Rosemount Information Bulletin: 700 and XA Series GC Power Source Wiring | Rosemount Manuals & Guides

Information Bulletin
CAUTION
15 amps is the maximum current for 14 AWG (wire).
May 23, 2011 Model 700, XA Series Gas Chromatographs
Power source wiring
Follow these precautions when installing power source wiring:
1. All wiring, as well as circuit breaker or power-disconnect switch locations, must conform to the CEC or NEC; all local, state, or other jurisdictions; and com­pany standards and practices.
2. Provide single-phase, three-wire
power at 120 or 240 VAC, 50-60 Hertz.
Note: You cannot use a three-phase AC
power source with the 1500XA; the transformer cannot support the current load.
Note: If you do not have a single phase,
three-wire AC power source, you must purchase an isolation transformer. See the schematic on the following page for more information.
3. Locate in a safe area.
6. Use multi-stranded copper conductor wire according to the following recommendations:
For power feed distances up to 250
feet (76 meters), use 14 AWG (American Wire Gauge) (18 Metric Wire Gauge), stranded.
For power feed distances 250 to 500
feet (76 to 152 meters), use 12 AWG (25 Metric Wire Gauge), stranded.
For power feed distances 500 to
1000 feet (152 to 305 meters), use 10 AWG (30 Metric Wire Gauge), stranded.
Cable entries are M32 in accordance
with ISO 965.
4. Provide the 700XA and any optionally installed devices with one 15-amp circuit breaker for protection.
5. Ensure that the 24 VDC input power is compliant with the Seperated Extra-Low Voltage (SELV) standard by ensuring suitable electrical separation from other circuits.