Rosemount Manual: 1085B-C Conductivity Microprocessor Analyzer | Rosemount Manuals & Guides

Model 1054B C
Conductivity Microprocessor Analyzer
Instruction Manual
PN 51-1054BC/rev.B April 2003
WARNING
ELECTRICAL SHOCK HAZARD
Relay contacts made to separate power sources must be disconnected before servicing.
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
Unused cable conduit entries must be securely sealed by non-flammable closures to provide enclosure integrity in compliance with personal safety and environmental protection requirements.
For safety and proper performance this instru­ment must be connected to a properly grounded three-wire power source.
Proper relay use and configuration is the respon­sibility of the user.
Do not operate this instrument without front cover secured. Refer installation, operation and servicing to qualified personnel.
Be sure to disconnect all hazardous voltage before opening the enclosure.
The unused conduit openings need to be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (IP65).
No external connection to the instrument of more than 43V peak allowed with the exception of power and relay terminals. Any violation will impair the safety protection provided.
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PRO-
CEEDING!
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international stan­dards. Because these instruments are sophisticated tech­nical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manual, telephone 1-949-757-8500 and the requested manual will be provided. Save this Instruction Manual for future reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper instal­lation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qual­ified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitu­tions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protec­tive covers are in place, except when maintenance is being performed by qualified persons, to prevent electri­cal shock and personal injury.
WARNING
This product is not intended for use in the light industrial, residential or commercial environment, per the instrument’s certification to EN50081-2.
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.RAuniloc.com
© Rosemount Analytical Inc. 2001
MODEL 1054B C TABLE OF CONTENTS
MODEL 1054B CONDUCTIVITY MICROPROCESSOR ANALYZER
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS........................................................... 1
1.1 Features and Applications ................................................................................. 1
1.2 Specifications .................................................................................................... 2
1.3 Ordering Information.......................................................................................... 4
2.0 INSTALLATION.................................................................................................. 5
2.1 General.............................................................................................................. 5
2.2 Unpacking and Inspection ................................................................................. 5
2.3 Mechanical Installations .................................................................................... 5
2.4 Electrical Wiring................................................................................................. 6
3.0 DESCRIPTION OF CONTROLS ....................................................................... 14
3.1 Keyboard Functions........................................................................................... 14
3.2 Item Selection and Value Adjustment Keys....................................................... 15
4.0 CONFIGURATION............................................................................................. 19
4.1 General ............................................................................................................. 19
4.2 Alarm 1 and 2 ................................................................................................... 21
4.3 Interval Timer .................................................................................................... 22
4.4 Temperature ..................................................................................................... 23
4.5 Current Output .................................................................................................. 23
4.6 Defaults ............................................................................................................. 24
4.7 Alarm Setpoint ................................................................................................... 25
4.8 Output Scale Expansion .................................................................................... 26
4.9 Simulate Output................................................................................................. 27
5.0 START-UP AND CALIBRATION........................................................................ 28
5.1 General.............................................................................................................. 28
5.2 Entering Cell Constant....................................................................................... 28
5.3 Temperature Calibration .................................................................................... 28
5.4 Initial Loop Calibration ....................................................................................... 28
5.5 Routine Standardization .................................................................................... 30
5.6 Sensor Maintenance.......................................................................................... 30
6.0 KEYBOARD SECURITY ................................................................................... 31
6.1 General.............................................................................................................. 31
6.2 Access Code ..................................................................................................... 31
7.0 THEORY OF OPERATION................................................................................ 31
8.0 DIAGNOSTICS AND TROUBLESHOOTING .................................................... 32
8.1 Diagnostics ........................................................................................................ 32
8.2 Troubleshooting................................................................................................. 33
8.3 Instrument Maintenance .................................................................................... 35
9.0 RETURN OF MATERIAL................................................................................... 37
i
MODEL 1054B C TABLE OF CONTENTS
LIST OF FIGURES
Figure No. Title Page
2-1 Panel Mounting Cutout...................................................................................... 7
2-2 Panel Mounting Tab Installation......................................................................... 8
2-3 Wall Mounting J-Box Installation........................................................................ 9
2-4 Wall Mounting J-Box Wiring............................................................................... 10
2-5 Pipe Mounting Installation ................................................................................. 11
2-6 Electrical Wiring................................................................................................. 12
2-7 Wall Mount Enclosure (Option -20).................................................................... 13
3-1 Function Select on Keypad................................................................................ 14
3-2 Accessing Editing Function ............................................................................... 14
3-3 Accessing Configuration Menus........................................................................ 14
3-4 LCD Display....................................................................................................... 15
3-5 Set Menu Items ................................................................................................. 18
4-1 Alarm 1 and 2 Configuration.............................................................................. 21
4-2 Interval Timer Configuration .............................................................................. 22
4-3 Timer Diagram for One Cycle ............................................................................ 22
4-4 Temperature Configuration................................................................................ 23
4-5 Current Output Configuration ............................................................................ 23
4-6 Default Configuration......................................................................................... 24
4-7 Alarm Setpoint................................................................................................... 25
4-8 Output Scale Expansion.................................................................................... 26
4-9 Simulate Current Output.................................................................................... 27
8-1 Simulate Conductivity Input............................................................................... 36
LIST OF TABLES
Table No. Title Page
1-1 Conductivity Range ........................................................................................... 3
1-2 Replacement Parts............................................................................................ 4
1-3 Accessories ....................................................................................................... 4
3-1 Key Description ................................................................................................. 16
3-2 Information Mnemonics ..................................................................................... 17
3-3 Set Function Mnemonics................................................................................... 17
4-1 Configuration Work Sheet ................................................................................. 20
4-2 Relay States for Various Conditions and Alarm/Default Configurations ............ 24
8-1 Fault Mnemonics ............................................................................................... 32
8-2 RTD Resistance Values..................................................................................... 32
8-3 Troubleshooting Guide ...................................................................................... 34
ii
1
MODEL 1054B C SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.1 FEATURES AND APPLICATIONS
The Model 1054B Microprocessor Analyzers, with the appropriate sensors, are designed to continuously meas­ure and control pH, ORP, conductivity, resistivity, ratio, percent concentration, dissolved oxygen, ozone or total free chlorine in industrial and municipal processes.
The Model 1054B Conductivity Analyzers are housed in a NEMA 4X (IP65) weatherproof, corrosion-resistant, flame retardant enclosure suitable for panel, pipe or wall mounting. All functions are accessed through the front panel membrane keyboard which features tactile feed­back. Measurement data may be read at any time. However, settings may be protected against accidental or unauthorized changes by a user selectable security code. The display indicates the measured value in engi­neering units as well as temperature, alarm status, hold output and fault conditions.
The 1054B transmits a user selected isolated current output which is continuously expandable over the meas­urement range for either direct or reverse action and can be displayed in milliamps or percent. Output dampening of 0-255 sec. is user selectable.
The output and relay default settings are user selectable for hold or fault mode operation. The hold output function allows manual control during routine sensor mainte­nance.
Continuous self diagnostics alert the operator to faults due to analyzer electronics, integral RTD failures, open wiring and process variable range problems. In the event of a fault condition or hold mode diagnosed by the ana­lyzer, the output will be set to a preset or last process value and the relays will be set to their default settings.
Dual alarms are a standard feature on the Model 1054B and are programmable for either high or low operation. Alarm 2 may be programmed as a fault alarm. Both alarms feature independent setpoints, adjustable hys­teresis and time delay action. The time delay is conven­ient when an alarm is used for corrective action, such as shutting down a demineralizer for regeneration. Time delay will ignore a temporary breakthrough and prevent shutting down a demineralizer unit prematurely. A dedi­cated interval timer with relay is also provided.
Automatic or manual temperature compensation is key­board selectable. The process temperature is accurately measured from an integral RTD in the sensor assembly and is read on the display. For greater accuracy, the tem­perature indication may be standardized to the process temperature. The temperature may be configured to read in °C or °F.
Calibration is easily accomplished by simply immersing the sensor in a known solution and entering the value. With a two point calibration, the Model 1054B will auto­matically calculate the temperature slope of the solution. Upon routine standardization a sensor cell factor value is calculated, and a trend of this value can be used to track sensor coating.
The Model 1054B Microprocessor Analyzer comes stan­dard with an LCD display. An LED display is available as an option.
• SELF DIAGNOSTICS with a user selectable fault alarm.
• KEYBOARD SECURITY is user selectable.
• NO BATTERY BACK-UP REQUIRED. Non-volatile EEPROM memory.
• DUAL ALARMS WITH PROGRAMMABLE LOGIC. A third relay is provided with timer func­tions.
• PROGRAMMABLE OUTPUT AND RELAY DEFAULTS for hold and fault modes.
• NEMA 4X (IP65) WEATHERPROOF CORROSION-RESISTANT ENCLOSURE.
2
Model 1054B C SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.2 SPECIFICATIONS -
Enclosure: Black, ABS, NEMA 4X, IP65,
CSA Enclosure 4 144 X 144 X 192 mm (5.7 X 5.7 X 7.6 in.)
Wall Mount Enclosure: NEMA 4X, Heavy duty
fiberglass, reinforced thermoplastic.
356.4 X 450.1 X 180.2 mm* (14 X 17.7 X 7.1 in.*)
Front Panel: Membrane keyboard with tactile feed-
back and user selectable security
Digital Display: LCD, black on grey
Optional, red LED Character Height: 18 mm (0.7 in.)
Electrical Classification:
FM Class I, Div. 2, Group A thru D
28 Vdc relays - 5.0 amps resistive only 150 mA - Groups A & B; 400 mA - Group C ;
540 mA - Group D; Ci = 0; Li = 0
CSA Class I, Div. 2, Group A thru D.
28 Vdc, 110 Vac & 230 Vac relays
5.0 Amps resistive only
Wall Mount Enclosure: General Purpose
Power: 100 - 127 VAC, 50/60 Hz ± 6%, 4.0 W
200 - 253 VAC, 50/60 Hz ± 6%, 4.0 W
Current Output: Isolated, 0-20 mA or 4-20 mA into
600 ohms maximum load at 115/230 Vac or 550 ohms maximum load at 100/200 Vac, Direct or Reverse Output Dampening: 0-255 seconds
Code -20 Wall Mount Enclosure does not meet CE requirements
*Includes latches and mounting feet
EMI/RFI: EN61326 LVD: EN61010-1 Ambient Temperature: -20 to 65°C (-4 to 149
°
F)
Ambient Humidity: LED max 95% RH
(LCD max 85% RH @ 50°C)
Alarms: Dual, field selectable High/Low, High/High,
Low/Low Alarm 2 configurable as a fault alarm Time Delay 0 to 254 seconds Dual Setpoints, continuously adjustable Hysteresis is adjustable up to 25% full scale for low side/High Alarm and high side/Low Alarm
Interval Timer: Interval: 10 min. to 2999 days
On Counts: 1 to 60 On Duration: 1 to 299.9 seconds Off Duration: 1 to 299.9 seconds Wait Duration: 1 to 299.9 seconds Controls dedicated relay
Relay Contacts: Epoxy Sealed Form A contacts,
SPST, Normally Open.
Resistive
Inductive
28 VDC 5.0 Amps 3.0 Amps 115 VAC 5.0 Amps 3.0 Amps 230 VAC 5.0 Amps 1.5 Amps
Weight/Shipping Weight: 1.1 kg/1.6 kg (2.5 lb/3.5 lb)
3
Model 1054B C SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
The Model 1054B Conductivity Analyzer measures over the range of 0-2 µS/cm to 0-1,000 mS/cm. Temperature slope may be adjusted anywhere between 0 and 5% to provide greater accuracy in chemical con­centration control. The temperature slope is factory set at 2% as a representative value, but each conductive solution has its own set of temperature vs. concentration curves. The Model 1054B C will automatically calcu­late the temperature slope for any given solution, or permit manual adjustment of the temperature slope if already known. On calibration the analyzer will also automatically correct for cell constant variations for better measurement accuracy.
ANALYZER SPECIFICATIONS @ 25°C Measurement Range: (See Table 1) Output Scale: Zero suppression: up to 90% full scale.
Span: from 10% to 100% full scale
Accuracy: ±0.5% of reading Repeatability: ±0.25% of reading Stability: ±0.25% month, non-cumulative Temperature Effect: 0.02% of reading/°C Temperature Compensation: -20 to 200°C
(-4 to 392°F) (automatic or manual)
Temperature Slope Adjustment: 0-5%/
°
C
RECOMMENDED SENSORS:
Model 140 Retractable Conductivity
Model 150 Insertion/Submersion Conductivity
Model 400 Screw-in Conductivity
Model 401-14 Screw-in Conductivity
Model 402 Retractable Conductivity
Model 403 Sanitary Flange Conductivity
Model 404 Low Flow Conductivity
FULL SCALE MICROSIEMENS/cm
TABLE 1-1. CONDUCTIVITY RANGE
Conductivity Sensor 150 140
Model Number 400 140 150 150
402/403 400/402/403
404
Cell Constant 0.1 0.2 0.5 1.0
Min. Range 2 4 100 200
Max. Range* 2,000 4,000 10,000 20,000
* Values shown are absolute conductivi-
ty. Maximum range will be reduced for compensated conductivity at elevated process temperatures.
4
MODEL 1054B C SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.3 ORDERING INFORMATION
The Model 1054B Microprocessor Analyzer: Housed in a corrosion resistant, weatherproof enclosure and oper-
ates on either 115 or 230 VAC, 50/60 Hz power. Standard features include digital display, isolated current output, dual alarms, and automatic and manual temperature compensation.
MODEL
1054B MICROPROCESSOR ANALYZER (3.5 lbs/1.5 kg)
Code Measurement
C Contacting Conductivity T Toroidal Conductivity
P/N DESCRIPTION
22966-00 PCB, LCD Digital Display
23025-01 Panel Mounting Kit
23739-00 PCB, Power Supply
23664-01 PCB, CPU, Conductivity
23245-01 PCB, LED Digital Display
23740-00 PCB, Motherboard
23695-04 Keyboard Overlay, LCD Version
23695-05 Keyboard Overlay, LED Version
33469-00 Enclosure, Body
33470-00 Enclosure, Rear Cover
32938-00 Gasket, Rear Cover
9100157 Fuse, 0.1A, 3AB, 250V, Slo-Blow
9100160 Fuse, .250A, 125V
9100189 Fuse, .750A, 125V
TABLE 1-2. Replacement Parts
CODE STANDARD ENCLOSURE OPTIONS
01 LCD Display 02 LED Display
CODE OPTIONS
20 Wall Mount Enclosure
1054BC 01 20 EXAMPLE
TABLE 1-3. Accessories
P/N DESCRIPTION
2001492 Tag, Stainless Steel, Specify
Marking
23053-00 Mounting Bracket, 2-inch Pipe
23054-01 Mounting Bracket, Wall, with
Junction Box
23268-01 Heater, 115 VAC, 50/60 Hz,
1054B (Code 20 Only)
23268-02 Heater, 230 VAC, 50/60 Hz,
1054B (Code 20 Only)
5
MODEL 1054B C SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
2.1 GENERAL. Installation must be performed by a
trained technician. This analyzer's enclosure is suit­able for outdoor use. However, it should be located in an area where temperature extremes and vibrations are minimized or absent. Installation must be per­formed by a trained technician.
2.2 UNPACKING AND INSPECTION. Inspect the
analyzer for shipping damage. If damaged, notify the carrier immediately. Confirm that all items shown on the packing list are present. Notify Rosemount Analytical if items are missing.
2.3 MECHANICAL INSTALLATION. Select an instal-
lation site that is at least one foot from any high volt­age conduit, has easy access for operating personnel, and is not in direct sunlight. Mount the analyzer as fol­lows:
1. Remove the four screws that secure the rear
cover of the enclosure.
2. Remove the four screws holding the front panel
assembly of the enclosure and carefully pull the front panel and connected printed circuit boards straight out.
3. Follow the procedure for the appropriate mounting
configuration: Section 2.3.1 for panel mounting, Section 2.3.2 for wall mounting, Section 2.3.3 for pipe mounting.
2.3.1 Panel Mounting (Standard). The Model 1054B
is designed to fit into a DIN standard 137.9 mm x
137.9 mm (5.43 in. x 5.43 in.) panel cutout (Refer to
Figure 2-1 and Figure 2-2).
1. Prepare the analyzer as described in Section 2.3.
2. Install the mounting latches as described in Figure
2-2 (latches are shown oversize for clarity). If the latches are not installed exactly as shown, they will not work correctly. The screws provided are self-tapping. Tap the screw the full depth of the mounting latch (refer to side view) leaving a gap greater than the thickness of the cutout panel.
3. Align the latches as shown and insert the analyz­er enclosure through the front of the panel cutout. Tighten the screws for a firm fit. To avoid damag­ing the mounting latches, do not use excessive force.
4. Replace the front panel assembly. Circuit boards must align with the slots on the inside of the enclosure. Assure that the continuity wire is con­nected to the rear cover and the interface board’s closest mounting screw. Replace the door and four front panel screws.
2.3.2 Wall Mounting Plate with Junction Box
(P/N 23054-01). Refer to Figure 2-3 and Figure 2-4.
1. Prepare the analyzer as described in Section 2.3.
2. Mount the junction box and bracket to the analyz­er with the hardware provided. All wiring can be brought to the terminal strip prior to mounting the analyzer.
3. Place the metal stiffener on the inside of the ana­lyzer and mount the two 1/2-inch conduit fittings using two each weather seals as shown. Mount NEMA 4X conduit plug (included) into center con­duit hole.
4. Mount the analyzer to the junction box using the 1/2-inch conduit fittings.
5. Complete wiring from the analyzer to the junction box (Refer to Figure 2-4).
NOTE
Run sensor wiring out of the left opening (From front view) to J-Box. All others out right opening to J-Box.
2.3.3 Pipe Mounting (P/N 23053-00). The 2-inch pipe
mounting bracket includes a metal plate with a cutout for the analyzer (Refer to Section 2.3 for mounting the analyzer into the plate). Mounting details are shown in Figure 2-5.
2.3.4 Wall Mount Enclosure (option -20). See
Figure 2-7 for installation details.
6
MODEL 1054B C SECTION 2.0
INSTALLATION
2.4 ELECTRICAL WIRING. The Model 1054B has
three conduit openings in the bottom rear of the ana­lyzer housing which will accommodate 1/2-inch con­duit fittings. From the front view, the conduit opening on the left is for sensor wiring; the center is for signal output and the opening on the right is for timer, alarm, and AC connections. Sensor wiring should always be run in a separate conduit from power wiring. AC power wiring should be 14 gauge or greater.
NOTE
For maximum EMI/RFI protection the output cable should be shielded and enclosed in an earth grounded, rigid metal conduit. When wiring directly to the instrument connect the output cable’s outer shield to the transmitter’s earth ground via terminal 8 on TB3. When wiring to the wall mounting junc­tion box connect the output cable’s outer shield to the earth ground terminal on TB-A. The sensor cable should also be shield­ed. When wiring directly to the instru­ment connect the sensor cable’s outer shield to the transmitter’s earth ground via terminal 8 of TB-2. If the sensor cable’s outer shield is braided an appro­priate metal cable gland fitting may be used to connect to braid to earth ground via the instrument case. When wiring to the wall mounting junction box connect the sensor cable’s outer shield to the earth ground terminal on TB-A. The user must provide a means to dis­connect the main power supply in the form of circuit breaker or switch. The cir­cuit breaker or the switch must be locat­ed in close proximity to the instrument and identified as the disconnecting device for the instrument.
2.4.1 Power Input Wiring. The Model 1054B can be configured for either 115 VAC or 230 VAC power.
Connect AC power to TB1-8 and -9 (115 VAC) or TB1­7 and -8 (230 VAC) ground to the ground terminal at TB3-8 (refer to Figure 2-6).
CAUTION
The sensitivity and stability of the analyzer will be impaired if the input wiring is not grounded. DO NOT apply power to the analyzer until all electrical connections are verified and secure. The following precau­tions are a guide using UL 508 as a safe­guard for personnel and property.
1. AC connections and grounding must be in compli­ance with UL 508 and/or local electrical codes.
2. The metal stiffener is required to provide support and proper electrical continuity between conduit fittings.
3. This type 4/4X enclosure requires a conduit hub or equivalent that provides watertight connect, REF UL 508-26.10.
4. Watertight fittings/hubs that comply with the requirements of UL 514B are to be used.
5. Conduit hubs are to be connected to the conduit before the hub is connected to the enclosure, REF UL 508-26.10.
6. If the metal support plate is not used, plastic fit­tings must be used to prevent structural damage to the enclosure. Also, appropriate grounding lug and awg conductor must be used with the plastic fittings.
2.4.2 Output Wiring. The signal output and alarm
connections are made to terminals 1 through 6 of TB1 and TB3-1 and 2. (Refer to Figure 2-6).
7
MODEL 1054B C SECTION 2.0
INSTALLATION
FIGURE 2-1. Panel Mounting Cutout
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
41054B01 B
8
MODEL 1054B C SECTION 2.0
INSTALLATION
FIGURE 2-2. Panel Mounting Tab Installation
DWG. NO. REV.
41054A26 A
Loading...
+ 32 hidden pages