Rosemount Manual: 400A Hydrocarbon Analyzer-Rev M (Cat # 194106 / Serial # beginning with 2000001) | Rosemount Manuals & Guides

Instruction Manual
748262-M May 2002
Model 400A
Hydrocarbon Analyzer
NOTE: This manual is applicable to Model 400A, Catalog Number 194106 with Serial Numbers beginning with 2000001.
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical representative
for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Teflon is a Registered Trademark of E.I. duPont de Nemours and Co., Inc. SNOOP is a registered trademark of NUPRO Co.
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Model 400A
PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-4
Documentation....................................................................................................................................P-5
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Description.............................................................................................................................1-1
1-2 Analyzer Mounting Options ...................................................................................................1-1
1-3 Fuel Gas Options...................................................................................................................1-1
1-4 Output Options ......................................................................................................................1-2
a. Isolated Remote Range Change and Identification ........................................................1-3
b. Range Trim Option..........................................................................................................1-3
1-5 Sample Pump Option ............................................................................................................1-3
1-6 Gas Safety Features..............................................................................................................1-3
1-7 Specifications ........................................................................................................................1-4
Instruction Manual
748262-M
May 2002

TABLE OF CONTENTS

2-0 INSTALLATION ....................................................................................................................2-1
2-1 Unpacking..............................................................................................................................2-1
2-2 Location .................................................................................................................................2-1
2-3 Voltage Requirements...........................................................................................................2-1
2-4 Fuel and Air Requirements....................................................................................................2-2
a. Fuel Gas..........................................................................................................................2-2
b. Air....................................................................................................................................2-2
2-5 Sample Handling ...................................................................................................................2-2
2-6 Gas Connection.....................................................................................................................2-3
2-7 Leak Check............................................................................................................................2-4
2-8 Electrical Connections ...........................................................................................................2-4
a. Line Power Connection...................................................................................................2-5
b. Voltage Output Selection and Cable Connections for Recorder ....................................2-5
c. Voltage to Current Output Board (Optional) ...................................................................2-5
d. Auxiliary Contacts ...........................................................................................................2-5
e. Remote Range Control and Indication............................................................................2-5
f. Sample Pump Accessory................................................................................................2-7
3-0 OPERATION .........................................................................................................................3-1
3-1 Initial Startup and Calibration ................................................................................................3-1
a. Selection of Calibration Method and Associated Standard Gas(es)...............................3-3
3-2 Calibration Procedure............................................................................................................3-4
3-3 Range Switch ........................................................................................................................3-5
3-4 Range Trim Option ................................................................................................................3-6
3-5 Routine Operation .................................................................................................................3-8
3-6 Recommended Calibration Frequency..................................................................................3-8
3-7 Shutdown...............................................................................................................................3-8
3-8 Obtaining Maximum Sensitivity .............................................................................................3-8
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Instruction Manual
748262-M May 2002
4-0 THEORY................................................................................................................................4-1
4-1 Principles of Operation ..........................................................................................................4-1
4-2 Burner ....................................................................................................................................4-1
a. Response to Different Hydrocarbons..............................................................................4-2
4-3 Analyzer Flow System ...........................................................................................................4-3
4-4 Preamplifier Board.................................................................................................................4-3
4-5 Main Electronics Board..........................................................................................................4-4
a. Post Amplifier ..................................................................................................................4-4
b. Digital Display .................................................................................................................4-4
c. Span ................................................................................................................................4-4
d. Range Trim (Option) .......................................................................................................4-4
e. Remote Range Control ...................................................................................................4-4
4-6 Heater/Fan Temperature Assembly ......................................................................................4-6
4-7 Ignition Circuit........................................................................................................................4-6
4-8 System Power Supplies.........................................................................................................4-6
4-9 Flame-Out Board ...................................................................................................................4-6
5-0 SERVICE AND TROUBLESHOOTING ................................................................................5-1
5-1 System Checkout ..................................................................................................................5-1
a. Amplifier Zero Adjustment...............................................................................................5-1
b. Zero Current Adjustment.................................................................................................5-1
c. Electrical Leakage Check ...............................................................................................5-1
d. Flame Ignition..................................................................................................................5-1
e. Noise Check....................................................................................................................5-1
f. Overall Sensitivity Check ................................................................................................5-1
g. Stability Check ................................................................................................................5-2
5-2 Servicing Flow System and Burner .......................................................................................5-2
a. Burner Disassembly and Cleaning..................................................................................5-2
b. Thermistor .......................................................................................................................5-3
c. Fuel and Air Restrictors ..................................................................................................5-3
d. Sample Capillary .............................................................................................................5-4
5-3 Troubleshooting.....................................................................................................................5-4
Model 400A
6-0 REPLACEMENT PARTS ......................................................................................................6-1
6-1 Circuit Board Replacement Policy .........................................................................................6-1
6-2 Recommended Replacement Parts List................................................................................6-1
6-3 Matrix .....................................................................................................................................6-2
7-0 RETURN OF MATERIAL ......................................................................................................7-1
7-1 Return Of Material .................................................................................................................7-1
7-2 Customer Service ..................................................................................................................7-1
7-3 Training..................................................................................................................................7-1
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Figure 2-1. Range Input-Output Board..................................................................................... 2-6
Figure 2-2. Range Control and Output Connections................................................................ 2-7
Figure 3-1. Power Switch Location..............................................................................................3-2
Figure 3-2. Typical Curves of Downscale Response vs. Time for Various Hydrocarbons.......... 3-3
Figure 3-3. Example 1 Display .................................................................................................... 3-6
Figure 3-4. Example 2 Display .................................................................................................... 3-6
Figure 3-5. Example 3 Display .................................................................................................... 3-6
Figure 3-6. Range Trim Option Schematic.................................................................................. 3-8
Figure 3-7. Typical Curve of Analyzer Response vs. Pressure Setting on Sample Pressure .... 3-9
Figure 3-8. Typical Curves of Analyzer Response vs. Pressure Setting on Fuel Pressure
Figure 3-9. Typical Curves of Analyzer Response vs. Pressure Setting on Air Pressure
Figure 4-1. Flame Ionization Detection Theory ........................................................................... 4-1
Figure 5-1. Effect of Analyzer Temperature on Background Signal – Typical Curve.................. 5-2
Instruction Manual
748262-M
May 2002

LIST OF ILLUSTRATIONS

Regulator ............................................................................................................... 3-9
Regulator ............................................................................................................. 3-10

LIST OF TABLES

Table 1-1. Fuel Gas vs. Analyzer Characteristics................................................................... 1-3
Table 2-1. Range Control........................................................................................................ 2-6
Table 3-1. Internal Pressure Regulators Settings................................................................... 3-1
Table 3-2. Range Switch Settings........................................................................................... 3-5
Table 4-1. Approximate Effective Carbon Numbers ............................................................... 4-2
Table 4-2. Input Sensitivities...................................................................................................4-4
Table 4-3. Fullscale Sensitivity ............................................................................................... 4-4
Table 5-1. Troubleshooting ..................................................................................................... 5-5

LIST OF DRAWINGS

(LOCATED IN REAR OF MANUAL)
620424 Schematic Diagram, Preamplifier 620429 Schematic Diagram, Main Electronics 620434 Schematic Diagram, Isolated V/I 622883 Flow Diagram, Model 400A 623190 Burner Parts List 624003 Schematic Diagram, Temperature Control Board 654328 Installation Drawing, Model 400A 654905 Schematic Diagram, Relay Board 654910 Schematic Diagram, Relay Board 654913 Assembly Instructions, I/O Board 654998 Assembly Instructions, Relay Board 655354 Pictorial Wiring Diagram, Model 400A 657546 Schematic Diagram, Flame-Out Board
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
748262-M May 2002
Model 400A
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 400A

PREFACE

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the 400A.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
DANGER .
748262-M
May 2002
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
748262-M May 2002
Model 400A

SAFETY SUMMARY

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the Model 400A Hydrocarbon Analyzer should be thoroughly familiar with and strictly follow the instructions in this manual. Save these
instructions.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
WARNING.
POSSIBLE EXPLOSION HAZARD
Ensure that all gas connections are made as labeled and are leak free. Improper gas connections could result in explosion or death.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 400A
WARNING .
POSSIBLE EXPLOSION HAZARD
Do not apply power to analyzer or ignite burner until all leak checks have been performed and until the environment of the analyzer has been determined to be non-hazardous. See Section 2-7, page 2-4 for leak check procedure.
This instrument uses a fuel containment hydrogen. The instrument is designed to protect against the formation of an explosive gas mixture within the enclosure. It must NOT be operated if the in­ternal ventilation fan is not functioning. Do NOT operate without factory installed fuel flow restric­tor in place.
Check the fuel supply and containment system for leaks, both inside and outside the analyzer, upon installation, before initial startup, during routine maintenance, or any time the integrity of the fuel containment system is broken, to assure that the system is leak-tight. Refer to Leak Check Procedure, Section 2-7, page 2-4.
The enclosure is protected by a continuous dilution air purge system, as recommended in IEC* Publication 79-2 (Third Edition - 1983). Do not operate unit unless the continuous dilution purge system is properly installed and functioning. Upon startup or loss of pressurization, purge for fif­teen minutes before restoring power unless the internal atmosphere is known to be non-explosive.
748262-M
May 2002
CAUTION .
HIGH PRESSURE GAS CYLINDERS
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
This analyzer requires periodic calibration with known zero and standard gases. See General Pre­cautions for Handling and Storing High Pressure Cylinders, page P-4 .
WARNING
TOPPLING HAZARD
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left side of the chassis.
When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise) orientation.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the latch; how­ever, further extraction must be done very carefully to insure the chassis does not fall out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling.
Failure to observe these precautions could result in personal injury and/or damage to the product.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
748262-M May 2002
Model 400A
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125 never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 400A

DOCUMENTATION

The following Model 400A instruction materials are available. Contact Customer Service Center or the local representative to order.
748262 Instruction Manual (this document)
748262-M
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
748262-M May 2002
Model 400A
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
SECTION 1

DESCRIPTION AND SPECIFICATIONS

1-1 DESCRIPTION

The Model 400A Hydrocarbon Analyzer automatically and continuously measures the concentration of hydrocarbons in a gas stream. Typical applications include monitor­ing atmospheric air for low-level hydrocarbon contaminants and determining the hydrocar­bon content of exhaust emissions from inter­nal combustion engines.
The analyzer utilizes the flame ionization method of detection. The sensor is a burner in which a regulated flow of sample gas passes through a flame sustained by regulated flows of a fuel gas and air. Within the flame, the hy­drocarbon components of the sample stream undergo a complex ionization that produces electrons and positive ions. Polarized elec­trodes collect these ions, causing current to flow through an electronic measuring circuit. The ionization current is proportional to the rate at which carbon atoms enter the burner, and is therefore a measure of the concentra­tion of hydrocarbons in the original sample. The analyzer provides readout on a front-panel digital display and a selectable output for an accessory recorder.
descriptions of the principal standard and op­tional features.

1-2 ANALYZER MOUNTING OPTIONS

WARNING
INSTALLATION
For safety, the analyzer should be installed in a non-confined, ventilated space.
The standard analyzer is housed in a case designed for bench-top use, or if desired, the analyzer may be mounted in a cabinet or rack using RETMA spaced mounting holes. Outline dimensions are shown on DWG 654328.

1-3 FUEL GAS OPTIONS

For burner fuel gas, the standard analyzer re­quires 40% hydrogen/60% nitrogen or helium. Through installation of the optional 400A hy­drogen fuel kit (P/N 622576), the analyzer may be converted to use 100% hydrogen. This kit may be ordered as a factory installed option or supplied as an option for installation by the user.
To ensure stable, drift-free operation, particu­larly in high-sensitivity applications, an internal temperature controller maintains the analyzer interior at a constant 50°C. This feature mini­mizes temperature-dependent variations in (a) electronic current-measuring circuitry, and (b) adsorption/desorption equilibrium of back­ground hydrocarbons within the internal flow system.
To minimize system response time, an inter­nal sample-bypass feature provides high-velocity sample flow through the ana­lyzer.
The Model 400A may be equipped with vari­ous optional features in addition to, or instead of, the standard features of the basic instru­ment. The following paragraphs provide brief
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
The preferred type of fuel depends on the particular application and the characteristics of the sample gas:
For measuring low-level hydrocarbons in am­bient air, or in other sample gas with relatively constant oxygen content, 100% hydrogen is preferable. It provides the highest obtainable sensitivity and the maximum stability. Zero drift caused by ambient temperature varia­tions of the fuel cylinder is somewhat lower for 100% hydrogen than for mixed fuel. (With ei­ther fuel, it is desirable to maintain cylinder temperature constant.)
For monitoring vehicular exhaust emissions, or other sample gas with varying oxygen content, mixed fuel is preferable; and a hy­drogen/helium mixture is more desirable than
Instruction Manual
748262-M May 2002
Model 400A
a hydrogen/nitrogen mixture. With this type of sample, the use of mixed fuel gas minimizes the error introduced by oxygen synergism. An effective way to reduce the effect of internal oxygen is to dilute it with an inert gas. This might be accomplished by a constant dilution of sample and calibration gases ahead of the burner but it is simpler and more accurate to provide that diluent in the form of premixed fuel. Both nitrogen and helium have been used as a diluent, but helium has proven to be most effective in improving the equality of re­sponse to the various species of hydrocar­bons.
As indicated earlier the flame output signal is optimum when the ratio of hydrogen flow to inert flow is about 40/60; therefore, this is the chosen composition for hydrogen/helium pre­mixed fuel.
The sample flow is kept low to maximize the dilution effect while still providing adequate sensitivity. The burner air flow is normally about four times the fuel flow, and changes have little effect on signal strength. For a given sample flow, the signal can be opti­mized by adjusting the fuel flow rate.
Typical flow rates with premixed fuel:
Fuel 100 cc/min
Sample 7 cc/min
Air 400 cc/min
It is worth noting that with a 40/60 premixed fuel, the above flows amount to 40 cc (8%) hydrogen, 67 cc (13%) inert plus sample and 400 cc (79%) air, which compare closely to the 30 cc (8%) hydrogen, 45 cc (12%) in­ert/sample and 300 cc (80%) air given earlier for straight hydrogen fuel.
Since the sample flow in the case of mixed fuel operation is only about one-sixth of that with straight hydrogen fuel, it is clear that higher sensitivity is obtained with straight hy­drogen fuel operation. However, in any appli­cation where the sample contains more than one species of hydrocarbon and/or a varying concentration of oxygen, the mixed fuel op­eration should be used.
The mixed fuel is recommended, not only for sample containing variable concentrations of oxygen, but also for two specific pure gas ap­plications. The first is the case of pure hydro­gen samples. The other is the case of pure oxygen samples. If straight oxygen samples are used with straight hydrogen fuel, the mix­ture entering the burner is essentially 40% H
/60% O2, which tends to produce an unsta-
2
ble signal. The mixed fuel works better. Note that the choice of fuel determines certain analyzer characteristics, as tabulated in Table 1-1.

1-4 OUTPUT OPTIONS

The standard analyzer provides (a) direct digital readout in percent of full scale on a front-panel display calibrated linearly from 0 to 100 %, (b) a selectable buffered output of 0 to
0.1 VDC, 1 VDC or 5 VDC suitable for a re­corder and (c) a 0 to 5 VDC un-buttered ac­cessory output.
An isolated output of 4 to 20 mA DC (max. load resistance 700 ohms) is obtainable through use of the optional current output board, P/N 620433, installed either during factory assembly or as a subsequent addition. When installed, this board uses the accessory 0 to 5 VDC output as an input signal and re­places this function at the output terminals.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
ANALYZER CHARACTERISTICS 100% H2
Fullscale Sensitivity
Fuel Consumption 35 to 40 cc/min 75 to 80 cc/min
Operating Range for SAMPLE Pres­sure Regulator
Table 1-1. Fuel Gas vs. Analyzer Characteristics
a. Isolated Remote Range Change and
Identification
This option provides a 24 VDC operation of remote range and identification as well as providing terminals for flame out indi­cation.

b. Range Trim Option

This option provides individual potenti­ometers for each range to allow adjust­ment of individual differences in bottled gas.

1-5 SAMPLE PUMP OPTION

To provide the required sample flow, the sample gas must be under adequate pressure when applied to the analyzer inlet. Refer to Section 2-4, page 2-2. To permit analysis of gases at atmospheric or sub-atmospheric pressure, the analyzer may be shipped with a sample pump accessory, P/N 621062.
Adjustable from 1 ppm CH4 to 2% CH4
4 to 5 psig (27 to 34.5 kPa)
restrictor, ventilation holes in the enclosure and an internal circulation fan serve to dilute and dissipate the hydrogen fuel for a worst case leak to a safety factor below 25% of the LEL of hydrogen. The design basis for this system presumes 100% hydrogen fuel at 50 psig inlet pressure. 40% hydrogen fuel and lower inlet pressure serve to further reduce hydrogen concentration in the event of a leak. In reality, an open fitting leak would never oc­cur. As a leak developed the burner would eventually be starved of fuel and flame-out would occur at a leak equivalent of a loss of about 20 psig fuel pressure, thus actuating the fuel shut-off solenoid valve system.
If the sample is flammable, accessory kit P/N 624080, must be utilized. This kit provides a restrictor to limit sample flow and a solenoid valve to shut-off sample in the event of burner flame-out. The design basis for this kit pre­sumes a maximum sample flow rate of 470 cc/min and a sample with LEL not below that of hydrogen (4% v/v in air).
40% H2/60% N2 or 40%
H2/He
Adjustable from 4 ppm CH4 to 10% CH4
1.5 to 5 psig
(10.3 to 34.5 kPa)

1-6 GAS SAFETY FEATURES

The Model 400A is designed to provide a high degree of operational safety. In all analyzers, a front-panel LED indicates that the burner flame is lit. In addition, fuel gas is automati­cally shut off when a flame-out condition oc­curs.
All tubing ahead of the burner is rigid metallic tubing made up with ferrule/nut type compres­sion fittings. However, should there be an in­ternal fuel leak, an inlet fuel flow limiting
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
POSSIBLE EXPLOSION HAZARD
Protection against explosion depends upon special fuel flow limiting restrictor in fuel inlet fitting. Do not remove fuel inlet restrictor. Replace only with factory sup­plied fitting.
WARNING
Instruction Manual
748262-M May 2002

1-7 SPECIFICATIONS

Power Requirements..................... 115 VAC ±10%, 50/60 ±3 Hz, 250 W
Operating Temperature ................. 32°F to 110°F (0°C to 43°C)
Case Temperature ........................ Controlled at 122°F (50°C)
Ambient Humidity .......................... 95% relative humidity, but not in excess of 34°C wet bulb
Dimensions.................................... 8.75 in (22.2 cm) H
Weight ........................................... 22 lbs (10 kg)
Repeatability.................................. 1% of fullscale for successive identical samples
Response Time ............................. 90% of fullscale in 0.6 seconds with sample bypass flow at 3
Analyzer Fullscale Sensitivity
Standard ............................... Adjustable from 4 ppm CH4 to 1% CH4. (Adjustable from 100 ppm
Equipped with 100%
Hydrogen Fuel Assembly ...... Adjustable from 1 ppm CH4 to 0.25% CH4
Analyzer Fuel Gas Requirements
Standard Analyzer ................. 75 to 80 cc/min premixed fuel consisting of 40% hydrogen and 60%
Equipped with 100%
Hydrogen Fuel Assembly ...... 35 to 40 cc/min of clean, zero grade hydrogen (THC <0.5 ppm) at
Sample Gas Requirements
Non-Flammable Samples ...... 0.35 to 3.0 liters/minute at 5 to 10 psig (34 to 69 kPa)
Flammable Samples.............. 470 cc/minute maximum for safety1
Burner Air Requirements............... 350 to 400 cc/minute of zero grade (THC <1 ppm) air, supplied at
Sample Bypass Flow..................... 0.3 to 3.0 liters/minute
Stability .......................................... Electronic stability at maximum sensitivity is 1% of fullscale
Range ............................................ Eight ranges: 1, 2.5, 10, 25, 100, 250, 1000 and REMOTE.
Model 400A
temperature.
18.75 in. (47.6 cm) W
15.88 in. (39.7 cm) D Recommended panel cutout is 17.75 in. X 8 25 in. (45.1 cm x 21.0 cm). May be mounted in standard 19 inch rack.
liters/minute
CH4 to 10% CH4 using high-range capillary.)
nitrogen or helium (THC <0.5 ppm) supplied at 45 to 50 psig (309 to 344 kPa) at instrument
45 to 50 psig (309 to 344 kPa) at instrument
25 to 50 psig (172 to 344 kPa)
throughout ambient temperature range of 32°F to 110°F (0°C to 43°C). Built-in temperature controller minimizes effect of ambient temperature variations on internal flow and electronic systems.
In addition SPAN control provides continuously variable adjustment within a dynamic range of 4:1
1
Safety design basis presumes flammable sample having LEL not less than that of hydrogen (4% v/v in air).
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Output............................................ 1) 0 to 5 VDC, 0 to 1 VDC, 0 to 0.1 VDC fully buffered - standard
Safety Features ............................. Flame-on indication and automatic flame-out fuel shutoff is
Contacts ........................................ Form A contact operates in parallel with flame-out fuel shut-off
Temperature Control ..................... Setpoint maintained at 122°F (50°C)
Data Display .................................. 3-1/2 digit LED, characters 0.52 inches high, range 0000 to 1999
Range Display ............................... 1 digit LED, character 0.52 inches high (1 to 7 normal ranges, 0 to
Remote Range Control.................. Standard, fully isolated range control and range ID is optional
Instruction Manual
748262-M
May 2002
(for 0 to 100.0%).
2) 4 to 20 mA isolated voltage to current - optional (maximum load resistance 700 ohms)
3) 0 to 5 VDC accessory output un-buffered - standard (for 0 to
100.0%) available when current option is not used.
standard.
All metal tubing with ferrule/nut compression fittings to minimize potential fuel leaks. Self-ventilated system maintains internal atmosphere below 25% of LEL for worst case internal leakage.
solenoid contact rating (24 VDC at 1 A) for sample shut-off by use of factory ordered kit (PN 624080) if sample is flammable (hydrogen).
remote control)
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
748262-M May 2002
Model 400A
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
SECTION 2

INSTALLATION

2-1 UNPACKING

Carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or its contents are damaged. Retain the carton and packing materials until the instrument is op­erational.

2-2 LOCATION

Install analyzer in a clean area, not subject to excessive vibration or extreme temperature variations. Preferably, the analyzer should be mounted near the sample stream, to minimize sample-transport time.
WARNING
INSTALLATION RESTRICTIONS
For safety, the analyzer should be installed in a non-confined, ventilated space. Do not block any of the vent holes at the top of each side panel of the instrument as they are part of the safety system.
A thermostatically controlled heating circuit holds internal temperature of the analyzer to the correct operating temperature for ambient temperatures in the range 32°F to 110°F (0°C to 43°C).

2-3 VOLTAGE REQUIREMENTS

WARNING
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified per­sonnel.
For safety and proper performance this in­strument must be connected to a properly grounded three-wire source of power.
This instrument was shipped from the factory configured to operate on 115 VAC, 50/60 Hz electric power. Verify that the power source conforms to the requirements of the individual instrument, as noted on the name-rating plate.
NOTE:
230 VAC operation requires an accessory transformer mounted external to the in­strument. If external 230 VAC power is provided, do not change the switch setting on the Temperature Control Board, which is factory-selected for 115 VAC. A 230 VAC switch setting will cause the case tempera­ture controller to malfunction.
The cylinders of fuel, air, and calibration gas(es) should be located in an area of rela­tively constant ambient temperature.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
748262-M May 2002
Model 400A

2-4 FUEL AND AIR REQUIREMENTS

WARNING
INSTALLATION RESTRICTIONS
Fuel, air and calibration gas cylinders are under pressure. Mishandling of gas cylin­ders could result in death, injury or prop­erty damage. Handle and store cylinders with extreme caution and in accordance with manufacturer's instructions. Refer to General Precautions for Handling and Storing Gas Cylinders, page P-4.
During normal operation, the analyzer uses fuel and air to maintain the burner flame. Cri­teria for selection of these gases are given in Sections 2-4a (page 2-2) and 2-4b (page 2-2). In addition, the analyzer requires suitable standard gas(es) for calibration. Refer to Sec­tion 3-1a (page 3-3).
Each gas used should be supplied from a tank or cylinder equipped with a clean, hydrocar­bon-free, two-stage regulator. In addition, a shutoff valve is desirable. Install the gas cylin­ders in an area of relatively constant ambient temperature.

a. Fuel Gas

NOTE:
Always assure the sample flow is pre­sent when using the 100% hydrogen fuel option. Absence of sample flow can result in burning of detector tip when using 100% hydrogen.

b. Air

Burner air should also be relatively free of hydrocarbons in order to assure a low background signal. Several grades of air are supplied by various gas vendors for this use. A maximum total hydrocarbon content of less than 1 ppm (THC < used as a zero standard).
An alternate source of pure air for burner and zero gas can be provided by a dia­phragm pump and heated palladium catalyst which effectively removes mod­erate amounts of both hydrocarbons and carbon monoxide from normal ambient air on a continuous basis.

2-5 SAMPLE HANDLING

CAUTION
BYPASS GAUGE PROTECTION
The standard analyzer is equipped to use only mixed fuel, i.e. 40% hydrogen/60% nitrogen or helium. Such blends are sup­plied by many gas vendors specifically for this use, with a guaranteed maximum to­tal hydrocarbon content of 0.5 ppm, measured as methane (THC < 0.5 ppm). This specification should be used when buying such mixtures.
When the analyzer is equipped with the optional hydrogen fuel kit, P/N 622576, 100% hydrogen fuel is to be used. This is also supplied by many gas vendors spe­cifically for this use, with the same guar­anteed total hydrocarbon content (THC <
0.5 ppm) which, again, should be speci­fied when buying the gas.
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
When applying sample pressures greater than 5 psig, insure that the bypass regula­tor is fully open to protect the bypass gauge.
Operating range for the internal sample pres­sure regulator is 4 to 5 psig (28 to 35 kPa) for an analyzer using 100% hydrogen fuel, and
1.5 to 5 psig (10 to 35 kPa) for an analyzer using mixed fuel. With either fuel, sample and calibration gas(es) must be supplied to the sample inlet at a pressure slightly, but not ex­cessively higher than the desired setting on the internal sample pressure regulator. The criterion for correct supply pressure is that the gas flow discharged from the by-pass outlet must be between 0.5 and 3.0 liters/minute to operate within the control range of the sample pressure regulator, and preferably should be between 2 and 3 liters/minute to minimize system response time.
Model 400A
Instruction Manual
748262-M
May 2002
Note that use of excessive bypass flow will not only cause the sample pressure regulator to operate outside its control range, but will also result in rapid depletion of sample and standard gases.
If the analyzer is equipped with the accessory 400A sample pump, P/N 621062, the accept­able pressure range at the pump inlet is ap­proximately -1 to +2.5 psig (7 to 17 kPa). If the pump is used, it will automatically provide a sample bypass flow within the correct range. If the analyzer is not equipped with sample pump, adjustment of the bypass flow is ob­tainable by inserting an external flow control­ler or throttle valve into the external sample line, upstream from the sample inlet. Flow may be measured by connecting a flowmeter to the by-pass outlet.
WARNING
POSSIBLE EXPLOSION HAZARD
In the event that flammable sample is to be introduced into this analyzer, it must be equipped with accessory kit PN 624080, which restricts sample flow and provides automatic sample shutoff in the event of burner flameout. DO NOT OPERATE WITHOUT SAMPLE FLOW RESTRICTOR IN PLACE. The sample containment system should also be thoroughly leak checked. This kit is designed considering applica­tion on hydrogen sample (LEL=4% v/v). The instrument must not be used on a sample having a LEL less than 4% in air.
WARNING
POSSIBLE EXPLOSION HAZARD
Do not apply power to analyzer or ignite burner until all leak checks have been per­formed and until the environment of the analyzer has been determined to be non­hazardous. See Section 2-7 (page 2-4) for leak check procedure.
This analyzer has been designed for use in environments that do not contain combus­tible or explosive materials.
This analyzer uses a fuel containing hy­drogen. Leakage from the fuel contain­ment system can result in an explosion. The fuel supply and containment system, both inside and outside the analyzer, should be carefully checked for leaks upon installation, before initial startup, during routine maintenance or any time the integ­rity of the sample containment system is broken.
If hazardous sample is to be introduced into this analyzer, the leak check proce­dure should also be applied to the sample containment system, both inside and out­side the analyzer.
Proceed as follows:
1. Check analyzer to make sure that plugs and caps are removed from all inlet and outlet fittings.

2-6 GAS CONNECTION

For external gas lines, the use of all new tub­ing throughout is strongly recommended. Copper refrigeration tubing is preferred. Stainless steel tubing is less desirable, be­cause it contains hydrocarbon contaminants, necessitating thorough cleaning before in­stallation.
NOTE
In connecting gas supply lines and asso­ciated fittings, use Teflon tape only. Do not use pipe thread compound or other sub­stance with an organic base.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
2. If a vent line is to be connected to exhaust outlet, use 1/2-inch ID tube slanted downward at least 10 degrees from hori­zontal.
NOTE:
Since water vapor is formed during oxida­tion of hydrogen, burner exhaust gas al­ways contains moisture, even if air and fuel entering the burner are completely dry. Unless exhaust line slants down, wa­ter may accumulate in line, causing back pressure and noisy readings. If exhaust line becomes blocked, water may back up in line and flood burner.