Rosemount Manual: 868 NDIR Analyzer-Rev N | Rosemount Manuals & Guides

Instruction Manual
748003-N November 2002
Model 868
Non-Dispersive Infrared Analyzer
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them
normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
to ensure they continue to operate within their
Read all instructions
prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical representative for clarification.
Follow all warnings, cautions, and instructions
marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to
the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel
to install, operate, update, program,
and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
.
The information contained in this document is subject to change without notice.
Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc. Alconox is a registered trademark of Alconox, Inc. SNOOP® is a registered trademark of NUPRO Co.
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Model 868
Instruction Manual
748003-N
November 2002

TABLE OF CONTENTS

PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-4
Documentation....................................................................................................................................P-5
Condensed Startup And Calibration Procedure .................................................................................P-6
1.0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Overview................................................................................................................................1-1
1-2 Specifications.........................................................................................................................1-2
2.0 INSTALLATION ....................................................................................................................2-1
2-1 Unpacking..............................................................................................................................2-1
2-2 Location .................................................................................................................................2-1
2-3 Facility Preparation................................................................................................................2-1
a. Power Requirements ......................................................................................................2-1
b. Gas Connections.............................................................................................................2-1
2-4 Outline And Mounting Dimensions ........................................................................................2-1
2-5 Power Connections ...............................................................................................................2-1
a. Voltage Select Board ......................................................................................................2-1
b. Heater Voltage Select.....................................................................................................2-2
c. Fuse Select Panel...........................................................................................................2-2
2-6 Recorder Connection.............................................................................................................2-2
2-7 Remote Range ......................................................................................................................2-2
2-8 Sample Inlet/Output Connections..........................................................................................2-2
2-9 Calibration Gas Requirements ..............................................................................................2-2
2-10 Sample Handling System ......................................................................................................2-3
2-11 Leak test ................................................................................................................................2-3
2-12 Sample Flow Rate .................................................................................................................2-3
2-13 Differential Analysis With Flow Through Reference Cell.......................................................2-4
3.0 DETAILED STARTUP...........................................................................................................3-1
3-1 Startup ...................................................................................................................................3-1
3-2 Current Source Adjustment ...................................................................................................3-1
3-3 Display Readouts Using CHECK Switch...............................................................................3-1
3-4 RANGE Switch Position (O.T.) Oscillator Tune.....................................................................3-2
3-5 Range Switch ........................................................................................................................3-2
3-6 Time Constant Selection .......................................................................................................3-2
3-7 Preamp Gain Adjustment ......................................................................................................3-3
3-8 Source Balance Shutter Adjustment .....................................................................................3-3
3-9 Zero Adjust ............................................................................................................................3-3
3-10 span adjustment ....................................................................................................................3-3
3-11 Calibration without linearizer .................................................................................................3-4
3-12 Calibration with Linearizer .....................................................................................................3-4
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Instruction Manual
748003-N November 2002
4.0 OPERATION .........................................................................................................................4-1
4-1 Routine Operation .................................................................................................................4-1
4-2 Recommended Calibration Frequency..................................................................................4-1
4-3 Shutdown...............................................................................................................................4-1
5.0 THEORY................................................................................................................................5-1
5-1 Overview................................................................................................................................5-1
5-2 Detection System ..................................................................................................................5-1
5-3 Electronic Circuitry.................................................................................................................5-2
a. Oscillator Circuit Board and Associated Elements of Amplitude-Modulation Circuit......5-2
b. Functioning of Modulation System in TUNE Mode.........................................................5-2
c. Functioning of Modulation System in Operating Mode...................................................5-4
d. Radio-Frequency Demodulator.......................................................................................5-4
e. Filter/Rectifier Board and Associated Elements .............................................................5-4
f. Remote Range Control ...................................................................................................5-6
g. Source Driver ..................................................................................................................5-7
h. Analog Signal Path..........................................................................................................5-8
5-4 Chopper Driver ......................................................................................................................5-9
a. Recorder Output..............................................................................................................5-9
b. Temperature Control.......................................................................................................5-9
c. Power Supply ..................................................................................................................5-9
Model 868
6.0 SERVICE AND TROUBLESHOOTING ................................................................................6-1
6-1 Cell Removal and Replacement............................................................................................6-2
6-2 Cell Cleaning .........................................................................................................................6-7
6-3 Cell Desiccant........................................................................................................................6-7
6-4 Source Housing Removal......................................................................................................6-7
a. Source Removal..............................................................................................................6-7
b. Chopper Motor Replacement..........................................................................................6-7
6-5 Detector Removal..................................................................................................................6-7
6-6 Troubleshooting Procedures .................................................................................................6-8
a. Chopper Drive.................................................................................................................6-8
b. Temperature Controller...................................................................................................6-8
c. Source Drive ...................................................................................................................6-8
d. Voltage Checks...............................................................................................................6-8
7.0 REPLACEMENT PARTS ......................................................................................................7-1
7-1 Circuit Board Replacement Policy.........................................................................................7-1
7-2 Recommended Replacement Parts ......................................................................................7-1
7-3 Linearization Kit .....................................................................................................................7-5
a. Installation.......................................................................................................................7-5
b. Electrical Connections.....................................................................................................7-5
c. Setup...............................................................................................................................7-5
d. Troubleshooting and Repair............................................................................................7-6
e. Board Functions (Linearizer)...........................................................................................7-6
f. Linearized Analyzer.........................................................................................................7-7
8.0 RETURN OF MATERIAL ......................................................................................................8-1
8-1 Return Of Material .................................................................................................................8-1
8-2 Customer Service ..................................................................................................................8-1
8-3 Training..................................................................................................................................8-1
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 868
Figure 2-1.
Figure 2-2. Heater Voltage Section.......................................................................................... 2-2
Figure 3-1. Source Current Adjustment ................................................................................... 3-1
Figure 3-2. Front Panel Controls.............................................................................................. 3-2
Figure 5-1. Functional Diagram of Detection System .............................................................. 5-1
Figure 5-2. Signal Circuitry....................................................................................................... 5-3
Figure 5-3. Modulation System ................................................................................................ 5-5
Figure 5-4. Remote Range Select Circuit ................................................................................ 5-6
Figure 5-5. Automatic Range Select Circuit............................................................................. 5-7
Figure 6-1. Configuration Assembly – Part 1 ........................................................................... 6-2
Figure 6-2. Configuration Assembly – Part 2 ........................................................................... 6-3
Figure 6-3. Configuration Assembly – Part 3 ........................................................................... 6-4
Figure 6-4. Configuration Assembly – Part 4 ........................................................................... 6-5
Figure 6-5. Motor/Source Assembly......................................................................................... 6-6
Figure 7-1. Functional Diagram of Linearizer Board ................................................................ 7-9
Figure 7-2. Installation of Linearizer Board .............................................................................. 7-9
Figure 7-3. Linearizer Setup Switches ................................................................................... 7-10
Figure 7-4. Block Diagram, Model 868 Electronics................................................................ 7-10
Figure 7-5. Behavior of Instrument within Block Diagram...................................................... 7-11
Figure 7-6. Example Showing Relationship Between Range 2 and Range 1 Responses
Figure 7-7. Example Showing Relationship Between Range 2 and Range 1 Responses
Instruction Manual
748003-N
November 2002

LIST OF ILLUSTRATIONS

Line Voltage Selection ........................................................................................... 2-2
When Sensitivity of Range 2 is 100/60 Times Sensitivity of Range 1................. 7-12
When Sensitivity of Range 3 is 100/37 Times Sensitivity of Range 1................. 7-12

LIST OF TABLES

Table P-1. Range Switch Settings........................................................................................... P-6
Table 2-1. Purging Time at Atmospheric Sample Pressure.................................................... 2-4
Table 3-1. Range Switch Readings ........................................................................................ 3-1
Table 7-1. Parts List - Configuration/Application 10, 12, 13, 18, 22, 23, 38, 42, 43 ............... 7-2
Table 7-2. Parts List – Configuration/Application 26............................................................... 7-3
Table 7-3. Parts List - Configuration/Application 11, 21, 31, 36, 41, 44, 54, 56 ..................... 7-4
Table 7-4. Linearized Configuration Selection........................................................................ 7-8
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
748003-N November 2002
Model 868
DRAWINGS
619803 Schematic Diagram, Oscillator Board 619839 Schematic Diagram, Front Panel Display 620126 Schematic Diagram, Power Supply 620249 Schematic Diagram, Filter/Rectifier Board 620363 Pictorial Wiring Diagram, Model 868 620391 Schematic Diagram, Recorder Range Output 620474 Installation Drawing, Model 868 622190 Schematic Diagram, Linearizer Board
(LOCATED IN REAR OF MANUAL)
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 868
November 2002

PREFACE

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the Model 868 Non-Dispersive Infrared Analyzer.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
748003-N
DANGER .
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
748003-N November 2002

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE IN­STRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Model 868
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
WARNING
POSSIBLE EXPLOSION HAZARD
This instrument is not designed for analysis of flammable gas samples. If flammable gas samples are introduced, internal leakage could create a flammable mixture which could result in an explosion.
WARNING .
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 868
WARNING.
HIGH PRESSURE GAS CYLINDERS
This instrument requires periodic calibration with a known standard gas. See Sections 2-9 on page 2-2, 3-11 on page 3-4 and 3-12 on page 3-4. See also General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-on page 4.
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
WARNING.
TOPPLING HAZARD
November 2002
748003-N
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left side of the chassis. When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise) orientation.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the latch; however, further extraction must be done very carefully to insure the chassis does not fall out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling.
Failure to observe these precautions could result in personal injury and/or damage to the product.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
748003-N November 2002
Model 868
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125 never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 868
November 2002

DOCUMENTATION

The following Model 868 instruction materials are available. Contact Customer Service Center or the local representative to order.
748003 Instruction Manual (this document)
748003-N
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
748003-N November 2002

CONDENSED STARTUP AND CALIBRATION PROCEDURE

Model 868
Prior to shipment, this analyzer was subjected to extensive factory performance testing, during which all necessary optical and electrical ad­justments were made. The following instruc­tions are recommended for initial startup and subsequent standardization of the analyzer. In most cases, these simple instructions are suffi­cient to operate the analyzer. The detailed in­structions given in Section 3 are required if the optical bench alignment has been disturbed, which could possible occur during shipment.

Leak Test Procedure

WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer is not designed for analysis of flammable gas samples. If flammable gas samples are introduced, internal leakage could create a flammable mixture which could result in an explosion.
The following test is designed for sample pres­sure up to 10 psig (69 kPa).
1. Supply air or inert gas such as nitrogen at 10 psig (69 kPa) to analyzer via a flow in-
SWITCH
POSITION
0 +5 VDC 4.75 To 5.25 VDC 1 -5 VDC -4.50 To -5.10 VDC 2 +15 VDC +14.25 To 15.75 VDC 3 -15 VDC -14.25 To -15.75 VDC 4 +12 VDC +11.4 To 12.6 VDC 5 REF Voltage 2.4 To 2.6 VDC 6 + Source V1 7 - Source V1 8 Source I1 9 Preamp Out1
Analyzer is now calibrated for Range 1. If the analyzer is to be used on auxiliary Range 2 and/or Range 3, perform the steps in Calibration of Range 2 and Calibration of Range 3 below.
FUNCTION READING
Table P-1. Range Switch Settings
dicator with a range of 0 to 250 cc/min. and set flow rate at 125 cc/min.
2. Use a suitable test liquid such as SNOOP (PN 837 801) to detect leaks. Cover all fittings, seals, or possible leak sources.
3. Check for bubbling or foaming which indicates leakage and repair as re­quired. Any leakage must be cor­rected before introduction of sample and/or application of electrical power.

Power Verification

1. Verify power switch settings are cor­rect for available power (115 VAC or 230 VAC).
2. Apply power. Verify that heater LED is on (decimal point of Mode display) and that chopper LED is blinking (decimal point of Range display).
3. Move RANGE switch to CH position and verify per Table P-1.
1
Determined by application. See Infrared Analyzer Calibration and Data Sheet, located in the rear of this manual.
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 868
Instruction Manual
748003-N
November 2002

Calibration

1. Allow system to warm-up a minimum of one hour.
2. Connect zero gas to the sample cell inlet at the back of the analyzer. Flow the gas at a flow rate of 500 cc/min, as read on a flowmeter, through the analyzer for at least two minutes. Adjust the ZERO con­trol on front panel so meter or recorder reads zero.
3. Connect the upscale gas to the sample cell inlet.
a. Verify that RANGE switch is in Posi-
tion 1 (least sensitive range).
b. Flow the upscale gas appropriate for
Range 1, at a flow rate of 500 cc/min., through the analyzer for at least two minutes.
c. Adjust the SPAN control on the front
panel so the display reads the desired output. Refer to the calibration curves in Section 8.

Calibration of Range 2

1. Place RANGE switch at Position 2.
2. Flow upscale gas appropriate to Range 2, at a flow rate of 500 cc/min., through the analyzer for at least two minutes.
3. Adjust switches S2, S3, and S4 for de­sired reading. (S4 is the least sensitive switch).

Calibration of Range 3

1. Place RANGE switch at Position 3.
2. Flow upscale gas appropriate to Range 3, at a flow rate of 500 cc/min., through the analyzer for at least two minutes.
3. Adjust switches S5, S6, and S7 for de­sired reading. (S7 is the least sensitive switch).

Time Constant Switch S9

Set switch S9 to the desired speed of re­sponse of the electronics (time constant).
The analyzer is now calibrated for all three ranges and is ready for sample analysis.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7
Instruction Manual
748003-N November 2002
Model 868
Preface Rosemount Analytical Inc. A Division of Emerson Process Management
P-8
Model 868

DESCRIPTION AND SPECIFICATIONS

1-1 OVERVIEW

The Model 868 Non-Dispersive Infrared Ana­lyzer (NDIR) continuously determines the con­centration of a particular component of interest in a flowing mixture. The analysis is based on a differential measurement of the absorption of in­frared energy. With the proper selection of de­tector, optics and cells, the instrument has a wide range of applications, subject only to the limitation that the analysis involve the determi­nation of a single component, which must ab­sorb infrared energy.
Within the analyzer, two equal energy infrared beams are directed through two parallel optical cells, a flow-through sample cell and a sealed reference cell. Solid-state electronic circuitry continuously measures the difference between the amount of infrared energy absorbed in the
Instruction Manual
748003-N
November 2002
SECTION 1
two cells. This difference is a measure of the concentration of the component of inter­est in the sample. Readout is on a front panel display with 0 to 199.9 digital display. In addition, a +5 VDC output for a potenti­ometric (voltage) recorder is provided as standard. Additional outputs (0 to 10 mV, 0 to 100 mV, 0 to 1 VDC) are available as op­tions.
A calibration curve may be used to convert display or recorder readings into concentra­tion values. The analyzer may utilize an op­tional linearizer circuit for linear readout of concentration values on the display and on a recorder.
The analyzer is equipped for remote selec­tion of ranges.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
748003-N November 2002

1-2 SPECIFICATIONS

Repeatability.................................. 1% of fullscale
Noise ............................................. 1% of fullscale
1
Zero/Span Drift
Response Time (Electronic).......... Variable, 90% of fullscale in 0.5 to 5 seconds, field selectable
Sensitivity ...................................... 500 p/10
Sample Cell Length ....................... 0.04 inches (1mm) to 9.5 inches (24mm)
Materials in Contact with Sample
Windows .............................. Sapphire, quartz, Irtran
Cells.....................................Gold-plated glass or stainless steel
Tubing..................................FEP Teflon
Fittings ................................. 316 stainless steel
O-rings .................................Viton-A
Sample Flow Rate ......................... Nominal 500 to 1000 cc/min.
Sample Pressure........................... Maximum 10 psig (69 kPa)
Ambient Temperature Range........ 50 to 104°F (10 to 40°C)
Analog Output
Standard .............................. 0 to 5 VDC
Optional ...............................0 to 10 mV, 0 to 100 mV, 0 to 1 VDC (selectable)
Linearization .................................. Optional microprocessor based
Power Requirements..................... 115/230 VAC ±10%; 50/60 Hz ±3%; 115 W
Enclosure....................................... General purpose for installation in weather protected area
Dimensions.................................... 8 11/16 inches (220 mm) H
Weight ........................................... 32 lbs (15 kg)
Shipping Weight ............................ Approximately 46 pounds (21 kg)
1
Performance specifications based on an ambient temperature variation of less than 20°F (11°C) per hour.
............................. ±1% of fullscale per 24 hours
6
fullscale carbon monoxide 200 p/10
dioxide
19 inches (483 mm) W 16 inches (406 mm) D (with handles)
6
Model 868
fullscale carbon
Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
1-2
Model 868
Instruction Manual
748003-N
November 2002
SECTION 2

INSTALLATION

2-1 UNPACKING

Carefully examine the shipping carton and con­tents for signs of damage. Immediately notify the shipping carrier if the carton or its contents are damaged. Retain the carton and packing material until the instrument is operational.

2-2 LOCATION

Preferably the analyzer should be mounted near the sample stream to minimize sample transport time. Of two or more alternative installation sites, select the one subject to the least vibra­tion.
The analyzer must be installed in a weather­protected, non-hazardous location.

2-3 FACILITY PREPARATION

a. Power Requirements

115 or 230 VAC ±10% single phase 47 to 53 or 57 to 63 Hz Less than 115 watts

b. Gas Connections

Sample and reference IN and OUT lines are ¼ inch tube.

2-4 OUTLINE AND MOUNTING DIMENSIONS

The Model 868 has three mounting varia­tions:
Panel mount
Rack mount (Standard RETMA 19 inch)
Table mount
Refer to installation drawing 620474 in the rear of this manual.
General Notes for Installation
General purpose enclosure
3-prong, six foot power cord supplied
with analyzer

2-5 POWER CONNECTIONS

WARNING
ELECTRICAL SHOCK HAZARD
For safety and proper performance, this analyzer must be connected to a prop­erly grounded 3-wire source of electrical supply.
CAUTION
Before applying power, perform the Volt­age Select procedure below.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1

a. Voltage Select Board

The voltage select board is located in the power receptacle on the rear panel. Check that line voltage and circuit board to insure that desired voltage is face up on trailing edge of board as it is inserted (Figure 2-1 on page 2-2).
Instruction Manual
748003-N November 2002
110–115 VAC
115–130 VAC

b. Heater Voltage Select

The heater voltage select switch is located inside the chassis in the lower right corner of the power supply board. Slide switch to position of desired voltage (Figure 2-2 below).
115V
Figure 2-2. Heater Voltage Select

c. Fuse Select Panel

The fuse select panel is located on the rear panel. Select the appropriate amperage:
207-230 VAC
230-254 VAC
Figure 2-1. Line Voltage Selection

2-7 REMOTE RANGE

230V
2-8 SAMPLE INLET/OUTPUT CONNECTIONS
Model 868
Input impedance greater than 5K ohms
The remote range connections are made to the rear panel. See drawings 620474 and
620126.
Maximum distance remote to analyzer – 1000 feet (305 meters)
Remote cable – minimum 20 gauge (customer supplied). Refer to Section 5­3f on page 5-6.
Sample inlet and outlet connections are lo­cated on the rear panel. This instrument is intended for atmospheric pressure opera­tion only. Vent must go to atmosphere. Connections are ¼ inch tubing.
230 VAC = 0.5 amp 115 VAC = 1.0 amp

2-6 RECORDER CONNECTION

The recorder connections are made to the rear panel at TB2. See drawings 620474 and
620126.
Maximum distance recorder to analyzer – 1000 feet (305 meters)
Recorder cable – minimum 20 gauge shielded (customer-supplied)
Installation Rosemount Analytical Inc. A Division of Emerson Process Management
2-2

2-9 CALIBRATION GAS REQUIREMENTS

Analyzer calibration consists of setting a zero point and one or more upscale points.
All applications require a zero standard gas to set the zero point on the display or re­corder chart. If the factory Calibration and Data Sheet (in the rear of this manual) specifies the background gas, use it as the zero gas. If background gas is not speci­fied, use dry nitrogen for the zero gas.
Model 868
Instruction Manual
748003-N
November 2002

2-10 SAMPLE HANDLING SYSTEM

Many different sample handling systems are available, either completely assembled or as loose components. The type used depends on the requirements of the particular application and the preferences of the individual user. Typically, the sample handling system incorpo­rates such components as pumps, valves to permit selection of sample, zero standard, or upscale standard gas; needle valve in sample­inlet line for flow adjustment; flowmeter for flow measurement and/or indication of flow stop­page; and filter(s) to remove particulate matter.

2-11 LEAK TEST

DANGER
POSSIBLE EXPLOSION HAZARD
This instrument is not designed for analysis of flammable gas samples. If flammable gas samples are introduced, internal leakage may result creating a flammable mixture which could result in an explosion.

2-12 SAMPLE FLOW RATE

For best results, the sample flow rate must be in the range of 1 to 2 SCFH (500 to 1000 cc/min.). A subnormal flow rate will result in undesirable time lag.
Assume that two cell columns are required to flush any cell. Table 2-1, on page 2-4, indicates approximate flushing time at at­mospheric sampling pressure, i.e., the out­let of the cell venting to a atmosphere for various cell lengths.
Flushing time is inversely proportional to flow rate.
The primary effect of flow rate, other than flushing time, is cell pressure. Due to the restriction of the exit tubing, increasing flow rate increases sample pressure in the cell. For a 9-inch (232 mm) cell venting to at­mosphere, the cell pressure rises from 0 psig (0 kPa) at no flow, essentially linearity, by 1 mm Hg per CFH flow up to at least 20 CFH (10 L/min.).
The following procedure is designed for sample pressure up to 10 psig (69 kPa).
1. Supply air or inert gas such as nitrogen at 10 psig (69 kPa) to analyzer via a flow in­dicator with a range of 0 to 250 cc/min.
2. Set flow rate at 125 cc/min.
3. Use a suitable test liquid such as SNOOP (PN 837801) to detect leaks. Cover all fittings, seals and all sources of possible leaks.
4. Check for bubbling or foaming which indi­cates leaks.
5. Correct any leakage for introduction of sample and/or application of electrical power.
At 7.5 to 8.0 CFH (3.8 to 4 L/min.), there­fore, the pressure is increased by about 1%, and the output signal is thereby increased by about the same 1% over static condi­tions. In all cases, the effect of pressure on readout is eliminated if the same flow rate is used for the measured sample as well as for the zero gas and span gas.
It should be noted that at higher flow rates, because of increase in sample cell pres­sure, the non-linearity of the calibration curve increases. Therefore, if higher flow rates are required, the calibration curve should be redrawn at these same flow rates.
At 2 CFH (1 L/min.) gaseous sample tem­peratures are equilibrated to instrument temperature regardless of stream tempera­ture. At extremely high flow rates, this may not be true, but no such effect has been noted up to 18 CFH (9 L/min.).
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
748003-N November 2002
Model 868
Cell Length
mm inch
4 .157 1.14 12 2 sec
8 .315 2.28 13 2 sec 16 .630 4.56 16 2 sec 32 1.25 9.12 20 2 sec 64 2.52 18.24 25 3 sec
128 5.03 36.48 44 3 sec 232 9.13 66.12 73 6 sec
2-13 DIFFERENTIAL ANALYSIS WITH FLOW
THROUGH REFERENCE CELL
In some applications the analyzer is used to measure the difference between the concentra­tion of the component of interest in two sample streams. If so, the reference side of the ana­lyzer, as well as the sample side, utilizes a flow-
Cell Volume (CC)
(Without inlet tube)
Table 2-1. Purging Time at Atmospheric Sample Pressure
Total Volume (CC)
(Cell with inlet tube)
through cell. The sample cell receives the sample stream, which contains the higher concentration of the component of interest. The reference cell receives the stream con­taining the lower concentrations of this component.
Time For Two Volumes @ 2
Scfh (1L/min)
@ 750 mm Hg
Installation Rosemount Analytical Inc. A Division of Emerson Process Management
2-4
Model 868

3-1 STARTUP

During final factory test all controls and ad­justments were set so that all field adjust­ments can be accomplished using the front panel display and the controls located be­hind the door in the front panel.
Instruction Manual
748003-N
November 2002
SECTION 3
DETAILED STARTUP
3. Move CHECK switch to Positions 6 and
7. Position 7 shows the negative output driver signal in volts. Position 6 shows the positive output driver signal. These should be within 50 mV of each other.

3-2 CURRENT SOURCE ADJUSTMENT

Refer to Figure 3-1 below.
1. RANGE switch in CH position.
2. CHECK switch in Position 8. This dis­plays the current through the sources. The reading must be multi­plied by 100 to obtain the current in milliamps. Clockwise adjustment of R5 will increase the value. Counter­clockwise will decrease the value. Adjust for desired current as indi­cated on data sheet for the applica­tion.
3-3 DISPLAY READOUTS USING CHECK SWITCH
CHECK SWITCH
POSITION
0 +5 VDC 4.75 to 5.25 VDC 1 -5 VDC -4.5 to -5.1 VDC 2 +15 VDC +14.25 to 15.75 VDC 3 -15 VDC -14.25 to -15.75 VDC 4 +12 VDC +11.4 to 12.6 VDC 5 REF Voltage 2.4 to 2.6 VDC 6 + Source V See note 7 - Source V See note 8 Source I See note 9 Preamp Out voltage will vary
Note: Determined by application. See Factory Data Sheet in the rear of this manual.
Table 3-1. Range Switch Readings
FUNCTION READING
Figure 3-1. Source Current Adjustment
R52 R5
SOURCE ADJUSTMENT ADC SPAN
FRONT PANEL
Rosemount Analytical Inc. A Division of Emerson Process Management Startup 3-1
Instruction Manual
748003-N November 2002
Model 868
3-4 RANGE SWITCH POSITION (O.T.) OSCILLA-
TOR TUNE
Refer to Figure 3-2 below.
1. This procedure is not necessary on a routine basis. Record initial number and check for any change exceeding ±5%. To check the tuning, turn the RANGE Switch to O.T.
2. Adjust coil (located on right side of de­tector housing) clockwise until a maxi­mum reading is obtained on the display.
3. Adjust coil counterclockwise until unit reads 75% and 80% of the maximum value. This positions the system for cor­rect operation.

3-5 RANGE SWITCH

REMOTE Position
Allows the range selection unit to be con­trolled from a remote source.

3-6 TIME CONSTANT SELECTION

Set by switch (S9) located behind the front panel door. This switch selects speed of re­sponse of the electronics. The base line will be noisier on 1/2-second response than on
2.5-second response. Refer to Figure 3-2 be­low.
Note
At least one switch must be depressed, but DO NOT energize more than one switch at a time. It is recommended that the system operate in the slowest mode that is required by the system.
Positions 1, 2, and 3
Allow manual selection of range.
ZERO
RANGE
#2
RANGE
#3
RMT
O.T.
CH
RANGE
USE THUMBWHEEL TO ADJUST
Figure 3-2. Front Panel Controls
1
2
3
CHECK
SWITCH
SPAN
TIME
CONSTANT
SWITCH
S9
5.0 SEC
2.5 SEC
1.0 SEC
0.5 SEC
Startup Rosemount Analytical Inc. A Division of Emerson Process Management
3-2
Loading...
+ 49 hidden pages