Rosemount Manual: 371 ORP Sensor | Rosemount Manuals & Guides

Model 371 ORP Sensor
13.5 EuroSenz
Instruction Sheet
PN 51A-371ORP/rev.C January 2011
SPECIFICATIONS
ORP Range: -1500 to 1500 mV
Glass, and Platinum
Process Connection: Standard PG
13.5 Thread
Electrode Connection: Single Pole
Euro Connector
Temperature Range: 0° - 85°C
(32° - 185°F)
Pressure Range: 100 - 790 kPa abs
(0 - 100 psig)
Specifications subject to change without notice.
FIGURE 1. Sensor Dimensions
MILLIMETER
INCH
STORAGE
1. It is recommended that electrodes be stored in their original shipping containers until needed.
2. Do not store at temperatures below -5°C (23°F).
3. Electrodes should be stored with a protective cap containing KCl solution (PN 9210342).
4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution.
ELECTRODE PREPARATION
1. Remove electrode from shipping container.
2. Remove the protective boot covering the electrode bulb.
NOTE
Do not allow lubricant to coat the platinum band or reference junction. If it does, wipe it clean before installation.
INSTALLATION
For sensor orientation, see Figure 2.
For wiring, see Figure 3.
FIGURE 2. Sensor Orientation
For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/.
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
WARNING
The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the
responsi-
bility of the user.
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
ATEX DIRECTIVE
Special Conditions for safe use
1. All pH/ORP sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the danger due to a build up of an electrostatic charge.
2. All pH/ORP sensor Models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. a.c. test to earth. This must be taken into consideration at installation.
FIGURE 3. Electrical Connection
1. INSTRUMENTS MUST BE USED IN MANUAL TEMPERATURE COMPENSATION MODE.
2. A SEPARATE THERMOCOUPLE MAY BE USED IF TEMPERATURE COMPENSA­TION IS DESIRED. PLEASE CONSULT APPROPRIATE ANALYZER/TRANSMITTER MANUAL FOR MORE WIRING DETAILS.
FOR USE WITH 1054 SERIES WITH INTEGRAL PREAMP IN ANALYZER, REQUIRES THE USE OF BNC ADAPTER (PN 9120516) TO BE PLACED ON THE BLACK WIRE.
3
3
ORP CALIBRATION
1. After making an electrical connection between the sensor and the instrument, obtain a standard solu­tion of saturated quinhydrone. This can also be made quite simply by adding a few crystals of quin­hydrone to either pH 4 or pH 7 buffer. Quinhydrone is only slightly soluble, therefore only a few crystals will be required.
2. Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize.
3. Adjust the standardize control of the instrument to the solution value shown in the table below. The resulting potentials, measured with a clean plat­inum electrode and saturated KCl/AgCl reference electrode, should be within +/- 20 millivolts of the value shown in the table below. Solution tempera­ture must be noted to ensure accurate interpreta­tion of results. The ORP value of saturated quinhy­drone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used.
4. Remove the sensor from the buffer, rinse, and install in the process.
MAINTENANCE
Electrodes should respond rapidly. Sluggishness, off­sets, and erratic readings are indicators that the elec­trodes may need cleaning or replacement.
1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution.
3. ORP (metallic) electrodes should be polished with moistened baking soda.
pH 4 Solution pH 7 Solution
Temp °C 20 25 30 20 25 30
mV Potential 268 264 260 94 87 80
ORP of Saturated Quinhydrone Solution (millivolts)
NOTE: NEWER CABLES MAY HAVE DIFFERENT WIRE COLORS. IN THIS CASE, THE BLACK WIRE IS REFER­ENCE, AND THE CLEAR/COPPER WIRE IS mV IN.
Emerson Process Management
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2011
Loading...