Rosemount Hx438 Sensor Amperometric Steam Sterilizable Dissolved Oxygen Sensors Quick Start Guide

Quick Start Guide
00825-0100-3494, Rev AA
June 2019
Rosemount™ Hx438
Amperometric Steam Sterilizable Dissolved Oxygen Sensors
Quick Start Guide June 2019
WARNING
High pressure and temperature
Failure to reduce the pressure and temperature may cause serious injury to personnel.
Before removing the sensor, reduce the process pressure to 0 psig and cool down the process temperature.
Do not exceed temperature and pressure limitations of 266 °F (130 °C) and 43 psig (400 kPa, 4 bar).
WARNING
The electrolyte is highly alkaline.
Protect your hands with gloves and use safety goggles. Avoid contact of the electrolyte with the skin, eyes, and mucous membranes.
2 Emerson.com
June 2019 Quick Start Guide

1 Plan

1.1 Unpack and inspect

Procedure
1. Inspect the outside of the carton for any damage.
2. If damage is detected, contact the carrier immediately.
3. Inspect the hardware.
4. Make sure all the items in the packing list are present and in good condition.
5. Notify the factory if any part is missing.

1.2 Specifications

Table 1-1: Rosemount Hx438 dissolved oxygen sensor specifications
Hx438 Dissolved Oxygen Sensor Material and Units
Operating Temperature Range 32 to 266 °F (0 to 130 °C)
Maximum Pressure 43 psig (400 kPa abs, 4 bar)
Measurement Range 0 to 20 ppm or 0 to 250% saturation,
Wetted Materials Stainless steel and EPDM
Process Connections PG 13.5 thread
Cable Connector Standard 4-pin
Cable Compatibility Standard 4-pin connector cable
Compatible Mounting Accessory Insertion Mounting Assembly
Compatible Transmitters Rosemount Transmitter Models 56,
Temperature Compensation 22K NTC
depending on instrument
1056, 5081, and 1066
Quick Start Guide 3
Quick Start Guide June 2019

2 Install

2.1 Preparing the sensor

Procedure
1. Remove the plastic cap carefully which covers the sensing end of sensor.
2. Avoid causing a vacuum by gently and slowly twisting the sensor and cap until it is removed.
3. Sensor is shipped dry, so electrolyte must be added prior to installation.
4. Remove membrane cartridge.
5. Add 1.5 ml of electrolyte to the membrane cartridge.
6. Screw membrane cartridge back onto sensor body.

2.2 Connecting the sensor to the trasmitter

Procedure
1. Wire the sensor to the transmitter.
2. Apply power to the transmitter.
3. Make the sensor ready for calibration after two hours. The polarization time is necessary to get stable signals from the
sensors
4. If the sensor is disconnected from the transmitter for a short time, allow the sensor to stabilize for at least two times longer than the time was disconnected.
Time needed for stabilization should not exceed two hours.
4 Emerson.com
June 2019 Quick Start Guide

2.3 Mount the sensor

Procedure
Use PG 13.5 process thread to mount the sensor.
NOTICE
Do not install the sensor upside down.
Quick Start Guide 5
Quick Start Guide June 2019
Figure 2-1: Dimensional Drawing for Rosemount Hx438-01
A.
Sensor part number
Hx438-01 4.725 120
B. 4-pin connector C. Free rotating PC 13.5 screw thread
D. O-ring
2.4
6 Emerson.com

Wire the sensor to the transmitter

Procedure
Wire the sensor to the transmitter. See wiring diagrams below.
A dimension
in. mm
June 2019 Quick Start Guide
Figure 2-2: Wiring for Rosemount™ 1056 and 56 Transmitter
A. Yellow B. Blue C. Yellow/green
D. Brown
E. Clear
F. Resistance temperature device return G. Resistance temperature device sense H. Resistance temperature device in
I. Resistance temperature device shield J. Anode shield
K. Anode
L. Cathode shield
M. Cathode
Quick Start Guide 7
Quick Start Guide June 2019
Figure 2-3: Wiring for Rosemount 1066 Transmitter
A. Hinge inside of front panel B. Anode C. Cathode
D. TB5/input-anode-brown
E. TB5/input-cathode-clear
F. TB3/resistance temperature device-return-yellow G. TB3/resistance temperature device/resistance temperature device in-blue H. TB2/reference and solution ground-solution ground-yellow/green
Follow recommended order.
a. TB4/anode and cathode
1. Anode
2. Cathode
b. TB3/resistance temperature device
1. Return
2. Sense
3. Resistance temperature device in
c. TB2/solution ground
1. No connection
2. No connection
8 Emerson.com
June 2019 Quick Start Guide
3. Solution ground
Note
TB1, TB4, TB6, and TB& not used for oxygen and ozone sensor wiring. TB1, TB2, and TB4 may be used for pH sensor wiring if free chlorine measurement requires live pH input.
Figure 2-4: Wiring for Rosemount 5081 Transmitter
A. Clear B. Brown C. Yellow/green
D. Blue
E. Yellow
F. Cathode G. Anode H. mV in
I. Drain J. Solution ground
K. Reference
L. Resistance temperature device in
M. Resistance temperature device sense
N. Resistance temperature device return
Quick Start Guide 9
Quick Start Guide June 2019

3 Calibrate

3.1 Sterilizing the sensor

Procedure
1. Before autoclaving the sensor, cover the connector end of the sensor with a tight protective cap (PN 242000).
2. If the connector end gets wet despite the protective cap, dry the connector with pressurized air or hair dryer.
Drying prevents corrosion and damage to insulation.
Note
No protection is necessary, when doing in-situ sterilization.

3.2 Dismounting the sensor

Procedure
1. Unscrew the free rotating PG 13.5 threaded connector.
2. Pull the sensor out of the process or mounting assembly.

3.3 Calibrate

Prerequisites
Before calibrating the sensor, refer to the transmitter instruction manual for more details on calibration.
Prerequisites
Make sure that the sensor is operated for at least two hours before zeroing and calibrating.
Procedure
1. Place the sensor in nitrogen gas or in water containing about 5% sodium sulfite to make it zero.
If using nitrogen gas, make sure that the membrane is dry.
2. Once the reading is stable, zero the sensor. Refer to transmitter instruction manual.
3. If sodium sulfite is used to zero the sensor, rinse the sensor with water and gently dry the membrane.
4. Place the sensor in water-saturated air. Once the reading is stable, complete the full scale calibration. Refer to transmitter instruction manual.
10 Emerson.com
June 2019 Quick Start Guide
5. Calibration can also be done using air-saturated water or air­saturated medium.
Altough the sensor has an extremely low oxygen consumption rate, the liquid should be gently stirred during calibration.
Quick Start Guide 11
Quick Start Guide June 2019

4 Maintaining and troubleshooting

4.1 Maintaining the sensor

Procedure
1. Periodically check the sensor response in air and nitrogen.
2. Place the sensor in air. Allow the reading to become stable and note the value.
Also, note the sensor current, which should be between 40 and 80 nA.
3. Place the sensor in nitrogen. A small plastic bag with a stream of nitrogen gas discharging into the bottom works well.
4. After one minute, the sensor current should be less than 2% of the value in air.

4.2 Replacing the electrolyte and the membrane cartridge

Replace the membrane cartridge (membrane kit PN 9160487) as follows:
Procedure
1. Hold the sensor vertically with the membrane pointing down. Carefully unscrew the membrane cartridge.
2. Carefully clean the tip of the glass body with the polishing cloth included with membrane kit, or clean with the polishing tool in one direction only.
CAUTION
Glass Body Damage Do not polish in circular motion. Failure to correctly polish the glass
body may result in permanent damage to the equipment.
NOTICE
When replacing the membrane cartridge, do not touch the anode wire.
12 Emerson.com
June 2019 Quick Start Guide
3. Rinse the sensor under running water and dry it carefully.
4. Check the small O-ring above the glass body. If it is damaged, replace it.
5. Use the plastic pipette in the membrane kit to add 2 ml of electrolyte solution to the new membrane cartridge.
6. Carefully screw the cartridge onto the sensor shaft. If too much electrolyte solution was used, the excess will be pushed
out.
7. Wash off the excess electrolyte with water.

4.3 Replace the cathode

If the sensor stil lhas a sluggish response to saturated air and/or too small current in air, then the cathode should be replaced.
Procedure
1. Put the sensor in an upright position and carefully unscrew the membrane cartridge.
Note
The membrane should be changed whenever the cathode is replaced.
2. Flush the replaceable cathode with deionized water then dry the metal parts. Do not touch anode and cathode.
Quick Start Guide 13
Quick Start Guide June 2019
Figure 4-1: Replacing the Cathode
A. Sensor shaft thread B. Cathode - metal part C. Sensor shaft
D. Cathode O-ring
3. Hold the cathode on the metal part (B) in front of the thread (A) and the sensor on the shaft (C) and pull apart. Just pull, do not turn.
4. Check that all contacts are dry and clean.
5. Replace cathode. (Part number 24262-00) Rotate the cathode until the correct position is found and then push cathode into the shaft.
14 Emerson.com
June 2019 Quick Start Guide
6. Check the small O-ring (D) above the cathode. Replace if damaged.
7. Add 1.5 ml of electrolyte into the new membrane cartridge.
NOTICE
The pipette must NOT touch the membrane itself.
8. Screw the membrane cartridge onto the sensor shaft. Any spillage of electrolyte should be rinsed away with water.

4.4 Troubleshoot

Table 4-1: Troubleshoot
Current in air too high (>80 nA at 25 °C)
Problem Solution Problem Solution Problem
Very thin or defective membrane
Defective glass body or connector
Poisoned anode
Cathode contaminated by silver
Replace with new membrane cartridge
Return to Emerson
Return to Emerson
Clean cathode with polishing paper. If cleaning does not restore the sensor, then replace the cathode
Sluggish response Current in air too low (<40
Contaminate d, fouled, or dirty membrane
Loose membrane
Dried out electrolyte film
Cathode contaminated by silver
Clean membrane or replace with new membrane cartridge
Replace with new membrane cartridge
Loosen membrane cartridge and tighten
Clean cathode with polishing paper. If cleaning does not restore the sensor, then replace the cathode
nA at 25 °C)
Contaminate d, fouled, or dirty membrane
Dried out electrolyte film
Cathode contaminated by silver
Exhausted electrolyte
Solution
Clean membrane or replace with new membrane cartridge
Loosen membrane cartridge and tighten
Clean cathode with polishing paper. If cleaning does not restore the sensor, then replace the cathode
Return to Emerson
Quick Start Guide 15
Quick Start Guide June 2019
Table 4-1: Troubleshoot (continued)
Current in air too high (>80 nA at 25 °C)
Problem Solution Problem Solution Problem
- - - - Defective
Sluggish response Current in air too low (<40
nA at 25 °C)
glass body or connector
FDA Approved Membrane replacement kit, which includes electrolyte solution and polishing paper, is PN 9160487.
Solution
Fill with new electrolyte
16 Emerson.com
June 2019 Quick Start Guide

5 Accessories

Part number Description
Cable accessories
9160493 5 m (16.4 ft). 4-pin connector, bare wire on transmitter
Mounting accessories
9160478 Insertion 70 mm insertion, use 120 mm sensor
9160484 Service kit for insertion mounting assembly
9160483 15° weld-in socket, G 1¼ in. thread, 44 mm
Servicing accessories
9160487 Service kit for sensors, includes three FDA approved
9160489 30 ml bottle electrolyte
242000 Third party labeled connector cap for 4-pin connector for
24262-00 Third party labeled replacement anode/cathode cartridge
end
membranes modules, O-rings, polishing tool, 20 ml electrolyte
autoclaving (compatible with Rosemount™ Hx438-01)
The 4-pin connector cable is offered in three lengths.
The insertion mounting assembly can be used to mount the Rosemount Hx438-01 into the process tanks or pipes.
The weld-in socket is used to mount the insertion mounting assembly into tanks or pipes.
Figure 5-1: Weld-in Socket
Quick Start Guide 17
Quick Start Guide June 2019
18 Emerson.com
June 2019 Quick Start Guide
Quick Start Guide 19
GLOBAL HEADQUARTERS
6021 Innovation Blvd. Shakopee, MN 55379
+1 866 347 3427 +1 952 949 7001 liquid.csc@emerson.com
*00825-0100-3494*
Quick Start Guide
00825-0100-3494, Rev. AA
June 2019
EUROPE
Emerson Automation Solutions Neuhofstrasse 19a PO Box 1046 CH-6340 Baar Switzerland
+41 (0) 41 768 6111 +41 (0) 41 768 6300 liquid.csc@emerson.com
ASIA-PACIFIC
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Republic of Singapore
+65 6 777 8211 +65 6 777 0947 liquid.csc@emerson.com
Linkedin.com/company/Emerson-
Automation-Solutions
twitter.com/rosemount_news
Facebook.com/Rosemount
youtube.com/RosemountMeasurement
MIDDLE EAST AND AFRICA
Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033
+971 4 811 8100 +971 4 886 5465 liquid.csc@emerson.com
©
2019 Emerson. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
Loading...