREFACE
SPECIFICATIONS
MEASUREMENT SPECIES:
R
ANGES
:
REPEATABILITY:
M
INIMUM DETECTABLE LEVEL
NOISE:
L
INEARITY
R
ESPONSE TIME
D
RIFT (ZERO AND SPAN
E
FFECT OF TEMPERATURE
E
NVIRONMENT
A
MBIENT TEMPERATURE
E
FFECT OF FLOW
P
OWER REQUIREMENTS
:
:
):
:
:
:
Oxygen
0 to 100% oxygen; four fullscale selections, including
suppressed zero ranges in 1% increments
±1% of fullscale (at constant temperature)
0.01% oxygen
:
<1% of fullscale, peak-to-peak; <±1% for suppressed ranges
±1% of fullscale
0 to 90% of fullscale in 20 seconds (±2 seconds)
<±1% of fullscale/24 hours, <±2% of fullscale/week at constant
temperature;
<±2% of fullscale/24 hours, <±4% of fullscale/week of range for
99 to 100% (at constant temperature)
:
:
<±1% of fullscale over any 10°C interval for rate of change no
greater than 10°C per hour
Location - Class B controlled, indoor, non-hazardous
0 to 45°C (32 to 113°F)
<±1% of range when sample flow rate is changed by 20 ml/min.
24 VDC ±5%, 50 W max.; ripple and noise: <100 mV
peak-to-peak; line and load regulations: <±1%
SPECIFICATIONS – SAMPLE
T
EMPERATURE
F
LOW RATE
E
XHAUST PRESSURE
P
ARTICLES
D
EWPOINT
M
ATERIALS IN CONTACT WITH
S
AMPLE
Paramagnetic Detector Analyzer Module
P6
Rosemount Analytical NGA 2000
:
:
:
:
:
Non-flammable;: 10 to 66°C (50 to 150°F)
800 to 1400 ml/min.
:
-345 to 690 hPa-gauge (-5 to 10 psig)
filtered to <2 microns
below 43°C (110°F), no entrained liquid
Glass, 316 stainless steel, titanium, Paliney No. 7, epoxy
resin, Viton A, platinum, nickel, rhodium and MgF
2
748419-B
June 2000
SPECIFICATIONS - PHYSICAL
PREFACE
C
ASE CLASSIFICATION
D
IMENSIONS
W
EIGHT
:
:
MOUNTING:
M
AXIMUM LENGTH OF
C
ABLE
:
:
LON
General purpose for installation in weather-protected
areas
See Outline and Mounting Dimensions, Figure 2-4
8 kg (17.6 lbs.
Inside a Platform or custom-installed in a panel
1600 m (1 mile) between Analyzer Module and Platform
See the Preface section of the Platform Components manual for specifications regarding
Platform-related components (e.g., case dimensions) and the Preface of the I/O Module
manual for specifications regarding I/O (e.g., relay outputs).
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P7
PREFACE
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE
For order administration, replacement Parts, application assistance, on-site or factory
repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when you
call: Model Number, Serial Number, and Purchase Order Number or Sales Order Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive or
other hazardous material or used in such a hazardous environment, the user must attach
an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the
material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
TRAINING
A comprehensive Factory Training Program of operator and service classes is available.
For a copy of the Current Operator and Service Training Schedule contact the Technical
Services Department at:
Rosemount Analytical Inc.
Phone: 1-718-986-7600
FAX: 1-714-577-8006
Paramagnetic Detector Analyzer Module
P8
Rosemount Analytical NGA 2000
748419-B
June 2000
PREFACE
9
6
DOCUMENTATION
The following Paramagnetic Detector Analyzer Module instruction materials are available.
Contact Customer Service or the local representative to order.
748419 Instruction Manual (this document)
COMPLIANCES
This product may carry approvals from several certifying agencies, including Factory Mutual
and the Canadian Standards Association (which is also an OSHA accredited, Nationally
Recognized Testing Laboratory), for use in non-hazardous, indoor locations
®
NRTL /C
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to
harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May
1993).
NAMUR
This product satisfies all obligations of all relevant standards of the EMC framework in
Australia and New Zealand.
97-C219
748419-B
June 2000
N
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P9
PREFACE
NOTES
P10
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
INTRODUCTION
1.1 OVERVIEW
This manual describes the Paramagnetic Detector (PMD) Analyzer Module of Rosemount
Analytical's NGA 200 Series of gas analysis components.
The PMD Analyzer Module is designed to continuously determine the concentration of
oxygen in a flowing gaseous mixture. The concentration is expressed in ppm or percent
volume O2.
The entire Analyzer Module is designed as a slide-in module (if configured in stand-alone
instrument fashion), removable from the front of the Platform, with gas connections made
from the rear. All electronics relative to sample detection and conditioning are included in
this module.
1.2 TYPICAL APPLICATIONS
1
PMD Analyzer Module applications include:
• process control
• continuous emissions monitoring systems (CEMS)
• industrial gas production
• fermentation process monitoring
1.3 THEORY OF TECHNOLOGY
Oxygen is strongly paramagnetic (i.e., capable of becoming a temporary magnet when
placed in a magnetic field) while most other common gases are weakly diamagnetic (i.e.,
tend to be non-magnetic). See Figure 1-1.
The Magnetic susceptibility of the flowing gas sample is sensed in the detector/magnet
assembly. As shown in Figure 1-2, a dumbbell shaped, nitrogen -filled, h ollow gas test body
is suspended on a platinum/nickel alloy ribbon in a non-uniform magnetic field.
Because of a "magnetic buoyancy" effect, the spheres of the test body are subjected to
displacement forces, resulting in a displacement torque proportional to the magnetic
susceptibility of the gas surrounding the test body.
Measurement is accomplished by a null-balance system, whereby the displacement torque
is opposed by an equal and opposite restorative torque. The restoring current is
automatically maintained at the correct level by an electro-optical feedback system. A beam
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
1
INTRODUCTION
1
of light from the source LED is reflected off the square mirror attached to the test body onto
a bi-cell (dual photodiode).
The current required to keep the test body to the null position is a linear function of the total
magnetic susceptibility of the sample gas.
See Figures 4-1, 4-2, and 4-3 for component configuration.
Shaded
Pole
Piece
Sphere
(Magnetic Susceptibility = ko )
F
k
Sample Gas
(Magnetic Susceptibility = k )
Note:
) increases.
k
IGURE
F
1-1. S
As percentage of oxygen in sample gas increases,
displacement force (F
PHERICAL BODY IN NON-UNIFORM MAGNETIC FIELD
1.4 FEATURES
Among the features incorporated into the PMD Analyzer Module is a flow splitter (≈2:1)
that allows for greater sample flow, decreased lag time and faster analyzer response.
The "Time Alignment" feature can be used to delay sending the Primary Variable from
the PMD Analyzer Module for up to 30 seconds in 0.1 second intervals. This feature
allows Primary Variables form more than one PMD Analyzer Module to be "time
aligned" if necessary.
Paramagnetic Detector Analyzer Module
2
Rosemount Analytical NGA 2000
748419-B
June 2000
q
y
Axis
A
Displacement
q
ASSEMBLY
Tor
INTRODUCTION
ue
1
Balancing
Weight
Electromagnetic
Platinum/Nickel Alloy
Suspension Ribbon
TEST BODY DETAIL
Displacement
Torque
Restoring
ue
Tor
Restoring
Current
Mirror
Restoring
Current
Restoring
Torque
Current Loop
Electromagnetic
xis
Balancing Weight
Nitrogen-Filled Hollow Glass
Test Bod
IGURE
F
Magnet
Shaded Pole Pieces (4)
Dual Photocell
1-2. P
ARAMAGNETIC DETECTOR TECHNOLOGY
Test Body
Restoring
LED Source
DETECTOR/MAGNET
Zero
NULL BALANCE SYSTEM
SIGNAL CONDITIONING
Span
SIGNAL OUTPUT
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
3
INTRODUCTION
1
NOTES
Paramagnetic Detector Analyzer Module
4
Rosemount Analytical NGA 2000
748419-B
June 2000
INSTALLATION
2.1 UNPACKING
If the Paramagnetic Analyzer Module is received as a separate unit, carefully examine the
shipping carton and contents for signs of damage. Immediately notify the shipping carrier if
the carton or contents is damaged. Retain the carton and packing material until all
components associated with the Analyzer Module are operational.
2.2 ASSEMBLY
If the Analyzer Module requires assembly with other components (e.g., the Platform and
associated I/O Modules), do so at this time. Following the guides on the bottom left and
bottom center of the Platform, carefully slide the Analyzer Module halfway into place.
2
WARNING: HAND INJURY HAZARD
Do not place hands or fingers in the Platform front handles when front panel is
open. Dropping the front panel of the Platform while hand or fingers are inside
either handle can cause serious injury.
ANALYZER MODULE GUIDES
PIN SEATS
IGURE
F
748419-B
June 2000
2-1. A
DISENGAGED FRONT PANEL
NALYZER MODULE INSTALLATION INTO INSTRUMENT PLATFORM
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
5
2
INSTALLATION
Lift the spring-loaded pins on the front of the Analyzer Module, and carefully slide it the rest
of the distance. Secure the module in position by releasing the pins, which seat in the
available holes in the bottom of the case (see Figure 2-1). If the module and Platform are
difficult to assemble, remove the module, ensure the top cover of the module is firmly
seated on the hold-down screws, and repeat the assembly procedure.
Install I/O Module(s) according to guidelines in the I/O manual. After startup and calibration
have been performed, secure the front panel with the six screws provided.
2.3 LOCATION
Install the Analyzer Module in a clean, non-hazardous, weather protected, vibration free
location free from extreme temperature variations. For best results, either install the module
near the sample stream to minimize sample transport time or supply a flow greater than
necessary and route only the appropriate amount through the Analyzer Module.
Observing these requirements are critical. Note the following:
• Excessive vibration can cause a noisy readout. To minimize vibration effects, the
detector/magnet assembly is enveloped in a shock-mounted compartment.
• The user should ensure, when making any internal electrical connections, that no
cables are placed in contact with the detector assembly or associated internal sample
inlet and outlet tubing.
• Magnetic susceptibilities and partial pressures of gases vary with temperature.
Permissible ambient temperature range is 32°F to 113°F (0°C to 45°C).
• The interior of the Detector Assembly is maintained at approximately 144°F (62°C) by
an electronically controlled heater. Prior to entering the detector assembly, the sample
is heated in a coiled tubing to match the detector's temperature.
Paramagnetic Detector Analyzer Module
6
Rosemount Analytical NGA 2000
748419-B
June 2000
2.4 GASES
INSTALLATION
2
2.4.1 R
Calibration Gases
Analyzer Module calibration requires the establishment of zero and span calibration points.
This requires a zero standard gas to set the zero point span gas to establish a calibration
point at or near the upper range limit.
An oxygen-free gas, typically nitrogen, is required for use as the zero standard gas.
Recommendations for span calibration gases, bases on various operating ranges, are
tabulated in Table 3-1. Air (20.93% oxygen) can be used as span gas regardless of the
ranges used for sampling, although very low ranges may lose accuracy.
Sample Gas
Sample gas should be non-flammable.
Temperature
Sample temperature at the inlet should be from 50°F to 150°F (10°C to 66°C). A maximum
entry temperature of 110°F (43°C) is recommended to prevent cooling of the sample and
possible internal condensation. Such condensation could damage some components of the
Analyzer Module. This recommendation can be ignored if a thoroughly dry sample is
examined.
EQUIREMENTS
Pressure
Sample exhaust pressure limits are -5 to 10 psig (-345 to 690 hPa-gauge). Normal
operation is in the positive range, between 0 and 10 psig (0 and 690 hPa-gauge). Negative
gauge pressures are not normally recommended, but may be used in certain special
applications.
To prevent over-pressurization, insert a pressure relief valve into the sample inlet line. A
check valve should also be placed in the outlet line if the Analyzer Module is connected to a
manifold associated with a flare or other apparatus that does not operate at atmospheric
pressure.
The outlet port is commonly vented to the atmosphere. Any change in barometric pressure
has a directly proportional effect on the indicated percent of oxygen, and should be
neutralized through manual or computer correction of data. Note the following example:
Range = 0% to 5% oxygen
Barometric pressure change after calibration = 1%
The error is more significant for suppressed range 99% to 100%.
An optional barometric pressure compensation board is available to automatically perform
this correction.
A general rule regarding calibration gas pressure is that it should be the same as the
expected sample gas pressure during routine operation.
The above requirement increases the difficulty of operation at negative gauge pressure. A
suction pump can be connected to the outlet port for drawing sample through the Analyzer
Module. Such operation necessitates special precautions to ensure accurate readout,
including the following:
The need for equilibrium between sample and gas calibration pressures.
Any leakage in the sample handling system will decrease readout accuracy.
Flow Rate
Recommended sample flow rate is 800 to 1400 ml/min., ±40 ml/min. Optimum flow rate is
1100 ml/min.
If flow is held to within tolerance and operating pressure remains constant, zero and span
drift will meet specified limits.
Paramagnetic Detector Analyzer Module
8
Rosemount Analytical NGA 2000
748419-B
June 2000
INSTALLATION
(
)
NETWORK 1
NETWORK 2
POWER
FUSE
2
IGURE
F
Note: Reference and purge gas connections are applicable only to certain applications.
IGURE
F
2-2. PMD F
2-3. PMD B
RONT PANEL CONNECTIONS
FAN
ACK PANEL CONNECTIONS
SAMPLE
10 PSI MAX
69 kPa MAX
OUT
IN
2.4.2 C
ONNECTIONS
(See Figure 2-3) Connect inlet and outlet lines for sample gas to appropriately labeled
fittings on the rear panel. Both connections are 1/4 inch ferrule-type compression fittings.
Zero and span gases use the same inlet and outlet as the sample. Figure 2-4 shows a
typical external sample handling manifold for gas selection. Particulates must be filtered
down to two microns, gases generally require pressurization, and flow measurement
metering MUST be present.
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
9
2
INSTALLATION
Needle
Valves
PMD Analyzer
Module
IGURE
F
Sample In
Zero
Standard
Gas
Span
Standard
Gas
2-4. I
Two Micron
Filter
Flowmeter
NTERCONNECTION OF TYPICAL GAS MANIFOLD TO
ODULE
M
Flow Splitter
(≈2:1)
PMD A
To Vent
NALYZER
10
2.4.3 L
EAK TEST
The Analyzer Module is thoroughly tested at the factory for gas leakage. The user is
responsible for testing for leakage only at the inlet and outlet fittings on the rear panel. The
user is also responsible for internal leak testing periodically and if any internal pneumatic
components are adjusted or replaced (with a test procedure selected by the user).
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
INSTALLATION
2.5 ELECTRICAL CONNECTIONS
NOTE
Electrical connections must be in compliance with National Electrical Code
(ANSI/NFPA 70) and/or any applicable national or electrical codes.
Two electrical connections are required on the Analyzer Module; POWER and NETWORK.
See Figure 2-2. On the Analyzer Module, two NETWORK connections are available, either
of which is appropriate for : 1) interconnection with Backplane of the Platform (see Platform
instruction manual) or 2) "daisy chaining" with other NGA 2000 components.
Connect Analyzer Module POWER 24 VDC power source, either the Platform or external
power source.
2
J5
J10
FAN
J11
J13
J14
J19
FLOW SENSOR CABLE
PRESSURE
COMPENSATION
BOARD
J6
COMPUTER ANALYSIS
J7
W1
PREAMP CABLE ASSEMBLY
CASE TEMP SENSOR
BOARD
J4
FLOW
SENSOR
J17
J1
J2
J3
J16
MAIN BOARD
J20
SENSOR
ASSEMBLY
J21J22
LED SOURCE
THERMAL CUTOFF
MAGNET HEATER AND
OPTICAL BENCH
ASSEMBLY
J9
J12
J15
IGURE
F
748419-B
June 2000
2-5. PMD W
IRING DIAGRAM
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
11
2
INSTALLATION
A
[213]
[208]
[109]
[28]
20.0
[508]
[157]
[152]
17.41
[442]
[15]
[13]
C
2.8
[71]
D
E
1.6
[40]
.6
[15]
8.4
[213]
E. FUSE.
D. POWER CABLE TO NETWORK.
C. NETWORK CABLE CONNECTIONS TO PLATFORM.
B. SAMPLE OUT: 1/4" O.D. TUBE FITTING.
A. SAMPLE IN: 1/4" O.D. TUBE FITTING.
5. MODULE TO BE INSTALLED WITHIN ±15° OF HO RIZONTAL.
4. POWER REQUIREMENTS: 24 VDC 3.5 A.
3. ELECTRICAL INSTALLATION MUST BE IN COMPLIANCE WITH NATIONAL ELECTRICAL
CODE (ANSI/NFPA 70) AND/OR ANY APPLICABLE NATIONAL OR LOCAL CODES.
2. MODULE IS NOT WEATHERPROOF.
1. APPROXIMATE WEIGHT: 15 LB (6.8 kg).
IGURE
F
2-6 O
UTLINE AND MOUNTING DIMENSIONS
6.7
[170]
B
10 PSI MAX
1.0
[25]
1.6
[41]
DIMENSIONS
INCH
[mm]
Paramagnetic Detector Analyzer Module
12
Rosemount Analytical NGA 2000
748419-B
June 2000
STARTUP AND OPERATION
3.1 OVERVIEW
Prior to initial startup, the user should perform the leak test procedure outlined in Section 2.
For the remainder of this section Analyzer Module interconnection with a Platform or some
interfacing component will be a ssumed. Display and keypad information shall ref er to that
which the user can expect to see and do with regard to the front panel of the Platform.
For a complete description of the Platform front panel controls and indicators, see the
Platform instruction manual.
3.2 DISPLAYS
Three kinds of Display screens are available to the user:
3
• Run Mode
• Menu
• Help
3.3 R
The Run Mode is the normal mode of operation. In this mode, the Display (see Figure 3-1)
will show current gas measurement, the component of interest, the current operations of the
softkeys, and a graphic bar representing the displayed concentration as ppm or as a
percent of oxygen.
If more than one Analyzer Module is connected to the system, the Run Mode display will
show as many as four gas measurements on screen. Alarm messages may also appear on
the display (See Table 3-1).
3.4 M
UN MODE DISPLAY
ENU DISPLAYS
The menu structure enables the user to access data and functions, and put information onto
the network.
The Main Menu (Figure 3-2) is subdivided into three levels of control based generally on
which personnel is likely to use it: Basic Controls, Expert Controls and Setup, and
748419-B
June 2000
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
13
Loading...
+ 65 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.