Rosemount Information Bulletin: XA Series GC Analog Wiring Diagrams | Rosemount Manuals & Guides

Information Bulletin
March 8, 2011 XA Series Gas Chromatographs
Analog wiring diagrams
Introduction
Our XA series gas chromatographs are furnished with one isolated 24V power supply to service six analog outputs. See the drawings on page 2 for more information. If the customer desires it, they can furnish power to each analog output while maintaining isolation between channels. See the drawings on page 4 for more information. Our XA series gas chromatographs are also capable of furnishing power to two analog inputs. These inputs are set at the factory to accept current (4-20 mA) sources, but they can be configured to accept voltage (0-10 VDC) sources. See the wiring diagram on page 5 for more information.
Information Bulletin XA Series Gas Chromatographs
Factory analog wiring and switch settings
Consult the following diagrams before wiring a device:
1. This drawing shows how to wire up to six devices to the analog outputs that are located on the back plane. It also shows how to wire up to two analog inputs.
2 Analog wiring diagrams