merson designs, manufactures and tests its products to meet many national and international stan-
E
dards. Because these sensors are sophisticated technical products, you MUST properly install, use,
and maintain them to ensure they continue to operate within their normal specifications. The
following instructions MUST be adhered to and integrated into your safety program when installing,
using, and maintaining Rosemount products. Failure to follow the proper instructions may cause any
one of the following situations to occur: loss of life; personal injury; property damage; damage to this
sensor; and warranty invalidation.
•Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson representative for
clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance
of the product.
•Install your equipment as specified in the Installation Instructions of the appropriate
Reference Manual and per applicable local and national codes. Connect all products to
the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update,
program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's
performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY.
Third-party substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and
personal injury.
The information contained in this document is subject to change without notice.
DANGER
Hazardous Area InstallationN
This sensor is not Intrinsically Safe. or Explosion Proof. Installations near flammable liquids or in hazardous
area locations must be carefully evaluated by qualified on site safety personnel.
To secure and maintain an intrinsically safe installation, an appropriate transmitter/safety barrier/sensor
combi nation must be used. The installation system must be in accordance with the governing approval
agency (FM, CSA or BASEEFA/CENELEC) hazardous area classification requirements. Consult your transmitter instruc tion manual for details.
Proper installation, operation and servicing of this sensor in a Hazardous Area Instal lation is entirely the
responsibility of the user.
CAUTION
Sensor/Process Application Compatibility
The wetted sensor materials may not be compatible with process composition and operating
conditions. Application compatibility is entirely the responsibility of the user.
Reference ManualTable of Contents
00809-0100-3396February 2019
Contents
Section 1: Description and Specifications
1.1 TUpH Features and Applications ..........................................................................1
EC Declaration of Conformity ...............................................................................41
Intrisicallly Safe Sensor Installation Drawing - FM ......................................47
Table of Contentsi
Table of ContentsReference Manual
February 201900809-0100-3396
iiTable of Contents
Reference ManualDescription and Specifications
00809-0100-3396February 2019
Section 1: Description and Specifications
1.1TUpH Features and Applications
The low maintenance and disposable Rosemount 396, 396VP, 397, 398, and 398VP TUpH sensors
offer long life and high performance in measuring pH or ORP in aqueous solutions in pipelines, open
tanks, or ponds.
These TUpH sensors feature a large area reference junction for maximum resistance to process
coatings, generally found in dirty, high solid applications. In addition, the secondary helical refer-
ence pathway serves as added protection from poisoning ions. The simplified construction,
designed with user convenience in mind, does not require electrolyte replenishment or any
component replacement. All models feature a large glass bulb for increased resistance to the
effects of aging for longer life.
1.1.1Rosemount 396 and 396VP Sensor Features and Applications
Rosemount 396/396VP sensors are constructed of
polypropylene and stainless steel and are completely
sealed by EP (ethylene propylene) to eliminate process
intrusion.
Sensors are specially designed for improved life in
harsh, dirty applications such as lime slurry waste treat-
ment and paper machine headbox and pigment/dye
applications where large quantities of suspended solids
are present. Installation is easily achieved through a
wide variety of mounting configurations.
1.1.2 Rosemount 397 Sensor Features and Applications
The Rosemount 397 is housed in a highly chemical
resistant polypropylene body and completely sealed with
EP to eliminate process intrusion. The Rosemount 397
body is specifically designed for use with the Quik-Loc Kit
which consists of an adapter and coupler. The PEEK (poly
etheretherke tone) adapter enables the Rosemount 397
sensor to fit into a 1 in. MNPT Dixon coupler for quick and
easy removal without sensor cable twisting. The 316 stain
less steel DIxon coupler is sealed with EP an features
locking arms. The Quik-Loc Kit is not recommended for
use in processes with hazardous, corrosive, or strong
oxidizing chemicals due to a risk of spray and bodily hazards.
1.1.2Rosemount 398 and 398VP Sensor Features and Applications
The chemical-resistant construction of Tefzel, titanium, and the TUpH reference junction make
the Rosemount 398 and 398VP the ideal sensors for measuring pH in harsh process liquids. Use
these sensors to measure pH in sour water strippers, in pulp bleaching towers that use chlorine
dioxide, and in process streams containing a variety of organic solvents.
Rosemount 398 and 398VP sensors use the highest quality materials to provide superior chem-
ical resistance. The sensors are housed in a titanium tube and features an optional 1 inch MNPT
process connector for insertion, submersion, or flow-through applications. The molded Tefzel
TUpH construction is offered with a choice of seals (Viton, EPDM, or Kalrez).
Description and Specifications1
Description and SpecificationsReference Manual
February 201900809-0100-3396
Figure 1-1: Cross Section Diagram of the TUpH Reference Technology
All TUpH sensors are designed with a large area reference junction, helical reference pathway, and an
AccuGlass pH glass bulb. This sensor technology ensures superior performance while only requiring
minimal maintenance.
For first time Rosemount 396/396VP/398/398VP installations, using the following guide is
recommended:
1.Process Connector Accessories (required for all first time installations with 1-inch process
connection threads)
Choose one: PN 23166-00, 316 SST, 1 in. x 1 in. NPT process connector, with EPDM O-ring
PN 23166-01, Titanium, 1 in. x 1 in. NPT process connector, with EPDM O-ring
PN 9510066, Nylon, 1 in. x 1 in. NPT process connector (submersion only)
Choose one (optional process connector o-rings)
PN 9550220, Kalrez o-ring, 2-214
PN 9550099, Viton o-ring, 2-214
2.Variopol Cable (required for all first time installations) of Models 396VP and 398VP
Choose one: PN 24281-00, 15 ft cable with mating VP connector
PN 24281-06, 10 ft cable with mating VP connector
3.Mounting Accessories (optional)
Choose one: PN 915240-03 PVC flow through tee, 3/4 in. NPT process connection
PN 915240-04 PVC flow through tee, 1 in. NPT process connection
PN 915240-05 PVC flow through tee, 1 1/2 in. NPT process connection
PN 11275-01 Sensor handrail mounting assembly
PN 2002011 1-1/2 in. CPVC Tee with 1 in. FNPT connection
PN 24091-00 Low Flow Cell
4.Remote Junction Boxes (optional, recommended for sensor to analyzer distances of more
than 15 ft)
Choose one: PN 23555-00 includes preamplifier
5.Extension cables (used with remote junction boxes)
Inspect the outside of the carton for any damage. If damage is detected, contact the carrier
immediately. Inspect the instrument and hardware. Make sure all the items in the packing list are
resent and in good condition. Notify the factory if any part is missing. If the sensor appears to be
p
in satisfactory condition, proceed to Section 2.2, Mounting.
Note: Save the original packing cartons and materials as most carriers require proof of damage due
to mishandling, etc. Also, if it is necessary to return the instrument to the factory, you must pack
the instrument in the same manner as it was received. Refer to Section 6 for return instructions. If
the sensor is to be stored, the vinyl boot should be filled with pH buffer solution and replaced on
sensor tip until ready to use.
WARNING
Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.
2.3Mounting
Each sensor has been designed to be located in industrial process environments. Temperature and
pressure limitations must not be exceeded at any time. A caution or warning label regarding this
matter is attached to each sensor. For insertion, transfer the label as shown on label instructions.
See Figure 2-2. For submersion applications, first note limits then remove and discard label.
Note: Before mounting the sensor, shake down the sensor to remove any air bubbles that may be
present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as
shipped and readings with an accuracy of ±0.6 pH may be obtained. To obtain greater accuracy or
to verify proper operation, the sensor must be calibrated as a loop with its compatible analyzer or
transmitter.
2.3.1Flow Through and Insertion Mounting for Rosemount 396, 396VP, 398,
and 398VP Sensors.
These sensors can be used with a 1 inch MNPT process connector at the front of the sensor for
mounting into a 1-1/2 inch tee or the process. See Figure 2-1 for instal lation configurations.
2.3.2Submersion Mounting for Rosemount 396, 396VP, 398, and 398VP
Sensors.
These sensors also have a 1 inch MNPT process connector available for use on the back of the sensor.
Tapered pipe threads in plastic fittings tend to loosen after installation. It is therefore recom
mended that Teflon tape be used on the threads and that the tightness of the connection be
checked frequently to assure that no loosening has occurred. To prevent rain water or condensation from running into the sensor, a weatherproof junction box is recommended (see Figure 2-4).
The sensor cable must be run through a protective conduit for isolation from electrical interference or physical abuse from the process. The sensor should be installed within 80° of vertical, with
the electrode facing down. The sensor’s cable should not be run with power or control wiring.
10Installation
Reference ManualInstallation
00809-0100-3396February 2019
2.3.2Quik-Loc Mounting for Rosemount 397
The Quik- Loc mounting is used with the Rosemount 397 sensor only.
WARNING
Once the Quik-Loc unit is installed the operator should wait for the process to cool to a safe temperature, use
the pressure drain valve to relieve all process pressure and observe the pressure on the pressure gauge for
proper removal of the sensor without spray or bodily injury. The Quik-Loc kit used with the Rosemount 397
sensor is not recommended for use with hazardous, corrosive, or strong oxidizing chemicals due to a risk of
spray or bodily injury.
WARNING
It is recommended that a thermometer, drain valve to relieve pressure and pressure gauge be inserted
near the Quik-Loc assembly (see Figure 2-5).
Wrap the pipe threads of the Twin-Kam Kamloc coupler with Teflon tape before placing it into the
process pipe. The coupler can be connected to any 1 in. process connection and must be mounted
within 80° of vertical, with the electrode facing down. Once the coupler is in place, the adapter
should be positioned onto the back end of the sensor. (See Figure 2-5) Remove the parafilm wrap
ping from the two 0-rings on the sensor, grease the o-rings with the lube provided, and feed the
sensor cable through the adapter. Once the adapter is slipped over the sensor’s back end, the
retaining ring (which is included with every Rosemount 397 sensor) should be installed on the black,
grooved, back end of the sensor body. The retaining ring secures the sensor into the adapter. The
adapter/sensor assembly is now ready to be inserted into the coupler. With both arms of the coupler
in the released position, insert the adapter/sensor assembly into the coupler.
Note: The adapter can not be inserted completely or properly unless both arms are in the fully
released position (see Figure 2-5).
Once the adapter has been properly placed in the coupler, both arms should be positioned in the
locked position. A Sur- Loc™ spring arm has been provided on one arm so that the arm cannot
be opened until the spring is released.
Note: The sensor may obstruct flow through smaller pipes.
Once the arms of the coupler have been locked in position, use tamper-evident safety wire on the
metal rings of the Twin-Kam arms to prevent unauthorized and/or untrained personnel from using
the Quik-Loc unit.
WARNING
It is the responsibility of each company using the Quik-Loc Kit/ 397 TUpH Sensor to train personnel of the
injury risks associated with using a quick-release coupler that is placed in a hot or pressurized process.
The Quik-Loc unit should be used only within the pressure and temperature limits stated for the
Rosemount 397 sensor in Section 1.2.
Installation11