Rosemount Manual: MLT2 CEMS Ex p Manual Addendum for Pressurized Analyzers | Rosemount Manuals & Guides

ATEX Instruction Manual
ETC01035 2015/10
MLT 2 CEMS Ex p
Manual Addendum for Pressurized Analyzers
intended to be used in Hazardous Areas Zone 2
www.EmersonProcess.com
ESSENTIAL INSTRUCTIONS
Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifi cations. The following instructions MUST be adhered to and integrated into your safety program when installing, using and maintaining Emerson Process Management (Rosemount Analytical) products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
 Read all instructions prior to installing, operating, and servicing the product.  If you do not understand any of the instructions, contact your Emerson Process
Management (Rosemount Analytical) representative for clari cation.
 Follow all warnings, cautions, and instructions marked on and supplied with the
product.
 Inform and educate your personnel in the proper installation, operation, and
maintenance of the product.
 Install your equipment as specifi ed in the Installation Instructions of the
appropriate Instruction Manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
 To ensure proper performance, use quali ed personnel to install, operate, update,
program, and maintain the product.
 When replacement parts are required, ensure that qualifi ed people use replacement
parts specifi ed by Emerson Process Management (Rosemount Analytical). Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fi re, electrical hazards, or improper operation.
 Ensure that all equipment doors are closed and protective covers are in place,
except when maintenance is being performed by qualifi ed persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice. 10th edition 2015-10
Emerson Process Management GmbH & Co. OHG
Rosemount Analytical Process Gas Analyzer Center of Excellence Industriestrasse 1 63594 Hasselroth, Germany T +49 6055 884 0 F +49 6055 884 209
Original Instruction Manual for the purpose of the European Directive 94/9/EC.
ATEX Instruction Manual
ETC01035 10/2015
MLT 2 CEMS Ex p
TABLE OF CONTENT
PREAMBLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-1
ADDITIONAL LITERATURE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-1
Defi nitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-2
Symbols used on and inside the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-3
Symbols used in this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-4
Terms used in this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-5
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-7
Intended use statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-7
General safety notice / Residual risk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-7
Authorized personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-8
CHAPTER 1 Technical Description 1-1
1.1 Application and Principle of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
1.2 Description of Pressurization System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.3.1 Installation Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.3.2 Explosion Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.3.3 Electrical and Mechanical Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.3.4 Purge Gas Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
1.3.5 Overpressure relief valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3.6 Sample gas specifi cations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Nameplate Label. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Table of Content
CHAPTER 2 Installation 2-1
2.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2 Installing the Intrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.3 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.4 Connecting Sample and Protective Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.5 Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
2.5.1 Cable Gland Assembly Instruction for Shielded Cables . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.5.2 Power Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
CHAPTER 3 Startup 3-1
3.1 Final Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
3.2 Switching On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
TOC-1Emerson Process Management GmbH & Co. OHG
ATEX Instruction Manual
ETC01035
MLT 2 CEMS Ex p
Table of Contents
10/2015
Chapter 4 Maintenance 4-1
4.1 Routine Tests in General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.1.1 Routine Tests Inteval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.1.2 Visual Inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.1.3 Close Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.1.4 Detail Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.2 Routine Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
4.2.1 Gasanalyzer, Analytical Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
4.2.2 Gasanalyzer, Ex Protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
4.2.3 Pressurization System and
other accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
4.2.4 Test the Relief Valve´s Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.3 Tests after Working on the Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
4.3.1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
4.3.2 Overpressure Test .......................................................................................................4-6
4.3.3 Leak Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
4.3.4 Test the Relief Valve´s Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.3.5 Removal of Modifi cations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.3.6 Special Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.4 Replacements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.5 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
4.5.1 Overpressure is not generated . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
TOC-2 Emerson Process Management GmbH & Co. OHG
ATEX Instruction Manual
ETC01035 10/2015
MLT 2 CEMS Ex p
PREAMBLE
This instruction manual provides additional information about installing, operating and maintai­ning/servicing pressurized ML T 2 series gas analyzers for zone 2 in hazardous (classifi ed) areas.
Operators must be familiar with operating these analyzers, before startup.
This instruction manual covers ML T 2 series gas analyzers variations and therefore may describe confi gurations and/or options not part of your specifi c analyzer.
This ATEX instruction manual is an addendum to the basic instruction manual and covers aspects of explosion protection and safety equipment only! To safely operate these instruments, you have to read and understand all associated manuals, including those of additional components!
If specifi cations or data provided by this manual differ from those given in the associated manuals, latter are superseded by this manual.
ADDITIONAL LITERATURE
This manual covers aspects specifi c for using pressurized MLT 2 series gas analyzers analyzers in hazardous (classifi ed) areas, only.
T o install, start-up, operate and maintain/service the instrument in a safe manner it is MAN­DATORY to read all additional instruction manuals shipped together with the instrument!
The following instruction manuals are available and/or referenced within this manual at hand and, depending on your system´s confi guration, shipped with your analyzer:
90002929 MLT Hardware instruction manual HAS3xE-IM-SW39 MLT Software instruction manual (rev. 3.9.x or later) User´s Manual F 840 Pressurized Enclosure System F840, Revision 2, 2012 or Revi-
sion 0, 2015
SAVE ALL INSTRUCTIONS FOR FUTURE USE!
Contact your local service center or sales offi ce when missing documents.
Safety Instructions
S-1Emerson Process Management GmbH & Co. OHG
ATEX Instruction Manual
ETC01035
MLT 2 CEMS Ex p
10/2015
DEFINITIONS
The following defi nitions apply to WARNINGS, CAUTIONS and NOTES found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. which, if not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. which, if not strictly observed, could result in minor or moderate injury or damage to or destruction of equipment, or loss of effectiveness.
REMARK
Highlights an essential operating procedure, condition or statement
NOTE
Highlights an important statement.
S-2 Emerson Process Management GmbH & Co. OHG
ATEX Instruction Manual
ETC01035 10/2015
MLT 2 CEMS Ex p
SYMBOLS USED ON AND INSIDE THE UNIT
Wherever one or more of the following symbols appear on or inside the instrument, be careful and read the instructions given in the accompanying manuals!
Follow these warnings and notes carefully
to minimize risks.
This symbol at the instrument ... ... indicates
dangerous voltages may be accessible. Remo­ving covers is permitted only, if the instrument is disconnected from power - and even in this case by qualifi ed personnel only!
hot surfaces may be accessible. Removing covers by qualifi ed personnel is permitted only, if the instrument is disconnected from power. Nevertheless several surfaces may remain hot for a limited time.
Safety Instructions
more detailled information available: see in-
struction manual before proceeding!
more detailled information available: see in­struction manual before proceeding!
S-3Emerson Process Management GmbH & Co. OHG
ATEX Instruction Manual
ETC01035
MLT 2 CEMS Ex p
10/2015
SYMBOLS USED IN THIS MANUAL
Wherever one or more of the following symbols are used in this instruction manual, read the accompanying information and instructions carefully.
Follow these warnings and notes carefully
to minimize risk.
This symbol used in the manual ... ... means
dangerous voltages may be exposed
hot surfaces may be exposed
possible danger of explosion
toxic substances may be present
substances harmful to health may be present
indicates notes relating to heavy instruments
electrical components may be destroyed by
electrostatic discharges
units must be disconnected from the power source
indicates special instructions or information for operation at low temperatures.
indicates basic conditions or procedures are being described.
This symbol may also indicate information impor­tant for achieving accurate measurements.
S-4 Emerson Process Management GmbH & Co. OHG
ATEX Instruction Manual
ETC01035 10/2015
TERMS USED IN THIS MANUAL
MLT 2 CEMS Ex p
ATEX
Directive 94/9/EC, (respectivly directive 2014/34/EU, valid from 20.04.2016) common­ly called the ATEX („Atmosphères Explosib­les“) products directive.
Area Classifi cation
Zone 1
Where ignitable concentrations of fl ammable gases can exist some of the time under nor­mal operating conditions.
Equipment to be used in Zone 1 has to be classifi ed Category 2.
Zone 2
Where ignitable concentrations of fl ammable gases are not likely to exist under normal operating conditions.
Equipment to be used in Zone 2 has to be classifi ed Category 3.
Containment System
part of the analyzer containing the fl ammable substance that may constitute an internal source of release
Pressurization
technique of guarding against the ingress of the external atmosphere into an enclosure by maintaining a protective gas therein at a pres­sure above that of the external atmosphere.
Pressurization system
grouping of safety devices and other compon­ents used to pressurize and monitor or control a pressurized enclsoure.
Safety Instructions
Explosive Gas(es)
Flammable gases and gas mixtures of a concentration within the explosion limits and present in mixture with air.
Flammable Gas(es)
Gases, vapors or mixtures thereof that are apable of being ignited
Purging
in a pressurized enclosure, the operation of passing a quantity of protective gas through the enclosure and ducts, so that the concen­tration of the explosive gas atmosphere is brought to a safe level.
Purge mode „Leakage Compensation“
In Ex p „leakage compensation mode“ just as much protective gas is used to hold an overpressure of > 1 mbar compared to at­mospheric.
S-5Emerson Process Management GmbH & Co. OHG
MLT 2 CEMS Ex p
Terms used in this manual
Lower Explosion Limit (LEL)
Volume ratio of fl ammable gas in air below which an explosive gas atmosphere will not be formed.
Upper Explosion Limit (UEL) Volume ratio of fl ammable gas in air above
which an explosive gas atmosphere will not be formed: the mixture of gas and air is too rich in fuel (defi cient in oxygen) to burn.
ATEX Instruction Manual
ETC01035
10/2015
Protective Gas
Air or inert gas used for purging and maintai­ning an overpressure and, if required, dilution and purging.
Overpressure
pressure above ambient pressure within a pressurized enclosure.
S-6 Emerson Process Management GmbH & Co. OHG
ATEX Instruction Manual
ETC01035 10/2015
MLT 2 CEMS Ex p
SAFETY INSTRUCTIONS
INTENDED USE STATEMENT
MLT 2 series gas analyzers are intended to be used as analyzers for industrial purposes. They must not be used in medical, diagnostic or life support applications nor as safety devices.
Using ML T 2 analyzers as safety devices, requiring redundant design or SIL classifi cation, is also not permitted.
No independent agency certifi cations or approvals are to be implied as covering such applications!
GENERAL SAFETY NOTICE / RESIDUAL RISK
If this equipment is used in a manner not specifi ed in these instructions, protective systems may be impaired or hazards can arise, resulting in explosion, loss of life, personal injury and damage to this equipment and on-site property.
Despite of incoming goods inspections, production control, routine tests and application of state-of-the-art measuring and test methods, an element of risk remains when operating a gas analyzer!
Even when operated as intended and observing all applicable safety instructions some residual risks remain, including, but not limited to, the following:
• Explosion protection measures may become ineffective on the occurence of
• 1 failure (for Category 3 instruments)!
• The emission of gases hazardous to health may even be possible when all gas con­nections have been correctly made.
Avoid exposure to the dangers of these residual risks by taking particular care when in­stalling, operating, maintaining and servicing the analyzer.
Safety Instructions
S-7Emerson Process Management GmbH & Co. OHG
ATEX Instruction Manual
ETC01035
MLT 2 CEMS Ex p
Safety Instructions
AUTHORIZED PERSONNEL
In-depth specialist knowledge is an absolutely necessary condition for working with and on the analyzer.
Authorized personnel for installing, operating, servicing and maintaining the analyzer are instructed and trained qualifi ed personnel of the operating company and the manufac- turer.
It is the responsibility of the operating company to
• train staff,
• observe safety regulations,
• follow the instruction manual.
Operators must
• have been trained,
• have read and understood all relevant sections of the instruction manual before commencing work,
• know the safety mechanisms and regulations.
10/2015
T o avoid explosions, loss of life, personal injury and damage to this equipment and on-site property, do not install, operate, maintain or service this instrument before reading and understanding this instruction manual and receiving appropriate training.
S-8 Emerson Process Management GmbH & Co. OHG
ATEX Instruction Manual
ETC01035 10/2015
Read all instruction manuals (including versions for auxiliary equipment) before installing this instrument!
Violating safety instruction can cause explosion and/or injury, death, or long-term health hazards of personnel!
MLT 2 CEMS Ex p
Safety Instructions
EXPLOSION HAZARD
EXPLOSION HAZARD WHEN OPEN
Do not open while explosive atmosphere may be present!
EXPLOSION HAZARD
Products covered by this manual are intended to be installed and operated in hazardous areas. This is permitted only, considering all relevant national legislative requirements and regulations, not listed within this manual.
Make sure, that all safety devices and the pressurization unit are working properly, to ensure safe operation of the analyzer!
Violating may cause explosions!
EXPLOSION HAZARD
Safety Instructions
Installing and wiring this instrument must comply with all relevant national legislative requirements and regulations.
Consider all safety instructions within this on hand manual and all associa­ted analyzer instruction manuals!
S-9Emerson Process Management GmbH & Co. OHG
MLT 2 CEMS Ex p
EXPLOSION HAZARD WHEN SHUT OFF / BATTERY
When the analyzer is out of order or if the pressurization unit shuts off due to a failure, all inputs and outputs connected to external equipment MUST be shut off too!
This ensures that no hazardous voltages are present within the analyzer enclosure, when not pressurized. Note that for data backup purposes some circuitry on the CPU board continuously is powered by a battery!
ATEX Instruction Manual
ETC01035
10/2015
Safety Instructions
EXPLOSION HAZARD DURING MAINTENANCE
Disconnect the system from voltage before open the doors. Don‘t open the analyzer when explosive atmosphere can be present
EXPLOSION HAZARD BY MODIFICATION
Any addition, substitution, or replacement of components installed on or in this device, must be certifi ed to meet the hazardous area classifi cation that the device was certifi ed to prior to any such component addition, substitution, or replacement. In addition, the installation of such device or devices must meet the requirements specifi ed and defi ned by the hazardous area classifi cation of the unmodifi ed device.
Any modifi cations to the device not meeting these requirements, will void the product certifi cation(s).
Contact Emerson Process Management‘s customer service center for return authorization.
S-10 Emerson Process Management GmbH & Co. OHG