The information contained in this document is subject to change without notice.
This manual is based on the production version of the Model 7D Thermal Conductivity Analyzer. Hardware
and/or software changes may have occurred since this printing.
Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc.
SNOOP® is a registered trademark of NUPRO Co.
Manual Part Number 748221-J
July 1997
Printed in U.S.A.
5-1 Range Switch Connections...................................................................... 5-2
D
RAWINGS (LOCATED IN REAR OF MANUAL
613561 Schematic Diagram, Bridge Power Supply - Regulated 5 to 15V
619710 Schematic Diagram, 15V Power Supply
624003 Schematic Diagram, Temperature Controller
652813 Schematic Diagram, Isolated Current Output
652863 Schematic Diagram, Linearizer Board
654616 Schematic Diagram, Master Board
654625 Installation Drawing, Model 7D
654642 Wiring Diagram, Model 7D
)
iv
July 1997 Rosemount Analytical 748221-JModel 7D Thermal Conductivity Analyzer
P
REFACE
PURPOSE/SAFETY SUMMARY
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site
property, all personnel authorized to install, operate and service the Model 7D Thermal
Conductivity Analyzer should be thoroughly familiar with and strictly follow the instructions
in this manual. Save these instructions.
If this equipment is used in a manner not specified in these instructions, protective
systems may be impaired.
DANGER is used to indicate the presence of a hazard which will cause severe
personal injury, death, or substantial property damage if the warning is ignored.
WARNING is used to indicate the presence of a hazard which can cause severe
personal injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor
personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation or maintenance information which is
important but not hazard-related.
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to
live parts which can cause death or serious injury. Refer servicing to qualified
personnel.
For safety and proper performance this instrument must be connected to a
properly grounded three-wire source of power.
Note:
Before supplying electrical power to the analyzer, remove power to the bridge
by disconnecting the red lead from the bridge to TB1-1 or TB1-2 (depending on
the bridge polarity). See drawing 654642. To safeguard against filament
damage, this lead should remain disconnected until proper gas flow has been
established.
748221-J Rosemount Analytical July 1997
Model 7D Thermal Conductivity Analyzer
P-1
REFACE
P
WARNING: EXPLOSION HAZARD
Do not operate the Model 7D Explosion-Proof Analyzer without the lens cover
and door in place with all bolts secured, unless location have been determined
to be non-hazardous.
WARNING: EXPLOSION HAZARD
This analyzer is of a type capable of analysis of sample gases which may be
flammable. If used for analysis of such gases, the instruments explosion-proof
enclosure must be suitable for the gas.
If explosive gases are introduced into this analyzer, the sample containment
system must be carefully leak-checked upon installation and before initial
startup, during routine maintenance and any time the integrity of the sample
containment system is broken, to ensure the system is in leak-proof condition.
Leak-check instructions are provided in Section 2.2.3.
Internal leaks resulting from failure to observe these precautions could result in
an explosion causing death, personal injury or property damage.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect
safety of this product. Use only factory documented components for repair.
WARNING: HIGH PRESSURE GAS CYLINDERS
This analyzer requires periodic calibration with known zero and standard gases.
Refer to General Precautions for Handling and Storing High Pressure Cylinders,
at the rear of this manual.
P-2
July 1997 Rosemount Analytical 748221-JModel 7D Thermal Conductivity Analyzer
SPECIFICATIONS
EPRODUCIBILITY
R
0.5% of fullscale
±
ERO DRIFT
Z
PAN DRIFT
S
OISE
N
ELL RESPONSE TIME
C
AMPLE FLOW
S
ALIBRATION GAS FLOW
C
EFERENCE GAS FLOW (IF REQUIRED
R
UPPLY PRESSURE
S
ETER
M
PERATING RANGES
O
MBIENT TEMPERATURE RANGE
A
UTPUT VOLTAGE (STANDARD NON-LINEARIZED
O
UTPUT VOLTAGE (OPTIONAL LINEARIZED
O
SOLATED CURRENT OUTPUT (OPTIONAL
I
UAL ALARMS (OPTIONAL
D
ELL MATERIALS (STANDARD CELL
C
OWER REQUIREMENTS
P
NCLOSURE
E
1
1% of fullscale per 24 hours
±
1
1% of fullscale per 24 hours
±
Less than ±0.5% of fullscale
2
30 seconds for 95% response, with sample flow of 250 cc/min.
Nominal, 50 to 350 cc/min; recommended, 250 cc/min.
Nominal, 50 to 350 cc/min; recommended, 250 cc/min.
5 to 50 cc/min.
10 to 50 psig (69 to 345 kPa)
Indicating analog meter is standard.
Various zero-based and zero-suppressed ranges, from 0% to 100%, are
available. Single range is standard; switch-selectable dual or triple range is
optional.
32°F to 100°F (0°C to 38°C). Case Temperature controlled at 117°F (47°C).
Switch selectable: 0 to 10 mV, 0 to 100 mV, 0 to 1V or 0 to 5V DC
Switch selectable: 0 to 10 mV, 0 to 100 mV, 0 to 1V or 0 to 5V DC
4 to 20 mA, maximum load 1500 ohms
)
Relay contact rating: 1.0 A, 120V AC; 5.0 A, 120V DC, resistive loads
316 stainless steel block with tungsten or Hitempco filaments. Corrosion-
resistant filament s available on order
115/230 VAC ±10%, 50/60 Hz, 250 Watts
Class I, Groups B, C and D, Division 1 hazardous locations (ANSI/NFPA 70)
REFACE
P
)
)
)
)
)
1
Zero and Span drift specifications based on ambient temperature shifts of less than 18 Fahrenheit degrees (10 Celsius
degrees) at a maximum rate of 18 Fahrenheit degrees (11 Celsius degrees) per hour.
2
Cell response time is less than 45 seconds for 95% response, with sample flow rate of 250 cc/min, for the following gas
combinations: Argon and air, nitrogen, or oxygen; carbon dioxide and argon, nitrogen, or oxygen; helium and methane;
hydrogen and methane.
748221-J Rosemount Analytical July 1997
Model 7D Thermal Conductivity Analyzer
P-3
REFACE
P
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE
For order administration, replacement Parts, application assistance, on-site or factory
repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when
you call: Model Number, Serial Number, and Purchase Order Number or Sales Order
Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive
or other hazardous material or used in such a hazardous environment, the user must
attach an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification
that the material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
TRAINING
A comprehensive Factory Training Program of operator and service classes is
available. For a copy of the Current Operator and Service Training Schedule contact
the Technical Services Department at:
Rosemount Analytical Inc.
Phone: 1-714-986-7600
FAX: 1-714-577-8006
D
OCUMENTATION
The following Model 7D Thermal Conductivity Analyzer instruction materials are
available. Contact Customer Service or the local representative to order.
748221 Instruction Manual (this document)
P-4
July 1997 Rosemount Analytical 748221-JModel 7D Thermal Conductivity Analyzer
COMPLIANCES
The explosion-proof Model 7D Thermal Conductivity Analyzer is approved by Factory
Mutual Research Corp. (FMRC) for installation in Class I, Groups B, C, and D, Division
1 hazardous locations as defined in the National Electrical Code (NEC), ANSI/NFPA-
70.
FM
APPROVED
REFACE
P
748221-J Rosemount Analytical July 1997
Model 7D Thermal Conductivity Analyzer
P-5
REFACE
P
NOTES
P-6
July 1997 Rosemount Analytical 748221-JModel 7D Thermal Conductivity Analyzer
I
NTRODUCTION
1
The Model 7D Thermal Conductivity Analyzer is designed to continuously measure the
concentration of a single component of interest in a flowing gas mixture. The
measurement is based on the different thermal conductivity's of the individual
components of the sample stream. The method is especially well suited to analysis
of two-component sample streams. However, analysis of multi-component streams is
possible if the various components of the background gas occur in relatively constant
ratio, or have similar thermal conductivity's.
Each Model 7D Analyzer is factory-assembled, as ordered, for determination of a
specified component, with specified range of concentration, contained in a
background component or background mixture of known composition. Typical
examples include: 0 to 100 % hydrogen in nitrogen; 20 to 50 % helium in methane;
and 0% to 3% carbon dioxide in air. If so ordered, the instrument is provided with two
or three ranges; selectable via a front-panel switch. Information specific to the
individual instrument is provided in the data sheet inserted in the back of this
instruction manual.
A Model 7D Analyzer consists of an analyzer module, Section 1.1, and, if ordered, an
accessory gas selector panel, Section 1.2.
1.1 ANALYZER MODULE
The analyzer module is supplied in an explosion-proof enclosure suitable for
installation in hazardous locations classified as Class I, Division 1, Groups B, C, and D
per the National Electrical Code (ANSI/NFPA 70) (See DWG 654625).
1.1.1 T
HERMAL CONDUCTIVITY CELL
The thermal conductivity cell is a metal block with separate passages for the sample
and reference gases. In all applications, the sample passage receives a continuous
flow of sample gas. Depending on the application, the reference passage may receive
a continuous flow of reference gas, or may have the reference gas sealed within.
The sample passage contains a pair of temperature-sensitive resistive filaments. The
reference passage contains a similar pair. Electrically, the filaments are connected as
legs of a Wheatstone bridge. An internal voltage-regulated power supply is
connected via a 20-ohm dropping resistor, to the bridge.
With the power supply output adjusted to provide an appropriate voltage across the
748221-J Rosemount Analytical July 1997
Model 7D Thermal Conductivity Analyzer
1-1
ODEL 7D THERMAL CONDUCTIVITY ANALYZER
M
bridge, an electric current flows through the filaments, heating them and thus
increasing their electrical resistance. The heat-dissipation rate for each filament
depends on the thermal conductivity of the surrounding gas.
Initially, with suitable downscale calibration gas flowing through the sample passage
(and also through the reference passage if of the flow-through configuration), the
bridge is balanced. Thereafter, any change in the relative proportions of the
components passing through the sample passage changes the thermal conductivity of
the gas mixture, causing a temperature differential between sample and reference
filaments. The resultant change in filament resistance unbalances the bridge,
applying a signal to the electronic circuitry (Section 1.1.2).
ZERO SPAN
ALARM 1 ALARM 2
50%
25%
75%
0% 100%
Rosemount Analytical
RANGE
1 3
2
25%
0% 100%
Model 7D
Thermal Conductivity
Analyzer
50%
75%
ZERO SPAN
ALARM 1 ALARM 2
50%
25%
0%100%
75%
1 3
Rosemount Analytical
RANGE
2
50%
25%
0%100%
Model 7D
Thermal Conductivity
Analyzer
75%
F
IGURE
1-2
1-1. M
ODEL
Rosemount Analytical
7D F
RONT PANEL CONTROLS
July 1997
748221-J
NTRODUCTION
I
1.1.2 E
LECTRONIC CIRCUITRY
The analyzer module contains solid-state circuitry that conditions the bridge-imbalance
signal as required to provide readout on the front-panel meter. In addition, a
field-selectable output for a voltage-type recorder is provided as standard. A
field-selectable output of 4 to 20 mA for a current-actuated recorder or other device is
obtainable through use of an optional plug-in circuit board. A calibration curve can be
used to convert meter or recorder readings into concentration values. Typical
calibration curves are supplied for standard ranges. Calibration curves for special
ranges are available as options.
To avoid use of a calibration curve in an application where it would otherwise be
required, the analyzer may be equipped with an optional linearizer board. If so, the
linearizer is factory set for a given range only, and is not usable on another range.
Note that a linearizer is usable only if nonlinearity at midscale does not exceed
approximately 20% of fullscale.
Isolated Current Output Board
Linearizer Board
±15V Power Supply
Bridge Power Supply
Master Board
Temperature Control Assembly
REF Hinge Side
AC Power Input
Terminal Block TB6
Fuse
Cell Enclosure
Alarm Kit
F
IGURE
748221-J Rosemount Analytical July 1997
1-2. M
ODEL
7D M
AJOR COMPONENT LOCATIONS
Model 7D Thermal Conductivity Analyzer
1-3
NTRODUCTION
1
1
I
1.2 GAS SELECTOR PANEL
If so ordered, the analyzer module is provided with an appropriate gas selector panel,
Figure 1-3. The gas selector panel permits selection, flow adjustment, and flow
measurement for the various gases: sample; flowing reference gas, if used; and
downscale and upscale calibration gases. Proper choice of a gas selector panel
depends on:
1. Configuration of the thermal conductivity cell, i.e., flowing or sealed-in reference
gas.
2. Composition of the sample stream. For non-corrosive streams, the gas selector
panel is assembled with brass components. For corrosive streams, stainless steel
is used.
F
IGURE
1 Reference Gas
Flow Meter
Downscale
Calibration Gas
Needle Valve
Upscale
Calibration Gas
Needle Valve
Provided only if thermal conductivity cell uses flowing reference gas.
1-3. T
YPICAL GAS SELECTOR PANEL
DOWNSCALE
CAL GAS
UPSCALE
CAL GAS
REF
SAMPLE
DESCRIPTION
Sample/Calibration
Gas Flow Meter
SAMPLE
REFERENCE
Sample Gas
Needle Valve
Reference Gas
Needle Valve
NUMBER
PART
B
RASS AND COPPER CONSTRUCTION FOR USE WITH SEALED REFERENCE
S
TAINLESS STEEL CONSTRUCTION FOR USE WITH SEALED REFERENCE
B
RASS AND COPPER CONSTRUCTION FOR USE WITH FLOWING REFERENCE
S
TAINLESS STEEL CONSTRUCTION FOR USE WITH FLOWING REFERENCE
T
ABLE
1-4
1-1. A
VAILABLE GAS SELECTOR PANELS
July 1997 Rosemount Analytical 748221-JModel 7D Thermal Conductivity Analyzer
113357
113920
117195
118210
I
NSTALLATION
2
2.1 FACILITY PREPARATION
Sections 2.1.1 through 2.1.4 provide information that may be required prior to
installation.
2.1.1 I
2.1.2 C
2.1.3 F
2.1.4 L
NSTALLATION DRAWINGS
For outline and mounting dimensions of the analyzer and gas selector panel modules,
refer to drawing 654625 and Figure 2-1or 2-2.
USTOMER ELECTRICAL CONNECTIONS
Customer electrical connections are shown in Figure 2-5.
LOW DIAGRAMS
For gas connections, refer to appropriate flow diagram:
• Analyzer using sealed reference gas, Figure 2-3
• Analyzer using flowing reference gas, Figure 2-4.
OCATION AND MOUNTING
OCATION
L
Proper location for the analyzer depends on two basic considerations:
• Accessibility to the samp ling point
• Protection of the instrument
Ideally, the analyzer should be located as close to the sampling point as possible. Short
sample lines reduce time lag in readings. In practice, however, protection of the
instrument sometimes calls for more remote placement.
The analyzer should be mounted in a clean, dry atmosphere. Ambient temperature
should be within the range of 32oF to 100 F (0oC to 38oC).
OUNTING
M
The analyzer is designed for surface mounting, utilizing the hardware provided. Refer
to drawing 654625.
748221-J Rosemount Analytical July 1997
Model 7D Thermal Conductivity Analyzer
2-1
NSTALLATION
I
F
IGURE
Span Gas In
2-1. G
S
To AnalyzerZero Gas In
Sample In
8.25
[210]
Mounting Holes (4)
NO. 8 Flat Head Screw
ZEROSPAN
10.25
[260]
SAMPLE
AS SELECTOR PANEL FOR THERMAL CONDUCTIVITY CELL WITH
EALED-IN REFERENCE GAS
F
IGURE
Sample Gas
2-2. G
To Analyzer
Sample Inlet
Ref Gas
To Analyzer
Ref Gas Inlet
Upscale
Calibration Gas
Downscale
Calibration Gas
3.25
[83]
.12
[3]
1.25
[32]
CALIBRATION GAS
8.25
[210]
CALIBRATION GAS
Mounting Holes (4)
NO. 8 Flat Head Screw
10.25
[260]
DOWNSCALE
REF SAMPLE
UPSCALE
AS SELECTOR PANEL FOR THERMAL CONDUCTIVITY CELL USING
F
LOWING REFERENCE GAS
SAMPLE
REF
2-2
July 1997 Rosemount Analytical 748221-JModel 7D Thermal Conductivity Analyzer