Rosemount Manual: 247 Economy Toroidal Conductivity Sensor | Rosemount Manuals & Guides

Model 247
Economy Toroidal Conductivity Sensor
Instruction Manual
PN 51-247/rev.A September 2002
CAUTION
SENSOR/PROCESS
APPLICATION COMPATIBILITY
DANGER
HAZARDOUS AREA INSTALLATION
Installations near flammable liquids or in haz­ardous area locations must be carefully eval­uated by qualified on site safety personnel. This sensor is not
Intrinsically Safe or
Explosion Proof.
To secure and maintain an intrinsically safe installation, the certified safety barrier, transmitter, and sensor combination must be used. The installation system must com­ply with the governing approval agency (FM, CSA or BASEEFA/CENELEC) hazardous area classification requirements. Consult your analyzer/transmitter instruction manual for details.
Proper installation, operation and servicing of this sensor in a Hazardous Area Instal­lation is entirely the responsibility of the user.
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international standards. Because these instruments are sophisticated technical prod­ucts, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifica­tions. The following instructions must be adhered to and inte­grated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property dam­age; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and serv­icing the product. If this Instruction Manual is not the cor­rect manual, telephone 1-800-654-7768 and the requested manual will be provided. Save this Instruction Manual for future reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installa­tion, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the prod­uct.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the prod­uct’s performance and place the safe operation of your process at risk. Look alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2002
CAUTION
BEFORE REMOVING THE SENSOR, be absolutely certain the process pressure is reduced to 0 psig and the process temperature is at a safe level!
MODEL 247 TABLE OF CONTENTS
MODEL 247
ECONOMY TOROIDAL CONDUCTIVITY SENSOR
TABLE OF CONTENTS
Section Title Page
1.0 MODEL 247 SENSOR............................................................................................. 1
1.1 General .................................................................................................................... 1
1.2 Specifications........................................................................................................... 1
1.3 Ordering Information ................................................................................................ 2
1.4 Accessories.............................................................................................................. 2
1.5 Sensor Installation ................................................................................................... 3
1.6 Sensor Wiring .......................................................................................................... 5
1.7 Sensor Calibration ................................................................................................... 7
1.8 Sensor Maintenance ................................................................................................ 7
1.9 Troubleshooting ....................................................................................................... 7
2.0 RETURN OF MATERIAL......................................................................................... 9
LIST OF TABLES
Table No. Title Page
1-1 Specifications: Sensor ............................................................................................. 1
1-2 Specifications: Adapters .......................................................................................... 1
1-3 Sensor Resistance Check ....................................................................................... 8
1-4 Resistance vs. Temperature .................................................................................... 8
i
MODEL 247 TABLE OF CONTENTS
TABLE OF CONTENTS (CONTINUED)
LIST OF FIGURES
Figure No. Title Page
1-1 Sensor Dimensions.................................................................................................. 2
1-2 Model 247 with Mounting Adapter: Kit 23986-00 .................................................... 3
1-3 Model 247 with CPVC Union: Kit 23986-01 ............................................................ 3
1-4 Model 247 in Custom Flow-Tee: Kit 23986-02 ........................................................ 4
1-5 Wire Functions......................................................................................................... 5
1-6 Wiring to Model 1055-10-21-31 ............................................................................... 5
1-7 Wiring to Model 1055-11-21-31 ............................................................................... 6
1
MODEL 247 SECTION 1.0
MODEL 247 SENSOR
SECTION 1.0
MODEL 247 SENSOR
1.1 GENERAL.
Rosemount Analytical Toroidal Conductivity Sensors are ideal for use in processes where conventional sensors -- those with electrodes exposed to the measured solution -- would corrode or become fouled.
The Model 247 Economy Toroidal Conductivity Sensor consists of wire-wound toroids that are isolated from the process. One toroid acts as a transmitter and the other as a receiver. Energizing the transmitter toroid induces an electric current into the process solution which induces an electric current into the receiver toroid. The strength of that induced current is directly proportional to the conductivity of the solution.
The Model 247 sensor is constructed with a solvent­welded CPVC body which makes it durable and chemically resistant to most dilute acids, bases, and inorganic salt solutions. The sensor’s integral Pt 100 RTD ensures accurate temperature compensation.
The Model 247 sensor is easy to install. The toroidal sensor is not sensitive to flow rate or direction. It can be installed using PVC* fittings readily available at local hardware stores or with CPVC fittings offered here as accessories.
All these features make the sensor ideal for applying to cooling water treatment, boiler blowdown, and metal plating processes.
The Model 247 CPVC Toroidal Conductivity Sensor is compatible with Rosemount Analytical instrument Model 1055-21 or -31.
*Carefully check the manufacturer’s pressure and tempera­ture limits for any adapters not supplied by Rosemount Analytical. PVC adapters typically have lower pressure and temperature ratings than those specified here for CPVC.
TABLE 1-2. Adapters
TABLE 1-1. Sensor
MODEL
Description Materials of Construction Temperature Pressure
23986-00 Mounting Adapter, CPVC 165°F (75°C) 100 psig (790 kPa abs)
1” slip x 1” MPT
23986-01 Union, CPVC, Viton
®
165°F (75°C) 100 psig (790 kPa abs)
1” slip x 1” slip
23986-02 Tee, Flow-through, CPVC, Viton
®
150°F (66°C) 100 psig (790 kPa abs)
2” with slip union
1.2 SPECIFICATIONS
Installation Type Submersion, Flow-Tee
Conductivity Range 500 microSiemen/cm to 1 Siemen/cm
Maximum Temperature 165ºF (75ºC)
Maximum Pressure 100 psig (790 kPa abs)
Wetted Materials CPVC
Shipping Weight 1.9 lbs (0.9 kg)
®
Viton is a registered trademark of DuPont Dow Elastomers.