Rosemount Hygienic Process Connection Adapters (FB-1001, FB-1002, FB-1010, FB-1011, FB-1020, FB-1021, FB-1022) Quick Start Guide

00825-0100-4103, Rev 02
Rosemount™ Hygienic Process Connection Adapter
Dairy Coupling, VARIVENT®, Tri Clamp
Quick Start Guide
August 2020

1 About this guide

This guide provides instructions for installing a hygienic process connection adapter on Rosemount 326, 327, and 1408 transmitters with G1 process connectors. The following adapters are covered by this guide:
Table 1-1: Hygienic Process Connection Adapters
Part number Description
FB-1001 1½-in. Tri Clamp (ISO 2852 / DIN 32676)
FB-1002 2-in. Tri Clamp (ISO 2852 / DIN 32676)
FB-1010 VARIVENT® Type F
FB-1011 VARIVENT Type N
FB-1020 DN32 Dairy Coupling (DIN 11851)
FB-1021 DN40 Dairy Coupling (DIN 11851)
FB-1022 DN50 Dairy Coupling (DIN 11851)
2

1.1 Safety instructions

Follow these guidelines to ensure safe use and extend the life of the product.
CAUTION
Read this document before using the product and keep it during
the entire service life.
The product must be suitable for the corresponding applications
and environmental conditions without any restrictions.
Only use the product for its intended purpose.
If the operating instructions or the technical data are not adhered
to, personal injury and/or damage to property might occur.
The manufacturer assumes no liability or warranty for any
consequences caused by tampering with the product or incorrect use by the operator.
Installation, electrical connection, set-up, operation and
maintenance of the product must be carried out by qualified personnel.

1.2 Hygienic adapter O-ring

The adapter comes with an EPDM O-ring. More O-rings are available as accessories. See each product's Product Data Sheet for more information.
3

2 Install the adapter

326P
1408
326L
327T
Prerequisites
Use a new adapter when replacing a PEEK sealing ring or changing from a PEEK seal to O-ring sealing.
Procedure
Place the O-ring in the groove of the transmitter.
1.
Note
Be careful not to scratch or damage the surfaces of the transmitter.
2. Grease the transmitter thread with lubricating paste.
Note
The paste must be approved for the application and compatible with the elastomers used.
4
3. (Only for DIN 11851) Slide the coupling nut up over the bottom of the transmitter.
4. Screw the adapter onto the transmitter until it is hand-tight.
5. Secure the transmitter and adapter using a vise.
CAUTION
Do not over-tighten the vise as it might damage the adapter.
5
6. Tighten the transmitter to the recommended torque.
34 mm / 39 mm
Torque 310 in-lb (35 N-m)
Note
Further tightening may affect the sealing effect.
6

3 Mount the transmitter

Procedure
Mount the transmitter and adapter assembly using a fixing element (coupling nut or clamp flange).
7

4 Product certifications

4.1 3-A® certification

FB-1001, FB-1002, FB-1010, and FB-1011 are authorized to display the 3-A symbol. Ensure gaskets and any other accessories selected for installation meet both the application and 3-A requirements. A certificate of compliance is available at

4.2 Other industry certifications

All connector surfaces and materials which come into contact with process medium comply with the following regulations:
(EC) No. 1935/2004
(EC) No. 2023/2006
CFR Title 21 (FDA) § 177.2600
CFR Title 21 (FDA) § 177. 2415
Emerson.com/Rosemount.
8
9
10
11
*00825-0100-4103*
Quick Start Guide
00825-0100-4103, Rev. 02
August 2020
Emerson Automation Solutions
6021 Innovation Blvd. Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888
+1 952 949 7001
RFQ.RMD-RCC@Emerson.com
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Youtube.com/user/RosemountMeasurement
©
2020 Emerson. All rights reserved.
Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
North America Regional Office
Emerson Automation Solutions 8200 Market Blvd. Chanhassen, MN 55317, USA
+1 800 999 9307 or +1 952 906 8888
+1 952 949 7001
RMT-NA.RCCRFQ@Emerson.com
Loading...