Rosemount Manual: IMPS 4000 Intelligent Multiprobe Test Gas Sequencer | Rosemount Manuals & Guides

IMPS 4000 Intelligent Multiprobe Test Gas Sequencer
HIGHLIGHTS OF CHANGES
Page Summary
Effective July 2016 - Rev. 3
General Removed Rosemount Analytical logo. Changed revision number and release date. Cover Updated revision number and release date. Removed Rosemount Analytical logo.
2-4 Revised step 3 to include information on Xi and 6888Xi. 2-5 Revised Figure 2-3 to include Xi and 6888Xi. 3-6 and 3-7 Revised Automatic Calibration instructions to include expanded autocalibration setup
details for Oxymitter and new setup details for X-STREAM.
3-7 Added new section: 6888A Oxygen Transmitter with 6888Xi. Updated Semi-Automatic Calibration introductory paragraph to include 6888Xi.
Added new section: 6888Xi Keypad. 4-1 Edited Overview paragraph to include 6888Xi.
Back Cover Updated company addresses, URL, and disclaimer. Added social media information and Rosemount logo.
4000 IMPS
Instruction Manual
IM-106-400 IMPS, Rev. 3 July 2016
Table of Contents
Essential
Instructions
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Technical Support Hotline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
SECTION 1 Description and Specifications
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Component Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
System
Overview
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Physical
Description
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Theory of
Operation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Z-Purge
Option
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
SECTION 2 Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Gas
Connections
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Electrical
Connections
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Switch Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
SECTION 3 Operation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Calibration
Requirements
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Operator Interface Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Membrane
Keypad
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Using The Operator
Interface
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Selecting a Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Changing a Typical
Parameter
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Test Gas Flow
Setup
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Automatic Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Semi-Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
SECTION 4 Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Troubleshooting
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Calibration
Troubleshooting
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
PLC I/O LED Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Instruction Manual
IM-106-4000 IMPS, Rev 3
July 2016
TOC-2
4000 IMPS
SECTION 5 Maintenance and Service
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Fuse Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
PLC Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Operator Interface Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Solenoid Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Pressure Regulator Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Pressure Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Condensation Drain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Flowmeter
Adjustments
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Reference Gas Flowmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Test Gas
Flowmeter
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
SECTION 6 Replacement Parts
Replacement Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
APPENDIX A Safety Data
Safety
Instructions
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
APPENDIX B Return of Material
Returning Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Instruction Manual
IM-106-400 IMPS, Rev. 3 July 2016
4000 IMPS
Intelligent Mutiprobe Test Gas Sequencer
READ THIS PAGE BEFORE PROCEEDING!
ESSENTIAL INSTRUCTIONS
Emerson Process Management designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Emersons Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the
product.
If you do not understand any of the instructions, contact your
Emerson Process Management representative for clarification.
Follow all warnings, cautions, and instructions marked on and
supplied with the product.
Inform and educate your personnel in the proper installation,
operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions
of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and
pressure sources.
To ensure proper performance, use qualified personnel to install,
operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use
replacement parts specified by Emerson Process Management. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers
are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Instruction Man
ual
IM-103-400A, Rev. 3
July 2016
ii
4000 IMPS
PREFACE The purpose of this manual is to provide information concerning the
components, functions, installation and maintenance of the IMPS 4000 Intelligent Multiprobe Test Gas Sequencer.
DEFINITIONS The following definitions apply to WARNINGS, CAUTIONS, and
NOTES
found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE Highlights an essential operating procedure, condition, or statement.
SYMBOLS
:
EARTH
(GROUND)
TERMINAL
:
PROTECTIVE CONDUCT
OR TERMINAL
:
RISK OF ELECTRICAL SHOCK
:
WARNING: REFER TO
INSTRUCTION
MANUAL
NOTE TO USERS The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, and is not related to the illustration in any technical manner.
TECHNICAL SUPPORT HOTLINE
For assistance with technical problems, please call the Customer Support Center (CSC). The CSC is staffed 24 hours a day, 7 days a week.
Phone: 1-800-433-6076 1-440-914-1261 In addition to the CSC, you may also contact Field Watch. Field Watch
coordinates Emerson Process Managements field service throughout the U.S. and abroad.
Phone: 1-800-654-RSMT (1-800-654-7768) Emerson Process Management may also be reached via the Internet through
e-mail and the World Wide Web: e-mail: GAS.CSC@emerson.com
Instruction Manual
IM-106-400 IMPS, Rev. 3 July 2016
4000 IMPS
Section 1 Description and Specifications
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-2
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-2
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-4
Z-Purge Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-5
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-5
INTRODUCTION The Rosemount Analytical IMPS 4000 Intelligent Multiprobe Test Gas
Sequencer is used with the Oxymitter 4000 or the X-STREAM Oxygen Transmitter. The IMPS 4000 has the intelligence to provide test gas sequencing of up to four oxygen transmitters to accommodate automatic and semi-automatic calibration routines. Table 1-1 lists the available IMPS 4000 versions.
NOTE The IMPS 4000 Intelligent Multiprobe Test Gas Sequencer operates the same when used with the Oxymitter 4000 or the X-STREAM O2 Transmitter with HART or FOUNDATION fieldbus Communications. References to the Oxymitter 4000 or the X-STREAM O2 Transmitter throughout this instruction manual also include the Oxymitter 5000 and the X-STREAM O2 Transmitter with FOUNDATION fieldbus Communications.
Table 1-1. IMPS 4000 Versions
Part Number Description Number of O2 Transmitters
3D39695G01 IMPS 1 3D39695G02 IMPS 2 3D39695G03 IMPS 3 3D39695G04 IMPS 4 3D39695G05 IMPS w/115 V Heater 1 3D39695G06 IMPS w/115 V Heater 2 3D39695G07 IMPS w/115 V Heater 3 3D39695G08 IMPS w/115 V Heater 4 3D39695G09 IMPS w/220 V Heater 1 3D39695G10 IMPS w/220 V Heater 2 3D39695G11 IMPS w/220 V Heater 3 3D39695G12 IMPS w/220 V Heater 4
Instruction Manual
IM-106-400 IMPS, Rev. 3
July 2016
4000 IMPS
1-2
COMPONENT CHECKLIST
A typical IMPS 4000 should contain the items shown in Figure 1-1. Record the part number, serial number, and order number for the IMPS 4000 on the back cover of this manual.
Figure 1-1. Typical System Package
1. Instruction Manual
2. IMPS 4000
SYSTEM OVERVIEW Physical Description The main components of the IMPS 4000 are housed in a NEMA 4X (IP56)
enclosure for use in non-hazardous areas. The main internal components include the flow panel and the inner assembly.
Flow Panel (Figure 1-2) The Programmable Logic Controller (PLC) operator interface and the
reference and test gas flowmeters are mounted to the flow panel.
1. The PLC operator interface allows you to set up time-sequenced calibration routines for up to four oxygen transmitters. It also allows you to initiate a semi-automatic calibration.
2. Four sets of test gas and reference gas flowmeters are mounted on the flow panel. One set of flowmeters is available for each of the four oxygen transmitters that can be attached to the IMPS 4000. The test gas flowmeter indicates the amount of test gas sent to the oxygen transmitter. The reference gas flowmeter indicates the amount of reference air flowing to the oxygen transmitter.
Instruction Manual
IM-106-400 IMPS, Rev. 3 July 2016
4000 IMPS
1-3
Figure 1-2. IMPS Components
Inner Assembly The inner assembly consists of the PLC, PC board, pressure regulator, gas
solenoids, test gas pressure switches, and heater.
1. The PLC consists of a power supply, Central Processing Unit (CPU), and three I/O modules.
a. The power supply supplies the voltage and current needed to
operate the IMPS 4000.
b. The CPU controls most system operations. It contains the program
memory in which the information entered into the PLC operator interface is stored, and it processes the I/O signals received from the I/O modules.
c. The I/O modules are terminated on the PC board. They receive and
send signals between the PLC and PC board for communication between the oxygen transmitter and the IMPS 4000.
2. The PC board contains all wiring connections for up to four oxygen transmitters.
3. The pressure regulator ensures the instrument air (reference gas) flowing to the oxygen transmitter is at a constant pressure. The regulator also has a drain valve that drains excess moisture from the internal gas circuit.
4. The manifold can have up to six solenoids: test gas 1 solenoid, test gas 2 solenoid, and a gas solenoid for each oxygen transmitter connected to the IMPS 4000. The solenoids activate and deactivate to allow test gas to flow between the IMPS 4000 and an oxygen transmitter.
Instruction Manual
IM-106-400 IMPS, Rev. 3
July 2016
4000 IMPS
1-4
5. The pressure switches detect if the pressure of a test gas is low, which can be caused by an empty gas bottle, a disconnected gas line, etc. Calibration is prohibited when test gas pressure is low.
6. The heater keeps the components in the ambient temperature range to prevent electronic malfunction.
THEORY OF OPERATION Reference Figure 1-3. When a calibration is initiated, via the IMPS,
HART/AMS, oxygen transmitter, or remote location, the signal is sent to the IMPS 4000 PLC. The PLC first energizes the oxygen transmitter solenoid and then energizes the test gas 1 (high gas) solenoid. Test gas 1 flows through the IMPS 4000 to the oxygen transmitter. The oxygen transmitter measures the oxygen content of test gas 1 and sends a signal to the IMPS 4000 indicating that it received test gas 1. When the IMPS 4000 receives the signal, the PLC deenergizes the test gas 1 solenoid.
Figure 1-3. IMPS Calibration Setup
1-5
Instruction Manual
IM-106-400 IMPS, Rev. 3 July 2016
4000 IMPS
After a time delay that allows test gas 1 to clear the system, the PLC energizes the test gas 2 (low gas) solenoid. Test gas 2 flows through the IMPS 4000 to the oxygen transmitter. The oxygen transmitter measures the oxygen content of test gas 2 and sends a signal to the IMPS 4000 indicating that it received test gas 2. After measuring the two test gases, the oxygen transmitter automatically makes the internal calibration adjustment and sends the signal to the IMPS 4000. When the IMPS receives the signal, the PLC deenergizes the test gas 2 solenoid and the oxygen transmitter solenoid. Whether or not the calibration passes or fails displays on the PLC operator interface.
Z-PURGE OPTION Some IMPS applications have area safety requirements (Class 1, Division 2).
These requirements may be satisfied with the installation of an optional Z-purge. The Z-Purge provides positive pressure within the IMPS 4000 enclosure to keep out dust and other foreign matter. Figure 1-4 shows the Z-purge unit and how it connects to the IMPS 4000.
SPECIFICATIONS
Feature Specification
Electrical Classification NEMA 4X (IP56) Humidity Range 95% Relative Humidity Ambient Temperature Range*
Input
Power*
w/o Heater Option
w/ Heater Option
w/o Heater Option
32° to 131°F (0° to 55°C)
-35° to 131°F (-37° to 55°C)
90 to 250 V, 50/60 Hz, 50 VA
w/ 115 V Nominal Heater (Optional) w/ 220 V Nominal Heater (Optional)
115 V Nominal, 50/60 Hz, 200 VA 230 V Nominal, 50/60 Hz, 200 VA
External Electrical Noise Minimum Interference Handshake Signal to/from Oxygen
Transmitter (Self-Powered)
Cal Initiate Contact Input from
5 V (5 mA Maximum)
Control Room (One per Probe) 24 VDC (Self-Powered)
Relay Outputs to Control
Room (One "In-Cal" per Probe, "Low Gas Flowing," "High Gas Flowing, "One "Cal Failed" per Probe, "Gas Pressure Low"
Cabling Distance Between IMPS 4000
and Oxygen Transmitter
Cabling Distance Between IMPS 4000
and Customer-Supplied Status Relay I/O in Control Room
Piping Distance Between IMPS 4000
and Oxygen Transmitter
5 to 30 VDC Form A (SPST),
1.5 A per point, 3 A total per common
1000 ft (303 m) Maximum
1000 ft (303 m) Maximum
300 ft (91 m) Maximum
Approximate Shipping Weight 40 lbs (18 kg)
*If using the heater option, the low temperature value of the ambient temperature range may vary depending on the input voltage supply.
1-6
Instruction Manual
IM-106-400 IMPS, Rev. 3
July 2016
4000 IMPS
Figure 1-4. IMPS 4000 with
Z-Purge
Instruction Manual
IM-106-400 IMPS, Rev. 3 July 2016
4000 IMPS
Section 2 Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Gas Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-3
Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-4
Switch Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-6
OVERVIEW This section describes the installation requirements for the IMPS 4000
Intelligent Multiprobe Test Gas Sequencer.
Before starting to install this equipment, read the “Safety Instructions for the Wiring and Installation of this Apparatus” in Appendix A of this Instruction Manual. Failure to follow the safety instructions could result in injury or death.
Install all protective equipment covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.
Instruction Manual
IM-106-400 IMPS, Rev. 3
July 2016
4000 IMPS
2-2
MECHANICAL INSTALLATION
The outline drawing in Figure 2-1 shows mounting centers and clearances for the IMPS 4000. The unit is designed to mount on a wall or bulkhead. Ensure the unit is installed according to the following instructions.
1. Install the module no further than 300 ft (91 m) from the probe and no further than 1000 ft (303 m) from a remote connection or from the status relay indicators.
2. Locate units without the optional heater where the ambient temperature is between 32° and 131°F (0° and 55°C). For units with the optional heater, install where the ambient temperature is between -35° and 131°F (-37° and 55°C).
Figure 2-1. Mounting Dimensions
Instruction Manual
IM-106-400 IMPS, Rev. 3 July 2016
4000 IMPS
2-3
GAS CONNECTIONS Gas connections are located on the bottom of the IMPS 4000 and are 1/8-27
NPT fittings (Figure 2-2). Use the following procedure to connect the test gases:
1. Connect the reference air supply to INSTR AIR. The pressure regulator valve is set at the factory to 20 psi (138 kPa). If necessary, readjust by turning the knob on the top of the valve until the desired pressure is obtained.
2. Connect the O2 test gas 1 to HIGH CAL GAS IN. The test gas pressure should be set at 20 psi (138 kPa).
3. Connect the O2 test gas 1 to LOW CAL GAS IN. The test gas pressure should be set at 20 psi (138 kPa).
4. Connect the oxygen transmitter REF GAS to REF AIR OUT.
5. Connect the oxygen transmitter CAL GAS to TEST GAS OUT.
Figure 2-2. Gas Connections
Instruction Manual
IM-106-400 IMPS, Rev. 3
July 2016
4000 IMPS
2-4
ELECTRICAL CONNECTIONS
All wiring must conform to local and national codes. Refer to Figure 2-3 and use the following procedure to connect the first oxygen transmitter. Repeat the procedure for each remaining oxygen transmitter.
1. Run the line voltage through the bulkhead fitting on the bottom of the IMPS 4000 marked LINE IN. Connect the line voltage to the J24 terminal. The power supply automatically adjusts to the input voltage. Tighten the cord grips to provide strain relief.
2. When connecting an Oxymitter O2 Transmitter, run the handshake logic I/O wires for each oxygen transmitter through a bulkhead fitting on the
bottom of the IMPS 4000 marked SIGNAL IN. Dedicate one fitting for each oxygen transmitter. Connect the oxygen transmitter logic I/O wires as shown in Figure 2-3.
3. When connecting an XI or 6888XI electronics, run the handshake signal wires from connector J3 of the Xi or 6888Xi I/O board through a bulkhead fitting on the bottom of the IMPS 4000 marked SIGNAL IN. Dedicate one fitting for each oxygen transmitter. Connect the oxygen transmitter handshake signal wires as shown in Figure 2-3.
4. To set up the IMPS 4000 to initiate a calibration from a remote location, run the remote contact input wires through the SIGNAL IN bulkhead fitting. Connect the remote contact input wires as shown in Figure 2-3.
5. Status relay connections are available on the IMPS 4000 PC board to signal when the oxygen transmitter is in or out of calibration and to indicate calibration gas status. Status relays can be connected to either indicator lights or a computer interface. Relay contacts are capable of handling up to 5 to 30 V, 1.5 A per point, 3.0 A total per common maximum power source. Cabling requirement is 1000 ft (303 m) maximum. Make the status relay switch connections as shown in Figure 2-3.
2-5
Instruction
Manual
IM-106-400 IMPS, Rev. 3 July 2016
4000 IMPS
Figure 2-3. Electrical
Connections
2-6
Instruction Manual
IM-106-400 IMPS, Rev. 3
July 2016
4000 IMPS
SWITCH
SETTINGS
Figure 2-4. RUN/TERM
Switch
Settings
The PLC RUN!TERM switch must be in the TERM position for
proper
operation. When the RUN!TERM switch is in the RUN position, the PLC
will
display an "E525 Keyswitch" error message. Refer to Figure 2-4 and
verify
that the RUN!TERM switch is in the RUN position.
RUN/TERM
SWITCH
Instruction Manual
IM-106-400 IMPS, Rev. 3 July 2016
4000 IMPS
Section 3 Operation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Calibration Requirements . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Operator Interface Description . . . . . . . . . . . . . . . . . . . . . page 3-2
Message Display Mode Parameters . . . . . . . . . . . . . . . . . page 3-5
Test Gas Flow Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-6
Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-6
Semi-Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . page 3-7
OVERVIEW This section specifies the requirements to set up an oxygen transmitter
calibration. It describes the PLC operator interface and explains how to set up time-sequenced calibration routines for up to four oxygen transmitters using the IMPS 4000. This section also explains the differences between automatic and semi-automatic calibrations and how to initiate them.
CALIBRATION REQUIREMENTS
The following components are required for an oxygen transmitter calibration using the IMPS 4000:
1. Two tanks of precision calibration gas mixtures are required. Recommended calibration gases are nominally 0.4% and 8.0% oxygen in nitrogen. Two sources of calibrated gas mixtures are:
LIQUID CARBONIC GAS CORP. SPECIALTY GAS LABORATORIES
700 South Alameda Street Los Angeles, California 90058 213/585-2154
767 Industrial Road San Carlos, California 94070 415/592-7303
9950 Chemical Road Pasadena, Texas 77507 713/474-4141
12054 S.W. Doty Avenue Chicago, Illinois 60628 312/568-8840
603 Bergen Street Harrison, New Jersey 07029 201/485-1995
255 Brimley Road Scarborough, Ontario, Canada 416/266-3161
Instruction Manual
IM-106-400 IMPS, Rev. 3
July 2016
4000 IMPS
3-2
SCOTT ENVIRONMENTAL TECHNOLOGY, INC. SCOTT SPECIALTY GASES
2600 Cajon
Blvd. San Bernardino, California 92411 714/887-2571 TWX:
910-390-1159
1290 Combermere Street Troy, Michigan
48084
314/589-2950 Route
611 Plumsteadville, Pennsylvania 18949 215/766
8861
TWX:
510-665-9344
2616 South Loop
West
Suite
100 Houston, Texas 77054 713/669-0469
2. A check valve is required at the oxygen transmitter (between the calibration fitting and the gas line) to prevent the migration of process gases down the calibration gas line. A typical calibration setup is shown in Section 1, Figure 1-3 of this manual.
OPERATOR INTERFACE DESCRIPTION
See Figure 3-1. The operator interface consists of a membrane keypad and a 4-line, 16-character, backlit LCD display. The LCD display is divided into two areas. The left side is a descriptive field that identifies the calibration parameter. The right side shows the current value or status of the calibration parameter.
Figure 3-1. Operator Interface
Loading...
+ 50 hidden pages