Rosemount Manual: 498CL pH-independent Free Chlorine Sensor Abridged | Rosemount Manuals & Guides

Model 498 CL-01
pH-independent Free Chlorine Sensor
Instruction Sheet
PN 51A-498CL01/rev.C June 2013
SPECIFICATIONS - SENSOR
Pressure: Sensor must be installed in a flow cell
that drains to open atmosphere.
flow cell, PN 24091-01 only.
Wetted materials: PVC, polyethersulfone, poly-
ester, Viton1, silicone, copper
Cathode: gold mesh
1
Viton is a registered trademark of DuPont Performance
Elastomers.
SPECIFICATIONS - FLOW CELL
Inlet pressure: 65 psig (549 kPa abs) maximum
Temperature: 122ºF (50ºC) maximum
Process connection: Compression fitting for
1/4 inch OD tubing
Wetted materials: polycarbonate, polyester,
silicone, 316 stainless steel
CAUTION
SENSOR/PROCESS APPLICATION
COMPATIBILITY
The wetted sensor materials may not be com­patible with process com position and operating conditions. Application compat ibility is entirely the responsibility of the user.
For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit our website at www.emersonprocess.com/raihome/liquid/.
INSTALLATION
The sensor leaves the factory with a shipping membrane in place. DO NOT REMOVE THE SHIPPING MEMBRANE UNTIL READY TO PUT THE SENSOR IN SERVICE.
FIGURE 1. Installation.
The plastic adapter shipped with the flow cell is
not used.
outlet
inlet
1. Remove the red protective cap from the sensor.
2. Holding the membrane end pointing up (cable end point­ing down), unscrew the retainer cap and remove the ship­ping membrane. See Figure 6. It is not necessary to remove the O-ring. Save the shipping membrane. It should be reinstalled when the sensor is not in use.
3. Still holding the membrane end pointing up, install the chlorine membrane. The chlorine membrane is in the plastic bag attached to the sensor. Screw the retainer back in place.
4. Place the sensor in the flow cell and start the flow. Do
not store or operate the sensor for more than about 45 minutes in a non-flowing sample. Sensor must
drain to open atmosphere. No back pressure. Do not leave the sensor in air.
Keep sample flow constant and between 1 and 2 gph (60-120 mL/min).
MAINTENANCE
MODEL 498 CL-01 WIRING
WIRING
FIGURE 3. Wiring Sensor to Model 1055 Analyzer
(wall/pipe mount version)
FIGURE 2. Wiring Sensor to Model 1055 Analyzer
(panel mount version)
FIGURE 4. Wiring Sensor to Model 54eA Analyzer
FIGURE 5. Pin Out Diagram for Model 498CL-01-VP
Sensor (top view of connector end of sensor)
The sensor requires routine maintenance, which includes cleaning the membrane as well as periodically replac­ing the membrane and fill slurry. The fill slurry and membrane should be replaced at approximately three-month intervals. The frequency of cleaning can be determined only by experience.
CLEANING THE MEMBRANE
Keep the membrane clean and free from dirt and algae. Clean the membrane with water sprayed from a wash bot­tle. Do not use tissues to clean the membrane. Touching the membrane may bend the mesh cathode and
permanently damage the sensor.
REPLACING THE ELECTROLYTE SLURRY AND MEMBRANE
1. Unscrew the membrane retainer and remove the membrane assembly and O-ring. See Figure 6.
2. Remove the fill plug.
CAUTION:
Fill solution and solid may cause irritation. Avoid contact with skin and eyes. May be harm­ful if swallowed. Read and follow manual.
2
Loading...
+ 2 hidden pages