Rosemount Manual: 1000A GC Hardware Reference 3-9000-750 Rev A | Rosemount Manuals & Guides

Model 1000A
Gas Chromatograph
Hardware Reference Manual
Applies to Both
Daniel Danalyzer Model 1000A
Rosemount Analytical Model 1000A
Part Number 3-9000-750
Revision A
Model 1000A Gas Chromatograph
Hardware Reference Manual
NOTICE
DANIEL MEASUREMENT AND CONTROL, INC. AND ROSEMOUNT ANALYTICAL, INC. (COLLECTIVELY, “SELLER”) SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS IN THIS MANUAL OR OMISSIONS FROM THIS MANUAL. SELLER MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL SELLER BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, ETC.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES.
THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
SELLER DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND MAINTENANCE OF ANY SELLER PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.
DANIEL AND THE DANIEL LOGO ARE REGISTERED TRADEMARKS OF DANIEL INDUSTRIES, INC. THE ROSEMOUNT AND ROSEMOUNT ANALYTICAL LOGO THE ARE REGISTERED TRADEMARKS OF ROSEMOUNT ANALYTICAL, INC. THE EMERSON LOGO IS A TRADEMARK AND SERVICE MARK OF EMERSON ELECTRIC CO.
COPYRIGHT
All rights reserved. No part of this work may be reproduced or copied in any form or by any
means - graphic, electronic, or mechanical — without first receiving the written permission of
©
2008 BY DANIEL MEASUREMENT AND CONTROL, INC., HOUSTON, TEXAS,
U.S.A.
Daniel Measurement and Control, Inc. Houston, Texas, U.S.A.
WARRANTY
1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise expressly provided herein, Daniel Measurement and Control, Inc. and Rosemount Analytical, Inc., (collectively“Seller”) warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of the Services. Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller rep­resentative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller. Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESSED OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. It
is understood that corrosion or erosion of materials is not covered by our guarantee.
2.
LIMITATION OF REMEDY AND LIABILITY: SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY HEREUNDER SHALL BE LIMITED TO REPAIR, CORRECTION, REPLACEMENT OR REFUND OF PURCHASE PRICE UNDER THE LIMITED WARRANTY CLAUSE IN SECTION 1 HEREIN. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED IN CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERWISE), SHALL SELLER'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED OR SERVICES PROVIDED BY SELLER GIVING RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES. THE TERM "CONSEQUENTIAL DAMAGES" SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE AND COST OF CAPITAL.
IMPORTANT INSTRUCTIONS
Read all instructions prior to installing, operating, and servicing this product.
Follow all warnings, cautions, and instructions marked on and supplied with this product.
Inspect the equipment packing case and if damage exists, notify your local carrier for liability.
Open the packing list and carefully remove equipment and spare or replacement parts
from the case. Inspect all equipment for damage and missing parts.
If items are missing, contact your local Product Services Department representative or
the sales office. Provide the equipment serial number and sales order number to the Product Services Department or sales representative.
All returned equipment or parts must have an RMA (Returned Materials Authorization) form obtained from the Products Services Department. Complete the Customer Problem Report or include a letter describing the problem and corrective action to be performed at the factory.
Phone: 1 (713) 827-5033
Physically attach the RMA, corrective action documentation, and a copy of the packing
list to the equipment and place inside the shipping case. An envelope with a copy of the packing list may be attached to the outside of the shipping case. Send to the address shown above.
Install equipment as specified per the installation instructions and per applicable local and
national codes. All connections shall be made to proper electrical and pressure sources.
Ensure that all equipment doors are closed and protective covers are in place, except
when maintenance is being performed by qualified persons, to prevent personal injury.
Use of this product for any purpose other than its intended purpose may result in property
damage and/or serious injury or death.
Before opening the flameproof enclosure in a flammable atmosphere, the electrical
circuits must be interrupted.
Repairs must be performed using only authorized replacement parts as specified by the
manufacturer. Use of unauthorized parts can affect the product's performance and place the safe operation of the product at risk.
When installing or servicing ATEX certified units, the ATEX approval applies only to
equipment without cable glands. When mounting the flameproof enclosures in a hazardous area, only flameproof cable glands certified to IEC 60079-1 must be used.
This page is intentionally left blank.
Model 1000A
DESCRIPTION 1.1 PURPOSE OF THIS MANUAL ..........................1-1
TABLE OF CONTENTS i

TABLE OF CONTENTS

1.2 INTRODUCTION............................................1-2
1.3 FUNCTIONAL DESCRIPTION...........................1-3
1.4 MINIMUM PC REQUIREMENTS .......................1-5
1.5 MODES OF OPERATION.................................1-6
1.5.1 User Interface ...............................................1-6
1.5.2 Capabilities...................................................1-7
1.6 THEORY OF OPERATION ...............................1-8
EQUIPMENT
DESCRIPTION
1.6.1 Analyzer Detector .........................................1-8
1.6.2 Data Acquisition .........................................1-10
1.6.3 Peak Detection ...........................................1-11
1.6.4 Basic Analysis Computations ........................ 1-13
1.7 GLOSSARY ................................................1-16
2.1 SAMPLING SYSTEM......................................2-1
2.1.1 Sampling Point Location.................................2-2
2.1.2 Sample Volume and Flow Rate........................2-2
2.1.3 Sample Conditioning......................................2-3
2.1.4 Contamination Precautions .............................2-3
2.1.5 Valving ........................................................2-3
2.1.6 Calibration Gas .............................................2-3
2.2 ANALYZER...................................................2-4
2.2.1 Physical Description ......................................2-4
2.2.2 Chromatograph Valves...................................2-5
2.2.3 Detector Subsystem ......................................2-6
MAY 2008 DESCRIPTION
ii TABLE OF CONTENTS
Model 1000A
2.2.4 Analyzer Preamplifier Unit ..............................2-6
2.2.5 Analyzer Specifications ..................................2-7
2.2.6 Utility Gas Requirements ................................2-8
2.3 ELECTRONIC ASSEMBLY ...............................2-8
2.3.1 Controller Hardware Configurations ................. 2-8
INSTALLATION AND
SETUP
3.1 PRECAUTIONS AND WARNINGS .................... 3-3
3.1.1 Hazardous Environments ................................3-3
3.1.2 Power Source Wiring .....................................3-4
3.1.3 Signal Wiring ................................................3-5
3.1.4 Electrical and Signal Ground ...........................3-6
3.1.5 Electrical Conduit ..........................................3-8
3.1.6 Sample Systems Requirements .......................3-9
3.2 PREPARATION............................................ 3-10
3.2.1 Introduction................................................ 3-10
3.2.2 Site Selection .............................................3-10
3.2.3 Unpacking the Unit...................................... 3-11
3.2.4 Necessary Tools and Components ................. 3-12
3.2.5 Optional Tools and Components.................... 3-13
3.3 INSTALLING THE ANALYZER .......................3-14
3.3.1 Analyzer AC Power Wiring ........................... 3-14
3.3.2 Sample and Gas Lines.................................. 3-15
3.4 SETTING THE COM ID ................................. 3-18
3.4.1 Inspect or Change the Com ID ...................... 3-18
3.4.2 Preparing for Serial Connections.................... 3-22
3.4.3 FTB Connection (RS-232)............................. 3-24
3.4.4 PC to GC Cable Short Distance Connection
(RS-232) .................................................... 3-25
3.4.5 Long Distance Connection (RS-422, RS-485).. 3-32
INSTALLATION AND SETUP MAY 2008
Model 1000A
TABLE OF CONTENTS iii
3.4.6 Ethernet Connection (Optional) ....................3-33
3.4.7 GC-Printer Wiring ........................................ 3-34
3.4.8 Discrete Digital I/O Wiring ............................3-35
3.4.9 Analog Input Wiring.....................................3-38
3.4.10 Analog Output Wiring .................................. 3-39
3.4.11 Optional Boards ..........................................3-41
3.5 ANALYZER LEAK CHECKS AND PURGING FOR
FIRST CALIBRATION ...................................3-43
3.5.1 Analyzer Leak Checks.................................. 3-43
3.5.2 Purging Carrier Gas Lines ............................. 3-44
3.5.3 Purging Calibration Gas Lines........................3-46
3.6 SYSTEM START-UP ....................................3-47
MAINTENANCE AND
TROUBLESHOOTING
4.1 HAZARDOUS ENVIRONMENTS.......................4-1
4.2 TROUBLESHOOTING AND REPAIR CONCEPT ...4-2
4.3 ROUTINE MAINTENANCE...............................4-2
4.3.1 Bimonthly Maintenance Checklist ....................4-2
4.3.2 Routine Maintenance Procedures.....................4-4
4.3.3 Contact Service ............................................4-4
4.4 ACCESS TO GC EQUIPMENT ELEMENTS .........4-4
4.4.1 Electrical/Electronic Components.....................4-4
4.4.2 Detector Elements, Heater Elements, Valves and
Columns ......................................................4-7
4.5 PRECAUTIONS FOR HANDLING PC
ASSEMBLIES ................................................4-9
4.6 GENERAL TROUBLESHOOTING.......................4-9
4.6.1 Hardware Alarms ..........................................4-9
4.6.2 Troubleshooting Checklist ............................ 4-12
4.6.3 Test Points Dual Methods Board and FTB .......4-15
MAY 2008 MAINTENANCE AND TROUBLESHOOTING
iv TABLE OF CONTENTS
Model 1000A
4.6.4 Preamplifier ................................................ 4-17
4.6.5 Flow Balance Check .................................... 4-17
4.6.6 Temperature............................................... 4-17
4.6.7 FID Configuration ........................................4-19
4.7 LEAK CHECKS............................................ 4-20
4.7.1 Field Service............................................... 4-20
4.7.2 Factory Level Leak Check............................. 4-21
4.7.3 Plugged Lines, Columns, or Valves ................ 4-23
4.8 CHROMATOGRAPH VALVES........................ 4-24
4.8.1 Required Tools............................................ 4-24
4.8.2 Chromatograph Valve Replacement Parts ....... 4-24
4.8.3 Valve Cleaning............................................4-25
4.8.4 Valve Overhaul ...........................................4-25
4.8.5 TCD Replacement ....................................... 4-27
4.8.6 Micro-FID Removal ...................................... 4-29
4.8.7 Micro-FID Maintenance ................................ 4-31
4.8.8 Micro-FID Re-assembly ................................4-32
4.9 TCD DETECTOR BRIDGE BALANCE............... 4-32
4.10 MEASURE VENT FLOW ...............................4-35
4.11 MODEL 1000A ELECTRICAL COMPONENTS ..4-36
4.11.1 DC Power Supply Replacement Procedures..... 4-39
4.12 COMMUNICATIONS ....................................4-40
4.13 ANALOG INPUTS/OUTPUTS......................... 4-43
4.13.1 Model 1000A Analog Inputs......................... 4-44
4.13.2 Analog Output Adjustment ...........................4-45
4.13.3 Model 1000A Analog Outputs ...................... 4-46
4.14 DISCRETE DIGITAL INPUTS/OUTPUTS ..........4-48
MAINTENANCE AND TROUBLESHOOTING MAY 2008
Model 1000A
TABLE OF CONTENTS v
4.15 RECOMMENDED SPARE PARTS.................... 4-49
4.16 UPGRADE PROCEDURES .............................4-49
4.16.1 Base Operating System ................................4-49
4.16.2 Applications ...............................................4-49
RECOMMENDED SPARE
PARTS
APPENDIX A,
COMMUNICATIONS
SPECIFICATIONS
5.1 ANALYZER SPARES ......................................5-2
5.1.1 Printed Circuit Card Assemblies (Analyzer) .......5-2
5.1.2 Electrical and Mechanical Assemblies (Analyzer)5-2
A.1 TCD SERIAL COMMUNICATIONS................... A-1
A.1.1 Model 1000A with TCD Communications Ports A-2
A.2 FID SERIAL COMMUNICATIONS .................... A-5
A.2.1 Connecting Serial Communications to the GC .. A-8
A.2.2 FTB Serial Communications ..........................A-10
A.3 WIRING LOCAL RS-232 COMMUNICATIONS..A-21
A.3.1 GC Serial Port and Cable Configurations.........A-21
A.3.2 GC DB 9-pin Serial Port to PC DB 9-pin Port ...A-24
A.3.3 GC DB 9-pin Serial Port to PC DB 25-pin Port .A-25
A.3.4 GC PHOENIX Plug Port to PC DB 9-pin Port ....A-26
A.3.5 GC PHOENIX Plug Port to PC DB 25-pin Port ..A-27
A.4 WIRING REMOTE RS-232
COMMUNICATIONS ....................................A-28
A.4.1 GC DB 9-pin Serial Port to Modem DB 25-pin
Port...........................................................A-28
A.4.2 GC PHOENIX Plug to Modem DB 25-pin Port ..A-29
A.5 EXAMPLE RS-422 PC-GC CONNECTION ........A-30
A.6 EXAMPLE RS-485 PC-GC CONNECTION ........A-32
MAY 2008 RECOMMENDED SPARE PARTS
vi TABLE OF CONTENTS
Model 1000A
APPENDIX B, MODEM
INSTALLATION
APPENDIX C,
MANIFOLD CARRIER
FOR GAS BOTTLES
APPENDIX D, LOCAL
OPERATOR INTERFACE
B.1 OPTIONAL INTERNAL MODEM .......................B-1
B.1.1 Optional Ethernet Board ................................. B-3
C.1 CARRIER GAS ..............................................C-1
C.2 INSTALLATION AND LINE PURGING................C-2
C.3 REPLACING CARRIER CYLINDER ....................C-3
C.4 CALIBRATION GAS .......................................C-3
D.1 INTERFACE COMPONENTS FOR DISPLAYING AND
ENTERING DATA ..........................................D-1
D.1.1 Light Emitting Diode Indicators........................D-1
D.1.2 LCD Screen ..................................................D-2
D.1.3 Keypad ........................................................D-2
D.1.4 Security Switch ............................................D-2
D.2 USING THE LOCAL OPERATOR INTERFACE .....D-3
APPENDIX E,
ENGINEERING
DRAWINGS
D.2.1 Navigating the Screen....................................D-4
D.2.2 Editing Numeric Data .....................................D-4
D.2.3 Editing Non-Numeric Data ..............................D-5
D.3 NAVIGATING THE LOI MENUS .......................D-7
D.3.1 The Ctrl Menu ..............................................D-9
D.3.2 The App Menu............................................D-14
D.3.3 The Chrom Menu ........................................D-23
D.3.4 The Logs Menu ...........................................D-29
D.3.5 The Manage Menu ......................................D-37
E.1 LIST OF ENGINEERING DRAWINGS ................. E-1
APPENDIX B, MODEM INSTALLATION MAY 2008
Model 1000A

DESCRIPTION

1.1 PURPOSE OF THIS MANUAL

The Emerson Process Management Model 1000A Gas Chromatograph System Hardware Reference Manual (P/N 3-9000-750) is intended as a
user's guide to accompany the MODEL 1000A GAS CHROMATOGRAPH SYSTEM.
NOTE: For software operation instructions, see the MON2000 Software for Gas Chromatographs User Manual (P/N 3-9000-522).
This manual provides the following information:
A general description of the Model 1000A Gas Chromatograph (GC) System and its components, their configurations and functions. (Section 1: Description)
DESCRIPTION 1-1
A brief description of the GC System's software, user interfaces, and capabilities. (Section 1: Description)
Introduction to GC theory of operation and terminology. (Section 1: Description)
Guidelines for sampling system and gas connections. (Section 2: Equipment Description)
Descriptions of Analyzer subsystems and components. (Section 2: Equipment Description)
Descriptions of GC Controller subsystems and components. (Section 2: Equipment Description)
Instructions for installing the GC System hardware. (Section 3: Installation and Startup)
Instructions for regular maintenance and care of the GC System hardware. (Section 4: Maintenance)
Instructions for troubleshooting, repair, and service of the GC System hardware. (Section 4: Maintenance)
List of boards, valves, and other components suggested as spare parts. (Section 5: Recommended Spare Parts)
JUNE 2008 PURPOSE OF THIS MANUAL
1-2 DESCRIPTION
Appendices with additional, helpful reference materials and drawings. (Appendices)

1.2 INTRODUCTION

The Emerson Process Management Model 1000A Gas Chromatograph is a high-speed GC system that is factory engineered to meet specific field application requirements based on stream composition and the anticipated concentration of the components of interest. The GC system typically consists of two major components, the Analyzer Assembly and the Sample Conditioning System:
Analyzer Assembly (Model 1000A Series)
Located near the sample tap in a freeze-protected shelter. The Analyzer includes columns, detectors, preamplifier, stream switching valves, solenoids, and the GC, which includes electronics and ports for signal processing, instrument control, data storage, personal computer (PC) interface, and telecommunications.
Model 1000A
Sample Conditioning System (SCS)
Located between the process stream and the Analyzer sample inlet, usually mounted on the lower portion of the Analyzer stand. The standard configuration SCS includes a mounting plate, block (or shutoff) valves, and filters. Optionally, the SCS can be configured with
Genie® bypass filters, liquid shut-off valves, and optional solenoids for stream switching; all of which can be enclosed in an electric (heat tape design) oven.
In its standard configuration, the Model 1000A series Analyzer can handle up to five streams: typically, four for sample and one for calibration. With an optional stream switch assembly added, the GC can switch up to twelve streams, maximum.
Although the GC is designed to be operated primarily from the LOI, you can also use a personal computer (PC) running MON2000. The PC option provides the user with the greatest capability, ease-of-use, and flexibility. One PC running MON2000 can connect with up to 32 chromatographs (via RS-485 serial communications links). The PC is used to display analysis chromatograms and reports, which can then be stored to files on the PC hard drive, or printed from either the PC's printer port or the GC's printer port.
INTRODUCTION JUNE 2008
Model 1000A
Since neither the PC nor a normal printer can be placed in a hazardous area, serial port and Modbus communications links are provided for connecting the GC to the PC, other computers, printers, and controllers.

1.3 FUNCTIONAL DESCRIPTION

A functional block diagram of a typical GC installation is shown in Figure
1-1. A sample of the gas to be analyzed is taken from the process stream
by a sample probe installed in the process line. The sample passes through a sample line to the sample conditioning system where it is filtered or otherwise conditioned. After conditioning, the sample flows to the Analyzer for separation and detection of the components of the gas.
The chromatographic separation of the sample gas into its components is accomplished in the Analyzer in the following manner. A precise volume of sample gas is injected into one of the unit's analytical columns. The column contains a stationary phase (packing) that is either an active solid (adsorption partitioning) or an inert solid support that is coated with a liquid phase (absorption partitioning). The gas sample is moved through the column by means of a mobile phase (carrier gas). Selective retardation of the components of the sample takes place in the column that causes each component to move through the column at a different rate. This action separates the sample into its constituent components.
DESCRIPTION 1-3
A detector located at the outlet of the analytical column senses the elution of components from the column and produces electrical outputs proportional to the concentration of each component. Outputs from the Analyzer detectors are amplified in the Analyzer electronics, then transmitted to the GC for further processing. See Section 1.6 for more information.
Although output from the GC is normally displayed on the LOI, it can also be displayed on a remotely located personal computer or a printer. Connection between the GC and the PC can be accomplished via a direct serial line or via the Modbus-compatible communication interface.
Multiple chromatograms may be displayed on the LOI and compared or contrasted with separate color schemes. This allows a stored chromatogram to be compared/contrasted with a current or another stored chromatogram. This could be of great assistance when changing parameters or isolating a problem.
JUNE 2008 FUNCTIONAL DESCRIPTION
1-4 DESCRIPTION
Model 1000A
In most instances, it is essential to use a PC for detailed troubleshooting procedures. Basic operations should be performed from the LOI that is built into the Model 1000A. With optional electronic boards, the PC can be connected remotely via ethernet, telephone, radio or satellite. Once installed and configured, the GC can operate independently for long periods of time.
Figure 1-1 GC System Functional Block Diagram
FUNCTIONAL DESCRIPTION JUNE 2008
Model 1000A

1.4 MINIMUM PC REQUIREMENTS

To achieve maximum performance when running the MON2000 software, ensure your PC system contains the following hardware and software equipment.
PC with a 486/90 MHz or higher processor (Pentium/100MHz or
higher recommended) running:
- Windows® 95 (service pack 1 or better) or later
NOTE: If running Windows® 95 with the optional ethernet card, the user must download Socket 2 from www.microsoft.com/windows95/ downloads to utilize MON2000’s ethernet feature.
- Windows® 98 version 1 or later
DESCRIPTION 1-5
- Windows® 2000 version 1 or later
- Windows® XP version 1 or later (see note for system requirements)
- Windows® Vista version 1 or later
NOTE: You must have administrator privileges to intall MON2000 because Vista will not allow a ‘standard’ user to install software. Even with administrator privileges, you will be prompted by Vista’s User Account Control feature to allow or cancel the installation. For more details, refer to Getting Started with User Account Control on Windows Vista (http://go.microsoft.com/fwlink/?LinkID=102562).
- Windows® NT version 4 (service pack 3 or later)
16 MB of RAM (32 MB or higher recommended)
5 MB of free hard disk space
Super VGA monitor with 800x600 resolution
Free serial port for remote/local connection to gas chromatograph (for
online operations)
Free parallel port for connection to printer
Windows®-compatible modem (for remote connection only)
JUNE 2008 MINIMUM PC REQUIREMENTS
1-6 DESCRIPTION
NOTE: Microsoft Internet Explorer 5.0 is required to view spreadsheets or reports saved in HTML format.
•Use the Settings Control Panel System General Page menu path to check the system version number.
- For Windows® 95, the version number should be 4.00.950A/B or
later.
- For Windows® 98 or Windows® 2000, the version number should
be 1 or later.
- To use Windows® XP you need a PC with 300 MHz or higher
processor clock speed recommended; 233 MHz minimum required (single or dual processor system);* Intel® Pentium®/Celeron®
family, or AMD K6®/Athlon™/Duron™ family, or compatible processor recommended.
- Memory 128 MB of RAM or higher recommended (64 MB minimum
supported; may limit performance and some features)
Model 1000A
- Hard Disk Minimum: 1.5 GB of available hard disk space
- For Windows NT4, the version number should be 4.00.1381 or
later.

1.5 MODES OF OPERATION

1.5.1 User Interface

You have two user interfaces from which to operate the gas chromatograph (GC) system: the LOI or a PC connected to the GC and running MON2000.
The LOI allows you to gather basic information and to perform maintenance repairs at the GC site.
A PC connected to the GC and running MON2000 offers the greatest amount of capability and flexibility.
Find complete user instructions for MON2000 in the program’s online HELP manual as well as in the MON2000 Software for Gas Chromatographs User Manual (P/N 3-9000-522).
MODES OF OPERATION JUNE 2008
Model 1000A

1.5.2 Capabilities

Some of the individual GC controller functions that can be initiated or controlled by the GC and its software, MON2000, include the following:
Valve activations
Timing adjustments
Stream sequences
Heater controls (when applicable)
Calibrations
Baseline runs
•Analyses
Halt operation
DESCRIPTION 1-7
Stream/detector assignments
Stream/component table assignments
Stream/calculation assignments
•Diagnostics
Alarm and event processing
Event sequence changes
Component table adjustments
Calculation adjustments
Alarm parameters adjustments
Analog scale adjustments
Some of the reports and logs that can be produced, depending upon the GC application in use, include the following:
Configuration report
Parameter list
Analysis chromatogram
Chromatogram comparison
JUNE 2008 Capabilities
1-8 DESCRIPTION
Alarm log (unacknowledged and active)
•Event log
Analysis raw data

1.6 THEORY OF OPERATION

NOTE: See Section 1.7 for definitions of some of the terminology used in the following explanations.

1.6.1 Analyzer Detector

The Analyzer detector subsystem is a thermal conductivity detector that consists of a balanced bridge network with heat-sensitive thermistors in each leg of the bridge. Each thermistor is enclosed in a separate chamber of the detector block. One thermistor is designated the reference element and the other the measurement element. A schematic diagram of the thermal conductivity detector is shown in Figure 1-2.
Model 1000A
THEORY OF OPERATION JUNE 2008
Model 1000A
In the quiescent condition (prior to injecting a sample), both legs of the bridge are exposed to pure carrier gas. In this condition, the bridge is balanced and the bridge output is electrically nulled. The bridge can be balanced by the fine and coarse adjustment potentiometers located on the preamplifier circuit board.
The analysis begins when a fixed volume of sample is injected into the column by operation of the sample valve. The sample is moved through the column by the continuous flow of carrier gas. As successive components elute from the column, the temperature of the measurement element changes. The temperature change unbalances the bridge and produces an electrical output proportional to the component concentration. The differential signal developed between the two thermistors is amplified by the preamplifier.
DESCRIPTION 1-9
Figure 1-2 Schematic Diagram of Analyzer Detector Bridge
Figure 1-3 illustrates the change in detector electrical output during
elution of a component.
3
1
detector bridge balanced
1
component begins to elute from column
2
and is measured by thermistor peak concentration of component
3
Figure 1-3 Detector output during component elution
2
1
JUNE 2008 Analyzer Detector
1-10 DESCRIPTION
In addition to amplifying the differential signal developed between the detector's two thermistors, the preamplifier also supplies drive current to the detector bridge. The preamplifier also supplies drive current to the detector bridge. The voltage signal is converted to a 4 to 20-milliamp (mA) current loop for transmission to the GC Controller.
The signal is proportional to the concentration of a component detected in the gas sample. The preamplifier provides four different gain channels as well as compensation for baseline drift.
The signals from the preamplifier are sent to the GC Controller for computation, recording on a printer, or viewing on a PC monitor or LOI.

1.6.2 Data Acquisition

Every second, exactly 40 equi-spaced data samples are taken for analysis by the GC (i.e., once every 25 milliseconds). Each data sample, after having been precision-amplified, is subjected to a twelve bit analog to digital (A/D) conversion. The sampling frequency of 40 Hertz (Hz) was chosen to reduce 60 Hz normal mode noise.
Model 1000A
After each point on the chromatograph signal is sampled, the resulting number is stored in a buffer area in the GC’s memory for processing. During the analysis, only the last 256 data points are available for processing. Because the data analysis is done as the signal is sampled (in real-time), only a limited number of past data samples is required to analyze any signal.
As a part of the data acquisition process, groups of incoming data samples are averaged together before the result is stored to the GC’s memory for processing. Non-overlapping groups of N samples are averaged and stored, and thus reduce the effective incoming data rate to 40/N samples/ second. For example, if N = 5, then a total of 40/5 or 6 (averaged) data samples are stored every second. The value for the variable N is determined by the selection of a Peak Width (PW) parameter. The relationship is:
N PW ondssec=
Data Acquisition JUNE 2008
Model 1000A
where PW is given in seconds. All the various details in the analysis process are independent of the value of N. Allowable values of N are 1 to 63, which corresponds to values of PW from 2 to 63 seconds.
The variable N is known as the integration factor. This term is used because N determines how many points are averaged or integrated to form a single value. The integration of data upon input, before storing, serves two purposes. First, the statistical noise on the input signal is reduced by the square root of N. In the case of N = 4, a noise reduction of two would be realized. Secondly, the integration factor controls the bandwidth of the chromatograph signal. It is necessary to match the bandwidth of the input signal to that of the analysis algorithms in the GC. This prevents small, short duration perturbations from being recognized as true peaks by the program. It is therefore important to choose a Peak Width corresponding to the narrowest peak in a group under consideration.
DESCRIPTION 1-11

1.6.3 Peak Detection

For normal area or peak height concentration evaluation, the determination of a peak's start, peak point, and end is automatic. The manual determination of start and end points is used only for area calculations in the Forced Integration mode. Automatic determination of peak onset or start is initiated whenever Integrate Inhibit is turned off. Analysis is started in a region of signal quiescence and stability, such that the signal level and activity can be considered as baseline values. It is important that this be the case because the assumption is made by the GC software.
Having initiated a peak search by turning Inhibit off, the GC performs a point by point examination of the signal slope. This is achieved by using a digital slope detection filter which is a combination low pass filter and differentiator. The output of this detector is constantly compared to a system constant entered by the operator called Slope Sensitivity. A default value of 8 is assumed if no entry is made. Lower values make peak onset detection more sensitive, and higher values make detection less sensitive. Higher values (20 to 100) would be appropriate for noisy signals, e.g. high amplifier gain.
Peak termination is determined by the same application of this detector to the signal, but in the reverse sense. Onset is defined where the
JUNE 2008 Peak Detection
1-12 DESCRIPTION
Model 1000A
detector output exceeds the baseline constant, but termination is defined subsequently where the detector output is less than the same constant.
Sequences of fused peaks are also automatically handled. This is done by testing each termination point to see if the region immediately following it satisfies the criteria of a baseline. A baseline region must have a slope detector value less than the magnitude of the baseline constant for a number of sequential points. When a baseline region is found, this terminates a sequence of peaks.
A zero reference line for peak height and area determination is established by extending a line from the point of the onset of the peak sequence to the point of the termination. The values of these two points are found by averaging the four integrated points just prior to the onset point and just after the termination points, respectively. The zero reference line will, in general, be non-horizontal, and thus compensates for any linear drift in the system from the time the peak sequence starts until it ends.
In a single peak situation, peak area is the area of the component peak between the curve and the zero reference line. The peak height is the distance from the zero reference line to the maximum point on the component curve. The value and location of the maximum point is determined from quadratic interpolation through the three highest points at the peak of the discrete valued curve stored in the GC.
For fused peak sequences, this interpolation technique is used both for peaks as well as valleys (minimum points). In the latter case, lines are dropped from the interpolated valley points to the zero reference line to partition the fused peak areas into individual peaks. The use of quadratic interpolation improves both area and height calculation accuracy and eliminates the effects of variations in the integration factor on these calculations.
For calibration, the GC may average several analyses of the calibration stream.
Peak Detection JUNE 2008
Model 1000A

1.6.4 Basic Analysis Computations

Two basic analysis algorithms are included in the GC. These are:
Area Analysis - Calculates area under component peak
Peak Height Analysis - Measures height of component peak
Concentration Analysis by Using Response Factor
Concentration calculations require a unique response factor foreach component in an analysis. These response factors may be manually entered by an operator or determined automatically by the system through calibration procedures (with a calibration gas mixture that has known concentrations).
Response factor calculation: (using the external standard)
DESCRIPTION 1-13
or
where:
ARF
n
HRF
n
Area
n
Ht
n
Cal
n
Area
n
-------------- -
=
ARF
n
Cal
n
Ht
n
-----------
=
HRF
n
Cal
n
Area response factor for component n in area per mole percent (%).
Height response factor for component n.
Area associated with component n in calibration gas.
Height associated with component n in mole percent in calibration gas.
Amount of component n in mole percent in calibration gas.
Calculated response factors are stored by the GC for use in the concentration calculations, and are printed out in the configuration and calibration reports.
JUNE 2008 Basic Analysis Computations
1-14 DESCRIPTION
Average response factor is calculated as follows:
k
i 1=
RFAVG
where:
RFAVGnArea or height average response factor for component n.
=
n
------------------
k
RF
Model 1000A
i
Rf
i
Area or height response factor for component n from the calibra­tion run.
k Number of calibration runs actually used to calculate the response
factors.
The percent deviation of new RF averages from old RF average is calculated in the following manner:
% deviation
new
------------------------------------- -
RF
old
old
100×=
RF
RF
where the absolute value of % deviation for alarm has been previously entered by the operator.
Concentration Calculations in Mole % without Normalization
Once response factors have been determined by the GC or entered by the operator, component concentrations are determined for each analysis by using the following equations:
Area
n
-------------- -
CONC
=
n
ARF
n
Basic Analysis Computations JUNE 2008
Model 1000A
or
where:
CONCnConcentration of component n in mole percent.
CONC
DESCRIPTION 1-15
Ht
n
--------------
=
n
HRF
n
Area
ARF
Area of component n in unknown sample
n
Response factor of component n calculated from area of calibration
n
sample. Units are area per mole percent.
Ht
n
HRF
Peak height of component n in unknown sample
Response factor of component n calculated from peak height of cal-
n
ibration sample. Units are height per mole percent.
Note that the average concentration of each component will also be calculated when data averaging is requested.
Component concentrations may be input through analog inputs 1 - 4 or may be fixed. If a fixed value is used, the calibration for that component is the mole % that will be used for all analyses.
Concentration Calculations with Normalization
CONC
CONCN
n
----------------------------
i 1=
k
CONC
n
100×=
i
where:
CONCN
CONC
CONC
k Number of components to be included in the normalization.
JUNE 2008 Basic Analysis Computations
Normalized concentration of component n in percent of total gas
n
concentration.
Non-normalized concentration of component n in mole percent.
n
Non-normalized concentration (in mole percent) from each of the k
i
components to be grouped into this normalization.
1-16 DESCRIPTION
NOTE: For additional information about other calculations that are performed by the GC and software, see the MON2000 Software for Gas Chromatographs User Manual (P/N 3-9000-522).

1.7 GLOSSARY

Auto Zero: Automatic zeroing of the preamplifier. May be entered into the Controller to take place at any time during the analysis when either the component is not eluting or the baseline is steady.
Chromatogram: A permanent record of the detector output. A chromatograph is obtained from the LOI or from a PC interfaced with the detector output through the GC. A typical chromatogram displays all component peaks, and gain changes. It may be viewed in color as it is processed on the LOI or a PC VGA display. Tick marks recorded on the chromatogram by the GC indicate where timed events take place.
Model 1000A
Component: Any one of several different gases that may appear in a sample mixture. For example, natural gas usually contains the following components: nitrogen, carbon dioxide, methane, ethane, propane, isobutane, normal butane, isopentane, normal pentane, and hexanes plus.
Condulet: Fitting resembling a pipe or a box with a removable cover for access to electric conduits.
CTS: Clear to send (a serial port pin assignment).
DCD: Data carrier detect; see also, RLSD (a serial port pin assignment).
DSR: Data set ready (a serial port pin assignment).
DTR: Data terminal ready (a serial port pin assignment).
LOI: Local operator interface; an integrated color display panel with
infrared touchkeys that allows you to interact with the GC.
Response Factor: Correction factor for each component as determined by the calibration. See “Concentration Analysis by Using Response
Factor” on page 13 for more information.
GLOSSARY JUNE 2008
Model 1000A
Retention Time: The time (in seconds) that elapses between start of analysis (0 seconds) and the sensing of the maximum concentration of each component by the Analyzer detector.
RI: Ring indicator (a serial port pin assignment).
RLSD: Received line signal detect (a digital simulation of carrier detect);
see also, DCD (a serial port pin assignment).
RTS: Request to send (a serial port pin assignment).
RxD, RD, or SIN: Receive data, or signal in (a serial port pin
assignment).
DESCRIPTION 1-17
TxD, TD, or S
assignment).
: Transmit data, or signal out (a serial port pin
OUT
JUNE 2008 GLOSSARY
1-18 DESCRIPTION
This page is intentionally left blank.
Model 1000A
GLOSSARY JUNE 2008
Loading...
+ 220 hidden pages