Rosemount Manual: NGA 2000 HFID Analyzer Module (combined with NGA 2000 Platform, MLT or CAT 200) | Rosemount Manuals & Guides

Instruction Manual
HAS64E-IM-SW39 9/2010
Software Version 3.9.x
NGA 2000 Software Manual for HFID Analyzer Module (combined with NGA 2000 Plat form, MLT or CAT 200)
www.EmersonProcess.com
READ THIS PAGE BEFORE PROCEEDING!
Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using and maintaining Emerson Process Management (Rosemount Analytical) products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson Process
Management (Rosemount Analytical) representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
T o ensure proper performance, use qualified personnel to inst all, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson Process Management (Rosemount Analytical). Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR W ARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
th
1
Edition 9/2010
Emerson Process Management GmbH & Co. OHG
Industriestrasse 1 D-63594 Hasselroth Germany T +49 (6055) 884 0 F +49 (6055) 884 209 www.emersonprocess.com

Contents

1 Introduction ………………………………………………………………………………………………… 1- 1 2 Menu Structure …………………………………………………………………………………………… 2- 1 3 Display and keyboard …………………………………………………………………………… ………. 3- 1
3.1 Starting and Initializing …………………………………………………………………………................. 3- 1
3.2 Display and operation ……………………………………………………………………………………… 3- 2
3.3 The keyboard ……………………………………………………………………………………................ 3 -2
3.4 Menu items ……………………………………………………………………………… ………………….. 3- 3
3.5 Common F-key functions ………………………………………………………………………………….. 3- 4
3.6 Entering and editing parameters ………………………………………………………………………….. 3- 5
3.7 Current measurement parameters ……………………………………………………………................. 3- 5
3.8 The main menu ……………………………………………………………………………………………... 3- 6
3.8.1 Control module manufacturing data ………………………………………………………………………. 3- 7
3.8.2 Analyzer module manufacturing data …………………………………………………………… ……….. 3- 8
4 Basic controls …………………………………………………………………………………………….. 4- 1
4.1 Measurement ……………………………………………………………………………………… ……….. 4- 3
4.1.1 Changing channels …………………………………………………… ……………………………………. 4- 5
4.1.2 Changing the order of the multi-component display ……………………………………………………. 4- 7
4.1.3 Changing the measurement range ……………………………………………………………………….. 4- 9
4.1.4 Automatic range change …………………………………………………………………………………… 4- 13
4.1.5 Lighting the flame ……………………………………………………………………… …………………… 4- 17
4.2 Calibration …………………………………………………………………………………………………… 4- 21
4.2.1 Zeroing ……………………………………………………………………………… ………………………. 4- 23
4.2.2 Spanning …………………………………………………………………………………………………….. 4- 27
5 Analyzer and I/O, expert controls and set-up ……………………………………………… ………... 5- 1
5.1 Analyzer module setup ..................................................................................................................... 5- 3
5.1.1 Calibration gas list ............................................................................................................................. 5- 5
5.1.2 Calibration parameters ...................................................................................................................... 5- 13
5.1.3 Concentration alarm setup ................................................................................................................ 5- 15
5.1.4 Gas measurement parameters.......................................................................................................... 5- 17
5.1.4.1 Linearization parameters .................................................................................................................. 5- 19
5.1.4.2 Response time/delay parameters...................................................................................................... 5- 23
5.1.4.3 Range setting..................................................................................................................................... 5- 25
5.1.4.4 Automatic range change control........................................................................................................ 5- 27
5.1.4.5 Units................................................................................................................................................... 5- 29
5.1.4.6 Linearization functions....................................................................................................................... 5- 31
5.1.5 Analyzer parameter list...................................................................................................................... 5- 33
5.1.6 Displayed parameters........................................................................................................................ 5- 35
5.2 System & network I/O module controls & setup ................................................................................ 5- 37
5.2.1 System SIO module........................................................................................................................... 5- 39
5.2.1.1 Analog output setup........................................................................................................................... 5- 41
5.2.1.2 Serial interface setup......................................................................................................................... 5- 47
5.2.1.3 Relay outputs setup........................................................................................................................... 5- 49
5.2.2 System DIO module .......................................................................................................................... 5- 51
5.2.3 E/A module ………........................................................................................................................... 5- 55
5.3 Analyzer module controls .................................................................................................................. 5- 57
5.3.1 Physical measurements .................................................................................................................... 5- 59
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
1
6 System configuration and diagnostics.......................................................................................... 6- 1
6.1 Diagnostic menus ............................................................................................................................. 6- 3
6.1.1 Control module diagnostics ............................................................................................................... 6- 5
6.1.2 Analyzer module diagnostics ............................................................................................................ 6- 7
6.1.2.1 Power supply voltages ...................................................................................................................... 6- 9
6.1.2.2 Primary variable parameters.............................................................................................................. 6- 11
6.1.2.3 Physical measurement parameters .................................................................................................. 6- 13
6.1.2.4 Temperature control parameters ...................................................................................................... 6- 15
6.1.2.5 Miscellaneous control parameters .................................................................................................... 6- 17
6.1.2.6 Auto ignition parameters.................................................................................................................... 6- 19
6.1.2.7 Self test results ................................................................................................................................. 6- 21
6.1.2.8 Software diagnostics.......................................................................................................................... 6- 23
6.1.2.9 Analyzer start up ............................................................................................................................... 6- 25
6.2 Load/Save configuration (CM/MCA) ................................................................................................. 6- 27
6.3 Date and time..................................................................................................................................... 6- 29
6.4 Security codes .................................................................................................................................. 6- 31
6.5 Network module management........................................................................................................... 6- 33
6.6 Measurement display setup............................................................................................................... 6- 37
6.7 Miscellaneous ................................................................................................................................... 6- 39
7 Display controls .............................................................................................................................. 7- 1
Supplement:
Calculator on Control Module Level (CM Calculator) ................................................................... Page 1 - 14
Programmable Logic Control on Control Module Level (CM PLC)...............................................Page 1 - 28
System Calibration ...........................................................................................................................Page 1 - 34
2
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10

Introduction: 1

This software handbook describes the individual steps for successfully operating the HFID analyzer module and HFID analyzer (analyzer module in a platform) from the Emerson Process Management NGA 2000 series.
Chapter 2 contains an overview of the HFID software menu structure. ⇒ Chapter 3 introduces the user interface (display and keys) and the main menu with its
submenus.
Chapter 4 describes the basic functions (measurement and calibration) in detail and step by
step.
Chapter 5 describes the expert level for the configuration of the module and any I/O modules. ⇒ Chapter 6 describes the system configuration and diagnostics. ⇒ Chapter 7 contains information about the display controls.
Not all the contents of Chapter 5 are relevant to all users. Which sub-chapters are relevant depends on the configuration of the NGA 2000 system with regard to the following components:
Control module CM Analyzer module AM Input/ Output Modules I/O (SIO = Standard Input/ Output;
DIO = Digital Input/ Output) Network I/O Modules Analog output with 3 alarms I/O; Auto-calibration I/O (“Autocal”); System Auto-calibration I/O (“Syscal”)
The following table gives the possible SIO/ DIO configurations:
System unit SIO/DIO Configuration Chapter
HFID analyzer module (AM)
No front panel, i.e. no control unit. Can be combined with a platform, an MLT analyzer,
a TFID analyzer or a customer-specific control unit
Platform (CM Software)
Control unit with front panel. No measurement channels.
HFID analyzer (CM and HFID analyzer module software)
HFID analyzer: HFID AM in a platform with front panel
HFID analyzer module combined with MLT/ TFID analyzer, i.e. all control unit and HFID analyzer module functions are displayed.
1 local SIO and 1 local DIO (or 2 local DIOs) can be built into an MLT AM or a TFID AM only.
SIO and DIO in the AM can only be configured for the MLT/TFID channel.
No local HFID I/Os
1 SIO and up to 4 DIOs can be built into the platform.
SIO and DIO can be configured for all AMs connected to the platform, e.g. for the HFID.
System I/Os
1 SIO and 1 DIO (or 2 DIOs) can be built in to the MLT/ TFID analyzer (CM I/Os).
SIO and DIO can be configured for all AMs connected to the MLT/ TFID analyzer, e.g. HFID
System I/Os
See MLT/ TFID manual.
5.2
5.2
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
1 - 1
NGA 2000 system with platform and HFID analyzer module
(additional manual)
NGA 2000 system with MLT analyzer and HFID analyzer module
(additional manual)
1- 2
NGA 2000
MLT ANALYZER
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10
Introduction: 1
This manual concerns all HFID analyzer modules combined with a platform or an MLT analyzer. This manual is also relevant for the CAT 200, which consists of an MLT 1 in an Ex d housing which is operated via a touch-screen front panel or a magnetically operated front panel. There is a separate manual for the CAT 200 which includes all CAT-related issues.
HFID analyzer modules operated via a customer-specific control unit are not described in this manual.
Network I/O modules have their own manuals which should be consulted for further detail. In this manual, these I/O modules are only very briefly described.
Further questions should be directed to our Service Support Center.
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
1 - 3
1- 4
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10

Menu structure: 2

- Channel => F3
- Reset … => F2
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
2 - 1
2- 2
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10

Display and keyboard: 3

3.1. Starting and initializing

On being powered up, the HFID analyzer module (whether on a platform or as part of an NGA network) runs through the initializing phase. This also includes a self-test of the analyzer(s) or analyzer module(s). The LCD screen displays the company name and software version number as well as a series of messages which indicate the status of the initialization cycle. Finally the single component display of the HFID is displayed, which serves as a starting point from which the main menu, the analyzer module status menu and, if appropriate, the multi-channel display of the analyzer network can be accessed.
(C) 2005 Emer s on P r oc ess Man agement
NGA 2000 Control-Modul e Rev.3. 9.4 / P017
Lang uage: P 017 / 01 / 00
Init ializing Network
Installing networ k interface
LCDReset Abort
F1
F2
F3 F4
F5
All the instructions for the basic controls (chapter 4) begin at the single component display. The actual display may vary from the illustrations in this manual, since it can be configured by the client.
HFID_WO1000001
6.28 ppm THC
0 Range: 2 25
Sample pressure: Oven temperature: Flame status: Raw signal:
Display Status... Main... Channel BasicCal
232.5 hPa 191 C
ON
531756
50 47
100000
900000
490
61
Analyzer tag
Can be set to any value. This is the factory setting: “WO” stands for “Work Order”
F1 F2 F3 F4 F5
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
3 - 1

3.2. Display and operation

4
The measurement display and all operational steps are carried out via the LCD screen. The keyboard itself comprises 5 function keys, 4 arrow keys and an Enter key. This handbook uses simplified sketches to represent the user interface:
HFID_WO1000001 6.28 ppm
Analyzer basic controls (calibration) & setu p...
Analyzer and I/O, expert controls & setup ... System
configu ration and diagnostics...
Display
controls...
______
_______________________________________
Time &
Date:
System
tag:
Measure Status... Channel Lock... MFG Data
F1 F2 F3 F
-- Main Menu --
03:11:05 August 16, 2009
Emerson
F5
The functions of the individual keys and the operation of the unit depend on:
the configuration of the analyzer or analyzer module the optional modules (e.g. I/O modules) installed the menu currently displayed.
If the power supply is interrupted, all user-specific module parameters are saved in a battery­operated buffer.

3.3 The keyboard

HFID_WO1000001
6.28 ppm THC
0 Range: 2 25
Sample pressure: Oven temperature: Flame status: Raw signal:
Display Status… Main... Channel BasicCal
F1 F2 F3 F4 F5
232.5 hPa 191 C
ON
531756
50 47
100000
490
61
900000
Arrow keys
-key and -key: Move up or down within a menu; Change or edit a setting.
-key and -key: Change to previous or next submenu or page; Select an individual digit or character.
Enter key
Confirm a selection; Execute the selected command; Change to the selected
Function keys: Function depends on the current menu and
menu.
is displayed on the LCD screen above each key.
3- 2
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10
Display and keyboard: 3

3.4 Menu items

A menu item can be selected with the -key or the -key. A selected item is shown in inverted colors (white text against a black background). There are four different types of menu items:
Submenu...
An item ending in three dots. Pressing the -key or the -key accesses a submenu.
Command line !
An item ending in an exclamation point. Pressing the -key or the -key executes the selected function.
Parameter:
An item ending in a colon. Displays a parameter. Some parameters can be changed, others are read-only. If the parameter is editable, pressing the -key or the -key puts the parameter into edit mode.
Text line
An item not ending in any of the above punctuation marks. For information only; cannot be edited.
Text line
Parameter
Submenu
Function key labels
HFID_WO1000001
Measurem ent range num ber:
Range upper limit: Span gas concentration: Aut omatic range change control: Sample flow: Ranges with valid calibrat io n: Cal ib ration status:
Flame condition: Light Flame...
HOME ESCAPE ZERO SPAN
F1 F2 F3 F4 F5
Basic Controls
6.28 ppm
1
10.0 ppm
8.6 ppm
Disabled
4301 ml/min
1&2&3&4
READY
ON
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
3 - 3
.
HFID_WO1000001
Command line
S yst em reset !
Measure.
F1
- -System R eset --
Are y ou sure ???
F2
F3 F4

3.5 Common F-key functions

Display
Change from the single component display to the multi-component display.
Measure
Change from the main menu to the single component display.
6.28 ppm
Back...
F5
Status
Change to the “Current measurement parameters” display for the most important parameters and status
information of the HFID module or the current channel.
Always F2, when this function is available.  ) ch. 3.7, page 3-5.
Main
Change from the single-component display to the main menu.
HOME
Change from a submenu to the main menu.
ESCAPE or Back
Change to the previous menu. Revert a changed, but not yet confirmed, parameter to its previous value.
Lock
Locks all levels for which security has been activated. F4 in the main menu.
Channel
Changes to a different channel within the same menu, if more than one channel is available. Each channel is
cycled through in turn.
In the main menu the channels of all analyzers are available, in submenus only the channel s of the currently
selected analyzer module.
3- 4
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10
Display and keyboard: 3
M

3.6 Entering and editing parameters

The Enter key
Pressing the Enter key when an editable parameter is selected will put that parameter into edit mode: only the value itself will be displayed white on black. Pressing the Enter key again will save the new value.
The ↑ -key and the ↓ -key:
Depending on the value being edited, these keys will:
change the numerical value scroll through a list of options change a letter or other character
Numbers with more than one digit will generally be incremented or decremented. Letters are changed one at a time.
The
-key and the → -key:
Select an individual digit or character. Certain numerical parameters can be edited one digit at a time.

3.7 Current measurement parameters

Pressing F2 (“Status”) in the single component display accesses the “Current measurement parameters” menu:
HFID_WO1000001
Current measurement parameters
St atus details. ..
Measurement range nu mb er : Rang e c ha nge contr ol: Flame c ondition: Linear iz ation mode:
Analyzer operati onal state: Analyzer alarm state: Alarm reporting lev el: Current total variable updates per second:
HOME ESCAPE
F1
F2
ORE
F3 F4
6.28 ppm
1
Local
ON
DISABLED
STANDB Y
NORMAL FAILURE
25
F5
In this menu the status of the HFID analyzer module can be checked. Pressing F3 (“MORE”) accesses a second page; press F2 (“ESCAPE”) or the -key to return to the first page.
Pressing the Enter key with the “Status details” line selected accesses further submenus listing error messages and many other details.
These menus are informative only. None of the settings here can be changed; this can only be done via the “Basic controls”, “Analyzer module setup” and “Analyzer module controls” menus (chapters 4, 5.1 and 5.3 respectively).
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
3 - 5

3.8 The main menu

Pressing F3 or the -key in the single component display accesses the main menu. From this menu all levels of the HFID analyzer or analyzer module can be accessed allowing measurement, calibration and data transfer parameters to be set.
HFID_W O1000001
Analyzer basic controls (calibration) & setup...
Analyzer and I/O, expert controls & setup... System configura tion and diagnostics... Display controls...
_____________________________________________ Time & Date: System tag:
Meas ure Status... Channel Lock... MFG Data
F1 F2 F3 F4 F5
-- Main Menu --
16:35:32 August 16, 2009
6.28 ppm
Emerson
See chapter 4
See chapter 5
See chapter 6
Siehe Kapitel 4
See chapter 7
To set: ch. 6.3, page 6-29
Factory setting
F1: Change to the single component display of the current channel F2: Change to the “Current measurement parameters” menu ) ch. 3.7, page 3-5 F3: Change to a different channel ) Channel Identifier F4: Lock level(s) using security codes ) ch. 6.4, page 6-31 F5: Change to the “Manufacturing Data” menu ) below
Pressing F5 accesses submenus in which important information about the control module and the analyzer module can be found:
HFID_WO1000001
Control modul e m anufacturing data.. .
Analyzer modul e m anuf ac turing data...
Measure <<< Back... >>>
F1
-- M anufacturing Data --
F2
F3 F4
6.28 ppm
F5
3- 6
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10

3.8.1 Control module manufacturing data

HFID_WO1000001
Control module manufacturing data...
An aly zer module manu factur ing da ta...
-- Manufacturing Data --
6.28 ppm
HFID_WO1000001
Display and keyboard: 3
-- Control Module Manufacturing Data --
Emerson Pr oc ess Management
Ma nufacturing GmbH & Co. OHG
D-635 94 Hasselroth / Germany
FAX. ( + 49) 6055 884 - 209
Copyright (c) 2005
Industriestrasse 1
Tel (+49) 6055 884 - 0
6.28 ppm
Measure <<< Back... >>>
F1
F2
F3 F4
F5
Measure Back... More
F1
HFI D_WO100 0001
-- Control Module Version Information --
Serial number: Manufacturing date:
Hardware revision: Soft ware revisio n: Revision date: Revision time:
Phrase dictionary version: Language:
Measure Back...
F1 F2 F3 F4 F5
F3 F4
F2
ACU02 R:3.9.0 D:Jul 17 2003
F5
6.28 ppm
CM1
20.05.2004
3.9.4 / P017
Nov 11 20 05
12:01:04
P017 / 01 / 00
English
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
3 - 7

3.8.2 Analyzer module manufacturing data

a
HFID_WO1000001
Control module manufacturing data...
An aly zer module manu factur ing da ta...
Measure <<< Back... >>>
F1
-- Manufacturing Data --
F3 F4
F2
HFID_WO1000001
-- Analyzer Module M anuf ac turing Data --
Emerson Pr oc ess Management
Ma nufacturing GmbH & Co. OHG
D-63594 Hasselroth / Germ any
FAX. ( + 49) 6055 884 - 209
Copyright (c) 2005
Industriestrasse 1
Tel. (+ 49) 6055 884 - 0
6.28 ppm
F5
6.28 ppm
HFID_WO1000001
More...
Minimum range: Maximum range: Measured gas: Capillary: User tag number:
HOME ESCAPE
F1 F2 F3 F4 F5
HFID_WO1000001
Serial number: Manufact uri n g date:
Hardwa re revisio n: Software revision: Revision date:
sion ti me:
Revi
Analyzer manufacturing dat
-- Analyzer Module Ver sion Information--
100 00 ppm
9.7 ml/min @ 3.5psig HFID_WO1000001
RESET STORE
02:28:1 0 Februar y 01, 2005
3.9.4 / P017
Nov 11 2005
6.28 ppm
10.0 ppm THC
6.28 ppm
HFD1
0.5
12:33:56
Measure Back... More...
F1
F2
F3 F4
F5
Measure Back...
F1 F2 F3 F4 F5
3- 8
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10
Basic controls: 4
Introduction
Chapter 4 describes the most important basic functions which are needed to configure a HFID analyzer module with an NGA front panel.
The following basic functions are available:
Measurement ) Chapter 4.1, page 4-3 Calibration ) Chapter 4.2, page 4-21
The basic settings are fully explained and illustrated step by step. The two-column layout shows the user interface and the LCD screen of the NGA front panel on the left. This enables the user to follow each step and compare the illustrations with the analyzer display screen. On the right, each step is briefly described. Each sub-chapter begins and ends with a single component display. The keys to be pressed for each step are highlighted in black.
See the example on the next page.
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/ 10
NGA 2000
4 - 1
Example: Changing from the single component display to the main menu
The first illustration shows the initial situation Pressing key F3 changes the display to that in the second illustration
HFID_WO1000001
6.28 ppm THC
0 Range: 1 10
Sample pressure: Oven temperature: Flame status: Raw signal:
Display Status... Main... Channel BasicCal
232.5 hPa 191 C
ON
531756
50 47
100000
F1 F2 F3 F4 F5
490
61
900000
1. Change to the main menu
Press F3 or the -key.
HFID_W O1000001
Analyzer basic controls (calibration) & setup...
Analyzer and I/O, expert controls & setup... System configura tion and diagnostics... Display controls...
_____________________________________________ Time & Date: System tag:
Meas ure Status... Channel Lock... MFG Data
-- Main Menu --
16:35:32 August 16, 2009
F1 F2 F3 F4 F5
6.28 ppm
Emerson
2. Any further instructions follow, e.g.:
Change to the basic controls
Press the -key or the -key.
4- 2
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/ 10
This subchapter describes the following functions:
Changing channels
) Chapter 4.1.1, page 4-5
Changing the order of the multi-component display
) Chapter 4.1.2, page 4-7
Changing the measurement range
) Chapter 4.1.3, page 4-9
Automatic range change
) Chapter 4.1.4, page 4-13
Lighting the flame
) Chapter 4.1.5, page 4-17
Basic controls: 4.1

Measurement

Chapters 4.1.1 and 4.1.2 are only relevant to systems with more than one measurement channel.
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
4 - 3
4- 4
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10
4
C
HFID_WO1000001
6.28 ppm THC
0 Range: 1 10
Sample pressure: Oven temperature: Flame status: Raw signal:
Display Status... Main... Channel BasicCal
F1 F2 F3 F4 F5
232.5 hPa 191 C
ON
531756
50 47
100000
490
61
900000
Basic controls: 4.1.1

Changing channels

1. Change to the multi­component display
Press F1.
Note: The multi-component display can be reached from any single component display by this method.
6.28
45.00
333.0
150.0
20.00
Select Status... Tags Off
F1 F2 F3 F
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
HFID
ppm TH
ppm CO
ppm SO2
ppm NO
%O2
6.28
>
45.00
333.0
150.0
20.00
M LT25/CH1/R2 M LT25/CH2/R2 M LT25/CH3/R2 M LT25/CH4/R2
HFID
ppm THC
ppm CO
ppm SO2
ppm NO
%O2
0 0
0 0
0
0 0 0 0
0
[1] [2]
[3] [4]
[5]
[1] [2]
[3] [4]
[5]
10 10
10 10
10
LCDReset
F5
10 10 10 10
10
2. Activate select function
Press F1 or the -key.
Note: If there are no other channels connected to the HFID, only the bargraph of the HFID itself will be displayed. A change of channels will then not be possible.
2. Select a channel
Press the or -key until the > symbol appears in the line you wish to select.
Select Status...
F1 F2 F3 F4 F5
Tags Off
LCDReset
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
4 - 5
3. Return to the single
6.28
45.00
>
333.0
150.0
20.00
HFID
ppm THC
M LT25/CH1/R2
ppm CO
M LT25/CH2/R2
ppm SO2
M LT25/CH3/R2
ppm NO
M LT25/CH4/R2
%O2
0 0 0 0
0
[1] [2]
[3] [4]
[5]
10 10 10 10
10
component display
Press F1 key .
Select Status...
F1 F2 F3 F4 F5
Tags Off
MLT 25/CH1/R2
45.28 ppm CO
0 Range: 2 250
Temperature: Failures: Check Requests: Operation:
Display Status... Main... Channel BasicCal
F1 F2 F3 F4 F5
51.4 C No No
Ready
0.0 100000
LCDReset
4. The single component display after selecting a new channel.
100
4- 6
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10

Changing the order of the multi-component display

HFID_WO1000001
6.28 ppm THC
0 Range: 1 10
Sample pressure: Oven temperature: Flame status: Raw signal:
Display Status... Main... Channel BasicCal
F1 F2 F3 F4 F5
232.5 hPa 191 C
ON
531756
50 47
100000
490
61
900000
Basic controls: 4.1.2
1. Change to the multi­component display
Press F1.
HFID
%O2
Tags Off
0 0 0 0
0
6.28
45.00
333.0
150.0
20.00
Select Status...
F1 F2 F3 F4 F5
ppm THC
M LT25/CH1/R2
ppm CO
M LT25/CH2/R2
ppm SO2
M LT25/CH3/R2
ppm NO
M LT25/CH4/R2
HFID
%O2
Tags Off
0 0 0 0
0
6.28
>
45.00
333.0
150.0
20.00
Select Status...
F1 F2 F3 F4 F5
ppm THC
M LT25/CH1/R2
ppm CO
M LT25/CH2/R2
ppm SO2
M LT25/CH3/R2
ppm NO
M LT25/CH4/R2
[1] [2]
[3] [4]
[5]
[1] [2]
[3] [4]
[5]
10 10 10 10
10
LCDReset
10 10 10 10
10
LCDReset
2. Activate select function
Press F1 or the -key.
Note: If no other channels are connected to the HFID, only the bargraph of the HFID itself will be displayed, and steps to change the order are superfluous.
2. Select a channel to appear in the first line
Press the or -key until the > symbol is next to the appropriate bargraph.
Example: MLT channel 2
) is to be displayed in the
(SO
2
first line.
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
4 - 7
O
C
3. Set the selected bargraph
6.28
45.00
333.0
>
150.0
20.00
HFID
ppm THC
M LT25/CH1/R2
ppm CO
M LT25/CH2/R2
ppm SO2
M LT25/CH3/R2
ppm NO
M LT25/CH4/R2
%O2
0 0 0 0
0
[1] [2]
[3] [4]
[5]
10 10 10 10
10
to the first line
Press the -key.
Select... Display
F1 F2 F3 F4 F5
Tags Off
333.0
6.28
45.00
150.0
20.00
Select Status...
F1
M LT25/CH2/R2
ppm SO2
ppm TH
M LT25/CH1/R2
ppm C
M LT25/CH3/R2
ppm NO
M LT25/CH4/R2
F2 F3 F4 F5
HFID
%O2
Tags Off
0 0 0 0
0
[1]
[2] [3]
[4]
[5]
LCDReset
10 10 10 10
10
LCDReset
4. The multi-component display after setting the order of bargraphs
It is now possible to select any single component display. See chapter 4.1.1 for further instructions.
4- 8
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10
Basic controls: 4.1.3

Changing the measurement range

HFID_WO1000001
6.28 ppm THC
0 Range: 1 10
Sample pressure: Oven temperature: Flame status: Raw signal:
Display Status... Main... Channel BasicCal
232.5 hPa 191 C
ON
531756
50 47
100000
F1 F2 F3 F4 F5
HFID_WO1000001
Analyzer basic controls (calibration) & setup...
Analyzer and I/O, ex pert controls & setup ... System configuration and diagnostics... Display controls...
_____________________________________________ Time & Date: System tag:
Meas ure Status... Channel Lock... MF G Data
-- Main Me nu --
16:35:32 August 16, 2009
490
61
900000
6.28 ppm
Emerson
1. Change to the main menu
Press F3 or the -key.
2. Change to the basic controls
Press the -key or the -key.
F1 F2 F3 F4 F5
HFID_WO1000001
Measurem ent range num ber:
Range upper limit: Span gas concentration: Aut omatic range change control: Sample flow: Ranges with valid calibrat io n: Cal ib ration status:
Flame condition: Light Flame...
HOME ESCAPE ZERO SPAN
Basic Controls
F1 F2 F3 F4 F5
6.28 ppm
1
10.0 ppm
8.6 ppm
Disabled
4296 ml/min
1&2&3&4
READY
ON
3. Select “Measurement range number”
Press the -key or the -key.
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
4 - 9
HFID_WO1000001
Basic Controls
Measu rement range number: Range upper limit: Span gas conc entration: Aut omati c range change control: Sample flow: Ranges with valid calibrat io n: Ca lib rati on st atus :
Flame condition: Light Flame...
Back...
F1 F2 F3 F4 F5
HFID_WO1000001
Basic Controls
Measu rement range number: Range upper limit: Span gas conc entration: Aut omati c range change control: Sample flow: Ranges with valid calibrat io n: Ca lib rati on st atus :
Flame condition: Light Flame...
Back...
6.28 ppm
10.0 ppm
8.6 ppm
Disabled
4296 ml/min
1&2&3&4
READY
ON
6.28 ppm
10.0 ppm
8.6 ppm
Disabled
4296 ml/min
1&2&3&4
READY
ON
4. Set the new range number
1
Press the ↑-key or the ↓ -key until the required range number appears.
Example: Changing from range 1 to range 2.
5. Confirm range number
2
Press the -key.
Note: To cancel the change and return to the previous value, press F2.
F1 F2 F3 F4 F5
HFID_WO1000001
Basic Controls
Measu rement range number:
Range upper limit: Span gas conc entration: Aut omati c range change control: Sample flow: Ranges with valid calibrat io n: Ca lib rati on st atus :
Flame condition: Light Flame...
HOME ESCAPE ZE RO SPAN
F1 F2 F3 F4 F5
6.28 ppm
100 ppm
96 ppm
Disabled
4296 ml/min
1&2&3&4
READY
ON
6 Change to the main menu
2
Press the -key or F1.
Note: The range upper limit (2nd line) automatically changes to the value previously set for the selected range.
4- 10
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10
Basic controls: 4.1.3
Changing the measurement range
HFID_WO1000001
Analyzer basic controls (calibration) & setup...
Analyzer and I/O, ex pert controls & setup ... System configuration and diagnostics... Display controls...
_____________________________________________ Time & Date: System tag:
Meas ure Status... Channel Lock... MF G Data
F1 F2 F3 F4 F5
-- Main Me nu --
16:35:44 August 16, 2009
HFID_WO1000001
6.28 ppm THC
0 Range: 2 25
Sample pressure: Oven temperature: Flame status: Raw signal:
Display Status... Main... Channel BasicCal
232.5 hPa 191 C
ON
531756
50 47
100000
6.28 ppm
Emerson
490
61
900000
7. Change to the single component display
Press F1.
8. The single component display showing the new measurement range
F1 F2 F3 F4 F5
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9.x)] 08/10
NGA 2000
4 - 11
4- 12
NGA 2000
HAS64E-IM-SW39(1) [NGA-e (HFID-Software 3.9 .x)] 08/10